Tektronix DC-502 INSTRUCTION MANUAL

INSTRUCTION MANUAL
Qservice ----- This Document is a complete scan from the Original Tektronix Manual ----- Qservice
Tektronix, lnc.
070-141
2-00
-e
P.O. Box 500
+
Beaverton, Oregon 7005
+
Phone: 644-0161 + Cables: Tektronix
872
TABLE
Qservice ----- Hyperlinked Page Scroll Up - Down Left click to select ----- Qservice
Qservice ----- Hyperlinked Page Scroll Up - Down Left click to select ----- Qservice
OF
CONTENTS
Qservice
-----
SECTION
SECTION
HY per
1
2
linked Page Scroll Up - Down Left click to select
OPERATING INSTRUCTIONS Page
Introduction
Description Installation
Operational Check
Display
Frequency Measurements
Totalizing
Using the Counter
Dl RECT INPUT Attenuation and Trigger
Signal Connection Measurement
Rear Connector Option 1-Precision Time Base
Specifications
THEORY OF OPERATION
Input Circuits Time Base and Control Circuit Counter Circuits Decode and Display Multiplex Power Supplies and InputIOutput Lines
Check
Level Adjustment
Interval and Display Time Controls
I10
Assignments
-----
Qservice
Qservice
SECTION
-----
SERVICE
3
Symbols
Electrical Parts List Controls and Connectors Illustration Adjustments Illustration Parts Location Grid
Block Diagram Counter Time Base and Control Schematic Counters and Display Schematic Mechanical Parts List
Fig.
Accessories
lNF ORMATION
&
Reference Designators
1
Exploded View
&
Repackaging
Hyperlinked Page Scroll Up - Down Left click to select
-----
Qservice
Section
Qservice ----- This Document is a complete scan from the Original Tektronix Manual ----- Qservice
1-DC
502
INTRODUCTION

Description

The DC 502 550 MHz Frequency Counter measures
frequencies from 10 Hz to 550 MHz or totalizes events to
10'
at
a
maximum rate of 550 MHz. The
in
a
TEKTRON
Frequency measurements are accomplished using one of two BNC inputs on the front panel. The DIRECT INPUT has a frequency range of 10 Hz to peak-to-peak sensitivity, selectable attenuators, and an adjustable trigger level range. The -10 PRE-SCALE INPUT has
a
frequency range of 50 MHz to 550 MHz with 500 mV peak-to-peak sensitivity and impedance. The same four measurement interval times are
selectable for each input.
IX
TM 500 Power Module only.
110 MHz with a 300 mV
DC
502 operates
a
50
i-2
input
SECTION
OPERATING
INSTRUCTIONS
Front panel controls reset the Counter and provide Start/Stop commands for the manual totalizing mode of each input.
Measurement display is accomplished with seven-segment
LED's in automatically positioned by the MEASUREMENT INTERVAL selected, and leading zeros (to the left of the most significant digit or the decimal point) are blanked. LED's indicate when the GATE is open, when the kHz or MHz units are displayed, and when OVERFLOW occurs.

Installation

a
The Referring to Fig. Module and turn on the power.
a
7-digit readout. The decimal point is
DC
502 is calibrated and ready to use as received.
1-1, install the DC 502 into the Power
1
Fig.
1-1.
Plug-in module installation/removal.
REV.
MAY
1974
Operating Instructions-DC
502
OPERATIONAL

Display Check

Press the RESET button to check the 7 character segments of each digit; the numerical display should be a row of eights. To check the decimal point position and the
units indicators,
switch as follows:
Switch
Position
In the MANUAL position, no decimal point will be displayed. Press the START button and check that the GATE indicator lights, then release the button (STOP) and check that the GATE light goes out. To check the OVERFLOW indicator,
INTERVAL switch to 10 INPUT connector. The length of time is determined by the DISPLAY TlME control, and will be
discussed in the next few paragraphs.
set
the MEASUREMENT INTERVAL
Numerical
Display Units
.OOOO
.OOOOO
.OOO
.OOOO
000 000
.OOO
.OOOOO
.OOOO
.OOO
set
the MEASUREMENT
s
and apply 15- or 20-MHz to the
a
display can be held
MHz MHz kHz' kHz
kHz MHz MHz
MHz
CHECK
Then turn the MEASUREMENT INTERVAL switch to the position that gives the desired reading. Generally, use the shorter measurement intervals for high-frequency, low­resolution measurements and the longer intervals for measurements requiring
The OVERFLOW indicator can be lit for high­resolution measurements, allowing the frequency to be indicated to
the end of this secifion for resolution and accuracy at each position of the MEASUREMENT INTER VAL switch.
The display is updated at a rate determined by the DISPLAY TlME control. Each time a sample of the input signal
is
taken, the GATE light will flash and the new
reading will be displayed. To change the display time, which is continuously variable from about 0.1 second to 10 seconds, or to hold a display indefinitely, turn the
DISPLAY TlME control.
+I0
Pre-Scale Input.
scaled, AC-coupled input to measure the average frequency
of signals from 50 MHz to 550 MHz. This input has
sensitivity of 500 mV, peak-to-peak, and a maximum input
voltage limit of 10
damage the diodes
may
10 V
ATTEN controls do not apply to this input.
a
high resolution.
NOTE
0, I Hz.
Refer to the Specifications at
The DC 502 also provides a pre-
\/,
peak-to-peak.
Signals greater than
of
the input circuit. The
a

Frequency Measurements

Direct Input.
of the average frequency of signals from about 10 Hz to 110 MHz. The input sensitivity is 300 mV peak-to-peak, so select the proper attenuation (XI, X5, X10, or X50) for the given signal.
The DC 502 provides direct measurement
CAUTION
a
The input signal must not exceed
a
Apply MEASUREMENT INTERVAL switch to the .01 SEC posi­tion and the DISPLAY TlME control fully CCW. Observe the numerical readout display. Adjust the TRIGGER LEVEL control for most significant digit in the display should be blanked.
signal to the INPUT connector. Set the
a
stable reading. The zeroes leading the
1-2
500
volts,
REV.
Apply a 50 MHz to 550 MHz signal of at least 500 mV
amplitude to the -10 PRE-SCALE INPUT. Set the
MEASUREMENT INTERVAL switch toJ .O1 SEC and observe the readout. Leading zeros should be blanked. Select a MEASUREMENT INTERVAL which gives the best accuracy and resolution. As with the DIRECT INPUT, shorter measurement intervals give higher frequency, lower resolution measu rernents; longer intervals offer greater resolution, especially when overflow is employed.

Totalizing

DIRECT INPUT.
accumulated number of signals (events) applied to the
DIRECT INPUT connector up to the register capacity of 9,999,999 during the time interval between STARTKTOP commands from the front-panel pushbutton. Input signal
rate should not exceed 1 10 MHz.
MAY
The DC 502 will count and display the
Operating Instructions-DC
502
Set the MEASUREMENT INTERVAL switch to MANUAL, apply the signal, and push the START button. The GATE indicator will light and the progressing count will be displayed. Adjust the LEVEL controls the counting, release the START button. The GATE light will go out and the displayed count will be held. The displayed count can continue when the START button depressed again. The counter can be cleared to zero
as
necessary for a steady count. To stop
ATTEN and TRIGGER
at
any
USING
THE
Dl R ECT l NPUT Attenuation and Trigger Level
Adjustment
Signals to be counted in the DIRECT INPUT channel
a
may have which are unsuitable to drive the counting circuits. Because of this, the signal applied to a signal-shaping circuit which converts
rectangular pulses of uniform amplitude. This circuit
includes volts to which the incoming signal the 300-millivolt sensitivity window of the signal-shaping circuit to be adjusted to
incoming waveform (see Fig. 1-2). Obtaining reliable reading input attenuation and proper adjustment of the TRIGGER
,EVE L control.
wide variety of shapes and amplitudes, many of
is
first passed through an attenuator, then
a
reference level adjustable between + and
is
compared, allowing
a
convenient amplitude on the
a
steady,
is
dependent upon the proper selection of
Noise Impulses 200-millivolt
it
to
-
time by pressing the RESET button or by MEASUREMENT INTERVAL switch to another position.
+I0
is
PRE-SCALE
to this input, the displayed count advances one count for every ten incoming events. The incoming events must have transition times and periods suitable for the input triggering requirements.
INPUT.
In the MANUAL mode related
COUNTER
Generally, the best point on a waveform for triggering the counter free of noise. On steepest slope occurs pulses or other signal com ponents of sufficient amp1 itude to produce unwanted trigger pulses will cause an erratic or incorrect count. Fig. 1-2 shows the TRIGGER LEVEL control adjusted to avoid error. applications, monitor the incoming signal with
2
oscilloscope.

Signal Connection

Coaxial cables and probes offer very convenient means of connecting the signals to the front-panel input BNC connectors. These devices are shielded to prevent pickup of electrostatic interference which can cause erroneous trig­gering and probe not only reduces the size of the signal, but also presents
test
to perform very close to normal operating conditions. For the +10 PRE-SCALE INPUT, the 50 careful impedance matching. If the signal must be atten­uated to avoid exceeding the maximum input limit of 10 use 50 impedance of the DC 502.
is
where the slope
a
sine-wave signal, for example, the
a
faulty count. For the DIRECT INPUT, a XI0
a
high input impedance to allow the circuit under
is
steep and therefore usually
at
the zero-crossing point. Noise
In critical measurement
s2
s2
attenuator pads terminated by the 50
moving
50
S2
input requires
the
pre-scale
a
test
V,
S'l
input
I
:A)
Erroneous count.
:B)
Correct count.
1-2.
Fig. proper adjustment of TRIGGER LEVEL control can avoid an erroneous count.
Two examples of triggering circuit output showing how
Measurement l nterval and Display Time Controls
The MEASUREMENT INTERVAL switch selects the time interval (also called gate time) during which the DC 502 counts. The internal time-base circuit derives gate times from an accurate 1-MHz reference signal to make frequency measurements. These gate times are 0.01 1
s,
or 10
s.
The measurement interval selected determines
the measurement range and resolution. Also, the displayed
is
decimal point measurement units (MHz or kHz) are indicated for the corresponding switch position.
The DISPLAY TIME control measurement can be held in the counter and displayed. The HOLD detent position allows indefinitely, or until the counter front-panel RESET button.
positioned correctly and the correct
sets
the length of time
a
measurement to be held
is
reset to zero by the
s,
0.1
s,
a
Operating Instructions-DC
502
Fig.
1-3.
Input/Output pin assignments at rear connector.

Rear Connector I/O Assignments

lnput and output data access to the DC 502 is available
at the rear of the main circuit board. Fig. 1-3 identifies the
contacts and their respective
Module mainframe option is available which provides a rear-panel, multi-pin connector to which these data can be hard-wired for external access. Also possible are intra­compartment connections with other plug-in modules when using a multi-compartment Power Module.
1/0
assignments. A Power
Contact
28B 27 A 27 B 26A 25A 258 24B 23B 226 20 B
20A 21 B 19A 19B
1
Option
The DC 502 can be ordered with compensated 5-MHz crystal oscillator to provide a highly stable (5 parts in option includes
1
-MHz clock.
-Precision Time Base
Second decimal point (D2) output.
Internal scan clock disable input. MHz light output. Reset inputloutput. TSg (Time Slot Zero) output. External scan clock input. Internal scan clock (2 kHz) output. Overflow output. MSD (Most Significant Digit) output.
2
Data good output.
lo7)
and precise internal time base. This
a
divide-by-five IC counter to produce the
BCD output, serial by digit.
Assignment
a
temperature-
SPECIFICATIONS
Measurement Ranges, Resolution, and Accuracy
Frequency, 10 Hz to 550 MHz; Gate times, 0.01
1
s,
and 10 Totalizing capacity, 0 to INPUT), 100 Hz
I
s,
INPUT), 1000 Hz and 1 Hz at 10
s;
Display time, about 0.1 s to 10 s to HOLD;
at
0.01 s gate time, 10 Hz
and 0.1 Hz at 10
at
0.01
s;
Accuracy,
lo7;
Resolution (DIRECT
at
s;
Resolution (+I0 PRE-SCALE
sf
100 Hz
+I
at
0.1
s,
count + time base accuracy.
Direct lnput
Frequency, 10 Hz to 11 0 MHz; Sensitivity, 300 mV peak-to-peak; Trigger level range, +2 V; Attenuators, XI, X5, X10, and X50; Coupling, AC; lnput impedance, approximately 1
input volts, 500 V (DC
+
10
Pre-Scale lnput
Frequency, 50 MHz to 550 MHz; Sensitivity, 500 mV
peak to peak; Coupling, AC; lnput impedance, approxi­mately 50
Ma paralleled by about 20 bF; Maximum
+
peak AC, or peak-to-peak AC).
Q;
Maximum input volts, 10 V peak-to-peak.
Data Inputs and Outputs
Available via plug-in connector to multi-pin connector rear of Power Module. lnput lines are available for internal and external scan clock control. Output lines are available
for BCD output (serial-by-digit), and to indicate status of timing, data good, reset, scale, decimal point and overflow.
0.1
10 Hz
s,
s,
1 Hz at
at
0.1
1
s,
s,
at
Internal Time Base
Option 1
5 MHz
Within 5 parts in
I
o7
1
part in
lo7
per month
within 5 parts in
1
lo9
Crystal Frequency
(OOC
Stability +50°c), after hour warm-up
Long-term Drift
Accuracy
to
I
/2
Standard
1 MHz
Within 1 part in
I
o5
1 part or less in
lo5
per month
Adjustable to Adjustable to within 1 part in
1
lo7
Other
Temperature Range, Operating:
operating: -40°c to +75O~.
Altitude Range, Operating: to 15,000 feet; Non-
operating: 50,000 feet.
OOC
to +50°c; Non-
REV.
MAY
1974
Section
Qservice ----- This Document is a complete scan from the Original Tektronix Manual ----- Qservice
2-DC
502

INPUT CIRCUITS

Direct lnput
Attenuators.
front-panel Dl R ECT INPUT connector J100 to the attenu­ators. The attenuators are frequency-compensated voltage dividers consisting of resistors R 102-R C102-C107. Switches SlOOA and SlOOB allow front-panel selection of XI, X5, X10, or X50 attenuation of the input signal. C110 provides AC coupling.
FET source follower Q115 and emitter follower 0122
present the base circuit of E. clamping network, which reduces the input signal to limits suitable for driving the shaping circuits. The clamping
diodes limit the voltage
dynamic range of about 1.2 volts.
with push-pull outputs, shapes the input signal into window"
states
threshold, then reverts to
voltage passes through the lower threshold. For this reason,
an input signal smaller in amplitude than the width of the
hysteresis window cannot activate the counting circuits.
adjusted to overcome some of the triggering difficulties
arising from various input-signal shapes and frequencies.
Integrated-circuit operational amplifier U135 and associated discrete components are connected follower. TR IGGE R LEVEL potentiometer R 135 selects voltage between ground and about -2 volts and applies to pin 3 of U135. This level hence, the input of U150B, through the action of the operational amplifier.
a
Signal-Shaping.
when the signal voltage passes through the upper
The quiescent level at the input of U150B can be
Signals to be counted are applied via
107 and capacitors
high impedance to the input signal. The diodes in
F.
0128 form a series-limiter and
at
the emitter of 0128 to
U150B1 an OR gate integrated circuit
is
connected
a
is
a
width of about 200 mV. The output changes
its
is
as
a
Schmitt trigger. It
square wave. Its "hysteresis
original
then established
state
when the signal
as
a
at
pin 2, and
voltage
its
SECTION
THEORY
OPERATION
phase inversion (to correspond with the input signal) and becomes the decade input.
+I
0
Pre-Scale lnput Circuit
50
MHz
to 550 AC-coupled into current,
and 188 in the diode clamping bridge turned on until the input signal amplitude reaches about bridge equivalent resistance of the bridge in series with R189, 2452, and the emitter resistance of Q190, about 552. As the signal amplitude exceeds k0.4 V, one pair of diodes
(CR 185, CR 188) begins to turn off during positive-going
excursions, while the other pair (CR 186, CR 187) turns off
a
during negative-going excursions. As these diode pairs
alternately turn off with signal amplitude changes, a
matched pair of diodes side of the coupling capacitor, C182, alternately turn on to maintain the CR181A levels which may be part of the input signal.
changing DC level
peak-to-peak. This signal via the high-frequency peaking network, C189-R189. C192 couples the amplified signal to pin 10, the high impedance input of the +I0 counter, U190. C192 circuit board with special lead dress to aid high frequency response mounted on one lead of C192 to suppress high frequency oscillations. R191 establishes C191, 193, and 194 decouple pins 14, 13, and 12,
a
respectively, right
it
board. L199 and C199 suppress power supply noise and decouple the Vcc input, pin 14.
set
by R185 and R187, keeps CR185, 186, 187
is
turned on, the signal source sees about 2052
&
CR181 B from being biased on or off by DC
The clamping action of the bridge diodes limits the
at
minimum signal amplitudes. Bead L192
MHz
signals applied to J180 are
a
5052 environment. A quiescent 10 mA
f0.4 V. While the
(CR181A, CR181 B) on the input
5052 input impedance. C181 prevents
at
the bridge output to a maximum 0.8 V
is
then coupled by C188 to Q190
is
mounted on the
a
bias
at
pin 10 of U190.
at
the terminal connections to the circuit
2
OF
is
Pin 4 of U190 produces one output cycle, a level change
The output of push-pull outputs drive Q160 and Q162, which are connec­ted
as
a
differential pair. This circuit provides a level shift to TTL level, and further shapes the signal to be counted. A waveform with fast rising and falling edges the collector of Q160. CR165 limits the amplitude of the count signal to 5 volts, clamping the negative-going portion
of the signal to ground. The signal emitter follower 0170 to U160B, where
U150B
is
applied to U150A, whose
is
produced
is
then passed through
it
receives a final
at
of about 0.5 V, for every ten input events to pin 10. The base of Q195 swings approximately between 3.4 V and
3.9 V. The emitter-follower action of 0195 provides healthy current drive to the base of Q198, which would saturate in no-signal conditions except for CR196. The collector of signal changes, producing pin 9 of U16OB in the gating logic to the Decade Counting
Units.
Ql98 can, therefore, respond immediately to
a
TTL-compatible output to drive
a
Theory of Operation-DC

TIME BASE AND CONTROL CIRCUIT

502
1
MHz
Clock
A precise one megahertz clock provides the reference for operation of the gate-generating circuits. The output of crystal oscillator Y 200 megahertz. The four parts of U200 form
is
adjustable by C201 to exactly one
a
shaper-buffer
stage to produce square-wave clock pulses and to isolate the
oscillator from the 1-MHz output line.

Optional Clock

An optional
5 MHz crystal oscillator and a divide-by-five counter. This
combination is shown on the schematic
1
MHz clock is available, using a very stable
as
Y201 and U201.

Decade Divider Units (DDU's)

The DDUrs consist of seven cascaded divided-by-ten counters, U209 through U215. They produce four gate
times, 0.01 sec, 0.1 sec, 1 sec and 10 sec, which are made available via the MEASUREMENT INTERVAL switch to the gate generator to establish the precise time interval the GATE is open. The of U209, whose output
1-MHz
clock signal is applied to pin 14
is
connected to the input of the
subsequent decade. Each decade is clocked with a negative­going transition. The DDUrs are reset by
which
a
count
in
U209
and
a
CLEAR pulse,
a
count
in
each
subsequent

Gate Generator

The gate generator produces the GATE control signal and initiates the CLEAR, CLEAR, and LATCH pulses. The generating portion consists of
U222B. The display time control portion consists of 0230, Q238, and 0240. The circuit will be described first in the normal gate mode (M EASU REM ENT INTERVAL switch in one of the four gate time positions).
Assume that the To conditions are
The
Q
outputs of U220A, U222A, U220Br and U222B are
all LO. Q230
is
off and the emitter of 0238 rises charges. At TI, Q238 reaches discharges the capacitor. This results in pulse on the direct-se_t input (pin 2) of U220A, forcing its Q output
NAND gate
HI and its Q output LO. With two HI
U230A,
its
U220A, U222Ar U220B, and
as
given in Fig. 2-1.
as
C235
its
firing potential and
a
short-duration LO
inputs on
output goes LO and the output of
Scan Clock
Firing Potential
0238
Emitter
0240
Collector
U220A
0
U222A
0
U220B
0
U222B
0
-
0
CLEAR
LATCH
0238
Output
Output
Output
Output Output
-
I
I
11 I
.
.
I (I I
I
II
Ill
I
'I
1
I'
I
II
I'
I I
I
II
I
Fig.
2-1.
I
1
I
I
I
I
I
I
I I
I
I Ill
I I
I
I
7-i
I I I
Time Base generator normal gating mode ladder diagram.
III
Ill
Ill
I
1
I I
I
I
I
I
I
Y
I
I
I
I]
1
I
I
I
I
I
I
I
I
I
I
I
I
I
I I
REV.
MAY
1974
NOR gate U230C goes HI, producing the CLEAR and CLEAR control signals. The next HI-to-LO transition from the I-MHz to go HI and
inputs, U230A reverts to the CLEAR and CLEAR pulses. The DDUrs then start counting from their 0999999 reset condition.
At the end of DDU's to count the first digit, plus negative transition from the DDU's via the MEASURE­M ENT INTERVAL switch toggles U220B. This corresponds
T3 in Fig. 2-1. U220Brs Q output goes HI and
to output goes
1-MHz clock (T4) toggles U2_22Br causing
go HI (GATE open) and current to the front-panel GATE indicator LED, DS225). The GATE signal saturating the transistor and preventing C235 from charging.
The GATE remains open (HI) for the time duration selected by the MEASUREMENT INTERVAL switch. At the end of this time, which corresponds to another negative transition from the-DDU1s toggles U220B.
U220Brs Q output goes LO and
next negative transition from the 1-MHz clock (T6) toggles
U222BI causing
Simultaneously, the Q output goes HI, removing current
from the GATE indicator LED.
When the GATE output goes LO, the negative transition
toggles U230D has two HI inputs, placing gate U230B and activating the LATCH control signal (HI
clock-(T2) toggles U222A, causing
its
Q to go LO. With a LO applied to one of
its
original condition, terminating
a
10-microsecond delay (time for the
a
propagation delay), a
LO. The next negative transition from the
its
its
Q output to go LO (supplying
is
also applied to the base of 0230,
its
Q output goes HI. The
its
Output to go LO, closing the GATE.
U220A. switching Q LO and
a
HI.
Now NAND gate
a
LO
at
the input of OR
its
Q output
its
its
Q output to
T5 in Fig. 2-1,
Theory
state).
One microsecond later (T,),
the 1-MHz clock toggles U222A, switching
placing reverts to signal.
(T6). When 0230 turns off, C235 begins to charge through
R232-R235 toward the Vcc supply. R235, DISPLAY TIME, provides an display time from about 0.1 second to about 10 seconds.
When the DISPLAY TIME control
h
(HOLD detent position), S235 opens, and C235 stops
charging. When S235 sufficiently to bring Q238 to
display time ends and the next GATE-opening sequence begins.
a
LO on the input of NAND gate U230D. U230D
its
original condition, terminating the LATCH
The display time begins when the GATE signal ends
adjustable time constant to vary the
is
of
Operation-DC
a
negative edge from
is
closed and C235 charges
its
firing potential (TI), the

Manual Gate

The manual mode of operation MEASUREMENT INTERVAL switch in the MANUAL position. The switch closure to ground (cam
switch) places
U222A, and
the
Q
outputs of U222A and U220B HI, and the Q output
of U220A LO. With both inputs of U230D held HI, the
LATCH output the display continuously. The GATE front-panel START button applying
GATE-open condition GATE out). To reset the counters in the manual mode, the RESET button must be pushed to activate to CLEAR, CLEAR and
RESET control signals.
is
a
LO on the set inputs of U220B and
a
LO on the clear input of U220A. This forces
is
held HI, allowing the counter to update
a
HI to the clear input of U222B. As before, the
is
HI
then closed when S210
is
selected by placing the
is
opened when the
is
pushed in, opening S210 and
at
the Q output of U222B. The
is
set
to STOP
502
its
outputs and
fully clockwise
5
of the
(button

COUNTER CIRCUITS

Decade Counter Units (DCU's)
The
lo0
through divide-by-ten counters. The first decade counter of four individual decade input (up to 100 MHz), and each subsequent DCU
a
single IC. U165A, U165B, U167, and U169 comprise the
first (lo0) decade counter, and U235 through U240 make
up the remaining six DCUrs.
When the
the counter
toggle input of U165B. On every tenth clock input counted by the first decade counter, the output of U169 goes LO, providing the second decade counter. Each subsequent decade divides
J
and K inputs of U165B are HI (GATE open),
is
enabled. The input signal
a
carry signal which becomes the clock input for
lo6
DCUrs are seven cascaded
J-K
flip-flops to accept the high-speed
is
applied to the
is
made up
by ten in connected from each DCU to latch. When the CLEAR (HI) and CLEAR (LO) signals are activated, all of the decade counters are reset to the zero-count
is
a
similar manner. Four BCD output lines are
its
associated storage-register
state.

Storage Register

The seven IC latches (U250 through U256) comprise storage register which stores the corresponding decade counter BCD output. The BCD output data inputs
respectively). The LATCH strobe input closure of the GATE or when the MEASUREMENT
INTERVAL switch
at
pins
I,
5,
7,
and 3 (2O, 2l , 22, and 23 bits
pulse
at
pin 2 of each latch immediately upon
is
placed in the MANUAL position,
is
is
applied to the data-
a
applied to the
as
Theory of Operation-DC
502
described in
LATCH input goes HI, the logic levels at the data inputs are
transferred to the associated BCD bit output to be scanned
by the multiplexing circuit.
the time base and control circuit. When the

Overflow Register

When the decade counters have counted to 9,999,999, the counters are full. At the next count, the 23 output of U240 goes LO, providing this occurs,
U2416, then when the LATCH pulse ends (goes LO), U241A pin 13 of U241A goes LO, CR241 and DS242 conduct.
a
LO
is
is
toggled and the LO
a
toggle input to U241 B. When
transferred from pin 10 to pin 8 of
is
transferred to pin 13. When

DECODE AND DISPLAY MULTIPLEX

Scan Clock

The scan rate of the multiplexing circuit by .the scan clock. The scan clock and U260D, which operate
at
an approximate 2-kilohertz rate. The scan-clock output passed through NOR gate U260A1 which can also accept an externally applied scan-clock signal. Other inputloutput lines provide internal scan-clock disable and internal scan-
clock output. The scan clock drives an eight-state counter
and
a
storage register for zero suppression.
+8
Counter and Time-Slot Decoder
The divide-by-eight counter U263A, and U262A, which are three halves of SN7474 type D flip-flops. The output of this counter drives U265, an SN74145 BCD-to-decimal decoder. U265 provides eight output lines (designated TSo through TS7 in the schematics and in Fig. 2-2) to simultaneously enable the output of each counter latch and sequentially. For example, when the TS1 line goes LO, Q280
is
turned on to supply anode voltage to DS280 same time inverter U267C applies U256, enabling allows the latches to share
its
output. Operation in a time sequence
as
its
a
common
is
a
free-running multivibrator
is
made up of U2626,
corresponding display LED
a
HI to pin 6 of latch
set
is
determined
composed of U260B
at
the
of output lines.
DS242 front-panel OVERFLOW indication.
achieved via 0242 and CR244. The emitter of Q242 grounded by a switch closure, then when pin 9 of U241 B goes HI on the first overflow count, 0242, CR244, and
DS242 turn on.
zero suppression during the overflow condition, the display­controlling circuits are notified via U245A that the count
in excess of that displayed by the LED readout.
is
an LED, and in
In the Manual counting mode, OVERFLOW indication
U241
is
reset by the CLEAR pulse. To prevent leading-
its
conduction state gives

Leading Zero Suppression

Decoder driver U270 also has a zero-blanking feature
which allows suppression of the zeroes leading the most
significant digit (MSD) in the display. At
is
applied to the direct-clear input of U2636, the zero-
suppression storage register. This sets U263B to the
zero-suppress state (HI at pin 8), allowing the Ripple­Blanking Input (RBI, pin 5) of U270 to be LO. When the output of U265 advances to the next time slot
LO
RBI of U270 remains propagation delays, which allows the first digit to arrive
frorn the latches while RBI
decoded inhibited and the Ripple Blanking Output (pin 4) will be LO. If the digit RBO goes HI. The ,RBO
of U263B and scan-clock HI-to-LO transition occurs. Thus, if the first digit output until the first non-zero digit comes through the decoder. When the first non-zero digit arrives, the outputs of U270 are enabled and the digit
RBO output and allowing the TS7 sequence.
is
a
zero, the output to the display LED will be
is
not a zero, the outputs are enabled and
is
transferred to the output when the next
is
a
zero, pin 5 of U270
at
pin 4
all
succeeding digits to be displayed through
for a few nanoseconds due to
is
LO. If this first digit being
is
applied to the D input (pin 12)
is
held LO, inhibiting the
is
displayed. Also, the
is
set HI, removing the RBI from pin 5
TSO,
a
(TS1
LO
),
a
is is
is
is
the
Seven-Segment Decoder and Display LED's
U270
is a BCD-to-seven decoder. output of the latches, then supplies current to the appropriate cathodes of the enabled LED to display the correct number. The display LED's are DS280 through
DS286. When looking DS280 controls the numerical digit displayed
(lo6),
DS281 controls the second (1 seven segments, arranged so that segments forms
lighted, an
"8"
is
at
the front panel of the DC 502,
a
number. When
formed.
It
accepts the BCD
at
the far left
o5
),
etc. Each LED has
a
combination of lighted
all
of the segments are
When the scan gets past the decimal point in the display, or if the display overflows, any zeroes arriving decoder should be displayed. This TS5
is
inverted by U267E and applied through OR gate
U245B pin 5 of U270 HI, preventing zero-blanking during the TS5, TS6, and TS7 time slots. The location of the decimal point in the display
INTERVAL switch. The proper information the closed contacts of the switch to either NAND gate
U246A or U246B. Then either
as
a
LO
at
the direct-set input of U263B. This holds
is
determined by the MEASUREMENT
TS3
is
achieved
or
TS4
is
enabled to the
as
follows:
is
applied via
at
the
1
MHz
clock
U
U
U
U
U
U
U
U
U
Theory
U
of
Operation-DC
U
502
RESET
RBI,
pin
5
U270
2-2.
Fig.
input of OR gate U245B via these NAND gates, setting U263B to the non-blank the case where the counter overflows, the HI output from non-blank and lamp-test functions to the decoder. This U245A non-blan k
is
applied to U245B, setting U263B to the causes
state.
state
Multiplexing circuit ladder diagram showing timing with an all-zero display.
at
the appropriate time. In goes LO, overriding the output of U263B, applying the
SUPPLIES

Regulated Power Supplies

The DC 502 operating power module mainframe and then electronically regulated to provide stable supplies of +I5 volts, +5 volts, -5.2 volts, and -10 volts. The U300, provides the reference for the remaining supplies. Its output
is
set
+15-volt supply, whose active device
to
+I
5 V by adjustment of R305.
is
obtained from the power
Dl D2
r----
?----‘I
AND
,
When the front-panel RESET button
all
seven segments in the display LED to be turned
on.
I
Forced by RESET
-
is
pushed, RESET
LINES
INT SCAN DISABLE: A LO applied to this line disables
the internal scan clock.
is
EXT SCAN: Provides input for an external scan clock.
Integrated circuit U320 regulates the +5-volt supply, and
transistors 0330 and 0340 regulate the -5.2-volt and
-1 0-volt supplies respectively. The series-pass transistors
for these supplies are located in the mainframe, where they can provide the proper heat dissipation.

Input/Output Lines

The following inputs and outputs are available via the
plug-in connector to external equipment. See Fig. 1-3, also.
INT SCAN CLOCK OUT: Provides output for the
internal scan clock.
TSO: A LO
state.
DATA GOOD: A HI
a
new reading
latches.
is
present on this output line in the TSo
is
present on this output line when
is
being transferred into the storage-register
Loading...
+ 29 hidden pages