Tektronix AM503B Reference manual

The OUTPUT DC LEVEL knob vertically positions the
signal on the oscilloscope screen.
SRQ, and REMOTE indicators give you GPIB status.
(AM5030 only.)
The CURRENT/DIVISION display shows the
current Amplifier scale factor in either
mA/division or A/division. Error codes and
output DC level also appear here.
The 20 MHz BW LIMIT button alternately selects
or deselects the 20 MHz bandwidth limit for
noise filtering.
Hold down the GPIB ADDRESS button while
adjusting the CURRENT/DIVISION buttons to
change the GPIB address. (AM 5030 only.)
The Amplifier output appears at the OUTPUT
connector. Connect this to a 50input of
your oscilloscope.
Pull the release lever to remove the
Amplifier from the power module.
Copyright 1993 by Tektronix, Inc. All rights reserved. Contents of this publication may not be reproduced in any form without the written permission of T ektronix, Inc.
First Printing: December 1993
First Time Operating Procedure
1. Turn off the power module that powers
the Amplifier.
2. Connect a current probe to the Amplifier
3. Turn on the power module.
4. On an oscilloscope, set the coupling to
50 DC, the vertical sensitivity to 10 mV/division, and turn off any band­width filters.
5. On the oscilloscope, adjust the ground
reference so that the trace appears at the center height of the screen.
6. Use a 50 coaxial cable to connect the
Amplifier OUTPUT to the 50 oscillo­scope input. If the oscilloscope has only 1M impedance inputs, install a 50 termination between the oscilloscope and the coaxial cable to the Amplifier.
PROBE OPEN and OVERLOAD indicate probe and Amplifier status.
The PROBE DEGAUSS AUTOBALANCE button removes residual magnetism from the attached current probe. The probe must be removed from the test circuit and locked.
The CURRENT/DIVISION buttons change the Amplifier scale factor.
The COUPLING button selects AC or DC probe coupling or a zero-current reference, as indicated by the lights.
The current probes connect to the Amplifier at the INPUT connector.
7. Unclamp the current probe from around
any conductor and lock the probe closed.
8. Press the COUPLING button until the
REF light comes on. (This step is not necessary for subsequent degauss opera­tions.)
9. Press the PROBE DEGAUSS AUTO-
BALANCE button. Wait until the Amplifier is finished clicking.
10. Press the COUPLING button until either
the AC or DC light (your choice) comes on.
11. Clamp the current probe around a current
carrying conductor. Observe the wave­form on the oscilloscope screen, and adjust it using the Amplifier controls: the OUTPUT DC LEVEL knob and the CURRENT/DIVISION
and but-
tons.
Operating Guidelines
To prevent equipment damage and increase measurement reliability, observe these guidelines:
Never insert the Amplifier into, or
remove it from, a power module unless the power module is turned off.
Never connect a current probe to, or
remove it from, the Amplifier while the power module is on, or while the current probe is clamped around a current-carry­ing conductor.
Always observe the frequency derating
curve, and never exceed the upper fre­quency indicated. Failure to do so may overheat the core of the current probe and damage it.
Always remove power from a bare
conductor before clamping a current probe around it.
Use an oscilloscope input channel hav-
ing an input impedance of 50 . If the oscilloscope has only 1 M impedance inputs, install a 50 termination be­tween the oscilloscope and the coaxial cable to the Amplifier. Do not install the termination at the Amplifier end of the cable.
Always set the oscilloscope vertical
sensitivity at 10 mV/division, and the coupling to DC.
Always move the oscilloscope wave-
form using the Amplifier OUTPUT LEVEL knob. Do not use the oscillo­scope vertical position controls.
Degauss the current probe frequently.
Always remove the current probe from around any conductors while degaus­sing.
Programming Commands (AM5030 only)
ALLEve?
Returns pending event codes.
ERRor?
Returns the most recent error or event code. Functions identically to EVent?
Reference
PROBEOPen?
Returns the status of the probe open indi­cator: ON or OFF.
AMPS <number> AMPS?
Sets or queries AM5030 resolution (am­peres per division).
BWLIMit ON BWLIMit OFF BWLIMit?
Sets or queries the status of 20 MHz band­width limit switch.
COUpling AC COUpling DC COUpling REF COUpling?
Sets or queries the status of input cou­pling.
DCLEVel <number> DCLEVel?
Sets or queries the DC offset level.
DEGAuss DEGAuss 0 DEGAuss 1 DEGAuss 2
Performs probe degauss/autobalance. No argument or 0 is full degauss/autobal­ance, 1 forces gain calibration even if temperature has not changed, 2 performs gain calibration without balancing the Hall device.
EVent?
Returns the most recent error or event code. Functions identically to ERRor?
EXit
Causes the AM5030 to exit test modes or recover from error conditions.
FPLock ON FPLock OFF FPLock?
Enables or disables the front panel con­trols.
HELp?
Returns the list of AM5030 commands.
ID?
Returns the identification string.
INIT
Initializes the AM5030.
OVerload?
Returns the status of the overload indica­tor: ON or OFF.
PATH ON PATH OFF PATH?
Controls inclusion of command names in query return strings.
PROBETRim <number> PROBETRim?
Sets or queries trim adjustment gain fac­tor.
PROBETYpe?
Returns the type of the connected current probe or NOPROBE.
RQS ON RQS OFF RQS?
Enables or disables system requests (SRQs).
SERIAL?
Returns AM5030 serial number .
SET?
Returns command string to return AM5030 to its present state.
TEST
Perform self-test operation and return re­sult.
UNIts?
Returns the units of output: A (amperes) normally or V (volts) in special circum­stances.
AM 503B & AM 5030 Current Probe Amplifier
070-8770-01