Tektronix AWG5002, AWG5004, AWG5012, AWG5014 Field Install Instructions

Instructions
065-0768-00 Hard Disk Drive Assembly Without Software AWG5000 Series Arbitrary Waveform Generators
075-0927-00
www.tektronix.com
*P075092700*
075092700
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered tra demarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-800-833-9200.
H Worldwide, visit www.tektronix.com to find contacts in your area.
Assembly Description
This document supports the hard disk drive assembly, Tektronix part number 065-0768-00.
NOTE. After you replace t his assembly, you will need to reinstall the system and application software.
Products
AWG5012 AWG5014 AWG5002 AWG5004
Minimum Tool and Equipment List
The following software packages are required to install the system and applica­tion software after you replace the hard disk drive assembly:
H AWG5000 & AWG7000 Series Operating System Restore Package,
Tektronix part number 020-2762-XX
H AWG5000 & AWG7000 Product Software Package, Tektronix part number
020-2763-XX
Hard Disk Drive Assembly Instructions
1
Assembly Description
Hard Disk Drive Assembly Instructions
2
Installation Instructions
This section describes how to install the hard disk drive assembly.
Installation Procedures
CAUTION. Before you remove or i nsert the hard disk drive assembly, make sure that the instrument is powered off. Failure to disconnect the power from the instrument may result in damage to the hard disk drive.
Removing the Hard Disk
Drive Assembly
1. Unplug the power cord from the power connector on the rear panel.
2. Position the instrument on the work surface with the rear panel facing you as
shown.
3. Loosen the two thumb screws securing the hard disk drive assembly to the instrument.
Thumb screws
AWG5000 series rear panel
4. Slowly pull the hard disk drive assembly towards you. Note the orientation of the assembly as you remove it.
Hard Disk Drive Assembly Instructions
3
Installation Instructions
Installing the Replacement
Hard Disk Drive Assembly
Installing the Instrument
Software
Adjusting the Touch
Screen
1. Orient the replacement assembly the same way as the original assembly. Install the replacement into the hard disk drive slot.
2. Push the hard disk drive assembly in until the data and power connectors are firmly engaged with the instrument.
3. Tighten the two thumbscrews to secure the hard disk drive assembly to the instrument.
After you have installed the replacement hard disk drive assembly into the instrument, you need to reinstall the Microsoft Windows XP operating system and application software onto the hard disk drive.
Refer to the following documents for information on how to install the software:
H AWG5000 & AWG7000 Series Windows XP Operating System Restore DVD
Instructions (Tektronix part number 071-2084-XX)
H AWG5000 & AWG7000 Series Product Software CD Instructions (Tektronix
part number 071-2085-XX)
1. Double-click the Touch screen configurator icon on the Windows desktop to start the touch screen program.
2. Follow the on-screen instructions to complete the adjustment using the stylus pen that was shipped with the instrument.
g End of document g