DPO2000/B and MSO2000/B Series Declassification and Security Instructionsi
Table of Contents
iiDPO2000/B and MSO2000/B Series Declassification and Security Instructions
Preface
If you have data security concerns, this document helps you to sanitize or remove
memory devices from the DPO2000/B and MSO2000/B Series Digital Phosphor
and Mixed Sig
The DPO2000/B and MSO2000/B series products h ave data storage (memory)
devices and
to clear or sanitize the memory devices, and also tell you how to declassify an
instrument that is not functioning.
Instrument code and calibration settings reside in nonvolatile flash memory.
Instrument setups and reference waveforms may also be stored in flashmemoryor
on USB drives connected to the instrument.
If you have any questions, contact the Tektronix Technical Support Center at
www.tektronix.com/support.
nal Oscilloscopes.
data output devices (USB ports). These instructions tell you how
Products
Related Documents
The following Tektronix products are covered by this document:
DPO20
DPO2004B
DPO2012/B
DPO2014/B
DPO2022B
DPO2024/B
MS
MSO2004B
MSO2012/B
MSO2014/B
MSO2022B
MSO2024/B
The DPO2000 and MSO2000 Series Digital Phosphor and Mixed SignalOscilloscopes Service Manual, part number 071-2331-XX, is available on the
Tektronix Web site at www.tektronix.com/manuals.
02B
O2002B
DPO2000/B and MSO2000/B Series Declassification and Security Instructionsiii
Preface
Terms
The following t
Clear. This removes data on media/memory before reusing it in a secured
area. All reus
information by standard means of ac cess.
Erase. This
Media storage/data export device. Any of several devices that can be used
to store or e
Nonvolatile memory. Data is retained when the instrument is powered off.
Power off. Some instruments have a “Standby” mode, in which power is
still supplied to the instrument. For the purpose of clearing data, putting the
instrume
products, you will need to either press a rear-panel OFF switch or remove the
power source from the instrument.
Remove. This is a physical means to clear the data by removing the memory
device from the instrument. Instructions are available in the product Service
Manual.
Sanitize. This eradicates the data from media/memory so that the data cannot
be recovered by other means or technology. This is typically used when the
device will be moved (temporarily or permanently) from a secured area to a
non-secured area.
ermsmaybeusedinthisdocument:
able memory is cleared to deny access to previously stored
is equivalent to clear.
xport data from the instrument, such as a USB port.
nt in Standby mode does not qualify as powering off. For these
Scrub. This is equivalent to sanitize.
User-modifiable. The user can write to the memory d evice during normal
instrument operation, using the instrument interface or remote control.
Volatile memory. Data is lost when the instrument is powered off.
ivDPO2000/B and MSO2000/B Series Declassification and Security Instructions
Clear and Sanitize Procedures
Memory Device
s
The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options.
FLASH: contains the
instrument firmware,
calibration constants,
reference waveforms,
and instrument setups
Real-time clock, with
battery backup
EEPROM: contains the
front end board serial
number and calibration
information
Microcontroller with
FLASH: contains the
front panel processor
code and power status
for soft power switch
EEPROM: contains
the application module
firmware
N/AN/A
Only by the
processor system
N/AN/A
Only by the
processor system
N/AN/ADPO2000/B and
Data input
methodLocationTo clear or sanitize
N/A
N/A
Acquisition boardTo clear: Not applicable
Acquisition boardNot applicable; no user
Front End BoardNot applicable; no user
Front Panel BoardNot applicable; no user
MSO2000/B application
module
for calibration constants.
Use the TekSecure
featuretoerase
reference waveforms
and instrument setups.
Use the procedure to
disable the LAN Ethernet
and to clear information,
such as IP addresses.
(See page 5, Disable the
LAN Port and Clear LAN
Ethernet Settings.)
To sanitize: Remove
the acquisition board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Return the product
to Tektronix and a
new acquisition board
will be installed and
the instrument will be
repaired and adjusted as
necessary.
data
data
data
Not applicable; no user
data
2DPO2000/B and MSO2000/B Series Declassification and Security Instructions
Data Export Devices
The following table lists the data export devices.
Table 3: Data Export Devices
port
mum size
FunctionUser modifiab
USB flash dri
storage of reference
waveforms, screen
images, and
setups
Supports r
and data transfer to a
PC
ve; user
instrument
emote control
Type an d min i
USB host portSupports the removable
USB device
Data input
le
methodLocationTo disable
Yes
YesRemote
SaveUSB host port on the
control v
USBTMC
ia
Clear and Sanitize Procedures
front of the
USB device
rear of the instrument
instrument
port on the
Files can be deleted
or overwritten on the
instrument
USB flash drive can be
removed and destroyed.
The USB hos
cannot be disabled.
Use the procedure to
disable t
port. (See page 6,
Disable the USB Device
Port.)
or a PC; the
t port
he USB device
DPO2000/B and MSO2000/B Series Declassification and Security Instructions3
Built-In Security Features
Built-In Secu
rity Features
When to Use TekSecure
What TekSecure Does
You can us e t
stored in internal flash memory.
The DPO2000
front panel. Any U SB flash devices can be r emoved, and stored or destr oyed.
The TekSecure function does the following:
Replace
Replaces the current front-panel setup and all stored setups with the default
setup v
Calculates the checksums of all reference waveform memory and setup
memor
erasure
he TekSecure function to erase setup and reference waveform data
/B and MSO2000/B series models have a USB host port on the
s all waveforms in all reference memories with null sample values
alues
y locations to verify successful completion of waveform and setup
To use TekSecure:
To reset the instrument
RAM:
ays a dialog indicating whether the secure erase was successful or
Displ
unsuccessful
NOTE. TekSecure does not erase or change factory calibration constants, user
protected data, or Ethernet settings.
1. Push the front-panel Utility button.
2. Push the Utility Page lower-bezel button and use the Multipurpose a knobto select Config.
3. Push the TekSecure Erase Mem ory lower-bezel button.
ush the OK Erase Setup & Ref Memory side-bezel button. Wait for the
4.P
“TekSecure operation complete” dialog box to display.
5. Push the Menu Off front-panel button to close the dialog box.
1. Power off the instrument for at least 20 seconds.
2. Power on the instrument.
4DPO2000/B and MSO2000/B Series Declassification and Security Instructions
Built-In Security Features
To overwrite protected
user data
This procedure
Send the following commands to the oscilloscope:
:PASSWORD “XYZZY” (or cur ren t password if not “XYZZY”)
:NEWPASS “ANYTHING”
:PASSWORD “ANYTHING”
*PUD #3300xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx xxxx[repeat
for a total of 300 x’s]
overwrites everything that is stored in the protected user data area.
Disable the LAN Port and Clear LAN Ethernet Settings
To disable the LAN port and clear LAN Ethernet settings, such as IP addresses,
follow these steps:
CAUTION. If you need to restore network connectivity at a later date, make sure to
write down all setting values before clearing them.
1. Remove the network cable from the LAN port on the rear of the instrument.
2. Push the front-panel Utility button.
3. Push the Utility Page lower-bezel button and use the Multipurpose a knobto select I/O.
4. Push the Ethernet Network Settings lower-bezel button.
5. Push the Change Instrument Settings side-bezel button.
6. Pus
7. Pu
8. Push the ↓ arrow side-bezel button to position the cursor on the Instrument
9. Push the Clear lower-bezel button.
10. Push the ↓ arrow side-bezel button to position the cursor on the Domain
11. Push the Clear lower-bezel button.
12. Push the ↓ arrow side-bezel button to position the cursor on the DNS IP
hthe↑ or ↓ arrow side-bezel button to position the cursor on the User
Password information.
sh the Clear lower-bezel button.
P Address information.
I
Name information.
Address information.
13. Push the Clear lower-bezel button.
DPO2000/B and MSO2000/B Series Declassification and Security Instructions5
Built-In Security Features
Disable t
14. Push the ↓ arrow
Address information.
15. Push the Clear
16. Push the ↓ arrow side-bezel button to position the cursor on the Subnet Mask
informatio
17. Push the Clear lower-bezel button.
18. Push the OK Accept side-bezel button.
The LAN system is disabled and no longer allows data traffic in or out. The
relevant LAN Ethernet settings are also cleared.
he USB Device Port
To disable the USB device port, follow these steps:
1. Remove any USB cable or device from the USB device port on the rear of the
instrument.
2. Push the front-panel Utility button.
3. Push the Utility Page lower-bezel button and use the Multipurpose a knob
to select I/O.
side-bezel button to position the cursor on the Gateway IP
lower-bezel button.
n.
4. Push the USB lower-bezel button.
5. Push the Disabled (Off Bus) side-bezel button to disable the USB device port.
USB device port is disabled and no longer allows data trafficinorout.
The
6DPO2000/B and MSO2000/B Series Declassification and Security Instructions
Clear and Sanitize a Non-Functional Instrum ent
Clear and Sani
Acquisition board
USB Flash Drive
tize a Non-Functional Instrument
If your instr
as follows:
Only qualified personnel should perform service procedures. Read the Service
Safety Sum
Series Digital Phosphor and Mixed Signal Oscilloscopes Service Manual before
performing any service procedures.
Remove the Acquisition board and return the product to Tektronix. A new
Acquisition board will be installed, and the instrument will be repaired and
adjusted as necessary.
For removal instructions, refer to the service manual, p art number 071-2331-XX,
available on the Tektronix Web site at www.tektronix.com/manuals.
After removal of the Acquisition board, refer to your company's internal policies
regarding han dling or disposal of the board.
Remove the USB flash drive and return the instrument to Tektronix for repair.
rremovaloftheUSBfla sh drive, refer to you r company's internal policies
Afte
regarding handling or disposal of the fl ash drive.
ument is not functioning and you need to clear or sanitize it, proceed
mary and the General Safety Summary in the DPO2000 and MSO2000
Charges
Replacement of any missing hardware will be charged according to the rate at
time of replacement.