The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
that in all previously published material. Specifications and price change privileges reserved.
Printed in the U.S.A.
T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000
TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.
Service Reference
This manual contains the Performance Verification and Adjustment procedures
for many handheld products. Table i lists the products supported by this manual.
T able i: Handheld products supported by this manual
Products
CMM150 AC Current ProbeDMM249 Digital Multimeter
DCM300 and 320 Digital Clamp MultimetersDMM254 Digital Multimeter
DCM330 Digital Clamp MeterDMM800 Series Digital Multimeters
DCM910 Digital Clamp MeterDMM912, 914, and 916 Digital Multimeters
DMM150 Digital MultimeterDTM510 and DTM520 Digital Thermometers
DMM157 Digital Multimeter (serial numbers
above TW80000)
DTM900 and DTM920 Digital Thermometers
Service information of products not provided in this manual may be available
under separate cover. Table ii provides a list of many handheld and benchtop
instruments and their service documentation.
T able ii: Handheld and benchtop instruments service manuals
Review the following safety precautions to avoid injury and prevent damage to
the products in this manual or any products connected to them. To avoid
potential hazards, use these products only as specified.
Only qualified personnel should perform service procedures.
While using these products, you may need to access other parts of the system.
Read the General Safety Summary in other system manuals for warnings and
cautions related to operating the system.
To Avoid Fire or
Personal Injury
Use Proper Power Cord. To avoid fire hazard, use only the power cord specified
for the product you are using.
Use Proper V oltage Setting. Before applying power, ensure that the line selector is
in the proper position for the power source being used.
Connect and Disconnect Properly . Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. Some of these products are grounded through the grounding
conductor or the power cord. To avoid electric shock, the grounding conductor
must be connected to earth ground. Before making connections to the input or
output terminals of the product you are using, ensure that the product is properly
grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product you are using. Consult the product manual for
further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Replace Batteries Properly. Replace batteries only with the proper type and rating
specified.
Recharge Batteries Properly. Recharge batteries for the recommended charge
cycle only.
Use Proper AC Adapter. Use only the AC adapter specified for the product you
are using.
Handheld Instruments Basic Service
iii
General Safety Summary
Do Not Operate Without Covers. Do not operate these products with covers or
panels removed.
Use Proper Fuse. Use only the fuse type and rating specified for the product you
are using.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate in Wet/Damp Conditions. To avoid electric shock, do not operate
these products in wet or damp conditions.
Do Not Operate in Explosive Conditions. To avoid injury or fire hazard, do not
operate these products in an explosive atmosphere.
Keep Product Surfaces Clean and Dry . To avoid electric shock and erroneous
readings, keep probe surfaces clean and dry.
Symbols and Terms
Provide Proper Ventilation. Refer to the product installation instructions for details
on installing the product so it has proper ventilation.
T erms in this Manual. These terms may appear in this manual.
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
T erms on the Products. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
iv
Handheld Instruments Basic Service
General Safety Summary
Symbols on the Product. The following symbols may appear on the product:
DANGER
High Voltage
Protective Ground
(Earth) T erminal
ATTENTION
Refer to Manual
Double
Insulated
Handheld Instruments Basic Service
v
General Safety Summary
vi
Handheld Instruments Basic Service
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, disconnect the main power by means
of the power cord or, if provided, the power switch.
Use Caution When Servicing the CRT. To avoid electric shock or injury, use
extreme caution when handling the CRT. Only qualified personnel familiar with
CRT servicing procedures and precautions should remove or install the CRT.
CRTs retain hazardous voltages for long periods of time after power is turned off.
Before attempting any servicing, discharge the CRT by shorting the anode to
chassis ground. When discharging the CRT, connect the discharge path to ground
and then the anode. Rough handling may cause the CRT to implode. Do not nick
or scratch the glass or subject it to undue pressure when removing or installing it.
When handling the CRT, wear safety goggles and heavy gloves for protection.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
X-Radiation. To avoid x-radiation exposure, do not modify or otherwise alter the
high-voltage circuitry or the CRT enclosure. X-ray emissions generated within
this product have been sufficiently shielded.
Handheld Instruments Basic Service
vii
Service Safety Summary
viii
Handheld Instruments Basic Service
Preventing Electrostatic Discharge
CAUTION. Static discharge can damage internal semiconductor components.
Follow the guidelines listed below to avoid product damage.
When performing service that requires internal access to an instrument, adhere to
the following precautions to avoid damaging internal modules or their components:
HAvoid handling modules or components in areas that have floors or work
surfaces capable of generating a static charge.
HSpray carpeted work areas with a solution of equal parts of water and fabric
softener.
HWear clothing made from materials that do not accumulate static charges.
Avoid Wool (and some artificial fibers) which build up static charges readily;
wear cotton which conducts electricity and resists static accumulation.
HMinimize the handling of static-sensitive devices.
HTransport and store static-sensitive devices in their protected containers or on
a metal rail. Label any package that contains static-sensitive parts.
HService instruments and modules at grounded, static-free work stations.
HDo not allow devices capable of generating a static charge on a work station
surface.
HWear a grounding strap while working with static-sensitive devices.
HHandle circuit boards by their edges, if possible.
HDo not slide static-sensitive components over any surface.
HDo not use high-velocity compressed air to clean or dry components or
modules.
Handheld Instruments Basic Service
ix
Preventing Electrostatic Discharge
x
Handheld Instruments Basic Service
Preface
This manual contains service information for a wide range of handheld products.
Each section covers a product or related series of products and includes the
following information:
HA product description that details instrument functions, capabilities, and
recommended uses
HA front panel illustration
HA set of electrical, mechanical, environmental, and physical specifications
HA performance verification procedure to ensure the instrument meets
specifications
HAn adjustment procedure to return the instrument to factory calibration (not
included for all instruments)
The information contained in this manual is current at the date of publication and
is typical or suggested, not guaranteed.
Some instruments have optional service information available under separate
cover. This manual does not duplicate information from optional service
manuals. Refer to the optional accessories list in your user manual for Tektronix
part numbers of optional service manuals. A list of many handheld and benchtop
instruments manuals are listed in Table ii on Page i.
NOTE. This manual provides the necessary service information to verify that
your instrument is working properly. Should you have service-related questions
not covered in either this manual or in an optional service manual, please
contact your Tektronix Service Center for additional information.
For product warranty information, refer to the user manual supplied with your
instrument.
Handheld Instruments Basic Service
xi
Preface
Contacting Tektronix
Product
Support
Service
Support
For other
information
To write usTektronix, Inc.
For application-oriented questions about a Tektronix measurement product, call toll free in North America:
1-800-TEK-WIDE (1-800-835-9433 ext. 2400)
6:00 a.m. – 5:00 p.m. Pacific time
Or contact us by e-mail:
tm_app_supp@tek.com
For product support outside of North America, contact your
local Tektronix distributor or sales office.
Contact your local Tektronix distributor or sales office. Or visit
our web site for a listing of worldwide service locations.
http://www.tek.com
In North America:
1-800-TEK-WIDE (1-800-835-9433)
An operator will direct your call.
The CMM150 measures AC current (up to 300 A) when used with the DMM150
Digital Multimeter.
The current probe acts as a current to voltage transformer. The probe captures the
current induced magnetic fields around a conductor (wire) and converts the fields
to a proportional voltage that is displayed on the DMM150.
Figure 1: CMM150 AC Current Probe with DMM150 Digital Multimeter
Handheld Instruments Basic Service
1
CMM150 AC Current Probe
2
Handheld Instruments Basic Service
CMM150 Specifications
Accuracies are ±(% reading + number of digits) at 23_ C ±5_ C at less than 80%
R.H. (relative humidity).
T able 1: General specifications
CharacteristicsDescription
Operating temperature0 _C to 45 _C <75% R.H.
Storage temperature–20 _C to +60 _C <75% R.H.
T emperature coef ficient0.15 (specified accuracy) /_C
(<18 _C or >28 _C)
Dimensions (W L D) with holster72 mm 102 mm 36 mm
(2.83 in. 4.02 in. 1.42 in.)
Weight130.1 g (0.29 lbs)
Maximum altitude2000 m
Maximum jaw opening30 mm (1.18 in)
Maximum conductor size29 mm dia. (1.14 in.)
T able 2: Measurement characteristics
CharacteristicsDescription
Current range0.1 A to 300 A AC
Output voltage10 mV/A
Accuracy±(1.9% + 0.5 A) 50X60 Hz
±(3.9% + 1 A) 40X400 Hz
Maximum bare wire voltage600VAC
Sensor typeInduction coil
Maximum output impedance120 W
RMS
RMS
CAT II
T able 3: Certifications and compliances
EC Declaration of ConformityMeets intent of Low Voltage Directive 73/23/EEC for Product Safety. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the
European Communities
Low Voltage Directive 73/23/EEC:
EN 61010-1 Safety requirements for electrical equipment for
measurement, control, and laboratory use
CertificationsCertified by CSA to CAN/CSA–C22.22 No. 231 and UL1244
Handheld Instruments Basic Service
3
CMM150 Specifications
T able 3: Certifications and compliances (cont.)
Overvoltage CategoryCategory:Examples of Products in this Category:
CA T IIIDistribution-level mains, fixed installation
Local-level mains, appliances, portable equipment
Signal levels in special equipment or parts of equipment,
telecommunications, electronics
Pollution Degree 2
CA T II
CA T I
As defined in EN 61010-1. Do not operate in environments where conductive
pollutants may be present.
4
Handheld Instruments Basic Service
CMM150 Performance Verification
Á
Á
Á
Á
Á
Á
This section contains the procedure to verify that the CMM150 AC Current
Probe performs as warranted when installed on a DMM150 Digital Multimeter.
If an instrument fails any of these checks, it needs adjustment and or repair.
The performance verification procedure provides a valid confirmation of
instrument electrical characteristics and function under the following conditions:
Test Equipment
HThe CMM150 and DMM150 operate in an 18_ to 28_
C (64_ to 82_F)
ambient environment with a relative humidity of less than 75%.
HThe DMM150 stabilizes at the stated ambient temperature for one hour.
HThe DMM150 warms up for five minutes.
HThe DMM150 is set to the 3 V AC mode.
HAllow the DMM150 to settle to its final value before taking the measure-
ment.
The performance verification procedure should be performed annually or after
every 2000 hours of operation if used infrequently.
The performance verification procedure uses external traceable test equipment to
directly check warranted characteristics.
Alternative test equipment must meet or exceed the intended minimum
requirements specified in Table 4. If you substitute equipment, you may need to
modify the procedures.
NOTE. Before beginning the performance verification procedure, warm up the
test equipment according to the manufacturer’s recommendations.
T able 4: Test equipment
DescriptionMinimum requirementsExample product
Universal Calibration
System
Handheld Instruments Basic Service
ББББББ
ББББББ
Resolution & accuracy 4 times
БББББББ
greater than the multimeter
display reading
БББББББ
Wavetek 9100 with option 200
БББББББ
current coil set
БББББББ
5
CMM150 Performance Verification
Set Up
To prepare for the performance verification checks, do the following steps.
1. Attach the CMM150 to the DMM150 and allow them to stabilize at the
ambient temperature for one hour before testing.
Verification Procedure
AC Ampere Check
2. Turn the DMM150 on by pushing the slide switch to V
NOTE. You need to keep the DMM150 powered on throughout the warm-up
period and throughout the entire verification procedure.
3. Warm up the DMM150 for five minutes.
4. Photocopy the test record on page 7 to record your test results.
Implement the following checks to verify the performance of your CMM150 AC
Current Probe.
WARNING. To avoid electric shock, avoid touching exposed connections.
Perform the following steps to verify the AC ampere measurement accuracy.
V .
1. Set the DMM150 slide switch to V
2. Push the Blue function button to select AC volts.
3. Press the RANGE button to select the 3 V range.
4. Select the appropriate coils as necessary to multiply the AC current
calibrator output to each of the values given in the Test Record. For more
information, refer to the user manual of your calibrator.
5. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
6. Verify that the multimeter reads within the specified Display minimum and
maximum limits.
7. Turn the calibrator output off.
8. Remove the clamp from the current loop.
6
V .
Handheld Instruments Basic Service
CMM150 Performance Verification
CMM150 Test Record
Serial numberProcedure performed byDate
CMM150 test record
Test inputToleranceDisplay minimumReadingDisplay maximum
AC ampere test (50 Hz or 60 Hz)
0.1 A±1.9% + 0.5 A–.005.007
10.0 A±1.9% + 0.5 A.094.106
100.0 A±1.9% + 0.5 A.9761.024
300.0 A±1.9% + 0.5 A2.9383.062
Handheld Instruments Basic Service
7
CMM150 Performance Verification
8
Handheld Instruments Basic Service
CMM150 Adjustment Procedures
This section contains procedures to adjust the CMM150 AC Current Probe.
Perform these procedures once a year or if the CMM150 Performance Verifica-tion procedure indicates the need for calibration.
In this section you will find the following information:
HA list of adjustments
HA list of test equipment needed to make the adjustments
HInstructions on how to prepare the instrument for adjustment
HStep-by-step adjustment procedures
The procedures in this section do not verify performance. To confirm that your
multimeter meets factory specifications, perform the procedures in the CMM150Performance Verification section.
Test Equipment
Test Accessories
The test equipment listed in Table 4 on page 5 is a complete list of equipment
needed for the adjustment procedures. This procedure assumes that the test
equipment is operating within tolerance.
Alternative test equipment must meet or exceed the intended minimum requirements specified in Table 4. If you substitute equipment, you may need to modify
the procedures.
In addition to the test equipment, some additional test accessories are required to
perform the adjustment procedure.
T able 5: Test accessories
DescriptionQtyExample product
Couplers
Patch cords2ITT Pomona B8
2
ITT Pomona 5635
Handheld Instruments Basic Service
9
CMM150 Adjustment Procedures
Preparation for Adjustment
The following guidelines apply to all CMM150 adjustments:
Adjustment
HPerform all adjustments in a 21_ to 25_
relative humidity of 75% or less.
HWarm up the DMM150 for at least 15 minutes.
HDo not alter any setting without reading the entire adjustment procedure first.
HDo not alter a setting unless a performance characteristic cannot be met at the
current setting.
HRead the Safety Summary at the beginning of this manual.
1. Connect the CMM150 to the DMM150 as shown in Figure 2.
VR1
C ambient environment with a
Coupler
Figure 2: Adjustment location
2. Set the DMM150 to measure AC volts.
3. Select the appropriate coil to multiply the AC current calibrator output to
100.0 Amps at 60 Hz.
4. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
5. Adjust VR1 for a reading of 1.000.
6. Turn the calibrator output off.
7. Remove the clamp from the current loop.
Patch cord
10
Handheld Instruments Basic Service
Instructions Manual
DCM300 and DCM320
Digital Clamp Multimeters
070-9847-01
Table of Contents
DCM300 and DCM320 Digital Clamp Multimeters1. . . . . . . . . . . . . .
The DCM300 and DCM320 Digital Clamp multimeters measure AC current, AC
voltage, and resistance/continuity. The meters use a current transformer to
measure current without opening the circuit.
The meters automatically select the correct measurement range and have a 4000
count resolution. (The maximum reading is 3999.)
The DCM320 meter provides true RMS readings for both AC volts and AC
current.
Figure 1: DCM300 Digital Clamp Multimeter
Handheld Instruments Basic Service
1
DCM300 and DCM320 Digital Clamp Multimeters
2
Handheld Instruments Basic Service
DCM300 and DCM320 Specifications
The characteristics listed in this section apply under the following conditions:
HThe instrument operates in a 0_ to 45_ C ambient environment unless
otherwise noted.
HThe instrument warms up for at least 20 minutes.
NOTE. All specifications are warranted unless marked “typical.” Typical
characteristics are not guaranteed but are provided for the convenience of the
user.
T able 1: Electrical characteristics
CharacteristicDescription
AC Voltage, Auto Ranging (nominal Input Impedance: 10 MW, <100 pF)
Ranges600 V and 400 V
Voltage Rating600 V
Resolution
400 V Range0.1 V
600 V Range1 V
Accuracy40 to 500 Hz: ±(1.2% of reading + 5 digits)
Crest Factor (DCM 320 only)1.4 to 2.0, add 0.6% to accuracy
2.0 to 2.5, add 2.0% to accuracy
AC Current, Auto Ranging
Ranges600 A and 400 A
Overload Protection800 A
Uninsulated Wire Voltage Rating600 V
Resolution
400 A Range0.1 A
600 A Range1 A
Accuracy50 to 60 Hz
400 A Range±(1.9% of reading + 5 digits)
600 A Range±(2.9% of reading + 5 digits)
RMS
RMS
CAT II
CAT II
Crest Factor (DCM 320 only)1.4 to 2.0, add 1.0% to accuracy
Handheld Instruments Basic Service
2.0 to 2.5, add 2.5% to accuracy
3
DCM300 and DCM320 Specifications
C
1010.1
L111-1M
T able 1: Electrical characteristics (cont.)
CharacteristicDescription
Resistance, Auto ranging (meter beeps if resistance is <100 W.)
Ranges4 kW and 40 kW
Overload Protection600 V
Resolution
4 kW Range1 W
40 kW Range10 W
Accuracy±(2.0% of reading + 9 digits)
Maximum Open Circuit Voltage1 V
RMS
T able 2: General specifications
CharacteristicDescription
Auto Power OffApproximately 30 minutes
Battery9 V, ANSI/NEDA1604A, IEC 6F22
Battery Life200 hours (alkaline)
Maximum Conductor Size40 mm
T able 3: Certifications and compliances
Certifications
Overvoltage CategoryCategory:Examples of Products in this Category:
Pollution Degree 2Do not operate in environments where conductive pollutants may be present.
4
anadian Standards Association certified to Standard CSA
UL31 11-1 for Electrical and Electronic Measuring and Testing Equipment, and
IEC1010-2-032 particular requirements for hand-held current clamps for electrical
measurement and test.
CA T IIIDistribution-level mains, fixed installation
CA T IILocal-level mains, appliances, portable equipment
CA T ISignal levels in special equipment or parts of equipment,
telecommunications, electronics
, Standard
Handheld Instruments Basic Service
T able 4: Environmental characteristics
CharacteristicDescription
Temperature
Operating
Nonoperating
T emperature Coef ficient
Maximum Altitude (Operating)2,200 m (7,218 ft.)
0_ to 45° C (32_ to 113° F), <75% relative humidity
–20_ to +60° C (–4_ to 140° F), <80% relative humidity
0.2% (specified accuracy) per _C at <18_ C (64_ F) or >28_ C (82_ F)
DCM300 and DCM320 Specifications
Handheld Instruments Basic Service
5
DCM300 and DCM320 Specifications
6
Handheld Instruments Basic Service
DCM300 and DCM320 Performance Verification
This section contains procedures to verify that the DCM300 and DCM320
Digital Clamp Multimeters perform as warranted. If an instrument fails any of
the checks, it needs adjustment and or repair.
The performance verification procedures provide a valid confirmation of
instrument electrical characteristics and function under the following conditions:
HThe instrument operates in an 18_ to 28_ C ambient environment with a
relative humidity of less than 75%.
HThe instrument warms up for 20 minutes.
HThe instrument remains fully assembled (do not remove the bottom cover).
The DCM300 and DCM320 performance verification consists of the checks
listed in Table 5.
T able 5: Performance verification checks
AC Current Check
Test Equipment
AC Voltage Check
Resistance and Continuity Check
The performance verification procedure should be performed annually or after
every 2000 hours of operation if used infrequently.
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics.
Alternative test equipment must meet or exceed the intended minimum
requirements specified in Table 6. If you substitute equipment, you may need to
modify the procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturer’s recommendations.
Handheld Instruments Basic Service
7
DCM300 and DCM320 Performance Verification
100
00
5500
5500
T able 6: Test equipment
DescriptionMinimum requirementsExample product
AC Current Calibrator>0.5% accuracy, 0 to 400 A
AC Voltage Calibrator>0.2% accuracy
Resistance Calibrator>0.3% accuracy
Set Up
To prepare for the performance verification checks, do the following.
1. Turn the DCM300/DCM320 Digital Clamp Multimeter on by sliding the
function switch to any position other than OFF.
2. Warm up the instrument for 20 minutes.
>0.7% accuracy, 400 to 600 A
Wavetek 9
current multiplier coils or Fluke
A with Wavetek X10 and
A with Wavetek X10 and
X50 Current multiplier Coils
with option 2
Verification Procedure
AC Current Check
3. Photocopy the test record on pages 11 and 12 to record your test results.
The following checks verify the performance of your DCM300 or DCM320
multimeter.
WARNING. The following procedures produce magnetic fields that may cause a
malfunction in heart pacemakers or damage to sensitive equipment.
To check the AC current accuracy, perform the following steps.
1. Set the multimeter function switch to the A
position.
2. Set up the AC current calibrator to output the values in the AC current test
record.
NOTE. Select the appropriate coils to multiply the AC Current calibrator output
for each of the values listed in the AC current test record.
8
3. For each of the conditions listed in the AC current test record, position the
clamp around the current loop of the AC current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
Handheld Instruments Basic Service
DCM300 and DCM320 Performance Verification
4. Verify that the multimeter display reads within the specified Display
minimum and maximum limits for each of the specified conditions.
5. Turn the calibrator output off.
6. Remove the clamp from the current loop.
AC Voltage Check
Resistance and Continuity
Check
To check the AC voltage accuracy, perform the following steps.
WARNING. To avoid electric shock, avoid touching the exposed connections on
the multimeter circuit board.
1. Set the multimeter function switch to the V
2. Connect the AC voltage calibrator output to the multimeter V–W and COM
input terminals.
3. Set the calibrator to each of the values listed in the AC voltage test record
and verify that the multimeter display reads within the specified Display
minimum and maximum limits.
4. Turn the calibrator output off.
5. Disconnect the calibrator.
To check the resistance accuracy and verify the continuity function, perform the
following steps.
1. Set the multimeter function switch to the W
position.
position.
2. Connect the resistance calibrator output to the multimeter V–W and COM
3. Set the calibrator to each of the values listed in the Resistance and continuity
4. Turn the calibrator output off.
5. Disconnect the calibrator.
Handheld Instruments Basic Service
input terminals.
test record and verify that the display reads within the specified Display
minimum and maximum limits.
9
DCM300 and DCM320 Performance Verification
10
Handheld Instruments Basic Service
DCM300 and DCM320 Performance Verification
DCM300 and DCM320 Test Record
Serial numberProcedure performed byDate
DCM300 and DCM320 test record
Test inputToleranceDisplay minimumReadingDisplay maximum
Test inputDisplay maximumReadingDisplay minimumTolerance
Resistance and continuity test
W±9 W000 W009 W
Buzzer must sound
W±11 W109 W131 W
Buzzer must sound
1.000 kW±0.029 kW971 W1.029 kW
3.700 kW±0.083 kW3.617 kW3.783 kW
39.00 kW±0.87 kW38.13 kW39.87 kW
12
Handheld Instruments Basic Service
DCM300 and DCM320 Adjustment Procedures
This section contains procedures to adjust DCM300 and DCM320 Digital Clamp
multimeters. If your instrument fails a performance requirement, use these
procedures to return it to factory specifications.
In this section you will find the following information:
HA list of adjustments
HA list of test equipment needed to make the adjustments
HInstructions on how to prepare the instrument for adjustment
HStep-by-step adjustment procedures
The procedures in this section do not verify performance. To confirm that your
multimeter meets factory specifications, implement the procedures in the
DCM300 and DCM320 Performance Verification section.
List of Adjustments
Test Equipment
Use the adjustments listed in Table 7 to return DCM300 and DCM320 multimeters to factory calibration.
T able 7: DCM300 and DCM320 adjustments
AC Current
AC Voltage
Resistance
Continuity
The test equipment listed in Table 6 on page 8 is a complete list of equipment
needed for the adjustment procedures. These procedures assume that all test
equipment is operating within tolerance. Detailed operating instructions for test
equipment are not given in this procedure. If you need operating information,
refer to the instruction manual of the test equipment.
Alternative test equipment must meet or exceed the intended minimum
requirements specified in Table 6. If you substitute equipment, you may need to
modify the procedures.
Handheld Instruments Basic Service
13
DCM300 and DCM320 Adjustment Procedures
Preparation for Adjustment
The following guidelines apply to all DCM300 & DCM320 adjustments.
HPerform all adjustments in a 21_ to 25_ C ambient environment with a
relative humidity of 75% or less.
HBefore making any adjustment, warm up the multimeter for 20 minutes.
HDo not alter any setting without reading the entire adjustment procedure first.
HDo not alter a setting unless a performance characteristic cannot be met at
the current setting.
HRead the Safety Summary at the beginning of this manual.
Open the Meter Case
Adjustment Procedure
You must open the meter case to gain access to the internal adjustments.
1. Lay the meter face down on a flat work surface.
2. Remove the two screws from the meter bottom with a Phillips-head
screwdriver.
3. Gently lift the end of the bottom cover until it unsnaps from the top cover.
Do not remove the circuit board mounting screws.
To reassemble the meter following the adjustments, perform steps 2 and 3 above
in reverse order.
To return your instrument to factory calibration, perform the following procedures.
WARNING. The following procedures produce magnetic fields that may cause a
malfunction in heart pacemakers or damage to sensitive equipment.
14
AC Current
To adjust the AC current calibration, perform the following steps.
1. Set the AC current calibrator to output 100 A at 50 Hz.
2. Set the multimeter function switch to the A position.
3. Select the appropriate coil to multiply the AC current calibrator output to
100 A at 50 Hz.
Handheld Instruments Basic Service
DCM300 and DCM320 Adjustment Procedures
4. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
5. Adjust VR2 with a small flat-tipped screwdriver; set the multimeter reading
to 100.0.
6. Turn the calibrator output off.
7. Remove the clamp from the current loop.
AC Voltage
Resistance
To adjust the AC voltage calibration, perform the following steps.
WARNING. To avoid electrical shock, avoid touching the exposed connections on
the multimeter circuit board.
1. Set the multimeter function switch to the V
2. Connect the AC voltage calibrator output to the multimeter V–W and COM
input terminals.
3. Set the AC voltage calibrator to output 300 V at 500 Hz (DCM300) or 300 V
at 50 Hz (DCM320).
4. Adjust VR1 with a small flat-tipped screwdriver; set the multimeter reading
to 300.0.
5. Turn the calibrator output off.
6. Disconnect the calibrator from the multimeter.
To adjust the resistance calibration, perform the following steps.
1. Set the multimeter function switch to the W
position.
position.
2. Connect the resistance calibrator output to the multimeter V–W and COM
3. Set the resistance calibrator to simulate a 1 kW resistance load.
4. Adjust VR3 with a small flat-tipped screwdriver; set the multimeter reading
5. Turn the calibrator output off.
6. Disconnect the calibrator from the multimeter.
Handheld Instruments Basic Service
input terminals.
to 1.000.
15
DCM300 and DCM320 Adjustment Procedures
Continuity
To adjust the continuity calibration, perform the following steps.
1. Connect the resistance calibrator output to the multimeter V–W and COM
input terminals.
2. Set the resistance calibrator to simulate a 150 W resistance load.
3. Place the bottom cover back on the meter and hold it in place. (The
continuity buzzer will not sound during the following adjustments without
the cover in place.)
4. If the buzzer does not sound, use a small flat-tipped screwdriver to adjust
VR4 until the buzzer sounds. You will have to remove the bottom cover to
make the adjustment and then replace the cover to make the test.
5. If the buzzer does sound, use a small flat-tipped screwdriver to adjust VR4
until the buzzer does not sound. After that, use the screwdriver to adjust VR4
until the buzzer sounds again. (Remove the cover to adjust; replace the cover
to test.)
6. When you complete all adjustments, turn the multimeter off and replace the
bottom cover. Do not pinch the battery leads between the case halves during
reassembly.
16
VR3
VR4
VR1
VR2
Figure 2: Adjustment locations
Handheld Instruments Basic Service
T able 8: Summary of adjustments
DCM300 and DCM320 Adjustment Procedures
Adjustment
name
AC Current100.0 A50 HzVR2±1.0 A99.0101.0
AC Volts300.0 V500 Hz (DCM300)
Ohm
Test valueRange setting
50 Hz (DCM320)
1.000 kWVR3±1 W999 W1.001 kW
150 WVR4Adjust VR4 until the buzzer just sounds
The DCM330 Digital Clamp Meter measures DC current, AC current, and
frequency. The meter uses a Hall-effect device to measure current without
opening the circuit.
The meter automatically selects the correct measurement range and has a 4000
count resolution. (The maximum reading is 3999.)
The DCM330 meter provides true RMS readings for AC current.
Figure 1: DCM330 Digital Clamp Meter
Handheld Instruments Basic Service
1
DCM330 Digital Clamp Meter
2
Handheld Instruments Basic Service
DCM330 Specifications
The characteristics listed in this section apply under the following conditions:
HThe instrument operates in a 0° to 50° C (32° to 122° F) ambient environ-
ment unless otherwise noted.
HThe instrument warms up for at least 20 minutes.
NOTE. All specifications are warranted unless marked “typical.” Typical
characteristics are not guaranteed but are provided for the convenience of the
user.
T able 1: Electrical characteristics
CharacteristicDescription
Overload Protection2000 A for one minute
Uninsulated Wire Voltage600 V
Measuring Rate2 times per second nominal
AC Current, Auto Ranging
Ranges400 A and 1000 A
Uninsulated Wire Voltage Rating600 V
Resolution
400 A Range0.1 A
1000 A Range1 A
Accuracy40 Hz to 400 Hz
0 A to 400 A±(1.9% of reading + 8 counts)
401 A to 1000 A±(2.9% of reading + 5 counts)
Crest Factor1.4 to 2.0, add 1.0% to accuracy
2.0 to 2.5, add 2.5% to accuracy
DC Current, Auto Ranging
Ranges400 A and 1000 A
Resolution
400 A Range0.1 A
1000 A Range1 A
RMS
RMS
CAT II
CAT II
Handheld Instruments Basic Service
3
DCM330 Specifications
T able 1: Electrical characteristics (cont.)
CharacteristicDescription
Accuracy
0 A to 20 A±(1.9% of reading + 10 counts)
20.1 A to 400 A±(1.9% of reading + 40 counts)
401 A to 1000 A±(2.9% of reading + 5 counts)
Frequency, Auto Ranging
Ranges4 kHz and 10 kHz
Sensitivity6 A
Resolution
4 kHz Range1 Hz
10 kHz Range10 Hz
Accuracy±(0.5% of reading + 3 counts)
Peak Hold
RangeLow, High
Resolution
Low0.1 A
High1.0 A
Accuracy±(3% of reading + 10 counts)
RMS
(10 A
RMS
, 1kHz to 10 kHz)
T able 2: General specifications
CharacteristicDescription
Auto Power OffApproximately 30 minutes
Battery9 V, ANSI/NEDA1604A, IEC 6F22
Battery Life40 hours (alkaline)
Maximum Conductor Size51 mm (2 inch) diameter or 24 × 60 mm (.95 × 2.36 inch) bus bar
4
Handheld Instruments Basic Service
T able 3: Certifications and compliances
C
1010.1
L111-1M
DCM330 Specifications
Certifications
Overvoltage CategoryCategory:Examples of Products in this Category:
Pollution Degree 2Do not operate in environments where conductive pollutants may be present.
anadian Standards Association certified to Standard CSA
UL31 11-1 for Electrical and Electronic Measuring and Testing Equipment, and
IEC1010-2-032 particular requirements for hand-held current clamps for electrical
measurement and test.
CA T IIIDistribution-level mains, fixed installation
CA T IILocal-level mains, appliances, portable equipment
CA T ISignal levels in special equipment or parts of equipment,
telecommunications, electronics
, Standard
T able 4: Environmental characteristics
CharacteristicDescription
Temperature
Operating
Nonoperating
T emperature Coef ficient
Maximum Altitude (Operating)2,000 m (6,562 ft)
0° to 50° C (32° to 122° F), <75% relative humidity
–20° C to +60° C (21° to 140° F), <80% relative humidity
0.2 × (specified accuracy) per °C at <18° C or >28° C
Handheld Instruments Basic Service
5
DCM330 Specifications
6
Handheld Instruments Basic Service
DCM330 Performance Verification
This section contains procedures to verify that the DCM330 Digital Clamp
Meter performs as warranted. If an instrument fails any of the checks, it needs
adjustment and or repair.
The performance verification procedures provide a valid confirmation of
instrument electrical characteristics and function under the following conditions:
HThe instrument operates in an 18_ to 28_ C (64_ to 82_ F) ambient
environment with a relative humidity of less than 75%.
HThe instrument warms up in the ambient environment for at least one hour.
HThe instrument remains fully assembled (do not remove the bottom cover).
The DCM330 performance verification consists of the checks listed in Table 5.
T able 5: Performance verification checks
AC Current Check
DC Current Check
Test Equipment
Frequency Check
The performance verification procedure should be performed annually or after
every 2000 hours of operation if used infrequently.
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics.
Alternative test equipment must meet or exceed the intended minimum
requirements specfied in Table 6. If you substitute equipment, you may need to
modify the procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturer’s recommendations.
Handheld Instruments Basic Service
7
DCM330 Performance Verification
100
T able 6: Test equipment
Set Up
To prepare for the performance verification checks, do the following.
1. Turn the DCM330 Digital Clamp Meter on.
2. Warm up the meter for 20 minutes.
3. Photocopy the test record on pages 11 and 12 to record your test results.
DescriptionMinimum requirementsExample product
AC/DC Current Calibrator>0.5 % accuracy 0 to 400 A
>0.7 % accuracy 400 to 1000 A
Wavetek 9
200 current multiplier coils
with Option
Verification Procedure
AC Current Check
The following checks verify the performance of your DCM330 meter.
WARNING. The following procedures produce magnetic fields that may cause a
malfunction in heart pacemakers or damage to sensitive equipment.
To check the AC current accuracy, perform the following steps.
1. Set the meter function to AC.
2. Select the appropriate coils as necessary to multiply the AC current
calibrator output to each of the test values in the AC current test record. For
more information, refer to the user manual of your calibrator.
3. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
4. Verify that the display reads within the specified Display minimum and
maximum limits.
5. Turn the calibrator output off.
8
6. Remove the clamp from the current loop.
Handheld Instruments Basic Service
DCM330 Performance Verification
DC Current Check
To check the DC current accuracy, perform the following steps.
1. Set the meter function to DC.
2. In the absence of any magnetic fields, press the DCA AUTO ZERO button
to zero the meter.
3. Select the appropriate coil(s) as necessary to multiply the DC current
calibrator output to each of the test values in the DC current test record. For
more information, refer to the user manual of your calibrator.
4. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
5. Verify that the display reads within the specified Display minimum and
maximum limits.
6. Before each measurement, set the calibrator output to off and press the DCA
AUTO ZERO button to zero the meter.
NOTE. Any time a measurement appears to be out of tolerance, turn the
calibrator output off, rezero the meter, and try again.
7. Turn the calibrator output off.
8. Disconnect the calibrator.
Handheld Instruments Basic Service
9
DCM330 Performance Verification
Frequency Check
To check the frequency accuracy, perform the following steps.
1. Set the meter function to Hz.
2. Select the appropriate coil as necessary to multiply the AC current calibrator
output to 20 A.
3. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
4. Verify that the display reads within the specified Display minimum and
maximum limits for each of the frequencies listed in the Frequency test
record.
5. Turn the calibrator output off.
6. Disconnect the calibrator.
10
Handheld Instruments Basic Service
DCM330 Performance Verification
DCM330 Test Record
Serial numberProcedure performed byDate
DCM330 test record
Test inputToleranceDisplay minimumReadingDisplay maximum
At these frequencies, the inductance of the DCM330 may shut down the output of some calibrators. If this happens,
decrease the calibrator output frequency until the output remains on for the duration of the test.
Handheld Instruments Basic Service
11
DCM330 Performance Verification
DCM330 test record (cont.)
Test inputDisplay maximumReadingDisplay minimumTolerance
This section contains procedures to adjust the DCM330 Digital Clamp Meter. If
your instrument fails a performance requirement, use these procedures to return
it to factory specifications.
In this section you will find the following information:
HA list of adjustments
HA list of test equipment needed to make the adjustments
HInstructions on how to prepare the instrument for adjustment
HStep-by-step adjustment procedures
The procedures in this section do not verify performance. To confirm that your
multimeter meets factory specifications, perform the procedures in the DCM330Performance Verification section.
List of Adjustments
Test Equipment
Use the adjustments listed in Table 7 to return the DCM330 clamp meter to
factory calibration.
T able 7: DCM330 adjustments
Position Error
AC Current
DC Current
Peak Hold
The test equipment listed in Table 6 on page 8 is a complete list of equipment
needed for the adjustment procedures. These procedures assume that all test
equipment is operating within tolerance. Detailed operating instructions for test
equipment are not given in this procedure. If you need operating information,
refer to the instruction manual of the test equipment.
Alternative test equipment must meet or exceed the intended minimum
requirements specfied in Table 6. If you substitute equipment, you may need to
modify the procedures.
Handheld Instruments Basic Service
13
DCM330 Adjustment Procedures
Preparation for Adjustment
The following guidelines apply to all DCM330 adjustments.
HPerform all adjustments in a 21_ to 25_ C ambient environment with a
HBefore making any adjustment, warm up the current meter for at least
HDo not alter any setting without reading the entire adjustment procedure first.
HDo not alter a setting unless a performance characteristic cannot be met at
HRead the Safety Summary at the beginning of this manual.
relative humidity of 75% or less.
30 minutes.
the current setting.
Open the Meter Case
Adjustment Procedure
You must open the meter case to gain access to the internal adjustments.
1. Lay the meter face down on a flat work surface.
2. Remove the two screws from the case bottom with a Phillips-head
screwdriver.
3. Gently lift the end of the case bottom until it unsnaps from the case top.
4. Remove the three screws that secure the circuit board assembly to the case
top. Do not remove the screws that secure the circuit boards to each other.
5. To access the adjustments, lift the circuit board assembly far enough out of
the top case to expose the adjustments. See Figure 2 and the procedure that
follows.
To reassemble the meter following the adjustments, perform steps 2 through 4
above in reverse order.
To return your instrument to factory calibration, implement the following
procedures.
14
Use a small flat-tipped screwdriver to make the adjustments. Refer to Figure 2
for adjustment locations.
WARNING. Magnetic fields are produced that may cause a malfunction in heart
pacemakers, or damage to sensitive equipment.
Handheld Instruments Basic Service
VR1
VR2
VR3
VR5
VR6
VR7
VR8
DCM330 Adjustment Procedures
Short for
Auto Zero
(VR2)
Short for
Peak Hold
(VR9)
VR9
VR10
Position Error
DC Auto Zero
Figure 2: Adjustment locations
To adjust the position error calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Select the appropriate coil to multiply the AC current calibrator output to
380 A at 50 Hz.
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Adjust VR1 to maintain the measurement error to less than 1% total while
positioning the coil in the clamp.
5. Turn the calibrator output off.
6. Remove the clamp meter from the coil.
To adjust the DC zero calibration, perform the following steps.
1. Set the clamp meter to the DC position.
2. Short the Auto Zero points indicated in Figure 2.
Handheld Instruments Basic Service
15
DCM330 Adjustment Procedures
3. Adjust VR2 until the display reads 00.0 ± 5 counts.
4. Remove the short.
5. Press the clamp meter DCA AUTO ZERO button to zero the display.
6. Adjust VR3 until the display reads 00.0.
DC 400 A Range
DC 1000 A Range
To adjust the DC 400 A range calibration, perform the following steps.
1. Set the clamp meter to the DC position.
2. Select the appropriate coil to multiply the DC current calibrator output to
200 A.
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Position the clamp to the center of the coil.
5. Adjust VR5 until the display reads 201.5.
6. Turn the calibrator output off.
7. Remove the clamp meter from the coil.
To adjust the DC 1000 A range calibration, perform the following steps.
1. Set the clamp meter to the DC position.
2. Select the appropriate coil to multiply the DC current calibrator output to
400 A.
16
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Press the clamp meter DCA AUTO ZERO button to zero the display.
5. Position the clamp to the center of the coil.
6. Adjust VR6 until the display reads 400.
7. Turn the calibrator output off.
8. Remove the clamp meter from the coil.
Handheld Instruments Basic Service
DCM330 Adjustment Procedures
AC 400 A Range
AC 1000 A Range
To adjust the AC 400 A range calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Select the appropriate coil to multiply the AC current calibrator output to
390 A at 400 Hz.
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Position the clamp to the center of the coil.
5. Adjust VR8 until the display reads 396.0. To keep the meter on the lower
range, it may be necessary to cycle the calibrator output off and on.
6. Turn the calibrator output off.
7. Remove the clamp meter from the coil.
To adjust the AC 1000 A range calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Select the appropriate coil to multiply the AC current calibrator output to
400 A at 400 Hz.
Peak Hold
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Position the clamp to the center of the coil.
5. Adjust VR7 until the display reads 400.
6. Turn the calibrator output off.
7. Remove the clamp meter from the coil.
To adjust the peak hold calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Short the Peak Hold points indicated in Figure 2.
3. Press PEAK HOLD to activate the function.
4. Adjust VR9 until the display reads 00.0.
5. Remove the short.
6. Press PEAK HOLD to cancel the function.
7. Press PEAK HOLD again to verify that the display reads 00.0.
Handheld Instruments Basic Service
17
DCM330 Adjustment Procedures
T able 8: Summary of adjustments
8. Press PEAK HOLD to cancel the function.
9. Repeat steps 2 through 8 above until the display reads 00.0.
10. Select the appropriate coil to multiply the AC current calibrator output to
200 A at 400 Hz.
11. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
12. Position the clamp to the center of the coil.
13. Adjust VR10 until the display reads 200.0.
14. Turn the calibrator output off.
15. Remove the clamp meter from the coil.
16. Reassemble the meter.
Adjustment
name
Position ErrorAC380 A50 HzVR1<5 counts0 count5 counts
DC ZeroDCVR2
The DCM910 Digital Clamp Meter measures DC current, AC current, and
frequency. The meter uses a Hall-effect device to measure current without
opening the circuit.
The meter automatically selects the correct measurement range and has a 4000
count resolution. (The maximum reading is 3999.)
The DCM910 meter provides true RMS readings for AC current.
Figure 1: DCM910 Digital Clamp Meter
Handheld Instruments Basic Service
1
DCM910 Digital Clamp Meter
2
Handheld Instruments Basic Service
DCM910 Specifications
The characteristics listed in this section apply under the following conditions:
HThe instrument operates in a 0° to 50° C (32° to 122° F) ambient environ-
ment unless otherwise noted.
HThe instrument warms up for at least 20 minutes.
NOTE. All specifications are warranted unless marked “typical.” Typical
characteristics are not guaranteed but are provided for the convenience of the
user.
T able 1: Electrical characteristics
CharacteristicDescription
Overload Protection2000 A for one minute
Uninsulated Wire Voltage600 V
Measuring Rate2 times per second nominal
AC Current, Auto Ranging
Ranges400 A and 1000 A
Uninsulated Wire Voltage Rating600 V
Resolution
400 A Range0.1 A
1000 A Range1 A
Accuracy40 Hz to 400 Hz
0 A to 400 A±(1.9% of reading + 8 counts)
401 A to 1000 A±(2.9% of reading + 5 counts)
Crest Factor1.4 to 2.0, add 1.0% to accuracy
2.0 to 2.5, add 2.5% to accuracy
DC Current, Auto Ranging
Ranges400 A and 1000 A
Resolution
400 A Range0.1 A
1000 A Range1 A
RMS
RMS
CAT II
CAT II
Handheld Instruments Basic Service
3
DCM910 Specifications
T able 1: Electrical characteristics (cont.)
CharacteristicDescription
Accuracy
0 A to 20 A±(1.9% of reading + 10 counts)
20.1 A to 400 A±(1.9% of reading + 40 counts)
401 A to 1000 A±(2.9% of reading + 5 counts)
Frequency, Auto Ranging
Ranges4 kHz and 10 kHz
Sensitivity6 A
Resolution
4 kHz Range1 Hz
10 kHz Range10 Hz
Accuracy±(0.5% of reading + 3 counts)
Peak Hold
RangeLow, High
Resolution
Low0.1 A
High1.0 A
Accuracy±(3% of reading + 10 counts)
RMS
(10 A
RMS
, 1kHz to 10 kHz)
T able 2: General specifications
CharacteristicDescription
Auto Power OffApproximately 30 minutes
Battery9 V, ANSI/NEDA1604A, IEC 6F22
Battery Life40 hours (alkaline)
Maximum Conductor Size51 mm (2 inch) diameter or 24 × 60 mm (.95 × 2.36 inch) bus bar
4
Handheld Instruments Basic Service
T able 3: Certifications and compliances
C
1010.1
L111-1M
DCM910 Specifications
Certifications
Overvoltage CategoryCategory:Examples of Products in this Category:
Pollution Degree 2Do not operate in environments where conductive pollutants may be present.
anadian Standards Association certified to Standard CSA
UL31 11-1 for Electrical and Electronic Measuring and Testing Equipment, and
IEC1010-2-032 particular requirements for hand-held current clamps for electrical
measurement and test.
CA T IIIDistribution-level mains, fixed installation
CA T IILocal-level mains, appliances, portable equipment
CA T ISignal levels in special equipment or parts of equipment,
telecommunications, electronics
, Standard
T able 4: Environmental characteristics
CharacteristicDescription
Temperature
Operating
Nonoperating
T emperature Coef ficient
Maximum Altitude (Operating)2,000 m (6,562 ft)
0° to 50° C (32° to 122° F), <75% relative humidity
–20° C to +60° C (21° to 140° F), <80% relative humidity
0.2 × (specified accuracy) per °C at <18° C or >28° C
Handheld Instruments Basic Service
5
DCM910 Specifications
6
Handheld Instruments Basic Service
DCM910 Performance Verification
This section contains procedures to verify that the DCM910 Digital Clamp
Meter performs as warranted. If an instrument fails any of the checks, it needs
adjustment and or repair.
The performance verification procedures provide a valid confirmation of
instrument electrical characteristics and function under the following conditions:
HThe instrument operates in an 18_ to 28_ C (64_ to 82_ F) ambient
environment with a relative humidity of less than 75%.
HThe instrument warms up in the ambient environment for at least one hour.
HThe instrument remains fully assembled (do not remove the bottom cover).
The DCM910 performance verification consists of the checks listed in Table 5.
T able 5: Performance verification checks
AC Current Check
DC Current Check
Test Equipment
Frequency Check
The performance verification procedure should be performed annually or after
every 2000 hours of operation if used infrequently.
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics.
Alternative test equipment must meet or exceed the intended minimum
requirements specified in Table 6. If you substitute equipment, you may need to
modify the procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturer’s recommendations.
Handheld Instruments Basic Service
7
DCM910 Performance Verification
100
T able 6: Test equipment
Set Up
To prepare for the performance verification checks, do the following.
1. Turn the DCM910 Digital Clamp Meter on.
2. Warm up the meter for 20 minutes.
3. Photocopy the test record on pages 11 and 12 to record your test results.
DescriptionMinimum requirementsExample product
AC/DC Current Calibrator>0.5 % accuracy 0 to 400 A
>0.7 % accuracy 400 to 1000 A
Wavetek 9
200 current multiplier coils
with Option
Verification Procedure
AC Current Check
The following checks verify the performance of your DCM910 meter.
WARNING. The following procedures produce magnetic fields that may cause a
malfunction in heart pacemakers or damage to sensitive equipment.
To check the AC current accuracy, perform the following steps.
1. Set the meter function to AC.
2. Select the appropriate coils as necessary to multiply the AC current
calibrator output to each of the test values in the AC current test record. For
more information, refer to the user manual of your calibrator.
3. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
4. Verify that the display reads within the specified Display minimum and
maximum limits.
5. Turn the calibrator output off.
8
6. Remove the clamp from the current loop.
Handheld Instruments Basic Service
DCM910 Performance Verification
DC Current Check
To check the DC current accuracy, perform the following steps.
1. Set the meter function to DC.
2. In the absence of any magnetic fields, press the DCA AUTO ZERO button
to zero the meter.
3. Select the appropriate coil(s) as necessary to multiply the DC current
calibrator output to each of the test values in the DC current test record. For
more information, refer to the user manual of your calibrator.
4. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
5. Verify that the display reads within the specified Display minimum and
maximum limits.
6. Before each measurement, set the calibrator output to off and press the DCA
AUTO ZERO button to zero the meter.
NOTE. Any time a measurement appears to be out of tolerance, turn the
calibrator output off, rezero the meter, and try again.
7. Turn the calibrator output off.
8. Remove the clamp from the current loop.
Handheld Instruments Basic Service
9
DCM910 Performance Verification
Frequency Check
To check the frequency accuracy, perform the following steps.
1. Set the meter function to Hz.
2. Select the appropriate coil as necessary to multiply the AC current calibrator
output to 20 A.
3. Position the clamp around the current loop of the current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
4. Verify that the display reads within the specified Display minimum and
maximum limits for each of the frequencies listed in the Frequency test
record.
5. Turn the calibrator output off.
6. Remove the clamp from the current loop.
10
Handheld Instruments Basic Service
DCM910 Performance Verification
DCM910 Test Record
Serial numberProcedure performed byDate
DCM910 test record
Test inputToleranceDisplay minimumReadingDisplay maximum
At these frequencies, the inductance of the DCM910 may shut down the output of some calibrators. If this happens,
decrease the calibrator output frequency until the output remains on for the duration of the test.
Handheld Instruments Basic Service
11
DCM910 Performance Verification
DCM910 test record (cont.)
Test inputDisplay maximumReadingDisplay minimumTolerance
This section contains procedures to adjust the DCM910 Digital Clamp Meter. If
your instrument fails a performance requirement, use these procedures to return
it to factory specifications.
In this section you will find the following information:
HA list of adjustments
HA list of test equipment needed to make the adjustments
HInstructions on how to prepare the instrument for adjustment
HStep-by-step adjustment procedures
The procedures in this section do not verify performance. To confirm that your
multimeter meets factory specifications, perform the procedures in the DCM910Performance Verification section.
List of Adjustments
Test Equipment
Use the adjustments listed in Table 7 to return the DCM910 clamp meter to
factory calibration.
T able 7: DCM910 adjustments
Position Error
AC Current
DC Current
Peak Hold
The test equipment listed in Table 6 on page 8 is a complete list of equipment
needed for the adjustment procedures. These procedures assume that all test
equipment is operating within tolerance. Detailed operating instructions for test
equipment are not given in this procedure. If you need operating information,
refer to the instruction manual of the test equipment.
Alternative test equipment must meet or exceed the intended minimum
requirements specified in Table 6. If you substitute equipment, you may need to
modify the procedures.
Handheld Instruments Basic Service
13
DCM910 Adjustment Procedures
Preparation for Adjustment
The following guidelines apply to all DCM910 adjustments.
HPerform all adjustments in a 21_ to 25_ C ambient environment with a
HBefore making any adjustment, warm up the current meter for at least
HDo not alter any setting without reading the entire adjustment procedure first.
HDo not alter a setting unless a performance characteristic cannot be met at
HRead the Safety Summary at the beginning of this manual.
relative humidity of 75% or less.
30 minutes.
the current setting.
Open the Meter Case
Adjustment Procedure
You must open the meter case to gain access to the internal adjustments.
1. Lay the meter face down on a flat work surface.
2. Remove the two screws from the case bottom with a Phillips-head
screwdriver.
3. Gently lift the end of the case bottom until it unsnaps from the case top.
4. Remove the three screws that secure the circuit board assembly to the case
top. Do not remove the screws that secure the circuit boards to each other.
5. To access the adjustments, lift the circuit board assembly far enough out of
the top case to expose the adjustments. See Figure 2 and the procedure that
follows.
To reassemble the meter following the adjustments, perform steps 2 through 4
above in reverse order.
To return your instrument to factory calibration, implement the following
procedures.
14
Use a small flat-tipped screwdriver to make the adjustments. Refer to Figure 2
for adjustment locations.
WARNING. Magnetic fields are produced that may cause a malfunction in heart
pacemakers, or damage to sensitive equipment.
Handheld Instruments Basic Service
VR1
VR2
VR3
VR5
VR6
VR7
VR8
DCM910 Adjustment Procedures
Short for
Auto Zero
(VR2)
Short for
Peak Hold
(VR9)
VR9
VR10
Position Error
DC Auto Zero
Figure 2: Adjustment locations
The adjust the position error calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Select the appropriate coil to multiply the AC current calibrator output to
380 A at 50 Hz.
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Adjust VR1 to maintain the measurement error to less than 1% total while
positioning the coil in the clamp.
5. Turn the calibrator output off.
6. Remove the clamp from the coil.
To adjust the DC zero calibration, perform the following steps.
1. Set the clamp meter to the DC position.
2. Short the Auto Zero points indicated in Figure 2.
Handheld Instruments Basic Service
15
DCM910 Adjustment Procedures
3. Adjust VR2 until the display reads 00.0 ± 5 counts.
4. Remove the short.
5. Press the clamp meter DCA AUTO ZERO button to zero the display.
6. Adjust VR3 until the display reads 00.0.
DC 400 A Range
DC 1000 A Range
To adjust the DC 400 A range calibration, perform the following steps.
1. Set the clamp meter to the DC position.
2. Select the appropriate coil to multiply the DC current calibrator output to
200 A.
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Position the clamp to the center of the coil.
5. Adjust VR5 until the display reads 201.5.
6. Turn the calibrator output off.
7. Remove the clamp from the coil.
To adjust the DC 1000 A range calibration, perform the following steps.
1. Set the clamp meter to the DC position.
2. Select the appropriate coil to multiply the DC current calibrator output to
400 A.
16
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Press the clamp meter DCA AUTO ZERO button to zero the display.
5. Position the clamp to the center of the coil.
6. Adjust VR6 until the display reads 400.
7. Turn the calibrator output off.
8. Remove the clamp from the coil.
Handheld Instruments Basic Service
DCM910 Adjustment Procedures
AC 400 A Range
AC 1000 A Range
To adjust the AC 400 A range calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Select the appropriate coil to multiply the AC current calibrator output to
390 A at 400 Hz.
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Position the clamp to the center of the coil.
5. Adjust VR8 until the display reads 396.0. To keep the meter on the lower
range, it may be necessary to cycle the calibrator output off and on.
6. Turn the calibrator output off.
7. Remove the clamp from the coil.
To adjust the AC 1000 A range calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Select the appropriate coil to multiply the AC current calibrator output to
400 A at 400 Hz.
Peak Hold
3. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
4. Position the clamp to the center of the coil.
5. Adjust VR7 until the display reads 400.
6. Turn the calibrator output off.
7. Remove the clamp from the coil.
To adjust the peak hold calibration, perform the following steps.
1. Set the clamp meter to the AC position.
2. Short the Peak Hold points indicated in Figure 2.
3. Press PEAK HOLD to activate the function.
4. Adjust VR9 until the display reads 00.0.
5. Remove the short.
6. Press PEAK HOLD to cancel the function.
7. Press PEAK HOLD again to verify that the display reads 00.0.
Handheld Instruments Basic Service
17
DCM910 Adjustment Procedures
T able 8: Summary of adjustments
8. Press PEAK HOLD to cancel the function.
9. Repeat steps 2 through 8 above until the display reads 00.0.
10. Select the appropriate coil to multiply the AC current calibrator output to
200 A at 400 Hz.
11. Position the clamp around the coil of the current calibrator and release the
clamp trigger. Ensure that the clamp is entirely closed.
12. Position the clamp to the center of the coil.
13. Adjust VR10 until the display reads 200.0.
14. Turn the calibrator output off.
15. Remove the clamp from the coil.
16. Reassemble the meter.
Adjustment
name
Position ErrorAC380 A50 HzVR1<5 counts0 count5 counts
DC ZeroDCVR2
The DMM150 is a rugged, handheld digital multimeter that allows you to make
accurate measurements quickly and easily. Whether you are a professional or
hobbyist, this instrument provides a useful range of features:
HSmall, pen-style case
1
@
H3
HAuto range (volts, ohms)
HMeasurement hold
HMeasures DC and AC voltages, resistance, diode voltage, and continuity
HDiode and continuity tester with audible signal
HOvervoltage protected
HRecessed input jacks for safety
HAutomatic power off after 10 minutes to prolong battery life
digit LCD display with bar graph
2
HLow-battery indicator
HUses standard AAA batteries
Figure 1: DMM150 Digital Multimeter
Handheld Instruments Basic Service
1
DMM150 Digital Multimeter
2
Handheld Instruments Basic Service
DMM150 Specifications
All specifications are warranted unless noted typical.
Stated accuracies are ±(% reading + number of counts) at 23_ C 5_ C, at less
than 75% R.H. (relative humidity).
T able 1: General specifications
CharacteristicsDescription
Display3!@2 digit liquid crystal display (LCD) with a maximum reading of 3200
65 segment analog bar graph
Polarity indicationAutomatic; positive implied, negative indicated
Overrange indication“OL” or “–OL”
Low battery indicationThe battery symbol is displayed when the battery voltage drops below the operating
voltage level
Sampling rate2 times/second for digital display
12 times/second for analog bar graph
Operating altitude2000 m (6561 ft.), maximum
Operating temperature0_ C to +50_ C, 0–80% R.H.
Storage temperature–20_ C to +60_ C, 0–80% R.H. with batteries removed from the meter
T emperature coef ficient0.15 (specified accuracy)/_C at <18°C or >28°C
Power supplyTwo standard AAA, IEC LR03, or ANSI/NEDA 24A 1.5 V batteries
Battery life (typical)Alkaline 800 hours
Dimensions (HxWxD)42 mm × 145 mm × 24 mm
(1.7 in. × 5.7 in. × 0.9 in.)
Maximum floating voltage600 VDC or 600 VAC
Maximum input voltage600 VDC or 600 VAC
30 MW10 kW
Continuity check thresholdThe beeper sounds if the resistance of the circuit measured is < 20 W
Diode test
Test current1.5 mA maximum
Test voltage3.3 V maximum open circuit
Resolution1mV
Accuracy±(1.5% + 5 counts) from 0.4 V to 0.8 V
Auto power offThe meter automatically shuts off approximately ten minutes from the last function or
mode change. The meter turns on again when another range is selected or any button
is pressed.
T able 3: Certifications and compliances
EC Declaration of ConformityMeets intent of Directive 89/336/EEC for Electromagnetic Compatibility and Low
Voltage Directive 73/23/EEC for Product Safety. Compliance was demonstrated to the
following specifications as listed in the Official Journal of the European Communities:
EMC Directive 89/336/EEC:
EN 55011 Class B Radiated and Conducted Emissions
EN 50082-1 Immunity:
IEC 801-2Electrostatic Discharge Immunity
IEC 801-3RF Electromagnetic Field Immunity
Low Voltage Directive 73/23/EEC as amended by 93/68/EEC:
EN 61010-1/A2 Safety requirements for electrical equipment for
measurement, control, and laboratory use
CertificationsListed UL31 11-1 and CAN/CSA C22.2 No. 1010.1.
Overvoltage CategoryCategory:Examples of Products in this Category:
CA T IIIDistribution-level mains, fixed installation
CA T IILocal-level mains, appliances, portable equipment
Signal levels in special equipment or parts of equipment,
telecommunications, electronics
Pollution Degree 2
CA T I
Do not operate in environments where conductive pollutants may be present.
Handheld Instruments Basic Service
5
DMM150 Specifications
6
Handheld Instruments Basic Service
DMM150 Performance Verification
This section contains procedures to verify that the DMM150 Digital Multimeter
performs as warranted. If an instrument fails any of these checks, it needs
adjustment and or repair.
The performance verification procedures provide a valid confirmation of
instrument electrical characteristics and function under the following conditions:
HThe multimeter operates in an 18_ to 28_
environment with a relative humidity of less than 75%.
HThe multimeter stabilizes in the stated ambient temperature for one hour.
HThe multimeter warms up for five minutes.
HFor AC measurements, allow the multimeter to settle to its final value before
taking the measurement.
HThe multimeter remains fully assembled.
The DMM150 performance verification consists of the checks listed in Table 4.
T able 4: Performance verification checks
AC Volts Check
DC Volts Check
W Check
Continuity Check
The performance verification procedure should be performed annually or after
every 2000 hours of operation if used infrequently.
C (64_ to 82_F) ambient
Handheld Instruments Basic Service
7
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.