Tektronix DMM4020 DATASHEET

Digital Multimeters
Tektronix DMM4020 Data Sheet
Available Functions and Features
Volts, Ohms, and Amps Measure m ents
Diode and Continuity Testing
DMM4020
Features & Benets
Key Performance Specications
5.5 Digit Resolution
Basic VDC Accuracy of up to 0.015% (1 yr.)
200 mV to 1000 V Voltage Range, with up to 1 μV Resolution
200 to 100 MRange, with up to 1 mResolution
CAT I 1000 V, CAT II 600 V
Frequency Me
2×4 Ohms 4-wire Measurement Technique
Dedicated DC Leakage Current M easurements
Six Dedicated Buttons for Fast Access to Instrument Setups
Limit Compare Mode for Pass/Fail Testing
Connectivi
Front 2×4 Measurement Inputs
RS-232 on Rear Panel for Quick PC Connectivity
Includes USB to RS-232 Interface Adapter Cable
Includes National Instrument’s LabVIEW SignalExpress™ TE Limited Edition for Connecting Your Bench
3-year Warranty
asurements
ty
Data Sheet
Measurements with the Push of a Button
As the circuits in embedded system designs become more sophisticated, you must measure a multitude of different parameters to validate your design. The Tektronix DMM4020 5.5 digit bench multimeter offers a broad range of functions in one easy-to-use instrument.
Typical multimeter measurements – volts, ohms, and amps – are made with a basic VDC accuracy of up to 0.015%, ensuring you have the performance you need for your design. You can also use the DMM4 020 to measure frequency, and to perform continuity and diode tests. This allows you to replace yo instrument, saving bench space and cost.
Measure Nano
Measuring standby current in today’s energy-efcient designs requires looking at very low currents, often in t he microamp or nanoamp range. Using a traditional multimeter for this measurement can lead to inaccurate results since they typically e mploy the shunt resistance technique for measuring current.
The DMM4020 offers an improved method for measuring low currents. By using a cur current measurements with 1 nA of resolution and with minimal loading affect on the circuit under test, giving results that reect real-world operation of the d evice.
ur counter, continuity tester, and DMM with one versatile
ampere Signals
rent-to-voltage op amp technique, the DMM4020 can make
Designed to Ma ke Your Work Easier
The DMM4020 multimeter is designed with the ease o f use and familiar operation you expect from Tektronix.
Limit compare mode on the DMM4020.
Setup Buttons for Your Common Measurements
With six setup buttons on the front panel, you can s ave the settings for your most common measurements. Simply set up the meter fo r a measurement, then press sh settings. Now each time you perform that measurement, you simply press the appropriate setup key.
ift followed b y a setup button to save the measurement
Dual Display
With the unique dual display, you can measure two different parameters of the same signal from one test connection .
Limit Compare
The limit compare mode provides pass/fail indicators to quickly show if a test passes or fails to help eliminate mistakes, especially for results that areclosetothelimit.
Simple and Accurate 4-wire Measurements
Patented split terminal jacks for the 2×4 ohms function allow you to perform 4-wire measurements using only two leads instead of four. Spe cial test lead accessories are available to enable you to establish the connection. You get excellent resolution and accuracy plus the convenience and ease of using a single pa
ir of leads.
Intuitive Operation
Dedicated front-panel buttons provide fast access to frequently used
ns and parameters, reducing setup time. You no longer need to
functio search through software menus to nd the function you need.
Easy Conn
The RS-232 port on the back panel can be used to conn ect to your PC. AUSBtoRS DMM4020 for connecting to your P C’s USB port.
ectivity to Your PC
-232 interface adapter cable is included standard with the
2 www.tektronix.com/dmm
Digital Multimeters — Tektronix DMM4020
SignalExpress acquiring data from Tektronix DMM4020 and DPO3052.
Connect Your Bench for Intelligent Debug
Easily capture, save, and analyze mea surement results from your multimeter with the special Tektronix Edition of National Instrument’s LabVIEW SignalExpress™ software. Every DMM4020 multimeter ships with a free copy of the Limited E dition version of SignalExpress for basic instrument control, data logging, and analysis. The optional Professional Edition offers over 200 built-in functions that provide additional signal processing, advanced analysis, sweep ing, limit testing, and user-dened step capabilities.
SignalExpress supports the range of Tektronix bench instruments* enabling you to connect your entire test bench. You can then access the
1
,
press used to export DMM4020 data into Excel.
SignalEx
feature-rich tools packed into each instrument from one intuitive software interface. This allows you to automate complex measurements requiring multiple instruments, log data for an extended period of time, time correlate data from multiple instruments, and easily capture and analyze your results,
your PC. Only Tektronix offers a connected test bench of intelligent
all from instruments to simplify and speed debug of your complex design.
Performance You Can Count On
In addition to industry-leading service and support, every DMM4020 multimeter comes backed with a three-year standard warranty.
*1NI LabVIEW SignalExpress supports the following Tektronix instruments: MSO/DPO4000/3000/2000
Series oscilloscopes, TDS3000C/2000B/1000B Series oscilloscopes, AFG3000 Series arbitrary/function generators, DMM4050/4040/4020 Series digital multimeters.
www.tektronix.com/dmm 3
Data Sheet
Characteristics
General Specications
Vol t a g e
Characteristic
100V Setting 120V Setting 220V Setting 240V Setting Frequency 47 Hz to 440 Hz Power Consumption 15 VA peak (10 W average)
Dimensions
Dimension
Height 88 3.46 Width 217 8.56 Depth 297 11.7
Weight kg lb.
Net 2.1 4.6
Display
Vacuum Fluorescent Display, segment
Environment
Characteristic
Temperature
Operating 0 °C to 50 °C Storage –40 °C to 70 °C Warm Up
Relative Humidity (noncondensing)
Operating Uncontrolled (<10 °C)
Storage <95% (–40 °C to 70 °C)
Altitude
Operating
Storage Vibration Safety Complies with IEC 61010-1:2001, ANSI/ISA 61010-1
EMC Des
Description
90 V to 110 V 108 V to 132 V 198 V to 242 V 216Vto264V
mm
Description
½ hour to full uncertainty specications
<90%(10°Cto28°C) <75%(28°Cto40°C) <45%(40°Cto50°C)
2,000 meters 12,000 meters Complies with MIL-PRF-28800F Class 3
(S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No. 61010.1:2004, CAT I 1000V / CAT II 600 V
ignedtocomplywithIEC
26-1:1997+A1:1998+A2:2000
613
in.
Trigg eri ng
Characteristic
Trigger Delay 400 ms External Trigger Delay <2 ms External Trigger Jitter <1 ms Trigger Input TTL Levels Trigger Output
Math Functio
Min/Max, relative, hold, compare, and dB functions
Description
5Vmax
ns
Electrical
Characteristic
Input Protection 1000 V all ranges Overrange 10% on the largest ranges of all functions except
Description
continuity and diode test
Remote Interfaces
RS-232C (RS-232 to USB Adapter cable included)
Warrant
Three years
y
4 www.tektronix.com/dmm
Electrical Specications
DC Voltage Specications
Specications are valid for 5½ digit mode and after at least a half-hour warm up.
Digital Multimeters — Tektronix DMM4020
Characteristic
Description
Maximum Input 1000 V on any range Common Mode
120 dB at 50 or 60 Hz 0.1% (1 kunbalance)
Rejection Normal Mode Rejection
80 dB at Slow Rate A/D Nonlinearity 15 ppm of range Input Bias Current <30 pA at 25 °C Settling Considerations Measurement settling times are affected by source
impedance, cable dielectric characteristics, and input
signal changes
Input Cha
200 mV 199.999 mV 1 μV10μV10μV 2 V 1.99999 V 10 μV100μV100μV 20 V 19.9999 V 100 μV1000μV 1000 μV 200 V 199.999 V 1 mV 10 mV 10 mV 1000 V 1000.00 V 10 mV 100 mV 100 mV
*2At som
racteristics
Full Scale
(5½ Digits)
e dual-display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 MΩ.
Slow
ResolutionRange
Medium Fast
Input Impedance
>10 GΩ*
>10 GΩ* 10 M±1% 10 M±1% 10 M±1%
Accuracy
Range
Uncertainty*
90 d ays 1 year
23 °C ±5 °C 23 °C ±5 °C
3
Temperature Coefcient/°C
Outside18-28°C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005 2 V 0.01 + 0.002 0.015 + 0.003 0.001 + 0.0005 20 V 0.01 + 0.003 0.015 + 0.004 0.0020 + 0.0005 200 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005 1000 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
*3Uncertainty given as ±(% of reading + % of range).
2
2
www.tektronix.com/dmm 5
Data Sheet
AC Voltage Specications
AC Voltage sp ec 1% to 5% of range and <50 kHz, add an additional error of 0.1% of range, and for 50 kHz to 100 kHz, add 0.13% of range.
ications are for AC sinewave signals >5% of range. For inputs from
Characteristic
Maximum Input 750 V Measurement Method
Description
or 1000 V peak or 8 × 107Volts-Hertz product
RMS
AC-coupled true RMS. Measures the AC component of
input with up to 1000 VDC bias on any range AC Filter Bandwidth Common Mode
20 Hz - 100 kHz
60 dB at 50 Hz or 60 Hz (1 kunbalance) Rejection
Maximum Crest Factor 3:1 at Full Scale Additional Crest Factor
Errors (<100 Hz)
Crest Factor 1-2, 0.05% of full scale
CrestFactor2-3,0.2%offullscale
Only applies for non-sinusoid signals
Input Characteristics
Full Sca
(5½ Digi
le
ts)
Slow
ResolutionRange
Medium Fast
200 mV 199.999 mV 1 μV10μV10μV 2 V 1.99999 V 10 μV100μV100μV 20 V 19.9999 V 100 μV1000μV 1000 μV 200 V 199.999 V 1 mV 10 mV 10 mV 750 V 750.00 V 10 mV 100 mV 100 mV
Accuracy
Range Frequ
200 mV
20 Hz -
45 Hz -
20 kH
z-100kHz
50 kH
2V
20 Hz
45 Hz
20 kH
Hz - 100 kHz
50 k
20 V
V
200
0V
75
*3Uncertainty given as ±(% of reading + % of range).
20 H
z-20kHz
45 H
Hz - 50 kHz
20 k
Hz - 100 kHz
50 k
20 H
Hz - 20 kHz
45
kHz - 50 kHz
20
kHz - 100 kHz
50
Hz - 45 Hz
20
Hz - 20 kHz
45
0kHz-50kHz
2
0kHz-100kHz
5
ency
45 Hz
20 kHz
z-50kHz
-45Hz
-20kHz
z-50kHz
z-45Hz
z-45Hz
Uncertainty*
s
90 day
23 °C ±5 °C 23 °C ±5 °C
.05
0.8 + 0
0.05
0.15 +
0.05
0.3 +
0.05
0.8 +
0.05
0.8 + +0.05
0.15
0.05
0.3 +
+0.05
0.8
+0.05
0.8
5 + 0.05
0.1 +0.05
0.3 +0.05
0.8 +0.05
0.8
15 + 0.05
0. 3 + 0.05
0. 8 + 0.05
0. 8 + 0.05
0.
15 + 0.05
0. .3 + 0.05
0
.8 + 0.05
0
3
1 year
0.9 + 0
0.2 + 0 +0.05
0.35
0.9 +
0.9 +
0.2 + +0.05
0.35 +0.05
0.9 +0.05
0.9 +0.05
0.2
5 + 0.05
0.3 +0.05
0.9 +0.05
0.9
2 + 0.05
0.
35 + 0.05
0. 9 + 0.05
0. 9 + 0.05
0. 2 + 0.05
0.
.35 + 0.05
0
.9 + 0.05
0
.05 .05
0.05
0.05
0.05
Input Impedance
1MΩ ±2% shunted by
<100 pf
Temperature Coefcient/°C
Outside 18 - 28 °C
0.005
0.01 +
0.005
0.01 + + 0.005
0.01
+0.01
0.05 + 0.005
0.01 + 0.005
0.01 + 0.005
0.01
5+0.01
0.0 1 + 0.005
0.0 1 + 0.005
0.0 1 + 0.005
0.0
5+0.01
0.0 1 + 0.005
0.0 01 + 0.005
0. 01 + 0.005
0.
05 + 0.01
0. 01 + 0.005
0. 01 + 0.005
0.
.01 + 0.005
0
.05 + 0.01
0
6 www.tektronix.com/dmm
Resistance
Specications REL is not used, add 0.2 for 2-wire resistance plus lead resistance.
are for 4-wire resistance function, or 2-wire resistance with REL. If
Digital Multimeters — Tektronix DMM4020
Characteristic
Measurement Method Max Lead Resistance
(4-wire ohms)
Description
Current source referenced to LO input 10% of range per lead for 200 ,2kΩ ranges. 1 kΩ
per lead on all other ranges
Input Protection 1000 V on all ranges
Input Characteristics
Full Scale
(5½ Digits)
Slow
200 199.999 0.001 0.01 0.01 2k 1.99999 kΩ 0.01 Ω 0.1 Ω 0.1 Ω 20 k 19.9999 k 0.1 1 Ω 1 Ω 200 k 199.999 k 1 Ω 10 10 2M 1.99999 MΩ 10 Ω 100 Ω 100 Ω 20 M 19.9999 M 100 1k 1k
ResolutionRange
Medium Fast
Current Source
0.8 mA
0.8 mA
0.08 mA
0.008 mA
0.9 μA
0.16 μA
100 M 100.000 M 1k 10 kΩ 10 kΩ 0.16 μA||10M
Accuracy
Range
200 2k 20 k 200 k 2M 20 M 100 M
*3Uncertainty given as ±(% of reading + % of range).
Uncertai
90 d ays 1 year
23 °C ±5 °
C
0.02 + 0.004 0.03 + 0.004 0.003 + 0.0006
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
0.03 + 0.003 0.04 + 0.004 0.004 + 0.0005
0.2 + 0.003 0.25 + 0.003 0.01 + 0.0005
1.5 + 0.004 1.75 + 0.004 0.2 + 0.0005
nty*
3
23 °C ±5 °
Temperat
C
ure Coefcient/°C
Outside 1
8-28°C
www.tektronix.com/dmm 7
Data Sheet
DC Current
Characteristic
Input Protection
Description
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V fuses
Shunt Resist
ance
0.01 for 2 A a 1 for 20 mA an Burden volta
nd 10 A ranges
d200mA
ge <5 mV for 200 μA and 2 mA range
Input Characteristics
Full Scale
(5½ Digits)
Slow
ResolutionRange
Medium Fast
Burden Voltage
200 μA 199.999 μA0.001μA0.01μA0.01μA<5mV 2 mA 1999.99 μA0.01μA0.1μA0.1μA<5mV 20 mA 19.9999 mA 0.1 μA1μA1μA <0.05 V 200 mA 19.9999 mA 1 μA10μA10μA<0.5V 2 A 1.99999 A 10 μA100μA100μA<0.1V 10 A 10.0000 A 100 μA1mA 1mA <0.5V
Accuracy
Range
Uncertainty*
90 d ays 1 year
23 °C ±5 °C 23 °C ±5 °C
3
Temperature Coefcient/°C
Outside18-28°C
200 μA 0.02 + 0.005 0.03 + 0.005 0.003 + 0.001 2 mA 0.015 + 0.005 0.02 + 0.005 0.002 + 0.001 20 mA 0.03 + 0.02 0.04 + 0.02 0.005 + 0.001 200 mA 0.02 + 0.005 0.03 + 0.008 0.005 + 0.001 2 A 0.05 + 0.02 0.08 + 0.02 0.008 + 0.001 10 A 0.18 + 0.01 0.2 + 0.01 0.008 + 0.001
*3Uncertainty given as ±(% of reading + % of range).
8 www.tektronix.com/dmm
AC Current
The following A greater than 5% of range. For inputs from 1% to 5% of range, add an additional error of 0.1% of range.
C current specications are for sinusoidal signals with amplitudes
Digital Multimeters — Tektronix DMM4020
Characteristic
Input Protection
Description
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V fuses
Measurement Method
AC-coupled true RMS
Shunt Resistance 0.01 for 2 A and 10 A ranges
1 for 20 mA and 200 mA
AC Filter Bandwidth
20 Hz - 100 kHz Maximum Crest Factor 3:1 at Full Scale Additional Crest Factor
Errors (<100 Hz)
Crest Factor 1-2, 0.05% of full scale
CrestFactor2-3,0.2%offullscale
Only applies to non-sinusoid signals
Input Characteristics
Full Scale
(5½ Digits)
Slow
ResolutionRange
Medium Fast
Burden Voltage
20 mA 19.9999 mA 0.1 μA1μA1μA <0.05 V 200 mA 199.999 mA 1 μA10μA10μA<0.5V 2 A 1.99999 A 10 μA100μA100μA<0.1V 10 A 10.0000 A 100 μA1mA 1mA <0.5V
Accuracy
Range Frequency
Uncertainty*
90 days 1 year
23 °C ±5 °C 23 °C ±5 °C
3
Temperature Coefcient/°C
Outside 18 - 28 °C
20 Hz - 45 Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.00520 mA 45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005 20 Hz - 45 Hz 0.8 + 0.05 1 + 0.06 0.015 + 0.005200 mA 45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005 20 Hz - 45 Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.0052A 45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005 20 Hz - 45 Hz 1 + 0.1 1.25 + 0.12 0.015 + 0.00510 A 45 Hz - 2 kHz 1 + 0.1 0.5 + 0.12 0.015 + 0.005
*3Uncertainty given as ±(% of reading + % of range).
www.tektronix.com/dmm 9
Data Sheet
Frequency
Characteristic Description
Gate Time
Measurement Method
Settling Considerations
Measurement Considerations
Input Characteristics
Range Frequency
100mVto750V*
*4Input >100 mV.
5
Limitedto8×107VHz.
*
131 ms
AC-coupled input using the AC voltage measurement function
When measuring frequency after a DC offset voltage change, errors may occur. For the most accurate measurement, wait up to 1 second to allow input-blocking RC time constant to settle
To minimize measurement errors, shield inputs from external noise when measuring low-voltage, low-frequency signals
4, 5
20 Hz - 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz - 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz - 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
200 kHz - 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
Uncertainty
90 days 1 year
23 °C ±5 °C 23 °C ±5 °C
Temperature Coefcient/°C
Outside 18 - 28 °C
Continuity
Characteristic
Continuity Threshold 20 Test Current Response Time Rate Fast Maximum Reading Resolution
Description
1mA
100 S/s with audible tone
199.99
0.01
Diode Test
Characteristic
Response Time Rate Fast Maximum Reading 1.9999 V Resolution 0.1 mV
Description
100 S/s with audible tone
10 www.tektronix.com/dmm
Digital Multimeters — Tektronix DMM4020
Ordering Information
Models
Model Description
DMM4020 5.5 Digit Multimeter
DMM4020 Includes: Meter, TL710 test leads, line cord, spare line fuse, statement of cal practices, Warranty statement, Safety and Installation Guide, Connectivity Installation Manual, CD-ROM with user manual (English, French, Italian, German, Spanish, Simplied Chinese, Traditional Chinese, Korean, Russian, Japanese), RS-232 to USB Adapter Cable, National Instruments LabVIEW SignalExpress™ Tektronix Edition, Limited Edition Software.
Please specify power plug when ordering.
Instrument
Power Plug Options
Option
Opt. A0 Opt. A1 Opt. A2 Opt. A3 Opt. A5 Switzerland Opt. A6 Opt. A10 China Opt. A11 Opt. E1
Options
Description
North America
Universal Euro
United Kingdom
Australia
Japan
India
Euro and UK power cords
Service Options*
Option
6
Description
Opt. CA1 Provides a single calibration event or coverage for the
designated calibration interval, whichever comes rst Opt. C3 Calibration Service 3 Years Opt. C5 Calibration Service 5 Years Opt. D1 Calibration Data Report Opt. R5 Repair Service 5 Years (including warranty)
*6Test Leads and accessories are not covered by the DMM warranty and Service Offerings. Refer to the
datasheet of each Test Lead and accessory model for its unique warranty and calibration terms.
Recommended Accessories and Software
Accessory Description
Calibration Manual 196-3520-xx TL705 TL725
077-0365-xx
Premium Test Leads (replacement/spare for TL710)
2×4 Wire Ohm 1000 V Precision Test Lead
2×4 Wire Ohm SMD Test Tweezers AC4000 Soft Transit Case HCTEK4321 Hard Carrying Case Y8846S Single Rackmount Kit Y8846D Dual Rackmount Kit 013-0369-xx
Calibration Fixture 4-terminal short SIGEXPTE NI LabVIEW SignalExpress Tektronix Edition Software –
Full Version
Product(s) are manufactured in ISO registered facilities.
Product(s) complies with IEEE Standard488.1-1987 and RS-232C.
www.tektronix.com/dmm 11
Data Sheet
Contact Tektronix:
ASEAN / Australa
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Central East Eu
Mexico, Central/South America & Caribbean 52 (55) 54247900
rope, Ukraine, and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Luxembourg +44 (0) 1344 392400
Middle East,
For other areas contact Tektronix, Inc at: 1 (503) 627-7111
Asia, and North Africa +41 52 675 3777
The N etherlands 090 02 021797
People’s Rep
United Kingdom & Ireland +44 (0) 1344 392400
ublic of China 86 (10) 6235 1230
Republic of
sia (65) 6356 3900
Austria +41 52 675 3777
Belgium 07 81 601
Brazil +55 (11) 3759-7627
Canada 1 (800) 661-5625
Denmark +4580881401
Finland +41526
France +33(0)169868181
Germany +49 (221) 94 77 400
Hong Kong (852
India (91) 80-42922600
Italy +39 (02) 25086 1
Japan 81 (3) 67
Poland +41 52 675 3777
Portugal 80 08 12370
Korea 82 (2) 6917-5000
Russia & CIS +7 (495) 7484900
South Africa +27 11 206 8360
Spain (+34)
Sweden 020 08 80371
Switzerland +41 52 675 3777
Tai wa n 886 (
USA 1 (800) 426-2200
75 3777
) 2585-6688
14-3010
Norway 800 16098
901 988 054
2) 2722-9622
Updated 5 August 2009
66
www.tektronix.com/dmm
For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com
t © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents,
Copyrigh
d pending. Information in this publication supersedes that in all previously published material.
issued an
tion and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of
Specica
x, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks
Tek t ro n i
espective companies.
of their r
16 Oct 2009 3MW-24431-1
Loading...