Tektronix A6303 Data Sheet

Page 1
DC to 100 MHz Bandwidth
SPECIFICATIONS
Simultaneous AC/DC Measurements
High Sensitivity Split Core Construction Enables
Easy Circuit Connection 20 Amps DC + Peak AC Current 50 Amps Peak Pulse Current
–6
100 x 10
Amp *Sec Product
1
A6312 DC-coupled Current Probe
UL, CSA, CE Marking
he A6312 is a 100 MHz
T
DC-coupled curre n t p robe for use with the A M503, AM 503A, AM5 0 3 B , and AM5030 current pro b e amplifiers. The c o m b i n a t i o n of an A6312, an AM 503B, and a TM502A provides a flexible curre n t - m e a s u re m e n t tool that can display curre n t s up to 20 Amps DC + peak AC and 50 Amps peak pulse on any oscilloscope. It has a split core construction that p e rmits easy connection to the circuit under test. It may be ord e red either separately as the A6312 or as Option 05 to the AM 503S.
STANDARD SPECIFICATIONS
for A6312/AM 503B/TM502A
Bandwidth
DC to 100 MHz
Maximum DC + Peak AC Current
20 Amps
Maximum Peak Pulse Current
50 Amps
Minimum Sensitivity
1 mA/div
DC Accuracy
±3% (±1.5% typical)
Maximum Barewire Voltage
300 Volts (Category I)
TYPICAL CHARACTERISTICS
Insertion Impedance
0.1 Ohms at 1 MHz
0.2 Ohms at 50 MHz
1.0 Ohm at 100 MHz
Propagation Delay
17 ns
Maximum Conductor Diameter
3.8 mm (0.15 in.) Electrostatic Immunity (per
IEC 801-2) >8 kV
Amp *Sec Product
100 x 10
Risetime
3.5 ns
-6
Amp *sec
PHYSICAL CHARACTERISTICS
Probe Weight
0.25 kg (0.56 lbs.)
Cable Length
2 m (6.6 ft.)
POWER REQUIREMENTS
The A6312 is powered by the AM 503B/TM502A. The perfor­mance specifications listed are for this combination of products.
A6312 SAFETY CERTIFICATIONS
UL 3111-1 CSA 1010.1 IEC 1010-1, 1010-2-032
Page 2
A6312 DC­coupled Curre n t P ro b e
For further information, contact Tektronix:
SYSTEM SPECIFICATIONS
A6312/AM 503B A6312/AM 5030 A6312/AM 503A A6312/AM 503
Bandwidth DC to 100 MHz DC to 100 MHz DC to 80 MHz DC to 100 MHz Risetime (10% to 90%) 3.5 ns 3.5 ns 4.4 ns 3.5 ns Aberrations (typical) 7% p-p 7% p-p 7% p-p 7% p-p DC Gain Accuracy 3% (<1.5% typical) 3% 3% 3% Noise, RMS (typical) 250 µA 250 µA 250 µA 250 µA
1
Requires adjustment of R364 (HFComp) in amplifier.
1
1
2
ASEAN Countries
(65) 356-3900
Australia & New Zealand
61 (2) 888-7066
Austria
43 (1) 70177-261
Belgium
32 (2) 725-96-10
Brazil and South America
55 (11) 3741 8360
Canada
1 (800) 661-5625
Denmark
45 (44) 850700
Finland
358 (0) 4783 400
France & North Africa
33 (1) 69 86 81 81
Germany, Eastern Europe, & Middle East
49 (221) 94 77-0
Hong Kong
(852) 2585-6688
India
91 (80) 2265470
Italy
39 (2) 250861
Japan
(Sony/Tektronix Corporation) 81 (3) 3448-4611
Mexico, Central America, & The Caribbean
52 (5) 666-6333
The Netherlands
31235695555
Norway
47 (22)070700
People’s Republic of China
(86) 10-62351230
Republic of Korea
82 (2) 528-5299
Spain & Portugal
34 (1) 372 6000
Sweden
46 (8) 629 6500
Switzerland
41 (42) 219192
Taiwan
886 (2) 765-6362
United Kingdom & Eire
44 (1628) 403300
USA
1-800-426-2200
From other areas, contact:
Tektronix, Inc. Export Sales P.O. Box 500, M/S 50-255 Beaverton, Oregon 97077-0001 USA (503)627-1916
Visit our Web site:
World Wide Web: http://www.tek.com
Copyright © 1996, Tektronix, Inc. All rights reserved. Printed in U.S.A. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks.
6/96 TD/XBS 51W–10939–1
A6312 DC-COUPLED CURRENT PROBE
Includes: Instruction Manual P/N 070-9603-00
ORDERING INFORMAT I O N
RELATED PRODUCTS
AM 503B
Current Probe Amplifier
AM 5030
Programmable Current Probe Amplifier
AM 503S
Current Measurement System
A6302
Current Probe – DC to 50 MHz, 20A
A6302XL
Current Probe – DC to 17 MHz, 20A, 8 m cable
A6303
Current Probe – DC to 15 MHz, 100A
A6303XL
Current Probe – DC to 10 MHz, 100A, 8 m cable
A6304XL
Current Probe – DC to 2 MHz, 500A
P5200
High Voltage Differential Probe
AM 503S
Current Measurement System Opt 01 – Adds A6303 Probe Opt 03 – Substitutes A6303
Probe for A6302 Probe
Opt 05 – Substitutes A6312
Probe for A6302 Probe
Opt 09 – Deletes A6302 Probe