6. In the TekScope application, click Utilities > Compensate and run
a compensation on the newly-installed module. You must also run a
compensation if a module is moved from one slot to another, or an extender
is installed, changed, or removed from a module.
0 instruments with TekScope application version 6.1.X and
280A00 and 82A00 Sampling Modules Read This First
Read This First 80A00 and 82A00 Series Modules
Remove a module
7. Use the Differe
DiffChAlign)toalignbothchannels(andTDRstepswhenapplicable)of
supported two-channel electrical modules, to the module input connectors.
This application steps through the connection setup to align both the
acquisition and the TDR steps. For more information, see the help file for
the application.
The Differential Channel Alignment application is available for DSA8300
instruments running TekScope version 6.2.X and later, and legacy instruments
running TekScope version 5.2.x and later.
CAUTION. Always use the module ejectors to eject a module from a compartment.
Do not pull the module by the connectors; you can damage the connectors if you
use them to pull out a module.
1. Connect an antistatic wrist strap to the instrument front-panel ground
connection. Place the strap on your wrist.
2. Use the power cycle that is appropriate for your instrument: