completed.
Seq/Wfm Advance - Continuous: Continuously outputs the
waveform/sequence in the Sequence file. The next trigger
advances to the next waveform/sequence. Master/Slave
Operation.
Step - Output quiescent until triggered; then execute the next
waveform/sequence in the Sequence file. When the loop count
reaches its value, output stops and waits for next trigger.
Auto Step - Continuous: Continuously outputs the
waveform/sequence in the Auto Step file. Step: Output
quiescent until triggered.
Master - Provides Point Rate Clock and Trigger to a slave
arbitrary waveform generator for phase synchronous parallel
operation.
Slave - Receives Clock and Trigger from a system clock for
parallel operation.
Main Output
Digital-to-analog Resolution - 12-Bits.
Output Impedance - 50 Ohm.
Amplitude (1 MHz Clock, 000 and FFF Waveform Data,
Norm, No Filter, No Offset, Excluding and ADD Operation)
-
Range: 0.05 V to 10 V
p-p
into 50 Ohm.
Resolution: 1 mV (4 digits).
DC Accuracy: 0.050 V to 0.999 V, ±(0.5% of amplitude +
5 mV); 1.000 V to 10.000 V, ±(1% of amplitude + 50 mV).
Offset -
Range: -5.0 V to +5.0 V into 50 Ohm (-200 mA to +200 mA).
Resolution: 5 mV (4 digits).
Accuracy (1 MHz clock, 7FF waveform data, norm, no filter,
amplitude range 0.05 V): -±(1% of offset + 10 mV).
Pulse Response (20 MHz clock, 000 and FFF waveform data,
norm, no filter, amplitude 5 V, no offset) -
Rise/Fall Time: < 35 ns.
Flatness: Within ±3% after 150 ns from rise/fall edges.
Aberrations: Within ±7%.
Cross Talk Between Channels - 512-point sine, 20 MHz clock,
norm, no filter, amplitude offset. <70 dBc.