Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverto
USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, c all 1-800-833-9200.
Worl dwid e, v isit www.tektronix.com to find contacts in your area.
Table of Contents
General safety summary .......... ................................ .................................. ...............ii
Service Safety Summary...................................... .................................. ..................iv
Series Logic Analyzer site considerations ......................................................2
Tektronix Logic Analyzer Family Product Safety & Compliance Instructionsi
General safety summary
General safet
To avoid fire or personal
injury
ysummary
Review the fo
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only quali
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions related to operating the system.
Use proper power cord. Use only the power cord specified for this product and
certified for the country of use.
Use proper voltage setting. Before applying power, ensure that the line selector
is in the proper position for the source being used.
Connect and disconnect properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input o r output
terminals of the product, ensure that the product is properly grounded.
llowing safety precautions to avoid injury and prevent damage to
fied personnel should perform service procedures.
Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the p roduct.
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Connect the probe reference lead to earth ground only.
Power disconnect. The power cord disconnects the product from the power source.
Donotblockthepowercord;itmustremain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Replace batteries properly. Replace batteries only with the specified type and
rating.
Use proper AC adapter. Use only the AC adapter specified for this product.
Use proper fuse. Use only the fuse type and rating specified for this product.
iiTektronix Logic Analyzer Family Product Safety & Compliance Instructions
General safety summary
Terms in this manual
Symbols and terms on the
product
Do not operate i
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide prop
on installing the product so it has proper ventilation.
These terms may appear in this manual:
WAR NI NG .
in injury or loss of life.
CAUTION
damage to this product or other property.
These t
erms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read
the ma
n wet/damp conditions.
er ventilation. Refer to the manual's installation instructions for details
Warning statements identify conditions or practices that could result
. Caution statements identify conditions or practices that could result in
rking.
WARNING indicates an injury hazard not immediately accessible as you
the marking.
read
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
Tektronix Logic Analyzer Family Product Safety & Compliance Instructionsiii
Service Safety Summary
Service Safet
y Summary
Only qualifieSafety Summary and the General Safety Summary before performing any service
procedures.
Do not service alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect power. To avoid electric shock, switch off the instrument p ower, then
disconnect the power cord from the mains power.
Use care when servicing with power on. Dangerous voltages or currents may exist
in this p
test leads before removing protective panels, soldering, or replacing components.
To avoi
d personnel should perform service procedures. Read this Service
roduct. Disconnect power, remove battery (if applicable), and disconnect
d electric shock, do not touch exposed connections.
ivTektronix Logic Analyzer Family Product Safety & Compliance Instructions
Compliance Information
This section lists the EMC (electromagnetic compliance), safety, and
environmental standards with which the instrument complies.
EMC Compliance
EC Declaration of
Conformity
–EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility.
Complianc
e was demonstrated to the following specifications as listed in the
Official Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment
for measurement, control, and laboratory use.
12
CISPR 11:2003. Radiated and conducted emissions, Group A
00-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
3
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, fluctuations, and flicker
European Contact.
Tektronix UK, Ltd.
Western Peninsula
estern Road
W
Bracknell, RG12 1RF
United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic
interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a
test object.
3
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test
levels (IEC 61000-4-11).
Tektronix Logic Analyzer Family Product Safety & Compliance Instructionsv
Compliance Information
Australia / New Zealand
Declaration of
Conformity – EMC
Safety Compliance
EC Declaration of
Conformity – Low Voltage
U.S. Nationally Recognized
Testing Laboratory Listing
Complies with t
he EMC provision of the Radiocommunications Act per the
following standard, in accordance with ACMA:
CISPR 11:2003
. Radiated and Conducted Emissions, Group 1, Class A, in
accordance with EN 61326-1:2006 and EN 61326-2-1:2006.
Compliance was demonstrated to the following specification as listed in the
Official Journal of the European Communities:
Low Voltage Directive 2006/95/EC.
EN 61010-1: 2001. Safety requirements for electrical equipment for
measurement control and laboratory use.
UL 61010
equipment.
1
6 Benchtop Mainframes comply with UL61010B-1:2003, 1
TLA701
and test equipment.
-1:2004, 2
1
nd
Edition. Standard for electrical measuring and test
st
edition, Standard for electrical measuring
Canadian Certification
Additional Compliances
Equipment Type
Safety Class
Safety Certification of
Plug-in or VXI Modules
CAN/CS
equipment for measurement, control, and laboratory use. Part 1.
1
TLA70
Safety requirements for measurement, control, and laboratory use. Part 1.
A-C22.2 No. 61010-1:2004. Safety requirements for electrical
16 Benchtop Mainframes comply with CAN/CSA C22.2 No. 61010.1–97, 1
st
edition, Second Amendment.
1
IEC 61010-1: 2001. Safety requirements for electrical equipment for
urement, control, and laboratory use.
meas
Test and measuring equipment.
Class 1 – grounded product.
e safety certification is valid only when installed in an appropriately approved
Th
(by a USA NRTL or a Canada Certified Organization) mainframe.
viTektronix Logic Analyzer Family Product Safety & C ompliance Instructions
Compliance Information
Pollution Degree
Description
Pollution Degree
A measure of the
and within a product. Typically the internal environment inside a product is
considered to be the same as the external. Products should be used only in the
environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs.
Products in this category are generally encapsulated, hermetically sealed, or
located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs.
Occasionally a temporary conductivity that is caused by condensation must
be expected. This location is a typical office/home environment. Temporary
condensa
Pollution Degree 3 . Conductive pollution, or dry, nonconductive pollution
that bec
where neither temperature nor humidity is controlled. The area is protected
from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through
conductive dust, rain, or snow. Typical outdoor locations.
tion Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor use only.
Pollu
contaminants that could occur in the environment around
tion occurs only when the product is out of service.
omes conductive due to condensation. These are sheltered locations
Installation (Overvoltage)
Category Descriptions
Overvoltage category
Terminals on this product may have different installation (overvoltage) category
designations. The installation categories are:
Measurement Category IV. For measurements performed at the source of
low-voltage installation.
Measurement Category III. For measurements performed in the building
installation.
Measurement Category II. For measurements performed on circuits directly
connected to the low-voltage installation.
Measurement Cate gory I. For measurements performed on circuits not
directly connected to MAINS.
Overvoltage Category II (as defined in IEC 61010-1)
Tektronix Logic Analyzer Family Product Safety & Compliance Instructionsvii
Compliance Information
Environmental considerations
This section provides information about the environmental impact of the product.
Product end-of-life
handling
Observe the f
ollowing guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and
use of natural resources. The equipment may contain substances that could be
harmful to the environment or human health if improperly handled at the product’s
end of life. In order to avoid release of such substances into the environment and
to reduce t
he use of natural resources, we encourage you to recycle this product
in an appropriate system that will ensure that most of the materials are reused or
recycled appropriately.
This symbol indicates that this product complies with the applicable European
Union re
on waste electrical and electronic equipment (WEEE) and batteries. For
information about recycling options, check the Support/Service section of the
Tektr on
quirements according to Directives 2002/96/E C and 2006/66/EC
ixWebsite(www.tektronix.com).
Mercury notification. This product uses an LCD backlight lamp that contains
mercury. Disposal may be regulated due to environmental considerations. Please
ct your local authorities or, within the United States, refer to the E-cycling
conta
Central Web page (www.eiae.org) for disposal or recycling information.
1
rcury notification does not apply to the TLA7016 Benchtop Mainframes.
The Me
1
striction of hazardous
Re
substances
Perchlorate materials. This product contains one or more type CR lithium
batteries. According to the state of California, CR lithium batteries are
classified as perchlorate materials and require special handling. See
www.dtsc.ca.gov/hazardouswaste/perchlorate for additional information.
2
The perchlorate materials information does not apply to the TLA6400 series products (TLA6401, TLA6402,
TLA6403, TLA6404)
2
This product has been classified as Monitoring and Control equipment, and is
outside the scope of the 2002/95/EC RoHS Directive.
viiiTektronix Logic Analyzer Family Product Safety & Compliance Instructions
Preface
Preface
This documen
t contains safety, environmental, and compliance information for
your Tektronix TLA6400 and TLA7000 Series Logic Analyzer products.
To prevent p
ersonal injury or damage, consider the following requirements before
starting service :
The proced
ures in this manual should be performed only by qualified service
personnel.
Read the Ge
neral Safety Summary and Service Safety Summary found at
the beginning of this manual.
For deta
iled installation information, refer to the installation manual specific
to your product, available on the TLA Documentation CD or on the Tektronix
Web site (www.tektronix.com/manuals).
Related documentation
ItemPurpose
TLA Quick Start User Manuals
Online Help
Installation Reference SheetsHigh-level installation information
Installation Manuals
XYZs of Logic Analyzers
Declassification and Securities instructionsData security concerns specific to sanitizing
Optional Service ManualsSelf-service documentation for modules and
High-level operational overview
In-depth operation and UI help
Detailed first-time installation information
Logic analyzer basics
or removing memory devices from Tektronix
products
Collection of logic analyzer application
specific notes
TLA Product specifications and performance
verification procedures
Upgrade information for your logic analyzer
mainframes
Tektronix Logic Analyzer Family Product Safety & Compliance Instructionsix
Preface
xTektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Read this section before installing the instrument. This section describes
environmental and power requirements for your logic analyzer product.
Verify that you have received all of the parts of your instrument and the following:
Correct power cords for your geographical area
CAUTION. This notice applies only to the TLA7016 Benchtop Mainframe with
Option A6.
Aspecialh
igh-current power cord set is provided and is for use exclusively with
this product.
Correct probes and modules
Standard accessories
All optional accessories that you ordered
TLA6400 Series Logic Analyzer environmental requirements
The following table describes the environmental requirements for the TLA6400
Series Logic Analyzer.
CAUTION. Allow a 15.3 cm (6-in) clearance at the back and sides of the instrument
low proper cooling. Avoid blocking any exhaust or intake fans or vents when
to al
usingtheinstrumentonacartorinaninstrumentrack. Avoidplacingloose
paper or other material near the bottom intake vents that could suck against the
instrument and block air flow. Inadequate clearances can damage the instrument..
Table 1: Environmental considerations
FeatureDescription
mperature
Te
Humidity
Altitude
Operating
Nonoperating
Operating20% to 80% relative humidity, non-condensing
Nonoperating
OperatingTo 3000 m (9842.5 ft.)
Nonoperating
+5 °C to +40 °C
–20 °C to +60 °C
8% to 80% relative humidity, non-condensing
To 12,000 m (39,370 ft.)
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions1
Operating requirements
TLA6400 Serie
s Logic Analyzer site considerations
Use the TLA6400 Series Logic Analyzer on a bench or on a cart in the normal
position (on the bottom feet). The logic analyzer also mounts in an instrument
rack.
Table 2: TLA6400 Series Logic Analyzer power requirements
FeatureDescription
Voltage range and
frequency
Power consumption400 W maximum
100 to 240 VAC ±10%, 50 Hz to 60 Hz ±5%
TLA7000 Series Logic Analyzer environmental requirements
The following table describes the environmental requirements for the TLA7000
Series Logic Analyzer.
CAUTION. For the TLA7012 Portable Mainframe, allow a 15.3 cm (6-in)
clearance at the top, back, and sides of the instrument to ensure proper cooling.
Keep the bottom of the instrument clear. Avoid blocking any exhaust fans or vents
when u
sing the instrument on a cart or in an instrument rack.
For the TLA7016 Benchtop Mainframe and the TLA7PC1 Benchtop Controller
low a 5.1 cm (2-in) clearance at the top, back, and sides of the instrument
PC, al
to ensure proper cooling.
he TL708EX TekLink 8-Port Hub allow a 15.3 cm (6-in) clearance at the
For t
back of the instrument to ensure proper cooling.
dequate clearances can cause the instrument to overheat and s hut down.
Ina
2Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Table 3: TLA700
0 Series Logic Analyzer environmental requirements
FeatureDescription
Temperature
Humidity
20% to 80%
Altitude
Operating
Nonoperating
+5 °C to +45 °C
–20 °C to +60 °C
Operating≤30 °C; 80% relative humidity (29 °C maximum wet bulb
temperature)
Nonoperating
Operating
8% to 80% (29 °C maximum wet bulb temperature)
To 3000 m (9843 ft.)
and
nonoperating
1
The operating temperature for the TL708EX TekLink 8-Port Hub is 0 °C to +50 °C
2
The operat
3000 m (9843 ft.)
ing humidity for the TL708EX TekLink 8-Port Hub is 5% to 95% relative humidity at an altitude up to
TLA7012 Portable Mainframe site considerations
Use the portable mainframe on a bench or on a cart in the normal position (on the
bottom feet). The front feet extend to give a better view of the instrument display.
The mainframe also mounts in an instrument rack.
Table 4: TLA7012 Portable Mainframe power requirements
1
2
FeatureDescription
Voltage range and
frequency
100 V
115 V
to 240 V
RMS
±10%, 400 Hz
RMS
Power consumption750 W maximum
TLA7016 Benchtop Mainframe site considerations
LA7016 Benchtop Mainframe is designed to operate on a bench, on a cart,
The T
or in a rackmount environment. If you need to stack more than two benchtop
mainframes, install the mainframes in a rack.
WAR NI NG . To avoid personal injury, never lift or move a benchtop mainframe
by yourself. The size and weight of the mainframe requires two people to lift or
move it.
±10%,50Hzto60Hz
RMS
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions3
Operating requirements
Do not stack mor
e than one benchtop mainframe on top of another benchtop
mainframe. Always use a rackmount kit to ensure that the mainframes are secure
and will not fall.
NOTE. The ratings in the following table are nominal line voltages and power
going into the power supply.
Table 5: TLA7016 Benchtop mainframe power requirements
FeatureDescription
Voltage range and
frequency
Power consumption1450 W maximum
CAUTION
. By placing too many modules in a TLA7016 Benchtop Mainframe,
100 V
120 V
115
RMS
to 120 V
RMS
to 240 V
RMS
, 400 Hz
, 50 Hz to 60 Hz, 1450 W maximum
RMS
, 50 Hz to 60 Hz, 1900 W maximum
RMS
you can overload the power distribution system and damage the mainframe
or modules. To avoid this, ensure that your module configurationinasingle
mainframe does not exceed 1000 W at line voltages less than 120 V nominal at
50 Hz or 60 Hz or at line voltages of 115 V at 400 Hz. Use the following table and
illustration to calculate the total power for your mainframe.
The power consumption depends on the number and type of modules installed
in the mainframes. To determine the total power consumption, perform the
owing steps:
foll
NOTE. The following calculations are based on the power from the power supply,
module power consumption, and worse case line voltages (nominal minus 10%
low-line).
1. Use the following table to determine the power consumption for each module.
dd the power for each module to determine the total power consumption
2.A
(include 170 W for each benchtop mainframe).
3. Ensure that the total power consumption is less than the limits shown in the
following table and illustration. (See Figure 1 on page 6.)
4Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Table 6: Powe r f
Module type
One benchtop mainframe
TLA7AA145
TLA7AA2, TL
TLA7AA3, TL
TLA7AA4, T
TLA7BB281
TLA7BB3119
TLA7BB4, TLA7BC4
TLA7NA145
TLA7NA258
TLA7NA371
TLA7NA
TLA7Q2
TLA7Q
TLA7P250
TLA7P474
TLA7
TLA7N258
TLA7N371
TLA
TLA7L1
TLA7L273
TL
TLA7L4109
TLA7M1
T
TLA7M399
TLA7M4116
TLA7D180
TLA7D2111
TLA7E190
TLA7E2121
TLA7PG2
TLA7S08
TLA7S16
4
4
N1
7N4
A7L3
LA7M2
or instrument modules
1
A7AB2 TLA7AC2
A7AC3
LA7AB4, TLA7AC4
Power (Watts)
170
65
85
105
158
82
51
75
45
82
55
94
7
5
6
7
110
65
105
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions5
Operating requirements
Table 6: Power for instrument modules (cont.)
Module type
TLA7SA08
TLA7SA16
1
Power for mainframe and controller module with fans operating at maximum speed.
Power (Watts)
90
120
Figure 1: Maximum power allowed to modules at various line voltages for
TLA7016 mainframes serial number B020000 and higher
6Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Support Hardw
aresiteconsiderations
The TLA7PC1 Benchtop PC Controller, TL708EX Hub, and GbE switch units
can operate separately on your workbench, equipment rack, or cart. You can also
mount them to
benchtop mainframe.
Table 7: TLA7PC1 Benchtop PC Controller power requirements
FeatureDescription
Voltage ra
frequency
Power cons
Table 8: TL708EX Hub power requirements
FeatureDescription
Voltage range and
frequency
Power consumption110 W maximum
your benchtop mainframe using the brackets provided with the
nge and
umption
100 V
350 W maxim
100 V
RMS
RMS
to 240 V
to 240 V
±10%,50Hzto60Hz
RMS
um
±10%,47Hzto63Hz
RMS
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions7
Operating requirements
8Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
TLA6400 Series basic installation
For detailed installation information, refer to the TLA6400 Series Logic Analyzer
Installation Manual available on the TLA Documentation CD or on the Tektronix
Web site (www
TLA6400 chassis ground connections
.tektronix.com/manuals).
Use the cha
(system-under-test) to the logic analyzer to verify a common ground connection
between instruments.
CAUTION. To reduce the risk of ground-loop noise, ground all of the instruments
in the system to the logic analyzer using the ground connections shown.
ssis ground connections to connect the grounds of the target system
gure 2: TLA6400 accessories and ground connection location
Fi
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions9
TLA6400 Series basic installation
Connect the ac
cessories
Connect the accessories such as an external monitor, keyboard, and printer. (See
Figure 2.)
Table 9: Add
ItemDescriptio
Monitor
LAN
Printer
itional TLA6400 accessory connection information
n
There are tw
(digital/analog out). Simulscan mode displays the same information on
the internal display and the external monitor connected to either Primary
or Seconda
To change the display settings, right-click the mouse on the desktop,
and select Properties > Settings > Advanced > Intel Graphics. Note
that some
external monitor.
To connect a VGA monitor to the DVI-I connector, use a VGA-to-DVI
adapter.
end of this manual.(See page 14, TLA7000 accessory connections.) If
you use a nonstandard monitor, you may need to change the Windows
display
Connect
connectors. Remotely control the logic analyzer through a LAN switch
using a PC (loaded with the TLA application software).
The instruments send printer information to the USB ports. Use any of
the fo
o display ports on the logic analyzer: VGA, and DVI-I
ry.
of the display settings may not function until you connect an
Part numbers for accessories and options are listed at the
settings to obtain the proper resolution.
the logic analyzer to your network through the Gbit LAN
ur US B ports for your printer.
10Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
TLA7000 Series basic installation
TLA7000 Serie
s basic installation
The TLA7000 S
the installation instructions for instruments in a network environment are beyond
the scope of this document.
For detailed network installation information, refer to the TLA7000 Series LogicAnalyzer Installation Manual available on the TLA Documentation CD or on the
Tektronix Web site (www.tektronix.com/manuals).
eries products are normally used in an network environment and
TLA7000 chassis ground connections
Use the chassis ground connections to connect the grounds of the target system to
the logic analyzer to ensure a common ground connection between instruments.
(See Figure 3 on page 11.)
CAUTIO
in the system to the logic analyzer mainframe using the ground connections shown.
N. To reduce the risk of ground-loop noise, ground all of the instruments
gure 3: Location of the ground connection on the TLA7000 logic analyzers
Fi
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions11
TLA7000 Series basic installation
Installing mo
dules in TLA7000 Series mainframes
TLA700 Modules or TLA7000 Modules are fully compatible with the TLA7000
Series mainframes.
CAUTION. To
modules while the instrument is powered on. Always power off the instrument
before installing or removing modules.
Merge modules together to create wider modules. If your application requires you
to merge modules, do so before installing the modules in the mainframe.
Use a screwdriver to tighten the retaining screws to 2.5 in-lbs after seating the
modules in place. (See Figure 4 on page 12.)
avoid damaging the instrument, do not install or remove any
Figure 4: Installing modules
12Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
TLA7000 Series basic installation
If you have any u
covers to meet EMC and cooling specifications. Install a blank-slot panel cover
for the empty slot in the portable mainframe. (See Figure 5.) Install the blank
slot-panels on the benchtop mainframe. (See Figure 6.) Install the blank-slot
panel covers before you install any modules. Use only Tektronix TLA covers;
otherwise the mainframe may not meet EMC and cooling requirements.
Make sure that the EMI shielding is in contact with the adjacent panel or module
cover, and that the airflow shutter activation arms protrude through the holes
in the blan
nused (empty) slots in your mainframe, install blank-slot panel
k-slot panel.
Figure 5: Installing panel covers on the portable mainframe
Figure 6: Installing panel covers on the benchtop mainframe
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions13
TLA7000 Series basic installation
TLA7000 accessory connections
After installing the mainframes and modules, connect the accessories such as
external monitors, keyboard, a nd printer.
Connect the accessories to the TLA7012 Portable Mainframe. (See Figure 7.)
If you have a TLA7016 Benchtop Mainframe, connect the accessories to the
external PC; connect other cables and connectors to the mainframe. (See Figure 8
on page 15.)
Figure 7: TLA7012 accessories connections
14Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Figure 8: TLA7016 accessories connections
TLA7000 Series basic installation
Connect additional accessories as needed. (See Table 9.)
Table 10: Additional accessory connection information
ItemDescription
Monitor
LAN
Printer
There are two display ports on the TLA7012: Primary (DVI-D, digital output only), and Secondary (DVI-I,
l/analog out). Simulscan mode displays the same information on the internal display and the
digita
external monitor connected to either Primary or Secondary.
To change the display settings, right-click the mouse on the desktop, select Properties, and go to the
ngs tab. Select the settings for your monitor. Note that some of the display settings may not
Setti
function until you connect an external monitor.
Use the VGA-to-DVI display adapter to connect a VGA monitor to the Secondary DVI-D output. If
se a nonstandard monitor, you may need to change the Windows display settings to achieve
you u
the proper resolution.
Connect the mainframes to your network through the Gbit LAN connectors. You can remotely control
the mainframes through a LAN switch using your PC (loaded with the TLA application software), or the
icated PC controller (TLA7PC1).
ded
instruments send printer information to the USB ports. Use any of the four USB ports for your printer.
The
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions15
Prepare the instrument for the first time operation
Prepare the in
strument for the first time operation
Connect the k
the logic analyzer. Complete the following steps to turn on the logic analyzer
for the first time:
1. Connect the power cord to your instrument.
2. If you have
monitor.
3. Refer to on
switch for your instrument.
eyboard, mouse, and other accessories before applying power to
an external monitor, connect the power cord and turn on the
e of the following illustrations, if needed, to locate the On/Standby
Figure 9: TLA6400 On/Standby switch location
Figure 10: TLA7000 On/Standby switch locations
16Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Prepare the instrument for the first time operation
4. Turn on the logi
a. Press the On/Standby switch on the front of the instrument.
b. Wait for the logic analyzer to complete the power-on self-tests and to
start Windows.
WAR NI NG . To
instrument or open the instrument to perform any internal cleaning procedures
that could allow moisture to enter while the power cord is connected. Always
remove the power cord to disconnect the instrument from the mains supply before
opening the instrument.
To power off the logic analyzer, press the On/Standby switch. The logic analyzers
have a built-in soft power-off function that safely powers off the instruments when
you press the On/Standby switch. Expansion mainframes automatically power off
when you
turn off the benchtop or portable mainframe.
c analyzer as follows:
avoid personal injury, do not remove any modules from the
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions17