Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Additional trademark statements can be added here.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 5 0 0
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worl dwid e, visi t www.tektronix.com to find contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1)
year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its
option, either will repair the defective product without charge for parts and labor, or will provide a replacement
in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty
work may be n
the property of Tektronix.
ew or reconditioned to like new performance. All replaced parts, modules and products become
In order to o
the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible
for packaging and shipping the defective product to the service center designated by Tektronix, with shipping
charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within
the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping
charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
result
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage
or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or
integrated with other products when the effect of such modification or integration increases the time or difficulty
of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
TEK
AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
btain service under this warranty, Customer must notify Tektronix of the defect before the expiration of
ing from attempts by personnel other than Tektronix representatives to install, repair or service the product;
Table of Contents
General safety summary ...................... ................................ ................................ ....iv
Service Safety Summary.............. ................................ .................................. ...........1
C offset..................... ................................ ................................ .............9
Table of Contents
AFG3000 Series Specifications and Performance Verificationiii
General safety summary
General safet
To avoid fire or personal
injury
ysummary
Review the fo
this product or any products connected to it.
To avoid pot
Only qualified personnel should perf orm service procedures.
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions r
Use proper power cord. Use only the power cord specified for this product and
certified for the country of use.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input o r output
terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To a void fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
llowing safety precautions to avoid injury and prevent damage to
ential hazards, use this product only as specified.
elated to operating the system.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source.
Donotblockthepowercord;itmustremain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual's installation instructions for details
on installing the product so it has proper ventilation.
ivAFG3000 Series Specifications and Performance Verification
General safety summary
Terms in this manual
Symbols and terms on the
product
These terms may
WAR NI NG . Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER in
the marking.
WARNING
read the marking.
CAUTIO
The following symbol(s) may appear on the product:
appear in this manual:
dicates an injury hazard immediately accessible as you read
indicates an injury hazard not immediately accessible as you
N indicates a hazard to property including the product.
AFG3000 Series Specifications and Performance Verificationv
General safety summary
viAFG3000 Series Specifications and P erformance Verification
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnec
of the power cord or, if provided, the power switch.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in this product. Disconnec t power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
compon
To avoid electric shock, do not touch exposed connections.
tPower. To avoid electric shock, disconnect the mains power by means
ents.
AFG3000 Series Specifications and Performance Verification1
Service Safety Summary
2AFG3000 Series Specifications and Performance Verification
Preface
This manual provides instructions to verify the performance of the AFG3000
Series Arbitrary Function Generators to the module level.
Unless noted otherwise, the term “AFG3000 Series” refers to the models in the
following table.
Table 1: Sup
AFG3011AFG3021BAFG3011C
AFG3101AFG3022BAFG3021C
AFG3102AFG3022C
AFG3251AFG3051C
AFG3252AFG3052C
To prevent personal injury or damage to the instrument, consider the following
before attempting service:
The procedures in this manual should b e performed only by a qualified service
person.
Read the General Safety Summary and the Service Safety Summary at the
beginning of this document.
When using this manual for servicing, be sure to follow all warnings, cautions,
and notes.
ported products
AFG3101C
AFG3102C
AFG3251C
AFG3252C
The manual consists of the following sections:
Specifications contains a description of the instrument and the characteristics
that apply to it.
Performance Verification contains procedures for confirming that the
instrument functions properly and meets warranted limits.
The procedures described in this document should be performed every 12
months or after module replacement.
If the instruments do not meet performance criteria, repair is nece ssary.
AFG3000 Series Specifications and Performance Verification3
Preface
Finding other
DocumentDescription
AFG3000 Series Quick Start User
Manual
AFG3000 Series Programmer
Manual
AFG3000 Series Online HelpA online help system, integrated with the User Interface application that ships with this
information
This manual focuses on the performance verification of the instrument. See the
following list for other documents supporting the instrument. All documents
except the On
CD-ROM that ship with instrument.
A quick reference to major features of the instrument and how they operate. It also provides
several tutorials to familiarize you with basic instrument features.
An encyclopedia of topics that describe the instrument interface and features, and gives
background information on how to use them. It provides Menu Structures, User Interface, and
Programming Information.
product. The help is preinstalled in the instrument.
line Help are on the AFG3000 Series Product Documentation
Manual conventions
This manual uses certain conventions that you should become familiar with.
Some se
instructions clear and consistent, this manual uses the following conventions:
ctions of the manual contain procedures for you to perform. To keep those
Modules
Safety
-panel controls and menu names appear in the same case (initial capitals,
Front
all uppercase, and so on) in the manual as is used on the instrument front-panel
and menus. Front-panel labels are all upper case letters (for example, MENU,
SELECT, PULSE GEN, and so on).
Instruction steps are numbered unless there is only one step.
roughout this manual, any replaceable component, assembly, or part of the
Th
instrument is referred to generically as a module. In general, a module is an
assembly (such as a circuit board). Sometimes a single component is a module;
for example, the chassis of the instrument is a module.
Symbols and terms related to safety appear in the Safety Summary near the
beginning of this manual.
4AFG3000 Series Specifications and Performance Verification
Specifications
These specifications apply to all AFG3000 Series Arbitrary Function Generators.
All specifications are guaranteed unless labeled “typical”. Typical specifications
are provided
for your convenience but are not guaranteed.
Specifications that are check marked with the
indirectly
All specifications apply to the arbitrary function generator unless noted otherwise.
These spec
) in the Performance Verification section.
ifications are valid under the following conditions:
The instrument must have been calibrated/adjusted at an ambient temperature
between +
The instrument must be operating at an ambient temperature between 0 °C
and +50 °
The instrument must have had a warm-up period of at least 20 minutes.
The instrument must be in an environment with temperature, altitude, and
humidity within the operating limits described in these specifications.
C.
Electrical specifications
Table 2: Operating mode
CharacteristicDescription
Run mode
Burst count
Internal trigger rate1.000 μs to 500.0 s
Continuous, Modulation, Sweep, and Burst
1 to 1,000,000 cycles or infinite
symbol are checked directly (or
20 °C and +30 °C.
AFG3000 Series Specifications and Performance Verification5
AFG3000 Series Specifications and Performance Verification9
Specifications
Table 8: DC offset (cont.)
PV reference
CharacteristicDescription
page
Output impedance50 Ω
1
AFG3011/3011C:Add1.0mVper°Cforoperationoutsidetherangeof20°Cto30°C.
AFG3021B / 3021C / 3022B / 3022C and AFG3101 / 31010C / 3102 / 3102C: Add 0.5 mV per °C for operation outside the range of 20 °C to 30 °C.
AFG3251 / 32
51C / 3252 / 3252C: Add 2.0 mV per °C for operation outside the range of 20 °C to 30 °C.
Table 9: Internal noise add
CharacteristicDescription
Range
Resolution
0% to 50% of amplitude setting
1%
Table 10
: Output characteristics
CharacteristicDescription
Sine wave
Flatness (at 1.0 V
AFG3011 / 3011C
amplitude (+4 dBm), relative to 100 kHz)
p-p
z: ±0.15 dB
<5MH
5MHz≤ frq ≤ 10 MHz: ±0.3 dB
z: ±0.15 dB
AFG3021B / 3021C /
3022B / 3022C
<5MH
5MHz≤ frq < 20 MHz: ±0.3 dB
20 MHz ≤ frq ≤ 25 MHz: ±0.5 dB