G DC Offset Accuracy ........................................................................................117
iiMDO3000 Series Specifications and Performance Verification
General safety s
ummary
General safet
Review the following safety precautions to avoid injury a nd prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as s pecified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other
component manuals for warnings and cautions related to operating the system.
To avoid fire or personal injury
Use p roper power cord. Use only the power cord specified for this product and certified for the country of use.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected
to a voltag
Connect a
probe.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric
shock, the grounding conductor must be connected to earth ground. Before m aking connections to the input or output
termina
Observe
product manual for further ratings information before making connections to the product.
e source.
nd disconnect properly.
ls of the product, ensure that the product is properly grounded.
all terminal ratings.
y summary
De-energize the circuit under test before connecting or disconnecting the current
To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the
Connect the probe reference lead to earth ground only.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source. Do not block the power cord; it
must remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by
fied service personnel.
quali
exposed circuitry.
Avoid
Do not touch exposed connections and components when power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
vide proper ventilation.
Pro
proper ventilation.
Refer to the manual's installation instructions for details on installing the product so it has
MDO3000 Series Specifications and Performance Verificationiii
General safety s
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the m arking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ummary
ivMDO3000 Series Specifications and Performance Verification
Specifications
Specification
This chapter contains specifications for the MDO3000 Series oscilloscopes. All specifications are guaranteed unless noted
as "typical." Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked
with the
All specifications apply to all MDO3000 models unless noted otherwise. To meet specifications, two conditions must
first be met:
The o sc illos cope must have been operating continuously for twenty minutes within the operating temperature range
specified.
You must perform the S ignal Path Compensation (SPC) operation described in this manual prior to evaluating
specifications. (See page 75.) If the operating temperature changes by more than 10 °C (18 °F), you must perform
the SPC operation again.
s
symbol have associated procedures listed in the Performance Verification section.
Analog Channel Input And Vertical Specifications
Table 1: Analog channel input and vertical specifications
CharacteristicDescription
Number of input channels
MDO3104, MDO3054, MDO3034, MDO3024,
MDO3014
MDO3102, MDO3052, MDO3032, MDO3022,
MDO3012
4 analog, digitized simultaneously
2 analog, digitized simultaneously
Input coupling
Input termination selection
Input termination,
1M, DC coupled
DC, AC
1M,50,or75.
The 7 5 setting is not available on MDO310X instruments.
1M,±1%
MDO3000 Series Specifications and Performance Verification1
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Input termination,
50 , DC coupl
ed
(See page 77.)
Input termination,
75 , DC coupled
age 77.)
(See p
Maximum input voltage
(50 and 75 )
50 ,±1%
For instruments with 1 GHz bandwidth
VSWR 1.5:1 from DC to 1 GHz, typical
(includes MDO310X models as well as
MDO305X/303X/302X/301X models with 1 GHz
upgrade):
For instruments with 500 MHz bandwidth
(includes MDO305X models as well as
VSWR 1.5:1 from DC to 500 MHz,
typical
MDO303X/302X/301X models with 500 MHz
upgrade):
For instruments with 350 MHz bandwidth
(includes MDO303X models as well as
VSWR 1.5:1 from DC to 350 MHz,
typical
MDO302X/301X models with 350 MHz upgrade):
For instruments with 200 MHz bandwidth
(includes MDO302X models as well as
VSWR 1.5:1 from DC to 200 MHz,
typical
MDO301X models with 200 MHz upgrade):
For instruments with 100 MHz bandwidth
(MDO301X models):
VSWR 1.5:1 from DC to 100 MHz,
typical
75 ,±1%
VSWR 1.3:1fromDCto30MHz,typical
VSWR 1.5:1 from 30 MHz to 60 MHz, typical
with peaks ±20 V, (DF<=6.25%)
5V
RMS
There is an over-voltage trip circuit, intended to protect against overloads that might damage
termination resistors. A sufficiently large impulse can cause damage regardless of the over-voltage
protection circuitry, due to the finite time required to detect the over-voltage condition and respond
to it.
Maximum input voltage
(1 M)
The m aximum input voltage at the BNC, 300 V
Installation Category II.
RMS
.
De-rate at 20 dB/decade between 4.5 MHz and 45 MHz, De-rate 14 db between 45 MHz and
450 MHz. Above 450 MHz, 5 V
RMS
.
Maximum peak input voltage at the BNC, ±424 V
2MDO3000 Series Specifications and Performance Verification
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Specifications
DC balance
(See page 78.)
Delay between channels,
full bandwidth, typical
Deskew range–125 ns to +125 ns
Crosstalk (channel
isolation), typical
TekVPI InterfaceThe probe interface allows installing, powering, compensating, and controlling a wide range of
0.2 div with the input DC-50 coupled and 50 terminated
0.25 div at 2 mV/div with the input DC-50 coupled and 50 terminated
0.5 div at 1 mV/div with the input DC-50 coupled and 50 terminated
0.2 div with the input DC-75 coupled and 75 terminated
0.25 div at 2 mV/div with the input DC-75 coupled and 75 terminated
0.5 div at 1 mV/div with input DC-75 coupled and 75 terminated
0.2 div with the input DC-1 M coupled and 50 terminated
0.3 div at 1 mV/div w ith the input DC-1 M coupled and 50 terminated
All the above specifications are increased by 0.01 divisions per °C above 40 °C.
100 ps between any two channels w ith input termination set to 50 , DC coupling
100 ps between any two channels w ith input termination set to 75 , DC coupling
Note: all settings in the instrument can be manually time aligned using the Probe Deskew function
100 MHz>100 MHz
1M
50
75
probes offering a variety of features.
The interface is available on CH1-CH4 front panel inputs. Aux In is available on the front of
two-channel instrument only and is fully VPI compliant. Four-channel instruments have no Aux In
input.
100:130:1
100:130:1
100:130:1
Number of digitized bits
Sensitivity range (fine)Allows continuous adjustment from 1 mV/div to 10 V/div, 1 M
Sensitivity resolution
(fine), typical
Position range±5 divisions
MDO3000 Series Specifications and Performance Verification3
8bits
Displayed vertically with 25 digitization levels (DL) per division, 10.24 divisions dynamic range
"DL" is the abbreviation for "digitization level." A DL is the smallest voltage level change that can
be resolved by an 8-bit A-D Converter. This value is also known as the LSB (least significant bit).
1M:1 mV/div to 10 V/div in a 1-2-5 sequenceSensitivity range (coarse)
50 and
75 :
Allows continuous adjustment from 1 mV/div to 1 V/div, 75
Allows continuous adjustment from 1 mV/div to 1 V/div, 50
1% of current setting
1 m V/div to 1 V/div in a 1-2-5 sequence
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Analog bandwidth,
50 input ter
(See page 80.)
Analog bandwidth, 75
input termination, typical
mination
The limits stated below are for ambient temperature of 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument
Bandwidth
1.00 GHzDC to 1.0 GHzDC to 500 MHzDC to 350 MHzDC to 150 MHz
500 MHz
350 MHz
200 MHz
100 MHz
The limits stated below are for ambient temperature of 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument
Bandwidth
500 MHz,
350 MHz,
and 200 MHz
100 MHz
10 mV/div to
1V/div
DC to 500 MHzDC to 350 MHzDC to 150 MHz
DC to 350 MHzDC to 350 MHzDC to 150 MHz
DC to 200 MHzDC to 150 MHz
DC to 100 MHz
10 mV/div to
1V/div
DC to 200 MHzDC to 140 MHzDC to 100 MHz
DC to 100 MHz
5 m V/div to
9.98 mV/div
5 m V/div to
9.98 mV/div
Vertical Scale Setting
2 mV/div to
4.98 mV/div
Vertical Scale Setting
2 mV/div to
4.98 mV/div
1 mV/div to
1.99 mV/div
1 mV/div to
1.99 mV/div
Analog bandwidth, 1 M
input termination. The
Analog Bandwidth when
the instrument is DC-1M
coupled, typical
The limits stated below are for ambient temperature of 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument
Bandwidth
1GHz,
500 MHz, or
350 MHz
200 MHz
100 MHz
Vertical Scale Setting
2mV/divto
10 V/div
DC to 350 MHzDC to 150 MHz
DC to 200 MHzDC to 150 MHz
DC to 100 MHz
1 m V/div to
1.99 mV/div
4MDO3000 Series Specifications and Performance Verification
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Specifications
Analog Bandwidth, 1 M
with Standard Probe,
typical
Calculated rise time,
typical
The limits stated below are for ambient temperature of 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument
Bandwidth
1 G HzDC to 1.00 GHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
500 MHz
350 MHz
200 MHz
100 MHz
The formula is calculated by measuring –3 dB bandw idth of the oscilloscope. The formula accounts
for the rise time contribution of the oscilloscope independent of the rise time of the signal source.
All values in the above table are in pS. 1 GHz BW models assume the TPP1000 probe. 500 MHz
and 350 MHz models assume the TPP0500B probe. 200 MHz and 100 MHz models assume
the TPP0250 probe.
Instrument
Bandwidth
1GHz
Vertical Scale Setting
100 mV/div to
100 V/div
DC to 500 MHzDC to 400 MHzDC to 250 MHzDC to 150 MHz
DC to 350 MHzDC to 250 MHzDC to 150 MHz
DC to 200 MHzDC to 150 MHz
DC to 100 MHz
1 mV/div to
1.99 mV/div
2,666 ps1,333 ps800 ps400 ps
50 mV/div to
99.8 mV/div
Vertical Scale Setting (50 )
2 m V/div to
4.98 mV/div
20 mV/div to
49.8 mV/div
5mV/divto
9.98 mV/div
10 mV/div to
19.9 mV/div
10 mV/div to 1 V/div
Analog bandwidth limit
filter selections
500 MHz2,666 ps1,333 ps800 ps800 ps
350 MHz2,666 ps1,333 ps1,143 ps1,143 ps
200 MHz2,666 ps2,000 ps2,000 ps2,000 ps
100 MHz4,000 ps4,000 ps4,000 ps4,000 ps
Instrument
Bandwidth
1GHz
500 MHz2,666 ps1,600 ps1,000 ps800 ps
350 MHz2,666 ps1,600 ps1,143 ps1,143 ps
200 MHz2,666 ps2,000 ps2,000 ps2,000 ps
100 MHz4,000 ps4,000 ps4,000 ps4,000 ps
For instruments w ith 1 GHz, 500 MHz or 350 MHz analog bandwidth: 20 MHz, 250 MHz, and Full
For instruments with 200 MHz and 100 MHz analog bandwidth: 20 MHz and Full
10 mV to
19.9 m V
2,666 ps1,600 ps1,000 ps400 ps
Vertical Scale Setting (TPPXXX0 probe)
20 mV to 49.8 mV50 mV to 99.8 mV100 mV to 100 V
MDO3000 Series Specifications and Performance Verification5
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Lower frequency limit, AC
coupled, typical
Upper frequency limit,
250 MHz bandwidth limit
filter, typical
Upper frequency limit,
20 MHz bandwidth limit
filter, typical
DC gain accuracy
(See pag
DC voltage
measurement accuracy
e 83.)
Sample
acquisition mode,
typical
< 10 Hz when AC to 1 M coupled
The AC coupled lower frequency limits are reduced by a factor of 10 when 10X passive probes
are used.
250 MHz, +25%, and –25% (all models, except 100 MHz and 200 MHz)
20 MHz, ±25% (all models)
±2.5% for 1 mV/Div, derated at 0.100%/°C above 30 °C
±2.0% for 2 mV/Div, derated at 0.100%/°C above 30 °C
±1.5% for 5 mV/Div and above, derated at 0.100%/°C above 30 °C
±3.0% Variable Gain, derated at 0.100%/°C above 30 °C
Measurement type
Any sample
Delta volts between any two
samples acquired with the
same oscilloscope setup and
ambient conditions
DC Accuracy (in volts)
±[DC gain accuracy X | reading – (offset – position) |
+ Offset Accuracy +0.15 div + 0.6 mV]
±[DC gain accuracy X | reading | + 0.15 div + 1.2 mV]
Average
acquisition mode
NOTE. Offset, position, and the constant offset term must be converted to volts by multiplying by
the appropriate volts/div term.
Average of 16 waveforms±[DC gain accuracy X | reading – (offset – position) |
+ Offset Accuracy + 0.1 div]
Delta volts between any two
averages of 16 waveforms
acquired with the same
oscilloscope setup and ambient
conditions
NOTE. Offset, position, and the constant offset term must be converted to volts by multiplying by
the appropriate volts/div term.
NOTE. The basic accuracy specification applies directly to any sample and to the following
measurements: High, Low, Max, Min, Mean, Cycle Mean, RMS, and Cycle RMS. The delta volt
accuracy specification applies to subtractive calculations involving two of these measurements.
The delta volts (difference voltage) accuracy specification applies directly to the following
measurements: Positive Overshoot, Negative Overshoot, P eak-P eak, and Amplitude.
±[DC gain accuracy X | reading | + 0.05 div]
6MDO3000 Series Specifications and Performance Verification
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
Specifications
Offset ranges
Offset accuracy
(See page 87.)
Offset rangeVolts/div
setting
1 mV/div to
1MΩ input50 Ω and 75 Ω input
±1 V±1 V
50 mV/div
50.5 mV/div
±0.5 V±0.5 V
to
99.5 mV/div
100 mV/div
±10 V±5 V
to
500 mV/div
505 mV/div
±5 V±5 V
to
995 mV/div
1 V/div to
±100 V±5 V
10 V/div
1
NOTE. The i nput signal cannot exceed the maximum input voltage for the 50 Ω and 75 Ω input
paths. Refer to the Maximum input voltage specifications (earlier in this table) for more information.
±[0.005 X | offset - position | + DC Balance]
NOTE. Both the position and constant offset term must be converted to volts by multiplying by the
appropriate volts/div term.
Random noise,
sample acquisition mode,
termination setting
50
50, RMS, unit in m V
(See page 90.)MDO31xxMDO305xMDO303xMDO302xMDO301x
1 mV, Full BW0.210.20.20.20.19
mV, Full BW
2
.33
0
.27
0
.25
0
.23
0
.21
0
5 mV, Full BW0.550.360.360.30.3
10 mV, Full BW0.70.50.50.450.45
20 mV, Full BW10.90.90.90.9
50 mV, Full BW32.752.752.752.25
100 mV, Full BW4.54.154.154.154.15
200 mV, Full BW98.158.158.158.15
500 mV, Full BW2120202020
1V,FullBW4040404040
MDO3000 Series Specifications and Performance Verification7
Specifications
Table 1: Analog channel input and vertical specifications (cont.)
CharacteristicDescription
1 mV, 250 MHzBW0.210.20.2
——
2 mV, 250 MHzBW0.250.230.23
5 mV, 250 MHzBW0.350.30.3
10 mV, 250 MHzBW0.50.50.5
20 mV, 250 MHzBW11 1
50 mV, 250 MHzBW2.752.752.75
100 mV, 250 MHzBW4.154.154.15
200 mV, 250 MHzBW101010
500 mV, 250 MHzBW202020
1 V, 250 MHz BW404040
1 mV, 20 MHz BW 0.120.120.120.120.12
2 mV, 20 MHz BW 0.150.150.150.150.15
5 mV, 20 MHz BW 0.250.250.250.250.25
——
——
——
——
——
——
——
——
——
10 mV, 20 MHzBW0.450.40.40.40.4
20 mV, 20 MHzBW0.80.750.750.750.75
50 mV, 20 MHzBW2.12.12.12.12.1
100 mV, 20 MHzBW4.14.14.14.14.1
200 mV, 20 MHzBW88 888
500 mV, 20 MHzBW2020202020
1 V, 20 MHz BW4040404040
8MDO3000 Series Specifications and Performance Verification
Table 1: Analog channel input and vertical specifications (cont.)
For 50 and 75 path, 1 V/div is the maximum vertical setting.
MDO3000 Series Specifications and Performance Verification9
Specifications
Digital Channel Acquisition System Specifications
Table 2: Digital channel acquisition system specifications
Characterist
Threshold voltage range–15 V to +25 V
Digital channel timing resolution
Threshold accuracy (See page 93.)
Minimum detectable pulse2.0 ns
Channel to channel skew (typical)
ic
Description
Minimum: 2 ns
Minimum: 121.2 ps for MagniVu memory
± [100 mV + 3%
Using MagniVu memory. Specified at the input to the P6316 probe with all
eight ground inputs connected to the user's ground. Use of lead sets, grabber
clips, ground extenders, or other connection accessories may compromise this
specification.
500 ps
Digital Channel to Digital Channel only
This is the propagation path skew, and ignores skew contributions due to
bandpass distortion, threshold inaccuracies (see Threshold Accuracy), and
sample binning (see Digital Channel Timing Resolution).
for the main memory
of threshold setting after calibration], after valid S PC
10MDO3000 Series Specifications and Performance Verification
Horizontal And Acquisition System Specifications
Table 3: Horizontal and acquisition system specifications
Specifications
Characterist
Long-term sample rate and delay
ic
Description
±10 ppm over any 1 ms time i nterval
time accuracy (See page 89.)
Seconds/division rangeMDO30XX models: 1 ns/div to 1000 sec/div
MDO310X models: 400 ps/div to 1000 sec/div
Peak detect or
envelope mode pulse response, typical
InstrumentMinimum pulse width
Models at 1 GHz BW
>1.5ns
Models at 500 MHz BW> 2.0 ns
Models at 350 MHz BW> 3.0 ns
Models at 200 MHz BW> 5.0 ns
Models at 100 MHz BW> 7.0 ns
Sample-rate rangeSee Sample Rate Range detail table. (See page 13.)
Record length range1K, 10K, 100K, 1M, 5M, 10M
Maximum triggered acquisition rate1 and 2 channels3 and 4 channels
BandwidthFastAcq
1GHz
> 280,000
DPO
> 80,000 wfm/sec> 230,000 wfm/sec> 50,000 wfm/sec
FastAcq
wfm/sec
<1GHz
> 230,000
> 50,000 wfm/sec> 230,000 wfm/sec> 50,000 wfm/sec
wfm/sec
Aperture uncertainty, typical
(also called “sample rate jitter”)
(5ps+1×10
Record duration = (Record Length) / (Sample Rate)
-6
X record duration)
, for records having duration 1 minute
RMS
DPO
Number of waveforms for average
acquisition mode
Delta time measurement accuracy
e page 91.)
(Se
2 to 512 waveforms
Default of 16 waveforms
The formula to calculate delta-time measurement accuracy (DTA) for a given
instrument setting and input signal is given below (assumes insignificant signal
content above Nyquist).
SR1=SlewRate(1stEdge) around the 1stpoint in the measurement
SR
=SlewRate(2ndEdge) around the 2ndpoint in the measurement
2
N = input-referred noise (volts
, refer to the Random Noise, Sample acquisition
rms
mode specification)
t
=1/(SampleRate)
sr
TBA = timebase accuracy (refer to the Long-term sample rate and delay time
accuracy specification above)
t
= delta-time measurement duration
p
RD = ( Record Length) / (Sample Rate)
MDO3000 Series Specifications and Performance Verification11
Specifications
Table 3: Horizontal and acquisition system specifications (cont.)
Assumes that error due to aliasing is insignificant.
The term under the square-root sign is the stability, and is related to the TIE
(Time Int
measurement. The second term is a result of both the absolute center-frequency
accuracy and the center-frequency stability of the timebase, and varies between
multiple
from the first single-shot measurement to the final single-shot measurement).
erval Error). The errors from this term occur throughout a single-shot
single-shot measurements over the observation interval (the amount of time
12MDO3000 Series Specifications and Performance Verification
Sample Rate Range
Table 4: Sample r ate range
(MDO310X with 3 or 4 channels enabled or all other MDO3000 with 1, 2, 3, or 4 channels enabled)
CharacteristicDescription
Specifications
Sample rate
range
(Analog
Channels)
Time/Div
1ns
2ns
4ns
10 ns
20 ns
40 ns
80 ns
100 ns
200 ns
400 ns
800 n
1 s
2 s
s
4
8 s
10 s
10 M record5 M record1 M record 100 K record10 K record1 K record
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
/s
2.5 GS
2.5 GS/s250 MS/s
s
GS/s
2.5
2.5 GS/s500 MS/s50 MS/s
2.5 GS/s250 MS/s25 MS/s
.25 GS/s
1
2.5 GS/s100 MS/s10 MS/s
1.25 GS/s
1.25 GS/s
/s
500 MS
MS/s
100
0 s
2
40 s
80 s
100 s
200 s
400 s
800 s
2.5 GS/s500 MS/s50 MS/s5 MS/s
2.5 GS/s250 MS/s25 MS/s2.5 MS/s
1.25 GS/s
2.5 GS/s100 MS/s10 MS/s1 MS/s
2.5 GS/s500 MS/s50 MS/s5 MS/s500 KS/s
2.5 GS/s1.25 GS/s250 MS/s25 MS/s2.5 MS/s250 KS/s
1.25 GS/s625 MS/s
MDO3000 Series Specifications and Performance Verification13
Specifications
Table 4: Sample rate range (MDO310X with 3 or 4 channels enabled or all other MDO3000 with 1, 2, 3, or 4
channels enabled) (cont.)
Characteris
Sample rate
range (Anal
Channels)
(Cont.)
tic
og
Description
Time/Div
1ms
2ms
4ms
10 ms
20 ms
40 ms
100 ms
200 ms
400 ms
1s
2s
4s
s
10
20 s
40 s
10 M record5 M record1 M record 100 K record10 K record1 K record
100 MS/s10 MS/s1 MS/s100 KS/s
500 MS/s250 MS /s50 MS /s5 MS/s500 KS/s50 KS/s
250 MS/s125 MS /s25 MS /s2.5 MS/s250 KS/s25 KS/s
100 MS/s50 MS/s10 MS/s1 MS/s100 KS/s10 KS/s
50 MS/s25 MS/s5 MS/s500 KS/s50 KS/s5 KS/s
25 MS/s12.5 MS/s2.5 MS/s250 KS/s25 KS/s2.5 KS/s
10 MS/s5 MS/s1 MS/s100 KS/s10 KS/s1 KS/s
250
5 KS/s
2
/s
KS/s
5MS/s2.5MS
2.5 MS/s1.25 MS/s250 KS/s25 KS/s2.5 KS/s250 S/s
1 M S/s500 KS/s100 KS/s10 KS/s1 KS/s100 S/s
KS/s
500
250 KS/s125 KS/s25 KS/s2.5 KS/s250 S/s25 S/s
100 KS/s50 KS/s10 KS/s1 KS/s100 S/s10 S/s
0KS/s
5
25 KS /s12.5 KS/s2.5 KS/s250 S/s25 S/s2.5 S/s
500 KS
S/s
50 K
KS/s
5
/s
50 KS/
5KS
00 S/s
5
s
/s
5 KS/s500 S/
S/s
500
0S/s
5
50 S
S/s
5
s
/s
100 s
200 s
400 s
1000 s
10 KS/s5 KS/s1 KS/s100 S/s10 S/s
5 KS/s2.5 KS/s500 S/s50 S/s5 S/s
2.5 KS/s1.25 KS/s250 S/s25 S/s2.5 S/s
1 KS/s500 S/s100 S/s10 S/s
14MDO3000 Series Specifications and Performance Verification
Table 5: Sample rate range, (MDO310X models with 1 or 2 channels enabled)
CharacteristicDescription
Specifications
Sample rate
range
(Analog
Channels)
Time/Div
400 ps
1ns
2ns
4ns
10 ns
20 ns
40 ns
100 ns
200 ns
400 ns
1 s
2 s
4 s
s
10
20 s
40 s
10 M record5 M record1 M record 100 K record10 K record1 K record
5GS/s
5GS/s
5GS/s
5GS/s
5GS/s
5GS/s
5 GS/s2.5 GS/s
5GS/s1GS/s
5GS/s500MS/
5 GS/s2.5 GS/s250 MS/s
5 GS/s1 GS/s100 MS/s
s
5GS/
5 GS/s2.5 GS /s250 MS/s25 MS/s
5 GS/s1 G S/s100 MS/s10 MS/s
S/s
5G
5 GS/s2.5 GS/s250 MS/s25 MS/s2.5 MS/s
0MS/s
50
500 M
MS/s
50
S/s
50 MS
5M
s
/s
S/s
00 s
1
200 s
400 s
5 GS/s1 GS/s100 MS/s10 MS/s1 MS/s
GS/s
5
2.5 GS/s1.25 GS/s250 MS/s25 MS/s2.5 MS/s250 KS/s
.5 GS/s
2
00 MS/s
5
0MS/s
5
5
MS/s
00 KS/s
5
MDO3000 Series Specifications and Performance Verification15
Specifications
Table 5: Sample rate range, (MDO310X models with 1 or 2 channels enabled) (cont.)
CharacteristicDescription
Sample rate
range (Analog
Channels)
(Cont.)
Time/Div
1ms
2ms
4ms
10 ms
20 ms
40 ms
100 ms
200 ms
400 ms
1s
2s
4s
10 s
20 s
40 s
100 s
10 M record5 M record1 M record 100 K record10 K record1 K record
1 G S/s500 MS/s100 MS/s10 MS/s1 MS/s100 KS/s
500 MS/s250 MS /s50 MS /s5 MS/s500 KS/s50 KS/s
250 MS/s125 MS /s25 MS /s2.5 MS/s250 KS/s25 KS/s
100 MS/s50 MS/s10 MS/s1 MS/s100 KS/s10 KS/s
50 MS/s25 MS/s5 MS/s500 KS/s50 KS/s5 KS/s
25 MS/s12.5 MS/s2.5 MS/s250 KS/s25 KS/s2.5 KS/s
10 MS/s5 MS/s1 MS/s100 KS/s10 KS/s1 KS/s
5 M S/s2.5 MS/s500 KS/s50 KS/s5 KS/s500 S/s
2.5 MS/s1.25 MS/s250 KS/s25 KS/s2.5 KS/s250 S/s
1 M S/s500 KS/s100 KS/s10 KS/s1 KS/s100 S/s
500 KS/s250 KS/s50 KS/s5 KS/s500 S/s50 S/s
250 KS/s125 KS/s25 KS/s2.5 KS/s250 S/s25 S/s
100 KS/s50 KS/s10 KS/s1 KS/s100 S/s10 S/s
50 KS/s25 KS/s5 KS/s500 S/s50 S/s5 S/s
25 KS /s12.5 KS/s2.5 KS/s250 S/s25 S/s2.5 S/s
10 KS/s5 KS/s1 KS/s100 S/s10 S/s
200 s
400 s
1000 s
5 KS/s2.5 KS/s500 S/s50 S/s5 S/s
2.5 KS/s1.25 KS/s250 S/s25 S/s2.5 S/s
1 KS/s500 S/s100 S/s10 S/s
16MDO3000 Series Specifications and Performance Verification
Trigger Specifications
Table 6: Trigger specifications
CharacteristicDescription
Specifications
Trigger level ranges
Trigger level
accuracy, DC
coupled, typical
Lowest frequency
for "Set Level to
50%" function,
typical
Trigger holdoff
range
SourceSensitivity
Any input channel
Aux In (External)
LineNot applicable
The line trigger level is fixed at about 50% of the line voltage.
This specification applies to logic and pulse thresholds.
For signals having rise and fall times 10 ns, the limits are as follows:
Source
Any channel±0.20 divisions
Aux In (external trigger)±(10% of setting + 25 mV)
LineNot applicable
45 Hz
20 ns minimum to 8 seconds maximum
±8 divisions from center of screen,
±8 divisions from 0 V when vertical LF reject trigger coupling
is selected
±8 V
Range
Trigger sensitivity
Edge trigger,
DC coupled,
typical
Edge trigger,
not DC coupled,
typical
Trigger SourceSensitivity
Any analog
channel
Aux in
(External)
LineNot applicable
Trigger
Coupling
AC1.5 times the DC Coupled limits for frequencies above 10 Hz.
NOISE REJ2.5 times the DC-coupled limits
HF REJ
LF REJ
1 mV/div to 4.98 m V/div: 0.75 div from DC to 50 MHz, increasing
to 1.3 div at instrument bandwidth.
5 m V/div: 0.40 divisions from DC to 50 MHz, increasing to
1 div at instrument bandwidth
200 mV from DC to 50 MHz, increasing to 500 mV at 200 MHz
Typical Sensitivity
Attenuates signals below 10 Hz
1.5 times the DC-coupled limit from DC to 50 kHz.
Attenuates signals above 50 kHz
1.5 times the DC-coupled limits for frequencies above 50 kHz.
Attenuates signals below 50 kHz
MDO3000 Series Specifications and Performance Verification17
Specifications
Table 6 : Trigger specifications (cont.)
CharacteristicDescription
Aux In (External
trigger)
Logic (pattern) trigger, DC coupled,
typical:
Trigger using a logic qualifier,
DC coupled, typical:
Delay-by-events sequence trigger,
DC coupled, typical:
Runt trigger, typical:
Pulse-width and glitch trigger, typical:
Video trigger,
typical
The limits for both delayed and main trigger are as follows:
SourceTypical Sensitivity
Any analog
input channel
Aux In
(External)
Maximum input voltage:
1.0 division from DC to maximum bandwidth
1.0 division from DC to maximum bandwidth
1.0 division from DC to maximum bandwidth
1.0 division from DC to maximum bandwidth
1.0 division, from DC to Max Bandwidth.
0.6 to 2.5 divisions of video sync tip
Video not supported through Aux In (External) input
At front panel connector, 300 V
Derate at 20 dB/decade above 3 MHz to 30 V
, Installation Category II;
RMS
at 30 MHz,
RMS
10 dB/decade above 30 MHz
Based upon sinusoidal or DC input signal. Excursion above
300 V should be less than 100 ms duration and the duty factor
is limited to < 44%. RMS signal level must be limited to 300 V.
If these values are exceeded, damage to the instrument may
result.
Edge, Pulse,
and Logic trigger
bandwidth, typical
Time accuracy
for Pulse-width
triggering
Bandwidth, typical:> 250 MHz
For instruments with 1 GHz bandwidth (includes MDO310X models
as well as MDO305X/303X/302X/301X models with 500 MHz upgrade):
For instruments with 500 MHz bandwidth (includes MDO305X models
as well as MDO303X/302X/301X models with 500 MHz upgrade):
For instruments with 350 MHz bandwidth (includes MDO303X models
as well as MDO302X/301X models with 350 MHz upgrade):
For instruments with 200 MHz bandwidth (includes MDO302X models
as well as MDO301X models with 200 MHz upgrade):
For instruments with 100 MHz bandwidth (MDO301X models):
Time rangeAccuracy
1 ns to 500 ns
520 ns to 8 s
±(20% of setting + 0.5 ns)
±(0.01% of setting + 100 ns)
1GHz
500 MHz
500 MHz
200 MHz
200 MHz
18MDO3000 Series Specifications and Performance Verification
Table 6: Trigger specifications (cont.)
CharacteristicDescription
Specifications
Video trigger
formats and field
rates
Logic trigger,
minimum logic or
re-arm time , typical
Setup/hold time
violation trigger
Minimum clock
pulse widths,
typical
Setup and hold
time ranges
Triggers from negative sync composite video, field 1 or field 2 for interlaced systems, any field, specific
line, or any line for interlaced or non-interlaced systems. Supported systems include NTSC, PAL,
SECAM.
Standard Video formats are: Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60,
1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/30, 1080p/50, 1080p/60, and custom
bi-level and tri-level sync video standards.
For all vertical settings, the minimums are:
Trigger typeMinimum pulse
width
Minimum re-arm
time
Minimum time
between channels
1
LogicNot applicable2 ns2 ns
Time Qualified Logic
4ns2ns2ns
For all vertical settings, the minimum clock pulse widths are:
Clock Active
2
User hold time + 2.5 ns
Clock Inactive
3
2ns
2
The limits are as follows;
FeatureMinMax
Setup time
Hold time
4
4
Setup + Hold time
–0.5 ns1.024 ms
1 ns1.024 ms
4
0.5 ns2.048 ms
Minimum pulse
width and
rearm time
NOTE. Input coupling on clock and data channels must be the same.
Trigger typeMinimum pulse
Minimum rearm time
width
Glitch
Pulse-width
4ns
4ns
2 ns + 5% of glitch width setting
2 ns + 5% of width upper limit setting
NOTE. For the pulse-width trigger class, pulse-width refers to the width of the
pulse being measured. The rearm time refers to the time between pulses.
4ns2nsRunt
NOTE. For the runt trigger class, pulse width refers to the width of the pulse being
measured. The rearm time refers to the time between pulses.
Time-qualified runt
Slew rate
4ns
4ns
8.5 ns + 5% of width setting
8.5 ns + 5% of delta time setting
NOTE. For the slew rate trigger class, pulse width refers to the delta time being
measured. The rearm time refers to the time it takes the signal to cross the two
trigger thresholds again.
MDO3000 Series Specifications and Performance Verification19
Specifications
Table 6 : Trigger specifications (cont.)
CharacteristicDescription
Rise/fall time trigger,
delta time range
Glitch, pulse-width,
or time-qualified runt
trigger,
time range
B trigger
(A/B sequence
trigger), time range
Standard serial bus
interface triggers
I2C
(Requires an
MDO3EMBD
app. module)
4 ns to 8 seconds
4 ns to 8 seconds
Trigger after events, minimum pulse width, typical:
Trigger after events, maximum event frequency, typical:
5
1 / (2 X Rated Instrument Bandwidth)
5
[Rated Instrument Bandwidth] or
500 MHz, whichever is lower
Minimum time between arm and trigger8 ns
B trigger after time, time range:
B trigger after events, event range:
8nsto8s
1 to 4,000,000
Maximum serial trigger bits:128 bits
Address Triggering:
7 and 10 bit user specified address, as well as General Call,
START byte, HS-mode, EEPROM, and CBUS
Data Trigger:
1 to 5 bytes of user specified data
Trigger On:Start
Repeated Start
Stop, Missing Ack
Address
Data
Address and Data
SPI
(Requires an
MDO3EMBD
app. module)
Maximum Data R ate:
Data Trigger:
10 Mb/s
1 to 16 bytes of user specified data
Trigger On:SS Active
MOSI
MISO
MOSI and MISO
Maximum Data R ate:
10 Mb/s
20MDO3000 Series Specifications and Performance Verification
Table 6: Trigger specifications (cont.)
CharacteristicDescription
Specifications
RS-232/422/
485/UART
(Requires a
MDO3COMP
app. module)
CAN
(Requires an
MDO3AUTO
app. module)
Data Trigger:Tx Data, Rx Data
Trigger On:Tx Start Bit
Rx Start Bit
Tx End of Packet
Rx End of Packet
Tx Data
Rx Data
Tx Parity Error
Rx Parity Error
Maximum Data Rate:
Data Trigger:
Trigger On:Start of Frame
10 Mb/s
1 to 8 bytes of user specified data, including qualifiers of equal
to (=), not equal to (<>), less than (<), greater than (>), less than
or equal to (<=), greater than or equal to (>=)
Type of Frame
Identifier, Data
Identifier and Data
End of Frame
Missing Ack
Bit Stuffing Error
LIN
(Requires a
MDO3AUTO
app. module)
Frame Type:
Identifier:Standard (11 bit) and Extended (29 bit) identifiers
Maximum Data Rate:
Data Trigger:
Trigger On :Sync
Maximum Data Rate:
Data, Remote, Error, Overload
1Mb/s
1 to 8 Bytes of user-specified data, including qualifiers of equal
to (=), not equal to (<>), less than (<), greater than (>), less than
or equal to (<=), greater than or equal to (>=)
Identifier
Data
Identifier & Data
Wakeup Frame
Sleep Frame
Error
1 M b/s (by LIN definition, 20 kbit/s)
MDO3000 Series Specifications and Performance Verification21
Specifications
Table 6 : Trigger specifications (cont.)
CharacteristicDescription
FlexRay
(Requires a
MDO3FLEX
app. module)
MIL-STD-1553
(Requires a
MDO3AERO
app. module)
Indicator Bits
Identifier Trigger11 bits of user-specified data, equal to (=),not equal to <>, less
Cycle Count Trigger6 bits of user-specified data, equal to (<=), greater than or equal
Header Fields Trigger
Data Trigger
End Of FrameUser-chosen types Static, Dynamic (DTS), and All.
Error
Trigger on
Trigger on:
Normal Frame, Payload Frame, Null Frame, Sync Frame,
Startup Frame
than (<), greater than (>), less than or equal to (<=), greater
than or equal to (>=), Inside Range, Outside Range.
to (>=), Inside Range, Outside Range.
40 bits of user-specified data comprising Indicator Bits, Identifier,
Payload Length, Header CRC, and Cycle Count, equal to (=).
1 – 16 Bytes of user-specified data, with 0 to 253, or “don’t care”
bytes of data offset, including qualifiers of equal to (=), not equal
to <>, less than (<), greater than (>), less than or equal to (<=),
greater than or equal to (>=), Inside Range, Outside Range.
Start of Frame, Indicator Bits, Identifier, Cycle Count, Header
Fields, Data, Identifier & Data, End of Frame, or Error
Sync
Word Type (Command, Status, Data)
Command Word (set the followi ng individually: RT Address
(trigger when equal to (=), not equal to <>, less than (<), greater
than (>), less than or equal to (<=), greater than or equal to (>=),
inside range, outside range), T/R, Sub-Address/Mode, Data
Word Count/Mode Code, And Parity)
Status Word (set the following individually: RT address (trigger
when equal to (=), not equal to <>, less than (<), greater than
(>), less than or equal to (<=), greater than or equal to (>=),
inside range, outside range), message error, instrumentation,
service request bit, broadcast command received, busy,
subsystem flag, dynamic bus control acceptance (DBCA),
terminal flag, and parity)
Data Word (user-specified 16-bit data value)
Error (Sync, Parity, Manchester, Non-Contiguous Data
Idle Time (minimum time selectable from 4 s to 100 s;
maximum time selectable from 12 s to 100 s; trigger on <
minimum, > maximum, inside range, outside range)
Maximum Data R ate:
22MDO3000 Series Specifications and Performance Verification
Up to 1 Mb/s (for automated decoding of bus)
Table 6: Trigger specifications (cont.)
CharacteristicDescription
Specifications
I2S
(Requires a
MDO3AUDIO
app. module)
Left Justified
(Requires a
MDO3AUDIO
app. module)
Right Justified
(Requires a
MDO3AUDIO
app. module)
Data Trigger:
Trigger on:
Maximum Data Rate:
Data Trigger:
Trigger on:
Maximum Data Rate:
Data Trigger:
Trigger on:
Maximum Data Rate:
32 bits of user-specified data in a left w ord, right word, or either,
including qualifiers of equal to (=), not equal to <>, less than (<),
greater than (>), less than or equal to (<=), greater than or equal
to (>=), inside range, outside range.
Word Select
Data
12.5 Mb/s
32 bits of user-specified data in a left w ord, right word, or either,
including qualifiers of equal to (=), not equal to (<>), less than
(<), greater than (>), l ess than or equal to (<=), greater than or
equal to (>=), inside range, outside range
Word Select
Data
12.5 Mb/s
32 bits of user-specified data in a left w ord, right word, or either,
including qualifiers of equal to (=), not equal to <>, less than (<),
greater than (>), less than or equal to (<=), greater than or equal
to (>=), inside range, outside range
Word Select
Data
12.5 Mb/s
TDM
(Requires a
MDO3AUDIO
app. module)
Data Trigger:
Trigger on:
32 bits of user-specified data in a channel 0-7, including
qualifiers of equal to (=), not equal to <>, less than (<), greater
than (>), less than or equal to (<=), greater than or equal to (>=),
inside range, outside range
Frame Sync
Data
Maximum Data Rate:
1
For logic, time between channels refers to the length of time a logic state derived from more than one channel must exist to be recognized. For Time
Qualified Logic events, the time is the minimum time between a main and delayed event that will be recognized if more than one channel is used.
2
An active pulse width is the width of the clock pulse from its active edge (as defined through the Define Inputs button on the lower menu and the
ClockEdgebuttononthesidemenu)toitsinactiveedge. Aninactivepulsewidthisthewidthofthepulsefromitsinactiveedgetoitsactiveedge.
3
The User hold time is the number selected by the user through the Setup and Hold trigger menu.
4
Setup + Hold time is the algebraic sum of the Setup Time and the Hold Time programmed by the user.
For Setup time, positive numbers mean a data transition before the clock.
For Hold time, positive numbers mean a data transition after the clock edge.
5
Trigger after events is the time between the last A trigger event and the first B trigger event.
Trigger after time is the time between the end of the time period and the B trigger event.
25 Mb/s
MDO3000 Series Specifications and Performance Verification23
Specifications
Display Specifications
Table 7: Display specifications
CharacteristicDescription
Display type
Display resolution
Minimum
Luminance,
typical
9” WVGA LCD display
Display Area: 198 mm (H) X 111.696 mm (V).
800 X 480 pixels, each made up of 3 vertical stripe sub-pixels colored red, green, and blue
2
300 cd/m
at IBL = 5.0 mA
Input/Output Port Specifications
Table 8: Input/Output port specifications
CharacteristicDescription
Ethernet interfaceStandard on all models: 10/100 Mb/s
GPIB interfaceAvailable as an optional accessory that connects to USB Device and USB Host port,
USB interface1 Device and 2 Host connectors (all models)
Device port
Host ports
/lamp
rms
with the TEK -USB-488 GPIB to USB Adapter
Control interface is incorporated in the instrument user interface
One USB 2.0 H igh Speed port. Also supports Full Speed and Slow Speed Modes
Two USB 2.0 High Speed ports. One on front, one on rear
Video signal output
Probe compensator output voltage and
frequency, typical
Auxiliary output (AUX OUT)
A 15 pin, VGA RGB-type connector
Output voltage:
Default Mode: 0 to 2.5 V amplitude, ±2% (Source Impedance of 1 K)
TPPXXXX Cal Mode: 0 to 2.5 V amplitude, ±5% (Source Impedance of 25 )
Frequency: 1 kHz ±25%
Selectable Output: Main Trigger, Event, or A FG
Main Trigger: HIGH to LOW transition indicates the trigger occurred
Event Out: The instrument will output a negative edge during a specified trigger
event in a test application.
A falling edge occurs when there is a specified event in a test application (i.e. the
waveform crosses the violation threshold in the limit / mask test application).
A rising edge occurs when the trigger system begins waiting for the next test
application event.
AFG: The trigger output signal from the AFG.
Characteristic
Vout (HI) 2.25 V open circuit; 0.9Vintoa50 load to ground
Vout (LO) 0.7 V into a load of 4mA; 0.25 V into a 50 load to ground
Limits
24MDO3000 Series Specifications and Performance Verification
Power Source Specifications
Table 9: Power source specifications
CharacteristicDescription
Source voltage100 V to 240 V ±10%
Specifications
Source frequency
Fuse ratingT3.15 A, 250 V
Data Storage Specifications
Table 10: Data storage specifications
CharacteristicDescription
Nonvolatile memory retention time,
typical
Real-time clockA programmable clock providing time in years, months, days, hours, minutes, and
Environmental Specifications
100 V to 240 V:
115 V:
The fuse is not customer replaceable.
No time limit for front-panel settings, saved waveforms, setups, and calibration
constants
seconds
50/60 Hz
400 Hz ±10%
Table 11: Environmental specifications
CharacteristicDescription
Temperature
Humidity
Altitude
Acoustic noise emission
Operating:–10 °C to +55 °C (+14 °F to +131 °F)
Nonoperating:
Operating:
5% to 90% relative humidity (% RH) at up to +40 °C,
5% to 60% RH above +40 °C up to +55 °C, non-condensing
Nonoperating:
5% to 90% RH (Relative Humidity) at up to +40 °C,
5% to 60% RH above +40 °C up to +55 °C
5% to 40% RH above +55 °C up to +71 °C, non-condensing
Operating:3,000 m (9,843 feet)
Nonoperating:
Sound power level: 32.0 dBA in accordance with ISO 9296
–40 °C to +71 °C (–40 °F to +160 °F)
12,000 m (39,370 feet)
MDO3000 Series Specifications and Performance Verification25
Specifications
Mechanical Specifications
Table 12: Mechanical specifications
Characterist
Dimensions
P6316 D
Table 13: P6316 Digital probe input characteristics
Characteristic
Numbe
Input resistance, typical
Inpu
Minimum Input Signal Swing, typical
Maximum Input Signal Swing, typical
ic
Description
Height
Handle down203.28.0
Handle up25410.3
Width416.616.4
Depth147.45.8
WeightkgLb.
Stand alone, no front cover
With accessories & carry case
Package
d for domestic shipment
igital Probe Input Characteristics
ption
Descri
r of input channels
t capacitance, typical
16 Digital Inputs
101 K to ground
1
8pF
p-p
V, –20 V
1
500 mV
+30
mm
In.
4.29.2
6.815.0
8.619.0
Maximum Input Dynamic Range50 V
Channel-to-channel skew
, dependent on threshold setting
p-p
0ps
50
Digital c hannel to digital channel only.
This is the propagation path skew. It ignores skew contributions due to bandpass
stortion, threshold inaccuracies, and sample binning.
di
1
Specified at the input to the P6316 probe with all eight ground inputs connected to the user's ground. Use of leadsets, grabber clips, ground
xtenders, or other connection accessories may compromise this specification.
e
26MDO3000 Series Specifications and Performance Verification
RF Input Specifications
The following table shows the RF input specifications for the MDO3000 Series oscilloscopes.
Table 14: RF input specifications
CharacteristicDescription
Center frequency range9 kHz to 3.0 GHz (with MDO3SA installed)
Resolution bandwidth range
Kaiser RBW shape factor60 db/3 db Shape factor 4:1
Reference frequency error
(cumulative)
Specifications
9 kHz to 1.0 GHz (Any model at 1 GHz BW without MDO3SA installed)
9 kHz to 500 MHz (Any model at 500 MHz BW without MDO3SA installed)
9 kHz to 350 MHz (Any model at 350 MHz BW without MDO3SA installed)
9 kHz to 200 MHz (Any model at 200 MHz BW without MDO3SA installed)
9 kHz to 100 MHz (Any model at 100 MHz without MDO3SA installed)
Includes allowances for Aging per Year, Reference Frequency Calibration
Accuracy, and Temperature Stability.
Valid over the recommended 1 year calibration interval, from –10 °C to +55 °C .
-6
Marker frequency measurement
accuracy
Phase noise
Displayed average noise level
(DANL)
NOTE. The RF and analog channels share the same reference frequency.
Reference frequency accuracy is tested by the Long-term Sample Rate and
Delay Time Accuracy checks.
±(([Reference Frequency Error] x [Marker Frequency]) + (span / 750 + 2)) Hz
Reference Frequency Error = 10 ppm (10 Hz/MHz)
Example, assuming the span is set to 10 kHz and the marker is at 1,500 MHz,
this would result in a Frequency Measurement Accuracy of ±((10 Hz/1 MHz x
1,500 MHz) + (10 kHz / 750 + 2)) = ±15.015 kHz.
Marker Frequency with Span/RBW 1000:1
Reference Frequency Error with Marker level to displayed noise level > 30 dB
with auto settings on and signals 10 dB below reference level
2nd harmonic distortion: 9 kHz to 100 MHz: < –55 dBc (< –60 dBc typical)
with auto settings on, signals 10 dB below reference level, and reference level
with auto settings on and signals 10 dB below reference level
3rd harmonic distortion: 9 kHz to 100 MHz: < –55 dBc (< –60 dBc typical)
with auto settings on, signals 10 dB below reference level, and reference level
DC example: an input channel set up with +2 V offset and 1 V/div measuring a –5 V
signal would have ±( 2 mV + [((( 4 * 1 ) / 5 ) + 1 )% of 5 V] + [0.5% of 2 V] ) = ±(
2 mV + [1.8% of 5 V] + [0.5% of 2 V] ) = ±( 2 mV + 90 mV + 10 mV ) = ±102 mV.
This is roughly ±2% of the input voltage.
AC:±2%(40Hzto1kHz)
AC (typical): ±2% (20 Hz to 10 kHz)
For AC measurements, the input channel vertical settings must allow the V
pp
input
signal to cover between 4 and 8 divisions.
Frequency resolution5 digits
quency accuracy
Fre
±(10 µHz/Hz + 1 count)
Frequency counter sourceAny analog input channel.
MHz for 100 MHz models
Frequency counter maximum input
frequency
100
150 MHz for all other models
Trigger Sensitivity limits must be observed for reliable frequency measurements.
32MDO3000 Series Specifications and Performance Verification
Performance Ver
ification
Performance V
This chapter contains performance verification procedures for the specifications marked with thesymbol. The following
equipment, or a suitable equivalent, is required to complete these procedures.
Table 18: Required equipment
DescriptionMinimum requirementsExamples
DC voltage source3 mV to 100 V, ±0.1% accuracy
Leveled sine wave generator
Time mark generator80 ms period, ±1 ppm accuracy, rise
50 BNC cable
BNC feed-through termination50
RF signal generator
Power meterUse with Power sensor
Power sensor–30 dBm to +10 dBm
erification
Fluke 9500B Oscilloscope Calibrator
9 kHz to 3,000 MHz, ±4% amplitude
accuracy
time < 50 ns
Male-to-male connectorsTektronix part number
9 kHz to 3 GHz, –20 dBm to + 10 dBmAnritsu MG3690C series with options 2,
with a 9530 Output Module
An appropriate BNC-to-0.1 inch pin
adapter between the Fluke 9530 and
P6316 probe
012-0057-01 (43 inch)
Tektronix part number
011-0049-02
3, 4, 15, 22
Rhode & Schwarz NRX
Rhode & Schwarz NRP-Z98
Frequency counter0.1 Hz to 50 MHz, 5 ppm accuracy
DMM
You may need additional cables and adapters, depending on the actual test equipment you use.
These procedures cover all MDO3000 models. Please disregard any checks that do not apply to the specific model you are
testing.
Print the test record on the following pages and use it to record the performance test results for your oscilloscope.
NOTE. Completion of the performance verification procedure does not update the stored time and date of the latest
successful adjustment. The date and time are updated only when the adjustment procedures in the service manual are
successfully completed.
The performance verification procedures verify the performance of your instrument. They do not adjust your instrument. If
your instrument fails any of the performance verification tests, you should perform the factory adjustment procedures as
described in the MDO3000 Series Service Manual.
DC Voltage: 0.1% accuracy
AC RMS Voltage: 0.2% accuracy
Tektronix FCA3000
Tektronix DMM4040
MDO3000 Series Specifications and Performance Verification33
Performance Ver
ification
Upgrade the Firmware
For the best functionality, you can upgrade the oscilloscope firmware. To upgrade the firmware, follow these steps:
1. Open a Web browser and go to www.tektronix.com/software/downloads to locate the most recent firmware upgrade.
2. Download the latest fi rmware for your oscilloscope onto your PC.
3. Unzip the files and copy the "firmware.img" file into the root folder of a USB flash drive.
4. Power off your oscilloscope.
5. Insert the USB flash drive into a USB Host port on the front or back of the oscilloscope.
6. Power on the oscilloscope. The oscilloscope automatically recognizes the replacement firmware and installs it.
If the instrument does not install the firmware, rerun the procedure. If the problem continues, contact qualified service
personnel.
NOTE. Do not power off the oscilloscope or remove the USB flash drive until the oscilloscope finishes installing the firmware.
7. Power off the oscilloscope and remove the USB flash drive.
8. Power on the oscilloscope.
9. Push the Utility button on the front-panel.
10. Push Utility Page on the lower menu.
11. Turn Multipurpose knob “ a” and select Config.
12. Push About on the lower menu. The oscilloscope displays the firmware version number.
13. Confirm that the version number matches that of the new firmware.
The o scilloscope displays a message when the installation is complete.
34MDO3000 Series Specifications and Performance Verification
Test Record
Print this section for use during the Performance Verification.
Model numberSerial numberProcedure performed byDate
TestPassedFailed
Self Test
Performance Ver
ification
MDO3000 Series Specifications and Performance Verification35
Channels 3 and 4 are only on four-channel oscilloscopes.
46MDO3000 Series Specifications and Performance Verification
DC Offset Accuracy Tests
Table 22: DC Offset Accuracy
Performance Ver
ification
Performance
checksVertical scaleVertical offset
All models
Channel 1
20 MHz BW, 1 M
Channel 2
20 MHz BW, 1 M
1 mV/div
1 mV/div
2 mV/div
2 mV/div
10 mV/div
10 mV/div
100 mV/div
100 mV/
1V/div
1V/div
1.01
1.01 V/div
1 mV/div
1m
2 mV/div
div
V/div
V/div
700 mV696.2 mV703.8 m V
–700 mV–703.8 mV–696.2 mV
700 m696.1 mV703.9 m V
–700 mV–703.9 mV–696.1 mV
1 V993 mV1007 mV
–1 V–1007 mV–993 mV
10.0 V9.930 V10.07 V
–10.0 V–10.07 V–9.930 V
100 V99.30 V100.7 V
–100 V–100.
100 V99.30 V100.7 V
–100 V–100.7 V–99.30 V
mV
700
–700 mV–703.8 mV–696.2 mV
700 mV696.1 mV703.9 m V
1
Low limitTest resultHigh limit
7V
.2 mV
696
–99.3
703
0V
.8 mV
2 mV/div
0mV/div
1
10 mV/div
100 mV/div
100 mV/div
1V/div
1V/div
1.01 V/div
1.01 V/div
700 mV
–
1 V993 mV1007 mV
–1 V–1007 mV–993 mV
10.0 V9.930 V10.07 V
–10.0 V–10.07 V–9.930 V
100 V99.30 V100.7 V
–100 V–100.7 V–99.30 V
100 V99.30 V100.7 V
–100 V–100.7 V–99.30 V
703.9 m V
–
696.1 mV
–
MDO3000 Series Specifications and Performance Verification47
Performance Ver
Table 22: DC Offset Accuracy (cont.)
ification
Performance
checksVertical scaleVertical offset
Channel 3
2
20 MHz BW, 1 M
Channel 4
2
20 MHz BW, 1 M
1 m V/div
1 m V/div
2 m V/div
2 m V/div
10 mV/div
10 mV/div
100 mV/div
100 mV/div
1V/div
1V/div
1.01 V/div
1.01 V/div
1 m V/div
1 m V/div
2 m V/div
2 m V/div
700 mV696.2 mV703.8 mV
–700 mV-703.8 mV–696.2 m V
700 mV696.1 mV703.9 mV
–700 mV–703.9 mV–696.1 mV
1 V993 mV1007 mV
–1 V–1007 mV–993 mV
10.0 V9.930 V10.07 V
–10.0 V–10.07 V–9.930 V
100 V99.30 V100.7 V
–100 V–100.7 V–99.30 V
100 V99.30 V100.7 V
–100 V–100.7 V–99.30 V
700 mV696.2 mV703.8 mV
–700 mV–703.8 mV–696.2 mV
700 mV696.1 mV703.9 mV
–700 mV–703.9 mV–696.1 mV
1
Low limitTest resultHigh limit
10 mV/div
10 mV/div
100 mV/div
100 mV/div
1V/div
1V/div
1.01 V/div
1.01 V/div
1
Use this value for both the calibrator output and the oscilloscope offset setting.
2
Channels 3 and 4 are only on four-channel oscilloscopes.
1 V993 mV1007 mV
–1 V–1007 mV–993 mV
10.0 V9.930 V10.07 V
–10.0 V–10.07 V–9.930 V
100 V99.30 V100.7 V
–100 V–100.7 V–99.30 V
100 V99.30 V100.7 V
–100 V–100.7 V–99.30 V
48MDO3000 Series Specifications and Performance Verification
Sample Rate and Delay Time Accuracy
Table 23: Sample Rate and Delay Time Accuracy
Performance Ver
ification
Performance c
Sample Rate and Delay Time
Accuracy
hecks
Low limitTest resultHigh limit
–2 division+2 division
MDO3000 Series Specifications and Performance Verification49
Performance Ver
Random Noise, Sample Acquisition Mode Tests
Table 24: Random Noise, Sample Acquisition Mode
ification
Random No ise,
Sample Acquisition Mode
For instruments with
1 GHz bandwidth
(includes MD
O310X
models as well as
MDO305X/303X/302
X/301X mode
ls with
1 GHz upgrade)
For instruments with
500 MHz bandwidth
des MDO305X
(inclu
models as well as
MDO303X/302X/301X
s with 500 MHz
model
upgrade)
Channel 1
Channel 2
Channel 3
Channel 4
Channel 1
Channel 2
Channel 3
Bandwidth Sel
ection
Test resultHigh limit
Full4.50 mV
250 MHz4.15 mV
20 MHz4.10 mV
Full4.50 mV
250 MHz4.15 mV
20 MHz4.10 mV
1
Full4.50 mV
250 MHz4.15 mV
20 MHz4.10 mV
1
Full4.50 mV
250 MHz4.15 mV
20 MHz4.10 mV
Full4.15 mV
250 MHz4.15 mV
20 MHz4.10 mV
Full4.15 mV
250 MHz4.15 mV
20 MHz4.10 m
1
Full4.15 mV
V
250 MHz4.15 mV
Channel 4
Hz
20 M
1
Full4.15 mV
4.1
0mV
250 MHz4.15 mV
20
MHz
10 mV
4.
50MDO3000 Series Specifications and Performance Verification
Performance Ver
Table 24: Random Noise, Sample Acquisition Mode (cont.)
Random Noise, Sample Acquisition ModeBandwidth SelectionTest resultHigh limit
ification
For instruments with
350 MHz bandwidth
(includes MDO303X
models as well as
MDO302X/301X
models with 350 MHz
upgrade)
For instruments with
200 MHz bandwidth
(MDO302X models
as well as MDO301X
models with 200 MHz
upgrade)
Channel 1
Channel 2
Channel 3
Channel 4
Channel 1
Channel 2
Channel 3
Full4.15 mV
250 MHz4.15 mV
20 MHz4.10 mV
Full4.15 mV
250 MHz4.15 mV
20 MHz4.10 mV
1
Full4.15 mV
250 MHz4.15 mV
20 MHz4.10 mV
1
Full4.15 mV
250 MHz4.15 mV
20 MHz4.10 mV
Full4.15 mV
20 MHz4.10 mV
Full4.15 mV
20 MHz4.10 mV
Full4.15 mV
Channel 4
For instruments with
Channel 1
100 MHz bandwidth
(MDO301X models)
Channel 2
Channel 3
Channel 4
1
Channels 3 and 4 are only on four-channel oscilloscopes.
1
1
20 MHz4.10 mV
Full4.15 mV
20 MHz4.10 mV
Full4.15 mV
20 MHz4.10 mV
Full4.15 mV
20 MHz4.10 mV
Full4.15 mV
20 MHz4.10 mV
Full4.15 mV
20 MHz4.10 mV
MDO3000 Series Specifications and Performance Verification51
Performance Ver
Delta Time Measurement Accuracy Tests (MDO301X and MDO302X models)
Table 25: Delta Time Measurement Accuracy
Channel 1
ification
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V309 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.04ns
100 mV800 mV1.43 ns
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, S ource frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV20.3ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V203 ns
100 mV800 mV141 ns
500 mV4 V136 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.03s
52MDO3000 Series Specifications and Performance Verification
Table 25: Delta Time Measurement Accuracy (cont.)
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
Channel 2
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
Performance Ver
ification
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V309 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.04ns
100 mV800 mV1.43 ns
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, Source frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV20.3ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V203 ns
100 mV800 mV141 ns
500 mV4 V136 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO3000 Series Specifications and Performance Verification53
Performance Ver
Table 25: Delta Time Measurement Accuracy (cont.)
ification
Channel 3
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.03s
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
1
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V309 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.04ns
100 mV800 mV1.43 ns
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, S ource frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV20.3ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V203 ns
100 mV800 mV141 ns
500 mV4 V136 ns
54MDO3000 Series Specifications and Performance Verification
Table 25: Delta Time Measurement Accuracy (cont.)
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
Performance Ver
ification
Channel 4
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.03s
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
1
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V309 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.04ns
100 mV800 mV1.43 ns
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, Source frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV20.3ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V203 ns
MDO3000 Series Specifications and Performance Verification55
Performance Ver
Table 25: Delta Time Measurement Accuracy (cont.)
ification
100 mV800 mV141 ns
500 mV4 V136 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.03s
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
1
Channels 3 and 4 are only on four-channel oscilloscopes.
Delta Time Measurement Accuracy Tests (MDO303X and MDO305X models)
Table 26: Delta Time Measurement Accuracy
Channel 1
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/DivSource V
pp
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
Test ResultHigh Limit
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V336 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.44ns
100 mV800 mV1.43 ns
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, S ource frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
56MDO3000 Series Specifications and Performance Verification
Table 26: Delta Time Measurement Accuracy (cont.)
5mV40mV24.2ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
Performance Ver
ification
Channel 2
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V242 ns
100 mV800 mV141 ns
500 mV4 V136 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.42s
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V336 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.44ns
100 mV800 mV1.43 ns
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO3000 Series Specifications and Performance Verification57
Performance Ver
Table 26: Delta Time Measurement Accuracy (cont.)
ification
MDO = 4 μs/Div, S ource frequency = 240 kHz
Channel 3
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV24.2ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V242 ns
100 mV800 mV141 ns
500 mV4 V136 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.42s
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
1
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V336 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.44ns
100 mV800 mV1.43 ns
58MDO3000 Series Specifications and Performance Verification
Table 26: Delta Time Measurement Accuracy (cont.)
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, Source frequency = 240 kHz
Performance Ver
ification
Channel 4
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV24.2ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V242 ns
100 mV800 mV141 ns
500 mV4 V136 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.42s
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
1
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV232 ps
500 mV4 V232 ps
1 V4 V233 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V336 ps
100 mV800 mV273 ps
500 mV4 V270 ps
1 V4 V355 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
MDO3000 Series Specifications and Performance Verification59
Performance Ver
Table 26: Delta Time Measurement Accuracy (cont.)
ification
5mV40mV2.44ns
100 mV800 mV1.43 ns
500 mV4 V1.38 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, S ource frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV24.2ns
100 mV800 mV14.1 ns
500 mV4 V13.6 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V242 ns
100 mV800 mV141 ns
500 mV4 V136 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV2.42s
100 mV800 mV1.41 s
500 mV4 V1.36 s
1V4V2.69s
1
Channels 3 and 4 are only on four-channel oscilloscopes.
Delta Time Measurement Accuracy Tests (MDO310X models)
Table 27: Delta Time Measurement Accuracy
Channel 1
MDO = 4 ns/Div, Source frequency = 240 MHz
MDO V/DivSource V
pp
100 mV800 mV117 ps
500 mV4 V117 ps
1V4V119ps
MDO = 40 ns/Div, Source frequency = 24 MHz
60MDO3000 Series Specifications and Performance Verification
Test ResultHigh Limit
Table 27: Delta Time Measurement Accuracy (cont.)
Performance Ver
ification
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V386 ps
100 mV800 mV192 ps
500 mV4 V185 ps
1 V4 V293 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.69ns
100 mV800 mV1.55 ns
500 mV4 V1.45 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, Source frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV36.8ns
100 mV800 mV15.3 ns
500 mV4 V14.3 ns
1V4V26.9ns
Channel 2
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V368 ns
100 mV800 mV153 ns
500 mV4 V143 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.68s
100 mV800 mV1.53 s
500 mV4 V1.43 s
1V4V2.69s
MDO = 4 n s/Div, Source frequency = 240 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV117 ps
500 mV4 V117 ps
MDO3000 Series Specifications and Performance Verification61
Performance Ver
Table 27: Delta Time Measurement Accuracy (cont.)
ification
1V4V119ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V386 ps
100 mV800 mV192 ps
500 mV4 V185 ps
1 V4 V293 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.69ns
100 mV800 mV1.55 ns
500 mV4 V1.45 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, S ource frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV36.8ns
100 mV800 mV15.3 ns
500 mV4 V14.3 ns
1V4V26.9ns
Channel 3
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V368 ns
100 mV800 mV153 ns
500 mV4 V143 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.68s
100 mV800 mV1.53 s
500 mV4 V1.43 s
1V4V2.69s
1
MDO = 4 ns/Div, Source frequency = 240 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
62MDO3000 Series Specifications and Performance Verification
Table 27: Delta Time Measurement Accuracy (cont.)
100 mV800 mV117 ps
500 mV4 V117 ps
1V4V119ps
MDO = 40 ns/Div, Source frequency = 24 MHz
Performance Ver
ification
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V386 ps
100 mV800 mV192 ps
500 mV4 V185 ps
1 V4 V293 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.69ns
100 mV800 mV1.55 ns
500 mV4 V1.45 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, Source frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV36.8ns
100 mV800 mV15.3 ns
500 mV4 V14.3 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V368 ns
100 mV800 mV153 ns
500 mV4 V143 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.68s
100 mV800 mV1.53 s
500 mV4 V1.43 s
1V4V2.69s
MDO3000 Series Specifications and Performance Verification63
Performance Ver
Table 27: Delta Time Measurement Accuracy (cont.)
ification
Channel 4
1
MDO = 4 ns/Div, Source frequency = 240 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
100 mV800 mV117 ps
500 mV4 V117 ps
1V4V119ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V386 ps
100 mV800 mV192 ps
500 mV4 V185 ps
1 V4 V293 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.69ns
100 mV800 mV1.55 ns
500 mV4 V1.45 ns
1 V4 V2.70 ns
MDO = 4 μs/Div, S ource frequency = 240 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV36.8ns
100 mV800 mV15.3 ns
500 mV4 V14.3 ns
1V4V26.9ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5 mV40 m V368 ns
100 mV800 mV153 ns
500 mV4 V143 ns
1 V4 V269 ns
MDO = 400 μs/Div, Source frequency = 2.4 kHz
MDO V/DivSource V
pp
Test ResultHigh Limit
5mV40mV3.68s
100 mV800 mV1.53 s
64MDO3000 Series Specifications and Performance Verification
Table 27: Delta Time Measurement Accuracy (cont.)
500 mV4 V1.43 s
1V4V2.69s
1
Channels 3 and 4 are only on four-channel oscilloscopes.
Performance Ver
ification
MDO3000 Series Specifications and Performance Verification65
Performance Ver
Digital Threshold Accuracy Tests (with MDO3MSO option)
Table 28: Digital Threshold Accuracy (with MDO3MSO option)
Digital Threshold Accuracy (w ith M DO3MSO option)
ification
Digital
channelThresholdV
0 V-0.1 V0.1 VD0
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD1
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD2
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD3
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD4
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD5
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD6
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD7
4 V3.78 V4.22 V
Test result
s-
V
s+
Low limit
V
sAvg
=(V
)/2High limit
s--+Vs+
0 V-0.1 V0.1 VD8
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD9
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD10
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD11
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD12
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD13
4 V3.78 V4.22 V
0 V-0.1 V0.1 VD14
4 V3.78 V4.22 V
66MDO3000 Series Specifications and Performance Verification
Table 28: Digital Threshold Accuracy (with MDO3MSO option) (cont.)
Digital Threshold Accuracy (with MDO3MSO option)
Performance Ver
ification
Digital
channelThresholdV
s-
0 V-0.1 V0.1 VD15
4 V3.78 V4.22 V
Displayed Average Noise Le vel Tests (DANL)
Table 29: Displayed Average Noise Level
Displayed Average Noise Level (DANL)
Performance checks
All models
9kHz–50kHz
50 kHz – 5 MHz
5MHz–BW
(MDO3SA not
installed)
5MHz–2GHz
(MDO3SA installed)
2GHz–3GHz
(MDO3SA installed)
Low limitTest resultHigh limit
N/A–109 dBm/Hz
N/A–126 dBm/Hz
N/A–136 dBm/Hz
N/A–136 dBm/Hz
N/A–126 dBm/Hz
Test result
V
s+
Low limit
V
sAvg
=(V
)/2High limit
s--+Vs+
Residual Spurio us Response Tests
Table 30: Residual Spurious Response
Residual Spurious Response
Performance checks
All models
9kHzto50kHz
50kHzto5MHz
5 MHz to 2 GHz (not 1.25 GHz)N/A
1.25 GHz (MDO3SA installed)N/A
2GHzto3GHz(not2.5GHz)
(MDO3SA installed)
2.5 GHz (MDO3SA installed)N/A
Low limitTest resultHigh limit
N/A
N/A
–78 dBm
–78 dBm
–78 dBm
–76 dBm
N/A
–78 dBm
–69 dBm
MDO3000 Series Specifications and Performance Verification67
Performance Ver
Level Measurement Uncertainty Tests
Table 31: Level Measurement Uncertainty
Level Measurement Uncertainty
Performance checks
+10 dBmAll models
0 dBmAll models
–15 dBmAll mod
ification
els
Low limitTest resultHigh limit
9 kHz–1.2 dB+1.2 dB
50 kHz–1.2 dB+ 1.2 dB
100 kHz – 900 kHz–1.2 dB+1.2 dB
1MHz–9MHz–1.2dB+1.2dB
10 MHz - 90 MHz–1.2 dB+1.2 dB
100 MHz – BW–1.2 dB+1.2 dB
9 kHz–1.2 dB+1.2 dB
50 kHz–1.2 dB+ 1.2 dB
100 kHz – 900 kHz–1.2 dB+1.2 dB
MHz
z
–1.2 dB+1.2 dB
–1.2 dB+1.2 dB
1MHz–9MH
10 MHz - 90 MHz–1.2 dB+1.2 dB
100 MHz – BW–1.2 dB+1.2 dB
9 kHz–1.2 dB+1.2 dB
50 kHz–1.2 dB+ 1.2 dB
100 kHz – 900 kHz–1.2 dB+1.2 dB
1MHz–9
10 MHz – 90 MHz–1.2 dB+1.2 dB
100 MHz – BW–1.2 dB+1.2 dB
tional check with a TPA-N-PRE Preamp Attached
Func
Table 32: Functional check with a TPA-N-PRE Preamp attached
Functional check with a TPA-N-PRE Preamp attached
Performance checks
All models
1.7 GHz
2.9 GHz
LimitTest result
1.5 dB
1.5 dB
68MDO3000 Series Specifications and Performance Verification
Displayed Average Noise Level (DANL) with a TPA-N-PRE Preamp Attached
Table 33: Displayed Average Noise Level (DANL) with a TPA-N-PRE Preamp Attached
Displayed Average Noise Level (DANL) with a TPA-N-PRE Preamp Attached
Performance checks
All models
9kHz-50kHz
50 kHz - 5 MHz
50 kHz - BW
(MDO3SA not
installed
5MHz-2GHz
(MDO3SA installed)
2GHz-3GHz
(MDO3SA installed)
)
Low limitTest resultHigh limit
N/A–117 dBm/Hz
N/A–138 dBm/Hz
N/A–148 dBm/Hz
N/A–148 dBm/H
N/A–138 dBm/Hz
Auxiliary (Trigger) Output Tests
Performance Ver
z
ification
Table 3
Auxiliary (Trigger) Output Tests
Performance checks
Trigger Output
AF
4: Auxiliary (Trigger) Output Tests
High 1 M
Low 1 M
High 50
Low 50
G Sine and Ramp Frequency Accuracy Tests
Low lim
2.25
—
0.9
—
Table35: AFGSineandRampFrequency Accuracy Tests
G Sine and Ramp Frequency Accuracy
AF
erformance checks
P
All models
Sine Wave at 10 kHz, 2.5 V,
50
Sine Wave at 50 MHz,
2.5 V, 50
Low lim itTest resultHigh limit
9.9987 kHz10.0013 kHz
49.9975 MHz50.0025 MHz
it
V
V
Test re
sult
High li
—
0.7
—
25 V
0.
mit
V
MDO3000 Series Specifications and Performance Verification69
Performance Ver
AFG Square and Pulse Frequency Accuracy Tests
Table 36: AFG Square and Pulse Frequency Accuracy Tests
AFG Square and Pulse Frequency Accuracy
Performance checks
All m odels
ification
Square Wave at 25 kHz,
2.5 V, 50
Square Wave
2.5 V, 50
at 25 MHz,
Low limitTest resultHigh limit
24.99875 kHz25.00125 kHz
24.99875 MHz25.00125 MHz
AFG Signal
Amplitude Accuracy Tests
Table 37: AFG Signal Amplitude Accuracy Tests
AFG Signa
Perform
All m odels
l Amplitude Accuracy
ance checks
Square W
1kHz,50,0VOffset
Square Wave 1 V
50 ,0.2VOffset
ave 20 mV
pp
pp
@ 1 kHz,
@
AFG DC Offset Accuracy Tests
e 38: AFG DC Offset Accuracy Tests
Tabl
AFG DC Offset Accuracy
Performance checks
models
All
20 mV DC offset @ 50
1 V DC offset @ 50
Low limitTest resultHigh limit
9.35 mV10.65 mV
490.5 m
imit
Low l
7mV
18.
mV
984
V
result
Test
509.5 m
limit
High
3mV
21.
16 V
1.0
V
70MDO3000 Series Specifications and Performance Verification
DVM Voltage Accuracy Tests (DC)
Table 39: DVM Voltage Accuracy (DC)
Channel 1
Performance Ver
ification
Vertical Scale
1
0.5–2–2–2.052–1.948
0.5–1–0.5–1.0345–0.9655
0.2–0.5–0.5–0.5175–0.4825
0.010.00200.000420.00442
0.20.50.50.48250.5175
0.510.50.96551.0345
0.5221.9482.052
1
Channel 2
Vertical Scale
1
0.5–2–2–2.052–1.948
0.5–1–0.5–1.0345–0.9655
0.2–0.5–0.5–0.5175–0.4825
0.010.00200.000420.00442
Input Voltage
–5–5
55
Input Voltage
–5–5
Offset Voltage
Offset Voltage
Low L imitTest ResultHigh Limit
–5.117–4.883
4.8835.117
Low L imitTest ResultHigh Limit
–5.117–4.883
0.20.50.50.48250.5175
0.510.50.96551.0345
0.5221.9482.052
1
Channel 3
Vertical Scale
1
0.5–2–2–2.052–1.948
0.5–1–0.5–1.0345–0.9655
0.2–0.5–0.5–0.5175–0.4825
0.010.00200.000420.00442
0.20.50.50.48250.5175
0.510.50.96551.0345
0.5221.9482.052
1
1
55
Input Voltage
–5–5
55
Offset Voltage
4.8835.117
Low L imitTest ResultHigh Limit
–5.117–4.883
4.8835.117
MDO3000 Series Specifications and Performance Verification71
Performance Ver
Table 39: DVM Voltage Accuracy (DC) (cont.)
ification
Channel 4
Vertical Scale
1
1
Input Voltage
Offset Voltage
–5–5
Low LimitTest ResultHigh Limit
–5.117–4.883
0.5–2–2–2.052–1.948
0.5–1–0.5–1.0345–0.9655
0.2–0.5–0.5–0.5175–0.4825
0.010.00200.000420.00442
0.20.50.50.48250.5175
0.510.50.96551.0345
0.5221.9482.052
1
1
Channels 3 and 4 are only on four-channel oscilloscopes.
55
4.8835.117
72MDO3000 Series Specifications and Performance Verification
DVM Voltage Accuracy Tests (AC)
Table 40: DVM Voltage Accuracy (AC)
Channel 1
Performance Ver
ification
Vertical ScaleInput S ignal
5mV20mVppat 1 kH
z
Low LimitTest ResultHigh Limit
9.800 mV10.200 mV
10 mV50 mVppat 1 kHz24.5 mV25.500 mV
100 mV0.5 Vppat 1 kHz245.000 mV255.000 mV
200 mV1 Vppat 1 kHz490.000 mV510.000 mV
1V5Vppat 1 kHz2.450 mV2.550 mV
Channel 2
Vertical ScaleInput S ignal
5mV20mVppa
t1kHz
Low LimitTest ResultHigh Limit
9.800 mV10.200 m
V
10 mV50 mVppat 1 kHz24.5 mV25.500 mV
100 mV0.5 Vppat 1 kHz245.000 mV255.000 mV
200 mV1 Vppat
1kHz
490.00
0mV
510.00
0mV
1V5Vppat 1 kHz2.450 mV2.550 mV
1
el 3
Chann
Vertical ScaleInput S ignal
5mV20mV
at 1 kHz
pp
Low LimitTest ResultHigh Limit
9.80
0mV
10.2
00 mV
10 mV50 mVppat 1 kHz24.5 mV25.500 mV
100 mV0.5 Vppat 1 kHz245.000 mV255.000 mV
200
mV
1V
at 1 kHz
pp
490
.000 mV
510
.000 mV
1V5Vppat 1 kHz2.450 mV2.550 mV
annel 4
Ch
Vertical ScaleInput S ignal
mV
5
1
Low LimitTest ResultHigh Limit
0mV
2
at 1 kHz
pp
.800 mV
9
0.200 mV
1
10 mV50 mVppat 1 kHz24.5 mV25.500 mV
100 mV0.5 Vppat 1 kHz245.000 mV255.000 mV
200 mV1 Vppat 1 kHz490.000 mV510.000 mV
1V5Vppat 1 kHz2.450 mV2.550 mV
1
Channels 3 and 4 are only on four-channel oscilloscopes.
MDO3000 Series Specifications and Performance Verification73
Performance Ver
DVM Frequency Accuracy Tests and Maximum Input Frequency
Table 41: DVM Frequency Accuracy
Channel 1
Channel 2
ification
NominalLow LimitTest ResultHigh Limit
9.0000 Hz8.9998 Hz9.0002 Hz
99.000 Hz98.998 Hz99.002 Hz
999.00 Hz998.98 Hz999.02 Hz
99.000 kHz98.998 kHz99.002 kHz
999.00 kHz998.98 kHz999.02 kHz
150 MHz
2
149.99 MHz150.01 MHz
9.0000 Hz8.9998 Hz9.0002 Hz
99.000 Hz98.998 Hz99.002 Hz
999.00 H
z
99.000 kHz98.998 kHz99.002 kHz
999.00 kHz998.98 kHz999.02 kHz
2
0Hz
Channel 3
150 MHz
1
9.000
99.000 Hz98.998 Hz99.002 Hz
999.00 Hz998.98 Hz999.02 Hz
00 kHz
99.0
999.00 kHz998.98 kHz999.02 kHz
2
Channel 4
150 MHz
1
9.0000 Hz8.9998 Hz9.0002 Hz
.000 Hz
99
999.00 Hz998.98 Hz999.02 Hz
99.000 kHz98.998 kHz99.002 kHz
99.00 kHz
9
150 MHz
1
Channels 3 and 4 are only on four-channel oscilloscopes.
2
Verifies the maximum frequency.
2
998.98 H
149.99
8Hz
8.999
98 kHz
98.9
z
MHz
999.02 H
150.01
9.000
99.0
z
MHz
2Hz
02 kHz
149.99 MHz150.01 MHz
.998 Hz
98
98.98 kHz
9
.002 Hz
99
99.02 kHz
9
149.99 MHz150.01 MHz
74MDO3000 Series Specifications and Performance Verification
Performance Verification Procedures
The following three conditions must be met prior to performing these procedures:
1. The oscilloscope must have been operating continuously for twenty (20) minutes in an environment that meets the
operating range specifications for temperature and humidity.
Performance Ver
ification
2. You must perform a signal path compensation (SPC). (See Self Tests — System Diagnostics and Sig
Compensation section below.) If the operating temperature changes by more than 10 °C (18 °F), you must perform the
signal path compensation again.
3. You must connect the oscilloscope and the test equipment to the same AC power circuit. Connect the oscilloscope
and test instruments into a common power strip if you are unsure of the AC power circuit distribution. Connecting the
oscilloscope and test instruments into separate AC power circuits can result in offset voltages between the equipment,
which can invalidate the performance verification procedure.
The time required to complete all the procedures is approximately one hour.
WARNING. Some procedures use hazardous voltages. To prevent electrical shock, always set voltage source outputs
to 0 V before making or changing any interconnections.
nal Path
Self Tests — System Diagnostics and Signal Pa th Compensation
These procedures use internal routines to verify that the oscilloscope functions and passes its internal self tests. No test
equipment or hookups are required. Start the self test with these steps:
Run the System Diagnostics (may take several minutes):
1. Disconnect all probes and cables from the oscilloscope inputs.
2. Push Default Setup on the front-panel to set the instrument to the factory default settings.
3. Push Utility .
4. Push Utility Page on the lower menu, and turn Multipur po se knob “a” to select Self Test.
5. Push Self Test on the lower menu. The Loop X Times side menu button will be set to Loop 1 Times.
6. Push OK Run Self Test on the side menu.
7. Wait while the s elf test runs. When the self test completes, a dialog box displays the results of the self test.
8. Cycle the oscilloscope power off and back on before proceeding.
NOTE. Remember to cycle the oscilloscope power off and back on before proceeding.
Run the signal path compensation routine (may take 5 to 15 m inutes):
1. Push Default Setup on the front panel.
2. Push Utility .
3. Push Utility Page on the lower menu.
4. Turn Multipurpose knob “a” to select Calibratio n.
5. Push Signal Path on the lower menu.
MDO3000 Series Specifications and Performance Verification75
Performance Ver
6. Push OK-Compensate Signal Paths on the side menu.
7. When the signal path compensation is complete, push Menu Off twice to clear the d ialog box and Self Test menu.
8. Check the Signal Path button on the lower menu to verify that the status is Pass. If it does not pass, run the test again.
If it still doe
ification
s not pass, recalibrate the instrument or have the instrument serviced by qualified service personnel.
This complete
s the procedure.
76MDO3000 Series Specifications and Performance Verification
Performance Ver
Check Input Termination, DC Coupled (Resistance)
This test checks the Input Termination for 1 M,75 or 50 settings.
NOTE. The 75 Ω setting is not available on MDO310X instruments.
1. Connect the output of the oscilloscope calibrator (for example, Fluke 9500) to the oscilloscope channel 1 input, as
shown below.
2. Push Default Setup on the front panel to set the instrument to the factory default settings.
ification
3. Push the channel button on the front panel for the oscilloscope channel that you are testing, as shown in the test
for example, 1, 2, 3,or4).
record (
4. Confirm t
setting is 1MΩ.
5. Turn the Vertical Scale knob to set the vertical scale, as shown in the test record (for example, 10 mV/div, 100 mV/div,
1 V/div). (See page 36, Input Termination Tests.)
6. Measure the input resistance of the oscilloscope with the calibrator. Record this value in the test record.
7. Repeat steps 5 and 6 for each volt/division setting in the test record.
8. Change the oscilloscope termination to 75 and calibrator impedance to 50 and repeat steps 5 through 7.
9. Change the oscilloscope termination to 50 and repeat steps 5 through 7.
10. Repeat steps 4 through 9 for each channel listed in the test record and relevant to the model of oscilloscope that you are
test
This
hat the oscilloscope termination and calibrator impedance are both set to 1 M. The default Termination
ing, as shown in the test record (for example, 2, 3,or4).
completes the procedure.
MDO3000 Series Specifications and Performance Verification77
Performance Ver
Check DC Balance
This test checks the DC balance.
You do not need to connect the oscilloscope to any equipment to run this test. The only piece of equipment needed is a
BNC feed-through 50 terminator.
1. For 50 coupling, attach a 50 terminator to the channel input of the oscilloscope being tested.
2. Push Default Setup on the front panel to set the instrument to the factory default settings.
3. Push the channel button on the front panel for the oscilloscope channel that you are testing, as shown in the test
record (for example, 1, 2, 3,or4).
4. Set the oscilloscope termination to 50 . Push Termination on the lower menu to select 50 Ω.
5. Push Bandwidth on the lower menu, and push the appropriate bandwidth button on the side menu for 20MHz, 150MHz,
or Full, as given in the test record.
ification
6. Turn the Horizontal Scale k
NOTE. Step 6 only needs to be done once, at the beginning of the test.
7. Turn the Vertical Scale knob to set the vertical scale, as shown in the test record (for example, 1 mV /div, 2 mV/div,
10 mV/div, 100 mV/div, 1 V
8. Push Acquire on the front panel.
NOTE. Steps 8, 9, and 10 only need to be performed once, at the beginning of this test.
9. Push Mode on the lower menu, and then, if needed, push Average on the side menu.
10. If needed, adjust the number of averages to 16 using M ultipu rpo se knob “a” .
11. Push the Trigger Menu button on the front panel.
NOTE. Steps 11, 12, and 13 only need to be performed once, at the beginning of this test.
12. Push Source on the lower menu.
13. Select the AC Line trigger source on the side menu using Mu ltip urp ose knob “a”. You do not need to connect an
external signal to the oscilloscope for this DC Balance test.
nob to 1 ms/division.
/div).
14. On the front panel, push the Measure button on the Wave Inspector.
78MDO3000 Series Specifications and Performance Verification
NOTE. Steps 14 though 17 must be performed once for each input channel under test.
Performance Ver
ification
15. Push Add Measu
16. Use Multipurp
the channel input being tested.
17. Push OK Add Measurement on the side menu, and then Menu Off on the front panel.
18. View the m ean measurement value in the display and enter that mean value as the test result in the test record. (See
page 38, DC Balance Tests.)
NOTE. Trans
scale value. (e.g.0.2 V / (1 V / division) = 0.2 divisions)
19. Repeat step 7 and step 18 for each volts/division value listed in the results table.
20. Push the channel button on the front panel, then change the oscilloscope bandwidth (for example, 20 MHz, 150 MHz, or
d repeat step 7, step 18, and step 19.
Full), an
21. For 1 M co
22. Repeat s
oscilloscope (for example, 1, 2, 3,or4).
NOTE. The BNC 50 Ω terminator needs to be moved to next input channel.
teps 3 through 20 for each channel combination listed in the test record and relevant to your model of
rement on the lower menu.
ose knob “b” to select the Mean measurement. If needed, use Multipurpose kno b “ b” to select
late the mean value into divisions for use in the test record. To do this, divide the voltage value by the vertical
upling, change the oscilloscope termination to 1 M and repeat steps 5 through 20.
23. For 75 coupling, change the oscilloscope termination to 75 and repeat steps 5 through 20.
The BNC 50 Ω terminator needs to be moved to next input channel.
NOTE.
24. Repe
NOTE. The BNC 50 Ω terminator needs to be moved to next input channel.
at steps 3 through 20 for each channel combination listed in the test record and relevant to your model of
oscilloscope (for example, 1, 2, 3,or4).
This completes the procedure.
MDO3000 Series Specifications and Performance Verification79
Performance Ver
Check A nalog Bandwidth, 50 Ω
This test checks the bandwidth at 50 for each channel.
1. Connect the output of the leveled sine wave generator (for example, Fluke 9500) to the oscilloscope channel 1 input as
shown below.
ification
2. Push Defa
3. Push chan
4. Set the ca
5. Set the o
6. Turn the
ult Setup on the front panel to set the instrument to the factory default settings.
nel button 1, 2,3,or4 for the channel that you want to check.
librator to 50 output impedance (50 source impedance) and to generate a sine wave.
scilloscope termination to 50 . Push Termination on the lower menu to select 50 Ω.
Vertical Scale knob to set the vertical scale, as shown in the test record (for example, 1 mV/div, 2 mV/div,
5mV/div).
7. Push Acquire on the front panel.
8. Confirm that the mode is set to Sample. If not, push Mode on the lower menu, if needed, and then push the Sample
side bezel button.
9. Adjust the signal source to at least 6 vertical divisions at the selected vertical scale with a set frequency of 50 kHz. For
e, at 5 mV/div, use a 30 mV
exampl
signal; at 2 mV/div, use a 12 mV
p-p
signal; at 1 mV/div, use a ³ 6 mV
p-p
p-p
signal. U se a s ine wave for the signal source.
10. Turn the Horizontal Scale knob to 40 s/division.
11. On the front panel, push the Measure button on the Wave Inspector, and then push Add Measurement on the lower
menu.
. Steps 11 though 14 must be performed once for each input channel under test.
NOTE
12. Use Multipurpose knob “b” to select the Peak-to-peak measurement. U se Multipurpo se knob "a" to select the input
channel being tested, and then push OK Add Measurement on the side menu.
13. Push More on the lower menu to select Gating, and then push Off (Full Record) on the side menu.
14. Push Menu Off on the front panel. This will allow you to see the display. Note the mean V
ading V
re
cord the mean value of V
Re
rn the Horizontal Scale knob to 10 ns/division.
15.Tu
in-pp
.
(for example, 816 mV) in the test record. (See page 44, Analog Bandwidth Tests, 50 Ω.)
in-pp
of the signal. Call this
p-p
80MDO3000 Series Specifications and Performance Verification
Performance Ver
16. Adjust the signal source to the maximum bandwidth frequency for the bandwidth and model desired, as shown in the
following work
mean V
. Call this reading V
p-p
sheet. Measure V
of the signal on the oscilloscope using statistics, as in the previous step, to get the
p-p
.
bw-pp
ification
Record the value of V
in the test record.
bw-pp
NOTE. For more information on the contents of this worksheet, refer to the Analog Channel Input and Vertical Specifications
table. (See p
age 1, Analog Channel Input And Vertical Specifications.)
Table 42: Maximum Bandwidth Frequency worksheet
Terminatio
n
Vertical Sc
For instruments with 1 GHz bandwidth
(includes
MDO310X models as well as MDO305X/303X/302X/301X models with 1 GHz upgrade):
50 Ω10 mV/div1 GHz
50 Ω5 mV/div500 MHz
50 Ω2 mV/div350 MHz
50 Ω1 mV/div150 MHz
truments with 500 MHz bandwidth
For ins
(includes MDO305X models as well as MDO303X/302X/301X models with 500 MHz upgrade):
50 5mV/d
50 2 mV/div
50 1 mV/div
ale
iv
Maximum Ban
500 MHz
dwidth Frequency
350 MHz
Hz
150 M
For instruments with 350 MHz bandwidth
ludes MDO303X models as well as MDO302X/301X models with 350 MHz upgrade):
(inc
505 mV/div
502 mV/div
50
1m
V/div
350 MHz
MHz
350
150 MHz
For instruments with 200 MHz bandwidth
(includes MDO302X models as well as MDO301X m odels with 200 MHz upgrade):
50 Ω2 mV/div200 MHz
50
Ω
1m
V/div
15
0MHz
For instruments with 100 MHz bandwidth
(MDO301X models):
00 MHz
50 1 mV/div
17. Use the values of V
bw-pp
andV
obtained above and stored in the test record to calculate the Gain at bandwidth
in-pp
1
with the following equation:
Gain = V
bw-pp/Vin-pp
.
18. To pass the performance measurement test, Gain should be 0.707. Enter Gain in the test record.
MDO3000 Series Specifications and Performance Verification81
Performance Ver
19. Repeat steps 9 through 17 for the other oscilloscope volts/div settings listed in the test record.
20. Repeat steps 3 through 18 for each channel combination listed in the test record and relevant to your model of
oscilloscope
ification
(for example, 1, 2, 3,or4).
This complete
s the procedure.
82MDO3000 Series Specifications and Performance Verification
Performance Ver
Check DC Gain Accuracy
This test checks the DC gain accuracy.
1. Connect the oscilloscope to a DC voltage source. If using the Fluke 9500 calibrator, connect the calibrator head to
the oscilloscope channel to test.
2. Push Default Setup on the front panel to set the instrument to the factory default settings.
3. Push channel button 1, 2,3 ,or4 to select the channel that you want to check.
4. Confirm that the oscilloscope termination and calibrator impedance are both set to 1 M. On the oscilloscope, push
Termination on the lower menu to select 1MΩ.
ification
5. Push 20 MHz on the side menu to select the bandwidth (push Bandwidth on the lower menu, if necessary, to activate
the Band
6. Push Ac
7. Push Mo
8. On the f
9. Use Mul
width menu).
quire on the front panel.
de on the lower menu, and then push Average on the side menu. Use the default number of averages (16).
ront panel, push the Measure button on the Wave Inspector, and then Add Measurement on the lower menu.
tipurpose knob “b” to select the Mean measurement. Use Multipurpose knob “a” to select the input
channel to be tested.
10. Push OK Add Measurement on the side menu.
11. Push the Trigger Menu button on the front panel.
12. Push Source on the lower menu.
13. Turn M ultipu rpose knob “a” to select AC Line as the trigger source. Push Menu Off on the front panel.
14. Turn the vertical Scale knob to the next setting to measure, as shown in the Gain Expected worksheet below.
15. Set the DC Voltage S ource to V
.PushMeasure on the front panel, then push More on the lower menu to select
negative
Statistics. Push Reset Statistics on the side menu, and then push Menu Off on the front panel.
16. Enter the mean reading into Gain Expected worksheet below as V
17. Set the DC Voltage Source to V
the side menu, and then push Menu Off on the front panel. Enter the mean reading into the Gain Expected
on
worksheet as V
positive-measured
.
.PushMore on the lower menu to select Statistics, push the Reset Statistics
positive
negative-measured
.
MDO3000 Series Specifications and Performance Verification83
Performance Ver
Table 43: Gain Expected worksheet - channel 1
Oscilloscope
Vertical
Scale SettingV
1 mV/div
ification
diffExpected
V
negative
V
positive
7 mV–3.5 mV+ 3.5 mV
V
negative-
measured
V
positive-
measured
Test Result
(Gain
V
diff
Accuracy)
2 mV/div
14 mV–7 mV+7 mV
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10 mV70 mV–35 mV+35 mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
50 mV350 mV–175 mV+175 mV
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
Table 44: Gain Expected worksheet - channel 2
Oscilloscope
cal
Verti
Scale SettingV
1 mV/div
diffExpected
7 mV–3.5
V
negative
mV
V
positive
+3.5
mV
V
negative-
measured
V
positive-
measured
esult
Test R
(Gain
V
diff
Accuracy)
2mV
/div
14 mV–7 mV+7 mV
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10
mV
70
mV
–3
5mV
+3
5mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
0mV
5
50 mV
3
175 mV
–
175 mV
+
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
84MDO3000 Series Specifications and Performance Verification
Table 45: Gain Expected worksheet - channel 3
Oscilloscope
Vertical
Scale SettingV
1 mV/div
diffExpected
7 mV–3.5 m V+3.5 mV
V
negative
V
positive
V
negative-
measured
V
positive-
measured
Performance Ver
V
diff
ification
Test Result
(Gain
Accuracy)
2 mV/div
14 mV–7 mV+7 mV
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10 mV70 mV–35 mV+35 mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
50 mV350 mV–175 mV+175 mV
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
Table 46: Gain Expected worksheet - channel 4
Oscilloscope
Vertical
Scale SettingV
1 mV/div
diffExpected
7 mV–3.5 m V+3.5 mV
V
negative
V
positive
V
negative-
measured
V
positive-
measured
Test Result
(Gain
V
diff
Accuracy)
2 mV/div
14 mV–7 mV+7 mV
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10 mV70 mV–35 mV+35 mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
50 mV350 mV–175 mV+175 mV
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
18. Calculate V
as follows:
diff
MDO3000 Series Specifications and Performance Verification85
Performance Ver
V
ification
=|V
diff
negative-measured
– V
positive-measured
|
Enter V
in the Gain Expected worksheet.
diff
19. Calculate GainAccuracy as follows:
GainAccuracy =((V
– V
diff
diffExpected
)/V
diffExpected
) X 100%
Write down GainAccuracy in the Gain Expected worksheet and in the test record. (See page 45, DC Gain AccuracyTests .)
20. Repeat steps
21. Repeat step
This comple
14 through 18 for each volts/division value in the test record.
s 3 through 19 for each channel of the oscilloscope that you want to check.
tes the procedure.
86MDO3000 Series Specifications and Performance Verification
Performance Ver
Check Offset Accuracy
This test checks the offset accuracy.
1. Connect the oscilloscope to a DC voltage source to run this test. If using the Fluke 9500 calibrator as the DC voltage
source, connect the calibrator head to the oscilloscope channel to test.
2. Push Default Setup on the front panel to set the instrument to the factory default settings.
3. Push channel button 1,2,3,or4 to select the channel you want to check.
ification
4. Confirm that the oscilloscope termination and calibrator
lower menu to select 1MΩ.
5. Set the calibrator to the vertical offset value shown in the test record (for example, 700 mV for a 1 mV/div setting). Set
the calibrator impedance to match the termination setting for the oscilloscope.
6. On the oscilloscope, push More on the lower menu repeatedly, to select Offset.
7. Set the oscilloscope to the vertical offset value shown in the test record (for example, 700 mV for a 1 mV/div setting).
8. Turn the vertical Scale knob to match the value in the test record (for example, 1 m V/division).
9. Turn the Horizontal Scale knob to 1 ms/div.
10. Push Bandwidth on the lower menu.
11. Push 20 MHz on the side menu.
12. Check that the vertical position is set to 0 divs. If not, turn the appropriate Vertical Position knob to set the position to
0 divs.
Or, push More on the lower menu repeatedly to select Position, and then push Set to 0 divs on the side menu.
13. Push Acquire on the front panel.
14. Push Mode on the lower menu, and then pus
15. Push the Trigger Menu button on the front panel.
16. Push Source on the lower menu.
h Average on the side menu. Use the default number of averages (16).
impedance are both set to 1 M.PushTermination on the
17. Turn Multipurpose knob “a” to select AC Line as the trigger source.
18. On the front panel, push the Measure button on the Wave Inspector.
19. Push
20. Use Multipurpose knob “b ” to select the Mean measurement. Use Multipurpose knob “a” to select the input
MDO3000 Series Specifications and Performance Verification87
Add Measurement on the lower menu.
channel to be tested.
Performance Ver
21. Push OK Add Measurement on the side menu, and then Menu Off on the front panel. The mean value should appear in
a measurement p
ification
ane at the bottom of the display.
22. Enter the meas
23. Repeat the pro
24. Repeat all ste
This complet
es the procedure.
ured value in the test record. (See page 47, DC Offset Accuracy Tests.)
cedure (steps 6, 7, 8 and 22) for each volts/division setting shown in the test record.
ps, starting with step 1, for each oscilloscope channel you want to check.
88MDO3000 Series Specifications and Performance Verification
Performance Ver
ification
Check Long-term Sample Rate and Delay Time Accuracy
This test checks the sample rate and delay time accuracy (time base).
1. Push Default Setup on the oscilloscope front panel to set the instrument to the factory default settings.
2. Connect the output of the time mark generator to the oscilloscope channel 1 input using a 50 cable. Use the time mark
generator with a 50 source with the oscilloscope set for internal 50 termination.
3. Set the time mark generator to 80 ms. Use a time m ark waveform with a fast rising edge.
4. Set the mark amplitude to 1 V
.
pp
5. Set the oscilloscope vertical Scale to 500 mV/div.
6. Set the Horizontal Scale to 20 ms/div.
7. Adjust the Trigger Level for a triggered display.
8. Adjust the vertical Position knob to center the time mark on center screen.
9. Adjust the Horizontal Position knob counterclockwise to set the delay to exactly 80 ms.
10. Set the Horizontal Scale to 400 ns/div.
11. Compare the rising edge of the marker to the center horizontal graticule. The rising edge should be within ±2 divisions of
nter graticule. Enter the deviation in the test record. (S ee page 49, Sample Rate and Delay Time Accuracy.)
the ce
NOTE. One division of displacement from graticule center corresponds to a 5 ppm time base error.
This completes the procedure.
MDO3000 Series Specifications and Performance Verification89
Performance Ver
Check Random Noise, Sample Acquisition Mode
This test checks random noise. You do not need to connect any test equipment to the o scilloscope for this test.
1. Disconnect everything connected to the oscilloscope inputs.
2. Push Default Setup on the front panel to set the instrument to the factory default settings. This sets the oscilloscope to
Channel 1, Full Bandwidth, 1 M input termination, 100 mV/div, and 4.00 s/div.
3. Set Horizontal to 10 ms/div.
4. Set CH1 Vertical Channel Setting to 50 termination and the desired bandwidth.
5. Set up the measurements to do RMS and Mean measurement of selected channel and record the measurement.
ification
6. Calculate RMS noise voltage = Square root of (RMS
2
– Mean2), and record the result.
7. The calculated RMS noise voltage from step 6 should be less than the high limit in the test record (the calculated
maximum RMS noise).
8. Repeat the above test for all other input channels. Channels 3 and 4 are only available on three or four channel
oscilloscopes.
This completes the procedure.
90MDO3000 Series Specifications and Performance Verification
Performance Ver
Check Delta Time Measurement Accuracy
This test checks the Delta time measurement accuracy (DTA) for a given instrument setting and input signal.
1. Set the sine wave generator output impedance to 50 .
2. Push the oscilloscope front-panel Default Setup button, and then push Menu Off.
3. Connect a 50 coaxial cable from the signal source to the oscilloscope channel being tested.
4. Push the channel 1 button to display the channel 1 m enu.
5. Push Termination on the lower menu to set the channel to 50 Ω.
ification
6. Push the Trigger Menu button on the front panel, and then, if necessary, set the trigger source to the channel being tested:
a. Push Source on the lower menu.
b. Use the Multipurpose a knob to select the channel being tested.
7. On the front panel, push the Measure button on the Wave Inspector, and then push Add Measurement on the lower
menu.
8. Use Multip urp ose Knob “b” to select the Burst Width measurement, and then push OK Add Measurement on the
nu. Use Multipurpose Knob “a” to select the input channel to be tested.
side me
9. Push M
10. Push Menu Off on the front panel to remove the Statistics menu.
11. Refer to the Test Record Delta Time Measurement Accuracy table. (See page 60, Delta Time Measurement Accuracy
12. Push More on the lower menu to select Statistics, and then push Reset Statistics. Wait fi ve or 10 seconds for the
13. Ver
14. Rep
15. Pus
ore on the lower menu to select Statistics and, if necessary, use Multipurpose Knob “a” to set the Mean &
Std Dev Samples to 100, as shown in the side menu.
Tests (MDO310X models).) Set the oscilloscope and the signal source as directed there.
lloscope to acquire all the samples before taking the reading.
osci
ify that the Std Dev is less than the upper limit shown for each setting, and note the reading in the Test Record.
eat steps 11 through 13 for each setting combination shown in the Test Record for the channel being tested.
h the channel button on the front panel for the current channel to shut off the channel. Push the channel button for
the next channel to be tested, and move the coaxial cable to the appropriate input on the oscilloscope. Only the channel
being tested should be enabled
16. Repeat steps 5 through 15 until all channels have been tested.
MDO3000 Series Specifications and Performance Verification91
Performance Ver
NOTE. For this test, enable only one channel at a time. If three or more channels are enabled at the same time, the
maximum sample
This completes the procedure.
ification
rate is reduced and the limits in the Test Record are no longer valid.
92MDO3000 Series Specifications and Performance Verification
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.