The 80A01 prescaled trigger amplifier was designed to be used with software
release versions 1.1.0 and later of the 8000 series sampling oscilloscope
application. If your instrument is running a release prior to version 1.1.0,
Tektronix recommends that you upgrade your instrument to the latest version.
Determining which
software version your
instrument has
How to get an upgrade
Using earlier versions
of software with your
80A01 module
To determine the software version running on your CSA8000 or TDS8000, select
About TDS/CSA in the Help menu from the user interface. This command
displays the About TDS/CSA dialog which lists the current software version
number.
If you find that you are running an older version of software, please visit the
Tektronix website at http://www.tektronix.com to find out how to get an upgrade
version mailed to you. When you visit the Tektronix website, select the Oscilloscopes page. You will find upgrade information there.
While the 80A01 was designed to be used with software release 1.1.0 or later, it
can be used with earlier versions of the software with the following limitations
and non-standard behavior:
1. If the 80A01 is inserted in the left-most small slot (labeled CH1 / CH2) of
the TDS/CSA8000 and at least one optical module is present, the 80A01 will
behave normally.
2. If the 80A01 is inserted in any other slot or there are no optical modules
present, the following behavior will be seen:
a. The instrument will behave as if the odd-numbered channel for the slot
in which the 80A01 is inserted is available for acquisition. Thus if the
module is inserted in the right-most small slot, the TDS/CSA8000 will
believe that it can acquire data on Channel 7. Turning on acquisition on
this ”pseudo-channel” will result in a waveform being acquired from a
floating input. This pseudo-waveform will appear as a DC level with all
points having amplitude approximately equal to zero.
b. The System Properties dialog (accessed by selecting System Properties
from the Utilities menu in the user interface) may contain incorrect
information.
In either scenario (1. or 2.), the actual performance of the 80A01 in amplifying
RF signals is not degraded.
– II –
WARRANTY
T ektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship
for a period of one (1) year from the date of shipment. If a product proves defective during this warranty period, T ektronix,
at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in
exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the
warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid.
T ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the
T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any
other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair
damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction
caused by the use of non-T ektronix supplies; or d) to service a product that has been modified or integrated with other
products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PAR TICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO
REP AIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO
THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE
LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE
OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
To Avoid Fire or
Personal Injury
Ground the Product. This product is indirectly grounded through the grounding
conductor of the mainframe power cord. To avoid electric shock, the grounding
conductor must be connected to earth ground. Before making connections to the
input or output terminals of the product, ensure that the product is properly
grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry .
Symbols and Terms
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
iii
General Safety Summary
T erms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
WARNING
High Voltage
Protective Ground
(Earth) Terminal
iv
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
Preface
Manual Structure
Related Manuals
This is the user manual for the 80A01 Trigger Prescale Preamplifier Module. It
covers capabilities of the module and how to install it, how to operate the
module, and lists specifications of the module.
This manual is composed of the following chapters:
HGetting Started shows you how to configure and install your module, lists
standard and optional accessories,
HOperating Basics describes controlling the module using the front panel and
the instrument user interface, system interaction, the programmer interface,
and user adjustments.
HSpecifications lists all typical and guaranteed specifications for this product.
This manual is part of a document set of standard-accessory manuals and online
documentation; this manual mainly focuses on installation and background
needed to use the module features. See the following list for other documents
supporting CSA8000 and TDS8000 instrument operation and service. The part
numbers of these documents are listed in the Accessories section of your
instrument user manual.
Manual nameDescription
CSA8000 & TDS8000 Online HelpAn online help system, integrated with the User Interface application that ships with the
CSA8000 and TDS8000 instruments
CSA8000 & TDS8000 ReferencesA quick reference to major features of the instrument and how they operate
CSA8000 & TDS8000 User Manual
80C00 Optical Module User Manual
CSA8000 & TDS8000 Programmer Online
Guide
CSA8000 & TDS8000 Service ManualDescribes how to service the instrument to the module level. This optional manual must
1
You can insert the 80A01 module user manual in Appendix A of the CSA8000 & TDS8000 User Manual.
1
1
The user manual for the CSA8000 and TDS8000 instruments
The user manual for the optical modules. Included as a standard accessory if you
ordered optical modules with your instrument
An alphabetical listing of the programming commands and other information related to
controlling the instrument over the GPIB
be ordered separately
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
v
Preface
Contacting Tektronix
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales support1-800-833-9200, select option 1*
Service support1-800-833-9200, select option 2*
Technical supportEmail: techsupport@tektronix.com
1-800-833-9200, select option 3*
1-503-627-2400
6:00 a.m. – 5:00 p.m. Pacific time
*This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
vi
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
Getting Started
Product Description
The Tektronix 80A01 Trigger Prescale Preamplifier Module is a one-channel
8-14 GHz AC coupled 50 W limiting preamplifier module that can be installed in
the CSA8000 Communications Signal Analyzer and TDS8000 Digital Sampling
Oscilloscope. It is designed to increase the sensitivity of the prescale trigger
input of the 8000 Series instruments to v200 mV
NOTE. Proper operation of the 80A01 module requires that the operating system
software installed on the main instrument is version 1.1.0 or greater. To display
the version installed, select About TDS/CSA8000 from the Help menu in the
user interface. See the section in this manual titled 80A01 Trigger PrescalePreamplifier Module Instrument Software Requirements for more information on
getting software upgrades if needed.
The module provides the features shown in Table 1. See Specifications for a full
list of product specifications.
pk-pk
.
T able 1: 80A01 module characteristics
CharacteristicValue or description
Input/output impedance50 W
Input/output connectorsSMA
CouplingAC
Trigger levelFixed at 0 V
Maximum input voltage swing1.1 V
Amplifier bandwidth8 to 14 GHz
Instrument trigger sensitivity with 80A01
installed
CAUTION. To prevent electrostatic damage to the 8000 Series instrument and
modules, follow the precautions described in this manual and the manuals
accompanying your instrument. (See Electrostatic Discharge starting on page 5.
p-p
v200 mV, 8 to 12.5 GHz (typical)
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
1
Getting Started
As shown in the block diagram in Figure 1, there are two identical SMA female
connectors on the front panel: one 50W RF input and one 50W RF output. Input
voltage is passed through the amplifier, allowing trigger sensitivity to drop as
low as 200 mV
in the 8–12.5 GHz range
pk-pk
The module receives power through the Tekprobe–Sampling interface. The
Power LED indicates the module is receiving power through the interface
connector.
CAUTION. To prevent electrostatic damage to the 8000 Series instrument and
module, follow the precautions described in this manual and the manuals
accompanying your instrument. (See Electrostatic Discharge starting on
page 5.)
Figure 1: 80A01 module block diagram
2
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
Options and Accessories
This section lists the standard and optional accessories available for the module,
as well as the product options.
Getting Started
Options
Standard Accessories
Optional Accessories
Option R3 (Three years of calibration services) is orderable with this module.
The following accessories are shipped with the instrument:
T able 2: Standard accessories
ItemPart Number
80A01 Trigger Prescale Preamplifier Module User Manual070-0873-xx
(2) SMA male 50 W termination caps015-1022-xx
Transit case, ESD protective004-5091-xx
The following accessories are orderable for use with the module at the time this
manual originally published. Consult a current Tektronix catalog for additions,
changes, and details:
T able 3: Optional accessories
ItemPart Number
2X attenuator (SMA male-to-female)015-1001-xx
5X attenuator (male-to-female)015-1002-xx
Coaxial cable015-0560-xx
Power divider015-0565-xx
Asymmetrical power divider015-0683-xx
SMA accessory kit020-1693-xx
CSA8000 & TDS8000 Service Manual071-0438-xx
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
3
Getting Started
Installation
The 80A01 module fits into one of the small module compartments on the front
panel of an 8000 Series instrument. Figure 2 shows the front panel of an
instrument and the locations of the small module compartments.
To install a module, first power off the instrument using the front-panel
On/Standby switch. Then place the module in a compartment and slowly push it
in with firm pressure. Once the module is seated, turn the hold-down screw on
the module to tighten the module into place. See Figure 2.
CAUTION. To prevent damage to the module or instrument, never install or
remove a module when the instrument is powered-on or when either input
connector is unprotected.
Hold-down screw
Small compartment ejectors
80A01 module
Figure 2: Installing a module
NOTE. When removing your module, first loosen the hold-down screw, then use
the module ejector on the main instrument to eject the module.
4
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
Getting Started
Electrostatic Discharge
To prevent electrostatic damage to the 8000 Series instrument and modules,
follow the precautions described in this manual and the manuals that come with
your instrument.
Circuitry in the module is very susceptible to damage from electrostatic
discharge or from overdrive signals. Be sure to only operate the module in a
static-controlled environment. Be sure to discharge to ground any electrostatic
charge that may be present on the center and outer connectors of cables before
attaching the cable to the module.
Know your signal source. If it is capable of delivering overvoltages, it is safer to
not depend on the signal source settings for protection, but instead use an
external attenuator that protects the input from the worst-case conditions. For
example, for a 10 V maximum source connected to a 1 V maximum module, use
a 10X attenuator. Where possible, connect your cables to the signal source first,
and to the module second.
CAUTION. To prevent damage from electrostatic discharge, install 50 W
terminations (Tektronix part number 015-1022-xx) on the I/O connectors before
removing them from an instrument, storing a module, or when not in use. Store
the module in a static-free container, such as the shipping container. Whenever
you move the module from one instrument to another, use a static-free container
to transport the module.
Static Controlled
Workstation
To prevent damage to the module, discharge to ground any electrostatic charge
that may be present on the center and outer conductors of cables before
attaching the cable to the module.
To prevent damage to the module, do not create an ESD antenna by leaving
cables dangling off the module input with the other end open.
To prevent damage to the module or instrument, never install or remove a
module when the instrument is powered-on.
Always use a wrist strap (provided with your instrument) when handling
modules or making signal connections. Wear anti-static clothing and work in a
static-free workstation when using modules.
To prevent damage to the module or instrument, do not apply a signal outside
the Maximum Input Voltage Swing for your module.
For information on creating a static-controlled workstation, consult the Electronic Industries Association document EIA-625; Requirements for Handling
Electrostatic-Discharge-Sensitive (ESDS) Devices.
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
5
Operating Basics
Usage
This chapter makes you familiar with the operation of your module. This chapter
describes the front-panel controls and connectors, interaction of the module with
your instrument, user adjustments, and the programmer interface.
Figure 3 shows the front panel of the module and identifies the connectors and
indicators.
CAUTION. To prevent damage to your module or instrument, do not apply a
signal outside the Maximum Input Voltage Swing for your module.
To prevent electrostatic damage to the instrument and modules, follow the
precautions described in this manual and the manuals accompanying your
instrument. (See Electrostatic Discharge starting on page 5.)
Always use a wrist strap (provided with your instrument) when handling
modules or making signal connections.
Front Panel
Hold-down screw
The input circuitry in your module is very susceptible to damage from overdrive
signals and electrostatic discharge. Never apply a DC or peak voltage greater
than the Maximum Input Voltage (see page NO TAG) of your module. Only
operate the instrument and module in a static-controlled environment.
The 80A01 module contains three primary features:
Use this to secure the module in the instrument module compartment once
connected. Turn clockwise to secure and counterclockwise to release. To prevent
damage to the module, use the compartment ejectors to remove the module from
the compartment (see Figure 2 on page 4 for the location of the ejectors).
6
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
Hold-down screw
TRIGGER PRESCALE
LIMITING PREAMPLIFIER 8-14 GHZ
OUTPUT
INPUT
POWER
AC COUPLED
50 W
1.1 V pk-pk MAX
Operating Basics
Input and Output
Connectors
Output connector
Input connector
Figure 3: 80A01 module front panel
The two connectors on the front panel of the 80A01 module provide connections
for the input and output signals. Both connectors are SMA female connectors.
Connector Care. Never attach a cable to a sampling-module connector if the cable
has a worn or damaged connector because you may damage the sampling-module connector. Use extra care when attaching or removing a cable from the
connectors. Turn only the nut, not the cable. When attaching a cable to a
sampling-module connector, align the connectors carefully before turning the
nut. Use light finger pressure to make this initial connection. Then tighten the
nut lightly with a wrench.
NOTE. For best repeatability and to prolong connector life, use a torque wrench
and tighten the connection to the range of 79-112 N@cm (7-10 inSlb).
If the module connectors will receive heavy use, such as in a production
environment, you should install adapters (such as a Tektronix 015-0549-xx for
3.5 mm connectors) on the module to make connections to the device under test.
Power LED
There is one LED on the front panel of the 80A01 module that indicates whether
or not the module is receiving power. The 80A01 module receives power through
the Tekprobe-Sample interface. When the module is receiving power the LED is
lit. If the module is installed but is not receiving power, check that the module is
seated correctly. Additionally, turn the hold-down screw clockwise to make sure
the module is installed tightly.
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
7
Operating Basics
System Interaction
The 80A01 module is part of a larger instrument system. This module functions
as a preamplifier for the prescaler trigger input. You connect a voltage to the
input connector and it is automatically passed through the amplifier. You then
send the output voltage where you want it to go for testing and evaluation.
The remainder of the system interaction for this module is contained in the
documentation set included with your TDS8000 or CSA8000 instrument. See the
8000 Series User Manual prescale trigger section for more information on how
to use signals output from the 80A01 module.
Programmer Interface Commands
The remote-programming commands for all modules are documented in the
CSA8000 & TDS8000 Programmer Online Guide.
User Adjustments
All module setups, parameters, and adjustments are controlled by the instrument.
To save, recall, or change any module settings, use the instrument menus or
front-panel controls or consult the CSA8000 & TDS8000 User Manual or
CSA8000 & TDS8000 Online Help
8
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
Specifications
This section contains specifications for the 80A01 Trigger Prescale Preamplifier
Module. All specifications are guaranteed unless noted as “typical.” Typical
specifications are provided for your convenience but are not guaranteed.
All specifications apply to all models of module unless noted otherwise. To meet
specifications, three conditions must first be met:
HThe instrument must have been calibrated/adjusted at an ambient tempera-
ture between +20_ C and +30_ C.
HThe oscilloscope must have been operating continuously for 20 minutes
within the operating temperature range specified.
HThe instrument must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifications
The 80A01 Trigger Prescale Preamplifier module is a single-channel 8-14 GHz
AC coupled limiting preamplifier. It increases the trigger sensitivity of the 8000
Series instruments to v200 mV
up to 12.5 GHz (typical).
pk-pk
T able 4: Module characteristics
SpecificationsCharacteristics
Mainframe interfaceTekprobe-Sampling, Level 3. Hot switching is not permitted.
Input and output
connectors
Input power
Output power+10 dBm maximum saturation
Output voltage
Input/output
impedance
Input/output VSWR2.0:1 maximum from 8-14 GHz (typical)
Power requirements2.35 W
Maximum input volt-
age swing
Noise figure5 dB maximum
1
2
1
2
See Input Voltage Swing specification for more information.
Amplifier functions to 14 GHz but prescale trigger is nominally limited to 12.5 GHz
for 200 mVpk-pk sensitivity.
Female SMA
–10 dBm to +5 dBm, AC coupled
8-12.5 GHz, input voltage 200 mV
12.5-14 GHz, input voltage 200 mV
50 W AC coupled (typical)
1.1 V
pk-pk
pk-pk
pk-pk
to 1 V
to 1 V
pk-pk
pk-pk
: w1 V
: w1 V
pk-pk
pk-pk
(typical)
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
9
Specifications
T able 5: Environmental specifications
SpecificationCharacteristics
TemperatureOperating: +10° C to +40° C
HumidityNon-operating: 5% to 90% relative humidity
Operating: 20% to 80% relative humidity
T able 6: Mechanical specifications
SpecificationsCharacteristics
Weight0.37 kg (13 oz.)
Overall dimensionsHeight:24 mm (0.95 in)
Width:74 mm (2.90 in)
Depth:114 mm (4.95 in)
Does not include connectors, connector savers, connector covers, push
buttons, or lock-down hardware protruding from the front or rear panels.
Construction materialChassisaluminum alloy;
Front panelplastic laminate;
Circuit boardsglass-laminate;
Cabinet sleevealuminum
Cabinet end coversaluminum
10
T able 7: Electromagnetic specifications
SpecificationCharacteristic
EmissionsThe instrument meets or exceeds the EMC requirements of the
following standards: Emissions European Community Requirements
(including EN 61326)
EN5501 1 Class A Radiated Emissions
EN5501 1 Class A Conducted Emissions
IEC 1000–3–2 Power Harmonic Current Emissions
SusceptibilityThe instrument meets or exceeds the EMC requirements of the
following standards:
EN61326–1 European Community Requirements
IEC 1000–4–2 Electrostatic Discharge Immunity
4 kV contact discharge
8 kV air discharge
Performance criteria B
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
Glossary
Attenuation
A decrease in magnitude of current, voltage, or power of a signal.
Attenuator
An electronic transducer that reduces the amplitude of a signal.
Bandwidth
The range of frequencies handled by a device or system. Bandwidth is a
measure of network capacity. Analogue bandwidth is measured in cycles per
second. Digital bandwidth is measured in bits of information per second.
Channel
A place to connect a signal or attach a network or transmission line to
modules.
dB
Decibel: a method of expressing power or voltage ratios. The decibel scale is
logarithmic. The formula for decibels is:
dB = 20 log (V
where Vi is the voltage of the incident pulse, V
and log is the decimal-based logarithmic function.
Dialog Box
A displayed box in which you enter instrument commands.
External Attenuation
Attenuation that is outside the module.
Impedance
The opposition to an AC signal in the wire. Impedance is very much like
resistance to a DC signal in a DC circuit. Impedance is made up of resistance
and inductive and capacitive reactance.
Trigger
An electrical event that initiates acquisition of a sample as specified by the
time base.
Waveform
The visible representation of an input signal or combination of signals.
i/Vref
)
is the voltage reference,
ref
80A01 Trigger Prescale Limiting Preamplifier Module User Manual
11
Index
A
Accessories, 3
optional, 3
standard, 3
Adjustments, 8
Attenuation, 11
B
Bandwidth, 11
C
Channel, 11
Circuitry, 2
Commands, Programmer interface, 8
Connectors, 7
Contacting T ektronix, vi
Power LED, 7
Product description, 1
Product support, contact information, vi
Programmer commands, 8
S
Safety Summary, iii
Service support, contact information, vi
Specifications, 9
Standard accessories, 3
Static controlled workstation, 5
System interaction, 8
F
Front panel controls, 6
G
Getting started, 1
Glossary, 11
I
Impedance, 11
Input Connector, 7
12
T
T echnical support, contact information, vi
T ektronix, contacting, vi
Trigger, 11
U
Usage, 6
W
Waveform, 11