Tektronix Logic Probe Deskew Fixture User manual

067-2298-00 Logic Probe Deskew Fixtur
ZZZ
Instructions
For best results up for 20 minutes, and then perform SPC (signal path compensati deskew procedure.
Fixture Preparation
You must solder the TriMode solder tip to the
e
xture before performing the deskew procedure. The following equipment is required:
ESD-approved soldering iron and tweezers
Wire, no-clean ux solder, and instructions (included with P75TLRST solder tips)
Cable tie or tape
M
1. Connect the TriMode solder tip. Refer to the gure below and the soldering instructions included with the solder tip.
, allow the oscilloscope to warm
on) before performing the
agnifying lens
2. Connect the prob on the xture.
3. Connect the logic probe to the oscilloscope, andthentothexture. Note the following:
You must connect single-ended probes with the signal side of the probe toward the short edge of the board as shown. (See Figure 3.)
Figure 3: Single-ended p ro be connections
e to the TriMode solder tip
*P071285000*
071-2850-00
Deskew Fixture
This xture provides an edge source to time-align trigger signals from the logic probe tips. It allows you to deskew a logic probe and one analog probe on the MSO70000 series oscilloscopes, resultinginimprovedreliabilityandaccuracy.
Figure 1: Deskew xture with TriMode solder tip attached
The xture supports P6717, P6717A, P6750 and P6780 oscilloscope logic probes and P7500 Series TriMode probes, and is powered by one of the USB ports of the oscilloscope. A USB cable is included with the xture.
A P7500 Series TriMode solder tip is also required for the alignment procedure, but is not included with the xture.
Equipment Required
Deskew xture and USB cable
One Logic Probe (P6717, P6717A , P6750 or P6780)
One analog P7500 Series TriMode Probe
One TriMode probe differential solder tip (tips with ground connections are recommended—the P75TLRST, for example)
Figure 2: P75TLRST solder tip connections
2. Check the Calibration Status of the Analog Probe: Select Probe Cal... from the Vertical menu. The Probe Setup screen appears.
On the left side of the Probe Setup screen, click the channel tab that corresponds to the channel that the analog probe is attached to, and then check the Probe Status readout:
Initialized. The probe has not been calibrated on the selected channel; perform the DC probe calibration procedure. Refer to the probe manual or oscilloscope online help for instructions.
Pass. The probe has been calibrated on the selected channel. Perform the deskew procedure.
Fail. The probe has not been calibrated or has failed; repeat the calibration procedure. If the test fails, troubleshoot the problem; do not proceed to the deskew procedure.
Deskew Procedure
1. Connect a P7500 Series TriMode probe to any channel (1–4) of the oscilloscope. Set the probe to A-B (differential).
Connect differential probes with the (+) side toward the short edge of the board. (See Figure 4.)
Figure 4: Differential probe connections
Connect P6750 logic probes to the color-keyed connector. Align the silver screw to the silver side. (See Figure 5.)
Figure 5: P6750 Probe connection
Probe connectio
n notes:
All 16 channels must be connected (in
any sequence).
Use the probe grouper or holder
accessory to simplify connecting to the board.
Leave the CLK channel disconnected
(unnecessary for the P6750 probe).
4. Plug the USB cable into the xture and into a USB port on the oscilloscope. The USB port only supplies board power; no communication occurs through the port.
5. Select Vertical | Probe Cal, and then select "D15-D0" on the left.
6. Select the proper Reference Channel from the drop down menu in the Trigger Path Alignment area of the Probe Setup menu.
7. ClicktheAlignTriggerPathbuttononthe Probe Setup screen to start the alignment.
8. Observe the Probe Status eld for results (Pass or Fail).
Probe Setup Screen Description
Trigger Path Alignment ? button. Click this
button to display additional information about the alignment paths used in the procedure.
Probe Status eld.
Initialized = ready for alignment
Pass = probes are aligned and ready to use
Fail = alignment failed; check connections and retry
Reference Channel drop-down menu. From this menu, select the oscilloscope channel that your analog reference probe is attached to.
Probe Propertie
s Screen Description
Probe drop-down menu. From this drop-down
menu, you can sel
ect the analog or logic probes
that are connected to the oscilloscope.
Probe Type and Serial Number elds. These elds identify
the probes that are connected to the digital channels and the reference channel of the oscillosco
pe. N/A indicates an alignment has
not been run.
Reference Channel, Type, and Serial Number
elds. These
elds display the properties of the
analog reference probe used for the procedure.
Analog to Digital Trigger Path Alignment
eld. This fiel
d s hows the relationship of the alignment between the digital and analog trigger paths:
(+) = analog path is rst
(–) = digital path is rst
N/A indicates a n alignment has not been run.
Restoration of Probe Constants
When you reconnect a probe to both the oscilloscope channel and to the same deskew xture channels that the probe was previously calibrated on, the oscilloscope recognizes the probe, recalls the alignment data for that probe/channel combination, and displays a dialog indicating that the stored alignment data is current. Another alignment procedure is not required for this probe/channel combination.
Safety Summary
To avoid potential hazards, use this product only as specied.
To avoid re or personal injury, do not operate in wet/damp conditions.
Keep product surfaces clean and dry.
Environmental Considerations
Equipment Recycling. This product complies with the European Union’s requirements according to D on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektron
irective 2002/96/EC
ix.com).
Warranty Information
For warranty information, go to www.tektronix.com\service, and then use the provided links to search for your product's warranty.
Contacting
Tektronix
Tektronix, Inc. 14200 SW Kar
lBraunDrive PO Box 500 Beaverton,
OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to nd contacts i
n your area.
Clear Trigger Path Alignment button. Click this button to clear the oscilloscope memory of previously-aligned probe data.
Align Trigger Path button. Click this button to start the calibration routine for the probe.
Probe Setup and Probe Properties screens