Tektronix DAQ6510 Service

www.tek.com/keithley
Model DAQ6510 Data Acquisition and Multimeter System
Calibration and Adjustment Manual
DAQ6510-905-01 Rev. D June 2022
*PDAQ6510-905-01D*
DAQ6510-905-01D
Data Acquisition and Multimeter System
DAQ6510
Calibration and Adjustment Manual
© 2022, Keithley Instruments, LLC
Cleveland, Ohio, U.S.A.
All rights reserved.
Any unauthorized reproduction, photocopy, or use of the information herein, in whole or in part,
without the prior written approval of Keithley Instruments, LLC, is strictly prohibited.
These are the original instructions in English.
TSPTM and TSP-LinkTM are trademarks of Keithley Instruments, LLC. All Keithley Instruments
product names are trademarks or registered trademarks of Keithley Instruments, LLC. Other brand
names are trademarks or registered trademarks of their respective holders.
The Lua 5.0 software and associated documentation files are copyright © 1994-2008, Tecgraf,
PUC-Rio. Terms of license for the Lua software and associated documentation can be accessed at
the Lua licensing site (http://www.lua.org/license.html).
Microsoft, Visual C++, Excel, and Windows are either registered trademarks or trademarks of
Microsoft Corporation in the United States and/or other countries.
Document number: DAQ6510-905-01 Rev. D June 2022

Safety precautions

The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient overvoltages often associated with local AC mains connections. Certain Keithley measuring instruments may be connected to mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are defined in the specifications and operating information and shown on the instrument panels, test fixture panels, and switching cards.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground) connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid interlock.
If a screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
The symbol on an instrument means caution, risk of hazard. The user must refer to the operating instructions located in the user documentation in all cases where the symbol is marked on the instrument.
The symbol on an instrument means warning, risk of electric shock. Use standard safety precautions to avoid personal contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains hazards that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the warranty.
The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or minor injury or damage the instrument. Always read the associated information very carefully before performing the indicated procedure. Damage to the instrument may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power transformer, test leads, and input jacks — must be purchased from Keithley. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley office for information.
Unless otherwise noted in product-specific literature, Keithley instruments are designed to operate indoors only, in the following environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2.
To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Safety precaution revision as of June 2018.

Table of contents

Introduction ............................................................................................................... 1-1
Welcome .............................................................................................................................. 1-1
Introduction to this manual ................................................................................................... 1-1
Extended warranty ............................................................................................................... 1-2
Contact information .............................................................................................................. 1-2
Performance verification .......................................................................................... 2-1
Introduction .......................................................................................................................... 2-1
Verification test requirements .............................................................................................. 2-2
Environmental conditions .......................................................................................................... 2-2
Warmup period .......................................................................................................................... 2-2
Line power................................................................................................................................. 2-3
Recommended test equipment ................................................................................................. 2-3
Calibration verification limits ................................................................................................ 2-4
Example reading limit calculation .............................................................................................. 2-4
Calculating resistance reading limits ......................................................................................... 2-4
Performing the verification test procedures ......................................................................... 2-5
Test summary ........................................................................................................................... 2-5
Test considerations ................................................................................................................... 2-5
Front-panel calibration verification ....................................................................................... 2-6
DC voltage verification .............................................................................................................. 2-6
AC voltage verification .............................................................................................................. 2-9
Digitize dc voltage verification ................................................................................................. 2-14
Frequency verification ............................................................................................................. 2-17
Simulated thermocouple type J temperature verification ......................................................... 2-18
Simulated RTD temperature verification ................................................................................. 2-20
Resistance verification ............................................................................................................ 2-24
DC current verification ............................................................................................................. 2-29
Digitize current verification ...................................................................................................... 2-33
AC current verification ............................................................................................................. 2-36
Capacitance verification .......................................................................................................... 2-39
Verifying zero values using a 4-wire short ............................................................................... 2-41
Rear-panel verification ....................................................................................................... 2-43
Adjustment ................................................................................................................ 3-1
Introduction .......................................................................................................................... 3-1
Environmental conditions ..................................................................................................... 3-2
Temperature and relative humidity ............................................................................................ 3-2
Line power................................................................................................................................. 3-2
Warmup period ..................................................................................................................... 3-2
Adjustment overview ............................................................................................................ 3-3
Recommended test equipment ............................................................................................ 3-3
General adjustment considerations ..................................................................................... 3-4
Initial instrument setup ......................................................................................................... 3-5
Table of contents
Calibration and Adjustment Manual
DAQ6510 Data Acquisition and Multimeter System
Select the correct terminals ....................................................................................................... 3-5
Select the TSP command set .................................................................................................... 3-5
Verify instrument date and time ................................................................................................. 3-6
Set up remote connections........................................................................................................ 3-6
Unlock calibration ...................................................................................................................... 3-6
Remote calibration adjustment procedures ......................................................................... 3-7
Disable temperature correction ................................................................................................. 3-7
Front-terminal adjustment with a 4-wire short ........................................................................... 3-7
Rear-terminal adjustment with a 4-wire short .......................................................................... 3-10
Front-terminal adjustment with open circuit inputs .................................................................. 3-11
Rear-terminal adjustment with open circuit inputs ................................................................... 3-13
Resistance adjustment ............................................................................................................ 3-14
DC voltage adjustment ............................................................................................................ 3-16
DC current adjustment ............................................................................................................ 3-17
AC voltage adjustment ............................................................................................................ 3-21
AC current adjustment ............................................................................................................ 3-23
Frequency adjustment ............................................................................................................. 3-25
Complete list of calibration commands .................................................................................... 3-26
Enable temperature correction ........................................................................................... 3-38
Save calibration and set the adjustment dates .................................................................. 3-39
Setting time, adjustment, and verification dates ................................................................ 3-39
Adjustment command timing and error checking ............................................................... 3-40
Handling events ................................................................................................................. 3-40
TSP command reference .......................................................................................... 4-1
TSP commands .................................................................................................................... 4-1
Introduction ............................................................................................................................... 4-1
cal.adjust.count ......................................................................................................................... 4-2
cal.adjust.date ........................................................................................................................... 4-3
cal.adjust.step.setup() ............................................................................................................... 4-4
cal.adjust.step.execute() ........................................................................................................... 4-5
cal.lock() .................................................................................................................................... 4-6
cal.password ............................................................................................................................. 4-7
cal.save()................................................................................................................................... 4-8
cal.unlock() ................................................................................................................................ 4-9
cal.verify.date .......................................................................................................................... 4-10
Contact information ...................................................................1-2
In this section:

Welcome

The DAQ6510 is a 6½-digit data acquisition and logging multimeter system that has a touchscreen user interface that enables faster setup time, real-time monitoring of test status, and detailed analysis on the instrument.
Section 1

Introduction

Welcome ...................................................................................1-1
Introduction to this manual ........................................................1-1
Extended warranty ....................................................................1-2
This manual provides information on completing verification and adjustment procedures for your DAQ6510.

Introduction to this manual

This manual provides instructions to help you calibrate and adjust your DAQ6510. In this manual, calibration refers to the process of verifying that the accuracy of the instrument is within its one-year accuracy specifications. Also, adjustment refers to the process of changing the calibration constants so that the accuracy of the instrument is within its one-year accuracy specifications.
This manual presents calibration information, adjustment information, and command descriptions for the calibration and adjustment commands.
For additional command descriptions, refer to the DAQ6510 Reference Manual available on the
Product Support and Downloads web page (tek.com/en/support/product-support
to the Release Notes for your instrument that contains relevant information on improvements, changes, and known issues.
). Additionally, refer
Section
Calibration and Adjustment Manual
1: Introduction DAQ6510 Data Acquisition and Multimeter System

Extended warranty

Additional years of warranty coverage are available on many products. These valuable contracts protect you from unbudgeted service expenses and provide additional years of protection at a fraction of the price of a repair. Extended warranties are available on new and existing products. Contact your local Keithley Instruments office, sales partner, or distributor for details.

Contact information

If you have any questions after you review the information in this documentation, please contact your local Keithley Instruments office, sales partner, or distributor. You can also call the Tektronix corporate headquarters (toll-free inside the U.S. and Canada only) at 1-800-833-9200. For worldwide contact numbers, visit tek.com/contact
.
1-2 DAQ6510-905-01 Rev. D June 2022
Rear-panel verification ............................................................2-43
In this section:
Introduction ...............................................................................2-1
Verification test requirements ....................................................2-2
Calibration verification limits......................................................2-4
Performing the verification test procedures ...............................2-5
Front-panel calibration verification ............................................2-6

Introduction

Use the procedures in this section to verify that DAQ6510 accuracy is within the limits stated in the instrument’s one-year accuracy specifications. Specifications and characteristics are subject to change without notice; refer to the Product Support and Downloads web page (tek.com/en/support/product-support) for the most recent specifications.
Section 2

Performance verification

You can use these verification procedures to:
Make sure that the instrument was not damaged during shipment.
Verify that the instrument meets factory specifications.
Determine if adjustment is required.
Verify that adjustment was done properly.
Although the following tests are based on the Model 7700 20-Channel Differential Multiplexer Module, the same general procedures can be used for other switching modules that have similar capabilities. Refer to the user's manual for your specific module for information on terminal connections.
The information in this section is intended for qualified service personnel only, as described by the types of product users in the Safety precautions pages, provided at the beginning of this document. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, that if contacted, could
ause personal injury or death. Use appropriate safety precautions when working with
c hazardous voltages.
Section
Calibration and Adjustment Manual
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
If the instrument is still under warranty and its performance is outside specified limits, please contact your local Keithley Instruments office, sales partner, or distributor. You can also call the Tektronix corporate headquarters (toll-free inside the U.S. and Canada only) at 1-800-833-9200. For worldwide contact numbers, visit tek.com/contact
.

Verification test requirements

Be sure that you perform these verification tests:
Under the proper environmental conditions.
After the specified warmup period.
Using the correct line voltage.
Using the proper test equipment.
Using the specified output signal and reading limits.

Environmental conditions

Conduct the calibration verification procedures in a test environment with:
An ambient temperature of 18 °C to 28 °C.
A relative humidity of less than or equal to 80 percent, unless otherwise noted.
No direct airflow on the input terminals.

Warmup period

Allow the DAQ6510 to warm up for at least 30 minutes before conducting the calibration verification procedures.
If the instrument has been subjected to temperature extremes (more than 5 °C above or below T allow additional time for the internal temperature of the instrument to stabilize. Typically, allow an additional 30 minutes to stabilize an instrument that is 10 °C outside the specified temperature range.
Also allow the test equipment to warm up for the time recommended by the manufacturer.
CAL
),
2-2 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Fluke
5720A or 5730A
High-Performance
DCV, ACV, ACI, and
See following Fluke
5725A
Amplifier
DCI and ACI
See following note.
Multimeter
note.
Keithley Instruments
3390
Function/Arbitrary Waveform Generator
Frequency
See following note.
IET Labs, Inc.
1423-A
Precision Decade
Capacitance, 1 nF to
See following
IET Labs, Inc.
HACS-Z-A-2E-1uF
Series HACS-Z High
Capacitance Box
Capacitance, 1 µF to
See following Keithley
8610 or 8620
4-Wire DMM Shorting
DCV, digitize DCV, and
See following
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:

Line power

The DAQ6510 requires a line voltage of 100 V to 240 V and a line frequency of 400 Hz, 50 Hz or 60 Hz. Calibration verification tests should be performed within this range.
The instrument automatically senses the line frequency at power-up.

Recommended test equipment

The following table summarizes the recommended calibration verification equipment. You can use alternate equipment if that equipment has specifications that meet or exceed those listed in the table below. Test equipment uncertainty adds to the uncertainty of each measurement. Generally, test equipment uncertainty should be at least four times more accurate than corresponding DAQ6510 specifications.
In this manual, the Model 8610 shorting plug is shown in the figures. However, you can use either the Mode
Manufacturer Model Description Used for Uncertainty
Fluke 8508A or 8588A 8.5-Digit Reference
Instruments
l 8610 or the Model 8620 shorting plug.
Multifunction Calibrator
Capacitor
Accuracy Decade
Plug
resistance
DCI See following
1 µF
100 µF
resistance
note.
note.
note.
note.
Refer to the manufacturer's specifications to calculate the uncertainty, which varies for each function and range test point.
DAQ6510-905-01 Rev. D June 2022 2-3
Section
Calibration and Adjustment Manual
2: Performance verification DAQ6510 Data Acquisition and Multimeter System

Calibration verification limits

The calibration verification limits stated in this section have been calculated using only the DAQ6510 one-year accuracy specifications and ambient temperature ±5 °C from T the instrument was calibrated). They do not include test equipment uncertainty. If a particular measurement falls outside the allowable range, recalculate new limits based on both the DAQ6510 specifications and corresponding test equipment specifications.
Specifications and characteristics are subject to change without notice; please refer to
tek.com/keithley

Example reading limit calculation

Assume you are testing the 10 V dc range using a 10 V input value. Using the DAQ6510 one-year accuracy specification for 10 V dc of ± (25 ppm of reading + 5 ppm of range), the calculated limits are:
Reading limits = 10 V ± [(10 V × 25 ppm) + (10 V × 5 ppm)]
for the most recent specifications.
(the temperature at which
CAL
Reading limits = 10 V ± (0.00025 + 0.00005) V
Reading limits = 10 V ± 0.00030 V
Reading limits = 9.99970 V to 10.00030 V

Calculating resistance reading limits

Resistance reading limits must be recalculated based on the actual calibration resistance values supplied by the equipment manufacturer. Calculations are performed in the same manner as shown in the preceding example. Use the actual calibration resistance values instead of the nominal values in the example when performing your calculations.
For example, assume that you are testing the 10 kΩ range using an actual 10.03 kΩ calibration resistance value. Using DAQ6510 one-year 10 kΩ range accuracy of ± (75 ppm of reading + 6 ppm of range), the calculated reading limits are:
Reading limits = 10.03 ± [(10.03 kΩ x 75 ppm) + (10 kΩ x 6 ppm)]
Reading limits = 10.03 kΩ ± [(0.7523) + (0.06)] Ω
Reading limits = 10.03 kΩ ± 0.8123 Ω
Reading limits = 10.029188 kΩ to 10.030812
2-4 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:

Performing the verification test procedures

The following topics provide a summary of calibration verification test procedures and items to consider before performing any calibration verification test.

Test summary

Front-panel tests:
DC voltage verification (on page 2-6)
AC voltage verification (on page 2-9)
Digitize voltage verification (on page 2-14)
Frequency verification (on page 2-17)
Simulated thermocouple type J temperature verification (on page 2-18)
Simulated RTD temperature verification (on page 2-20)
Resistance verification (on page 2-24)
DC current verification (on page 2-29)
Digitize current verification (on page 2-33)
AC current verification (on page 2-36)
Capacitance verification (on page 2-39)
Verifying zero values using a 4-wire short (on page 2-41)

Test considerations

When performing the calibration verification procedures:
Be sure to restore factory front-panel defaults. From the front panel, select the MENU key, select
Info/Manage, and select System Reset.
Make sure that the test equipment is warmed up for the time recommended by the manufacturer
and is connected to the DAQ6510 input/output terminals.
Make sure that the correct DAQ6510 terminals are selected with the TERMINALS FRONT/REAR
switch.
Make sure the test equipment is set up for the proper function and range.
Do not connect test equipment to the DAQ6510 through a scanner, multiplexer, or other switching
DAQ6510-905-01 Rev. D June 2022 2-5
equipment.
Section
Calibration and Adjustment Manual
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
The front and rear terminals of the instrument are rated for connection to circuits rated Measurement Category II up to 300 V, as described in International Electrotechnical Commission (IEC) Standard IEC 60664. This range must not be exceeded. Do not connect the instrument terminals to CAT III or CAT IV circuits. Connection of the instrument terminals to circuits higher than CAT II can cause damage to the equipment and severe personal injury.

Front-panel calibration verification

The following topics describe verification procedures that are done with connections attached to the terminals on the DAQ6510 front panel.

DC voltage verification

The maximum input voltage between INPUT HI and INPUT LO is 1000 V dc and 750 V ac. Exceeding this value may create a shock hazard.
The maximum common-mod 500 V
. Exceeding this value may cause a breakdown in insulation that can create a shock
PEAK
e voltage (the voltage between INPUT LO and chassis ground) is
hazard.
Verify dc voltage accuracy for the 100 mV to 1000 V ranges
To verify 100 mV to 1000 V dc voltage accuracies, you will:
Apply accurate dc voltages from the calibrator to the DAQ6510 front-panel terminals.
Verify that the displayed readings are within specified limits.
Use the values in the tables following the steps below to verify the performance of the DAQ6510.
Actual values depend on the published specifications (see Example reading limit calculation (on page 2-4)).
Use shielded low-ther caused by noise or thermal effects. Connect the shield to the output LO terminal of the calibrator.
mal connections when testing the 100 mV and 1 V ranges to avoid errors
2-6 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
To verify dc voltage accuracy:
1. Use a low-thermal cable to connect the DAQ6510 HI and LO INPUT terminals to the calibrator HI and LO terminals as shown in the following figure.
Figure 1: DC voltage 100 mV to 1000 V ranges verification connections
2. On the DAQ6510, press the FUNCTION key and select DC voltage.
3. On the home screen, select the button next to Range and select 100 mV.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Set Input Impedance to Auto.
7. Set the calibrator output to 0 V.
8. Set the calibrator to OPERATE.
9. Allow 5 minutes of settling time.
10. Press the MENU key.
11. Select Calculations.
12. [Only for the 100 mV range] Select Rel Acquire.
13. Source positive and negative full-scale and half-scale voltages and allow for proper settling.
14. Select each range on the DAQ6510, allow for proper settling, and verify the ranges according to the following tables.
DAQ6510-905-01 Rev. D June 2022 2-7
Section
Calibration and Adjustment Manual
Perform relative offset
0
-0.01
0.01 Full scale (+)
100
99.9935
100.0065
Half scale (–)
-50
-50.005
-49.995
Full scale (–)
-100
-100.0065
-99.9935
Full scale (+)
1
0.999969
1.000031
Half scale (+)
0.5
0.499979
0.500021
Half scale (–)
-0.5
-0.500021
-0.499979
Full scale (–)
–1
–1.000031
-0.999969
Full scale (+)
10
9.9997
10.0003
Half scale (+)
5
4.99982
5.00018
Half scale (–)
–5
–5.00018
–4.99982
Full scale (–)
–10
–10.0003
–9.9997
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Verify the dc voltage 100 mV range
Description Nominal value Lower limit Upper limit
Half scale (+) 50 49.995 50.005
Verify the dc voltage 1 V range
Description Nominal value Lower limit Upper limit
Verify the dc voltage 10 V range
Description Nominal value Lower limit Upper limit
2-8 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Full scale (+)
100
99.954
100.0046
Half scale (–)
-50
-50.00315
-49.99685
Full scale (–)
-100
-100.0054
-99.9946
Full scale (+)
1000
999.944
1000.056
Half scale (+)
500
499.974
500.026
Half scale (–)
-500
-500.026
-499.974
Full scale (–)
-1000
-1000.069
-999.931
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
Verify the dc voltage 100 V range
The information in this section is intended for qualified service personnel only, as described by the types of product users in the Safety precautions pages, provided at the beginning of this document. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, that if contacted, could c
ause personal injury or death. Use appropriate safety precautions when working with
hazardous voltages.
Description Nominal value Lower limit Upper limit
Half scale (+) 50 49.99685 50.00315
Verify the dc voltage 1000 V range
Description Nominal value Lower limit* Upper limit*
*For each additional volt over ±500 V, add 0.02 mV of uncertainty.

AC voltage verification

To verify ac voltage accuracy:
For the 100 mV to 100 V ranges, apply accurate voltages from the calibrator to the DAQ6510
front-panel terminals.
For the 750 V range, connect the Fluke 5725A Amplifier to the calibrator. Apply accurate voltages
from the calibrator terminals to the terminals on the front panel of the DAQ6510.
Verify that the displayed readings are within specified limits.
Use the values in the tables following the steps below to verify the performance of the DAQ6510. Actual values depend on the published specifications (see Example reading limit calculation (
on page
2-4)).
DAQ6510-905-01 Rev. D June 2022 2-9
Section
Calibration and Adjustment Manual
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
The maximum input voltage between INPUT HI and INPUT LO is 750 V dc. Exceeding this value may create a shock hazard.
The maximum common-mo 500 V
. Exceeding this value may cause a breakdown in insulation that can create a shock
PEAK
de voltage (the voltage between INPUT LO and chassis ground) is
hazard.
Verify ac voltage accuracy for the 100 mV to 100 V ranges
Use shielded, low-capacitance cabling. For the 100 mV to 100 V ranges, avoid loading that exceeds 1000 pF.
Excessive capacitance may result in additional load regulation uncertainties and could cause the cal
ibrator output to open (go into standby).
To verify ac voltage accuracy:
1. Connect the DAQ6510 HI and LO INPUT connectors to the calibrator as shown in the following figure.
Figure 2: Connections for ac voltage verification 100 mV to 100 V ranges
2-10 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
0.1
1 kHz
99.91
100.09
0.1
100 kHz
99.32
100.68
1
1 kHz
.9991
1.0009
1
100 kHz
.9932
1.0068
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
2. On the DAQ6510, press the FUNCTION key and select AC voltage.
3. On the home screen, select the button next to Range and select 100 mV.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Make sure that detector bandwidth is set to 30 Hz.
AC voltage is specified for the detector bandwidth setting of 3 Hz. Three Hz measures accurately for input signals from 3 Hz to 300 kHz, with reading rates ≈ 2 readings/s. To improve verification throughput to ≈ 20 readings/s, set detector bandwidth to 30 Hz for frequencies of 30 Hz to 300 kHz. To verify frequencies 1 kHz and higher, set the detector bandwidth to 300 Hz for faster ≈ 200 readings/s throughput.
7. Source ac voltages for each of the frequencies listed in the Verify the ac voltage 100 mV range (on page 2-11) table.
8. Repeat these steps for each range and frequency listed in the tables below. For each voltage setting, be sure that the reading is within low and high limits.
Verify the ac voltage 100 mV range
Nominal value Frequency Lower limit Upper limit
0.1 20 Hz 99.91 100.09
0.1 50 kHz 99.83 100.17
Verify the ac voltage 1 V range
Nominal value Frequency Lower limit Upper limit
1 20 Hz .9991 1.0009
1 50 kHz .9983 1.0017
DAQ6510-905-01 Rev. D June 2022 2-11
Section
Calibration and Adjustment Manual
10
20 Hz
9.991
10.009
10
1 kHz
9.991
10.009
10
50 kHz
9.983
10.017
10
100 kHz
9.932
10.068
100
20 Hz
99.910
100.09
100
50 kHz
99.830
100.17
100
100 kHz
99.320
100.68
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Verify the ac voltage 10 V range
Nominal value Frequency Lower limit Upper limit
Verify the ac voltage 100 V range
Nominal value Frequency Lower limit Upper limit
100 1 kHz 99.910 100.09
Verify ac voltage accuracy for the 750 V range
Use shielded low capacitance cabling. For the 750 V range, avoid cable capacitances of >150 pF.
Excessive capacitance may result in additional load regulation uncertainties and could cause the
alibrator output to open (go into standby).
c
2-12 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
To verify ac voltage accuracy for the 750 V range:
1. Put the calibrator in Standby.
2. Connect the DAQ6510 HI and LO INPUT connectors to the calibrator as shown in the following figure.
3. For 750 V at 50 kHz and 100 kHz outputs, connect the calibrator to the Fluke 5725A amplifier.
Figure 3: Connections for ac voltage accuracy verification 750 V range
4. On the DAQ6510, press the FUNCTION key and select AC voltage.
5. On the home screen, select the button next to Range and select 750 V.
6. Press the MENU key.
7. Under Measure, select Settings.
8. Ensure that detector bandwidth is set to 30 Hz.
AC voltage is specified for the detector bandwidth setting of 3 Hz. Three Hz measures accurately for input signals from 3 Hz to 300 kHz, with reading rates ≈ 2 readings/s. To improve verification throughput to ≈ 200 readings/s, set detector bandwidth to 30 Hz for frequencies of 30 Hz to 300 kHz. To verify frequencies 1 kHz and higher, set the detector bandwidth to 300 Hz for faster ≈ 200 readings/s throughput.
DAQ6510-905-01 Rev. D June 2022 2-13
Section
Calibration and Adjustment Manual
740
50 Hz
739.33
740.67
740
50 kHz
738.66
741.34
740
100 kHz
734.96
745.04
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
9. Set the calibrator to OPERATE.
10. Source ac voltages for each of the frequencies listed in the "Verify the ac voltage 750 V range" table, below. Be sure that the readings are within low and high limits.
Verify the ac voltage 750 V range
Nominal value Frequency Lower limit Upper limit
740 1 kHz 739.33 740.67

Digitize dc voltage verification

To verify digitize dc voltage accuracy, you will:
Apply accurate voltages from the calibrator to the terminals on the front panel of the DAQ6510.
Verify that the displayed readings are within specified limits.
Use the values in the tables following the steps below to verify the performance of the DAQ6510. Actual values depend on the published specifications (see Example reading limit calculation (on page 2-4)).
The maximum input voltage between INPUT HI and INPUT LO is 1000 V dc and 750 V ac. Exceeding this value may create a shock hazard.
The maximum common-mo 500 V
. Exceeding this value may cause a breakdown in insulation that can create a shock
PEAK
de voltage (the voltage between INPUT LO and chassis ground) is
hazard.
Verify the digitize voltage 100 mV to 1000 V ranges
Use shielded low-thermal connections when testing the 100 mV and 1 V ranges to avoid errors caused by noise or thermal effects. Connect the shield to the output LO terminal of the calibrator.
2-14 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
To verify digitize voltage accuracy:
1. Connect the DAQ6510 HI and LO INPUT connectors to the calibrator as shown in the following figure.
Figure 4: Connections for digitize voltage verification 100 mV to 1000 V ranges
2. On the DAQ6510, press the FUNCTION key, select the Digitize Functions tab, and select Digitize Voltage.
3. On the home screen, select the button next to Range and select 100 mV.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Set the Sample Rate to 1000.
7. Set the Aperture to Auto.
8. Set the Count to 100.
9. Set the calibrator output to 0.00000 mV dc and allow the reading to settle.
10. Press the MENU key.
11. Under Measure, select Calculations.
12. Source positive and negative full-scale and half-scale voltages, as listed in the following table. Verify the 100 mV to 100 V range settings listed in the tables below. For each voltage setting, verify that the STATISTICS swipe screen reading for Average is within low and high limits.
The Fluke 5720A or 5730A calibrator 1000 V range 0.0 V setting is not verified.
DAQ6510-905-01 Rev. D June 2022 2-15
Section
Calibration and Adjustment Manual
Full scale (+)
100
99.94
100.06
Half scale (+)
50
49.95
50.05
Half scale (–)
-50
-50.05
-49.95
Full scale (–)
-100
-100.06
-99.94
Full scale (+)
1
0.9996
1.0004
Half scale (–)
-0.5
-0.50025
-0.49975
Full scale (–)
-1
-1.0004
-0.9996
Full scale (+)
10
9.996
10.004
Half scale (+)
5
4.9975
5.0025
Half scale (–)
-5
-5.0025
-4.9975
Full scale (–)
-10
-10.004
-9.996
Full scale (+)
100
99.96
100.04
Half scale (+)
50
49.975
50.025
Half scale (–)
-50
-50.025
-49.975
Full scale (–)
-100
-100.04
-99.96
Full scale (+)
1000
999.6
1000.4
Half scale (+)
500
499.75
500.25
Half scale (–)
–500
-500.25
-499.75
Full scale (–)
–1000
-1000.4
-999.6
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Verify the digitize voltage 100 mV range
Description Nominal value Lower limit Upper limit
Verify the digitize voltage 1 V range
Description Nominal value Lower limit Upper limit
Half scale (+) 0.5 0.49975 0.50025
Verify the digitize voltage 10 V range
Description Nominal value Lower limit Upper limit
Verify the digitize voltage 100 V range
Description Nominal value Lower limit Upper limit
Verify the digitize voltage 1000 V range
Description Nominal value Lower limit Upper limit
2-16 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:

Frequency verification

To verify frequency accuracy, you will:
Apply accurate frequencies from the function generator to the terminals on the front panel of the
DAQ6510.
Verify that the displayed readings are within specified limits.
Use the values in the table following the steps below to verify the performance of the DAQ6510. Actual values depend on the published specifications (see Example reading limit calculation 2-4)).
1. Connect the Keithley Instruments Model 3390 function generator to the DAQ6510 INPUT HI and LO terminals as shown in the following figure.
Figure 5: Connections for frequency verification and adjustment
(on page
2. On the DAQ6510, press the FUNCTION key, select the Measure Functions tab, and select Frequency.
3. Select the MENU key.
4. Under Measure, select Settings.
5. Set the Aperture to 250 ms.
6. Set the Threshold Range to 10 V.
7. Press the HOME key.
8. Source the voltage and frequency values as listed in Verify the frequency (on page 2-18
). For
each setting, be sure that the reading is within low and high limits.
DAQ6510-905-01 Rev. D June 2022 2-17
Section
Calibration and Adjustment Manual
10 Hz at 10 V
10
9.997
10.003
10 kHz at 5 V
10
9.9991
10.0009
100 kHz at 5 V
100
99.991
100.009
300 kHz at 5 V
300
299.973
300.027
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Verify the frequency
Use the following values to verify the performance of the DAQ6510. Actual values depend on published specifications (see Example reading limit calculation (on page 2-4)).
Description Frequency (Hz) Lower limit (Hz) Upper limit (Hz)
1 kHz at 5 V 1 0.9999 1.0001

Simulated thermocouple type J temperature verification

To verify thermocouple accuracy, you will:
Apply accurate voltages from the calibrator to the terminals on the front panel of the DAQ6510.
Verify that the displayed readings are within specified limits.
Thermocouple accuracy is verified by using a dc voltage calibrator to output values from standard thermocouple tables available from the National Institute of Standards and Technology (NIST) or other sources.
In the table following the steps below, three representative values are listed from a type J thermocouple table for temperatures –190 °C, 0 °C, and 750 °C, with their respective thermocouple voltages listed in the “Uncompensated calibrator source value” column. The calibrator source values are based on NIST Monograph 175, reference data 60, version 2.0.
Verify thermocouple accuracy
Because the cable connecting the calibrator to the DAQ6510 can have non-trivial thermal offset voltages, you must first correct for these to verify the DAQ6510 specifications.
2-18 DAQ6510-905-01 Rev. D June 2022
DAQ6510
cation
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2: Performance verifi
To verify the simulated thermocouple type J temperature:
1. Connect the DAQ6510 HI and LO INPUT terminals to the calibrator HI and LO terminals as shown in the following figure.
Figure 6: Connections for thermocouple verification
2. On the DAQ6510, press the FUNCTION key and select DC voltage.
3. Press the MENU key.
4. Under Measure, select Settings.
5. Set the range to 100 mV.
6. Set Input Impedance to Auto.
7. Set autozero to On.
8. Select Integration Rate. The Integration Rate dialog box opens.
9. Set the unit to NPLC.
10. Set NPLC to 1 PLC.
11. Select OK and press the HOME key to return to the Home Screen.
12. Set the calibrator to output 0 V and enable the output.
13. Allow five minutes for settling of the thermal voltage.
14. Record the measured offset voltage to 1 µV precision. If necessary, use the DAQ6510 filter settings to reduce the noise of this measurement (for filter settings, go to MENU > Measure Calculations).
15. Press the DAQ6510 FUNCTION key and select Temperature.
16. Press the MENU key.
17. Under Measure, select Settings.
DAQ6510-905-01 Rev. D June 2022 2-19
Section
Calibration and Adjustment Manual
–190 °C
–7.659 mV
–190.2 °C
–189.8 °C
0 °C
0.000 mV
–0.2 °C
0.2 °C
750 °C
42.281 mV
749.8 °C
750.2 °C
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
18. On the Measure Settings screen, set the following values:
Units: °C
Transducer: TC
Thermocouple: J
Temperature (simulated reference temperature): 0 °C
Integration Rate: 1 PLC
Auto Zero: On
19. Set the calibrator to the simulated thermocouple voltage you want (from the following table), first correcting for the offset voltage measured in step 14. For example, if the measured offset voltage was –2 µV, set the calibrator to –7.659 mV – (–0.002 mV), which equals –7.657 mV, to simulate – 190 °C.
20. Verify that the DAQ6510 reading is within lower and upper limits.
21. Repeat steps 18 and 19 for each value in the following table.
Use the following values to verify the performance of the DAQ6510. Actual values depend on published specifications (see Example reading limit calculation (on page 2-4
)).
Simulated temperature
Uncompensated calibrator source value (V)
Lower limit Upper limit

Simulated RTD temperature verification

Use the following information to verify the performance of the DAQ6510. Actual calibrator source values will vary. RTD verification is based on the calibrator sourcing resistance and the DAQ6510 conversion of the resistance measurement to calculated temperature based on the Callendar-Van Dusen equation.
To verify RTD temperature accuracy, you will:
Apply accurate resistance from the calibrator to the terminals on the front panel of the DAQ6510.
Verify that the displayed readings are within specified limits.
2-20 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
RTD equations
The temperature versus resistance readings listed in the RTD reference tables are calculated using the Callendar-Van Dusen equation. There are two equations that are based on different temperature ranges. There is an equation for the –200 °C to 0 °C range and one for the 0 °C to 850 °C range.
Equation for –200 °C to 0 °C temperature range
R
= R0 [1 + AT + BT2 + CT3(T – 100)]
RTD
where:
R
R
is the calculated resistance of the RTD
RTD
is the known RTD resistance at 0 °C
0
T is the temperature in °C
A = alpha [1 + (delta/100)]
B = –1 (alpha)(delta)(1E-4)
C = –1 (alpha)(beta)(1E-8)
The alpha, beta, and delta values are listed in the following table.
Equation for 0 °C to 850 °C temperature range
R
= R0 (1 + AT + BT2)
RTD
where:
R
R
is the calculated resistance of the RTD
RTD
is the known RTD resistance at 0 °C
0
T is the temperature in °C
A = alpha [1 + (delta/100)]
B = –1 (alpha)(delta)(1E-4)
The alpha and delta values are listed in the following table.
DAQ6510-905-01 Rev. D June 2022 2-21
Section
Calibration and Adjustment Manual
PT100
ITS-90
0.00385055
0.10863
1.49990
100.0000
D100
0.003920
0.10630
1.49710
F100
0.003900
0.11000
1.49589
PT385
IPTS-68
0.003850
0.11100
1.50700
PT3916
0.003916
0.11600
1.50594
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
RTD parameters for equations
The RTD parameters for the Callendar-Van Dusen equations are listed in the following table.
DAQ6510 resistance to temperature device (RTD)
Type Standard Alpha Beta Delta R0 at 0 °C (Ω)
Verify the simulated RTD temperature
Use the values in the tables following the steps below to verify the performance of the DAQ6510. Actual values depend on the published specifications (see Example reading limit calculation 2-4)).
(on page
To verify RTD accuracy:
1. For 4-wire accuracy, connect the DAQ6510 INPUT and SENSE terminals to the calibrator as shown in the following figure.
Figure 7: Connections for 4-wire RTD accuracy verification
2-22 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
2. For 3-wire accuracy, connect the DAQ6510 INPUT and SENSE terminals to the calibrator as shown in the following figure.
The SENSE HI wire is not required for 3-wi
re RTD measurements. For 3-wire RTD, accuracy is for
< 0.1 Ω lead resistance mismatch for input HI and LO. Add 0.25 °C per 0.1 Ω of HI-LO lead resistance mismatch.
Figure 8: Connections for 3-wire RTD accuracy verification
3. On the DAQ6510, press the FUNCTION key and select Temperature.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Select Transducer.
7. Set the Type to 4-wire RTD or 3-Wire RTD.
8. Set the RTD Type to PT100.
9. Press the HOME key.
10. On the calibrator, select 19 Ω source resistance.
11. Select the OPER and EX SNS keys.
12. Record DAQ6510 accuracies.
13. Refer to the table for PT100 accuracies.
Fluke 5720 and 5730 resistance source values vary and may require new resistance-to-temp target accuracy values.
erature
DAQ6510-905-01 Rev. D June 2022 2-23
Section
tment Manual
1.000000E+02
3.850550E-03
1.086300E-01
-5.775440E-07
-4.182852E-12 100
99.99707
-0.0075
0.0235
0.2932
190
189.99234
238.6775
0.0218
0.2725
2: Performance verification DAQ6510 Data Acquisition and Multimeter System Calibration and Adjus
14. Repeat for 100 Ω and 190 Ω source values.
Example PT100
R0
alpha
beta
delta
A
B
C
1.499900E+00
3.908304E-03
Nominal calibrator value (Ω)
19 18.999520 -198.8900 0.0259 0.3241
Actual calibrator value (Ω)

Resistance verification

Use the following information to verify the performance of the DAQ6510 resistance functions.
Four-wire resistance verification
To verify the 4-wire resistance function, you will:
Use shielded, Teflon-insulated or equivalent cables in a 4-wire configuration.
Characterize the calibrator 1 Ω and 10 Ω nominal values with an external reference digital
multimeter (DMM); verify accuracy from the reference DMM readings.
Temperature (°C)
4-wire RTD 3-wire RTD
±0.06 °C accuracy
(±Ω from actual
calibrator value)
±0.75 °C accuracy (±Ω from actual calibrator value)
For the 100 Ω to 100 MΩ ranges, verify accuracy from actual calibrator source values.
2-24 DAQ6510-905-01 Rev. D June 2022
Verify that the displayed readings are within specified limits.
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
To verify 4-wire resistance accuracy:
1. Connect the DAQ6510 INPUT and SENSE terminals to the calibrator as shown in the following figure.
Figure 9: Connections for 4-wire resistance accuracy verification
2. Set the calibrator for 4-wire resistance with external sense on.
3. On the DAQ6510, press the FUNCTION key and select 4W Res.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Set Offset Compensation On.
7. Verify that Open Lead Detector is Off.
8. On the home screen, select the button next to Range and select 1 Ω.
9. Source the nominal zero and full-scale resistance values for the 1 Ω to 10 kΩ ranges. Source the nominal zero value for the 100 kΩ range. Refer to the tables in Calculated limits (on page 2-26).
10. For the 100 kΩ range, only verify 0 Ω with Offset Compensation set to On.
11. Set Offset Compensation to Off.
12. Verify full-scale 100 kΩ on the 100 kΩ range and 0 and full-scale for the 1 MΩ and 10 MΩ ranges.
13. Verify that the readings are within calculated limits.
When Offset Compensation is set to On, ranges are limited to 1
Ω to 10 kΩ. When Offset
Compensation is set to Off, all ranges (1 Ω to 100 MΩ) are available from all interfaces.
DAQ6510-905-01 Rev. D June 2022 2-25
Section
Calibration and Adjustment Manual
1
0 0 -0.1
0.1 1 0.999973
0.999688
1.000258
10
9.99983
9.99878
10.00088
100
0 0 -0.1
0.1
100
99.9975
99.987
100.008
1
0.99999
0.99991
1.00007
10 K
0 0 -1 1 10
9.99982
9.99901
10.00063
100 K
0 0 -10
10
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
You can use either the front-panel controls or remote interface commands to set measurement parameters for verification. For calibration, you must use remote interface commands. The example below is an example of remote interface commands that will generate event messages.
To do the same steps over the remote interface, send the commands:
dmm.measure.func = dmm.FUNC_4W_RESISTANCE dmm.measure.offsetcompensation.enable = dmm.OCOMP_ON dmm.measure.range = 1e6
The following warning message is displayed:
1131, Parameter, measure range, expected value from 1 to 100000
Set dmm.measure.offsetcompensation.enable = dmm.OCOMP_OFF, and then set
dmm.measure.range = 1e6 to run without warnings.
Verify that the readings are within calculated limits.
The values and limits in the following tables are for example only. You must calculate test limits based on the actual resistance values output by your calibrator or resistance source (see Example
reading limit calculation (on page 2-4)).
Range (Ω)
10 0 0 -0.1 0.1
Nominal calibrator values (Ω)
Typical reference DMM reading (Ω)
Lower limit (Ω)
Upper limit (Ω)
Range (Ω)
1 K 0 0 -1 1
Nominal calibrator values (Ω)
Actual calibrator value (Ω)
Lower limit (Ω)
Upper limit (Ω)
2-26 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
100 K
100
100.0012
99.9928
100.0098
1 M
0 0 -1 1 1
0.999966
0.99986
1.000072
10 M
0 0 -1
1
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
Four-wire resistance verification with offset compensation off
The values and limits in the following tables are for example only. You must calculate test limits based on the actual resistance values output by your calibrator or resistance source (see Example
reading limit calculation (on page 2-4)).
For 10 MΩ verification, the Sense HI cable is optional. Measurement is with Input HI and LO and
Sense LO only.
Range (Ω)
Nominal calibrator values (Ω)
10 9.99931 9.99521 10.00341
Verify 2-wire resistance accuracy
To verify the 2-wire resistance function 100 MΩ range, you will:
Use shielded, Teflon-insulated or equivalent cables in a 2-wire configuration.
Apply accurate resistance from the calibrator to the terminals on the front panel of the DAQ6510.
Verify that the displayed readings are within specified limits.
Verify resistance 100 MΩ range
To verify the 100 MΩ range:
Actual calibrator value (Ω)
Lower limit (Ω)
Upper limit (Ω)
1. Connect the DAQ6510 INPUT to the calibrator as shown in the following figure.
DAQ6510-905-01 Rev. D June 2022 2-27
Section
Calibration and Adjustment Manual
100 M
0 0 -10
10
100
100.001
99.798
100.204
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Figure 10: Connections for 100 MΩ verification
2. Set the calibrator for 2-wire resistance with external sense off.
3. On the DAQ6510, press the FUNCTION key and select 2W Res.
4. On the home screen, select the button next to Range and select 100 MΩ.
5. Source the nominal full-scale resistance values for the 100 MΩ range as shown in the following table.
The values and limits in the following tables are for example only. You must calculate test limits based on the actual resistance values output by your calibrator or resistance source (see
Example
reading limit calculation (on page 2-4)).
Range (Ω) Nominal calibrator
values (Ω)
Actual calibrator
(Ω)
Lower limit (Ω) Upper limit (Ω)
2-28 DAQ6510-905-01 Rev. D June 2022
DAQ6510
mance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2: Perfor

DC current verification

The DAQ6510 dc current ranges can be verified using several methods, depending on the level of measurement uncertainty required. This manual describes the verification procedure using a Fluke 8508A or 8588A reference digital multimeter (DMM) in series with the DAQ6510 to determine the nominal test current value for the 10 µA to 100 mA ranges. For the 1 A to 10 A ranges, this manual describes using direct output from a Fluke Model 5720A or 5725A calibrator.
These configurations are adequate for most purposes, but may not provide sufficient test uncertainty
io (TUR) for some users. You must evaluate the measurement uncertainties and ensure that they
rat are adequate for your use.
DC current 10 µA to 100 mA range verification
When verifying dc current on the 10 µA to 100 mA ranges, systematic calibrator and cable offsets must be compensated and test limits calculated based on reference digital multimeter (DMM) readings.
In the following section, offset measurements may exceed DAQ6510 zero-current measurement specifications due to systematic source offset current from the test setup.
To verify the DAQ6510 specifications with zero input current, disconnect all cables and calibrators
om the DAQ6510 input. This is a separate setup from that used in the procedure below for
fr mid-scale and full-scale readings.
DAQ6510-905-01 Rev. D June 2022 2-29
Section
Calibration and Adjustment Manual
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
To prepare the DAQ6510 for dc current accuracy verification:
1. Set up the DAQ6510 for dc current and the range being tested. Make sure relative offset is disabled.
2. Connect the calibrator, DAQ6510, and reference DMM as shown in the following figure.
Figure 11: Connection for dc current
To verify DAQ6510 accuracy for each range:
1. Set the calibrator to source zero current.
2. Set the reference DMM to DC Current and select the appropriate range to be verified. Use the Model 8508A or 8588A 200 µA range to verify the DAQ6510 10 µA and 100 µA ranges. Use the Model 8508A or 8588A 2 mA, 20 mA, and 200 mA ranges to verify the DAQ6510 1 mA, 10 mA, and 100 mA ranges, respectively.
3. On the calibrator, select the OPR/STBY key. Make sure that the front panel displays OPERATE.
4. On the DAQ6510, press the MENU key.
5. Select Calculations. The Calculation Settings screen is displayed.
6. On the reference DMM, zero the range for system offset.
7. Set the calibrator to source the current for the range you are verifying (listed in the 1 mA verification table in Test limit calculation for 10 µA to 100 mA ranges).
2-30 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
10 µA
10
0.00500
100 µA
100
0.0500
5
0.00275
50
0.0275
–5
0.00275
–50
0.0275
–10
0.00500
–100
0.0500
1 mA
1
0.000500
10 mA
10
0.00250
0.5
0.000275
5 0.00150
–0.5
0.000275
–5
0.00150
–1
0.000500
–10
0.00250
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
8. Note the offset-compensated reference DMM reading, and calculate limits based on DAQ6510 specifications (use the reference DMM reading as the expected value and verify the DAQ6510 accuracy from the calculated reference DMM current).
Test limit calculation for 10 µA to 100 mA ranges
The following tables list nominal test current for 10 µA to 100 mA ranges. Test limits must be calculated relative to actual current, as determined by the reference digital multimeter (DMM) measurement. For example, using a specification of 60 ppm of reading + 9 ppm of range on the 10 mA range, the reference DMM measures 5.00012 mA on the nominal 5 mA test.
Specification tolerance = 5.00012 (mA) × 60 ppm + 10 (mA) × 9 ppm = 0.000390072 mA
Lower test limit = 5.00012 – 0.000390072 = 4.999729928 mA
Upper test limit = 5.00012 + 0.000390072 = 5.000510072 mA
Although the specification tolerance calculated above from the actual test current differs slightly from the values listed in the table (based on nominal value), this difference is generally much smaller than the measurement uncertainty and can be ignored. As a result, the test limits can be calculated from the table specification tolerance as:
Lower test limit = 5.00012 – 0.00039 = 4.99973 mA
Upper test limit = 5.00012 + 0.00039 = 5.00051 mA
9. Repeat steps 1 through 8 for all ranges (10 µA through 100 mA).
Range Nominal
input (µA)
Specification tolerance (µA) (based on nominal)
Range Nominal
input (µA)
Specification tolerance (µA) (based on nominal)
Range Nominal
input (µA)
Specification tolerance (µA) (based on nominal)
Range Nominal
input (µA)
Specification tolerance (µA) (based on nominal)
DAQ6510-905-01 Rev. D June 2022 2-31
Section
Calibration and Adjustment Manual
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
DC current 100 mA to 3 A range verification
To verify dc current accuracy on the 100 mA to 3 A ranges, you will:
Apply accurate current from the dc current calibrator directly to the DAQ6510 front-panel
terminals.
Verify that the displayed readings are within specified limits.
To verify dc current accuracy:
1. Set up the DAQ6510 for dc current and the range being tested. Make sure that relative offset is disabled.
2. Connect the DAQ6510 and calibrator as shown in the following figure.
Figure 12: Connections for 100 mA to 3 A range verification
Zero verify the DAQ6510:
1. On the calibrator, select the OPR/STBY key. Make sure that the front panel displays STANDBY.
2. Set the ranges to 100 mA.
3. Verify the DAQ6510 zero reading for each range.
4. Source dc current from the following table. For each setting, be sure that the reading is within stated limits.
5. Repeat these steps for the 1 A and 3 A ranges.
2-32 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Half scale (+)
50
49.985
50.015
Half scale (–)
-50
-50.015
-49.985
Full scale (–)
-100
-100.025
-99.975
Full scale (+)
1
0.99955
1.00045
Half scale (+)
0.5
0.49975
0.50025
Half scale (–)
–0.5
-0.50025
-0.49975
Full scale (–)
–1
-1.00045
-0.99955
Half scale (+)
1.5
1.49913
1.50087
Half scale (–)
–1.5
–1.50087
–1.49913
* The 3 A range full-scale test points are limited to 2 A in this table because of the accuracy limitations of Fluke Models 57xxA and 5725A series calibrators at currents above 2.2 A.
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
Verify dc current 100 mA range
Description Calibrator
Full scale (+) 100 99.975 100.025
Verify dc current 1 A range
Description Calibrator
Verify dc current 3 A range
Description Calibrator
*Full scale (+) 2 1.99888 2.00112
Lower limit Upper limit
setpoint (A)
Lower limit Upper limit
setpoint (A)
Lower limit Upper limit
setpoint (A)
*Full scale (–) –2 –2.00112 –1.99888

Digitize current verification

The following topics describe how to verify digitized dc current on the DAQ6510.
DAQ6510-905-01 Rev. D June 2022 2-33
Section
on and Adjustment Manual
Full scale (+)
99.998
99.878
100.118
Half scale (+)
49.9991
49.9141
50.0841
Half scale (–)
-49.993
-50.0843
-49.9143
Full scale (–)
-99.9985
-100.1185
-99.8785
2: Performance verification DAQ6510 Data Acquisition and Multimeter System Calibrati
Verify digitize current 100 µA to 3 A ranges
To verify digitize dc current accuracy:
1. Connect the DAQ6510 and calibrator as shown in the following figure.
Figure 13: Connections for digitize dc current 10 µA to 3 A range verification
2. Press the FUNCTION key.
3. Select the Digitize Functions tab and select Digitize Current.
4. Press the HOME key.
5. Set the Range to 10 µA.
6. Press the MENU key.
7. Under Measure, select Settings.
8. Set the Sample Rate to 1000.
9. Set the Aperture to Auto or 1 ms.
10. Set the Count to 100.
11. Source positive and negative full-scale and half-scale currents, as listed in the following tables.
12. Repeat these steps for the 1 mA to 3 A range settings listed in the following tables.
Verify digitize current 100 µA range
Description Calibrator setpoint
(A)
Lower limit Upper limit
2-34 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Full scale (+)
0.999993
0.998993
1.000993
Half scale (+)
0.499998
0.499348
0.500648
Half scale (–)
-0.500009
-0.500659
-0.499359
Full scale (–)
-1.000017
-1.001017
-0.999017
Half scale (+)
5.00004
4.99454
5.00554
Half scale (–)
-5.00003
-5.00553
-4.99453
Full scale (–)
-10.00006
-10.00806
-9.99206
Full scale (+)
100
99.92
100.08
Half scale (+)
50
49.945
50.055
Half scale (–)
-50
-50.055
-49.945
Full scale (–)
-100
-100.08
-99.92
Full scale (+)
1
0.999
1.001
Half scale (+)
0.5
0.49935
0.50065
Half scale (–)
–0.5
-0.50065
-0.49935
Full scale (–)
–1
-1.001
-0.999
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
Verify digitize current 1 mA range
Description Calibrator setpoint
(A)
Verify digitize current 10 mA range
Description Calibrator setpoint
(A)
Full scale (+) 10.00004 9.99204 10.00804
Verify digitize current 100 mA range
Description Calibrator setpoint
(A)
Lower limit Upper limit
Lower limit Upper limit
Lower limit Upper limit
Verify digitize current 1 A range
Description Calibrator setpoint
(A)
DAQ6510-905-01 Rev. D June 2022 2-35
Lower limit Upper limit
Section
Calibration and Adjustment Manual
Half scale (+)
1.5
1.49745
1.50255
Half scale (–)
–1.5
–1.50255
–1.49745
*Full scale (–)
–2
–2.003
–1.997
* The 3 A range full-scale test points are limited to 2 A in this table because of the accuracy limitations of Fluke Models 57xxA and 5725A series calibrators at currents above 2.2 A.
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Verify digitize current 3 A range
Description Calibrator setpoint
(A)
*Full scale (+) 2 1.997 2.003
Lower limit Upper limit

AC current verification

The following topics describe how to verify ac current.
Verify ac current on the 100 µA to 3 A ranges
To verify ac current accuracy, you will:
Apply accurate voltages from the Fluke 5720A or 5730A multifunction calibrator to the DAQ6510
front-panel terminals.
Verify that the displayed readings fall within specified limits.
Use the values in the following tables to verify the performance of the DAQ6510. Actual values depend on the published specifications (see Example reading limit calculation (on page 2-4
)).
To verify ac current accuracy:
1. On the DAQ6510, press the FUNCTION key and select AC Current.
2. Press the HOME key.
3. Set the range you are verifying.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Make sure that Detector Bandwidth is set to 30 Hz.
AC current is specified for the detector bandwidth setting of 3 Hz. 3 Hz measures accurately for input
signals from 3 Hz to 10 kHz, with reading rates of ≈ 0.5 readings/s. To improve verification throughput to ≈ 3.3 readings/s, set detector bandwidth to 30 Hz for frequencies of 30 Hz to 10 kHz. To verify frequencies 1 kHz and higher, set the detector bandwidth to 300 Hz for faster ≈ 55
readings/s throughput.
2-36 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
100 µA at 20 Hz
100
99.83
100.17
100 µA at 1 kHz
100
99.83
100.17
1 mA at 20 Hz
1
0.9986
1.0014
1 mA at 1 kHz
1
0.9986
1.0014
1 mA at 4.9 kHz
1
0.9986
1.0014
10 mA at 40 Hz
10
9.986
10.14
10 mA at 1 kHz
10
9.986
10.14
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
7. Connect the DAQ6510 to the calibrator as shown in the following figure.
Figure 14: Connections for ac current verification
8. Source ac current for each of the frequencies listed in the following tables.
9. For each setting, make sure that the reading is within low and high limits.
Verify ac current 100 µA range
Description Verification point Lower limit Upper limit
Verify ac current 1 mA range
Description Verification point Lower limit Upper limit
Verify ac current 10 mA range
DAQ6510-905-01 Rev. D June 2022 2-37
Description Verification point Lower limit Upper limit
10 mA at 4.9 kHz 10 9.986 10.14
Section
Calibration and Adjustment Manual
100 mA at 40 Hz
100
99.86
100.14
100 mA at 1 kHz
100
99.86
100.14
100 mA at 4.9 kHz
100
99.86
100.14
1 A at 40 Hz
1.000
0.9986
1.0014
1 A at 1 kHz
1.000
0.9986
1.0014
1 A at 4.9 kHz
1.000
0.9986
1.0014
2 A at 40 Hz
2.000
1.9952
2.0048
2 A at 1 kHz
2.000
1.9952
2.0048
2 A at 4.9 kHz
2.000
1.9952
2.0048
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Verify ac current 100 mA range
Description Verification point Lower limit Upper limit
Verify ac current 1 A range
Description Verification point Lower limit Upper limit
Verify ac current 3 A range
Description Verification point Lower limit Upper limit
2-38 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:

Capacitance verification

To compensate for capacitance offset of the cable and 1 µF thru 100 µF decade box:
1. Connect the DAQ6510, shielded banana cable, banana to dual BNC cable, and 1 µF through 100 µF decade capacitor box as shown in the following diagram.
Figure 15: Connect DAQ6510 to decade capacitor box
2. Set the decade capacitor box to 0 F.
3. On the DAQ6510, press the FUNCTION key, select the Measure Functions tab, and select Capacitance.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Set the Range to 1 nF.
7. Press the MENU key.
DAQ6510-905-01 Rev. D June 2022 2-39
Section
Calibration and Adjustment Manual
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
8. Select Calculations and select Rel, then Acquire.
Cabling could be as high as ≈ 300 pF, which could prevent full-s
cale verification due to the large cable capacitance offset. Cable lengths should be minimized to reduce cable capacitance as much as possible.
9. Connect the shielded banana cable to the 1 nF to 1 µF Decade Capacitance Box as shown in the
figure below.
Figure 16: Capacitance verification connections
10. Verify capacitance following the verification points and accuracies from the table below.
2-40 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Performance verification
10% 1 nF range
0.1
0.0942
0.1058
70% 1 nF range
0.7
0.6894
0.7106
10% 10 nF range
1
0.986
1.014
100% 10 nF range
10
9.95
10.05
100% 100 nF range
100
99.5
100.5
10% 1 µF range
0.1
0.0986
0.1014
100% 1 µF range
1
0.995
1.005
10% 10 µF range
1
0.986
1.014
10% 100 µF range
10
9.86
10.14
100% 100 µF range
100
99.5
100.5
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
Verify the capacitance
Description Verification point Lower limit (F) Upper limit (F)
1 nF range cable REL 0.375 0.25 0.5
10% 100 nF range 10 9.86 10.14
100% 10 µF range 10 9.95 10.05

Verifying zero values using a 4-wire short

Four-wire short verifications are not included in the Customer Calibration Data Report.
To verify zero values using a 4-wire short, you will:
Check the zero values of various test points with 4-wire connections to the DAQ6510 front
terminals.
Verify that the displayed readings are within specified limits.
DAQ6510-905-01 Rev. D June 2022 2-41
Section
Calibration and Adjustment Manual
1
-0.0002
0.0002
10
-0.0002
0.0002
100
-0.002
0.002
1 k
-0.006
0.006
10 k
-0.06
0.06
100 k
-1
1
2: Performance verification DAQ6510 Data Acquisition and Multimeter System
Verify resistance zero values using a 4-wire short
To verify resistance zero values:
1. Select the 4W Res function.
2. Set the DAQ6510 to the range.
3. Press the MENU key.
4. Under Measure, select Settings.
5. Set the Offset Compensation to On.
6. Connect the Model 8610 or 8620 4-wire short to the front panel as shown in the following figure.
7. Allow to settle for 5 minutes. Do not use relative offset.
Figure 17: Front panel 4-wire shorting plug orientation
8. Verify that the 1 Ω range is within specification (see the following table).
9. Repeat verification for the 10 Ω to 100 kΩ ranges.
Verify 4-wire resistance zero values
Range (Ω) Lower limit Upper limit
2-42 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 2:
Performance verification
1000
-6000
6000
100
-600
600
10
-50
50 1 -66 100 mV
-3.5
3.5
Verify dc voltage zero values using the 4-wire short
To verify dc voltage zero values:
1. Leave the short connected as described in Verify resistance zero values using a 4-wire short (o
page 2-42).
2. Press the FUNCTION key.
3. Select the DC Voltage function.
4. Press the HOME key.
5. Set the range to 1000 V.
DC voltage verification is done in descending range order, starting with the 1000 V range and fi
nishing on the 100 mV range.
6. Verify that the 1000 V range zero is within specification. See the table below.
7. Verify that the 100 V to 100 mV range zero is within specification.
Verify dc voltage zero values
Range Lower limit Upper limit
n

Rear-panel verification

The DAQ6510 does not have rear-panel measurement inputs. For more detail, see the DAQ6510 Datasheet by going to tek.com and search for DAQ6510. Next you can select the Datasheet found in the Filter by Type section.
DAQ6510-905-01 Rev. D June 2022 2-43
Handling events ......................................................................3-40
Section 3

Adjustment

In this section:
Introduction ...............................................................................3-1
Environmental conditions ..........................................................3-2
Warmup period .........................................................................3-2
Adjustment overview .................................................................3-3
Recommended test equipment .................................................3-3
General adjustment considerations ...........................................3-4
Initial instrument setup ..............................................................3-5
Remote calibration adjustment procedures ...............................3-7
Enable temperature correction ................................................3-38
Save calibration and set the adjustment dates........................3-39
Setting time, adjustment, and verification dates ......................3-39
Adjustment command timing and error checking ....................3-40

Introduction

Use the procedures in this section to adjust the DAQ6510 calibration.
DAQ6510 performance is specified for a period of 90 days, 1 year, or 2 years from adjustment. Adjustment should be performed at one of these intervals, depending on your specification requirements.
Performance is specified relative to calibration adjustment temperature (T adjustment is performed at 23 °C ± 1 °C.
The information in this section is intended for qualified service personnel only, as described by the types of product users in the Safety precautions pages, provided at the beginning of this document. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, that if contacted, could cause p hazardous voltages.
ersonal injury or death. Use appropriate safety precautions when working with
). Keithley factory
CAL
Section
t Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System Calibration and Adjustmen

Environmental conditions

To make sure you get accurate results, the environment must meet the following conditions.

Temperature and relative humidity

Conduct the adjustment procedures in a test environment with:
A stable ambient temperature controlled to vary less than ±1 °C during the period of adjustment.
Keithley Instruments recommends a calibration adjustment temperature (T
different nominal temperature is used, it should be noted on the calibration report. A relative humidity of less than or equal to 40 percent, unless otherwise noted.
No direct airflow on the input terminals.

Line power

The DAQ6510 requires a line voltage of 100 V to 240 V and a line frequency of 400 Hz, 50 Hz or 60 Hz.
Calibration adjustments should be performed within this range.
The instrument automatically senses the line frequency at power-up.

Warmup period

Allow the DAQ6510 to warm up for at least 30 minutes before conducting the adjustment procedures.
) of 23 °C. If a
CAL
If the instrument has been subjected to temperature extremes (more than 5 °C above or below T
CAL
), allow additional time for the internal temperature of the instrument to stabilize. Typically, allow an additional 30 minutes to stabilize an instrument that is 10 °C outside the specified temperature range.
Also allow the test equipment to warm up for the time recommended by the manufacturer.
3-2 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Fluke
5720A or 5730A
High Performance
DCV, 10 A DCI, ACV,
resistance
DCI and ACI 10 A ranges
Fluke
8508A or 8588A
8.5 Digit Reference Multimeter
DCI
Keithley Instruments
3390
Function/Arbitrary Waveform Generator
Frequency
Keithley Instruments
8610 or 8620
4-Wire DMM Shorting Plug
DCV, digitize DCV, and resistance
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:

Adjustment overview

DAQ6510 adjustment is performed using a remote connection through either an optional GPIB, LAN, or USB interface. The calibration adjustment commands provided in this manual use the Test Script
®
Processor (TSP
You can use the Keithley Test Script Builder to send you adjustment commands. See "Test Script Builder (TSB)" in the Model DAQ6510 Reference Manual (DAQ6510-901-01).
) command language. There is no front-panel method for full adjustment.
To use GPIB with your DAQ6510, you must use the KTTI-G Accessory.

Recommended test equipment

The following table summarizes the recommended calibration equipment. Specified accuracy of all functions and ranges is dependent on the precision of reference signals used during the adjustment process. To achieve specified performance, adjustment reference uncertainties must be at least four times smaller than the best corresponding DAQ6510 one-year accuracy specification for measuring that signal.
Manufacturer Model Description Used for:
Multifunction Calibrator
Fluke 5725A Amplifier
PIB Communication and Digital I/O
ACI, and 10 kΩ
DAQ6510-905-01 Rev. D June 2022 3-3
Refer to the manufacturer's specifications to calculate the uncertainty, which varies for each function and r
ange test point.
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System

General adjustment considerations

DAQ6510 dc voltage performance is sensitive to errors from thermoelectric potentials generated
by test cables and connections. Be sure to use high-quality cables and connection techniques. When changing a connection during the adjustment process, be sure to allow time (up to five minutes) for thermal settling. Also allow thermal settling time, typically 60 seconds, when changing the TERMINALS FRONT/REAR switch position during adjustment.
Calibration steps are performed on both rear and front inputs as described in sections below. Be
sure that the TERMINALS FRONT/REAR switch is in the proper position before sending adjustment commands.
The Keithley Models 8610 and 8620 4-wire shorts connect all four terminals electrically, but the
layout of the board traces makes DAQ6510 adjustment sensitive to their orientation. Note the HI/LO terminal markings and be sure to insert connections in the correct orientation. Note that the DAQ6510 rear input terminals are rotated 90° compared to front-panel inputs.
During adjustment steps using a calibrator reference source, be sure the calibrator reference
signal has fully settled before sending adjustment commands to the DAQ6510.
The front and rear terminals of the instrument are rated for connection to circuits rated Measurement Category II up to 300 V, as described in International Electrotechnical Commission (IEC) Standard IEC 60664. This range must not be exceeded. Do not connect the instrument terminals to CAT III or CAT IV circuits. Connection of the instrument terminals to circuits higher than CAT II can cause damage to the equipment and severe personal injury.
The maximum input voltage between INPUT HI and INPUT LO is 1000 V dc and 750 V ac. Exceeding this value may create a shock hazard.
The maximum common-mode voltage (the voltage between INPUT LO and chassis ground) is 500 V
. Exceeding this value may cause a breakdown in insulation that can create a shock
PEAK
hazard.
The information in this section is intended for qualified service personnel only, as described by the types of product users in the Safety precautions pages, provided at the beginning of this document. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, that if contacted, could cau
se personal injury or death. Use appropriate safety precautions when working with
hazardous voltages.
3-4 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:

Initial instrument setup

Before adjusting calibration, make sure that the instrument is set up:
For remote operation and TSP commands.
To the correct date and time.
You must also unlock calibration.

Select the correct terminals

On the DAQ6510, you must adjust calibration from both the front and rear terminals. You can verify calibration on either the front or rear terminals. To set the instrument to the rear-panel terminals, press the TERMINALS switch on the front panel of the instrument.

Select the TSP command set

Calibration adjustment must be performed by remote control using LAN, USB, or optional GPIB interfaces. No front-panel calibration commands are available.
Calibration is only available using TSP commands. See the instructions below to change the command set.
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. For Command Set, select TSP.
4. You are prompted to reboot.
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
DAQ6510-905-01 Rev. D June 2022 3-5
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System

Verify instrument date and time

Before adjusting calibration, verify the system date of the DAQ6510.
From the front panel:
1. Press the MENU key.
2. Under System, select Settings. The SYSTEM SETTINGS menu opens.
3. Verify the date and time.
4. If necessary, correct the date and time.

Set up remote connections

For detail on remote communications, refer to the DAQ6510 Reference Manual section "Remote communications interfaces."
Calibration adjustment is performed by connecting reference signals to the DAQ6510 and sending a series of adjustment commands using one of the remote interfaces. The adjustment procedure may be done interactively, programmatically, or using a combination of the two.
Interactive: Manually set up and connect each reference signal, then send the appropriate
calibration adjustment command.
Programmatic: A computer performs setup of calibrator source instruments and controls
appropriate settling delays before automatically sending calibration adjustment commands to the DAQ6510. The control program must determine when the adjustment step is complete and check for adjustment errors before continuing to the next step.
The following section describes the DAQ6510 adjustment commands that are used for a manual interactive adjustment procedure. Techniques for controlling automated program timing and error checking are described in Adjustment command timing and error checking (on page 3-40
).

Unlock calibration

To start making adjustments, you must unlock it by sending the calibration password:
cal.unlock("KI000CAL")
KI000CAL is the default password. You can change the password. Refer to cal.password (on page
4-7) for details.
3-6 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:

Remote calibration adjustment procedures

The front-panel display does not show calibration progress or completion.
The following sections provide the preparation and command parameters that you need to complete adjustments to you DAQ6510. The preparation sections provide information necessary for making connections and other equipment needed for that adjustment. The command parameter tables are meant to run through in any order allowing you to adjust just the parameters you need. For a complete adjustment, run through these sections in order, making the preparations and running all the command parameters.
You will send each command parameter twice. The first time you use the setup command, and the second time you will use the execute command. More information on these commands can be found in the TSP command reference (on page 4-4
cal.adjust.step.setup cal.adjust.step.execute
). The two commands used are:

Disable temperature correction

Before you start your adjustment, you must turn off temperature correction. Run the following commands to turn off temperature correction.
cal.adjust.step.setup("TC_EN") cal.adjust.step.execute("TC_EN", 0)

Front-terminal adjustment with a 4-wire short

The following procedure provides instructions for completing a front-terminal adjustment using a 4-wire short. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
Keithley Model 8610 or Model 8620 Low Thermal Shorting Plug.
DAQ6510-905-01 Rev. D June 2022 3-7
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Prepare your DAQ6510 for a front-terminal adjustment with a 4-wire
short
To prepare the DAQ6510 for a front-terminal adjustment with a 4-wire short:
1. Set the TERMINALS switch to FRONT.
2. Install the Keithley Model 8610 or 8620 shorting plug on the front terminals of the DAQ6510 as
shown in the figure below.
The shorting plug terminals must be connected so that HI and LO are correctly aligned. Zero ac
curacy will be affected if the shorting plug terminals are not aligned correctly.
Figure 18: Connection for a front-panel adjustment with a 4-wire short
3. Allow to settle for five minutes.
Command parameters for a front-terminal adjustment with a 4-wire
short
When calibrating your DAQ6510 for a front-terminal adjustment with a 4-wire short, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first command parameter from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_DCV_100mV_zero_front") cal.adjust.step.execute("cal_DCV_100mV_zero_front")
3-8 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
cal_DCV_100mV_zero_front
cal_DCV_DIGI_100mV_zero
cal_DCV_1V_zero_front
cal_DCV_DIGI_1V_zero
cal_DCV_10V_zero_front
cal_sense_10V_zero_front
cal_DCV_DIGI_10V_zero
cal_DCV_100V_zero_front
cal_DCV_DIGI_100V_zero
cal_DCV_1kV_zero_front
cal_DCV_DIGI_1kV_zero
cal_4W_1ohm_zero_front
cal_2W_10ohm_zero_front
cal_4W_10ohm_zero_front
cal_2W_100ohm_zero_front
cal_4W_100ohm_zero_front
cal_2W_1kohm_zero_front
cal_3W_1kohm_zero_front
cal_3W_1kohm_Hi_zero_front
cal_3W_1kohm_SLO_zero_front
cal_4W_1kohm_zero_front
cal_2W_10kohm_zero_front
cal_3W_10kohm_zero_front
cal_3W_10kohm_SLO_zero_front
cal_4W_10kohm_zero_front
cal_2W_100kohm_zero_front
cal_3W_100kohm_SLO_zero_front
cal_4W_100kohm_zero_front
cal_2W_1Mohm_zero_front
cal_4W_1Mohm_zero_front
cal_2W_HiOhm_zero_front
cal_4W_HiOhm_zero_front
cal_4W_HiOhm_zero_sense
cal_diode_10mA_zero_front
cal_diode_1mA_zero_front
cal_diode_100uA_zero_front
cal_diode_10uA_zero_front
cal_ACV_1V_zero
cal_ACV_10V_zero
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
Command parameters
DAQ6510-905-01 Rev. D June 2022 3-9
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System

Rear-terminal adjustment with a 4-wire short

The following procedure provides instructions for completing a rear-terminal adjustment using a 4­wire short. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
Keithley Model 7797 Calibration System
Prepare your DAQ6510 for a rear-terminal adjustment with a 4-wire
short
To prepare the DAQ6510 for rear-terminal adjustment:
1. Set the TERMINALS switch to REAR.
2. Install the Keithley Model 7797 Calibration System into your DAQ6510 rear slot 1.
The "Model 7797 Calibration System Instructions" can be found at tek.com/keithley
3. Allow the DAQ6510 to settle for five minutes.
Command parameters for a rear-terminal adjustment with a 4-wire
short
When calibrating your DAQ6510 for a rear-terminal adjustment with a 4-wire short, use the following command parameters.
.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
3-10 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
cal_DCV_100mV_zero_rear
cal_DCV_1V_zero_rear
cal_DCV_10V_zero_rear
cal_sense_10V_zero_rear
cal_DCV_100V_zero_rear
cal_DCV_1kV_zero_rear
cal_4W_1ohm_zero_rear
cal_2W_10ohm_zero_rear
cal_4W_10ohm_zero_rear
cal_2W_100ohm_zero_rear
cal_4W_100ohm_zero_rear
cal_2W_1kohm_zero_rear
cal_3W_1kohm_zero_rear
cal_3W_1kohm_SLO_zero_rear
cal_4W_1kohm_zero_rear
cal_2W_10kohm_zero_rear
cal_3W_10kohm_zero_rear
cal_3W_10kohm_SLO_zero_rear
cal_4W_10kohm_zero_rear
cal_2W_100kohm_zero_rear
cal_3W_100kohm_SLO_zero_rear
cal_4W_100kohm_zero_rear
cal_2W_1Mohm_zero_rear
cal_4W_1Mohm_zero_rear
cal_2W_HiOhm_zero_rear
cal_4W_HiOhm_zero_rear
cal_diode_10mA_zero_rear
cal_diode_1mA_zero_rear
cal_diode_100uA_zero_rear
cal_diode_10uA_zero_rear
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
cal.adjust.step.setup("cal_DCV_100mV_zero_rear") cal.adjust.step.execute("cal_DCV_100mV_zero_rear")
Command parameters

Front-terminal adjustment with open circuit inputs

The following procedure provides instructions for completing a front-terminal adjustment with open circuit inputs. The following section provides a command parameter table to complete the adjustment.
DAQ6510-905-01 Rev. D June 2022 3-11
Section
Calibration and Adjustment Manual
cal_DCI_3A_zero_front
cal_DCI_1A_zero_front
cal_DCI_100mA_zero_front
cal_DCI_10mA_zero_front
cal_DCI_1mA_zero_front
cal_DCI_100uA_zero_front
cal_DCI_10uA_zero_front
cal_DCI_DIGI_3A_zero
cal_DCI_DIGI_1A_zero
cal_DCI_DIGI_100mA_zero
cal_DCI_DIGI_10mA_zero
cal_DCI_DIGI_1mA_zero
cal_DCI_DIGI_100uA_zero
cal_ACI_1A_zero
cal_ACI_3A_zero
cal_4W_HiOhm_7Vref_open
cal_CAP_zero
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Prepare your DAQ6510 for a front-terminal adjustment with open circuit
inputs
To prepare the DAQ6510 for a front-terminal adjustment with open circuit inputs:
1. Set the TERMINALS switch to FRONT.
Command parameters for a front-terminal adjustment with open circuit
inputs
When calibrating your DAQ6510 for a front-terminal adjustment with open circuit inputs, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_DCI_3A_zero_front") cal.adjust.step.execute("cal_DCI_3A_zero_front")
Command parameters
2. Remove all connections from the front terminals of your DAQ6510.
3-12 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
cal_DCI_10A_zero_rear
cal_DCI_DIGI_10A_zero_rear
cal_DCI_3A_zero_rear
cal_DCI_1A_zero_rear
cal_DCI_100mA_zero_rear
cal_DCI_10mA_zero_rear
cal_DCI_1mA_zero_rear
cal_DCI_100uA_zero_rear
cal_DCI_10uA_zero_rear
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:

Rear-terminal adjustment with open circuit inputs

The following procedure provides instructions for completing a rear-terminal adjustment with open circuit inputs. The following section provides a command parameter table to complete the adjustment.
Prepare your DAQ6510 for a rear-terminal adjustment with a 4-wire
short
To prepare the DAQ6510 for rear-terminal adjustment:
1. Install a Keithley Model 7797 Calibration System into the rear panel of your DAQ6510.
For information on installing a Keithley Model 7797 Calibration System, refer to the Model 7797 Calibration System Installation Instructions manual at tek.com/keithley
2. Set the TERMINALS switch to REAR. Make sure that the orange R is displayed.
.
Command parameters for a rear-terminal adjustment with open circuit
inputs
When calibrating your DAQ6510 for a rear-terminal adjustment with open circuit inputs, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_DCI_10A_zero_rear") cal.adjust.step.execute("cal_DCI_10A_zero_rear")
Command parameters
3. Allow the DAQ6510 to settle for five minutes.
DAQ6510-905-01 Rev. D June 2022 3-13
Section
Calibration and Adjustment Manual

Resistance adjustment

The following procedure provides instructions for completing a resistance adjustment. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Fluke 5720 calibrator
Prepare your DAQ6510 for a resistance adjustment
To prepare the DAQ6510 for a front-terminal adjustment:
1. Set the TERMINALS switch to FRONT.
2. Connect the DAQ6510 to the calibrator as shown in the following figure.
Figure 19: Connection for a resistance accuracy adjustment
3. Allow the instruments and cables to settle for five minutes.
4. On the calibrator, enable the OPR key and the EX SNS key.
5. Make sure that the OPERATE display and EX SNS keys are illuminated.
3-14 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
cal_TS7
n/a
n/a
n/a
n/a
cal_4W_10ohm_fs
4-wire Ω
10
10
actual from calibrator
cal_4W_10ohm_OCOMP_fs
4-wire Ω
10
10
actual from calibrator
cal_4W_100ohm_fs
4-wire Ω
100
100
actual from calibrator
cal_4W_100ohm_OCOMP_fs
4-wire Ω
100
100
actual from calibrator
cal_4W_1kohm_fs
4-wire Ω
1000
1000
actual from calibrator
cal_4W_1kohm_OCOMP_fs
4-wire Ω
1000
1000
actual from calibrator
cal_source_1mA_1V
4-wire Ω
1000
1000
actual from calibrator
cal_4W_10kohm_fs
cal_4W_10kohm_OCOMP_fs
4-wire Ω
10000
10000
actual from calibrator
cal_source_100uA_1V
4-wire Ω
10000
10000
actual from calibrator
cal_4W_100kohm_fs
4-wire Ω
100000
100000
actual from calibrator
cal_source_10uA_1V
4-wire Ω
100000
100000
actual from calibrator
cal_4W_1Mohm_fs
4-wire Ω
1000000
1000000
actual from calibrator
cal_source_1uA_1V
4-wire Ω
1000000
1000000
actual from calibrator
cal_4W_HiOhm_halfV_10Meg
4-wire Ω
10000000
10000000
actual from calibrator
cal_user_10Meg_value
4-wire Ω
10000000
10000000
actual from calibrator
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
Command parameters for a resistance adjustment
When calibrating your DAQ6510 for a resistance adjustment, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command along with the calibrator stimulus value.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_4W_10ohm_fs") cal.adjust.step.execute("cal_4W_10ohm_fs", <calibrator stimulus value>)
Since the temperature sensor is adjusted here, you are advised to perform all resistance adjustment steps in the same calibration session. If only a partial calibration is performed and the original T_cal differs from the new T_cal, the ranges that are not adjusted may less accurate. Completing an adjustment for all the resistance steps will remove these errors.
You must send the cal_TS7 first.
Command parameters Calibrator
function
4-wire Ω 10000 10000 actual from calibrator
Calibrator range
Calibrator value
Reference DMM dc current range
DAQ6510-905-01 Rev. D June 2022 3-15
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System

DC voltage adjustment

The following procedure provides instructions for completing a front-terminal adjustment for a dc voltage accuracy. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
Fluke 5720 calibrator
Prepare your DAQ6510 for a dc voltage adjustment
To prepare the DAQ6510 for a dc voltage adjustment:
1. Connect a cable between the calibrator and the DAQ6510 as shown in the figure below.
Figure 20: Connection for a dc voltage accuracy adjustment
2. Allow the DAQ6510 to settle for five minutes.
3-16 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
cal_DCV_DIGI_1kV_fs_pos
dc voltage
1100
1000
cal_DCV_DIGI_1kV_fs_neg
cal_DCV_100V_fs_pos
dc voltage
220
100
cal_DCV_DIGI_100V_fs_pos
dc voltage
220
100
cal_DCV_100V_fs_neg
dc voltage
220
-100
cal_DCV_DIGI_100V_fs_neg
dc voltage
220
-100
cal_DCV_10V_fs_pos
dc voltage
22
10
cal_DCV_DIGI_10V_fs_pos
dc voltage
22
10
cal_DCV_10V_fs_neg
dc voltage
22
-10
cal_DCV_DIGI_10V_fs_neg
dc voltage
22
-10
cal_DCV_DIGI_1V_fs_pos
dc voltage
2.2
1
cal_DCV_DIGI_1V_fs_neg
dc voltage
2.2
-1
cal_DCV_DIGI_100mV_fs_pos
dc voltage
0.22
0.1
cal_DCV_DIGI_100mV_fs_neg
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
Command parameters for a dc voltage adjustment
When calibrating your DAQ6510 for a dc voltage adjustment, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_DCV_DIGI_1kV_fs_pos") cal.adjust.step.execute("cal_DCV_DIGI_1kV_fs_pos")
Command parameters Calibrator
function
dc voltage 1100 -1000
Calibrator range
Calibrator value
dc voltage 0.22 -0.1

DC current adjustment

The following procedure provides instructions for completing a front-terminal adjustment for dc current accuracy. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
Fluke 5720 calibrator
DAQ6510-905-01 Rev. D June 2022 3-17
Fluke 8508A or 8588A Reference DMM
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Prepare your DAQ6510 for a dc current adjustment
To prepare the DAQ6510 for dc current adjustment:
Connect the Model DAQ6510 to the calibrator as shown in the following figure.
Figure 21: Connection for dc current
3-18 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
cal_TS6
n/a
n/a
n/a
n/a
cal_DCI_10uA_fs_pos
dc current
0.00022
0.00001
8508A or 8588A Reading
10 µA
cal_DCI_10uA_fs_neg
dc current
0.00022
-0.00001
8508A or 8588A Reading
100 µA
cal_DCI_100uA_fs_pos
8508A or 8588A Reading
cal_DCI_DIGI_100uA_fs_pos
dc current
0.00022
0.0001
8508A or 8588A
100 µA
cal_DCI_100uA_fs_neg
dc current
0.00022
-0.0001
8508A or 8588A Reading
100 µA
cal_DCI_DIGI_100uA_fs_neg
dc current
0.00022
-0.0001
8508A or 8588A Reading
100 µA
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
Command parameters for a dc current adjustment
When calibrating your DAQ6510 for a dc current adjustment, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command along with the calibrator stimulus value.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_DCI_10uA_fs_pos") cal.adjust.step.execute("cal_DCI_10uA_fs_pos", <calibrator stimulus value>)
You must send the cal_TS6 first.
Command parameters Calibrator
function
Calibrator range
Calibrator value
Calibrator stimulus value*
Reference DMM dc current range
10 µA range (+) full scale nominal
10 µA range (+) full scale nominal
100 µA range (+) full scale nominal
dc current 0.00022 0.0001
Reading
100 µA range (-) full scale nominal
*Characterized step measured by the Fluke 8508A or 8588A
100 µA
DAQ6510-905-01 Rev. D June 2022 3-19
Section
Calibration and Adjustment Manual
cal_DCI_1mA_fs_pos
dc current
0.0022
0.001
8508A or 8588A Reading
1 mA
cal_DCI_DIGI_1mA_fs_pos
Reading
cal_DCI_1mA_fs_neg
dc current
0.0022
-0.001
8508A or 8588A Reading
1 mA
cal_DCI_DIGI_1mA_fs_neg
8508A or 8588A Reading
cal_DCI_10mA_fs_pos
dc current
0.022
0.01
8508A or 8588A Reading
10 mA
cal_DCI_DIGI_10mA_fs_pos
Reading
cal_DCI_10mA_fs_neg
dc current
0.022
-0.01
8508A or 8588A Reading
10 mA
cal_DCI_DIGI_10mA_fs_neg
cal_DCI_100mA_fs_pos
dc current
0.22
0.1
n/a
20 A
cal_DCI_DIGI_100mA_fs_pos
cal_DCI_100mA_fs_neg
dc current
0.22
-0.1
n/a
20 A
cal_DCI_DIGI_100mA_fs_neg
dc current
0.22
-0.1
n/a
20 A
cal_TS5
dc current
n/a
n/a
n/a
20 A
cal_DCI_1A_fs_pos
dc current
2.2 1 n/a
20 A
cal_DCI_DIGI_1A_fs_pos
dc current
2.2 1 n/a
20 A
cal_DCI_1A_fs_neg
dc current
2.2
-1
n/a
20 A
cal_DCI_DIGI_1A_fs_neg
dc current
2.2
-1
n/a
20 A
cal_DCI_DIGI_3A_fs_pos
dc current
2.2 2 n/a
20 A
cal_DCI_DIGI_3A_fs_neg
dc current
2.2
-2
n/a
20 A
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Command parameters Calibrator
function
Calibrator range
Calibrator value
Calibrator stimulus value*
1 mA range (+) full scale nominal
Reference DMM dc current range
dc current 0.0022 0.001
8508A or 8588A
1 mA range (-) full scale nominal
dc current 0.0022 -0.001
10 mA range (+) full scale nominal
dc current 0.022 0.01 8508A or 8588A
10 mA range (-) full scale nominal
dc current 0.022 -0.01 8508A or 8588A
Reading
*Characterized step measured by the Fluke 8508A or 8588A
Command parameters Calibrator
function
Calibrator range
Calibrator value
Calibrator stimulus value
1 mA
1 mA
10 mA
10 mA
Reference DMM dc current range
3-20 DAQ6510-905-01 Rev. D June 2022
dc current 0.22 0.1 n/a 20 A
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:

AC voltage adjustment

The following procedure provides instructions for completing a front-terminal adjustment for ac voltage. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
Fluke 5720 calibrator
Keithley Model 3390 function generator
Prepare your DAQ6510 for an ac voltage adjustment
To prepare to run ac adjustments:
1. Set the TERMINALS switch to FRONT.
2. Connect a cable between the calibrator and the DAQ6510 as shown in the figure below.
Figure 22: Connection for an ac voltage adjustment
3. Allow the instrument and cables to settle for 30 seconds.
4. On the calibrator, source 10 mV 1.0 kHz and allow the calibrator and DAQ6510 to settle properly.
5. Enable the OPR key.
6. Ensure that the OPERATE display is illuminated.
7. Allow the calibrator and cable to settle properly.
DAQ6510-905-01 Rev. D June 2022 3-21
Section
Calibration and Adjustment Manual
cal_ACV_100mV_1kHz_1pct
cal_ACV_100mV_10Hz_fs
ac voltage
0.22
0.1
10 Hz
cal_ACV_100mV_1kHz_fs
ac voltage
0.22
0.1
1 kHz
cal_ACV_100mV_50kHz_fs
ac voltage
0.22
0.1
50 kHz
cal_ACV_100mV_100kHz_fs
cal_ACV_100mV_200kHz_fs
ac voltage
0.22
0.1
200 kHz
cal_ACV_100mV_300kHz_fs
ac voltage
0.22
0.1
300 kHz
cal_ACV_1V_1kHz_1pct
ac voltage
0.022
0.01
1 kHz
cal_ACV_1V_10Hz_fs
cal_ACV_1V_1kHz_fs
ac voltage
2.2 1 1 kHz
cal_ACV_1V_50kHz_fs
ac voltage
2.2 1 50 kHz
cal_ACV_1V_100kHz_fs
ac voltage
2.2 1 100 kHz
cal_ACV_1V_200kHz_fs
ac voltage
2.2 1 200 kHz
cal_ACV_1V_300kHz_fs
cal_ACV_10V_1kHz_1pct
ac voltage
0.22
0.1
1 kHz
cal_ACV_10V_10Hz_fs
ac voltage
22
10
10 Hz
cal_ACV_10V_1kHz_fs
ac voltage
22
10
1 kHz
cal_ACV_10V_50kHz_fs
ac voltage
22
10
50 kHz
cal_ACV_10V_100kHz_fs
ac voltage
22
10
100 kHz
cal_ACV_10V_200kHz_fs
ac voltage
22
10
200 kHz
cal_ACV_10V_300kHz_fs
ac voltage
22
10
300 kHz
cal_ACV_100V_10Hz_1pct
ac voltage
2.2 1 10 Hz
cal_ACV_100V_1kHz_1pct
ac voltage
2.2 1 1 kHz
cal_ACV_100V_50kHz_1pct
ac voltage
2.2 1 50 kHz
cal_ACV_100V_100kHz_1pct
ac voltage
2.2 1 100 kHz
cal_ACV_100V_200kHz_1pct
ac voltage
2.2 1 200 kHz
cal_ACV_100V_300kHz_1pct
ac voltage
2.2 1 300 kHz
cal_ACV_100V_1kHz_fs
ac voltage
220
100
1 kHz
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Command parameters for an ac voltage adjustment
When calibrating your DAQ6510 for an ac voltage adjustment, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_ACV_100mV_1kHz_1pct") cal.adjust.step.execute("cal_ACV_100mV_1kHz_1pct")
Command parameter Calibrator
function
ac voltage 0.0022 0.001 1 kHz
Calibrator range
Calibrator value
Calibrator frequency
ac voltage 0.22 0.1 100 kHz
ac voltage 2.2 1 10 Hz
ac voltage 2.2 1 300 kHz
3-22 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
cal_ACV_700V_10Hz_1pct
ac voltage
22 7 10 Hz
cal_ACV_700V_1kHz_1pct
ac voltage
22 7 1 kHz
cal_ACV_700V_50kHz_1pct
ac voltage
22 7 50 kHz
cal_ACV_700V_100kHz_1pct
ac voltage
22 7 100 kHz
cal_ACV_700V_200kHz_1pct
cal_ACV_700V_300kHz_1pct
ac voltage
22 7 300 kHz
cal_ACV_700V_1kHz_fs
ac voltage
1100
750
1 kHz
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
Command parameter Calibrator

AC current adjustment

The following procedure provides instructions for completing a front-terminal adjustment for ac current. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
Fluke 5720 calibrator
function
Calibrator range
Calibrator value
Calibrator frequency
ac voltage 22 7 200 kHz
Prepare your DAQ6510 for an ac current adjustment
To prepare the DAQ6510 for an ac current adjustment:
Connect the Model DAQ6510 to the calibrator as shown in the following figure.
Figure 23: Connection for ac current
DAQ6510-905-01 Rev. D June 2022 3-23
Section
Calibration and Adjustment Manual
cal_ACI_100uA_1kHz_tenth
cal_ACI_100uA_1kHz_fs
ac current
0.00022
0.0001
1 kHz
cal_ACI_1mA_1kHz_tenth
ac current
0.00022
0.0001
1 kHz
cal_ACI_1mA_1kHz_fs
ac current
0.0022
0.001
1 kHz
cal_ACI_10mA_1kHz_tenth
cal_ACI_10mA_10Hz_fs
ac current
0.022
0.01
10 Hz
cal_ACI_10mA_10kHz_fs
ac current
0.022
0.01
10 kHz
cal_ACI_10mA_1kHz_fs
ac current
0.022
0.01
1 kHz
cal_ACI_100mA_1kHz_tenth
cal_ACI_100mA_1kHz_fs
ac current
0.22
0.1
1 kHz
cal_ACI_100mA_10Hz_fs
ac current
0.22
0.1
10 Hz
cal_ACI_200mA_1kHz_fs
ac current
0.22
0.2
1 kHz
cal_ACI_200mA_10kHz_fs
ac current
0.22
0.2
10 kHz
cal_ACI_1A_1kHz_tenth
ac current
0.22
0.1
1 kHz
cal_ACI_1A_10Hz_fs
ac current
2.2 1 10 Hz
cal_ACI_1A_1kHz_fs
ac current
2.2 1 1 kHz
cal_ACI_3A_1kHz_tenth
ac current
2.2
0.3
1 kHz
cal_ACI_3A_1kHz_fs
ac current
2.2 2 1 kHz
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Command parameters for an ac current adjustment
When calibrating your DAQ6510 for an ac current adjustment, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_ACI_100uA_1kHz_tenth") cal.adjust.step.execute("cal_ACI_100uA_1kHz_tenth")
Command parameters Calibrator
function
ac current 0.00022 0.00001 1 kHz
Calibrator range
Calibrator value
Calibrator frequency
ac current 0.0022 0.001 1 kHz
ac current 0.022 0.01 1 kHz
3-24 DAQ6510-905-01 Rev. D June 2022
DAQ6510
stment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3: Adju

Frequency adjustment

The following procedure provides instructions for completing a front-terminal adjustment for frequency. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
50 Ω coaxial cable
Keithley Model 3390 function generator
Prepare your DAQ6510 for a frequency adjustment
To prepare the DAQ6510 for a frequency adjustment:
1. Connect the Keithley Instruments Model 3390 function generator to the DAQ6510 INPUT HI and
LO terminals as shown in the following figure.
2. Use the BNC to banana adapter at the UUT connection.
Figure 24: Connection for a frequency adjustment
3. Set the function generator output impedance to high, amplitude of 5 V
, waveform to
RMS
square wave.
4. Enable output.
DAQ6510-905-01 Rev. D June 2022 3-25
Section
Calibration and Adjustment Manual
cal_FREQ_1kHz
Frequency
22
10
1 kHz
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
Command parameters for a frequency adjustment
When calibrating your DAQ6510 for a frequency adjustment, use the following command parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command parameters using this command.
cal.adjust.step.setup("cal_FREQ_1kHz") cal.adjust.step.execute("cal_FREQ_1kHz")
Command parameter Ca
librator
function
Calibrator range
Calibrator value
Calibrator frequency

Complete list of calibration commands

To make your adjustment procedure easier, you can copy and paste the code examples below into the Keithley Test Script Builder (TSB software). The commands can be copied and pasted two lines at a time.
All parameters in italics must be replaced with the specified value. For example,
cal.adjust.step.execute("cal_4W_10ohm_fs",
10ohm_value_read_from_calibrator
) must have the italics replaced with the calibrator value.
3-26 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_100mV_zero_front") cal.adjust.step.execute("cal_DCV_100mV_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_DIGI_100mV_zero") cal.adjust.step.execute("cal_DCV_DIGI_100mV_zero")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_1V_zero_front") cal.adjust.step.execute("cal_DCV_1V_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_DIGI_1V_zero") cal.adjust.step.execute("cal_DCV_DIGI_1V_zero")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_10V_zero_front") cal.adjust.step.execute("cal_DCV_10V_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_sense_10V_zero_front") cal.adjust.step.execute("cal_sense_10V_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_DIGI_10V_zero") cal.adjust.step.execute("cal_DCV_DIGI_10V_zero")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_100V_zero_front") cal.adjust.step.execute("cal_DCV_100V_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_DIGI_100V_zero") cal.adjust.step.execute("cal_DCV_DIGI_100V_zero")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_1kV_zero_front") cal.adjust.step.execute("cal_DCV_1kV_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_DIGI_1kV_zero") cal.adjust.step.execute("cal_DCV_DIGI_1kV_zero")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_1ohm_zero_front") cal.adjust.step.execute("cal_4W_1ohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_10ohm_zero_front") cal.adjust.step.execute("cal_2W_10ohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_10ohm_zero_front") cal.adjust.step.execute("cal_4W_10ohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_100ohm_zero_front") cal.adjust.step.execute("cal_2W_100ohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_100ohm_zero_front") cal.adjust.step.execute("cal_4W_100ohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_1kohm_zero_front") cal.adjust.step.execute("cal_2W_1kohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_1kohm_zero_front") cal.adjust.step.execute("cal_3W_1kohm_zero_front")
-- Use front-terminal 4-wire short setup
DAQ6510-905-01 Rev. D June 2022 3-27
Section
ration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System Calib
cal.adjust.step.setup("cal_3W_1kohm_Hi_zero_front") cal.adjust.step.execute("cal_3W_1kohm_Hi_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_1kohm_SLO_zero_front") cal.adjust.step.execute("cal_3W_1kohm_SLO_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_1kohm_zero_front") cal.adjust.step.execute("cal_4W_1kohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_10kohm_zero_front") cal.adjust.step.execute("cal_2W_10kohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_10kohm_zero_front") cal.adjust.step.execute("cal_3W_10kohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_10kohm_SLO_zero_front") cal.adjust.step.execute("cal_3W_10kohm_SLO_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_10kohm_zero_front") cal.adjust.step.execute("cal_4W_10kohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_100kohm_zero_front") cal.adjust.step.execute("cal_2W_100kohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_100kohm_SLO_zero_front") cal.adjust.step.execute("cal_3W_100kohm_SLO_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_100kohm_zero_front") cal.adjust.step.execute("cal_4W_100kohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_1Mohm_zero_front") cal.adjust.step.execute("cal_2W_1Mohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_1Mohm_zero_front") cal.adjust.step.execute("cal_4W_1Mohm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_HiOhm_zero_front") cal.adjust.step.execute("cal_2W_HiOhm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_HiOhm_zero_front") cal.adjust.step.execute("cal_4W_HiOhm_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_HiOhm_zero_sense") cal.adjust.step.execute("cal_4W_HiOhm_zero_sense")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_10mA_zero_front") cal.adjust.step.execute("cal_diode_10mA_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_1mA_zero_front") cal.adjust.step.execute("cal_diode_1mA_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_100uA_zero_front") cal.adjust.step.execute("cal_diode_100uA_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_10uA_zero_front")
3-28 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
cal.adjust.step.execute("cal_diode_10uA_zero_front")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_ACV_1V_zero") cal.adjust.step.execute("cal_ACV_1V_zero")
-- Use front-terminal 4-wire short setup cal.adjust.step.setup("cal_ACV_10V_zero") cal.adjust.step.execute("cal_ACV_10V_zero")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_100mV_zero_rear") cal.adjust.step.execute("cal_DCV_100mV_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_100mV_zero_rear") cal.adjust.step.execute("cal_DCV_100mV_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_1V_zero_rear") cal.adjust.step.execute("cal_DCV_1V_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_10V_zero_rear") cal.adjust.step.execute("cal_DCV_10V_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_sense_10V_zero_rear") cal.adjust.step.execute("cal_sense_10V_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_100V_zero_rear") cal.adjust.step.execute("cal_DCV_100V_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_DCV_1kV_zero_rear") cal.adjust.step.execute("cal_DCV_1kV_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_1ohm_zero_rear") cal.adjust.step.execute("cal_4W_1ohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_10ohm_zero_rear") cal.adjust.step.execute("cal_2W_10ohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_10ohm_zero_rear") cal.adjust.step.execute("cal_4W_10ohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_100ohm_zero_rear") cal.adjust.step.execute("cal_2W_100ohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_100ohm_zero_rear") cal.adjust.step.execute("cal_4W_100ohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_1kohm_zero_rear") cal.adjust.step.execute("cal_2W_1kohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_1kohm_zero_rear") cal.adjust.step.execute("cal_3W_1kohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_1kohm_SLO_zero_rear") cal.adjust.step.execute("cal_3W_1kohm_SLO_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_1kohm_zero_rear") cal.adjust.step.execute("cal_4W_1kohm_zero_rear")
DAQ6510-905-01 Rev. D June 2022 3-29
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_10kohm_zero_rear") cal.adjust.step.execute("cal_2W_10kohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_10kohm_zero_rear") cal.adjust.step.execute("cal_3W_10kohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_10kohm_SLO_zero_rear") cal.adjust.step.execute("cal_3W_10kohm_SLO_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_10kohm_zero_rear") cal.adjust.step.execute("cal_4W_10kohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_100kohm_zero_rear") cal.adjust.step.execute("cal_2W_100kohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_3W_100kohm_SLO_zero_rear") cal.adjust.step.execute("cal_3W_100kohm_SLO_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_100kohm_zero_rear") cal.adjust.step.execute("cal_4W_100kohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_1Mohm_zero_rear") cal.adjust.step.execute("cal_2W_1Mohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_1Mohm_zero_rear") cal.adjust.step.execute("cal_4W_1Mohm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_2W_HiOhm_zero_rear") cal.adjust.step.execute("cal_2W_HiOhm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_4W_HiOhm_zero_rear") cal.adjust.step.execute("cal_4W_HiOhm_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_10mA_zero_rear") cal.adjust.step.execute("cal_diode_10mA_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_1mA_zero_rear") cal.adjust.step.execute("cal_diode_1mA_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_100uA_zero_rear") cal.adjust.step.execute("cal_diode_100uA_zero_rear")
-- Use rear-terminal 4-wire short setup cal.adjust.step.setup("cal_diode_10uA_zero_rear") cal.adjust.step.execute("cal_diode_10uA_zero_rear")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_3A_zero_front") cal.adjust.step.execute("cal_DCI_3A_zero_front")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_1A_zero_front") cal.adjust.step.execute("cal_DCI_1A_zero_front")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_100mA_zero_front") cal.adjust.step.execute("cal_DCI_100mA_zero_front")
-- Use front-terminal 4-wire open circuit setup
3-30 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
cal.adjust.step.setup("cal_DCI_10mA_zero_front") cal.adjust.step.execute("cal_DCI_10mA_zero_front")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_1mA_zero_front") cal.adjust.step.execute("cal_DCI_1mA_zero_front")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_100uA_zero_front") cal.adjust.step.execute("cal_DCI_100uA_zero_front")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_10uA_zero_front") cal.adjust.step.execute("cal_DCI_10uA_zero_front")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_DIGI_3A_zero") cal.adjust.step.execute("cal_DCI_DIGI_3A_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_DIGI_1A_zero") cal.adjust.step.execute("cal_DCI_DIGI_1A_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_DIGI_100mA_zero") cal.adjust.step.execute("cal_DCI_DIGI_100mA_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_DIGI_10mA_zero") cal.adjust.step.execute("cal_DCI_DIGI_10mA_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_DIGI_1mA_zero") cal.adjust.step.execute("cal_DCI_DIGI_1mA_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_DIGI_100uA_zero") cal.adjust.step.execute("cal_DCI_DIGI_100uA_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_ACI_1A_zero") cal.adjust.step.execute("cal_ACI_1A_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_ACI_3A_zero") cal.adjust.step.execute("cal_ACI_3A_zero")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_4W_HiOhm_7Vref_open") cal.adjust.step.execute("cal_4W_HiOhm_7Vref_open")
-- Use front-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_CAP_zero") cal.adjust.step.execute("cal_CAP_zero")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_10A_zero_rear") cal.adjust.step.execute("cal_DCI_10A_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_DIGI_10A_zero_rear") cal.adjust.step.execute("cal_DCI_DIGI_10A_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_3A_zero_rear") cal.adjust.step.execute("cal_DCI_3A_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_1A_zero_rear") cal.adjust.step.execute("cal_DCI_1A_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_100mA_zero_rear")
DAQ6510-905-01 Rev. D June 2022 3-31
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
cal.adjust.step.execute("cal_DCI_100mA_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_10mA_zero_rear") cal.adjust.step.execute("cal_DCI_10mA_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_1mA_zero_rear") cal.adjust.step.execute("cal_DCI_1mA_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_100uA_zero_rear") cal.adjust.step.execute("cal_DCI_100uA_zero_rear")
-- Use rear-terminal 4-wire open circuit setup cal.adjust.step.setup("cal_DCI_10uA_zero_rear") cal.adjust.step.execute("cal_DCI_10uA_zero_rear")
-- Use resistance setup cal.adjust.step.setup("cal_TS7") cal.adjust.step.execute("cal_TS7")
-- Use resistance setup cal.adjust.step.setup("cal_4W_10ohm_fs") cal.adjust.step.execute("cal_4W_10ohm_fs", 10ohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_10ohm_OCOMP_fs") cal.adjust.step.execute("cal_4W_10ohm_OCOMP_fs", 10ohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_100ohm_fs") cal.adjust.step.execute("cal_4W_100ohm_fs", 100ohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_100ohm_OCOMP_fs") cal.adjust.step.execute("cal_4W_100ohm_OCOMP_fs",100ohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_1kohm_fs") cal.adjust.step.execute("cal_4W_1kohm_fs", 1kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_1kohm_OCOMP_fs") cal.adjust.step.execute("cal_4W_1kohm_OCOMP_fs", 1kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_source_1mA_1V") cal.adjust.step.execute("cal_source_1mA_1V", 1kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_10kohm_fs") cal.adjust.step.execute("cal_4W_10kohm_fs", 10kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_10kohm_OCOMP_fs") cal.adjust.step.execute("cal_4W_10kohm_OCOMP_fs",
10kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_source_100uA_1V") cal.adjust.step.execute("cal_source_100uA_1V", 10kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_100kohm_fs") cal.adjust.step.execute("cal_4W_100kohm_fs", 100kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_source_10uA_1V") cal.adjust.step.execute("cal_source_10uA_1V", 100kohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_1Mohm_fs") cal.adjust.step.execute("cal_4W_1Mohm_fs", 1Mohm_value_read_from_calibrator)
3-32 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
-- Use resistance setup cal.adjust.step.setup("cal_source_1uA_1V") cal.adjust.step.execute("cal_source_1uA_1V", 1Mohm_value_read_from_calibrator)
-- Use resistance setup cal.adjust.step.setup("cal_4W_HiOhm_halfV_10Meg") cal.adjust.step.execute("cal_4W_HiOhm_halfV_10Meg",10Mohm_value_read_from_calibrato
r)
-- Use resistance setup cal.adjust.step.setup("cal_user_10Meg_value") cal.adjust.step.execute("cal_user_10Meg_value", 10Mohm_value_read_from_calibrator)
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_1kV_fs_pos") cal.adjust.step.setup("cal_DCV_DIGI_1kV_fs_pos")
-- Use dc voltage setupcal.adjust.step.setup("cal_DCV_DIGI_1kV_fs_neg") cal.adjust.step.setup("cal_DCV_DIGI_1kV_fs_neg")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_100V_fs_pos") cal.adjust.step.setup("cal_DCV_100V_fs_pos")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_100V_fs_pos") cal.adjust.step.setup("cal_DCV_DIGI_100V_fs_pos")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_100V_fs_neg") cal.adjust.step.setup("cal_DCV_100V_fs_neg")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_100V_fs_neg") cal.adjust.step.setup("cal_DCV_DIGI_100V_fs_neg")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_10V_fs_pos") cal.adjust.step.setup("cal_DCV_10V_fs_pos")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_10V_fs_pos") cal.adjust.step.setup("cal_DCV_DIGI_10V_fs_pos")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_10V_fs_neg") cal.adjust.step.setup("cal_DCV_10V_fs_neg")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_10V_fs_neg") cal.adjust.step.setup("cal_DCV_DIGI_10V_fs_neg")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_1V_fs_pos") cal.adjust.step.setup("cal_DCV_DIGI_1V_fs_pos")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_1V_fs_neg") cal.adjust.step.setup("cal_DCV_DIGI_1V_fs_neg")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_100mV_fs_pos") cal.adjust.step.setup("cal_DCV_DIGI_100mV_fs_pos")
-- Use dc voltage setup cal.adjust.step.setup("cal_DCV_DIGI_100mV_fs_neg") cal.adjust.step.setup("cal_DCV_DIGI_100mV_fs_neg")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_10uA_fs_pos") cal.adjust.step.execute("cal_DCI_10uA_fs_pos", value_read_from_ 8508A/8558A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_10uA_fs_neg")
DAQ6510-905-01 Rev. D June 2022 3-33
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
cal.adjust.step.execute("cal_DCI_10uA_fs_neg",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_100uA_fs_pos") cal.adjust.step.execute("cal_DCI_100uA_fs_pos",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_100uA_fs_pos") cal.adjust.step.execute("cal_DCI_DIGI_100uA_fs_pos",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_100uA_fs_neg") cal.adjust.step.execute("cal_DCI_100uA_fs_neg",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_100uA_fs_neg") cal.adjust.step.execute("cal_DCI_DIGI_100uA_fs_neg",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_1mA_fs_pos") cal.adjust.step.execute("cal_DCI_1mA_fs_pos",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_1mA_fs_pos") cal.adjust.step.execute("cal_DCI_DIGI_1mA_fs_pos",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_1mA_fs_neg") cal.adjust.step.execute("cal_DCI_1mA_fs_neg",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_1mA_fs_neg") cal.adjust.step.execute("cal_DCI_DIGI_1mA_fs_neg",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_10mA_fs_pos") cal.adjust.step.execute("cal_DCI_10mA_fs_pos",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_10mA_fs_pos") cal.adjust.step.execute("cal_DCI_DIGI_10mA_fs_pos",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_10mA_fs_neg") cal.adjust.step.execute("cal_DCI_10mA_fs_neg",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_10mA_fs_neg") cal.adjust.step.execute("cal_DCI_DIGI_10mA_fs_neg",value_read_from_ 8508A/8588A)
-- Use dc current setup cal.adjust.step.setup("cal_DCI_100mA_fs_pos") cal.adjust.step.execute("cal_DCI_100mA_fs_pos")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_100mA_fs_pos") cal.adjust.step.execute("cal_DCI_DIGI_100mA_fs_pos")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_100mA_fs_neg") cal.adjust.step.execute("cal_DCI_100mA_fs_neg")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_100mA_fs_neg") cal.adjust.step.execute("cal_DCI_DIGI_100mA_fs_neg")
-- Use dc current setup cal.adjust.step.setup("cal_TS5") cal.adjust.step.execute("cal_TS5")
-- Use DC current setup cal.adjust.step.setup("cal_DCI_1A_fs_pos") cal.adjust.step.execute("cal_DCI_1A_fs_pos")
3-34 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_1A_fs_pos") cal.adjust.step.execute("cal_DCI_DIGI_1A_fs_pos")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_1A_fs_neg") cal.adjust.step.execute("cal_DCI_1A_fs_neg")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_1A_fs_neg") cal.adjust.step.execute("cal_DCI_DIGI_1A_fs_neg")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_3A_fs_pos") cal.adjust.step.execute("cal_DCI_DIGI_3A_fs_pos")
-- Use dc current setup cal.adjust.step.setup("cal_DCI_DIGI_3A_fs_neg") cal.adjust.step.execute("cal_DCI_DIGI_3A_fs_neg")
-- Use dc current 10 A setup cal.adjust.step.setup("cal_DCI_10A_fs_pos") cal.adjust.step.execute("cal_DCI_10A_fs_pos")
-- Use dc current 10 A setup cal.adjust.step.setup("cal_DCI_DIGI_10A_fs_pos") cal.adjust.step.execute("cal_DCI_DIGI_10A_fs_pos")
-- Use dc current 10 A setup cal.adjust.step.setup("cal_DCI_10A_fs_neg") cal.adjust.step.execute("cal_DCI_10A_fs_neg")
-- Use dc current 10 A setup cal.adjust.step.setup("cal_DCI_DIGI_10A_fs_neg") cal.adjust.step.execute("cal_DCI_DIGI_10A_fs_neg")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100mV_1kHz_1pct") cal.adjust.step.execute("cal_ACV_100mV_1kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100mV_10Hz_fs") cal.adjust.step.execute("cal_ACV_100mV_10Hz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100mV_1kHz_fs") cal.adjust.step.execute("cal_ACV_100mV_1kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100mV_50kHz_fs") cal.adjust.step.execute("cal_ACV_100mV_50kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100mV_100kHz_fs") cal.adjust.step.execute("cal_ACV_100mV_100kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100mV_200kHz_fs") cal.adjust.step.execute("cal_ACV_100mV_200kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100mV_300kHz_fs") cal.adjust.step.execute("cal_ACV_100mV_300kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_1V_1kHz_1pct") cal.adjust.step.execute("cal_ACV_1V_1kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_1V_10Hz_fs") cal.adjust.step.execute("cal_ACV_1V_10Hz_fs")
-- Use ac voltage setup
DAQ6510-905-01 Rev. D June 2022 3-35
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System
cal.adjust.step.setup("cal_ACV_1V_1kHz_fs") cal.adjust.step.execute("cal_ACV_1V_1kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_1V_50kHz_fs") cal.adjust.step.execute("cal_ACV_1V_50kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_1V_100kHz_fs") cal.adjust.step.execute("cal_ACV_1V_100kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_1V_200kHz_fs") cal.adjust.step.execute"cal_ACV_1V_200kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_1V_300kHz_fs") cal.adjust.step.execute("cal_ACV_1V_300kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_10V_1kHz_1pct") cal.adjust.step.execute("cal_ACV_10V_1kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_10V_10Hz_fs") cal.adjust.step.execute("cal_ACV_10V_10Hz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_10V_1kHz_fs") cal.adjust.step.execute("cal_ACV_10V_1kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_10V_50kHz_fs") cal.adjust.step.execute("cal_ACV_10V_50kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_10V_100kHz_fs") cal.adjust.step.execute("cal_ACV_10V_100kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_10V_200kHz_fs") cal.adjust.step.execute("cal_ACV_10V_200kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_10V_300kHz_fs") cal.adjust.step.execute("cal_ACV_10V_300kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100V_10Hz_1pct") cal.adjust.step.execute("cal_ACV_100V_10Hz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100V_1kHz_1pct") cal.adjust.step.execute("cal_ACV_100V_1kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100V_50kHz_1pct") cal.adjust.step.execute("cal_ACV_100V_50kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100V_100kHz_1pct") cal.adjust.step.execute("cal_ACV_100V_100kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100V_200kHz_1pct") cal.adjust.step.execute("cal_ACV_100V_200kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100V_300kHz_1pct") cal.adjust.step.execute("cal_ACV_100V_300kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_100V_1kHz_fs")
3-36 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:
cal.adjust.step.execute("cal_ACV_100V_1kHz_fs")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_700V_10Hz_1pct") cal.adjust.step.execute("cal_ACV_700V_10Hz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_700V_1kHz_1pct") cal.adjust.step.execute("cal_ACV_700V_1kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_700V_50kHz_1pct") cal.adjust.step.execute("cal_ACV_700V_50kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_700V_100kHz_1pct") cal.adjust.step.execute("cal_ACV_700V_100kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_700V_200kHz_1pct") cal.adjust.step.execute("cal_ACV_700V_200kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_700V_300kHz_1pct") cal.adjust.step.execute("cal_ACV_700V_300kHz_1pct")
-- Use ac voltage setup cal.adjust.step.setup("cal_ACV_700V_1kHz_fs") cal.adjust.step.execute("cal_ACV_700V_1kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_100uA_1kHz_tenth") cal.adjust.step.execute("cal_ACI_100uA_1kHz_tenth")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_100uA_1kHz_fs") cal.adjust.step.execute("cal_ACI_100uA_1kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_1mA_1kHz_tenth") cal.adjust.step.execute("cal_ACI_1mA_1kHz_tenth")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_1mA_1kHz_fs") cal.adjust.step.execute("cal_ACI_1mA_1kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_10mA_1kHz_tenth") cal.adjust.step.execute("cal_ACI_10mA_1kHz_tenth")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_10mA_10Hz_fs") cal.adjust.step.execute("cal_ACI_10mA_10Hz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_10mA_10kHz_fs") cal.adjust.step.execute("cal_ACI_10mA_10kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_10mA_1kHz_fs") cal.adjust.step.execute("cal_ACI_10mA_1kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_100mA_1kHz_tenth") cal.adjust.step.execute("cal_ACI_100mA_1kHz_tenth")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_100mA_1kHz_fs") cal.adjust.step.execute("cal_ACI_100mA_1kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_100mA_10Hz_fs") cal.adjust.step.execute("cal_ACI_100mA_10Hz_fs")
DAQ6510-905-01 Rev. D June 2022 3-37
Section
djustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System Calibration and A
-- Use ac current setup cal.adjust.step.setup("cal_ACI_200mA_1kHz_fs") cal.adjust.step.execute("cal_ACI_200mA_1kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_200mA_10kHz_fs") cal.adjust.step.execute("cal_ACI_200mA_10kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_1A_1kHz_tenth") cal.adjust.step.execute("cal_ACI_1A_1kHz_tenth")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_1A_10Hz_fs") cal.adjust.step.execute("cal_ACI_1A_10Hz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_1A_1kHz_fs") cal.adjust.step.execute("cal_ACI_1A_1kHz_fs")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_3A_1kHz_tenth") cal.adjust.step.execute("cal_ACI_3A_1kHz_tenth")
-- Use ac current setup cal.adjust.step.setup("cal_ACI_3A_1kHz_fs") cal.adjust.step.execute("cal_ACI_3A_1kHz_fs")<CT6500_only_start>
-- Use ac current 10 A setup cal.adjust.step.setup("cal_ACI_10A_400Hz_tenth") cal.adjust.step.execute("cal_ACI_10A_400Hz_tenth")
-- Use ac current 10 A setup cal.adjust.step.setup("cal_ACI_10A_400Hz_fs") cal.adjust.step.execute("cal_ACI_10A_400Hz_fs")
-- Use ac current 10 A setup cal.adjust.step.setup("cal_ACI_10A_5kHz_fs") cal.adjust.step.execute("cal_ACI_10A_5kHz_fs")
-- Use ac current 10 A setup cal.adjust.step.setup("cal_ACI_10A_400Hz_2A") cal.adjust.step.execute("cal_ACI_10A_400Hz_2A")
-- Use ac current 10 A setup cal.adjust.step.setup("cal_ACI_10A_10Hz_2A") cal.adjust.step.execute("cal_ACI_10A_10Hz_2A")<CT6500_only_end>
-- Use frequency setup cal.adjust.step.setup("cal_FREQ_1kHz") cal.adjust.step.execute("cal_FREQ_1kHz")

Enable temperature correction

After your adjustments are complete, turn on temperature correction. Run the following command to turn on temperature correction.
cal.adjust.step.setup("TC_EN") cal.adjust.step.execute("TC_EN", 1)
3-38 DAQ6510-905-01 Rev. D June 2022
DAQ6510
Adjustment
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 3:

Save calibration and set the adjustment dates

Use the following commands to save and lock calibration adjustments. These steps will take seconds to complete.
cal.save() cal.lock()
Calibration is temporary until you send the cal.save() command. Also, calibration data is not
saved if calibration is locked or if invalid data exists.
This completes the remote calibration adjustment procedure.

Setting time, adjustment, and verification dates

The DAQ6510 calibration adjustment date variable is set automatically to the present system time when the cal.save() command is executed. However, you can also use Test Script Processor
®
) commands to set the time, verification date, and adjustment date.
(TSP
To set the DAQ6510 time, adjust date, and verify date using TSP commands:
-- Unlock calibration (if not already unlocked). cal.unlock("KI000CAL")
-- Set the time (year, month, date, hour, minutes, seconds). localnode.settime(2018, 1, 9, 10, 4, 38)
-- Set the adjustment and verification dates. cal.adjust.date = os.time({year = 2018, month = 1, day = 9}) cal.verify.date = os.time({year = 2018, month = 1, day = 9})
The cal.verify.date command is used to record the date of the last verification that was done
independently of DAQ6510 adjustment. Typically, this date is set at the completion of a performance verification procedure.
Calibration must be unlocked to change the adjustment and verification dates, but changing these dates does not require using the cal.save() command and does not affect the cal.count
command.
DAQ6510-905-01 Rev. D June 2022 3-39
Section
Calibration and Adjustment Manual
3: Adjustment DAQ6510 Data Acquisition and Multimeter System

Adjustment command timing and error checking

Before each adjustment step, the input of the DAQ6510 must be connected to an appropriate reference signal, as documented in Remote calibration adjustment procedures (on page 3-7
You must make sure that the correct signal is connected and fully settled before sending the associated cal.adjust command. Failing to wait for a signal to settle completely may result in a
poor-quality adjustment that could cause the DAQ6510 to fail later performance verification.
Once the reference signal is stable, sending the calibration adjustment command initiates internal measurement operations in the DAQ6510. These steps can take from a few seconds to 30 seconds to complete.
Another method of getting calibration status feedback is to use the TSP prompts mode of operation. This is especially useful when you are entering cal.adjust commands interactively from an
ethernet Telnet session.
).
To enable prompts mode, send the localnode.prompts = 1 command. This causes the
DAQ6510 to return a TSP> prompt to the computer screen when it has completed a step and is ready
for the next command. Prompts mode also returns a TSP? prompt if an event message is available
(for example, if an error occurs during an adjustment step). See Handling events (on page 3-40 more information.
Although not prohibited, prompts mode is not recommended for automated test programs. If prompts mode is enabled in an automated test, the control program must be constructed to expect and read each prompt that is sent after each TSP command is processed (including queries). Distinguishing prompt messages from normal query responses will cause unnecessary complications to the control program and should be avoided.

Handling events

If an error occurs while performing a calibration adjustment step, the DAQ6510 displays an error message on the front panel. Typically, errors occur during calibration adjustment only if something is wrong with the test setup or reference signal (for example, if the calibrator output is not enabled before sending a cal.adjust command). After correcting the cause, the cal.adjust command
that generated an error can be sent again without having to restart the entire calibration adjustment sequence. An automated test program can check for errors using the eventlog.next() function.
Refer to the DAQ6510 Reference Manual for details on eventlog commands. The same text that is
displayed after a cal.adjust error is returned in the eventlog.next() message.
) for
3-40 DAQ6510-905-01 Rev. D June 2022
TSP commands ........................................................................4-1
In this section:

TSP commands

The TSP commands available for the instrument are listed in alphabetical order.

Introduction

Section 4

TSP command reference

This section contains detailed information on the DAQ6510 remote calibration commands.
Section
Calibration and Adjustment Manual
Attribute (R)
Yes
Not applicable
Nonvolatile memory
Not applicable
adjustments
The number of adjustments
Count = cal.adjust.count
Assign the number of times the instrument has been adjusted
This shows that the instrument has been adjusted 3 times.
4: TSP command reference DAQ6510 Data Acquisition and Multimeter System

cal.adjust.count

This attribute returns the number of times the instrument has been adjusted.
Type TSP-Link accessible Affected by Where saved Default value
Usage
adjustments = cal.adjust.count
Details
You can use this command if calibration is locked or unlocked.
The adjust count is read-only. The count is automatically incremented by one when the cal.save() command is sent with calibration unlocked.
Example
Also see
print(Count)
cal.adjust.date (on page 4-3)
to a user variable named Count. Output the value. Example output:
3
4-2 DAQ6510-905-01 Rev. D June 2022
DAQ6510
ence
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
Not applicable
adjustDate
The date when the last adjustment occurred
year
Year; must be more than 1970
month
Month (1 to 12)
day
Day (1 to 31)
hour
Hour in 24-hour time format (0 to 23)
minute
Minute (0 to 59)
second
Second (0 to 59)
lastCal = cal.adjust.date
Assign the last adjustment date of the instrument to a user
Sep 23 2018 10:07:51.447
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 4: TSP command refer

cal.adjust.date

This attribute contains the date when the instrument was last adjusted.
Type TSP-Link accessible Affected by Where saved Default value
Usage
adjustDate = cal.adjust.date cal.adjust.date = (os.time{year = year, month = month, day = day}) cal.adjust.date = (os.time({year = year, month = month, day = day, hour = hour, min
= minute, sec = second}))
Details
Example
Also see
The date and time are returned in the format:
MMM DD YYYY HH:MM:SS.NNN
Where:
MMM DD YYYY is the month, date, and year
HH:MM:SS.NNN is the hour, minute, second, and fractional second
You can read this command if calibration is locked or unlocked. To set the date and time, calibration must be unlocked.
print(lastCal)
cal.adjust.count (on page 4-2)
variable named lastCal. Output the value. Example output:
DAQ6510-905-01 Rev. D June 2022 4-3
Section
Calibration and Adjustment Manual
stepname
The adjustment step to start
value
The value for this adjustment step. If using the default, then value is optional
4: TSP command reference DAQ6510 Data Acquisition and Multimeter System

cal.adjust.step.setup()

This function sets up the specified adjustment step.
Type TSP-Link accessible Affected by Where saved Default value
Function Yes
Usage
cal.adjust.step.setup(stepname) cal.adjust.step.setup(stepname, value)
Details
This command generates an error if:
Calibration is locked
The step does not complete successfully
Also see
The value that is passed is invalid for the step, out of range, or not needed
cal.lock() (on page 4-6) cal.unlock() (on page 4-9)
4-4 DAQ6510-905-01 Rev. D June 2022
DAQ6510
TSP command reference
stepname
The adjustment step to start
value
Value for this adjustment step. If using the default, then value is optional
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 4:

cal.adjust.step.execute()

This function executes the specified adjustment step.
Type TSP-Link accessible Affected by Where saved Default value
Function Yes
Usage
cal.adjust.step.execute(stepname) cal.adjust.step.execute(stepname, value)
Details
This command generates an error if:
Calibration is locked
The step does not complete successfully
Also see
The value that is passed is invalid for the step, out of range, or not needed
cal.lock() (on page 4-6) cal.unlock() (on page 4-9)
DAQ6510-905-01 Rev. D June 2022 4-5
Section
Calibration and Adjustment Manual
Function
Yes
cal.unlock("KI000CAL")
Unlock the calibration for the instrument using the default password.
Lock the calibration data.
4: TSP command reference DAQ6510 Data Acquisition and Multimeter System

cal.lock()

This function prevents access to instrument calibration.
Type TSP-Link accessible
Usage
cal.lock()
Details
Calibration data is locked during normal operation. To perform calibration, you must unlock calibration.
This command does not save calibration data.
Calibration data is lost if it is you do not save it before sending cal.lock(). Use cal.save() to
save the data.
An error is generated if this command is issued when calibration is already locked.
Example
-- Perform operations to generate the calibration data cal.save() cal.lock()
Affected by Where saved Default value
Save the calibration data.
Also see
cal.save() (on page 4-8)
4-6 DAQ6510-905-01 Rev. D June 2022
cal.unlock() (on page 4-9)
DAQ6510
TSP command reference
Attribute (W)
Yes
Not applicable
Nonvolatile memory
KI000CAL
password
A string that contains the password to unlock calibration; maximum of 10 characters
cal.unlock("KI000CAL")
To change the default calibration password, unlock
subsequent unlocks.
Data Acquisition and Multimeter System Calibration and Adjustment Manual Section 4:

cal.password

This attribute sets the password that you send when you unlock calibration.
Type TSP-Link accessible Affected by Where saved Default value
Usage
cal.password = "password"
Details
This command can only be sent when calibration is unlocked.
The password is not saved until calibration is saved with cal.save().
Be sure to record the password; there is no command to retrieve the password once it is set.
Example
Also see
cal.password = "XYZCorp" cal.lock() cal.unlock("XYZCorp")
cal.save() (on page 4-8) cal.unlock() (on page 4-9)
the calibration with the default password. Sets the password to XYZCorp. Lock calibration. Use the password XYZCorp for
DAQ6510-905-01 Rev. D June 2022 4-7
Section
Calibration and Adjustment Manual
Function
Yes
cal.unlock("KI000CAL")
Unlock the calibration for the instrument using the default password.
Lock the calibration data.
4: TSP command reference DAQ6510 Data Acquisition and Multimeter System

cal.save()

This function saves the calibration constants.
Type TSP-Link accessible
Usage
cal.save()
Details
This command stores the internally calculated calibration constants that were derived during the comprehensive calibration procedure. It also sets the adjustment date and increments the adjustment count. Calibration constants are retained indefinitely once saved.
Calibration is temporary unless the changes are saved. Calibration data is not saved if:
Calibration is locked.
Invalid data exists (for example, if a calibration step failed or was aborted).
Example
-- Perform operations to generate the calibration data cal.save() cal.lock()
Save the calibration data.
Affected by Where saved Default value
Also see
cal.adjust.count (on page 4-2) cal.adjust.date (on page 4-3) cal.lock() (on page 4-6)
4-8 DAQ6510-905-01 Rev. D June 2022
cal.unlock() (on page 4-9)
Loading...