Tektronix Keithley AMM1 Analog Measurement Module Rev. B User manual

AMMI Analog Measurement Module
The AMMl Analog Measurement Module combines two important Series 500 functions into a single module: the AMMl performs analog signal conditioning and switching, and A/D conversion. The analog section of the module provides signal selection and programmable gain for both local and global analog signals connected to the Series 500. After analog conditioning, signals are routed to the AID converter section of the module for the analog-to-digital conversion process.
For A/D conversion, the AMMl utilizes a K&bit successive approximation converter that provides fast, accurate measurement and conversion. A maximum conversion time of only 25psec and a sample-and-hold acquisition time of 3~s allow sampling rates as high as 35.7kHz. To maximize resolution, the AMMl has five A/D converter ranges (three bipolar, two unipolar) that can be selected by on-card DIP switches.
The AMMl is designed to be used only in slot 1 of the system baseboard. To install the module, first remove the baseboard top cover and install the module in slot 1 with the component side facing the power supply.
CAUTION: Always turn off the system power before installing or removing modules. To minimize the possibility of EM1 radiation, always operate the system with the top cover in place and properly secured.
User-Configured Components
User-configured components for the AMMl include the input screw terminals and the switches that control A/D converter ranges, as surnmarized in Table 1. For the locations of these components, refer to Figure 1.
Table 1. AMMl User-Configured Components Description Designation Function
Screw Terminals DIP Switch Set SlOl A/D Converter range
All local input signals are applied to screw terminals, which are designed to accept 16-24 gage wire stripped 3116 of an inch.
Local inputs, channels O-7
Document Number: 500-910-01 F&v. B
AMMl-1
A/D CONVERTER
RANGE SELECTION
I
6
ii
D
AMMI LOCAL
INPUT TERMINALS
END C’H7 ‘i-IS c’ti5 C’H4
dH3
di2 C-H 1 C’HO
END
PI77
\ /
PI78
Switch SlOl controls the input range of the A/D converter located on the module. Available bipolar ranges include unipolar ranges are 0 to +lOV and 0 to +5V.
Connection Local input signals for channels 0 through 7 of the AMMl are applied to screw ter-
minals located at the back edge of the board. The channel numbers are marked on the board and are shown in Figure 1. Typical connections for channel 0 are shown in Figure 2. Note that the high side of the input signal is applied to the channel 0 ter­minal, and the low side of the signal is connected to module ground.
-10 to +lOV, -5 to +5V, and -2.5 to +2.5V. The
0
0 0 0 0 0 0 0
I
Figure 2. Typical Connection (Channel 0 Shown)
CAUTION: AMMl inputs are non-isolated, meaning that one side of the input is con­nected to power line ground. Any signal connected to the AMMl must also be referenced to power line ground, or module or system damage may occur. Also note the inaccuracies on other channels may result.
In many situations, shielded cable may be required to minimize EM1 radiation, or to keep noise to a minimum. If shielded cable is used, connect the shield to ground only,
and do not use the shield as a signal carrying lead. Usually, a module ground terminal should be used, but in some cases better results may be obtained by using one of the baseboard ground posts instead. Use the configuration that results in the lowest noise.
For shielding to be effective, the shield must contain both high and low signal wires,
and must not carry any other signals. If a number of AMMl signal input lines are shielded, all shields should be connected to the same ground terminal.
AMMl-3
A/D Converter Range Selection As shipped, the Ah&I1 is set up for the i-1OV range, but the module may be recon-
figured to one of four other ranges by setting the five DIP switches located on SlOl to the correct positions, as summarized in Table 2. To set the A/D converter to a specific range, first turn off system power and then set the switches to the correct positions, either open (off) or closed (on). For example, for the 0 to +5V range, switches 1, 3 and 4 should be closed (on), and switches 2 and 5 should be open (off).
NOTE: The module must be recalibrated if the range is changed. Turn to the Calibra­tion Section of this chapter for AMMl calibration information.
Table 2. 901 Settings for the A/D Converter Ranges
DIP Switches
Input Range 1
-10 to +lov*
-5 to +5v Open Open Closed Open Closed
-2.5 to +2.5V Open 0 to +5v 0 to +lov
*Factory default value NOTE: A/D converter must be recalibrated if range is changed.
Open Closed Open Open Closed
Closed Open Closed Open Closed Open Open
2
Open Closed Closed Closed
3
Closed Closed Open
4 5
Signal Conditioning A simplified block diagram of the AMMl is shown in Figure 3. The module is divided
into four general sections: a local multiplexer, a global multiplexer, a programmable gain amplifier (PGA) and a XI-bit AID converter.
Local input signals from channels 0 to 7 are applied to the local multiplexer for selec­tion. At any given time, only one channel will be selected, as determined by the SELECT CHANNEL command (covered later in this section). The signal from the selected channel is then routed to the global multiplexer for further signal selection and conditioning.
The global multiplexer selects a single signal from among the 10 slots in the signal. In this manner, signals from any of the Xl slots can be selected by software. The global multiplexer is controlled by the SELECT SLOT command, discussed later in this section.
After the signal is selected, the PGA applies software-selectable gains of xl, x2, x5, or x10. When this signal conditioning process is complete, the signal is routed to the l2-bit AID converter for digitization. After the conversion process, digital data representing the applied signal travels via the baseboard and interface card to the host computer.
AMMl-4
INPUTS
FROM OTHER AIM MODULES (9 MAXIMUM)
AMMI LOCAL
INPUTS (8)
1.
\‘.
\
\
r \
/
\
\-
GLOBAL
- MUX
,
xl, x2, x5OR x10
SOFTWARE
CONTROLLED GAIN
DIGITAL INFORMATION TO COMPUTER
Input Filtering Noise introduced into the input signal can corrupt the accuracy of the measurement.
Such noise will usually be seen as an unsteady reading that jumps around, or, in some cases, as a constant offset. In the former case, the presence of noise will usually be quite obvious, but its effects may not be noticeable in the steady-state offset situation. Regardless of the type of noise, however, such unwanted signals can degrade measure­ment accuracy considerably if enough of the unwanted signal is present.
Frequently, noise is introduced into the signal from 50 or 6OHz power sources. In many cases, such noise can be attenuated by shielding the input signal lines, as discussed earlier. In more difficult situations, however, it may be necessary to filter the input signal to achieve the necessary noise reduction.
When noise is a problem, a single-pole low-pass filter like the one shown in Fiie 4
can be conntected between the input signal and the corresponding AMMl channel. Note that the filter is made up of a single capacitor and resistor with the capacitor con­nected between the AMMl channel input terminal and the module
The resistor is then placed in series with the high input signal lead.
ground terminal.
1
27iRC
TO AMMl INPUT
7
FROM SIGNAL C
f
-3dB = -
T
Figure 4. Input Filtering
A common yardstick for a simple filter like the one in Figure 4 is the -3dB or half­power point, which is given as follows:
1
f-J, = -
2nRC
AMMl-6
where f is in Hz, C is in farads, and R is in ohms. Above this frequency, filter response will roll off (decrease) at a rate of -2OdB per decade. Thus, each time the frequency in­creases by a factor of lo, the filter output voltage decreases by a factor of 10 (-20dB).
Although such filtering can quiet down a noisy signal, there is a trade-off in the form of increased response time. This response time may be important in the case of a rapid­ly changing input signal. For the filter in Figure 4, the response time to 1% of final value is 4.6RC, while the response time to 0.1% and 0.01% of final value are 6.9RC and
9.2RC, respectively.
As an example, assume that 10 counts of 6OHz noise is present in the input signal. To reduce the noise to one count, an attenuation factor of I.0 (-20dB) at 6OHz will be necessary. Thus, the filter should have a -3dB point of 6Hz.
To determine the relative RC values, the above equations can be rearrange to solve for either R or C. If we wish to choose a nominal capacitor value and then solve for the resistance, we have:
1
R=
PnCf+fB
Choosing a nominal value of 2$ for C, the necessary resistance is:
1
R=
2n(2
x 10-6)x 6Hz
I? = 13.263k3
The resulting response times with these R and C values would be:
t(l%) = 4.6RC = 122ms t(O.l%) = 6.9RC = 183ms t(O.Ol%) = 9.2RC = 244ms
Note that there are a number of RC values that can be used in a given situation. To minimize the effects of the series resistance, however, it is recommended that the value of R be kept under 20kQ.
Current-to-Voltage Conversion AMMl local inputs are designed to accept voltages in the range of flOV. Thus, the
AMMl can be directly connected to many signal sources. Some transducers and in­strumentation, however, provide current outputs that must be converted into voltages in order to be measured through an AMMl input channel.
When connecting current inputs to the AMMl, a resistor should be installed across the input to make the necessary current-to-voltage conversion. One end of the resistor should be connected to the channel input terminals and the other end of the resistor should be connected to module ground.
AMMl-7
Loading...
+ 15 hidden pages