Tektronix CA-573 User manual

Model CA-573
Keithley Instruments, Inc. 4-Pin Interlock Cable
28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY
http://www.keithley.com

Description

The Model CA-573 is a 4-pin interlock cable assembly that is unterminated on one end.
Figure 1: Model CA-573-3 interlock cable
This interlock cable assembly is used with the M odel 2657A-LIM-3 Interconnect Module. Connect this cable to t he INTERLOCK OUT connector on the Model 2657A-LIM-3 and connect t he unterminated end to a user-supplied, normally-open switch.
Install the interlock switch on a test fixture, on the enclosure of a semiconductor prober or device handler, or on the door or doors of a test equipment rack. The switch must be installed so that it opens when the lid or door of the test enclosure is opened. The switch must close whe n the lid or door of the test enclosure is closed.
With the safety interlock properly configured f or a source-measure unit (SMU) test system using the Model 2657A-LIM-3, power from all the SMUs is disabled when the lid or door of the test enclosure is opened.
You can order the cable in four different length s (s ee table below).
Model number Cable length
CA-573-1 40 in. (1.0 m) CA-573-2 80 in. (2.0 m) CA-573-3 120 in. (3.0 m) CA-573-4 160 in. (4.0 m)
PA-1039 Rev. A / March 2012 1
Model CA-573

Electrical characteristics

Maximum working voltage: 32 V DC

Connector pin identification

The following pin-out only applies when the Model CA-573 is connected to the Model 2657A-LIM-3:
4-pin connector* Wire color
Pin 1: + 5 V connected to Red wire Pin 2: Interlock connected to Black wire * Pins 3 and 4 are not connected.
2 PA-1039 Rev. A / March 2012