Tektronix Logic Analyzer Family Combined User/Service

xx
Tektronix Logic Analyzer Family
ZZZ
Product Safety & Compliance
Instructions
*P071259103*
071-2591-03
xx
Tektronix Logic Analyzer Family
ZZZ
Product Safety & Compliance
Instructions
071-2591-03
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, c all 1-800-833-9200. Worl dwid e, v isit www.tektronix.com to nd contacts in your area.
Table of Contents
General safety summary .......... ................................ .................................. ............... ii
Service Safety Summary...................................... .................................. .................. iv
Compliance Information........................................................................................... v
EMC Complia
Safety Compliance............................................................................................ vi
Environmental considerations.............................................................................. viii
Preface .............................................................................................................. ix
Operating requirements ........... ................................ .................................. ............... 1
TLA6400 Series Logic Analyzer environmental requirements.................... ....................... 1
TLA6400
TLA7000 Series Logic Analyzer environmental requirements.................... ....................... 2
TLA7012 Portable Mainframe site considerations.............. ................................ ........... 3
TLA7016 Benchtop Mainframe site considerations .............. .................................. ....... 3
Support Hardware site considerations .................... ................................ ................... 7
TLA6400 Series basic installation................................................................................ 9
400 chassis ground connections...................................... ................................. 9
TLA6
Connect the accessories ...................................................................................... 10
TLA7000 Series basic installation............................................................................... 11
TLA7000 chassis ground connections................................ ................................ ...... 11
Installing modules in TLA7000 Series mainframes ... . ..... . ..... . ..... . ... . . . .... . . .... . ..... . ..... . ... 12
TLA7000 accessory connections.................................... .................................. ...... 14
epare the instrument for the rst time operation................. ................................ ............ 16
Pr
nce............... ................................ ................................ ............... v
Series Logic Analyzer site considerations ...................................................... 2
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions i
General safety summary
General safet
To avoid re or personal
injury
ysummary
Review the fo this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only quali
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
Use proper power cord. Use only the power cord specied for this product and certied for the country of use.
Use proper voltage setting. Before applying power, ensure that the line selector is in the proper position for the source being used.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input o r output terminals of the product, ensure that the product is properly grounded.
llowing safety precautions to avoid injury and prevent damage to
ed personnel should perform service procedures.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the p roduct.
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Connect the probe reference lead to earth ground only.
Power disconnect. The power cord disconnects the product from the power source. Donotblockthepowercord;itmustremain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Replace batteries properly. Replace batteries only with the specied type and rating.
Use proper AC adapter. Use only the AC adapter specied for this product.
Use proper fuse. Use only the fuse type and rating specied for this product.
ii Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
General safety summary
Terms in this manual
Symbols and terms on the
product
Do not operate i
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide prop
on installing the product so it has proper ventilation.
These terms may appear in this manual:
WAR NI NG .
in injury or loss of life.
CAUTION
damage to this product or other property.
These t
erms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the ma
n wet/damp conditions.
er ventilation. Refer to the manual's installation instructions for details
Warning statements identify conditions or practices that could result
. Caution statements identify conditions or practices that could result in
rking.
WARNING indicates an injury hazard not immediately accessible as you
the marking.
read
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions iii
Service Safety Summary
Service Safet
y Summary
Only qualifie Safety Summary and the General Safety Summary before performing any service procedures.
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the instrument p ower, then disconnect the power cord from the mains power.
Use care when servicing with power on. Dangerous voltages or currents may exist in this p test leads before removing protective panels, soldering, or replacing components.
To avoi
d personnel should perform service procedures. Read this Service
roduct. Disconnect power, remove battery (if applicable), and disconnect
d electric shock, do not touch exposed connections.
iv Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Compliance Information
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument complies.
EMC Compliance
EC Declaration of Conformity
–EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Complianc
e was demonstrated to the following specications as listed in the
Ofcial Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment for measurement, control, and laboratory use.
12
CISPR 11:2003. Radiated and conducted emissions, Group A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic eld immunity
IEC 610
00-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
3
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European Contact.
Tektronix UK, Ltd. Western Peninsula
estern Road
W Bracknell, RG12 1RF United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels (IEC 61000-4-11).
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions v
Compliance Information
Australia / New Zealand
Declaration of
Conformity – EMC
Safety Compliance
EC Declaration of
Conformity – Low Voltage
U.S. Nationally Recognized
Testing Laboratory Listing
Complies with t
he EMC provision of the Radiocommunications Act per the
following standard, in accordance with ACMA:
CISPR 11:2003
. Radiated and Conducted Emissions, Group 1, Class A, in
accordance with EN 61326-1:2006 and EN 61326-2-1:2006.
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC.
EN 61010-1: 2001. Safety requirements for electrical equipment for measurement control and laboratory use.
UL 61010 equipment.
1
6 Benchtop Mainframes comply with UL61010B-1:2003, 1
TLA701 and test equipment.
-1:2004, 2
1
nd
Edition. Standard for electrical measuring and test
st
edition, Standard for electrical measuring
Canadian Certication
Additional Compliances
Equipment Type
Safety Class
Safety Certication of
Plug-in or VXI Modules
CAN/CS equipment for measurement, control, and laboratory use. Part 1.
1
TLA70 Safety requirements for measurement, control, and laboratory use. Part 1.
A-C22.2 No. 61010-1:2004. Safety requirements for electrical
16 Benchtop Mainframes comply with CAN/CSA C22.2 No. 61010.1–97, 1
st
edition, Second Amendment.
1
IEC 61010-1: 2001. Safety requirements for electrical equipment for
urement, control, and laboratory use.
meas
Test and measuring equipment.
Class 1 – grounded product.
e safety certication is valid only when installed in an appropriately approved
Th (by a USA NRTL or a Canada Certied Organization) mainframe.
vi Tektronix Logic Analyzer Family Product Safety & C ompliance Instructions
Compliance Information
Pollution Degree
Description
Pollution Degree
A measure of the and within a product. Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical ofce/home environment. Temporary condensa
Pollution Degree 3 . Conductive pollution, or dry, nonconductive pollution that bec where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
tion Degree 2 (as dened in IEC 61010-1). Note: Rated for indoor use only.
Pollu
contaminants that could occur in the environment around
tion occurs only when the product is out of service.
omes conductive due to condensation. These are sheltered locations
Installation (Overvoltage)
Category Descriptions
Overvoltage category
Terminals on this product may have different installation (overvoltage) category designations. The installation categories are:
Measurement Category IV. For measurements performed at the source of low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Cate gory I. For measurements performed on circuits not directly connected to MAINS.
Overvoltage Category II (as dened in IEC 61010-1)
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions vii
Compliance Information
Environmental considerations
This section provides information about the environmental impact of the product.
Product end-of-life
handling
Observe the f
ollowing guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. In order to avoid release of such substances into the environment and to reduce t
he use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union re on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Support/Service section of the Tektr on
quirements according to Directives 2002/96/E C and 2006/66/EC
ixWebsite(www.tektronix.com).
Mercury notication. This product uses an LCD backlight lamp that contains mercury. Disposal may be regulated due to environmental considerations. Please
ct your local authorities or, within the United States, refer to the E-cycling
conta Central Web page (www.eiae.org) for disposal or recycling information.
1
rcury notication does not apply to the TLA7016 Benchtop Mainframes.
The Me
1
striction of hazardous
Re
substances
Perchlorate materials. This product contains one or more type CR lithium
batteries. According to the state of California, CR lithium batteries are classied as perchlorate materials and require special handling. See www.dtsc.ca.gov/hazardouswaste/perchlorate for additional information.
2
The perchlorate materials information does not apply to the TLA6400 series products (TLA6401, TLA6402, TLA6403, TLA6404)
2
This product has been classied as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS Directive.
viii Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Preface
Preface
This documen
t contains safety, environmental, and compliance information for
your Tektronix TLA6400 and TLA7000 Series Logic Analyzer products.
To prevent p
ersonal injury or damage, consider the following requirements before
starting service :
The proced
ures in this manual should be performed only by qualied service
personnel.
Read the Ge
neral Safety Summary and Service Safety Summary found at
the beginning of this manual.
For deta
iled installation information, refer to the installation manual specific to your product, available on the TLA Documentation CD or on the Tektronix Web site (www.tektronix.com/manuals).
Related documentation
Item Purpose
TLA Quick Start User Manuals
Online Help
Installation Reference Sheets High-level installation information
Installation Manuals
XYZs of Logic Analyzers
Declassication and Securities instructions Data security concerns specic to sanitizing
Application notes
Product Specications & Performance Verication Procedures
Field upgrade kits
Optional Service Manuals Self-service documentation for modules and
High-level operational overview
In-depth operation and UI help
Detailed rst-time installation information
Logic analyzer basics
or removing memory devices from Tektronix products
Collection of logic analyzer application specic notes
TLA Product specications and performance verication procedures
Upgrade information for your logic analyzer
mainframes
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions ix
Preface
x Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Read this section before installing the instrument. This section describes environmental and power requirements for your logic analyzer product.
Verify that you have received all of the parts of your instrument and the following:
Correct power cords for your geographical area
CAUTION. This notice applies only to the TLA7016 Benchtop Mainframe with
Option A6.
Aspecialh
igh-current power cord set is provided and is for use exclusively with
this product.
Correct probes and modules
Standard accessories
All optional accessories that you ordered
TLA6400 Series Logic Analyzer environmental requirements
The following table describes the environmental requirements for the TLA6400 Series Logic Analyzer.
CAUTION. Allow a 15.3 cm (6-in) clearance at the back and sides of the instrument
low proper cooling. Avoid blocking any exhaust or intake fans or vents when
to al usingtheinstrumentonacartorinaninstrumentrack. Avoidplacingloose paper or other material near the bottom intake vents that could suck against the instrument and block air ow. Inadequate clearances can damage the instrument..
Table 1: Environmental considerations
Feature Description
mperature
Te
Humidity
Altitude
Operating
Nonoperating
Operating 20% to 80% relative humidity, non-condensing
Nonoperating
Operating To 3000 m (9842.5 ft.)
Nonoperating
+5 °C to +40 °C
–20 °C to +60 °C
8% to 80% relative humidity, non-condensing
To 12,000 m (39,370 ft.)
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 1
Operating requirements
TLA6400 Serie
s Logic Analyzer site considerations
Use the TLA6400 Series Logic Analyzer on a bench or on a cart in the normal position (on the bottom feet). The logic analyzer also mounts in an instrument rack.
Table 2: TLA6400 Series Logic Analyzer power requirements
Feature Description
Voltage range and frequency
Power consumption 400 W maximum
100 to 240 VAC ±10%, 50 Hz to 60 Hz ±5%
TLA7000 Series Logic Analyzer environmental requirements
The following table describes the environmental requirements for the TLA7000 Series Logic Analyzer.
CAUTION. For the TLA7012 Portable Mainframe, allow a 15.3 cm (6-in)
clearance at the top, back, and sides of the instrument to ensure proper cooling. Keep the bottom of the instrument clear. Avoid blocking any exhaust fans or vents when u
sing the instrument on a cart or in an instrument rack.
For the TLA7016 Benchtop Mainframe and the TLA7PC1 Benchtop Controller
low a 5.1 cm (2-in) clearance at the top, back, and sides of the instrument
PC, al to ensure proper cooling.
he TL708EX TekLink 8-Port Hub allow a 15.3 cm (6-in) clearance at the
For t back of the instrument to ensure proper cooling.
dequate clearances can cause the instrument to overheat and s hut down.
Ina
2 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Table 3: TLA700
0 Series Logic Analyzer environmental requirements
Feature Description
Temperature
Humidity
20% to 80%
Altitude
Operating
Nonoperating
+5 °C to +45 °C
–20 °C to +60 °C
Operating 30 °C; 80% relative humidity (29 °C maximum wet bulb
temperature)
Nonoperating
Operating
8% to 80% (29 °C maximum wet bulb temperature)
To 3000 m (9843 ft.) and nonoperating
1
The operating temperature for the TL708EX TekLink 8-Port Hub is 0 °C to +50 °C
2
The operat 3000 m (9843 ft.)
ing humidity for the TL708EX TekLink 8-Port Hub is 5% to 95% relative humidity at an altitude up to
TLA7012 Portable Mainframe site considerations
Use the portable mainframe on a bench or on a cart in the normal position (on the bottom feet). The front feet extend to give a better view of the instrument display. The mainframe also mounts in an instrument rack.
Table 4: TLA7012 Portable Mainframe power requirements
1
2
Feature Description
Voltage range and frequency
100 V
115 V
to 240 V
RMS
±10%, 400 Hz
RMS
Power consumption 750 W maximum
TLA7016 Benchtop Mainframe site considerations
LA7016 Benchtop Mainframe is designed to operate on a bench, on a cart,
The T or in a rackmount environment. If you need to stack more than two benchtop mainframes, install the mainframes in a rack.
WAR NI NG . To avoid personal injury, never lift or move a benchtop mainframe
by yourself. The size and weight of the mainframe requires two people to lift or move it.
±10%,50Hzto60Hz
RMS
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 3
Operating requirements
Do not stack mor
e than one benchtop mainframe on top of another benchtop mainframe. Always use a rackmount kit to ensure that the mainframes are secure and will not fall.
NOTE. The ratings in the following table are nominal line voltages and power
going into the power supply.
Table 5: TLA7016 Benchtop mainframe power requirements
Feature Description
Voltage range and frequency
Power consumption 1450 W maximum
CAUTION
. By placing too many modules in a TLA7016 Benchtop Mainframe,
100 V
120 V
115
RMS
to 120 V
RMS
to 240 V
RMS
, 400 Hz
, 50 Hz to 60 Hz, 1450 W maximum
RMS
, 50 Hz to 60 Hz, 1900 W maximum
RMS
you can overload the power distribution system and damage the mainframe or modules. To avoid this, ensure that your module congurationinasingle mainframe does not exceed 1000 W at line voltages less than 120 V nominal at 50 Hz or 60 Hz or at line voltages of 115 V at 400 Hz. Use the following table and illustration to calculate the total power for your mainframe.
The power consumption depends on the number and type of modules installed in the mainframes. To determine the total power consumption, perform the
owing steps:
foll
NOTE. The following calculations are based on the power from the power supply,
module power consumption, and worse case line voltages (nominal minus 10% low-line).
1. Use the following table to determine the power consumption for each module.
dd the power for each module to determine the total power consumption
2.A
(include 170 W for each benchtop mainframe).
3. Ensure that the total power consumption is less than the limits shown in the
following table and illustration. (See Figure 1 on page 6.)
4 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Table 6: Powe r f
Module type
One benchtop mainframe
TLA7AA1 45
TLA7AA2, TL
TLA7AA3, TL
TLA7AA4, T
TLA7BB2 81
TLA7BB3 119
TLA7BB4, TLA7BC4
TLA7NA1 45
TLA7NA2 58
TLA7NA3 71
TLA7NA
TLA7Q2
TLA7Q
TLA7P2 50
TLA7P4 74
TLA7
TLA7N2 58
TLA7N3 71
TLA
TLA7L1
TLA7L2 73
TL
TLA7L4 109
TLA7M1
T
TLA7M3 99
TLA7M4 116
TLA7D1 80
TLA7D2 111
TLA7E1 90
TLA7E2 121
TLA7PG2
TLA7S08
TLA7S16
4
4
N1
7N4
A7L3
LA7M2
or instrument modules
1
A7AB2 TLA7AC2
A7AC3
LA7AB4, TLA7AC4
Power (Watts)
170
65
85
105
158
82
51
75
45
82
55
94
7
5
6
7
110
65
105
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 5
Operating requirements
Table 6: Power for instrument modules (cont.)
Module type
TLA7SA08
TLA7SA16
1
Power for mainframe and controller module with fans operating at maximum speed.
Power (Watts)
90
120
Figure 1: Maximum power allowed to modules at various line voltages for TLA7016 mainframes serial number B020000 and higher
6 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Operating requirements
Support Hardw
aresiteconsiderations
The TLA7PC1 Benchtop PC Controller, TL708EX Hub, and GbE switch units can operate separately on your workbench, equipment rack, or cart. You can also mount them to benchtop mainframe.
Table 7: TLA7PC1 Benchtop PC Controller power requirements
Feature Description
Voltage ra frequency
Power cons
Table 8: TL708EX Hub power requirements
Feature Description
Voltage range and frequency
Power consumption 110 W maximum
your benchtop mainframe using the brackets provided with the
nge and
umption
100 V
350 W maxim
100 V
RMS
RMS
to 240 V
to 240 V
±10%,50Hzto60Hz
RMS
um
±10%,47Hzto63Hz
RMS
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 7
Operating requirements
8 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
TLA6400 Series basic installation
For detailed installation information, refer to the TLA6400 Series Logic Analyzer Installation Manual available on the TLA Documentation CD or on the Tektronix
Web site (www
TLA6400 chassis ground connections
.tektronix.com/manuals).
Use the cha (system-under-test) to the logic analyzer to verify a common ground connection between instruments.
CAUTION. To reduce the risk of ground-loop noise, ground all of the instruments
in the system to the logic analyzer using the ground connections shown.
ssis ground connections to connect the grounds of the target system
gure 2: TLA6400 accessories and ground connection location
Fi
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 9
TLA6400 Series basic installation
Connect the ac
cessories
Connect the accessories such as an external monitor, keyboard, and printer. (See Figure 2.)
Table 9: Add
Item Descriptio
Monitor
LAN
Printer
itional TLA6400 accessory connection information
n
There are tw (digital/analog out). Simulscan mode displays the same information on the internal display and the external monitor connected to either Primary or Seconda
To change the display settings, right-click the mouse on the desktop, and select Properties > Settings > Advanced > Intel Graphics. Note that some external monitor.
To connect a VGA monitor to the DVI-I connector, use a VGA-to-DVI adapter. end of this manual.(See page 14, TLA7000 accessory connections.) If you use a nonstandard monitor, you may need to change the Windows display
Connect connectors. Remotely control the logic analyzer through a LAN switch using a PC (loaded with the TLA application software).
The instruments send printer information to the USB ports. Use any of the fo
o display ports on the logic analyzer: VGA, and DVI-I
ry.
of the display settings may not function until you connect an
Part numbers for accessories and options are listed at the
settings to obtain the proper resolution.
the logic analyzer to your network through the Gbit LAN
ur US B ports for your printer.
10 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
TLA7000 Series basic installation
TLA7000 Serie
s basic installation
The TLA7000 S the installation instructions for instruments in a network environment are beyond the scope of this document.
For detailed network installation information, refer to the TLA7000 Series Logic Analyzer Installation Manual available on the TLA Documentation CD or on the Tektronix Web site (www.tektronix.com/manuals).
eries products are normally used in an network environment and
TLA7000 chassis ground connections
Use the chassis ground connections to connect the grounds of the target system to the logic analyzer to ensure a common ground connection between instruments. (See Figure 3 on page 11.)
CAUTIO
in the system to the logic analyzer mainframe using the ground connections shown.
N. To reduce the risk of ground-loop noise, ground all of the instruments
gure 3: Location of the ground connection on the TLA7000 logic analyzers
Fi
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 11
TLA7000 Series basic installation
Installing mo
dules in TLA7000 Series mainframes
TLA700 Modules or TLA7000 Modules are fully compatible with the TLA7000 Series mainframes.
CAUTION. To
modules while the instrument is powered on. Always power off the instrument before installing or removing modules.
Merge modules together to create wider modules. If your application requires you to merge modules, do so before installing the modules in the mainframe.
Use a screwdriver to tighten the retaining screws to 2.5 in-lbs after seating the modules in place. (See Figure 4 on page 12.)
avoid damaging the instrument, do not install or remove any
Figure 4: Installing modules
12 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
TLA7000 Series basic installation
If you have any u covers to meet EMC and cooling specications. Install a blank-slot panel cover for the empty slot in the portable mainframe. (See Figure 5.) Install the blank slot-panels on the benchtop mainframe. (See Figure 6.) Install the blank-slot panel covers before you install any modules. Use only Tektronix TLA covers; otherwise the mainframe may not meet EMC and cooling requirements.
Make sure that the EMI shielding is in contact with the adjacent panel or module cover, and that the airow shutter activation arms protrude through the holes in the blan
nused (empty) slots in your mainframe, install blank-slot panel
k-slot panel.
Figure 5: Installing panel covers on the portable mainframe
Figure 6: Installing panel covers on the benchtop mainframe
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 13
TLA7000 Series basic installation
TLA7000 accessory connections
After installing the mainframes and modules, connect the accessories such as external monitors, keyboard, a nd printer.
Connect the accessories to the TLA7012 Portable Mainframe. (See Figure 7.) If you have a TLA7016 Benchtop Mainframe, connect the accessories to the external PC; connect other cables and connectors to the mainframe. (See Figure 8 on page 15.)
Figure 7: TLA7012 accessories connections
14 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Figure 8: TLA7016 accessories connections
TLA7000 Series basic installation
Connect additional accessories as needed. (See Table 9.)
Table 10: Additional accessory connection information
Item Description
Monitor
LAN
Printer
There are two display ports on the TLA7012: Primary (DVI-D, digital output only), and Secondary (DVI-I,
l/analog out). Simulscan mode displays the same information on the internal display and the
digita external monitor connected to either Primary or Secondary.
To change the display settings, right-click the mouse on the desktop, select Properties, and go to the
ngs tab. Select the settings for your monitor. Note that some of the display settings may not
Setti function until you connect an external monitor.
Use the VGA-to-DVI display adapter to connect a VGA monitor to the Secondary DVI-D output. If
se a nonstandard monitor, you may need to change the Windows display settings to achieve
you u the proper resolution.
Connect the mainframes to your network through the Gbit LAN connectors. You can remotely control the mainframes through a LAN switch using your PC (loaded with the TLA application software), or the
icated PC controller (TLA7PC1).
ded
instruments send printer information to the USB ports. Use any of the four USB ports for your printer.
The
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 15
Prepare the instrument for the rst time operation
Prepare the in
strument for the rst time operation
Connect the k the logic analyzer. Complete the following steps to turn on the logic analyzer for the rst time:
1. Connect the power cord to your instrument.
2. If you have
monitor.
3. Refer to on
switch for your instrument.
eyboard, mouse, and other accessories before applying power to
an external monitor, connect the power cord and turn on the
e of the following illustrations, if needed, to locate the On/Standby
Figure 9: TLA6400 On/Standby switch location
Figure 10: TLA7000 On/Standby switch locations
16 Tektronix Logic Analyzer Family Product Safety & Compliance Instructions
Prepare the instrument for the rst time operation
4. Turn on the logi
a. Press the On/Standby switch on the front of the instrument.
b. Wait for the logic analyzer to complete the power-on self-tests and to
start Windows.
WAR NI NG . To
instrument or open the instrument to perform any internal cleaning procedures that could allow moisture to enter while the power cord is connected. Always remove the power cord to disconnect the instrument from the mains supply before opening the instrument.
To power off the logic analyzer, press the On/Standby switch. The logic analyzers have a built-in soft power-off function that safely powers off the instruments when you press the On/Standby switch. Expansion mainframes automatically power off when you
turn off the benchtop or portable mainframe.
c analyzer as follows:
avoid personal injury, do not remove any modules from the
Tektronix Logic Analyzer Family Product Safety & Compliance Instructions 17