Tektronix MDO4000C Series User Manual

MDO4000C Series
Preliminary
Mixed Domain Oscilloscope
User Manual
*P077116700*
077-1167-00
Preliminary
MDO4000C Series
Preliminary
Mixed Domain Oscilloscope
User Manual
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Supports Firmware V1.02 and above
www.tek.com
077-1167-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
Preliminary
or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to find contacts in your area.

Table of Contents

Preliminary
Important safety information ............................................................................................................... v
General safety summary ................................................................................................................. v
Service safety summary ................................................................................................................. ix
Terms in the manual ...................................................................................................................... ix
Terms on the product ...................................................................................................................... x
Symbols on the product .................................................................................................................. x
Compliance Information ..................................................................................................................... xi
EMC compliance ........................................................................................................................... xi
Safety compliance ........................................................................................................................ xii
Environmental compliance ........................................................................................................... xv
Preface ............................................................................................................................................. xvii
Key features ................................................................................................................................ xvii
Conventions used in this manual ............................................................................................... xviii
Warranty .................................................................................................................................... xviii
Installation
Before installation ........................................................................................................................... 1
Operating positions ......................................................................................................................... 8
Connecting probes .......................................................................................................................... 9
Securing the oscilloscope ............................................................................................................. 11
Power-On ...................................................................................................................................... 11
Powering off the oscilloscope
Functional check ........................................................................................................................... 13
Compensating a TPP0500B or TPP1000 passive voltage probe .................................................. 15
Compensating a non-TPP0500B or non-TPP1000 passive voltage probe ................................... 17
Application module free trial ........................................................................................................ 18
Installing an application module ................................................................................................... 18
Upgrading bandwidth ................................................................................................................... 19
Changing the language of the user interface or keyboard ............................................................ 21
Changing the date and time .......................................................................................................... 23
Signal path compensation for time and frequency domains ......................................................... 24
Upgrading firmware ..................................................................................................................... 26
Connecting your oscilloscope to a computer ................................................................................ 29
Using VISA ............................................................................................................................. 29
Using e*Scope ......................................................................................................................... 29
Using the LXI web page and e*Scope .................................................................................... 31
....................................................................................................... 12
MDO4000C Series Oscilloscope User Manual i
Table of Contents
Preliminary
Using a socket server ............................................................................................................... 31
Connecting a USB keyboard to your oscilloscope ....................................................................... 34
Get acquainted with the instrument
Front-panel menus, controls, and connectors ............................................................................... 35
Rear-Panel Connectors ................................................................................................................. 36
Acquire the signal
Using the default setup ................................................................................................................. 39
Using autoset ................................................................................................................................ 40
Using FastAcq .............................................................................................................................. 41
How the analog acquisition modes work ...................................................................................... 43
Changing the Acquisition Mode, Record Length, and Delay Time ............................................. 44
Using Roll Mode .......................................................................................................................... 46
Act on Event ................................................................................................................................. 46
Setting up digital channels ............................................................................................................ 48
When and why to turn on MagniVu ............................................................................................. 50
Using MagniVu ............................................................................................................................ 51
Trigger setup
Choosing a trigger type ................................................................................................................. 53
Selecting triggers .......................................................................................................................... 54
Triggering on buses ...................................................................................................................... 57
Parallel bus trigger ........................................................................................................................ 58
I2C bus trigger .............................................................................................................................. 58
SPI bus trigger .............................................................................................................................. 59
RS-232 bus trigger ........................................................................................................................ 59
CAN bus trigger
LIN bus trigger ............................................................................................................................. 60
FlexRay bus trigger ...................................................................................................................... 61
Audio bus trigger .......................................................................................................................... 61
USB bus trigger ............................................................................................................................ 61
Ethernet bus trigger ...................................................................................................................... 61
MIL-STD-1553 bus trigger .......................................................................................................... 62
Data value matching ..................................................................................................................... 62
Parallel bus trigger data matching ................................................................................................ 62
Checking trigger settings .............................................................................................................. 63
Starting and stopping an acquisition ............................................................................................. 64
............................................................................................................................ 60
ii MDO4000C Series Oscilloscope User Manual
Table of Contents
Preliminary
Display waveform or trace data
Adding and removing a waveform ............................................................................................... 65
Setting waveform intensity ........................................................................................................... 66
Scaling and positioning a waveform ............................................................................................ 67
Setting input parameters ............................................................................................................... 69
Positioning, scaling, and grouping digital channels ..................................................................... 72
Viewing digital channels .............................................................................................................. 73
Annotating the screen ................................................................................................................... 75
Viewing the trigger frequency ...................................................................................................... 76
Analyze waveform or trace data
Taking automatic measurements in the time domain ................................................................... 77
Selecting automatic measurements in the time domain ................................................................ 78
Taking automatic measurements in the frequency domain .......................................................... 82
Taking manual measurements with cursors
Using math waveforms ................................................................................................................. 87
Using FFT ..................................................................................................................................... 89
Using advanced math .................................................................................................................... 92
Using spectrum math .................................................................................................................... 93
Using reference waveforms and traces ......................................................................................... 94
Using wave inspector to manage long record length waveforms ................................................. 96
Panning a waveform ................................................................................................................ 96
Playing and pausing a waveform ............................................................................................. 97
Searching and marking waveforms ......................................................................................... 98
Auto-magnify .............................................................................................................................. 100
Taking automated power measurements .................................................................................... 101
.................................................................................. 83
Save and recall information
External file structure ................................................................................................................. 103
Naming your file ......................................................................................................................... 104
Editing file directory reference waveform or instrument setup names ...................................... 105
Saving a screen image ................................................................................................................ 107
Saving and recalling setups ........................................................................................................ 109
Saving with one button push ...................................................................................................... 111
Managing drives directories and files ......................................................................................... 113
Mounting a network drive .......................................................................................................... 114
Erasing oscilloscope memory ..................................................................................................... 116
MDO4000C Series Oscilloscope User Manual iii
Table of Contents
Preliminary
Use the Arbitrary Function Generator
How to access the AFG .............................................................................................................. 120
Using application modules
Appendix A: MDO4000C Series specifications
.................................................................................................................................................... 125
Appendix B: TPP0500B and TPP1000 500 MHz and 1 GHz 10X
passive probes information
Operating information ................................................................................................................ 127
Connecting the probe to the oscilloscope ................................................................................... 127
Compensating the probe with the MDO4000C series oscilloscope ........................................... 128
Standard accessories ................................................................................................................... 128
Optional accessories ................................................................................................................... 129
Replacing the probe tip ............................................................................................................... 130
Specifications .............................................................................................................................. 130
Performance graphs ............................................................................................................... 131
Safety summary .......................................................................................................................... 132
iv MDO4000C Series Oscilloscope User Manual

Important safety information

Preliminary
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, see the Service safety summary that follows the General safety summary.

General safety summary

Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
Comply with local and national safety codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
MDO4000C Series Oscilloscope User Manual v
Important safety information
Preliminary
To avoid fire or personal
injury
Use proper power cord. Use only the power cord specified for this product and
certified for the country of use.
Use proper power cord. Use only the power cord specified for this product and certified for the country of use. Do not use the provided power cord for other products.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded. Do not disable the power cord grounding connection.
Power disconnect. The power cord disconnects the product from the power source. See instructions for the location. Do not position the equipment so that it is difficult to operate the power cord; it must remain accessible to the user at all times to allow for quick disconnection if needed.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source. Use only insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be suitable for the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and markings on the product. Consult the product manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product, probe, or accessory. Use caution when using 1:1 test leads because the probe tip voltage is directly transmitted to the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not float the common terminal above the rated voltage for that terminal.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualified service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
vi MDO4000C Series Oscilloscope User Manual
Important safety information
Preliminary
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Use proper fuse. Use only the fuse type and rating specified for this product.
Wear eye protection. Wear eye protection if exposure to high-intensity rays or
laser radiation exists.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
Provide proper ventilation. Refer to the manual's installation instructions for details on installing the product so it has proper ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Use care when lifting and carrying the product. This product is provided with handles for lifting and carrying.
Use only the Tektronix rackmount hardware specified for this product.
MDO4000C Series Oscilloscope User Manual vii
Important safety information
Preliminary
Probes and test leads
Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
Keep fingers behind the finger guards on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads, and adapters for any measurement.
Beware of high voltages. Understand the voltage ratings for the probe you are using and do not exceed those ratings. Two ratings are important to know and understand:
The maximum measurement voltage from the probe tip to the probe reference lead.
The maximum floating voltage from the probe reference lead to earth ground.
These two voltage ratings depend on the probe and your application. Refer to the Specifications section of the manual for more information.
WARNING. To prevent electrical shock, do not exceed the maximum measurement or maximum floating voltage for the oscilloscope input BNC connector, probe tip, or probe reference lead.
Connect and disconnect properly. Connect the probe output to the measurement product before connecting the probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement product.
Connect and disconnect properly. De-energize the circuit under test before connecting or disconnecting the current probe.
Connect the probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages above the current probe voltage rating.
Inspect the probe and accessories. Before each use, inspect probe and accessories for damage (cuts, tears, or defects in the probe body, accessories, or cable jacket). Do not use if damaged.
Ground-referenced oscilloscope use. Do not float the reference lead of this probe when using with ground-referenced oscilloscopes. The reference lead must be connected to earth potential (0 V).
viii MDO4000C Series Oscilloscope User Manual
Important safety information
Preliminary

Service safety summary

The Service safety summary section contains additional information required to safely perform service on the product. Only qualified personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains power before removing any covers or panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.

Terms in the manual

Verify safety after repair. Always recheck ground continuity and mains dielectric
strength after performing a repair.
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
MDO4000C Series Oscilloscope User Manual ix
Important safety information
Preliminary

Terms on the product

These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.

Symbols on the product

When this symbol is marked on the product, be sure to consult the manual to find out the nature of the potential hazards and any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.)
The following symbols may appear on the product:
x MDO4000C Series Oscilloscope User Manual

Compliance Information

Preliminary
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument complies.

EMC compliance

EC Declaration of
Conformity – EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities:
EN 61326-1, EN 61326-2-1. EMC requirements for electrical equipment for measurement, control, and laboratory use.
CISPR 11. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2. Electrostatic discharge immunity
IEC 61000-4-3. RF electromagnetic field immunity
IEC 61000-4-4. Electrical fast transient / burst immunity
IEC 61000-4-5. Power line surge immunity
IEC 61000-4-6. Conducted RF immunity
IEC 61000-4-8. Power frequency magnetic field immunity test
IEC 61000-4-11. Voltage dips and interruptions immunity
EN 61000-3-2. AC power line harmonic emissions
EN 61000-3-3. Voltage changes, fluctuations, and flicker
1 2 3
4
5
6
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
Equipment may not meet the immunity requirements of applicable listed standards when test leads and/or test probes are connected due to coupling of electromagnetic interference onto those leads/probes. To minimize the influence of electromagnetic interference, minimize the loop area between the unshielded portions of signal and associated return leads, and keep leads as far away as possible from electromagnetic disturbance sources. Twisting unshielded test leads together is an effective way to reduce loop area. For probes, keep the ground return lead as short as possible and close to the probe body. Some probes have accessory probe tip adapters to accomplish this most effectively. In all cases, observe all safety instructions for the probes or leads used.
4
For compliance with the EMC standards listed here, high quality shielded interface cables should be used.
5
≤ 4.0 division waveform displacement or ≤ 8.0 division increase in peak-to-peak noise.
Residual spurious signals in the RF section can typically increase to -65 dBm when the instrument is subjected to electromagnetic interference per the IEC 61000-4-3 test for frequencies up to 1 GHz, and to -45 dBm for frequencies above 1GHz.
6
≤ 4.0 division waveform displacement or
Residual spurious signals in the RF section can typically increase to -80 dBm when the instrument is subjected to electromagnetic interference per the IEC 61000-4-6 test.
MDO4000C Series Oscilloscope User Manual xi
≤ 8.0 division increase in peak-to-peak noise.
Compliance Information
Preliminary
European contact.
Tektronix UK, Ltd. Western Peninsula Western Road Bracknell, RG12 1RF United Kingdom
Australia / New Zealand
Declaration of Conformity
– EMC
Russian federation

Safety compliance

EU declaration of
conformity – low voltage
Complies with the EMC provision of the Radiocommunications Act per the following standard, in accordance with ACMA:
CISPR 11. Radiated and conducted emissions, Group 1, Class A, in accordance with EN 61326-1 and EN 61326-2-1.
Australia / New Zealand contact.
Baker & McKenzie Level 27, AMP Centre 50 Bridge Street Sydney NSW 2000, Australia
This product is approved by the Russian government to carry the GOST mark.
This section lists the safety standards with which the product complies and other safety compliance information.
Compliance was demonstrated to the following specification as listed in the Official Journal of the European Union:
Low Voltage Directive 2006/95/EC.
EN 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
EN 61010-2-030. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and measuring circuits.
xii MDO4000C Series Oscilloscope User Manual
Compliance Information
Preliminary
U.S. nationally recognized
testing laboratory listing
Canadian certification
Additional compliances
UL 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
UL 61010-2-030. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and measuring circuits.
CAN/CSA-C22.2 No. 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
CAN/CSA-C22.2 No. 61010-2-030. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and measuring circuits.
IEC 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
IEC 61010-2-030. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and measuring circuits.
Equipment type
Safety class
Pollution degree
description
Test and measuring equipment.
Class 1 – grounded product.
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical office/home environment. Temporary condensation occurs only when the product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
MDO4000C Series Oscilloscope User Manual xiii
Compliance Information
Preliminary
Pollution degree
IP rating
Measurement and
overvoltage category
descriptions
Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor, dry location use only.
IP20 (as defined in IEC 60529).
Measurement terminals on this product may be rated for measuring mains voltages from one or more of the following categories (see specific ratings marked on the product and in the manual).
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category IV. For measurements performed at the source of low-voltage installation.
NOTE. Only mains power supply circuits have an overvoltage category rating. Only measurement circuits have a measurement category rating. Other circuits within the product do not have either rating.
Mains overvoltage
category rating
Overvoltage Category II (as defined in IEC 61010-1)
xiv MDO4000C Series Oscilloscope User Manual
Compliance Information
Preliminary

Environmental compliance

This section provides information about the environmental impact of the product.
Product end-of-life
handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Tektronix Web site (www.tektronix.com/
productrecycling).
Perchlorate materials. This product contains one or more type CR lithium batteries. According to the state of California, CR lithium batteries are classified as perchlorate materials and require special handling. See www.dtsc.ca.gov/ hazardouswaste/perchlorate for additional information.
MDO4000C Series Oscilloscope User Manual xv
Compliance Information
Preliminary
xvi MDO4000C Series Oscilloscope User Manual

Preface

Preliminary

Key features

This manual provides operating information the following oscilloscopes:
MDO4024C MDO4034C MDO4054C MDO4104C
MDO4000C Mixed Domain Oscilloscopes include up to six built-in instruments, each with exceptional performance to address tough challenges. Every oscilloscope features powerful triggering, search and analysis, and these are the only scopes to offer synchronized analog, digital, and RF signal analysis at the same time. The MDO4000C is completely customizable and fully upgradeable.
Models are available with bandwidths from 2
00 MHz to 1 GHz
Sample rates of either 2.5 or 5 GS/s on all analog channels depending on the model
20 M points record length on all channels
>340,000 waveforms/second maximum waveform capture rate
A dedicated RF input channel for frequency domain measurements (optional)
Time-synchronized acquisition of analog, digital, and RF signals in a single instrument
50 MHz arbitrary/function generator (optional)
16 digital channels and four analog channels for time domain measurements (optional)
Advanced triggering and analysis: I2C, SPI, USB 2.0, CAN, LIN, FlexRay, RS-232, RS-422, RS-485, UART, I2S, Left Justified (LJ), Right Justified (RJ), TDM, Ethernet, MIL-STD-1553 (with the appropriate application module), and Parallel (optional)
Power analysis, and limit and mask testing application modules (optional)
MDO4000C Series Oscilloscope User Manual xvii
Preface
Preliminary

Conventions used in this manual

The following icons are used throughout this manual.

Warranty

Sequence Step Front panel
power
Three-year warranty covering all parts and labor, excluding probes.
Connect power Network USB
xviii MDO4000C Series Oscilloscope User Manual

Installation

Preliminary

Before installation

Unpack the oscilloscope and check that you received all items listed as standard accessories. The following pages list recommended accessories and probes, instrument options, and upgrades. Check the Tektronix Web site (www.tektronix.com) for the most current information.
All products are shipped with a printed Installation and Safety manual that is in English, Japanese, Simplified Chinese, and French. Full user manuals translated in each language listed below are available in PDF format from www.tektronix.com/manuals.
Table 1: Standard accessories
Accessory Description Tektronix part number
MDO4000C Series Oscilloscopes User Manual
Please find these manuals on www.tektronix.com/manuals
Calibration certificate documenting traceability to national metrology institute(s), and ISO9001 quality system registration.
English (Option L0) 077-1167-XX
French (Option L1) 077-1168-XX
Italian (Option L2) 077-1170-XX
German (Option L3) 077-1169-XX
Spanish (Option L4) 077-1171-XX
Japanese (Option L5) 077-1177-XX
Portuguese (Option L6) 077-1172-XX
Simplified Chinese (Option L7) 077-1174-XX
Traditional Chinese (Option L8) 077-1175-XX
Korean (Option L9) 077-1176-XX
Russian (Option L10) 077-1173-XX
– –
MDO4000C Series Oscilloscope User Manual 1
Installation
Preliminary
Accessory Description Tektronix part number
Front Panel Overlay When selecting a language option, you will receive two front-panel overlays in the language selected. Use the overlay that matches your instrument.
Probes For 200, 350 and 500 MHz
Adapter, for models with options SA3 or SA6
Front Cover Hard plastic cover to help
French (Option L1) SA version Non-SA version
Italian (Option L2) SA version Non-SA version
German (Option L3) SA version Non-SA version
Spanish (Option L4) SA version Non-SA version
Japanese (Option L5) SA version Non-SA version
Portuguese (Option L6) SA version Non-SA version
Simplified Chinese (Option L7) SA version Non-SA version
Traditional Chinese (Option L8) SA version Non-SA version
Korean (Option L9) SA version Non-SA version
Russian (Option L10) SA version Non-SA version
models, one 500 MHz, 10X passive probe per channel
For 1 GHz models, one 1 GHz, 10X passive probe per channel
N Male to BNC Female 103-0045-00
protect the instrument
335-3598-XX 335-3608-XX
335-3600-XX 335-3610-XX
335-3601-XX 335-3611-XX
335-3602-XX 335-3612-XX
335-3603-XX 335-3613-XX
335-3604-XX 335-3614-XX
335-3605-XX 335-3615-XX
335-0306-XX 335-3616-XX
335-3607-XX 335-3617-XX
335-3599-XX 335-3609-XX
TPP0500B
TPP1000
200-5130-00
2 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
Accessory Description Tektronix part number
Power Cord North America (Option A0) 161-0348-00
Universal Euro (Option A1) 161-0343-00
United Kingdom (Option A2) 161-0344-00
Australia (Option A3) 161-0346-00
Switzerland (Option A5) 161-0347-00
Japan (Option A6) 161-0342-00
China (Option A10) 161-0341-00
India (Option A11) 161-0349-00
Brazil (Option A12) 161-0356-00
No power cord or AC adapter (Option A99)
Logic probe, with Option MDO4MSO
Probe and accessory pouch Bag to hold probes and related
One, 16-channel logic probe, with accessories
accessories
– –
P6616
016-2030-XX
Table 2: Optional accessories
Accessory Description Tektronix part number
Aerospace serial triggering and analysis application module
Audio serial triggering and analysis application module
Automotive serial triggering and analysis application module
This module enables triggering on MIL-STD-1553 serial buses. Also, it provides digital views of the signal, bus views, bus decoding, search tools, and decode tables with time stamp information.
This module enables triggering on I2S, Left Justified (LJ), Right Justified (RJ), and TDM audio buses. Also, it provides digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with time stamp information
This module enables triggering on packet level information on CAN and LIN serial buses. Also, it provides a digital view of the signal, bus view, packet decoding, search tools, and a packet decode table with time stamp information.
DPO4AERO
DPO4AUDIO
DPO4AUTO
MDO4000C Series Oscilloscope User Manual 3
Installation
Preliminary
Accessory Description Tektronix part number
FlexRay, CAN, and LIN serial triggering and analysis application module
Bundle application module This module enables
Computer triggering and analysis application module
Embedded serial triggering and analysis application module
Ethernet serial triggering and analysis application module
This module enables triggering on packet level information in FlexRay, CAN, and LIN buses. Also, it provides digital views of the signal, bus views, packet decoding, search tools, packet decode tables with time stamp information and eye diagram analysis software.
functionality for DPO4AERO, DPO4AUDIO, DPO4AUTO, DPO4COMP, DPO4EMBD, DPO4ENET, DPO4LMT, DPO4PWR, DPO4USB, and DPO4VID
This module enables triggering on RS-232, RS-422, RS-485 and UART serial buses. Also, it provides digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with time stamp information.
This module enables triggering on packet level information on I2C and SPI serial buses. Also, it provides digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with time stamp information.
This module enables triggering on 10BASE-T and 100BASE­TX buses. Also, it provides search tools, bus views, and decode tables with time stamp information.
DPO4AUTOMAX
DPO4BND
DPO4COMP
DPO4EMBD
DPO4ENET
NOTE. ≥350 MHz bandwidth models are recommended for 100BASE-TX.
4 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
Accessory Description Tektronix part number
Limit and mask test application module
This module supports limit testing and testing on telecom standard masks or custom masks.
NOTE. ≥350 MHz bandwidth models are recommended for Telecomm standards >55 Mb/s. 1 GHz bandwidth models are recommended for high-speed (HS) USB.
DPO4LMT
Power analysis application module
Universal Serial Bus triggering and analysis application module
Extended video application module
Advanced RF triggering application module (for instruments with options SA3 or SA6)
NEX-HD2HEADER Adapter that routes the
TEK-USB-488 Adapter GPIB to USB Adapter TEK-USB-488
Rackmount kit Adds rackmount brackets RMD5000
Soft transit case Case for carrying instrument ACD4000B
This module supports measurements of power quality, switching loss, harmonics, ripple, modulation, safe operating area, and slew rate (dV/dt and dI/dt).
This module enables triggering on packet level information on USB 2.0 serial buses. Also, it provides digital views of the signal, bus views, bus decoding data in hex, binary, and ASCII, search tools, and packet decode tables with time stamp information.
NOTE. 1 GHz bandwidth models are required for high­speed (HS) USB.
This module enables triggering on a variety of standard HDTV signals, as well as on custom (non-standard) bilevel and trilevel video signals with 3 to 4,000 lines.
This module enables triggering with RF power as the source for Pulse Width, Timeout, Runt, Logic, and Sequence triggers.
channels from a Mictor connector to 0.1 inch header pins.
DPO4PWR
DPO4USB
DPO4VID
MDO4TRIG
NEX-HD2HEADER
MDO4000C Series Oscilloscope User Manual 5
Installation
Preliminary
Accessory Description Tektronix part number
Hard transit case Traveling case, which requires
use of the soft transit case (ACD4000B).
MSO4000B, DPO4000B, MDO4000/B/C, and MDO3000 Series Oscilloscopes Programmer Manual
MDO4000C Series Oscilloscopes Specifications and Performance Verification Technical Reference Manual
MDO4000C Series Oscilloscopes Service Manual
MDO4000C Series Application Module Installation Instructions
DPO3PWR and DPO4PWR Power Measurement Module User Manual
MDO4000C Series Oscilloscopes Declassification and Security Instructions
TekVPI Probes Visit the Oscilloscope Probe
Describes commands for remote control of the oscilloscope. Available for download from www.tektronix.com/manuals.
Describes the oscilloscope specifications and performance verification procedure. Available for download from www.tektronix.com/manuals.
Service information on MDO4000C Series oscilloscopes.
Describes how to install application modules on your oscilloscope.
English (Option L0) 071-2631-XX
French (Option L1) 077-0235-XX
Italian (Option L2) 077-0236-XX
German (Option L3) 077-0237-XX
Spanish (Option L4) 077-0238-XX
Japanese (Option L5) 077-0239-XX
Portuguese (Option L6) 077-0240-XX
Simplified Chinese (Option L7) 077-0241-XX
Traditional Chinese (Option L8) 077-0242-XX
Korean (Option L9) 077-0243-XX
Russian (Option L10) 077-0244-XX
Describes how to sanitize or remove memory devices from the Tektronix MDO4000C Series oscilloscopes.
and Accessory Selector Tool on the Tektronix website at www.tektronix.com.
HCTEK54
077-0510-XX
077-1178-XX
077-1179-XX
071-3253-XX
077-1180-00
NOTE. A subset of TekVPI probes can be used on the RF input as well. These probes require the use of the TPA-N­VPI adapter listed below.
6 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
Accessory Description Tektronix part number
TPA-N-VPI adapter Adapter from N connection (RF
input) to TekVPI probe.
TPA-BNC adapter TekVPI to TekProbe II BNC
Adapter
The MDO4000C Series oscilloscopes support a multitude of optional probes. See
Connecting probes on page 9 . Visit the Oscilloscope Probe and Accessory
Selector Tool on the Tektronix Web site (www.tektronix.com) for the most up to date information.
TPA-N-VPI
TPA-BNC
MDO4000C Series Oscilloscope User Manual 7
Installation
Preliminary

Operating positions

Use the handle and front flip feet to place the oscilloscope in a convenient operating position. When the feet are extended, always have the handle in a down position.
8 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary

Connecting probes

The oscilloscope supports probes with the following:
1. Tektronix Versatile Probe Interface (TekVPI)
These probes support two-way communication with the oscilloscope through on-screen menus and remotely through programmable support. The remote control is useful in applications like ATE where you want the system to preset probe parameters.
2. Tektronix Versatile Probe Interface (TekVPI) for Passive Probes
These probes build upon the functionality of the TekVPI interface. Each probe is matched with its corresponding oscilloscope channel, allowing the oscilloscope to optimize the signal input path. This provides AC compensation across the frequency band.
3. TPA-BNC Adapter
The TPA-BNC Adapter allows you to use TEKPROBE II probe capabilities, such as providing probe power, and passing scaling and unit information to the oscilloscope.
MDO4000C Series Oscilloscope User Manual 9
Installation
Preliminary
4. BNC Interfaces
Some of these use TEKPROBE capabilities to pass the waveform signal and scaling to the oscilloscope. Some only pass the signal and there is no other communication.
5. Logic Probe Interface
The P6616 probe provides 16 channels of digital (on or off state) information.
6. The TPA-N-VPI Adapter allows you to use TekVPI probes in the RF input.
For more information on the many probes available for use with MDO4000C Series oscilloscopes, visit the Oscilloscope Probe and Accessory Selector Tool on the Tektronix website at www.tektronix.com.
10 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary

Securing the oscilloscope

1. Use a standard laptop computer style security lock to secure your oscilloscope to your location.

Power-On

1. Connect the supplied power cord to the rear-panel power connector.
2. Push the power button on the instrument front-panel and the instrument will
turn on.
NOTE. The Standby button on the front-panel does not disconnect mains power. Only the power cord at the rear of the product can disconnect mains power.
MDO4000C Series Oscilloscope User Manual 11
Installation
Preliminary

Powering off the oscilloscope

To power off the oscilloscope, remove the power cord.
12 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary

Functional check

Perform this quick functional check to verify that your oscilloscope is operating correctly.
1. Connect the oscilloscope power cable. See Power-On on page 11.
2. Power on the oscilloscope.
3. Connect the probe connector to oscilloscope channel 1 and the probe tip and
reference lead to the PROBE COMP terminals on the oscilloscope front panel.
4. Push Default Setup.
MDO4000C Series Oscilloscope User Manual 13
Installation
Preliminary
5. Push Autoset. The screen should now display a square wave, approximately
2.5 V at 1 kHz.
If the signal appears but is misshapen, perform the procedures for compensating the probe.
If no signal appears, rerun the procedure. If this does not remedy the situation, have the instrument serviced by qualified service personnel.
14 MDO4000C Series Oscilloscope User Manual
5
6
Installation
Preliminary

Compensating a TPP0500B or TPP1000 passive voltage probe

Your Tektronix oscilloscope can automatically compensate TPP0500B and TPP1000 probes. This eliminates the need for manual probe compensation, as is typically performed with other probes.
Each compensation generates values for a specific probe and channel combination. If you want to use the probe on another channel and desire to compensate the new probe-channel pair, you must run a new set of compensation steps for that new combination.
1. Connect the oscilloscope power cable. See Power-On on page 11.
2. Power on the oscilloscope.
3. Connect the probe connector to the oscilloscope channel, and the probe tip
and reference lead to the PROBE COMP terminals on the oscilloscope front panel.
NOTE. Connect only one probe at a time to the probe comp terminals.
4. Push a front panel button for an input channel connected to the probe you wish to compensate. (1, 2, 3, or 4)
5. Notice on the lower menu that the oscilloscope has automatically set the probe termination value
Coupling
DC| AC
MDO4000C Series Oscilloscope User Manual 15
Termination set
by TPP1000
Invert
On| Off
Bandwidth
Full
Label
More
7
8
Installation
Preliminary
6. Push More repeatedly to select Probe Setup from the resulting pop-up menu.
7. Notice that the compensation status starts as Default.
8. Push Compensate probe and follow the instructions that appear on the
display.
TPP1000
Probe Setup
SN:
000001 Atten:
10X
Compensation
Status
Default
Compensate
probe for 1
Measure
Current Yes| No
When compensating TPP0500B/TPP1000 probes on the oscilloscope:
Each compensation generates values for a specific probe and channel combination. If you want to use the probe on another channel and desire to compensate the new probe-channel pair, you must run a new set of compensation steps.
Each channel can store compensation values for 10 individual probes. If you try to compensate an 11th probe on a channel, the oscilloscope will delete the values for the least recently used probe and add the values for the new probe.
The oscilloscope will assign default compensation values to a TPP0500B or TPP1000 probe connected to the Aux In channel.
NOTE. A factory calibration will delete all stored compensation values
NOTE. A probe compensation failure is most likely due to intermittent connection
of the probe tip or ground connection during the probe compensation operation. If a failure occurs, the oscilloscope will re-use the old probe compensation values, if they existed prior to the failed probe compensation operation.
16 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary

Compensating a non-TPP0500B or non-TPP1000 passive voltage probe

Whenever you attach a passive voltage probe for the first time to any input channel, compensate the probe to match it to the corresponding oscilloscope input channel.
If you are interested in using the automatic probe compensation procedure described above for the TPP0500B and TPP1000 probes See Compensating a
TPP0500B or TPP1000 passive voltage probe on page 15. on a non-TPP0500B/
TPP1000 Tektronix passive probe, check the instruction manual for your probe to see if it qualifies. Otherwise, to properly compensate your passive probe:
1. Follow the steps for the functional check. See Functional check on page
2. Check the shape of the displayed waveform to determine if your probe is
properly compensated.
3. If necessary, adjust your probe. Repeat as needed.
Quick Tips
Use the shortest possible ground lead and signal path to minimize probe-induced ringing and distortion on the measured signal.
13.
Signal with a short ground lead
MDO4000C Series Oscilloscope User Manual 17
Signal with a long ground lead
Installation
Preliminary

Application module free trial

A 30-day free trial is available for all application module licenses not installed in your oscilloscope. The trial period begins when you power on the oscilloscope for the first time.
After 30 days, you must purchase the module if you want to continue using the application. To see the date when your free trial period expires, push Utility on the front panel, push Utility Page on the lower menu, use multipurpose knob a to select Config, push About on the lower menu, and then push Application Modules on the side menu to see the expiration date.

Installing an application module

CAUTION. To avoid damage to the oscilloscope or application module, observe ESD (electrostatic discharge) precautions. See Power-On on page 11.
Turn off the oscilloscope power while removing or adding an application module.
See Powering off the oscilloscope on page 12.
Optional application module packages extend the capability of your oscilloscope.
You can physically install up to four application modules at one time. Application modules go into the two slots with windows in the upper right corner of the front panel. Two additional slots are directly behind the two that you can see. To use these slots, install the module with the label facing away from you.
Some of the modules have licenses which allow you to transfer the license between your application modules and the oscilloscope. You can keep each license in the module, which will allow you to move the module from one instrument to another. Alternatively, you can move the license from the module to the oscilloscope. This approach will allow you to store the module separately from the oscilloscope for safe keeping. This approach will also allow you to use more than four applications on your oscilloscope simultaneously. To transfer a license from a module to your oscilloscope or from your oscilloscope to a module:
1. Turn off the power to the oscilloscope. Insert the application modules into the oscilloscope. Turn on the power.
2. Push Utility on the front panel. If needed, push Utility Page on the lower menu and turn multipurpose knob a to select Config. Push Manage Modules and Options on the lower menu, and then push License Type on the side menu until “Modules” is selected. The licenses contained in the oscilloscope will be listed in the side menu. Push the button next to the appropriate license to transfer. You may transfer up to four licenses at one time.
3. After you turn off the power to the oscilloscope, you can remove the physical application module from the oscilloscope.
18 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
Refer to the MDO3000 and MDO4000 Series Application Module Installation Manual that came with your application module for instructions on installing and
testing an application module.
NOTE. If you transfer a license from a module to an oscilloscope, the module will not work on another oscilloscope until you transfer the license back from the oscilloscope to the module. Consider putting the physical module in an envelope or other storage with a label recording the date, module name, model and serial number of the oscilloscope which holds the license. This will help prevent confusion later if someone finds the module, installs it in some other oscilloscope, and wonders why it does not work.

Upgrading bandwidth

You may increase the bandwidth of an instrument when your project requirements demand higher performance. Do this by purchasing an upgrade.
All upgrades require that you send your instrument to a Tektronix Service Center and require a full calibration.
Model to be upgraded
MDO4024C No 200 MHz 350 MHz MDO4BW2T34
MDO4034C No 350 MHz 500 MHz MDO4BW3T54
MDO4054C No 500 MHz 1 GHz MDOBW5T104
Option SA3 or Option SA6 (spectrum anlayzer)
Bandwidth before upgrade
200 MHz 500 MHz MDO4BW2T54
200 MHz 1 GHz MDO4BW2T104
350 MHz 500 MHz MDO4BW3T54
350 MHz 1 GHz MDO4BW3T104
500 MHz 1 GHz MDO4BW5T104
350 MHz 1 GHz MDO4BW3T104
500 MHz 1 GHz MDO4BW5T104
Bandwidth after upgrade
Order product
MDO4000C Series Oscilloscope User Manual 19
Installation
Preliminary
Model to be upgraded
MDO4024C Yes 200 MHz 350 MHz MDO4BW2T34-
MDO4034C Yes 350 MHz 500 MHz MDO4BW3T54-
MDO4054C Yes 500 MHz 1 GHz MDO4BS5T104-
Option SA3 or Option SA6 (spectrum anlayzer)
Bandwidth before upgrade
200 MHz 500 MHz MDO4BW2T54-
200 MHz 1 GHz MDO4BW2T104-
350 MHz 500 MHz MDO4BW3T54-
350 MHz 1 GHz MDO4BW3T104-
500 MHz 1 GHz MDO4BW5T104-
350 MHz 1 GHz MDO4BW3T104-
500 MHz 1 GHz MDO4BW5T104-
Bandwidth after upgrade
Order product
SA
SA
SA
SA
SA
SA
SA
SA
SA
SA
20 MDO4000C Series Oscilloscope User Manual
2
5
6
Installation
Preliminary

Changing the language of the user interface or keyboard

To change the language of the oscilloscope user interface or keyboard, and to change the front panel button labels through the use of an overlay:
1. Push Utility.
2. Push Utility Page.
Utility Page
3. Turn multipurpose knob a and select Config.
4. Push Language from the resulting lower menu.
5. Push Menus from the resulting side menu and turn multipurpose knob a to
select the desired language for the user interface.
6. Push USB Keyboard from the resulting side menu and turn multipurpose knob a to select the desired language version of keyboard to use.
Language
Menus
(a) English
USB Keyboard
English
MDO4000C Series Oscilloscope User Manual 21
Installation
Preliminary
7. If you choose to use an English user interface, be sure that the plastic front panel overlay is removed.
If you choose a language other than English, place the plastic overlay for the language that you desire over the front panel to display labels in that language.
NOTE. If you order a language option, you will receive two front-panel overlays in the language you ordered: One for models with RF options (options SA3 or SA6) and one with no RF option. Follow the instructions that come with the front-panel overlays.
22 MDO4000C Series Oscilloscope User Manual
2
Installation
Preliminary

Changing the date and time

To set the internal clock with the current date and time:
1. Push Utility.
2. Push Utility Page.
Utility Page
3. Turn multipurpose knob a and select Config.
4. Push Set Date & Time.
5. Push the side menu buttons and turn both multipurpose knobs (a and b) to set
the time and date values.
6. Push OK Set Date & Time.
MDO4000C Series Oscilloscope User Manual 23
3
Installation
Preliminary

Signal path compensation for time and frequency domains

Signal Path Compensation (SPC) corrects for DC inaccuracies caused by temperature variations and/or long-term drift. Run the compensation whenever the ambient temperature has changed by more than 10 °C (18 °F) or once a week if you use vertical settings of 5 mV/division or less. Failure to do so may result in the instrument not meeting warranted performance levels at those volts/div settings.
To compensate the signal path:
1. Warm up the oscilloscope for at least 20 minutes. Remove all input signals (probes and cables) from channel inputs. Input signals with AC components adversely affect SPC.
2. Push Utility.
3. Push Utility Page.
Utility
4. Turn multipurpose knob a and select Calibration.
24 MDO4000C Series Oscilloscope User Manual
4
5
6
7
7
Installation
Preliminary
5. Push Signal Path on the lower menu.
Utility Page
Calibration
6. Push OK Compensate Signal Paths on the resulting side menu.
OK Compensate Signal Paths
The calibration will take approximately 10 minutes to complete.
7. After calibration, verify that the status indicator on the lower menu displays Pass.
If it does not, then recalibrate the instrument or have the instrument serviced by qualified service personnel.
Service personnel use the factory calibration functions to calibrate the internal voltage references of the oscilloscope using external sources. Contact your Tektronix field office or representative for assistance with factory calibration.
Signal Path
Pass
Factory
Pass
Utility Page
Calibration
NOTE. Signal Path Compensation does not include calibration to the probe tip.
Signal Path
Pass
Factory
Pass
MDO4000C Series Oscilloscope User Manual 25
Installation
Preliminary

Upgrading firmware

To upgrade the firmware of the oscilloscope:
1. Open up a Web browser and go to www.tektronix.com/software/downloads. Proceed to the software finder. Download the latest firmware for your oscilloscope on your PC.
Unzip the files and copy the firmware.img file into the root folder of a USB flash drive or USB hard drive.
2. Power off your oscilloscope.
26 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
3. Insert the USB flash or hard drive into the USB port on the front panel of your oscilloscope.
4. Power on the oscilloscope. The instrument automatically recognizes the replacement firmware and installs it.
If the instrument does not install the firmware, rerun the procedure. If the problem continues, try a different model of USB flash or hard drive. Finally, if needed, contact qualified service personnel.
NOTE. Do not power off the oscilloscope or remove the USB drive until the oscilloscope finishes installing the firmware.
MDO4000C Series Oscilloscope User Manual 27
8
Installation
Preliminary
5. Power off the oscilloscope and remove the USB flash or hard drive.
6. Power on the oscilloscope.
7. Push Utility.
8. Push Utility Page.
Utility Page
9. Turn multipurpose knob a and select Config.
10. Push About. The oscilloscope displays the firmware version number.
28 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
11. Confirm that the version number matches that of the new firmware.

Connecting your oscilloscope to a computer

Connect your oscilloscope directly to a computer to let the PC analyze your data, collect screen images, or to control your oscilloscope.
Three ways to connect your oscilloscope to a computer are through the VISA drivers, the e*Scope Web-enabled tools, and a socket server. Use VISA to communicate with your oscilloscope from your computer through a software application, such as Tektronix OpenChoice Desktop®. Use e*Scope to communicate with your oscilloscope through a Web browser, such as Microsoft Internet Explorer. For best results, use a browser that supports html 5.

Using VISA

Using e*Scope

VISA lets you use your MS-Windows computer to acquire data from your oscilloscope for use in an analysis package that runs on your PC, such as Microsoft Excel, National Instruments LabVIEW, Tektronix OpenChoice Desktop software, or a program of your own creation. You can use a common communications connection, such as USB, Ethernet, or GPIB, to connect the computer to the oscilloscope.
For VISA, load the VISA drivers on your computer. Also, load your application, such as OpenChoice Desktop. You will find the drivers and OpenChoice Desktop software on the appropriate CD that comes with your oscilloscope or at the Tektronix software finder Web page (www.tektronix.com).
With e*Scope, you can access and control any Internet-connected MDO3000 Series oscilloscope from a web browser on your computer.
Connect the oscilloscope to your network using the LAN port. The built-in LXI web interface (Core 2011, Version 1.4) provides network configuration information, which you can edit and customize. It also provides remote instrument control through the e*Scope user interface. There you can control instrument settings, save screen images, save instrument data or setups, and much more. Do all this through a password-protectable web-interface.
To set up VISA communication between your oscilloscope and a computer.
1. Load the VISA drivers on your computer. Also, load your application, such as OpenChoice Desktop.
You will find the drivers and OpenChoice Desktop software on the appropriate CD that comes with your oscilloscope or at the Tektronix software finder Web page (www.tektronix.com).
2. Connect the oscilloscope to your computer with the appropriate USB or Ethernet cable.
MDO4000C Series Oscilloscope User Manual 29
Installation
Preliminary
To communicate between the oscilloscope and a GPIB system, connect the oscilloscope to the TEK-USB-488 GPIB-to-USB Adapter with a USB cable. Then connect the adapter to your GPIB system with a GPIB cable. Cycle the power on the oscilloscope.
3. Push Utility.
4. Push Utility page.
5. Turn the Multipurpose a and select I/O.
6. If you are using USB, the system sets itself up automatically for you, if USB
is enabled.
Check USB on the lower menu to be sure that USB is enabled. If it is not enabled, push USB. Then push Connect to Computer on the side menu.
7. To use Ethernet, push Ethernet & LXI on the lower menu.
Use the side menu buttons to adjust your network settings, as needed. For more information, see the e*Scope setup information below.
8. If you want to change socket server parameters, push Socket Server and enter new values through the resulting side menu.
9. If you are using GPIB, push GPIB. Enter the GPIB address on the side menu, by turning the Multipurpose a knob.
This will set the GPIB address on an attached TEK-USB-488 Adapter.
10. Run your application software on your computer.
TIP.
Your oscilloscope shipped with a CD that contains a variety of Windows­based software tools for efficient connectivity between your oscilloscope and your computer. These include tool bars that speed connectivity with Microsoft Excel and Word. There is also a standalone acquisition program called Tektronix OpenChoice Desktop.
The rear-panel USB 2.0 device port is the correct USB port for computer connectivity. Use the rear and front panel USB 2.0 host port to connect your oscilloscope to USB flash drives. Use the USB Device port to connect your oscilloscope to a PC or a PictBridge printer.
30 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
Using the LXI web page
and e*Scope
With e*Scope, you can access any Internet-connected MDO4000C oscilloscope from a web browser on your computer or tablet.
To set up e*Scope communications between your oscilloscope and a Web browser running on a remote computer:
1. Connect the oscilloscope to your computer network with an appropriate Ethernet cable.
2. Push Utility.
3. Push Utility Page
4. Turn Multipurpose knob and select I/O.
5. Push Ethernet and LXI.
6. View the top item on the side menu to determine the condition of the LAN.
An indicator turns green for good status and red if the device detects a fault.
7. Push LAN Settings to display the network parameters configured on your oscilloscope.
8. Push LAN Reset to restore the LAN defaults to your oscilloscope.
9. Push Test Connection to check if your oscilloscope can find an attached
network.

Using a socket server

10. Push more to see another page of side menu items.
11. Push Change Names to change the name of the oscilloscope, the network
domain, or the service name.
12. Push Change Ethernet & LXI Password to change the name of the password.
13. Push Change e*Scope Password to use the LXI password to also protect your
oscilloscope from changes made to LAN settings from a Web browser.
14. Start your browser on your remote computer. In the browser address line, enter the host name, a dot, and the domain name together. Alternatively, just enter the IP address of the instrument. Either way, you should then see the LXI Welcome page on your Web browser on your computer screen.
15. Click “Network Configuration” to view and edit the network configuration settings. If you are using a password and changing your settings, you need to know that the default user name is “lxiuser”.
16. For e*Scope, click the Instrument Control (e*Scope) link on the left side of the LXI Welcome page. You should then see a new tab (or window) open in your browser with e*Scope running.
A socket server provides two-way communication over an Internet Protocol­based computer network. You can use your oscilloscope’s socket server feature to let your oscilloscope communicate with a remote-terminal device or computer.
MDO4000C Series Oscilloscope User Manual 31
Installation
Preliminary
To set up and use a socket server between your oscilloscope and a remote terminal or computer:
1.
2. Push Utility.
3. Push Utility Page
4. Turn Multipurpose a and select I/O.
5. Push Socket Server.
6. On the resulting Socket Server side menu, push the top entry to highlight
Enabled.
7. Choose whether the protocol should be None or Terminal.
A communication session run by a human at a keyboard typically uses a terminal protocol. An automated session might handle its own communications without such protocol from the oscilloscope.
8. If required, change the port number by turning Multipurpose a.
9. If required, press OK to set the new port number.
10. After you have set up the socket server parameters, the computer is now
ready to communicate with the oscilloscope. If you are running an MS Windows PC, you could run its default client Telnet, which has a command interface. One way to do this is by typing “ Telnet ” in the Run window. The Telnet window will open on the PC.
NOTE. On MS Windows 7, you must first enable Telnet in order for Telnet to work.
11. Start a terminal session between your computer and your oscilloscope by typing in an open command with the oscilloscope's LAN address and port #.
You can obtain the LAN address by pushing the Ethernet & LXI lower menu button and the resulting LAN Settings side menu button to view the resulting Ethernet and LXI Settings screen. You can obtain the port # by pushing Socket Server on the lower menu and viewing Current Port on the side menu.
For example, if the oscilloscope IP address was 123.45.67.89 and the post # was the default of 4000, you could open a session by writing into the MS Windows Telnet screen: o 123.45.67.89 4000.
The oscilloscope will send a help screen to the computer when it has finished connecting.
32 MDO4000C Series Oscilloscope User Manual
Installation
Preliminary
12. You can now type in a standard query, such as, *idn?.
The Telnet session window will respond by displaying a character string describing your instrument.
You can type in more queries and view more results using this Telnet session window. You can find the syntax for relevant commands, queries and related status codes in the MDO3000 Series Programmer Manual.
NOTE. Do not use the computer’s backspace key during an MS Windows Telnet session with the oscilloscope.
MDO4000C Series Oscilloscope User Manual 33
Installation
Preliminary

Connecting a USB keyboard to your oscilloscope

You can connect a USB keyboard to a USB Host port on the rear or front panel of the oscilloscope. The oscilloscope will detect the keyboard, even if it is plugged in while the oscilloscope is powered on.
You can use the keyboard to quickly create names or labels. You can bring up the Label button on the lower menu by pushing a Channel or Bus button. Use the arrow keys on the keyboard to move the insertion point, and then type in a name or label. Labeling channels and buses makes the information on the screen easier to identify.
To choose whether to use a keyboard with a United States (US) key layout or one with an alternative layout:
1. Push Utility.
2. Push Utility Page.
3. Turn multipurpose knob a and select Config.
4. Push Language on the lower menu.
5. Push USB Keyboard on the side menu.
6. Turn multipurpose knob a and select the desired keyboard layout style from
the resulting menu.
34 MDO4000C Series Oscilloscope User Manual

Get acquainted with the instrument

Preliminary

Front-panel menus, controls, and connectors

The front panel has buttons and controls for the functions that you use most often. Use the menu buttons to access more specialized functions.
1. Traditional oscilloscope front panel controls
2. 10-digit keypad
3. Application module slots
4. Ground strap connector
5. Ground
6. PROBE COMP
7. Auxiliary Input (only on instruments without option SA3 or SA6)
8. Dedicated spectral analysis controls, available with Option SA3 and SA6
9. Dedicated RF input with N-connector, available with Option SA3 and SA6
10. Analog channel (1, 2, 3, 4) inputs with TekVPI versatile probe interface
MDO4000C Series Oscilloscope User Manual 35
Get acquainted with the instrument
Preliminary
11. Digital channel input
12. Display: shows frequency and/or time domain
13. Arbitrary/function generator (AFG) enable button

Rear-Panel Connectors

1. AFG OUT
function generator.
2. AUX OUT. Use to output Trigger, Signal Out, or AFG Sync Out. Also used to generate a signal on a main trigger pulse, as a 10 MHz reference signal, or to output a signal when other events happen, such as mask-limit test events.
3. REF IN. Use the 10 MHz external reference input to provide an external timebase.
4. LAN. Use the LAN (Ethernet) port (RJ-45 connector) to connect the oscilloscope to a 10/100/1000 Base-T local area network.
5. Video Out. Use the Video Out port (DB-15 female connector) to show the oscilloscope display on an external monitor or projector.
36 MDO4000C Series Oscilloscope User Manual
. Use the AFG OUT port to transmit signals from the arbitrary
Get acquainted with the instrument
Preliminary
6. USB 2.0 Device port. Use the USB 2.0 High Speed Device port to connect a PictBridge compatible printer, or for direct PC control of the oscilloscope using USBTMC protocol.
NOTE. The cable connected from the USB 2.0 Device port to the host computer must meet the USB2.0 specification for high speed operation when connected to a high speed host controller.
7. USB 2.0 Host port. Use the USB 2.0 High Speed Host port to connect a USB memory device or USB keyboard.
8. Power
9. Lock. Use to secure the oscilloscope.
10. VESA mounts. Use to secure the oscilloscope.
input. Attach to an AC power line with integral safety ground.
MDO4000C Series Oscilloscope User Manual 37
Get acquainted with the instrument
Preliminary
38 MDO4000C Series Oscilloscope User Manual

Acquire the signal

2
Preliminary
This section describes concepts of and procedures for setting up the oscilloscope to acquire the signal as you want it to.

Using the default setup

To return the oscilloscope to its default settings:
1. Push Default Setup.
2. If you change your mind, push Undo Default Setup to undo the last default
setup.
Undo Default
Setup
MDO4000C Series Oscilloscope User Manual 39
3
Acquire the signal
Preliminary

Using autoset

Autoset adjusts the instrument (acquisition, horizontal, trigger, and vertical controls) such that it displays four or five waveform cycles for analog channels with the trigger near the midlevel, and ten cycles for digital channels.
Autoset works with both the analog and digital channels.
1. To autoset an analog channel, connect the analog probe, and then select the input channel.
To autoset a digital channel, connect the logic probe and select the input channel. See Setting up digital channels on page 48.
2. Push Autoset to execute an Autoset.
3. If desired, push Undo Autoset to undo the last Autoset.
Undo Autoset
You can also disable the Autoset function. To disable or enable the Autoset function:
1. Push and hold Autoset.
2. Push and hold Menu Off.
40 MDO4000C Series Oscilloscope User Manual
4
4
Acquire the signal
Preliminary
3. Release Menu Off, and then release Autoset.
4. Select the desired setting using the side menu.
Autoset
Enabled
Autoset
Disabled
Quick Tips
To position the waveform appropriately, Autoset may change the vertical position. Autoset always sets vertical offset to 0 V.
If you use Autoset when no channels are displayed, the instrument turns on channel 1 and scales it.
If you use Autoset and the oscilloscope detects a video signal, the oscilloscope automatically sets the trigger type to video and makes other adjustments to display a stable video signal.

Using FastAcq

FastAcq™ provides high-speed waveform capture. It is helpful in finding elusive signal anomalies. Fast acquisition mode reduces the dead time between waveform acquisitions, enabling the capture and display of transient events such as glitches and runt pulses. Fast acquisition mode can also display waveform phenomena at an intensity that reflects their rate of occurrence.
To use FastAcq:
1. Push Acquire on the front panel.
2. Push FastAcq on the lower menu.
3. Toggle FastAcq on the side menu to select On.
4. Push Waveform Palette on the side menu and use the multipurpose knob to
select the desired display palette.
MDO4000C Series Oscilloscope User Manual 41
Acquire the signal
Preliminary
The display palette lets you enhance the visibility of events. This choice uses intensity grading to indicate how often rare transients occur relative to normal signals. The choices are Temperature, Spectral, Normal and Inverted.
Temperature uses color-grading to indicate frequency of occurrence with hot colors like red/yellow indicating frequently occurring events and colder colors like blue/green indicating rarely occurring events.
Spectral uses color-grading to indicate frequency of occurrence with colder colors like blue indicating frequently occurring events and hot colors like red indicating rarely occurring events.
Normal uses the default channel color (like yellow for channel one) along with gray-scale to indicate frequency of occurrence where frequently occurring events are bright.
Inverted uses the default channel color along with gray scale to indicate frequency of occurrence where rarely occurring events are bright.
These color palettes help highlight the events that over time occur more often or, in the case of infrequent anomalies, occur less often.
42 MDO4000C Series Oscilloscope User Manual
Acquire the signal
Preliminary

How the analog acquisition modes work

Sample mode retains the first sampled point from each acquisition interval. Sample is the default mode.
Peak Detect mode uses the highest and lowest of all the samples contained in two consecutive acquisition intervals. This mode only works with real-time, noninterpolated sampling and is useful for catching high frequency glitches.
Hi Res mode calculates the average of all the samples for each acquisition interval. This mode also only works with real-time, noninterpolated sampling. Hi-Res provides a higher-resolution, lower-bandwidth waveform.
Envelope mode finds the highest and lowest record points over all acquisitions. Envelope uses Peak Detect for each individual acquisition.
Average mode calculates the average value for each record point over a user­specified number of acquisitions. Average uses Sample mode for each individual acquisition. Use average mode to reduce random noise.
MDO4000C Series Oscilloscope User Manual 43
3
3
3
3
Acquire the signal
Preliminary

Changing the Acquisition Mode, Record Length, and Delay Time

Use this procedure to change the acquisition mode.
1. Push Acquire.
2. Push Mode.
3. Then choose the acquisition mode from the side menu. You can chose from:
Sample, Peak Detect, Hi Res, Envelope, or Average.
Acquisition
Mode
Sample
Peak Detect
Hi Res
Envelope
Average
16
4. If you chose Average, turn multipurpose knob a to set the number of waveforms to average over.
5. Push Record Length.
6. Scroll through the available choices. Choose among 1000, 10 k, 100 k, 1 M,
5M, 10 M, and 20 M points.
44 MDO4000C Series Oscilloscope User Manual
Acquire the signal
Preliminary
7. Push Delay on the lower menu to select On when you want to delay the acquisition relative to the trigger event.
With Delay set to On, turn the Horizontal Position knob counterclockwise to increase the delay. The trigger point will move to the left and ultimately outside of the acquired waveform. Then you can adjust the Horizontal Scale knob to acquire more detail around the area of interest at the center of the screen.
When this delay is on, the trigger point separates from the horizontal expansion point. The horizontal expansion point stays at the center of the screen. The trigger point can move off the screen. When this happens, the trigger marker turns to point in the direction of the trigger point.
Use the delay feature when you want to acquire waveform detail that is separated from the trigger event by a significant interval of time. For example, you can trigger on a sync pulse that occurs once every 10 ms and then look at high-speed signal characteristics that occur 6 ms after the sync pulse.
When the delay feature is set to Off, the expansion point is tied to the trigger point so that scale changes are centered around the trigger point.
MDO4000C Series Oscilloscope User Manual 45
Acquire the signal
Preliminary

Using Roll Mode

Roll mode gives a display similar to a strip chart recorder for low-frequency signals. Roll mode lets you see acquired data points without waiting for the acquisition of a complete waveform record.
Roll mode is enabled when the trigger mode is auto and the horizontal scale is set to 40 ms/div or slower.
Quick Tips
Switching to Envelope or Average acquisition mode, using digital channels, using math waveforms, turning on a bus, or switching to Normal trigger will disable Roll mode.
Roll mode is disabled when you set the horizontal scale to 20 ms per division or faster.
Push Run/Stop to halt Roll mode.

Act on Event

Have the oscilloscope perform a defined action after a defined event has occurred. The event can be a trigger or a certain number of acquisitions. The action can be to:
Stop acquisitions
Save a waveform or a screen image to a file
Print
Send a pulse out the AUX OUT port
Generate a remote interface SRQ
Send an e-mail notification
Display a message on the oscilloscope display
1. Push Test on the front panel.
2. Push Application on the lower menu.
3. Turn the multipurpose knob to select Act on Event.
4. Push Event on the lower menu, which will display the Event Type side
menu. Select the desired event type.
46 MDO4000C Series Oscilloscope User Manual
Acquire the signal
Preliminary
5. Push Actions on the lower menu, which will display the Actions side menu. Use this menu to select a action to occur upon the event just selected.
6. Select the action type from the pop-up menu.
7. Select whether or not to enable the action.
8. If you selected Email Notification from the list of actions, you can now select Configure E-mail from the side menu to define the parameters of the email.
NOTE. There is a single set of SMTP server settings saved for both Act on Event e-mail notification and for the e-mail printer (set through the Utility > Print Setup > Select Printer > Add E-mail Printer menus). If you modify the SMTP settings in one of these two locations, they will be similarly modified in the other location.
9. Push Repeat from the lower menu to set how many times you want this event and action pair to repeat.
MDO4000C Series Oscilloscope User Manual 47
Acquire the signal
Preliminary

Setting up digital channels

Use front panel buttons and knobs to set up your instrument to acquire signals using the digital channels.
1. Connect the P6616 16-channel logic probe to the input signal source.
2. Connect the ground lead or leads to the circuit ground.
You can connect a separate lead for each channel or a common ground lead for each group of 8 wires.
3. If needed, connect the appropriate grabber for each probe to the probe tip.
4. Connect each probe to the desired circuit test point.
5. Push D15 - D0 on the front panel to display the menu.
48 MDO4000C Series Oscilloscope User Manual
6891011
12
Acquire the signal
Preliminary
6. Push D15 - D0 on the lower menu to access the D15 - D0 On or Off menu.
D15 – D0
On/Off
7. Turn multipurpose knob a to scroll through the list of digital channels. Turn multipurpose knob b to position the selected channel.
As you position channels close to each other on the display, the oscilloscope groups the channels, and adds the group to the pop-up list. You can select the group from the list to move all the channels in the group instead of individual channels.
8. Push Thresholds on the lower menu. You can assign a different threshold value to each channel.
9. Push Edit Labels on the lower menu and create the label. You can create labels through the front panel or with an optional USB keyboard.
10. Push Monitor to show activity on digital channels at a glance.
Thresholds Edit Labels Monitor On | Off MagniVu
On |
Off
Height
S | M L
11. Push MagniVu on the lower menu to increase the timing resolution.
12. Push Height on the lower menu repeatedly to set the signal height. You only
need to do this once to set the height for all of the digital channels.
Quick Tips
Use the zoom feature to see multiple cycles of the signal in the upper part, and a single cycle in the lower part of the display. See Using wave inspector
to manage long record length waveforms on page 96.
When setting up the logic probe, the first set of eight leads (pins 7 to 0) on the logic probe are marked GROUP 1 on the lead box. The second set (pins 15 to 8) is marked GROUP 2.
The lead for the first channel in each group is colored blue for easy identification while you connect the logic probe to the device under test. The other leads are gray.
Digital channels store a high or low state for each sample. The threshold that separates high from low can be set separately for each channel.
MDO4000C Series Oscilloscope User Manual 49
Acquire the signal
Preliminary

When and why to turn on MagniVu

Tektronix MagniVu acquisition technology allows you to have higher timing resolution so that you can more accurately determine edge placement and make more precise timing measurements on digital edges. Using MagniVu, you can see up to 32 times more detail than you can using normal digital channel sampling.
The MagniVu record is acquired in parallel to the main digital acquisition and is available at any time, running or stopped. MagniVu provides an ultra high resolution view of your data sampled at a maximum resolution of 60.6 ps for 10,000 points centered around the trigger.
NOTE. MagniVu centers itself around the trigger point. If you turn MagniVu on while using a large record length and you are looking somewhere other than the trigger point, then the digital signal may be off screen. In most of these cases, you can find the digital record by looking for the digital signal in the upper overview and panning accordingly.
NOTE. You should turn on MagniVu when light gray shading is displayed to indicate the uncertainty of the edge position. If the shading is not displayed, you do not need to use MagniVu. See Viewing digital channels on page 73.
50 MDO4000C Series Oscilloscope User Manual
Acquire the signal
Preliminary

Using MagniVu

1. Push D15 – D0.
2. Push MagniVu and select On.
Quick Tips
If you think you need more timing resolution, turn on MagniVu to increase the resolution.
MagniVu is always acquired. If the oscilloscope is in a stopped state, you can turn on MagniVu and still get the resolution without taking another acquisition.
The serial bus features do not use data acquired in MagniVu mode.
MDO4000C Series Oscilloscope User Manual 51
Acquire the signal
Preliminary
52 MDO4000C Series Oscilloscope User Manual

Trigger setup

24444
Preliminary

Choosing a trigger type

This section contains concepts and procedures for setting up the oscilloscope to trigger on your signal.
To select a trigger:
1. Push Trigger Menu.
2. Push Type to bring up the Trigger Type side menu.
NOTE. The bus trigger in the MDO4000C Series works on parallel buses even without an application module. Using the bus trigger on other buses requires use of a DPO4AERO, DPO4AUDIO, DPO4AUTO, DPO4AUTOMAX, DPO4COMP, DPO4EMBD, DPO4ENET, or DPO4USB application module.
3. Turn multipurpose knob a to select the desired trigger type.
4. Complete the trigger setup using the lower menu controls displayed for the
trigger type. The controls to set up the trigger vary depending on the trigger type.
Type
Edge
Source
1
Coupling
DC
Slope
Level
100 mV
Mode
Auto
& Holdoff
MDO4000C Series Oscilloscope User Manual 53
Trigger setup
Preliminary

Selecting triggers

Trigger Type Trigger Conditions
Edge Trigger on a rising edge, a falling edge, or both
edges as defined by the slope control. Coupling choices are DC, LF Reject, HF Reject, and Noise Reject. Edge triggers are the simplest and most commonly used trigger type, with both analog and digital signals. An edge trigger event occurs when the trigger source passes through a specified voltage level in the specified direction.
Sequence (B Trigger)
Combine an edge A Event (Main) trigger with the B Event (Delayed) trigger to capture more complex signals. Time. After the A Event occurs, the trigger system waits the specified amount of time, and then looks for the B Event before triggering and displaying the waveform.
Events. After the A Event occurs, the trigger system looks for a specified number of B Events before triggering and displaying the waveform.
Pulse Width Trigger on pulses that are less than, greater
than, equal to, or not equal to a specified time. Additionally, you can trigger when a pulse width is within or outside a range of two different specified times. You can trigger on positive or negative pulses. Pulse width triggers are primarily used on digital signals.
Timeout Trigger when no pulse is detected within a
specified time. The signal stays above or below (or either above or below) a set value for a set amount of time.
Runt Trigger on a pulse amplitude that crosses one
threshold but fails to cross a second threshold before recrossing the first. You can detect positive or negative (or either) runts, or only those wider than, less than, greater than, equal to, or not equal to a specified width. Runt triggers are primarily used on digital signals.
54 MDO4000C Series Oscilloscope User Manual
Trigger setup
Preliminary
Trigger Type Trigger Conditions
Logic Trigger when all channels transition to the
specified state. Use multipurpose knob a to select a channel. Push the appropriate button on the side menu to set that channel's state to High (H), Low (L), or Don't Care (X). Push Clock on the side menu to enable clocked (state) triggering. You can have at most a single clock channel. Push Clock Edge on the lower menu to change the polarity of the clock edge. Turn off clocked triggering and return to unclocked (pattern) triggering by selecting the clock channel and setting it to high, low, or don't care. For unclocked triggering, by default, triggering occurs when the selected condition goes true. You can also select triggering when the condition goes false, or time-qualified triggering. You can use up to 21 channels for a Logic trigger (4 analog, 16 digital, and 1 RF).
NOTE. To use the RF input in a Logic trigger, you must first install the MDO4TRIG application module.
NOTE. Optimum Logic trigger performance is achieved by using only analog channels or only digital channels.
Setup and Hold Trigger when a logic data input changes state
inside of the setup or hold time relative to a clock edge. Setup is the amount of time that data should be stable and not change before a clock edge occurs. Hold is the time that data should be stable and not change after a clock edge occurs. MDO4000C Series oscilloscopes are capable of multiple channel Setup and Hold triggering, and can monitor the state of an entire bus for setup and hold violations. You can use up to 20 channels for a Setup and Hold trigger (4 analog and 16 digital). Push Clock on the side menu to select the clock channel. Push the Select control, Data, and Not used buttons to select one or more channels you want to monitor for setup and hold violations.
NOTE. Optimum Setup and Hold trigger performance is achieved by using only analog channels or only digital channels.
MDO4000C Series Oscilloscope User Manual 55
Trigger setup
Preliminary
Trigger Type Trigger Conditions
Rise/Fall Time Trigger on rise and fall times. Trigger on pulse
edges that traverse between two thresholds at faster or slower rates than the specified time. Specify pulse edges as positive or negative or either.
Video Trigger on specified fields or lines of a
composite video signal. Only composite signal formats are supported. Trigger on NTSC, PAL, or SECAM. Works with Macrovision signals. With the DPO4VID module, trigger on a variety of HDTV video standard signals, as well as custom (non-standard) bilevel and trilevel video signals with 3 to 4,000 lines.
Bus Trigger on various bus conditions.
I2C requires a DPO4EMBD module. SPI requires a DPO4EMBD module. CAN requires a DPO4AUTO or DPO4AUTOMAX module. RS-232, RS-422, RS-485, and UART require a DPO4COMP module. LIN requires either a DPO4AUTO or a DPO4AUTOMAX module. FlexRay requires a DPO4AUTOMAX module. Audio requires a DPO4AUDIO module. USB requires a DPO4USB module. Ethernet requires a DPO4ENET module. MIL-STD-1553 requires a DPO4AERO module. See Application module free trial on page 18.
56 MDO4000C Series Oscilloscope User Manual
Trigger setup
Preliminary

Triggering on buses

You can use your oscilloscope to trigger on multiple data buses, if you have the appropriate application module installed. The MDO4000C Series with Option MDO4MSO can trigger on parallel buses without an application module. The oscilloscope can display both the physical layer (as analog waveforms) and the protocol level information (as digital and symbolic waveforms).
To set up the bus trigger:
1. If you have not already defined your bus using the front panel B1, B2, or B3 buttons, do so now.
2. Push Trigger Menu.
3. Push Type.
4. Turn multipurpose knob a to scroll through the trigger type side menu until you select Bus.
5. Push Source Bus and use the Source Bus side menu to select the bus that you
want to trigger on.
6. Push Trigger On and select the desired trigger on feature from the side menu.
MDO4000C Series Oscilloscope User Manual 57
Trigger setup
Preliminary

Parallel bus trigger

(Requires option MDO4MSO.)
You can trigger on a binary or hex data value. Push Data on the lower menu and enter the parameters of interest with multipurpose knobs a and b.

I2C bus trigger

You can trigger on Start, Repeated Start, Stop, Missing Ack, Address, Data, or Address/Data.
If you are setting up an I2C trigger and have made a Trigger On selection of Address or Address/Data, push Address on the lower menu to access the I2C Address side menu.
Push Addressing Mode on the side menu and select 7 bit or 10 bit. Push Address on the side menu. Enter the address parameters of interest with multipurpose knobs a and b.
Then push Direction on the lower menu and select the direction of interest: Read, Write, or Read or Write.
If you have made a Trigger On selection of Data or Address/Data, push Data on the lower menu to access the I2C Data side menu.
Push Number of Bytes and enter the number of bytes with multipurpose knob a.
Push Addressing Mode on the side menu and select 7-bit or 10–bit. Push Data on the side menu. Enter the data parameters of interest with multipurpose knobs a and b.
For more information on the I2C address formats, refer to item 2 under Setting Up Bus Parameters.
58 MDO4000C Series Oscilloscope User Manual
Trigger setup
Preliminary

SPI bus trigger

You can trigger on SS Active, MOSI, MISO, or MOSI & MISO.
If you are setting up an SPI trigger and have made a Trigger On selection of MOSI or MISO, push Data on the lower menu, push MOSI or MISO on the side menu, and enter the data parameters of interest using multipurpose knobs a and b.
Then push Number of Bytes and enter the number of bytes with multipurpose knob a.
If you select MOSI & MISO, push Data on the lower menu and enter the parameters of interest in the side menus.

RS-232 bus trigger

You can trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, or Rx Data.
If you are setting up an RS-232 trigger and have made a Trigger On selection of Tx Data or Rx Data, push Data on the lower menu.
Push Number of Bytes and enter the number of bytes with multipurpose knob a.
Push Data on the side menu, and enter the parameters of interest with multipurpose knobs a and b.
MDO4000C Series Oscilloscope User Manual 59
Trigger setup
Preliminary

CAN bus trigger

You can trigger on Start of Frame. Type of Frame, Identifier, Data, Id & Data, End of Frame, and Missing Ack.
If you are setting up a CAN trigger and have made a Trigger On selection of Type of Frame, push Frame Type on the lower menu, and select Data Frame, Remote Frame, Error Frame, or Overload Frame.
If you have made a Trigger On selection of Identifier, push Identifier on the lower menu and select a Format. Then push Identifier on the side menu, and enter a binary or hex value with multipurpose knobs a and b.
Push Direction on the lower menu and select the direction of interest: Read, Write, or Read or Write.
If you have made a Trigger On selection of Data, push Data on the lower menu and enter the parameters of interest.

LIN bus trigger

You can trigger on Sync, Identifier, Data, Id & Data, Wakeup Frame, Sleep Frame, or Error.
If you are setting up an LIN trigger and have made a Trigger On selection of Identifier, Data, or Identifier & Data, push Identifier or Data on the lower
menu and enter the parameters of interest on the resulting side menu.
If you have made a Trigger On selection of Error, push Error Type on the lower menu and enter the parameters of interest on the side menu.
60 MDO4000C Series Oscilloscope User Manual
Trigger setup
Preliminary

FlexRay bus trigger

You can trigger on Start of Frame, Type of Frame, Identifier, Cycle Count, Header Fields, Data, Id & Data, End of Frame or Error.

Audio bus trigger

If you are using an I2C, Left Justified (LJ), or Right Justified (RJ) audio bus, you can trigger on Word Select or Data.
If you are using an TDM audio bus, you can trigger on Frame Sync or Data.

USB bus trigger

You can trigger on Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet, Handshake Packet, Special Packet, or Error.

Ethernet bus trigger

NOTE. For triggering on a high speed USB bus (480 MB/s), use an oscilloscope with 350 MHz or higher bandwidth.
You can trigger on Start Frame Delimiter, MAC Addresses, MAC Length/ Type, TCP/IPv4 Client Data, End of Packet, Idle, or an FCS (CRC) Error. If you turn on Q-(VLAN) Tagging, you can also trigger on MAC Q-Tag Control Information.
MDO4000C Series Oscilloscope User Manual 61
Trigger setup
Preliminary

MIL-STD-1553 bus trigger

You can trigger on Sync, Command, Status, Data, Time (RT/IMG), or Error.
If you are setting up a MIL-STD-1553 trigger and have made a Trigger On selection of Command, push RT Address on the lower menu to enter specific values of RT Address to trigger on. Push Command Word Details on the lower menu to enter the T/R bit value, Subaddress/Mode value, Word Count/Mode Code value, and Parity value.
If you are setting up a MIL-STD-1553 trigger and have made a Trigger On selection of Status, push RT-Address on the lower menu to enter specific values of RT Address to trigger on. Push Status Word Bits on the lower menu to enter values for Message Error (bit 9), Instr. (bit 10), Service Req. (bit 11), BCR
(bit 15), Busy (bit 16), Subsystem Flag (bit 17), DBCA (bit 18), Terminal Flag (bit 19) and Parity.
If you are setting up a MIL-STD-1553 trigger and have made a Trigger On selection of Data, push Data on the lower menu to enter specific Data values, and the Parity value.
If you are setting up a MIL-STD-1553 trigger and have made a Trigger On selection of Time (RT/IMG), push Trigger When on the lower menu to set the trigger condition. Push Times on the lower menu to set the Maximum and Minimum times.
If you are setting up a MIL-STD-1553 trigger and have made a Trigger On selection of Error, push l Error Type on the lower menu to select the type of error to trigger on.

Data value matching

You can trigger on a specific data value for RS-232 bytes. If you defined an end­of-packet character to use for RS-232 bus decoding, you can use the same end-of­packet character as a data value for trigger data matching. To do so, choose the Tx End of Packet or the Rx End of Packet character as the Trigger On selection.
You can also trigger on a specific data value for other buses.

Parallel bus trigger data matching

Optimum parallel bus trigger performance is achieved by using only analog channels or only digital channels.
62 MDO4000C Series Oscilloscope User Manual
Trigger setup
Preliminary

Checking trigger settings

To quickly determine the settings of some key trigger parameters, check the Trigger readout at the bottom of the display. The readouts differ for edge and the advanced triggers.
1. Trigger source = channel 1.
2. Trigger slope = rising.
3. Trigger level = 0.00 V.
MDO4000C Series Oscilloscope User Manual 63
Trigger setup
Preliminary

Starting and stopping an acquisition

After you have defined the acquisition and trigger parameters, start the acquisition with Run/Stop or Single.
Push Run/Stop to start acquisitions. The oscilloscope acquires repeatedly until you push the button again to stop the acquisition.
Push Single to take a single acquisition.
If any RF trace and any other analog or digital waveform are active, then pressing Run/Stop to stop acquisitions will cause the oscilloscope to wait for one more trigger event before stopping. While waiting for the trigger event, the Run/Stop button will turn yellow and the Single button will turn green. Once the acquisition occurs, the Run/Stop button will turn red and the Single button will become unlit.
Single sets the trigger mode to Normal for the single acquisition.
If the Trigger Mode is set to Auto and another trigger event does not occur within the Auto trigger timeout period then an acquisition will be made and the instrument will stop.
If the Trigger mode is set to Normal, the oscilloscope will continue waiting for a trigger event to occur as long as necessary
64 MDO4000C Series Oscilloscope User Manual

Display waveform or trace data

Preliminary
This section contains concepts and procedures for displaying the acquired waveform or trace.

Adding and removing a waveform

1. To add or remove a waveform from the display, push the corresponding front panel channel button or the D15-D0 button.
You can use the channel as a trigger source whether or not it is displayed.
MDO4000C Series Oscilloscope User Manual 65
Display waveform or trace data
Preliminary

Setting waveform intensity

1. Push Intensity on the front panel.
This will bring up the intensity readout on the display.
2. Rotate multipurpose knob a to select the desired waveform intensity.
3. Rotate multipurpose knob b to select the desired intensity for the graticule.
4. Push Intensity again to clear the intensity readout from the display.
66 MDO4000C Series Oscilloscope User Manual
Display waveform or trace data
Preliminary

Scaling and positioning a waveform

Use the horizontal controls to adjust the time base, adjust the trigger point, and to examine waveform details more closely. You can also use the Wave Inspector Pan and Zoom controls to adjust the display of waveforms. See Using wave
inspector to manage long record length waveforms on page 96.
If you push the Horizontal Position knob and Delay is set to On, the horizontal position is set to 0 seconds. If you push it and Delay is set to Off, the horizontal position is set to 10%.
Original waveform Scaled horizontally Positioned horizontally
Use the vertical controls to select waveforms, adjust the waveform vertical position and scale, and set input parameters. Push a channel menu button (1, 2, 3, or 4), button as many times as needed and the associated menu items to select, add, or remove a waveform.
Pushing the Vertical position control will center the waveform on the screen vertically, and pushing the Vertical scale control will switch from Coarse to Fine scaling.
Original waveform
Original waveform
Scaled vertically
Scaled vertically
Positioned vertically
Positioned vertically
Quick Tips
Preview. If you change the Position or Scale controls when the acquisition is stopped or when it is waiting for the next trigger, the oscilloscope rescales and repositions the relevant waveforms in response to the new control settings. It simulates what you will see when you next push the RUN button. The oscilloscope uses the new settings for the next acquisition.
You may see a clipped waveform if the original acquisition went off the screen.
MDO4000C Series Oscilloscope User Manual 67
Display waveform or trace data
Preliminary
The math waveform, cursors, and automatic measurements remain active and valid when using preview.
68 MDO4000C Series Oscilloscope User Manual
2345678
Display waveform or trace data
Preliminary

Setting input parameters

Use the vertical controls to select waveforms, adjust the waveform vertical position and scale, and set input parameters.
1. Push a channel menu button 1, 2, 3, or 4 to bring up the vertical menu for the designated waveform. The vertical menu only affects the selected waveform.
Pushing a channel button will also select or cancel that waveform selection.
2. Push Coupling repeatedly to select the coupling to use.
Use DC coupling to pass both AC and DC components.
Use AC coupling to block the DC component and show only the AC signal.
Coupling
DC| AC
3. Push Termination repeatedly to select the input impedance to use.
Set the input impedance (termination) to 50 Ω or 1 M Input impedance is automatically set to 1 MΩ when using AC coupling.
For more information on input impedance, see Quick Tips.
4. Push Invert to invert the signal.
Select Off for normal operation and On to invert the polarity of the signal in the preamplifier.
Termination
1MΩ | 50Ω
Invert
On | Off
Bandwidth
Full
(1) Label
Ω if using DC coupling.
More
MDO4000C Series Oscilloscope User Manual 69
Display waveform or trace data
Preliminary
5. Push Bandwidth, and select the desired bandwidth on the side menu.
The set choices are: Full, 250 MHz, and 20 MHz. Additional choices may appear, depending on the probe that you use.
Full to set the bandwidth to the full oscilloscope bandwidth.
Select
Select 250 MHz to set the bandwidth to 250 MHz.
Select 20 MHz to set the bandwidth to 20 MHz.
NOTE. 200 MHz model oscilloscopes do not include a 250 MHz option on the menu.
6. Push Label to create a label for the channel.
7. For some types of probes, you can push this button to instruct the
oscilloscope to perform an AC calibration on the entire signal path from the probe tip to the specific oscilloscope channel. This can result in a flatter frequency response over the entire frequency range.
8. Push More to access additional side menus.
10. Select Offset to enable multipurpose knob a to make vertical offset
adjustments.
Push Set to 0 V on the side menu to set the vertical offset to 0 V.
For more information on offset, see Quick Tips.
11. Select Probe Setup to define probe parameters.
On the side menu:
Select Voltage or Current to set the probe type for probes that do not have a TekProbe Level 1, TekProbe II (requires a TPA-BNC adapter) or TekVPI interface.
For probes that do not have a Tek interface, when Probe Type is set to Voltage, use multipurpose knob a to set Attenuation to match the probe
For probes that do not have a Tek interface, when Probe Type is set to Current, use multipurpose knob a to set the Amps/Volts ratio (Attenuation) to match the probe.
If you are measuring current by probing the voltage drop across a resistor, set Measure Current to Yes. Push A/V ratio on the side menu and turn multipurpose knob a to set the Amps/Volts or Volts/Amp ratio of your setup. For example, if you are measuring the drop across a 2 Ω resistor, set the V/A ratio to 2.
70 MDO4000C Series Oscilloscope User Manual
Display waveform or trace data
Preliminary
12. Select Deskew to make display and measurement adjustments for probes that have differing propagation delays. This is especially important when using a current probe in conjunction with a voltage probe.
For best results, use a deskew fixture, such as the Tektronix 067-1686-xx.
If you do not have a deskew fixture, you can use the controls in the Deskew menu to set the oscilloscope's deskew parameters to recommended values, based on the nominal propagation delay of each probe. The oscilloscope automatically loads the nominal propagation delay values of TekVPI and TekProbe II (requires use of a TPA-BNC adaptor) probes. For other common probes, first push Select on the side menu, and select the channel to which the probe is attached. Then
Probe Model on the side menu, and select the probe model. If your probe is
push not in the list, set probe model to Other, and push Propagation Delay on the side menu and dial in its propagation delay with multipurpose knob a.
To display the recommended deskew values calculated by the oscilloscope, set Show rec. deskews on the side menu to Yes.
To set the deskew values of each channel to the recommended values, push Set
all deskews to recommended values on the side menu.
Quick Tips
Using Probes with the TekProbe II and TekVPI Interfaces. When you attach a probe with the TekProbe II or the TekVPI interface, the oscilloscope sets the channel sensitivity, coupling, and termination resistance automatically to match the probe requirements. Tek Probe II probes require use of the TPA-BNC Adapter.
The Difference Between Vertical Position and Offset. Adjust the vertical position to place the waveforms where you want to see them. The waveform baseline indicators indicate the zero Volts (or Amps) level for each waveform. If you adjust the channel's Vertical Scale, the waveform expands or contracts around the waveform's baseline indicator.
When you use the Channel<x> > More >Offset > Vertical Offset control to move a waveform, the baseline indicator no longer represents zero. Instead, it represents the level of the offset. If you adjust the channel's Vertical Scale, the waveform expands or contracts around the waveform's baseline indicator.
50 Ω Protection. If you select 50 Ω termination, the maximum vertical scale factor is limited to 1 V/div, except that with a 10X probe the scale factor is 10 V. If you apply excessive input voltage, the oscilloscope automatically switches to 1 MΩ termination to protect the internal 50 Ω termination. For more details, refer to the specifications in the MDO4000C Series Oscilloscopes Technical Reference.
MDO4000C Series Oscilloscope User Manual 71
2
6
Display waveform or trace data
Preliminary

Positioning, scaling, and grouping digital channels

1. Push the front panel D15–D0 button.
2. Pushl D15–D0 on the lower menu.
D15 – D0
On/Off
4. Turn multipurpose knob a to select the channel to move.
5. Turn multipurpose knob b to move the selected channel.
NOTE. The display of the channel (or group) only moves after you stop rotating the knob.
Thresholds Edit Labels Monitor MagniVu
On | Off
Height
S | M L
72 MDO4000C Series Oscilloscope User Manual
Display waveform or trace data
Preliminary
6. To change the scale (height) of the digital channels, push Height on the lower menu.
NOTE. The S (Small) selection will display each waveform at 0.2 divisions tall. The M (Medium) selection will display each waveform at 0.5 divisions tall. The L (Large) selection will display each waveform at 1 division tall. L only works if there is enough room in the display to display the waveforms. You can display up to 10 L waveforms at one time.
7. You can label individual digital channels for easier identification. See
GUID-168896E1-E133-4460-8160-3A2D65209ADD#GUID-168896E1­E133-4460-8160-3A2D65209ADD.
8. To group some or all of the digital channels together, move the channels right next to each other. All the channels that are next to each other automatically form a group.

Viewing digital channels

The various ways of displaying data from the digital channels help you analyze the signals. Digital channels store a high or low state for each sample.
Logic high levels are displayed in green. Logic low levels are displayed in blue. When a single transition occurs during the time represented by one pixel column, the transition (edge) is displayed in gray.
You can see the groups by pushing Select on the side menu and turning multipurpose knob a.
When a group is selected, turn multipurpose knob b to move the whole group.
MDO4000C Series Oscilloscope User Manual 73
Display waveform or trace data
Preliminary
When multiple transitions occur during the time represented by one pixel column, the transition (edge) is displayed in white. When the display shows a white edge, indicating multiple transitions, you may be able to zoom in and see the individual edges.
When you are zoomed in so far that there is more than one pixel column per sample, the uncertainty of the edge position is indicated by light gray shading.
NOTE. When the light gray shading is displayed, use MagniVu.
74 MDO4000C Series Oscilloscope User Manual
2
2
4
Display waveform or trace data
Preliminary

Annotating the screen

You can add your own text to the screen by doing the following:
1. Push Utility.
2. Push Utility Page.
Utility Page
3. Turn multipurpose knob a and select Display.
4. Push Screen Annotation on the resulting lower menu.
Utility Page
Display
5. Push Display Annotation to select On on the side menu.
The annotation window now appears. Position it by turning multipurpose knobs a and b.
6. Push Edit Annotation on the side menu
7. Turn multipurpose knob a to scroll through the list of letters, numbers, and
other characters to select each desired character.
Alternatively, use a USB keyboard to type in characters. See Connecting a
USB keyboard to your oscilloscope on page 34.
To reposition the annotated text, push Position on the side menu and turn multipurpose knobs a and b, as desired.
Backlight
Intensity
High
Graticule
Full
Screen
Annotation
Trigger
Frequency
Readout
MDO4000C Series Oscilloscope User Manual 75
2
Display waveform or trace data
Preliminary

Viewing the trigger frequency

You can display a readout of trigger frequency. It counts all of the triggerable events, whether the oscilloscope triggered on them or not, and displays the number of times per second that they occur. To display this readout, do the following:
1. Push Utility.
2. Push Utility Page.
Utility Page
3. Turn multipurpose knob a and select Display.
4. Push Trigger Frequency Readout from the resulting lower menu.
5. Push On on the side menu.
The trigger frequency now appears in the Trigger readout, toward the lower right of the display.
76 MDO4000C Series Oscilloscope User Manual

Analyze waveform or trace data

Preliminary
After having properly set up the acquisition, triggering, and display of your desired waveform or trace, you can analyze the results. Select from features such as cursors, automatic measurements, statistics, waveform histograms, math, and FFT.

Taking automatic measurements in the time domain

To take an automatic measurement in the time domain:
1. Push Measure.
2. Push Domain to select Time domain measurements.
3. Push Add Measurement.
4. Turn multipurpose knob b to select the specific measurement. If needed, then turn multipurpose knob a to select the channel to measure on.
5. To remove a measurement, push Remove Measurement, turn multipurpose knob a to select the specific measurement, and push OK Remove Measurement on the side menu.
Quick Tips
To remove all measurements, select Remove All Measurements.
A symbol appears instead of the expected numerical measurement if a vertical clipping condition exists. Part of the waveform is above or below the display. To obtain a proper numerical measurement, turn the vertical scale and position knobs to make all of the waveform appear in the display.
If the oscilloscope displays a Low Resolution message, increase the record length of the acquisition so that the oscilloscope has more points from which to calculate the measurement.
MDO4000C Series Oscilloscope User Manual 77
Analyze waveform or trace data
Preliminary

Selecting automatic measurements in the time domain

The following tables list each automatic measurement by category: time or amplitude. See Taking automatic measurements in the time domain on page 77.
Table 3: Time measurements
Measurement Description
Frequency The first cycle in a waveform or gated region. Frequency is the
reciprocal of the period; it is measured in hertz (Hz) where one Hz is one cycle per second.
Period The time required to complete the first cycle in a waveform or
gated region. Period is the reciprocal of frequency and is measured in seconds.
Rise Time The time required for the leading edge of the first pulse in the
waveform or gated region to rise from the low reference value (default = 10%) to the high reference value (default = 90%) of the final value.
Fall Time The time required for the falling edge of the first pulse in the
waveform or gated region to fall from the high reference value (default = 90%) to the low reference value (default = 10%) of the final value.
Delay The time between the mid reference (default 50%) amplitude
point of two different waveforms. See also Phase.
Phase The amount of time that one waveform leads or lags another
waveform, expressed in degrees where 360° makes up one waveform cycle. See also Delay.
Positive Pulse Width
Negative Pulse Width
Positive Duty Cycle
Negative Duty Cycle
Burst Width The duration of a burst (a series of transient events) and is
The distance (time) between the mid reference (default 50%) amplitude points of a positive pulse. The measurement is made on the first pulse in the waveform or gated region.
The distance (time) between the mid reference (default 50%) amplitude points of a negative pulse. The measurement is made on the first pulse in the waveform or gated region.
The ratio of the positive pulse width to the signal period expressed as a percentage. The duty cycle is measured on the first cycle in the waveform or gated region.
The ratio of the negative pulse width to the signal period expressed as a percentage. The duty cycle is measured on the first cycle in the waveform or gated region.
measured over the entire waveform or gated region.
78 MDO4000C Series Oscilloscope User Manual
Loading...