H600, SA2600 Declassification and Security Instructionsi
Table of Contents
iiH600, SA2600 Declassification and Security Instructions
Preface
This document helps customers with data security concerns to sanitize or remove
memory devices from the H600 RFHawk or the SA2600 Spectrum Analyzer
instrument.
This product has data storage (memory) devices and data output devices (USB
ports). The
Clear or sanitize the memory devices
Clear or sanitize an instrument that is not functioning
se instructions tell how to do the following:
Products
Terms
The following Tektronix product is covered by this document:
H600 RFHawk
SA2600 S
The following terms may be used in this document:
Clear. This removes data on media/memory before reusing it in a secured area.
All reusable memory is cleared to deny access to previously stored information
by standard means of access.
Erase. This is equivalent to clear.
Media storage/data export. Various devices that are used to store or export data
from the instrument, such as a USB port.
Nonvolatile memory. Data is retained when the instrument is powered off.
device from the instrument. Instructions are available in the product service
manual.
Sanitize. This eradicates the data from media/memory so that the data cannot be
recovered by other means or technology. This is typically used when the device
will be moved (temporarily or permanently) from a secured area to a non-secured
area.
pectrum Analyzer
Scrub. This is equivalent to sanitize.
User-modifiable. The m emory device can be written to by the user during normal
instrument operation, using the instrument’s user interface or remote control.
Volatile memory. Data is lost when the instrument is powered off.
H600, SA2600 Declassification and Security Instructionsiii
Preface
ivH600, SA2600 Declassification and Security Instructions
Clear and Sanitize Procedures
Memory Device
Table 1: Vol
atile Memory Devices
s
Type and
minimum sizeFunction
SDRAM (2)
16M x 16
SRAM (3)
256K x 16
SRAM,
2M x 18
Tabl e 2
Type an
: Nonvolatile Memory D evices
d
System memory
DSP program and data
memory
Acquisition me moryNo
minimum sizeFunction
M_System
DiskOnChip
32M x 8 or
Sandisk 512M x 8
Flash (2)
16M x 8
EEPROM
128 x 8
EEPROM
128 x 8
Flash
8M x 16
File systemYesUser writeable,
Operating
system and
register
settings
EthernetNoFactory
Storing internal
calibration data
FPGA
configuration,
DSP
acquisition and
measurement
control
The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options.
User
modifiable
YesWritten
No
User
modifiableInput methodLocationProcess to clear
YesUser input
NoWritten during
NoFactory
Input
methodLocationProcess to clear
by
system
processor
Read/write
by DSP
processor
Read/write
by DSP
processor
default location
for application
software,
application
settings and
measurement
result files (My
Device\BuiltInDisk
folder)
programmed
factory or service
calibration
programmed,
software updates,
and used by
application
Platform processor boardRemove power from the
Module digital signal processor boardPower down the module
Module digital signal processor boardPower down the module
Mainframe processor
board
Mainframe processor
board
Mainframe processor
board
Module RF board
Module digital signal
processor board
System reset (See page 2, Perform aSystem Reset.)
System reset (See page 2, Perform aSystem Reset.)
Store the instrument in a secure area
Store the instrument in a secure area
Store the instrument in a secure area, or
destroy the instrument
instrument for at least
20 seconds
by exiting the H600 or
SA2600 application
by exiting the H600 or
SA2600 application
H600, SA2600 Declassification and Security Instructions1
Clear and Sanitize Procedures
Perform a System Reset
Data Export Devices
Table 3: Data Export Devices
TypeFunction
USB
host port
(supports
removable
USB flash
drive)
USB slave
port
PCMCIASupport of PCMCIA-compliant
LAN
Ethernet
connector
User storage of measurement
data, screen images, and
instrument setups; remote control
anddatatransfertoaPC
Remote control and data transfer
to a PC
devices including ATA flash
memory cards
Remote control and data transfer
to a PC
This procedure
does not erase or change factory calibration constants. Resetting
the system puts the instrument back to factory specifications.
1. Power on the in
strument.
2. Tap the Start button.
3. Select Programs > Tektronix Utilities > System Reset.
NOTE. After a system reset, the user settings and data are cleared, but the
instrumen
t operating software is preserved.
The following table lists the data export devices in the standard instrument and
listed options.
User
modifiable
Yes
Yes
YesUser writable
YesRemote
Input
methodLocationProcess to disable
Software
operations,
remote
control and
data transfer
Software
operations,
remote
control and
data transfer
control and
data transfer
USB host port on top
of instrument
USB device port on
top of instrument
Right side of
instrument
Ethernet port on top
of instrument
The USB flash drive can be removed and
destroyed.
The USB host port cannot be disabled.
The USB device port cannot be disabled.
Remove all devices attached to the port.
Remove all memory devices and format
them, store them in a secure area, or
destroy them. The PCMCIA device port
cannot be disabled.
The Ethernet port cannot be disabled.
2H600, SA2600 Declassification and Security Instructions
Troubleshooting
How to Clear or Sanitize a Non-Functional Instrument
If your instrument is not functioning and you need to clear or sanitize it, proceed
as follows:
H600, SA2600
USB Flash Drive
Charges
There are no customer-removable internal memory devices or boards in the H600
or SA2600. Refer to your company’s internal policies regarding handling or
disposal of the module.
Remove the USB flash drive, and then refer to your company’s internal policies
regarding handling or disposal of the flash drive.