DTG5078 & DTG5274
Data Timing Generators
Performance Verification & Specifications
071-1280-02
This document supports firmware version 1.0.14
and above.
www.tektronix.com
Copyright E Tektronix Japan, Ltd. All rights reserved.
Copyright E Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Tektronix Japan, Ltd. 5--9--31 Kitashinagawa, Shinagawa--ku, Tokyo 141 --0001 Japan
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077--0001
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship
for a period of one (1) year from the date of shipment. If a product proves defective during this warranty period, Tektronix,
at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in
exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the
warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid.
Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the
Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any
other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair
damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction
caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other
products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO
REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO
THE CUSTOMER FOR BREACH O F THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE
LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE
OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NO TICE OF THE POSSIBILITY OF SUCH
DAMAGES.
Table of Contents
List of Figuresii............................................
List of Tablesiii.............................................
General Safety Summaryv...................................
Table 2--36: Power Supply2--34.................................
Table 2--37: Certifications and compliances2--35...................
Table 2--38: Installation category and Pollution degree2--36..........
iv
DTG5078 & DTG5274 Data Timing Generator Technical Reference
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
The common terminal is at ground potential. Do not connect the common
terminal to elevated voltages.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
v
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on the Product. These terms may appear on the product :
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
WARNING
High Voltage
Protective Ground
(Earth) Terminal
CAUTION
Refer to Manual
Double
Insulated
vi
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Preface
This manual provi des information necessary for users or service technicians to
verify the performance of the DTG5000 Series Data Timing Generator.
Manual Structure
The DTG5000 Series Data Timing Generator Technical Reference contains the
following sections:
The Performance Verifications section contains an introduction, a list of
equipment required, and procedures that, when passed, ensure that the product
meets its specifications.
The Specifications section contains a brief product description and characteristics
tables. These tables cover the electrical, mechanical, environmental characteristics and certification.
Related Manuals and Online Documents
This manual is part of a document set of standard-accessory manuals and online
documentation; this manual mainly focuses on the performance verification and
specifications information needed to verify the product performance. See the
following list for other documents supporting the data timing generator operation
and service. (Manual part numbers are listed in Accessories & Options section of
User Manual.)
Document nameDescription
DTG5000 Series Online HelpAn online help system, integrated with the User Interface application that ships with this
product. The help is preinstalled in the instrument.
DTG5000 Series User Manual, volume 1A quick reference to major features of the instrument and how they operate. It also
provides several tutorials to familiarize the user with basic instrument features.
DTG5000 Series User Manual, volume 2A comprehensive usage information on how to operate the instrument including the
descriptions of functions and menu operations.
DTG5000 Series Programmer ManualProvides complete information on programming commands and remote control of the
instrument.
DTG5000 Series Service ManualDescribes how to service the instrument to the module level. This optional manual must
be ordered separately.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
vii
Preface
Contacting Tektronix
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales support1-800-833-9200, select option 1*
Service support1-800-833-9200, select option 2*
Technical supportEmail: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. -- 5:00 p.m. Pacific time
*This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
viii
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Performance Verification
Two types of Performance Verification procedures can be performed on this
product: Self Tests and Performance Tests. You may not need to perform all of
these procedures, depending on what you want to accomplish.
HVerify that the DTG5000 Series Data Timing Generator is operating correctly
by running the Self Tests, which begin on page 1--3.
Advantages: These procedures require minimal time to perform, and test the
internal hardware of the DTG5000 Series Data Timing Generator.
HIf a more extensive confirmation of performance is desired, complete the self
tests, and t hen do the Performance Tests beginning on page 1--10.
Advantages: These procedures add direct checking of warranted specifications. These procedures require suitable test equipment and more time to
execute. (Refer to Equipment Required on page 1--11).
Conventions in this manual
Throughout these procedures the following conventions apply:
HEach test procedure uses the following general format:
Title of Test
Equipment Required
Prerequisites
Procedure
HEach procedure consists of as many steps, substeps, and subparts as required
to do the test. Steps, substeps, and subparts are sequenced as follows:
1. First Step
a. First Substep
HFirst Subpart
HSecond Subpart
b. Second Substep
2. Second Step
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--1
Performance Verification
HWhere instructed to use a control in the display or a front-panel button or
knob, the name of the control, button or knob appears in boldface type.
Menu Selections
User Manual
Install the Output Modules
Instructions for menu selection use the following format:
Menu buttonLeft or right Allow button (by using this button, you can move
to desired menu category)Upper or down Allow button (by using this button,
you can move to desired menu item)SELECT or Enter key (this completes
the selection).
You can use a mouse as a pointer, use keyboard shortcuts for quick operation, or
use front panel knob instead of sticking to above menu selection format.
The DTG5078 & DTG5274 Data Timing Generators User Manual is strongly
recommended to familiarize the first-time user with instrument controls and
features.
Any output modules ordered are shipped separately. For complete instructions on
how to install the output modules, refer to the User Manual. (Output modules do
not ship preinstalled.)
CAUTION. Do not install or remove any output modules while the instrument is
powered on.
Always power the instrument down before attempting to remove or install any
output module.
1--2
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Self Tests
Performance Verification
There are two types of tests in this section that provide a quick way to confirm
basic functionality and proper adjustment:
HDiagnostics
HCalibration (You must perform this calibration before the performance tests.)
These procedures use internal diagnostics to verify that the instrument passes the
internal circuit tests, and calibration r outines to check and adjust the instrument
internal calibration constants.
NOTE. To perform the Self Tests, at least one output module (DTGM10,
DTGM20, or DTGM30) must be installed in the DTG5000 series Data Timing
Generator mainframe. You can select any slot when you perform the tests even
though the descriptions below are assuming the Slot A is used.
Diagnostics
This procedure uses internal routines to verify that the instrument is operating
correctly. No test equipment or hookups are required.
The instrument automatically performs the internal diagnostics when powered
on; you can also run the internal diagnostics using the menu selections described
in this procedure. The difference between these two methods of initiating the
diagnostics is that the menu method does a more detailed memory check than the
power-on method.
Do the following steps to run the internal routines that confirm basic functionality and proper adjustment.
Equipment
required
Prerequisites
None
First, at least one output module must be installed properly in the
mainframe.
Second, power on the instrument and allow a twenty-minute warmup
before doing this procedure.
1. Set up the instrument:
HConfirm that there is no output being performed by verifying that the
RUN button indicator is not on. If the indicator is on, push the RUN
buttontoturnitoff.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--3
Performance Verification
HVerify that the output module LEDs are not on. If any output module
LEDs are on, push the ALL OUTPUTS ON/OFF buttontoturnthe
LEDs off.
2. Internal diagnostics: Perform these substeps to verify internal diagnostics.
a. Display the Diagnostics dialog:
HFrom the application menu bar, select System , and then select
Diagnostics.... The following dialog appears if you have changed
the settings.
HSelect OK to display the Diagnostics dialog. See Figure 1--1.
Figure 1--1: Diagnostics dialog
1--4
HVerify that the Loop until Fail box is not checked. If it is checked,
click the box to remove the check mark.
HSelect Execute All Items to start the diagnostics.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Performance Verification
b. Wait: The internal diagnostics takes one to six minutes. When complete,
the resulting status appears in the diagnostics control window.
c. Verify that no failures are found and reported: All tests should pass.
Confirm that the word Pass appears in all the Status fields. If any
failures occur, record the failure information and contact your local
Tektronix service personnel for more information.
3. Select Close to exit the diagnost ics.
Calibration
Two types of calibrations are provided in the DTG5000 Series Data Timing
Generator.
HThe Level Calibration uses internal calibration routines that check electrical
characteristics s uch as DC accuracy of data output, and then adjust the
internal calibration constants as necessary.
HThe Skew Calibration checks the delay time of data output, and then adjust
the i nternal calibration constants as necessary. The calibration is performed
by connecting each channel output to Skew Cal In.
NOTE. Level Calibration and Skew Calibration are not valid until the instrument
reaches a valid temperature.
Level Calibration.
Equipment
required
Prerequisites
None
Power on the instrument and allow a 20 minute warmup at an ambient
temperature between +20_ C and +30_ C before doing this procedure.
The calibration routine must be performed whenever the ambient
temperature changes by 5_ C or more.
1. Set up the instrument:
HConfirm that there is no output being performed by verifying that the
RUN button indicator is not on. If the indicator is on, push the RUN
buttontoturnitoff.
HVerify that the output module LEDs are not on. If any output module
LEDs are on, push the ALL OUTPUTS ON/OFF button to turn the
LEDs off.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--5
Performance Verification
2. Perform the calibration suite:
a. From the application menu bar, select System , and then select Level
Calibration.... The Level Calibration dialog appears. See Figure 1--2.
Figure 1--2: Level Calibration dialog
b. Select Execute.
c. All the Status fields must be Pass. If any failures occur, record the
failure information and contact your local Tektronix service personnel
for more information.
3. Select Close to exit the calibration.
1--6
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Skew Calibration.
Performance Verification
Equipment
required
Prerequisites
One 50 Ω SMA coaxial cable, Tektronix part number 174-1427-00
Refer to test equipment list on page 1--12.
Power on the instrument and allow a 20 minute warmup at an ambient
temperature between +20_ C and +30_ C before doing this procedure.
The calibration routine must be performed whenever the ambient
temperature changes by 5_ C or more.
1. Set up the instrument:
HConfirm that there is no output being performed by verifying that the
RUN button indicator is not on. If the indicator is on, push the RUN
buttontoturnitoff.
HVerify that the output module LEDs are not on. If any output module
LEDs are on, push the ALL OUTPUTS ON/OFF button to turn the
LEDs off.
2. Perform the calibration suite:
a. From the application menu bar, select System, and then select Skew
Calibration. The following dialog appears if you have changed thesettings. Click OK.
b. The Skew Calibration dialog appears. See Figure 1--3.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--7
Performance Verification
Figure 1--3: Skew Calibration dialog
c. Attach an SMA coaxial cable to the SKEW CAL IN at the front panel
of the data timing generator mainframe.
d. Select Execute to display the dialog box shown below.
e. Connect the opposite end of the SMA coaxial cable to the CH1
connector of output module and select OK to start t he calibration. Wait
until the calibration completes.
f.Follow the on-screen instruction to continue the calibration:
HDisconnect the cable from the channel and reconnect it to next
channel.
1--8
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Performance Verification
HRepeat the same calibration procedure for all channels.
NOTE. When you connect the output module and Skew Cal In, use the identical
cable. If you use different cables, the calibration result may be affected.
g. When complete, the resulting status appears on the screen. See Figure
1--4.
Figure 1--4: Skew Calibration results screen
h. Verify that no failures are found and reported on the screen.
i.If any f ailures occur, record the failure information and contact your
local Tektronix service personnel for more information.
3. Select Close to exit the calibration.
NOTE. The calibration data in the memory may be lost if the instrument is
powered off while the calibration is executed.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--9
Performance Verification
Performance Tests
The Performance Tests include functional test items, such as the interface
functional test, in this manual.
HThe Functional Tests verify the functions, that is, they verify that the
DTG5000 Series Data Timing Generator features operate. They do not verify
that they operate within limits.
HThe Performance Tests verify that the DTG5000 Series Data Timing
Generator performs as warranted. The Performance Tests check all the
characteristics that are designated as checked in Specifications.(The
characteristics that are checked appear with a n in Specifications.)
Table 1--1: Performance test items
TitlesTest itemsReference page
DTG5000 series mainframe
1
Sync outputOutput levelPage 1--15
Internal clock frequencyInternal clock output frequency accuracyPage 1--18
External clock outputExternal clock output amplitude, rise time/ fall time, and aberrationPage 1 --20
External clock inputExternal clock input function and external clock input frequency accuracy Page 1--22
5.Function GeneratorOutput voltage: --5 V to +5 V,
Frequency accuracy: <0.01%
Agilent 53181A op.050/010Checks clock frequency.
Fluke 8842AMeasures voltage. Used in
Tektronix TDS7104Checks output signals. Used
Tektronix CSA8000B,
80E03
Tektronix AFG320Generates external input
DTG5078 & DTG5274 Data Timing Generator Technical Reference
multiple procedures.
in multiple procedures.
2
Checks output signals. Used
in multiple procedures.
signals. Used in multiple input
signal test procedures.
1--11
Performance Verification
Table 1--2: Test equipment (cont.)
Item number and
description
6.SMA Coaxial Cable
(3 required)
7.BNC Coaxial Cable
(3 required)
8.Adapter
(2 required)
9.Adapter
(2 required)
10. AdapterN (male) to SMA (male), 50 ΩTektronix part number
11. AdapterSMA (female) to SMA (female), 50 Ω Tektronix part number
12. Lead set for DC output16-CON twisted pair, 60 cm (24in)Tektronix part number
13. Dual-Banana PlugBNC (female) to dual bananaTektronix part number
14. BNC-T ConnectorBNC (male) to BNC (female) to BNC
15. Feed-through Termination
16. SMA Termination50 Ω,SMATektronix part number
17. Attenuator
(2 required)
2
For best repeatability and to prolong the life of both connectors, use a torque wrench (5/16 in) and tighten the connection
to the range of 7-10 lb-in (79-112 N-cm) when you connect an SMA cable to a sampling module. For more information,
refer to your sampling module user manual.
Minimum requirements
50 Ω, male to male SMA connectorTektronix part number
50 Ω, male to male BNC connectorTektronix part number
SMA (male) to BNC (female), 50 ΩTektronix part number
SMA (female) to BNC (male), 50 ΩTektronix part number
(female)
50 Ω,0.1%,BNCTektronix part number
12 dB, SMA-- -- --Signal attenuation
Recommended equipment
or equivalent
174-1427-00
012-0076-00
015-0554-00
015-0572-00
015-0369-00
015-1012-00
012-A229-00
103-0090-00
Tektronix part number
103-0030-00
011-0129-00
015-0706-00
Purpose
Signal interconnection
Signal interconnection
Signal interconnection
Signal interconnection
Signal interconnection
Signal interconnection
Signal interconnection
Signal interconnection
Signal interconnection
Signal termination
Signal termination
1--12
CAUTION. Sampling modules are inherently vulnerable to static damage. Always
observe static-safe procedures and cautions as outlined in the sampling module
user manual.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Performance Verification
Loading Files
The following steps explain how to load files from the DTG5000 Series Data
Timing Generator.
1. From the application menu bar, select File, and then select Open Setup.The
Open Setup dialog appears. See Figure 1--5.
2. Specify C:\Program Files\Tektronix\DTG5000\PV\DTG5078 (or
C:\Program Files\Tektronix\DTG5000\PV\DTG5274)toLook in field.
Figure 1--5: Open Setup dialog
3. Select the necessary file in the File name:, and then click Open.
4. The Open Setup Dialog automatically disappears, and then the selected
waveform and sequence file are loaded.
If your data timing generator mainframe is not equipped with maximum
output module configuration, the following dialog box appears.
5. Click OK to complete the i nstrument setup.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--13
Performance Verification
Performance Check Files
Table 1--3 lists the setup files on the internal hard disk drive that are used in
these performance tests. A specified file must be loaded each time you execute
Performance Test procedure. Test pattern data and setup information are included
in the file.
Table 1--3: Performance check files
No.File nameClock frequencyTest item
1SYNCOUT.dtgInternal: 10 MHzSync output
2INTCLK.dtgInternal: 100 MHzInternal clock frequency
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Mainframe
Performance Verification
The following procedures check those characteristics that relate to the mainframe
that are checked under Mainframe in Specifications. Refer to page 2--3.
NOTE. To perform the Performance Tests, at least one output module must be
installed in the DTG5000 Series Data Timing Generator mainframe. You can
select any slot when you perform the tests even though the descriptions below
are assuming the Slot A is used.
Sync Output
This test verifies that the DTG5000 series mainframe sync output is functional.
Equipment
required
Prerequisites
One oscilloscope (TDS7104) (item 3)
Two 50 Ω SMA coaxial cables (item 6)
Two SMA (female)-BNC (male) adapters (item 9)
DTG5000 Series Data Timing Generator
The
prerequisites listed on page 1 --11.
must meet the
1. Install the test hookup and preset the instrument controls:
a. Hook up the oscilloscope:
HAttach SMA (female)-BNC (male) adapters to the oscilloscope CH1
input and CH2 input connectors.
HConnect an SMA coaxial cable from t he CH2 connector of output
module, which is inserted in the slot A of DTG5000 series mainframe, to the SMA-BNC adapter (CH2 input of oscilloscope).
HConnect an SMA coaxial cable from t he SYNC OUT at the front
panel of the DTG5000 series mainframe to the SMA-BNC adapter
(CH1 input of the oscilloscope). See Figure 1--6.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--15
Performance Verification
DTG5000 series
CH2 OUTPUT
SYNC OUT
50 Ω SMA coaxial cables
Figure 1--6: Sync output tests
b. Set the oscilloscope controls as follows:
Vertical .
CH1andCH2coupling ......... DC
CH1scale .................. 100mV/div
CH2scale .................. 200mV/div
CH1 and CH2 input impedance . . .50 Ω
CH1offset .................. --200mV
Horizontal
Scale ...................... 100ns/div
Acquisition
Mode ...................... Average
Number of running averages ..... 32
Trigger
Source ..................... CH2
Coupling.................... DC
Slope ...................... Positive
Level ...................... 500mV
Mode ...................... Auto
Measurement .................... CH1High
Oscilloscope (TDS7104)
CH1 and CH2 Inputs
+ SMA (female)-BNC
(male) adapter
CH1 Low
1--16
2. Set the data timing generator controls and load the setup file:
a. Load the setup file (SYNCOUT.dtg). Refer to Loading Files on page
1--13.
b. After the file is loaded, the Frequency of data timing generator is set to
10MHz.
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Performance Verification
3. Push the RUN button of the data timing generator to light the RUN LED,
and then push the ALL OUTPUTS ON/OFF button to activate the output.
4. Confirm the oscilloscope screen: Verify that the 400 ns width square
waveform appears in the CH1 display.
5. Using the oscilloscope Measurement functions, verify that the High Level
and Low Level values of Sync Out are within the following range.
HHigh Level: 0 V ¦ 50 mV
HLow Level: -- 0.4 V ¦ 50 mV
DTG5078 & DTG5274 Data Timing Generator Technical Reference
1--17
Performance Verification
Internal Clock Frequency
This test verifies the frequency accuracy of internal clock.
Equipment
required
Prerequisites
One frequency counter (item 1)
One 50 Ω SMA coaxial cable (item 6)
One SMA (female) - BNC (male) adapter (item 9)
One N (male) - SMA (male) adapter (item 10)
One SMA (female) - SMA (female) adapter (item 11)
DTG5000 Series Data Timing Generator
The
prerequisites listed on page 1 --11.
1. Install the test hookup and preset the instrument controls:
a. Hook up the frequency counter:
HAttach an SMA (female)-BNC (male) adapter to the CHANNEL 1
input of frequency counter.
HAttach a N (male)-SMA (male) adapter to the CHANNEL 2 input of
frequency counter, and then attach an SMA (female)-SMA (female)
adapter to the N-SMA adapter.
must meet the
HConnect an SMA coaxial cable from the CLOCK OUT at the rear
panel of DTG5000 series mainframe through the SMA-SMA and
N-SMA adapters to the frequency counter CHANNEL 2 input. See
Figure 1--7.
CHANNEL 2
+ N (male)-SMA (male) adapter
+ SMA (female)-SMA (female) adapter
Frequency Counter
CHANNEL 1
+ SMA (female)-BNC (male) adapter
CLOCK OUT
DTG5000 series rear
50 Ω SMA coaxial cable
Figure 1--7: Internal Clock Frequency tests
b. Power on the frequency counter, and verify that the frequency counter is
set to frequency measurement mode (default setting).
1--18
DTG5078 & DTG5274 Data Timing Generator Technical Reference
Loading...
+ 88 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.