Tektronix DPO70000 Series, DPO70000B Series, DSA70000 Series, DSA70000B Series, MSO70000 Series Technical Reference

...
xx
MSO70000 Series Mixed Signal Oscilloscopes DSA70000/B Series Digital Signal Analyzers DPO70000/B Series Digital Phosphor Oscilloscopes DPO7000 Series Digital Phosphor Oscilloscopes
ZZZ
Specications and Performance Verication
Technical Reference
077-0063-04
xx
MSO70000 Series Mixed Signal Oscilloscopes DSA70000/B Series Digital Signal Analyzers DPO70000/B Series Digital Phosphor Oscilloscopes DPO7000 Series Digital Phosphor Oscilloscopes
ZZZ
Specications and Performance Verication
Technical Reference
This document applies to rmware version 5.2.0 and above.
Warning
The servicing instructions are for use by qualied personnel only. To avoid per unless you are qualied to do so. Refer to all safety summaries prior to performi
sonal injury, do not perform any servicing
ng service.
www.tektronix.com
077-0063-04
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Pinpoint, TekLink, and RT-Eye are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O . Bo x 50 0 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. World wide , vi s it www.tektronix.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be n the property of Tektronix.
ew or reconditioned to like new performance. All replaced parts, modules and products become
In order to o the warranty period a nd make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage result b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
TEK AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
btain service under this warranty, Customer must notify Tektronix of the defect before the expiration of
ing from attempts by personnel other than Tektronix representatives to install, repair or service the product;
Table of Contents
General Safety Summary .......................................................................................... v
Service Safety Summary........................ ................................ ................................ . vii
Specications
Specications .......... .................................. ................................ ......................... 1-1
Specication Tables ......................................................................................... 1-1
Performance Verication
Performance Verication ... ................................ ................................ ..................... 2-1
Conventions .................................................................................................. 2-2
Brief Procedures ........... ................................ .................................. ..................... 2-4
Self Tests............................... ................................ .................................. ..... 2-4
Functional Tests.............................................................................................. 2-5
rmance Tests ......... ................................ .................................. ................... 2-18
Perfo
Prerequisites ......................... .................................. ................................ ..... 2-18
Equipment Required....................................................................................... 2-18
MSO70000 Series, DSA/DPO70000/B, and DPO7000 Series Test Record ........................ 2-21
Signal Acquisition System Checks .... .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . . 2-49
Time Base System Checks................................................................................ 2-86
gger System Checks................................... ................................ ................. 2-97
Tri
Output Signal Checks .......................... .................................. ........................ 2-111
Serial Trigger Checks (Optional on Some Models)..... .................................. ............ 2-116
Sine Wave Generator Leveling Procedure ................... ................................ .......... 2-126
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication i
Table of Contents
List of Figure
Figure 2-1: Toolbar and menu bar (< 4 GHz models shown)................................................ 2-3
Figure 2-2: Universal test hookup for functional tests - Ch 1 shown....................................... 2-6
Figure 2-3: Channel button location..................................... ................................ ....... 2-7
Figure 2-4: Setup for time base test ....... ................................ ................................ ... 2-10
Figure 2-5: Setup for trigger test........... ................................ ................................ ... 2-12
Figure 2-6: Setup for the le system test .................................................................... 2-14
Figure 2-7: Setup for the digital channels test.......... ................................ ..................... 2-15
Figure 2-8: Initial test hookup . . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . . 2-50
Figure 2-9: Measurement of DC accuracy at maximum offset and position...................... ....... 2-54
Figure 2-10: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-55
Figure 2-11: Measurement of DC gain accuracy............. ................................ ............... 2-68
Figure 2-12: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-69
Figure 2-13: Measurement of offset accuracy.......................... .................................. ... 2-72
Figure 2-14: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-74
Figure 2-15: Measurement of analog bandwidth .. ................................ ......................... 2-76
Figure 2-16: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-79
Figure 2-17: Measurement of analog bandwidth .. ................................ ......................... 2-83
Figure 2-18: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-84
Figure 2-19: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-86
Figure 2-20: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-87
Figure 2-21: Final test hookup ................................................................................ 2-88
Figure 2-22: Delta time accuracy test hookup............................................................... 2-89
Figure 2-23: Delta time accuracy test hookup............................................................... 2-93
Figure 2-24: Initial test hookup . . .. ... .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-98
Figure 2-25: Measurement of time accuracy for pulse and glitch triggering .. . .. .. . .. .. . .. .. . .. .. . .. .. . 2-99
Figure 2-26: Initial test hookup .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . . 2-103
Figure 2-27: Measu
Figure 2-28: Initial test hookup .. .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. . 2-111
Figure 2-29: Measurement of trigger out limits...................... ................................ ...... 2-112
Figure 2-30: Initial test hookup .. .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. . 2-113
Figure 2-31: Measurement of probe compensator frequency............................................. 2-114
Figure 2-32: Subsequent test hookup ......... .................................. ............................ 2-115
Figure 2-33: Measurement of probe compensator amplitude. .. . .. .. . .. .. .. . .. .. . .. .. .. . .. .. . . . .. .. . .. .. . 2-115
Figure 2-34: Initial test hookup .. .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. . 2-117
Figure 2-35: Isolated 0 triggering.................. ................................ .......................... 2-119
Figure 2-36: Isolated 1 triggering.................. ................................ .......................... 2-121
Figure 2-37: Initial test hookup .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. ... .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . . 2-122
Figure 2-38: Clock recovery ................................................................................. 2-125
s
rement of trigger sensitivity - 50 MHz results shown . .. .. . .. .. . .. .. . .. .. . .. . .. .. . . 2-104
ii MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Table of Contents
Figure 2-39: Si
Figure 2-40: Equipment setup for maximum amplitude ... .. . .. . .. .. . .. .. . .. . .. .. . .. .. . .. . .. .. . .. .. . .. . .. 2-128
ne wave generator leveling equipment setup ............. ................................ 2-126
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication iii
Table of Contents
List of Tables
Table 1-1: Channel input and vertical specications ............ ................................ ............. 1-1
Table 1 - 2 : H
Table 1-3: Trigger specications.............................................................................. 1-31
Table 1-4: Serial Trigger specications (optional on < 4 GHz models DPO7000 Series).............. 1-38
Table 1-5: Digital acquisition specications (MSO70000 Series) ........................................ 1-40
Table 1-6: Input/output port specications .................................................................. 1-42
Table 1-7: Data storage specications........................................................................ 1-44
Table 1 - 8
Table 1-9: Mechanical specications............ ................................ ............................. 1-45
Table 1-10: Environmental specications .................. ................................ ................. 1-46
Table 2-1: Vertical settings ... .. . .. .. . .. .. . .. ... .. . .. .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. . 2-8
Table 2-2: Test equipment................ ................................ .................................. ... 2-19
Table 2-3: MSO70000, DPO/DSA70000B, DPO/DSA70000, DPO7000 Series Test Record. .. ... .. . 2-22
Table
Table 2-5: Gain accuracy ... ................................ .................................. ................. 2-56
Table 2-6: Offset accuracy............................................ .................................. ....... 2-70
Table 2-7: Analog bandwidth, < 3.5 GHz models .......................................................... 2-75
Table 2-8: Analog bandwidth 4 GHz B models and MSO70000 Series....................... ......... 2-80
Table 2-9: Analog bandwidth 4 GHz non-B models...................................................... 2-80
ble 2-10: Analog bandwidth 3.5 GHz model ............................................................. 2-81
Ta
Table 2-11: Delta time measurement settings .. .. . .. .. . .. . .. .. . .. .. . .. . .. .. . .. .. . .. ... .. . .. .. . .. . .. .. . .. .. . . 2-91
Table 2-12: Delta time measurement settings . . .. .. . .. . .. .. . .. .. . .. .. . .. .. . .. .. . .. . .. .. . .. .. . .. .. . .. . .. .. . .. 2-94
Table 2-13: Trigger settings for 4 GHz models .............................. ............................ 2-107
Table 2-14: Serial pattern data ............................................................................... 2-117
Table 2-15: Word recognizer data ........................................................................... 2-120
Table 2-16: Clock recovery input frequencies and baud rates...................................... ...... 2-124
orizontal and acquisition system specications .............................................. 1-26
: Power source specications ...................................................................... 1-44
2-4: DC Voltage measurement accuracy....................... ................................ ....... 2-51
iv MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
General Safety Summary
General Safet
To Avoid Fire or Personal
Injury
ySummary
Review the fo this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
Use proper power cord. Use only the power cord specied for this product and certied for the country of use.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
llowing safety precautions to avoid injury and prevent damage to
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source. Do not block the power cord; it must remain access ible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Wear eye protection. Wear eye protection if exposure to high-intensity rays or laser radiation exists.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
.
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication v
General Safety Summary
TermsinThisManual
Symbols and Terms on the
Product
These terms may
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER in the marking.
WAR N IN G read the marking.
CAUTIO
The following symbol(s) may appear on the product:
appear in this manual:
dicates an injury hazard immediately accessible as you read
indicates an injury hazard not immediately accessible as you
N indicates a hazard to property including the product.
vi MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Service Safety Summary
Service Safet
y Summary
Only qualifie Safety Summary and the General Safety Summary before performing any service procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay exist in disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
d personnel should perform service procedures. Read this Service
this product. Disconnect power, remove battery (if applicable), and
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication vii
Service Safety Summary
viii MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Specications
Specications
This chapter contains the specications for the instrument. All specications are guaranteed unless labeled "typical." Typical specications are provided for your convenience symbol are checked in this manual. All specications apply to all models unless noted otherwise.
4 GHz models specications apply to DPO70000, DSA70000, DPO70000B, DSA70000B, and MSO70000 Series instruments unless noted otherwise.
To meet specications, the following conditions must be met:
but are not guaranteed. Specications that are marked with the
The instr 18°Cand28°C(64°Fand82°F).
The inst page 1-46.)
The ins specications.(Seepage1-44.)
The in within the specified operating temperature range.
You m 20-minute w arm-up period, and the ambient temperature must not change more than 5 °C without rst repeating the procedure.
ument must have been calibrated in an ambient temperature between
rument must be operating within the environmental limits. (See
trument must be powered from a source that meets the
strument must have been operating continuously for at least 20 minutes
ust perform the Signal Path Compensation procedure after the
Specication Tables
Table 1-1: Channel input and vertical specications
Characteristic Description
Number of channels
Input connector
4 GHz models TekConnect.
< 4 GHz models BNC and VPI probe
Input coupling
4 GHz models DC 50 and GND. GND coupling disconnects the input connector from all channel input
< 4 GHz models DC, AC, or GND. GND coupling approximates ground reference by measuring an unused
4, all identical
circuitry and connects a g round reference to the channel input circuitry.
preamplier input that has been connected to ground. The signal being measured is not disconnected from the channel input load.
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication 1–1
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Input resist
4GHzBmode MSO70000 models
4 GHz non-B models 50 ±1.5% at 25 ºC (77 ºF)
Input impedance, < 4 GHz models
1MΩ , DC coupled 1 MΩ ± 1% in parallel with 13 pF ± 2 pF
50 , DC coupled, typical 50 ±1%
Maximum RMS input voltage
4 GHz non-B models <1 V
4 GHz B models and
MSO70000 models
Maximum Input Voltage, < 4 GHz
els
mod
1M coupled or GND coupled
50 5 V RMS, with peaks ±24V
ance
ls and
-DC coupled, 1 M-AC
100 mV FS to 9
95 mV FS:
50 ±0.75 Ω at 25 ºC (77 ºF) 50 ±1Ω over 10 to 45 ºC (50 to 113 ºF)
1VFSto5VF
S:
50 ±2.2 over 10 to 45 ºC (50 to 113 ºF)
50 ±2%over10to45ºC(50to113ºF)
for <1.0 V/Full Scale settings and
RMS
<5.5 V
for 1.0 V/Full Scale settings
RMS
Full scale is the scale setting times 10.
<1 V
<5.0 V
150
for <1.0 V/Full Scale settings and
RMS
for 1.0 V/Full Scale settings
RMS
V. Derate at 20 dB/decade to 9 V
The maximum input voltage at the BNC, between center conductor and ground is 400 V peak. The RMS voltage is limited to <150 V for arbitrary waveshapes including DC. The maximum
lse width for impulses with peaks over 150 V is 50 μs. Example: At 0 V to 400 V peak,
pu rectangular wave, the duty factor is 14%.
The maximum transient withstand voltage is ± 800 V peak.
above 200 kHz.
rms
1–2 MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Input VSWR, typical
4 GHz B models and MSO70000 models Input Frequency
<2.5 GHz
<6 GHz
<14 GHz
<15 GHz
<20 GHz
4 GHz non-B models Input Frequency
<2.5 GHz
<6 GHz
<14 GHz
< 4 GHz models fin<3.5 GHz
Number of digitized bits
Digitizer nonlinearity, typical
Sensitivity range
50 Ω , ≥ 4 GHz B models and MSO70000 models
50 Ω , ≥ 4 GHz non-B models 100 mV/Full Scale to 10 V/Full Scale, in a 1-2-5 sequence. Below 100 mV/Full Scale, Full
50 , < 4 GHz models 1 mV/div to 1 V/division, in a 1-2-5 sequence
1MΩ , < 4 GHz models 1 mV/div to 10 V/division, in a 1-2-5 sequence
DC gain accuracy, sample o r average acquisition mode, 4 GHz models
DC gain accuracy, sample or average acquisition mode, < 4 GHz models
VSWR < 1 V/Full Scale VSWR 1V/FullScale
1.25
1.5
2.1
2.5
3.2
1.2
1.2
1.5
1.5
1.9
Measured with a TekConnect 2.92 mm adapter
VSWR < 1 V/Full Scale VSWR 1V/FullScale
1.25
1.5
2.1
2.5
1.2
1.2
1.5
1.5
Measured with a TekConnect 2.92 mm adapter
3.0
<2.5 GHz
f
in
<2 GHz
f
in
f
<1 GHz
in
2.0
1.5
1.2
8bits
< 1 digitization level (DL), differential; 1 DL integral, independently based
100 mV/Full Scale to 5 V/Full Scale. Below 100 mV/Full Scale, Full Scale (FS) is software zoom.
Scale (FS) is software zoom.
Fine adjustment available with 1% resolution
Fine adjustment available with 1% resolution
±2%
± 1.0% with 0 V net offset Add 0.5% for ranges <2 mV/div Add 1.5% x | net offset/Max offset | for ranges <5 mV/div Add 0.5% x | net offset/Max offset | for ranges 5 mV/div Add 0.5% for ranges 1 V/div in 1 Mcoupling and with offset >10 V
Specications
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication 1–3
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
DC voltage me accuracy, 4 GHz B models and MSO70000 models
Average acquisition mode ( 16 averages)
Delta voltage measurement between of 16 waveforms acquired under the same setup and ambient
DC voltage measurement accuracy 4 GHz non-B models
Average acquisition mode ( 16 averages)
Delta voltage measurement between any two averages of 16 waveforms acquired under the same setup and ambient conditions
asurement
any two averages
conditions
Measurement type
100 mV/FS to 995 mV/FS ±[(2% | reading - net offset | ) +
1 V/FS to 5 V/FS ±[(2% | reading - net offset | ) +
100 mV/F
Measurement type
100 mV/FS to 995 mV/FS ±[(2% | reading - net offset | ) +
1 V/FS to 10 V/FS ±[(2% | reading - net offset | )
100 mV/FS to 10 V/FS ±[(2% | reading | ) + 0.016 FS]
Sto5V/FS
DC accuracy (in volts)
0.35% | net
0.014 FS]
0.35% | net offset | +7.5 mV +
0.014 FS]
±[(2% | r
DC accu
0.35% | net offset | +1.5 mV +
0.014 FS]
+ 0.35% | net offset | +15 mV +
0.014 FS]
offset | +1.5 mV +
eading | ) + 0.016 FS]
racy (in volts)
1–4 MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
DC voltage me
accuracy, < 4 GHz models
Average acq ( 16 averages)
Delta voltage measurement between an of 16 waveforms acquired under the same setup and ambient c
Sample a typical
Delta vo between any two samples acquired under the same setup a typical
Position range ± 5 divisions
asurement
uisition mode
y two averages
onditions
cquisition mode,
ltage measurement
nd ambient conditions,
DC accuracy (in volts)
±(DC Gain Accuracy | reading -(offset - position) | + offset accuracy + 0.1 division)
For <5 mV/division settings: ±(DC Gain A
±(DC Gain A
For <5 mV/division settings: ±(DC Gain Accuracy | reading | + 0.05 division)
±(DC Gai
±(DC Gain Accuracy | reading | + 0.15 division + 1.2 mV)
Convert offset, position and the constant offset term to volts by multiplying by the appropriate volts/div.
Speci and Cycle RMS measurements. Delta volts specication applies to subtractive c alculations involving two of these measurements. Delta volts specication applies to the Positive Overs
ccuracy | reading -(offset - position) | + offset accuracy + 0.1 division)
ccuracy | reading | + 0.05 division)
n Accuracy | reading -(offset - position) | + offset accuracy + 0.15 division + 0.6 mV)
cation applies to any sample and to the High, Low, Max, Min, M ean, Cycle Mean, RMS,
hoot, Negative Overshoot, Pk-Pk, and amplitude measurements.
Specications
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication 1–5
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Offset range
4 GHz B models and MSO70000 models Offset is reduced to allow for position control according to the following formulas: 10 mV/div to 99.5 mV/div Offset range = ±(0.500 V ­(V/div setting × 5)) 100 mV/div to 500 mV/div Offset range = ±(2.5 V - (V/div setting * 5))
4 GHz non-B models Offset is reduced to allow for position control according to the following formulas: 10 mV/div to 99.5 mV/div Offset range = ±(0.500 V ­(V/div setting × 5)) 100 mV/div to 1 V/div Offset range = ±(0.500 V ­(V/div setting * 0.5))
SCALE range Offset range
10 mV/div
12 mV/div
14 mV/div
16 mV/div
18 mV/div
19.9 mV/div
20 mV/div
30 mV/div
40 mV/div
49.8 mV/div
50 mV/div
60 mV/div
70 mV/div
80 mV/div
90 mV/div
99.5 mV/div
100 mV/div
120 mV/div
140 mV/div
160 mV/div
180 mV/div
200 mV/div
248 mV/div
250 mV/div
300 mV/div
400 mV/div
500 mV/div
SCALE range Offset range
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1V/div
±0.450 V
±0.440 V
±0.430 V
±0.420 V
±0.410 V
±0.4005 V
±0.400 V
±0.350 V
±0.300 V
±0.251 V
±0.250 V
±0.200 V
±0.150 V
±0.100 V
±0.050 V
±0.0025 V
±2.00 V
±1.90 V
±1.80 V
±1.70 V
±1.60 V
±1.50 V
±1.260 V
±1.250 V
±1.00 V
±0.50 V
±0.00 V
±0.450 V
±0.400 V
±0.250 V
±4.50 V
±4.00 V
±2.50 V
±0.00 V
1–6 MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
< 4 GHz models, 50 coupling
< 4 GHz models, 1MΩ coupling
Offset accuracy Net offset is the nominal voltage that must be applied to the channel to bring the trace to
4 GHz B models and
O70000 models
MS
4 GHz non-B models
< 4 GHz models
SCALE range Offset range
1 mV/div to 50 mV/div
50.5 mV/div to 99.5 mV/div
100 mV/div to 500 mV/div
505 mV/div to 1 V/div
SCALE range Offset range
1 mV/div to 50 mV/div
50.5 mV/div to 99.5 mV/div
100 mV/div to 500 mV/div
505 mV/div to 995 mV/div
1.0 V/div to 5 V/div
5.05 V/div to 10 V/div
center screen. Net offset = offset - ( position × volts/division). Offset accuracy is the accuracy of this voltage level.
SCALE range Offset accuracy
mV/div to 99.5 mV/div
10
100 mV/div to 0.5 V/div ±(0.35% | net offset | + 7.5 mV +
SCALE range Offset accuracy
10 mV/div to 99.5 mV/div ±(0.35% | net offset | + 1.5 mV +
100 mV/div to 1 V/div ±(0.35% | net offset | + 15 mV +
SCALE range Offset accuracy
1 mV/div to 9.95 mV/div ±(0.2% | net offset | + 1.5 mV +
10 mV/div to 99.5 mV/div ±(0.35% | net offset | + 1.5 mV +
100 mV/div to 1 V/div ±(0.35% | net offset | + 15 mV +
1.01 V/div to 10 V/div ±(0.25% | net offset | + 150 mV +
±1.0 V
±1.5 V - 10 divisions
±10 V
±15V-10divisions
±1 V
±1.5 V - 10 divisions
±10 V
±15V-10divisions
±100 V
±150 V - 10 divisions
0.35% | net offset | + 1.5 mV +
±( 1% x Full Scale)
1% x Full Scale)
1% x Full Scale)
1% x Full Scale)
0.1 div V/div setting)
0.1 div V/div setting)
0.1 div V/div setting)
0.1 div V/div setting)
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication 1–7
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Analog bandw
idth
Instrument
DPO72004B
and MSO72004
20 GHz BWE
100 mV full scale (10 mV/div)
20 mV, 50 mV
, 100 mV, and
250 mV/div
DPO71604B and MSO71604 No DSP
DPO71254B and MSO71254 No DSP
DPO708
04B and MSO70804
DPO70604B and MSO70604 No DSP
DPO70404B and MSO70404
DC 50 coupling, Full bandwidth, TCA-292mm or T CA-N adapter
BW Settings
1
No DSP
Enhanced BW
Enhanced B
W
Enhanced BW
Enhance
dBW
No DSP
Enhanced BW
Enhanced BW
No DSP
nced BW
Enha
Temperature Variation
uency
Freq
11 GHz
DC -
12.5 GHz
16 GHz
GHz
18
20 GHz
Enhanced bandwidth of 12.5 GHz, 16 GHz and 20 GHz is only available at 50 GS/s.
Bandwidth
>16 GHz, ty
pical
>18 GHz
>20 GHz
>16 GHz, typical
>16 GHz
>12.5 GHz
>12.5 GH
z
>8 GHz
>8 GHz
>6 GHz
>6 GHz
>4 GHz
z
>4 GH
dB/ °C)
TC, (
0
02
-0.
-0.04
-0.09
.09
-0
Temp Constr
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<45 °C
<45 °C
<45 °C
<45 °C
<45 °C
C
<45 °
aint
3
1–8 MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Instrument
DPO72004 20 GHz BWE
100mVand1Vfullscale (10 mv/div and 0.1 V/div)
20 mV, 50 mV, 200 mV, and 500 mV/div
18 GHz BWE
DPO71604 No DSP
DPO71254 No DSP
DPO70804 No DSP
DPO70604 No DSP
DPO70404
Analog bandwidth with P7313 active probe, typical
BW Settings
No DSP
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
No DSP
Enhanced BW
Temperature Variation Frequency
DC - 11 GHz
12.5 GHz
16 GHz
18 GHz
20 GHz
Enhanced bandwidth of 12.5 GHz, 16 GHz and 20 GHz is only available at 50 GS/s.
DPO72004/B, DPO71604/B, and MSO72004:
DPO71254/B and MSO71254:
DPO70804/B and MSO70804:
DPO70604/B and MSO70604:
DPO70404/B and MSO70404:
2
Bandwidth
>16 GHz, typical
>18 GHz
>20 GHz
>18 GHz
>16 GHz, typical
>16 GHz
>12.5 GHz
>12.5 GHz
>8 GHz
>8 GHz
>6 GHz
>6 GHz
>4 GHz
>4 GHz
TC, (dB/ °C)
0
-0.02
-0.04
-0.08
-0.08
DC >12.5 GHz
DC >11 GHz
DC >8 GHz
DC >6 GHz
DC >4 GHz
Temp Constraint
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<40 °C
<40 °C
<45 °C
<45 °C
<45 °C
<45 °C
<45 °C
<45 °C
Specications
3
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication 1–9
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Analog bandw
DPO7354 10 mV/div to 1 V/div DC to 2.5 GHz DC to
DPO7254 10 mV/div to 1 V/div DC to 2.5 GHz
Analog bandwidth, 1 M,DC coupled, typical, < 4 GHz models
Analog bandwidth with TAP2500 VPI probe, typical, DPO7254
idth
DC 50 coupling, Full bandwidth, operating ambient of 30 °C (86 °F), derated by 1% for each °C above 30 °C (86 °F)
SCALE range
5 mV/div to 9.9 mV/div DC to 2.0 GHz
2 mV/div to 4.99 mV/div DC to 500 MHz,
1 mV/div to 1.99 mV/div DC to 200 MHz,
5 mV/div to 9.9 mV/div DC to 2.0 GHz
2 mV/div to 4.99 mV/div DC to 500 MHz,
1 mV/div to 1.99 mV/div DC to 200 MHz,
5 mV/div to 1 V/div DC to 1.0 GHzDPO7104
2 mV/div to 4.99 mV/div DC to 500 MHz,
1 mV/div to 1.99 mV/div DC to 200 MHz,
5 mV/div to 1 V/div DC to 500 MHzDPO7054
2 mV/div to 4.99 mV/div DC to 400 MHz,
1 mV/div to 1.99 mV/div DC to 200 MHz,
Full bandwidth, operating ambient of 30 °C (86 °F), derated by 1% for each °C above 30 °C (86 °F)
SCALE range
5 mV/div to 10 V /div DC to 500 MHz
2 mV/div to 4.98 mV/div DC to 350 MHz
1 mV/div to 1.99 mV/div DC to 175 MHz
Full bandwidth, operating ambient of 30 °C (86 °F), derated by 1% for each °C above 30 °C (86 °F)
SCALE range
100 mV/div DC to 2.5 GHz
50 mV/div to 99.58 mV/div DC to 2.0 GHz
20 mV/div to 49.8 mV/div DC to 500 MHz
10 mV/div to 19.9 mV/div DC to 200 MHz
Bandwidth
BWE off
typical
typical
typical
typical
typical
typical
typical
typical
Bandwidth
Bandwidth
BWE on
3.5 GHz
5
1–10 MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Analog bandwidth with X10 passive probe, typical, < 4 GHz models
Analog bandwidth selections, < 4 GHz models
Lower frequency limit, AC coupled, < 4 GHz models
Upper frequency limit, 250 MHz bandwidth limited, < 4 GHz models
Upper frequency limit, 20 MHz bandwidth limited, < 4 GHz models
Full bandwidth, operating ambient of 30 °C (86 °F), derated by 1% for each °C above 30 °C (86 °F)
SCALE range
5 mV/div to 10 V/div DC to 500 MHz
2 mV/div to 4.98 mV/div DC to 300 MHz
1 mV/div to 1.99 mV/div DC to 175 MHz
20 MHz, 250 MHz and Full
10 Hz when 1 M, AC coupled. The limit is reduced by a factor of 10 when 10X, passive probes are used.
250 MHz
20 MHz
Specications
Bandwidth
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication 1–11
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Calculated rise time, 50 ,typical
DPO72004B and MSO72004
DPO71604B and MSO71604
DPO71254B and MSO71254
DPO70804B and MSO70804
DPO70604B and MSO70604
DPO70404B and MSO70404
DPO72004
DPO71604
DPO71254
DPO70804
DPO70604
DPO70404
DPO7354
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div, BWE off
10 mV/div - 1 V/div, BWE on
DPO7254
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div
DPO7104
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div
DPO7054
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div
4
DSP On 10%
- 90%
18.0 ps NA NA
24.5ps 30ps 20ps
32.0ps 33ps 23ps
49 ps 34 ps 47 ps 32 ps
65 ps 45 ps 62 ps 43 ps
98 ps 68 ps 93 ps 65 ps
22.5 ps NA NA
27.5ps 30ps 20ps
34.3ps 33ps 23ps
47 ps 33 ps 47 ps 32 ps
62 ps 43 ps 62 ps 43 ps
93 ps 65 ps 93 ps 65 ps
1.2 ns
650 ps
180 ps
145 ps
115 ps
1.2 ns
650 ps
180 ps
160 ps
1.2 ns
580 ps
300 ps
300 ps
1.2 ns
680 ps
460 ps
460 ps
20% ­80%
DSP Off 10% ­90%
20% ­80%
1–12 MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
Step response settling time, typical, 4 GHz B models and MSO70000 models
Instrument
DPO72004B and MSO72004
DPO71604B and MSO71604
DPO71254B and MSO71254
DPO70804B and MSO70804
DPO70604B and MSO70604
DPO70404B and MSO70404
The time by which the step response enters and stays below the indicated % error. Step transition occurs at the 50% amplitude point of the step leading edge.
DSP off
Gain s etting (FS)
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
± Step amplitude Settling
100% FS,
50% FS overdrive
100% FS,50% FS overdrive
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
Error Amount
<9%
<5.5%
<2.5%
< 0.15%
<9%
<5.5%
<2.5%
<0.15%
<9%
<5.5%
<2.5%
< 0.15%
<6%
<3%
<2.5%
<0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
Specications
Time
150 ps
400 ps
3ns
1ms
150 ps
400 ps
3ns
1ms
150 ps
400 ps
3ns
1ms
150 ps
600 ps
3ns
1ms
200 ps
800 ps
3ns
1ms
300 ps
1.2 ns
3ns
1ms
MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication 1–13
Specications
Table 1-1: C hannel input and vertical specications (cont.)
Characteristic Description
DSP on
Instrument
DPO72004B and MSO72004
DPO71604B and MSO71604
DPO71254B and MSO71254
DPO70804B and MSO70804
DPO70604B and MSO70604
DPO70404B and MSO70404
Gain setting (FS)
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
± Step amplitude Settling
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
100% FS,
50% FS overdrive
Error Amount
<6%
<3%
<2.5%
<0.15%
<6%
<3%
<2.5%
<0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
Time
100 ps
400 ps
3ns
1ms
100 ps
400 ps
3ns
1ms
100 ps
400 ps
3ns
1ms
150 ps
600 ps
3ns
1ms
200 ps
800 ps
3ns
1ms
300 ps
1.2 ns
3ns
1ms
1–14 MSO70000, DSA/DPO70000/B, DPO7000 Series Specications and Performance Verication
Loading...