Tektronix DSA70404B Data Sheet

Digital Phosphor Oscilloscopes/Digital Serial Analyzers
DPO/DSA70000B Series Data Sheet
DPX and Fast Acquisition - Quickly nd intermittent e vents with the industry’s highest waveform capture rate
Long Record Lengths – Capture long-term trends or infrequent events then analyze the entire acquired record with built-in measurements
P7500 TriMode™ Probing System - Perfectly matched signal connectivity from 4 GHz to 20 GHz
Application Support – Enables standard-speciccertification, measurement automation, and ease of use
Serial Data Link Analysis (SDL A) - For high-speed serial data streams, emulate and equalize the channel as well as de-embedding the xture.
Applications
Signal Integrity, Jitter, and Timing Analysis
Verication, Debug, Characterization, or Ce rti cation of Sophisticated Designs
Feature
On All Four Channels Simultaneously
Highes high-speed serial standards
Superior Signal Integrity and Excellent Signal-to-Noise Ratio – Observe the truest representation of your waveform
Pinpoint®Triggering – M inimize time spen t trying to acquire problem signals for efcient troubleshooting and shortened debug time
5 Gbps R of a specied NRZ or 8b/10b pattern to allow isolation of pattern dependent effect s
Search & Mark – Provides waveform pattern matching and software trigg
s & Benets
20, 16, 12.5, 8, 6 , and 4 GHz Bandwidth Models Up to 50 GS/s Real-Time Sample Rate Up to 250 Feature for Quick Navigation Fastest Waveform Capture Rate with >300,000 wfms/s Maximum per Channel
ers for signals of interest
t Bandwidth – Up to 20 GHz enables measurement on the latest
eal-time Serial Trigger – Assures triggering on the rst instance
Debugging and Compliance Testing of Se rial Data Streams for Industry Standards
Memory Bus Analysis
Switching Power Supply Verication
Investigation of Transient Phenomena
Spectral Analysis of Transient or Wide-bandwidth RF Signals
Unmatched Performance for Greater Insight Gets Your Work Done Faster
The DPO70000B and the DSA70000B Series real-time digital phosphor oscilloscopes are the industry’s best solution to the challenging signal integrity issues faced by designers verifying, characterizing, debugging, and testing sophisticated electronic designs. The specialized DSA70000B Series provides a complete and dedicated solution to address the challenges of high-speed serial designs.
The family features exceptional performance in signal acquisition and analysis, operational simplicity, and unmatched debugging tools to accelerate your day-to-day tasks. A large display and the intuitive user
ace provide easy access to the maximum amount of information.
interf
Data Sheet
Unmatched Acquisition Performance
Superior Signal Integrity and Excellent Signal-to-Noise Ratio of the DP Ensures Condence in Your Measurement Results
High bandwidth up to 20 GHz matched across 2, 3, or 4 channels and enabled b user-selectable lter for each channel provides magnitude and phase correction for more accurate representation of extremely fast signals. In addition, enhancement for applications needing the highest measurement throughput.
Bandwidth enhancement eliminates imperfections in frequency response all the way t
Simultaneo details (transients, imperfections, fast edges)
50 GS/s on all channels for the 12.5, 16, and 20 GHz models 25 GS/s on a
With high s DPO70000B and DSA70000B series enable you to perform precise characterization measurements. When debugging a DUT, a low noise
oor and mnding the smallest anomalies that might affect the DUT’s performance.
For RF signal verication, a lower noise oor translates into a higher dynamic r applications.
Lowest jitter noise oor and highest vertical accuracy provide additional margin in your measurement s.
Long record length provides high resolution and extended-duration
rm capture
wavefo
Standard 10 M samples per channel on the DPO70000B Series and 20 M on the DSA70000B Series Option 8GHzmodels Optional up to 250 M samples on all four channels for the 12.5, 16, and 20 MultiView Zoom helps you manage long records, compare and analyze multiple waveform segments
Widest Range of Probing Solutions – Whether you need to measure
serial data or switching currents from your new power supply
8 Gbps design, Tektronix offers a vast array of probing solutions, including active single ended, differential, high voltage, current, optical, and a wide range
obe and oscilloscope accessories.
of pr
O70000B and D SA70000B Series
y Tektronix proprietary bandwidth enhancement. The
only Tektronix allows the user to disable the bandwidth
o the probe tip
us high sample rate on all channels captures more signal
ll channels for the 4, 6, and 8 GHz models
ignal-to-noise ratio and low internal noise oor, the
aximum signal delity of the measurement instrument allows
ange, opening the DPO/DSA70000B for a wider range of
al up to 100 M samples on all four channels f or the 4, 6, and
GHz models
User-selectable bandwidth limiting choices
Zoom in on four areas of interest simultaneously to quickly compare them
2 www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
Low-cost solder tip features quick connection so moving probe to various solder points is fast and easy
Maximize t FastAcq ac
he probability of capturing elusive glitches and other infrequent events with
quisition mode.
knob you can clearly “see a world others don’t see”, displaying the complete picture of your circuit’s operation. Some oscilloscope vendors claim high waveform capture rates for short bursts of time, but only DPO70000B and DSA70000B oscilloscopes, enabled by DPX technology, can deliver these
eform capture rates on a sustained basis — saving minutes, hours,
fast wav or even days by quickly revealing the nature of faults so sophisticated trigger modes can be applied to isolate them.
P7500 TriMode probes simplify complex measurement setups
P7500 TriMode probing system – Perfectly matched to the DPO/DSA70000B, the P7500 TriMode probes allow you to switch among differential, single ended, and common-mode measurements without moving the probe from its connection points. The P7500 series offers probes for performance from 4 GHz to 20 GHz.
Accelerate the Debug of Comple x Electrical Designs
FastAcq Acquisition Mode Expedites Debugging by
ly Showing Imperfections
Clear
More than just color-grading, FastAcq’s proprietary DPX®acquisition
ology captures signals at more than 300,000 waveforms per second
techn on all four channels simultaneously, dramatically increasing the probability of discovering infrequent fault events. And w ith a simple turn of the intensity
The Abil
ity to Trigger an Oscilloscope on Events of Interest is Paramount in Complex Signal Debug and Verication
Whether you’re trying to nd a problem signal or need to isolate a section of a complex signal for further analysis, like a DDR read or write burst. Tektronix’ Pinpoint
uses Silicon Germanium (SiGe) technology to provide very high
system
®
triggering provides the solution. The Pinpoint trigger
trigger sensitivity with very low trigger jitter and ability to capture very narrow glitches. Pinpoint triggering allows selection of virtually all trigger types on both A and B trigger events. Other trigger systems offer multiple trigger types only on a single event (A event), with delayed trigger (B event) selection limited to edge type triggering and often do not provide a way
et the trigger se quence if the B event doesn’t occur. But Pinpoint
to res triggering provides the full suite of advanced trigger types on both A and B triggers, logic qualication to control when to look for these events, and reset triggering to begin the trigger sequence again after a specied time, state, or transition so that even events in the most complex signals can be captured. Other oscilloscopes typically offer less than 20 trigger
nations; Pinpoint triggering offers over 1400 combinations, all at full
combi performance.
nhanced Triggering, trigger jitter is reduced to <100 fs. With this
With E stability at the trigger point, the trigger point can be used as a measurement reference.
www.tektronix.com/oscilloscopes 3
Data Sheet
riggering can isolate glitches down to 100-ps wide
Pinpoint t
NRZ Pattern Lock Triggering of a 640-bit long SATA serial pattern at 6 Gbps
of a long serial test pattern with outstanding time base accura cy. Pattern lock triggering can be used to remove random jitter from long serial data patterns. Effects of specic bit transitions can be investigated, and averaging can be used with mask testing. This feature supports up to
6.25 Gbps NRZ serial data stream and is standard on the DSA70000B instruments, or included as part of Option PTH on the DPO70000B models.
Isolate only the valid glitches
col and Serial Pattern Triggering
Proto
To debug serial architectures, use the serial p attern triggering for NRZ serial
tream with built-in clock recovery and correlate events across physical
data s and link layers. This feature comes standard on the DSA70000B series and is available on DPO70000B models as Opt. PTH. The instrument can recover the clock signal, identify transitions, and allow you to set the desired encoded words for the serial pattern trigger to capture. Opt. PTH and the DSA70000B Series cover serial standard s up to 3.1 25 Gbps. For highe r bit
tandards like USB 3.0, option PTU on the DPO70000B s e ries and
rate s option STU on the DSA70000B series extend 8b10b triggering and decode to5Gbps.
Pattern Lock Triggering adds an extra dimension to NRZ serial pa ttern triggering by enabling the oscilloscope to take synchronized acquisitions
Large 10-Division XGA Display Screen
The DPO/DSA70000B Series have a large 12.1” (307.3 mm) XGA display with a touch screen with 10 vertical divisions give you 25% more vertical measurement resolution than other oscilloscopes.
Unmatched Usability
TekLink™ allows yo u to acquire more than four channels by synchronizing multiple oscilloscopes and debug multiple-lane h igh-spee d applications such as PCI Express X16 or DDR memory.
The DPO/DSA70000B Series instruments excel in usability with a s uite of productivity features, such as a touch screen, at menu structures, intuitive graphical icons, knob-per-channel vertical controls, right clicks, mouse wheel operation, and familiar Windo ws-based controls.
Interoperability with Logic A nalyzers for Digital Design and Debug
Tektronix’ Integrated View (iView™) data display enables digital designers to solve signal integrity challenges and effectively debug and verify their systems more quickly and easily. This integration allows designers to view time-correlated digital and analog data in the same display window, and
te the analog characteristics of the digital signals t hat are causing
isola system failures. No user calibration is required. And, once set up, the iView feature is completely automated.
4 www.tektronix.com/oscilloscopes
3 modes of operation of the horizontal time base
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
Unmatched
Get the Mos
Versatility
t of Your Oscilloscope by Fully Controlling
its Waveform Acquisition and Display Parameters.
Choose the operation you need to do your job the fastest: Automatic, Constant Sample Rate, or Manual. When you a re doing signal exploration and want a lively signal, the default Automatic mode provides you with the liveliest display update rate. If you want the highest real-time sample rate that will give you the most measurement accuracy, then the Constant
Sample Ra
provide the best real-time resolution. Finally the Manual mode ensures direct and independent contro l of the sample rate and record length for applications requiring specic settings.
te mode is for you. It will maintain the highest sample rate and
With the MyScope®Feature, Create Your Own Control Windows With Only the Controls, Features, and Capabilities that You Care About
reate your own personalized "toolbox" of oscilloscope features
Easily c in a matter of minutes using a simple, visual, drag-and-drop process. Once created, these custom control windows are easily accessed through a dedicated MyScope button and menu selection on the oscilloscope button/menu bar, just like any other control window. You can make an unlimited number of custom control windows, enabling each person who
e oscilloscope in a shared environment to have their own unique
uses th control window. MyScope control windows will benet all oscilloscope users, eliminating the ramp-up time that many face when returning to the lab after not using an oscilloscope for a while, and enables t he power user to be
Drag and drop menu items of interest to create the MyScope control window
far more efcient. Everything you need is found in one control window rather than navigating t hrough multiple menus to repeat similar tasks.
With OpenChoice®Software, Customize Your Test and Measurement System with Familiar Analysis Tools
The analysis and networking features of Open Choice sof tware add more exibility to Tektronix’ Windows XP oscilloscopes: Using the fast embedded bus, waveform data can be mo ved directly from acquisition
sis applications on the Windows desktop at much faster spe eds
to analy than conventional GPIB transfers. Tektronix’ implementation of industry standard protocols, such as TekVISA™ interface and ActiveX controls, are included for using and enhancing Windows applications for data analysis and documentation. IVI instrument drivers are included to enable easy communication with the oscilloscope using GPIB, RS-232, and LAN
ions from programs running on the instrument or an external PC. Or,
connect use the Software Developer’s Kit (SDK) to help create custom software to automate multistep processes in waveform collection and analysis with Visual BASIC, C, C++, MATLAB, LabVIEW, LabWindows/CVI, and other common Application Development Environments (ADE). Integration of the oscilloscope with exte rnal PCs and non-Windows hosts is also supported.
tion, the OpenChoice architecture provides a comprehensive
In addi software infrastructure for faster, more versatile operations. Data transfer utilities, such as the Excel or Word toolbar plug-ins can be used to simplify analysis and documentation on the Windows desktop or on an external PC.
www.tektronix.com/oscilloscopes 5
Data Sheet
More Insight into Your Complex Electrical Design for Characterization and Certication Testi ng
The DPO/DSA70000B series offers the industry’s most comprehensive set of analysis and certication tools, such as a simple m at h expression, waveform mask testing, a pass/fail tests, event searching, an d event marking. Cust development environments such as MATLAB set.
A Wide Range of Built-In Advanced Waveform Analysis Tools
Waveform cursors make it easy to measure trace-to-trace timing characteristics, while cursors that link between YT and XY display modes make it easy to investigate phase relationships and Safe Operating Area violations. palette that logically organizes measurements into Amplitude, Time, Combination, Histogram, and Communications categories. Gather further insight into your measurement results with statistical data such as mean, min, max, standard deviation, and population.
Dene and apply math expressions to waveform data for on-screen results in terms that you can use. Access common waveform m ath functions with the touch of a button. Or, for advanced applications, create algebraic expression functions, measurement values, scalars, and user-adjustable variables with an easy-to-use calculator-style editor.
FFT – To analyze your signal in the frequency domain, use the built-in
spectral an you can directly control the frequency span, center frequency, reference level, and resolution bandwidth. Vector signal analysis and Ultra-Wideband (UWB) verication and characterization are provided as part of the SignalVu and UWB options.
Filtering – Enhance your ability to isolate or remove a component of your
signal (noise or specic harmonics of the signal) by creating your ownltersorbyusingthefilters provided as standard with the instrument.
These customizable FIR lters can be used to implement today’s preferred signal-ltering techniques, including removing signal pre-emphasis or minimizin test.
om applications that you develop yourself using application
®
can further extend this tool
Select from 53 automatic measurements using a graphical
s consisting of live waveforms, reference waveforms, math
alysis function. Using the oscilloscope front-panel controls,
g the effects of xtures and cables connected to the device under
Capture da reports us
Basic
ta into Microsoft Excel using the unique Excel toolbar, and create custom
ing the Word toolbar
spectral UI control window
6 www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
Application Specic Solutions –
Enable standard speciccertification, measurement automation, and extended signal analysis
DPOJET Jitter And Eye Diagram Analysis (Opt. DJA) – Simplify identifying signal integrity concerns, jitter, and their related sources with DPOJET software. DPOJET provides the highest sensitivity and accuracy available for real-time oscilloscopes.
Advanced Event Search and Mark (Opt. ASM) – Ease the tedious task of examining data by highlighting important events, skipping the unimportant ones, and enhancing the comprehension of event relationships. Navigate between the events of interest effortlessly.
DDR Memory Bus Analysis (Opt. DDRA) – Automatically identify DDR1, LP-DDR1, DDR2, DDR3, and GDDR3 Reads and Writes and clearly see and analyze how analog anomalies are affecting your DDR/Memory. DDRA with DPOJET, combined with the DPO/DSA70000B DPX issues.
®
technology, is the fastest way to solve complex memory signaling
SDLA - Serial Data Link Analysis (options SLE and SLA) – Gain insight into serial data links with the capability to emulate the serial data channel from its S-parameters, de-embed the xture or other network, and add or remove transmitter equalization (de-emphasis/pre-emphasis). Option SLA adds processing of waveforms with FFE and DFE equalizations and automatic equalizer training. DPOJET provides advanced measurement and jitter analysis of the resulting waveforms.
www.tektronix.com/oscilloscopes 7
Data Sheet
DisplayPort Compliance Test Solution (Opt. DSPT) – Support DisplayPort Compliance Test Standard (CTS) source test with four-line simultaneous testing using the Tektronix P7300SMA Series probes and DisplayPort software. Detailed test reports with waveform plots, pass/fail results, and margin analysis are included.
DVI Compliance Test Solution (Opt. DVI) – Obtain quick and dependable results with the DVI compliance test software. Automated testing based on objective pass/fail detection without human intervention will dramatically enhance your productivity.
Ethernet Compliance Test Solution (Opt. ET3) – Receive full PHY layer support for Ethernet variants 10BASE-T, 100BASE-TX, and 1000BASE-T with Tektronix’ comprehensive, integrated Ethernet tool set. Analog verication, automated compliance software, and device characterization solutions are all included.
8 www.tektronix.com/oscilloscopes
FB-DIMM Compliance Test Solution (Opt. FBD) – Test the Receiver, Transmitter, and Reference clock test points for compliance as per FB-DIMM standards.
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
HDMI Compliance Test Solution (Opt. HT3) – Fast, efcient solution for HDMI compliance measurement challenges, no matter if you are working on a Source, Cable, or Sink solution. This application provides all the HDMI compliance test solutions you need to ensure quality and interoperability.
Power Measurement and Analysis Software (Opt. PWR) –Improvetheefficiency of switching power supplies with increased power densities. Measure and analyze power dissipation in power supply switching devices and magnetic components, and generate detailed test reports in customizable formats.
PCI Expres
PCI-Expre Option PCE or RT-Eye.
TekExpress™ SATA Automated Compliance Test Software – Reduce your compliance test time by approximately 70% with the simple, efcient testing and automation of all required testing suites provided by TekExpress software. Also included is auto-recognition of all required test equipment, precise DUT/Host control one-button testing, and complete support for SATA Gen1 and SATA Gen2 dened test suites.
s Analysis Test Solution (Opt. PCE) – Analyze the performance of your
ss Rev 1.0, 2.0, or 3.0 (draft spec) design with comprehensive test support.
enables tests that conform to PCI-SIG standards using either DPOJET
www.tektronix.com/oscilloscopes 9
Data Sheet
USB 2.0 Compliance Test Solution (Opt. USB) – Ensure your design has sufcient margins by characterizing your device with variable test limits. The USB compliance test solution provides “one touch” compliance testing by automatically setting up the oscilloscope for all compliance test congurations including probe deskew.
Ultra-Wideband Spectral Analysis Software (Opt. UWB) – Analyze hundreds of packet, TFC, and data-rate combinations without adjusting the UI. In addition, the automatic WiMedia modulation analysis conguration will analyze how complex wideband signals change frequency and amplitude with time using real-time spectrograms spanning 20 GHz.
0 Transmitter Testing – Verify, debug, and certify to the specications of
USB 3.
B 3.0 PHY layer using native capabilities in and additional applications for the
the US
DSA70000B series. Math function FIR lters and SDLA software, allow emulation
DPO/
e Cable and Reference channels dened in the USB 3.0 spec. With DPOJET, jitter
of th
urements needed to complete the Normative compliance test can be done.
meas
10 www.tektronix.com/oscilloscopes
SignalVu™ Vector Signal Analysis – Easily verify wide-bandwidth designs such as wideband radar, high data rate satellite links, or frequency-hopping radios and characterize wideband spectral events. SignalVu combines the functionality of a vector signal analyzer, a spectrum analyzer, and the powerful triggering capabilities of the DPO/DSA70000B – all in a single package.
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
DSA70000B – A dedicated solution congured for today’s high-speed serial design challenges
The DSA70000B Digital Serial Analyze r is specially congured to address high-speed serial data designs by encapsulating many of the serial domain features needed for high-speed serial verication and characterization. These standar DPO70000B series.
Serial Pattern Triggering – Real-time serial pattern triggering and protocol
decode with b uilt-in clock recovery recovers the clock signal, identies the transitions, the 8b/10b bit sequences decoded into their words for convenient analysis, or you can set the desired encoded words for the serial pattern trigger to capture. With Pattern Lock Triggering, the DSA70000B can synchronize to long serial test patterns up to 6.25 Gbps and remove random jitter. The DSA70000B Series covers serial standards up to 3.125 Gbps with an option to support up
Jitter, Timing, and Eye-diagram Analysis – The DSA70000B Series
features the highest-accuracy jitter and timing measurements as well as comprehensive analysis algorithms. Tight timing margins demand stable, low-jitter d clock cycles on every valid pulse in a single-shot acquisition. Multiple measurements and trend plots quickly show system timing under variable conditions, including Random and Deterministic Jitter separation.
Communicat
for verifying compliance to serial communications standards. 156 masks for the following standards are supported – PCI Express, ITU-T/ANSI T1.102, Ethernet IEEE 802.3, ANSI X3.263, Sonet/SDH, Fiber Channel, InniBand, USB, Serial ATA, Serial Attached SCSI, IEEE 1394b, RapidIO, OIF Standards.
20 MSample Record Length – 20 Msamples on all four chan nels provides
a longer time sequence at high resolution. Optional record lengths up to 100 Msamples for the 4, 6, and 8 GHz models, 250 Msamples for the 12.5, 16, and 20 GHz models e xtend the acquisition time sequence.
Advanced Event Search and Mark – Finding important events such as
fast or slow transitions, setup and hold violations, or logic patterns within a long-duration capture is made easy with the pattern matching and software triggering functionality of Event Search and Mark.
Accurate, Simple, and Customizable Physical Layer Certication Testing –
For designers with industry-standard certi cation needs, standard-specific compliance and analysis modules that congure the pass/fail waveform mask an d measurement limit testing are available as options to the DSA70000B and SAS (Opt. SST), FB-DIMM, (Opt. FBD), InniBand (Opt. IBA), HDMI (Opt. HT3), Ethernet (Opt. ET3), DisplayPort (Opt. DSPT), DVI (Opt. DVI), and USB (Opt. USB).
d features on the DSA70000B Series are options on the
and decodes characters and other protocol data. You can see
to 5 Gbps (Opt. STU)
esigns. You can make jitter measurements over contiguous
ions Mask Testing –
. Modules are available for PCI E xpress (Opt. PCE), Serial ATA
Provides a complete portfolio of masks
ming, and Eye-diagram Analysis
Jitter, Ti
Communications Mask Testing
standard features that extend the functionality of the Tektronix
With DPO70000B series to address high-speed serial signal analysis and certication, the DSA70000B series offers a specialized instrument that efciently addresses y ou r design challenges.
www.tektronix.com/oscilloscopes 11
Data Sheet
Characteristics
Verti cal S yst
DPO/DSA Models
Bandwidth (user
em
70404B 70604B 70804B 71254B 71604B 72004B
4 GHz 6 GHz 8 GHz 12.5 GHz 16 GHz 20 GHz selectable DSP enhance)
Hardware Analog
4 GHz 6 GHz 8 GHz 12.5 GHz 16 GHz (typical) 16 GHz (typical) Bandwidth (-3 dB)
Input Channels Rise Time 10% to 90%
444444
98 ps 65 ps 49 ps 32 ps 24.5 ps 19 ps
(typical) Rise Time 20% to 80%
68 ps 45 ps 34 ps 22 ps 17 ps 14 ps
(typical) Vertical Noise (% of full
scale) (typical)*
16
Bandwidth Limits
0.28% 0.32% 0.35% 0.38% 0.43% 0.77%
Dependingoninstrumentmodel: 19GHz,18GHz,17GHz,16GHz,15GHz,14GHz,13GHz,12GHz,11GHz,10GHz,
9GHz,8GHz,7GHz,6GHz,5GHz,4GHz,3GHz,2GHz,1GHz,or500MHz
Channel-to-channel Isolation (Any Two Channels at Equal Vertical Scale Settings)
120:1 (for input frequency 0 to 10 GHz)80:1 (for input frequency >10 GHz to 12 GHz.50:1 (for input frequency >12 GHz to 15 GHz)
25:1 (for input frequency >15 GHz)
DC Gain Accuracy ±2% (of reading) Delay between any two
channels (typical) Effective Number of
100 ps for any two channels with equal V/div and coupling settings
50 ps with BW enhance enabled (BW+)
14
5.4 bits*
Bits (typical) Input Coupling DC (50 ), GND Input Impedance Input Sensitivity
18 GHz and below 20 GHz and 19 GHz
Max Input Voltage, 50
<5.0 V
50 ±2%, 1 Mwith TCA-1MEG adapter
10 mV/div to 500 mV/div (100 mV to 5 V full scale)
20to500mV/div(200mVto5Vfullscale)
for 100 mV/div; also determined by TekConnect accessory
RMS
1.0 V
for <100 mV/div
RMS
Offset Accuracy 10 mV/div – 99.5 mV/div. ±(0.35% (offset value-position) + 1.5 mV + 1% of full scale)
100 mV/div – 500 mV/div. ±(0.35% (offset value-position) + 7.5 mV + 1% of full scale)
Offset Range 10 mV/div: ±450 mV
20 mV/div: ±400 mV 50 mV/div: ±250 mV
100 mV/div: ±2.0 V 200 mV/div: ±1.5 V 500 mV/div: ±0.0 V
Passband Flatness (20,
±0.5 dB to 50% of nominal bandwidth at 25 °C
50, 100, 250 mV/div) (typical)
Position Range ±5 div Vertical Resolution
*1450 mV/div, bandwidth lteron,maxbandwidthupto13GHz,maxsamplerate
16
50 mV/div, bandwidth lter on
*
8 bit (11 bit with averaging)
12 www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
Time Base System
DPO/DSA Models
Time Base Range Time Resolution (in
ET/IT mode) Channel-to-channel
Deskew Delta Time
Measurement Accuracy RMS Over <100 ns duration; single shot; with signal rise time = 1.2X scope rise time
Jitter Noise Floor (typical) (With BW+ bandwidth enhance enabled)
Time Base Accuracy ±1.5 ppm initial accuracy, aging <1 ppm per year Time Base Delay
Time Range TriggerJitter(RMS) 1ps
70404B 70604B 70804B 71254B 71604B 72004B
20 ps/div to 1000 s/div 10 ps/div to 1000 s/div
200 fs 100 fs
Range ±75 ns
1.61 ps 1.29 ps 1.14 ps
940 fs 900 fs
450 fs 450 fs 450 fs 300 fs 300 fs 400 fs
-5.0 ks to 1.0 ks
(typical) with enhanced t riggering OFF
RMS
<100 fs
with enhanced triggering ON
RMS
1.02 ps
Acquisition System
DPO/DSA Models 70404B / 70604B / 70804B 71254B / 71604B / 72004B
Sample Rates
Real-time mode 1, 2, 3, or 4 channel (max) 25 GS/s 50 GS/s ET/IT Mode (max) 5 TS/s 10 TS/s
Maximum Record Length per Channel
With Standard Conguration 10 M on all four channels (DPO70000B Series only)
20 M on all four channels (DSA70000B Series only) With Record Length Opt. 2XL 20 M on all four channels (DPO70000B Series only) With Record Length Opt. 5XL 50 M on all four channels With Record Length Opt. 10XL 100 M on all four channels With Record Length Opt. 20XL N/A 250 M on all four channels
Maximum Duration at Highest Real-Time Resolution
DPO/DSA Models 70404B / 70604B / 70804B 71254B / 71604B / 72004B
Resolution Max Duration with Standard Memory 0.4 ms DPO70000B Series; 0.8 ms for
40 ps (25 GS/s) 20 ps (50 GS/s)
0.2 ms DPO70000B Series; 0.4 ms for
DSA70000B Series
DSA70000B Series
Max Duration with Opt. 2XL 0.8 ms (DPO70000B Series only) 0.4 ms (DPO70000B Series only) Max Duration with Opt. 5XL Max Duration with Opt. 10XL Max Duration with Opt. 20XL N/A
2.0 ms 1.0 ms
4.0 ms 2.0 ms
5.0 ms
www.tektronix.com/oscilloscopes 13
Data Sheet
Acquisition Modes
Mode Description
Averaging Envelope FastAcq Acquisition Mode Maximum FastAcq Waveform Capture Rate >300,000 wfms/s on all 4 channels simultaneously FastFrame™ Acquisition
Hi-Res Real-time boxcar averaging reduces random noise and increases resolution Peak Detect Roll Mode Sample Waveform Database Accumulate waveform database providing three-dimensional array of amplitude, time, and counts
Pinpoint®Trigg er System
Sensitivity
Internal DC Coup led 4% of full scale from DC to 50 MHz
External (Auxiliary Input) 50 250 mV from DC to 50 MHz, increasing to 350 mV at 1.0 GHz.
Trigger Characteristics
A Event and Delayed B Event Trigger Types
Main Trigger Modes Trigger Sequences Main, Delayed by Time, Delayed by Events, Reset by Time, Reset by State, Reset by Transition. All sequences
Clock Recovery System
ock Recovery Phase Locked Loop Bandwidth
Cl
ock Recovery Frequency Range
Cl
8b10b Max Baud Rate
Communications-related Triggers
Serial Pattern Trigger
AUX Trigger Clock Recovery Jitter (RMS) <0.25% bit period + 2 ps
Enhanced Triggering
Line Fixed at 0 V Minimum Signal Amplitude needed for Clock Recovery 1 div
Tracking/Acquisition Range ±2% of requested baud Trigger Coupling DC, AC (attenuates <100 Hz),
Trigger Holdoff Range Trigger Level Range Internal
From 2 to 10,000 waveforms included in average From 1 to 2×10
9
waveforms included in min-max envelope
FastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events
Acquisition memory divided into segments; maximum trigger rate >310,000 waveforms per second. Time of arrival recorded with each event. Frame nder tool helps to visually identify transients.
Captures narrow glitches at all real-time sampling rates: 1 ns at 125 MS/s; 1/sample rate at 250 MS/s Up to 10 MS/s with a maximum record length of 40 MS Acquire sampled values
70404B / 70604B / 70804B / 71254B /
DPO Models
71604B / 72004B
70404B / 70604B / 70804B / 71254B /
DSA Models
71604B / 72004B
10% of full scale at 4 GHz 20% of full scale at 8 GHz
50% of full scale at 11 GHz
Edge, Glitch, Runt, Width, Transition Time, Time-out, Pattern, State, Setup/Hold, Window—all except
Edge, Pattern, and State can be Logic State qualied by up to two channels
Auto, Normal, and Single
can include separate horizontal delay after the trigger event to position the acquisition window in time.
Requires Opt. PTH or Opt. MTH Standard
xed at FBaud/1600
Fi
5 MBaud to 3.125 GBaud
1.
Requires Opt. PTU Requires Opt. STU
5GBaud
Requires Opt. MTH Standard
Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded communications signals. Select among isolated
positive or negative one, zero pulse form or eye patterns as applicable to the standard.
Requires Opt. PTH Stan dard
Up to 64 bit serial word recognizer, bits specied in binary (high, low, don’t care) or hex format.
Trigger on NRZ-encoded data up to 1.25 GBaud.
Trigger on 8b/10b-encoded data from 1.25 to 3.125 GBaud (40 bits)
Requires Opt. PTU Requires Opt. STU
Trigger on 8b/10b-encoded data up to 5 GBaud (40 bits)
TekConnect interface: ±5 V
for PRBS data patterns
<0.25% bit period + 1.5 ps
RMS
for repeating "0011” data pattern
RMS
User-selectable; enhanced triggering corrects the difference in timing between the trigger path and the acquired data
path (supports all Pinpoint trigger types on both A- and B-Events except pattern trigger); Not available in FastAcq.
up to 1.25 Gbaud
p-p
1.5 div
above 1.25 Gbaud
p-p
HF Rej (attenuates >20 kHz),
LF Rej (attenuates <200 kHz),
Noise Reject (reduces sensitivity)
250 ns min to 12 s max
±120% of full scale from center of screen
14 www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
Trigger Modes
Mode Description
Comm Standard feature on the DSA70000B, provided as part of Opt.
Edge
Glitch Trigger on or reject glitches of positive, negative, or either
Pattern
Runt
Serial Pattern Trigger on NRZ-encoded data up to 3.125 Gbaud (5 Gbaud
Setup/Hold Trigger on violations of both setup time and hold time
State Any logical pattern of channels (1, 2, 3) clocked by edge on
Time-out
Transition
Trigger Delay by Events
Trigger Delay by Time
Width
Window
MTH on the DPO70000B Series. Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded signals.
Positive or negative slope on any channel or front panel auxiliary input. Coupling includes DC, AC, noise reject, HF reject, and LF reject.
polarity. Minimum glitch width isdownto150ps(typical)with rearm time of 300 ps
Trigger when pattern goes false or stays true for specied period of time. Pattern (AND, OR, NAND, NOR) specied for four input channels dened as high, low, or don’t care.
Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the rst again. Event can be time- or logic-qualied.
with Opt. PTU or Opt. STU); above 1.25 Gbaud requires 8b/10b encoded data. Extended with pattern lock triggering to capture repeated acquisitions of long serial test patterns up to 6.25 Gbps.
between clock and data present on any two input channels.
channel 4. Trigger on rising or falling clock edge. Trigger on an event which remains high, low, or either, for a
specied time period. Selectable from 300 ps. Trigger on pulse edge rates that are faster or slower than
specied. Slope may be positive, negative, or either. 1to2Gevents
3.2nsto3Ms
Trigger on width of positive or negative pulse either within or out of selectable time limits (down to 150 p s).
Trigger on an event that enters or exits a window dened by two user-adjustable thresholds. E vent can be time- or logic-qualied.
Search and Mark Events
Event Description
Basic
Advanced
Mark any events and document waveforms. Search positive, negative slopes or both on any channels. Event table summarizes all found events. All events are time stamped in reference to trigger position. Users can choose to stop acquisitions when an event is found.
Search glitches or runts, as well as transition rate, pulse width, setup and hold, time-out, window violations, or nd any logic or state pattern on any number of channels. Search DDR read or write bursts with Opt. DDRA.
Waveform Measurements
Measurement Description
Automatic Measurements
Amplitude Related Amplitude, High, Low, Maximum, Minimum, Peak-to-Peak,
Combination Area, Cycle Area, Phase, Burst Width Eye-pattern Related
Histogram Related
Time Related Rise Time, Fall Time, Positive Width, Negative Width,
53, of which 8 can be displayed on screen at any one time; measurement statistics, user-denable reference levels, measurement within gates isolating the specic occurrence within an acquisition to take measurements on.
Mean, Cycle Mean, RMS, Cycle RMS, Positive Overshoot, Negative Overshoot
Extinction Ratio (absolute, %, dB), Eye Height, Eye Width, Eye Top, Eye Base, Crossing %, Jitter(p-p,RMS,6sigma), Noise (p-p, RMS), Signal/Noise Ratio, Cycle Distortion, Q-Factor
Waveform Count, Hits in Box, Peak Hits, Median, Maximum, Minimum, Peak-to-Peak, Mean (μ), Standard Deviation (sigma), μ+1sigma, μ+2sigma, μ+3sigma
Positive Duty Cycle, Negative Duty Cycle, Period, Frequency, Delay
Waveform Processing/Math
Processing Type Description
Algebraic Expressions
Arithmetic Filtering Functions
Frequency Domain Functions
Mask Function
Math Functions
Relational Vertical Units
Waveform Denition Window Functions Rectangular, Hamming, Hanning, Kaiser-Bessel,
Dene extensive algebraic expressions including Waveforms, Scalars, User-adjustable Variables and Results of Parametric Measurements e.g. (Integral (CH.1–Mean(CH.1))×1.414×VAR1)
Add, Subtract, Multiply, Divide Waveforms and Scalars User-denable lters. Users specify a le containing the
coefcients of the lter. Several example lter les are provided.
Spectral Magnitude and Phase, Real and Imaginary Spectra
unction that generates a Waveform Database pixmap from
Af
ample waveform. Sample count can be dened.
as Average, Invert, Integrate, Differentiate, Square Root,
Exponential, Log Cos, Tan, ASin, ACos, ATan, Sinh, Cosh, Tanh
Boolean result of comparison >, <, ≥, ≤,==,!= Magnitude: Linear, dB, dBm Phase: Degrees, radians, group delay IRE and mV units As an arbitrary math expression
Blackman-Harris, Gaussian, Flattop2, Tek Exponential
, Loge, Abs, Ceiling, Floor, Min, Max, Sin,
10
Display Characteristics
Characteristic
Color Palettes Normal, Green, Gray, Temperature, Spectral, and
Display Format YT, XY Display Resolution Display Size Diagonal: 307.3 mm (12.1 in.) Display Type Liquid crystal active-matrix color display Horizontal Divisions 10 Vertical Divisions 10 Waveform Styles Vectors, Dots, Variable Persistence, Innite Persistence
Description
User-dened
XGA 1024 horizontal × 768 vertical pixels
www.tektronix.com/oscilloscopes 15
Data Sheet
Computer System and Peripherals
Item Description
Operating System
Windows XP CPU Intel Pentium 4, 3.4-GHz processor PC System Memory 2 GB Hard Disk Drive
Rear-panel, removable hard disk drive, 160 GB capacity CD-R/W Drive Front-panel CD-R/W drive with CD creation software
application DVD Drive Read only Mouse Keyboard
Optical wheel mouse, USB interface
USB interface
Input/Output Ports
Front Panel
Port Description
AUX Trigger Input AUX Trigger Output BNC connector, provides a TTL-compatible,
DC Probe Calibration Output
Fast Edge Output SMA connector provides fast edge signal. 1 kHz ±20%;
Recovered Clock SMA connector, 1.25 Gbps, Output swing 130 mV
Recovered Data
USB 2.0 Port One in front. Allows connection of USB keyboard, mouse,
See trigger specications
polarity-switchable pulse when the oscilloscope
triggers.
BNC connector, ±10 V DC for DC probe calibration. (Signal
available only during probe calibration.)
810 mV (base to top) ±20% into 10 kload; 440 mV ±20%
into a 50 load..
50 at 1.25 Gbps. Requires Opt. PTH or Opt. MTH to
p-p
enable on DPO70000B, standard on DSA70000B.
SMA connector, 1.25 Gbps, Output swing of 1010 repeating
pattern 200 mV into 50 at 1.25 Gbps. Requires Opt.
PTH or Opt. MTH to enable on DPO70000B, standard on
DSA70000B.
or storage device.
into
Rear Panel
Port Description
Audio Ports
Miniature phone jacks for stereo microphone input and stereo line output.
AUX Trigger Output BNC connector, 0 to 3 V; default output is A-Event Trigger
low true
External Tim Reference I
eBase
n
BNC connector; allows time base system to phase lock to external 10/100 MHz reference. Optimized (by using a software switch) for either a highly stable clock or tracking mode.
GPIB Port Keyboard Port
IEEE 488.2 standard PS/2 compatible
LAN Port RJ-45 connector, supports 10Base-T, 100Base-T, and
1000Base-T
Mouse Port
PS/2 compatible Parallel Port IEEE 1284, DB-25 connector Power
Scope X Port
GA Video
100to240V
<870 Watts, 400 Hz; CAT II, <1100 VA typical
15 pin D
compat
ing live waveforms on an external monitor or projector.
includ
The pri
nal monitor using this port.
exter
, ±10%, 50/60 Hz; 115 V
RMS
RMS
±10%,
-sub connector on the rear panel, video is IBM XGA
ible. Connects to show the oscilloscope display,
mary Windows desktop can also be displayed on an
Serial Port DB-9 COM1 port TekLink ™
Synchronizes multiple Tektronix scopes together to a llow
simultaneous acquisition of more than four channels. Time Base
Reference Out
BNC connector; provides TTL-compatible output of internal
10 MHz reference oscillator USB 2.0 Ports Four in back. Allow connection of USB keyboard, mouse,
or storage device. Windows Video Port 15 pin D-sub connector on the rear panel; connects a second
monitor to use dual-monitor display mode allowing analysis
results and plots to be viewed along with the oscilloscope
display. Video is DDC2B compliant.
16 www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
Physical Characteristics
Dimensions
Benchtop Conguration
mm
in.
Height 298 11.74 Width 451 17.75 Depth 489.97 19.29
Weight kg lbs.
Net 20 44 Shipping
Rackmount Conguration
34
mm in.
75
Height 311 12.25 Width 480.1 18.9 Depth (from rack mounting
546.1 21.5
ear to back of instrument)
Weight kg lbs.
Net 20 44 Kit 2.7 6
Mechanical
Cooling — Required Clearance
mm
Top 0 0 Bottom 0 0 Left side
76 3 Right side 76 3 Front 0 0 Rear 0 0
in.
Environmental
Temp e r ature
Operating 5 °C to +45 °C Nonoperating
Humidity
–20°Cto+60°C
Operating 8% to 80% relative humidity (RH) at up to 32 °C. 5% to 45%
RH above +32 °C up to +45 °C
Nonoperating
5% to 95% relative humidity (RH). Upper limit derated to 45% RH above +30 °C up to +60 °C
Altitude
Operating 10,000 ft. (3,048 m) Nonopera
Regulatory
Electromagnetic
ting
40,000 ft. (12,190 m)
93/68/EEC; EN61326:1997 +A1 1998+A2:2000
Compatibility Certi
cations
-1, CSA1010.1, ISO11469,EN61010-1, IEC 61010-1
UL 3111
Ordering Information
Model Description
DPO70404B 4 GHz Digital Phosphor Oscilloscope DPO70604B 6 GHz Digital Phosphor Oscilloscope DPO70804B 8 GHz Digital Phosphor Oscilloscope DPO71254B 12.5 GHz Digital Phosphor Oscilloscope DPO71604B 16 GHz Digital Phosphor Oscilloscope DPO72004B 20 GHz Digital Phosphor Oscilloscope DSA70404B 4 GHz Digital Serial Analyzer DSA70604B 6 GHz Digital Serial Analyzer DSA70804B 8 GHz Digital Serial Analyzer DSA71254B 12.5 GHz Digital Serial Analyzer DSA71604B 16 GHz Digital Serial Analyzer DSA72004B 20 GHz Digital Serial Analyzer
All Models Include: Accessory pouch, front cover, mouse, keyboard, quick start
user manual (071-173x-xx), DPO70000B Series product software CD/DVD-ROM, Optional applications software CD/DVD-ROM, performance verication procedure PDF le, GPIB programmer’s reference (on product software CD/DVD-ROM), calibration certicate documenting NIST traceability, Z 540-1 compliance and ISO9001, power cord, one year warranty. Note: Please specify quick-start u ser manual language and power plug when ordering. (4) TekConnect adapter (TCA-BNC)
®
to 2.92 mm adapters (TCA-292MM) and (1) Tekconnect to BNC
www.tektronix.com/oscilloscopes 17
Data Sheet
Options
Instrument Options
Option
Record Length Options for DPO70000B Series
Opt. 2XL 20 MSamples/ch Opt. 5XL 50 MSamples/ch Opt. 10XL 100 MSamples/ch Opt. 20XL*
8
Record Length Options for DSA70000B Series
Opt. 5XL 50 MSamples/ch Opt. 10XL 100 MSamples/ch Opt. 20XL*
8
Software Options for DPO70000B Series
Opt. ASM Advanced Event Search and Mark Opt. MTH Mask testing for Serial Standards up to 4.25 Gbps. Includes
Opt. PTH Protocol Triggering and Decoding for 8b/10b-encoded Serial
Opt. PTU Protocol Triggering and Decoding for 8b/10b-encoded Serial
ware Options for DSA70000B Series
Soft
Opt. STU Protocol Triggering and Decoding for 8b/10b-encoded Serial
Software Options for DPO70000B Series and DSA70000B Series
Opt. DDRA* Opt. DJA DPOJET Jitter and Eye Diagram Analysis Opt. DSPT*
11
Opt. DVI DVI Compliance Test Solution Opt. ET3* Opt. FBD*
2
5
Opt. HT3 HDMI Compliance Test Software Opt. IBA*
5
Opt. LT Waveform Limit Testing Opt. PCE* Opt. PWR*
15
4
Opt. RTE Serial Data Compliance and Analysis Software Opt. SLA Serial Data Link Analysis Advanced (with Equalization) Opt. SLE Serial Data Link Analysis Essentials (no Equalization) Opt. SST*
5
Opt. SVE SignalVu™ Essentials – Vector Signal Analysis Software Opt. SVM* Opt. SVP*
Opt. USB*
10
10
3
Opt UWB Ultra-Wideband Spectral Analysis (includes WiMedia Alliance
Description
250 MSamples/ch
250 MSamples/ch
hardware clock recovery.
Signals up to 3.125 Gbps. Includes hardware clock recovery and pattern lock triggering.
Signals up to 5 Gbps.
Signals up to 5 Gbps.
9
DDR Memory Bus Analysis
DisplayPort Compliance Test Solution
Ethernet Compliance Test Software FB-DIMM Compliance Module for RT-Eye Ser ial Data
Compliance and Analysis Software
InniBand Compliance Module for RT-Eye Serial Data Compliance and Analysis Software
PCI Express™ Analysis Software Power Measurement and Analysis Software
SATA and SAS Analysis Software Module for RT-Eye Serial Data Compliance and Analysis Software
General Purpose Modulation Analysis. Requires option SVE Advanced Signal Analysis (including pulse measurements).
Requires option SVE USB 2.0 Compliance Test Software only
PHY interf. testing)
Option
Description
Opt UWBE Ultra-Wideband Spectral Analysis Essentials
*2Requires Ethernet Test Fixture.
3
Requires TDSUSBF (USB Test Fixture).
*
4
At least Opt. 2XL and a TCA-1MEG TekConnect 1 Mbuffer amplier are recommended for use.
*
5
Requires Opt. RTE on DPO70000B Series.
*
8
For models of bandwidth 12.5 GHz only.
*
9
Requires DJA and ASM.
*
10
Requires Opt. SVE, SVEH, or SVEU.
*
11
For models of bandwidth 8GHzonly.
*
15
Requires Opt. RTE or Opt. DJA on DPO70000B Series
*
User Manual Options
Option
Opt. L0 Opt. L1
Description
English
French Opt. L3 German Opt. L5
Japanese Opt. L7 Simple Chinese Opt. L8 Standard Chinese Opt. L9 Opt. L10 Opt. L99
Korean
Russian
No manual
Power Plug Options
Option
Opt. A0 Opt. A1 Opt. A2 Opt. A3
Description
North America
Universal European Union
UK
Australia Opt. A5 Switzerland Opt. A6
Japan Opt. A10 China Opt. A11 Opt. A99
India
No power cord
Service Options
Option
Description
Opt. CA1 Provides a single calibration event or coverage for the
designated calibration interval, whichever comes rst Opt. C3 Calibration Service 3 Years Opt. C5 Calibration Service 5 Years Opt. D1 Calibration Data Report Opt. D3 Calibration Data Report 3 Years (with Opt. C3) Opt. D5 Calibration Data Report 5 Years (with Opt. C5) Opt. R3 Repair Service 3 Years Opt. R5 Repair Service 5 Years
18 www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
Recommended Accessories
Probes
Probe Description
P7520 P7516 P7513A P7313SMA 13 GHz TekConnect®differential SMA probe P7508 P7380SMA 8 GHz TekConnect®differential SMA probe P7506 P7504 P6251
P6250
TCPA300/TCPA400 Series current measurement systems P5200/P5205/P5210 High-voltage differential probes
Adapter
Adapter Description
TCA-1MEG TekConnect high-impedance buffer amplier. Includes
TCA-292MM TekConnect to 2.92 mm connectors TCA-BNC TekConnect-to-BNC adapter TCA-N TekConnect-to-N adapter TCA-SMA TekConnect-to-SMA adapter TCA75 8 GHz precision TekConnect 75 to 50 adapter with 75
s
20 GHz TriMode™ probe 16 GHz TriMode™ probe 13 GHz TriMode™ probe
8 GHz TriMode™ probe
6 GHz TriMode™ probe 4 GHz TriMode™ probe DC to 1 GHz, 42V, Differential Probe (requires TCA-BNC
adapter) DC to 500 MHz, 42V, Differential Probe (requires TCA-BNC
adapter)
P6139A passive probe
BNC input connector
Cables
Cable Order
Centronics Cable GPIBCable(1m) GPIBCable(2m) RS-232 Cable
012-1214-00
012-0991-01
012-0991-00
012-1298-00
Accessories
Accessory
Ethernet Test Fixture
Instrumented DIMM for DDR3
Oscilloscope Cart Probe Calibration
and Deskew Fixture (4 GHz)
Probe Deskew Fixture (>4 GHz)
Power Deskew Fixture
Rackmount Kit 016-1985-00 Service Manual TDSUSBF Test xtureforusewithOpt. USB Transit Case
Order
Order through Crescent Heart Software
(http://www.c-h-s.com)
Order Scope NEXVu card for UDIMM Raw Card E. (Contact
http://www.nexustechnology.com)
K4000
067-0484-xx
067-1586-xx
067-1686-xx
071-1740-xx
016-1977-00
www.tektronix.com/oscilloscopes 19
Data Sheet
Instrument Up grades
To upgrade your DPO70000B Series Oscilloscope or your DSA70000B Series Serial Analyzer, order DPO7UP and option as noted.
Option Description
To upgrade re
XL02 XL25 XL210 XL220*
cord length on DPO70000B Series from:
Standard conguration to Opt. 2XL conguration Opt. 2XL conguration to Opt. 5XL conguration Opt. 2XL conguration to Opt. 10XL conguration
8
Opt. 2XL conguration to Opt. 20XL conguration (only available on instruments of bandwidth >= 12.5 GHz)
To upgrad
XL05 XL010 XL020*
erecordlengthfromstandardconfiguration to:
Opt. 5XL conguration Opt. 10XL conguration
8
Opt. 20XL conguration (only available on instruments of bandwidth >= 12.5 GHz)
ade record length on DPO70000B Series or DSA70000B
To upgr
Series from:
XL510 XL520*
8
Opt. 5XL conguration to Opt. 10XL conguration Opt. 5XL conguration to Opt. 20XL conguration (only
available on instruments of bandwidth >= 12.5 GHz)
8
XL1020*
Opt. 10XL conguration to Opt. 20XL conguration (only available on instruments of bandwidth >= 12.5 GHz)
To upgrade bandwidth on DPO70000B Series or DSA70000B Series
from:
BWU4T6 BWU4T8 BWU4T12 BWU4T16 BWU4T20 BWU6T8 BWU6T12 BWU6T16 BWU6T20 BWU8T12 BWU8T16 BWU8T20 BWU12T16 BWU12T20 BWU16T20
4GHzto6GHzconfiguration 4GHzto8GHzconfiguration 4GHzto12.5GHzconfiguration 4GHzto16GHzconfiguration 4GHzto20GHzconfiguration 6GHzto8GHzconfiguration 6GHzto12.5GHzconfiguration 6GHzto16GHzconfiguration 6GHzto20GHzconfiguration 8GHzto12.5GHzconfiguration 8GHzto16GHzconfiguration 8GHzto20GHzconfiguration
12.5 GHz to 16 GHz conguration
12.5 GHz to 20 GHz conguration 16GHzto20GHzconfiguration
To up grad e:
ASM DPO70000B with Opt. ASM MTH PTH DJAH DJAU
DPO70000B with Opt. MTH DPO70000B with Opt. PTH DPO70404B, 70604B, 70804B with Opt. DJA DPO71254B, 71604B, 72004B with Opt. DJA
To upgrade DPO
6
CP2*
70000B Series or DSA70000B Series with:
TDSCPM2 ANSI/ITU Telecom pulse co mpliance testing
software (requires Opt. MTH on DPO70000B Series)
9
DDRA* DVI
13
EQ* ET3 FBD* HT3
5
IBA* J2 LT PCE* PWR
5
15
Opt. DDRA
Opt. DVI
From Opt SLE to Opt. SLA
Opt. ET3
Opt. FBD
Opt. HT3
Opt. IBA
TDSDDM2 disk-drive analysis software
Opt. LT
Opt. PCE
Opt. PWR SLA Opt. SLA SLE Opt. SLE
5
SST* SVEH* SVEU* SVM* SVP*
11
12
10
10
Opt. SST
Opt. SVE
Opt. SVE
Opt. SVM
Opt. SVP USB Opt. USB UWB UWBE VNM*
7
Opt. UWB
Opt. UWBE
TDSVNM CAN and LIN Timing and Protocol Decode (no
CAN triggering included)
*5Requires Opt. RTE on DPO70000B Series.
6
Requires Opt. MTH on DPO70000B Series
*
7
*
Requires ATM1 CAN/LIN trigger module - Order through Crescent Heart Software
8
For models of bandwidth 12.5 GHz only
*
9
Requires DJA and ASM
*
10
*
Requires Opt. SVE, SVEH, or SVEU
11
DPO/DSA70404B, DPO/DSA70604B, DPO/DSA70804B only
*
12
DPO/DSA71254B, DPO/DSA71604B, DPO/DSA72004B only
*
13
*
Request Opt. SLE
15
Requires Opt. RTE or Opt. DJA on DPO70000B Series
*
Product(s) are manufactured in ISO registered facilities.
20 www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes/Digital Serial Analyzers — DPO/DSA70000B Series
www.tektronix.com/oscilloscopes 21
Data Sheet
Contact Tektronix:
ASEAN / Australa
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Central East Eu
Mexico, Central/South America & Caribbean 52 (55) 54247900
rope, Ukraine, and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Luxembourg +44 (0) 1344 392400
Middle Eas t,
For other areas contact Tektronix, Inc at: 1 (503) 627-7111
Asia, and North Africa +41 52 675 3777
The N etherlands 090 02 021797
People’s Rep
United Kingdom & Ireland +44 (0) 1344 392400
ublic of China 86 (10) 6235 1230
Republic of
sia (65) 6356 3900
Austria +41 52 675 3777
Belgium 07 81 601
Brazil +55 (11) 40669400
Canada 1 (800) 661-5625
Denmark +4580881401
Finland +41526
France +33(0)169868181
Germany +49 (221) 94 7 7 400
Hong Kong (852
India (91) 80-42922600
Italy +39 (02) 25086 1
Japan 81 (3) 67
Poland +41 52 675 3777
Portugal 80 08 12370
Korea 82 (2) 6917-5000
Russia & CIS +7 (495) 7484900
South Africa +27 11 206 8360
Spain (+3 4)
Sweden 020 08 80371
Switzerland +41 52 675 3777
Tai wa n 886 (
USA 1 (800) 426-2200
75 3777
) 2585-6688
14-3010
Norway 800 16098
901 988 05 4
2) 2722-9622
Updated 30 October 2008
66
www.tektronix.com/oscilloscopes
For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com
t © Tektronix, Inc. All rights r eserved. Tektronix products are covered by U.S. and foreign patents,
Copyrigh
d pending. Information in this publication supersedes that in all previously published material.
issued an
tion and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of
Specica
x, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks
Tek tro ni
espective companies.
of their r
29 Jan 2009 55W-23032-1
Loading...