CSA7000, TDS7000, TDS6000, TLA600, &
OTS9000 Series Instruments
071-0716-04
Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered tradem arks of Tektronix, Inc.
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary in the product service manual
or the instruction manual.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
To prevent the instrument and rack from falling onto the operator, two or more
installers should install the instrument into the rack cabinet. After completing the
installation procedure, the installers should verify that the instrument and rack
cabinet will not tip forward while the instrument is in the extended position.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
CSA7000, TDS7000, TDS6000, TLA600 & OTS9000 Series Rackmount (Option 1R)
1
Service Safety Summary
2
CSA7000, TDS7000, TDS6000, TLA600 & OTS9000 Series Rackmount (Option 1R)
Kit Description
Products
Kit Parts List
This introduction describes the installation of option 1R (rackmount kit) to your
standard bench-top instrument.
The rackmount kit is a collection of parts that, once installed, configure the
instrument for mounting into a standard 19-inch equipment rack.
CSA7000, TDS7000, TDS6000, TLA600 & OTS9000 Series Rackmount (Option 1R)
3
Kit Description
Warranted Characteristics
Hot air out
When the instrument is installed according to the instructions in this document,
the rackmounted instrument meets all warranted requirements listed in the
instrument specification except for those listed in Environmental Requirementson page 5. Instruments mounted using methods other than those described in
these instructions may cause the instrument to not meet its warranted requirements.
See Specification in the user or service manual that applies to your instrument
model for tables of the warranted characteristics.
Cooling air enters on the bottom and right sides as shown in Figure 1. You
assume the responsibility to provide adequate cool air to meet the ambient
temperature requirements listed in Table 1.
Measure
input air
temperature
Figure 1: instrument cooling
Cool air in
4
CSA7000, TDS7000, TDS6000, TLA600 & OTS9000 Series Rackmount (Option 1R)
Kit Description
Environmental
Requirements
The following environmental characteristics supercede those listed in the user or
service manual for your instrument. The following table does not apply to the
OTS9000 Series instruments.
Table 1: Warranted characteristics
CharacteristicDescription
Temperature, Inside Rack Cabinet
Operating+10_ Cto+45_ C
Non-operating--22_ Cto+60_ C
Vibration
Operating0.24 g rms, from 5 to 500 Hz, 10 minutes
each axis
Non-operating2.22 g rms, from 5 to 500 Hz, 10 minutes
each axis
Shock
Operating and Non-operating20 g, 11 ms, half sine, each axis, 3 drops
each
Clearance Requirements
The rack in which the rack adapted instrument is mounted must provide the
following clearance requirements:
HA minimum of 266.7 mm (10.5 in) of vertical space.
HA minimum width of 448 mm (17 5/8 in) between the left- and right-front
HA minimum inside height depth of at least 508 mm (20 in).
WARNING. Adhering to these clearance requirements mounts the rack-adapted
instrument with sufficient clearance for air circulation and accommodation of
the power cord and mounting hardware. Failure to provide these clearances can
result in overheating and can cause the instrument to not operate properly
and/or fail.
rails in the rack.
CSA7000, TDS7000, TDS6000, TLA600 & OTS9000 Series Rackmount (Option 1R)
5
Kit Description
531.925 mm
(20.925 in)
424.053 mm
(16.695 in)
421.64 mm
(16.80 in)
471.170 mm
(18.550 in)
50.80 mm
(2.00 in)
500.024 mm
(19.686 in)
482.600 mm
(19.000 in)
161.925 mm
(6.375 in)
465.480 mm
(18.326 in)
Figure 2: instrument with Rack Adapter Installed
60.325 mm
(2.375 in)
30.480 mm
(1.200 in)
265.938 mm
(10.470 in)
258.639 mm
(10.160 in)
29.72 mm
(1.170 in)
6
CSA7000, TDS7000, TDS6000, TLA600 & OTS9000 Series Rackmount (Option 1R)