Tektronix CSA8000B, CSA8000, TDS8000, TDS8000B, 80A01 User manual

...
Read This First 80A00 and 82A0
0 Series Modules

Supported instruments

The 80A00 and Digital Serial Analyzer, as well as the legacy TDS/CSA8000, TDS/CSA8000B, and TDS/CSA/DSA8200 series instruments.
82A00 Series Modules are for use with the Tektronix DSA8300
CAUTION. Electrostatic discharge (ESD) can damage sampling module electrical
inputs an
Always use a grounded wrist strap (provided with your instrument) when installing o r removing modules from the mainframe and when connecting cables.
Wear anti-static clothing and work in a static-free workstation when u sing sampling modules.
Discharge cables to ground before connecting them to modules.
Do not input with the other end of the cable open.
d outputs.
create an ESD antenna by leaving cables attached to the sampling-module
www.tektronix.com Copyright © Tektronix
Use a T measurements.
Do no assembly can result.
Alw terminations installed.
To connectors clean at all times. Check that all connectors, jumpers, and protective caps are clean before connecting them to the module.
Do not transport or ship the instrument with the modules installed in the module compartments.
ektronix 80A02 EOS/ESD Protection Module when taking TDR
t drop the module since damage and misalignment of the photodiode optical
ays store and transport modules in a static-free container, with all supplied
prevent loss of optical power or damage to the optical connectors, keep the
*P071138613*
071-1386-13
Read This First 80A00 and 82A00 Series Modules

Install a module

Figure 1: Module insertion
1. Connect an antistatic wrist strap to the instrument front-panel ground connection. Place the strap on your wrist.
2. For optical modules, turn the module lock screws counterclockwise to rotate the retaining tabs to a vertical position before inserting the module.
3. Use the power cycle that is appropriate for your instrument:
DSA830
greater: Click Tools > View/Change Module Cong and follow the on-screen instructions.
Legacy instruments and DSA8300 with TekScope application version
6.0.X: Power off the instrument using the front-panel On/Standby switch.
4. Push the module into a compartment, using rm pressure. When the module
is se ated, turn the lock screw(s) clockwise to tighten the module in the slot.
5. Power on the instrument (legacy instruments or DSA8300 with TekScope
6.0.X software) or complete the View/Change Module Cong process (TekScope version 6.1.X and later). Let the instrument warm up for 20 minutes.
NOTE. The warm-up cycle is not required if you installed the module using the
iew/Change Module Congfunction(TekScopeversion6.1.Xandlater).
V
6. In the TekScope application, click Utilities > Compensate and run a compensation on the newly-installed module. You must also run a compensation if a module is moved from one slot to another, or an extender is installed, changed, or removed from a module.
0 instruments with TekScope application version 6.1.X and
2 80A00 and 82A00 Sampling Modules Read This First
Read This First 80A00 and 82A00 Series Modules

Remove a module

7. Use the Differe DiffChAlign)toalignbothchannels(andTDRstepswhenapplicable)of
supported two-channel electrical modules, to the module input connectors. This application steps through the connection setup to align both the acquisition and the TDR steps. For more information, see the help le for the application.
The Differential Channel Alignment application is available for DSA8300 instruments running TekScope version 6.2.X and later, and legacy instruments running TekScope version 5.2.x and later.
CAUTION. Always use the module ejectors to eject a module from a compartment.
Do not pull the module by the connectors; you can damage the connectors if you use them to pull out a module.
1. Connect an antistatic wrist strap to the instrument front-panel ground connection. Place the strap on your wrist.
2. Use the power cycle that is appropriate for your instrument:
ntial Channel Alignment application (Applications >
DSA8300 instruments with TekScope application version 6.1.X and
greater: Click Tools > View/Change Module Cong and follow the on-screen instructions.
Legacy instruments and DSA8300 with TekScope application version
6.0.X: Power off the instrument using the front-panel On/Standby switch.
3. Remove all cables or probes from the module connectors.
4. Turn the module lock screws counterclockwise until loose or they stop turning.
5. Push the module ejector lever to unseat the module. The arrow near an ejector
lever shows which compartment that ejector controls.
6. Use the module lock screws to pull the module partly out of the compartment. Then hold the module chassis and pull the module from the instrument.
80A00 and 82A00 Sampling Modules Read This First 3
Read This First 80A00 and 82A00 Series Modules

Moving modules while powered on

The TekScope View/Change Module Cong function lets you remove, install, or move modules in the instrument without powering down the mainframe. See the DSA8300 Digital Serial Analyzer Quick Start User Manual for information on how to use this function.
CAUTION. In
requires DSA8300 TekScope software version 6.1.X or greater. For DSA8300 TekScope software versions earlier than 6.1.X, or all legacy instruments (TDS/CSA/DSA8200, TDS/CSA8000B, and TDS/CSA8000B), always power off the mainframe to remove, install, or move modules. Failure to do so may result in damage to the module or mainframe.
You can download and install the latest DSA8300 software from www.tek.com/software.

User documentation for modules

To download the most recent versions of m anuals from the Tektronix Web site:
1. Go to w
2. Enter the module nomenclature, or the manual part number, in the Downloads
Ente
3. Select Manuals from the Select Download Type list and click GO.
4. Click Sort by DATE to place the newest les at the top of the list.
5. Click the document link to download the le.
ww.tek.com.
rModelNumbereld.
stalling or changing modules wh ile the instrument is powered on
Contact Tektronix Customer Support to purchase printed copies of manuals.
4 80A00 and 82A00 Sampling Modules Read This First
Read This First 80A00 and 82A00 Series Modules
Getlatestsof
tware
To determine if you need a newer TekScope application software for your sampling module, do the following:
1. Click Help > About in the TekScope application to show the current software version.
2. Use the Module software compatibility tables to nd the TekScope application software version required for your module and instrument. (See page 6.)
3. To download a newer version of t he application software:
a. Go to www.tek.com.
b. Enter the model number for your instrument in the Downloads Enter
Model Number eld.
c. Select Software from the Select Download Type list and click GO.
d. Select the appropriate software and follow the on-screen instructions to
download and install the software.
NOTE. Newer software versions support all previously-released modules.
80A00 and 82A00 Sampling Modules Read This First 5
Read This First 80A00 and 82A00 Series Modules
Module softwa
re compatibility tables
Use the following tables to verify that your TekScope application software version supports a specic sampling m odule or is appropriate for the version of Microsoft Windows load
DSA8300 instrument software and module compatibility.
Supported modules
TekScope software version 6.0.3.X:
80A01, 80A02, 80A03, 80A05, 80C01, 80C01, 80C02, 80C03, 80C04, 80C05, 80C06, 80C07/B, 80C08/B/C/D, 80C09, 80C10/B/C, 80C11/B, 80C12/B, 80C14, 80C25GBE, 80E01, 80E02, 80E03, 80E04, 80E06, 80E07, 80E08, 80E09, 80E10, 82A04
TekScope software version 6.2.X:
80E07B, 80E08B, 80E09B, 80E10B, 80E11, 80E11X1, 82A04B
1
Listed v operating system support all previously-released modules.
2
Listed modules include all available options unless noted otherwise.
3
The DSA8300 does not support the 80A06 module.
ed on your mainframe.
ersion number or greater. Newer versions of the TekScope application software and the DSA8300
1,2,3
Required operating system
Microsoft Windows 7 Ultimate (32 bit) v0.9
Microsoft Windows 7 Ultimate (32 bit) v0.9
6 80A00 and 82A00 Sampling Modules Read This First
Read This First 80A00 and 82A00 Series Modules
Legacy instrum
ents software and module compatibility (TDS/CSA8000,
TDS/CSA8000B and TDS/CSA/DSA8200).
TekScope software version
1.0
1.1
1.2
1.3
1.4
1.5
2.0.1.
2.0.1.
2.1
2.4 80A06
2.5 80E0
5.0
5.1.2
5.2.X 80E07B, 80E08B, 80E09B, 80E10B, 80E11,
1
2
3
1
Supported mo
80C01, 80C01
dules
-CR, 80C02/-CR, 80C03/-CR,
Required ope
rating system
80E01, 80E02, 80E03, 80E04, 80A01, 80A02
80C04-CR1
80C04-CR2, 80C05, 80C06
80C07/-CR1, 80C08/-CR1, 80C09/-CR1,
Microsoft Windows 98
80A03
80C08B/
80C07B/
-CR1-CR2, 80C10, 80E06
-CR1, 80C08C/-CR1/-CR2/-CR4,
80C11/-CR1/-CR2/-CR3/-CR4
2
3
5
80C12 (limited), 80A05 (limited)
80C12, 80A05
82A04 (Must be installed in a DSA8200, CSA8200, or TDS8200 mainframe)
7, 80E08, 80E09, 80E10
02 Option 01, 80C08C Option 01,
80C
3
Microsoft Windows 2000
80C10B, 80C11 Option 01, 80C12 Option 01, 80C25GBE
80C08D, 80C10C, 80C11B, 80C12B, 80C14
Microsoft Windows XP
80E11X1, 82A04B
Newer versions of the TekScope application software and instrument operating system software support all previously-released modules.
Replace product application software version 2.0.1.3 with the latest available version from the Tektronix Web.
Using the 80E07, 80E08, 80E09, or 80E10 modules in mainframes with older versions of software may prevent the application from starting.