Tektronix A6312 100 MHz AC/DC Current Probes User manual

Instructions
A6312 100 MHz AC/DC Current Probe
070-9603-02
.
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
Copyright © Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered tradem arks of Tektronix, Inc.

WARRANTY

Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If a product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and a ny other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representative s to install, re pair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS W ARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER W ARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Service Assurance

If you have not already purchased Service Assurance for this product, you may do so at any time during the product’s warranty period. Service Assurance provides Repair Protection and Calibration Services to meet your needs.
Repair Protection extends priority repair services beyond the product’s warranty period; you may purchase up to three years of Repair Protection.
Calibration Services provide annual calibration of your product, standards compliance and required audit documentati on, recall assurance, and reminder notification of scheduled calibration. Coverage begins upon registration; you may purchase up to five years of Calibration Services.
Service Assurance Advantages
H Priced well below the cost of a single repair or calibration
H Avoid delays for service by eliminating the need for separate purchase authorizations from your company
H Eliminates unexpected service expenses
For Information and Ordering
For more information or to order Service Assurance, contact your Tektronix re p resentative and provide the information below. Service Assurance may not be available in locations outside the United States of America.
Name VISA or Master Card number and expiration Company date or purchase order number Address Repair Protection (1,2, or 3 years) City, State, Postal code Calibration Services (1,2,3,4, or 5 years) Country Instrument model and serial number Phone Instrument purchase date

Table of Contents

General Safety Summary v...................................
Preface vii...................................................
Contacting Tektronix viii.............................................
Getting Started 1............................................
Probe Installation 1................................................
Operating the Current Probe Slide 2...................................
Degaussing and Autobalancing the Current Probe 3......................
Maximum Current Limits 4..........................................
Specifications 7.............................................
Performance Verification 11...................................
Test Procedure Conditions 11.........................................
Equipment Preparation 11............................................
Required Test Equipment 12..........................................
Bandwidth 14......................................................
Rise Time 16......................................................
DC Gain Accuracy 19...............................................
Probe Trim Adjust for Gain Accuracy (AM503B and AM5030 Only) 22.......
DC Offset Adjustment 24......................................
Required Test Equipment 24..........................................
Equipment Connections 25...........................................
AM503B or AM5030 25.............................................
AM503A 27.......................................................
AM503 29........................................................
Maintenance 30..............................................
Cleaning 30.......................................................
Disassembly Instructions 30..........................................
Obtaining Replacement Parts 34.......................................
Preparation for Shipment 34..........................................
Replaceable Parts 35..........................................
Parts Ordering Information 35.........................................
Using the Replaceable Parts List 35....................................
A6312 Instructions
i
Table of Contents

List of Figures

Figure 1: Connecting a current probe to the current
probe amplifier 2........................................
Figure 2: Operating the probe slide 3...........................
Figure 3: Applying the amp-second product rule 5................
Figure 4: A6312 specified operating area 8......................
Figure 5: A6312 frequency derating curve for maximum
input current 8..........................................
Figure 6: A6312 insertion impedance curve 9....................
Figure 7: A6312 Probe jaw dimensions (nominal) 9...............
Figure 8: Bandwidth test setup for the A6312 current probe 14......
Figure 9: Rise time test setup for A6312 17.......................
Figure 10: DC gain accuracy test setup for A6312 20...............
Figure 11: A6312 DC offset adjustment location 24................
Figure 12: Removing the strain relief boot 31.....................
Figure 13: Removing the top half of the probe 32..................
Figure 14: Removing the probe slide 32..........................
Figure 15: Removing the current transformer 33..................
Figure 16: A6312 exploded view 37..............................
ii
A6312 Instructions

List of Tables

Table of Contents
Table 1: Warranted A6312 specifications 7......................
T able 2: Electrical characteristics 7............................
T able 3: Mechanical characteristics 9..........................
Table 4: Environmental characteristics 10.......................
Table 5: Certifications and compliances 10.......................
Table 6: Required test equipment 12............................
Table 7: Equipment settings for bandwidth 14....................
Table 8: Equipment settings for rise time 17.....................
T able 9: Equipment settings for DC gain accuracy 20..............
Table 10: DC gain accuracy test for A6312 21....................
Table 11: DC gain accuracy test work sheet for A6312 23...........
Table 12: Required test equipment 24...........................
Table 13: AM503B/AM 5030 settings for DC offset adjustment 25...
Table 14: AM 503B/AM 5030 error codes requiring
DC offset adjustment 26...................................
Table 15: AM503A settings for DC offset adjustment 27............
Table 16: AM503A error codes requiring DC offset adjustment 28...
Table 17: AM 503 settings for DC offset adjustment 29.............
A6312 Instructions
iii
Table of Contents
iv
A6312 Instructions

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
Only qualified personnel should perform service procedures.
Injury Precautions
Product Damage
Precautions
Avoid Electric Overload. To avoid electric shock or fire hazard, do not apply a voltage to a terminal that is outside the range specified for that terminal.
Avoid Electric Shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is indirectly grounded through the grounding conductor of the mainframe power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Do Not Operate in Wet/Damp Conditions. To avoid electric shock, do not operate this product in wet or damp conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fire hazard, do not operate this product in an explosive atmosphere.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Immerse in Liquids. Clean the probe using only a damp cloth. R efer to cleaning instructions.
A6312 Instructions
v
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
Certifications and
Compliances
CAUTION
Refer to Manual
Breakable.
Do not drop.
WARNING
High Voltage
Use only on an
insulated wire.
Double
Insulated
Protective Ground
(Earth) Terminal
Do not connect to or
remove from an
uninsulated conductor that
is HAZARDOUS LIVE.
Refer to the specifications section for a listing of certifications and compliances that apply to this product.
vi
A6312 Instructions

Preface

This instruction manual supports the operation and maintenance of the A6312 Current Probe with any of the AM503 series current probe amplifiers.
You can find additional documentation supporting the operation and maintenance of the AM 503 series amplifiers in the following manuals:
H AM503 Instruction Manual (070-2052-XX)
H AM503S (AM503A) User Manual (070-8170-XX)
H AM503S (AM503A) Service Manual (070-8174-XX)
H AM503B & AM5030 Instruction Manual (070-8766-XX)
A6312 Instructions
vii
Preface

Contacting Tektronix

Phone 1-800-833-9200*
Address Tektronix, Inc.
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
Web site www.tektronix.com
Sales support 1-800-833-9200, select option 1*
Service support 1-800-833-9200, select option 2*
Technical support Email: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. -- 5:00 p.m. Pacific time
* This phone number is toll free in North America. After office hours, pl ease leave a
voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.
viii
A6312 Instructions

Getting Started

Probe Installation

The A6312 is a DC to 100 MHz current probe designed for use with the AM503 series current probe amplifiers. With the AM503A amplifier, you can measure frequencies up to 80 MHz. With the AM503, AM503B, and AM5030 amplifiers, you can measure frequencies up to 100 MHz. The A6312 current probe can measure currents to 20 A (DC + peak AC), and up to 50 A peak current (while not exceeding the amp-second rating).
This section explains how to install and operate the A6312 current probe.
To connect the current probe to the amplifier input connector, align the tab of the probe connector with the slot in the amplifier input connector as shown in Figure 1(a). Align the dot on the probe connector with the groove opening of the input connector as shown in Figure 1(b). Push the probe connector in while twisting the barrel clockwise to lock the connector.
CAUTION. Handle the current probe with care. Do not drop the probe or subject it to impact, or the core may crack. Do not connect or disconnect the current probe while the probe is clamped around a live conductor or while the amplifier is powered on, or the probe may suffer electrical damage.
A6312 Instructions
1
Getting Started
Amplifier
Push connector in and twist to lock
Current probe connector
Tab
Slot
(a) Align the tab with the connector slot
Amplifier
Figure 1: Connecting a current probe to the current probe amplifier

Operating the Current Probe Slide

The current probe has a slide mechanism that opens and closes the probe jaw. This allows you to clamp the probe around a conductor under test. The slide must be locked to accurately measure current or to degauss the probe. If a probe is unlocked, the PROBE OPEN indicator on the amplifier lights.
WARNING. When the probe slides are open, the exposed ferrite core pieces are not insulated. To avoid injury or equipment damage, remove power from an uninsulated wire before clamping the current probe around it. Also, never disconnect the probe from the amplifier when the probe is connected to a live conductor.
Current probe
connector
Groove
(b) Insert the connector into the amplifier
Alignment dot
2
A6312 Instructions
Getting Started
Figure 2 illustrates the slide operation of the probe. To open the probe, pull the slide back until the jaw is open. To lock the probe, push the slide forward until the detent snaps into place.
Probe open Probe locked
Figure 2: Operating the probe slide

Degaussing and Autobalancing the Current Probe

Degaussing the probe removes any residual magnetization from the probe core. Such residual magnetization can induce measurement error. Autobalancing removes unwanted DC offsets in the amplifier circuitry.
Failure to degauss the probe is a leading cause of measurement errors. To maintain measurement accuracy, degauss your probe in each of these cases:
H After turning on the amplifier and allowing a 20-minute warm-up period
H Before connecting the probe to a conductor, or changing conductors under
test
H Whenever an overload condition occurs
H Whenever the probe is subjected to a strong external magnetic field
H Periodically during normal use
Degauss and autobalance the current probe as follows
1. Verify that the current probe is connected to the amplifier.
2. Remove the current probe from the conductor under test.
A6312 Instructions
3. Lock the probe slide closed (see Figure 2).
4. Press the amplifier PROBE DEGAUSS AUTOBALANCE button.
3
Getting Started
NOTE. The degauss procedure will fail if the amplifier is not properly connected toa50Ω termination impedance.
After you have completed the oscilloscope adjustments and the degauss/autoba­lance procedure, your system is ready to measure current.

Maximum Current Limits

Current probes have three maximum current ratings: continuous, pulsed, and Ampere-second product. Exceeding any of these ratings can saturate the probe core and cause measurement errors. The section titled Specifications on page 7 lists the maximum current ratings of the probe.
H Maximum Continuous Current refers to the maximum current that can be
continuously measured at DC or at a specified AC frequency. The maximum continuous current value is derated with frequency; as the frequency increases, the maximum continuous current rating decreases.
Procedure A
H Maximum Pulsed Current refers to the maximum peak value of pulsed
current the probe can accurately measure, regardless of how short (within bandwidth limitations) the pulse duration.
H Ampere-Second Product defines the maximum width of pulsed current that
you can measure when the pulse amplitude is between the maximum continuous and maximum pulsed current specifications. The maximum continuous specification itself varies by frequency.
NOTE. Always degauss the probe after measuring a current that exceeds the maximum continuous current, maximum pulsed current, or Ampere-second product rating of the probe. Exceeding these ratings can magnetize the probe and cause measurement errors.
To determine if your measurement exceeds the Ampere-second product, perform either Procedure A or Procedure B:
To determine the maximum allowable pulse width, measure the peak current of the pulse (see Figure 3a). Divide the Ampere-second (or Ampere-microsecond) specification of your probe by the measured peak current of the pulse. The quotient is the maximum allowable pulse width; the pulse width at the 50% point of the measured signal must be less than this value.
For example, the A6312 current probe has a maximum Ampere-second product of 100 As. If a pulse measured with an A6312 has a peak current of 40 A, the maximum allowable pulse width would be 100 As divided by 40 A, or 2.5 s.
4
A6312 Instructions
Getting Started
I
max
p
Pulse width at 50%
50%
I
max
c
0A
(a) Maximum allowable pulse width (b) Maximum allowable pulse amplitude
Figure 3: Applying the amp-second product rule
Do not exceed
Maximum pulsed current
Pulse width at 50%
50%
Maximum continuous current
A6312 Instructions
5
Getting Started
Procedure B
To determine the maximum allowable pulse amplitude, measure the pulse width at the 50% points (see Figure 3b). Divide the Ampere-second (or Ampere­microsecond) specification of your probe by the pulse width. The quotient is the maximum allowable current; the peak amplitude of the measured pulse must be less than this value.
For example, the A6312 current probe has a maximum Ampere-second product of 100 As. If a pulse measured with an A6312 probe has a width of 3 s, the maximum allowable peak current would be 100 As divided by 3 s, or 33.3 A.
6
A6312 Instructions

Specifications

This section lists the specifications, characteristics, certifications, and com­pliances for the A6312 current probe.
Warranted specifications, Table 1, are guaranteed performance specifications unless specifically designated as typical or nominal.
Table 1: Warranted A6312 specifications
Installed amplifier
Parameter
AM503
AM503A AM503B/AM5030
Bandwidth DC to 100 MHz, --3 dB DC to 80 MHz, --3 dB DC to 100 MHz, --3 dB
P-P
1
1
4.4 ns 3.5 ns
7%
P-P
7%
P-P
Rise time, 10% to 90% 3.5 ns
Aberrations (typical) 7%
DC gain accuracy 3% 3% 3% (<1.5% typical)
System noise (typical) 250 A
1
You can optimize the pulse response by adjusting R364 (HF COMP) located inside the AM503 amplifier. Refer to the
RMS
3
250 A
RMS
3
250 A
RMS
3
AM503 Instruction Manual (070-2052-XX) for instructions on how to access this adjustment.
2
On the AM503B and AM5030, the DC gain accuracy is correctable to < 0.2% when using the AM503B and AM5030 probe trim procedure described on page 22.
3
The bandwidth of the measurement instrument must be 500 MHz.
Mechanical, electrical, and environmental characteristics for the A6312 current probe are listed in Tables 2 through 4 and Figures 4 through 7.
Table 2: Electrical characteristics
Frequency derating 2 A at 20 MHz
Maximum bare wire working voltage 300 V
Maximum continuous current 20 A (DC + peak AC)
RMS
2
Maximum pulsed current 50A
Ampsecond product 1 ¢ 10-4As (100 As)
Insertion impedance 0.1 at 1 MHz
0.5 at 50 MHz 1 at 100 MHz
A6312 Instructions
7
Specifications
50
Maximum peak pulse (≤50 A)
40
30
20
Amperes (peak)
10
Allowable pulse width (microseconds)
Figure 4: A6312 specified operating area
Amp-second product limit = 100 A s
Maximum continuous input (20 A)
Any width
54321
200
100
50
10
Maximum input current (amperes peak)
1
1kHz
10 kHz
Frequency
Figure 5: A6312 frequency derating curve for maximum input current
100 MHz10 MHz1MHz100 kHz
8
A6312 Instructions
1.0
Specifications
0.5
0.2
0.1
.05
Typical impedance Impedance deviation
Insertion impedance (ohms)
.02
.01
0.1
0.2
Frequency (MHz)
Figure 6: A6312 insertion impedance curve
105210.5
100
5020
Table 3: Mechanical characteristics
Probe dimensions Length: 20 cm (7.77 inches)
Width: 1.6 cm (0.625 inches) Height: 3.2 cm (1.25 inches)
Cable length 2 m (6.6 feet)
Weight 255 g (0.56 lbs)
Maximum Wire Size
11.9 mm (0.470 in)
18.9 mm (0.745 in)
3.8 mm (0.15 in)
Figure 7: A6312 Probe jaw dimensions (nominal)
A6312 Instructions
9
Specifications
c
ala
Und
Labora
ifiedtoStand
ardIE
C
P
arti
cul
a
Table 4: Environmental characteristics
Operating temperature 0° Cto50° C
Storage temperature -- 4 0 ° Cto75° C
Humidity
Nonoperating
Operating
Transportation Qualifies under National Safe Transit Procedure 1A, category II, 36 in. drop
Mechanical shock 500 g. Half sine. Three shocks on three axes of the probe for 1 ms duration. Total of 9 shocks
Vibration 0.025 in. pk--pk displacement. 10 -- 50 Hz in 1min. cycles. Hold 9 min. at any major
Random vibration
Operating
30° Cto60° Cat90to95%RH
30° Cto50° Cat90to95%RH
resonance, or if none, at 55 Hz. Total time, 54 min
0.31 g
Tektronix Std. 062--2858--00, Rev. B, Class 3
, 5 to 500 Hz, 10 minutes on each axis
RMS
Table 5 lists the product certifications and compliances.
Table 5: Certifications and compliances
EC Declaration of Conformity Meets intent of Low Voltage Directive 73/23/EEC for Product Safety. Compliance was
demonstrated to the following specification as listed in the Official Journal of the European Communities:
Low Voltage Directive 73/23/EEC as amended by 93/68/EEC: EN 61010-1 Safety requirements for elect ric al equipment for measurement,
control, and laboratory use
EN 61010-2-032:1994 Particular requirements for hand-held current clamps for
electrical measurement and test
Certifications Underwriters Laboratories certified to Standard UL3111-1 and CSA/CAN C22.2 No. 1010.1 for
Electri
nd ElectronicMeasuringand Testing Equipment.
erwriters
hand-held current clamps for electrical measurement and test.
Overvoltage category Category: Examples of Products in this Category:
CAT III Distribution-level mains, fixed installation
CAT II
CAT I
Pollution degree 2
Do not operate in environments where conductive pollutants may be present.
10
toriescert
Local-level mains, appliances, portable equipment
Signal levels in special equipment or parts of equipment, telecommunications, electronics
1010-2-032,
r requirementsfor
A6312 Instructions

Performance Verification

The Performance Verification tests allow you to demonstrate that the A6312 current probe meets its specified levels of performance.
Tolerances that are specified in these procedures apply to the AM503 series current probe amplifiers and the A6312 current probe and do not include test equipment error.
The recommended calibration interval is one year.

Test Procedure Conditions

These procedures are valid only under the following conditions:
H The system has been calibrated at an ambient temperature of 23_ ± 5_ C.
H The system is operating in an environment whose limits are described in
Table 4 on page 10.
H The system, including probe, has had a warm-up period of at least
20 minutes.

Equipment Preparation

H The probe degauss/autobalance routine has been performed after the
20-minute warm-up period.
Before starting these procedures, read completely through each procedure to ensure proper completion.
Before performing the verification tests, install all plug-in units into the power module and turn the power on. Turn any remaining equipment on and allow the entire system, including the attached probe, to warm up for a minimum of 20 minutes.
A6312 Instructions
11
Performance Verification
NOTE. Before performing any verification procedure, properly degauss the probe. Remove the probe from any current-carrying conductor, lock the probe, and press the amplifier PROBE DEGAUSS AUTOBALANCE button. The degauss/autobalance routine is complete when the indicator light turns off.
The amplifier front panel displays an error code 266 during the degauss/ autobalance routine if the amplifier is not properly terminated into 50 Ω. Verify that the oscilloscope input is 50 and set to DC coupling. Use a 50 feedthrough termination, attached at the oscilloscope input, if necessary.

Required Test Equipment

To perform the acceptance tests in this section, you will need the test equipment listed in Table 6. The test equipment must meet or exceed the specifications listed. The test procedures may need to be modified if the recommended equipment is not used.
Table 6: Required test equipment
Recommended
Qty Item Description
1 Oscilloscope 500 MHz bandwidth TDS520B
1 Current probe
amplifier
1 Leveled sine wave
generator
1 Calibration generator 1 MHz square wave, rise time <1 ns,
1 Digital multimeter 0.25% 31/2digit resolution at 50 mV DM2510/G or
1 Current source 0.3% accuracy, 0 to ±2A Fluke 5700A or
1 Calibration fixture 50 , BNC Connector 015-0601-50
1 Current loop 20 turns 27 AWG coated wire Refer to page 19.
1 Termination 50 Ω, ±0.1%, 0.5 W 011-0129-00
AM503 series Select the amplifier
3 MHz to 100 MHz, ±4% flatness Wavetek 9100 with
1V
into 50
p-p
Equipment
as noted on page 13.
Option 250 or SG503 equivalent.
Wavetek 9100 with Option 250 or PG506A equivalent.
Fluke 850x/884x
Wavetek 9100 with Option 250
12
1 BNC cable 50 , 1.05 m (42 in) long 012-0057-01
1 BNC-to-Dual
Banana Adapter
103-0090-00
A6312 Instructions
Performance Verification
Performance and interchangeability of the A6312 probe depends upon the type of amplifier you select to verify the probe. When selecting one of the AM503 series amplifiers, please note the following:
H If you select the AM503B or AM5030 amplifier, the procedures verify the
A6312 probe to 100 MHz and the probe is then interchangeable between either amplifier.
H If you select the AM503 amplifier, you must verify the probe and AM503 as
a system. The procedures verify the A6312 probe to 100 MHz and the probe is not interchangeable with any other amplifier or system.
H If you select the AM503A amplifier, you must verify the probe and
AM503A as a system. The procedures verify the A6312 probe to 80 MHz and the probe is not interchangeable with any other amplifier or system.
A6312 Instructions
13
Performance Verification

Bandwidth

This procedure tests the bandwidth of the A6312 current probe. In this test you measure a signal at a relatively low frequency and again at the rated bandwidth of the probe. The two measurements are compared to verify that the signal amplitude does not fall below --3 dB at the probe bandwidth. Refer to Figure 8 when making equipment connections.
Amplifier in TM
Test oscilloscope
50 oscilloscope input (or add 50
termination here if oscilloscope has
only high-impedance input).
Series Power
Module
Leveled sine wave generator
Current probe
Output
Calibration
fixture
Equipment Connections
Equipment Settings
Figure 8: Bandwidth test setup for the A6312 current probe
1. Using a 50 BNC cable, connect the amplifier output to a 50 oscillo- scope input. If the input impedance of your oscilloscope is 1 M, connect a 50 feedthrough termination at the oscilloscope input. Do not connect the termination at the amplifier output.
2. Connect the current probe to the amplifier input.
3. Connect the current loop calibration fixture to the output of the leveled sine
wave generator.
Make or verify the equipment settings in Table 7.
Table 7: Equipment settings for bandwidth
Oscilloscope
Vertical input impedance 50 Vertical gain 10 mV/division Time base 200 ns/division Record length 500 Coupling DC Offset 0 V (mid-scale)
14
A6312 Instructions
Table 7: Equipment settings for bandwidth (Cont.)
Oscilloscope
Trigger type Edge Trigger mode Auto Trigger position 50% Acquisition mode Average Number of waveforms to average 8 Measurement type Peak-to-peak
Leveled sine wave generator
Frequency 3MHz Amplitude 3V
Current probe amplifier
Coupling DC BW limit Off (full bandwidth) Current/division 10 mA/division
p-p
Performance Verification
Procedure
1. Do not clamp the current probe around any conductor, but make sure the jaws are locked shut.
2. Press the amplifier PROBE DEGAUSS AUTOBALANCE button. Wait for the degauss/autobalance routine to complete before proceeding. The routine is complete when the indicator light turns off.
3. Clamp the current probe around the calibration fixture.
4. Adjust the signal generator output so that the amplifier output is approxi-
mately 60 mV
, or six graticule divisions on the oscilloscope.
p-p
5. Using the peak-peak measurement capability of the oscilloscope, measure and record the peak-peak reading as M
.
1
6. Set the oscilloscope time base to 5 ns/division. Increase the signal generator frequency to the warranted bandwidth. Refer to Table 1 on page 7.
7. Using the peak-peak measurement capability of the oscilloscope, measure and record the peak-peak reading as M
.
2
A6312 Instructions
15
Performance Verification
8. The probe meets the bandwidth specification if the ratio of the signal amplitude at the warranted bandwidth is at least 70.7% of the signal amplitude at 3 MHz. Using the following calculation, verify probe band­width:
M
2
> 0.707
M
1
NOTE. The impedance of the calibration fixture used in this test changes between 3 MHz and 100 MHz. Typically the impedance changes from 50 at 3 MHz to 59 at 100 MHz. Thus you can substitute the following equation to make this test more accurate:
At 100 MHz,
M
(
1.18
For the AM503A at 80 MHz,
2
)
> 0.707
M
1

Rise Time

M
(
1.12
9. Verify that the bandwidth is greater than the warranted specification. Refer to
This procedure measures the rise time of the A6312 current probe. In this test you directly measure the rise time of a step input. Refer to Figure 9 when making equipment connections.
2
)
> 0.707
M
1
Table 1 on page 7.
16
A6312 Instructions
Performance Verification
Equipment Connections
Test oscilloscope
50 input
Amplifier in TM
Series Power
Module
Current probe
Calibration
generator
Fast rise
output
Calibration
fixture
Figure 9: Rise time test setup for A6312
1. Using a 50 BNC cable, connect the amplifier output to a 50 oscillo- scope input. If the input impedance of your oscilloscope is 1 M, connect a 50 feedthrough termination at the oscilloscope input. Do not connect the termination at the amplifier output.
2. Connect the current probe to the amplifier input.
Equipment Settings
3. Connect the current loop calibration fixture to the calibration generator 50 fast rise output.
NOTE. If your oscilloscope cannot trigger on the pulse, use another BNC cable to connect the trigger output of the pulse generator to the trigger input of the oscilloscope. Configure the oscilloscope for an external trigger.
Make or verify the equipment settings in Table 8:
Table 8: Equipment settings for rise time
Oscilloscope
Vertical input impedance 50 Vertical gain 10 mV/division Time base 2 ns/division Record length 500 Coupling DC Offset 0 V (mid-scale) Trigger type Edge Trigger mode Auto
A6312 Instructions
17
Performance Verification
Table 8: Equipment settings for rise time (Cont.)
Oscilloscope
Trigger position 50% Acquisition mode Average Number of waveforms to average 32 Measurement type Rise time
Calibration generator
Period 1 s Output Fast rise Amplitude Maximum
Current probe amplifier
Coupling DC BW Limit Off (full bandwidth) Current/division 5 mA/division
Procedure
1. Do not clamp the current probe around any conductor, but make sure the jaws are locked shut.
2. Press the amplifier PROBE DEGAUSS AUTOBALANCE button. Wait for the degauss/autobalance routine to complete before proceeding. The routine is complete when the indicator light turns off.
3. Clamp the current probe around the calibration fixture. Verify that the arrow-shaped indicator on the probe points away from the pulse source.
4. Using the measurement capability of the oscilloscope, measure the rise time of the displayed pulse from 10% to 90% amplitude.
5. Calculate the rise time of the probe (t
trprobe = trmeasured2− trsystem
probe) using the formula below:
r
2
The measured rise time (trmeasured) is the value calculated in step 4.
The system rise time (t
system) is the rise time of the displayed signal when
r
output of the pulse generator is connected directly to the oscilloscope input. (The current probe and amplifier are excluded.)
18
A6312 Instructions

DC Gain Accuracy

Performance Verification
6. Verify that the probe rise time is less than the warranted specification. Refer to Table 1 on page 7.
NOTE. When using the A6312 on an AM503, you can optimize the pulse response by adjusting R364 (HF COMP) located inside the AM503 amplifier. Refer to the AM503 Instruction Manual (070-2052-XX) for instructions on how to access this adjustment. This information is in the service section of the manual and the adjustment should be performed by qualified service personnel only.
This procedure tests the DC gain accuracy of the A6312 current probe. In this test you compare the voltage output of the amplifier to a reference input.
Current Loop for DC Gain
Accuracy Check
Equipment Connections
You will need to construct a simple current loop in order to complete the DC gain accuracy tests.
WARNING. Magnetic fields are produced that may cause a malfunction in heart pacemakers, or damage to sensitive equipment.
Construct the loop using a cylindrical form approximately 3 inches in diameter, wind exactly 20 turns of #27 coated wire.
NOTE. Ensure that the current loop has exactly 20 turns. A significant error will result for each turn variance from 20 turns.
1. Using a BNC cable, connect the amplifier output to the 50 feedthrough termination. Attach the termination to a BNC-to-dual banana adapter. Taking care to observe and maintain polarity, insert the dual banana adapter into the digital multimeter DC voltage input.
2. Connect the current loop to the current source as shown in Figure 10.
A6312 Instructions
19
Performance Verification
Amplifier in TM
Series Power
Module
Equipment Settings
DMM
+--
BNC-to-dual
banana adapter
50 termination
50 coaxial
cable
20 turns
Probe
Figure 10: DC gain accuracy test setup for A6312
Make or verify the equipment settings in Table 9:
Table 9: Equipment settings for DC gain accuracy
Digital multimeter
Measurement type DC volts Range Autoranging
Current source
Output Off
Current probe amplifier
Coupling DC BW limit On Current/division 1 mA/division
Output
Current flow
Current
source
+--
20
Procedure
1. Do not clamp the current probe around any conductor, but make sure the jaws are locked shut.
2. Press the amplifier PROBE DEGAUSS AUTOBALANCE button. Wait for the degauss/autobalance routine to complete before proceeding. The routine is complete when the indicator light turns off.
3. Clamp the current probe around the 20 turn current loop, as shown in Figure 10. Verify that the arrow-shaped indicator on the probe points away from the current source.
A6312 Instructions
Performance Verification
4. For each of the amplifier current/division settings in Table 10, perform the following steps:
a. Set the Amplifier to the desired current/division setting from Table 10.
b. Set the current source to the correct positive test current using Table 10.
c. Turn on the current source.
d. Record the exact measurement of the digital multimeter as M
. (You can
1
copy Table 11 on page 23 to record the results of your measurements.)
e. Set the current source for the correct negative test current using Table 10.
Table 10: DC gain accuracy test for A6312
Turns of current loop passing through probe
20 1mA ±250 A 10 mA
20 2mA ±500 A 20 mA
20 5mA ±1.25 mA 50 mA
20 10 mA ±2.5 mA 100 mA
20 20 mA ±5mA 200 mA
20 50 mA ±12.5 mA 500 mA
20 100 mA ±25 mA 1A
20 200 mA ±50 mA 2A
20 500 mA ±125 mA 5A
20 1A ±250 mA 10 A
Current probe amplifier current/division
Current source output
Test current
I
t
A6312 Instructions
20 2A ±500 mA 20 A
20 5A ±1A 40 A
f. Record the digital multimeter reading as M2.
g. Calculate the measured current (I
M
Im=
1–M2
× (currentdivision)
0.01
) using the following formula:
m
21
Performance Verification
For example, you might have obtained values of 50 mV for M 48 mV for M you can compute I
(50.0x10
Im=
. If you are using an amplifier setting of 10 mA/division,
2
as:
m
3
) – (–48x10–3)
0.01
× (10x103) = 98 mA
h. Verify that the measured current (Im) is within ±3% of the test current (It)
by computing %Error as follows:
%Error =
m
t
× 100
I
t
I
I
For example, using a test current Itof 100 mA and a measured current I of 98 mA, the %Error would be:
%Error =
98 – 100
100
× 100 = –2%

Probe Trim Adjust for Gain Accuracy (AM503B and AM5030 Only)

After the PROBE DEGAUSS AUTOBALANCE routine has been run, the probe and amplifier system will meet all published specifications; however, if you want to improve the tolerance of the system gain accuracy, or to intentionally offset the gain accuracy to make up for total system errors, the probe trim adjust routine may be performed.
and
1
m
Probe trim adjust is a multiplicative factor that you can use to adjust the gain of the current amplifier system. You can set this multiplier in increments of 0.001 from 0.750 through 1.250. Probe trim adjust is used for an optional calibration of some current probes. If you are not performing such an adjustment, leave probe trim adjust to the factory-default of unity gain (1.000).
To set probe trim adjust, press and hold the 20MHz BW LIMIT button while pressing and releasing the COUPLING button. Use the
and buttons to adjust the setting that is displayed in the CURRENT/DIVISION display. When finished, press either the 20MHz BW LIMIT or COUPLING button to restore normal operation.
The display shows the last three significant digits of the display adjust setting; the leading 0. or 1. are omitted. If the first digit displayed is 7, 8, or 9, then the leading digit must be 0. If the first digit displayed is 0, 1, or 2, then the leading digit must be 1.
22
A6312 Instructions
Table 11: DC gain accuracy test work sheet for A6312
Performance Verification
Turns of current loop passing through probe
Current probe amplifier current/division
Current source output
Test current
I
t
20 1mA ±250 A 10 mA
20 2mA ±500 A 20 mA
20 5mA ±1.25 mA 50 mA
20 10 mA ±2.5 mA 100 mA
20 20 mA ±5mA 200 mA
20 50 mA ±12.5 mA 500 mA
20 100 mA ±25 mA 1A
20 200 mA ±50 mA 2A
20 500 mA ±125 mA 5A
20 1A ±250 mA 10 A
20 2A ±500 mA 20 A
20 5A ±1A 40 A
1
Im=
2
%Error =
M
– M
1
2
× (currentdivision)
0.01
I
I
m
I
t
t
× 100
M
1
M
2
1
I
m
%Error
2
A6312 Instructions
23

DC Offset Adjustment

The following adjustment procedures describe how to adjust the DC offset of the A6312 using the AM503, AM503A, AM503B or AM5030 amplifiers. Toler­ances that are specified in these procedures apply to the current probes and do not include test equipment error.
Refer to Figure 11 for the location of the DC offset adjustment.
Figure 11: A6312 DC offset adjustment location

Required Test Equipment

To perform the adjustment procedures in this section, you will need the test equipment listed in Table 12. The test equipment must meet or exceed the specifications listed. The test procedure may need to be changed if the recom­mended equipment is not used.
DC offset
24
Table 12: Required test equipment
Recommended
Qty Item Description
1 Oscilloscope 150 MHz bandwidth TDS460
1 Termination
3 BNC cable
1
Provided as a standard accessory with the AM503 series amplifiers.
1
1
50 Ω, BNC connector, feedthrough
50 , 1.05 m (42 in) long 012--0057--01
equipment
011--0049--01
A6312 Instructions

Equipment Connections

1. Connect the amplifier output to a 50 oscilloscope input using a 50 BNC
2. Connect the current probe to the amplifier input connector.
3. Do not clamp the current probe around a conductor, but make sure the jaws

AM503B or AM5030

Use the following procedure to adjust the DC offset of the A6312 current probe when using an AM503B or AM5030 current probe amplifier.
DC Offset Adjustment
cable. If the input impedance of your oscilloscope is 1 MΩ, first connect a 50 feedthrough termination to the oscilloscope input. Do not connect the termination at the amplifier output.
are locked shut.
Equipment Settings
Procedure
Make or verify the equipment settings in Table 13:
Table 13: AM503B/AM 5030 settings for DC offset adjustment
AM503B and AM5030
Coupling DC BW Limit On (20 MHz bandwidth limit) Current/division 10 mA/division
Oscilloscope
Vertical Gain 10 mV/division Time Base Auto triggered
1 ms/division,
Input Coupling Ground
1. Move the oscilloscope trace to the center horizontal graticule line using the vertical position control.
2. Set the oscilloscope input coupling to DC.
3. Do not clamp the current probe around a conductor, but make sure the jaws
are locked shut.
A6312 Instructions
4. Press the amplifier PROBE DEGAUSS AUTOBALANCE button. Wait for the degauss/autobalance routine to complete before proceeding. The routine is complete when the indicator light turns off.
25
DC Offset Adjustment
NOTE. If the amplifier is not properly terminated into 50 Ω, the amplifier front panel displays error code 266 after the degauss/autobalance routine completes. Verify that the oscilloscope input is 50 and set to DC coupling. If necessary, use a 50 Ω termination at the oscilloscope input.
5. If no error codes are displayed after the degauss/autobalance routine completes, no offset adjustment is necessary. If any of the error codes listed in Table 14 are displayed, continue with the procedure:
Table 14: AM 503B/AM 5030 error codes requiring DC offset adjustment
Error code Meaning
580 Unable to complete negative offset adjustment
581 Unable to complete positive offset adjustment
6. Press and hold the 20 MHz BW LIMIT button, and while holding it, press the
PROBE DEGAUSS AUTOBALANCE button. This sets the front panel
display to --00 and puts the AM503B and AM5030 into an internal test mode.
7. Press the CURRENT/DIVISION
button until the front panel display
reads --52.
8. Press and release the 20 MHz BW LIMIT button. The amplifier degausses itself. Wait until the DEGAUSS light goes out.
9. Adjust the DC Offset control so that the oscilloscope trace is on the center graticule line (zero offset), ± 2 divisions.
10. Press and release the 20 MHz BW LIMIT button, then press and release the COUPLING button. This exits the AM503B and AM5030 test mode.
26
A6312 Instructions

AM503A

DC Offset Adjustment
Use the following procedure to adjust the DC offset of the A6312 current probe when using an AM503A current probe amplifier.
Equipment Settings
Procedure
Make or verify the equipment settings in Table 15:
Table 15: AM503A settings for DC offset adjustment
AM503A
Coupling DC BW limit On (20 MHz bandwidth limit) Current/division 10 mA/division
Oscilloscope
Vertical gain 10 mV/division Time base Auto triggered
1 ms/division,
Input coupling Ground
1. Move the oscilloscope trace to the center horizontal graticule line using the vertical position control.
2. Set the oscilloscope input coupling to DC.
3. Do not clamp the current probe around a conductor, but make sure the jaws
are locked shut.
A6312 Instructions
4. Press the AM503A PROBE DEGAUSS AUTOBALANCE button. Wait for the degauss/autobalance routine to complete before proceeding. The routine is complete when the indicator light turns off.
NOTE. If the AM503A is not properly terminated into 50 Ω,
the AM503A front
panel displays error code 54 after the degauss/autobalance routine completes. Verify that the oscilloscope input is 50 and set to DC coupling. If necessary, use a 50 Ω termination at the oscilloscope input.
5. If no error codes are displayed after the degauss/autobalance routine completes, no offset adjustment is necessary. If any of the error codes in Table 16 are displayed, continue with the rest of this procedure.
27
DC Offset Adjustment
Table 16: AM503A error codes requiring DC offset adjustment
Error code Meaning
46 Unable to complete positive coarse offset adjustment
47 Unable to complete negative coarse offset adjustment
48 Unable to complete positive fine offset adjustment
49 Unable to complete negative fine offset adjustment
6. Press and hold the 20 MHz BW LIMIT button, and while holding it, press the
PROBE DEGAUSS AUTOBALANCE button. This sets the front panel
display to 00 and puts the AM503A into an internal test mode.
7. Rotate the CURRENT/DIVISION knob until the front panel display reads 21.
8. Press and release the 20 MHz BW LIMIT button. This sets the AM503A internal offsets to zero.
9. Adjust the A6312 DC Offset control so that the oscilloscope trace is on the center graticule line (zero offset), ± 2 divisions.
10. Press and release the 20MHz BW LIMIT button, then press and release the DC COUPLING button. This exits the AM503A test mode.
28
A6312 Instructions

AM503

DC Offset Adjustment
Use the following procedure to adjust the DC offset of the A6312 current probe when using an AM503 current probe amplifier.
Equipment Settings
Procedure
Make or verify the equipment settings in Table 17:
Table 17: AM 503 settings for DC offset adjustment
AM503
Coupling CAL DC LEVEL BW limit On (5 MHz bandwidth limit) Current/division 10 mA/division
Oscilloscope
Vertical gain 10 mV/division Time base Auto triggered
1 ms/division
Input coupling Ground
1. Move the oscilloscope trace to the center horizontal graticule line using the vertical position control.
2. Set the oscilloscope input coupling to DC.
3. Move the oscilloscope trace to the center horizontal graticule line using the
AM503 DC LEVEL control.
A6312 Instructions
4. Do not clamp the current probe around a conductor, but make sure the jaws are locked shut.
5. Press and release the DEGAUSS button.
6. Set the AM503 input coupling to DC.
7. Set the AM503 BALANCE adjustment to mid--position.
8. Adjust the A6312 DC Offset control so that the oscilloscope trace is on the
center graticule line (zero offset), ± 2 divisions.
9. Press and release the DEGAUSS button.
10. If necessary, readjust the A6312 DC Offset control so that the oscilloscope
trace is on the center graticule line (zero offset), ± 2 divisions.
29

Maintenance

Cleaning

This section explains how to clean the A6312 current probe and, if necessary, disassemble the probe for maintenance or repair. Also included are instructions for preparing the probe for shipment.
WARNING. Probe disassembly should only be performed by qualified service personnel.
To clean the probe body, use a soft cloth dampened in a solution of mild detergent and water. To clean the core, open the jaw and clean the exposed core surfaces with a cotton swap dampened with isopropyl alcohol (isopropanol) or ethyl alcohol (fotocol or ethanol).
Do not lubricate the jaws mating surfaces. Any lubricant between the core pieces should be removed with a recommended solvent.
Do not use chemicals containing benzine, benzene, toluene, xylene, acetone, or similar solvents.
Do not use a petroleum based lubricant on the plastic. If the plastic slide assembly requires lubrication, use a silicone based grease sparingly.
Do not immerse the probe in liquids or use abrasive cleaners.

Disassembly Instructions

WARNING. Probe disassembly should only be performed by qualified service personnel.
1. Remove the two screws from the bottom of the probe and pull the strain relief boot back as shown in Figure 12.
30
A6312 Instructions
Strain relief boot
Figure 12: Removing the strain relief boot
Maintenance
Probe body
Screws
2. Move the probe slide assembly to the open position.
NOTE. The probe slide contains a tiny metal ball. In step 3, be careful not to lose the ball by accidentally letting it fall out.
3. Hold the probe in a top-up horizontal position and slide the top half of the probe body off as shown in Figure 13.
4. Remove the metal ball.
5. Turn the probe upside down, push the slide back slightly, and remove the
slide (see Figure 14).
A6312 Instructions
31
Maintenance
Be careful to keep
this tiny metal ball
from falling out
(b)
(a)
(1) Hold the bottom half of the probe body in one hand
Figure 13: Removing the top half of the probe
(1) Hold the bottom half of the probe body in one hand
(2) Grasp the top half of the probe body here with your other hand, then
(a) pivot the back end up, and
(b)slidethetopforward off the end of the bottom half of the probe body
(a)
(b)
32
(2) Grasp the top half of the probe body here with your other hand, then
(a) push the slide back slightly, and
(b) withdraw the slide from the probe body
Figure 14: Removing the probe slide
A6312 Instructions
Maintenance
6. If you want to replace the current transformer, lift the front edge of the circuit board and transformer out of the probe and then pull the transformer out of the circuit board socket.
7. To remove the circuit board, unsolder the two connections then lift the strain relief and circuit board from the body half. Refer to Figure 15.
Circuit board
Current
transformer
Body half
Figure 15: Removing the current transformer
8. Before reassembling the probe, be sure that the gap between the stationary and moveable core pieces is clean. If necessary, use isopropyl alcohol or a similar cleaning agent to clean the pieces. Also, clean the contacts of the slide switch, if necessary. Should the plastic slide assembly require lubrication, sparingly apply silicone-based grease to the parts.
Probe reassembly is the reverse of steps 1 through 7.
NOTE. Exercise care when fitting the slide back into the probe body; aligning the switch contacts can require patience.
A6312 Instructions
33
Maintenance

Obtaining Replacement Parts

Replacement parts may be obtained through your local Tektronix field office or representative. Refer to the Replaceable Parts List on page 35 for more informa­tion.

Preparation for Shipment

If you must ship your Tektronix product, please use the original packaging if possible. If the original packaging is unfit for use or not available, use the following packaging guidelines:
1. Use a corrugated cardboard shipping carton having inside dimensions at least one inch greater than the probe dimensions. The box should have a carton test strength of at least 200 pounds.
2. Put the probe into a plastic bag or wrap to protect it from dampness.
3. Place the probe into the box and stabilize it with light packing material.
4. Seal the carton with shipping tape.
34
A6312 Instructions

Replaceable Parts

This section contains a list of the components that are replaceable for the Current Probe. As described below, use these lists to identify and order replacement parts.

Parts Ordering Information

Replacement parts are available from or through your local Tektronix, Inc., service center or representative.
Changes to Tektronix instruments are sometimes made to accommodate improved components as they become available and to give you the benefit of the latest circuit improvements. Therefore, when ordering parts, it is important to include the following information in your order:
H Part number
H Instrument type or model number
H Instrument serial number
H Instrument modification number, if applicable
If a part you order has been replaced with a different or improved part, your local Tektronix service center or representative will contact you concerning any change in the part number.

Using the Replaceable Parts List

The tabular information in the Replaceable Parts List is arranged for quick retrieval. Understanding the structure and features of the list will help you find all the information you need for ordering replacement parts.
Item Names
In the Replaceable Parts List, an Item Name is separated from the description by a colon (:). Because of space limitations, an Item Name may sometimes appear as incomplete. For further Item Name identification, U.S. Federal Cataloging Handbook H6--1 can be used where possible.
A6312 Instructions 35
Replaceable Parts
Indentation System
Abbreviations
This parts list is indented to show the relationship between items. The following example is of the indentation system used in the Description column:
12345 Name&Description
Assembly and/or Component Attaching parts for Assembly and/or Component
(END ATTACHING PARTS)
Detail Part of Assembly and/or Component Attaching parts for Detail Part
(END ATTACHING PARTS)
Parts of Detail Part Attaching parts for Parts of Detail Part
(END ATTACHING PARTS)
Attaching parts always appear at the same indentation as the item it mounts, while the detail parts are indented to the right. Indented items are part of, and included with, the next higher indentation. Attaching parts must be purchased separately, unless otherwise specified.
Abbreviations conform to American National Standards Institute (ANSI) standard Y1.1.
36 A6312 Instructions
12
Replaceable Parts
3
1
2
4
11
10
5
9
8
Figure 16: A6312 exploded view
6
7
A6312 Instructions 37
Replaceable Parts
Replaceable parts list
Fig. & Index Number
16-- -- -- -- -- ------ ------ -- 1 PROBE, CURRENT:A6312
--1 204--0288--03 1 BODY HALF,PROBE:UPPER BODY
--2 214--0835--00 1 SPRING,HLCPS:0.127 OD X 2.65 L,SST 91260 ORDER BY
--3 214--0849--00 1 RTNR RETURN SPR:BRS CD PL P6042 80009 214--0849--00
--4 352--0106--00 1 HOLDER,SPR RTNR:DELRIN P6042 TK2565 352--0106--00
--5 650--3496--00 1 CABLE W/BD ASSY:CABLE W/A6312 PROBE
--6 213--0087--00 2 SCREW,TPG,TC:2-- 32 X 0.5,T YPE
--7 334--8976--00 1 MARKER,IDENT:PROBE IDENT LABEL,A6312, 80009 334-- 8976-- 00
--8 204--0714--06 1 BODY,HALF:LOWER BODY HALF
--9 120--1988--00 1 XFMR SUBASSY:UPPER & LOWER TRANSFORMER
--10 214--0854--00 1 CONTACT,ELEC:UPPER SHELF,CU BE TK1947 214--0854--00
--11 351--0121--01 1 CONT ASSY,ELEC:PROBE SLIDE ASSY,A6302, 80009 351-- 0121-- 01
--12 214--0997--00 1 BALL,BEARING:0.094,SST 05469 ORDER BY
Tektronix Part Number
Serial No. Effective
Serial No. Discont’d
Qty Name & Description Mfr. Code Mfr. Part Number
HALF,BLACK,POLY,A6302,
CIRCUIT BOARD,A6312
BT,PANHEAD,STEEL,CADIUM PLATED,POZIDRIVE
W/CONTACTS,A6302,
SUBASSY,A6312
80009 204-- 0288-- 03
DESCRIPTION
80009 650-- 3496-- 00
3M099 ORDER BY
DESCRIPTION
80009 204-- 0714-- 06
80009 120-- 1988-- 00
DESCRIPTION
Standard Accessories
020--0167--01 1 ACCESSORY PKG:ACCESSORY PACKAGE,P6021 80009 020--0167--01
070--9603--XX 1 MANUAL,TECH:INSTRUCTIONS,A6312,DP 80009 070--9603-- XX
38 A6312 Instructions
Manufacturers cross index
Mfr. Code
05469 BEARINGS INC 3634 EUCLID
3M099 PORTLAND SCREW COMPANY 6520 N BASIN AVE PORTLAND, OR 97217
80009 TEKTRONIX INC 14150 SW KARL BRAUN DR
91260 CONNOR FORMED METAL PRODUCTS 1729 JUNCTION AVENUE SAN JOSE, CA 95112
TK1947 NORTHWEST ETCH TECHNOLOGY 2601 S HOOD ST
TK2565 VISION PLASTICS INC 26000 SW PARKWAY CENTER DRIVE WILSONVILLE, OR 97070
Manufacturer Address City, State, Zip Code
PO BOX 6925
PO BOX 500
PO BOX 110610
CLEVELAND, OH 44101
BEAVERTON, OR 97077--0001
TACOMA, WA 98411--0610
Replaceable Parts
A6312 Instructions 39
Replaceable Parts
40 A6312 Instructions
Loading...