Tektronix A6304XL 500 Amp AC/DC Current Probe User manual

Instructions
A6304XL 500 Amp AC/DC Current Probe
070-8939-03
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Copyright © Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070--1000
TEKTRONIX and TEK are registered tradem arks of Tektronix, Inc.
WARRANTY
Tektronix warrants that the products that it m anufactures and sells will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If a product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair dam age resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS W ARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Contacting Tektronix
Product Support
Service Support
For other informa­tion
To write us
For questions about using Tektronix measurement products, call toll free in North America: 1-800-TEK-WIDE (1-800-835-9433 ext. 2400) 6:00 a.m. -- 5:00 p.m. Pacific time
Or contact us by e-mail: tm_app_supp@tek.com
For product support outside of North America, contact your local Tektronix distributor or sales office.
Tektronix offers extended warranty and calibration programs as options on many products. Contact your local Tektronix distributor or sales office.
For a listing of worldwide service centers, visit our web site.
In North America: 1-800-TEK-WIDE (1-800-835-9433) An operator will direct your call.
Tektronix, Inc. P.O. Box 1000 Wilsonville, OR 97070-1000 USA
Website
Tektronix.com
Table of Contents
General Safety Summary ii...................................
Service Safety Summary iv....................................
Getting Started 1............................................
Probe Installation 1................................................
Operating the Current Probe Slide 2...................................
Degaussing and Autobalancing the Current Probe 3......................
Why Degauss the Current Probe? 3...................................
Maximum Current Limits 4..........................................
Specifications 6.............................................
Maintenance 11..............................................
Cleaning 11.......................................................
Disassembly Instructions 11..........................................
Obtaining Replacement Parts 13.......................................
Preparation for Shipment 14..........................................
Replaceable Parts List 15......................................
Parts Ordering Information 15.........................................
Using the Replaceable Parts List 15....................................
A6304XL Instructions
i
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
ToAvoidFireor
Personal Injury
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is indirectly grounded through the grounding conductor of the mainframe power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and marking on the product. Consult the product manual for further ratings information before making connections to the product.
The common terminal is at ground potential. Do not connect the common terminal to elevated voltages.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
ii
A6304XL Instructions
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
Breakable.
Do not drop.
WARNING
High Voltage
Double
Insulated
Protective Ground
(Earth) Terminal
Do not connect to or
remove from an
uninsulated conductor that
is HAZARDOUS LIVE.
A6304XL Instructions
iii
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
iv
A6304XL Instructions
Getting Started
This section covers the installation and operation of the A6304XL with the AM 503B and AM 5030 amplifiers. The probe does not need to be matched to a dedicated amplifier for operation or calibration.
The A6304XL is a DC to 2 MHz current probe designed for use with the AM 503B and AM 5030 current probe amplifiers. The A6304XL can measure currents to 500 A (DC + peak AC), and up to 700 A peak current (while not exceeding the amp-second rating).
The A6304XL can only be used with the AM 503B and AM 5030 and will not be recognized by other current probe amplifiers.
NOTE. If the A6304XL is not recognized by your AM503B or AM5030, contact your nearest Tektronix Service Center concerning an amplifier firmware upgrade.
Please refer to the amplifier documentation for verification procedures, adjust­ments, and additional specifications.
Probe Installation
If you need assistance using your probe, please call our Customer Support Center at 1-800-TEK-WIDE (1-800-835-9433). If you are outside the United States or Canada, please contact your nearest Tektronix Service Center.
To connect a current probe to the Amplifier input connector, align the tab of the probe connector with the slot in the Amplifier INPUT connector as shown in Figure 1(a). Align the dot on the probe connector with the groove opening of the input connector as shown in Figure 1(b). Push the probe connector in while twisting the barrel clockwise to lock the connector.
CAUTION. Handle current probes with care. Do not drop a probe or subject it to impact, or the core may crack. Do not connect or disconnect a current probe while the probe is clamped around a live conductor or while the Amplifier is powered on, or the probe may suffer electrical damage.
A6304XL Instructions
1
Getting Started
Amplifier
Push connector in and twist to lock
Current Probe Connector
Tab
Slot
(a) Align the Tab With the Connector Slot
Figure 1: Connecting a Current Probe to the Amplifier
Operating the Current Probe Slide
Each current probe has a slide that opens and closes the probe jaw. This allows you to clamp the probe around a conductor under test. The slide must be locked to accurately measure current or to degauss the probe. If a probe is unlocked, the PROBE OPEN indicator on the Amplifier will light.
WARNING. When the probe slides are open, the exposed ferrite core pieces are not insulated.
To avoid injury or equipment damage, remove power from an uninsulated wire before clamping the current probe around it.
Current Probe
Connector
Amplifier
(b) Insert the Connector Into the Amplifier
Groove
Alignment Dot
Never disconnect the probe from the Amplifier when the probe is connected to a live conductor.
Allow your hands to contact the probe handle and lock only.
Figure 2 illustrates the slide operation of the probe. To open the probe, press the bottom of the lock button and squeeze the handle until the core is open. To lock the probe, release the squeeze handle and press the top of the lock button.
2
A6304XL Instructions
(1) Unlock the probe
(2) Squeeze
the handle
(a) Opening the Probe
Getting Started
(2) Lock the probe
(1) Release the handle
(b) Closing and Locking the Probe
Figure 2: Operating the Probe Slide
Degaussing and Autobalancing the Current Probe
1. Verify that the current probe is connected to the Amplifier.
2. Remove the current probe from the conductor under test.
3. Lock the probe slide closed as shown in Figure 2(b).
4. Press the Amplifier PROBE DEGAUSS AUTOBALANCE button.
NOTE. The degauss procedure will fail if the Amplifier is not properly connected toa50Ω termination impedance.
After you have completed the oscilloscope adjustments and the Amplifier degauss/autobalance procedure, your system is ready to measure current.
Why Degauss the Current Probe?
Degaussing the probe removes any residual magnetization from the probe core. Such residual magnetization can induce measurement error. Autobalancing removes unwanted DC offsets in the amplifier circuitry.
A6304XL Instructions
3
Getting Started
Failure to degauss the probe is a leading cause of measurement errors. To maintain measurement accuracy, degauss your probe in each of these cases:
H After turning on the Amplifier and allowing a 20-minute warm-up period.
H Before connecting the probe to a conductor, or changing conductors under
H Whenever an overload condition occurs.
H Whenever the probe is subjected to a strong external magnetic field.
H Periodically during normal use.
Maximum Current Limits
Current probes have three maximum current ratings: continuous, pulsed, and Ampere-second product. Exceeding any of these ratings can cause measurement errors or probe heating. See Specifications on page 6 for probe current ratings.
test.
H Maximum Continuous Current refers to the maximum current that can be
continuously measured at DC or at a specified AC frequency. The maximum continuous current value is derated with frequency; as the frequency increases, the maximum continuous current rating decreases. Higher frequency currents may be measured at limited duty cycles. The duty cycle is the ratio of the maximum intermittent current to the continuous current.
H Maximum Pulsed Current refers to the maximum peak value of pulsed
current the probe can accurately measure, regardless of how short (within bandwidth limitations) the pulse duration is.
H Ampere-Second Product defines the maximum width of pulsed current that
you can measure when the pulse amplitude is between the maximum continuous and maximum pulsed current specifications. The maximum continuous specification itself varies with frequency.
NOTE. Always degauss the probe after measuring a current that exceeds the maximum continuous current, maximum pulsed current, or Ampere-second product rating of the probe. Exceeding these ratings can magnetize the probe and cause measurement errors.
To determine if your measurement exceeds the rating of the probe, determine the maximum allowable pulse width and pulse amplitude.
4
A6304XL Instructions
Getting Started
Maximum Allowable
Pulse Width
Maximum Allowable
Pulse Amplitude
To determine the maximum allowable pulse width, refer to figure 3 on page 7 in the Specifications section. Pulse width is measured at 50% of the peak pulse amplitude.
To determine the maximum allowable peak pulse amplitude, refer to figure 3 on page 7 in the Specifications section.
A6304XL Instructions
5
Specifications
Mechanical, electrical, and environmental characteristics unique to the probe are listed in tables 1 through 3. Please refer to the amplifier documentation for additional specifications.
Table 1: Electrical Characteristics
Bandwidth (--3 dB) DC to 2MHz
Rise Time (10% to 90%) 175 ns
Maximum Continuous Current 500 Amps (DC + peak AC)
Maximum Pulsed Current 700 Amps (DC + peak AC)
(See Figure 3)
Pulse AmpSecond Product 0.4 As (400,000 As)
DC Accuracy ±3%
Frequency Derating SeeFigure4
Maximum Bare Wire Working Voltage 600 V
300 V
Insertion Impedance SeeFigure5
Typical Aberrations <±10%
RMS RMS
,CATII , CAT III
6
A6304XL Instructions
2,000 A
1000 A
Specifications
500 A
100 A
10 s
Bipolar
Unipolar
100 s 1 ms 10 ms
(Measured at 50% of pulse peak)
Figure 3: Typical Maximum Pulsed Current
1000 A
Pulse Width
A6304XL Instructions
100 A
Continuous
Intermittent
10 A
10 Hz 100 Hz 1 kHz 10 kHzDC 100 kHz 1 MHz
Figure 4: Typical Continuous Current vs. Frequency Derating
7
Specifications
100 m
10 m
|Z|
100 Ω
1
1m
10 Ω
1 Ω
10 Hz 100 Hz 1 kHz 10 kHz1 Hz 100 kHz 1 MHz
Figure 5: Typical Insertion Impedance Magnitude
8
A6304XL Instructions
Specifications
Table 2:
Probe Weight
Mechanical Characteristics
3.2kg(7lbs)
Cable Length 8m(26ft)
Jaw Dimensions SeeFigure6
21 mm
(0.83 in)
25 mm (1.0 in)
67 mm (2.7 in)
41 mm (1.6 in)
Maximum Wire Size
21 mm (0.83 in)
Figure 6: A6304XL Probe Jaw Dimensions (Nominal)
Table 3: Environmental Characteristics
Operating Temperature 0° Cto50° C(32° F to 122° F)
Storage Temperature -- 4 0 ° Cto75° C(--40° F to 167° F)
Humidity
Operating Storage
30° Cto50° Cat90to95%RH 30° Cto60° Cat90to95%RH
Altitude
Operating Storage
2,000 m (6,416 ft) 15,380 m (50,000 ft)
Transportation The packaged product qualifies under the
Distribution Cycle 1, Assurance Level II for packaged products 0 -- 20 lbs. Test 2 f or Warehouse and Vehicle Stacking (Compression) is omitted.
Tektronix Design Standard 062-2858-00, Rev. B, Section 17.
Mechanical Shock 500 g. Half sine. (200 g in axis towards handle
and nose) two axes. Three shocks on each axes for 1 ms duration. Total of 18 shocks.
Random Vibration Class 3: 3 Axes, 60 Min. Total, 20 Min/Axis
(10 Min. Operating, 10 Min. Nonoperating)
A6304XL Instructions
9
Specifications
Table 4: Certifications and Compliances
EC Declaration of Conformity -­Low Voltage
Approvals UL3111-1 -- Standard for electrical measuring and test equipment
Installation Category Descriptions Terminals on this product may have different installation category designations. The installation
Compliance was demonstrated to the fol lowing specification as listed in the Official Journal of the European Communities:
Low Voltage Directive 73/23/EEC
EN 61010-1:1993 Safety requirements for elect ric al equipment for measurement,
control, and laboratory use
EN 61010-2-032:1995 Particular requirements for hand-held current clamps for electrical
measurements and test
CAN/CSA C22.2 No. 1010.1 -- Safety requirements for electrical equipment for measurement, control and laboratory use
IEC1010-2-032 -- Safety requirements for electrical equipm ent f or measurement, control, and laboratory use
categories are:
CAT III Distribution-level mains (usually permanently connected). Equipment at this l evel is
typically in a fixed industrial location
CAT II Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected
CAT I Secondary (signal level) or battery operated circuits of electronic equipment
10
A6304XL Instructions
Maintenance
Cleaning
This section explains how to clean the A6304XL current probe and, if necessary, disassemble a probe for maintenance or repair. Also included are instructions for preparing a probe for shipment.
The A6304XL current probe should require only routine cleaning. Performance verification and probe adjustment procedures may be found in the AM 503B & AM 5030 Instruction Manual.
WARNING. Probe disassembly should only be performed by qualified service personnel.
To clean the probe body, use a soft cloth dampened in a solution of mild detergent and water. To clean the core, open the jaw and clean the exposed core surfaces with a cotton swap dampened with isopropyl alcohol (isopropanol) or ethyl alcohol (fotocol or ethanol).
Do not lubricate the mating surfaces of the jaws. Any lubricant between the core pieces should be removed with a recommended solvent.
Do not use chemicals containing benzine, benzene, toluene, xylene, acetone, or similar solvents.
Do not use a petroleum based lubricant on the plastic. If the plastic slide assembly requires lubrication, use a silicone based grease sparingly.
Do not immerse the probe in liquids or use abrasive cleaners.
Disassembly Instructions
The following procedures explain how to disassemble the probe body and replace the current transformer.
WARNING. Probe disassembly should only be performed by qualified service personnel.
1. Unlock the probe slide.
A6304XL Instructions
11
Maintenance
2. Place the probe on a flat surface with the screw heads facing up.
3
3. Using a
/32inch Allen wrench, remove the eight retaining screws from the
case.
4. While holding the squeeze handle and slide, lift the top part of the probe body off.
5. Remove the slide.
WARNING. There is a spring in the squeeze handle that can pop out and cause personal injury unless care is exercised when disassembling the probe.
6. Lift the cable out of the probe handle and insert a pair of pliers in the handle as shown in Figure 7.
Torsion Spring
Body Half
Squeeze
Handle
Figure 7: Removing the Handle
7. While maintaining a firm grip on the pliers, carefully lift the squeeze handle and gear out of the probe. See Figures 7 and 9.
8. To remove the current transformer, lift the assembly out of the probe as shown in Figure 8 and unplug it from the circuit board.
12
A6304XL Instructions
Current Transformer
Circuit Board/Cable Assembly
Maintenance
Figure 8: Removing the Current Transformer
9. To remove the circuit board, desolder the two leads to the slide switch and lift the circuit board and cable clear .
NOTE. Observe the routing of the switch leads through the circuit board shield. It is important that the leads be routed in the same way when the board is replaced.
10. Before reassembling the probe, be sure that the gap between the stationary and moveable core pieces is clean. If necessary, clean the core pieces using a lint-free cloth and water or isopropyl alcohol (IPA). Also, clean the contacts of the slide switch, if necessary.
11. Should the plastic slide assembly require lubrication, sparingly apply silicone-based grease to the parts.
12. Probe reassembly is the reverse of steps 1 through 9. You may need to squeeze the handle slightly to align the gear teeth with the slide rack.
Obtaining Replacement Parts
A6304XL Instructions
Replacement parts may be obtained through your local Tektronix field office or representative. Refer to the Replaceable Parts List on page 15 for more information.
13
Maintenance
Preparation for Shipment
If you must ship your Tektronix product, please use the original packaging if possible. If the original packaging is unfit for use or not available, use the following packaging guidelines:
1. Use a corrugated cardboard shipping carton having inside dimensions at least
2. Put the probe into a plastic bag or wrap to protect it from dampness.
3. Place the probe into the box and stabilize it with light packing material.
4. Seal the carton with shipping tape.
one inch greater than the probe dimensions. The box should have a carton test strength of at least 200 pounds.
14
A6304XL Instructions
Replaceable Parts List
This section contains a list of the components that are replaceable for the Current Probe. As described below, use these lists to identify and order replacement parts.
Parts Ordering Information
Replacement parts are available from or through your local Tektronix, Inc., service center or representative.
Changes to Tektronix instruments are sometimes made to accommodate improved components as they become available and to give you the benefit of the latest circuit improvements. Therefore, when ordering parts, it is important to include the following information in your order:
H Part number
H Instrument type or model number
H Instrument serial number
H Instrument modification number, if applicable
If a part you order has been replaced with a different or improved part, your local Tektronix service center or representative will contact you concerning any change in the part number.
Using the Replaceable Parts List
The tabular information in the Replaceable Parts List is arranged for quick retrieval. Understanding the structure and features of the list will help you find all the information you need for ordering replacement parts.
Item Names
Abbreviations
In the Replaceable Parts List, an Item Name is separated from the description by a colon (:). Because of space limitations, an Item Name may sometimes appear as incomplete. For further Item Name identification, U.S. Federal Cataloging Handbook H6--1 can be used where possible.
Abbreviations conform to American National Standards Institute (ANSI) standard Y1.1.
A6304XL Instructions 15
Replaceable Parts List
1
12
13
3
2
11
A1
10
4
5
6
7
8
9
Figure 9: A6304XL Exploded View
16 A6304XL Instructions
Replaceable Parts List
Replaceable Parts List
Fig. & Index Number
9-- 1 A6304XL,500A CNT PROBE:DC CURRENT PROBE
--1 204--0713--01 1 BODY HALF,PROBE:LEFT W/CONTACTS P6303 TK1163 ORDER BY
--2 351--0955--00 1 SLIDE ASSEMBLY:WITH WARNING LABEL 80009 351--0955--00
--3 334--8759--01 1 MARKER,IDENT:PROBE ID LABEL,A6304XL 0KB05 334--8759--01
--4 214--2422--00 1 SPRING,FLAT:UPPER CAN TK1326 214--2422--00
--5 120--1962--00 B011499 1 TRANSFORMER:SUBASSEMBLY, TOP AND
--6 204--0712--01 1 BODY HALF,PROBE:RIGHT W/CONTACTS P6303 TK1163 ORDER BY
--7 211--0093--00 6 SCREW,CAP:4--40 X 0.75,SCH,STL,CD PL,HEX REC 0KB01 ORDER BY
--8 348--0023--00 4 PLUG,HOLE:U/W0.14 DIA HOLE,WHT PLSTC 02768 207-- 090201-- 00-- 0101
--9 211--0183--00 2 SCREW,CAP:4--40 X 0.5,SCH,STL,CD PL,HEX REC 93907 ORDER BY
--10 174--3223--01 B011499 1 CABLE ASSEMBLY:CABLE ASSY W/CIRCUIT
--11 214--2446--00 1 SPR,HLCL,TRSN:0.1 OD X 0.5 L,MUSIC WIRE 8X345 ORDER BY
--12 367--0218--00 1 HANDLE,SQUEEZE:PROBE P6301 TK2565 367-- 0218-- 00
--13 401--0352--00 1 GR CLUSTER,SPUR:(2)18 AND (1) 24 T,PLASTIC TK2565 401--0352--00
Tektronix Part Number
120--1962--01 B011500 1 TRANSFORMER:SUBASSEMBLY, TOP AND
174--3223--02 B011500 1 CABLE ASSEMBLY:CABLE ASSY W/CIRCUIT
070--8939--03 1 MANUAL,TECH:INSTRUCTIONS,A6304XL,DP TK2548 070--8939--02
016--0622--08 1 CASE,CRYG,PROBE:15.5 X 11.0 X 4.5.PLASTIC
Serial No. Effective
Serial No. Discont’d
Qty Name & Description Mfr. Code Mfr. Part Number
DESCRIPTION
80009 120-- 1962-- 00
BOTTOM, A6304XL
80009 120-- 1962-- 01
BOTTOM, A6304XL
DESCRIPTION
DESCRIPTION
DESCRIPTION
80009 174-- 3223-- 01
BOARD,A6304XL
80009 174-- 3223-- 02
BOARD,A6304XL
DESCRIPTION
Standard Accessories
80009 016-- 0622-- 08
W/A6304XL LABEL
A6304XL Instructions 17
Replaceable Parts List
Manufacturers Cross Index
Mfr. Code
02768 ILLINOIS TOOL WORKS INC 195 ALGONQUIN RD DES PLAINES, IL 60016--6103
0KB01 STAUFFER SUPPLY CO 810 SE SHERMAN PORTLAND, OR 97214--4657
0KB05 NORTH STAR NAMEPLATE INC 5750 NE MOORE COURT HILLSBORO, OR 97124-- 6474
80009 TEKTRONIX INC 14150 SW KARL BRAUN DR
8X345 NORTHWEST SPRING MFG CO 5858 SW WILLOW LANE LAKE OSWEGO, OR 97035
93907 CAMCAR DIV OF TEXTRON INC ATTN: ALICIA SANFORD
TK1163 POLYCAST INC 9898 SW TIGARD ST TIGARD, OR 97223
TK1326 NORTHWEST FOURSLIDE INC 18500 SW TETON AVENUE TUALATIN, OR 97062
TK2548 XEROX CORPORATION DIV OF XEROX CORPORATION
TK2565 VISION PLASTICS INC 26000 SW PARKWAY CENTER DRIVE WILSONVILLE, OR 97070
Manufacturer Address City, State, Zip Code
PO BOX 500
516 18TH AVE
14181 SW MILLIKAN WAY
BEAVERTON, OR 97077--0001
ROCKFORD, IL 611045181
BEAVERTON, OR 97005
18 A6304XL Instructions
Loading...