Tektronix MSO2024B, MSO2022B, MSO2002B, MSO2014B, DPO2024B User Manual

...
x
MSO2000B and DPO2000B Series Oscilloscopes
ZZZ
User Manual
*P077072400*
077-0724-00
MSO2000B and DPO2000B Series Oscilloscopes
ZZZ
User Manual
www.tek.com
077-0724-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by na
tional copyright laws and international treaty provisions.
Tektronix pro previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
e*Scope, FilterVu, OpenChoice, TekSecure, and TekVPI and Wave Inspector are registered trademarks of Tektronix, Inc.
PictBridge is a registered trademark of the Standard of Camera & Imaging Products Association CIPA DC-001-2003 Digital Photo Solutions for Imaging Devices.
Contacting
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
ducts are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all
Tektronix
MSO2000B and DPO2000B Series Oscilloscopes
Warranty
Tektronix warrants that the product will be free from defects in materials and workmanship for a period of ve (5) years from the date of original purc option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are e xcluded from this warranty. Parts, m odules and replacem ent products used by Tektronix for warranty work may be ne
hase from an authorized Tektronix distributor. If the product proves defective during this warranty period, Tektronix, at its
w or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to ob period and make suitable arrangements for the performance of s ervice. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, shipping charges prepaid, and with a copy of customer proof of purchase. T the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tekt other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to serv increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS IS THE SOLE AND E XCLU S IVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPE DAMAGES.
[W19 – 03AUG12]
tain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty
ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which
ronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel
ice a product that has been modied or integrated with other products when the effect of such modication or integration
FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRO DUCTS
CTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
TPP0200 and TPP0100 Probes
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If a product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modul
ny such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective
es and products become the property of Tektronix.
In order to ob and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to C be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or integrated with other products when the effect of such modication or integration increase
THIS WARR EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SO AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE PO SSIBILITY OF SUCH DAMAGES
[W2 – 15
tain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period
ustomer if the shipment is to a location w ithin the country in which the Tektronix service center is located. Customer shall
Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or
s the time or difculty of s ervicing the product.
ANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO T HE PRODUCT IN LIEU OF ANY OTHER WARRANTIES,
LE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FO R BREACH OF THIS WARRANTY. TEKTRONIX
.
AUG04]
P6316 Probe
Warranty
Tektronix warrants that the product will be free from defects in materials and workmanship for a period of one (1) year from the date of original purc option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are e xcluded from this warranty. Parts, m odules and replacem ent products used by Tektronix for warranty work may be ne
hase from an authorized Tektronix distributor. If the product proves defective during this warranty period, Tektronix, at its
w or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to ob period and make suitable arrangements for the performance of s ervice. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, shipping charges prepaid, and with a copy of customer proof of purchase. T the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tekt other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to serv increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS IS THE SOLE AND E XCLU S IVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPE DAMAGES.
[W15 – 15AUG04]
tain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty
ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which
ronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel
ice a product that has been modied or integrated with other products when the effect of such modication or integration
FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRO DUCTS
CTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
Table of Contents
General safety summary ...................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................ .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... iii
Compliance Information............................................................................................................... v
EMC Compliance................................................................................................................. v
Safety Compliance.............................................................................................................. vii
Environmental Considerations................................................................................................... ix
Preface................................................................................................................................. x
Key Features .....................................................................................................................x
Conventions Used in This Manual................................ .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................... .. . . . . . . . . . . xi
Installation.............................................................................................................................. 1
Before Installation................................................................................................................ 1
Operating Considerations........................................................................................................ 5
Connecting Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............... 8
Securing the Oscilloscope ....................................................................................................... 9
Powering On the Oscilloscope ................................................................................................. 10
Powering Off the Oscilloscope.................................................................................................. 11
Functional Check.......... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................ .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........ 11
Compensating a Passive Voltage Probe . . . . . . . .................. .. . . . . . . . . . . . . . . . . . . . . . . . . . .................. . . . . . . . . . . . . . . . . . 12
Application M odule Free Trial. . . . . . . ............... . . . . . . . . . . . . . . . . . . . . ............. .. . . . . . . . . . . . . . . . . . . ............. .. . . . . . . . . . 13
Installing an Application Module ................ . . . . . . . . . . . . . . . . . . . . . ............... .. . . . . . . . . . . . . . . . . . . . . . ............... . . . . . . 14
Changing the User Interface Language .................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................... .. 14
Changing the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . .................................... ... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Signal Path Compensation . . . . . . . . . . . . . . . . . . . . ................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . ................... .. . . . . . . . . . . . . 17
Upgrading Firmware ............................................................................................................ 18
Connecting Your Oscilloscope to a Computer . . . . . . . . . . . . . . . . .................. .. . . . . . . . . . . . . . . . . . . . . . . . . . .................. .. 21
Connecting a USB Keyboard to Your Oscilloscope................................ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .............. 25
Getting Acquainted with the Oscilloscope .............. . . . . . . . . . . . . . . . . . . . . . . . . ................. . . . . . . . . . . . . . . . . . . . . . . . . . ........... 26
Front-Panel Menus and Controls ......... .. . . . . . . . . . . . . . . . . . . . . . . .................. .. . . . . . . . . . . . . . . . . . . . . . . .................. . . 26
Front-Panel Connectors . . . ................. . . . . . . . . . . . . . . . . . . . . . . . . . ................. . . . . . . . . . . . . . . . . . . . . . . . . . ................. 38
Side-Panel Connector. . . ........................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................... .. . . . . . . . . . 38
Rear-Panel Connectors ...... . . . . . . . . . . . . . . . . . . . . . . . . . ................. . . . . . . . . . . . . . . . . . . . . . . . . . ............... .. . . . . . . . . . . . . . . . 39
ire the Signal . . . . . . . . ................ .. . . . . . . . . . . . . . . . . . . . . . . . . ................ .. . . . . . . . . . . . . . . . . . . . . . . . . ................ .. . . . . . .40
Acqu
Setting Up Analog Channels...................................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................ .. 40
Using the Default Setup......................................................................................................... 43
Using Autoset ................................................................................................................... 44
Acquisition Concepts............................................................................................................ 45
How the Analog Acquisition Modes Work . . . ..................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................... .. . . . . . . . . . 47
Changing the Acquisition Mode, Record Length, and Delay Time............................ .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Using Roll Mode................................................................................................................. 49
Setting Up a Serial or Parallel Bus ............................................................................................. 49
Setting Up Digital Channels (MSO2000B Series Only). . . . . . . . . . . . . . . . . . ...................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Reducing Unwanted Noise With FilterVu ...................................................................................... 60
Using FilterVu ...................................................................................................................62
Table of Content
s
MSO2000B and DP O2000B Series Oscilloscopes User Manual i
Table of Content
Trigger Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................... .. . . . . . . . . . . . . . . . . . . . . . . . . . ..................... . . . . . . . . . . . . . . . . . . . . ...... 64
Display Wave
Analyze Waveform Data................................. .. . . . . . . . . . . . . . . . . . . . . . . . . . ..................... . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 89
Save and Recall Information........................................................................................................ 111
g Application Modules .................. . . . . . . . . . . . . . . . . . . . . . . . . . . . .................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............ 124
Usin
Appendix: Warranted Specications............................................................................................... 125
Index
s
Triggering Con
Choosing a Trigger Type. . . . . . . . . . . . . . . . . . . . ...................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................... .. . . . . . . 67
Selecting Triggers ............. .. . . . . . . . . . . . . . . . . . . . . . . .................. .. . . . . . . . . . . . . . . . . . . . . . . .................. . . . . . . . . . . . . . .68
Triggering on
Checking Trigger Settings ................... . . . . . . . . . . . . . . . . . . . . . ............... .. . . . . . . . . . . . . . . . . . . . . . ............... . . . . . . . . . 75
Starting and Stopping an Acquisition........ .. . . . . . . . . . . . . . . . . . . . . . . . . . .................. .. . . . . . . . . . . . . . . . . . . . . . . . . . ........... 75
Adding and Removing a Waveform . . ................... .. . . . . . . . . . . . . . . . . . . . . . . . . . ..................... . . . . . . . . . . . . . . . . . . . . . . . 76
Setting the Display Style and Persistence ..................................................................................... 76
Setting Wav
Scaling and Positioning a Waveform ........................................................................................... 80
Setting Input Parameters ....... . . . . . . . . . . . . . . . . . . . . . . . . . . . . .................. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .................. .. 81
Positioni
Positioning, Scaling, and Grouping Digital Channels... . . . . . . . . . . . . . . . . . . . . . . . . . . .................. . . . . . . . . . . . . . . . . . . . . . . . . . .. 84
Viewing Digital Channels .. .. . . . . . . . . . . . . . . . . . . . . . . . . ................ .. . . . . . . . . . . . . . . . . . . . . . . .................. .. . . . . . . . . . . . . . . . 87
Annotati
Taking Automatic Measurements............................................................................................... 89
Selectin
Customizing an Automatic Measurement...................................................................................... 93
Taking Manual Measurements with Cursors . . . . . . . . . . . . ........................... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........ 96
Using Ma
Using FFT ..................................................................................................................... 100
Using Reference Waveforms ................................................................................................. 103
Using W
Saving a Screen Image ....................................................................................................... 113
g and Recalling Waveform Data........................................................................................ 114
Savin
Saving andRecalling Setups................................................................................................. 116
Saving with One Button Push ................................................................................................ 118
g Setup, Screen Image, and Waveform Files. . . . . . . . . . . . . . . . . . . . . . . . . . ........................... .. . . . . . . . . . . . . . . . . . . . 119
Savin
Printing a Hard Copy.......................................................................................................... 120
Erasing Oscilloscope Memory................................................................................................ 122
cepts............................................................................................................. 64
Buses............................................................................................................. 70
form Data .............................................................................................................. 76
eform Intensity ..................................................................................................... 79
ng and Labeling Bus Signals . ....................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................ .. . . . . . . . . . 84
ng the Screen .......................................................................................................... 87
g Automatic Measurements............................................................................................ 90
th Waveforms......................................................................................................... 99
ave Inspector to Manage Long Record Length Waveforms........ .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............... 105
ii MSO2000B and DPO 2000B Series Oscilloscopes User Manual
General safety s
ummary
General safet
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
To avoid re or personal injury
Use proper power cord. Use only the power cord specied for this product and certied for the country of use.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected
to a voltage source.
Connect and disconnect properly. De-energize the circuit under test before connecting or disconnecting the c urrent
probe.
Ground th
shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the
product
Connect
e product.
manual for further ratings information before making connections to the product.
the probe reference lead to earth ground only.
y summary
This product is grounded through the grounding conductor of the power cord. To avoid electric
Do not ap
Power d
must remain accessible to the user at all times.
ply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
isconnect.
The power cord disconnects the product from the power source. Do not block the power cord; it
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by
qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive a tmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation.
per ventilation.
pro
Refer to the manual's installation instructions for details on installing the product s o it has
MSO2000B and DPO2000B Series Oscilloscopes User Manual iii
General safety s
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify c onditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ummary
iv MSO2000B and DPO 2000B Series Oscilloscopes User Manual
Compliance Info
rmation
Compliance In
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument complies.
EMC Compliance
EC Declaration of Conformity – EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following specications as listed in the Ofcial Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment for measurement, control, and
laboratory use.
CISPR 11:2003. Radiated and conducted emissions, G roup 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic eld immunity
IEC 61000-4-4:2004. Electrical fast transient/bur
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
123
formation
4
st immunity
5
6
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European Contact.
Tektronix UK, Ltd. Western Peninsula Western Road Bracknell, RG12 1RF United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interfe
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
Trace bloom not exceeding 4 divisions pk-to-pk may be induced under the conditions of the IEC 61000-4-3 test.
5
Trace bloom not exceeding 1 division pk-to-pk may be induced under the conditions of the IEC 61000-4-6 test.
6
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels (IEC 61000-4-11).
rence.
Australia / New Zealand Declaration of Conformity – EMC
Complies with the EMC provision of the Radiocommunications Act per the following standard, in accordance with ACMA:
MSO2000B and DP O2000B Series Oscilloscopes User Manual v
Compliance Info
CISPR 11:2003. Radiated and Conducted Emissions, Group 1, Class A, in accordance with EN 61326-1:2006 and EN 61326-2-1:2
rmation
006.
vi MSO2000B and DPO 2000B Series Oscilloscopes User Manual
Safety Compliance
EC Declaration of Conformity – Low Voltage
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC.
EN 61010-1: 2001. Safety requirements for electrical equipment for m easurem ent control and laboratory use.
Compliance Info
rmation
U.S. Nation
UL 61010-1:2004, 2ndEdition. Standard for electrical measuring and test equipment.
ally Recognized Testing Laboratory Listing
Canadian Certication
CAN/CSA­laboratory use. Part 1.
C22.2 No. 61010-1:2004. Safety requirements for electrical equipment for measurement, control, and
Additional Compliances
IEC 61010-1: 2001. Safety requirements for electrical equipment for measurement, control, and laboratory use.
Equipment Type
Test and measuring equipment.
y Class
Safet
Class 1 – grounded product.
Pollution Degree Description
asure of the contaminants that could occur in the environ ment around and within a p roduct. Typically the internal
Ame environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally
capsulated, hermetically sealed, or located in clean rooms.
en
llution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is
Po caused by condensation m ust be expected. This location is a typical ofce/home environment. Temporary condensation occurs only when the product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation.
hese are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct
T sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
MSO2000B and DP O2000B Series Oscilloscopes User Manual vii
Compliance Info
Pollution Degree
Pollution Degree 2 (as dened in IEC 61010-1). Note: Rated for indoor use only.
Installation (Overvoltage) Category Descriptions
Terminals on this product may have different installation (overvoltage) category designations. The installation categories are:
Measurement Category IV. For measurements performed at the source of low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Category I. For measurements perform ed on circuits not directly connected to M AINS.
rmation
Overvolta
Overvoltage Category II (as dened in IEC 61010-1).
ge Category
viii MSO2000B and DPO 2000B Series Oscilloscopes User Manual
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life Handling
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. Production of this equipment required the extraction and use of natural resources. The
equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. In order to avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies w ith the applicable European Union requirements according to Directives 2002/96/EC and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
Compliance Info
rmation
Mercury N
to environmental considerations. Please contact your local authorities or, within the United States, refer to the E-cycling Central Web page (www.eiae.org) for disposal or recycling information.
Restric
This product has been classied as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS Directive.
otication.
tion of Hazardous Substances
This product uses an LCD backlight lamp that contains mercury. Disposal may be regulated due
MSO2000B and DP O2000B Series Oscilloscopes User Manual ix
Preface
Preface
This manual describes the installation and operation of the following oscilloscopes:
MSO2024B MSO2022B MSO2014B MSO2012B MSO2004B MSO2002B DPO2024B DPO2022B DPO2014B DPO2012B DPO2004B DPO2002B
Key Features
MSO2000B and DPO2000B series oscilloscopes can help you verify, debug, and characterize electronic designs. Key features include:
200 MHz, 100 MHz, and 70 MHz bandwidths
2 channel and 4 channel models
Sample rates up to 1 GS/s on all analog channels
1 M points record length on all channe
5,000 waveforms/second waveform capture rate
I2C, SPI, CAN, LIN, R S-232, RS-422, RS-485, and UART bus triggering and analysis (with the appropriate ap module and model oscilloscope)
Wave Inspector controls for managing long record lengths, with zoom and pan, play and pause, search and mark
Large 178 mm (7 inch) WQVGA wide screen color display
Small and lightweight, at 140 mm (5.5 inch) deep and 3.6 kg (7 lbs, 14 oz)
FilterVu provides a variable low pass lter to block unwanted noise while still displaying high frequency events
USB ash drive port for quick and easy storage of measurement results
Direct printing to any PictBridge compatible printer
Ethernet port for remote programmability with the optional connectivity module
Video Out port to display the oscilloscope screen on an external monitor with the optional connectivity module
USB 2.0 Device port for direct PC control of the oscilloscope using USBTMC protocol
OpenChoice documentation software for simple transfer of screen shots and waveform data to a PC
Remote viewing and control with e*Scope
Remote control with VISA connectivity
ls
plication
TekVPI Versatile P robe Interface supports active, differential, and current probes for automatic scaling and units
MSO2000B series of mixed signal oscilloscopes also offer:
16 digital channels
Parallel bus triggering and analysis
Easy connection to your device-under-test through the convenient design of the P6316 digital probe
x MSO2000B and DPO2000B Series Oscilloscopes User Manual
Conventions Used in This Manual
The following icons are used throughout this manual.
Preface
Sequence Step
Front panel power
Connect power
Network
USB
MSO2000B and DP O2000B Series Oscilloscopes User Manual xi
Preface
xii MSO2000B and DPO2000B Series Oscilloscopes User Manual
Installation
Before Installation
Unpack the oscilloscope and check that you received all items listed as standard accessories. The following pages list recommended accessories and probes, instrument options, and upgrades. Check the Tektronix Web site (www.tektronix.com) for the most current information.
Standard Accessories
Accessory Description
MSO2000B and DPO2000B Series Oscilloscopes Installation and Safety Manual
MSO2000B and DPO2000B Series Oscilloscopes Documentation Browser CD
Tektronix OpenChoice Desktop PC Communications CDs
Calibration certicate documenting traceability to national metrology institute(s), and ISO9001 quality system registration.
Front panel overlay
For MSO2000B and DP O2000B series: Probes
For MSO2000B series: Digital probe One, 16-channel digital probe For M SO2000B series: Accessories pouch Pouch that attaches to the handle for carrying
Five year warranty
Safety and compliance information along with hardware installation instructions.
Electronic versions of documents, including the User Manual, Programmer Manual and the Technical Reference.
Connectivity and documentation software.
French (Option L1) Italian (Option L2) German (Option L3) Spanish (Option L4) Japanese (Option L5) Portuguese (Option L6) Simple Chinese (Option L7) Traditional Chinese (Option L8) Korean (Option L9) Russian (Option L10) For models 100 MHz bandwidth: One
200 MHz, 10X passive voltage probe with 10 Minput resistance per channel
For models < 100 MHz bandwidth: One 100 MHz, 10X passive voltage probe with 10 M input resistance per channel
probes and other accessories. For details, refer to the warranty in the front of
this manual
Installation
Tektronix part number
071-3078-XX
063-4472XX
063-4402XX
——
335-2020-00 335-2021-00 335-2022-00 335-2023-00 335-2024-00 335-2025-00 335-2026-00 335-2027-00 335-2028-00 335-2029-00 TPP0200
TPP0100
P6316 016-2008-00
——
MSO2000B and DP O2000B Series Oscilloscopes User Manual 1
Installation
Standard Accessories (cont.)
Accessory Description
Power cord
North America (Opti o n A0) Universal Euro (Option A1) United Kingdom (Option A2) Australia (Option A3) Switzerland (Option A5) Japan (Option A6) China (Option A10) India (Option A11) Brazil (Option A12) No power cord or AC adapter (O ption A99)
Tektronix part number
161-0348-00 161-0343-00 161-0344-00 161-0346-00 161-0347-00 161-0342-00 161-0341-00 161-0349-00 161-0356-00 ——
2 MSO2000B and DPO2000B Series O scilloscopes User Manual
Installation
Optional Accessories
Accessory Description Tektronix part number
DPO2EMBD
The embedded serial triggering and analysis module enables triggering on packet level information on
2
I
C and SPI ser
ial buses, as well as bus views, bus decoding, search tools, and packet decode tables with timestamp information
DPO2AUTO
The automotive serial triggering and analysis module enables triggering on packet level information on CAN and LIN se
rial buses, as well as bus views, bus decoding, search tools, and packet decode tables with timestamp information
DPO2COMP
The computer triggering and analysis module enables triggering on RS-232, RS-422, RS-485 and UART seria
l buses, search tools, bus views, bus decoding in hex, binary, and ASCII, and decode tables with timestamp information
DPO2CONN
The connectivity module adds an Ethernet port for remote programmability and a Video Out port to displa
y the oscilloscope screen on an external
monitor
NEX-HD2H
EADER
Adapter that routes the channels from a Mictor connector to 0.1 inch header pins
TekVPI probes that work with MSO/DPO2000B Series
oscopes
oscill
Visit the Oscilloscope Probe and Accessory Selector
the Tektronix website at www.tektronix.com.
Tool on
Note: These probes require the use of the TekVPI external power
r listed below.
adapte TekVPI external power adapter
NC
TPA-B
ies external power to a TekVPI probe
Suppl
I to TekProbe II BNC Adapter
TekVP
Note: This adapter requires the use of the TekVPI external power
ter listed above.
adap Deskew pulse generator Deskew pulse generator and signal source with
PI oscilloscope interface
TekV
Power measurement deskew and
ibration xture
cal
verts TEK-DPG pulse generator output into a
Con series of test point connections
TEK-USB-488 Adapter GPIB to USB Adapter TEK-USB-488
ckmount kit
Ra
oft transit case
S Hard transit case
ds rackmount brackets
Ad
ase for carrying an oscilloscope
C
raveling hard case, which requires use of the soft
T transit case (ACD2000)
DPO2EMBD
DPO2AUTO
DPO2COMP
DPO2CONN
NEX-HD2H
1197465XX
NC
TPA-B
DPG
TEK-
067-1686-00
D2000
RM
CD2000
A
CTEK4321
H
EADER
MSO2000B and DP O2000B Series Oscilloscopes User Manual 3
Installation
Optional Accessories (cont.)
Accessory Description Tektronix part number
MSO2000B and DPO2000B Series Oscilloscopes Service manual
MSO2000B/DPO2000B and MSO2000/DPO2000 Series Oscilloscopes Application Module Installation
The MSO2000B and DPO2000B series oscilloscopes work with multiple optional probes. (See page 8, Connecting Probes.) Check the Tektronix Web site (www.tektronix.com) for the most current information.
Service information on MSO2000B and DPO2000B series oscilloscopes
Describes how to install application modules in MSO/DPO2000B and MSO/DPO2000 series oscilloscopes
077-0737XX
071-2330XX
Related Documentation
Accessory Description Tektronix part number
MSO2000B and DPO2000B Series Oscilloscopes User Manual
MSO2000B and DPO2000B Series Oscilloscopes Programmer Manual
MSO2000B and DPO2000B Series Oscilloscopes Technical Reference Manual
English 077-0724-XX French 077-0725-XX Italian 077-0726-XX German Spanish Japanese 077-0729-XX Portuguese 077-0730-XX Simple Chinese Traditional Chinese Korean 077-0733-XX Russian 077-0735-XX Describes commands for remote control
of the oscilloscope; available electronically on the D ocumentation Browser CD or for download from www.tektronix.com/manuals
Describes the oscilloscope specications and performance verication procedure; available electronically on the Documentation Browser CD or for download from www.tektronix.com/manuals
077-0727-XX 077-0728-XX
077-0731-XX 077-0732-XX
077-0738XX
077-0735XX
4 MSO2000B and DPO2000B Series O scilloscopes User Manual
Operating Considerations
MSO2000B and DPO2000B Series Oscilloscopes
Power Supply Input Voltage: 100 V to 240 V ± 10%
Power Supply Input Power Frequency: 50/60 Hz at 100 V to 240 V 400 Hz at 115 V
Power Consumption: 80 W maximum
Weight: 3.6 kg (7 lbs 14 oz), stand-alone oscilloscope
Height, including the feet but not the handle: 175 mm (6.885 inch)
Installation
Width: 377 mm (14.85 inch)
Depth: from the feet to the front of the knobs: 134 mm (5.3 inch) Depth: fr
Clearan
Input V 300 V Installation Category II - for measurements performed
on circ installation
Tempe Operating: 0 °C to +50 °C Nonoperating: -20 °C to +60 °C
Humidity: Operating: High: 40 °C to 50 °C, 10% to 60% RH Operating: Low: 0 °C to 40 °C, 10% to 90% RH Non-operating: High: 40 °C to 60 °C, 5% to 60% RH Non-operating: Low: 0 °C to 40 °C, 5% to 90% RH
om the feet to the front of the front cover: 139 mm (5.47 inch)
ce: 50 mm (2 inch)
oltage (between the signal and reference):
CAT II
RMS
uits directly connected to the low-voltage
rature:
DPO2000B series
MSO2000B series
Altitude: Operating: 3,000 m (9,842 ft) Nonoperating Altitude: 12,000 m (39,370 ft)
MSO2000B and DP O2000B Series Oscilloscopes User Manual 5
Installation
Random Vibrati Operating: 0.31 G Non-operating: 2.46 G
on:
, 5 - 500 Hz, 10 minutes per axis, 3 axes (30 minutes total)
RMS
, 5 - 500 Hz, 10 minutes per axis, 3 axes (30 minutes total)
RMS
Pollution Degree: 2, Indoor use only
CAUTION. To ensure proper cooling, keep the sides and rear of the oscilloscope clear of obstructions.
TPP0200/TPP0100 Pa
Single-ended volt 300 V
RMS
age probe (ground referenced):
CAT II safety requirements
ssive Probe
Installation Category II - for measurem ents performed on circuits directly connected to the low-voltage installation
Temperature: Operating: -10 °C to +55 °C (+14 °F to +131 °F) Nonoperating: -5
1 °C to +71 °C ( -60 °F to +160 °F)
Pollution Degree: 2, Indoor use only
Humidity: 5% to 95% RH
MSO2000B Series Oscilloscope with a P6316 Digital Probe
Threshold Accuracy: ±(100 mV + 3% of threshold)
Threshold Range: ±20 V
Maximum nond
Minimum signal swing: 500 mV
Input resistance: 101 k
Input capacitance: 8.0 pF
Temperat Operating: 0 °C to +50 °C (+32 °F to +122 °F) Nonoperating: -40 °C to +71 °C (-40 °F to +160 °F)
Altitude: Operating: 3,000 m (9,843 ft) maximum Nonoper
Pollution Degree: 2, Indoor use only
estructive input signal to probe: ±40 V
peak-to-peak
ure:
ating: 12,000 m (39,370 ft) maximum
6 MSO2000B and DPO2000B Series O scilloscopes User Manual
Installation
Humidity: 5% to 95% relative humidity
Cleaning
Inspect the oscilloscope and probes as often as operating conditions require. To clean the exterior surface, perform the following steps:
1. Remove loose dust on the outside of the oscilloscope and probes with a lint-free cloth. Use care to avoid scratching the clear glass display lter.
2. Use a soft cloth dampened with water to clean the oscilloscope. Use an aqueous solution of 75% isopropyl alcohol for more efcient cleaning.
CAUTION. To avoid damage to the surface of the oscilloscope or probes, do not use any abrasive or chemical cleaning
agents.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 7
Installation
Connecting Probes
The oscilloscope supports probes with the following:
1. Tektronix Versatile Probe Interface (TekVPI)
These probes support two-way communication with the oscilloscope through on-screen menus and remotely through programmable support. The remote control is useful in applications like an ATE (automated test environment) where you want the system to preset probe parameters.
NOTE. For more information on the many
probes available for use with MSO2000B and DPO2000B Series oscilloscopes, visit the Oscilloscope Probe and Accessory Selector tool on the Tektronix website.
2. TPA-BNC Adapter The TPA-BNC Adapter allows you to
use Tek Probe II probe capabilities, such as providing probe power, and passing scaling and unit information to the oscilloscope.
NOTE. To use a TekVPI probe and a
TPA-BNC adapter, connect a TekVPI external power adapter (Tektronix part number 119 Probe Power connector.
3. Plain BNC Interfaces
4. Digital Probe Interface
For more information on the many probes available for use with MSO2000B and DPO2000B series oscilloscopes, refer to www.tektronix.com.
7465‑XX) to the side panel
Some probes use TekProbe capabilities to pass the waveform signal and scaling to the oscilloscope. Other probes only pass the signal and there is no communication.
(MSO2000B series only) The P6316 probe provides 16 channels
of digital (on or off state) information.
8 MSO2000B and DPO2000B Series O scilloscopes User Manual
Securing the Oscilloscope
1. Use a standard laptop computer style
security lock to your location.
This photo also shows the optional DPO2CONN mod module provides an Ethernet port and a Video Out port for the o scilloscopes.
to secure your oscilloscope
ule installed. The
Installation
MSO2000B and DP O2000B Series Oscilloscopes User Manual 9
Installation
Powering On the Oscilloscope
Ground the Oscilloscope and Yourself
Before pushing the power switch, connect the oscilloscope to an electrically neutral reference point, such as earth ground. Do this by plu
gging the three-pronged power cord into an outlet grounded to earth ground.
Grounding th same ground as any circuits that you are testing.
To connect the power cord and power on the oscilloscope:
e oscilloscope is necessary for safety and to take accurate measurements. The oscilloscope needs to share the
Quick Tips
Ifyouareworkingwithstaticsensitive components, ground yourself. Static electricity that builds up on your body can damage static-sensitive components. Wearing a grounding strap safely sends static charges on your body to earth ground.
10 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Powering Off the Oscilloscope
To power off the oscilloscope and remove the power cord:
Installation
Functional Check
Perform this quick functional check to verify that your oscilloscope is operating correctly.
1. Connect the oscilloscope power cable
as described in Powering On the Oscilloscope. (See page 10.)
2. Power on the oscilloscope.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 11
Installation
3. Connect the TPP0200/TPP0100 probe tip and reference lead to the PROBE
COMP connecto
4. Push Default Setup .
5. Push Autoset. The screen should now
display a square wave, approximately 5Vat1kHz.
rs on the o scilloscope.
NOTE. For best performance, it is
recommended that you set the Vertical scale to1V.
If the signal appears but is misshapen, perform the procedures for compensating the probe. (See page 12.)
If no signal appears, rerun the procedure. If this does not remedy the situation, have the oscilloscope serviced by qualied service personnel.
Compensating a Passive Voltage Probe
ever you attach a passive voltage probe for the rst time to any input channel, compensate the probe to match it to
When the corresponding oscilloscope input channel.
To properly compensate your passive probe:
1. Follow the steps for the functional check. (See page 11.)
2. Check the shape of the displayed waveform to determine if your probe is properly compensated.
Properly compensated
Under compensated Over compensated
12 MSO2000B and DPO2000B Series Oscilloscopes User Manual
3. If necessary, adjust your probe.
Repeat as needed.
Quick Tips
Use the shortest possible ground lead and signal path to minimize probe-induced ringing and distortion on the measured signal.
Installation
Application Module Free Trial
A 30-day power on the oscilloscope for the rst time.
After 30 days, you must purchase the module if you want to continue using the application. To see the date when your free trial period expires, push the front panel Utility button, push the lower-bezel Utility Page button, use multipurpose knob a to select
free trial is available for all application modules not installed in your oscilloscope. The trial period begins when you
Cong, and push the lower-bezel About button.
Signal wi
th a short ground lead
Signal with a long ground lead
MSO2000B and DP O2000B Series Oscilloscopes User Manual 13
Installation
Installing an Application Module
CAUTION. To avoid damage to the oscilloscope or application module, observe ESD (electrostatic discharge) precautions.
(See page 10, Pow
ering On the Oscilloscope.)
Turn off the os
(See page 11, Po
cilloscope power while removing or adding an application module.
wering Off the O scilloscope.)
Optional application module packages extend the capability of your oscilloscope. You can install one or two application modules at one time. An application module goes into the slot with a window in the upper right corner of the front panel. Another slo
For more in
t is directly behind the one that you can see. To use this slot, install the module with the label facing away from you.
formation on how to install and test application modules, refer to the MSO/DPO2000B and MSO/DPO2000
Series Oscilloscopes Application Module Installation manual.
NOTE. If you remove an application module, the features provided by the application module become unavailable. To
restore th
e features, turn off the oscilloscope power, r einstall the module and turn on the oscilloscope power.
Changing the User Interface Language
To change the language of the o scilloscope user interface, and to change the front-panel button labels through the use of an overlay:
1. Push Utility.
2. Push Utility Page.
3. Turn multipurpose knob a and select Cong.
Utility Page
Utility Page
Config
Language
English
Set Date &
Time
TekS ecu r e
Erase
Memory
About
4. Push Language from the resulting lower-bezel menu.
14 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Installation
5. Turn multipurp
desired language. Choose among: English, French, German, Italian, Spanish, Brazilian Portuguese, R Simplied Chinese, and Traditional Chinese.
6. If you choose to use English, be sure that
the plastic front-panel overlay is removed. If you choose a language other than English,
place the plastic overlay for the language that you desire over the front panel to display labels in that language.
ose knob a and select the
ussian, Japanese, Korean,
Changing the Date and Time
he internal clock w ith the current date and time:
To set t
1. Push Ut
2. Push Utility Page.
ility.
Utility
Page
MSO2000B and DP O2000B Series Oscilloscopes User Manual 15
Installation
3. Turn multipurpose k nob a and select Cong.
4. Push Set Date &
5. Push the side
Time.
-bezel buttons and use
multipurpose knobs a and b to set the Day,
Month, Year, Hour, and Minute values.
6. Push Display and turn multipurpose knob a to choose Date & Time, Date Only, Time Only,orNo
ne.
7. Push OK Enter Date & Time.
System
Config
Set Date &
Time
Display
Time Only
Select
Day
Day
3
OK Enter
Date &
Time
Language
English
Set Date &
Time
TekSecure
Erase
Memory
About
16 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Signal Path Compensation
Signal Path Compensation (SPC) corrects for DC inaccuracies caused by temperature variations and/or long-term drift. You should run the SPC whenever the ambient temperature has changed by more than 10 °C or once a week if you use vertical settings of 5 mV per division or less. Failure to do so may result in the oscilloscope not meeting warranted performance levels at those volts per division settings.
To compensate the signal path:
1. Warm up the oscilloscope for at least
20 minutes. Remove all input signals (probes and cables) from channel inputs. Input signals with AC components adversely affect SPC.
Installation
2. Push Utility.
3. Push Utility Page.
4. Turn multipurpose knob a and select
Calibration.
5. Push Signal Path from the lower-bezel
menu.
6. Push OK Com pensate Signal Paths from
the resulting side-bezel menu.
Utility Page
Utility Page
Calibration
OK Com-
pensate
Signal
Paths
Signal
Path
Pass
Factory
Pass
MSO2000B and DP O2000B Series Oscilloscopes User Manual 17
Installation
The oscillosco
pe displays a message when the calibration is complete. Push Menu Off to remove the message.
7. After calibr
ation, v e rify that the status indicator on the lower-bezel menu displays Pass.
Utility Page
Calibration
Signal
Path
Pass
Factory
Pass
If it does not, then recalibrate the oscilloscope or have the oscilloscope serviced b
y qualied service personnel.
Service personnel use the factory calibration
s to calibrate the internal voltage
function references of the oscilloscope using external sources. Refer to your Tektronix eld ofc
e or representative for assistance
with factory calibration.
NOTE. Signal Path C ompensation does not include calibration to the probe tip. (See page 12, Compensating a Passive
Voltage Probe.)
Upgrading Firmware
To upgrade the rmware of the oscilloscope:
1. Open up a Web browser and go to www.tektronix.com/software. Proceed to the software nder. Download the latest rmware for your oscilloscope on your PC.
ip the les and copy the rmware.img
Unz le into the r oot folder of a USB ash drive.
18 MSO2000B and DPO2000B Series Oscilloscopes User Manual
2. Power off your oscilloscope.
3. Insert the USB ash drive into the front-panel
USB port o
n your oscilloscope.
Installation
4. Power on the oscilloscope. The oscilloscope
automatically recognizes and installs the replacement rmware.
If the oscilloscope does not install the rmware, rerun the procedure. If the problem continues, try a different model of USB ash drive. Finally, if needed, contact qualied service personnel.
NOTE. Do not power off the oscilloscope or
remove the USB ash drive until the oscilloscopenishes installing the rmware.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 19
Installation
5. Power off the oscilloscope and remove the USB ash drive.
6. Power on the oscilloscope.
7. Push Utility.
ity
8. Push
Utility Page.
Util Page
20 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Installation
9. Turn multipurpose knob a and select Cong.
Utility
Page
Config
Language
English
10. Push About. The oscilloscope displays the
rmware versi
11. Conrm that t
on number.
he version number matches
that of the new rmware.
Connecting Your Oscilloscope to a Computer
You may want to document your work for future reference. Instead of saving screen images and waveform data to a USB ash drive and generating a report later, you may want to get a copy of the image or waveform data directly from a remote PC for analysis. You may also want to control an oscilloscope at a remote location from your computer. (See page 113, Saving a Screen Image.) (See page 114, Saving and Recalling Waveform Data.)
Two ways to connect your oscilloscope to a computer are through the VISA (Virtual Instrument Software Architecture) drivers and the e*Scope Web-enabled tools. Use VISA to communicate with your oscilloscope from your computer through a software application. Use e*Scope to communicate with your oscilloscope through a Web browser.
Using VISA
s you use your MS-Windows computer to acquire data from your oscilloscope for use in an analysis package that
VISA let runs on your PC, such as Microsoft Excel, National Instruments LabVIEW, or a program of your own creation. You can use a common communications connection, such as USB or Ethernet, to connect the computer to the oscilloscope.
Set Date &
Time
TekSecure
Erase
Memory
About
To set up VISA communications between your oscilloscope and a computer:
1. Load th
e VISA drivers on your computer.
You w ill nd the drivers on the appropriate CD that comes with your oscilloscope or
Tektronix software nder Web page
at the (www.tektronix.com).
MSO2000B and DP O2000B Series Oscilloscopes User Manual 21
Installation
2. Connect the oscilloscope to your computer with the appropriate USB or Ethernet cable.
NOTE. You need
a DPO2CONN module to
make the Ethernet connection.
To communicate between the oscilloscope and a GPIB system, connect the oscilloscope to the TEK-USB-488 GPIB-to-USB Adapter with a USB cable. Then connect the adapter to your GPIB system with a GP IB cable. Cycle the power on the oscilloscope.
3. Push Utility.
4. Push Utility Page.
5. Turn multipurpose knob a and select I/O.
6. If a USB cable is connected between
the oscilloscope and your computer, the oscilloscope automatically sets itself up for you.
Check the USB on the lower-bezel menu to be sure it is enabled. If it is not enabled, push USB and make an appropriate selection in the side-bezel menu.
Utility Page
Utility Page
I/O
USB
Computer
Ethernet Network Settings
GPIB
1
22 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Installation
7. To use Ethernet
, push Ethernet Network
Settings.
8. On the side-be
zel menu, if you are on a DHCP Ethernet network and using a through cable, set DHCP to On. If you are using a cross-over c
able, set it to Off and set a hard
coded TCPIP address.
9. If you are using GPIB, push GPIB. Enter the GPIB ad
dress on the side-bezel menu,
using multipurpose knob a.
This will
set the GPIB address on an
attached TEK-USB-488 Adapter.
10. Run your
application software on your
computer.
Quick Tips
Change
Instrument
Settings
DHCP/
BOOTP
On|Off
Talk/List
Address
(a) 1
en
The CDs that are shipped with your oscilloscope include a variety of Windows-based software tools designed to ensure
nt connectivity between your oscilloscope and your computer. There are toolbars that enhance connectivity with
efcie Microsoft E xcel and Word. There is also a stand-alone acquisition program called the OpenChoice Desktop.
USB Host port
e front-panel USB 2.0 Host port for USB ash drives and keyboards.
Use th
USB Device port
he rear-panel USB 2.0 Device port for PCs or PictBridge printers.
Use t
ng e*Scope
Usi
e*Scope lets you access any Internet-connected MSO2000B or DPO2000B series oscilloscope from a browser on your workstation, PC, or laptop computer. No matter where you are, your oscilloscope is as close as the nearest browser.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 23
Installation
To set up e*Scope communications between your oscilloscope and a Web browser running on a remote computer:
1. Connect the oscilloscope to your computer network with the appropriate Ethernet cable.
NOTE. You need a DPO2CONN module to
make the Ethernet connection.
If you are connecting directly to your computer, you need a Crossover Ethernet Cable. If you are connecting to a network or a hub, you need a Straight Through Ethernet Cable.
2. Push Utility.
3. Push Utility Page.
4. Turn multipurpose knob a and select I/O.
5. Push Ethernet Network Settings.
6. On the side-bezel menu, if you are on a
DHCP Ethernet network and using dynamic addressing, set DHCP to On. If you are using static addressing, set it to Off.
7. Push Change Instrument Settings.Ifyou are using DHCP, note the Ethernet address and instrument name. If you are using Static addressing, enter the Ethernet address you will be using.
Utility Page
Utility Page
I/O
DHCP/
BOOTP
On|Off
Change
Instrument
Settings
USB
Enabled
Ethernet Network Settings
GPIB
1
NOTE. Depending on the type and speed of network to which your oscilloscope is connected, you may not see the
DHCP/BOOTP eld update instantaneously after pressing the DHCP/BOOTP button. It may take a few seconds to update.
24 MSO2000B and DPO2000B Series Oscilloscopes User Manual
8. Start your browser on y our remote computer. In the browser address line, enter the IP address or, if DHCP is set to On
in the oscilloscope, simply enter the instrument name. You should now
work, rerun the procedure. If it still does not work, contact qualied service personnel.
see the e*Scope screen showing the oscilloscope display on your Web browser. If e*Scope does not
Connecting a USB Keyboard to Your Oscilloscope
You can connect a USB keyboard to the USB Host port on the front panel of the oscilloscope. The oscilloscope will detect the keyboard, even if it is plugged in while the oscilloscope is powered on. (See page 41, Labeling Channels and Buses.)
Installation
MSO2000B and DP O2000B Series Oscilloscopes User Manual 25
Getting Acquain
ted with the Oscilloscope
Getting Acqua
inted with the Oscilloscope
Front-Panel Menus and Controls
The front panel has buttons and controls for the functions that you use most often. Use the menu buttons to access more specialized functions.
Using the Menu System
To use the menu system:
1. Push a front-panel menu button to display the menu that you want to use.
2. Push a lower-bezel button to select a menu item. If a pop-out menu appears, turn multipurpose knob a to select the desired choice. If a pop-up menu appears, press the button again to select the desired choice.
26 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Getting Acquain
ted with the Oscilloscope
3. Push a side-bez
el button to choose a
side-bezel menu item. If the menu item contains more than
one choice, pu
sh the side-bezel button
repeatedly to cycle through the choices. If a pop-out menu appears, turn
multipurpos
e knob a to select the desired
choice.
4. Toremoveas
ide-bezel menu, push the lower-bezel button again or push Menu Off.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 27
Getting Acquain
5. Certain menu choices require you to set a numeric value to complete the setup. Use the upper a knobs a and b to adjust values.
6. Push Fine to turn off or on the ability to make smaller adjustments.
Using the Menu Buttons
ted with the Oscilloscope
nd lower multipurpose
Use the me
1. Measure. Push to perform automated
2. Search.
3. Test. Push to activate advanced or
4. Acquire. Push to set the acquisition
5. Autoset. Push to perform an automatic
6. Tri
7. Uti
nu buttons to perform many functions in the oscilloscope.
measurements on waveforms or to
ecursors.
congur
Push to search through an acquisition for user-dened events/criteria.
application-specic testing features.
mode and adjust the record length.
p of oscilloscope settings.
setu
gger Menu. Push to specify trigger
settings.
lity. Push to activate the system utility functions, such as selecting a language or setting the date/time.
8. Save / Recall Menu. Push to save and call setups, waveforms, and screen
re images to internal memory, or a USB ash drive.
28 MSO2000B and DPO2000B Series Oscilloscopes User Manual
9. Channel 1, 2, 3,or4Menu. Push to set vertical parameters for input waveforms and
to display or remove the corresponding waveform from the display.
Getting Acquain
ted with the Oscilloscope
MSO2000B and DP O2000B Series Oscilloscopes User Manual 29
Getting Acquain
10. B1 or B2. Pushtodefine and display a serial bus if you have the appropriate m The DPO2AUTO module supports CAN and LIN buses. The DPO2EMBD module supports I module supports RS-232, RS-422, RS-485, and UART buses.
Parallel bus MSO2000B products.
Also, push the B1 or B2 button to display or remove th the display.
11. R. Push to manage reference waveforms, including the display or removal of each referenc
ted with the Oscilloscope
odule application keys.
2
and SPI. The DPO2COMP
C
support is available on
e corresponding bus from
e waveform from the display.
12. M . Push to including the display or removal of the math waveform from the display.
Using O
These buttons and knobs control waveforms, cursors, and other data input.
1. Turn the upper multipurpose knob a, when activated, to move a cursor, to set a numerical parameter value for a menu item, or to select from a pop-out list of choices. Push the Fine button to toggle between coarse and ne adjustment.
Screen icons tell you when a or b are active.
2. Cursors. Push once to activate the two vertical cursors. Push again to turn on the two vertical and two horizontal cursors. Push again to turn off all cursors.
When the cursors are on, you can turn the multipurpose knobs to control their position.
manage the math waveform,
ther Controls
30 MSO2000B and DPO2000B Series Oscilloscopes User Manual
3. Select. Push to activate special
functions. For example, w
cursors (and no horizontal ones are visible), you can push this button to link or unlink the vertical and two horizontal cursors are both visible, you can push this button to make either t horizontal cursors active.
4. FilterVu. Push to lter unwanted noise
from your signal and still capture glitches.
5. Fine. Push to toggle between making
coarse and ne adjustments with the vertical trigger level knob, and many operations of multipurpose knobs a and b.
6. Inten sity. Push to enable multipurpose
knob a to control waveform display
ty and knob b to control graticule
intensi intensity.
hen using the two vertical
cursors. When the two
he vertical cursors or the
and horizontal position knobs, the
Getting Acquain
ted with the Oscilloscope
7. Turn th
8. Zoom b
9. Pan (outer knob). Turn to scroll the zoom
10. Zoom (inner knob). Turn to control the
11. Play-pause button. Push to start or stop
e lower multipurpose knob b, when activated, to move a cursor or set a numerical parameter value for a menu
ush Fine to make adjustments
item. P more slowly.
utton. Push to activate zoom
mode.
window through the acquired waveform.
m factor. Turning it clockwise zooms
zoo in further. Turning it counterclockwise zooms out.
the automatic panning of a waveform.
ntrol the speed and direction with the
Co pan knob.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 31
Getting Acquain
ted with the Oscilloscope
12. Prev.Pushtoj waveform mark.
13. Set/Clear Mark. Push to establish or delete a waveform mark.
14. Next. Push to jump to the next waveform mark.
15. Horizontal Position. Turn to adjust the trigger point location relative to the acquired wa smaller adjustments.
16. Horizontal Scale. Turn to adjust the horizontal scale (time/division).
17. Run/Stop . Push to start or stop acquisitions.
18. Single. Push to make a single acquisition.
ump to the previous
veforms. Push Fine to make
19. Autoset. Push to automatically set the vertical, horizontal, and trigger controls for a usable, stable display.
20. Trigger Level. Turn to adjust the trigger level.
Push Level to Set 50%. Push the Trigger level knob to set the trigger level to the midpoint of the waveform.
21. Force Trig. Push to force an immediate trigger event.
22. Vertical Position. Turn to adjust the vertical position of the corresponding
eform. Push Fine to make smaller
wav adjustments.
2, 3, 4 Menu. Push to display or
23.1,
remove the corresponding waveform from the display and access the vertical
nu.
me
32 MSO2000B and DPO2000B Series Oscilloscopes User Manual
24. Vertical Scale. Turn to adjust the
vertical scale factor of the corresponding waveform (vol
25. Print. Push to print to a PictBridge
printer.
ts/division).
Getting Acquain
ted with the Oscilloscope
26. Power switc
the oscilloscope.
27. USB 2.0 Host port. Insert a USB
peripheral to the oscilloscope, such as a keyboard o
28. Save. Pus
save operation. The save operation uses the current save parameters, as dened in the Sav
29. Default Setup. Push to perform an
immediate restore of the oscilloscope to the default settings.
30. D15 - D0. Push to display or remove the
digital channels from the display, and to access the digital channel setup menu (MSO2000B series only).
h. Push to power on or off
raash drive.
h to perform an immediate
e / Recall menu.
31. Menu Off. Push to clear a displayed
menu from the screen.
32. Waveform Only. Push to remove menu
and readout information from the screen so the oscilloscope only displays the waveform or bus. Push a second time to recall the previous menu and readout information.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 33
Getting Acquain
Identifying Items in the Display
The items shown to the right may appear in the display. N at any given time. Some readouts move outside the graticule area when menus are turned off.
1. The acquisition readout shows when an acquisition is running, stopped, or when acquisition preview is in effect. Icons are:
ted with the Oscilloscope
ot all of these items are visible
Run: Acquisitions enabled
Stop: Acquisitions not enabled
Roll: In Roll mode (40 ms per division
or slower)
PreVu: In this state, the o scilloscope
is stopped or between triggers. You can change the horizontal or vertical position or scale to see approximately what the next acquisition will look like.
2. The trigger status readout shows trigger status. Status conditions are:
Trig'd: Triggered
Auto: Acquiring untriggered data
PrTrig: Acquiring pretrigger data
Trig?: Waiting for trigger
34 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Getting Acquain
ted with the Oscilloscope
3. The trigger pos
the trigger occurred in the acquisition.
4. The expansion point icon (an orange
triangle) horizontal scale expands and compresses around.
5. The waveform record view shows the
trigger l record. The line color corresponds to the selected waveform color.
6. The FilterVu indicator shows if the
e low pass lter is active.
variabl
ition icon shows where
shows the point that the
ocation relative to the waveform
7. The cursor readout shows time,
ude, and delta () values for each
amplit cursor.
For FFT measurements, it shows
ncy and magnitude.
freque For serial buses, the readout shows the
decoded values.
8. The trigger level icon shows the trigger
level on the waveform. The icon color
esponds to the trigger source color.
corr
MSO2000B and DP O2000B Series Oscilloscopes User Manual 35
Getting Acquain
ted with the Oscilloscope
9. The trigger rea source, slope, level, and frequency for an Edge trigger. The trigger readouts for other trig parameters.
10. The horizontal position/scale readout shows on the top line the horizontal scale (adjust with
With Delay Mode on, the bottom line shows the time from the T symbol to the expansion p Horizontal Position knob).
Use horizontal position to insert added delay betw and when you actually capture the data. Insert a negative time to capture more pretrigge
With Delay Mode off, the bottom line shows the time location of the trigger within th
11. Measurement readouts show the selecte up to four measurements to display at one time.
A expected numerical measurement if a vertical clipping condition exists. Part of the w display. To obtain a proper numerical measurement, turn the vertical scale and po waveform appear in the display.
dout shows the trigger
ger types show other
the Horizontal Scale knob).
oint icon (adjust with the
een when the trigger occurs
r information.
e acquisition, as a percentage.
d measurements. You can select
l appears instead of the
symbo
aveform is above or below the
sition knobs to make all of the
12. The auxiliary waveform readouts show the vertical and horizontal scale factors
e math and reference waveforms.
of th
13. The c hannel readout shows the channel
le factor (per division), coupling,
sca invert, and bandwidth status. Adjust with the Vertical Scale knob and the channel
, 3,or4 menus.
1, 2
36 MSO2000B and DPO2000B Series Oscilloscopes User Manual
14. For digital channels (MSO2000B series
only), the baseline indicators label the channel, and p levels. The color s follow the color code used on resistors. The D0 indicator is black, the D1 D2 indicator is red, and so on.
The bus display shows decoded packet level inform parallel buses (MSO2000B series only). The bus indicator shows the bus number and bus type
Not shown in this illustration, the Timing Resolutio resolution of the digital channels. To see the readout, push the D15-D0 front panel button.
oint to the high and low
indicator is brown, the
ation for serial buses or for
.
n readout shows the timing
Getting Acquain
ted with the Oscilloscope
15. For math c
baseline indicator shows the zero-volt level of a waveform.
16. For analog channels, the waveform
baseli level of a waveform (ignoring the effect of offset). The icon colors correspond to the wav
hannels, the waveform
ne indicator shows the zero-volt
eform colors.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 37
Getting Acquain
ted with the Oscilloscope
Front-Panel Connectors
1. Digital Probe
(MSO2000B series only).
2. Channel 1, 2,(3, 4). Channel inputs with TekVPI Versatile Probe Interface.
3. Aux In. Trigger level range is adjustable from +12.5 V t
4. PROBE COMP. source to compensate probes.
Output voltage: 0 V to 5 V Frequency
5. Ground.
6. Applicati
Connector
o –12.5 V.
Square wave signal
:1kHz
on Module Slots.
Side-Panel Connector
1. TekVPI external power supply connector.
Use the connector for the TekVPI
l power supply (Tektronix part
externa number 1197465XX) when additional power is needed for TekVPI probes.
38 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Rear-Panel Connectors
Getting Acquain
ted with the Oscilloscope
1. LAN.UsetheLA
connector) to connect the oscilloscope to a 10/100 Base-T local area network. The port is ava connectivity module (DPO2CONN).
2. Lock. Use to secure the oscilloscope
and optional connectivitiy module.
3. Video Out. Use the Video Out port
(DB-15 fema oscilloscope display on an external monitor or projector. The port is available on the opti (DPO2CONN).
4. USB 2.0 Device port.UsetheUSB
2.0 Full Speed Device port to connect aPictBri direct PC control of the oscilloscope using USBTMC protocol.
NOTE. Th
2.0 Device port to the host computer must meet the USB2.0 specication for high speed operati host controller.
dge compatible printer, or for
e cable connected from the USB
on when connected to a high speed
N (Ethernet) port (RJ-45
ilable on the optional
le connector) to show the
onal connectivity module
5. Power input. Attach to an AC power line
ntegral safety ground. (See page 5,
with i Operating Considerations.)
MSO2000B and DP O2000B Series Oscilloscopes User Manual 39
Acquire the Sign
al
Acquire the Si
This section describes concepts of and procedures for setting up the oscilloscope to acquire the signal as you want it to.
gnal
Setting Up Analog Channels
Use front-panel buttons and knobs to set up your oscilloscope to acquire signals u sing the analog channels.
1. Connect a TPP0200/TPP0100 probe or a TekVPI prob
2. Push Default Setup.
NOTE. If you are using a probe that does not
supply p (probe factor) on the oscilloscope vertical menu to match the probe. The default
tion for the oscilloscope is 10X and is
attenua set in the Probe Setup lower-bezel menu of any analog channel.
e to the input signal source.
robe encoding, set the attenuation
3. Select the input channel by pushing the front-panel buttons.
4. Push Autoset.
40 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Acquire the Sign
al
5. Push the desire
adjust the vertical position and scale.
6. Adjust the horizontal position and scale.
The horizontal position determines the number of pretrigger and posttrigger samples.
The horizontal scale determines the size of the acquisition window relative to the waveform. You can scale the window to contain a waveform edge, a cycle, several cycles, or thousands of cycles.
d channel button. Then
Quick Tip
Use the zoom feature to see multiple cycles of your signal in the upper part, and a single cycle in the lower part of the
display. (See page 105, Using Wave Inspector to Manage Long Record Length Waveforms.)
Labeling Channels and Buses
You c an waveform baseline indicator in the left side of the screen. The label can have up to 32 characters.
To label a channel, push a channel input button for an analog channel to label a channel.
add a label to the channels and buses shown on the display for easy identication. The label is placed on the
1. Push a front panel button for an input el or a bus.
chann
MSO2000B and DP O2000B Series Oscilloscopes User Manual 41
Acquire the Sign
al
2. Push a lower-b
ezel button to create a label,
such as for channel 1 or B1.
3. Turn multipu
rpose knob b to scroll through the list to nd a suitable label. You can edit the label after you insert it if necessary.
4. Push Insert Preset Label to add the label.
If you are using a USB keyboard, use the arrow keys to position the insertion point and edit the inserted label, or type in a new label. (See page 25, Connecting a USB Keyboard to Your Oscilloscope.)
Label
Insert
Preset
Label
5. If you do not have a USB keyboard connected, push the side- and lower-bezel
eys to position the insertion point.
arrow k
6. Turn multipurpose knob a to scroll through the list of letters, numbers, and other characters to nd the character in the name that you want to enter.
ABCDEFGHIJKLMNOPQRSTUVWXYZ
cdefghijklmnopqrstuvwxyz
ab 0123456789_=+-!@#$%^&*()[]{}<>/~'”\|:,.?
42 MSO2000B and DPO2000B Series Oscilloscopes User Manual
7. Push Select or Enter Character to let the
oscilloscope know that you have picked the proper charac
ter to use.
Acquire the Sign
al
8. Continue scrolling and pushing Select until
9. Push Display Labels and select On to see
Using th
To return the oscilloscope to its default settings:
1. Push Default Setup.
You can use the lower-bezel buttons to edit the label as n
you have ent
eeded.
ered all the desired characters.
For another label, push the side- and lower-bezel arrow keys to reposition the insertion
the label
point.
.
e Default Setup
Enter
Character
Display
Labels
On|Off
Back
Space
Delete
Clear
2. If you change your mind, push Undo Default Setup to undo the last default
Undo
Default
Setup
setup.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 43
Acquire the Sign
al
Using Autoset
Autoset adjusts the oscilloscope (acquisition, horizontal, trigger, and vertical controls) such that it displays four or ve waveform cycles for analog channels with the trigger near the middle level, and ten cycles for digital channels.
Autoset works with both the analog and digital channels.
1. Connect the analog probe, and then select the input channel. (See page 40, Setting Up Analog Channels.)
Connect the digital probe and select the input chann
Up Digital Channels (MSO2000B Series Only).)
2. Push Autoset to execute an Autoset.
el. (See page 59, Setting
3. If desired, push Undo Autoset to undo
Autoset.
the last
You can also disable the Autoset function if you want to set up a waveform manually. To disable or enable the Autoset function:
1. Push and hold Autoset.
2. Push a
3. Release Menu Off, and then release
nd hold Menu Off.
set.
Auto
Undo
Autoset
44 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Acquire the Sign
al
4. Select the desired side-bezel setting.
Autoset
Enabled
Autoset
Disabled
Quick Tips
To position the waveform appropriately, Autoset may change the vertical position. Autoset always sets vertical offset
to0V.
If you use Autoset when no channels are displayed, the oscilloscope turns on channel one (1) and scales it.
If you use Autoset and the oscilloscope detects a video signal, the oscilloscope automatically sets the trigger type to
video and makes other adjustments to display a stable video signal.
To manually set the oscilloscope to view a video signal in an IRE graticule:
1. Set the Trigger Type to Video.
2. Set the Vertical Scale to 143mV/div.
3. Select the proper input channel Attenuation for the probe or cable being used to feed the video signal to the oscilloscope.
4. If necessary, connect a 75 ohm pass-through terminator to the oscilloscope input. For example, you need to add a terminator when you use a 75-ohm cable between the video signal and the oscilloscope.
Acquisition Concepts
Before a signal can be displayed, it must pass through the input channel where it is scaled and digitized. Each channel has a de extracts waveform records.
Sampling Process
Acqui analog signal, converting it into digital data, and assembling it into a waveform reco memory.
dicated input amplier and digitizer. Each channel produces a stream of digital data from which the oscilloscope
sition is the process of sampling an
rd, which is then stored in acquisition
Input signal
Sampled points
Digital values
MSO2000B and DP O2000B Series Oscilloscopes User Manual 45
Acquire the Sign
Real-Time S ampling
al
MSO2000B and D oscilloscopes use real-time sampling. In real-time sampling, the oscilloscope digitizes all of the poin trigger event.
Waveform R
The oscilloscope builds the waveform record through the use of the following parameters:
Sample interval: The time between recorded sample points. Adjust this b y turning the Horizontal Scale knob or changing the record length with the bezel buttons.
Record length: The number of samples required to ll a waveform record. Set this by pushing the Acquire button and using the resulting lower- and side-bezel menus.
PO2000B series
ts it acquires using a single
ecord
Record points
Sampling rate
Trigger point: The zero time reference in a waveform record. It is shown on the screen by an orange T.
Horizontal position: When the Delay mode is on, this is the time from the trigger point to the expansion point. Adjust this
ning the Horizontal Position knob.
by tur When the Delay mode is off, the expansion point is xed to the trigger point. (Push the Acquire front panel button
to set the Delay mode.)
positive time to acquire the record after the trigger point. Use a negative time to acquire it before the trigger point.
Use a
Expansion point: The point that the horizontal scale expands and contracts around. It is shown by an orange triangle.
46 MSO2000B and DPO2000B Series Oscilloscopes User Manual
How the Analog Acquisition Modes Work
FilterVu Noise Filter Foreground mode retains the rst sampled point from each acquisition i Sample in other oscilloscopes. T his is the default mode.
FilterVu Glitch Capture Background mode uses the high samples contained in two consecutive acquisition intervals. At fast time per division settings, t mode is unavailable. This mode is also called Peak Detect in other oscilloscopes.
Average mode calculates the average value for each record point over a user-specied number of a Sample mode for each individual acquisition. Use Average mode to reduce random noise.
nterval. This mode is also called
est and lowest of all the
he Glitch Capture Background
cquisitions. Average uses
Acquire the Sign
al
Changin
Use this procedure to change the acquisition mode.
1. Push Acquire.
2. Push Average.
g the Acquisition Mode, Record Length, and Delay Time
Average
Off
Record Lengt
100k
Delay
h
On |
Set Horiz.
Position to
ff
O
10%
Waveform
Display
XY Display
Off
Acquisition
ls
Detai
MSO2000B and DP O2000B Series Oscilloscopes User Manual 47
Acquire the Sign
al
3. Set the Average acquisition mode from the side-bezel menu. You can chose how many samples on whi
ch to average: 2, 4, 8, 16,
32, 64, 128, 256, or 512.
4. Turn multipurpose knob a to set the number of waveforms
to average over.
5. Push Record Length. Choose 100 k or 1.00 M points. The
selections depend on the horizontal time per division setting. 125 k and 1.25 M records lengths are available at slower time per division settings.
6. Push the lower-bezel Delay buttontoselect On when you want to delay the acquisition
relative to the trigger event.
Average
Average
On |Off
With Delay set to On, turn the Horizontal P osition knob counterclockwise to increase the delay. The trigger point will move
left and ultimately outside of the acquired waveform. Then you can adjust the Horizontal Scale knob to acquire more
to the detail around the area of interest at the center of the screen.
When this delay is on, the trigger point separates from the horizontal expansion point. T he horizontal expansion point stays at the center of the screen. The trigger point can move off the screen. When this happens, the trigger marker turns
int in the direction of the trigger point.
to po
he delay feature when you w ant to acquire waveform detail that is separated from the trigger event by a signicant
Use t interval of time. For example, you can trigger on a sync pulse that occurs once every 10 ms and then look at high-speed signal characteristics that occur 6 ms after the sync pulse.
When the delay feature is s et to Off, the expansion point is tied to the trigger point so that scale changes are centered
und the trigger point.
aro
48 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Using Roll Mode
Roll mode gives a display similar to a strip chart recorder fo r low-frequency signals. Roll mode lets you see acquired data points without waiting for the acquisition of a complete waveform record.
Roll mode is enabled when the trigger mode is auto and the horizontal scale is set to 40 ms/division or slower.
Quick Tips
Roll mode is disabled when you switch t o Average acquisition mode, use digital channels, use math waveforms, turn on
abus,switc
h to Normal trigger, or set the horizontal scale to 20 ms per division or faster.
Acquire the Sign
al
Zoom is disa
Push Run/Stop to halt Roll mode.
bled when you use Roll mode.
SettingUpaSerialorParallelBus
Your oscilloscope can decode and trigger on:
I2C and SPI serial buses if the DPO2EMBD application module is installed
CAN and LIN serial buses if the DPO2AUTO application module is i nstalled
RS-232, RS-422, RS-485, and UART serial buses if the DPO2COMP application module is installed
Parallel buses if you are using an MSO 2000B series oscilloscope
(See page 13, Application Module Free Trial.)
Using
Buses in Two Steps
To quickly use serial bus triggering:
1. Push B1 or B2 and enter parameters of the bus to trigger on.
You can separately use B1 and B2 to view two different buses.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 49
Acquire the Sign
al
2. Push Trigger Me parameters. (See page 67, Choosing a Trigger Type.)
You can displa triggering on the bus signal.
nu and enter trigger
y bus information without
SettingUpBusParameters
NOTE. For all serial bus sources, use any combination of channels 1 through 4, and D15 through D0.
To trigger on serial or parallel bus conditions, refer to Triggering on Buses. (See page 70, Triggering on Buses.)
50 MSO2000B and DPO2000B Series Oscilloscopes User Manual
To set up bus parameters:
1. Push B1 or B2 to bring up the lower-bezel bus menu.
Acquire the Sign
al
2. Push Bus. Turn multipurpose knob a to scroll through the list of bus types and select the desired bus: Parallel (MSO2000B series
2
only), I
C, SPI, CAN, RS-232, or LIN.
The actual menu items shown will depend on your model oscilloscope and the application modules installed.
3. Push Dene Inputs. The choices depend on the selected bus.
Use the side-bezel buttons to dene parameters for the inputs, such as specific signals to an analog or digital channel.
If you select Parallel, push the side-bezel button to enable or disable Clocked Data .
Push the side-bezel button to select the Clock Edge on which to clock data: rising edge, falling edge, or both edges.
Turn multipurpose knob a to select the Number of Data Bits in the parallel bus.
Turn multipurpose knob a to select the desired bit to dene.
Turn multipurpose knob b to select the desired analog or digital channel as the source for the bit.
B1
Parallel
Dene Inputs
Clocked
Data
No
Yes |
Clock Edge
Number of
Data Bits
(a) 16
Dene Bits
(a) Bit 15
(b) D15
Dene Inputs
Thresholds B1 Label
Parallel
Bus
Display
Event Table
MSO2000B and DP O2000B Series Oscilloscopes User Manual 51
Acquire the Sign
al
4. Push Threshol
You c an set the
ds.
threshold for all channels in the parallel or serial bus from a list of preset values. The values are based on common types of inte
grated circuits. The preset
values are:
1.4 V for TTL
2.5 V for 5.0
VCMOS
1.65 V for 3.3 V CMOS
1.25 V for 2.5 V CMOS
-1.3 V for E
CL
3.7 V for PECL 0V
Alternate
ly, you can set the threshold to a specic value for the signals that makes up the parallel or serial bus. To do so, push the S
elect side-bezel button and turn multipurpose knob a to select a Bit or a Channel number (Signal name).
Bus
Parallel
Dene Inputs
Thresholds B1 Label
Parallel
Bus
Display
Event Tab l e
Then, turn multipurpose knob b to dene the voltage level above which the oscilloscope treats the signal as high and below which as low.
5. Push B1 Label to edit the label for the bus. (See page 41, Labeling Channels and
s.)
Buse
Bus
Parallel
Dene Input
holds
Thres
s
B1 Lab
Parallel
el
Bus
Display
Event Tab l e
52 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Acquire the Sign
al
6. Push Bus Displ
ay and use the side-bezel
menu to dene how to display the parallel or serial bus.
Push the desir
ed side-bezel menu choice to display the bus data in hex, binary, or ASCII (RS-232 only) format.
7. Push Event Table and select On to display alistofI with times
For a cloc
2
C, SPI, CAN, or LIN bus packets
tamps.
ked parallel bus, the table lists the value of the bus at each clock edge. For an unclocked parallel bus, the table lists the value of t
he bus whenever any one of its bits
changes. For an RS-232 bus, the table lists decoded
r packets.
bytes, o
Hex
Binary
ASCII
Event
Tab l e
On |Off
Save
Event
Tab l e
8. Push Sav
e Event Table to save the event
table data in a .csv (spreadsheet) formatted le on the c urrently selected storage device.
This example of an Event Table is from an RS-232 bus.
2 event tables display one line for each
RS-23 7- or 8-bit byte when Packets are set to Off. RS-232 event tables display one line for each
t when Packets are set to On.
packe
2
I
C, SPI, CAN, and LIN event tables display
one line for each packet.
9. Push B1 or B2 and turn multipurpose knob a to move the bus display up or down on the
screen.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 53
Acquire the Sign
To acquire data from an I2C bus, you need to also set up these items:
al
1. If you select I
2C, push Dene Inputs and
the desired side-bezel menu choices.
B1
I2C
Dene Inputs
Thresholds Include
R/W in
Address
No
B1 Label
I2C
Bus
Display
You can assign the predened SCLK Input or SDA Input to the channel connected to the signal.
2. Push Includ e R/W in Address and then push the des
ired side-bezel button.
This control determines how the oscilloscope shows the I traces, cur
2
C addresses in bus decode
sor readouts, Event Table listings,
and trigger settings.
If you sele
If you sel
The oscil bit is the R/~W bit. The last eight bits are the eight LSBs of the address. (In the physical layer of the I
ct Yes, the oscilloscope displays 7-bit addresses as eight bits, where the eighth bit (LSB) is the R/~W bit.
ect N o, the oscilloscope displays 7-bit addresses as seven bits, and 10-bit addresses as ten bits.
loscope also displays 10-bit addresses as 11 bits. The rst two bits are the two MSB s of the address. The next
2
C protocol, 10 bit I2C
addresses are preceded by the ve bit code, 11110. The oscilloscope never includes these ve bits in address readouts.)
Event Tabl e
54 MSO2000B and DPO2000B Series Oscilloscopes User Manual
To acquire data from an SPI bus, you need to also set up these items:
Acquire the Sign
al
1. If you selected SPI, push Dene Inputs and
the desired side-bezel menu choices.
You c an set the Framing to SS (Slave Select) or to Idle Time.
You can assign the predened SCLK, SS,
MOSI,orMISO signals to any channel.
2. Push Congure and the desired side-bezel
menu choices.
3. Push SCLK to set the edge of the signal to
match the SPI bus being acquired.
4. Set the level of the SS, MOSI, and MISO
signals to match the SPI bus.
Active High means that a signal is considered active when the signal is greater than the threshold value.
Active Low means that the signal is considered active when the signal is lower than the threshold value.
Bus
SPI
SCLK
SS
Active
High
Active Low
MOSI
Active
High
Active Low
MISO
Active
High
Active Low
-more­10f2
Dene
Inputs
Thresholds
Congure
B1 Label
SPI
Bus
Display
Event Tab l e
5. Use multipurpose knob a to set the number of bits to match the Word Size of the SPI bus.
6. Push either side bezel button to set the bit order to match the SP I bus.
Word Size
(a) 8 bits
Bit Order
MS First
Bit Order
LS First
To acquire data from a CAN bus, you need to also set up these items:
1. If you selected CAN, push Dene Inputs and the desired side-bezel menu choices.
Bus
CAN
Dene
Inputs
Thresholds Bit Rate
500 Kbps
B1 Label
CAN
Bus
Display
Event Tab l e
MSO2000B and DP O2000B Series Oscilloscopes User Manual 55
Acquire the Sign
al
2. Turn multipur
pose knob a to select the
channel connected to the CAN bus source.
3. Turn multipurpose knob a to select the type of CAN signal to match the source of the CAN bus: CAN_H, C
AN_L, Rx, Tx, or Differential.
4. Turn multipurpose knob a to set the Sample Point from 5
% to 95% of the position within
the bit period or the unit interval.
5. Push Bit Rate and turn multipurpose knob a to select f
rom the list of predened bit rates.
Alternately, you can set the bit rate to a specic value. To do so, select Custom, and then turn
multipurpose knob b to set the bit
rate from 10000 to 1000000.
To acqui
re data from a LIN bus, you need to also set up these items:
1. If you selected LIN, push Dene Inputs and the desired side-bezel menu choices.
CAN Input
(a) 1
Signal
Type
CAN_H
Sample
Point
50%
Bus
CAN
Bus
LIN
Dene Inputs
Dene Inputs
Thresholds Bit Rate
500 Kbps
Thresholds
Congure
B1 Label
CAN
B1 Label
LIN
Bus
Display
Bus
Display
Event Tabl e
Event Tab l e
2. Turn multipurpose knob a to select the channel connected to the LIN bus source.
3. Turn multipurpose knob a to set the Sample Point from 5% to 95% of the position within
the bit period or the unit interval.
4. Select the Polarity to match the LIN bus being acquired.
5. Push Congure and the desired side-bezel menu choices.
LIN Input
(a) 1
Sample
Point
50%
Polarity Normal
(High=1)
Polarity
Inverted
(High=0)
Bus
LIN
Dene Inputs
Thresholds
Congure
B1 Label
LIN
Bus
Display
Event Tab l e
56 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Acquire the Sign
al
6. Push Bit Rate,
and turn multipurpose knob a
to s e lect from the list of predened bit rates. Alternately, you can set the bit rate to a
specic value
.Todoso,selectCustom, and
Bit Rate
(a)
19.2K bps
then turn multipurpose knob b to set the bit rate from 800 bps to 100000 bps.
7. Push LIN Standard, and turn multipurpose
knob a to select the appropriate standard.
8. Push In clud
e Parity Bits with Id to select
whether or not to include parity bits.
LIN
Standard
v1.x
Include
Parity Bits
with Id
On|
Off
To acquire data from a RS-232 bus, you need to also set up these items:
1. If you sel
ected RS-232, push Congure and
the desired side-bezel menu choices.
Use the si
de-bezel menu to congure the
Bus
RS-232
Dene
Inputs
bus. Use Normal polarity to trigger on RS-232 signals and Inverted polarity to
on RS-422, RS-485, and UART
trigger signals.
Thresholds
Congure
9600-8-N
B1 Label
RS-232
Bus
Display
Event Tab l e
2. Push Bit Rate, and turn multipurpose knob a
ct the appropriate bit rate.
to sele
3. Push Data Bits and select the number to
he bus.
match t
4. Push Parity and turn multipurpose knob a to
the polarity used by the bus as None,
match Odd, or Even.
5. Push
Packets and select On or Off.
6. Turn multipurpose knob a to select an
of-packet character.
end-
32 decoding displays a stream of bytes.
RS-2 You can organize the stream into packets with an end-of-packet character
Bit Rate
9600 b
Data Bits
7|
8
Parity
one
(a) N
Packets
Off
On|
End of Packet
0A
(Linefeed)
ps
MSO2000B and DP O2000B Series Oscilloscopes User Manual 57
Acquire the Sign
Physical Layer Bus Activity
Oscilloscope waveform traces from analog channels 1 to 4, digital channels D15 to D0, and the traces y ou see when you choose to display a bus always show the physical layer bus activity. In the physical layer display, bits that were transmitted earlier are to the left, and bits that were transmitted later are to the right.
I2C, and CAN buses transmit the MSB (most signicant bit) rst
SPI buses do not specify a bit order
RS-232 and LIN buses transmit the LSB (least s ignicant bit) rst
NOTE. The oscilloscope displays the decode traces and event tables for all buses with the MSB on the left and LSB
on the right.
For example, an RS-232 signal (after the start bit) might be high, high, high, low, high, low, low, and high. Since the RS-232 protocol uses high for zero and low for one, this value would be 0001 0110.
Since the decode displays the MSB rst, the oscilloscope reverses the order of the bits and displays 0110 1000. If the bus display i s s et to hex, the value displays as 68. If the bus display is set to ASCII, the value displays as h.
al
RS-232
If you dened an end-of-packet character to use for RS-232 decoding, the stream of bytes will be displayed as packets.
When decoding an RS-232 bus in ASCII mode, a large dot indicates that the value represents a character outside the printable ASCII range.
58 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Setting Up Digital Channels (MSO2000B Series Only)
Use front-panel buttons and knobs to set up your oscilloscope to acquire signals using the digital channels available only on M SO 2000B models.
1. Connect the P6316 digital probe to the input
signal source.
Acquire the Sign
al
2. Connect the ground lead or leads to the
circuit ground. You can connect one or two of the common
ground leads for each group of 8 channels (wires).
3. If needed, connect the appropriate grabber
for each channel to the probe tip.
4. Connect each channel to the desired circuit
test point.
5. Push the D15 - D0 front panel button to
display the menu.
6. Push the lower-bezel D15 - D0 button to
access the D15 - D0 On or Off menu.
D15–D0
On/Off
Thresholds Edit Labels Height
S |ML
MSO2000B and DP O2000B Series Oscilloscopes User Manual 59
Acquire the Sign
al
7. Turn multipurp the list of digital channels. Turn multipurpose knob b to position the selected channel.
As you positio on the display, the oscilloscope groups the channels, and adds the group to the pop-up list. You can move all the channels in the group instead of individual channels.
8. Push the low You can assign a different threshold value to each pod.
9. Push the lo and create the label. You can create labels through the front panel or with an optional USB keyboa Channels and Buses.)
10. Push the lower-bezel Height button repeated need to do this once to set the height for all of the digital channels.
ose knob a to scroll through
n channels close to each other
select the group from the list to
er-bezel Thresholds button.
wer-bezel Edit Labels button
rd. (See page 41, Labeling
ly to set the signal height. You only
Quick Tip
Use the zoom feature to see multiple cycles of the signal in the upper part, and a single cycle in the lower part of the display. (See page 105, Using Wave Inspector to Manage Long Record Length Waveforms.)
The lead for each digital channel has a color-coded band for easy identication. The ground leads are shorter and have black wires.
Digital channels store a high or low state for each sample. The threshold that separates a high from a low can be set for all the channels in GROUP 1 or in GROUP 2. The threshold can not be set for individual channels.
Reducing Unwanted Noise With FilterVu
You can use F ilterVu to lter unwanted noise from your signal and still capture glitches. The oscilloscope does this by
rimposing a ltered foreground waveform over a glitch capture background waveform.
supe
ltered waveform uses a variable low-pass lter to block out noise, yielding a cleaner waveform. When the noise lter is
The adjusted to the minimum bandwidth setting, no more than 1% of high-frequency content that could cause the oscilloscope to alias will pass through the lter.
The glitch capture waveform shows signal details up to the full bandwidth of the oscilloscope. The oscilloscope captures
ses as narrow as 5 ns, using peak detect min/max sampling.
pul
en your signal is ltered, the glitch capture waveform changes color to distinguish it from the ltered waveform. When
Wh the signal is ltered, you can independently adjust the intensity of the glitch capture waveform. When the signal is not ltered, the intensity o f the glitch capture waveform depends on the overall intensity (under the Intensity button on the
cilloscope front panel).
os
60 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Acquire the Sign
FilterVu, Bandwidth Limit, and Averaging
You can use FilterVu, Bandwidth Limits, or Average mode to reduce noise in your signal. Each has benets.
NOTE. FilterVu does not apply to triggering. To reduce the effects of noise on your signal while triggering, use Bandwidth
Limits or use Coupling settings.
Use FilterVu when you want to do the following:
Adjust your lter frequency with more granularity than the Bandwidth Limit provides.
Adjust all channels to the same frequency.
Acquire a non-repetitive or a single-shot signal.
al
Use Bandwidth Limit when you w
Filter only one channel.
Trigger on a ltered signal.
Acquire a non-repetitive or a single-shot signal.
Use Average mode when you want to do the following:
Acquire a repetitive signal.
Noise on your signal is random (not correlated with the trigger).
Increase the vertical precision of your acquisition.
NOTE. You can use FilterVu with Bandwidth Limit. You cannot use FilterVu with Average mode.
ant to do the following:
Quick Tips
You can choose to save to a le with full resolution (every point that was captured) or reduced resolution (fewer points). When your signal is ltered, only the reduced-resolution waveform is guaranteed to be ltered to the chosen frequency. Sometimes, the lter is unable to produce a full-resolution waveform with a frequency as low as the chosen frequency. The frequency of the full-resolution waveform is shown in the saved le.
When the glitch capture background is on, a saved waveform le includes both the foreground and the background data.
Reference waveforms can be ltered; the chosen noise lter frequency also applies to references. Reference waveforms also have a glitch capture background.
Reference waveforms are always saved at full resolution. Unlike saving a waveform to a le, there is no reduced resolution option.
When the signal is not ltered, most measurements are taken on the foreground waveform. The min, max, and peak-to-peak measurements are taken on the glitch capture background, because they measure the smallest and largest amplitudes in the waveform.
When the signal is ltered, all measurements are taken from the ltered waveform.
Dual-waveform math can be ltered; the chosen noise lter frequency also applies to math. Dual-waveform math does not have a glitch capture background.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 61
Acquire the Sign
al
Search uses the glitch capture background, when available. This lets you better search for spikes and other anomalies.
Serial bus decode uses a full-resolution sample (foreground) waveform. When ltering is active, that waveform might not be ltered to t
Using FilterVu
By default, the noise lter cutoff frequency is set to the full bandwidth of the oscilloscope when you use FilterVu. To invoke the variable low-pass lter (noise lter), turn the multipurpose knob counterclockwise. The oscilloscope passes lower frequency signals which results in a cleaner-looking waveform. The noise frequency readout displays frequencies that the oscilloscope includes in the foreground waveform. The oscilloscope displays higher frequency components in the background waveform, up to the bandwidth of the MSO2000B or DPO2000B model.
NOTE. The o scilloscope always displays the noise lter cutoff frequency in the upper right part of the screen. The cutoff
frequency reects the -3 dB point.
1. Push the FilterVu button.
he chosen frequency.
2. Turn multipurpose knob a to adjust the Noise Filter (Foreground) value.
eFil-
Nois ter (Fore-
ground)
(a)
200MHz
62 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Acquire the Sign
al
3. Push Glitch Capture Background to turn
off the background waveform that contains the higher fre
quency components of the
signal. With Glitch Capture set to Off, the
oscilloscop
e displays only the noise lter
(foreground) waveform.
4. With Glitch
Capture set to On, turn multipurpose knob b to adjust the intensity of the Glitch Capture (background) waveform. You c an only
adjust the intensity w hen
ltering.
Glitch
Capture
Back-
ground
On |Off
Back-
ground
Intensity
Auto
MSO2000B and DP O2000B Series Oscilloscopes User Manual 63
Trigger Setup
Trigger Setup
This section contains concepts and procedures for setting up the oscilloscope to trigger on your signal.
Triggering Concepts
Trigger Event
The trigger event establishes the time-reference point in the waveform record. All waveform record data is located in time with respect to that point. The oscilloscope continuously acquires and retains enough sample points to ll the pretrigger portion of the waveform record. That is the part of the waveform that is displayed before, or to the left of, the triggering event on screen. When a trigger event occurs, the oscilloscope starts acquiring samples to build the posttrigger portion of the waveform record, that is, the part displayed after or to the right of the trigger event. After a trigger is recognized, the oscilloscope will not accept another trigger until the acquisition is complete and the holdoff time has expired.
Untrigg
ered display
Trigger
ed display
Trigger Modes
The trigger mode determines how the oscilloscope behaves in the absence of a trigger event:
Normal trigger mode enables the oscilloscope to acquire a waveform only when it is triggered. If no trigger occurs, the last waveform record acquired remains on the display. If n o last waveform exists, no waveform is displayed.
Auto trigger mode enables the oscilloscope to acquire a waveform even if a trigger does not occur. Auto mode uses a timer that starts when the acquisition is started, and the pretrigger information is obtained. If a trigger event is not detected before the timer times out, the oscilloscope forces a trigger. The length of time it waits for a trigger event depends on the time base setting.
Auto mode, when forcing triggers in the absence of valid triggering events, does not synchronize the waveform on the display. The waveform will appear to roll across the screen. If valid triggers occur, the display will become stable.
You can also force the oscilloscope to trigger. To do so, push the front-panel Force Trig button.
64 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Trigger Holdoff
Trigger Setup
Adjust holdof when the oscilloscope is triggering on undesired trigger events.
Trigger holdo since the oscilloscope does not recognize new triggers during the holdoff time. When the oscillos it disables the trigger system until acquisition is complete. In addition, the trigger system remains dis that follows each acquisition.
f to obtain stable triggering
ff can help stabilize triggering,
cope recognizes a trigger event,
abled during the holdoff period
Holdoffs
Trigger Coupling
Trigger coupling determines what part of the signal is passed to the trigger circuit. Edge triggering can use all available coupling types: DC, Low Frequency Rejection, High Frequency Rejection, and Noise Rejection. All other trigger types use DC coupling only.
The cutoff frequency for low frequency rejection is 65 kHz. The cutoff frequency for high frequency rejection is 85 kHz.
Horizontal Position
When Delay Mode is on, use horizontal position to acquire waveform detail in a region that is separated from the trigger location by a signicant interval of time.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 65
Trigger Setup
1. Adjust the position (delay) time by
2. Turn the horizontal Scale knob to acquire
The part of the record that occurs before the trigger is the pretrigger portion. The part that occurs after the trigger is the posttrigger portion. Pretrigger data can help you troubleshoot. For example, to nd the cause of an unwanted glitch in your test circuit, you can trigger on the glitch and make the pretrigger period large enough to capture data before the glitch. By analyzing what happens before the glitch, you may uncover information that helps you nd the source of the glitch. Alternatively, to see what is happening in your system as a result of the trigger event, make the posttrigger period large enough to capture data after the trigger.
Slope and Level
rotating the Horizontal Position knob.
the detail th position (delay) expansion point.
at you need around the
The slope control determines whether the oscilloscope nds the trigger point on the rising or the falling edge of a signal.
The level control determines where on that edge the trigger point occurs.
The oscilloscope provides a long horizontal bar or bars across the graticule to temporarily show the trigger level.
1. Turn the front-panel Trigger Level knob just the trigger level without going
to ad to a menu.
2. Push the Force Trig button to force the
oscilloscope to trigger.
66 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Choosing a Trigger Type
To select a trigger:
1. Push Trigger Menu.
2. Push Type to bring up the list of Trigger
Types.
Trigger Setup
NOTE. The bu
s trigger in the MSO2000B series works on parallel buses even without an application module. Using the bus trigger on other buse
s requires use of the DPO2EMBD, DPO2AUTO, or DPO2COMP application module.
3. Turn multipurpose knob a to select the
desired trigger type.
ger types are: Edge, Pulse Width,
The trig Runt, Logic, Setup & Hold, Rise/Fall Time, Video, and Bus.
4. Complete the trigger setup using the
lower-bezel menu controls displayed for the trigger type. The controls to set up the trigger vary depending on the trigger type.
Type
Edge
Source
1
Coupling
DC
Slope
Level
100 mV
Mode
Auto
& Holdoff
MSO2000B and DP O2000B Series Oscilloscopes User Manual 67
Trigger Setup
Selecting Triggers
Trigger type Trigger conditions
Edge
Pulse Width Trigger on pulses that are less than, greater than, equal
Runt Trigger on a pulse amplitude that crosses one threshold
Trigger on a rising or falling edge, as dened by the slope control. Coupling choices are DC, LF Reject, HF Reject, and N
Edge triggers are the simplest and most commonly used trigger type, with both analog and digital signals. An edge trigg passes through a specied voltage level in the specied direction.
to, or not equal to a specied time. You can trigger on positive o primarily used on digital signals.
but fails the rst. You can detect positive or negative (or either) runts, or only those wider than, less than, greater than, equal to are primarily used on digital signals.
oise Reject.
er event occurs when the trigger source
r negative pulses. Pulse width triggers a re
to cross a second threshold before r ecrossing
, or not equal to a specied width. Runt triggers
Logic
Trigger when all channels transition to the specied state. Use multipurpose knob a to select a channel. Push the appropriate side-bezel button to set that channel's state to High (H), Low (L),orDon't Care (X).
Use the Clo c k side-bezel button to enable clocked (state) triggering. You can have at most a single clock channel. Push the Clock Edg e bottom bezel button to change the polarity of the clock edge. Turn off clocked triggering and return to unclocked (pattern) triggering by selecting the clock channel and setting it to high, low, or don't care.
For unclocked triggering, by default, triggering occurs when the selected condition goes true. You can also select triggering when the condition goes false, or time-qualied triggering.
You can use up to 20 channels for a logic trigger (4 analog and 16 digital) with MSO2000B series oscilloscopes.
68 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Trigger Setup
Trigger type Trigger conditions
Setup and Hold Trigger when a logic data input changes state inside of
the setup or ho
ld time relative to a clock edge.
Setup is the amount of time that data should be stable and not change before a clock edge occurs. Hold is the time that data
should be stable and not change after
a clock edge occurs. MSO2000B series oscilloscopes are capable o f multiple
channel Setu
p and Hold triggering, and can monitor the state of an entire bus for setup and hold violations. You can use up to 20 channels for a Setup and Hold trigger (4 a
nalog and 16 digital) with MSO2000B series
oscilloscopes. Use the Clock side-bezel button to select the clock
channel. U
se the Select control, Data,andNot u sed buttons to select one or more channels you want to monitor for setup and hold violations.
Rise/Fall Time Trigger on rise and fall times. Trigger on pulse edges
erse between two thresholds at faster or slower
that trav rates than the specied time. Specify pulse edges as positive or negative or either.
Video
Trigger on specied elds or lines of a composite video
Only composite signal formats are supported.
signal. Trigger on NTSC, PAL, or SECAM. Works with
Macrovision signals.
Bus Trigger on various bus conditions.
2
quires a DPO2EMBD module.
Cre
I SPI requires a DPO2EMBD module. CAN requires a DPO2AUTO module.
equires a DPO2AUTO module.
LIN r RS-232, RS-422, RS-485, and UART require a
DPO2COMP module.
llel requires an MSO2000B series oscilloscope.
Para (See page 13, Application Module Free Trial.)
MSO2000B and DP O2000B Series Oscilloscopes User Manual 69
Trigger Setup
Triggering on Buses
You can use your oscilloscope to trigger on CAN, I2C, SPI, RS-232, RS-422, RS-485, UART, and LIN buses, if you have the appropriate DPO2AUTO, DPO2EMBD, or DPO 2CO M P application m odule installed. The MSO2000B series can trigger on parallel buses without an application module. The oscilloscope can display both physical layer (as analog waveforms) and protocol level information (as digital and symbolic waveforms).
To set up the bus trigger:
1. If you have not already dened your bus using the front-panel B1 and B2 buttons, do so now. (See page 49, Setting Up a Serial or Parallel Bus.)
2. Push Trigger Menu.
3. Push Type.
4. Turn multipurpose knob a to scroll through the list of trigger types until you select Bus.
5. Push Source Bus and turn multipurpose
knob a to scroll through the source bus side menu until you select the bus that you want to trigger on.
Type
Bus
B1 (I2C)
B2
(Parallel)
Source
Bus
B1 (I2C)
Trigger On
Address
Address
07F
Direction
Write
Mode
Auto
& Holdoff
70 MSO2000B and DPO2000B Series Oscilloscopes User Manual
6. Push Trigger On and turn multipurpose
knob a to scroll through the side-bezel menu until you
select the desired trigger on
feature. If you are using the parallel bus trigger, you
can trigger o
n a binary or hex data value. Push the lower-bezel Data button and enter the parameters of interest with multipurpose knobs a and b.
IfyouareusingtheI
2
C bus trigger, you
can trigger on Start, Repeated Start,
Stop, Missi
ng Ack, Address, Data,or
Address/Data.
If you are using the SPI bus trigger, you can trigger on
SS Active, MOSI, MISO,orMOSI
&MISO.
If you are using the CAN bus trigger, you can trigger on
Start of Frame, Type of Frame, Identier, Data, Id & Data, End of Frame, Missing Ack,orBit Stufng Error.
If you are using the RS-232 bus trigger, you can trigger on Tx Start Bit, Rx Start Bit, Tx
End of Pa
cket, Rx End of Packet,Tx Data, Rx Data, Tx Parity Error,orRx Parity Error.
Trigger Setup
If you are using the LIN bus trigger, you can trigger on Sync, Identier, Data, Id
Wakeup Fram e, Sleep Frame,or
&Data, Error.
7. If you
are setting up an I
2
C trigger and have made a Trigger On selection of Address or Address/Data, push the lower-bezel
ess button to access the I
Addr
2
CAddress
side-bezel menu. Push the side-bezel Addressing Mode
on and select 7bitor 10 bit. Push
butt the side-bezel Address button. Enter the address parameters of interest with
tipurpose knobs a and b.
mul Then push the lower-bezel menu Direction
button and select the direction of interest:
d, Write,orRead or Write.
Rea
MSO2000B and DP O2000B Series Oscilloscopes User Manual 71
Trigger Setup
8. If you are setting up an SPI trigger and have
If you have made a Trigger On selection of Data or Address/Data, push the lower-bezel D
ata button to access the I
2
C
Data side-bezel menu. Push the Number of Bytes button and
enter the num
ber of bytes with multipurpose
knob a. Push the side-bezel Addressing Mode
button and se
lect 7bitor 10 bit. Push the side-bezel Data button. Enter the data parameters of interest with multipurpose knobs a and b
For more information on the I
.
2
C address
formats, refer to item 2 under Setting Up
Bus Parame
made a Tri
ters.
gger On selection of MOSI or
MISO, push the lower-bezel Data button,
push the side-bezel MOSI (or MISO) button, and enter
the data parameters of interest
with multipurpose knobs a and b. Then push the Number of Bytes button and
e number of bytes with multipurpose
enter th knob a.
If you select MOSI & MISO, push the
bezel Data button and enter the
lower­parameters of interest in the side-bezel menus.
9. If you are setting up a CAN trigger and have made a Trigger On selection of Type of
, push the lower-bezel Frame Type
Frame
button, and select Data Frame, Remote Frame, Error F rame, or Overload Frame.
u have made a Trigger On selection of
If yo Identier, push the lower-bezel Identier button, and select a Format. Then push the
tier side-bezel button, and enter a
Iden
binary o r hex value with multipurpose knobs a and b.
h the lower-bezel menu Direction button
Pus and select the direction of interest: Read, Write,orRead or Write.
ou have made a Trigger On selection
If y of Data. Push the lower-bezel Data button and enter the parameters of interest with
ltipurpose knobs a and b.
mu
72 MSO2000B and DPO2000B Series Oscilloscopes User Manual
10. If you are setting up an RS-232 trigger and
have made a Trigger On selection of Tx Data or Rx Data
, push the lower-bezel Data
button. Push the Number of Bytes button and
enter the num
ber of bytes with multipurpose
knob a. Push the side-bezel Data button, and enter
the paramete
rs of interest with multipurpose
knobs a and b.
Trigger Setup
MSO2000B and DP O2000B Series Oscilloscopes User Manual 73
Trigger Setup
11. If you are setting up an LIN trigger and
I2C, SPI, CAN, and LIN Bus Trigger Data Matching
Rolling window byte matching for I2C, and SPI. To use a rolling window to trigger on data, you dene the number
of bytes to match. Then the oscilloscope uses a rolling window to nd any match within a packet, with the window rolling one byte at a time.
For example, if the number of bytes is one, the oscilloscope will attempt to match the rst byte, second byte, third, and so on within the packet.
If the number of bytes is two, the oscilloscope will try to match any two consecutive bytes, such as one and two, two and three, three and four, and so on. If the oscilloscope nds a match, it will trigger.
have made a Trigger On selection of Identier, Da the lower-bezel Identier or Data buttons and enter the parameters of interest on the resulting si
If you have made a Trigger On selection of Error, push the lower-bezel Error Type button and en on the resulting side-bezel menu.
ta,orIdentier & Data,push
de-bezel menu.
ter the parameters of interest
Specific and CAN.
byte matching (non-rolling window matching for a specic position in the packet) for I
You can trigger on a specicbyteforI2C, SPI, and CAN in several ways:
For I2C and SPI, enter the number of bytes to match the number of bytes in the signal. Then use don't cares (X) to mask the bytes that you are not interested in.
For I2C, push the lower-bezel Trigger On to trigger on Address/Data. Push Address. On the side-bezel menu, push
s, and turn multipurpose knobs a and b as needed. Set the address to don't cares (X) if you want to mask the
Addres
address. The data will be matched starting at the rst byte without using a rolling window.
For CAN, triggering occurs when the user-selected data input matches the data and qualier in the signal starting at the rst byte. Set the number of bytes to match the number of bytes of interest. Use the data qualier to perform: =,
>, >=, and <= operations. Triggering on identier and data always matches the identier and data selected by
!=, <, the user, with the data starting at the rst byte. No rolling window is used.
2
C, SPI,
RS-232 Bus Trigger Data Matching
You can trigger on a specic data value for RS-232 bytes. If you d ened an end-of-packet character to use for RS-232 bus decoding, you can use the same end-of-packet character as a data value for trigger data matching. To do so, choose the Tx End of Packet or the Rx End of Packet character as the Trigger On selection.
ParallelBusTriggerDataMatching
Optimum parallel bus trigger performance is achieved by using only analog channels or only digital channels (MSO2000B series only).
74 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Checking Trigger Settings
Trigger Setup
To quickly det key trigger parameters, check the Trigger readout at the bottom of the display. The readouts diff triggers.
1. Trigger source = channel 1.
2. Trigger slope = rising.
3. Trigger level = 1.84 V.
4. 6-Digit Trigger Frequency Readout =
68.901 Hz.
Starting a
After you have dened the acquisition and trigger parameters, start the acquisition with Run/Stop or Single.
Push Run/Stop to start acquisitions. The oscilloscope acquires repeatedly until you push the button again to stop the acquisition.
Push Single to take a single acquisition. Single sets the trigger mode to Normal
for the single acquisition.
ermine the settings of some
Edge trigger readout
er for edge and the advanced
nd Stopping an Acquisition
MSO2000B and DP O2000B Series Oscilloscopes User Manual 75
Display Wavefor
mData
Display Wavef
orm Data
This section contains concepts and procedures for displaying the acquired waveform.
Adding and Removing a Waveform
1. To add or remove a waveform from
the display, push the corresponding front-panel channel button or the D15-D0 button.
You can use the channel as a trigger source whether or not it is displayed.
Setting the Display Style and Persistence
1. To set the display style, push Acquire.
2. Push Waveform Display.
Average
Off
Record
Length
100k
Delay
On|Off
Set Hor
Position to
0s
Wavefo
Display
rm
iz.
XY Display
Off
Acquisition
Details
76 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Display Wavefor
Waveform
Display
3. Push Persist Time, and turn multipurpose
knob a to have waveform data remain on screen for a us
er-specied amount of time.
4. Push Set to Auto to have the oscilloscope
automatical
ly determine a persistence time
Persist
Time
(a) Auto
Set to Auto
for you.
5. Push Clear P
persistence information.
ersistence to reset the
Clear Per-
sistence
Quick Tips
Variable persistence accumulates sampled waveform points for a specied time interval. Each sampled waveform decays independently according to the time interval. Use variable persistence for displaying infrequently appearing signal anomalies, such as glitches.
Innite persistence continuously accumulates record points until you change one of the acquisition display settings. Use innite persistence for displaying uniqu e signal anomalies, such as glitches.
mData
XY Display Mode
The XY Display mode graphs the data in xed pairs of waveforms against one another. You can use CH1 versus CH2, and REF1 versus REF2. On four-channel models, you can also use CH3 versus CH4.
ng the Graticule Style
Setti
1. To s e t
2. Push Utility Page.
the graticule style, push Utility.
Utility
Pag
e
MSO2000B and DP O2000B Series Oscilloscopes User Manual 77
Display Wavefor
mData
3. Turn multipur
pose knob a and select
Display.
4. Push Graticule from the lower-bezel menu.
5. Select the desired style from the resulting
side-bezel menu. Use the Full
graticule for quick estimates of
waveform parameters. Use the Grid graticule for full-screen
measureme
nts with cursors and automatic
readouts when cross hairs are not needed. Use the Cross Hair graticule for making
quick est
imates of waveforms while leaving more room for automatic readouts and other data.
Use the F
rame graticule with automatic readouts and other screen text when display features are not needed.
Utility Page
Display
Backlight
Intensity
High
Graticule
Full
Screen
Annotation
Quick Tips
n display IRE and mV graticules. To do this, set the trigger type to video and set the vertical scale to
You ca 143 mV/division. (The 143 mV/division selection is available in the coarse vertical scale settings for the channel when you s et the trigger type to video.) The oscilloscope will automatically display the IRE graticule for NTSC signals, and the
ticule for other video signals (PAL, SECAM, and custom).
mV gra
Setting the LCD Backlight
1. Push Utility.
2. Push Utility Page.
Utility Page
78 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Display Wavefor
mData
3. Turn multipur
4. Push Backligh
5. Select the intensity level from the resulting
Setting W
1. Push the front-panel Intensity button.
pose knob a and select
Display.
t Intensity.
side-bezel
menu. Choices are: High,
Medium, and Low.
aveform Intensity
Utility
Page
Display
Backlight
Intensity
High
Medium
Low
Backlight
Intensity
High
Graticule
Full
Screen
Annotation
This will bring up the intensity r eadout on
lay.
the disp
2. Turn multipurpose knob a to select the
desired intensity for the waveform.
3. Turn multipurpose knob b to select the
desired intensity for the graticule.
MSO2000B and DP O2000B Series Oscilloscopes User Manual 79
Display Wavefor
mData
4. Push Intensity readout from the display.
again to clear the i ntensity
Scaling and Positioning a Waveform
Use the horizontal controls to adjust the time base, adjust the trigger point, and to examine waveform details more closely. You can also use the Wave Inspector Pan and Zoom controls to adjust the display of waveforms. (See page 105, Using Wave Inspector to Manage Long Record Length Waveforms.)
Original waveform Scaled horizon tally Positioned horizontally
Use the vertical controls to select waveforms, adjust the waveform vertical position and scale, and set input parameters. Push a channel menu button (1, 2, 3,or4) as many times as needed and the associated menu items to select, add, or remove a waveform.
Original waveform Scaled vertically Positioned vertically
80 MSO2000B and DPO2000B Series Oscilloscopes User Manual
Loading...