Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Additional trademark statements can be added here.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to find contacts in your area.
Table of Contents
General safety summary..........................................................................................iv
C offset................................................................................................. 10
Table of Contents
AFG3000C and AFG3000 Series Specifications and Performance Verificationiii
General safety summary
General safet
To avoid fire or personal
injury
ysummary
Review the fo
this product or any products connected to it.
To avoid pot
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions r
Use proper power cord. Use only the power cord specified for this product and
certified for the country of use.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
llowing safety precautions to avoid injury and prevent damage to
ential hazards, use this product only as specified.
elated to operating the system.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source.
Donotblockthepowercord;itmustremain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual's installation instructions for details
on installing the product so it has proper ventilation.
ivAFG3000C and AFG3000 Series Specifications and Performance Verification
General safety summary
Terms in this manual
Symbols and terms on the
product
These terms may
WAR NI NG . Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER in
the marking.
WARNING
read the marking.
CAUTION indicates a hazard to property including the product.
The following symbols may appear on the product:
appear in this manual:
dicates an injury hazard immediately accessible as you read
indicates an injury hazard not immediately accessible as you
AFG3000C and AFG3000 Series Specifications and Performance Verificationv
General safety summary
viAFG3000C and AFG3000 Series Specifications and Performance Verification
Service safety summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnec
of the power cord or, if provided, the power switch.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
compon
To avoid electric shock, do not touch exposed connections.
tPower. To avoid electric shock, disconnect the mains power by means
ents.
AFG3000C and AFG3000 Series Specifications and Performance Verification1
Service Safety Summary
2AFG3000C and AFG3000 Series Specifications and Performance Verification
Preface
This manual provides instructions to verify the performance of the AFG3000C
and AFG3000 Series Arbitrary Function Generators to the module level.
Unless noted otherwise, the term “AFG3000 Series” refers to the models in the
following table.
Table 1: Sup
AFG3011AFG3021BAFG3011C
AFG3101AFG3022BAFG3021C
AFG3102AFG3022C
AFG3251AFG3051C
AFG3252AFG3052C
ported products
AFG3101C
AFG3102C
AFG3151C
AFG3152C
AFG3251C
AFG3252C
To prevent personal injury or damage to the instrument, consider the following
before attempting service:
The procedures in this manual should be performed only by a qualified service
person.
Read the General Safety Summary and the Service Safety Summary at the
beginning of this document.
When using this manual for servicing, be sure to follow all warnings, cautions,
and notes.
The manual consists of the following sections:
Specifications contains a description of the instrument and the characteristics
that apply to it.
Performance Verification contains procedures for confirming that the
instrument functions properly and meets warranted limits.
The procedures described in this document should be performed every 12
months or after module replacement.
If the instruments do not meet performance criteria, repair is necessary.
AFG3000C and AFG3000 Series Specifications and Performance Verification3
Preface
Finding other
DocumentDescription
AFG3000C and AFG3000 Series
Quick Start User Manual
AFG3000C and AFG3000 Series
Programmer Manual
AFG3000C and AFG3000 Series
Online Help
information
This manual focuses on the performance verification of the instrument. See the
following
A quick reference to major features of the instrument and how they operate. It also provides
several tutorials to familiarize you with basic instrument features.
An encyclopedia of topics that describe the instrument interface and features, and gives
background information on how to use them. It provides Menu Structures, User Interface, and
Programming Information.
A online help system, integrated with the User Interface application that ships with this
product. The help is preinstalled in the instrument.
Manual conventions
This manual uses certain conventions that you should become familiar with.
Some se
instructions clear and consistent, this manual uses the following conventions:
ctions of the manual contain procedures for you to perform. To keep those
list for other documents supporting the instrument on www.tek.com.
Modules
Safety
Front-panel controls and menu names appear in the same case (initial capitals,
all uppercase, and so on) in the manual as is used on the instrument front panel
and menus. Front-panel labels are all upper case letters (for example, MENU,
SELECT, PULSE GEN, and so on).
Instruction steps are numbered unless there is only one step.
roughout this manual, any replaceable component, assembly, or part of the
Th
instrument is referred to generically as a module. In general, a module is an
assembly (such as a circuit board). Sometimes a single component is a module;
for example, the chassis of the instrument is a module.
Symbols and terms related to safety appear in the Safety Summary near the
beginning of this manual.
4AFG3000C and AFG3000 Series Specifications and Performance Verification
Specifications
These specifications apply to all AFG3000C and AFG3000 Series Arbitrary
Function Generators, unless otherwise stated. All specifications are guaranteed
unless label
but are not guaranteed.
ed “typical”. Typical specifications are provided for your convenience
Specificati
indirectly) in the Performance Verification section.
All specifi
These specifications are valid under the following conditions:
The instr
between +20 °C and +30 °C.
The inst
and +50 °C.
The ins
The instrument must be in an environment with temperature, altitude, and
humid
Electrical specifications
e 2: Operating mode
Tabl
CharacteristicDescription
mode
Run
st count
Bur
ternal trigger rate
In
Continuous, Modulation, Sweep, and Burst
1 to 1,000,000 cycles or infinite
000 μs to 500.0 s
1.
ons that are check marked with the
cations apply to the arbitrary function generator unless noted otherwise.
ument must have been calibrated/adjusted at an ambient temperature
rument must be operating at an ambient temperature between 0 °C
trument must have had a warm-up period of at least 20 minutes.
ity within the operating limits described in these specifications.
symbol are checked directly (or
AFG3000C and AFG3000 Series Specifications and Performance Verification5
Specifications
tion
rms
2C / 3011/
3052C
/ 3251C / 3252
Exponential D ecay, and Haversine)
2 to 131,072
250 MS/s
250 MS/s or 1 G S/s (rate changes automatically at ≤16K)
AFG3021B / 3021C / 3022B / 3022C and AFG3101 / 31010C / 3102 / 3102C / 3151C / 3152C: Add 0.5 mV per °C for operation outside the range of 20 °C to
30 °C.
AFG3251 / 3251C / 3252 / 3252C: Add 2.0 mV per °C for operation outside the range of 20 °C to 30 °C.
page
(See page 54,
DC Offset test.)
10AFG3000C and AFG3000 Series Specifications and Performance Verification
Table 9: Internal noise add
CharacteristicDescription
Range
Resolution
0% to 50% of amplitude setting
1%
Table 10: Output characteristics
CharacteristicDescription
Sine wave
Flatness (at 1.0 V
AFG3011 / 3011C
AFG3021B / 3021C /
3022B / 3022C
AFG3051C / 3052C
AFG3101 / 3101C / 3102
/ 3102C
AFG3151C / 3152C
AFG3251 / 3251C / 3252
/ 3252C
Harmonic distortion (at 1.0 V
AFG3011 / 3011C10 Hz ≤ frq < 20 kHz: < –60 dBc
AFG3021B / 3021C /
3022B / 3022C
amplitude (+4 dBm), relative to 1 kHz)
p-p
<5MHz: ±0.15dB
5MHz≤ frq ≤ 10 MHz: ±0.3 dB
<5MHz: ±0.15dB
5MHz≤ frq < 20 MHz: ±0.3 dB
20 MHz ≤ frq ≤ 25 MHz: ±0.5 dB
<5MHz: ±0.15dB
5MHz≤ frq < 45 MHz: ±0.3 dB
45MHz ≤ frq ≤ 50 MHz: ±0.5 dB