Tektronix DPO70000SX Series, DPO77002SX, DPO75002SX, DPO73304SX, DPO72304SX User Manual

...
DPO70000SX Series Oscilloscopes
User
*P071335707*
071-3357-07
DPO70000SX Series Oscilloscopes
User
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Supports DPO70000SX Series Product Firmware V10.0 and above
www.tek.com
071-3357-07
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
TekScope, TekConnect, and FastAcq are registered trademarks of Tektronix, Inc.
FastFrame, OpenChoice, MyScope, MultiView Zoom, SignalVu, TekExpress, TriMode, TekSecure, TekProbe, TekVPI, TekVISA, UltraSync, and PinPoint are trademarks of Tektronix.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to find contacts in your area.

Table of Contents

Important safety information ............................................................................................................................... vii
General safety summary ............................................................................................................................... vii
Service safety summary ................................................................................................................................ ix
Terms in the manual ...................................................................................................................................... ix
Terms on the product .................................................................................................................................... ix
Symbols on the product ................................................................................................................................. ix
Compliance Information ...................................................................................................................................... xi
EMC compliance ........................................................................................................................................... xi
Safety compliance ........................................................................................................................................ xii
Environmental compliance ........................................................................................................................... xiii
Preface ............................................................................................................................................................... xv
Install your instrument
Standard accessories ..................................................................................................................................... 1
Operating requirements ................................................................................................................................. 2
Power supply requirements ............................................................................................................................ 3
Preventing instrument damage ...................................................................................................................... 3
Preventing ESD ........................................................................................................................................ 3
Observe maximum input voltage .............................................................................................................. 5
Selecting the proper attenuator ................................................................................................................ 5
Connector cleaning .................................................................................................................................. 5
Proper connection technique .................................................................................................................... 6
Power on the instrument ................................................................................................................................ 9
Power off the instrument .............................................................................................................................. 10
Multi-instrument configuration ...................................................................................................................... 10
Instrument stacking ................................................................................................................................ 10
Before startup ......................................................................................................................................... 13
UltraSync™ bus cable ............................................................................................................................ 13
UltraSync™ bus cable connection order ................................................................................................ 13
Master and extension connection order ................................................................................................. 15
Multi-instrument power on ...................................................................................................................... 16
Switching between multi-instrument modes ........................................................................................... 20
ATI versus TekConnect channels .......................................................................................................... 21
Multi-instrument status displays ............................................................................................................. 22
Available features ................................................................................................................................... 22
DPO7AFP Auxiliary Front Panel (optional) .................................................................................................. 23
DPO70000SX Series User
i
Table of Contents
Inspect the instrument .................................................................................................................................. 23
Verify internal diagnostics pass .............................................................................................................. 23
Activating Windows 10 ........................................................................................................................... 24
Windows interface guidelines ................................................................................................................. 25
Signal path compensation ............................................................................................................................ 26
Connecting to a network .............................................................................................................................. 29
Adding a second monitor ............................................................................................................................. 30
Activating Windows 10 ................................................................................................................................. 30
Restoring instrument operating system and product software ..................................................................... 31
Operating system restore ............................................................................................................................. 31
Internal recovery utility ................................................................................................................................. 31
Product software installation ........................................................................................................................ 32
Getting acquainted with your instrument
Front-panel connectors ................................................................................................................................ 33
Rear-panel connectors ................................................................................................................................. 35
Interface and display .................................................................................................................................... 36
Control panel ................................................................................................................................................ 38
Accessing online help .................................................................................................................................. 39
Accessing menus and control windows ....................................................................................................... 40
Inspect your instrument
Verify internal diagnostics pass .................................................................................................................... 41
Acquisition
Signal path compensation ............................................................................................................................ 43
Setting up analog signal input ...................................................................................................................... 46
Using default setup ...................................................................................................................................... 48
Using autoset ............................................................................................................................................... 49
Probe compensation and deskew ................................................................................................................ 50
Deskew Tool ................................................................................................................................................ 50
Acquisition concepts .................................................................................................................................... 55
Acquisition hardware .............................................................................................................................. 55
Sampling process ................................................................................................................................... 56
Real-Time sampling ............................................................................................................................... 56
Interpolated Real-Time sampling ........................................................................................................... 56
Equivalent-Time sampling ...................................................................................................................... 56
Waveform record .................................................................................................................................... 57
ii DPO70000SX Series User
Table of Contents
Interpolation ............................................................................................................................................ 57
How the acquisition modes work ............................................................................................................ 57
Enable enhanced effective number of bits ................................................................................................... 58
Changing the acquisition mode .................................................................................................................... 59
Starting and stopping an acquisition ............................................................................................................ 60
Selecting the horizontal mode ...................................................................................................................... 61
Using FastAcq .............................................................................................................................................. 63
Using DSP enhanced bandwidth ................................................................................................................. 64
Setting the termination voltage ..................................................................................................................... 66
Using roll mode ............................................................................................................................................ 67
Setting up a bus ........................................................................................................................................... 68
Set up a serial bus ....................................................................................................................................... 70
Set up a parallel bus .................................................................................................................................... 71
Set up bus display ........................................................................................................................................ 73
Using FastFrame mode ................................................................................................................................ 74
Using FastFrame frame finder ..................................................................................................................... 76
Pinpoint triggers
Triggering concepts ...................................................................................................................................... 79
Trigger event .......................................................................................................................................... 79
Trigger modes ........................................................................................................................................ 79
Trigger holdoff ........................................................................................................................................ 79
Trigger coupling ...................................................................................................................................... 80
Horizontal position .................................................................................................................................. 80
Slope and level ....................................................................................................................................... 80
Delayed trigger system ........................................................................................................................... 80
Choosing a trigger type ................................................................................................................................ 81
Trigger selections ......................................................................................................................................... 83
Checking trigger status ................................................................................................................................ 84
Using A (Main) and B (Delayed) triggers ..................................................................................................... 85
Trigger on B event .................................................................................................................................. 86
B trigger after delay time ........................................................................................................................ 86
B triggers after arm on A ........................................................................................................................ 86
Arm-on-A then Trigger-on-B (Horizontal delay on) ................................................................................. 87
Vertical setup control window (M Chx tab) ............................................................................................. 88
Triggering with reset ............................................................................................................................... 89
Correcting trigger position ...................................................................................................................... 90
Triggering with B-Event scan ....................................................................................................................... 91
Triggering on a parallel bus .......................................................................................................................... 94
DPO70000SX Series User iii
Table of Contents
Triggering on a serial bus ............................................................................................................................. 96
Triggering using visual triggers (Visual triggering) ....................................................................................... 98
Setting up action on event .......................................................................................................................... 100
Sending E-Mail on trigger ........................................................................................................................... 101
Setting up E-Mail on event ......................................................................................................................... 102
Using horizontal delay ................................................................................................................................ 104
Display a waveform
Setting the display style ............................................................................................................................. 105
Setting the display persistence .................................................................................................................. 106
Setting the display format ........................................................................................................................... 107
Selecting the waveform interpolation ......................................................................................................... 108
Adding screen text ..................................................................................................................................... 109
Setting the graticule style ........................................................................................................................... 110
Setting the trigger level marker .................................................................................................................. 111
Displaying the date and time ...................................................................................................................... 111
Using the color palettes .............................................................................................................................. 112
Setting reference waveform colors ............................................................................................................. 113
Setting math waveform colors .................................................................................................................... 114
Using MultiView zoom ................................................................................................................................ 115
Zooming in multiple areas .......................................................................................................................... 116
Lock and scroll zoomed waveforms ........................................................................................................... 118
Hide waveforms in the zoomed window ..................................................................................................... 119
Searching and marking waveforms ............................................................................................................ 119
To manually set and clear (delete) marks: ................................................................................................. 120
To automatically set and clear (delete) search marks ................................................................................ 122
Using a visual search ................................................................................................................................. 127
Analyzing waveforms
Taking automatic measurements ............................................................................................................... 129
Automated measurement selections .......................................................................................................... 130
Customizing an automatic measurement ................................................................................................... 133
Gating ................................................................................................................................................... 134
Statistics ............................................................................................................................................... 134
Snapshot .............................................................................................................................................. 135
Annotate measurements ...................................................................................................................... 135
Reference levels ................................................................................................................................... 136
Taking cursor measurements ..................................................................................................................... 138
iv DPO70000SX Series User
Table of Contents
Setting up a histogram ............................................................................................................................... 140
Using math waveforms ............................................................................................................................... 142
Using spectral analysis .............................................................................................................................. 145
Using the error detector ............................................................................................................................. 148
Using mask testing ..................................................................................................................................... 152
Using limit testing ....................................................................................................................................... 155
MyScope
Creating a new MyScope control window .................................................................................................. 157
Using MyScope control windows ............................................................................................................... 161
Saving and recalling information
Saving screen captures .............................................................................................................................. 163
Saving waveforms ...................................................................................................................................... 165
Recalling waveforms .................................................................................................................................. 167
Saving instrument setups ........................................................................................................................... 168
Recalling instrument setups ....................................................................................................................... 169
Saving measurements ............................................................................................................................... 170
Saving user masks ..................................................................................................................................... 171
Saving histogram data ............................................................................................................................... 172
Saving timestamps ..................................................................................................................................... 173
Copying your results to the clipboard ......................................................................................................... 174
Printing a hard copy ................................................................................................................................... 176
Run application software
Application examples
Capturing intermittent anomalies ............................................................................................................... 179
Using the extended desktop and OpenChoice architecture for efficient documentation ............................ 182
Triggering on buses ................................................................................................................................... 184
Specifications
Vertical system analog channels ................................................................................................................ 187
Horizontal and acquisition system .............................................................................................................. 191
Trigger specifications ................................................................................................................................. 192
Input-output port specifications .................................................................................................................. 194
DPO70000SX Series User v
Table of Contents
Power source specification ........................................................................................................................ 194
Mechanical specifications .......................................................................................................................... 195
Environmental specifications ...................................................................................................................... 195
Appendix A, Maintenance
Maintenance ............................................................................................................................................... 197
Cleaning ..................................................................................................................................................... 197
Exterior cleaning ........................................................................................................................................ 197
Adjustment interval ..................................................................................................................................... 198
Adjustment ................................................................................................................................................. 198
Flat panel display cleaning ......................................................................................................................... 198
Returning the instrument for service .......................................................................................................... 199
TekScope recovery report utility ................................................................................................................. 199
Replaceable parts ...................................................................................................................................... 201
Parts ordering information .......................................................................................................................... 202
Appendix B, Version releases
Obtaining the latest advanced analysis application and version releases ................................................. 203
vi DPO70000SX Series User

Important safety information

This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, see the Service safety summary that follows the General safety summary.

General safety summary

Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
To avoid fire or personal injury
Use proper power cord. Use only the power cord specified for this product and certified for the country of use. Do not use the
provided power cord for other products.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded. Do not disable the power cord grounding connection.
Power disconnect. The power cord disconnects the product from the power source. See instructions for the location. Do not position the equipment so that it is difficult to operate the power cord; it must remain accessible to the user at all times to allow for quick disconnection if needed.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source. Use only insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be suitable for the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
The measurement terminals on this product are not rated for connection to mains or Category II, III, or IV circuits.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
DPO70000SX Series User
vii
Important safety information
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualified service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean the product.
Provide proper ventilation. Refer to the installation instructions in the manual for details on installing the product so it has
proper ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Use care when lifting and carrying the product. This product is provided with a handle or handles for lifting and carrying.
WARNING. The product is heavy. To reduce the risk of personal injury or damage to the device get help when lifting or carrying the product.
Use only the Tektronix rackmount hardware specified for this product.
viii DPO70000SX Series User
Important safety information

Service safety summary

The Service safety summary section contains additional information required to safely perform service on the product. Only qualified personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering
first aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains power before removing any covers or panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains dielectric strength after performing a repair.

Terms in the manual

These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.

Terms on the product

These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.

Symbols on the product

When this symbol is marked on the product, be sure to consult the manual to find out the nature of the potential hazards and any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.)
The following symbols may appear on the product:
DPO70000SX Series User ix
Important safety information
x DPO70000SX Series User

Compliance Information

This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument complies.

EMC compliance

EC Declaration of Conformity – EMC
Meets intent of Directive 2014/30/EU for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities:
EN 61326-1, EN 61326-2-1. EMC requirements for electrical equipment for measurement, control, and laboratory use.
CISPR 11. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2. Electrostatic discharge immunity
IEC 61000-4-3. RF electromagnetic field immunity
IEC 61000-4-4. Electrical fast transient / burst immunity
IEC 61000-4-5. Power line surge immunity
IEC 61000-4-6. Conducted RF immunity
IEC 61000-4-11. Voltage dips and interruptions immunity
1 2 3
4
5
5
EN 61000-3-2. AC power line harmonic emissions
EN 61000-3-3. Voltage changes, fluctuations, and flicker
Australia / New Zealand Declaration of Conformity – EMC
Complies with the EMC provision of the Radiocommunications Act per the following standard, in accordance with ACMA:
EN 61326-1 and EN 61326-2-1. Radiated and conducted emissions, Group 1, Class A.
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
If interconnect cables are used, they must be high quality, low-EMI shielded cables.
4
Equipment may not meet the immunity requirements of this standard when test leads and/or test probes are connected due to coupling of electromagnetic interference onto those leads/probes. To minimize the influence of electromagnetic interference, minimize the loop area between the unshielded portions of signal and associated return leads, and keep leads as far away as possible from electromagnetic disturbance sources. Twisting unshielded test leads together is an effective way to reduce loop area. For probes, keep the ground return lead as short as possible and close to the probe body. Some probes have accessory probe tip adapters to accomplish this most effectively. In all cases, observe all safety instructions for the probes or leads used.
5
The performance criterion for when the oscilloscope is subjected to the continuously present electromagnetic phenomenon: 10 mV/division to 1 V/division: ≤0.4 division waveform displacement or ≤0.8 division increase in peak–to–peak noise.
DPO70000SX Series User xi
Compliance Information

Safety compliance

This section lists the safety standards with which the product complies and other safety compliance information.
EU declaration of conformity – low voltage
Compliance was demonstrated to the following specification as listed in the Official Journal of the European Union:
Low Voltage Directive 2014/35/EU.
EN 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
U.S. nationally recognized testing laboratory listing
UL 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
Canadian certification
CAN/CSA-C22.2 No. 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
Additional compliances
IEC 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
Equipment type
Test and measuring equipment.
Safety class
Class 1 – grounded product.
Pollution degree description
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical office/home environment. Temporary condensation occurs only when the product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
xii DPO70000SX Series User
Compliance Information
Pollution degree
Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor, dry location use only.
IP rating
IP20 (as defined in IEC 60529).
Measurement and overvoltage category descriptions
Measurement terminals on this product may be rated for measuring mains voltages from one or more of the following categories (see specific ratings marked on the product and in the manual).
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category IV. For measurements performed at the source of low-voltage installation.
NOTE. Only mains power supply circuits have an overvoltage category rating. Only measurement circuits have a measurement category rating. Other circuits within the product do not have either rating.
Mains overvoltage category rating
Overvoltage Category II (as defined in IEC 61010-1)

Environmental compliance

This section provides information about the environmental impact of the product.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Tektronix Web site (www.tek.com/productrecycling).
Perchlorate materials. This product contains one or more type CR lithium batteries. According to the state of California, CR lithium batteries are classified as perchlorate materials and require special handling. See www.dtsc.ca.gov/hazardouswaste/ perchlorate for additional information.
DPO70000SX Series User xiii
Compliance Information
xiv DPO70000SX Series User

Preface

This manual describes the installation and basic operation of DPO70000SX series instruments. For additional operating information, refer to the instrument help on your instrument. The following instruments are supported by this document:
DPO77002SX
DPO75902SX
DPO75002SX
DPO73304SX
DPO72304SX
DPO71604SX
DPO71304SX
Key features
DPO70000SX Series instruments can help you verify, debug, and characterize electronic designs. Key features include:
70 GHz bandwidth and 200 GS/s on 1 analog channel and 33 GHz bandwidth and 100 GS/s on 2 analog channels, DPO77002SX
59 GHz bandwidth and 200 GS/s on 1 analog channel and 33 GHz bandwidth and 100 GS/s on 2 analog channels, DPO75902SX
50 GHz bandwidth and 200 GS/s on 1 analog channel and 33 GHz bandwidth and 100 GS/s on 2 analog channels, DPO75002SX
33 GHz bandwidth and 100 GS/s on 2 analog channels or 4 channels at 50 GS/s, DPO73304SX
23 GHz bandwidth and 100 GS/s on 2 analog channels or 4 channels at 50 GS/s, DPO72304SX
16 GHz bandwidth and 100 GS/s on 2 analog channels or 4 channels at 50 GS/s, DPO71604SX
13 GHz bandwidth and 100 GS/s on 2 analog channels or 4 channels at 50 GS/s, DPO71304SX
Enhanced Bandwidth capability that, when enabled, applies Digital Signal Processing (DSP) filters that can extend the bandwidth and flatten the passband. Enhanced Bandwidth provides a matched response across enabled channels when they are at maximum sample rate. You can limit the bandwidth down to 500 MHz to optimize the signal to noise ratio. Enhanced bandwidth is extended to the probe tip for some high performance probes and tips.
Record lengths up to 1,000,000,000 samples, depending on model and option
Up to 1.0% DC vertical gain accuracy, depending on model
Combine up to 4 instruments using UltraSync™ interface for increased channel count.
Display user interface on an optional external monitor
Up to 4 analog input channels, depending on model (each with 8-bit resolution when not in Hi-Res mode), auxiliary trigger input and output
Full programmability, with an extensive command set and a message-based interface
PinPoint triggering with flexible A and B trigger events
Edge trigger >20 GHz, Aux trigger > 10 GHz.
Selectable trigger position correction to more accurately place the trigger and reduce jitter
DPO70000SX Series User
xv
Preface
Powerful built-in measurement capability, including histograms, automatic measurements, eye pattern measurements and measurement statistics
Mathematically combine waveforms to create waveforms that support your data-analysis task. Use arbitrary filters in math equations. Use spectral analysis to analyze waveforms in the frequency domain.
Ability to control sample rate and record length separately from horizontal scale
An intuitive, graphical user interface (UI), with online help that is built in and available on screen
Internal, removable disk storage
Documentation
Review the following table to locate more information about this product.
To read about Use these documents
Installation and operation (overviews) User manual.
Operation and user interface Instrument help from the Help menu.
Programmer commands Programmer manual. This manual is available on the Tektronix
Web site (www.tektronix.com/manuals).
Conventions used in this manual
The following icons are used throughout this manual.
Step Front panel
power
Connect power Network PS-2 SVGA USB
xvi DPO70000SX Series User

Install your instrument

Unpack the instrument and check that you received all items listed as Standard Accessories. Recommended accessories, probes, instrument options, and upgrades are listed in the online help. Check the Tektronix Web site (www.tektronix.com) for the most current information.

Standard accessories

Accessory Tektronix part number
User manual -- depends on language option 071-3357-xx
Front protective cover , TekConnect instruments ATI instruments
PCIe Host Port protective plug 200-5344-00
2nd ethernet port plug 200-5389-00
50 Ω term on Fast Edge (2X) 015-1022-01
TCA-292D (5X) (3X on ATI instruments) 090-0044-00
Windows compatible keyboard 119-7275-xx
Windows compatible mouse 119-7054-xx
Static protection wrist strap 006-3415-05
Deskew cable (M2.92 to M2.92) 174-6793-00
Deskew adapter (1.85F to 2.92F), ATI instruments only 103-0483-00
Attenuator 2.92 mm female to 2.92 mm male, 50 Ω,10 DB 011-0221-00
ATI connector saver (1.85mm), ATI instruments only 103-0474-00
ATI protective cap, ATI instruments only 016-2101-00
Torque wrench, ATI instruments only 067-2787-00
Backing wrench, ATI instruments only 003-1942-00
Accessories pouch 016-2045-00
Best Practices manual 071-2989-04
ROHS info 071-2185-04
Calibration certification 001-1179-00
Cal cert envelope 006-8018-01
Power cord – one of the following:
200-5337-00 200-5358-00
DPO70000SX Series User 1
Install your instrument
Accessory Tektronix part number
North America (Option A0) 161-0213-00
Universal Euro (Option A1) 161-0209-00
United Kingdom (Option A2) 161-0210-00
Australia (Option A3) 161-0211-01
Switzerland (Option A5) 161-0212-01
Japan (Option A6) 161-0213-00
China (Option A10) 161-0320-00
India (Option A11) 161-0325-00
Brazil (Option A12) 161-0358-00
No power cord (Option A99)

Operating requirements

WARNING.
To prevent injury and instrument damage, do not operate the instrument while the instrument is resting on its rear feet.
1. Place the instrument on a cart or bench. The instrument should rest on its bottom feet. An optional rack mounting kit is available. Observe the following clearance requirements and dimensions:
Top: 0 in (0 mm)
Left side: 3 in (76 mm)
Right side: 3 in (76 mm)
Rear: 3 in (76 cm) beyond the rear feet to protect any cables connected to the rear panel
Bottom: 0 in (0 mm) standing on feet, flip stands down
CAUTION. To ensure proper cooling, keep the bottom and sides of the instrument clear of obstructions.
To ensure proper cooling, if the instruments are stacked on top of each other, the bottom feet must remain on the instruments.
2. Width: 19.0 inches (483 mm)
3. Height: 6.0 inches (152 mm)
4. Before operating the instrument, verify the ambient temperature: 5 °C to +45 °C (+41 °F to +113 °F)
5. Verify the operating humidity: 8% to 80% relative humidity at up to +32 °C (+90 °F)
5% to 45% relative humidity above +32 °C (+90 °F) up to +45 °C (+113 °F), noncondensing, and is limited by a maximum wet-bulb temperature of +29.4 °C (+85 °F) (derate relative humidity to 32% at +45 °C (+113 °F))
6. Verify the operating altitude: 3,000 m (9,843 feet), derate maximum operating temperature by 1 °C per 300 meters (984.25 feet) above 1500 meters (4921.25 feet) altitude.
7. Maximum input voltage:
TekConnect channels: ≤1.2 V/FS settings:
±1.5 V relative to the termination bias (30 mA maximum).
2 DPO70000SX Series User
Install your instrument
±5 V absolute maximum input.
>1.2 V/FS settings: ±8.0 V. Limited by maximum Vterm current and the attenuator power.
ATI Channel: ± 0.75 V
Aux channel: ±5.0 V
pk
pk

Power supply requirements

The power supply requirements for your instrument are listed in the following table.
Source Voltage and Frequency Power Consumption
100 VAC to 240 VAC, 50/60 Hz or 115 VRMS, 400 Hz.
WARNING.
To reduce the risk of fire and shock, ensure that the mains supply voltage fluctuations do not exceed 10% of the operating voltage range.

Preventing instrument damage

Preventing ESD

980 W
CAUTION. A direct electrostatic discharge can damage the instrument input. To learn how to avoid this damage, read the following information.
Electrostatic discharge (ESD) is a concern when handling any electronic equipment. The instrument is designed with robust ESD protection, however it is still possible that large discharges of static electricity directly into the signal input may damage the instrument. To avoid damage to the instrument, use the following techniques to prevent electrostatic discharge to the instrument.
1. Discharge the static voltage from your body by wearing a grounded antistatic wrist strap while connecting and disconnecting cables and TekConnect adapters. The instrument provides a front panel connection for this purpose.
DPO70000SX Series User 3
Install your instrument
2. A cable that is left unconnected on a bench, or carried across a room, can develop a large static charge. Discharge the static voltage from all cables before connecting them to the instrument or device under test by momentarily grounding the center conductor of the cable, or by connecting a 50 Ω termination to one end, before attaching the cable to the instrument.
CAUTION. Do not use a tool (such as a screwdriver, wrench end, etc) to short the center conductor to the ground shell on the connector, as any sort of scratch or nick may damage the RF response of the cable.
4 DPO70000SX Series User
Install your instrument

Observe maximum input voltage

WARNING. Observe all terminal ratings. To avoid damaging the instrument, observe all rating and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
The maximum input voltage for the ATI (Asynchronous Time-Interleaved) input is ±0.75 Vpk. This is the maximum voltage before damage occurs.
The maximum input voltage for the TekConnect inputs is ±1.5 V relative to the termination bias (30 mA maximum) on ≤ 1.2 V full scale settings and 8 V on >1.2 V full scale settings.

Selecting the proper attenuator

The ATI input has a maximum voltage range of 300 mVFS. Use of an attenuator can increase the maximum voltage range.
Table 4: Maximum ATI voltage range
Attenuator Maximum ATI voltage range
None 300 mV
3 dB 420 mV
6 dB 600 mV
10 dB 950 mV
16 dB (6 dB + 10 dB attenuators) 1.88 V
20 dB 3 V
FS
FS
FS
FS
FS
FS

Connector cleaning

All connectors must be kept clean. Dirt in connectors can damage the connector and any connectors the dirty connector is used with. Dirt will also likely impair the RF performance. All cables, attenuators, and adapters should be stored with connector covers to keep dust out.
Before each use do the following:
DPO70000SX Series User 5
Install your instrument
1. Verify that the connectors are free of dirt, metal particles, scratches and deformations.
2. Verify the connectors are the correct size.
WARNING. To avoid eye damage, ware eye protection when using compressed air.
3. With the connector pointed down, use low pressure compressed air, at an angle to the connector, to clean the connectors.
4. If the connector still needs cleaning, use an appropriately sized swab moistened with isopropyl alcohol to clean the connector threads and mating surfaces. Do not snag the center conductor.
CAUTION. Never put pressure on the center pin of the connectors. Applying pressure to the center pin could damage the connector.

Proper connection technique

Use 8 inch-pounds of torque when making connections to the ATI 1.85 mm connector, SMA connectors, adapters, DC Blocks, or cables. Using the improper torque or connection technique may degrade your signals and damage the connectors.
Use the following procedure to make connections to the ATI input:
CAUTION. Failure to use the supplied backing wrench and torque wrench when making connections to the ATI input may damage your instrument.
6 DPO70000SX Series User
Install your instrument
1. Discharge the static voltage from your body by wearing a grounded antistatic wrist strap while connecting and disconnecting cables and adapters. The instrument provides a front panel connection for this purpose. Use the ground connection with the black plastic grommet, as this has a 1 MΩ series isolation resistance, so that there is no chance of shock, but the static voltage can be discharged.
Figure 1: Wear an antistatic wrist strap
2. Use the supplied connector saver to protect the instrument connector. When not in use, protect against dirt and ESD
damage by installing the end cap. Alternate locations to store the ATI protective cover include the upper right corner of the front panel and to the right of the audio connectors on the rear panel. When installing the connector saver tighten to 8 inch­pounds.
CAUTION. The connector saver can appear to tighten/loosen opposite to expectations because the oscilloscope ATI input connector is a male connector.
Figure 2: Installing a connector saver
3. Use the backing wrench on the connector saver to prevent damage caused by rotation of the connectors when installing a
connector.
DPO70000SX Series User 7
Install your instrument
4. Use the supplied torque wrench to tighten (8 inch-pounds) your cable connector to the connector saver.
Figure 3: Installing a cable
CAUTION. Failure to prevent rotation of the center conductors when making connections to the ATI input will damage your instrument. If necessary use a second backing wrench on the cable, to prevent its rotation. Use the supplied torque wrench to tighten (8 inch-pounds) all connectors.
5. If necessary use a second backing wrench on the cable, to prevent its rotation.
Figure 4: Use a second backing wrench
8 DPO70000SX Series User
Install your instrument

Power on the instrument

1. Connect the AC power cord to the rear of the instrument.
2. Use the front-panel power button to switch the instrument on.
The power button indicates instrument power states:
No light – no power applied
Green – powered on
DPO70000SX Series User 9
Install your instrument

Power off the instrument

1. Press the front-panel power button to shut down the instrument.
NOTE. If powering off a multi-instrument configuration, power off the master first.
The shutdown process takes approximately 30 seconds to complete, placing the instrument in standby mode. Alternatively, use the Windows Shutdown menu.
NOTE. You can force an immediate shutdown by pressing and holding the power button for four seconds. Unsaved data is lost.
2. To completely remove power to the instrument, perform the shutdown just described, and then remove the power cord from the instrument.

Multi-instrument configuration

Connect instruments in a multi-instrument configuration to create an up to 4 channel, highest bandwidth, and highest sample rate system with synchronized trigger and sample clock.

Instrument stacking

Multiple instruments may be stacked to save space and allow shorter cables and more convenient connections.
WARNING. When using an instrument upside down, be careful to avoid pinching your fingers.
10 DPO70000SX Series User
Install your instrument
DPO70000SX Series User 11
Install your instrument
Standard orientation of the top feet is useful when placing an instrument inverted on the work bench. Standard orientation of the top feet is also use when stacking an OM4000 on the instrument.
Use the rotated orientation of the top feet to stack two instruments on top of each other.
When stacking two instruments, the bottom feet sit in the holes exposed when the top feet are rotated.
12 DPO70000SX Series User
Install your instrument

Before startup

All instruments in a multi-Instrument configuration must be connected properly before powering on the instruments. If the configuration is changed after the instruments are powered on, a system restart may be required.
If an instrument is not connected in a multi-instrument configuration, the instrument powers on in its stand alone mode, not multi­instrument mode.

UltraSync™ bus cable

An UltraSync bus cable will connect any multi-instrument configuration capable instrument.
All cables must be connected to the instruments prior to power up. A restart will be required if the instruments are turned on before the cables are connected.

UltraSync™ bus cable connection order

Connect UltraSync bus cable bundles to the Master:
1. Select which instrument will be the Master instrument.
At the Master instrument connect each UltraSync cable bundle set to the designated ports (start with the To Extension B ports; then if connecting a third instrument use the To Extension C ports, and then if connecting a fourth instrument use the To Extension D ports):
2. Connect the clock cable to the 12.5 GHz Clock Out connector (8 in-lbs torque).
3. Connect the Trigger cable to the Trigger connector.
4. Connect the PCIe cable to the PCIe connector.
DPO70000SX Series User 13
Install your instrument
Connect UltraSync bus cable bundles to the Extension.
At the Extension, connect each UltraSync cable bundle set to the designated ports:
1. Connect the clock cable to the 12.5 GHz Clock In connector (8 in-lbs torque).
2. Connect the trigger cable to the To Master Trigger connector.
3. Connect the PCIe cable to the PCIe connector
14 DPO70000SX Series User
Install your instrument

Master and extension connection order

Connect each extension instrument to the master instrument sequentially from left to right (Extension B, C, and then D).
NOTE. Connect the extensions to the master in a star pattern, do not daisy chain the instruments.
If powering off a multi-instrument system, power off the master first.
DPO70000SX Series User 15
Install your instrument

Multi-instrument power on

During the start-up process the instrument checks to see if instruments are connected in a Multi-instrument configuration. If instruments are not connected, the instrument powers up in stand-alone mode. If instruments are connected, the configuration is validated. If the configuration is not valid, guidance is provided to help you create a valid configuration.
NOTE. After successful power on of your multi-instrument system, wait for the warm-up time to complete, and then run Signal Path Compensation.
The following image shows an extension with a connection that is invalid and provides guidance to correct the connection (provide the missing connection to the master).
The following image shows a master connected incorrectly and provides guidance to fix the problem (provide the missing connection to the extension).
16 DPO70000SX Series User
Install your instrument
The following images show the master and extension with valid connections.
DPO70000SX Series User 17
Install your instrument
After the multi-instrument connections are verified and the instruments display their operating status displays, the master displays its status and timer display. Pause the timer if you have additional changes to make before the oscilloscope application starts. Press Start Scope to bypass the timer. When the timer runs down, the oscilloscope application starts.
18 DPO70000SX Series User
Install your instrument
If your multi-instrument configuration includes ATI instruments, you have the opportunity to select either ATI Channels or non-ATI (TekConnect) Channels. In this example ATI Channels are selected.
NOTE. You also have the option to select Time Synchronized mode of all TekConnect channels. In this mode acquisitions are time synchronized and the Master controls the horizontal, record length, and trigger settings of all TekConnect channels. To ensure all channels are synchronized, operate the multi-instrument configuration in single sequence mode. Retrieve all TekConnect channel data using programmable interface commands. Waveform data is not sent to the Master. If you plan to use Time Synchronized mode, contact your Tektronix representative for additional information.
After the multi-instrument connections are verified, the instruments display their operating status displays. In this example the Master (A) is acquiring on channel 2 (ATI). The Extension (B) is acquiring on channel 2 (ATI) and is displayed as channel 1.
The master display identifies itself as the Master and identifies the channels being acquired.
To switch between the Configuration Application and the oscilloscope display (on the oscilloscope), use Alt-Tab. To close the Configuration Application, click the x. If the x is not visible, make the x visible by double clicking in the upper right hand corner of the Configuration Application.
DPO70000SX Series User 19
Install your instrument
After the multi-instrument system powers on, perform a signal path compensation. See Signal path compensation on page 26.

Switching between multi-instrument modes

To switch among ATI channels, TekConnect channels, and Time Synchronized modes, use the following procedure:
1. From the Vertical > Configuration menu select either ATI, TekConnect, or TimeSync.
20 DPO70000SX Series User
Install your instrument
2. Click Yes to continue.
Restart of a multi-instrument system. To restart your multi-instrument system without powering down the oscilloscopes, do the
following step:
1. To shut down the Configuration Manager and all of the Scope applications and then restart them without changing the configuration mode, select File > Restart.

ATI versus TekConnect channels

ATI channels provide up to 70 GHz bandwidth and 200 GS/s sample rate and the lowest noise. When using the ATI channel, the TekConnect channels are not available.
TekConnect channels have up to 33 GHz bandwidth and 100 GS/s sample rate. TekConnect channels offer multiple connection options with TCA adapters, and are compatible with a wide range of performance probes.
DPO70000SX Series User 21
Install your instrument

Multi-instrument status displays

In a multi-instrument configuration, the display shows status information such as: channel numbers, on/off, run/stop, Arm/trig'd, and UltraSync connectivity information. The master instrument also displays the trigger setup and the channels being acquired and displayed.

Available features

Features discussed in this manual that are available on some instruments or configurations are noted in the tables.
Table 5: Stand-Alone instrument
Feature DPO77002SX
DPO75902SX DPO75002SX
State (Clocked
Available on TekConnect channels Available
Pattern) trigger
XYZ mode Available on TekConnect channels Available
Equivalent time
Available on TekConnect channels Available
acquisitions
HiRes and peak
Available on TekConnect channels Available
detect
FastAcq Available on TekConnect channels Available
Roll mode Available on TekConnect channels Available
IRE and MV graticule
Available on TekConnect channels Available
modes
DPO73304SX DPO72304SX DPO71604SX DPO71304SX
22 DPO70000SX Series User
Install your instrument

DPO7AFP Auxiliary Front Panel (optional)

The DPO7AFP is an optional plug-in panel that provides physical knobs and buttons to control the DPO70000SX series oscilloscope. To connect the DPO7AFP:
1. Close the TekScope application.
NOTE. The TekScope application must be closed before connecting the DPO7AFP.
2. Plug the DPO7AFP into any USB port on the DPO70000SX. Wait until the Windows OS recognizes and loads the required drivers.
3. Start the TekScope application. The panel LEDs will light when TekScope connects to the panel.
4. To verify that the DPO7AFP is working, push any channel button and observe that the TekScope application enables or disables the selected channel
NOTE. If you unplug the DPO7AFP while TekScope is running, you will need to close TekScope, reconnect the DPO7AFP, and restart TekScope to use the DPO7AFP.

Inspect the instrument

Every time you power on the instrument, the instrument automatically performs the power-on self test diagnostics.

Verify internal diagnostics pass

Use the following procedures to verify the functionality of your instrument.
DPO70000SX Series User 23
Install your instrument
1. Power on the instrument.
2. Select Instrument Diagnostics....
3. Click Run. The test results appear in the
diagnostics control window.
4. Verify that all tests pass. If diagnostic failures occur, contact your local Tektronix service personnel.

Activating Windows 10

The copy of Microsoft Windows 10 that ships with this product is shipped in Deferred Activation mode. Microsoft has a default activation method that requires a connection to the Internet. If you are not able to connect to the Internet, you can activate Windows 10 by phone or you can continue to operate in Deferred Activation mode.
First time activation. You can activate Windows 10 either of the following ways.
1. Connect the instrument to the Internet. Windows will automatically activate. You can also go to System Properties to
connect and activate.
2. Call Microsoft and speak with a representative to activate Windows 10. Contact information and a unique activation code (installation ID) will appear in a pop up window.
Activation with removable drives. If the SSD is removed from one instrument and inserted into another instrument, then the activation process will need to run again on the second instrument. If the second instrument is not connected to the Internet, the activation will fail and an Activate Windows watermark will display on the screen. To activate Windows and remove the watermark, you must connect to the Internet or call Microsoft.
24 DPO70000SX Series User
Install your instrument

Windows interface guidelines

Because the instrument uses the Microsoft Windows interface, you have open access to the Windows operating system. You can access the Windows desktop to load and run other Windows-based applications such as Microsoft Excel.
Follow these guidelines to avoid making operating system changes that might cause problems while using the instrument:
Be careful when making changes in the Control Panel. Avoid making changes to any controls with which you are unfamiliar.
Do not delete or change any system fonts; this can affect the quality of the display.
Be careful when making changes to the system Display properties. Changing settings such as resolution, text size, fonts, and orientation affects the usability of the display and the touchscreen.
Do not change the contents of the Windows folder or the Program Files\Tektronix\AWG70000\ folder.
Do not change the BIOS settings; this can affect the overall operation of the instrument.
DPO70000SX Series User 25
Install your instrument

Signal path compensation

Perform the Signal Path Compensation (SPC) regularly to ensure that your measurements have the highest level of accuracy. Tektronix considers it a best practice to run SPC when using the instrument to measure signals with higher sensitivity (10 mV/div and lower) settings regardless of temperature shift or time since it was last run. Failure to perform SPC may result in the instrument not meeting warranted performance levels.
SPC corrects for DC inaccuracies caused by temperature variation or by long-term drift. SPC optimizes the acquisition system, corrects DC offset, and interleave calibration. SPC is adversely affected by input signals with AC components.
Use this procedure to optimize the acquisition system:
If the temperature has changed more than 5 °C (9 °F) since the last signal path compensation (SPC)
If using the instrument to measure signals with higher sensitivity (10 mV/div and lower) run SPC at least once a week
If the front panel SPC status icon is not green
If you replace or insert drive media
If you change the configuration of your multi-instrument system, such as changing which instrument is the master or an extension.
1. Prerequisites:
Instrument must be powered on until Utility > Instrument Calibration > Temperature Status is Ready.
All channel input signals must be removed.
If the timebase external reference mode is selected, leave the external reference signal connected and active.
26 DPO70000SX Series User
Install your instrument
2. Select Instrument Calibration.
3. When the Temperature Status changes
to Ready, click Run SPC to start the calibration. Calibration may take 10 to 15 minutes.
NOTE. Before running the SPC calibration, remove all channel input signals.
NOTE. If you use a drive that has not had SPC run while in the current instrument, you will see a no prior SPC warning message. If you see this warning, run SPC.
Utility menu on DPO70000SX
instruments
Calibration menu on DPO70000SX instruments
DPO70000SX Series User 27
Install your instrument
4. If the instrument does not pass, recalibrate the instrument, or have the instrument serviced by qualified service personnel.
NOTE. To always show the SPC status icon or display a warning when SPC has not been run in more than a month, click the corresponding check box.
Calibration menu on DPO70000SX instruments
5. If the SPC needed icon is red, perform a signal path compensation.
Check the color of the SPC Status icon:
Green indicates that SPC successfully passed and the temperature is stable.
Yellow indicates that the instrument is in the warm up state or that it has been over 30 days since SPC was last run.
Red indicates that SPC needs to be run (the temperature has varied more than 5 °C, SPC failed, or SPC has not been run).
28 DPO70000SX Series User
Install your instrument

Connecting to a network

Connect your instrument to a network for printing, file sharing, internet access, and other functions. Consult with your network administrator and use the standard Windows utilities to configure the instrument for your network.
Use the Remote Desktop Connection of Windows to control and view the instrument remotely.
DPO70000SX Series User 29
Install your instrument

Adding a second monitor

You can operate the instrument while using Windows and installed applications on an external monitor. Follow the procedure below to set up a dual monitor configuration.
1. Turn power off.
2. Connect second monitor.
3. Connect keyboard.
4. Connect mouse.
5. Turn instrument power on.
6. Turn monitor power on.

Activating Windows 10

The copy of Microsoft Windows 10 that ships with this product is shipped in Deferred Activation mode. Microsoft has a default activation method that requires a connection to the Internet. If you are not able to connect to the Internet, you can activate Windows 10 by phone or you can continue to operate in Deferred Activation mode.
First time activation
You can activate Windows 10 either of the following ways.
1. Connect the instrument to the Internet. Windows will automatically activate. You can also go to System Properties to connect and activate.
2. Call Microsoft and speak with a representative to activate Windows 10. Contact information and a unique activation code (installation ID) will appear in a pop up window.
30 DPO70000SX Series User
Install your instrument
Activation with removable drives
If the SSD is removed from one instrument and inserted into another instrument, then the activation process will need to run again on the second instrument. If the second instrument is not connected to the Internet, the activation will fail and an Activate Windows watermark will display on the screen. To activate Windows and remove the watermark, you must connect to the Internet or call Microsoft.

Restoring instrument operating system and product software

You can restore the instrument’s Windows Operating System directly from the instrument. The instrument does not ship with an operating system restore disc.
To restore or update the product software, you need to download the current version from the Tektronix web site.
NOTE. Restoring or updating the product software does not require that you restore the Windows operating system.

Operating system restore

CAUTION. Using the restore process reformats the hard drive and reinstalls the operating system. All saved data is lost. If possible, save important files to external media before performing a system restore.
The Windows operating system installed is designed for this instrument’s hardware and product software. Installing a version other than the one provided will not function properly.
After the operating system is installed, you need to download the product software installation package from the Tektronix web
site and reinstall the software.

Internal recovery utility

This is the preferred method to restore a damaged Windows Operating System. This method restores the Windows Operating System using the Acronis Startup Recovery Manager. The Acronis software reinstalls the operating system using restore images preinstalled on the hard drive.
This method retains the restore images, providing the ability to repeat the restore process.
1. Attach the keyboard to the instrument.
2. Restart the instrument. During the boot-up process you will see the following message at the top of the screen:
Starting Acronis Loader... press F5 for Acronis Startup Recovery Manager
NOTE. To successfully complete the system restore, you must use the Windows version of the Acronis software. Using a generic MAC keyboard starts the DOS version of the Acronis software. Do not use a MAC keyboard.
3. Repeatedly press the F5 key until the Acronis True Image Tool opens. There is a 15-second time period from when the message appears until the instrument proceeds with the normal instrument startup. If the instrument does not open the Acronis application, power off the instrument, then power on the instrument and try again.
4. Click Restore.
5. In the Confirmation dialog box, click Yes to restore the instrument operating system, or No to exit the restore process. The
restore process takes approximately 30 minutes; the actual time depends on the instrument configuration.
DPO70000SX Series User 31
Install your instrument

Product software installation

NOTE. For DPO77002SX instruments you must order the product software from Tektronix.
1. Download the product software installation package. The installation package contains:
Instructions
Product software installer
2. Install the product software following all on-screen prompts.
32 DPO70000SX Series User

Getting acquainted with your instrument

Front-panel connectors

Figure 5: ATI with TekConnect channels
DPO70000SX Series User 33
Getting acquainted with your instrument
Figure 6: TekConnect channels
Table 6: Front-panel connectors
Identifier Connector Description
1 USB 2.0 Four USB connectors.
2 Chassis ground Banana type ground connection.
Ground through 1 MΩ resistor Banana type ground connection.
Discharge the static voltage from your body by wearing a grounded antistatic wrist strap while connecting and disconnecting cables and TekConnect adapters.
3 DC Probe Calibration Probe calibration output
4 1, 2, 3, and 4(Analog inputs) These connectors supply analog signals.
5 Aux In Auxiliary trigger input
6 Differential Fast Edge Differential fast edge step output.
34 DPO70000SX Series User
Getting acquainted with your instrument

Rear-panel connectors

Table 7: Rear-panel connectors
Connector Description
PCIe PCIe port to connect external devices or to configure multi-instrument systems.
12.5 GHz Clock In SMA type connector to provide an external clock signal used in multi-instrument configurations.
12.5 GHz Clock Out SMA type connector to provide a high speed clock that is related to the sample rate. This clock is used in multi-instrument configurations.
Trigger UltraSync trigger bus.
LAN RJ-45 connectors to connect the instrument to a network.
USB 3.0 Four USB 3.0 Host connectors (type A) to connect devices such as a mouse, keyboard, or other USB
devices. Tektronix does not provide support or device drivers for USB devices other than the supplied mouse and keyboard.
USB USB device connector
VGA VGA port to connect a monitor for extended desktop operation. To connect a DVI monitor to the VGA
connector, use a DVI-to-VGA adapter.
DVI-D Use the DVI-D Video port to send the instrument display to a projector or to a flat-panel LCD monitor.
Display port These connectors provide digital display interfaces
PS/2 mouse This connector is for a PS/2 mouse.
PS/2 keyboard This connector is for a PS/2 keyboard
Audio These connectors provide microphone input, line in, and line out.
Auxiliary output SMA type connector provides a TTL-compatible, negative polarity pulse when the instrument triggers or
when certain other events occur, such as a mask test failure or completion.
Reference output SMA type connector to provide a synchronization signal to external devices.
External reference input SMA type connector to provide an external reference clock input.
Power Power cord input.
DPO70000SX Series User 35
Getting acquainted with your instrument

Interface and display

The menu bar mode provides access to commands that control all of the instrument features and functions. The toolbar mode provides access to the most common features.
1. Menu Bar: Access to data I/O, printing, online help, and instrument functions
2. Buttons/Menu: Click to toggle between toolbar and menu bar modes and to customize your toolbar
3. Multipurpose Knob Readouts: Adjust and display parameters controlled by the multipurpose knobs
4. Display: Live, reference, math, and bus waveforms display here, with cursors
5. Waveform Handle: Click and drag to change vertical position of a waveform or bus. Click the handle and change the position and scale using the multipurpose knobs.
6. Controls Status: Quick reference to vertical selections, scale, offset, and parameters. Probe status is provided for some probe tips.
7. Readouts: Display cursor and measurement readouts in this area. Measurements are selectable from the menu bar or toolbar. If a control window is displayed, some combinations of readouts move to the graticule area.
WARNING. If there is vertical clipping, there may be a dangerous voltage on the probe tip, but the readout will indicate a low
voltage. A symbol appears in the measurement readout if a vertical clipping condition exists. Automatic amplitude-related measurements where the signal is vertically clipped produce inaccurate results. Clipping also causes inaccurate amplitude values in waveforms that are stored or exported for use in other programs. If a math waveform is clipped, it will not affect amplitude measurements on that math waveform.
8. Status: Display of acquisition status, mode, and number of acquisitions; trigger status; date; time; and quick reference to record length and horizontal parameters
36 DPO70000SX Series User
Getting acquainted with your instrument
1. Buttons/Menu: Click to toggle between toolbar and menu bar modes and to customize the toolbar
2. Drag cursors to measure waveforms on screen
3. Drag the position icons to reposition a waveform
4. Click the icon to assign the multipurpose knobs to waveform vertical position and scale
5. Drag across the waveform area to create a box for zooming, enabling/disabling histograms, gating measurements, and adding and controlling visual trigger areas
6. Drag icon to change the trigger level
DPO70000SX Series User 37
Getting acquainted with your instrument

Control panel

NOTE. These controls are available on the Auxiliary Front Panel accessory. The functions these controls perform are available in the instrument menus.
1. Push to automatically set up the vertical, horizontal, and trigger controls based on selected channels.
2. Push to return settings to default values.
3. Push to make a hard copy or save a
screen capture.
4. Push to turn on MultiView Zoom and add a magnified graticule to the display.
5. Horizontally scale, position, delay, and set record length (resolution) of all waveforms.
6. Use to start and stop acquisition, start a single acquisition sequence, clear data, or start fast acquisitions.
7. Use to set the trigger parameters. Push Advanced to display additional trigger functions. The Arm, Ready, and Trig'D lights show the acquisition status.
8. Turn to adjust waveform intensity.
9. Turn to adjust parameters selected from
the screen interface. Push to toggle between normal and fine adjustment.
10. Push to turn cursors on or off.
11. Use to search and mark waveforms.
12. Push to turn the touch screen on and off.
13. Turn channel displays on and off.
Vertically scale, position, or offset the waveform. Toggle between position and offset.
38 DPO70000SX Series User
Getting acquainted with your instrument

Accessing online help

In-depth information is available in the online help on all the features of your instrument.
To access context-sensitive help on the active window, select Help > Help on Window... or press F1.
1. To access any topic in the help system,
select Help > Contents and Index....
2. Use the Contents, Index, Search, or Favorites tab to select the topic, and then click Display.
To navigate within the help system:
Click a button in a help window to navigate between the Overview and specific topics.
Click the Minimize button in the help window to move help out of the way so you can operate the instrument.
Click Alt and Tab to see the last help topic again.
DPO70000SX Series User 39
Getting acquainted with your instrument

Accessing menus and control windows

Access menus and control windows using the following techniques:
Click a menu, and then select a command.
For a shortcut menu, right-click anywhere in the graticule or on an object. The shortcut menu is context sensitive and varies with the area or object where you right-clicked. Some examples are shown in the figure at right.
In the toolbar mode, click a button to quickly access a setup control window. See Interface and display.
40 DPO70000SX Series User

Inspect your instrument

Use the following procedures to verify the functionality of your instrument.

Verify internal diagnostics pass

1. Power on the instrument.
2. Select Instrument Diagnostics....
3. Click Run. The test results appear in the
diagnostics control window.
4. Verify that all tests pass. If diagnostic failures occur, contact your local Tektronix service personnel.
DPO70000SX Series User 41
Inspect your instrument
42 DPO70000SX Series User

Acquisition

This section contains concepts of and procedures for using the acquisition system.

Signal path compensation

Perform the Signal Path Compensation (SPC) regularly to ensure that your measurements have the highest level of accuracy. Tektronix considers it a best practice to run SPC when using the instrument to measure signals with higher sensitivity (10 mV/div and lower) settings regardless of temperature shift or time since it was last run. Failure to perform SPC may result in the instrument not meeting warranted performance levels.
SPC corrects for DC inaccuracies caused by temperature variation or by long-term drift. SPC optimizes the acquisition system, corrects DC offset, and interleave calibration. SPC is adversely affected by input signals with AC components.
Use this procedure to optimize the acquisition system:
If the temperature has changed more than 5 °C (9 °F) since the last signal path compensation (SPC)
If using the instrument to measure signals with higher sensitivity (10 mV/div and lower) run SPC at least once a week
If the front panel SPC status icon is not green
If you replace or insert drive media
If you change the configuration of your multi-instrument system, such as changing which instrument is the master or an extension.
1. Prerequisites:
Instrument must be powered on until Utility > Instrument Calibration > Temperature Status is Ready.
All channel input signals must be removed.
If the timebase external reference mode is selected, leave the external reference signal connected and active.
DPO70000SX Series User 43
Acquisition
2. Select Instrument Calibration.
3. When the Temperature Status changes
to Ready, click Run SPC to start the calibration. Calibration may take 10 to 15 minutes.
NOTE. Before running the SPC calibration, remove all channel input signals.
NOTE. If you use a drive that has not had SPC run while in the current instrument, you will see a no prior SPC warning message. If you see this warning, run SPC.
Utility menu on DPO70000SX
instruments
Calibration menu on DPO70000SX instruments
44 DPO70000SX Series User
Acquisition
4. If the instrument does not pass, recalibrate the instrument, or have the instrument serviced by qualified service personnel.
NOTE. To always show the SPC status icon or display a warning when SPC has not been run in more than a month, click the corresponding check box.
Calibration menu on DPO70000SX instruments
5. If the SPC needed icon is red, perform a signal path compensation.
Check the color of the SPC Status icon:
Green indicates that SPC successfully passed and the temperature is stable.
Yellow indicates that the instrument is in the warm up state or that it has been over 30 days since SPC was last run.
Red indicates that SPC needs to be run (the temperature has varied more than 5 °C, SPC failed, or SPC has not been run).
DPO70000SX Series User 45
Acquisition

Setting up analog signal input

Use the drop down menus or the Auxiliary Front Panel to set up your instrument to acquire the signal.
If using the drop down menus, do the following steps:
1. Connect the probe or cable to the input signal source.
CAUTION. To prevent damage to the instrument, always wear an antistatic wrist strap when making connections to the instrument and observe the maximum input voltage ratings for input connectors.
2. Select the input channel: Select Vertical > Vertical Setup. Select the tab for the
channel you want to select, and then press the Display button to toggle the channel on or off.
3. Select Horiz/Acq > Autoset.
4. Select Vertical > Vertical Setup. Adjust
the vertical position, scale, and offset: Double click the Position, Scale, and Offset controls and use the pop-up keypad make the adjustment.
5. If you want to invert the input signal, click the Invert button to toggle invert on and off. The Invert button is in the Vertical Setup menu or the Deskew/Attenuation/ Invert menu.
46 DPO70000SX Series User
Acquisition
6. Select Horiz/Acq > Horizontal/ Acquisition Setup. Adjust the horizontal
position and scale: Double click the Position and Scale controls and use the pop-up keypad to make the adjustment.
The horizontal position determines the number of pretrigger and posttrigger samples.
If using the Auxiliary Front Panel, do the following steps:
1. Connect the probe or cable to the input signal source.
CAUTION. To prevent damage to the instrument, always wear an antistatic wrist strap when making connections to the instrument and observe the maximum input voltage ratings for input connectors.
2. Select the input channel.
3. Select Horiz/Acq then Autoset.
4. Adjust the vertical position, scale, and
offset.
DPO70000SX Series User 47
Acquisition
5. Adjust the horizontal position and scale.
The horizontal position determines the number of pretrigger and posttrigger samples.

Using default setup

1. To quickly return to the factory default
settings, select Recall Default Setup from the file menu or push DEFAULT SETUP on the Auxiliary Front Panel.
48 DPO70000SX Series User
Acquisition

Using autoset

Use Autoset to quickly and automatically set up the instrument (acquisition, horizontal, trigger, and vertical) based on the characteristics of the input signal. Autoset makes adjustments to the signal such that the waveform displays two or three cycles with the trigger level near the midlevel.
1. Connect the probe, and then select the input channel. See Setting up analog
signal input on page 46.
2. To execute an Autoset, select Horiz/Acq and then Autoset or push the AUTOSET button on the Auxiliary Front Panel.
3. Click Undo to undo the last Autoset. Parameters that are not affected by Autoset retain their settings.
Quick tips
Autoset automatically sets up the analog channels.
On instruments with iCapture, autoset automatically sets up the iCapture channels.
To position the waveform appropriately, Autoset may change the vertical position. Autoset may also adjust vertical offset.
If you use Autoset when one or more channels are displayed, the instrument selects the lowest numbered channel for horizontal scaling and triggering. You can individually control the vertical scaling of each channel.
If you use Autoset when no channels are displayed, the instrument turns on channel one (Ch 1) and scales it.
Close Autoset Undo control window by clicking the X. After Autoset Undo closes, you can still undo the last Autoset by selecting the Undo Last Autoset command from the Edit menu.
Stop the Autoset Undo control window from opening automatically by changing the User Preferences in the Utilities menu.
DPO70000SX Series User 49
Acquisition

Probe compensation and deskew

To optimize measurement accuracy, see the instrument online help to perform the following procedures:
Compensate passive probes
Compensate active probes
Deskew input channels

Deskew Tool

From the Analyze menu, select Deskew Tool.
Overview
Use the Deskew Tool to compensate for propagation delays in your input signal path.
The deskew tool can be used either on a stand alone oscilloscope or between multiple DPO70000SX oscilloscopes connected via ultrasync cables.
The deskew tool uses the oscilloscope trigger point to align all channels in time. This tool requires that a live signal be connected to each channel. The live signals must be time synchronized to the trigger channel. The signals used for deskew can be the same or opposite polarity compared to the trigger channel. For signals with the same polarity, the nearest edge with same polarity as the trigger will be used as the reference edge for alignment. For differential signals, the nearest edge with the opposite polarity as the trigger will be used for alignment.
50 DPO70000SX Series User
Acquisition
1. Display Mode where the modes can be:
Standalone for a standalone oscilloscope
Multiscope for multiple ATI/TekConnect instruments
Time Sync mode
2. Channel selection - channels are grayed out when the channel is off. Press the channel button to toggle the channel on or off.
3. Histogram view - Mean, StdDev and Population are shown.
4. Slope and Level - define the slope and level to use for deskewing.
Trigger uses the trigger slope and level as the reference polarity and level for the channel's alignment.
Rise uses the rising edge nearest the trigger point and allows you to adjust the level.
Fall uses the falling edge nearest the trigger point and allows you to adjust the level.
Either uses either the rising or falling edge and level.
NOTE. When using the deskew tool to align a differential pair select the trigger slope for the first channel and an edge with its polarity opposite of the trigger's polarity.
5. Deskew To Mean - adds the measured skew value for each selected channel to the channel deskew value for each respective channel.
6. Set 0 - resets all channel deskew values to 0.
7. Histogram Reset - clears the histograms of their population data.
8. Set Default - sets the Slope to Trigger on all channels and resets the histograms.
The tool works by precisely computing the difference between the trigger location and each channel's selected edge. The computed values are shown in the histogram view. To align the oscilloscope channels, select the slope and level, wait for a sufficient population to be accumulated (wait for the indicator near the To Mean button to turn green), and then press the To Mean button.
Typically on DPO70000SX oscilloscopes it takes two accumulations and To Mean button presses to get femtosecond (fs) level alignment. This is because when deskew values are set to 0 in the oscilloscope, the skew values can be in the picosecond (ps) range. The difference in resolution between ps and fs is large. So the first try gets you close and the second gets into the fs range.
DPO70000SX Series User 51
Acquisition
To use
Use the Deskew Tool to compensate for propagation delays in your input signals.
1. Set the trigger mode to edge and adjust the trigger level on the reference channel until you have a stable trigger.
2. Select the channels to deskew. Ensure that each channel has a live signal attached and it has an edge that is within 0.5 UI
of the edge on the trigger channel.
3. Select the slope and level for each channel.
4. Wait for a sufficient population to be accumulated (the indicator near the To Mean button turns green)
5. Press the To Mean button.
6. Repeat steps 4 and 5 multiple times. The difference in resolution between ps and fs is large. So the first try gets you close
and the second gets into the fs range.
This figure is an example of fast edge signals before deskew:
This figure is an example of fast edge signals after deskew:
52 DPO70000SX Series User
Acquisition
This figure is an example of differential signals before deskew:
DPO70000SX Series User 53
Acquisition
This figure is an example of differential signals after deskew:
54 DPO70000SX Series User
Acquisition

Acquisition concepts

Acquisition hardware

Before a signal is displayed, it must pass through the input channel where it is scaled and digitized. Each channel has a dedicated input amplifier and digitizer. Each channel produces a stream of digital data from which the instrument extracts waveform records.
DPO70000SX Series User 55
Acquisition

Sampling process

Acquisition is the process of sampling an analog signal, converting it into digital data, and assembling it into a waveform record, which is then stored in acquisition memory.

Real-Time sampling

In real-time sampling, the instrument digitizes all of the points it acquires using one trigger event. Use real-time sampling to capture single-shot or transient events.

Interpolated Real-Time sampling

In interpolated real-time sampling, the instrument digitizes all of the points it acquires using one trigger event. If the instrument cannot acquire enough samples for a complete waveform at the maximum real-time sample rate, it interpolates. Use interpolated real-time sampling to capture single-shot or transient events.

Equivalent-Time sampling

The instrument uses equivalent-time sampling to extend its sample rate beyond its real-time maximum sampling rate. Equivalent-time sampling is only used if Equivalent Time is selected and the time base is set to a sampling rate that is too fast to create a waveform record using real-time sampling.
The instrument makes multiple acquisitions of a repetitive waveform to obtain the sample density required for one complete waveform record. Thus, equivalent time sampling should only be used with repetitive signals.
56 DPO70000SX Series User
Acquisition

Waveform record

The instrument builds the waveform record through use of the following parameters:
Sample interval: The time between sample points.
Record length: The number of samples required to fill a waveform record.
Trigger point: The zero time reference in a waveform record.
Horizontal position: When horizontal delay is off, the horizontal position is a percentage of the waveform record between 0 and 99.9 percent. The trigger point and the horizontal reference are at the same time in the waveform record. For example, if the horizontal position is 50 percent, then the trigger point is in the middle of the waveform record. When horizontal delay is on, the time from the trigger point to the horizontal reference is the horizontal delay.

Interpolation

Your instrument can interpolate between the samples it acquires when it does not have all of the actual samples it needs to fill the waveform record. Linear interpolation computes record points between actual acquired samples by using a straight line fit.
Sin(x)/x interpolation computes record points using a curve fit between the actual values acquired. Sin(x)/x interpolation is the default interpolation mode because it requires fewer actual sample points than linear interpolation to accurately represent the waveform.

How the acquisition modes work

Acquisition mode
Sample mode retains the first sampled point from
each acquisition interval. Sample is the default mode.
Peak Detect mode uses the highest and lowest of all the samples contained in two consecutive acquisition intervals. This mode only works with real-time, noninterpolated sampling and is useful for catching high frequency glitches.
Hi Res mode calculates the average of all the samples for each acquisition interval. Hi-Res provides a higher-resolution, lower-bandwidth waveform.
DPO70000SX Series User 57
Acquisition
Acquisition mode
Envelope mode finds the highest and lowest
record points over many acquisitions. Envelope uses Peak Detect for each individual acquisition.
Average mode calculates the average value for each record point over many acquisitions. Average uses Sample mode for each individual acquisition. Use average mode to reduce random noise.
Waveform Database mode is a three-dimensional accumulation of source waveform data over several acquisitions. In addition to amplitude and timing information, the database includes a count of the number of times a specific waveform point (time and amplitude) was acquired.

Enable enhanced effective number of bits

Use this control, in the Acquisition tab, to turn the Enhanced Effective Number of Bits (ENOB) on or off to optimize waveform detail.
1. Click the Enhanced Effective Number of Bits check box to turn ENOB On or Off. You will find this control in the Acquisitions tab of the Horizontal/Acquisition control window.
Enhanced Effective Number of Bits enables interleave correction DSP that improves the effective number of bits. This control is only available in some instrument models. This control is not available in Hi Res mode.
When Enhanced Effective Number of Bits is enabled, EB+ is displayed in the horizontal/acquisition readout.
58 DPO70000SX Series User
Acquisition

Changing the acquisition mode

Use this procedure to change the acquisition mode.
1. Select Horiz/Acq > Acquisition Mode.
2. To select an acquisition mode, do one of
the following:
Select an acquisition mode directly from the menu.
Click Mode..., and then select an acquisition mode.
3. For Average or Envelope acquisition modes, click the # of Wfms control, and then set the number of waveforms with the multipurpose knob. For WfmDB mode, click the Samples control, and then set the number of samples with the multipurpose knob.
Quick tip
Click the keypad icon to set the number of waveforms or samples.
WfmDB mode is not available in multi-unit configurations.
DPO70000SX Series User 59
Acquisition

Starting and stopping an acquisition

After the channels that you want to acquire are selected, use the following procedure.
1. To start the acquisition, select Horiz/Acq > Run/Stop and click Run/Stop or push the RUN/STOP button on the Auxiliary Front Panel.
2. To stop the acquisition, click Run/Stop again or push the RUN/STOP button.
3. To take a single acquisition, click Single Sequence or push the Single button.
60 DPO70000SX Series User
Acquisition

Selecting the horizontal mode

Your instrument has three horizontal modes. Automatic is the default mode. Select the horizontal mode that works best for your test setup.
To set the horizontal mode, select Horiz/Acq > Horizontal/Acquisition Setup to display the horizontal control window. Choose one of the modes described below.
In Automatic mode you can set the Scale and Sample Rate. Record length is a dependent variable. If changing the scale would cause the record length to exceed the Record Length Limit, the sample rate is decreased to the next available setting.
If the sampling mode is real time and the sample rate is at the real time limit, attempting to increase the sample rate has no effect.
To set the Record Length Limit, click Record Length Limit and set the limit using the buttons or keypad. The default maximum limit depends on your instrument model and record length option.
In Constant Sample Rate mode you can set the Sample Rate and Scale. The default sample rate ensures bandwidth filter operation. Record length is a dependent variable. The maximum record length depends on your instrument model and record length option.
The Auxiliary Front Panel resolution knob changes the sample rate in both automatic and constant sample rate modes.
DPO70000SX Series User 61
Acquisition
In Manual mode you can set the Sample Rate and Record Length. Horizontal Scale is a dependent variable calculated from the sample rate and record length.
The Horizontal Scale knob changes record length in manual mode.
All three modes interact with sample rate, scale, and record length as shown. The horizontal line is the maximum real-time sample rate. Each staircase shows that as you increase the scale, the sample rate must decrease when either the maximum record length or the record length limit you set is reached. Manual mode uses the maximum record length.
Automatic and Constant Sample Rate modes are identical. However, Constant Sample Rate mode keeps the sample rate constant at a rate that guarantees bandwidth enhancement filters are used.
62 DPO70000SX Series User
Acquisition

Using FastAcq

Fast acquisition mode reduces the dead time between waveform acquisitions, enabling the capture and display of transient events such as glitches or runt pulses. Fast acquisition mode can also display waveform phenomena at an intensity that reflects their rate-of-occurrence. FastAcq is not available on ATI channels and on some instrument configurations.
1. Push FastAcq on the optional front pannel or select Horiz/Acq > Fast
Acquisition.
2. Find glitches, transients, or other random
events.
When you have identified an anomaly, set the trigger system up to look for it. See Capturing intermittent anomalies on page 179.
Quick tips
To optimize for capturing details or rare events, select Horiz/Acq > Horizontal/Acquisition Setup > Acquisition> Fast Acq, and then select Optimize For Capturing Details or Capturing rare events.
DPO70000SX Series User 63
Acquisition

Using DSP enhanced bandwidth

If your instrument has the enhanced bandwidth feature, use the DSP (digital signal processing) enhanced bandwidth for more accurate rise time measurements, to extend the bandwidth, and flatten the passband at the full sample rate. The enhanced bandwidth provides a matched response across enabled channels so you can perform channel-to-channel comparison and differential measurements.
1. Use AUTOSET to set the horizontal, vertical, and trigger controls or set the controls manually.
2. Select Vertical > Bandwidth
Enhanced....
64 DPO70000SX Series User
Acquisition
3. Click Digital Filters (DSP) Enabled to turn the enhanced bandwidth on. You must set the sample rate correctly to enable DSP.
4. To force a constant sample rate that enables DSP filters, check Force Constant Sample Rate.
NOTE. If not already set, selecting Constant Sample rate sets the horizontal mode to constant sample rate, sets the sample rate to allow DSP, and selects a DSP bandwidth.
5. Select the desired bandwidth from the Bandwidth list.
The available bandwidth selections depend on your instrument, probe, and probe tip.
Selecting Analog Only selects a hardware (HW) bandwidth.
6. To apply your selections to all channels, check Apply To All Channels.
When different probing makes it impossible for the instrument to set all channels the same, the instrument sets each channel to the closest bandwidth value possible.
The bandwidth indicator appears in the vertical readout when the enhanced bandwidth is turned on.
Quick tips
Right click the waveform handle to display a menu where you can select the channel bandwidth and other bandwidth enhanced settings.
The DSP enhanced bandwidth occurs at maximum sample rate.
Use the DSP enhanced bandwidth when your signals have rise times less than 50 ps.
Select Analog Only for higher waveform throughput, overdriven signals, and when you prefer to use your own DSP post­processing.
You can limit the instrument bandwidth by selecting Vertical > Bandwidth Limit and then selecting the bandwidth.
DPO70000SX Series User 65
Acquisition

Setting the termination voltage

Oscilloscope inputs have traditionally been terminated to ground. The signals being measured often are not ground referenced. Pulling signals to ground can impair measurement results or potentially damage the DUT. Termination voltage is available on TekConnect channels, but not ATI channels.
The instrument delivers a variable termination voltage up to ±3.5 volts to the device under test (DUT), and supports a large offset range. This allows you to adjust the oscilloscope to mirror the conditions and behavior of the DUT, and measure high-speed signals in an environment similar to the one in which they operate.
Using the combination of termination voltage and offset to center the oscilloscope reference point in the operating range of the DUT maximizes the available dynamic range and minimizes the noise of the measurement system.
To set the termination voltage of a channel, do these steps:
1. Select Vertical > Vertical Setup.
2. Click in the Termination Voltage field.
Set the termination voltage using the keypad or multipurpose knobs.
66 DPO70000SX Series User
Acquisition

Using roll mode

Roll mode gives a display similar to a strip chart recorder for low-frequency signals. Roll mode displays acquired data points without waiting for the acquisition of a complete waveform record.
1. Select Horiz/Acq > Horizontal/
Acquisition Setup....
2. If not selected, click the Acquisition tab. Click Auto to turn on Roll mode.
NOTE. Roll mode requires Sample, Peak Detect, or Hi Res acquisition mode.
3. To stop acquisitions in Roll mode:
If you are not in Single Sequence, select Horiz/Acq > Run/Stop and then click Run/Stop or push RUN/ STOP on the Auxiliary Front Panel to stop Roll mode.
If you are in Single Sequence, Roll mode acquisitions stop automatically when a complete record is acquired.
Quick tips
Switching to Envelope, Average, or WfmDB acquisition mode will turn off Roll mode.
Roll mode is disabled when you set the horizontal scale to 50 ms per division or faster.
Roll mode is disabled when using a multi-unit configuration.
DPO70000SX Series User 67
Acquisition

Setting up a bus

You can set up serial (optional) and parallel buses.
1. Select Vertical > Bus Setup.
2. Select the Config tab.
3. Scroll to, and then select the bus to set
up.
4. To toggle the display of the bus on or off, click the Bus button.
5. To change the default label of the bus, click Label and enter a new label using the keyboard.
6. To change the displayed position of the bus, click Bus Position and use the keyboard or multipurpose knob to enter a new bus position.
68 DPO70000SX Series User
Acquisition
7. To select the bus type, click Bus Type Serial or Parallel.
You can set up each bus type.
See Set up a serial bus on page 70. or See Set up a parallel bus on page 71.
Quick tips
Position a bus or waveform by clicking on the bus or waveform handle and dragging the handle to the desired position.
DPO70000SX Series User 69
Acquisition

Set up a serial bus

You can set serial (optional) bus parameters.
1. To set up a serial bus, select Bus Type Serial and select the type of serial bus
from the drop-down list.
Each bus type has its own collection of parameters that you set up. Set up other bus types as appropriate for the selected bus.
See the online help for additional help setting up a bus.
2. To select the input for the displayed components of the bus, click the component Input and select from the displayed list.
3. To set an input threshold, double click Threshold and enter the threshold using the keypad.
NOTE. To set thresholds independently when they are currently shared, go to
Trig > A Event (Main) Trigger Setup and change Settings to Independent.
4. To select the polarity for the displayed components of the bus, click the component Polarity and select from the displayed list.
Quick tips
To use a custom serial decoder, see the online help.
70 DPO70000SX Series User
Acquisition

Set up a parallel bus

You can set parallel bus parameters.
1. To set up a parallel bus, select Bus Type Parallel.
See the online help for additional help setting up a bus.
2. To add sources to the bus, click the Add Sources Select button.
3. Click the button for each channel you want in the bus. The order of the parallel bus is determined by the order in which you select the channels.
4. Click the OK button.
5. To move a channel within the bus, scroll
to and select the channel you want to move, and then click the up or down arrow buttons.
6. To remove a channel within the bus, scroll to and select the channel you want to remove, and then click the Remove button.
7. To set channel thresholds, click the Thresholds Setup button.
DPO70000SX Series User 71
Acquisition
8. For each threshold you want to set, double click the channel Threshold, and enter the threshold using the keypad or multipurpose knobs on the optional Auxiliary Front Panel.
9. When the desired thresholds have been set, click the OK button.
10. To set the clock source and polarity, double click the item and select from the displayed list. To set the clock level, double click the Clock Level, and enter the level using the keypad or multipurpose knobs on the optional Auxiliary Front Panel.
72 DPO70000SX Series User
Acquisition

Set up bus display

You can set the bus style and decoding.
1. Select Vertical > Bus Setup.
2. Select the Display tab.
3. To select the bus, scroll to display the
bus, and then select the bus.
4. Select the decoding from the bus decoding list.
See the online help for information on the Results Table.
5. Click a + box to expand (display additional) the views, or click a - box to collapse (remove) views.
6. Bus views may also be added or removed from the display by clicking on the + box or - box in the display. Available views may include Transaction/transport, link/framing/packets, physical/bytes/ symbols, and bit level.
DPO70000SX Series User 73
Acquisition
7. If using a symbol table, check Use Symbol File. Click Browse and browse
to the symbol table file.
For some buses other settings are available. Set the other settings as appropriate for your bus.
See the online help for additional help setting up a bus.

Using FastFrame mode

FastFrame allows you to capture many trigger events as single records in a larger record, and then view and measure each record individually. Time Stamps display the absolute trigger time for a specific frame and the relative time between triggers of two specified frames.
1. Use AUTOSET to set the horizontal, vertical, and trigger controls or set the controls manually.
2. Select Horiz/Acq > FastFrame Setup....
74 DPO70000SX Series User
Acquisition
3. Push FastFrame to toggle it On.
4. Select Frame Size and # of Events Frames. Then use the multipurpose
knobs to set each one. Number of frames represents the number of trigger events that are captured. Frame size is the number of samples that are stored with each trigger event (or frame). The number of frames is reduced if there is insufficient memory to store all the records. The shorter the frame length, the more frames you can acquire.
5. Use the Frame Viewing controls to select the frame that you want to view.
6. To view multiple frames superimposed over each other, select Overlay.
7. Use the Time Stamps controls to select the source for, and frame number of, the reference frame. The reference frame is the starting point when measuring the relative time between two frames.
Quick tips
Use FastFrame when you want to preserve the data associated with each trigger event for further analysis or visual inspections.
Use FastFrame when you want to capture multiple events that have long dead times between them that are of no interest to you.
Multiple frames are best viewed with the Normal, Green, or Gray color palettes because the dark blue selected frame may be hard to distinguish if you use Temp or Spectral.
DPO70000SX Series User 75
Acquisition
If you select an Average Summary Frame, you can set Math functions (Averaging) to apply to All Frames or have it apply to Only the Summary Frame. If the summary data is of primary interest, selecting Summary Frame Only can substantially improve system throughput.

Using FastFrame frame finder

Frame finder allows you to find a FastFrame frame that is different from other frames.
1. Select the source of the FastFrame frames.
2. Set the Search Range by entering the From Frame # and the To Frame #.
3. Stop acquisitions by selecting Horiz/Acq
> Run/Stop and clicking Run/Stop or by pushing Run/Stop on the Auxiliary Front Panel.
4. Push Start to start the search.
76 DPO70000SX Series User
Acquisition
Frame Finder searches for and displays a different frame.
5. If the anomaly you are looking for is not in the displayed frame, push No. Frame Finder searches for another different frame.
6. If the anomaly you are looking for is in the displayed frame, push Yes.
7. If you are done searching, push Stop.
8. Select Horiz/Acq > Run/Stop and click Run/Stop or push Run/Stop on the
Auxiliary Front Panel to restart acquisitions.
DPO70000SX Series User 77
Acquisition
78 DPO70000SX Series User

Pinpoint triggers

The Pinpoint trigger system comes with advanced trigger types that are usable on both A and B triggers, and it can reset the trigger sequence if the B event does not occur after a specific number of events or a specific time. Pinpoint triggers support capture of events based on the most complex trigger event or sequence of trigger events.
This section contains concepts and procedures for using the trigger system.

Triggering concepts

Trigger event

The trigger event establishes the time-zero point in the waveform record. All waveform record data are located in time with respect to that point. The instrument continuously acquires and retains enough sample points to fill the pretrigger part of the waveform record. When a trigger event occurs, the instrument starts acquiring samples to build the posttrigger part of the waveform record (displayed after, or to the right of, the trigger event). Once a trigger is recognized, the instrument will not accept another trigger until the acquisition is complete and the holdoff time has expired.

Trigger modes

The trigger mode determines how the instrument behaves in the absence of a trigger event:
Normal trigger mode enables the instrument to acquire a waveform only when it is triggered. If no trigger occurs, the last waveform record acquired remains on the display. If no last waveform exists, no waveform is displayed.
Auto trigger mode enables the instrument to acquire a waveform even if a trigger does not occur. Auto mode uses a timer that starts after a trigger event occurs. If another trigger event is not detected before the timer times out, the instrument forces a trigger. The length of time it waits for a trigger event depends on the time base setting.
Auto mode, when forcing triggers in the absence of valid triggering events, does not synchronize the waveform on the display. The waveform will appear to roll across the screen. If valid triggers occur, the display will become stable.
In edge-trigger mode, force the instrument to trigger by pushing the Force Trigger button on the Trigger Setup control window.
Select the trigger mode in the Trig > Mode menu. For more information, see the instrument online help.

Trigger holdoff

Trigger holdoff can help stabilize triggering, by extending the period in which further trigger recognition is suppressed after initiating an acquisition. Such an extension can help the system skip over the remaining events of a recurring burst of events, and thus always trigger on the first event in each burst. Adjust holdoff to obtain stable triggering when the instrument is triggering on undesired trigger events.
Set the trigger holdoff in the Trig > Holdoff menu. For more information, see the instrument online help.
DPO70000SX Series User
79
Pinpoint triggers

Trigger coupling

Trigger coupling determines what part of the signal is passed to the trigger circuit. Edge triggering can use all available coupling types: AC, DC, Low Frequency Rejection, High Frequency Rejection, Noise Rejection, and RF coupling. All other trigger types use DC coupling only. Your instrument may not offer all coupling types.
Select the trigger coupling in the Trig > A Event (Main) Trigger Setup menu. For more information, see the instrument online help.

Horizontal position

Horizontal position defines where the trigger occurs on the waveform record. It lets you choose how much the instrument acquires before and after the trigger event. The part of the record that occurs before the trigger is the pretrigger part. The part that occurs after the trigger is the posttrigger part.
Pretrigger data is valuable when troubleshooting. For example, if you are trying to find the cause of an unwanted glitch in your test circuit, trigger on the glitch and make the pretrigger period large enough to capture data before the glitch. By analyzing what happens before the glitch, you may uncover information that helps you find the source of the glitch. Alternatively, to see what is happening in your system as a result of the trigger event, make the posttrigger period large enough to capture data after the trigger.

Slope and level

The slope control determines whether the instrument finds the trigger point on the rising or the falling edge of a signal. The level control determines where on that edge the trigger point occurs.

Delayed trigger system

You can trigger with the A (Main) trigger system alone or combine the A (Main) trigger with the B (Delayed) trigger to trigger on sequential events. When using sequential triggering, the A trigger event arms the trigger system, and the B trigger event triggers the instrument when the B trigger conditions are met. A and B triggers can (and typically do) have separate sources. The B trigger condition is based on a time delay or a specified number of events. See Using A (Main) and B (Delayed) triggers on page 85.
80 DPO70000SX Series User
Loading...