Tektronix CSA8UP, TDS8UP, CSA8BUP, TDS8BUP, CSA82UP,,TDS82UP Option XP Field Install Instructions

Instructions
CSA8UP, TDS8UP, CSA8BUP, TDS8BUP, CSA82UP, and TDS82UP Option XP
TDS8000 and CSA8000 Series Instruments Microsoft Windows XP Upgrade Kit
071-2156-00
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
*P071215600*
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered tradem arks of Tektronix, Inc.
Windows is a registered trademark of Microsoft Corporation in the United Stat es and other countries.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-800-833-9200.
H Worldwide, visit www.tektronix.com to find contacts in your area.
Kit Description and Installation
This kit includes hardware and software to upgrade your CSA8000, TDS8000, CSA8000B, TDS8000B, CSA8200, and TDS8200 instruments from Microsoft Windows 2000 to Windows XP.
NOTE. Installation of the Microsoft Windows XP operating system requires that the instrument currently has the Microsoft Windows 2000 operating system installed.
Tektronix installation service (Option IF) is available for this upgrade. If installation service is desired, please contact the service center in your region to schedule installation.
Products
CSA8000 All serial numbers (if previously upgraded to Windows 2000) TDS8000 All serial numbers (if previously upgraded to Windows 2000) CSA8000B All serial numbers (if previously upgraded to Windows 2000) TDS8000B All serial numbers (if previously upgraded to Windows 2000) TDS8200 All serial numbers CSA8200 All serial numbers
Kit Parts List
Quantity Part number Description
1ea 071--2156--00 Kit Instructions: Windows 2000, CSA8000 and TDS8000
Series
1ea 650--5047--xx Hard drive: Windows XP and 8000 Series product sof tware
pre-installed
1ea 020--2786--xx 8000 Series product software CD-ROM designed for
Microsoft Windows XP
1ea 020--2787--xx 8000 Series Windows XP Operating System Restore:
5 CD-ROMs, Document
1ea 063--3860--xx BIOS Update CD--ROM
1ea 063--3868--xx Microsoft Windows XP Certificate of Authenticity Label
(This label must be applied to the instrument after installation is complete.)
CSA/TDS8UP, CSA/TDS8BUP, and CSA/TDS82UP Option XP Upgrade Kit
1
Kit Description and Installation
Installation
These instructions are for personnel who are familiar with servicing the product. If you need further details for disassembling or reassembling the product, refer to the appropriate service manual. Contact your nearest Tektronix, Inc., Service Center or Tektronix Factory Service for installation assistance.
CAUTION. To prevent static discharge damage, service the pr oduct only in a static-free environment. Observe standard handling precautions for static-sensi­tive devices while installing this kit. Always wear a grounded wrist strap, grounded foot strap, and static resistant apparel while installing this kit.
Check Instrument BIOS
Before installing the new hard disk drive, you must have the correct BIOS installed. Do the following procedure to check the BIOS version and update if necessary.
1. Power on the instrument.
2. While the instrument is booting, press F2 to enter the BIOS setup.
3. Check the first line of the BIOS display. The first line of the BIOS display
should be the BIOS version number. The BIOS version must be 1.03.03.
4. If the BIOS is 1.03.03 or greater:
a. Exit the BIOS setup without making any changes.
b. Power off the instrument.
5. If the BIOS is a version less than 1.03.03:
a. Exit the BIOS setup without making any changes.
b. After the instrument is finishes the boot cycle, insert the BIOS CD
contained in this kit.
c. Follow the update prompts provided with the CD. BIOS update
instructions are also located on the Readme file located on the CD.
d. After the BIOS update is finished, power off the instrument.
2
CSA/TDS8UP, CSA/TDS8BUP, and CSA/TDS82UP Option XP Upgrade Kit
Kit Description and Installation
Replace Hard Disk Drive
Access and remove the old hard drive using the procedure below. If you need additional instructions, see the service manual for the product you are servicing.
Follow these procedures to remove the hard drive:
CAUTION. Do not remove the replaceable hard disk drive when the instrument is powered on.
The replaceable hard disk drive may be permanently damaged if it is removed while the instrument is powered on.
1. Set the instrument so that its bottom is down on the work surface and its rear
panel is facing you.
2. Prepare to remove the hard disk drive:
a. Verify that the chassis is powered off and unplugged.
b. Push the hard disk drive cover in and the hard drive will disconnect from
the latch.
c. Grasp the hard disk drive assembly and slide it out of the instrument.
Depress the latch
Remove the
hard disk drive
3. Lay the hard disk drive assembly face down.
4. Remove the four #0 phillips screws that fasten the hard disk drive to the
cartridge.
CSA/TDS8UP, CSA/TDS8BUP, and CSA/TDS82UP Option XP Upgrade Kit
3
Kit Description and Installation
Remove 4 screws
5. Carefully remove the hard disk drive from the cartridge, disconnecting the
ribbon cable from the hard disk drive.
6. Insert the new hard disk drive (from this kit) into the cartridge, connecting
the ribbon cable to the hard disk drive.
7. Reinstall the four phillips screws to fasten the hard disk drive to the
cartridge.
8. Push the hard disk drive chassis back into the instrument.
Verify Operation
Successful startup and product software launch verifies operation. No further verification is needed. If the New Device Found dialog box appears, refer to the Possible Driver Problems section for information on how to proceed.
Apply Windows XP Authenticity Label
Place the Windows XP Certificate of Authenticity label on the rear panel of the instrument. This is an important step to show that the instrument is licensed to run the Windows XP operating system.
Possible Driver Problems
In some instances, the new hard drive may cause the New Device Found dialog message to display. The two most likely causes are the need for reinstallation of device drivers.
National Instruments GPIB
4
If the New Device Found dialog box displays and requires the drivers for the National Instruments GPIB card, follow these steps along with the dialog box.
1. Select No -- Not at this time to not search the internet for drivers.
CSA/TDS8UP, CSA/TDS8BUP, and CSA/TDS82UP Option XP Upgrade Kit
2. Select Next.
3. Select Install Software Automatically.
4. Select Next.
5. Select Finish.
Kit Description and Installation
Tektronix Power PC Board
If the New Device Found dialog box displays and requires the drivers for the Tektronix Power PC Board, follow these steps along with the dialog box.
1. Select No -- Not at this time to not search the internet for drivers.
2. Select Next.
3. Select Install Software Automatically.
4. Select Next.
NOTE. If the driver cannot be located automatically, you will need to browse to its location at c:\Tektronix\Tektronix custom.
5. Select Finish.
g End of document g