Tektronix Logic Analyzer Family Reference manual

Technical Reference
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures Version 4.0 Software
071-0862-00
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
Copyright © T ektronix, Inc. All rights reserved. Licensed software products are owned by Tektronix or its suppliers and are protected by United States copyright laws and international treaty provisions.
Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in T echnical Data and Computer Software clause at DFARS 252.227-7013, or subparagraphs (c)(1) and (2) of the Commercial Computer Software – Restricted Rights clause at F AR 52.227-19, as applicable.
T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
T ektronix, Inc., 14200 SW Karl Braun Drive, Beaverton, OR 97077 TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.

WARRANTY

T ektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If a product proves defective during this warranty period, T ektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-T ektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUST OMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

General Safety Summary ix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Safety Summary xi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preventing Electrostatic Discharge xiii. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preface xv. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Naming Conventions xvi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Offerings xvi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contacting T ektronix xviii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Introduction
Introduction 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional Verification Procedures 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Certification T est Procedures 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification Procedures 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Procedures 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 1–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Software Installation and Removal Instructions 1–8. . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software 1–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration Data Reports 1–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TLA600 Series Logic Analyzer
TLA600 Series Logic Analyzer Functional Verification 2–1. . . . . . . . . . .
Tests Performed 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedure 2–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Logic Analyzer Probe Functional Verification Procedure 2–4. . . . . . . . . . . . . . . . . .
TLA600 Series Certification 2–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration Data Report 2–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TLA600 Series Logic Analyzer Performance Verification 2–11. . . . . . . . .
Prerequisites 2–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T ests Performed 2–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 2–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedures 2–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software 2–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting 2–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification Tests 2–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Completing the Performance Verification 2–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TLA600 Series Logic Analyzer Adjustment 2–23. . . . . . . . . . . . . . . . . . . . .
Prerequisites 2–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software 2–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 2–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Instructions 2–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T ests Performed 2–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Procedures 2–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Self Calibration 2–27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Completing the Adjustment Steps 2–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
i
Table of Contents
Portable Mainframe
Portable Mainframe Functional Verification 3–1. . . . . . . . . . . . . . . . . . . .
Tests Performed 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedure 3–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Portable Mainframe Certification 3–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration Data Report 3–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Portable Mainframe Performance Verification 3–9. . . . . . . . . . . . . . . . . .
Tests Performed 3–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedures 3–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Portable Mainframe Adjustment 3–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Benchtop Chassis and Expansion Mainframe
Benchtop Chassis and Expansion Mainframe Functional
Verification 4–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Benchtop Chassis 4–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Expansion Chassis 4–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Benchtop Chassis and Expansion Mainframe Certification 4–3. . . . . . . .
Benchtop Chassis and Expansion Mainframe Performance
Verification 4–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedure 4–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Benchtop Chassis and Expansion Mainframe Adjustment 4–7. . . . . . . . .
Benchtop Controller
Logic Analyzer Module
Benchtop Controller Functional Verification 5–1. . . . . . . . . . . . . . . . . . . .
Tests Performed 5–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedure 5–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Benchtop Controller Certification 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration Data Report 5–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Benchtop Controller Performance Verification 5–7. . . . . . . . . . . . . . . . . .
Tests Performed 5–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedures 5–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Benchtop Controller Adjustment 5–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Logic Analyzer Module Functional Verification 6–1. . . . . . . . . . . . . . . . .
T est Equipment 6–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup 6–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Module Self T ests and Power-On Diagnostics 6–2. . . . . . . . . . . . . . . . . . . . . . . . . . .
Single LA Module Functional Verification Procedure 6–3. . . . . . . . . . . . . . . . . . . .
Merged LA Module Functional Verification Procedure 6–4. . . . . . . . . . . . . . . . . . .
Logic Analyzer Probe Functional Verification Procedure 6–5. . . . . . . . . . . . . . . . . .
Logic Analyzer Module Certification 6–7. . . . . . . . . . . . . . . . . . . . . . . . . .
Logic Analyzer Module Performance Verification 6–9. . . . . . . . . . . . . . .
Prerequisites 6–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Merged Modules 6–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T ests Performed 6–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 6–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software 6–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting 6–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification Tests 6–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Completing the Performance Verification 6–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Logic Analyzer Module Adjustment Procedure 6–21. . . . . . . . . . . . . . . . . .
Prerequisites 6–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Merged Modules 6–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software 6–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 6–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Instructions 6–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T ests Performed 6–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Procedures 6–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Self Calibration 6–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Completing the Adjustment Steps 6–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Logic Anlayzer Adjustment/V erification Fixture
Adjustment/Verification Fixture Functional Verification 7–1. . . . . . . . . .
T est Procedure 7–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment/Verification Fixture Certification 7–5. . . . . . . . . . . . . . . . . . .
Calibration Data Report 7–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment/Verification Fixture Performance Verification 7–9. . . . . . . .
Prerequisites 7–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 7–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T ests Performed 7–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Procedures 7–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment/Verification Fixture Adjustment 7–21. . . . . . . . . . . . . . . . . . . .
Adjustment Procedure 7–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment/Verification Fixture Adjustment 7–22. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table of Contents
Oscilloscope Module
DSO Module Functional Verification 8–1. . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 8–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup 8–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Module Self T ests and Power-On Diagnostics 8–2. . . . . . . . . . . . . . . . . . . . . . . . . . .
Extended Diagnostics 8–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Self Cal 8–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Calibration 8–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DSO Module Certification 8–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DSO Module Performance Verification 8–7. . . . . . . . . . . . . . . . . . . . . . . .
Prerequisites 8–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software 8–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 8–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification Instructions 8–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification Tests 8–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Completing the Performance Verification 8–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
iii
Table of Contents
DSO Module Adjustment 8–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Prerequisites 8–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software 8–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 8–27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Instructions 8–28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Procedures 8–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Completing the Adjustment Steps 8–38. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pattern Generator Module
Pattern Generator Module Functional Verification 9–1. . . . . . . . . . . . . . .
T est Equipment 9–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup 9–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Module Self T ests and Power-On Diagnostics 9–4. . . . . . . . . . . . . . . . . . . . . . . . . . .
Extended Diagnostics 9–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Software Setup Files 9–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting 9–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional Verification Tests 9–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pattern Generator Module Certification 9–29. . . . . . . . . . . . . . . . . . . . . . .
Pattern Generator Module Performance Verification 9–31. . . . . . . . . . . . .
Prerequisites 9–31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T ests Performed 9–31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T est Equipment 9–31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification Tests 9–32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pattern Generator Module Adjustment Procedure 9–37. . . . . . . . . . . . . . .
Appendices
Appendix A: Specifications A–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix B: Replaceable Parts B–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Parts Ordering Information B–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Replaceable Parts List B–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iv
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

List of Figures

Table of Contents
Figure 1–1: Calibration/certification procedure flow chart 1–4. . . . . . . .
Figure 2–1: Probe functional verification test setup 2–4. . . . . . . . . . . . . .
Figure 2–2: Activity Monitor 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–3: P6434 probe detail 2–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–4: Adjustment/verification fixture connections and jumper
locations 2–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–5: FPV_DC_Threshold test setup 2–18. . . . . . . . . . . . . . . . . . . . .
Figure 2–6: Initial FPV_Setup_0F test setup 2–19. . . . . . . . . . . . . . . . . . . .
Figure 2–7: Initial FPV_Hold_0F test setup 2–20. . . . . . . . . . . . . . . . . . . . .
Figure 2–8: FPV_Maxsync test setup 2–21. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–9: P6434 probe detail 2–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–10: Initial deskew test setup 2–28. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–1: Location of J1 and J2 pins in the portable mainframe 3–11. .
Figure 4–1: Benchtop and Expansion Chassis 25-pin rear panel
connector 4–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–1: Probe functional verification test setup 6–5. . . . . . . . . . . . . .
Figure 6–2: Activity Monitor 6–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–3: P6434 probe detail 6–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–4: Adjustment/verification fixture connections and jumper
locations 6–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–5: FPV_DC_Threshold test setup 6–15. . . . . . . . . . . . . . . . . . . . .
Figure 6–6: Initial FPV_Setup_0F test setup 6–16. . . . . . . . . . . . . . . . . . . .
Figure 6–7: Initial FPV_Hold_0F test setup 6–17. . . . . . . . . . . . . . . . . . . . .
Figure 6–8: FPV_Maxsync test setup 6–18. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–9: P6434 probe detail 6–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–10: Initial deskew test setup 6–28. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 7–1: Probe tip adapter detail 7–10. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 7–2: Adjustment/verification fixture detail 7–12. . . . . . . . . . . . . . . .
Figure 7–3: Adjustment/verification fixture circuit board layout 7–23. . .
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
v
Table of Contents
Figure 8–1: Calibration and Diagnostics property sheet 8–2. . . . . . . . . . .
Figure 8–2: Initial setup for the DC gain accuracy test 8–12. . . . . . . . . . . .
Figure 8–3: Initial setup for the analog bandwidth test 8–14. . . . . . . . . . . .
Figure 8–4: Using a power meter to monitor the amplitude of the
sine wave generator during the analog bandwidth test 8–15. . . . . . . . .
Figure 8–5: Initial setup for the delay between channels test 8–17. . . . . . .
Figure 8–6: Initial setup for the timebase accuracy test 8–18. . . . . . . . . . .
Figure 8–7: Initial setup for the main trigger accuracy test 8–20. . . . . . . .
Figure 8–8: Initial setup for the main trigger sensitivity test 8–21. . . . . . .
Figure 8–9: Initial setup for the probe compensation test 8–22. . . . . . . . . .
Figure 8–10: Initial setup for the base calibration adjustment 8–30. . . . . .
Figure 8–11: Initial setup for the frequency response adjustment 8–32. . . Figure 8–12: Using a power meter to monitor the amplitude of the
sine wave generator during the frequency response adjustment 8–33. Figure 8–13: Initial setup for the channel-to-channel skew
adjustment 8–36. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 8–14: Initial setup for the probe compensation test 8–37. . . . . . . . .
Figure 9–1: Termination Board 9–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9–2: Internal Clock Frequency connections 9–8. . . . . . . . . . . . . . .
Figure 9–3: External Clock Input connections 9–11. . . . . . . . . . . . . . . . . . .
Figure 9–4: Sequence and Data Output connections 9–14. . . . . . . . . . . . . .
Figure 9–5: Timing Chart 9–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9–6: Timing Chart 9–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9–7: Merge operation Connections 9–18. . . . . . . . . . . . . . . . . . . . . .
Figure 9–8: Merged and unmerged modules in the Merge Modules
window 9–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9–9: Timing Chart 9–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9–10: Deskew function connections 9–22. . . . . . . . . . . . . . . . . . . . .
Figure 9–11: Inhibit function connections 9–25. . . . . . . . . . . . . . . . . . . . . . .
Figure 9–12: Maximum Operating Frequency Connections 9–33. . . . . . . .
Figure 9–13: Timing chart for the maximum operating frequency
check 9–34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure B–1: Adjustment/verification fixture exploded view B–5. . . . . . . .
vi
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

List of Tables

Table of Contents
Table 1–1: Levels of testing and adjustment 1–2. . . . . . . . . . . . . . . . . . . .
Table 1–2: Typical order of procedures 1–2. . . . . . . . . . . . . . . . . . . . . . . . .
Table 1–3: Test equipment 1–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 2–1: TLA600 Functional verification procedures 2–1. . . . . . . . . . .
Table 2–2: TLA600 PV/Adjust software performance verification
tests 2–12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 2–3: TLA600 characteristics indirectly checked by the PV/Adjust
software performance verification tests 2–12. . . . . . . . . . . . . . . . . . . . .
Table 3–1: Functional verification procedures 3–1. . . . . . . . . . . . . . . . . .
Table 3–2: Performance verification procedures 3–9. . . . . . . . . . . . . . . .
Table 3–3: Power supply voltages and backplane connector pins 3–11. . .
Table 4–1: Power supply voltages at the 25-pin connector 4–6. . . . . . . . .
Table 5–1: Functional verification procedures 5–1. . . . . . . . . . . . . . . . . .
Table 5–2: Performance verification procedures 5–7. . . . . . . . . . . . . . . .
Table 6–1: Logic Analyzer Module functional verification
procedures 6–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 6–2: LA Module performance verification tests 6–10. . . . . . . . . . . .
Table 6–3: LA Module characteristics indirectly checked by the
performance verification tests 6–11. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 7–1: Adjustment/verification fixture performance
verification tests 7–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 7–2: Adjustment/verification fixture jumper settings 7–12. . . . . . .
Table 8–1: DSO PV/Adjust software performance verification tests 8–8
Table 8–2: DSO module fault isolation 8–10. . . . . . . . . . . . . . . . . . . . . . . . .
Table 8–3: DSO Adjustments and dependencies 8–27. . . . . . . . . . . . . . . . .
Table 8–4: Fault isolation 8–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 9–1: Pattern generator module functional verification
procedures 9–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 9–2: Front panel indicators 9–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 9–3: Extended Diagnostic test Items and faulty component 9–5. .
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
vii
Table of Contents
Table A–1: Adjustment/verification fixture specifications A–1. . . . . . . . .
viii
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
To Avoid Fire or
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use. Connect and Disconnect Properly . Do not connect or disconnect probes or test
leads while they are connected to a voltage source. Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
The common terminal is at ground potential. Do not connect the common terminal to elevated voltages.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product. Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present. Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry .
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
ix
General Safety Summary
Provide Proper Ventilation. Refer to the manual’s installation instructions for details on installing the product so it has proper ventilation.

Symbols and Terms

T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
T erms on the Product. These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the
marking. WARNING indicates an injury hazard not immediately accessible as you read the
marking. CAUTION indicates a hazard to property including the product. Symbols on the Product. The following symbols may appear on the product:
WARNING
High Voltage
x
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Protective Ground
(Earth) T erminal
CAUTION
Refer to Manual
Double
Insulated

Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures.

Do Not Service Alone

Disconnect Power

Use Care When Servicing
With Power On
Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
To avoid electric shock, disconnect the main power by means of the power cord or, if provided, the power switch.
Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xi
Service Safety Summary
xii
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Preventing Electrostatic Discharge

CAUTION. Static discharge can damage internal semiconductor components. Follow the guidelines listed below to avoid product damage.
When performing service that requires internal access to an instrument, adhere to the following precautions to avoid damaging internal modules or their compo­nents:
H Avoid handling modules or components in areas that have floors or work
surfaces capable of generating a static charge. H Spray carpeted work areas with a solution of equal parts of water and fabric
softener. H Wear clothing made from materials that do not accumulate static charges.
Wool (and some artificial fibers) build up static charges readily; cotton
conducts electricity and resists static accumulation.
H Minimize the handling of static-sensitive devices. H Transport and store static-sensitive devices in their protected containers or on
a metal rail. Label any package that contains static-sensitive parts.
H Service instruments and modules at grounded static-free work stations. H Do not allow any device capable of generating a static charge on work
station surfaces.
H Wear a grounding wrist strap while working with static-sensitive devices. H Handle circuit boards by their edges, if possible. H Do not slide static-sensitive components over any surface. H Do not use high-velocity compressed air to clean or dry components or
modules.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xiii
Preventing Electrostatic Discharge
xiv
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Preface

This manual contains performance verification and adjustment information for the Tektronix Logic Analyzer Family hardware components. Each section covers an individual hardware component such as a mainframe, module, or adjustment fixture and includes the following information:
H An introduction that describes the procedures H Equipment lists required to perform the procedures H Functional verification procedures to quickly determine that the component
operates properly (as for a brief incoming inspection)
H Certification test procedures and Calibration Data Reports H Performance verification procedures to ensure the component operates
properly and meets warranted specifications H Adjustment procedures used to adjust the component to meet its warranted
specifications This manual also includes an introduction with important overview information,
and an appendix with a parts list and an exploded view of the Logic Analyzer Adjustment and Verification fixture.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xv
Preface

Naming Conventions

This manual uses the following abbreviated names for the full product name throughout the text:
Full product name Abbreviated name
TLA600 Series Logic Analyzer TLA600

Service Offerings

TLA714 Portable Mainframe TLA720 Benchtop Chassis TLA720 Benchtop Controller benchtop controller TLA7XM Expansion Mainframe expansion mainframe
TLA7Lx, TLA7Mx, TLA7Nx, TLA7Px, or TLA7Qx Logic Analyzer Module
TLA7Dx and TLA7Ex Digital Storage Scope DSO module TLA7PG2 Pattern Generator Module pat gen module Device Under Test / System Under Test DUT / SUT
portable mainframe benchtop chassis
LA module or LA
Throughout this manual, the term “module” refers to a LA module, a DSO module, an expansion module, or a benchtop controller module.
Tektronix provides service to cover repair under warranty as well as other services that are designed to meet your specific service needs.
Whether providing warranty repair service or any of the other services listed below, Tektronix service technicians are well equipped to service the Tektronix Logic Analyzer Family products. Tektronix technicians train on Tektronix products; they have access to the latest information on improvements to the products as well as the latest new product upgrades. Services are pr ovided at
Tektronix Service Centers and on-site at your facility, depending on your location.

Warranty Repair Service

xvi
Tektronix technicians provide warranty service at most Tektronix service locations worldwide. The Tektronix product catalog lists all service locations worldwide or you can visit us on our Customer Services World Center web site at Tektronix.com/Measurement/Service. See our latest service offerings and contact us by email.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Preface
Calibration and Repair
Service
In addition to warranty repair, Tektronix Service offers calibration and other services which provide cost-effective solutions to your service needs and quality­standards compliance requirements. Our instruments are supported worldwide by the leading-edge design, manufacturing, and service resources of Tektronix to provide the best possible service.
The following services can be tailored to fit your requirements for calibration and/or repair of the Tektronix Logic Analyzer Family products.
Service Options. Tektronix Service Options can be selected at the time you purchase your instrument. You select these options to provide the services that best meet your service needs. These service options are listed on the Tektronix Service Options page at the beginning of this manual.
Service Agreements. If service options are not added to the instrument purchase, then service agreements are available on an annual basis to provide calibration services or post-warranty repair coverage for the Tektronix Logic Analyzer Family products. Service agreements may be customized to meet special turn-around time and/or on-site requirements.
Service on Demand. Tektronix also offers calibration and repair services on a per-incident basis that is available with standard prices for many products.
Self Service. Tektronix supports repair to the replaceable-part level by providing for circuit board exchange.
Use this service to reduce down-time for repair by exchanging circuit boards for remanufactured ones. Tektronix ships updated and tested exchange boards. Each board comes with a 90-day service warranty.
For More Information. Contact your local Tektronix service center or sales engineer for more information on any of the Calibration and Repair Services just described.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xvii
Preface

Contacting Tektronix

Phone 1-800-833-9200*
Address Tektronix, Inc.
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
Web site www.tektronix.com
Sales support 1-800-833-9200, select option 1*
Service support 1-800-833-9200, select option 2*
Technical support Email: techsupport@tektronix.com
1-800-833-9200, select option 3* 1-503-627-2400
6:00 a.m. – 5:00 p.m. Pacific time
* This phone number is toll free in North America. After office hours, please leave a
voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.
xviii
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Introduction

Introduction

This manual contains instructions for testing the performance and adjustment of the following instruments:
H TLA600 Series Logic Analyzers H TLA714 Portable Mainframe H TLA720 Benchtop Chassis H TLA720 Benchtop Controller H TLA7Lx, TLA7Mx, TLA7Nx, TLA7Px, and TLA7Qx Logic Analyzer
Modules
H Logic analyzer probes (tested with the associated logic analyzer) H Logic Analyzer Module adjustment/verification fixture H TLA7Dx and TLA7Ex Digitizing Oscilloscope Module H TLA7PG2 Pattern Generator Modules H Pattern generator probes (tested with the associated pattern generator
module)
Four levels of testing are provided:
H Functional verification procedures H Performance verification procedures H Certification test procedures and Calibration Data Reports H Adjustment procedures
Not all of the components of the Tektronix Logic Analyzer Family require all four levels of testing. Table 1–1 on page 1–2 lists the components of the logic analyzer family and the levels of testing described in this manual.
Certain instruments, such as the Termination Board (used with pattern generator modules), do not require any testing or certification.
Table 1–2 lists the typical order of procedures needed for verifying the perfor­mance and adjustment of the instruments. These procedures assume that you are familiar with operating the instruments.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–1
Introduction
T able 1–1: Levels of testing and adjustment
Component/product Functional verification Certification
Performance verification
Adjustment
TLA600 Series Logic
Yes Yes Yes Yes, software controlled
Analyzers TLA714 Portable
Yes Yes Yes No
Mainframe TLA720 Benchtop
Yes No Yes No
Chassis TLA720 Benchtop
Yes Yes Yes No
Controller Logic analyzer modules
Yes Yes, software controlled Y es, software controlled Y es, software controlled
(all versions) Logic Analyzer Adjust-
Yes Yes Yes Yes ment and Verification fixture
TLA7Dx/TLA7Ex
Yes Yes, software controlled Y es, software controlled Y es, software controlled Digitizing Oscilloscope Module
TLA7PG2 Pattern
Yes No Yes No Generator modules
T able 1–2: Typical order of procedures
Instrument type
Mainframes and
Annual performance verification, calibration, re-certification
1
Run performance verification procedure Run performance verification procedure
Adjustment, performance verification, calibration, re-certification
benchtop controller TLA600 Logic
Analyzers and all LA modules
1. Run performance verification procedure
2. If module fails performance verification procedure, then run the adjustment procedure
1. Run adjustment procedure
2. Run performance verification procedure
3. Rerun performance verification if step 2 is done
Logic Analyzer
Run performance verification procedure
2
Run performance verification procedure Adjustment and Verification fixture
Pattern Generator Modules
DSO modules 1. Run performance verification procedure
1
Refer to the appropriate certification sections in this document for all certification procedures
2
The Logic Analyzer Adjustment and Verification fixture requires calibration every two years
1–2
Run the functional verification and the performance verification procedures
2. If module fails performance verification procedure, then run the adjustment procedure
3. Rerun performance verification if step 2 is done
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Run the functional verification and the performance
verification procedures
1. Run performance verification procedure
2. If module fails performance verification procedure,
then run the adjustment procedure
3. Rerun performance verification if step 2 is done
1
following board repair

Functional Verification Procedures

Functional verification procedures verify the basic functionality of the instru­ment inputs, outputs, and basic instrument actions. These procedures include power-on and extended diagnostics, self calibration, as well as semi-automated or manual check procedures. These procedures can be used for incoming inspection purposes.

Certification Test Procedures

Certification procedures are used to certify the accuracy of an instrument and provide a traceability path to national standards. Certification data is recorded on Calibration Data Reports provided with this manual. Calibration Data Reports are produced for the TLA600 Logic Analyzers, LA modules, and DSO modules as an output from the PV/Adjust software.
Introduction

Performance V erification Procedures

Performance verification procedures confirm that a product meets or exceeds the performance requirements for the published specifications documented in the Tektronix Logic Analyzer Family User Manual. Refer to Figure 1–1 on page 1–4 for a graphic overview of the procedures.

Adjustment Procedures

Adjustment procedures check for and, if necessary, correct any adjustment errors discovered when performing functional or performance verification procedures. The adjustment procedures documented in this manual are controlled by software, except for the procedures for the adjustment/verification fixture.
Some procedures require user intervention to move probes or change test equipment settings. The adjustment software is part of the self-calibration routines in the logic analyzer application software and the PV/Adjust software provided with the product CD-ROM.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–3
Introduction
Determine equipment
Gather required test
Yes
certified include an LA
to certify.
Go to appropriate
procedure(s).
equipment.
Does
equipment to be
or DSO module?
No
Load PV/Adjust
software.
Perform procedure
using software.
Allow all mainframes, modules and
equipment to warm up (30 minutes).
Yes
certified an LA or DSO
Is
equipment to be
module?
No
Use
Tektronix-supplied
probes.
Photocopy the required
Calibration Data Report(s).
Enter product model and serial number(s) on the
Calibration Data Report(s).
Is it OK
No
with customer to
remove/reconnect probes
from System
Under T est?
Select procedure.
Configure customer’s
mainframe and modules as
described in procedure.
Set up test equipment.
Yes
Do tests
pass?
Yes
Print Calibration
Data Report.
No
Repair or adjust
equipment as
required.
Repair or adjust
equipment as
required.
Create necessary
Calibration Certificate.
No
all certifications
Are
complete?
Perform procedure.
No
measurements
tolerance?
Enter measurements on
the appropriate
Calibration Data Report.
Are
within
Yes
Figure 1–1: Calibration/certification procedure flow chart
1–4
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Yes
Done

Test Equipment

The procedures in this manual use external, traceable signal sources to directly test characteristics that are designated as checked (n) in the Specifications section of the Tektronix Logic Analyzer Family User Manual. Table 1–3 shows the required equipment list. Always warm up the equipment for 30 minutes before beginning the procedures.
T able 1–3: Test equipment
Item number and description Minimum requirements Example Where used
Introduction
1. Mainframe TLA700 Series Mainframe Benchtop Chassis or Portable Mainframe
2. Controller TLA700 Series Controller Benchtop Controller used with the Benchtop Chassis or the embedded controller in the Portable Mainframe
3. TLA700 Series Logic Analyzer
Performance Verification & Adjustment Software
4. Adjustment/verification fixture,
with one of the following Power Supplies: USA/CAN Europe Japan United Kingdom
5. DSO probe calibration fixture One required Tektronix part number
6. Oscilloscope 1 GHz bandwidth Delay time accura-
7. DSO probes Three required, with < one-inch
8. Logic analyzer probes Two required Tektronix P6417 or P6418
9. High density logic analyzer
probe (optional)
10. Pattern Generator module One required TLA7PG2 Merge and Deskew pattern
11. Pattern Generator probe
cable
Included with the Tektronix Logic Analyzer Family CD-ROM
Tektronix part number 671-3599-XX Tektronix part numbers:
12 V , 1.5 A 12 V , 1.5 A 12 V , 1.5 A 12 V , 1.5 A
cy ±25 ppm over any 1 ms interval
ground leads
One required Tektronix P6434 Logic
Eight required Tektronix part number
119-4855-XX 119-4856-XX 119-4859-XX 119-4857-XX
671-3930-XX Tektronix TDS 784D Adjustment/verification
Tektronix P6243 or P6245 probe, with std. accessories
Logic Analyzer probes
Analyzer probe
012-A212-00
All procedures, except TLA600 procedures
All procedures, except TLA600 procedures
All procedures for the TLA600 Logic Analyzers, LA modules and DSO modules
Most procedures for the LA module
DSO functional verification procedure
fixture procedures and pattern generator procedures
Most procedures for the DSO and Pattern Generator modules
Most procedures for TLA600 and the LA module
Most procedures for TLA600 and the LA module
generator procedures Pattern generator procedures
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–5
Introduction
T able 1–3: Test equipment (Cont.)
Item number and description Where usedExampleMinimum requirements
12. TTL/CMOS Pattern Genera­tor probes
13. ECL Pattern Generator probes
14. Termination board Two required Tektronix part number
15. Frequency counter Frequency accuracy: <0.0025%
16. Function generator Amplitude: 4V
17. Power supply Voltage: –2 V
18. BNC Cable Impedance: 50 W,
19. T-Connector One required Tektronix part number
20. SMB-to-BNC cable One required Tektronix P6041 PV procedures for the
21. Adapter, male BNC-to-male BNC
22. Miniature probe-to-square pin adapters
23. 0.025-inch square pin 3/4-inch length, (two required) Tektronix part number
24. Digital multimeter with test leads
25. Connector, dual-banana Female BNC-to-dual banana Tektronix part number
26. Voltage reference Accuracy: 0.01% Data Precision 8200 PV procedures for the
27. Capacitor
28. Cable, dual-input Female BNC-to-dual male BNC Tektronix part number
29. Adapter, N-to-BNC Male type N-to-female BNC Tektronix part number
30. Adapter, SMA-to-BNC Female type SMA-to-female BNC Tektronix part number
1
Four required Tektronix P6470 Pattern
Generator probes
Four required Tektronix P6471 Pattern
Generator probes
067-A016-00 Hewlett Packard 5314A Portable Mainframe and
Frequency range: 1 kHz to 100 MHz
T ektronix AFG2020 Pattern generator procedures Offset: 2 V (50 W termination) Frequency: 1 MHz or higher
METRONIX 524B Pattern generator procedures Current: 1 A
Tektronix part number Length: 24 in (two required)
One required Tektronix part number
Two required Tektronix part number
DCV accuracy: 0.1% from –10 V to +100 V
0.1 mF, 200 V Tektronix part number
012-1342-XX
103-0030-XX
103-0029-XX
103-0177-XX
131-1426-XX
Tektronix DMM 900 Series All procedures except
103-0090-XX
283-0189-XX
067-0525-XX
103-0045-XX
015-1018-XX
Most pattern generator procedures
Most pattern generator procedures
Most pattern generator procedures
Benchtop Controller procedures
Pattern generator procedures
Pattern generator procedures
Benchtop Controller DSO procedures
Portable Mainframe (power supply check)
Portable Mainframe (power supply check)
Benchtop Controller PV procedures for the
LA and DSO module
LA and DSO module PV procedures for the
LA and DSO module PV procedures for
the DSO module PV procedures for
the DSO module DSO procedures
1–6
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
T able 1–3: Test equipment (Cont.)
Item number and description Where usedExampleMinimum requirements
Introduction
31. Cable, precision 50 W coaxial
32. Generator, function Frequency range: 1 Hz to 20 MHz
33. Generator, sine wave Frequency range: 100 kHz to
50 W, 36 in, male-to-male BNC connectors
Frequency accuracy: 0.1% Amplitude range: 10 mV to 20 V
1005 MHz
Tektronix part number 012-0482-XX
All procedures except Benchtop Chassis
Hewlett Packard 3325B PV procedures for
the DSO module
p-p
Gigatronics 6061 PV procedures for the
DSO and LA module Frequency accuracy: <6 ppm Amplitude: 2 mV to 1.5 V
RMS
Accuracy: <0.35 dB
34. Power meter with sensor Bandwidth: >1.2 GHz Accuracy:
0.2 dB Sensitivity: 500 pW to 20 mW
35. Adapter, N-to-BNC (used
Female type N-to-male BNC Tektronix part number
only with power meter)
36. 50 W power divider (used only with power meter)
37. SMA-to-BNC adapters (used only with power meter)
38. Printer with cable (optional, used to print Calibration Data Reports)
1
The capacitor is installed across the Data Precision 8200 output terminals to reduce noise. If your voltage reference
Provide load isolation between equipment Maximum VSWR: 1.50
For use with the 50 W power divider (three required)
PC-compatible, continuous feed, prints ASCII text, connects to EPP connector
Rohde & Schwarz NRVS with model NRV-Z4 Sensor
DSO adjustment procedure #4
DSO adjustment
103-0058-XX Tektronix part number
015-0565-XX Tektronix part number
015-1018-XX
procedure #4 DSO adjustment
procedure #4 DSO adjustment
procedure #4
Any general purpose printer Printing test results for
the LA and DSO module
produces <4 mVp-p of noise, external noise reduction is not necessary.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–7
Introduction

Software Installation and Removal Instructions

These procedures describe loading and unloading the PV/Adjust software. It is recommended you have 10 MB of free space on the hard drive before installing the software. The Performance Verification software is located on Disc 1 of the Tektronix Logic Analyzer Family Application Software CD.
NOTE. This installation program uses parameters you supply to create a custom start-up file in your hard disk directory.
The batch file enables the software to configure your instrument properly before it runs the program.
1. Power on the instrument.
2. Exit the Application.
Verify PV/Adjust Software
Version

Verify Directories

If your logic analyzer already has PV/Adjust software loaded on it, you must verify that the version is the same as the version printed on Disc 1 of the Tektronix Logic Analyzer Family Application Software CD.
If the version of the PV/Adjust software loaded on your logic analyzer is an earlier version, you must delete the earlier version before you can load the newer version.
If your logic analyzer already has a directory named Tekcats or Temptek on the hard drive, the software installation cannot be completed. Follow these instruc­tions to verify the directory is not present:
1. Select Start Programs Windows Explorer.
2. From the Exploring window select Tools Find Files or Folders.
3. Select the Name & Location tab, and search for the following directories:
H Tekcats H Temptek
4. If either directory is found follow the instructions under Removing the
Software beginning on page 19 to remove the software and the directories.
1–8
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Introduction
Install the PV/Adjust
Software
Follow these instructions to install the PV/Adjust software and the pattern generator setup files.
1. Close all open windows on the desktop.
2. Insert Disc 1 of the Tektronix Logic Analyzer Family Application Software
CD in the CD-ROM drive.
3. Click the My Computer Icon and double-click the CD-ROM drive.
4. Double-click the TLA Performance Verification folder.
5. Double-click on the Logic Analyzer PV folder and then double-click the
Disk1 folder.
6. Double-click the Setup.exe icon to begin the installation program.
7. Follow the on-screen instructions to install the software on the hard disk.
8. After the installation is complete, go back to the TLA Performance
Verification folder on the CD.
9. Double-click on the Pattern Generator PV folder and then double-click the
Disk1 folder.
10. Double-click the Setup.exe icon to begin the installation program.

Removing the Software

11. Follow the on-screen instructions to install the pattern generator setup files
on the hard disk.
This completes the software installation procedure.
Use the following procedure to remove the PV/Adjust software from the instrument. These steps are necessary when you want to upgrade the PV software.
1. Select Start Programs Windows Explorer.
2. Using Windows Explorer, click the C:\ Tekcats folder.
3. Go to the File menu and select Delete to delete the folder.
4. Repeat steps 1 through 3 to delete the Temptek folder if it exists.
5. Select Start Settings Taskbar & Start Menu.
6. Click the Advanced tab.
7. Open the following directory path under Documents and Settings:
All Users Start Menu Programs
8. Locate and delete the TLA Performance Verification item.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–9
Introduction

Using the Software

The PV/Adjust software is required to complete the performance verification, adjustment and calibration/certification procedures for the TLA600 Logic Analyzers, the LA modules, and for the DSO modules. A separate series of program files are required for the pattern generator modules. Software is not required to perform any checks on the mainframes or any other subcomponents.
Pattern Generator Setup
Files
The PV/Adjust software used to verify the pattern generator modules consists of setup files rather than executable software. To use the pattern generator setup files complete the following steps:
1. Verify that your module configuration matches the setup as called for in the
writtern procedure (for some of the setups you will need to merge or unmerge modules).
2. Select Load Module from the File menu; the Load Module dialog box
appears.
3. Click the Browse button and navigate to the C:\Program Files\Tektronix
Pattern Generator\PV folder.
4. Double-click on the file name; the Load Module dialog box reappears with
the file name and module name under the Module list.
5. Click OK to load the module setup. A dialog box may appear reminding you
that the current module settings and data will be lost.
6. Click Yes to confirm your choice.
7. Follow the remaining written procedures.
1–10
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Introduction
LA and DSO Performance
Verification and
Adjustment Software
Running the LA and
DSO Software
The logic analyzer and DSO software consists of executable software files. Use the software in conjunction with this manual to set up and perform the adjust­ment procedures or the performance verification procedures. Use this manual in the following manner:
H To provide an overview of each test H To identify the prerequisites for each test H To identify the required test equipment H To connect the external test equipment for each test
Use the following steps to start and run the logic analyzer and DSO software:
1. Allow the instruments to warm up for at least 30 minutes before beginning
the procedure.
2. Quit all applications including the TLA application.
3. Select Start Programs TLA Performance Verification.
4. To run the PV/Adjust software, select one of the following:
H For the TLA7Dx and TLA7Ex DSO modules, select DSO PV. H For the TLA700 logic analyzer modules, select LA 7XX PV. H For the TLA600 Logic Analyzer, select LA 6XX PV.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–11
Introduction
5. Follow the instructions on the screen to enter the name you want to appear in
the User Name field as shown below. This name will appear on the Calibration Data Reports.
6. Click Enter to continue.
7. The program lists several different modules, referred to as DUT (Device
Under Test). Enter the number corresponding to the module type that you want to test; then click Enter to continue.
The screen will display an error message if the DUT chosen does not match the installed DUT.
8. Enter the complete serial number of the DUT (for example, B010100). Click
Enter to continue.
9. Verify the serial number entry; then click Enter to continue.
If you select no, a prompt asks you to enter the serial number again.
10. The program lists sequences for PV (performance verification) and ADJ
(adjustments). Enter a number to select which sequence you want and click Enter to continue.
11. If an instrument is being tested, the program lists the different probe types
available for testing. Enter the appropriate number corresponding to your probe and then click Enter to continue.
12. Enter the operating temperature in degrees C (entries in the range of 20 to 30
degrees are valid). Click Enter to continue.
1–12
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Introduction
13. Enter the operating humidity as a percentage (0% to 100% entries are valid).
Click Enter to continue.
14. Determine which sequence to run:
H RUN FULL SEQUENCE runs the entire sequence from beginning to
end. This is the recommended selection.
H RUN PARTIAL SEQUENCE runs part of the full sequence. The
sequence runs from the selected starting point to the end of the sequence.
H SELECT TEST(S) runs only the selected tests. To run a single test, enter
the test number. To run multiple tests, enter a comma-separated list of numbers or a hyphen-separated list of numbers.
Enter the number next to your choice and click Enter to continue.
15. Follow the on-screen instructions to connect and adjust test equipment. You
can also refer to the connection diagrams provided in this manual for each test.

Troubleshooting

Using the Interrupt Button

16. When testing is completed, disconnect the test equipment.
If the test you are attempting to run fails, do the following: H Check the settings and connections of the test equipment used in the
procedure.
H If you are using the adjustment/verification fixture, verify that the LED is
lighted and that the jumper positions and external connections are correct.
H Rerun mainframe or module diagnostics and module self-cal, then rerun the
test(s) which failed.
If the test(s) still do not pass, the mainframe or module might need further troubleshooting.
While the program is running, you can interrupt the program to rerun a test, start over, or to exit the program by clicking the Interrupt button (shown below).
The program will then provide a list of choices. Enter the number next to the choice that you want and click Enter.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–13
Introduction
NOTE. If you interrupt a test before it has completed, you must restart the test to obtain valid test data.
Some tests such as Internal Cal do not allow interrupts. If you stop these tests using more aggressive methods, you may have to reboot the instrument.

Obtaining Test Results

The results of all tests can be stored in a file on the hard disk. You can view the test results, print the test results to a printer, or save the test results in another file on the hard disk.
After running the performance verification procedures, you should create a calibration data report for certification purposes. Refer to Calibration Data Reports for details on using the resultant test data with the Calibration Data Report.

Calibration Data Reports

The Calibration Data Reports in this manual are intended to be copied and used for calibration/certification procedures. Some of the certification procedures require you to record data from manual measurements on the Calibration Data Reports directly. The TLA600 Logic Analyzer, the LA module, and the DSO module procedures require that you obtain the test results from the file created by the PV/Adjust software.
After completing the performance verification procedures or the certification procedures, you can fill out a Calibration Data Report to keep on file with your instrument. Calibration Data Reports are provided at the end of the certification procedures in each chapter of this manual, except for the TLA600 Logic Analyzer, the LA module, and DSO module. Calibration Data Reports for these instruments are generated by the PV/Adjust software. The Calibration Data Reports list those product attributes and specifications upon which product accuracy depends.
Retrieving Test Data from
the PV/Adjust Software
1–14
The program stores the test results in a file containing the module name and serial number (for example, TLA7E2.B020123). The file is located under the following path: C:\Tekcats\Rpt.
NOTE. If you want to save the content of the Report file, you must rename or copy the Report file using the Windows file utilities such as Explorer.
The Report file will be overwritten the next time you run the PV/Adjust software and print or view a new set of test data.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Introduction
After completing a full or partial test sequence (or just before you exit the program) you can generate the test data and write it to a file. You have the option of printing the file, viewing the file on screen, or transferring the file to another directory or host computer.
You can print the test data directly from the program. Ensure that a printer is connected to your logic analyzer and follow the on-screen instructions to print the test results.
If a printer is not available, you can view the test results directly from the screen, or you can copy the test results to a different file or folder/host computer for future use.
Field Adjust/PV Software
Housekeeping
The PV/Adjust software creates data log files (.dlf files) that store program data. The .dlf files are used by the PV/Adjust software to generate the view data and print out options. Each .dlf file is identified by the product serial number; example, B010100.dlf. The files are stored in the Tekcats folder under each TLA7xx folder. To conserve disk space, you must occasionally delete the .dlf files.
for
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–15
Introduction
1–16
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
TLA600 Series Logic Analyzer

TLA600 Series Logic Analyzer Functional Verification

This section contains instructions for performing the functional verification procedure. This procedure provides an easy way to check the basic functionality of the TLA600 Series Logic Analyzer.

Tests Performed

Table 2–1 lists the functional verification procedures that are available for the TLA600 Logic Analyzer and probes.
T able 2–1: TLA600 Functional verification procedures
Adjustment/verification
Instrument Procedure
Logic Analyzer Power-up diagnostics No Logic Analyzer Extended diagnostics No Logic Analyzer Mainframe diagnostics No Logic Analyzer QA+Win32 diagnostics No
fixture required
P6417, P6418, or P6434 logic analyzer probes
NOTE. Running the extended diagnostics will invalidate any acquired data. If you want to save any of the acquired data, do so before running the extended diagnostics.
If any check within this section fails, refer to the TLA600 Series Logic Analyzer Service Manual for corrective action.
The functional verification procedure consists of the following parts:
H Module self tests and power-on diagnostics H Probe verification
This procedure provides a functional check only. If more detailed testing is required, perform the Performance Verification Procedure after completing this procedure.
Signal Input check Yes
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–1
TLA600 Series Logic Analyzer Functional Verification

Test Procedure

You will need a TLA600 Series Logic Analyzer to complete this test procedure.
Equipment required Cable, 50 coaxial (item 31) Prerequisites Warm-up time: 30 minutes
Perform the following tests to complete the functional verification procedure.
Extended Diagnostics
Do the following steps to run the extended diagnostics:
NOTE. Perform the following steps to complete the functional verification procedures. Before beginning this procedure, be sure that no active signals are applied to the logic analyzer. Certain diagnostic tests will fail if signals are applied to the probe during the test.
1. Power on the instrument and wait for the system application to start.
2. Go to the System menu and select Calibration and Diagnostics.
3. Verify that all power-on diagnostics pass.
4. Click the Extended Diagnostics tab.
5. Select All Modules, All Tests (top line) and click the Run button on the
property sheet. The diagnostics software performs each one of the tests listed in the menu
under the module selection. All tests that displayed an Unknown status will change to a Pass or Fail status depending on the outcome of the tests.
6. Scroll through the test results for the instrument and verify that all tests pass.
2–2
NOTE. If Extended Diagnostics fail, run Self Cal for the logic analyzer and then rerun Extended Diagnostics.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
TLA600 Series Logic Analyzer Functional Verification
Mainframe Diagnostics
QA+Win32
The TLA Mainframe Diagnostics check the functionality of the instrument. To run these diagnostics, do the following steps:
1. Turn off all other applications.
2. Click the Windows Start button.
3. Select Programs from the Start menu.
4. Select Tektronix Logic Analyzer from the Programs menu.
5. Select Tektronix TLA Mainframe Diagnostics from the Tektronix Logic
Analyzer menu.
6. Run the mainframe diagnostics. Connect the BNC cable when requested. When the TLA Mainframe Diagnostics are run, a Mainframe Diagnostics
Warning dialog box appears. Clicking OK will initiate a SYSRESET, which will invalidate any acquired data.
QA+Win32 is a comprehensive software application used to check and verify the operation of the PC hardware in the instrument. To run QA+Win32, you must have either a working keyboard or a working mouse or other pointing device and have Windows running.
Checking the Cooling Fan
Operation
CAUTION. Before running the QA+Win32 tests, check the release notes for possible problems and workarounds.
To run QA+Win32, follow these instructions:
1. Turn off all other applications.
2. Click the Windows Start button.
3. Select Programs.
4. Select QA+Win32.
5. Run the QA+Win32 diagnostics.
Power on the instrument and inspect the side of the instrument to verify that all six cooling fans are rotating.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–3
TLA600 Series Logic Analyzer Functional Verification

Logic Analyzer Probe Functional V erification Procedure

The following procedure checks the basic operation of the probes by verifying that the probes recognize signal activity at the probe tips.
Equipment required Adjustment/verification fixture (item 4 ) Prerequisites Warm-up time: 30 minutes for instrument and test equipment
P6417, P6418, or P6434 probe connected to the instrument Test equipment connected as shown in Figure 2–1 Power-up diagnostics pass SELF_CAL passes
1
Do not mix probes; only one type of probe can be functionally verified at a time.
1
Test Procedure
C2/C3/CK3
Channel/Group
Clock channel
Adjustment/verification
J5 J3
J1
J2
fixture
Fixture supply
J26
Figure 2–1: Probe functional verification test setup
Perform the following steps to complete the probe functional verification procedure:
1. Ensure that the jumper at J15 on the adjustment/verification fixture is in the
INT position to select the internal 50.065 MHz clock.
2. Open the Setup window and click the Set Thresholds button to display the
Probe Threshold dialog box.
3. Adjust the threshold level to 700 mV for all channels.
2–4
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
TLA600 Series Logic Analyzer Functional Verification
NOTE. These procedures assume a P6418 or P6417 probe is used. If you are use a P6434 probe, use J5, the Data Out connector on the adjustment/
verification fixture to verify probe functionality. Observe proper polarity: pin 1 to pin 1.
4. Refer to Figure 2–1 and connect the podlets of the acquisition probe to J1
and J2 on the fixture. Ensure that you connect the ground side of the podlets to the ground side of the connectors.
5. Connect the single clock (CK n) or the qualifier (Q n) channel to one of the
J3 CLK OUT connector pairs on the fixture.
6. Return to the Setup window and click the Show Activity button to display
the Activity Monitor.
7. Verify that the Activity Monitor shows activity on all probe channels
connected to the test fixture. Figure 2–2 shows an example of the Activity Monitor. Note the signal
activity for clock CK3 and data channels for the C3(7-0) and C2(7-0) groups. Also note that there is no activity on the other groups because the probe podlets are not connected to a signal source (the channels are all high).
Figure 2–2: Activity Monitor
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–5
TLA600 Series Logic Analyzer Functional Verification
8. Verify that none of the connected channels are stuck high or stuck low.
9. Disconnect the probe from the fixture and the instrument.
10. Repeat the steps for any remaining probes.
11. Close the Activity Monitor.
12. Return the threshold levels to their former values in the Probe Threshold
dialog box.
13. This completes the probe functional verification procedure.
2–6
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

TLA600 Series Certification

Using the performance verification procedures, check the accuracy of the system clock and perform the DC Threshold test. The instrument is certifiable if these parameters meet specifications. Other specifications can also be verified by running the complete performance verification procedures.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–7
TLA600 Series Certification
2–8
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Calibration Data Report

TLA600 Logic Analyzer

Instrument model number: Serial number: Certificate number: Verification performed by: Verification date:

System Clock Test Data

Procedure
Characteristic Specification Tolerance
Clock frequency 10 MHz ±1 kHz (9.9990 MHz-10.0010 MHz) Page 2–13, Step 5
reference
Incoming data Outgoing data

TLA600 Series Logic Analyzer Performance Verification

This section contains procedures to verify that the TLA600 series instruments perform as warranted. Verify instrument performance whenever the accuracy or function of your instrument is in question, or as part of an annual calibration/cer­tification.
As a general rule, these tests only need to be done once a year.

Prerequisites

These procedures ask for the serial number of the TLA600 series instrument under test. The serial number and state speed to the instrument is noted on a label on the right side of the back panel.
The tests in this section comprise an extensive, valid confirmation of perfor­mance and functionality when the following requirements are met:
H The Tektronix Logic Analyzer application must not be running. H The PV/Adjust software must be loaded. Refer to Software Installation and
Removal Instructions on page 18.

Tests Performed

H The logic analyzer, must have been operating for a warm-up period of at
least 30 minutes, and must be operating at an ambient temperature between +20_ C and +30_ C.
H The logic analyzer must have been last adjusted at an ambient temperature
between +20_ C and +30_ C.
H The logic analyzer must be in an operating environment within the limits
described in the Specifications section of the Tektronix Logic Analyzer
Family User Manual.
The PV/Adjust software contains the tests shown in Table 2–2. Each test verifies one or more parameters. All of the tests check characteristics that are designated as checked (n) in the Specifications section of the Tektronix Logic Analyzer
Family User Manual.
By running a full PV sequence, you will verify the performance of the logic analyzer.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–11
TLA600 Series Logic Analyzer Performance Verification
T able 2–2: TLA600 PV/Adjust software performance verification tests
Performance verification test name Specification tested
1. System clock (CLK10)
1
2. FPV_DC_THRESHOLD
3. FPV_SETUP_0F Setup time 4 FPV_HOLD_0F Hold time
5. FPV_MAXSYNC Maximum synchronous clock rate
1
Certifiable parameter
Table 2–3 lists the additional characteristics that are designated as checked (n) in the Specifications section of the Tektronix Logic Analyzer Family User Manual.
These characteristics are indirectly tested by the PV/Adjust software tests named in the table.
Signal output check
1
Threshold accuracy
T able 2–3: TLA600 characteristics indirectly checked by the PV/Adjust software performance verification tests
Performance verification test name Specification tested
1. FPV_SETUP_0F Minimum recognizable word 2 FPV_HOLD_0F Minimum recognizable word
1,2 1,2
3. All tests Trigger state sequence rate
4. All tests, and Extended Diagnostics Internal sampling period
1
When the setup and hold time tests are both performed, the setup and hold
3
window size is indirectly verified.
2
When the setup and hold time tests are both performed, the channel-to-channel skew is indirectly verified.
3
When all of the tests are performed, including Extended Diagnostics, the internal sampling period is indirectly verified.
The procedures in this document assume that you will run a full sequence of tests. Each test includes a simple illustration of the test equipment setups.
2–12
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Test Equipment

Test Procedures

TLA600 Series Logic Analyzer Performance Verification
These procedures use external, traceable signal sources to directly test the characteristics that are designated as checked (n) in the Specifications section of the Tektronix Logic Analyzer Family User Manual.
In addition to the basic system setup, you will need some of the equipment shown in Table 1–3 on page 1–5 in the Introduction section to complete the performance verification procedures. A condensed list of test equipment is provided at the beginning of each test.
Use the following steps to complete the performance verification procedure. Each procedure includes a table that calls out the equipment used; use Table 1–3 beginning on page 1–5 for equipment specifications. If you substitute equip­ment, always choose instruments that meet or exceed the minimum requirements specified.
Do the following to check the accuracy of the 10 MHz system clock.
Equipment required
Prerequisites Warm-up time: 30 minutes for the logic analyzer and test equipment
1. Connect the frequency counter to the External Signal Out TTL BNC
connector on the back of the instrument.
2. Verify that the Tektronix Logic Analyzer application is running.
3. Go to the System window and select System Configuration from the System
menu.
4. In the System Configuration dialog box, select 10 MHz Clock from the list
of routable signals in the External Signal Out selection box and click OK.
5. Verify that the output frequency at the External Signal Out TTL connector is
10 MHz ±1 kHz. Record the measurement on the Calibration Data Report.
6. In the system configuration dialog box, reset the External Signal Out signal
to None.
Frequency counter (item 15) Precision BNC cable (item 31)
Power-on, Mainframe, and QA+WIN32 diagnostics pass
7. Disconnect the frequency counter.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–13
TLA600 Series Logic Analyzer Performance Verification

Using the Software

This section provides a brief overview on using the PV/Adjust software.
NOTE. If you are testing an instrument that has been repaired, you must complete the adjustment procedures before running the performance verification.
When using the PV/Adjust software, you will connect external test equipment to the probe inputs in response to prompts on the screen. You will connect the test signals and then instruct the program to continue. The PV/Adjust software automatically selects the LA settings and determines the results of each test.
The results of the tests are recorded in a temporary file and are available upon test completion for completing test records for certification. To obtain partial test information you can also run individual tests or selected groups.
NOTE. The SELF_CAL test must run successfully before the other tests are performed. The remaining tests can then be performed in any order.
The PV/Adjust software contains online instructions for performing the performance verification procedure. The basic steps for completing the proce­dures follow:
1. Start the program, enter user and product identification information.
2. Select either PV or Adj.
3. Select temperature and humidity.
4. Select the probes.
NOTE. If you are using P6417 or P6418 probes, label one probe as the Reference Probe, and the other probe as the Probe Under Test.
If you are using a P6434 probe, label the probe channel group identified as the pin 38 side as Probe A. Label the probe channel group identified as the pin 1 side as Probe B; refer to Figure 2–3.
2–14
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Beveled corner
Pin 1 label
Recess on
pin 1 side
TLA600 Series Logic Analyzer Performance Verification
Pin 38 side
Pin 1 side
Pin 38 side
Recessed edge
Beveled corners
End view
Figure 2–3: P6434 probe detail
5. Select a full test sequence.
6. Connect the test equipment.
7. Set up the test equipment for the output signals described by on-screen
instructions and by the connection illustration for each test.
8. Run each test as instructed.
NOTE. Some tests prompt you for input on whether to “1:Do Section or 2:Skip Section. It is recommended that you select Do Section unless you use the software for troubleshooting.
9. After completing all the tests, view the test results and print them out if
performing a certification. Refer to Retrieving Test Data from the PV/Adjust Software on page 1–14 for instructions on obtaining test data.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–15
TLA600 Series Logic Analyzer Performance Verification

Troubleshooting

If any tests fail, use the following steps to troubleshoot problems:
1. Check all test equipment for improper or loose connections.
2. Check that all test equipment is powered on and has the proper warm-up
time.
3. If you are using the adjustment/verification fixture, verify the LED is
lighted, jumper positions match the on-screen instructions, and the external connections are correct. (See Figure 2–4 for jumper locations.)
4. Rerun all diagnostics and adjustment procedures.
5. Run the tests a second time to verify the failure.
6. If tests continue to fail, refer to the TLA600 Series Logic Analyzer Service
Manual for corrective action.

Performance V erification Tests

Use the following tables and figures to set up and execute each procedure.
NOTE. The illustrations in the following procedures assume that you are using the P6418 or P6417 probes.
If you have a P6434 probe, use J14 on the adjustment/verification fixture for the DC Threshold test; all other procedures using the P6434 probe use J5, the Data Out connector.
When using either type of probe, always observe correct polarity (GND to GND, pin 1 to pin 1).
2–16
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
TLA600 Series Logic Analyzer Performance Verification
Threshold
External
Clock in
P6434
Threshold out
J14
J15
J10 J8
J22
J18
J26
J19
J17
J9
J11
DC in
in
P6417 and P6418 Threshold out
J6 J7
J20
J1
J16
J13
J2
J3
J5
J4
P6417 and P6418 Data out
P6434
Data out
J17
Hold/Setup
select
SetupHold
Jumper installed:
clocked on falling edge
Jumper removed:
clocked on rising edge
J19
Ext/Int
Clock select
J15
IntExt
1 2 3 4 5
J16
Deskew
Figure 2–4: Adjustment/verification fixture connections and jumper locations
Deskew
Probe
setup
6
J13
P6434
Deskew/Probe setup
1
Clock Clock
GND GND
J4
Setup/Hold
Clock out
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–17
TLA600 Series Logic Analyzer Performance Verification
TLA600 Procedure 1:
FPV_DC_Threshold
This procedure verifies the DC Threshold Accuracy of the logic analyzer. This test is performed once and applies to all channels of the instrument.
SW test name FPV_DC_Threshold Equipment
required
Prerequisites Warm-up time: 30 minutes for the TLA600 and test equipment
Adjustment/verification fixture and fixture supply (item 4) Voltage reference (item 26) Precision BNC cable (item 31) Dual banana-to-BNC adapter (item 25) Capacitor, 0.1 mF (item 27)
Test equipment connected as shown in Figure 2–5 Diagnostics and SELF_CAL pass
Voltage reference
+ –
Dual banana-to­BNC adapter
BNC cable
0.1 mF Capacitor Fixture supply
C2/C3/CK3 Channel/Group
Adjustment/verification
J7
fixture
J14 Threshold out P6434
J6
J22
J26
Figure 2–5: FPV_DC_Threshold test setup
1. If the logic analyzer application is running, quit the application.
2. Verify that all of the prerequisites listed previously are met for the procedure.
3. Run the PV/Adjust software
as described in Running the LA and DSO
Software on page 111. Run the C:\Tekcats\Tla_la, program, and select the
correct options.
4. Follow the on-screen instructions to run each portion of the test for each
parameter of the instrument.
2–18
5. Verify that all of the tests pass.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
TLA600 Series Logic Analyzer Performance Verification
TLA600 Procedure 2:
FPV_Setup_0F
This procedure verifies the setup time of the instrument.
SW test name FPV_Setup_0F Equipment
required Prerequisites Warm-up time: 30 minutes for the TLA600 and test equipment
C2/C3/CK3
Channel/Group
Adjustment/verification fixture and fixture supply (item 4)
Test equipment connected as shown in Figure 2–6 Diagnostics and SELF_CAL pass
Adjustment/verification
J4
Clock channel
J5
J2
J1
fixture
J26
Fixture supply
Figure 2–6: Initial FPV_Setup_0F test setup
1. If the logic analyzer application is running, quit the application and verify
that all of the prerequisites listed previously are met for the procedure.
2. Follow the on-screen instructions to run each portion of the test for each
parameter of the instrument.
3. Verify that all tests pass. If a test fails, run the Deskew routine as described
in the TLA600 Series Logic Analyzer Adjustment section beginning on page 2–23, then rerun the test.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–19
TLA600 Series Logic Analyzer Performance Verification
TLA600 Procedure 3:
FPV_Hold_0F
This procedure verifies the hold time of the instrument.
SW test name FPV_Hold_0F Equipment
required Prerequisites Warm-up time: 30 minutes for the TLA600 and test equipment
C2/C3/CK3
Channel/Group
Adjustment/verification fixture and fixture supply (item 4)
Test equipment connected as shown in Figure 2–7 Diagnostics and SELF_CAL pass
Adjustment/verification
J4
Clock channel
J5
J2
J1
fixture
Fixture supply
J26
Figure 2–7: Initial FPV_Hold_0F test setup
1. If the logic analyzer application is running, quit the application and verify
that all of the prerequisites listed previously are met for the procedure.
2. Follow the on-screen instructions to run the test for each channel on the
system.
3. Verify that all tests pass. If a test fails, run the Deskew routine as described
in the TLA600 Series Logic Analyzer Adjustment section beginning on page 2–23, then rerun the test.
2–20
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
TLA600 Series Logic Analyzer Performance Verification
TLA600 Procedure 4:
FPV_Maxsync
This procedure checks the Maximum Synchronous Clock Rate and the Trigger State Sequence Rate. This test is performed once and applies to all channels of the instrument.
SW test name FPV_Maxsync Equipment
required
Prerequisites Warm-up time: 30 minutes for LA instrument and test equipment
Adjustment/verification fixture and fixture supply (item 4) Sine wave generator (item 33) BNC cable (item 31) Adapter, N-to-BNC (item 29)
Test equipment connected as shown in Figure 2–8 Diagnostics and SELF_CAL pass
Sine wave generator
Adjustment/verification
J4
J5
fixture
J18
Type N-to-BNC adapter
BNC cable
Fixture supply
C2/C3/CK3 Channel/Group
Clock channel
J2
J26
J1
Figure 2–8: FPV_Maxsync test setup
1. If the TLA system application is running, quit the application and verify that
all of the prerequisites listed previously are met for the procedure.
2. Follow the on-screen instructions.
3. Verify that all of the tests pass.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–21
TLA600 Series Logic Analyzer Performance Verification

Completing the Performance Verification

After completing the performance verification procedures, obtain a copy of the test results and verify that all parameters are within the allowable specifications as listed in the Tektronix Logic Analyzer Family User Manual.
Refer to Retrieving Test Data from the PV/Adjust Software on page 1–14 for instructions on obtaining test data.
2–22
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

TLA600 Series Logic Analyzer Adjustment

This section contains procedures which use the PV/Adjust software to adjust the TLA600 Series Logic Analyzers to within factory specifications. The PV/Adjust software contains instructions and control programs for adjusting the logic analyzer. The software describes test equipment connections and settings, selects setup parameters, and loads calibration constants into memory.
These procedures adjust the logic analyzer for conformance with the warranted characteristics listed in the Specifications section of the Tektronix Logic Analyzer Family User Manual.
Adjustments should be done after repair or when performance verification test fail.

Prerequisites

These procedures ask for the serial number of the instrument under test. You can access the serial number through the logic analyzer application. In the applica­tion, go to the System menu, select System Properties, and click on the LA tab. However, you must quit the application before continuing with the PV/Adjust software procedures.
Only trained service technicians should perform this procedure after meeting the following requirements:
H The logic analyzer application must not be running. H The PV/Adjust software must be loaded. Refer to Software Installation and
Removal Instructions beginning on page 18.
H The logic analyzer requires a 30-minute warm-up time in a +20_ C to
+30_ C environment before it is adjusted. Adjustments performed before the operating temperature has stabilized may cause errors in performance.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–23
TLA600 Series Logic Analyzer Adjustment

Using the Software

This section describes how to perform adjustments using the PV/Adjust software.
Performing the
Adjustments
Adjustment Sequences
and Dependencies
Adjustment After Repair

Test Equipment

There are no manual adjustments for the logic analyzer. Instead, the PV/Adjust software adjusts the instrument hardware using external test equipment connec­tions that you provide in response to prompts on the screen.
Upon successful completion of each adjustment, the PV/Adjust software automatically loads the new calibration data into memory.
The PV/Adjust software allows you to run groups of adjustments, or sequences. A sequence consists of one or more individual adjustments. Normally you will perform a RUN FULL SEQUENCE, which executes each adjustment in the proper order.
The PV/Adjust software also provides instructions for running each adjustment individually. However, you should only perform individual adjustments while troubleshooting.
You must perform a full adjustment sequence following replacement of any circuit board. Refer to the Maintenance Section of the TLA600 Series Logic Analyzer Service Manual for more details.
2–24
In addition to the basic system setup, you will need some of the equipment shown in Table 1–3 on page 1–5 in the Introduction section.
Each procedure includes a table that calls out the equipment used. Use Table 1–3 to identify required equipment specifications. If you substitute equipment, always choose instruments that meet or exceed the minimum requirements specified.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Adjustment Instructions

TLA600 Series Logic Analyzer Adjustment
This section describes how to perform adjustments using the PV/Adjust software.
Using the PV/Adjust
Software
The PV/Adjust software contains instructions for performing the adjustments. The basic steps for completing the procedures follow:
1. Start the program, enter user and product identification information and
temperature and humidity.
2. If you are using P6417 or P6418 probes, label one probe as the Reference
Probe, and the other probe as the Probe Under Test.
3. If you are using a P6434 probe, label the probe channel group identified as
the pin 38 side as Probe A. Label the probe channel group identified as the pin 1 side as Probe B, refer to Figure 2–9.
Pin 1 label
Recess on
pin 1 side
Beveled corners
End view
Pin 1 side
Pin 38 side
Beveled corners
Figure 2–9: P6434 probe detail
4. Select a full adjustment sequence.
5. Connect the test equipment.
6. Set up the test equipment for the output signals described by on-screen
instructions and by the connection illustration for each test.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Pin 38 side
2–25
TLA600 Series Logic Analyzer Adjustment
7. Run each adjustment step as instructed by following the on-screen prompts.
8. After completing all the adjustment steps, view the results to confirm that
the adjustment was successful.
Troubleshooting

Tests Performed

If any adjustments fail, use the following steps to troubleshoot the problems:
H Check all test equipment for improper or loose connections. H Check that all test equipment is powered on and has the proper warm-up
time.
H Verify that the adjustment/verification fixture LED is lighted, the jumper
positions match the on-screen instructions, and the external connections are correct.
H Rerun the diagnostics and self-cal. H Run the adjustment procedures a second time to verify the failure. H If the adjustment procedures continue to fail, refer to the TLA 600 Series
Logic Analyzer Service Manual for corrective action.
The adjustment procedures check and adjust the following parameters of the logic analyzer:
H SELF_CAL, an internal routine in the application software that adjusts
acquisition thresholds, internal module and signal timing.

Adjustment Procedures

2–26
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
H Deskew, an adjustment routine in the PV/Adjust software which time-aligns
all channels.
NOTE. Do not mix probe types (P6417, P6418 and P6434) when performing the deskew procedure.
Refer to the following procedures to identify the initial setup for each adjust­ment. Then follow the program instructions to complete the adjustments.

Self Calibration

TLA600 Series Logic Analyzer Adjustment
Self calibration (SELF_CAL) is an internal routine that optimizes performance at the current ambient temperature to maximize measurement accuracy. No external equipment or user actions are needed to complete the procedure. The logic analyzer saves data generated by the self calibration in memory. Passing the self calibration provides a higher level of confidence of functionality.
NOTE. Performing the self calibration does not guarantee that all parameters operate within limits.
Operation within limits is achieved by performing the adjustment procedures. Verification of operation within limits is accomplished by performing the
performance verification procedures.
When to Perform the Self Calibration. You can run the self calibration at any time during normal operation. To maintain measurement accuracy, perform the self calibration if the following conditions occur:
TLA600 Procedure 1:
SELF_CAL
H After repair and replacement of any circuit boards. H It has been a year since the last self calibration was run
Perform the following steps to run the SELF_CAL routine. Before beginning this procedure, be sure that no active signals are applied to the instrument. Self calibration can fail if signals are applied to the probe during the procedure.
Prerequisites Warm-up time: 30 minutes
Power-up diagnostics pass
1. Ensure that the logic analyzer has had a 30-minute warm up before attempt-
ing the self calibration, and that the logic analyzer application is running.
2. Disconnect any probes connected to the logic analyzer.
3. Select the Calibration and Diagnostics property sheet from the System menu.
4. Select the Self Calibration tab.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–27
TLA600 Series Logic Analyzer Adjustment
5. Click the Run button to start the self calibration.
The self calibration takes several minutes to complete, depending on the number of channels in the instrument. Upon successfully completing the self calibration, the status changes from Running to Calibrated, and the Date and Time field is set to the present.
TLA600 Procedure 2:
Deskew
The deskew procedure is used to calibrate and adjust the timing alignment of the probe data channels and receivers.
Perform the deskew procedure:
H Once a year. H If you have replaced the acquisition board (the probe constants are stored on
the acquisition board).
This procedure checks and adjusts the time alignment of all channels. There are no manual adjustments.
SW test name Deskew Equipment
required
Prerequisites Warm-up time: 30 minutes for the TLA600 and test equipment
Adjustment/verification fixture and fixture supply (item 4 ) Two P6417 or P6418 Logic Analyzer probes (item 8), OR One P6434 Logic Analyzer probe (item 9)
Test equipment connected as shown in Figure 2–10 Diagnostics and SELF_CAL pass
2–28
Fixture supply
J26
C2/C3/CK3 Channel/Group
Clock channel
Adjustment/verification
J5 J3
J1
fixture
J2
Figure 2–10: Initial deskew test setup
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
TLA600 Series Logic Analyzer Adjustment
1. If the logic analyzer application is running, quit the application.
2. Verify that all of the prerequisites listed previously are met for the procedure.
3. Run the PV/Adjust software as described in the Running the LA and DSO
Software on page 1–11. Run the C:\Tekcats\Tla_la program and select the correct options.
NOTE. These procedures assume that a P6418 or P6417 probe is being used. If you are using a P6434 probe, use J5, the Data Out connector on the adjustment/ verification fixture for performing the deskew test. Observe proper polarity: pin 1 to pin 1.
4. Follow the on-screen prompts to perform the adjustment procedure.
5. Verify no failures occur for each test.
If desired, you can print or save the results of the deskew adjustment operation. However, this is not required for calibration. The file will be over written when the next adjustment or performance verification sequence is run.
If failures do occur, confirm that diagnostics and SELF_CAL pass.

Completing the Adjustment Steps

After completing the adjustments, obtain a copy of the test results and verify that all tests passed. Refer to Retrieving Test Data from the PV/Adjust Software on page 1–14 for instructions on obtaining test data.
Run the Performance Verification Procedures to verify that the all parameters are within the allowable specifications as listed in the Tektronix Logic Analyzer
Family User Manual.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2–29
TLA600 Series Logic Analyzer Adjustment
2–30
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Portable Mainframe

Portable Mainframe Functional Verification

This section contains instructions for performing the functional verification of the Portable Mainframe.

Tests Performed

Table 3–1 lists the functional verification procedures that are available for the Portable Mainframe.
T able 3–1: Functional verification procedures
Instrument Procedure

Portable Mainframe Power-up diagnostics Portable Mainframe Extended diagnostics Portable Mainframe TLA700 Mainframe diagnostics Portable Mainframe QA+Win32 diagnostics

NOTE. Running the extended diagnostics will invalidate any acquired data. If you want to save any of the acquired data, do so before running the extended diagnostics.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
3–1
Portable Mainframe Functional Verification

Test Procedure

You will need a Portable Mainframe with an LA or DSO module installed.
Extended Diagnostics
Equipment required
Prerequisites Warm-up time: 30 minutes
Perform the following tests to complete the functional verification procedure.
Do the following steps to run the extended diagnostics:
1. Power on the instrument and wait for the logic analyzer application to start.
2. Go to the System menu and select Calibration and Diagnostics.
3. Verify that all power-on diagnostics pass.
4. Click the Extended Diagnostics tab.
5. Select All Modules, All Tests and then click the Run button on the property
sheet. All tests that displayed an “Unknown” status will change to a Pass or Fail
status depending on the outcome of the tests.
6. Scroll through the tests and verify that all tests pass.
Cable, 50 W coaxial (item 31)
3–2
TLA700 Mainframe
Diagnostics
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
The TLA700 Mainframe Diagnostics are a comprehensive software test that checks the functionality of the portable mainframe. To run the TLA700 Mainframe Diagnostics, do the following steps:
1. Quit all other applications.
2. Click the Windows Start button.
3. Select Programs from the Start menu.
4. Select Tektronix TLA700 from the Programs menu.
5. Select Tektronix TLA700 Mainframe Diagnostics from the Tektronix
TLA700 menu.
6. Run the mainframe diagnostics.
Portable Mainframe Functional Verification
NOTE. If modules are installed in the mainframe when the TLA700 Mainframe Diagnostics run, a Mainframe Diagnostics Warning dialog box appears. Clicking OK will initiate a SYSRESET, which will invalidate any acquired data in the modules. You may want to save acquired data before continuing.
QA+Win32
Cooling Fan Operation
QA+Win32 is a comprehensive software application used to check and verify the operation of the PC hardware in the portable mainframe. To run QA+Win32, you must have either a keyboard, mouse, or other pointing device.
NOTE. Before running the QA+Win32 tests, check the release notes to become aware of possible problems and workarounds.
To run QA+Win32, follow these instructions:
1. Quit all other applications.
2. Click the Windows Start button.
3. Select QA+Win32.
4. Run the QA+Win32 diagnostics.
Power on the instrument and inspect the rear of the instrument to verify that all six cooling fans are rotating.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
3–3
Portable Mainframe Functional Verification
3–4
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Portable Mainframe Certification

The system clock is checked for accuracy. The instrument is certifiable if this parameter meets specifications. Do the performance verification procedures and record the values on the Calibration Data Report provided on the following page.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
3–5
Portable Mainframe Certification
3–6
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Calibration Data Report

TLA 714 Color Portable Mainframe

Instrument model number: Serial number: Certificate number: Verification performed by: Verification date:

System Clock Test Data

Procedure
Characteristic Specification Tolerance
Clock frequency 10 MHz ±1 kHz (9.9990 MHz-10.0010 MHz) Page 3–12,
reference
Step 5
Incoming data Outgoing data

Portable Mainframe Performance Verification

This section contains procedures to verify that the Portable Mainframe performs as warranted. Verify instrument performance whenever the accuracy or function of your instrument is in question, or as part of an annual calibration/certification.

Tests Performed

Table 3–2 lists the performance verification procedures available for the Portable Mainframe.
T able 3–2: Performance verification procedures
Parameter Procedure
Power supply voltages Voltage level check System clock (CLK10)
1
Certifiable parameter
1
Signal output check
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
3–9
Portable Mainframe Performance Verification

Test Procedures

Use the following steps to complete the performance verification procedure. You will need some of the equipment shown in Table 1–3 on page 1–5 to complete the performance verification procedures. If you substitute equipment, always choose instruments that meet or exceed the minimum requirements specified.
Checking Power
Supply Voltages
Do the following to verify power supply performance:
Equipment required
Prerequisites Warm-up time: 30 minutes for the portable mainframe and test
DMM with test leads (item 24) Two miniature probe-to-square pin adapters (item 22)
0.025-inch square pin (item 23) Frequency counter (item 15) Precision BNC cable (item 31)
equipment Power-on, mainframe, and QA+WIN32 diagnostics pass
1. Power off the portable mainframe and remove any modules installed in it.
2. Power on the portable mainframe and connect the reference lead of a digital
voltmeter to chassis ground, such as the top of the power supply.
3. Attach a 0.025-inch square pin to the probe tip of the other lead and insert it
into one of the pins on the backplane connectors listed in Table 3–3.
3–10
CAUTION. Use of a square pin of other dimensions will damage the Portable Mainframe.
The pins that should be carrying voltages are listed in Table 3–3. The arrange­ment of J1 and J2 connectors on the backplane is shown in Figure 3–1. Measure the power supply voltages with the voltmeter and compare each reading to the values listed in the tables. If the voltages are within the ranges listed in the table, the power supply is operating properly.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Portable Mainframe Performance Verification
T able 3–3: Power supply voltages and backplane connector pins
J1 pin Voltage (range) J2 pin Voltage (range)
Row A Row A
31 –12 V (–12.60 V to –1 1.64 V) 2 –2 V (–2.10 V to –1.90 V)
25 +5 V (4.875 V to 5.250 V)
Row C Row C
31 +12 V (11.64 V to 12.60 V) 4 –5.2 V (–5.460 V to –5.044 V)
31 +24 V (23.28 V to 25.20 V) 32 –24 V (–25.20 V to –23.28 V)
J2 rows of pins J1 rows of pins
Figure 3–1: Location of J1 and J2 pins in the portable mainframe
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
3–11
Portable Mainframe Performance Verification
Checking the 10 MHz
System Clock (CLK10)
Do the following to check the accuracy of the system clock.
1. Connect the frequency counter to the External Signal Out TTL BNC
connector on the rear of the portable mainframe.
2. Verify that the logic analyzer application is running.
3. Go to the System window and select System Configuration from the System
menu.
4. In the System Configuration dialog box, select 10 MHz Clock from the list
of routable signals in the External Signal Out selection box and click OK.
5. Verify that the output frequency at the External Signal Out TTL connector is
10 MHz ±1 kHz. Record the measurement on the Calibration Data Report.
6. In the system configuration dialog box, reset the External Signal Out signal
to None.
7. Disconnect the frequency counter and power off the portable mainframe.
8. Reinstall any modules that were removed at the beginning of the perfor-
mance verification procedure.
3–12
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Portable Mainframe Adjustment

There are no adjustments for the Portable Mainframe.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
3–13
Portable Mainframe Adjustment
3–14
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Benchtop Chassis and Expansion
Mainframe

Benchtop Chassis and Expansion Mainframe Functional Verification

This section contains instructions for performing the functional verification of the TLA720 Benchtop Chassis and the TLA7XM Expansion Mainframe (with an expansion module). Because these products are basically identical the functional checks are the same for both products.

Benchtop Chassis

Power on the benchtop chassis and observe that the On/Standby switch illuminates. Verify that the fan spins without undue noise.

Expansion Chassis

Ensure that you have an expansion module installed in the expansion mainframe. Refer to the Tektronix Logic Analyzer Family User Manual for installation information if necessary.
The expansion mainframe automatically powers on when the attached benchtop or portable mainframe powers on. If everything is properly connected and operational, you should see the expansion mainframe and the installed modules in the System window.
Observe that the On/Standby switch illuminates. Verify that the fan spins without undue noise.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
4–1
Benchtop Chassis and Expansion Mainframe Functional Verification
4–2
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Benchtop Chassis and Expansion Mainframe Certification

There are no certifiable parameters for the TLA720 Benchtop Chassis and for the TLA7XM Expansion Mainframe. There are also no certifiable parameters for the expansion module.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
4–3
Benchtop Chassis and Expansion Mainframe Certification
4–4
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Benchtop Chassis and Expansion Mainframe Performance Verification

This section contains procedures to verify that the TLA720 Benchtop Chassis and the TLA7XM Expansion Mainframe (with the expansion module) perform as warranted. Verify instrument performance whenever the accuracy or function of your instrument is in question.
NOTE. There are no certifiable parameters on the TLA720 Benchtop Chassis or for the TLA7XM Expansion Mainframe and expansion module. Refer to the benchtop controller for certification requirements for the chassis.

Test Procedure

Do the following tests to verify the performance of the TLA720 Benchtop Chassis or the TLA7XM Expansion Mainframe. You will need some of the equipment shown in Table 1–3 on page 1–5 to complete the performance verification procedures. If you substitute equipment, always choose instruments that meet or exceed the minimum requirements specified.
Check Power Supply
Voltages
Use a digital voltmeter to check the voltages on the pins of the 25-pin connector and compare the results against the range listed in Table 4–1. The connector is located on the right rear panel of the chassis (see Figure 4–1).
Equipment required
Prerequisites Warm-up time: 30 minutes
CAUTION. To prevent damaging the 25 pin connector, use care when probing the connector with the square pins.
Digital multimeter with test leads (item 24) Two 0.025-inch square pins (item 23) Two miniature probe-to-square pin adapters (item 22)
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
4–5
Benchtop Chassis and Expansion Mainframe Performance Verification
Pin 1
Pin 25
Figure 4–1: Benchtop and Expansion Chassis 25-pin rear panel connector
1. Connect the reference lead of the digital voltmeter to one of the ground pins.
2. Connect the other lead of the digital voltmeter to the supply pins of the
25-pin connector on the right rear panel of the chassis.
3. Measure the power supply voltages with the voltmeter and compare each
reading to the values listed in Table 4–1. If the voltages are within the specified ranges, your chassis is operating properly.
T able 4–1: Power supply voltages at the 25-pin connector
Pin Supply Acceptable voltage range
1 +5 V 4.875 V to 5.250 V 2 –12 V –12.60 V to –11.64 V 3 –24 V –25.20 V to –23.28 V 4 –2 V –2.10 V to –1.90 V 14 +12 V 11.64 V to 12.60 V 15 +24 V 23.28 V to 25.20 V
4–6
16 –5.2 V –5.460 V to –5.044 V 9, 17–20, 22, 24 Logic Ground
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Benchtop Chassis and Expansion Mainframe Adjustment

There are no adjustments for the TLA720 Benchtop Chassis and the TLA7XM Expansion Mainframe (and expansion module).
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
4–7
Benchtop Chassis and Expansion Mainframe Adjustment
4–8
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Benchtop Controller

Benchtop Controller Functional Verification

This section contains instructions for performing the TLA720 Benchtop Controller functional verification procedure.

Tests Performed

Table 5–1 lists the functional verification procedures available for the benchtop controller.
T able 5–1: Functional verification procedures
Instrument Procedure

Benchtop Controller Power-up diagnostics Benchtop Controller Extended diagnostics Benchtop Controller TLA700 Mainframe diagnostics Benchtop Controller QA+Win32 diagnostics

Test Procedure

Extended Diagnostics
NOTE. Running the extended diagnostics will invalidate any acquired data. If you want to save any of the acquired data, do so before running the extended diagnostics.
You will need a benchtop chassis with an LA module or DSO module installed.
Equipment Required
Prerequisites Warm-up time: 30 minutes
Do the following steps to run the extended diagnostics:
1. Power on the instrument and wait for the logic analyzer application to start.
2. Go to the System menu and select Calibration and Diagnostics.
3. Verify that all power-on diagnostics pass.
Cable, 50 W coaxial (item 31)
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
5–1
Benchtop Controller Functional Verification
4. Click on the Extended Diagnostics tab.
5. Select All Modules, All Tests and then click the Run button on the property
sheet. All tests that displayed an “Unknown” status will change to a Pass or Fail
status depending on the outcome of the tests.
6. Scroll through the tests and verify that all tests pass.
TLA700 Mainframe
Diagnostics
The TLA700 Mainframe Diagnostics are a comprehensive software test that checks the functionality of the instrument. To run the TLA700 Mainframe Diagnostics, do the following steps:
1. Quit all other applications.
2. Click on the Windows Start button.
3. Select Programs from the Start menu.
4. Select Tektronix TLA700 from the Programs menu.
5. Select Tektronix TLA700 Mainframe Diagnostics from the Tektronix
TLA700 menu.
6. Run the mainframe diagnostics.
NOTE. If modules are installed in the mainframe when the TLA700 Mainframe Diagnostics are run, a Mainframe Diagnostics Warning dialog box appears. Clicking OK will initiate a SYSRESET, which will invalidate any acquired data in the modules. You may want to save acquired data before continuing.
5–2
QA+Win32
QA+Win32 is a comprehensive software application used to check and verify the operation of the PC hardware. To run the software, you must have either a keyboard, mouse, or other pointing device.
To run the software, follow these instructions:
1. Quit all other applications.
2. Click the Windows Start button.
3. Select QA+Win32 software.
4. Run the diagnostics.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Benchtop Controller Certification

The system clock is checked for accuracy. The instrument is certifiable if this parameter meets specifications. The procedure is described in the Performance Verification section. Record the results in the Calibration Data Report on the following page.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
5–3
Benchtop Controller Certification
5–4
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Calibration Data Report

TLA720 Color Benchtop Controller

Instrument model number: Serial number: Certificate number: Verification performed by: Verification date:

System Clock Test Data

Procedure
Characteristic Specification Tolerance
Clock frequency 10 MHz ±1 kHz (9.9990 MHz-10.0010 MHz) Page 5–8,
reference
Step 5
Incoming data Outgoing data

Benchtop Controller Performance Verification

This section contains procedures to verify that the TLA720 Benchtop Controller performs as warranted. Verify instrument performance whenever the accuracy or function of your instrument is in question.

Tests Performed

Table 5–2 lists the performance verification procedures available for the TLA720 Benchtop Controller.
T able 5–2: Performance verification procedures
Parameter Procedure
System clock (CLK 10)
1
Certifiable parameter
1
10 MHz system clock test

Test Procedures

Perform the following steps to complete the performance verification procedure. You will need some of the equipment shown in Table 1–3 on page 1–5 to complete the performance verification procedures. If you substitute equipment, always choose instruments that meet or exceed the minimum requirements specified
The following procedure checks the accuracy of the 10 MHz system clock.
Equipment required
Prerequisites Warm-up time: 30 minutes
1. Verify that all of the prerequisites above are met for the procedure.
2. Connect the frequency counter to the External Signal Out TTL SMB
connector on the benchtop controller.
3. Go to the System window and select System Configuration from the System
menu.
Frequency counter (item 15) SMB-to-BNC cable (item 20)
Power-up, mainframe, and QA+WIN32 diagnostics pass
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
5–7
Benchtop Controller Performance Verification
4. In the System Configuration dialog box, select 10 MHz Clock from the list
of routable signals in the External Signal Out selection box and click OK.
5. Verify that the output frequency at the External Signal Out TTL connector is
10 MHz ±1 kHz. Record the measurement on the Calibration Data Report and disconnect the frequency counter.
6. In the System Configuration dialog box, reset the External Signal Out signal
to None.
5–8
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures

Benchtop Controller Adjustment

There are no adjustments for the benchtop controller.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
5–9
Loading...