Current (Amplitude Mode)
(Used to calibrate current probes.)
Range — 1 mA to 100 mA(1-2-5
sequence).
Multipliers — 1, 2, 3, 4, 5, 6, 8 or 10.
Aberrations — Less than 5% of period
and less than ±15% of amplitude (±100 µA).
Accuracy — ±(0.25% +2 µA).
Frequency — DC or 10 Hz to 1 MHz
(decade steps).
Droop — ≤1% p-p amplitude.
Variable Range — ±9.9%.
Low Edge (Amplitude Mode)
(Used to test input amplifier and attenuator
compensation.)
Range — 20 mV to 1 V p-p, 50 Ω load
only (1-2-5 steps with multipliers).
Multipliers — 1, 2, 3, 4, 5, 6, 8 or 10.
Aberrations — ±2% of square wave
amplitude.
Accuracy — ±3%.
Polarity — Positive or negative transi-
tions to ground.
Risetime/Fall time — <1.3 ns.
Long T erm Flatness —±0.5% after first 10 ns.
Frequency —10 Hz to 1 MHz (decade steps).
Variable Amplitude Range — ≥±9.9%
from nominal.
High Edge (Amplitude Mode)
(Used to test input amplifier and attenuator
compensation.)
Range — 1.2 V to 100 V p-p, 1 M Ω load
(1-2-5 steps with multipliers).
Multipliers — 1, 2, 3, 4, 5, 6, 8 or 10.
Aberrations — ±2% of square wave
amplitude.
Accuracy — ±3%.
Polarity — Positive transition only
(negative voltage rising to ground).
Risetime — <100 ns.
Long Term Flatness — ±0.5% after first
500 ns.
Frequency — 10 Hz to 100 kHz (decade
steps).
Variable Amplitude Range — ≥±9.9%
from nominal.
MODULAR TEST INSTRUMENTS
CG 5011 Specifications
Voltage (Amplitude Mode)
(Used to calibrate vertical display accuracy .)
Range — 40 µ V to 200 V, 1 M Ωload;
40 µ V to 5 V , 50 Ωload (1-2-5 steps with
multiplier).
Multipliers — 1,2,3,4, 5, 6, 8 or 10.
Polarity — Positive from ground.
Aberrations — Less than ± 15% of
amplitude (± 10 mV).
Accuracy — ±(0.25% +1 µ V).
Frequency —
40 µ V to 80 mV 10 Hz to 10 kHz
100 mV to 10 V 10 Hz to 100 kHz
+dc or -dc
12 V to 200 V 10 Hz to 10 kHz
+dc or -dc
Variable Range — ±9.9%.
TM 500 / 5000