The servicing instructions are for use by qualified personnel
only. To avoid personal injury, do not perform any servicing
unless you are qualified to do so. Refer to all safety summaries
before performing service.
Tektronix products are c overed by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O . B ox 50 0
Beaverto
USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to find contacts in your area.
Table of Contents
General safety summary ........................ ................................ ................................ ...v
Service safety summary..........................................................................................vii
system .................. .................................. ................................ ..20
AWG5200 Series Technical Referenceiii
Table of Contents
Table 39: Marke
r low level accuracy.............................................. ..............................61
ivAWG5200 Series Technical Reference
General safety summary
General safet
To avoid fir
e or personal
injury
y summary
Review the fo
this product or any products connected to it.
To avoid pot
Only qualified personnel should perform service procedures.
Use proper
certified for the country of use.
Ground th
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
llowing safety precautions to avoid injury and prevent damage to
ential hazards, use this product only as specified.
power cord. Use only the power cord specified for this product and
e product. This product is grounded through the grounding conductor
Power disconnect. The power cord disconnects the product from the power source.
Do not block the power cord; it must remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual's installation instructions for details
on installing the product so it has proper ventilation.
AWG5200 Series Technical Referencev
General safety summary
Termsinthismanual
Symbols and terms on the
product
These terms may
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may a ppear on the product:
DANGER in
the marking.
WAR NI NG
read the marking.
CAUTIO
The following symbol(s) may appear on the product:
appear in this manual:
dicates an injury hazard immediate l y accessible as you read
indicates an injury hazard not immediately accessible as you
N indicates a hazard to property including the product.
viAWG5200 Series Technical Reference
Service safety summary
Service safet
ysummary
Only qualifiesafety summary and the General safety summary before performing any service
procedures.
Do not service alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
Use care when servicing with power on. Dangerous voltages or currents may exist
in this p
test leads before removing protective panels, soldering, or replacing components.
To avoi
d personnel should perform service procedures. Read this Service
roduct. Disconnect power, remove battery (if applicable), and disconnect
d electric shock, do not touch exposed connections.
AWG5200 Series Technical Referencevii
Service s afety summary
viiiAWG5200 Series Technical Reference
Preface
Related documents
This manual contains specifications and performance verification procedures for
the AWG5200 Series Arbitrary Waveform Generators.
The following documents are also available for this product and can be
downloade
AWG5200 Series Installation and Safety Manual. This document provides
safety in
071-3529-xx.
d from the Tektronix website www.tek.com/manual/downloads.
formation and how to install the generator. Tektronix part number:
AWG5200 S
programming commands to remotely control the generator. Tektronix part
number: 077-1337-xx.
AWG5200 User Manual. This document is a printable version of the
AWG5200 help system. Tektronix part number: 077-1334-xx.
eries Programmer Manual. This document provides the
AWG5200 Series Technical Referenceix
Preface
xAWG5200 Series Technical Reference
Specifications
Performance conditions
This section contains the specifications for the AWG5200 series Arbitrary
Waveform Generators.
All specifications are typical unless noted as warranted. Warranted specifications
that are marked with the
To meet speci fications, the following conditions must be met:
The instrument must have been calibrated/adjusted at an ambient temperature
between +20 °C and +30 °C.
The instrument must be operating within the environmental limits. (See
Table 31 on page 23.)
The instrument must be powered from a sourc e that meets the specifications.
(See Table 29 on page 20.)
The instrument must have been operating continuously for at least 20 minutes
within the specified operating temperature range.
symbol are checked in this manual.
AWG5200 Series Technical Reference1
Specifications
Electrical sp
ecifications
Table 1: Run mode
Characteris
Continuous m
Triggered mode
Triggered
tics
ode
continuous mode
Description
An arbitrary
An arbitrar
is output, the instrument waits for the next trigger signal.
An arbitrary waveform is output continuously after a trigger signal is applied.
waveform is output continuously.
y waveform is output only once when a trigger signal is applied. After the waveform
Table 2: Arbitrary waveform
CharacteristicsDescription
Waveform memoryReal Waveforms: 2 Gs/channel
Complex waveforms: 1 Gs/channel
Minimum waveform size
Continuous run mode
Triggered run modes or sequence
Waveform granularity
Continuous run mode
Triggered run modes1 sample
IQ (Complex) waveform supportIQ waveforms, referred to as “Complex waveforms”, are supported for use with real time
1 sample
Real waveform: 2400 samples
Complex waveform: 1200 samples
Real waveforms are waveforms that have a single input value for each sample point. IQ
waveforms, referred to as “Complex waveforms”, use 2 values for each sample point.
1 sample
digital up-conversion and play out. The carrier signal is generated independently of the
waveform with an NCO (Numerically Controlled Oscillator).
The w aveform requires 2 values for each sample point. In the IQ waveform, I and Q
samples alternate in pairs or groups depending on the interpolation selection. The
format depends on the interpolation rate selected (2x or 4x)
2AWG5200 Series Technical Reference
Specifications
Table 3: Real ti
CharacteristicsDescription
Double Data Rate Interpolation (DDR
Mode)
Digital Up-conversion (DIGUP license
required)
Waveform interpolationReal time interpolation of IQ (complex) waveforms is supported independently on each
Inverse SINC filterReal time correction of the sinx/x frequency roll off can be enabled or disabled
me digital signal processing
Enabling DDR mode increases the output sample rate to 5 to 10 GS/s (2*fclk) and
interpolates the input sample data by 2X to match the output rate. 2X interpolation is
required for
With DDR enabled, the output image moves from ( fclk - fout) to (2*fclk - fout). Because
the input data rate does not increase, the output bandwidth remains (fclk/2).
DDR is most u
to specify the output center frequency up to the DDR Nyquist frequency. When the
waveform is a traditional, real valued, waveform (not IQ), enabling DDR applies a low
pass filter a
(fclk/2) and (2*fclk - fclk/2).
The DAC system in each channel includes a digital IQ modulator and numerically
controlled oscillator (NCO ) that provides digital up-conversion to a specified carrier
frequenc
Digital up-conversion requires an IQ input waveform. In the IQ waveform I and Q
samples alternate in pairs or groups depending on the interpolation selection.
Digital u
interpolation w hen a lower waveform sample rate is needed.
channel during play out.
Suppor
Only IQ (complex) waveforms can be interpolated. The interpolation factor refers to the
sample rate of the complex pair of points relative to the global instrument sample rate
set by t
factor is 2, then the waveform sample rate of both I and Q samples is 2.5 GS/s. DDR
interpolation offers an additional doubling o f the sample rate.
indep
sample rates above 5.0 GS/s.
seful when combined with digital up-conversion which allows the user
t a frequency just below (fclk/2) so that no signal is generate between
y
p-conversion can only be used with sample rates between 2.5 and 5 GS/s. Use
ted interpolation rates are 2x and 4x.
he clock. For example if the sample rate is set to 5 GS/s and the interpolation
endently on each channel.
Table 4: Sequencer
CharacteristicsDescription
Number of steps
Maximum repeat count
16,384
14 address bits. Numbers are zero-0 based in HW (0 to -16383).
1048576 (2
20
)
AWG5200 Series Technical Reference3
Specifications
Table 5: Sample
CharacteristicsDescription
Sample rateThe sample clock frequency is a global parameter that applies to all channels. DDR can be
DDR enabled:
DDR disabled
Sample rate resolution
Jitter Reduction Mode (PLL
integer mode)
Without Jitter Reduction (PLL
FracN mode)
Sample rate frequency
accuracy
10 MHz reference accuracy
clock generator
enabled on a per channel basis allowing the sam ple rate to be doubled on selected channels.
The sample cl
lower than 2.5 GS/s, the system replicates points. The number of replicated points increases
by powers of 2, therefore the clock frequency is SR×2
a frequency
When using complex waveforms digital up conversion, the sample rate is limited to 2.5 GS/s to
5 GS/s. To achieve lower sample rates, use waveform interpolation.
Real waveforms: 596 S/s to 10 GS/s
Complex (I
Real wave
Complex (IQ) waveforms: 2.5 GS/s to 5 GS/s
3 digits with jitter reduction (50 MHz sample clock frequency steps from 2.5 GHz to 5 GHz).
With DDR enabled, the resolution is 100 MHz Sample rates below the clock range are a power
of 2 div
the 50 MHz stepped frequencies.
8digit
Sample Rate * 10 MHz Ref Accuracy/10 MHz
Examp
10 MHz ± 20 Hz
perature between 0 to 50 °C; includes aging within 1 y ear of calibration.)
(Tem
ock frequency is always between 2.5 GHz and 5 GHz. To achieve sample rates
n
,. where n is an integer that results in
between 2.5 GHz and 5 GHz.
Q)waveforms: 5GS/sto10GS/s
forms: 298 S/s to 5 GS/s
ision of the clock frequency so Low Jitter sample rates are a power of 2 divisions of
s
le:5GS/s*(±20Hz)/10MHz=10kHz
Table 6: Analog output skew
CharacteristicsDescription
Skew between (+) and (–) outputs
Skew between channels
(DChighBWmodeonly)
Delay change from DC High BW
output path to other output paths
DC High Volt
(Option HV)
AC Direct
AC Amplified
(Option AC)
Skew adjustment range
±15 ps
±25 ps
Skew is calibrated using the (+) outputs of the DC High BW output path for each channel.
Channel delay will change when a different path is selected or when various DAC features are
enabled.
1.2 ns
340 ps
740 ps
±2 ns
Used to adjust skew between channels in a single instrument.
4AWG5200 Series Technical Reference
Specifications
Table 6: Analog output skew (cont.)
CharacteristicsDescription
Skew adjustment resolution250 fs
Skew stability between channels
Sync out to channel< ±0.5 ps/ °C
Channel to channel< ±0.5 ps/ °C (±0.18 ps/ ° C @ 1 GHz)
Phase adjustmentUsed to adjust skew between all channels in an instrument relative to another instrument.
Range
Resolution
-8,640° to +8,640° of the DAC clock.
1° of the DAC clock.
Table 7
CharacteristicsDescription
Connector type2 SMA connectors per channel.
Number of outputsAWG5202: 2.
DAC
Type of outputs
ON/OFF controlIndependent control for each analog output channel.
Output impedance50 Ω
: Signal output characteristics
AWG5204: 4.
208: 8.
AWG5
resolution
Output path
DC High BW(+) and (–) complementary (differential).
DC High Volt
(Option HV)
AC DirectSingle ended output from the (+) connector.
AC Amplified
(Option AC)
16, 15, 14, 13 or 12 bits.
bling markers degrades resolution.
Ena
16-bit mode: 0 markers available.
15-bit mode: 1 marker, M1.
bit mode: 2 markers, M1, M2.
14-
13-bit mode: 3 markers, M1, M2, M3.
12-bit mode: 4 markers, M1, M2, M3, M4.
Includes a variable gain, high bandwidth, DC coupled amplifier in the signal path.
(+) and (–) complementary (differential).
An additional amplifier adds high amplitude with reduced bandwidth.
A direct connection to the DAC output including a balun to reduce common mode distortion.
The AC Direct path offers the lowest noise and distortion performance.
Single ended output from the (+) connector.
Includes an amplified path and a passive variable attenuator path to provide a large output
amplitude range.
In NRZ mode, each sample is held for the entire sam ple period (1/sample rate). This results
in the familiar sin(x)/x frequency response. With DDR mode enabled, the sin(x)/x bandwidth
doubles.
In RZ mode, each sample is held for half of the sample period. This doubles the sin(x)/x
bandwidth, but reduces the amplitude by half. This may be useful when playing a real
waveform with the signal in the second Nyquist zone. For real waveforms, DDR mode filters
the signal in the 2nd and 3rd Nyquist zones and is not useful in this case.
In Mix mode, each sample is inverted for the second half of the sample period. This is
effectively like mixing the output waveform with the sample clock. This boosts the signal in
the s econd Nyquist zone, but zeros the DC component of the waveform and reduces low
frequency components. This may be useful when playing a real waveform with the signal in
the second Nyquist zone. For real waveforms, DDR mode filters the signal in the 2nd and 3rd
Nyquist zones and is not useful in this case.
fsample = sample rate.
The sin(x)/x bandwidth can be solved by using the following equation:
20 * log (sin(x)/x) = –3.
x=π * fout ÷ fsample.
fsample = sample rate.
fout = sin(x)/x bandwidth.
Independent amplitude control for all channels.
Units of dBm or V can be selected.
50 mV
25 mV
50 mV
10 mV
20 mV
to 750 mV
p-p
to 1.5 V
p-p
to 1.5 V
p-p
to 3.0 V
p-p
to 5 V
p-p
to 10.0 V
p-p
into 50 Ω single-ended.
p-p
into 100 Ω differential.
p-p
into 50 Ω single-ended.
p-p
into 100 Ω differential.
p-p
into 50 Ω single-ended.
p-p
into 100 Ω differential.
p-p
6AWG5200 Series Technical Reference
Table 7: Signal output characteristics (cont.)
CharacteristicsDescription
AC Direct
AC Amplified
(Option AC)
Amplitude adjustment resolution
Output paths
DC High BW
DC High Volt
(Option HV)
AC Direct
AC Amplified
(Option AC)
DC amplitude accuracy
Output path
DC High BWAmplitude < 100 mV: ±(5% of amplitude).
igh Volt
DC H
(Option HV)
AC amplitude accuracy
Output path
AC Direct0.5 dB at 100 MHz (0 °C to 45 °C)
AC Amplified
(Option AC)
DC Offset range
Output path
DC High BW± 2 V into 50 Ω to ground.
DC High Volt
(Option HV)
DC Offset resolution
Output path
DC High BW
DC High Volt
(Option HV)
–17 dBm to –5 dBm.
10 MHz to 3.5 GHz.
–85 dBm to 10 dBm (10 MHz to 3.5 GHz.)
–50 dBm to 10 dB m (3.5 GHz to 5 GHz.)
Amplitude accuracy and flatness degrades at frequencies beyond 3.5 GHz and below –50 dBm
output amplitude. It is not recommended to operate in this region.
1.1 mV or 0.1 dB.
1.1 mV or 0.1 dB.
0.1 dB
0.1 dB
Within ±5 °C of internal self calibration temperature.
Amplitude 100 mV to 750 mV: ±(2% of amplitude).
litude 100 mV to 1.5 V (Option DC): ±(2% of amplitude).
Amp
litude < 160 mV: ±(5% of amplitude).
Amp
Amplitude 160 mV to 5 V: ±(2% of amplitude).
1 dB at 100 MHz (45 °C to 50 °C)
0.5dBat100MHz(0°Cto45°C)
1 dB at 100 MHz (45 °C to 50 °C)
± 4 V into high resistance or m atching voltage termination.
±2Vinto50Ω to ground.
± 4 V into high resistance or m atching voltage termination.
1mV
1mV
Specifications
AWG5200 Series Technical Reference7
Specifications
Table 7: Signal output characteristics (cont.)
CharacteristicsDescription
DC Offset accuracyDifferential offset is sensitive to output amplitude setting.
Within ±5 °C of internal self calibration temperature.
Common mode = ((OutP + OutN)/2).
Differential Mode = (OutP - OutN).
Output path
DC High BW
Common mo
(Warranted)
Differential mode±25 mV; into 100 Ω differential.
DC High Volt
(Option HV)
Common mode
(warranted)
Differential mode± 88 mV; Into 100 Ω differential.
AC output DC bias range
Output path
AC Direct
AC Amplified
(Option AC)
AC DC bias resistance
Output path
AC Direct1 Ω
AC Amplified
(Option AC)
AC DC bias accuracy
(warranted)
Output path
AC Direct±(2% of bias + 20 mV); into an open circuit (zero load current).
AC Amplified
(Option AC)
de
±(2% of |offset| + 10 mV); into 50 Ω to G nd.
±(2% of |offset| + 1% of amplitude + 20 mV).
± 5 V at 150 mA.
± 5 V at 150 mA.
1 Ω
±(2% of bias + 20 mV); into an open circuit (zero load current).
8AWG5200 Series Technical Reference
Table 7: Signal output characteristics (cont.)
CharacteristicsDescription
Analog bandwidth
Output path
DC High BW
DC High BW
(Option DC)
DC High Volt
(Option HV)
AC Direct10 MHz - 2 GHz (–3 dB bandwidth).
AC Amplified
(Option AC)
Rise/fall timeRise and fall times only apply to DC output paths.
Output path
DC High BW
DC High BW
(Option DC)
DC High Volt
(Option HV)
Step response aberrationsStep response aberrations only apply to DC output paths.
Output path
DC High BW< 16%
DC High BW
(Option DC)
DC High Volt
(Option HV)
Analog bandwidth is measured with the ideal sin(x)/x response curve of the DAC
mathematically removed from the measured data.
At 750 mV
single ended:
pp
DC - 2 GHz (–3 dB bandwidth).
DC - 4 GHz (–6 dB bandwidth).
At 1.5 V
single ended:
pp
DC - 1.3 GHz (–3 dB bandwidth).
The analog bandwidth degrades as the amplitude is increased beyond 750 mV.
At 2 V
single-ended:
pp
DC – 370 MHz (–3 dB bandwidth).
At 4 V
single-ended:
pp
DC – 200 MHz (–3 dB bandwidth).
10 MHz - 4 GHz (–6 dB bandwidth).
10 MHz - 2 GHz (–3 dB bandwidth).
10 MHz - 4 GHz (–6 dB bandwidth).
< 110 ps at 750 mV
< 180 ps at 1.5 V
<1.3ns,at5V
<1.1ns,at4V
<0.8ns,at3V
<0.6ns,at2V
, at 750 mVppsingle ended.
pp
< 16%
,at1.5Vppsingle ended.
pp
single ended.
pp
single ended.
pp
single-ended.
pp
single-ended.
pp
single-ended.
pp
single-ended.
pp
< 10%pp,at5Vppsingle ended.
Specifications
AWG5200 Series Technical Reference9
Specifications
Table 7: Signal output characteristics (cont.)
CharacteristicsDescription
Harmonic distortion
Output path
DC High BW(See Table 8 on page 11.)
AC Direct(See Table 9 on page 12.)
AC Amplified
(Option AC)
ENOB
(See Table 10 on page 12.)
SFDRSFDR is the difference in dB between a CW carrier signal and the largest spur, excluding
harmonics, within a defined frequency range around the carrier. Measured with a balun and
Table 15: Phase noise operating at 5.0 GS/s or 10 GS/s with DDR enabled
Analog output frequency
Offset100 MHz1 GHz2 GHz4 GHz
100 Hz
1kHz
10 kHz
100 kHz
1MHz
10 MHz
–112 dBc/Hz–92 dBc/Hz–86 dBc/Hz–80 dBc/Hz
–132 dBc/Hz–110 dBc/Hz–105 dBc/Hz–99 dBc/Hz
–138 dBc/Hz–118 dBc/Hz–112 dBc/Hz–106 dBc/Hz
–138 dBc/Hz–118 dBc/Hz–112 dBc/Hz–106 dBc/Hz
–148 dBc/Hz–128 dBc/Hz–122 dBc/Hz–116 dBc/Hz
–160 dBc/Hz–150 dBc/Hz–140 dBc/Hz–140 dBc/Hz
Table 16: Marker outputs
CharacteristicsDescription
Connector typeSMA on rear panel.
Number of outputs
Type of outputSingle ended.
ON/OFF ControlIndependent control for each marker.
Output impedance50 Ω
Output voltageIndependent control for each m arker.
Amplitude range
Window
Resolution0.1 mV
External termination voltage–1.0 V to +3.5 V.
Maximum output current60 mA
DC accuracy (warranted)
Rise/fall time< 150 ps (20% to 80% of swing when High = 0.4 V, Low = –0.4 V).
Aberrations
Random jitter
4 per channel.
Output voltage into RLOAD [Ω] to GND is approximately (2 * RLOAD / (50 + RLOAD) ) times of
voltage setting.
0.2 V
to 1.75 V
p-p
into 50 Ω.
p-p
–0.5 V to 1.7 V into 50 Ω.
±(10% of |output high or low setting| + 25 mV) into 50 Ω.
< 20%
for the first 1 ns following the step transition w ith 100% reference at 10 ns.
p-p
5ps
14AWG5200 Series Technical Reference
Table 16: Marker outputs (cont.)
CharacteristicsDescription
Sample rate2.5 GS/s to 5 GS/s.
Minimum pulse width400 ps
2 Samples at 5 GS/s.
Maximum data rate
Skew between markers
(From the same channel)
Variable delay control
Range±2 ns
Resolution1 ps
Accuracy
2.5 Gb/s.
Minimum pulse width does not support data output at maximum sample rate.
±25 ps
Independent control for each marker.
±25 ps from delay value.
Table 17: 10 MHz Ref Out (reference output)
CharacteristicsDescription
Connector typeSMA on rear panel.
Output impedance50 Ω (AC coupled).
Amplitude
Frequency (warranted)
+4 dBm, ±2 dBm. Sine wave output.
Within ±(1 ppm + Aging), Aging: ±1 ppm per year.
(Temperature between 0 °C to 50 °C.)
Specifications
Table 18: Ref In (reference input)
CharacteristicsDescription
Connector typeSMA on rear panel.
Input impedance
Input amplitude–5 dBm to +5 dBm.
Fixed frequency range
Variable frequency range
50 Ω (AC coupled).
10 MHz, ±40 Hz.
35 MHz to 240 MHz.
Acceptable frequency drift while the instrument is operating is ± 0.1%.
Table 19: Clock Out
CharacteristicsDescription
The external clock output is a copy of an internal clock generator that is used to create the
DAC sample clock. This clock always operates in the
multiplied and divided to create the effective DAC sampling rate.
Connector typeSMA on rear panel.
Output impedance50 Ω AC coupled.
Output amplitude
+3 dBm to +10 dBm.
octave range specified below. It is
AWG5200 Series Technical Reference15
Specifications
Table 19: Clock Out (cont.)
CharacteristicsDescription
Frequency range
2.5GHzto5GHz.
For sample rates lower than 2.5 GS/s the output frequency is: Fout = SR * 2n ; where n is an
integer that gives Fout between 2.5 GHz and 5 GHz.
Frequency resolution
Internal and fixed reference
clock operation
External variable reference
clock operation
With jitter reduction: 50 MHz.
Without jitter reduction: 100 MHz ÷ 2
With jitter reduction: Fref ÷ R.
Without jitter reduction: Fref ÷ R ÷ 2
Fref = reference clock frequency
R = 4 when 140 MHz < Fref ≤ 240 MHz
R = 2 when 70 MHz < Fref ≤ 140 MHz
R = 1 when 35 MHz ≤ Fref ≤ 70 MHz
Table 20: Clock In
20
.
20
CharacteristicsDescription
The external clock input can be used to create the DAC
operate in the octave range specified below. It is multiplied and divided to create the actual
DAC sample clock.
Connector typeSMA on rear panel.
Input impedance
50 Ω (AC coupled).
Input amplitude0 dBm to +10 dBm.
Frequency range
2.5GHzto5GHz.
Acceptable frequency drift while the instrument is operating is ±0.1%.
Table 21: Sync In
CharacteristicsDescription
Connector typeSMA on rear panel.
Input impedance
Input amplitude2.5 V
Frequency
500 Ω (AC coupled)
Max
p-p
Clock output ÷ 32.
Table 22: Sync Out
CharacteristicsDescription
Connector typeSMA on rear panel.
Output impedance50 Ω (AC coupled).
Output amplitude1 V
Frequency
, ±20% into 50 Ω.
p-p
Clock output ÷ 32.
sample clock. This clock must always
16AWG5200 Series Technical Reference
Specifications
Table 23: Sync C
lock Out
CharacteristicsDescription
Connector typeSMA on rear panel.
Output impedance50 Ω (AC coupled).
Output amplitude0.85 V to 1.25 V
Frequency
Clock output ÷ 32.
p-p
Table 24: Trigger Inputs
CharacteristicsDescription
Number of inputs2 (A and B)
On 2 and 4 channel instruments, only one trigger is usable for asynchronous triggering. On 8
channel instruments, both triggers can be used.
ConnectorSMA on rear panel.
Trigger modes
Input impedance
Slope / Polarity
Input voltage range
1kΩ selected
50 Ω selected
Input voltage minimum amplitude0.5 V
Threshold control
Range–5.0 V to 5.0 V.
Resolution0.1 V
Accuracy
Minimum pulse width20 ns
Synchronous and Asynchronous, selectable.
When asynchronous trigger mode is selected, playback starts on the next qualified sample
clock edge. If the trigger pulse has no fixed timing relationship with the sample clock, then
delay jitter w ill vary by 1 clock cycle.
When synchronous mode is selected, playback starts on the next qualified Sync Clock edge
(Clock ÷ 32). If the trigger pulse is made synchronous with the Sync Out clock, then very low
delay jitter is possible. Using the Sync Out clock provides a larger setup time for the trigger
pulse so that stable triggering can be achieved.
The DAC sampling clock frequency is displayed on the clock settings tab when the external
clock output is enabled.
8760/ fclk +68 ns ± 20 ns.
(1.820 μsat5GS/s)
(3.572 μsat5GS/s)
fclk is the frequency of the DAC sampling clock. The DAC sampling clock frequency is
displayed on the clock settings tab when the external clock output is enabled.
(1.685 μs at 5 GS/s.)
(3.340 μs at 2.5 GS/s.)
fclk is the frequency of the DAC sampling clock. The DAC sampling clock frequency is
displayed on the clock settings tab when the external clock output is enabled.
>2 μs
Trigger hold off is the amount of delay required at the end of a waveform before another trigger
pulse can be processed.
The asynchronous jitter performance is directly proportional the frequency of the DAC sampling
clock. The DAC sampling clock frequency is displayed on the clock settings tab when the
external clock output is enabled.
for 2.5 GHz DAC sampling clock.
p-p
240 ps
220 ps
300 fs
400 fs
1.7 ps
for 5 GHz DAC sampling clock.
p-p
,24ps
p-p
,14ps
p-p
rms
rms
rms
for 2.5 GHz DAC sampling clock.
rms
for 5 GH z DAC sampling clock.
rms
18AWG5200 Series Technical Reference
Specifications
Table 25: Patte
CharacteristicsDescription
Connector type15-pin D-sub female connector on rear panel.
Input signal
put levels
In
Input impedance
Number of jump destinations
Strobe
PolarityData is clocked in on negative edge.
Minimum pulse width64 ns
Setup and holdSetup: 5 ns.
Holdoff time
rn Jump In connector
pin assignment
Pin assignm
1
2Data bit 0, input
3Data bit 1, input
4Databit2
5
6
7
8
9
10Data bit 4, input
11Da ta b
12Data bit 6, input
13Data bit 7, input
14
15
3.3 V LVCMOS.
5VTTLcompliant.
1kΩ resistor pull down to GND.
256
Hold: 5 ns.
>18 μs
Strobe hold off is the amount of delay required at the end of a waveform before another strobe
pulse can be processed.
46.35 lb (21.02 kg)47.75 lb (21.66 kg)53 lb (24.04 kg)
153.6 mm (6.05 in)
460.5 mm (18.13 in)
603 mm (23.76 in)
50 mm (2 in)
Specifications
AWG5200 Series Technical Reference21
Specifications
22AWG5200 Series Technical Reference
Environmental characteristics
Table 31: Environmental characteristics
CharacteristicsDescription
Temperature
Operating0 °C to +50 °C (+32 °F to 122 °F)
Nonoperati
Relative humidity
Operating5% to 90% relative humidity at up to +30 °C (+86 °F).
Nonoperating
Altitude
OperatingUp to 3,000 m (approximately 10,000 feet).
Nonope
ng
rating
–20 °C to +60 °C (-4 °F to 140 °F) w ith 30 °C/hour (86 °F/hour) maximum gradient, with
no media installed in disc drives.
5% to 45% relative humidity above +30 °C (+ 86 °F) up to +50 °C (122 °F) noncondensing.
5% to 90% relative humidity at up to 30 °C.
5% to 45% r
Maximu
(4921 ft).
Up to 12,000 m (approximately 40,000 feet).
elative humidity above +30 °C (+86 °F) up to +60 °C (140 °F) noncondensing.
m operating temperature decreases 1 °C (34 °F) each 300 m (984 ft) above 1.5 km
Specifications
AWG5200 Series Technical Reference23
Specifications
24AWG5200 Series Technical Reference
Performance verification procedures
Two types of performance verification procedures can be performed on the
instrument: Brief Procedures and Perform ance Tests. You may not need to
perform all o
To rapidly confirm that the instrument functions and was adjusted properly,
perform Dia
Advantages: These procedures are quick to do and require no external equipment
or signal s
testing to provide high confidence that the instrument will perform properly.
f these procedures, depending on what you want to accomplish.
gnostics and Calibration.
ources. These procedures perform extensive functional and accuracy
To furthe
and then perform Functional Test.
Advanta
requires minimal equipment. The procedure can be used when the instrument is
first received.
If more extensive confirmation of performance is desired, complete the self
tests and functional test, and then do the Performance Tests.
Advantages: These procedures add d irect checking of warranted specifications.
These procedures require specific test equipment. (See page 42, Requiredequipment.)
If you are not familiar with operating this instrument, refer to the online help or
the user information supplied with the instrument.
r check functionality, first perform Diagnostics and Calibration,
ges: The procedure requires minimal additional time to perform, and
AWG5200 Series Technical Reference25
Performance verification procedures
Input and output options
The instrument has two USB ports on the front panel, and four USB ports on the
rear panel. (See Figure 1.) These ports can be used for an external mouse and/or
keyboard. Additionally, an external video display can be connected to the VGA
display port on the rear panel.
Figure 1
: Peripheral connections
26AWG5200 Series Technical Reference
Brief procedures
Brief procedu
Diagnostics
res
There are thr
basic functionality and proper adjustment:
Diagnostics
Calibration
Functional test
The following steps run the internal routines that confirm basic functionality
and proper adjustment.
EquipmentPrerequisites
NoneNone
1. Disconnect all the cables from the output channels.
2. From the Utilities tab, select Diag & Cal.
3. Click the Diagnostics & Calibration button and then select Diagnostics.
ee procedures in this section that provide a quick way to confirm
4. In the Diagnostics dialog box, confirm that all the check boxes are selected. If
they are not all selected, click the Select all tests button.
AWG5200 Series Technical Reference27
Brief procedures
Calibration
5. Click the Start
The internal diagnostics perform an exhaustive verification of proper
instrument fu
verification is completed, the resulting status will appear in the dialog box.
6. Verify that
complete.
7. Click the Cl
EquipmentPrerequisites
NonePower on the instrument and allow a
1. From the Utilities tab, select System.
2. From the Utilities tab, select Diag & Cal.
Click the Diagnostics & Calibration button and then select Calibration.
button to execute the diagnostics.
nction. This verification may take several minutes. When the
Pass appears as Status in the dialog box when the diagnostics
ose button.
20 minute warm-up before doing this
procedure.
3. Click the Start button to start the routine.
28AWG5200 Series Technical Reference
Brief procedures
4. Verify that Pas
calibration completes.
5. Click the Clos
s appears in the Summary column for all items when the
e button.
AWG5200 Series Technical Reference29
Brief procedures
Functional test
The purpose of the procedure is to confirm that the instrument functions properly.
The procedures use “AWG” when referring to the AWG5200 series instruments.
The required
equipment is listed below.
Table 32: Required equipment for the functional test
ItemQty.Minimum requirementsRecommended equipment
Oscilloscope
Function generator1 ea.1 kHz, square wave, 5 V
Signal analyzer
Adapter4 ea
50 Ω SMA cable
50 Ω SMA terminator
50 Ω BNC cable
SMA-BNC adapter
Planar Crown RF Input
Connector – 7005A-1 SMA
Female
1 ea.
1 ea.
4 ea.
3 ea.
1 ea.Male connectors both endsTektronix part number 012-0057-01
3 ea.
1 ea.
Bandwidth: 4 GHz or higher
4 channels
output
p-p
Bandwidth: 14 GHz or higherTektronix RSA5126B
TekConnect oscilloscope input to SMA inputTektronix TCA-SMA
DC to 20 GHz
DC to 18 GHz
SMA female to BNC male connector
Planar Crown RF Input Connector – Type N
to SMA Female
For use with Tektronix RSA5126B signal
analyzer
Tektronix DPO70404C
Tektronix AFG3021C
Tensolite 1-3636-465-5236
Tektronix part number 136-7162-xx
(supplied with AWG).
Tektronix part number 015-0572-00
Tektronix part number 131-8689-00
Functional check
requisites
pre
1. Click the Reset to Default Setup button in the toolbar.
2. Load the test waveform PV_Square.wfmx into the Waveform List.
Test waveforms are located at C:\Program
Files\Tektronix\AWG5200\Samples\PV.
30AWG5200 Series Technical Reference
Brief procedures
Checking the a
channel outputs
nalog
Required equipmentPrerequisites
Oscilloscope
One TCA-SMA adapter
One 50 Ω SMA cable
One 50 Ω SMA terminator
None
1. Set the test oscilloscope as follows:
a. Vertical scale: 200 mV/div (CH 1 and CH 2)
b. Horizontal scale: 100 ns/div
c. Input coupling: DC (CH 1 and CH 2)
d. Input impedance: 50 Ω (CH 1 and CH 2)
e. Position: +2 div (CH 1 and CH 2, if necessary)
f.Trigge
r source: CH 1
g. Trigger level: 0 mV
h. Trigger slope: Positive
i.Trigger mode: Auto
2. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
3. Connect CH 1 (+) of the AWG to channel 1 of the test oscilloscope using a
50 Ω SMA cable and a TCA-SMA adapter.
4. Connect CH 1 (–) of the AWG to channel 2 of the test oscilloscope using a
50 Ω SMA cable and a TCA-SMA adapter.
5. Click the Home tabonthedisplay.
AWG5200 Series Technical Reference31
Brief procedures
6. From the Wavefo
to Channel 1.
7. Click the Setu
8. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home
9. Click the Play button on-screen or press the button on the front panel of the
AWG.
10. Check that the channel’s waveform is properly displayed on the test
scope screen.
oscillo
rm List window, assign the waveform PV_Square.wfmx
p -> Channel tab and enable Channel 1.
screen) to enable the outputs (front panel light off).
11. Press AWG the front panel All Outputs Off button (or click All Outputs Off
he Home screen) to disable the outputs (front panel light on).
on t
12. Repeat steps 3 through 11 until all channels are checked, modifying the
structions with the channel number under test.
in
13. Disconnect the test setup.
32AWG5200 Series Technical Reference
Brief procedures
Checking the m
outputs
arker
Required equipmentPrerequisites
Oscilloscope
Four TCA-SMA adapters
Four 50 Ω SMA cables
None
1. Set the test oscilloscope as follows:
a. Vertical scale: 1 V/div (CH 1 through CH 4)
b. Horizontal scale: 100 ns/div
c. Input coupling: DC
d. Input impedance: 50 Ω
e. CH 1 through CH 4 position: adjust as necessary to display all four traces
f.Trigger
source: CH1
g. Trigger level: 0 mV
h. Trigger slope: Positive
i.Trigger mode: Auto
2. If needed, press the AWG front panel All Outputs Off button (or click All
Outputs Off on the Home screen) to disable the outputs (front panel light on).
3. Connect the AWG’s Channel 1 markers to the test oscilloscope using a
50 Ω SMA cable and a TCA-SMA adapter.
Connect marker CH1:1 to channel 1 of the test oscilloscope.
Connect marker CH1:2 to channel 2 of the test oscilloscope.
nnect marker CH1:3 to channel 3 of the test oscilloscope.
Co
Connect marker CH1:4 to channel 4 of the test oscilloscope.
OTE. If a channel’s marker is not connected to the test oscilloscope, it must
N
be terminated with a 50 Ω SMA terminator.
AWG5200 Series Technical Reference33
Brief procedures
4. Click the Home tabonthedisplay.
5. From the W
to Channel 1.
6. Click th
7. In the Setup -> Channel tab, select Output Settings and set the Channel 1
Resol
8. Click the Play button on-screen or on the front panel.
9. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
10. Check that the CH1:1 through CH1:4 waveforms are properly displayed on
the test oscilloscope screen.
aveform List window, assign the waveform PV_Square.wfmx
e Setup -> Channel tab and enable the select Channel 1 output.
ution to 12+4 Mkrs.
34AWG5200 Series Technical Reference
Brief procedures
11. Press th
on the Home screen) to disable the outputs (front panel light on).
12. Repeat s
instructions with the channel number under test.
Discon
e AWG front panel All Outputs Off button (or click All Outputs Off
teps 3 through 11 until all channels are checked, modifying the
nect the test setup.
AWG5200 Series Technical Reference35
Brief procedures
Checking the A
Coutput
Required equipmentPrerequisites
Signal analyzer
One 50 Ω SMA cable
Planar Crown RF Input Connector – Type N
to SMA Female
Two 50 Ω SMA terminators
None
1. If needed, press the AWG front panel All Outputs Off button (or click All
Outputs Off on the Home screen) to disable the outputs (front panel light on).
2. Create a 1 GHz test waveform from the AWG using the Basic Waveform
plug-in.
a. Click the Waveform Plug-in tabonthedisplay.
b. Select Basic Waveform from the Waveform Plug-ins drop down list.
c. Click the Reset Plug-in button.
d. Set the Function to Sine.
e. Set the Frequency to 1 GHz.
NOTE. Leave all other settings at their default settings.
f.Click the Compile Settings icon to open the compile settings dialog screen.
g. In the Name field, change the name to Waveform_1 GHz.
36AWG5200 Series Technical Reference
Brief procedures
h. Close the compi
i.Click Compile.
3. Set the spectrum analyzer as follows:
a. Press the Preset button to set the analyzer to its default settings.
b. Display the Spectrum measurement.
c. Set Center Frequency to 1 GHz.
4. Use a 50 Ω SM
the RF input of the signal analyzer.
le settings dialog screen.
A cable to connect the CH 1 AC connector (+) on the AWG to
5. Click the Setup -> Channel tab and click the Output Settings tab.
a. Select Channel 1.
b. Set the Output Path to AC Direct.
c. Enable the Channel 1 output.
AWG5200 Series Technical Reference37
Brief procedures
6. Click the Home tabonthedisplay.
7. In the Waveform List window, assign the Waveform_1 GHz waveform to
the Channel 1.
8. Press the Play button, or click Play on the display.
9. Press the A
on the Home screen) to enable the outputs (front panel light off).
10. Check tha
WG front panel All Outputs Off button (or click All Outputs Off
t the waveform is properly displayed on the signal analyzer screen.
Figure 2: 1 GHz output waveform
11. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
12. Repeat steps 4 through 11 until all channels are checked, modifying the
instructions with the channel number under test.
Disconnect the test setup.
38AWG5200 Series Technical Reference
Brief procedures
Checking the t
riggered
outputs
Required equipmentPrerequisites
Oscilloscope
Function Generator (AFG3021C or
equivalent)
One TCA-SMA adapter
Two 50 Ω SMA cables
One SMA female to BNC male adapter
None
1. Set the oscilloscope as follows:
a. Vertical scale: 200 mV/div (CH 1)
b. Horizontal scale: 20 ns/div
c. Trigger source: CH 1
d. Trigger level: 100 mV
2. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
3. Connect a BNC to SMA adapter to the output of the function generator.
4. Connect an SMA cable between the output of the function generator and the
Trigger A input on the rear panel of the AWG.
5. Connect CH 1 (+) of the AWG to channel 1 of the test oscilloscope using a
50 Ω SMA cable and a TCA-SMA adapter.
7. Click the Home tabonthedisplay.
AWG5200 Series Technical Reference39
Brief procedures
8. Click the Reset to Default Setup button in the toolbar.
9. Set the Function Generator to output a 1 kHz square wave at 5 V
p-p
.
10. Turn on the output of the Function Generator.
11. Load the test waveform PV_Square.wfmx into the Waveform List.
Test waveforms are located at C:\Program
Files\Tektronix\AWG5200\Samples\PV.
12. From the Waveform List window, assign the waveform PV_Square.wfmx
to Channel 1.
13. Click the Home tab and set the AWG’s Channel 1 as follows:
Run Mode to Trig g e re d
Trigger Input to A
Enable the Channel
14. In the Setup -> Trigger tab, set the External Trigger Level to 1.0 V (A and
B). Leave all other settings to their default settings.
40AWG5200 Series Technical Reference
Brief procedures
15. Click the Play b
16. Click the Home tab and verify that the squarewave output is displayed on
the AWG.
17. Verify that the squarewave output is displayed on the test oscilloscope.
utton on-screen or on the front panel of the AWG.
18. Move the cable from the Trigger A input to the Trigger B input.
19. Click the Home tab and set the trigger input to B.
20. Verify that the output is displayed on the test oscilloscope (as in step 17).
21. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
22. Disconnect the test setup.
AWG5200 Series Technical Reference41
Performance tests
Performance t
Prerequisites
ests
This section
listed below.
10 MHz reference frequency acc uracy
Analog amplitude accuracy
Marker high and low level accuracy
The tests in this section provide confirmation of performance and functionality
The following requirements and conditions must be met:
The cabinet must be installed.
The AWG must have been last adjusted at an ambient temperature between
+20 °C and +30 °C, must have been operating for a warm-up period of at
least 20 minutes, and must be operating at an ambient temperatures between
+10 °C and +40 °C.
You must have performed and passed the procedure Diagnostics and
Calibration, and the procedure Functional Tests.
contains performance verification procedures for the specifications
Required equipment
The following table lists the test equipment required to perform the performance
verification procedures. The table identifies examples of recommended equipment
and lists the required precision where applicable. If you substitute other test
equipment for the listed examples, the equipment must meet or exceed the listed
tolerances.
Table 33: Required equipment for performance tests
ItemQty.Minimum requirementsRecommended equipment
Frequency counter1 ea.Frequency accuracy: within ± 0.01 ppm
Digital multimeter1 ea.
Adapter3 ea.
50 Ω SMA cable
50 Ω SMA terminator
50 Ω BNC feed-through
terminator
1 ea.
3 ea.
1 ea.
DC accuracy: within ± 0.01%
TekConnect oscilloscope input to SMA inputTektronix TCA-SMA
DC to 20 GHz
DC to 18 GHz
DC to 1 GHz, feedthrough
Tektronix MCA 3040
Keithley 2000 DMM or
Tektronix DMM4040/4050
Tensolite 1-3636-465-5236
Tektronix part number 136-7162-xx
(supplied with AWG).
Tektronix part number 011-0049-02
42AWG5200 Series Technical Reference
Performance tests
Table 33: Required equipment for performance tests (cont.)
ItemQty.Minimum requirementsRecommended equipment
SMA-BNC adapter
SMA-BNC adapter
BNC-dual banana adapter
3 ea.
1 ea.
1 ea.
SMA female to BNC male connector
SMA male to BNC female connector
BNC to dual banana plugs
Tektronix part number 015-0572-00
Tektronix part number 015-0554-00
Tektronix part number 103-0090-00
Test record
Photocopy the test records and use them to record the performance test results.
(See pag
e64,Test record.)
AWG5200 Series Technical Reference43
Performance tests
Te rmination resistance measurement
Many of the performance tests use a BNC-dual banana adapter and 50 Ω BNC
feed-through terminator connected to a DMM.
For accuracy, the termination resistance of this connection is used in the
calculations.
Use this procedure and note the measured value for use in these procedures.
1. Connect the BNC-dual banana adapter and 50 Ω BNC feed-through terminator
to the HI and LO inputs of the digital multimeter.
Figure 3: Equipment connection to measure terminator resistance
2. Set the digital multimeter to the Ω 2wiresmode.
3. Measure the resistance and note the value as Term_R.
Keep this value available for use in several performance chec k calculations.
4. Set t
NOTE. Lead resistance is not included in the measurement results when using
four wire ohms. The accuracy is higher especially for small resistances. Use a
four wire method if necessary.
he digital multimeter to the DCV mode.
44AWG5200 Series Technical Reference
Analog amplitude accuracy
Required equipmentPrerequisites
Digital multimeter
BNC-dual banana adapter
50 Ω BNC feed-through terminator
SMA female-B NC male adapter
50 Ω SMA terminator
Before starting this procedure, ensure you have the “Term R” value used in the
calculations. (See page 44, Termination resistance measurement.)
Performance tests
AWG preparation and load test waveforms
(See page 42.)
Termination resistance measurement
procedure (Term R) (See page 44.)
DC High BW output path
1. Click the Reset to Default Setup button in the toolbar.
2. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
3. Load the test waveforms PV_DC_Plus.wfmx and PV_DC_Minus.wfmx
into the Waveform List.
NOTE. Test waveforms are located at
ogram Files\Tektronix\AWG5200\Samples\PV.
C:\Pr
4. From the Waveform List window, assign the test waveform
C_Plus.wfmx to Channel 1.
PV_D
5. Connect the CH 1 (+) c onnector from the AWG to the HI and LO inputs of
digital multimeter.
the
Use a 50 Ω SMA cable, a BNC-SMA adapter, a 50 Ω BNC feed-through
rminator, and a BNC dual banana adapter.
te
6. Terminate the CH 1 (–) connector on the AWG with a 50 Ω SMA terminator.
AWG5200 Series Technical Reference45
Performance tests
7. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
8. Click the Setup -> Channel tab and click the Output Settings tab.
a. Select Channel 1.
b. Set the Output Path to DC High BW.
c. Enable the Channel 1 output.
9. Set the Amplitude of the AWG as shown in the following table. (See
Table 34.)
Table 34: Analog amplitude accuracy (DC High BW output path)
Amplitude settingsAccuracy lim its
25 mV
100 mV
p-p
p-p
23.75 mV to 26.25 mV
98 mV to 102 mV
46AWG5200 Series Technical Reference
Performance tests
Table 34: Analog amplitude accuracy (DC High BW output path) (cont.)
Amplitude settingsAccuracy lim its
200 mV
p-p
500 mV
p-p
750 mV
p-p
1V
p-p
(Requires option DC)
1.5 V
p-p
(Requires option DC)
196 mV to 204 mV
490 mV to 510 mV
735 mV to 765 mV
980 mV to 1.02 V
1.47Vto1.53V
10. Press the Play button, or click Play on the display.
11. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_1.
12. Use the following formula to compensate the voltage for the 50 Ω BNC
Where Term_R is the resistance of the 50 Ω BNC feed-through terminator.
(See page 44, Termination resistance measurement.) procedure.
16. Verify that the voltage difference |(V_high – V_low)| falls within the limits
given in the table. (See Table 34 on page 46.)
17. Repeat steps 9 through 16 for each Amplitude setting in the table. (See
Table 34 on page 46.)
18. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
19. Move the SMA cable from the CH 1 (+) connector to the CH 1 (–) connector
and move the 50 Ω SMA terminator from the CH 1 (–) connector to the CH
1(+)connector.
AWG5200 Series Technical Reference47
Performance tests
DC High Volt output path
20. Press the A
on the Home screen) to enable the outputs (front panel light off).
21. Repeat st
22. Repeat steps 4 through 21 until all channels are checked, modifying the
instruc
23. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the H
24. Disconnect the test setup.
NOTE. This is the start of testing the optional DC High Volt output path
(Option HV).
If option HV is not licensed, skip this procedure.
1. Click the Reset to Default Setup button in the toolbar.
2. Load the te
into the Waveform List.
WG front panel All Outputs Off button (or click All Outputs Off
eps 9 through 17 for the CH1 (–) connector.
tions for the channel under test.
ome screen) to disable the outputs (front panel light on).
st waveforms PV_DC_Plus.wfmx and PV_DC_Minus.wfmx
NOTE. Test waveforms are located at
C:\Program Files\Tektronix\AWG5200\Samples\PV.
3. From the Waveform List window, assign the waveform PV_DC_Plus.wfmx
to Channel 1.
4. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
48AWG5200 Series Technical Reference
Performance tests
5. Connect the CH 1
the digital multimeter.
Use a 50 Ω SMA ca
terminator, and a BNC dual banana adapter.
6. Terminat e t
7. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
he CH 1 (–) connector on the AWG with a 50 Ω SMA terminator.
(+) connector from the AWG to the HI and LO inputs of
ble, a BNC-SMA adapter, a 50 Ω BNC feed-through
8. Click the Setup -> Channel tab and click the Output Settings tab.
a. Select Channel 1.
b. Set th
c. Enable the Channel 1 output.
e Output Path to DC High Volt.
AWG5200 Series Technical Reference49
Performance tests
9. Set the Amplitu
de of the AWG as shown in the following table. (See
Table 35.)
Table 35: Analog amplitude accuracy (DC High Volt output path)
Amplitude settingsAccuracy lim its
10 mV
p-p
100 mV
p-p
500 mV
p-p
5 V4.90 V to 5.1
9.5 mV to 10.5
95 mV to 105 mV
490 mV to 510 mV
mV
0V
10. Press the Play button, or click Play on the display.
11. Measure t
he output voltage on the digital multimeter and note the value as
Measured_voltage_1.
12. Use the fo
llowing formula to compensate the voltage for the 50 Ω BNC
feed-through terminator:
V_high =
[(Term_R + 50) / (2 Term_R)] Measured_voltage_1
Where Term_R is the resistance of the 50 Ω BNC feed-through terminator.
(See pa
ge 44, Termination resistance measurement.) procedure.
13. From the Waveform List window, assign the waveform PV_DC_Minus.wfmx
nnel 1.
to Cha
14. Measure the output voltage on the digital multimeter and note the value as
ured_voltage_2.
Meas
15. Use the following formula to compensate the voltage for the 50 Ω BNC
Where Term_R is the resistance of the 50 Ω BNC feed-through terminator.
(See page 44, Termination resistance measurement.) procedure.
16. Verify that the voltage difference |(V_high – V_low)| falls within the limits
given in the table. (See Table 35 on page 50.)
17. Repeat steps 9 through 16 for each amplitude setting in the table. (See
Table35onpage50.)
18. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
19. Move the SMA cable from the CH 1 (+) connector to the CH 1 (–) connector
and move the 50 Ω SMA terminator from the CH 1 (–) connector to the CH
1 (+) connector.
50AWG5200 Series Technical Reference
Performance tests
20. Repeat ste
ps 9 through 17 for the CH1 (–) connector.
21. Repeat steps 3 through 20 until all channels are checked, modifying the
instruct
ions for the channel under test.
22. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Ho
me screen) to disable the outputs (front panel lig ht on).
23. Disconnect the test setup.
Analog offset accuracy (DC output paths)
red equipment
Requi
Digital multimeter
BNC-dual banana adapter
50 Ω BNC feed-through terminator
SMA female-B NC male adapter
50 Ω SMA terminator
Before starting this procedure, ensure you have the “Term R” value used in the
lculations. (See page 44, Termination resistance measurement.)
ca
quisites
Prere
eparation and load test
AWG pr
waveforms(See page 42, Prerequisites.)
Termination resistance measurement
edure
proc
1. Click the Reset to Default Setup button in the toolbar
.
2. Load the test waveform PV_DC_Zero.wfmx into the Waveform List.
NOTE. Test waveforms are located at
C:\Program Files\Tektronix\AWG5200\Samples\PV.
3. From the Waveform List window, assign the waveform PV_DC_Zero.wfmx
to Channel 1.
AWG5200 Series Technical Reference51
Performance tests
4. Press the AWG fr
on the Home screen) to disable the outputs (front panel light on).
5. Connect the CH 1 (+) connector from the AWG to the HI and LO inputs of the
digital multimeter. Use a 50 Ω SMA cable, a BNC-SMA adapter, a 50 Ω BNC
feed-through terminator, and a BNC dual banana adapter.
6. Termin
7. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
ate the CH 1 (–) connector on the AWG using a 50 Ω SMA terminator.
ont panel All Outputs Off button (or click All Outputs Off
8. Click the Setup -> Channel tab and click the Output Settings tab.
a. Select Channel 1.
b. Set the Output Path to DC High BW.
c. Enable the Channel 1 output.
52AWG5200 Series Technical Reference
Performance tests
9. Set the Offset o
f the AWG as shown in the first row of the following table.
(See Table 36.)
Table 36: Offset accuracy
Offset settingsAccuracy limits
2V1.95Vto2.05V
0 V–10 mV to 10 mV
–2 V–2.05 V to –1.95 V
10. Press the Play button, or click Play on the display.
11. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage.
12. Use the following formula to compensate the voltage for the 50 Ω BNC
feed-through terminator:
V = [(Term_R + 50) / (2×Term_R)] × Measured_voltage
Where Term_R is the resistance of the 50 Ω BNC feed-through terminator.
(See page 44, Termination resistance measurement.) procedure.
13. Verify that the voltage V falls within the limits given in the table. (See
Table 36 on page 53.)
14. Repeat steps 9 through 13 for each offset setting in the table. (See Table 36
on page 53.)
15. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel
light on).
16. Move the SMA cable from the CH 1 (+) connector to the CH 1 (–) connector
and move the 50 Ω SMA terminator from the CH 1 (–) connector to the CH
1(+)connector.
17. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
AWG5200 Series Technical Reference53
Performance tests
18. Repeat steps 9 t
hrough 14 for the (–) output.
19. Repeat steps 3 through 18 until all channels are checked, modifying the
instructions for the channel under test.
NOTE. This is the start of testing the optional DC High Volt output path.
If option HV is licensed, continue with this procedure. If not, skip to step 22.
20. Click the Setup -> Channel tab and click the Output Settings tab.
a. Select Channel 1.
b. Set the Output Path to DC High Volt.
c. Enable the Channel 1 output.
21. Repeat steps 9 through 19 for the DC High Volt path.
22. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
23. Disconnect the test setup.
Analog DC Bias accuracy (AC output paths)
Required equipmentPrerequisites
Digital multimeter
BNC-dual banana adapter
SMA female-BNC male adapter
50 Ω SMA terminator
AWG preparation and load test
waveforms(See page 42, Prerequisites.)
Termination resistance measurement
procedure
54AWG5200 Series Technical Reference
Performance tests
Before startin
calculations. (See page 44, Termination resistance measurement.)
1. Click the Reset to Default Setup button in the toolbar
2. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
3. Connect the CH 1 (+) c onnector from the AWG to the HI and LO inputs of
the digital multimeter. Use a 50 Ω SMA cable, a BNC-SMA adapter, and
a BNC dual banana adapter.
4. Terminate the CH 1 (–) connector on the AWG using a 50 Ω SMA terminator.
g this procedure, ensure you have the “Term R” value used in the
.
5. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
6. Click the Setup -> Channel tab and click the Output Settings tab.
a. Select Channel 1.
b. Set the Output Path to AC Direct.
c. Enable the Channel 1 output.
AWG5200 Series Technical Reference55
Performance tests
7. Set the DC Bias of the AWG as shown in the first row of the following table.
(See Table 37.)
Table 37: Analog DC bias accuracy
Bias settingsAccuracy limits
5 V4.88 V to 5.12 V
0 V–20 mV to 20
–5 V–5.12 V to –4.88 V
mV
8. Measure the output voltage on the digital multimeter and note the value as
V_1.
9. Press the front panel All Outputs Off button (or click All Outputs Off on the
Home scr
een) to disable the outputs (front panel light on).
10. Measure the output voltage on the digital multimeter and note the value as
V_2 (DM
M residual voltage).
11. Verify that the voltage difference (V_1 – V_2) falls within the limits given in
ble. (See Table 37 on page 56.)
the ta
12. Repeat steps 7 through 11 for each bias setting in the table. (See Table 37
ge 56.)
on pa
13. Repeat steps 3 through 12 until all channels are checked, modifying the
tructions for the channel number under test.
ins
NOTE. This is the start of testing the optional AC Amplified output path.
option AC is licensed, continue with this procedure. If not, skip to step 17.
If
14. Click the Setup -> Channel tab and click the Output Settings tab.
a. Select Channel 1.
b. Set the Output Path to AC Amplified.
56AWG5200 Series Technical Reference
Performance tests
15. Repeat steps 7 through 13 until all channels are checked, modifying the
instructi
16. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Hom
17. Disconnect the test setup.
ons for the channel number under test.
escreen)todisable the outputs (front panel light on).
AWG5200 Series Technical Reference57
Performance tests
Marker high and low level accuracy
Required equipmentPrerequisites
Digital multimeter
BNC-dual banana adapter
50 Ω BNC cable
50 Ω BNC feed-through terminator
SMA male-BNC female adapter
50 Ω SMA terminator (three)
(See page 42, Prerequisites.)
Termination resistance measurement
procedure
Marker high level accuracy
Before s
tarting this procedure, ensure you have the “Term R” value used in the
calculations. (See page 44, Termination resistance measurement.)
1. Click the Reset to Default Setup button in the toolbar
.
2. Load the test waveform PV_DC_Plus.wfmx into the Waveform List.
NOTE. Test waveforms are located at
C:\Program Files\Tektronix\AWG5200\Samples\PV.
3. From the Waveform List window, assign the waveform PV_DC_Plus.wfmx
to Channel 1.
4. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
5. Connect the AWG’s CH1:1 marker connector to the HI and LO inputs on the
digital multimeter. Use a 50 Ω SMA cable, a BNC-SMA adapter, a 50 Ω BNC
feed-through terminator, and a B NC-Banana adapter.
6. Terminate the AWG’s CH1:2 through CH1:4 markers using 50 Ω SMA
terminators.
58AWG5200 Series Technical Reference
Performance tests
7. In the Setup -> Channel tab, select Output Settings and set the Channel 1
Resolution to 12+4 Mkrs.
8. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
9. Enable the Channel 1 output.
10. In the Setup -> Channel tab, select Output Settings and set the Channel 1
Marker High Level setting as shown in the first row of the following table.
(See Table 38.)
AWG5200 Series Technical Reference59
Performance tests
Table 38: Marke
High level set
+ 1.75 V1.55 V to 1.95 V
0.0 V–25 mV to +25 mV
–0.3 V–0.355 V to –0
r high level accuracy
tings
Accuracy limi
ts
.245 V
11. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_1.
12. Use the following formula to compensate the voltage for the 50 Ω BNC
m_R is the resistance of the 50 Ω BNC feed-thought terminator.
(See page 44, Termination resistance measurement.) procedure.
13. Verify t
hat the Marker_High level falls within the limits given in the table.
(SeeTable38onpage60.)
14. Repeat
steps 10 through 13 for the remaining rows in the table. (See Table 38
on page 60.)
15. Repea
t steps 5 through 13 for each CH1 marker (markers CH1:2 through
CH1:4), always terminating the unused markers for the channel under test.
rker low level accuracy
Ma
16. Repe
at steps 5 through 15 until all channels are checked, modifying the
instructions with the channel number under test.
17. Pres
s the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
NOTE. This is a continuation of the procedure Marker high level accuracy check.
18. Load the test waveform PV_DC_Minus.wfmx into the Waveform List.
NOTE. Test waveforms are located at
C:\Program Files\Tektronix\AWG5200\Samples\PV.
19. From the Waveform List window, assign the waveform PV_DC_Minus.wfmx
to Channel 1.
20. Connect the AWG’s CH1:1 marker connector to the HI and LO inputs on the
digital multimeter. Use a 50 Ω SMA cable, a BNC-SMA adapter, a 50 Ω BNC
feed-through terminator, and a B NC-Banana adapter.
60AWG5200 Series Technical Reference
Performance tests
21. Terminate the A
WG’s CH1:2 through CH1:4 markers using 50 Ω SMA
terminators.
22. Press the front panel All Outputs Off button (or click All Outputs Off on the
Home screen) to enable the outputs (front panel light off).
23. Enable the Channel 1 output.
24. In the Setup -> Channel tab, select Output Settings and set the Channel 1
Marker Low Level setting as shown in the first row of the following table.
(See Table 39.)
Table 39: Marker low level accuracy
Low level settingsAccuracy limits
+1.55V1.37Vto1.73V
0.0 V–25 mV to +25 mV
–0.5 V–0.575 V to –0.425 V
25. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_2.
26. Use the following formula to compensate the voltage for the 50 Ω BNC