Tektronix DC 503A Technical Manual

TM 9-6625-474-14&P-3
TECHNICAL MANUAL
OPERATOR, ORGANIZATIONAL,
DIRECT SUPPORT, AND GENERAL SUPPORT
MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS)
UNIVERSAL
TEKTRONIX,
COUNTER/TIMER,
MODEL DC 503A
(NSN 6625-01-114-4890)
DEPARTMENT OF THE ARMY
26 DECEMBER 1984

WARNING

DANGEROUS VOLTAGE
is used in the operation of this equipment
DEATH ON CONTACT
TM 9-6625-474-14&P-3
personnel fail to observe safety precautions
t if
. .
Never work
may resul
on
electronic equipment unless there is another person nearby who is familiar with the operation and hazards of the equipment and who is competent in administering first aid. When the technician is aided by operators, he must warn them about dangerous areas.
Whenever possible, the power supply to the equipment must be shut off before beginning work on the equipment. Take particular care to ground every capacitor likely to hold a dangerous potential. When working inside the equipment, after the power has been turned off, always ground every part before touching it.
Be careful not to contact high-voltage connections when installing or operating this equipment. Whenever the nature of the operation permits, keep one hand away from the equipment to reduce the
hazard of current flowing through vital organs of the body.
WARNING
Do not be misled by the term “low voltage.”
Potentials as low as 50 volts may cause death under adverse
conditions.
COMMON
any is present on each -
and probe ground straps are electrically connected. Herefore, an elevated reference applied to
as indicated by the yellow warning bands under the probe retractable hook tips.
For Artificial Respiration, refer to FM 21-11,
Power Source
This product is intended to operate in a power module connected to a power source that will not apply more than 250 volts rms between the supply conductors or between either supply conductor and ground. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
a/(b blank)
Copyright 1980 Tektronix, Inc. All rights reserved.
Reproduced by permission of copyright owner.
TM 9-6625-474-14&P-3
TECHNICAL MANUAL
DEPARTMENT OF THE ARMY
No. 9-6625-474-14&P-3
Washington, D. C.,
OPERATOR, ORGANIZATIONAL, DIRECT SUPPORT, AND GENERAL SUPPORT MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS)
FOR
UNIVERSAL COUNTER/TIMER,
TEKTRONIX, MODEL DC 503A
(NSN 6625-01-114-4890)
REPORTING OF ERRORS
You can help improve this manual. If you find any mistakes or if you know of a way to improve the procedures, please let us know. Mail your Ietter or DA Form 2028
(Recommended Changes to Publications and Blank Forms), direct to: Command­er, US Army Missile Command, ATTN: DRSMI-SNPM, Redstone Arsenal, AL
35898-5238. A reply will be furnished to you.
TABLE OF CONTENTS
LIST OF ILLUSTRATIONS
LIST OF TABLES . . . . . . . . . . . . . . . . . . .
SECTION 0
SECTION 1. SECTION 2.
SECTION 3. SECTION 4.
SECTION 5.
GENERAL NFORMATION Scope indexes of Publications Forms, Records, and Reports
Reporting Equipment Improvement Recommendations (EIRs) Administrative Storage Destruction of Army Electronics Materiel.
SPECIFICATION . . . . . . . . . . . . . . . . . . .
THEORY OF OPERATION.
MAINTENANCE
HEADQUARTERS
26 December 1984
Page
iii
IV
0-1 0-1 0-1 0-1 0-1 0-1 0-1
1-1 2-1
2-1 2-2 2-4 2-4 2-5 2-7
3-1 4-1
4-1 4-8
5-1
This manual is, in part, authenticated manufacturer’s commercial literature. Recommended Spare Parts List has been added to supplement the commercial
literature. The format of this manual has not been structured to consider levels of maintenance.
TM 9-6625-474-14&P-3
SECTION 6. SECTION 7. SECTION 8. SECTION 9. APPENDIX A. APPENDIX B. APPENDIX C. APPENDIX D.
OPTIONS REPLACEABLE ELECTRICAL PARTS DIAGRAMS AND CIRCUIT BOARD ILLUSTRATIONS
REPLACEABLE MECHANICAL PARTS REFERENCES MAINTENANCE ALLOCATION CHART (MAC) RECOMMENDED SPARE PARTS LIST MANUAL CHANGE INFORMATION
TABLE OF CONTENTS (CONT)
Page
6-1 7-1 8-1 9-1 A-1 B-1 C-1 D-1
ii
LIST OF ILLUSTRATIONS
TM 9-6625-474-14&P-3
Fig. No.
0-1 2-1 2-2 2-3
2-4 2-5
2-6
3-1 4-1
4-2 4-3 4-4
5-1 5-2 5-3 5-4 5-5 5-6 5-7 5-8 5-9 5-10
Title
DC 503A Universal Counter/Timer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Plug-ln lnstallation/Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Controls and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Triggering Circuit Response to Improper (A) and
Proper (B) Level Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Representation of lnterval Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . .
Typical CHAand CH B Level out Voltage Settings for
Various Time interval Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration ofCH A Events Counted from Portion of
CH A Signal Pulses during the Counter Gate Open
Time (Controlled by the CH B Signal) . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Typical DC 503A Timing Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Check Set-Up for the High Frequency Sensitivity
Using X1 and X5 Attenuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Check Set-Up for Low Frequency ac and dc Sensitivity . . . . . . . . . . . . . . .
Check Set-Up for Minimum Pulse Width Signals . . . . . . . . . . . . . . . . . . . . .
Check Set-Up for Trigger Level Range (3.5V) and
Accuracy
(Y20mVf0.59’0
of Reading) . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Typical Square Pin Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Exploded View of Circuit Board Pin and Ferrule . . . . . . . . . . . . . . . . . . . . . .
Dual Entry Circuit Board Pin Socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Bottom Entry Circuit Board Pin Socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Orientation and Disassembly of Multipin Connectors . . . . . . . . . . . . . . . . .
Rear Frame Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Circuit Board Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pushbutton Switch Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Rear Interface Connector Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Rear Interface Timing for a Display of 1079.0674 . . . . . . . . . . . . . . . . . . .
Page
0-2 2-1 2-3
2-5 2-9
2-10
2-11
3-5
4-4 4-6 4-8
4-12 5-3 5-3 5-4 5-4 5-4 5-5 5-5 5-5 5-7 5-8
iii
TM 9-6625-474-14&P-3
LIST OF TABLES
Table No.
1-1 1-2 1-3 1-4
2-1 2-2
2-3
2-4 2-5 3-1 4-1 5-1
Title
Electrical Characteristics . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Environmental . . . ....... . . . . . . . . . . . . . . . . . . . . . . . . . . .
Physical Characteristics ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency A Display Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Page
1-2 1-7 1-7 1-8
2-5
Period B, Time A-B, Width B (Timing Mode)
Display Check
2-6
Period B, Time A- B, Width B (Averaging Mode)
Display Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ratio A/B and Events A During B Display Check . . . . . . . . . . . . . . . . . . . . .
Gate Time vs Measurement Resolution . . . . . . . . . . . . . . . . . . . . . .
PROM Selection Code . . . . . . . . . . . . . . . . . . . . . . . . . . . .
List of Test Equipment Requirements . . . . . . . . . . . . . . . . . . . .
Relative Susceptibility to Static Discharge Damage
2-6 2-6 2-8
3-10
4-2 5-1
iv

SECTION 0

GENERAL INFORMATION
TM 9-6625-474-14&P-3
0-1. Scope. This manual contains instructions for
the operator, organizational, direct support, and gen­eral support maintenance of and calibration proce­dures for Tektronix Universal Counter/Timer, Model DC 503A. Throughout this manual, Tektronix Universal Counter/Timer, Model DC 503A is referred to as the DC 503A.
0-2. Indexes of publications. a. DA Pam 310-4.
Refer to the latest issue of DA Pam 310-4 to deter­mine whether there are new editions, changes, or additional publications pertaining to Tektronix Univer­sal Counter/Timer, Model DC 503A.
b. DA Pam 310-7. Refer to the latest issue of DA
Pam 310-7 to determine whether there are modifica­tion work orders (MWO’S) pertaining to Tektronix Universal Counter/Timer, Model DC 503A.
0-3. Forms, Records, and Reports. Department of Army forms and procedures used for equipment main­tenance and calibration are those prescribed by TM 38-750, The Army Maintenance Management System. Accidents involving injury to personnel or damage to
materiel will be reported on DA Form 285, Accident
Report, in accordance with AR 385-40.
0-4. Reporting Equipment Improvement Recom­mendations (EIR). If your DC 503A needs im-
provement, let us know. Send us an EIR. You, the user, are the only one who can tell us what you don’t like about your equipment. Let us know why you don’t like the design. Tell us why a procedure is hard to perform. Put it on an SF 368 (Quality Deficiency Report). Mail it to Commander, U.S. Army Missile
Command, ATTN: DRSMI-QMD, Redstone Arsenal, AL 35898-5290. We’ll send you a reply.
0-5. Administrative Storage. To prepare the Tek­tronix Universal Counter/Timer, Model DC 503A for placement into and removal from administrative stor­age, refer to Section 3, Chapter 4, AR 750-25-1, Maintenance of Equipment and Supplies. Temporary storage should be accomplished in accordance with
TB 750-25-1, Section 2, Maintenance of Supplies and
Equipment.
0-6. Destruction of Army Electronics
Materiel. Destruction of Tektronix Universal Counter/
Timer, Model DC 503A to prevent enemy use shall be
in accordance with TM 750-244-2.
0-1
TM 9-6625-474-14&P-3
0-2
Fig. 0-1. DC 503A Universal Counter/Timer

SECTION 1

SPECIFICATION
TM 9-6625-474-14&P-3
Instrument Description
The DC 503A Universal Counter/Timer is designed to
operate in a TM 500-series power module.
The instrument has two input channels, CH A and CH B, each with 125 MHz capability. Each channel has separate triggering level, triggering slope, attenuator, and coupling mode controls.
The DC 503A has eight measurement functions: FRE­QUENCY A, PERIOD B, WIDTH B, TIME A - B, RATlO A/B, EVENTS A DURING B, TOTALIZE A, and TIME MANUAL. All of the modes except FREQUENCY A, TOTALIZE A, and TIME MANUAL have the capability of averaging the selected measurement over a range of 1 to
10*
times the input signal. The signals to be counted or measured can be applied via front panel bnc connectors or
through the rear Interface.
The triggering level for each channel can be monitored
via the front panel or the rear i nterface connections. The buffered voltage available at these connectors cor­responds to the trigger levels set by the front panel controls.
The output of the internal signal shaping circuits can
also be monitored via front panel connectors. These shaped signal outputs are useful in setting the triggering points on complex waveforms.
Instrument Options
Option 01 replaces
(clock) circuit with temperature controlled oven oscillator for increased ac­curacy and stability.
Standard Accessories
1 Instruction Manual 1 Cable assembly, bnc-to-tip jack.
Refer to the tabbed Accessories page at the rear of this manual for more information.
Performance Conditions
The limits stated in the Performance Requirements
columns of the following tables are valid only if the
DC 503A has been calibrated at an ambient temperature
between +20°C and +30°C and is operating at an ambient temperature between stated.
the internal 10 MHz time base a self-contained proportional
NOTE
O“C
and
+50”C,
unless otherwise
Measurement results are displayed in an eight digit LED readout. The decimal point is automatically position­ed and leading zeros are blanked. Single annunciators (LEDs) are used to indicate register overflow, active gating interval, and the frequency or time units associated with
the measurement being made.
The DC 503A can be equipped with an optional, oven
controlled, 10 MHz crystal oscillator to obtain a highly stable and precise internal time base. Both the optional oscillator and the standard 10 MHz time bases provide
100 ns single shot resolution.
Information given in the Supplemental Information and
Description columns of the following tables is provided for
user information only and should not be interperted as
Performance Check requirements.
The DC 503A must be operated or stored in an environment whose limits are described under En­vironmental Characteristics.
Allow at least 20 minutes warm-up time for operation to specified accuracy, 60 minutes after storage in a high humidity environment.
1-1
TM 9-6625-474-14&P-3

Table 1-1

1-2
Table 1-1
TM 9-6625-474-14&P-3
1-3
TM 9-6625-474-14&P-3
Table 1-1
1-4
Table 1-1
TM 9-6625-474-14&P-3
1-5
TM 9-6625-474-14&P-3
Table 1-1
1-6
Table 1-1

Table 1-2

TM 9-6625-474-14&P-3

Table 1-3

1-7
TM 9-6625-474-14&P-3
Table 1-3

Table 1-4

1-8
TM 9-6625-474-14&P-3

SECTION 2

OPERATING INSTRUCTIONS
INTRODUCTION
This section of the manual provides installation and
removal instructions and the operating information re-
quired to obtain the most effective performance from the instrument. Also included is the function of all front panel controls and a general description of the operating modes,
which also describes procedures for making basic
measurements.
INSTALLATION AND REMOVAL
The DC 503A is calibrated and ready to use when
received. It operates in one compartment of a TM 500-
Series power module. Refer to the power module instruc-
tion manual for line voltage requirements and power
module operation.
To prevent damage to the DC 503A, turn the power module off before installation or removal of the instrument from the mainframe. Do not use ex­cessive force to install or remove.
press firmly to seat the circuit board edge connector in the power module interconnecting jack. Apply power to the DC 503A by operating the power switch on the power
module.
To remove the DC 503A from the power module, pull the release latch (located in the lower left corner) until the interconnecting jack disengages. The DC 503A will now slide straight out.
Check to see that the plastic barriers on the inter-
connecting jack of the selected power module compart-
ment match the cutouts in the DC 503A circuit board edge connector. If they do not match, do not insert the instrument until the reason is investigated. When the units are properly matched, align the DC 503A chassis with the
upper and lower guides of the selected compartment (see
Fig. 2-1). Insert the DC 503A into the compartment and
Fig. 2-1. Plug-in installation/removal.
2-1
TM 9-6625-474-14&P-3
CONTROLS AND CONNECTORS
Even though the DC 503A is fully calibrated and ready
to use, the functions and actions of the controls and
connectors should be reviewed before attempting to use it.
With the exception of the TOTALIZEA/TIME MANUAL
jumper, which is described in the maintenance section, all
controls for operation of the DC 503A are located on the front panel. A brief functional description of these controls
follows (refer to Fig. 2-2).
NOTE
Because the Channel A and Channel B controls are identical, only Channel A will be described.
DISPLAY AND UNIT INDICATORS
DISPLAY READOUT: eight-digit, seven segment
LED readout with automatically positioned decimal point.
OVERFLOW: when illuminated indicates register
overflow.
MODE SELECTION AND
CONTROL FUNCTIONS
FUNCTION: selects the measurement, events, or
time counting modes for the counter.
NOTE
The TOTALIZE A/TIME MANUAL position is an
"either/or" MANUAL is selected and set by positioning an Internal jumper. Placement of this jumper is discuss­ed in the maintenance section.
Unless you are qualified to do so, refer positioning of this jumper to qualified personnel.
AVGS/TIMING: depending on the position of the
FUNCTION switch, this switch selects the clock rate which will be counted or the number of measurements to be averaged.
function.
TOTALIZE A or TIME
GATE: indicates the state of the main gate. When lit,
the main gate is open (the DC 503A is in the process
of making a measurement). When the light is off, the gate is closed.
GHz/nSEC: when illuminated, indicates the dis­played number is gigahertz (GHz) in FREQ A mode or nanoseconds (nSEC) in a time mode.
MHz/uSEC:
played number is Megahertz (MHz) in FREQ A mode or microseconds
kHz/mSEC: when illuminated, indicates the dis­played number is kilohertz (kHz) in FREQ A mode or milliseconds (m SEC) in a time mode.
Hz/SEC: when illuminated, indicates the displayed number is Hertz (Hz) in FREQ A mode or seconds (SEC) in a time mode.
when illuminated, indicates the dis-
(HSEC)
in a time mode.
2-2
DISPLAY TIME: sets the length of time the reading will be displayed after the count is made and before the next measurement is taken, Display time can be varied from about 0.1 second, fully counter­clockwise (ccw), to about 10 seconds fully clockwise (cw). The HOLD position provides continuous dis­play until reset by pushing the RESET button,
RESET: momentary switch resets the count to zero when operating in the TOTALIZE A mode. Also acts as a master reset, ensuring that the readout has been cleared before the next measurement. Provides a check of all display LED’s; when pressed, a row of 8’s will be displayed in the readout window.
START/STOP: push-push switch acts as a manual gate when the FUNCTION switch is in the TOTALIZE A/TIME MANUAL position. Button in Starts the measurement interval gate; button out
terminates the gate.
TM 9-6625-474-14&P-3
Fig. 2-2. Controls and connectors.
2-3
TM 9-6625-474-14&P-3
CHANNEL A INPUT AND LEVEL
FUNCTIONS
CH A INPUT: bnc connector for Channel A signal
input. Input impedance is 1 shunted by ap­proximately 20 pF.
LEVEL: selects the amplitude point on the positive or negative slope of the input signal at which the
triggering window is placed.
SLOPE: push-push switch selects the slope of the input signal on which triggering will occur, Button out selects plus (+) slope; button in minus (–) slope.
ATTEN: push-push switch selects X1 (button out) or X5 (button in) attenuation of the input signal.
COUPL: push-push switch selects DC (button out) or AC (button in) coupling of the input signal to the attenuator circuit.
SOURCE: push-push switch selects the source of the input signal. Button out, EXT, selects the front panel connector as a signal source. Button in, INT, routes the input signal to the counter via the rear interface connections.
SHAPED OUT A: provides a shaped output signal derived from the output of the Channel A signal shaper circuitry.
SHAPED OUT GND: common connector for Channel A shaped output signals.
TRIG LEVEL A: pin jack permits monitoring of the Channel A triggering voltage level.
RELEASE LATCH: pull to disengage and remove DC 503A from the power module.
INPUT CONSIDERATIONS
Input Sources
NOTE
Maximum input voltage limited to 200 V peak.
The SOURCE switch for each channel selects either the front panel bnc connector (external), or the rear interface connector (internal) pins. The external inputs present impedances of approximately 1 paralleled by about 27 pF. The internal input circuits present nominal 50 impedances to match typical coaxial cable signal connec­tions.
Input Coupling
Front panel pushbuttons select ac (capacitive) or dc (direct) coupling for the input signal of each channel. This coupling takes place before the signals are passed into the
attenuator circuits.
Attenuators and Maximum Input Volts
For either attenuation factor, X1 or X5, the maximum
safe input voltatage that can be applied to the front panel
bnc connectors is 200 V (peak) from dc to 50 kHz. At
frequencies above 50 kHz, the maximum safe peak-to-
peak input voltage tothe front panel bnc connectors must
be calculated (see Specification section). The maximum safe input voltage to the rear interface input connectors is equal to or less than 4 V (dc plus peak ac) from dc to 50 MHz.
Sensitivity and Frequency Range
CH A and CH B will respond to a signal amplitude of 20 mV rms sinewave, times attenuation, to lOO MHzandto a sinewave of 35 mV rms, times attenuation, to 125 MHz.
[)
Depending on the coupling mode selected, the low
frequency limit for each channel is either zero (dc
coupled) or 10 Hz (ac coupled).
Slope and Level
The SLOPE pushbuttons for each channel determine whether the trigger circuits will respond to the negative or positive transition of the input signal.
Refer to Fig. 2-3. The LEVEL control for each channel allows the operator to move the hysteresis window of the trigger circuit to an optimum level on the input signal to
ensure stable triggering The LEVEL control adjusts over
+3.5 V, times attenuation, of the input signal. This level
can bemoniiored atthefront panel TRIG LEVEL pin jacks.
2-4
TM 9-6625-474-14&P-3
Fig. 2-3. Triggering circuit response to improper (A) and proper (B) level settings.
OPERATORS FAMILIARIZATION
PREPARATION
Turn on the power module to apply power to the DC 503A. One or more characters in the display should be visible. Allow twenty minutes warm-up time for operation
to specified accuracy.
DISPLAY TESTS
With no signal applied, test the DC 503A readout
displays and switching logic. The following checks will
test most of the major circuits of the counter and ensure its
readiness to make measurements. If any malfunctions are encountered, refer personnel.
Readout Segment
Press the RESET segments of each digit. A row of 8's should be displayed. This check of the display devices can be done at anytime.
the condition to qualified service
Test
button to check the seven character

Table 2-1

FREQUENCY A DISPLAY CHECK
With the DISPLAY TIME control in the fully counterclockwise position, observe that the GATE in­dicator flashes rapidly for short gate times and more slowly for longer gate times. Using a short gate time (100 ms), rotate the DISPLAY TIME control slowly clockwise. Observe that the GATE indicator stays off for a longer and longer time, until the control clicks into the HOLD (detent) position, holding off the gate indefinitely. Return counterclockwise position.
the DISPLAY TIME control to the
Frequency A Displays
Set the FUNCTION switch to FREQUENCY A. With the AVGS/TIMING switch, select a gate time of 100 ns. Check the decimal point location, leading zero suppression, and units indicators according to Table 2-1.
Period B, Width B, and Time A - B Displays
Timing Mode. Set the FUNCTION switch to PERIOD B
in the blue area of the front panel and the AVG/TIMING switch to 100 ns. Observe the correct readout displays as shown in Table 2-2.
2-5
TM 9-6625-474-14&P-3

Table 2-2

PERIOD B, TIME A - B, WIDTH B
(TIMING MODE) DISPLAY CHECK
Set the FUNCTION switch to WIDTH B in the blue area
of the front panel while retaining the setting of the
AVG/TIMING switch; observe the correct readout display.
Set the FUNCTION switch to TIME A - B in the blue
area of the front panel while retaining the setting of the
AVG/TIMING switch; observe the correct readout display.

Table 2-4

RATIO A/B AND EVENTS A
DURING B DISPLAY CHECK
Time Manual Displays
Verify that the jumper located on the Auxiliary Circuit Board is in the TIME MANUAL position. Set the FUNC­TION switch to the TIME MANUAL Position and the AVGS/TIMING switch to 1 sec.
Averaging Mode. Repeat the preceding checks for these functions in the dark grey area of the front panel. Observe the correct readout display for each switch setting as shown in Table 2-3.

Table 2-3

PERIOD B, TIME A - B, WIDTH B
(AVERAGING MODE) DISPLAY CHECK
Events A During B and Ratio A/B Displays
Set the FUNCTION switch to EVENTS A DURING B and the AVGS/TIMING switch to 1. Check the readout displays according to Table 2-4.
Set the FUNCTION switch to RATlO A/B and the AVGS/TIMING switch to 1. Again check the readout displays using Table 2-4.
The GATE indicator should light and an advancing count should be displayed when the START/STOP button is pushed in. The GATE indicator should go out when the count is stopped by releasing the START/STOP button. Check the overflow display by setting the AVGS/TIMING switch to 100 ns pressing the START/STOP button in, and
letting the count advance, When the last decade (eighth digit) goes from nine to zero the OVERFLOW indicator will light. Release the START/STOP button and observe that the OVERFLOW indicator remains on, but the count does
not change. Pressing the RESET button clears the overflow condition, sets the count to zero, and ex­tinguishes the OVERFLOW indicator.
Totalize A Display
For this check, the jumper located on the Auxiliary
Circuit Board must be in the Totalize position.
Unless you are qualified to do so, refer placement of this jumper to qualified personnel.
Set the FUNCTION switch to the TOTALIZE A/TIME MANUAL position. Observe a zero at the right of the readout display. The GATE indicator should light when the START/STOP button is pushed in, and go out when the button is released. The units indicators and decimal points should remain off.
2-6
TM 9-6625-474-14&P-3
Channel A Slope
Verify that the TOTALIZE/TIME MANUAL jumper is in the TOTALIZE position. With the FUNCTION switch set to TOTALIZE A/TIME MANUAL and CH A to + SLOPE (button out), press the START/STOP button. Turn the CH A LEVEL control fully clockwise. The readout display should increase one count each time the control is rotated from clockwise to counterclockwise (past center posi­tion). Verify that the count does not increase when the control is turned from counterclockwise to clockwise.
Change to - SLOPE (button in) and push the RESET button to clear the display. The readout should now increase one count each time the CH A LEVEL control is rotated from counterclockwise to clockwise (past center). Turning the control from clockwise to counterclockwise should not increment the display.
OPERATING MODES
GENERAL
The following discussion provides general information about each mode of operation and instructions on making measurements for FREQUENCY A, RATIO A/B, TIME INTERVAL (WIDTH B and TIME A - B), EVENTS A DURING B, and TOTALIZE.
FREQUENCY A MODE
In this mode the input signal isconnectedto CH A Input only, either through the rear interface or the front paneI connector. Use ac coupling for most frequency measurements to avoid readjusting the LEVEL control because of changing dc levels. The repetitive nature of the signals makes slope selection unnecessary for frequency measurements. Signals less than 3 volts peak-to-peak need not be attenuated; larger signals should be attenuated to within the range of 60 mV to 3 V peak-to­peak.
Channel B Slope
Set the FUNCTION switch to PERIOD B, CH B to + SLOPE (button out), and the AVGS/TIMING switch to 1. Push the RESET button. Check that the GATE indicator
turns on when the CH B LEVEL control is rotated from clockwise to the counterclockwise position. Turning the
control back to clockwise should have no effect on the GATE indicator. Another turn from clockwise to counterclockwise turns the GATE indicator off.
Change to - SLOPE (button in) and press the RESET button. Observe that rotating the CH B LEVEL control
from counterclockwise to clockwise and back produces an action that is just opposite that described in the
preceding paragraph.
changes unreasonably, the DC 503A is not being triggered properly, either because the controls are not correctly set or the signal is beyond the capabilities of the counter.
Measurement Intervals. To adjust the trigger controls,
choose a short gate time such as .1 second or .01 seconds. This gives rapid feedback via the display whether or not the counter is being triggered. Final selection of gate time
depends upon the frequency being measured, desired
resolution, and willingness of the operator to wait for a
measurement.
Resolution. A 10 second gate time means the operator
must wait 10 seconds for a measurement to be made and
displayed. This will give 0.1 Hz resolution. A 10 second
count will display fewer than the available eight digits for any signal below 10 MHz.
Set the FUNCTION switch to FREQUENCY A and, with the AVGS/TIMING switch, select one of the shorter gate times. Set the DISPLAY TIME control fully coun­terclockwise. Connect the signal to be measured to the input and adjust the LEVEL control for a stable display. The LEVEL control setting should not be critical unless the signal amplitude and frequency are close to the specified limits.
If the count varies from reading to reading, it is
probably caused by jitter in the signal source. If the count
Overflow. Through intentional use of “overflow” dis-
plays, it is possible to improve the resolution of the counter. Select a gate time that displays the most signifi­cant digit as far to the left as possible. Note the numbers displayed to the right of the decimal. Move the decimal to the left, by selecting longer gate times, until the desired resolution is achieved. The OVERFLOW indicator will light when the most significant number overflows the last storage register. The relationship between gate time,
measured frequency, displayed digits, and overflow is
shown in Table 2-5.
2-7
TM 9-6625-474-14&P-3
GATE TIME

Table 2-5

VS MEASUREMENT RESOLUTION
Measurement Rate. Once a stable measurement is
obtained, the rate at which measurements are made can
be controlled by the DISPLAY TIME control. Turning the control clockwise holds off the gate and stores the display for a longer time before a new measurement is made and displayed. Display time and gate time together complete a measurement-display cycle.
The DISPLAY TIME control is uncalibrated and variable from about 0.1 second (in the MIN position) to about 5 seconds. At the extreme clockwise end of the control is a detent position called HOLD. In HOLD, the last count taken will be stored and displayed for an indefinite period. A new count and display can be initiated by pressing the RESET button, moving the DISPLAY TIME control out of the detent, or changing the gate time.
PERIOD MODES
The period and period average modes allow single
period measurements or multiple period averages to be
made with input frequencies into CH B. These modes are
useful for making low frequency measurements where maximum resolution is desired without waiting for long measurement time. Simply stated, the PERIOD B mode reverses the functions of signal and clock as compared to
FREQUENCY A mode. Refer to Fig. 2-4A.
Averaging. Resolution and accuracy is improved by averaging the signal value over a large number of signal events. This increases the total time to take a measure­ment, i.e., similar to selecting a longer gate time in
FREQUENCY A mode. Refer to Fig. 2-4B.
preferred. Other wave shapes can be accurately measured
if the amplitude is kept high.
TIME INTERVAL MODES
Two modes of time interval measurement can be selected: WIDTH B, and TIME A - B. The WIDTH B mode measures the time between two points on a waveform, These two points are selected by the CH B triggering controls such that the counter main gate turns on at the point selected by the CH B SLOPE and LEVEL controls, and turns off at the same level but the opposite slope, Refer to Fig. 2-4C.
The TIME A - B mode measures the time between two points on two waveforms. These two points are controlled such that the CH A triggering controls select the point at
which the main gate turns on, andthe CH B controls select the point at which the main gate turns off. Refer to Fig. 2-
4D.
Triggering. The voltage levels necessary to establish the triggering points on any selected slope are monitored and set with digital voltmeter readings at the CH A/CH B TRIG LEVEL pin jacks on the front panel or rear interface connections. Fig. 2-5 illustrates typical TRIG LEVEL voltage settings for various time interval measurements. When making these measurements, each channel must be dc coupled and coaxial cables must be properly ter­minated in order to maintain signal fidelity.
Low Frequencies. Period Measurements of signals
below 10 Hz, and particularly in the lowest decade from
0.1 Hz to 1.0 Hz, become rather sensitive to wave shape and amplitude. Since it is desirable for the signals to pass
through the trigger hysteresis abruptly, square waves are
2-8
WIDTH B Mode. In order to measure pulse duration (Fig. 2-5, waveform 3), the 50% level must be determined. Set the FUNCTION switch to WIDTH B and the CH B
LEVEL control fully counterclockwise. Apply the input signal to the CH B input connector. The GATE indicator must be off.
TM 9-6625-474-14&P-3
Fig. 2-4. Representation of Interval measurements.
Rotate the LEVEL control until the GATE indicator just come on and record the digital voltmeter reading. Continue rotating the LEVEL control until the GATE indicator just goes off and record the digital voltmeter
reading. Subtract the first digital voltmeter reading from
the second and divide by 2; this is the 50% level.
Reset the CH B LEVEL control so that the digital voltmeter indicates the 50% level. Read the pulse duration from the DC 503A display.
Time A -
B Mode. This measurement requires input
signals to both CH A and CH B, but the peak-to-peak signal amplitude should first be determined using the WIDTH B mode instructions. For TIME A - B measurements, follow these steps:
1.
Set the FUNCTION switch to WIDTH B.
Referring to WIDTH B mode instructions, determine
2. the peak-to-peak amplitude and desired triggering level of the signal to be applied to the Channel B input.
If the signal to be applied to Channel A input is
3. different than that being applied to Channel B, repeat Step 2 for this signal.
4.
Set the Channel B LEVEL control to the desired triggering level as calculated in Step 2.
5.
Set the FUNCTION switch to TIME A - B.
Set the Channel A LEVEL control to the desired
6. triggering level as calculated in Step 3.
7.
With signals connected to the proper channels, read the elapsed time interval between the triggering level of Channel A and the subsequent triggering level of Channel B from the DC 503A display.
Time Interval Averaging. Averaging can be used to increase the accuracy and resolution of repetitive signal measurements. The basic reason for averaging is the
error is truly random, then as more intervals are averaged,
the measurement will tend to approach the true value of the time interval. For time interval averaging to work, the time interval being measured must be repetitive and have a
repetition frequency that is nonsynchronous to the
counter clock rate. The DC 503A will measure up to 10”
averages in both Width B averaging and TIME A - B
averaging.
2-9
TM 9-6625-474-14&P-3
Fig. 2-5. Typical CH A and CH B Level Out voltage settings for various time interval measurements.
EVENTS A DURING B MODE
In the EVENTS A DURING B mode, the events applied to Channel A are counted. The count is gated by the signal applied to Channel B input. The accumulated total of events A that arrived during the time signal B was
triggered is displayed in the readout. Refer to Fig. 2-6.
The following procedure can be used to make a measurement like that shown in Fig. 2-6.
1. Apply the signal to be counted to Channel A. With the FUNCTION switch at FREQUENCY A, set Channel A SLOPE switch to + SLOPE. Adjust the LEVEL control for a stable display.
2. Apply the control signal to Channel B. With the FUNCTION switch at PERIOD B, set Channel B SLOPE switch to + SLOPE, Adjust the LEVEL control for a stable display.
3. Set the FUNCTION switch to EVENTS A DURING B.
When the Channel B signal excursion occurs, Channel B is triggered andthe gate opens, allowing the Channel A pulses to be counted.
Averaging. Averaging can be used to Increase the accuracy and resolution of repetitive event per interval measurements. As more events are averaged, the measurement tends to approach the true value of the number of events per interval.
RATlO MODE
The DC 503A may be used to measure the ratio of two signals, where one signal is applied to Channel A input and the other signal is applied to Channel B input.
2-10
TM 9-6625-474-14&P-3
Fig. 2-6. Illustration of CH A events counted from portion of CH A signal pulses during the counter gate open time (controlled by CH B signal).
In the Ratio A/B mode, the frequency of the signal applied to Channel A is divided by the frequency of the signal applied to Channel B, and the resultant ratio is displayed.
measurements, the smallest number of averages that produces a useful number of digits should be considered.
TIME MANUAL MODE
This mode is a manual analog of the TIME A - B mode.
Triggering. The operation of Channel A and Channel B trigger controls is the same as for frequency and period measurements. Set the trigger controls as follows:
1.
Go to the FREQUENCY A mode and adjust the Channel A trigger controls for a normal frequency measurement.
Go to the PERIOD B mode and adjust the Channel B
2. trigger controls for a normal period measurement.
Leaving the Channel A and Channel B trigger
3. controls as they are, gotothe RATIO A/B mode. The correct ratio should be displayed.
Resolution. The AVGS/TIMING switch, which controls
the number of averages of the Channel B signal, may now
be set to display maximum resolution. For most
In this mode, only the AVGS/TIMING switch and START/STOP switch affect the display.
Starting and Stopping. The TIME MANUAL mode may bethought of as a “stop-watch” type of operation. With the FUNCTION switch inthe TIME MANUALposition (and the internal jumper properly positioned), the display starts counting time-base pulses when the START/STOP switch is depressed. It will continue to count and display the accumulated total until the START/STOP switch is releas­ed. The last count will then be held in the display until another START command is given (in which case the count will again advance), or other controls are actuated.
Pressing the RESET button will return the display to zero. Changing the setting of the AVGS/TIMING switch will change the frequency of the time-base pulses being
counted and reset the display to zero. The start/stop function can also be performed remotely via the rear
interface connections.
2-11
TM 9-6625-474-14&P-3
Clocking Rate. When the AVGS/TIMING switch is in
the 1 s position, one-second pulses are being counted and the display accumulation advances one count per second, and so on.
Whenever the accumulated count is above 99,999,999, the OVERFLOW indicator will light to indicate register overflow; however, the accumulation continues at the
not displayed.
TOTALIZE A MODE
This mode is a manual analog of the FREQUENCY A mode. lnthismode, signal event sapplied tothe Channel A input are counted and the accumulated total displayed during the time the START/STOP button is depressed to the START position. The main application of this mode is to accumulate the count of relatively infrequent and irregular events.
Operation. Apply the signal to Channel A input and set the trigger controls the same as for a frequency measure­ment. Only the Channel A trigger controls, the RESET button, and the START/STOP button affect the display in this mode.
Starting the Count. Press the START/STOP button and adjust the Channel A LEVEL control until a count begins to advance. The accumulated count is displayed in whole numbers.
will continue to be displayed. No more incoming events will be added to the total.
Restarting and Resetting. When the START/STOP button is again depressed, incoming events will advance the displayed total. Resetting the count to zero can be done at any time by pressing the RESET button.
Remote start/stop. Starting and stopping the count can
be accomplished remotely via connections to the rear
interface.
REPACKAGING FOR SHIPMENT
If the Tektronix instrument is to be shipped to a Tektronix Service Center for service or repair, attach a tag showing: owner (with address) and the name of an individual at your firm that can be contacted. Include
complete instrument serial number and a description of
the service required.
If the original package is not fit for use or not available,
repackage the instrument as follows:
Surround the instrument with polyethylene sheeting, or
other suitable material, to protect the exterior finish.
Obtain a carton of corrugated cardboard of adequate
strength and having inside dimensions no Iess than six
inches more than the instrument dimensions. Cushion
the instrument by tightly packing dunnageor urethane
foam between the carton and the instrument,
sides, Seal the carton with shipping tape or
dustrial stapler.
on all
an in-
Stopping the Count. If the START/STOP button is
released and no other controls are actuated, the last total
2-12
The carton test strength for your instrument
pounds.
is 200

SECTION 3

TM 9-6625-474-14&P-3
THEORY
OF OPERATION
BLOCK DIAGRAM DESCRIPTION
Introduction
For the following block diagram description refer to the
Block Diagram foldout page at the rear of this manual.
Channel A and Channel B Amplifiers
There are two inputs, CH A and CH B. Signals to be
counted or timed are applied to either or both channels via front panel bnc connectors or via the rear interface. The front panel inputs for both channels are terminated with an
paralleled with approximately 27 pF. The rear interface
inputs to both channels are terminated with a resistance of
Signal Flouting circuits.
Each channel contains an ac/dc coupling switch, a X1 or X5 attenuation network, a buffer amplifier circuit acting as a comparator that compares the incoming signal level against the triggering level as a reference, and amplifier/Schmitt circuits driving the signal slope selec-
tion functions in the Signal Routing circuits. Each channel
also contains an operational amplifier serving as a X1 buffer circuit, supplying a buffered version of the trigger level at the front panel tip jacks or rear interface connec-
tions.
NOTE
panel with the proper decimal point location and correct annunciator illuminated.
The Measurement Cycle Timing circuit determines the Display Time, clears the Gate Generator circuits, loads (latches) the decade counters, and resets the counters for the next measurement cycle in all modes of operation.
PERIOD B (Variable Clock)
In this TIMING mode, the CH B signal is passed through the Signal Routing circuits to the Gate Generator
Input of the Decade Accumulators and the Measurement Gate is generated by a single period of the signal from Channel B. As before, the accumulated count for this mode and all subsequent modes is latched, decoded from
BCD data to 7-segment information, and displayed on the front panel with the correct annunciator illuminated and the proper decimal point location.
PERIOD B (Averageable—100 ns Clock)
For this AVGS mode, the Time Base signal
(10 MHz = 100 ns) is not divided; it is applied through the Signal Routing circuits directly to the Count Input of the
Decade Accumulators. The CH B signal is routed to the
The remainder of this block diagram description discusses the signal paths through the remaining circuit blocks and the typical events related to each mode of operation (FUNCTION) listed on the front panel.
FREQUENCY A (Variable Gate)
For this mode of operation the CH A signal passes
directly through the Signal Routing circuits to the Decade
Accumulators. The signal is counted by the 1st DCU, then the 2nd DCU, and then by the 6-Decade Counter (a total of eight decades). In the FREQUENCY A mode the Time
generate a Measurement Gate (via the Gate Generator) for the desired measurement time. At the end of the Measure-
ment Gate interval, the accumulated count is latched in the 8-Decade Latch/Multiplexer circuits, converted from
BCD to 7-segment information and displayed on the front
of the CH B signal. The Time Base is counted for 10”
periods before the accumulated count is latched for
display.
WIDTH B (Variable Clock)
In this TIMING mode, the 10 MHz Time Base is routed
routed directly to the Count Input of the Decade Ac­cumulators. The CH B signal is used to generate the
Measurement Gate (via the Gate Generator). A single pulse width at the output of the Channel B amplifier
generates the gate.
WIDTH B (Averageable—lOO ns Clock)
In WIDTH B, AVGS mode, the Time Base signal is not
divided by N; it is routei directly to the Count Input of the
3-1
TM 9-6625-474-14&P-3
Counts are accumulated in the decade counters during either the positive portions of the pulse widths or the negative portions, dependent on the SLOPE polarity selection for the Channel B signal.
signals and the Channel A signals are counted during that
time.
TOTALIZE A
In the TOTALIZE A mode the Measurement Gate is generated by the START/STOP switch on the front panel or via the Remote Start/Stop line at the rear interface. The
Circuit, and Gate Generator circuits are not used for this mode. Instead of accumulating clock signals from the
signals are accumulated during the START/STOP inter­val.
TIME MANUAL (Variable Clock)
For this mode there are no inputs to Channel A or
Circuit whose output is routed directly to the Count Input
of the Decade Accumulators. The Measurement Gate interval is generated either by using the START/STOP switch on the front panel or by changing the voltage level
on the Remote Start/Stop input at the rear interface.
EVENTS A DURING B (Averageable)
This mode is exactly like the WIDTH B (Averageable) mode, except that the output of the Time Base is disabled and the output of Channel A is applied directly to the Count Input of the Decade Accumulators. For this mode,
the pulse width at the output of Channel B is routed to the
produce the Measurement Gate Interval. The Channel A events are averaged for 10” pulse widths from Channel B.
RATIO A/B (Averageable)
Decade Accumulators, 6-Decade Counter/8­Decade Latch
The 1st DCU consists of ECL flip-flops, requiring
ECL/TTL conversion to drive the first decade latch The 2nd DCU operates at TTL levels and drives the second decade latch directly. From that point, there are six more
internal counters and six more decades of latch, all
contained in one integrated circuit. This arrangement
provides a total of eight decades of count and eight decades of latch.
The 6-Decade Counter/8-Decade Latch circuit has its own internal oscillator to generate the Time Slot informa­tion. It also generates the Scan Clock, Overflow, and BCD output data. Between the time slots and BCD data there is enough information to drive the Display. The zero blank­ing function is also provided internally.
Measurement Cycle Timing
The display timing, reset, clear, and load (latch) functions for the decade counters are provided by the Measurement Cycle Timing circuit.
Decimal Point and Annunciator Encoder
The Time Base output is disabled and not used for this
mode; Channel A signals are routed directly to the Count
3-2
The decimal point location is determined by encoding circuits using the time slot information and information
Loading...
+ 92 hidden pages