MSO70000 Series Mixed Signal Oscilloscopes
DSA70000/B Series Digital Signal Analyzers
DPO70000/B Series Digital Phosphor Oscilloscopes
DPO7000 Series Digital Phosphor Oscilloscopes
ZZZ
Specifications and Performance Verification
Technical Reference
*P077006304*
077-0063-04
xx
MSO70000 Series Mixed Signal Oscilloscopes
DSA70000/B Series Digital Signal Analyzers
DPO70000/B Series Digital Phosphor Oscilloscopes
DPO7000 Series Digital Phosphor Oscilloscopes
ZZZ
Specifications and Performance Verification
Technical Reference
This document applies to firmware version 5.2.0 and above.
Warning
The servicing instructions are for use by qualified personnel
only. To avoid per
unless you are qualified to do so. Refer to all safety summaries
prior to performi
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Pinpoint, TekLink, and RT-Eye are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O . Bo x 50 0
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
World wide , vi s it www.tektronix.com to find contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1)
year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its
option, either will repair the defective product without charge for parts and labor, or will provide a replacement
in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty
work may be n
the property of Tektronix.
ew or reconditioned to like new performance. All replaced parts, modules and products become
In order to o
the warranty period a nd make suitable arrangements for the performance of service. Customer shall be responsible
for packaging and shipping the defective product to the service center designated by Tektronix, with shipping
charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within
the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping
charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
result
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage
or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or
integrated with other products when the effect of such modification or integration increases the time or difficulty
of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
TEK
AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
btain service under this warranty, Customer must notify Tektronix of the defect before the expiration of
ing from attempts by personnel other than Tektronix representatives to install, repair or service the product;
Table of Contents
General Safety Summary ..........................................................................................v
Service Safety Summary........................ ................................ ................................ .vii
Table 2-14: Serial pattern data ...............................................................................2-117
Table 2-15: Word recognizer data ...........................................................................2-120
Table 2-16: Clock recovery input frequencies and baud rates...................................... ......2-124
orizontal and acquisition system specifications ..............................................1-26
: Power source specifications ......................................................................1-44
2-4: DC Voltage measurement accuracy....................... ................................ .......2-51
ivMSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
General Safety Summary
General Safet
To Avoid Fire or Personal
Injury
ySummary
Review the fo
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions related to operating the system.
Use proper power cord. Use only the power cord specified for this product and
certified for the country of use.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
llowing safety precautions to avoid injury and prevent damage to
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source.
Do not block the power cord; it must remain access ible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Wear eye protection. Wear eye protection if exposure to high-intensity rays or
laser radiation exists.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
.
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verificationv
General Safety Summary
TermsinThisManual
Symbols and Terms on the
Product
These terms may
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER in
the marking.
WAR N IN G
read the marking.
CAUTIO
The following symbol(s) may appear on the product:
appear in this manual:
dicates an injury hazard immediately accessible as you read
indicates an injury hazard not immediately accessible as you
N indicates a hazard to property including the product.
viMSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
Service Safety Summary
Service Safet
y Summary
Only qualifieSafety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
d personnel should perform service procedures. Read this Service
this product. Disconnect power, remove battery (if applicable), and
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verificationvii
Service Safety Summary
viiiMSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
Specifications
Specifications
This chapter contains the specifications for the instrument. All specifications are
guaranteed unless labeled "typical." Typical specifications are provided for your
convenience
symbol are checked in this manual. All specifications apply to all models unless
noted otherwise.
≥ 4 GHz models specifications apply to DPO70000, DSA70000, DPO70000B,
DSA70000B, and MSO70000 Series instruments unless noted otherwise.
To meet specifications, the following conditions must be met:
but are not guaranteed. Specifications that are marked with the
The instr
18°Cand28°C(64°Fand82°F).
The inst
page 1-46.)
The ins
specifications.(Seepage1-44.)
The in
within the specified operating temperature range.
You m
20-minute w arm-up period, and the ambient temperature must not change
more than 5 °C without first repeating the procedure.
ument must have been calibrated in an ambient temperature between
rument must be operating within the environmental limits. (See
trument must be powered from a source that meets the
strument must have been operating continuously for at least 20 minutes
ust perform the Signal Path Compensation procedure after the
Specification Tables
Table 1-1: Channel input and vertical specifications
CharacteristicDescription
Number of channels
Input connector
≥ 4 GHz modelsTekConnect.
< 4 GHz modelsBNC and VPI probe
Input coupling
≥ 4 GHz modelsDC 50 Ω and GND. GND coupling disconnects the input connector from all channel input
< 4 GHz modelsDC, AC, or GND. GND coupling approximates ground reference by measuring an unused
4, all identical
circuitry and connects a g round reference to the channel input circuitry.
preamplifier input that has been connected to ground. The signal being measured is not
disconnected from the channel input load.
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification1–1
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
Input resist
≥ 4GHzBmode
MSO70000 models
≥ 4 GHz non-B models50 Ω ±1.5% at 25 ºC (77 ºF)
Input impedance, < 4 GHz models
1MΩ , DC coupled1 MΩ ± 1% in parallel with 13 pF ± 2 pF
50 Ω , DC coupled, typical50 Ω ±1%
Maximum RMS input voltage
≥ 4 GHz non-B models<1 V
≥ 4 GHz B models and
MSO70000 models
Maximum Input Voltage, < 4 GHz
els
mod
1MΩ
coupled or GND coupled
50Ω5 V RMS, with peaks ≤ ±24V
ance
ls and
-DC coupled, 1 MΩ -AC
100 mV FS to 9
95 mV FS:
50 Ω ±0.75 Ω at 25 ºC (77 ºF)
50 Ω ±1Ω over 10 to 45 ºC (50 to 113 ºF)
1VFSto5VF
S:
50 Ω ±2.2 Ω over 10 to 45 ºC (50 to 113 ºF)
50 Ω ±2%over10to45ºC(50to113ºF)
for <1.0 V/Full Scale settings and
RMS
<5.5 V
for ≥ 1.0 V/Full Scale settings
RMS
Full scale is the scale setting times 10.
<1 V
<5.0 V
150
for <1.0 V/Full Scale settings and
RMS
for ≥ 1.0 V/Full Scale settings
RMS
V. Derate at 20 dB/decade to 9 V
The maximum input voltage at the BNC, between center conductor and ground is 400 V peak.
The RMS voltage is limited to <150 V for arbitrary waveshapes including DC. The maximum
lse width for impulses with peaks over 150 V is 50 μs. Example: At 0 V to 400 V peak,
pu
rectangular wave, the duty factor is 14%.
The maximum transient withstand voltage is ± 800 V peak.
above 200 kHz.
rms
1–2MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
Input VSWR, typical
≥ 4 GHz B models and
MSO70000 models
Input Frequency
<2.5 GHz
<6 GHz
<14 GHz
<15 GHz
<20 GHz
≥ 4 GHz non-B models
Input Frequency
<2.5 GHz
<6 GHz
<14 GHz
< 4 GHz modelsfin<3.5 GHz
Number of digitized bits
Digitizer nonlinearity, typical
Sensitivity range
50 Ω , ≥ 4 GHz B models and
MSO70000 models
50 Ω , ≥ 4 GHz non-B models100 mV/Full Scale to 10 V/Full Scale, in a 1-2-5 sequence. Below 100 mV/Full Scale, Full
50 Ω , < 4 GHz models1 mV/div to 1 V/division, in a 1-2-5 sequence
1MΩ , < 4 GHz models1 mV/div to 10 V/division, in a 1-2-5 sequence
DC gain accuracy, sample o r
average acquisition mode,
≥ 4 GHz models
DC gain accuracy, sample or
average acquisition mode,
< 4 GHz models
100 mV/Full Scale to 5 V/Full Scale. Below 100 mV/Full Scale, Full Scale (FS) is software zoom.
Scale (FS) is software zoom.
Fine adjustment available with ≥1% resolution
Fine adjustment available with ≥1% resolution
±2%
± 1.0% with 0 V net offset
Add 0.5% for ranges <2 mV/div
Add 1.5% x | net offset/Max offset | for ranges <5 mV/div
Add 0.5% x | net offset/Max offset | for ranges ≥ 5 mV/div
Add 0.5% for ranges ≥ 1 V/div in 1 MΩ coupling and with offset >10 V
Specifications
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification1–3
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
DC voltage me
accuracy, ≥ 4 GHz B models and
MSO70000 models
Average acquisition mode
( ≥16 averages)
Delta voltage measurement
between
of ≥16 waveforms acquired
under the same setup and
ambient
DC voltage measurement
accuracy ≥ 4 GHz non-B models
Average acquisition mode
( ≥16 averages)
Delta voltage measurement
between any two averages
of ≥16 waveforms acquired
under the same setup and
ambient conditions
asurement
any two averages
conditions
Measurement type
100 mV/FS to 995 mV/FS±[(2% | reading - net offset | ) +
1 V/FS to 5 V/FS±[(2% | reading - net offset | ) +
100 mV/F
Measurement type
100 mV/FS to 995 mV/FS±[(2% | reading - net offset | ) +
Convert offset, position and the constant offset term to volts by multiplying by the appropriate
volts/div.
Specifi
and Cycle RMS measurements. Delta volts specification applies to subtractive c alculations
involving two of these measurements. Delta volts specification applies to the Positive
Overs
cation applies to any sample and to the High, Low, Max, Min, M ean, Cycle Mean, RMS,
hoot, Negative Overshoot, Pk-Pk, and amplitude measurements.
Specifications
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification1–5
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
Offset range
≥ 4 GHz B models and
MSO70000 models
Offset is reduced to allow for
position control according to
the following formulas:
10 mV/div to 99.5 mV/div
Offset range = ±(0.500 V (V/div setting × 5))
100 mV/div to 500 mV/div
Offset range = ±(2.5 V - (V/div
setting * 5))
≥ 4 GHz non-B models
Offset is reduced to allow for
position control according to
the following formulas:
10 mV/div to 99.5 mV/div
Offset range = ±(0.500 V (V/div setting × 5))
100 mV/div to 1 V/div
Offset range = ±(0.500 V (V/div setting * 0.5))
SCALE rangeOffset range
10 mV/div
12 mV/div
14 mV/div
16 mV/div
18 mV/div
19.9 mV/div
20 mV/div
30 mV/div
40 mV/div
49.8 mV/div
50 mV/div
60 mV/div
70 mV/div
80 mV/div
90 mV/div
99.5 mV/div
100 mV/div
120 mV/div
140 mV/div
160 mV/div
180 mV/div
200 mV/div
248 mV/div
250 mV/div
300 mV/div
400 mV/div
500 mV/div
SCALE rangeOffset range
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1V/div
±0.450 V
±0.440 V
±0.430 V
±0.420 V
±0.410 V
±0.4005 V
±0.400 V
±0.350 V
±0.300 V
±0.251 V
±0.250 V
±0.200 V
±0.150 V
±0.100 V
±0.050 V
±0.0025 V
±2.00 V
±1.90 V
±1.80 V
±1.70 V
±1.60 V
±1.50 V
±1.260 V
±1.250 V
±1.00 V
±0.50 V
±0.00 V
±0.450 V
±0.400 V
±0.250 V
±4.50 V
±4.00 V
±2.50 V
±0.00 V
1–6MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
< 4 GHz models,
50 Ω coupling
< 4 GHz models,
1MΩ coupling
Offset accuracyNet offset is the nominal voltage that must be applied to the channel to bring the trace to
≥ 4 GHz B models and
O70000 models
MS
≥ 4 GHz non-B models
< 4 GHz models
SCALE rangeOffset range
1 mV/div to 50 mV/div
50.5 mV/div to 99.5 mV/div
100 mV/div to 500 mV/div
505 mV/div to 1 V/div
SCALE rangeOffset range
1 mV/div to 50 mV/div
50.5 mV/div to 99.5 mV/div
100 mV/div to 500 mV/div
505 mV/div to 995 mV/div
1.0 V/div to 5 V/div
5.05 V/div to 10 V/div
center screen. Net offset = offset - ( position × volts/division). Offset accuracy is the accuracy
of this voltage level.
SCALE rangeOffset accuracy
mV/div to 99.5 mV/div
10
100 mV/div to 0.5 V/div±(0.35% | net offset | + 7.5 mV +
SCALE rangeOffset accuracy
10 mV/div to 99.5 mV/div±(0.35% | net offset | + 1.5 mV +
100 mV/div to 1 V/div±(0.35% | net offset | + 15 mV +
SCALE rangeOffset accuracy
1 mV/div to 9.95 mV/div±(0.2% | net offset | + 1.5 mV +
10 mV/div to 99.5 mV/div±(0.35% | net offset | + 1.5 mV +
100 mV/div to 1 V/div±(0.35% | net offset | + 15 mV +
1.01 V/div to 10 V/div±(0.25% | net offset | + 150 mV +
±1.0 V
±1.5 V - 10 divisions
±10 V
±15V-10divisions
±1 V
±1.5 V - 10 divisions
±10 V
±15V-10divisions
±100 V
±150 V - 10 divisions
0.35% | net offset | + 1.5 mV +
±(
1% x Full Scale)
1% x Full Scale)
1% x Full Scale)
1% x Full Scale)
0.1 div V/div setting)
0.1 div V/div setting)
0.1 div V/div setting)
0.1 div V/div setting)
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification1–7
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
Analog bandw
idth
Instrument
DPO72004B
and MSO72004
20 GHz BWE
100 mV full scale (10 mV/div)
20 mV, 50 mV
, 100 mV, and
250 mV/div
DPO71604B and MSO71604No DSP
DPO71254B and MSO71254No DSP
DPO708
04B and MSO70804
DPO70604B and MSO70604No DSP
DPO70404B and MSO70404
DC 50 Ω coupling, Full bandwidth, TCA-292mm or T CA-N adapter
BW Settings
1
No DSP
Enhanced BW
Enhanced B
W
Enhanced BW
Enhance
dBW
No DSP
Enhanced BW
Enhanced BW
No DSP
nced BW
Enha
Temperature Variation
uency
Freq
11 GHz
DC -
12.5 GHz
16 GHz
GHz
18
20 GHz
Enhanced bandwidth of 12.5 GHz, 16 GHz and 20 GHz is only available at 50 GS/s.
Bandwidth
>16 GHz, ty
pical
>18 GHz
>20 GHz
>16 GHz, typical
>16 GHz
>12.5 GHz
>12.5 GH
z
>8 GHz
>8 GHz
>6 GHz
>6 GHz
>4 GHz
z
>4 GH
dB/ °C)
TC, (
0
02
-0.
-0.04
-0.09
.09
-0
Temp Constr
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<45 °C
<45 °C
<45 °C
<45 °C
<45 °C
C
<45 °
aint
3
1–8MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
Instrument
DPO72004
20 GHz BWE
100mVand1Vfullscale
(10 mv/div and 0.1 V/div)
20 mV, 50 mV, 200 mV, and
500 mV/div
18 GHz BWE
DPO71604No DSP
DPO71254No DSP
DPO70804No DSP
DPO70604No DSP
DPO70404
Analog bandwidth with P7313
active probe, typical
BW Settings
No DSP
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
Enhanced BW
No DSP
Enhanced BW
Temperature Variation
Frequency
DC - 11 GHz
12.5 GHz
16 GHz
18 GHz
20 GHz
Enhanced bandwidth of 12.5 GHz, 16 GHz and 20 GHz is only available at 50 GS/s.
DPO72004/B,
DPO71604/B, and
MSO72004:
DPO71254/B and
MSO71254:
DPO70804/B and
MSO70804:
DPO70604/B and
MSO70604:
DPO70404/B and
MSO70404:
2
Bandwidth
>16 GHz, typical
>18 GHz
>20 GHz
>18 GHz
>16 GHz, typical
>16 GHz
>12.5 GHz
>12.5 GHz
>8 GHz
>8 GHz
>6 GHz
>6 GHz
>4 GHz
>4 GHz
TC, (dB/ °C)
0
-0.02
-0.04
-0.08
-0.08
DC >12.5 GHz
DC >11 GHz
DC >8 GHz
DC >6 GHz
DC >4 GHz
Temp Constraint
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<30 °C
<40 °C
<40 °C
<45 °C
<45 °C
<45 °C
<45 °C
<45 °C
<45 °C
Specifications
3
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification1–9
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
Full bandwidth, operating ambient of ≤ 30 °C (86 °F), derated by 1% for each °C above
30 °C (86 °F)
SCALE range
5 mV/div to 10 V/divDC to 500 MHz
2 mV/div to 4.98 mV/divDC to 300 MHz
1 mV/div to 1.99 mV/divDC to 175 MHz
20 MHz, 250 MHz and Full
10 Hz when 1 M, AC coupled. The limit is reduced by a factor of 10 when 10X, passive probes
are used.
250 MHz
20 MHz
Specifications
Bandwidth
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification1–11
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
Calculated rise time, 50 Ω,typical
DPO72004B and MSO72004
DPO71604B and MSO71604
DPO71254B and MSO71254
DPO70804B and MSO70804
DPO70604B and MSO70604
DPO70404B and MSO70404
DPO72004
DPO71604
DPO71254
DPO70804
DPO70604
DPO70404
DPO7354
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div, BWE
off
10 mV/div - 1 V/div, BWE
on
DPO7254
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div
DPO7104
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div
DPO7054
1 mV/div - 1.99 mV/div
2 mV/div - 4.99 mV/div
5 mV/div - 9.9 mV/div
10 mV/div - 1 V/div
4
DSP On 10%
- 90%
18.0 psNANA
24.5ps30ps20ps
32.0ps33ps23ps
49 ps34 ps47 ps32 ps
65 ps45 ps62 ps43 ps
98 ps68 ps93 ps65 ps
22.5 psNANA
27.5ps30ps20ps
34.3ps33ps23ps
47 ps33 ps47 ps32 ps
62 ps43 ps62 ps43 ps
93 ps65 ps93 ps65 ps
1.2 ns
650 ps
180 ps
145 ps
115 ps
1.2 ns
650 ps
180 ps
160 ps
1.2 ns
580 ps
300 ps
300 ps
1.2 ns
680 ps
460 ps
460 ps
20% 80%
DSP Off
10% 90%
20% 80%
1–12MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
Step response settling time,
typical, ≥ 4 GHz B models and
MSO70000 models
Instrument
DPO72004B and MSO72004
DPO71604B and MSO71604
DPO71254B and MSO71254
DPO70804B and MSO70804
DPO70604B and MSO70604
DPO70404B and MSO70404
The time by which the step response enters and stays below the indicated % error. Step
transition occurs at the 50% amplitude point of the step leading edge.
DSP off
Gain s etting
(FS)
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
± Step amplitudeSettling
≤ 100% FS,
≤50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
Error
Amount
<9%
<5.5%
<2.5%
< 0.15%
<9%
<5.5%
<2.5%
<0.15%
<9%
<5.5%
<2.5%
< 0.15%
<6%
<3%
<2.5%
<0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
Specifications
Time
150 ps
400 ps
3ns
1ms
150 ps
400 ps
3ns
1ms
150 ps
400 ps
3ns
1ms
150 ps
600 ps
3ns
1ms
200 ps
800 ps
3ns
1ms
300 ps
1.2 ns
3ns
1ms
MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification1–13
Specifications
Table 1-1: C hannel input and vertical specifications (cont.)
CharacteristicDescription
DSP on
Instrument
DPO72004B and MSO72004
DPO71604B and MSO71604
DPO71254B and MSO71254
DPO70804B and MSO70804
DPO70604B and MSO70604
DPO70404B and MSO70404
Gain setting
(FS)
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
100 mV - 5 V
± Step amplitudeSettling
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
≤ 100% FS,
≤ 50% FS overdrive
Error
Amount
<6%
<3%
<2.5%
<0.15%
<6%
<3%
<2.5%
<0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
<6%
<3%
<2.5%
< 0.15%
Time
100 ps
400 ps
3ns
1ms
100 ps
400 ps
3ns
1ms
100 ps
400 ps
3ns
1ms
150 ps
600 ps
3ns
1ms
200 ps
800 ps
3ns
1ms
300 ps
1.2 ns
3ns
1ms
1–14MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification