MSO3000 and DPO3000 Series Specifications and Performance Verificationi
Table of Content
s
iiMSO3000 and DPO3000 Series Specifications and Performance Verification
General safety s
ummary
General safet
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other
component manuals for warnings and cautions related to operating the system.
To avoid fire or personal injury
Use proper power cord. Use only the power cord specified for this product and certified for the country of use.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected
to a voltag
Connect a
probe.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric
shock, the grounding c onductor must be connected to earth ground. Before making connections to the input or output
termina
Observe
product manual for further ratings information before making connections to the product.
e source.
nd disconnect properly.
ls of the product, ensure that the product is properly grounded.
all terminal ratings.
y summary
De-energize the circuit under test before connecting or disconnecting the current
To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the
Connect the probe reference lead to earth ground only.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source. Do not block the power cord; it
must remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by
fied service personnel.
quali
exposed circuitry.
Avoid
Do not touch exposed connections and components when power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
vide proper ventilation.
Pro
proper ventilation.
Refer to the manual's installation instructions for details on installing the product so it has
MSO3000 and DPO3000 Series Specifications and Performance Verificationiii
General safety s
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ummary
ivMSO3000 and DPO3000 Series Specifications and Performance Verification
Specifications
Specification
This chapter contains specifications for the MSO3000 and DPO3000 Series oscilloscopes. All specifications are guaranteed
unless noted as "typical." Typical specifications are provided for your convenience but are not guaranteed. Specifications
that are marked with the
All specifications apply to all MSO3000 and DPO3000 models unless noted otherwise. To meet specifi cations, two
conditions must first be met:
The oscilloscope must have been operating continuously for twenty minutes within the operating temperature range
specified.
You must perform the Signal Path Compensation (SPC) operation described in this manual prior to evaluating
specifications. (See page 48.) If the operating temperature changes by more than 10 °C (18 °F), you must perform
the SPC operation again.
s
symbol have associated procedures listed in the Performance Verification section.
Analog Channel Input And Vertical Specifications
Table 1: Analog channel input and vertical specifications
GND coupling approximates a ground reference by switching the channel's input relay to an
internal DAC set to 0V. The signal connected to the input BNC is not disconnected from the
channel's input load when the input is set to GND coupling.
1MΩ,50Ω,or75Ω
For instruments with 500MHz bandwidth
(includes MSO/DPO305X models as well as
MSO/DPO303X/301X models with 500 MHz
upgrade):
For instruments with 300MHz bandwidth
(includes MSO/DPO303X models as well as
MSO/DPO301X models with 300 MHz upgrade):
For instruments with 100MHz bandwidth
(MSO/DPO301X models):
5V
with peaks ≤ ±20 V
RMS
There is an overvoltage trip circuit, int ended to protect against overloads that might damage
termination resistors. A sufficiently large impulse can cause damage regardless of the overvoltage
protection circuitry, due to the finite time required to detect the overvoltage condition and respond
to it.
At front panel connector, 300 V
CAT II
, Installation Category II; Apply UL specified voltages for 300 V
RMS
2 analog, digitized simultaneously
VSWR ≤ 1.5:1 from D C to 500 MHz,
typical
VSWR ≤ 1.5:1 from D C to 350 MHz,
typical
VSWR ≤ 1.5:1 from D C to 100 MHz,
typical
MSO3000 and DPO3000 Series Specifications and Performance Verification1
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
< 100mV/divDerate at 20 dB/decade above 100 kHz to 30 V
≥ 100mV/divDerate at 20 dB/decade above 3 MHz to 30 V
DC balance
(See page 51.)
0.2 div with the input DC-50Ω coupled and 50 Ω terminated
0.25 div at 2 mV/div with the input DC-50 Ω coupled and 50 Ω terminated
at 1 MHz, 10 dB/decade above 1 MHz
RMS
at 30 MHz, 10 dB/decade above 30 MHz
RMS
0.5 div at 1 mV/div with the input DC-50 Ω coupled and 50 Ω terminated
0.2 div with the input DC-75 Ω coupled and 75 Ω terminated
0.25 div at 2mV/div with the input DC-75 Ω coupled and 75 Ω terminated
0.5 div at 1mV/div with input DC-75 Ω coupled and 75 Ω terminated
0.2 div with the input DC-1 MΩ coupled and 50 Ω terminated
0.3 div at 1 mV/div with the input DC-1 MΩ coupled and 50 Ω terminated
All the above specifications are increased by 0.01 divisions per ° C above 40 ° C.
Delay between channels,
full bandwidth, typical
≤ 100 ps between any two channels with input termination set to 50 Ω, DC coupling
≤ 100 ps between any two channels with input termination set to 75 Ω, DC coupling
Note: all settings in the instrument can be manually time aligned using the Probe Deskew function
Deskew range-125 ns to +125 ns
Crosstalk (channel
isolation), typical
≥ 100:1 at ≤ 100 MHz and ≥ 30:1 at > 100 MHz up to the rated bandwidth for any two channels
having equal Volts/Div settings
TekVPI InterfaceThe probe interface allows installing, powering, compensating, and controlling a wide range of
probes offering a variety of features.
The interface is available on all front panel inputs including Aux In. Aux In only provides 1 M Ω
input termination and does not offer 50 Ω or 75 Ω as do the other input channels.
Total probe power,
typical
Three (MSO30X2 and DPO30X2) or five (DPO30X4 and MSO30X4) TekVPI compliant probe
interfaces, 1 per channel
20 W internally available probe power
Provision for 50 W external power from rear panel
Number of digitized bits
8bits
Displayed vertically with 25 digitization levels (DL) per division, 10.24 divisions dynamic range
"DL" is the abbreviation for "digitization level." A DL is the smallest voltage level change that can
be resolved by an 8-bit A-D Converter. This value is also known as the LSB (least significant bit).
1MΩ:1 mV/div to 10 V/div in a 1-2-5 sequenceSensitivity range (coarse)
50 Ω and 75Ω:1 mV/div to 1 V/div in a 1-2-5 sequence
Sensitivity range (fine)Allows continuous adjustment from 1 mV/div to 10 V/div, 1 MΩ
Allows continuous adjustment from 1 mV/div to 1 V/div, 75 Ω
Allows continuous adjustment from 1 mV/div to 1 V/div, 50 Ω
Sensitivity resolution
≤ 1% of current setting
(fine), typical
2MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Position range±5 divisions
For firmware versions 2.04 and 2.07 only, the position range is ±4 divisions.
Specifications
Analog bandwidth,
50 Ω input termination
(See page 53.)
Instrument ModelInstrument
MSO/DPO3054
DPO3052
MSO/DPO3034
PO3032
MSO/D
MSO/DPO3014
MSO/DPO3012
The limits stated below are for ambient temperature of ≤ 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Vertical Scale Setting
Bandwidth
500 MHz
500 MHz
500 MHz
300 MHz
500 MHz
300 MHz
Hz
500 M
300 MHz
100 MHz
0MHz
50
300 MHz
100 MHz
10 mV/div to
1V/div
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 500
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 300 MHzDC to 300 MHzDC to 250 MHzDC to 150 MHz
00 MHz
DC to 5
DC to 300 MHzDC to 300 MHzDC to 250 MHzDC to 150 MHz
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
o 300 MHz
DC t
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
C to 300 MHzDC to 300 MHz
D
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
5mV/divto9.98
mV/div
MHz
DC to 400
DC to 4
DC t
MHz
00 MHz
o 300 MHz
2mV/divto
4.98mV/div
DC to 250
DC to 2
DC t
C to 250 MHz
D
MHz
50 MHz
o 250 MHz
1 m V/div to
1.99 mV/div
DC to 150
50 MHz
DC to 1
o 150 MHz
DC t
C to 150 MHz
D
MHz
MSO3000 and DPO3000 Series Specifications and Performance Verification3
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Analog bandwidth, 75 Ω
input termination, typical
Instrument ModelInstrument
MSO/DPO3054
DPO3052
MSO/DPO3034
MSO/DPO3032
MSO/DPO3014
MSO/DPO3012
The limits stated below are for ambient temperature of ≤ 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Vertical Scale Setting
Bandwidth
500 MHz
500 MHz
500 MHz
300 MHz
500 MHz
300 MHz
500 MHz
300 MHz
100 MHz
500 MHz
300 MHz
100 MHz
10 mV/div to
1V/div
DC to 230 MHzDC to 190 MHzDC to 140 MHzDC to 100 MHz
DC to 230 MHzDC to 190 MHzDC to 140 MHzDC to 100 MHz
DC to 230 MHzDC to 190 MHzDC to 140 MHzDC to 100 MHz
DC to 230 MHzDC to 190 MHzDC to 140 MHzDC to 100 MHz
DC to 230 MHzDC to 190 MHzDC to 140 MHzDC to 100 MHz
DC to 230 MHzDC to 190 MHzDC to 140 MHzDC to 100 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
5 mV/div to 9.98
mV/div
2 m V/div to
4.98mV/div
1mV/divto
1.99 mV/div
4MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Specifications
Analog bandwidth, 1 M Ω
input termination with
P6139A or P6139B 10X
probe, typical
Instrument ModelInstrument
MSO/DPO3054
DPO3052
MSO/DPO3034
MSO/DPO3032
MSO/DPO3014
MSO/DPO3012
The limits stated below are for ambient temperature of ≤ 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Vertical Scale Setting
Bandwidth
500 MHz
500 MHz
500 MHz
300 MHz
500 MHz
300 MHz
500 MHz
300 MHz
100 MHz
500 MHz
300 MHz
100 MHz
100 mV/div to
100 V/div
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 300 MHzDC to 300 MHzDC to 250 MHzDC to 150 MHz
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 300 MHzDC to 300 MHzDC to 250 MHzDC to 150 MHz
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 300 MHzDC to 300 MHzDC to 250 MHzDC to 150 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 300 MHzDC to 300 MHzDC to 250 MHzDC to 150 MHz
DC to 100 MHzDC to 100 MHzDC to 100 MHzDC to 100 MHz
5mV/divto9.98
mV/div
20 mV/div to
49.8mV/div
10 mV/div to
19.9 mV/div
MSO3000 and DPO3000 Series Specifications and Performance Verification5
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Calculated rise time,
typical
Instrument ModelInstrument
MSO/DPO3054
DPO3052
MSO/DPO3034
MSO/DPO3032
MSO/DPO3014
MSO/DPO3012
filter selections
The formula is calculated by measuring -3 dB bandwidth of the oscilloscope. The formula accounts
for the rise time contribution of the oscilloscope independent of the rise time of the signal source.
Vertical Scale Setting
Bandwidth
500 MHz700 ps750 ps1.2 ns2 ns
500 MHz700 ps750 ps1.2 ns2 ns
500 MHz700 ps750 ps1.2 ns2 ns
300 MHz1.2 ns1.2 ns1.2 ns2 ns
500 MHz700 ps750 ps1.2 ns2 ns
300 MHz1.2 ns1.2 ns1.2 ns2 ns
500 MHz700 ps750 ps1.2 ns2 ns
300 MHz1.2 ns1.2 ns1.2 ns2 ns
100 MHz3.5 ns3.5 ns3.5 ns3.5 ns
500 MHz700 ps750 ps1.2 ns2 ns
300 MHz1.2 ns1.2 ns1.2 ns2 ns
100 MHz3.5 ns3.5 ns3.5 ns3.5 ns
For instruments with 500MHz or 300MHz analog bandwidth: 20 M Hz, 150 MHz, and FullAnalog bandwidth limit
For instruments w ith 100MHz analog bandwidth:20 MHz
10 mV/div to 1
V/div
5 mV/div to 9.98
mV/div
2 m V/div to
4.98 mV/div
1 mV/div to 1.99
mV/div
Lower frequency limit, AC
coupled, typical
Upper frequency limit, 150
MHz bandwidth limit filter,
typical
Upper frequency limit, 20
MHz bandwidth limit filter,
typical
DC gain accuracy
(See page 55.)
< 10 Hz when AC to 1 MΩ coupled
The AC coupled lower frequency limits are reduced by a factor of 10 when 10X passive probes
are used.
150 MHz, +40%, and -20%
20 MHz, ±35% (all models)
±2.5% for 1 mV/Div, derated at 0.100%/°C above 30 °C
±2.0% for 2 mV/Div, derated at 0.100%/°C above 30 °C
±1.5% for 5 mV/Div and above, derated at 0.100%/°C above 30 °C
±3.0% Variable Gain, derated at 0.100%/°C above 30 °C
6MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Specifications
DC voltage
measurement accuracy
Sample
acquisition mode,
typical
Average
acquisition mode
Measurement type
Any sample
DC Accuracy (in volts)
±[DC gain accuracy X | reading - (offset - position) |
+ Offset Accuracy +0.15 div + 0.6 mV]
Delta volts between any two
±[DC gain accuracy X | reading | + 0.15 div + 1.2 mV]
samples acquired with the
same oscilloscope setup and
ambient conditions
NOTE. Offset, position, and the constant offset term must be converted to volts by multiplying by
the appropriate volts/div term.
Average of ≥ 16 waveforms±[DC gain accuracy X | reading - (offset - position) |
+ Offset Accuracy + 0.1 div]
Delta volts between any two
±[DC gain accuracy X | reading | + 0.05 div]
averages of ≥ 16 waveforms
acquired with the same
oscilloscope setup and ambient
conditions
NOTE. Offset, position, and the constant offset term must be converted to volts by multiplying by
the appropriate volts/div term.
NOTE. The basic accuracy specification applies directly to any sample and to the following
measurements: High, Low, Max, Min, Mean, Cycle Mean, RMS, and Cycle RMS. The delta volt
accuracy specification applies to subtractive calculations involving two of these measurements.
Offset ranges
The delta volts (difference voltage) accuracy specification applies directly to the following
measurements: Positive Overshoot, Negative Overshoot, Peak-Peak, and Amplitude.
Offset rangeVolts /div
setting
1 mV/div to
1MΩ input50 Ω and 75 Ω input
±1 V±5 V
1
99.5 mV/div
100 mV/div
±10 V±5 V
to 995
mV/div
1 V/div to 10
±100 V±5 V
V/div
NOTE. The input signal cannot exceed the maximum input voltage for the 50 Ω and 75 Ω input
paths. Refer to the Maximum input voltage specifications (earlier in this table) for more information.
MSO3000 and DPO3000 Series Specifications and Performance Verification7
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Offset accuracy
(See page 57.)
Random noise,
sample acq
uisition mode
(See page 59.)
±[0.005 X | offset - position | + DC Balance]
NOTE. Both the position and constant offset term must be converted to volts by multiplying by the
appropriate volts/div term.
ModelInstrument
Bandwidth
MSO/DPO-
500 MHz
3054 and
DPO3052
MSO303X
500 MHz
and
DPO303X
300 MHz
MSO301X
500 MHz
and
DPO301X
Bandwidth Limit
Filter Selection
Full
150 MHz
20 MHz
Full
150 MHz
20 MHz
Full
150 MHz
20 MHz
Full
150 MHz
20 MHz
RMS Noise
<(170 μV + 8% of V/div setting)
<(90 μV + 6% of V/div setting)
<(25 μV + 6% of V/div setting)
<(170 μV + 8% of V/div setting)
<(90 μV + 6% of V/div setting)
<(25 μV + 6% of V/div setting)
<(140 μV + 6% of V/div setting)
<(80 μV + 6% of V/div setting)
<(30 μV + 5% of V/div setting)
<(170 μV + 8% of V/div setting)
<(90 μV + 6% of V/div setting)
<(25 μV + 6% of V/div setting)
1
For 50 Ω path, 1 V/div is the maximum vertical setting.
300 MHz
100 MHz
Full
150 MHz
20 MHz
Full
20 MHz
<(140 μV + 6% of V/div setting)
<(80 μV + 6% of V/div setting)
<(30 μV + 5% of V/div setting)
<(100 μV + 6% of V/div setting)
<(100 μV + 6% of V/div setting)
8MSO3000 and DPO3000 Series Specifications and Performance Verification
Digital Channel Acquisition System Specifications
Table 2: Digital channel acquisition system specifications
Specifications
Characterist
Threshold voltage range-15 V to +25 V
Digital channel timing resolution
Threshold accuracy (See page 61.)
Minimum detectable pulse2.0 ns
Channel to channel skew
ic
Description
2nsforthema
± [100 mV + 3% of threshold setting after calibration], after valid SP C
Using MagniVu memory. Specified at the input to the P6316 probe with all
eight ground inputs connected to the user's ground. Use of leadsets, grabber
clips, ground extenders, or other connection accessories may compromise this
specification.
500 ps
Digital Channel to Digital C hannel only
This is the propagation path skew, and ignores skew contributions due to
bandpass distortion, threshold inaccuracies (see Threshold Accuracy), and
sample binning (see Digital Channel Timing Resolution).
in memory, and 121.2 ps for MagniVu memory
Horizontal And Acquisition System Specifications
Table 3: Horizontal and acquisition system specifications
CharacteristicDescription
Long-term sample rate and delay
time accuracy (See page 64.)
Seconds/division range1 ns/div to 1,000 s/div
Peak detect or
ope mode pulse response, typical
envel
Sample-rate rangeSee Sample Rate Range detail table. (See page 11.)
MSO3000 and DPO3000 Series Specifications and Performance Verification9
Specifications
Table 3: Horizontal and acquisition system specifications (cont.)
CharacteristicDescription
Number of waveforms for average
acquisition mode
Delta time measurement accuracy
(See page 65.)
2 to 512 waveforms
Default of 16 waveforms
The formula to calculate delta-time measurement accuracy (DTA) for a given
instrument setting and input signal is given below (assumes insignificant signal
content above Nyquist).
SR1=SlewRate(1stEdge) around the 1stpoint in the measurement
SR
=SlewRate(2ndEdge) around the 2ndpoint in the measurement
2
N = input-referred noise (volts
, refer to the Random Noise, Sample acquisition
rms
mode specification)
t
= 1 / (Sample Rate)
sr
TBA = timebase accuracy (refer to the Long-term sample rate and delay time
accuracy specification above)
t
= delta-time measurement duration
p
RD = (Record Length) / (Sample Rate)
Assumes that error due to aliasing is insignificant.
erm under the square-root sign is the stability, and is related to the TIE
The t
(Time Interval Error). The errors from this term occur throughout a single-shot
measurement. The second term is a result of both the absolute center-frequency
uracy and the center-frequency stability of the timebase, and varies between
acc
multiple single-shot measurements over the observation interval (the amount of time
from the first single-shot measurement to the final single-shot measurement).
10MSO3000 and DPO3000 Series Specifications and Performance Verification
Sample Rate Range Detail
Table 4: Sample rate range detail
Specifications
Characterist
Sample rate range
(Analog Cha
ic
nnels)
Description
Time/Div
1ns
2ns
4ns
10 ns
20 ns
40 ns
80 ns
100 ns
200 ns
400 ns
800 ns
1 μs
2 μs
4 μs
8 μs
5 M record1 M record100 K record10 K record1 K record
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
1.25 GS/s
s
2.5 GS/
2.5 GS/s500 MS/s
2.5 GS/s250 MS/s
GS/s
1.25
2.5 GS/s100 MS/s
2.5 GS/s500 MS/s50 MS/s
5GS/s
2.
1.25 GS/s
0MS/s
25
25
MS/s
0 μs
1
20 μs
40 μs
80 μs
100 μs
200 μs
400 μs
800 μs
2.5 GS/s100 MS/s10 MS/s
2.5 GS/s500 MS/s50 MS/s5 MS/s
2.5 GS/s250 MS/s25 MS/s2.5 MS/s
1.25 GS/s
2.5 GS/s100 MS/s10 MS/s1 MS/s
2.5 GS/s500 MS/s50 MS/s5 MS/s500 KS/s
1.25 GS/s250 MS/s25 MS/s2.5 MS/s250 KS/s
625 MS/s
MSO3000 and DPO3000 Series Specifications and Performance Verification11
Specifications
Table 4: Sample rate range detail (cont.)
CharacteristicDescription
Sample rate range
(Analog Channels)
(Cont.)
Time/Div
1ms
2ms
4ms
10 ms
20 ms
40 ms
100 ms
200 ms
400 ms
1s
2s
4s
10 s
20 s
40 s
100 s
5 M record1 M record100 K record10 K record1 K record
100 MS/s10 MS/s1 MS /s100 KS/s
250 MS/s50 MS/s5 MS/s500 KS/s50 KS/s
125 MS/s25 MS/s2.5 MS/s250 KS/s25 KS/s
50 MS/s10 MS/s1 MS/s100 KS/s10 KS/s
25 MS/s5 MS/s500 KS/s50 KS/s5 KS/s
12.5 MS/s2.5 MS/s250 KS/s25 KS/s2.5 KS/s
5 MS/s1 MS/s100 KS/s10 KS/s1 KS/s
2.5 MS/s500 KS/s50 KS/s5 KS/s500 S/s
1.25 MS/s250 KS/s25 KS/s2.5 KS/s250 S/s
500 KS/s100 KS/s10 KS/s1 KS/s100 S/s
250 KS/s50 KS/s5 KS/s500 S/s50 S/s
125 KS/s25 KS/s2.5 KS/s250 S/s25 S/s
50 KS/s10 KS/s1 KS/s100 S/s10 S/s
25 KS/s5 KS/s500 S/s50 S/s5 S/s
12.5 KS/s2.5 KS/s250 S/s25 S/s2.5 S/s
5 KS/s1 KS/s100 S/s10 S/s
200 s
400 s
1000 s
2.5 KS/s500 S/s50 S/s5 S/s
1.25 KS/s250 S/s25 S/s2.5 S/s
500 S/s100 S/s10 S/s
12MSO3000 and DPO3000 Series Specifications and Performance Verification
Trigger Specifications
Table 5: Trigger specifications
CharacteristicDescription
Specifications
Trigger level ranges
Trigger level
accuracy, DC
coupled, typical
Lowest frequency
for "Set Level to
50%" function,
typical
Trigger holdoff
range
Trigger sensitivity
SourceSensitivity
Any input channel
Aux In (External)
LineNot applicable
The line trigger level is fixed at about 50% of the line voltage.
This specification applies to logic and pulse thresholds.
For signals having rise and fall times ≥ 10 ns, the limits are as follows:
Source
Any channel±0.20 divisions
Aux In (external trigger)±(10% of setting + 25 mV)
LineNot applicable
45 Hz
20 ns minimum to 8 s maximum
Edge trigger,
DC coupled,
typical
Trigger SourceSensitivity
Any input
channel
±8 divisions from center of screen, ±8 divisions from 0 V when
vertical LF reject trigger coupling is selected
±8 V
Range
0.50 div from DC to 50 MHz, increasing to 1 div at oscilloscope
bandwidth
Aux in
(External)
LineFixed
Edge trigger,
not DC coupled,
typical
Logic (pattern) trigger, DC coupled,
typical:
MSO3000 and DPO3000 Series Specifications and Performance Verification13
Trigger
Coupling
AC1.5 times the DC Coupled limits for frequencies above 10 Hz.
NOISE REJ2.5 times the DC-coupled limits
HF REJ
LF REJ
200 mV from DC to 50 MHz, increasing to 500 mV at 250 MHz
Typical Sensitivity
Attenuates signals below 10 Hz
1.5 times the DC-coupled limit from D C to 50 kHz.
Attenuates signals above 50 kHz
1.5 times the DC-coupled limits for frequencies above 50 kHz.
Attenuates signals below 50 kHz
1.0 division from DC to maximum bandwidth
Specifications
Table 5: Trigger specifications (cont.)
CharacteristicDescription
Aux In (External
trigger)
Edge, Pulse,
and Logic trigger
bandwidth, typical
Trigger using a logic qualifier,
1.0 division from DC to maximum bandwidth
DC coupled, typical:
Delay-by-events sequence trigger,
1.0 division from DC to maximum bandwidth
DC coupled, typical:
Runt trigger, typical:
1.0 division from DC to maximum bandwidth
Pulse-width and glitch trigger, typical:1.0 division
Video trigger,
typical
The limits for both delayed and main trigger are as follows:
SourceSensitivity
Any input
0.6 to 2.5 divisions of video sync tip
channel
Aux In
Video not supported through Aux In (External) input
(External)
Maximum input voltage:
At front panel connector, 300 VRMS, Installation Category II;
Derate at 20 dB/decade above 3 MHz to 30 V
10 dB/decade above 30 MHz
Input termination, typical:
1MΩ ±1% in parallel with 10 pF ±2 pF
Bandwidth, typical:> 200 MHz
For instruments with 500 MHz bandwidth (includes MSO/DPO305X models
as well as MSO/DPO303X/301X models with 500 MHz upgrade):
For instruments with 300 MHz bandwidth (includes MSO/DPO303X models
as well as MSO/DPO301X models with 300 MHz upgrade):
at 30 MHz,
RMS
500 MHz
300 MHz
Time accuracy for
Pulse-width or Glitch
triggering
Video trigger
formats and field
rates
Logic trigger,
minimum logic or
re-arm time, typical
For instruments with 100 MHz bandwidth (MSO/DPO301X models):
100MHz
Time rangeAccuracy
4 ns to 500 ns
520 ns to 8 s
±(20% of setting + 0.5 ns)
±(0.01% of setting + 100 ns)
Triggers from negative sync composite video, field1orfield 2 for interlaced systems, on any field,
specific line, or any line for interlaced or non-interlaced systems. Supported systems include NTSC,
PAL, and SECAM, which are standard, as well as HDTV and custom scan rates, which require a
DPO3VID application module.
For all vertical settings, the minimums are:
Trigger typeMinimum pulse
width
Minimum re-arm
time
Minimum time
between channels
1
LogicNot applicable2 ns2 ns
Time Qualified Logic
4ns2ns2ns
14MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 5: Trigger specifications (cont.)
CharacteristicDescription
Setup/hold time
violation trigger
Specifications
Minimum clock
pulse widths,
typical
Setup and hold
time ranges
Minimum pulse
width and
rearm time
For all vertical settings, the minimum clock pulse widths are:
Clock Active
2
User hold time + 2.5 ns
Clock Inactive
3
2ns
2
Analog ChannelsDigital Channels
MinMaxMinMax
Setup time
Hold time
4
4
Setup + Hold time
0ns8s-0.5ns1.0ms
4 ns8 s1 ns1.0 ms
4
4ns16s0.5ns2.0ms
NOTE. Input coupling on clock and data channels must be the same.
Trigger typeMinimum pulse
Minimum rearm time
width
Glitch
Pulse-width
4ns
4ns
2 ns + 5% of glitch width setting
2 ns + 5% of width upper limit setting
NOTE. For the pulse-width trigger class, pulse-width refers to the width of the
pulse being measured. The rearm time refers to the time between pulses.
4ns2nsRunt
NOTE. For the runt trigger class, pulse width refers to the width of the pulse being
measured. The rearm time refers to the time between pulses.
Rise/fall time trigger,
delta time range
Glitch, pulse-width,
or time-qualified runt
trigger,
time range
B trigger
(A/B sequence
trigger), time range
Time-qualified runt
Rise/fall time
4ns
4ns
8.5 ns + 5% of width setting
8.5 ns + 5% of delta time setting
NOTE. For the rise/fall time trigger class, pulse width refers to the delta time
being measured. The rearm time refers to the time it takes the signal to cross the
two trigger thresholds again.
4nsto8s
4nsto8s
Trigger after events, minimum pulse width, typical:
Trigger after events, maximum event frequency, typical:
5
1 / (2 X Rated Instrument Bandwidth)
5
Rated Instrument Bandwidth
Minimum time between arm and trigger, typical:8 ns
B trigger after time, time range:
B trigger after events, event range:
8nsto8s
1 to 4,000,000
MSO3000 and DPO3000 Series Specifications and Performance Verification15
Specifications
Table 5: Trigger specifications (cont.)
CharacteristicDescription
Standard serial bus
interface triggers
I2C
(Requires a
DPO3EMBD
app. module)
SPI
(Requires a
DPO3EMBD
app. module)
RS-232/422/
485/UART
(Requires a
DPO3COMP
app. module)
Maximum serial trigger bits:128 bits
Address Triggering:
Data Trigger:
Trigger On:Start
Maximum Data Rate:
Data Trigger:
Trigger On:SS Active
Maximum Data Rate:
Data Trigger:Tx D ata, Rx Data
Trigger On:Tx Start Bit
7 and 10 bit user specified address, as well as General Call,
START byte, HS-mode, EEPROM, and CBUS
1 to 5 bytes of user specified data
Repeated Start
Stop, Missing Ack
Address
Data
Address and Data
10 Mb/s
1 to 16 bytes of user specified data
MOSI
MISO
MOSI and MISO
10 Mb/s
Rx Start Bit
Tx End of Packet
Rx End of Packet
Tx Data
Rx Data
Tx Parity Error
Rx Parity Error
Maximum Data Rate:
16MSO3000 and DPO3000 Series Specifications and Performance Verification
10 Mb/s
Table 5: Trigger specifications (cont.)
CharacteristicDescription
Specifications
CAN
(Requires a
DPO3AUTO
app. module)
LIN
(Requires a
DPO3AUTO
app. module)
Data Trigger:
Trigger On:Start of Frame
Frame Type:
Identifier:Standard (11 bit) and Extended (29 bit) identifiers
Maximum Data Rate:
Data Trigger:
Trigger O n :Sync
1 to 8 bytes of user specified data, including qualifiers of equal
to (=), not equal to (<>), less than (<), greater than (>), less than
or equal to (<=), greater than or equal to (>=)
Type of Frame
Identifier, D ata
Identifier and Data
End of Frame
Missing Ack
Bit Stuffing Error
Data, Remote, Error, Overload
1Mb/s
1 to 8 Bytes of user-specified data, including qualifiers of equal
to (=), not equal to (<>), less than (<), greater than (>), less than
or equal to (<=), greater than or equal to (>=)
Identifier
Data
Identifier & Data
Wakeup Frame
Sleep Frame
Error
Maximum Data Rate:
1 Mb/s (by LIN definition, 20 kbit/s)
MSO3000 and DPO3000 Series Specifications and Performance Verification17
Identifier (trigger when equal to (=), not equal to <>, less than
(<), greater than (>), less than or equal to (<=), greater than or
equal to (>=), inside range, outside range)
Cycle Count (trigger when equal to (=), not equal to <>, less
than (<), greater than (>), less than or equal to (<=), greater
than or equal to (>=), inside range, outside range)
Data (number of bits 1–16; byte offset ‘don’t care’ – 253; trigger
10 Mb/swhen equal to (=), not equal to <>, less than (<), greater
than (>), less than or equal to (<=), greater than or equal to (>=),
inside range, outside range)
Command Word (set the following individually: RT Address
(trigger when equal to (=), not equal to <>, less than (<), greater
than (>), less than or equal to (<=), greater than or equal to (>=),
inside range, outside range), T/R, Sub-Address/Mode, Data
Word Count/Mode Code, And Parity)
Status Word (set the following individually: RT address (trigger
when equal to (=), not equal to <>, less than (<), greater than
(>), less than or equal to (<=), greater than or equal to (>=),
inside range, outside range), message error, instrumentation,
service request bit, broadcast command received, busy,
subsystem flag, dynamic bus control acceptance (DBCA),
terminal flag, and parity)
Data Word (user-specified 16–bit data value)
Error (Sync, Parity, Manchester, Non-Contiguous Data
Idle Time (minimum time selectable from 4 μs to 100 μs;
maximum time selectable from 12 μs to 100 μs; trigger on <
minimum, > maximum, inside range, outside range)
Maximum Data Rate:
I2S
(Requires a
DPO3AUDIO
app. module)
18MSO3000 and DPO3000 Series Specifications and Performance Verification
Trigger on:
Maximum Data Rate:
Up to 1Mb/s (for automated decoding of bus)
Word Select
Data
12.5 Mb/s
Table 5: Trigger specifications (cont.)
CharacteristicDescription
Specifications
Left Justified
(Requires a
DPO3AUDIO
app. module)
Data Trigger:
Trigger on:
32 bits of user-specified data in a left word, right word, or either,
including qualifiers of equal to (=), not equal to (<>), less than
(<), greater than (>), less than or equal to (<=), greater than or
equal to (>=), inside range, outside range
Word Select
Data
Right Justifi ed
(Requires a
DPO3AUDIO
app. module)
Maximum Data Rate:
Data Trigger:
Trigger on:
12.5 Mb/s
32 bits of user-specified data in a left word, right word, or either,
including qualifiers of equal to (=), not equal to <>, less than (<),
greater than (>), less than or equal to (<=), greater than or equal
to (>=), inside range, outside range
Word Select
Data
TDM
(Requires a
DPO3AUDIO
app. module)
Maximum Data Rate:
Data Trigger:
Trigger on:
12.5 Mb/s
32 bits of user-specified data in a channel 0-7, including
qualifiers of equal to (=), not equal to <>, less than (<), greater
than (>), less than or equal to (<=), greater than or equal to (>=),
inside range, outside range
Frame Sync
Data
Maximum Data Rate:
1
For logic, time between channels refers to the length of time a logic state derived from more than one channel must exist to be recognized. For Time
Qualified Logic events, the time is the minimum time between a main and delayed event that will be recognized if more than one channel is used.
2
active pulse width is the width of the clock pulse from its active edge (as defined through the Define Inputs button on the lower menu and the
An
ClockEdgebuttononthesidemenu)toitsinactiveedge. Aninactivepulsewidthisthewidthofthepulsefromitsinactiveedgetoitsactiveedge.
3
The User hold time is the number selected by the user through the Setup and Hold trigger menu.
4
Setup + Hold time is the algebraic sum of the Setup Time and the Hold Time programmed by the user.
For Setup time, positive numbers mean a data transition before the clock.
For Hold time, positive numbers mean a data transition after the clock edge.
5
igger after events is the time between the last A trigger event and the first B trigger event.
Tr
Trigger after time is the time between the end of the time period and the B trigger event.
25 Mb/s
MSO3000 and DPO3000 Series Specifications and Performance Verification19
Specifications
Display Specifications
Table 6: Display specifications
Characterist
Display type
Display resolution
Luminance,
typical
ic
Description
9” WVGA LCD display with CCFL backlight
Display Area
800 X 480 pixels, each made up of 3 vertical stripe sub-pixels colored red, green, and blue
400 cd/m
- 196.8 mm (H) X 118.08 mm (V). 230 mm diagonal
2
at IBL = 5.0 mA
Input/Output Port Specifications
Table 7: I
CharacteristicDescription
Ethernet interfaceStandard on all models: 10/100 Mb/s
GPIB in
USB interface1 Device and 2 Host connectors (all models)
nput/Output port specifications
terface
Device port
Host ports
/lamp
rms
ble a s an optional accessory that connects to USB Device and USB Host port,
Availa
with the TEK-USB-488 GPIB to USB Adapter
Control interface is incorporated in the instrument user interface
0 High Speed; also supports Full Speed and Slow Speed Modes
USB 2.
Two USB 2.0 High Speed ports; one on front, one on rear
o signal output
Vide
Probe compensator output voltage and
quency, typical
fre
Trigger (Auxiliary) output (AUX OUT)
A 15 pin, SVGA RGB-type connector
put voltage: 0 V to 2.5 V ±1% behind 1 kΩ ±2%
Out
Frequency: 1 kHz ±20%
W T RUE. A negative pulse indicates that a trigger has occurred.
LO
The logic levels are:
aracteristic
Ch
Vout (HI)≥ 3.25 V open circuit; ≥ 2.2Vintoa50Ω load to ground
Vout (LO)≤ 0.4 V into a load of ≤ 4mA;≤ 0.30 V into a 50 Ω load to ground
Limits
20MSO3000 and DPO3000 Series Specifications and Performance Verification
Power Source Specifications
Table 8: Power source specifications
CharacteristicDescription
Source voltage100 V to 240 V ±10%
Specifications
Source frequency
Fuse ratingT3.15 A, 250 V
Data Storage Specifications
Table 9: Data storage specifications
CharacteristicDescription
Nonvolatile memory retention time,
typical
Real-time clockA programmable clock providing time in years, months, days, hours, minutes, and
Environmental Specifications
100 V to 240 V:
100 V to 132 V:
The fuse is not customer replaceable.
No time limit for front-panel settings, saved waveforms, setups, and calibration
constants
seconds
50/60 Hz
400 Hz ±10%
Table 10: Environmental specifications
CharacteristicDescription
Temperature
Humidity
Pollution DegreePollution Degree 2, indoor use only
Altitude
Operating:0°C to +50 °C (+32 °F to +122 °F)
Nonoperating:
Operating:
5% to 95% relative humidity (% RH) at up to +30 °C,
5% to 45% RH above +30 °C up to +50 °C, non-condensing, and as limited by a Maximum
Wet-Bulb Temperature of +38 °C (derates relative humidity to 45 % RH at +50 °C)
Nonoperating:
5% to 95% RH (Relative Humidity) at up to +30 °C,
5% to 45% RH above +30 °C up to +50 °C, non-condensing, and as limited by a Maximum
Wet-Bulb Temperature of +38 °C (derates relative humidity to 27% RH at +60 °C)
Operating:3,000 m (9,843 ft)
Nonoperating:
-40 °C to +71 °C (-40 °F to +160 °F)
12,000 m (39,370 ft)
MSO3000 and DPO3000 Series Specifications and Performance Verification21
Specifications
Mechanical Specifications
Table 11: Mechanical specifications
Characterist
Dimensions
P6316 D
Table 12: P6316 Digital probe input characteristics
Characteristic
Numbe
Input resistance, typical
Inpu
ic
Description
Height
Handle down203.28.0
Handle up25410.3
Width416.616.4
Depth147.45.8
WeightkgLb.
Stand alone, no front cover
With accessories & carry case
Package
d for domestic shipment
igital Probe Input Characteristics
ption
Descri
r of input channels
t capacitance, typical
16 Digital Inputs
101 KΩ to ground
1
8pF
mm
4.29.2
6.815.0
8.619.0
In.
Minimum Input Signal Swing, typical
DC Input Voltage Range
Maximum Input Dynamic Range50 Vp-p, dependant on threshold setting
1
Specified at the input to the P6316 probe with all eight ground inputs connected to the user's ground. Use of leadsets, grabber clips, ground
extenders, or other connection accessories may compromise this specification.
500 mV
V, -20V
+30
p-p
22MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
ification
Performance V
This chapter contains performance verification procedures for the specifications marked with thesymbol. The following
equipment, or a suitable equivalent, is required to complete these procedures.
DescriptionMinimum requirementsExamples
DC voltage source3 mV to 4 V, ±0.1% accuracy
Leveled sine wave generator
Time mark generator80 ms period, ±1 ppm accuracy, rise
One 50Ω BNC cable
One BNC feed-through terminator50 Ω
You may need additional cables and adapters, depending on the actual test equipment you use.
These procedures cover all MSO/DPO3000 models. Please disregard any checks that do not apply to the specific model
you are testing.
Print the test record on the following pages and use it to record the performance test results for your oscilloscope.
NOTE. Completion of the performance verification procedure does not update the stored time and date of the latest
successful adjustment. The date and time are updated only when the adjustment procedures in the service manual are
successfully completed.
erification
Fluke 9500 Oscilloscope Calibrator with
50 kHz to 1000 MHz, ±4% amplitude
accuracy
time < 50 ns
Male-to-male connectorsTektronix part number
a 9510 Output Module
An appropriate BNC-to-0.1 inch pin
adapter between the Fluke 9500 and
P6316 probe
012-0057-01 (43 inch)
The performance verification procedures verify the performance of your instrument. They do not adjust your instrument. If
your instrument fails any of the performance verification tests, you should perform the factory adjustment procedures as
described in the MSO3000 and DPO3000 Series Service Manual.
MSO3000 and DPO3000 Series Specifications and Performance Verification23
Performance Ver
ification
Upgrade the Firmware
For the best functionality, you can upgrade the oscilloscope firmware. To upgrade the firmware, follow these steps:
1. Open a Web browser and go to www.tektronix.com/software/downloads to locate the most recent firmware upgrade.
2. Download the latest firmware for your oscilloscope onto your PC.
3. Unzip the files and copy the "firmware.img" file into the root folder of a USB flash drive.
4. Power off your oscilloscope.
5. Insert the USB flash drive into a USB Host port on the front or back of the oscilloscope.
6. Power on the oscilloscope. The oscilloscope automatically recognizes the replacement firmware and installs it.
If the instrument does not install the firmware, rerun the procedure. If the problem continues, contact qualified service
personnel.
NOTE. Do not power off the oscilloscope or remove the USB flash drive until the oscilloscope finishes installing the firmware.
7. Power off the oscilloscope and remove the USB flash drive.
8. Power on the oscilloscope.
9. Push the Utility button on the front-panel.
10. Push Utility Page on the lower menu.
11. Turn Multipurpose knob “a” and select Config.
12. Push About on the lower menu. The oscilloscope displays the firmware version number.
13. Confirm that the version number matches that of the new firmware.
The oscilloscope displays a message when the installation is complete.
24MSO3000 and DPO3000 Series Specifications and Performance Verification
Test Record
Print this section for u se during the Performance Verification.
Model numberSerial numberProcedure performed byDate
TestPassedFailed
Self Test
Performance Ver
ification
MSO3000 and DPO3000 Series Specifications and Performance Verification25
Performance Ver
Input Termination Tests
Table 13: Input termination
ification
Performance c
hecks
Channel 1
Termination, 1 MΩ
Channel 1 Input
Terminati
on, 75Ω
Channel 1 Input
Terminat
ion, 50Ω
Channel 2
Termination, 1 MΩ
Channel 2 Input
nation, 75Ω
Termi
Channel 2 Input
nation, 50Ω
Termi
Channel 3
1
Termination, 1 MΩ
Channel 3 Input
rmination, 75Ω
Te
Vertical scal
e
Low limitTest resultHigh limit
10 mV/div990 kΩ1.01 MΩChannel 1 Input
100 mV/div990 kΩ1.01 MΩ
100 mV/div74.25Ω75.75Ω
100 mV/div49.5Ω50.5Ω
10 mV/div990 kΩ1.01 MΩChannel 2 Input
100 mV/
div
990 kΩ1.01 MΩ
100 mV/div74.25Ω75.75Ω
100 mV/div49.5Ω50.5Ω
10 mV/div990 kΩ1.01 MΩChannel 3 Input
100
mV/div
990
kΩ
1.0
1MΩ
100 mV/div74.25Ω75.75Ω
Channel 3 Input
ermination, 50Ω
T
Channel 4
1
100 mV/div49.5Ω50.5Ω
10 mV/div990 kΩ1.01 MΩChannel 4 Input
Termination, 1 MΩ
Channel 4 Input
100 mV/div990 kΩ1.01 MΩ
100 mV/div74.25Ω75.75Ω
Termination, 75Ω
Channel 4 Input
100 mV/div49.5Ω50.5Ω
Termination, 50Ω
1
Channels 3 and 4 are only on four-channel oscilloscopes.
26MSO3000 and DPO3000 Series Specifications and Performance Verification
DC Balance Tests
Table 14: DC Balance
Performance Ver
ification
Performance c
Channel 1
Channel 1 DC Balance,
50Ω,20MHzBW
Channel 1 DC Balance,
75Ω,20M
nel 1 DC Balance,
Chan
1MΩ,20MHzBW
hecks
Hz BW
Vertical scal
1 mV/div
2 mV/div
10 mV/div
100 mV/div
1V/div
1 mV/div
2 mV/div
10 mV/d
100 mV/div
1V/div
1mV/
2 mV/div
10 mV/div
10
1V/div
iv
div
0 mV/div
e
Low limitTest resultHigh limit
-0.5 mV0.5 mV
-0.5 mV0.5 mV
-2.0 mV2.0 mV
-20.0 mV20.0 mV
-200.0 mV200.0 mV
-0.5 mV0.5 mV
-0.5 mV0.5 mV
-2.0 mV2.0 mV
-20.0 mV20.0 mV
0mV
-200.
-0.3 mV0.3 mV
-0.4 mV0.4 mV
0mV
-2.
-20.0 mV20.0 mV
-200.0 mV200.0 mV
200.0
2.0
mV
mV
Channel 1 DC Balance,
50Ω,150MHzBW
Channel 1 DC Balance,
75Ω,150MHzBW
Channel 1 DC Balance,
1MΩ, 150 MHz BW
1 mV/div
mV/div
2
10 mV/div
100 mV/div
1V/div
1 mV/div
2 mV/div
10 mV/div
100 mV/div
1V/div
1 mV/div
2 mV/div
10 mV/div
100 mV/div
1V/div
0.5 mV
-
-0.5 mV0.5 mV
-2.0 mV2.0 mV
-20.0 mV20.0 mV
-200.0 mV200.0 mV
-0.5 mV0.5 mV
-0.5 mV0.5 mV
-2.0 mV2.0 mV
-20.0 mV20.0 mV
-200.0 mV200.0 mV
-0.3 mV0.3 mV
-0.4 mV0.4 mV
-2.0 mV2.0 mV
-20.0 mV20.0 mV
-200.0 mV200.0 mV
.5 mV
0
MSO3000 and DPO3000 Series Specifications and Performance Verification27
Channels 3 and 4 are only on four-channel oscilloscopes.
MSO3000 and DPO3000 Series Specifications and Performance Verification35
Performance Ver
DC Offset Accuracy Tests
Table 17: DC Offset Accuracy
ification
Performance
checksVertical scaleVertical offset
All models
Channel 1
20 MHz BW, 1 MΩ
Channel 2
20 MHz BW, 1 MΩ
1 mV/div
1 mV/div
2 mV/div
2 mV/div
10 mV/div
10 mV/div
100 mV/div
100 mV/
1V/div
1V/div
1.01
1.01 V/div
1 mV/div
1m
2 mV/div
div
V/div
V/div
700 mV696.2 mV703.8 mV
-700 mV-703.8 mV-696.2 mV
700 m696.1 mV703.9 mV
-700 mV-703.9 mV-696.1 mV
1 V993 mV1007 mV
-1 V-1007 mV-993 mV
10.0 V9.930 V10.07 V
-10.0 V-10.07 V-9.930 V
100 V99.30 V100.7 V
-100 V-100.
100 V99.30 V100.7 V
-100 V-100.7 V-99.30 V
mV
700
-700 mV-703.8 mV-696.2 mV
700 mV696.1 mV703.9 mV
1
Low limitTest resultH igh limit
7V
.2 mV
696
-99.3
703
0V
.8 mV
2 mV/div
0 mV/div
1
10 mV/div
100 mV/div
100 mV/div
1V/div
1V/div
1.01 V/div
1.01 V/div
700 mV
-
1 V993 mV1007 mV
-1 V-1007 mV-993 mV
10.0 V9.930 V10.07 V
-10.0 V-10.07 V-9.930 V
100 V99.30 V100.7 V
-100 V-100.7 V-99.30 V
100 V99.30 V100.7 V
-100 V-100.7 V-99.30 V
703.9 mV
-
696.1 mV
-
36MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 17: DC Offset Accuracy (cont.)
Performance Ver
ification
Performance
checksVertical scaleVertical offset
Channel 3
2
20 MHz BW, 1 MΩ
Channel 4
2
20 MHz BW, 1 MΩ
1 mV/div
1 mV/div
2 mV/div
2 mV/div
10 mV/div
10 mV/div
100 mV/div
100 mV/div
1V/div
1V/div
1.01 V/div
1.01 V/div
1 mV/div
1 mV/div
2 mV/div
2 mV/div
700 mV696.2 mV703.8 mV
-700 mV-703.8 mV-696.2 mV
700 mV696.1 mV703.9 mV
-700 mV-703.9 mV-696.1 mV
1 V993 mV1007 mV
-1 V-1007 mV-993 mV
10.0 V9.930 V10.07 V
-10.0 V-10.07 V-9.930 V
100 V99.30 V100.7 V
-100 V-100.7 V-99.30 V
100 V99.30 V100.7 V
-100 V-100.7 V-99.30 V
700 mV696.2 mV703.8 mV
-700 mV-703.8 mV-696.2 mV
700 mV696.1 mV703.9 mV
-700 mV-703.9 mV-696.1 mV
1
Low limitTest resultHigh limit
10 mV/div
10 mV/div
100 mV/div
100 mV/div
1V/div
1V/div
1.01 V/div
1.01 V/div
1
Use this value for both the calibrator output and the oscilloscope offset setting.
2
Channels 3 and 4 are only on four-channel oscilloscopes.
1 V993 mV1007 mV
-1 V-1007 mV-993 mV
10.0 V9.930 V10.07 V
-10.0 V-10.07 V-9.930 V
100 V99.30 V100.7 V
-100 V-100.7 V-99.30 V
100 V99.30 V100.7 V
-100 V-100.7 V-99.30 V
MSO3000 and DPO3000 Series Specifications and Performance Verification37
Performance Ver
Performance Checks
Table 18: Performance
ification
Performance c
Sample Rate and Delay Time
Accuracy
Auxiliary (
hecks
Low limitTest resultHigh limit
-0.5 division+2 division
Trigger) Output Tests
Table 19: Auxiliary (Trigger) Output
Auxiliary (Trigger) Output
Trigger O
utput
Low 1 MΩ
High 1 MΩ
Low 50Ω
High 50Ω
Low limitTest resul
—
≥ 3.25 V
—
≥ 2.2 V
t
High limit
≤ 0.4 V
—
≤ 0.30 V
—
38MSO3000 and DPO3000 Series Specifications and Performance Verification
Random Noise, Sample Acquisition Mode Tests
Table 20: Random Noise, Sample Acquisition Mode
Performance Ver
ification
Random Noise,
Sample Acquisition Mode
For instruments with
500 MHz bandwidth
(includes
MSO/DPO305X modelsaswellasMSO/DPO303X/301
X models
with 500 MHz upgrade)
For instruments with
300 MHz bandwidth
des
(inclu
MSO/DPO303X
models as well
as MSO
/DPO301X
models with 300 MHz
upgrade)
Channel 1
Channel 2
Channel 3
Channel 4
Channel 1
Channel 2
Channel 3
Bandwidth Sel
ection
Test resultHigh limit
Full8.170 mV
150 MHz6.090 mV
20 MHz6.025 mV
Full8.170 mV
150 MHz6.090 mV
20 MHz6.025 mV
1
Full8.170 mV
150 MHz6.090 mV
20 MHz6.025 mV
1
Full8.170 mV
150 MHz6.090 mV
20 MHz6.025 mV
Full6.140 mV
150 MHz6.080 mV
20 MHz5.030 m
V
Full6.140 mV
150 MHz6.080 mV
20 MHz5.030
1
Full6.140 mV
mV
150 MHz6.080 mV
Channel 4
Hz
20 M
1
Full6.140 mV
5.0
30 mV
150 MHz6.080 mV
20
MHz
030 mV
5.
MSO3000 and DPO3000 Series Specifications and Performance Verification39
Performance Ver
Table 20: Random Noise, Sample Acquisition Mode (cont.)
Random Noise, Sample Acquisition ModeBandwidth SelectionTest resultHigh limit
ification
For instruments with
Channel 1
100 MHz bandwidth
(MSO/DPO301X
models)
1
Channels 3 and 4 are only on four-channel oscilloscopes.
Channel 2
Channel 3
Channel 4
1
1
Full6.100 mV
20 MHz6.100 mV
Full6.100 mV
20 MHz6.100 mV
Full6.100 mV
20 MHz6.100 mV
Full6.100 mV
20 MHz6.100 mV
40MSO3000 and DPO3000 Series Specifications and Performance Verification
Delta Time Measurement Accuracy Tests
Table 21: Delta Time Measurement Accuracy
Channel 1
MSO/DPO = 4 ns/Div, Source frequency = 240 MHz
Performance Ver
ification
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV240 ps
100 mV800 mV240 ps
500 mV4 V240 ps
1 V4 V240 ps
MSO/DPO = 40 ns/Div, Source frequency = 24 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV450 ps
100 mV800 mV360 ps
500 mV4 V360 ps
1 V4 V590 ps
MSO/DPO = 400 ns/Div, Source frequency = 2.4 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.8ns
100 mV800 mV2.8 ns
500 mV4 V2.8 ns
1V4V5.4ns
MSO/DPO = 4 μs/Div, Source frequency = 240 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV38ns
100 mV800 mV28 ns
500 mV4 V28 ns
1V4V54ns
MSO/DPO = 40 μs/Div, Source frequency = 24 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV380 ns
100 mV800 mV280 ns
500 mV4 V280 ns
1 V4 V540 ns
MSO/DPO = 40 μs/Div, Source frequency = 2.4 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
MSO3000 and DPO3000 Series Specifications and Performance Verification41
Performance Ver
Table 21: Delta Time Measurement Accuracy (cont.)
Channel 2
ification
5mV40mV3.8μs
100 mV800 mV2.8 μs
500 mV4 V2.8μs
1V4V5.4μs
MSO/DPO = 4 ns/Div, Source frequency = 240 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV240 ps
100 mV800 mV240 ps
500 mV4 V240 ps
1 V4 V240 ps
MSO/DPO = 40 ns/Div, Source frequency = 24 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV450 ps
100 mV800 mV360 ps
500 mV4 V360 ps
1 V4 V590 ps
MSO/DPO = 400 ns/Div, Source frequency = 2.4 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.8ns
100 mV800 mV2.8 ns
500 mV4 V2.8 ns
1V4V5.4ns
MSO/DPO = 4 μs/Div, Source frequency = 240 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV38ns
100 mV800 mV28 ns
500 mV4 V28 ns
1V4V54ns
MSO/DPO = 40 μs/Div, Source frequency = 24 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV380 ns
100 mV800 mV280 ns
500 mV4 V280 ns
42MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 21: Delta Time Me asurement Accuracy (cont.)
1 V4 V540 ns
MSO/DPO = 40 μs/Div, Source frequency = 2.4 kHz
Performance Ver
ification
Channel 3
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.8μs
100 mV800 mV2.8 μs
500 mV4 V2.8 μs
1V4V5.4μs
1
MSO/DPO = 4 ns/Div, Source frequency = 240 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV240 ps
100 mV800 mV240 ps
500 mV4 V240 ps
1 V4 V240 ps
MSO/DPO = 40 ns/Div, Source frequency = 24 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV450 ps
100 mV800 mV360 ps
500 mV4 V360 ps
1 V4 V590 ps
MSO/DPO = 400 ns/Div, Source frequency = 2.4 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.8ns
100 mV800 mV2.8 ns
500 mV4 V2.8 ns
1V4V5.4ns
MSO/DPO = 4 μs/Div, Source frequency = 240 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV38ns
100 mV800 mV28 ns
500 mV4 V28 ns
1V4V54ns
MSO/DPO = 40 μs/Div, Source frequency = 24 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
MSO3000 and DPO3000 Series Specifications and Performance Verification43
Performance Ver
Table 21: Delta Time Measurement Accuracy (cont.)
ification
5 mV40 mV380 ns
100 mV800 mV280 ns
500 mV4 V280 ns
1 V4 V540 ns
MSO/DPO = 40 μs/Div, Source frequency = 2.4 kHz
Channel 4
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.8μs
100 mV800 mV2.8 μs
500 mV4 V2.8 μs
1V4V5.4μs
1
MSO/DPO = 4 ns/Div, Source frequency = 240 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV240 ps
100 mV800 mV240 ps
500 mV4 V240 ps
1 V4 V240 ps
MSO/DPO = 40 ns/Div, Source frequency = 24 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 mV450 ps
100 mV800 mV360 ps
500 mV4 V360 ps
1 V4 V590 ps
MSO/DPO = 400 ns/Div, Source frequency = 2.4 MHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.8ns
100 mV800 mV2.8 ns
500 mV4 V2.8 ns
1V4V5.4ns
MSO/DPO = 4 μs/Div, Source frequency = 240 kHz
MSO/DPO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV38ns
100 mV800 mV28 ns
500 mV4 V28 ns
44MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 21: Delta Time Me asurement Accuracy (cont.)
1V4V54ns
MSO/DPO = 40 μs/Div, Source frequency = 24 kHz
Performance Ver
ification
MSO/DPO V/DivSource V
5 mV40 mV380 ns
100 mV800 mV280 ns
500 mV4 V280 ns
1 V4 V540 ns
MSO/DPO = 40 μs/Div, Source frequency = 2.4 kHz
MSO/DPO V/DivSource V
5mV40mV3.8μ3
100 mV800 mV2.8 μ3
500 mV4 V2.8 μ3
1V4V5.4μ3
1
Channels 3 and 4 are only on four-channel oscilloscopes.
pp
pp
Test ResultHigh Limit
Test ResultHigh Limit
MSO3000 and DPO3000 Series Specifications and Performance Verification45
Performance Ver
Digital Threshold Accuracy Tests
Table 22: Digital Threshold Accuracy
Digital Threshold Accuracy (MSO3000 series only)
ification
Digital
channelThresholdV
0 V-0.1 V0.1 VD0
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD1
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD2
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD3
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD4
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD5
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD6
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD7
4 V3.78 V4.22 V
Test result
s-
V
s+
Low limit
V
sAvg
=(V
)/2High limit
s--+Vs+
0 V-0.1 V0.1 VD8
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD9
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD10
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD11
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD12
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD13
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD14
4 V3.78 V4.22 V
46MSO3000 and DPO3000 Series Specifications and Performance Verification
Table 22: Digital Threshold Accuracy (cont.)
Digital Threshold Accuracy (MSO3000 series only)
Performance Ver
ification
Digital
channelThresholdV
0 V-0.1 V0.1 VD15
4 V3.78 V4.22 V
Test result
s-
V
s+
Low limit
V
sAvg
=(V
)/2High limit
s--+Vs+
MSO3000 and DPO3000 Series Specifications and Performance Verification47
Performance Ver
ification
Performance Verification Procedures
The following three conditions must be met prior to performing these procedures:
1. The oscilloscope must have been operating continuously for twenty (20) minutes in an environment that meets the
operating range specifications for temperature and humidity.
2. You must perform a signal path compensation (SPC). (See Self Tests — System Diagnostics and Signal Path
Compensation section below.) If the operating temperature changes by more than 10 °C (18 °F), you must perform the
signal path compensation again.
3. You must connect the oscilloscope and the test equipment to the same AC power circuit. Connect the oscilloscope
and test instruments into a common power strip if you are unsure of the AC power circuit distribution. Connecting the
oscilloscope and test instruments into separate AC power circuits can result in offset voltages between the equipment,
which can invalidate the performance verification procedure.
The time required to complete all the procedures is approximately one hour.
WARNING. Some procedures use hazardous voltages. To prevent electrical shock, always set voltage source outputs
to 0 V before making or changing any interconnections.
Self Tests — System Diagnostics and Signal Path Compensation
These procedures use internal routines to verify that the oscilloscope functions and passes its internal self tests. No test
equipment or hoo
Run the System Di
1. Disconnect all
2. Push Default Se
3. Push Utility .
4. Push Utility P
5. Push Self Test
6. Push OK Run Se
7. Wait while th
8. Cycle the osc
Run the sign
1. Push Defaul
2. Push Utilit
3. Push Utili
4. Turn Multi
kups are required. Start the self test with these steps:
agnostics (may take several minutes):
probes and cables from the oscilloscope inputs.
tup on the front-panel to set the instrument to the factory default settings.
age on the lower menu, and turn Multipurpose knob “a” to select Self Test.
on the lower menu. The Loop X Times side menu button will be set to Loop 1 Times.
lf Test on the side menu.
e s elf test runs. When the self test completes, a dialog box displays the results of the self test.
illoscope power off and back on before proceeding.
al path compensation routine (may take 5 to 15 minutes):
t Setup on the front panel.
y .
ty Page on the lower menu.
purpose knob “a” to select Calibration.
5. Push Signa
6. Push OK-C
48MSO3000 and DPO3000 Series Specifications and Performance Verification
l Path on the lower menu.
ompensate Signal Paths on the side menu.
Performance Ver
ification
7. When the signal path c ompensation is complete, push Menu Off twice to clear the dialog box and Self Test menu.
8. Check the Signal Path button on the lower menu to verify that the status is Pass. If it does not pass, then recalibrate the
instrument or
have the instrument serviced by qualified service personnel.
This complete
s the procedure.
MSO3000 and DPO3000 Series Specifications and Performance Verification49
Performance Ver
Check Input Termination, DC Coupled (Resistance)
This test checks the Input Termination for 1 MΩ,75Ω or 50Ω settings.
1. Connect the output of the oscilloscope calibrator (for example, Fluke 9500) to the oscilloscope channel 1 input, as
shown below.
ification
2. Push Defa
3. Push the c
record (for example, 1, 2, 3,or4).
4. Confirm that the oscilloscope termination and calibrator impedance are both set to 1 MΩ. The default Termination
setting is 1MΩ.
5. Turn the Vertical Scale knob to set the vertical scale, as shown in the te st record (for example, 10 mV/div, 100 mV/div, 1
V/div).
6. Measur
7. Repeat
8. Change
9. Change
10. Repea
testing, as shown in the test record (for example, 2, 3,or4).
This completes the procedure.
ult Setup on the front panel to set the instrument to the factory default settings.
hannel button on the front panel for the oscilloscope channel that you are testing, as shown in the test
(See page 26, Input Termination Tests.)
e the input resistance of the oscilloscope with the calibrator. Record this value i n the test record.
steps 5 and 6 for each volt/division setting in the test record.
the o scilloscope termination to 75 Ω and calibrator impedance to 50 Ω and repeat steps 5 through 7.
the oscilloscope termination to 50 Ω and repeat steps 5 through 7.
t steps 4 through 9 for each channel listed in the test record and relevant to the model of oscilloscope that you are
50MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
ification
Check DC Balance
This test checks the DC balance.
You do not need to connect the oscilloscope to any equipment to run this test. The only piece of equipment needed is a
BNC feed-through 50 Ω terminator.
1. For 50 Ω coupling, attach a 50 Ω terminator to the channel input of the oscilloscope being tested.
2. Push Default Setup on the front panel to set the instrument to the factory default settings.
3. Push the channel button on the front panel for the oscilloscope channel that you are testing, as shown in the test
record (for example, 1, 2, 3,or4).
4. Set the oscilloscope termination to 50 Ω.PushTermination on the lower menu to select 50 Ω.
5. Push Bandwidth on the lower menu, and push the appropriate bandwidth button on the side menu for 20MHz, 150MHz,
or Full, as given in the test record.
6. Turn the Horizontal Scale k
NOTE. Step 6 only needs to be done once, at the beginning of the test.
7. Turn the Vertical Scale knob to set the vertical scale, as shown in the test record (for example, 1 mV/div, 2 mV/div,
10 mV/div, 100 mV/div, 1 V
8. Push Acquire on the front panel.
NOTE. Steps 8, 9, and 10 only need to be performed once, at the beginning of this test.
9. Push Mode on the lower menu, and then, if needed, push Average on the side menu.
10. If needed, adjust the number of averages to 16 using Multipurpose knob “a” .
11. Push the Trigger Menu button on the front panel.
NOTE. Steps 11, 12, and 13 only need to be performed once, at the beginning of this test.
12. Push Source on the lower menu.
13. Select the AC Line trigger source on the side menu using Multipurpose knob “a”. You do not need to connect an
external signal to the oscilloscope for this DC Balance test.
nob to 1 ms/division.
/div).
14. On the front panel, push the Measure button on the Wave Inspector.
MSO3000 and DPO3000 Series Specifications and Performance Verification51
Performance Ver
NOTE. Steps 14 though 17 must be performed once for each input channel under test.
ification
15. Push Add Measu
16. Use Multipurp
the channel input being tested.
17. Push OK Add Measurement on the side menu, and then Menu Off on the front panel.
18. View the mean measurement value in the display and enter that mean value as the test result in the test record. (See
page 27, DC Balance Tests.)
19. Repeat step 7 and step 18 for each volts/division value listed in the results table.
20. Push the channel button on the front panel, then change the oscilloscope bandwidth (for example, 20 MHz, 150 MHz, or
Full), and
21. For 1 MΩ cou
22. Repeat ste
oscilloscope (for example, 1, 2, 3,or4).
NOTE. The BNC 50
23. For 75 Ω coupling, change the oscilloscope termination to 75 Ω and repeat steps 5 through 20.
NOTE. T
he BNC 50
rement on the lower menu.
ose knob “b” to select the Mean measurement. If needed, use Multipurpose knob “b” to select
repeat step 7, step 18, and step 19.
pling, change the oscilloscope termination to 1 MΩ and repeat steps 5 through 20.
ps 3 through 20 for each channel combination listed in the test record and relevant to your model of
Ω
terminator needs to be moved to next input channel.
Ω
terminator needs to be moved to next input channel.
24. Repea
NOTE. The BNC 50
This completes the procedure.
t steps 3 through 20 for each channel combination listed in the test record and relevant to your model of
oscilloscope (for example, 1, 2, 3,or4).
Ω
terminator needs to be moved to next input channel.
52MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
Check Analog Bandwidth, 50Ω
This test checks the bandwidth at 50 Ω for each channel.
1. Connect the output of the leveled sine wave generator (for example, Fluke 9500) to the oscilloscope channel 1 input as
shown below.
ification
2. Push Defa
3. Push chan
4. Set the ca
5. Set the o
6. Turn the
ult Setup on the front panel to set the instrument to the factory default settings.
nel button 1,2,3,or4 for the channel that you want to check.
librator to 50 Ω output impedance (50 Ω source impedance) and to generate a sine wave.
scilloscope termination to 50 Ω.PushTermination on the lower menu to select 50 Ω.
Vertical Scale knob to set the vertical scale, as shown in the test record (for example, 1 mV/div, 2 mV/div, 5
mV/div).
7. Push Acquire on the front panel.
8. Confirm that the mode is set to Sample. If not, push Mode on the lower menu, if needed, and then push the Sample
side bezel button.
9. Adjust the signal source to at least 6 vertical divisions at the selected vertical scale with a set frequency of 50 kHz. For
e, at 5 mV/div, use a ≥ 30 mV
exampl
signal; at 2 mV/div, use a ≥ 12 mV
p-p
signal; at 1 mV/div, use a ³ 6 mV
p-p
p-p
signal. Use a sine wave for the signal source.
10. Turn the Horizontal Scale knob to 40 μs/division.
11. On the front panel, push the Measure button on the Wave Inspector, and then push Add Measurement on the lower
menu.
. Steps 11 though 14 must be performed once for each input channel under test.
NOTE
12. Use Multipurpose knob “b” to select the Peak-to-peak measurement. Use Multipurpose knob "a" to select the input
channel being tested, and then push OK Add Measurement on the side menu.
13. Push More on the lower menu to select Gating, and then push Off (Full Record) on the side menu.
14. Push Menu Off on the front panel. This will allow you to see the display. Note the mean V
ading V
re
cord the mean value of V
Re
rn the Horizontal Scale knob to 10 ns/division.
15.Tu
in-pp
.
(for example, 816 mV) in the test record. (See page 33, Bandwidth Tests.)
in-pp
of the signal. Call this
p-p
MSO3000 and DPO3000 Series Specifications and Performance Verification53
Performance Ver
16. Adjust the signal source to the maximum bandwidth frequency for the bandwidth and model desired, as shown in the
following work
mean V
ification
sheet. Measure V
. Call this reading V
p-p
of the signal on the oscilloscope using statistics, as in the previous step, to get the
p-p
.
bw-pp
Record the value of V
in the test record.
bw-pp
NOTE. For more information on the contents of this w orksheet, refer to the Analog Channel Input and Vertical Specifications
table. (See p
age 1, Analog Channel Input And Vertical Specifications.)
Table 23: Maximum Bandwidth Frequency worksheet
Terminatio
n
Vertical Sc
For instruments with 500MHz bandwidth
(includes MSO/DPO305X models as well as MSO/DPO303X/301X models with 500 MHz upgrade):
50Ω10 mV/div
50Ω5 mV/div
50Ω2 mV/div
50Ω1 mV/div
For instruments with 300MHz bandwidth
es MSO/DPO303X models as well as MSO/DPO301X models with 500 MHz upgrade):
(includ
50Ω5 mV/div
50Ω2 mV/div
50Ω1mV/d
ale
Maximum Ban
dwidth Frequency
500 MHz
400 MHz
250 MHz
150 MHz
300 MHz
250 MHz
iv
150 MHz
For instruments with 100MHz bandwidth
(MSO/DPO301X models):
z
50Ω1 mV/div
17. Use t
he values of V
bw-pp
andV
obtained above and stored in the test record to calculate the Gain at bandwidth
in-pp
100 MH
with the following equation:
Gain = V
bw-pp/Vin-pp
.
18. To pass the performance measurement test, Gain should be ≥ 0.707. Enter Gain in the test record.
19. Repeat steps 9 through 17 for the other oscilloscope volts/div settings listed in the test record.
20. Repeat steps 3 through 18 for each channel combination listed in the test record and relevant to your model of
oscilloscope (for example, 1, 2, 3,or4).
This completes the procedure.
54MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
Check DC Gain Accuracy
This test checks the DC gain accuracy.
1. Connect the oscilloscope to a DC voltage source. If using the Fluke 9500 calibrator, connect the calibrator head to
the oscilloscope channel to test.
2. Push Default Setup on the front panel to set the instrument to the factory default settings.
3. Push channel button 1,2,3,or4 to select the channel that you want to check.
4. Confirm that the oscilloscope termination and calibrator impedance are both set to 1 MΩ. On the oscilloscope, push
Termination on the lower menu to select 1MΩ.
ification
5. Push 20 MHz on the side menu to select the bandwidth (push Bandwidth on the lower menu, if necessary, to activate
the Band
6. Push Ac
7. Push Mo
8. On the f
9. Use Mul
width menu).
quire on the front panel.
de on the lower menu, and then push Average on the side menu. Use the default number of averages (16).
ront panel, push the Measure button on the Wave Inspector, and then Add Measurement on the lower menu.
tipurpose knob “b” to select the Mean measurement. Use Multipurpose knob “a” to select the input
channel to be tested.
10. Push OK Add Measurement on the side menu.
11. Push the Trigger Menu button on the front panel.
12. Push Source on the lower menu.
13. Turn Multipurpose knob “a” to select AC Line as the trigger source. Push Menu Off on the front panel.
14. Turn the vertical Scale knob to the next setting to measure, as shown in the Gain Expected worksheet below.
15. Set the DC Voltage Source to V
.PushMeasure on the front panel, then push More on the lower menu to select
negative
Statistics. Push Reset Statistics on the side menu, and then push Menu Off on the front panel.
16. Enter the mean reading into Gain Expected worksheet below as V
17. Set the DC Voltage Source to V
the side menu, and then push Menu Off on the front panel. Enter the mean reading into the Gain Expected
on
worksheet as V
positive-measured
.
.PushMore on the lower menu to select Statistics, push the Reset Statistics
positive
negative-measured
.
MSO3000 and DPO3000 Series Specifications and Performance Verification55
Performance Ver
Table 24: Gain Expected worksheet
Oscilloscope
Vertical
Scale SettingV
1 mV/div
ification
diffExpected
V
negative
V
positive
7 mV-3.5 mV+3.5 mV
V
negative-
measured
V
positive-
measured
Test Result
(Gain
V
diff
Accuracy)
2 mV/div
14 mV-7 mV+7 mV
4.98 mV34.86 mV-17.43 mV+17.43 mV
5 mV35 mV-17.5 mV+17.5 mV
10 mV70 mV-35 mV+35 mV
20 mV140 mV-70 mV+70 mV
49.8 mV348.6 mV-174.3 mV+174.3 mV
50 mV350 mV-175 mV+175 mV
100 mV700 mV-350 mV+350 mV
200 mV1400 mV-700 mV+700 mV
500 mV3500 mV-1750 mV+1750 mV
1.0 V7000 mV-3500 mV+3500 mV
ate V
18. Calcul
V
diff
Enter
19. Calcu
Gain
=|V
late GainAccuracy as follows:
Accuracy =((V
as follows:
diff
negative-measured
in the Gain Expected worksheet.
V
diff
- V
positive-measured
- V
diff
diffExpected
|
)/V
diffExpected
) X 100%
edownGainAccuracy in the Gain Expected worksheet and in the test record. (See page 34, DC Gain Accuracy
Writ
Tes t s .)
20. Repeat steps 14 through 18 for each volts/division value in the test record.
21. Repeat steps 3 through 19 for each channel of the oscilloscope that you want to check.
This completes the procedure.
56MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
Check Offset Accuracy
This test checks the offset accuracy.
1. Connect the oscilloscope to a DC voltage source to run this test. If using the Fluke 9500 calibrator as the DC voltage
source, connect the calibrator head to the oscilloscope channel to test.
2. Push Default Setup on the front panel to set the instrument to the factory default settings.
3. Push channel button 1,2,3,or4 to select the channel you want to check.
ification
4. Confirm that the oscilloscope termination and calibrator
lower menu to select 1MΩ.
5. Set the calibrator to the vertical offset value shown in the test record (for example, 700 mV for a 1 mV/div setting). Set
the calibrator impedance to match the termination setting for the oscilloscope.
6. On the oscilloscope, push More on the lower menu repeatedly, to select Offset.
7. Set the oscilloscope to the vertical offset value shown in the test record (for example, 700 mV for a 1 mV/div setting).
8. Turn the vertical Scale knob to match the value in the test record (for example, 1 mV/division).
9. Turn the Horizontal Scale knob to 1 ms/div.
10. Push Bandwidth on the lower menu.
11. Push 20 MHz on the side menu.
12. Check that the vertical position is set to 0 divs. If not, turn the appropriate Vertical P osition knob to set the position to
0 divs.
Or, push More on the lower menu repeatedly to select Position, and then push Set to 0 divs on the side menu.
13. Push Acquire on the front panel.
14. Push Mode on the lower menu, and then pus
15. Push the Trigger Menu button on the front panel.
16. Push Source on the lower menu.
h Average on the side menu. Use the default number of averages (16).
impedance are both set to 1 MΩ.PushTermination on the
17. Turn Multipurpose knob “a” to select AC Line as the trigger source.
18. On the front panel, push the Measure button on the Wave Inspector.
19. Push
20. Use Multipurpose knob “b” to select the Mean measurement. Use Multipurpose knob “a” to select the input
MSO3000 and DPO3000 Series Specifications and Performance Verification57
Add Measurement on the lower menu.
channel to be tested.
Performance Ver
21. Push OK Add Measureme nt on the side menu, and then Menu Off on the front panel. The mean value should appear in
a measurement p
ification
ane at the bottom of the display.
22. Enter the meas
23. Repeat the pro
24. Repeat all ste
This complet
es the procedure.
ured value in the test record. (See page 36, DC Offset Accuracy Tests.)
cedure ( steps 6, 7, 8 and 22) for each volts/division setting shown in the test record.
ps, starting with step 1, for each oscilloscope channel you want to check.
58MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
Random Noise, Sample Acquisition Mode
This test checks random noise. You do not need to connect any test equipment to the oscilloscope for this test.
1. Disconnect everything connected to the oscilloscope inputs.
2. Push Default Setup on the front panel to set the instrument to the factory default settings. This sets the oscilloscope to
Channel 1, Full Bandwidth, 1 MΩ input termination, 100 mV/div, and 4.00 μs/div.
3. Set the input termination to 50 Ω:
ification
a. Push channel 1, 2, 3, or 4
1
to display the channel input menu.
b. Push Termination on the lower menu to select 50 Ω .
4. Set Gating to Off:
a. On the front panel, push the Measure button on the Wave Inspector.
b. Push More on the lower menu to select Gating.
c. Push Off (Full Record) on the side menu.
5. Select the RMS measurement:
a. Push Add Measurement on the lower menu.
b. Use Multipurpose knob “b” to select the RMS measurement.
c. If necessary, use Multipurpose knob “a” to select the channel being tested as the source for the RMS measurement.
d. Push OK Add Measurement on the side menu.
6. Push More on the lower menu to select Statistics, and then push Reset Statistics on the side menu.
7. Push Menu Off on the front panel to remove the menus from the display.
8. Read the RMS Mean value. This is the Sampled Mean Value (SMV).
9. Push the Horizontal Acquire button and then, if necessary, push Mode on the lower menu to display the Acquisition
Mode menu.
10. Push Average on the side menu, and if necessary, use Multipurpose knob “a” to set the number of averages to 16.
11. On the front panel, push the Measure button on the Wave Inspector. Then, if necessary, push More on the lower menu
to select Statistics, and then push Reset Statistics on the side menu.
12. Push Menu Off on the front panel to remove the menus from the display.
13. Read the RMS Mean value. This is the Averaged Mean Value (AMV).
14. Calculate the RMS noise (RMS noise = SMV - AMV), and enter the calculated RMS noise into the appropriate segment of
the Random Noise, Sample Acquisition Mode test record. (See page 39, Random Noise, Sample Acquisition Mode Tests.)
Set the Acquisition Mode to Sample:
15.
a. Push the Horizontal Acquire button on the front panel.
b. If necessary, push Mode on the lower menu to display the Acquisition Mode menu.
c. Push Sample on the side menu.
MSO3000 and DPO3000 Series Specifications and Performance Verification59
Performance Ver
16. Set the bandwidth to 150 MHz:
a. Push channel 1, 2, 3, or 4
b. Push Bandwidth on the lower menu, and then push 150MHz on the side m enu.
17. On the front panel, push the Measure button on the Wave Inspector, and repeat steps 6 through 15.
18. Set the channel under test bandwidth to 20 MHz:
a. Push channel 1, 2, 3, or 4
b. Push Bandwidth on the lower menu, and then push 20MHz on the side menu.
19. On the front panel, push the Measure button on the Wave Inspector, and repeat steps 6 through 15.
20. Select the next channel to test.
21. Repeat steps 3 through 20 until all channels have been tested.
1
Channels 3 and 4 are only available on three or four channel oscilloscopes.
This completes the procedure.
ification
1
to select the channel you want to test.
1
.
60MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
ification
Check Digital Threshold Accuracy (MSO3000 models only)
For the MSO3000 series only, this test checks the threshold accuracy of the digital channels. This procedure applies to
digital channels D0 through D15, and to channel threshold values of 0 V and +4 V.
1. Connect the P6316 digital probe to the MSO3000 series instrument.
2. Connect the P6316 Group 1 pod to the DC voltage source to run this test. You will need a BNC-to-0.1 inch pin adapter to
complete the connection.
NOTE. If using the Fluke 9500 calibrator as the DC voltage source, connect the calibrator head to the P6316 Group 1 pod.
You will need a BNC-to-0.1 inch pin adapter to complete the connection.
3. Push Default Setup on the front panel to set the instrument to the factory default settings.
4. Push D15-D0 on the front panel.
5. Push D15-D0 On/Off on the lower menu.
6. Push Turn On D7 - D0 and Tur n On D15 - D8 on the side menu. The instrument will display the 16 digital channels.
7. Push Thresholds on the lower menu.
8. Before you change the threshold value, push Fine on the front panel to turn off the fine adjustment and make adjusting
the value quicker.
Turn M ultipu rpo se knob “a” (for channels D7 - D0) or Multipurpose knob “b” (for channels D15 - D8) to set the
threshold value to 0.00 V (0 V/div).
The thresholds are set for the 0 V threshold check. You need to record the test values in the test record row for 0 V
for each digital channel. (See page 46, Digital Threshold A ccuracy Tests.)
9. Push the Trigger Menu button on the front panel.
10. Push Source on the lower menu, and then turn Multipurpose knob “a” to select the appropriate channel, such as D0.
By default, the Type is set to Edge, Coupling is set to DC, Slope is set to Rising, Mode is set to Auto, and Level is
set to match the threshold of the channel being tested.
11. Set the DC voltage source (Vs) to -400 mV. Wait 3 seconds. Check the logic level of the corresponding digital channel in
the display.
MSO3000 and DPO3000 Series Specifications and Performance Verification61
Performance Ver
If the channel is a static logic level high (green), change the DC voltage source Vs to -500 mV.
12. Increment Vs by +20 mV. Wait 3 seconds and check the logic level of the corresponding digital channel in the display. If
the channel is
ification
at a static logic level high (green), record the Vs value as in the 0 V row of the test record.
If the channel
is a logic level low (blue) or is alternating between high and low, repeat this step (increment Vs by 20 mV,
wait 3 seconds, and check for a static logic high). Continue until a value for Vs- is found.
NOTE. In this procedure, the channel might not change state until after you pass the set threshold level.
13. Push Slope o
14. Set the DC vo
n the lower m enu to change the slope to Falling.
ltage source (Vs) to +400 mV. Wait 3 seconds. Check the logic level of the corresponding digital channel in
the display.
If the channel is a static logic level low (blue), change the DC voltage source V s to +500 mV.
15. Decrement Vs by -20 mV. Wait 3 seconds and check the logic level of the corresponding digital channel in the display. If
the channel is at a static logic level low, record the Vs value as Vs+ in the 0 V row of the test record.
If the channel is a logic level high (green) or is alternating between high and low, repeat this step (decrement Vs by 20
mV,wait3
16. Find the a
Compare
seconds, and check for a static logic low). Continue until a value for Vs+ is found.
verage, V
= (Vs- + Vs+)/2. Record the average as the test result in the test record.
sAvg
the test result to the limits. If the result is between the limits, continue with the procedure to test the channel at
the +4 V threshold value.
17. The remaining part o f this procedure is for the +4 V threshold test. Push D15-D0 on the front panel. The Thresholds
menu should display.
18. With the Fine button on the front panel turned off, turn Multipurpose knob “a” (for channels D7 - D0) or Multipurpose
knob “b
” (for channels D15 - D8) to set the threshold value to 4.00 V (+4.0 V/div). To remove the menu from the
display, push Menu Off on the front panel.
19. Set the DC voltage source (Vs) to +4.4 V. Wait 3 seconds. Check the logic level of the corresponding digital channel in
the display.
If the channel is a static logic level low (blue), change the DC voltage source Vs to +4.5 V.
20. Decrement Vs by -20 mV. Wait 3 seconds and check the logic level of the corresponding digital channel in the display. If
annel is at a static logic level low, record the Vs value as Vs+ in the 4 V row of the test record.
the ch
e channel is a logic level high (green) or is alternating between high and low, repeat this step (decrement Vs by 20
If th
mV, wait 3 seconds, and check for a static logic low). Continue until a value for Vs+ is found.
21. Push the Trigger Menu button on the front panel.
22. Push the Slope lower-bezel button to change the slope to Rising.
23. Set the DC voltage source (Vs) to +3.6 V. Wait 3 seconds. Check the logic level of the corresponding digital channel in
the display.
If the channel is a static logic level high (green), change the DC voltage source Vs to +3.5 V.
24. Increment Vs by +20 mV. Wait 3 seconds and check the logic level of the corresponding digital channel in the display. If
e c hannel is at a static logic level high, record the Vs value as in the 4 V row of the test record.
th
62MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
If the channel is a logic level low (blue) or is alternating between high and low, repeat this step (increment Vs by 20 mV,
wait 3 seconds,
and check for a static logic high). Continue until a value for Vs- is found.
ification
25. Find the avera
ge, V
Compare the te
26. Push D15-D0 on
27. Repeat the pr
28. Disconnect t
29. Repeat the p
This comple
ocedure starting with step 8 for each remaining digital channel in the pod.
he P6316 Group 1 pod from the BNC-to-0.1 inch pin adapter and connect the Group 2 pod in its place.
rocedure starting with step 8 for each digital channel in the Group 2 pod.
tes the procedure.
= (Vs- + Vs+)/2. Record the average as the test result in the test record.
sAvg
st result to the limits. If the result is between the limits, the channel passes the test.
the front panel. The Thresholds menu should display.
MSO3000 and DPO3000 Series Specifications and Performance Verification63
Performance Ver
Check Long-term Sample Rate and Delay Time Accuracy
This test checks the sample rate and delay time accuracy (time base).
1. Connect the output of the time mark generator to the oscilloscope channel 1 input using a 50 Ω cable.
2. Set the time mark generator period to 100 ms. Use a time mark waveform with a fast rising edge.
3. Push Default Setup on the front panel to set the instrument to the factory default settings.
4. Push the channel 1 button.
5. Set the termination to 50 Ω. Push Termination on the lower menu to select 50 Ω.
ification
6. If adjustable, set the time mark amplitude to approximately 1V
.
p-p
7. Set the Vertical SCALE to 500 mV.
8. Set the Horizontal SCALE to 20 ms.
9. Adjust the Vertical POSITION knob to center the time mark signal on the screen.
10. Adjust the Trigger LEVEL knob as necessary for a triggered display.
11. Adjust the Horizontal POSITION knob to move the trigger location to the center of the screen (50%).
12. Turn the Horizontal POSITION knob counterclockwise to set the delay to exactly 100 ms.
13. Set the Horizontal Scale to 2 μs/div.
14. Compare the rising edge of the marker with the center horizontal graticule line. The rising edge should be within 0.5
ion o f the center graticule. Enter the deviation in the test record. (See page 38, Performance Checks.)
divis
NOTE. One division of displacement from graticule center corresponds to a 5 ppm time base error.
This completes the procedure.
64MSO3000 and DPO3000 Series Specifications and Performance Verification
Performance Ver
Delta Time Measurement Accuracy
This test checks the Delta time measurement accuracy (DTA) for a given instrument setting and input signal.
1. Set the sine wave generator output impedance to 50 Ω.
2. Push the oscilloscope Default Setup button on the front panel, and then push Menu Off.
3. Connect a 50 Ω coaxial cable from the signal source to the oscilloscope channel being tested.
4. Push the channel 1 button to display the channel 1 menu.
5. Push Termination on the lower menu to set the channel to 50 Ω.
ification
6. Push the Trigger Menu button on the front panel, and then, if necessary, set the trigger source to the channel being tested:
a. Push Source on the lower menu.
b. Use Multipurpose Knob “a” to select the channel being tested.
7. On the front panel, push the Measure button on the Wave Inspector, and then push Add Measurement on the lower
menu.
8. Use M u ltipurpose Kno b “b” to select the Burst Width measurement, and then push OK Add Measurement on the
nu. Use Multipurpose Knob “a” to select the input channel to be tested.
side me
9. Push M
10. Push Menu Off on the front panel to remove the Statistics menu.
11. Refer to the Test Record Delta Time Measurement Accuracy table. (See page 65, Delta Time Measurement Accuracy.)
12. Push More on the lower menu to select Statistics, and then push Reset Statistics. Wait five or 10 seconds for the
13. Ver
14. Rep
15. Pus
ore on the lower menu to select Statistics and, if necessary, use Multipurpose Knob “a” to set the Mean &
Std Dev Samples to 100, as shown in the side menu.
Set the oscilloscope and the signal source as directed there.
lloscope to acquire all the samples before taking the reading.
osci
ify that the Std D ev is less than the upper limit shown for each setting, and note the reading in the Test Record.
eat steps 11 through 13 for each setting combination shown in the Test Record for the channel being tested.
h the channel button on the front panel for the next channel to be tested, and move the coaxial cable to the
appropriate input on the oscilloscope.
16. Repeat steps 5 through 15 until all channels have been tested.
This completes the procedure.
MSO3000 and DPO3000 Series Specifications and Performance Verification65
Performance Ver
Check Aux Out
This test checks the Auxiliary Output.
1. Connect the Aux Out signal from the rear of the instrument to the channel 1 input using a 50 Ω cable.
ification
2. Push Defau
3. Push the ch
4. Set the os
5. Set the ho
6. On the fro
7. Push Add
8. Use Mult
side menu.
9. Use Multipurpose Knob “b” to select High in the Measurements menu, and then push OK Add Measurement
on the side menu.
10. Push Menu Off on the front panel.
11. Record the high and low measurements in the test record (for example, low = 200 mV and high = 3.52 V). (See page 38,
Auxil
12. Repea
ompletes the procedure.
This c
lt Setup on the front panel to set the instrument to the factory default settings.
annel 1 button.
cilloscope termination to 1 MΩ. The default Termination setting is 1M Ω.
rizontal to 4 uS/div and the vertical to 1 V/div.
nt panel, push the Measure button on the Wave Inspector.
Measurement on the lower menu.
ipurpose Knob “b ” to select Low in the Measurements menu, and then push OK Add Measurement on the
iary (Trigger) Output Tests.)
t the procedure, using 50 Ω instead of 1MΩ in step 4.
66MSO3000 and DPO3000 Series Specifications and Performance Verification
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.