Thank you very much for your shopping on us, if you need
any other manual, email me at
HTUjasonsyj@hotmail.comUTH , I
will do my best to help you.
If you find any others selling the manual made by me, please
email me also.
Thanks and enjoy reading.
Jasonsyj on Ebay
ARMY TM 11-6625-2925-24
AIR FORCE TO 33A1-10-242-2
ORGANIZATIONAL, DIRECT SUPPORT, AND GENERAL SUPPORT
MAINTENANCE MANUAL
DUAL TIME BASE, TEKTRONIX MODEL 7B92A
(NSN 6625-01-027-0265)
DEPARTMENTS OF THE ARMY AND THE AIR FORCE
21 JULY 1980
WARNING
DANGEROUS VOLTAGES EXIST IN THIS EQUIPMENT
Dangerous potentials exist at several points throughout this equipment. When the equipment is operated
with the covers removed DO NOT touch exposed connections or components. Some trans istors have
voltages present on their cases. Disconnect power before cleaning the equipment or replacing parts.
DON’T TAKE CHANCES!
Copyright 1974 (new material 1978) by Tektronix Inc.
REPRODUCED BY PERMISSION OF THE COPYRIGHT OWNER, TEKTRONIX INC.
TM 11-6625-2925-24
TO 33A1-10-242-2
Technical Manual)DEPARTMENTS OF THE ARMY
No. 11-6625-2925-24)AND THE AIR FORCE
TO 33A1-10-242-2)Washington, DC, 21 July 1980
ORGANIZATIONAL, DIRECT SUPPORT, AND GENERAL SUPPORT
MAINTENANCE MANUAL
DUAL TIME BASE, TEKTRONIX MODEL 7B92A
(NSN 6625-01-027-0265)
REPORTING ERRORS AND RECOMMENDING IMPROVEMENTS
You can help improve this manual. If you find any mistakes or if you know of a way to improve the
procedures, please let us know. Mail your letter or DA Form 2028 (Recommended Changes to
Publications and Blank Forms) direct to: Commander, US Army Communications and Electronics
Materiel Readiness Command and Fort Monm outh, AT T N: DRSEL-ME- MQ, Fort Monm outh, New Jer sey
07703.
For Air Force, submit AFTO Form 22 (T echnical Order System Publication Improvem ent Report and
Reply) in accordance with paragraph 6-5, Section VI, T.O. 00-5-1. Forward direct to prime ALC/MST.
In either case, a reply will be furnished direct to you.
THE FOLLOWING SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY.
TO AVOID PERSONAL INJURY, DO NOT PERFORM ANY SERVICING OTHER THAN THAT
CONTAINED IN OPERATING INSTRUCTIONS UNLESS YOU ARE
QUALIFIED TO DO SO.
3-1 7B92A Readout Character Selection........................................................................................................3-8
5-1 Test Equipment.........................................................................................................................................5-2
5-3 Fast Timing...............................................................................................................................................5-12
5-4 Fast Sweep Linearity.................................................................................................................................5-12
5-5 Differential Delay Time Accuracy..............................................................................................................5-13
5-6 Fast Differential Delay Time Accuracy......................................................................................................5-14
5-8 Fast Timing...............................................................................................................................................5-37
5-9 Fast Sweep Linearity.................................................................................................................................5-37
5-10 Differential Delay Time Accuracy..............................................................................................................5-39
5-11 Fast Differential Delay Time Accuracy......................................................................................................5-39
vi
@
7B92A
SAFETY SUMMARY
This manual contains safety information that the user must follow to ensure safe operation of this instrument.
WARNING information is intended to protect the operator; CAUTION information is intended to protect the instrument.
The following are general safety precautions that must be observed during all phases of operation and maintenance.
Ground the Instrument
To reduce electrical-shock hazard, the mainframe (oscilloscope) chassis must be properly
grounded. Refer to the mainframe manual for grounding information.
Do Not Operate in Explosive Atmosphere
Do not operate this instrument in an area where flammable gases or fumes are present. Such
operation could cause an explosion.
Avoid Live Circuits
Electrical-shock hazards are present in this instrument. The protective instrument cov ers must
not be removed by operating personnel. Component replacement and internal adjustments must
be referred to qualified service personnel.
Do Not Service or Adjust Alone
Do not service or make internal adjustments to this instrument unless another person, capable of
giving first aid and resuscitation, is present.
Warning Statements
Warning statements accompany potentially dangerous procedures in this manual. The following
warnings appear in this manual and are listed here for additional emphasis.
To avoid electrical shock, disconnect the instrument from the power source before soldering.
To avoid electrical shock, disconnect the instrument from the power source before replacing
components.
Handle silicone grease with care. Avoid getting silicone grease in eyes. Wash hands thoroughly
after use.
@
vii
7B92A
viii
REV B, JUN 1978
TM 11-6625-2925-24/T0 33A1-10-242-2
SECTION 0
INTRODUCTION
0-1.Scope
This manual contains instructions for organizational, direct support, and general support maintenance of Dual
Time Base, Tek tronix Model 7B92A. Throughout this manual Dual T ime Base, Tektr onix Model 7B92A is referred to as
the 7B92A.
0-2.Indexes of Publications
a.DA Pam 310-4
changes, or additional publications pertaining to the equipment.
b.DA Pam 310-7
pertaining to the equipment.
0-3.Maintenance Forms, Records, and Reports
a.Reports of Maintenance and Unsatisfac tor y Equipment
for equipment m aintenance will be those prescribed by TM 38-750, T he Arm y Maintenance Management System (Arm y).
Air Force personnel will use AFM 66-1 for maintenance reporting and TO-00-35D54 for unsatisfactory equipment
reporting.
b.Report of Packaging and Handling Deficiencies
Report) as prescribed in AR 735-11-2/NAVSUPINST 4440.127E/AFR 400-54/MCO 4430.3E and DSAR 4140.55.
c.Disc repancy in Shipment Report (DISREP) (SF 361)
(DISREP) (SF 361) as prescribed in AR 55-38/NAVSUPINST 4610.33B/AFR 75-18/MCO P4610.19C and DLAR 4500.15.
a.Army
can tell us what you don’t like about your equipment. Let us know why you don’t like the design. Tell us why a procedure
is hard to perform. Put it on an SF 368 (Quality Deficiency Report). Mail it to Comm ander, US Ar my Communications and
Electronics Materiel Readiness Command and Fort Monmouth, ATTN: DRSEL-ME-MQ, Fort Monmouth, New Jersey
07703. We’ll send you a reply.
. If your 7B92A needs improvement, let us k now. Send us an EIR. You, the user, are the only one who
. Refer to the latest issue of DA Pam 310-4 to determine whether there are new editions,
. Refer to DA Pam 310-7 to determ ine whether there are modific ation work orders (MWO ’s)
. Department of the Army form s and pr ocedures us ed
. Fill out and forward DD Form 6 (Pac kaging Improvement
. Fill out and forward Discrepancy in Shipment Report
b.Air Force
0-5.Destruction of Army Electronics Materiel
Destruction of Army electronics materiel to prevent enemy use shall be in accordance with TM 750-244-2.
. Air Force personnel are encouraged to submit EIR’s in accordance with AFM 900-4.
0-1
OPERATING INSTRUCTIONS
Section 1-7B92A
INTRODUCTION
The 7B92A Dual Time Base unit provides
normal, delayed, intensified, and alternate sweep
operation for TEKTRONIX 7000-Series Oscilloscopes.
Calibrated sweep rates from 0.2 second to 0.5
nanosecond and triggering to 500 megahertz are
provided. The 7B92A is intended for use with highfrequency 7000-Series Oscilloscope systems; however,
most 7B92A functions are compatible with all 7000Series Oscilloscopes.
Other features include lighted pushbutton
switches, compatibility with indicator oscilloscopes
having an alphanumeric readout system, and 0 to 9.8
times continuous sweep delay. A VARIABLE control
allows continuously variable sweep rates between
calibrated steps. Also, when operating in the AUTO
MAIN TRIGGERING Mode, a bright base line is
displayed in the absence of a trigger signal.
This section describes the operation of the frontpanel controls and connectors, and provides a functional
check and general operating information for this
instrument.
NOTE
The ALT OFF position, which allows
the Intensified mode of operation, is
applicable to instruments SN
B060000-above.
INSTALLATION
The 7B92A is designed to operate in the
horizontal plug-in compartment of the oscillosc ope. This
instrument can also be installed in a vertical plug-in
compartment to provide a vertical sweep on the crt.
However, when used in this manner, there are no
internal triggering or retrace blanking provisions, and the
unit may not meet the specifications given in the
Specification section of this manual.
Before proceeding with installation, check the
settings of the Time/Div Variable Selector multi-pin
connector and the Mainframe Selector multi-pin
connector (see Fig. 1-1). The Time/Div Variable
Selector determines whether the front-panel Time/Div
VARIABLE control operates in conjunction with the
delaying or delayed sweeps. The Mainframe Selector
adapts the 7B92A to the oscillsocope mainframe being
used. The two mainframe selections are:
1.7800 and 7900-Series Oscilloscopes.
2.All other 7000-Series Oscilloscopes.
NOTE
The 7B71 will not delay the 7B92A for
displaying alternating Delaying and
Delayed sweeps. Instead, it causes a
sweep lock-up to occur under the
following conditions; with the 7B71 in
the A horizontal compartment
REV B, JUN 1978
Fig. 1-1. Location of Variable and Mainframe Selector multi-pin conectors.
1-1
Operating Instructions-7B92A
and the 7B92A in the B horizontal
compartment, the sweep locks up if
the 7B92A is in its Alternate mode
and the Mainframe Horizontal mode
is Alt. Lock-up can be avoided by
changing the Mainframe Horizontal
mode to Chop or B. This allows both
7B92A sweeps to be displayed.
To install the 7B92A in a plug-in compartment,
push it in until it fits firmly into the compartment. The
front panel of the 7B92A should be flush with the front
panel of the oscilloscope. Even though the gain of the
oscilloscope is standardized, the sweep calibra tion of the
7B92A should be checked when installed. The
procedure for checking the unit is given under Sweep
Calibration in the Operating Checkout procedur e In this
section.
To remove the 7B92A, pull the release latch (see
Fig. 1-2) to disengage the unit from the os c illos cope, and
pull it out of the plug-in compartment.
Information. Fig. 1-1shows the Time/Div Variable and
Mainframe Selectors. Fig. 1-3 shows the front-panel
controls, connectors, and indicators.
MAIN TRIGGERING Controls
(1) LEVEL Control
Selects amplitude point on trigger signal where sweep
triggering occurs when the MAIN TRIGGERING MODE
AUTO, NORM, or SINGLE SWEEP switches are
pressed. When the MAIN TRIGGERING MODE HF
SYNC switch is pressed, the LEVEL control adjusts the
frequency of the trigger generator to synchronize with the
frequency (or sub-harmonic) of the triggering signal to
provide a stable display.
(2) SLOPE Control
Permits triggering on the positive or negative s lope of the
trigger signal (except in HF SYNC).
(3) TRIG’D Indicator
When lit, indicates that the sweep is triggered and will
produce a display with correct setting of the POSITION
control and the controls on the assoc iated am plif ier plugin unit(s) and oscilloscope.
Fig. 1-2. Location of release latch.
FRONT-PANEL CONTROLS,
CONNECTORS, AND INDICATORS
All controls, connectors, and indicators r equired
for the operation of the 7B92A, except the Time/Div
Variable and Mainframe Selectors, are located on the
front panel. A brief description of the front-panel
controls, connectors, and indicators is given here. More
detailed information is given under General Operating
(4) MODE Switches
Select the desired MAIN TRIGGERING mode. Selec ted
mode is indicated by lighted pushbutton.
AUTO: Selects a triggered sweep initiated by the applied
trigger signal at a point determined by the LEVEL control
and SLOPE control when the trigger signal repetition rate
is above 30 hertz and within the frequency range
selected by the COUPLING switches. When the LEVEL
control is outside the amplitude range, the trigger
repetition rate is outside the frequency range selected by
the COUPLING switches, or the trigger signal is
inadequate, the sweep free-runs to provide a reference
trace.
HF SYNC: Sweep initiated by trigger signals with
repetition rates above 100 megahertz and within the
range selected by the COUPLING switch. Stable display
can be obtained when the LEVEL control adjusts the
frequency of the trigger generator to the frequency (or
sub-harmonic) of the trigger signal. When the LEVEL
control is adjusted to frequencies between subharmonics, the sweep free-runs.
REV B, JUN 1978
1-2
Operating Instructions-7B92A
Fig. 1-3. Front-panel controls and connectors.
NORM: Sweep initiated by the applied trigger signal at a
point selected by the LEVEL control and SLOPE contr ol
over the frequency range selected by the COUPLING
switches. Triggered sweep can be obtained only over
the amplitude range of the applied trigger signal. W hen
the LEVEL control is either outside the am plitude range,
the trigger repetition rate is outside the fr equency range
selected by the COUPLING switches, or the trigger
signal is inadequate, there is no trace.
SINGLE SW EEP-RESET: When the SINGLE SWEEPRESET switch is pressed, a single trace will be
presented when the next trigger pulse is received. The
SINGLE SW EEP-RESET switch rema ins lit until REV B,
REV B, JUN 1978
JUN 1978 a trigger is received and the sweep is
completed. The SINGLE SWEEP-RESET switch must
be pressed again before another sweep can be
displayed.
(5) COUPLING Switches
Select trigger signal coupling. Selected coupling is
indicated by lighted pushbutton.
AC: Rejects dc and attenuates ac signals below about
30 hertz. Accepts signals between 30 hertz and 500
megahertz.
1-3
Operating Instructions-7B92A
AC LF REJ: Rejects dc and attenuates signals below30
kilohertz. Accepts signals between 30 k ilohertz and 500
megahertz.
AC HF REJ. Rejects dc and attenuates signals above
50 kilohertz. Accepts signals from 30 hertz to 50
kilohertz.
DC: Accepts all signals from dc to 500 megahertz.
(6) SOURCE Switches
Select the triggering source. Selected source is
indicated by lighted pushbutton.
INT: Trigger signal obtained internally from amplifier
plug-in unit of oscilloscope.
LINE: Trigger signal obtained internally from the line
voltage applied to the oscilloscope.
EXT: Trigger signal obtained from an external source
applied to the MAIN TRIG IN connector.
EXT Þ 10: Trigger signal obtained from an external
source applied to the MAIN TRIG IN c onnector. In this
position, the external signal is attenuated before it is
applied to the trigger circuit.
Sweep Controls
(7) TIME/DIV OR DLY TIME
Selects the basic sweep rate for norm al s weep operation
and selects the delay time (multiplied by the DELAY
TIME MULT dial setting) when operating in the Alternate
or Delayed sweep modes. The VARIABLE control m ust
be in the CAL (knob in) position for calibrated sweep
rate.
(8) DLY’D Time/Division
Selects the delayed sweep rate for operation in Delayed
and Alternate sweep modes. The VARIABLE control
must be in the CAL (knob in) position for calibrated
sweep rate.
Four display modes can be selected by the following
switch settings:
Normal Sweep: A normal sweep is selected when the
TIME/DIV OR DLY TIME switch and the DLY’D
Time/Division switches are lock ed together at the same
sweep rate. The DLY’D Time/Division knob must be
pressed in for normal sweep mode. Calibrated sweep
rates from 0.2 second/division to 0.5 nanosecond/division can be selected
ALT Sweep: The Alternate mode is s elected when the
DLY’D Time/Division switch is pulled out and rotated
clockwise and the TRACE SEP control is turned
clockwise from the ALT OFF pos ition. In this mode, the
delaying sweep is displayed (with an intensified zone
during the time that the delayed sweep runs) alternately
with the delayed sweep.
Intensified Sweep: The Intensified mode, a function of
the delaying and delayed sweeps is selected when the
DLY’D Time/Division switch is pulled out and rotated
clockwise, and the TRACE SEP control is rotated fully
counterclockwise to the ALT OFF pos ition (see Fig. 1-3).
In this mode, a portion of the delaying sweep is
intensified during the time that the delayed sweep
generator runs.
DLY’D Sweep: The Delayed sweep mode is selected
when the DLY’D Time/Division switch is pulled out,
rotated for the desired delayed sweep rate, and then
pushed in. In this mode, the delayed sweep is displayed
at a rate determined by the DLY’D Time/Division switch
at the end of each delay period, as selected by the
TIME/DIV OR DLY TIME switch and the DELAY TIME
MULT dial setting.
(9) VARIABLE
Two-position switch actuated by the VARIABLE control
to select calibrated or uncalibrated sweep rates. In the
CAL position (knob in) the VARIABLE control is
inoperative and the sweep rate is calibrated. When
pressed and released, the knob moves out to activate
the VARIABLE control for uncalibrated sweep rates. The
sweep rate in each TIME/DIV OR DLY TIME switch
position can be reduced at least to the sweep rate of the
next slower position. The VARIABLE control will operate
with either the delaying or delayed sweep by means of
the internal Time/Div Variable Selector.
(10) POSITION Control
Positions the display horizontally on the graticule.
(11) CONTRAST Adjustment
Varies the relative brightness of the intensified por tion of
the delaying trace when in the Alternate mode.
(12) TRACE SEP/ALT OFF Control
This control vertically positions the delaying sweep
display up to 3.5 divisions above the delayed sweep
display when in the Alternate mode. The ALT OFF
position allows the Intensified sweep mode of operation,
permitting the display of an intensified portion of the
delaying sweep.
1-4
REV B, JUN 1978
Operating Instructions—7B92A
(13) INTENSITY Control
Varies the intensity of the Intensified sweep only, when
operating in the Intensified or Alternate mode (must be
set near clockwise end when 7B92A is operating in some
7000-Series Oscilloscopes).
(14) SWP CAL Adjustment
Screwdriver adjustment sets the basic timing of the
7B92A to compensate for slight differences in input
sensitivity when changing indicator oscilloscopes.
(15) DELAY TIME MULT Dial
Provides variable delay of 0 to 9.8 times the basic delay
time selected by the TIME/DIV OR DLY TIME switch.
Delayed Triggering Controls
(16) LEVEL Control
Determines the delayed trigger mode and the delayed
trigger level.
RUNS AFTER DELAY TIME (LEVEL control turned fully
clockwise into detent): Delayed sweep runs immediately
following the delay time selected by the TIME/DIV OR
DLY TIME switch and the DELAY TIME MULT dial.
Delayed Slope, Coupling, and Source functions are
inoperative.
INT: The delayed trigger signal is obtained from the
vertical amplifier of the oscilloscope.
EXT: The delayed trigger signal is obtained from an
external source connected to the DLY’D TRIG IN
connector.
(19) COUPLING Switch
Two-position pushbutton switch to determ ine the m ethod
of coupling the trigger signal to the delayed trigger circuit.
AC: Rejects dc and attenuates signals below 30 hertz.
Accepts trigger signals from 30 hertz to 500 megahertz.
DC: Accepts trigger signals from dc to 500 megahertz.
Front-Panel Inputs
(20) MAIN TRIG IN Connector
Serves as an external trigger input for the main tr iggering
circuit when the MAIN TRIGGERING SOURCE EXT or
EXT Þ 10 pushbutton switches are pressed.
(21) TERM Switch
Two-position pushbutton switch to select 50 ohms (out
position) or 1 megohm (in position) input impedance for
the MAIN TRIG IN and DLY’D TRIG IN connectors.
Delayed Sweep Triggerable: When the Delayed
Triggering LEVEL control is turned counterclock wise out
of detent, the delayed sweep is triggerable. The Delayed
Triggering LEVEL control can now be rotated to select
the amplitude point on the trigger signal at which the
delayed sweep is triggered. In the Delayed Sweep
Triggerable mode, the delayed Slope, Coupling, and
Source functions are activated.
(17) SLOPE Switch
Two-position pushbutton switch to select the slope of the
trigger signal which starts the delayed sweep.
+: The delayed sweep can be triggered on the positive
slope of the trigger signal.
-: The delayed sweep can be triggered on the negative
slope of the trigger signal.
(18) SOURCE Switch
Two-position pushbutton switch to select the source of
the delayed trigger signal.
REV A, JUN 1978
DLY’D TRIG IN Connector
Serves as an external trigger input for the delayed
triggering circuit when the Delayed Triggering SOURCE
switch is set to EXT.
Do not exceed 7 volts (rms) of
external signal when 50 n termination
is selected.
FUNCTIONAL CHECK
The following procedures may be used for
familiarization or as a check of basic instrument
operation. The procedure is divided into two parts,
Sweep Functions and Triggering Functions. A complete
operating check of the 7B92A functions c an be made by
performing both parts, or each part may be performed
separately. If performing the functional chec k procedur e
reveals a malfunction or
1-5
Operating Instructions-7B92A
possible improper adjustment; first check the operation
of the associated plug-in units, then refer to the
instruction manual for maintenance and adjustment
procedures.
NOTE
For optimum high-frequency
performance, the 7B92A should be
installed in an oscilloscope system
with similar frequency and sweeprate capabilities.
Setup Procedure
1.Install the 7B92A in a horizontal
compartment of the oscilloscope.
2.Install the amplifier plug-in unit in a
vertical compartment.
3.Turn on the oscilloscope and allow at least
20 minutes warm up.
4.Set the 7B92A controls as follows:
MAIN TRIGGERING
SLOPE(+)
MODEAUTO
COUPLINGAC
SOURCEINT
Delayed Triggering
LEVELRUNS AFTER DELAY
TIME
SLOPE(+)
COUPLINGAC
SOURCEINT
Sweep Controls
POSITIONMidrange
INTENSITYAs desired
TIME/DIV OR1 ms
DLY TIME
DLY’D Time/Division1 ms (knob in)
VARIABLECAL
Time/Div VariableDelayed Sweep
Selector (internal)
DELAY TIME MULT1.00
TRACE SEPMidrange
5.Set the oscilloscope to display the plug-in units
and adjust for a well-defined display. See oscilloscope
and amplifier plug-in unit instruction m anuals for detailed
operating instructions.
Sweep Functions
Normal Sweep. Perform the following
procedure to obtain a normal sweep and demonstrate
the function of the related sweep controls:
1.Perform the Setup Procedure.
2.Connect a 4 volt, one-kilohertz signal from
the oscilloscope calibrator to the amplifier plug-in unit
input.
3.Adjust the amplifier plug-in unit
volts/division switch for two divisions of display.
4.Rotate the MAIN TRIGGERING LEVEL
control for a stable display.
5.Rotate the POSITION control and note that
the trace moves horizontally.
6.Check the crt display for one complete
cycle per division. If necessary, adjust the SWP CAL
screwdriver adjustment for one complete cycle per
division over the center eight graticule divisions. Be sure
that the timing of the calibrator signal is accurate within
0.5%.
Alternate and Delayed Sweep. Perform the
following procedure to obtain alternate-delayed sweeps
and demonstrate the function of the related sweep
controls:
7.Pull out the DLY’D Time/Division knob and
rotate clockwise to 0.1 ms f or the Alternate s weep mode.
Note both an intensified trace and a normal-intensity
delayed sweep trace on the crt. Increased oscilloscope
intensity may be required for viewing the delayed sweep.
8.Rotate the INTENSITY control and note that
it varies the intensity of the delaying sweep.
9.Rotate the CONTRAST adjustment for
adequate identification on the intensified portion of the
delaying sweep when alternating with Delayed sweep.
NOTE
When operating in the Intensified
mode, the intensified zone is
controlled by the intensity of the test
oscilloscope instead of the 7B92A
CONTRAST and INTENSITY controls.
REV B, JUN 1978
1-6
Operating Instructions—7B92A
10. Rotate the TRACE SEP control to vertically
position the delaying sweep trace with respect to the
delayed sweep trace. With the TRACE SEP control in
the ALT OFF position, only the Intensified sweep will be
displayed.
11. Rotate the DELAY TIME MULT dial and
note that the amount of delay time before the intensif ied
portion of the display is controlled by the DELAY TIME
MULT dial.
12. Press the DLY’D Time/Division switch in for
the Delayed sweep mode. Note the delayed display with
sweep rate determined by the DLY’D Time/Division
switch.
13. Press and release the VARIABLE control.
Rotate the VARIABLE control and note that the sweep
rate indicated by the DLY’D Time/Division switch can be
varied to at least the sweep rate of the next adjacent
position (0.2 ms). The internal Time/Div Variable
Selector must be set to the Delayed Sweep position.
Return the VARIABLE control to the CAL position (knob
in).
Triggering Functions
Main and Delayed Triggering. Perform the
following procedure to obtain a triggered alternate,
normal, or delayed sweep and demonstrate the function
of the related controls:
14. Perform the Setup Procedure. Connect the
one-kilohertz calibrator signal from the oscilloscope to
the amplifier plug-in unit input and adj ust for about four
divisions of vertical display.
15. Set the DLY’D Time/Division switch and the
TIME/DIV OR DLY TIME switch to 1 ms, and press in the
DLY’D Time/Division knob (normal sweep mode).
Rotate the MAIN TRIGGERING LEVEL control for a
stable display.
16. Check that a stable display can be obtained
with the MAIN TRIGGERING COUPLING switch set to
AC, AC HF REJ, and DC, for both the positive and
negative positions of the SLOPE control (MAIN
TRIGGERING LEVEL control may be adjusted as
necessary to obtain a stable display). Remove all
connections from the oscilloscope system.
17. Connect a 0.4 volt, one-kilohertz signal from
the oscilloscope calibrator to the amplifier plug-in unit
and to the MAIN TRIG IN connector. Set the MAIN
TRIGGERING SOURCE switch to EXT. Set the
amplifier plug-in unit volts/div switch for about four
divisions of display. Check that a stable display can be
obtained with the MAIN TRIGGERING COUPLING
switch set to AC, AC HF REJ, and DC, for both the
positive and negative positions of the SLOPE control
(MAIN TRIGGERING LEVEL control m ay be adjusted as
necessary for a stable display).
18. Change the MAIN TRIGGERING SOURCE
switch to EXT Þ 10. Set the oscilloscope calibratorfor
four volts at
one kilohertz and adjust the amplifier plug-in unit volts/div
switch for about four divisions of display. Check that a
stable display can be obtained with the MAIN
TRIGGERING COUPLING switch set to AC, AC HF REJ,
and DC, for both the positive and negative positions of
the SLOPE control (MAIN TRIGGERING LEVEL contr ol
may be adjusted as necessary to obtain a stable
display). Remove all connections from the oscilloscope
system.
19. Set the MAIN TRIGGERING COUPLING
switch to AC and SOURCE switch to INT. Connect a
one-kilohertz calibrator signal from the oscilloscope to
the amplifier plug-in unit input and adj ust for about four
divisions of display amplitude. Adjust the MAIN
TRIGGERING LEVEL control for a stable display. Set
the MODE switch to NORM and check for a stable
display. Change the MODE switch to AUTO and adjust
the LEVEL control for a free-running display. Change
the MODE switch to NORM and check for no display.
20. Adjust the MAIN TRIGGERING LEVEL
control for a stable display. Change the MAIN
TRIGGERING MODE switch to SINGLE SWEEP. Press
the RESET button and check for one sweep as the
RESET button is pressed. Remove the one-kilohertz
signal from the amplifier plug-in unit and press the
RESET button. Check for no dis play and RESET button
light on. Connect the one-kilohertz signal to the amplifier
plug-in unit input and check for one sweep as the signal
is applied. Remove all connections from the
oscilloscope system.
21. Set the MAIN TRIGGERING MODE switch
to AUTO and SOURCE switch to AC. Turn the Delayed
Triggering LEVEL control fully clockwise to the RUNS
AFTER DELAY TIME position. Pull out the DLY’D
Time/Division switch and rotate to 0.2 ms (Alternate
mode). Connect a 0.4 volt, one-kiloher tz signal from the
oscilloscope calibrator to the am plifier plug-in unit input
and adjust for about two divisions of display amplitude.
Rotate the MAIN TRIGGERING LEVEL control for a
stable intensified display. The INTENSITY control may
need to be adjusted to view the intensified display.
Rotate the DELAY TIME MULT dial and note that the
delay time before the intensified portion of display is
continuously variable.
22. Set the Delayed Triggering SLOPE,
COUPLING, and SOURCE switches to (+), AC, and INT.
Rotate the Delayed Triggering LEVEL control
counterclockwise out of the detent and adjust for a stable
display. Rotate the
REV A, JUN 1978
1-7
Operating Instructions-7B92A
DELAY TIME MULT dial and note that the intensified
sweep does not start at the completion of the delay time
but waits for the next trigger pulse.
23. Check that a stable display can be obtained
with the Delayed Triggering COUPLING switch set to AC
and DC for both the (+) and (-) positions of the SLOPE
switch (Delayed Triggering LEVEL control may be
adjusted as necessary for a stable display).
24. Change the Delayed Triggering SOURCE
switch to EXT. Connect a 0.4 volt, one-kilohertz signal
from the oscilloscope calibrator to the DLY’D TRIG IN
connector. Check that a stable dis play can be obtained
with the Delayed Triggering COUPLING switch set to AC
and DC, for both the (+) and (-) positions of the SLOPE
switch (Delayed Triggering LEVEL control may be
adjusted as necessary for a stable delayed sweep
display).
High-Frequency Synchronization. Perform
the following procedure to obtain a triggered alternate,
normal, or delayed sweep with a 100 megahertz to 500
megahertz input signal:
NOTE
To check HF sync operation, a signal
source frequency between 100
megahertz and 500 megahertz is
required. If a 100 to 500 megahertz
signal source is not available, or if it
is not desired to check the HF sync
operation, the last two steps of this
procedure may be deleted.
25. Change the MAIN TRIGGERING MODE
switch to HF SYNC. Connect a high-frequency signal
source (100 to 500 megahertz) to the amplifier plug-in
unit input and adjust for four divisions of vertical
deflection. Adjust the TIME/DIV OR DLY TIME switch
and DLY’DTime/Division switch for about six cycles of
display. Rotate the MAIN TRIGGERING LEVEL control
throughout its range and note that the sweep is
alternately stable, then free-running, several times during
the rotation (stable display indicates that the triggergenerator frequency is adjusted to a sub-har m onic of the
trigger signal frequency). Check that stable dis plays can
also be obtained with the MAIN TRIGGERING
COUPLING switch set to AC, AC LF REJ, and DC (MAIN
TRIGGERING LEVEL control may be adjusted, as
necessary, for a stable display).
26. Pull out the DLY’D Time/Division switch and
rotate it to the next fastest sweep rate fr om the setting of
the TIME/DIV OR DLY TIME switch (Alternate mode).
When the LEVEL contr ol is in the detent (RUNS AFTER
DELAY TIME) the display should be stable. Rotate the
LEVEL control counterclockwise out of the detent and
adjust for a stable display. Check that stable crt dis plays
can be obtained with the Delayed Triggering COUPLING
switch set to AC and DC. Disconnect the high-frequency
signal from the oscilloscope system.
GENERAL OPERATING INFORMATION
MAIN TRIGGERING
The MAIN TRIGGERING MODE, COUPLING,
and SOURCE pushbutton switches are arranged in a
sequence which places the most-often used position at
the top of each series of pushbuttons. With this
arrangement, a stable display can usually be obtained by
pressing the top pushbuttons: AUTO, AC, and INT.
When an adequate trigger signal is applied and the
LEVEL control is set correctly, the TRIG’D indicator will
light. If the TRIG’D indicator is not lit: (1) the LEVEL
control is at a setting outside the range of the trigger
signal from the amplifier plug-in unit, ( 2) the tr igger s ignal
is inadequate, or (3) the trigger-signal f requenc y is below
the lower frequency limit of the COUPLING switch
position. If the desired display is not obtained with these
pushbuttons, other selections must be made. Refer to
the following discussions or the instruction manuals for
the oscilloscope and amplifier plug-in unit for more
information.
Main Trigger Modes
The MODE pushbutton switch selects the mode
in which the main sweep is triggered.
AUTO. The AUTO pushbutton provides a
triggered display with the correct setting of the LEVEL
control (see Trigger Level discussion) whenever an
adequate trigger signal is applied. The TRIG’D indicator
lights when the display is triggered.
When the trigger repetition rate is outside the
frequency range selected by the COUPLING switch or
the trigger signal is inadequate, the sweep free-runs at
the sweep rate indicated by the TIME/DIV OR DLY TIME
switch (TRIG’D indicator off). An adequate trigger signal
ends the free-running condition and a triggered dis play is
presented. When the LEVEL control is at a setting
outside the amplitude range of the trigger signal, the
sweep also free runs at the sweep rate indicated by the
TIME/DIV OR DLY TIME switch. This type of freerunning display can be useful when it is desired to
measure only the maximum peak -to-peak amplitude of a
signal without observing the waveshape (such as in
bandwidth measurements).
HF SYNC. The HF SYNC pushbutton permits
stable displays of repetitive signals with only 0.5 division
of internal trigger signal (100 millivolts external signal)
required for frequencies between 100 megahertz and
500 megahertz.
REV C, JUN 1978
1-8
Operating Instructions—7B92A
A triggered display is obtained when the LEVEL
control adjusts the frequency of the trigger gener ator to
the frequency or sub-harmonic of the trigger signal.
Stable displays may be obtained several times between
the limits of the LEVEL control, depending on the
amplitude and frequency of the trigger signal. The
LEVEL control should be set for optimum display. (In the
High-Frequency Synchronization mode, the display is not
necessarily stable when the TRIG’D indicator is lit.)
When the LEVEL control is adjusted between
subharmonics of the trigger signal frequency, the trigger
repetition rate is below 100 megahertz or outside the
frequency range selected by the COUPLING switch, or
the trigger signal amplitude is inadequate, the sweep
free-runs at the sweep rate determ ined by the TIME/DIV
OR DLY TIME switch.
NORM. The NORM pushbutton provides a
triggered display with the correct setting of the LEVEL
control whenever an adequate trigger signal is applied.
The TRIG’D indicator lights when the display is triggered.
The NORM trigger mode must be used to
produce triggered displays with trigger repetition rates
below about 30 hertz. When the TRIG’D indicator is off ,
no trace will be displayed.
SINGLE SWEEP. When the signal to be
displayed is not repetitive or varies in amplitude,
waveshape, or repetition rate, a conventional repetitive
type display may produce an unstable presentation. A
stable display can often be obtained under these
circumstances by using the single-sweep feature of this
unit. The Single Sweep mode is also useful to
photograph non-repetitive or unstable displays.
To obtain a single-sweep display of a repetitive
signal, first obtain the best possible display in the Norm
mode. Then, without changing the other MAIN
TRIGGERING controls, press the SINGLE SWEEPRESET pushbutton. A single trace is presented each
time the pushbutton is pressed. Fur ther sweeps cannot
be presented until the SINGLE SWEEP-RESET
pushbutton is pressed again. If the displayed signal is a
complex waveform composed of pulses of varying
amplitude, successive single-sweep displays may not
start at the same point on the waveform. To avoid
confusion due to the crt persistenc e, allow the display to
disappear before pressing the SINGLE SW EEP-RESET
pushbutton again. At fast sweep rates, it may be difficult
to view the single-sweep display. The apparent trace
intensity can be increased by reducing the ambient light
level or using a viewing hood as recommended in the
oscilloscope instruction manual.
When using the Single Sweep mode to
photograph waveforms, the graticule must be
photographed separately in the normal manner to
prevent over-exposing the film. Be sure the camera
system is well protected against stray light, or operate
the system in a darkened room. For repetitive
waveforms, press the SINGLE SWEEP-RESET
pushbutton only once for each waveform unless the
signal is completely symmetrical. Otherwise, multiple
waveforms may appear on the film . For random signals,
the lens can be left open until the signal triggers the unit.
Further information on photographic techniques is given
in the appropriate camera instruction manual.
Main Trigger Coupling
The MAIN TRIGGERING COUPLING switches
select the method in which the trigger signal is
connected to the trigger circuits. Each position permits
selection or rejection of som e frequency components of
the trigger signal which trigger the sweep.
AC. The AC pushbutton blocks the dc
component of the trigger signal. Signals with lowfrequency components below about 30 hertz are
attenuated. In general, ac coupling can be used for most
applications. However, if the signal contains unwanted
frequency components, or if the sweep is to be triggered
at a low repetition rate or dc level, one of the other
COUPLING switch positions will provide a better display.
AC LF REJ. The AC LF REJ pushbutton rejects
dc, and attenuates low-frequency trigger signals below
about 30 kilohertz. Therefore, the sweep is triggered
only by the higher-frequency components of the trigger
signal. This position is particularly useful for providing
stable triggering if the trigger signal contains linefrequency components. Also, the AC LF REJ position
provides the best Alternate mode vertical dis plays at fas t
sweep rates when comparing two or more unrelated
signals.
AC HF REJ. The AC HF REJ pushbutton
passes all low-frequency signals between about 30 hertz
and 50 kilohertz. The dc component is rejected and
signals outside the above range are attenuated. When
triggering from complex waveforms, this position is
useful to provide a stable display of the low-frequency
components. AC HF REJ coupling should not be used
when operating in the HF SYNC triggering mode.
DC. The DC pushbutton can be used to provide
stable triggering from low-f requency signals which would
be attenuated in the other modes. It can also be us ed to
trigger the sweep when the trigger signal reaches a dc
level set by the LEVEL control. When using internal
triggering, the setting of the amplifier plug- in unit position
control affects the triggering point.
REV A, JUN 1978
1-9
Operating Instructions-7B92A
Main Trigger Source
The MAIN TRIGGERING SOURCE pushbutton
switches select the source of the trigger signal that is
connected to the main trigger circuits.
INT. The INT pushbutton connects the trigger
signal from the am plif ier plug-in unit. Fur ther s elec tion of
the internal trigger signal may be provided by the
amplifier plug-in unit or oscilloscope; s ee the instruction
manuals for these ins trum ents f or m ore inf orm ation. For
most applications, the internal source can be used.
However, some applications require special triggering
that cannot be obtained in the INT position. In such
cases, the LINE or EXT positions of the SOURCE
switches must be used.
LINE. The LINE pushbutton connects a sample
of the power-line voltage from the oscilloscope to the
trigger circuit. Line triggering is useful when the input
signal is time-related (multiple or s ubmultiple) to the line
frequency. It is also useful for providing a s table display
of a line-frequency component in a complex waveform.
Line triggering cannot be used when operating in the HF
sync triggering mode.
EXT. The EXT pushbutton connects the signal
from the MAIN TRIG IN connector to the trigger cir cuit.
The external signal mus t be tim e-related to the displayed
waveform for a stable display. An external trigger signal
can be used to provide a triggered display when the
internal signal is too low in amplitude for correct
triggering, or contains signal components on which
triggering is not desired. It is also useful when signal
tracing in amplifiers, phas e-shift network s, wave-shaping
circuits, etc. The signal from a single point in the circuit
can be connected to the MAIN TRIG IN connector
through a probe or cable. The sweep is then triggered
by the same signal at all times and allows amplitude,
time relationship, or waveshape changes of signals at
various points in the circuit to be examined without
resetting the MAIN TRIGGERING controls.
EXT Þ 10. The EXT Þ 10 pushbutton operates
the same as described for EXT except that the external
signal is attenuated. Attenuation of high-amplitude
external trigger signals is desirable to increase the
effective range of the LEVEL control.
Input Impedance
The input Impedance of the MAIN TRIG IN and
DLY’D TRIG IN connectors may be selected by the frontpanel TERM switch.
FRONT-PANEL IN -- 1 MΩ. The 1 MΩ position is
suitable for most low- and medium-frequency
applications or when using a 10X probe. The 1 M
position provides a high input impedance for minimum
loading on the trigger signal source.
FRONT-PANEL OUT -- 50 Ω. The 50 Ω position is
recommended for high-frequency applications requiring
maximum overal bandwidth. The 50 ohm termination
should be used when externally triggering from a 50 ohm
system.
Trigger Slope
The MAIN TRIGGERING SLOPE control
(concentric with the MAIN TRIGGERING LEVEL control)
determines whether the trigger circuit responds on the
positive-going or negative-going portion of the trigger
signal. The trigger slope cannot be selected when
operating in the high-frequency synchronization mode.
When the SLOPE control is in the + (positive-going)
position, the display starts on the positive-going portion
of the waveform; in the - (negative-going) position, the
display starts on the negative-going portion of the
waveform (see Fig. 1-4). When several cycles of a
signal appear in the display, the setting of the SLOPE
control is often unimpor tant. However, if only a certain
portion of a cycle is to be displayed, correct setting of the
SLOPE control is important to provide a display which
starts on the desired slope of the input signal.
Trigger Level
The MAIN TRIGGERING LEVEL control
determines the voltage level on the trigger signal at
which the sweep is triggered when operating in the Auto,
Norm, or Single Sweep modes. When the LEVEL
control is set in the + region, the trigger cir cuit responds
at a more positive point on the trigger signal. W hen the
LEVEL control is set in the - region, the trigger circuit
responds at a more negative point on the trigger signal.
Fig. 1-4 illustrates this effect with different settings of the
SLOPE switch.
To set the LEVEL control, first s elect the MAIN
TRIGGERING MODE, COUPLING, SOURCE, and
SLOPE. Then set the LEVEL control fully
counterclockwise and rotate it c lockwise until the display
starts at the desired point. Less selection of the
triggering level is available as the trigger signal frequency
exceeds 150 megahertz.
When operating in the MAIN TRIGGERING HF
sync mode, the LEVEL control synchronizes the trigger
generator frequency to a sub-harmonic of the trigger
signal frequency. Trigger slope and level cannot be
selected.
REV A, JUN 1978
Ω
1-10
Operating Instructions—7B92A
REV B, JUN 1978
Fig. 1-4. Effect of MAIN TRIGGERING LEVEL and SLOPE controls on crt display.
1-11
Selecting Sweep Rates
The TIME/DIV OR DLY TIME switch selects
calibrated sweep rates for the delaying sweep. The
DLY’D Time/Division switch selects calibrated sweep
rates for the delayed sweep. The sweep rate for the
delaying sweep is bracketed by the black lines on the
clear plastic flange of the TIM E/DIV OR DLY TIME
switch. Sweep rate of the delayed sweep is indicated by
the white line on the DLY’D Time/Division knob. W hen
the white line on the outer knob is set to the same
position as the lines on the inner knob, the two knobs
lock together and the sweep rate of both generators is
changed at the same time. However, when the DLY’D
Time/Division knob is pulled outward, the clear plastic
flange is disengaged and only the delayed sweep rate is
changed. This allows changing the delayed sweep rate
without changing the delaying sweep rate. The
TIME/DIV OR DLY TIME switch and the DLY’D
Time/Division switch also select display modes. See
Display Mode discussion in this section for further
information.
A VARIABLE control is provided concentric with
the TIME/DIV OR DLY TIME and DLY’D Time/Division
switches (see Fig. 1-3). This control can be used with
either the delaying or delayed sweeps as determined by
the Time/Div Variable Selector multi-pin connector
(internal, see Fig. 1-1 for location). The VARIABLE
control also incorporates a two-position switch to
determine if the applicable sweep is calibrated or
uncalibrated. W hen the VARIABLE c ontr ol is pres s ed in,
it is inoperative. However, when pressed and released,
the VARIABLE control is activated for uncalibrated
sweep rates. The sweep rate can be returned to the
calibrated position by pressing the VARIABLE knob in.
This feature is useful when a spec if ic unc alibrated s weep
rate has been obtained and it is desired to switch
between calibrated and uncalibrated sweep rates.
Switching from uncalibrated to calibrated and vice-versa
does not affect the setting of the VARIABLE control. The
VARIABLE control allows the sweep rate in each
Time/Division switch pos ition to be increased to at least
the next adjacent switch position.
Time Measurement
When making time measurements from the
graticule, the area between the second and tenth vertic al
lines of the graticule provides the most linear time
measurements (see Fig. 1-5). Position the start of the
timing area to the second vertical line and adjust the
TIME/DIV OR DLY TIME switch so the end of the timing
area falls between the fourth and tenth vertical lines.
Display Modes
Four display modes can be selected by
appropriate settings of the TIME/DIV OR DLY TIME and
DLY’D Time/Division switches.
Operating Instructions-7B92A
Fig. 1-5. Area of graticule used for most accurate time
measurements.
Normal Sweep Operation. To select the
Normal sweep display mode, press in the DLY’D
Time/Division switch and set it to the same sweep rate
as the TIME/DIV OR DLY TIME switch. Rotate the
Delayed Triggering LEVEL control clockwise into the
detent to the RUNS AFTER DELAY TIME position.
Calibrated sweep rates in the Normal sweep
display mode are 0.2 s/Div to 0.5 ns/Div. By using the
VARIABLE control (Time/Div Variable Selector connec tor
set for variable Delayed Sweep rates) uncalibrated
sweep rates to 0.5 s/Div are available. Triggering in the
Normal sweep display mode is controlled by the MAIN
TRIGGERING controls.
Alternate Sweep Display. To select the
Alternate display mode, pull out the DLY’D Time/Division
knob and rotate it to a desired sweep rate faster than the
TIME/DIV OR DLY TIME switch setting. In this mode,
both an intensified sweep and a delayed sweep are
displayed (see Fig. 1-6).
The intensified trace of the Alternate sweep
display provides an intensified portion on the delaying
sweep during the time the delayed sweep is running.
The amount of delay time between the start of the
delaying sweep and the intensified portion is determined
by the TIME/DIV OR DLY TIME switch and the DELAY
TIME MULT dial. Triggering for the delaying sweep
portion of the intensified trace is controlled by the MAIN
TRIGGERING controls; triggering for the intensified
portion of the delayed sweep trace is controlled by the
Delayed Triggering controls.
REV B, JUN 1978
1-12
Fig. 1-6. Typical alternate sweep display.
The TRACE SEP control vert ically positions the
intensified trace up to 3.5 divisions above the delayed
sweep trace. The brightness of the intens ified zone may
be varied by the CONTRAST adjustment. The
brightness of the intensified sweep m ay be varied by the
7B92A INTENSITY control.
Intensified Sweep Mode. The Intensified
sweep mode is selected when the DLY’D Time/Division
switch is pulled out, rotated to the desired sweep rate
faster than the TIME/DIV OR DLY TIME switch setting,
and the TRACE SEP control is rotated counterclock wise
to the ALT OFF position.
Delayed Sweep Display. The Delayed sweep
display mode is selected when the DLY’D Time/Division
switch is pulled out, rotated to the desired sweep rate,
and then pushed in. In this mode, only the delayed
sweep is displayed.
Calibrated sweep rates in the Delayed sweep
mode are available from 0.2 s/Div to 0.5 ns/Div. By
using the VARIABLE control (Tim e/Div Variable Selector
connector set for variable delayed sweep rates),
uncalibrated delayed sweep rates to 0.5 s/Div are
available. Triggering for the delayed sweep is controlled
by the Delayed Triggering controls.
@
Operating Instructions—7B92A
Delay Time Multiplier
The DELAY TIME MULT dial (functional in the
Delayed, Intensified, Alternate, or Mainframe Delaying
modes) provides 0 to 9.8 tim es c ontinuous sweep delay.
The amount of tim e that the delaying sweep runs before
the start of the delayed sweep is determined by the
settings of the TIME/DIV OR DLY TIME switch and the
DELAY TIME MULT dial.
For example, a DELAY TIME MULT dial setting
of 3.55 corresponds to 3.55 crt divisions of delaying
sweep. Thus, 3.55 multiplied by the delaying sweep rate,
indicated by the TIME/DIV OR DLY TIME switch, gives
the calibrated delay time before the start of the delayed
sweep.
Delayed Sweep Triggering
The Delayed Triggering LEVEL control
determines the delayed triggering mode, and delayed
triggering level. W hen the LEVEL control is in the RUNS
AFTER DLY TIME detent position (fully clockwise), the
delayed sweep starts immediately after the delay time
(determined by the TIME/DIV OR DLY TIME switch and
DELAY TIME MULT dial). This m ode permits selection
of continuously variable delay times by rotating the
DELAY TIME MULT dial. The Delayed Triggering
LEVEL control and SLOPE, COUPLING, and SOURCE
switches are inoperative.
When the delayed sweep is triggerable (Delayed
Triggering LEVEL out of the RUNS AFTER DELAY
detent), the delayed sweep does not start at the
completion of the delay time. Instead, it waits until a
trigger pulse is received by the Delayed Triggering
circuits. The delay time in this mode is dependent not
only on the settings of the delay time controls, but on the
Delayed Triggering controls and the occurrence of the
delayed-sweep triggering signal as well. The primary
purpose of this mode is to eliminate jitter from the
delayed sweep is triggered by the input waveform, jitter is
eliminated from the delayed sweep display even though it
may be inherent in the input waveform. When jitter in the
delayed sweep display is not a problem, the Runs After
Dly Time mode should be used.
In the Delayed Sweep Triggerable mode, the
Delayed Triggering LEVEL control is rotated to select the
amplitude point on the trigger signal at which the delayed
sweep is triggered. The Slope, Coupling, and Source
functions are the same for delayed triggering as for
MAIN TRIGGERING (see MAIN TRIGGERING SLOPE,
COUPLING, SOURCE, and TERM switch discussions in
this section).
1-13
Section 2-7B92A
SPECIFICATION
This instrument will meet the electrical characteristics listed under Performance Requirement in Table 2-1,
following complete calibration. The following electrical characteristics apply over an ambient temperature range of 0° C to
+50° C, except as otherwise indicated. Warm-up time for given accuracy is 20 minutes.
Table 2-1.
ELECTRICAL CHARACTERISTICS
Characteristic Performance Requirement Supplemental Information
MAIN TRIGGERING
Trigger Sensitivity Triggering Frequency Minimum Triggering
Range Signal Required
Operating in AUTO, NORM, or INT
2
EXT
3
SINGLE SWEEP MODE (div) (mV)
COUPLING
AC 30 Hz to 20 MHz 0.5 100
20 MHz to 500 MHz 1.0 500
30 kHz to 20 MHz 0.5 100
AC LF REJ
’
20 MHz to 500 MHz 1.0 500
AC HF REJ 30 Hz to 50 kHz 0.5 100
DC Dc to 20 MHz 0.5 100
20 MHz to 500 MHz 1.0 500
Operating in HF SYNC MODE
AC 100 MHz to 500 MHz 0.5 100 Use NORM or SINGLE
AC LF REJ SWEEP MODE for signals
DC below about 30 Hz
AC HF REJ Not recommended for HF
SYNC MODE
External Trigger Input
Level Range
EXT At least + and - 3.5 volts Not applicable in HF SYNC
MAIN TRIGGERING
EXT Þ 10 At least + and -35 volts MODE
Maximum Safe Input
1 MΩ Input 250 V (dc + peak ac)
50 Ω Input 1 W average (7 V rms)
Input R and C
1 MΩ Input Approximately 1 MΩ
paralleled by approximately 20 pF
REV B, JUN 1978
2-1
Specification--7B92A
Table 2-1 (cont)
CharacteristicPerformance RequirementSupplemental Information
Normal, Alternate (Delayed
Sweep Trace) and Delayed
Sweep
Sweep Rates0.2 s/div to 0.5 ns/div in 27 calibrated stepsSelected by TIME/DIV
OR DELAY TIME switch.
Steps in a 1-2-5
sequence
REV B, JUN 1978
2-2
Speclfication-7B92A
Table 2-1 (cont)
CharacteristicPerformance RequirementSupplemental Information
Sweep AccuracyMeasured in 7900-Series Oscilloscope
Over Center 8 Div+15°C to +35°C0°C to +50°C
.2 s/Div to 20 ns/DivWithin 2%Within 3%
10 ns/Div to 5 ns/DivWithin 3%Within 4%
2 ns/Div to 1 ns/DivWithin 4%Within 5%
.5 ns/DivWithin 5%Within 6%
Over Any 2 Div Portion Within
Center 8 Div
.2 s/Div to 10 ns/DivWithin 5%Within 5%
5 ns/Div to .5 ns/DivWithin 10%Within 10%
Variable Sweep RateContinuously variable between calibratedExtends sweep rate to
sweep ratesat least 0.5 s. VARIA-
BLE control internally
switchable between Delaying and Delayed
Sweeps. Variable range
at least 2.5:1
Intensified Sweep (DelayingSelected by TIME/DIV
Sweep Trace of AlternateOR DLY TIME switch.
Display)Steps in a 1-2-5 sequence
Sweep Rates0.2 s/div to 10 ns/div in 23 calibrated steps
Sweep AccuracyMeasured in 7900-Series Oscilloscope
Over Center 8 Div+15°C to +35°C0°C to +50°C
.2 s/Div to 20 ns/DivWithin 2%Within 3%
10 ns/DivWithin 3%Within 4%
Over Any 2 Div Portion WithinWithin 5%Within 5%
8 Div
Variable Sweep RateContinuously variable between calibrated sweepExtends sweep rate to at
ratesleast 0.5 s. Variable con-
trol internally switchable
between Delaying and
Delayed Sweeps
Trace SeparationIntensified sweep can be positioned at least 3.5
div above the delayed sweep
ALT OFFIntensified sweep of the delaying sweep is dis-Allow Intensified mode of
played when the TIME/DIV OR DELAY TIME switchoperation
is pulled out and rotated clockwise, and the TRACE
SEP control is in ALT OFF position
Variable Time Delay
Delay Time Range
DLY TIME/DIV Settings
.2 s/Div to 10 ns/Div0 to 9.8 times the DLY TIME switch setting (0
to 1.96 s)
REV B, JUN 1978
2-3
Specification-7B92A
Table 2-1 (cont)
CharacteristicPerformance RequirementSupplemental Information
Differential Delay Time Measurement
Accuracy
+15°C to +35°C
.2 s/Div to .1 µs/Div
Both DELAY TIME MULTWithin 0.75% of measurement +0.25% of fullFull scale is 10 times the
dial settings at 0.50 orscaleTIME/DIV OR DLY TIME
greatersetting
One or both DELAY TIMEWithin 0.75% of measurement +0.5% of full scale
MULT dial settings less than+5 ns
0.50
50 ns/Div to 10 ns/Div
Both delay times equal to orWithin 1.0% of measurement +0.5% of full scale
greater than 25 ns
One or both delay times lessWithin 1.0% of measurement +1.0% of full scale
than 25 ns+5 ns
Delay Time JitterJitter specification does not
apply to the first 2% of the
maximum available delay
time (DELAY TIME MULT
dial setting less than 0.20)
0.2 s/Div to 50 µs/Div1 part or less in 50,000 of the maximum available
delay time (10 times the TIME/DIV OR DLY
TIME switch setting)
20 µs/Div to 10 ns/Div1 part or less in 50,000 of the maximum available
delay time (10 times the TIME/DIV OR DLY
TIME switch setting +0.5 ns)
1
Will not trigger on the sine waves of 8 div amplitude or less ( internal), or 3 V or less (external) at
60 Hz or below.
2
For Internal Triggering only, the specified -3 dB frequency of the Vertical System replaces any
frequencies in the above table when the number in the table is greater than the -3 dB frequency of
the Vertical System.
3
Triggering signal amplitude requirements increased by factor of 10 for EXT - 10 operation.
4
The specified -3 dB frequency of the Vertical System replaces any frequencies in the above table
when the number in the table is greater than the -3 dB frequency of the Vertical System.
Table 2-2.Table 2-3.
ENVIRONMENTALPHYSICAL
Refer to the Specification for the associated oscilloscope.Net Weight3.062 lbs (1.372 kg)
DimensionsSee Fig. 2-1, Dimensional Drawing
2-4
REV A, JUN 1978
Specification—7B92A
REV C, JUN 1978
Fig. 2-1. Dimensional drawing.
2-5
WARNING
THE FOLLOWING SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED
PERSONNEL ONLY. TO AVOID PERSONAL INJURY, DO NOT PERFORM ANY
SERVICING OTHER THAN THAT CONTAINED IN OPERATING INSTRUCTIONS
UNLESS YOU ARE QUALIFIED TO DO SO.
THEORY OF OPERATION
Section 3-7B92A
This section of the m anual contains a description of the
circuitry used in the 7B92A. The description begins with
a discussion of the instrument using the block diagram in
the Diagrams section. Each c ircuit is then described in
detail with a block diagram provided to show the major
interconnections between circuits, and the relationship of
the front-panel controls to each circuit.
BLOCK DIAGRAM DESCRIPTION
The Main Trigger Generator ensures a stable crt
display by starting each sweep at the same point on the
waveform. The output of the Main Trigger Generator is a
fast-rise pulse which starts the Delaying Sweep
Generator.
The Delaying Sweep Generator produces a
linear voltage ramp. This ramp is displayed when the
time base is in the Intensified or Alternate mode. The
Delaying Sweep ramp is also used as a delay-time
reference when the time base is in the Alternate or
Delayed mode. The delay time is set by the slope of the
Delaying Sweep ramp and the Delay Pick off com parator
voltage. When the time base is in the Normal Sweep
mode, the comparator voltage is set to 0 (zero) and the
Delay Pick off outputs a pulse when the Delaying Sweep
ramp starts.
The Delayed Trigger Generator produces a fast-
rise pulse to start the Delayed Sweep Generator. When
the Delayed Trigger Generator is in the Runs Af ter Delay
Time mode, the puls e from the Delay Pick off produces
the Delayed Trigger output pulse. When the Delayed
Trigger Generator is in the Tr iggerable After Delay Time
mode, the pulse from the Delay Pick off enables the
Delayed Trigger Generator, which then processes the
input signal in the same way as the Main Trigger
Generator.
The Delayed Sweep Generator produces a linear
voltage ramp that is displayed as either the Normal or
Delayed sweep.
The Output Amplifier horizontally positions the
crt display and couples the proper sweep ramp(s) to the
oscilloscope.
DETAILED CIRCUIT DESCRIPTION
Main Trigger Generator
The Main Trigger Gener ator provides a stable display by
starting the Delaying Sweep Generator at a selected
point on the input waveform. The triggering point can be
varied by the LEVEL control and may be on either the
positive or negative slope of the waveform. The input
signal may be the waveform being displayed (INT), a
waveform from an exter nal sourc e (EXT or EX T ’10), or a
sample of the power-line voltage (LINE).
The bandwidth of the Main Trigger Generator is
set by the COUPLING switches. Dc coupling provides a
bandwidth of dc to 500 megahertz. Ac coupling blocks
dc and frequencies below about 30 hertz. AC LF REJ
(ac coupling, low-frequency rejection) passes
frequencies above 30 kilohertz. AC HF REJ (ac
coupling, high-frequency rejection) passes frequencies
between 30 hertz and 30 kilohertz.
External Source (SN B070000-above). The
external trigger signal is connected to the Main Trigger
Generator through the MAIN TRIG IN connector, J100.
The input impedance at J100 can be set to either 1
megohm or 50 ohms by TERM switch S205.
If the SOURCE switch is set to EXT, relay K6
energizes and applies the trigger signal to C11 and R11.
Signals below 30 kilohertz are connected to the gate of
Q22A through R14, C12, R12, and R11. Signals
between 30 kilohertz and 100 megahertz are connected
to the gate of Q22A through R17 and C11. These
signals pass through Q22A and Q24 to pin 3 of U74.
Signals above 100 megahertz are connected to pin 4 of
U74 through C20. (Pins 3 and 4 of U74 are internally
connected.)
The Horizontal Logic controls the Main Trigger
Generator, the Delayed Trigger Generator, and the
Output Amplifier. The Horizontal Logic produces a
Trigger Disable pulse which resets the trigger generator s
and allows the sweep generators to reset and stabilize
before starting another ram p. The Horizontal Logic also
controls which sweep ramp is pass ed thr ough the O utput
Amplifier to be displayed.
If the SOURCE switch is set to EXT . 10, relay
K6 de-energizes and applies the input signal to C10 and
R9. The signal is then divided by 10 before being
applied to the gate of Q22A.
3-1
Theory of Operation-7B92A
External Source (SN B069999-below). The external
trigger signal is connected to the Main Trigger Generator
through the MAIN TRIG IN connector, J100. The input
impedance at J100 can be set to either 1 megohm or 50
ohms by TERM switch S205.
If the SOURCE switch is set to EXT, relay K6
energizes and applies the trigger signal to C11 and R11.
Signals below 30 kilohertz are connected to the gate of
Q22A through R14, C12, R12, and R11. Signals
between 30 kilohertz and 100 megahertz are connected
to the gate of Q22A through R17 and C11. These
signals pass through Q22A and 024 to pin 3 of U64.
Signals above 100 megahertz are connected to pin 4 of
U64 through C20. (Pins 3 and 4 of U64 are internally
connected.) .
f the SOURCE switch is set to EXT ’ 10, relay K6
deenergizes and applies the input signal to C10 and R9.
The signal is then divided by 10 before being applied to
the gate of Q22A.
Internal Source (SN B070000-above). The
internal trigger signal from the vertical channel of the
oscilloscope is connected to the Main T rigger Generator
through J150. Signals below 30 kilohertz are amplified
by U44B and connected, along with the offset from the
LEVEL control, to pin 1 of U74 and to the base of Q86.
Signals above 30 kilohertz are coupled through C46 to
pin 14 of U74. (Pins 13 and 14 of U74 are internally
connected.)
Internal Source (SN B069999-below). The
internal trigger signal from the vertical channel of the
oscilloscope is connected to the Main T rigger Generator
through J150. Signals below 30 kilohertz are amplified
by U128B and connected, along with the offset from the
LEVEL control, to pin 1 of U64 and to the base of 072.
Signals above 30 kilohertz are coupled through C46 to
pin 14 of U64. (Pins 13 and 14 of U64 are internally
connected.)
Internal-External Amplifier (SN B070000above). Amplifier U74 is a dual-channel, differential
amplifier with a common reference voltage for both
channels. The input channel to be amplif ied is selected
by the bias on pins 2 and 15. When R77 is connected to
-15 volts, the internal trigger signal (pins 13 and 14) is
amplified. When R79 is connected to -15 volts, the
external trigger signal (pins 3 and 4) is amplified. The
push-pull outputs (pins 8 and 9) are connected to the
inputs (pins 5 and 8) of U122.
When AC H F REJ coupling is selected, R77 and R79
are both disconnected from -15 volts, disabling U74.
The low-frequency signals (below 30 kilohertz) are then
connected to U122 through Q86 (internal signals) or 082
(external or line signals).
Internal-External Amplifier (SN B069999-
below). Amplifier U64 is a dual-channel, differential
amplifier with a common reference voltage for both
channels. The input channel to be amplif ied is selected
by the bias on pins 2 and 15. When R66 is connected to
-15 volts, the internal trigger signal (pins 13 and 14) is
amplified. When R67 is connected to -15 volts, the
external trigger signal (pins 3 and 4) is amplified. The
push-pull outputs (pins 8 and 9) are connected to the
inputs (pins 3 and 14) of U84.
When AC HF REJ coupling is selec ted, R66 and
R67 are both disconnected from - 15 volts, disabling U64.
The low-frequency signals (below 30 kilohertz) are then
connected to U84 through Q72 (internal signals) or Q74
(external or line signals).
Slope Selector and Trigger Generator (SN
B070000above). U122 selects the slope of the input
waveform on which triggering occ ur s . If pin 3 is high, the
time base will trigger on the negative slope of an inter nal
signal or the positive slope of an external signal. If pin 3
is low, the time base will trigger on the positive slope of
an internal signal or the negative slope of an external
signal. (The internal signal is inverted by U74 or Q86
before being applied to U122.)
The delay mode control signal into U122, pin 4 is
functional only when the unit is operating as a delayed
sweep unit in the B horizontal compartment of a
mainfram e with two horizontal compartments . When the
unit is operating in the Independent or Triggerable After
Delay Time modes (as determined by the delaying
sweep time-base unit in the A horizontal com partment),
there is no affect on the Trigger Generator circuits.
However, when the unit is operating in the B Starts Af ter
Delay Time mode, a HI level at U122, pin 4 causes the
trigger IC to generate a gate pulse at pin 15 when the
trigger disable input goes low.
Slope Selector Amplifier (SN B069999-below).
Amplifier U84 select s the slope of the input wavef orm on
which triggering occurs. If pin 6 is high, the time base
will trigger on the positive slope of an inter nal signal or
the negative slope of an external signal. If pin 6 is low,
the time base will trigger on the negative slope of an
internal signal or the positive slope of an external signal.
(The internal signal is inverted by U64 or Q72 before
being applied to U84.)
The push-pull outputs (pins 7, 8 and 9, 10) of
U84 are connected to the inputs (pins 3 and 13) of U104.
Output Amplifier (SN B069999-below).
Amplifier U104 provides a final gain stage befor e driving
the trigger-output tunnel diodes. The push-pull output of
U104 (pins8 REV B, JUN 1978
3-2
Theory of Operation-7B92A
and 9) drives the emitter and base of Q0112. Transis tor
Q112 converts the push-pull output of U104 to a singleended output to drive tunnel diode CR114. As the
emitter-base voltage of Q112 increases, the current
through CR114 increases. When the current through
CR114 reaches 10 milliamperes, it switches to its high
level. The fast rise-time of CR114 is coupled through
C141 to the Arming Tunnel Diode, CR172, which also
switches high. The high level at the anode of CR172
increases the current through the Gate Tunnel Diode,
CR176. The rise-time of CR114 is also coupled thr ough
C124 (and a 1 nanosecond delay line) to CR176. The
combination of the added current from CR172 and the
pulse from CR114 (1 nanosec ond later) s witches CR176
high. The fast-rise pulse from CR176 is connected to
the Sweep Start Comparator of the Delaying Sweep
Generator.
Trigger Disable (SN B070000-above). At the
end of each sweep, the Logic circuits supply a Trigger
Disable pulse to U122, pin 2. A HI level disables the
trigger generator to allow enough time for the sweep
generator to stabilize before another trigger pulse s tarts
the next sweep.
Trigger Disable (SN B069999-below). At the
end of each sweep, the Horizontal Logic produces a
Trigger Disable pulse to reset the Arming and Gate
Tunnel Diodes to a low level and allow enough tim e for
the sweep generator to reset and stabilize before another
trigger pulse starts the next sweep. T he Trigger Disable
signal is connected to the base of 0151 through
interconnecting pin CA. W hen the Trigger Dis able pulse
is high, 0148 and Q158 turn on, removing the bias
current from CR172 and CR176, respectively. The rest
of the trigger generator operates normally, but CR172
and CR176 will not switch to a high level.
High-Frequency Synchronization (SN
B070000above). The HF sync mode increases the
sensitivity of the trigger generator and is useful at
frequencies above 10 megahertz. In the HF sync mode,
the trigger generator (U122) is caused to free run by
reducing the hysteresis to zero (pin 9). The LEVEL
control, R50, adjusts the hysteresis around zero, varying
the oscillating frequency and enabling U122 to
synchronize with the input signal or a sub-harmonic of
the input signal.
The inputs to U122 pins 5 and 8 are forced to be
balanced by negative feedback loop U132B, U44A, and
U74. This enables the HF sync circuit to function
regardless of the dc level of the input signal. U44 is
used to select phase of feedback required f or the input
channel selected.
High-Frequency Synchronization (SN B069999below).
The H F sync mode increases the sensitivity of the
trigger generator at frequencies above 100 megahertz.
In the HF sync mode, the trigger generator free runs at a
frequency determined by the LEVEL control, R50. The
LEVEL control adjusts the free-r unning frequency of the
trigger generator to be close enough to the signal
frequency, or sub-harmonic, to synchronize with it.
The positive feedback loop required to maintain
oscillation is through Q104 to input pin 4 of U104,
through Q0112 to CR114, and back to Q104. A negative
feedback loop is also present to maintain synchronization
of the oscillator with the trigger signal. The negative
feedback loop is through U128Ato input pin 1 of U64,
through U84, U104, and Q0112 to CR114, and back to
U128A.
Delaying Sweep Generator
The Delaying Sweep Generator produces a linear ramp
waveform when gated by the Main Trigger Generator.
The Delaying Sweep ramp is displayed as the intensif ied
sweep of the Alternate display. The Delaying Sweep
ramp is also the time reference for the Delay Pick off
comparator.
The linear ramp waveform is produced by
charging a capacitor from a constant current source.
The slope of the ramp determ ines the time/division of the
displayed trace and the delay time set by the DELAY
TIME MULT dial.
Ramp Generator. When a trigger pulse is
received from the Main Trigger Generator , Q402 cuts off
and Q404 conducts, driving the base of 0410 high.
When Q410 turns on, Q412 turns off and the timing
current from 0436 starts to charge the tim ing capacitors
in a positive ramp. The tim ing current is determined by
the timing resistors in the em itter circuit of Q436 and the
reference voltage at the base of Q436. The reference is
set by the SWP CAL adjustment, R750. Integrated
circuit U752 is a unity-gain voltage follower. Diode
CR753 compensates for the base- emitter voltage drop of
Q436, Q494, and Q536. Transistor s Q446A and B and
0450 form a unity-gain voltage-follower for the delaying
sweep signal. The output of Q450 is attenuated by
divider R468, R469, and R458 before it is connected to
the Output Amplifier. T he output of Q450 also dr ives the
Sweep Stop Comparator, the Baseline Stabilizer, and the
Delay Pick off.
Sweep Stop Comparator. Transistors Q462
and 0464 control the sweep length. W hen the Delaying
Sweep ramp exceeds the voltage on the base of 0464,
Q462 turns off and Q464 provides a positive pulse to end
the Auxiliary Gate and produce the Hold-off Start pulse.
3-3
Theory of operation-7B92A
Auxiliary Gate Generator. When the trigger
pulse cuts off Q402, the negative pulse at the base of
Q472 causes the Auxiliary Gate at the emitter of 0474 to
go high. The Auxiliary Gate signal remains high until a
positive pulse from the Sweep Stop Comparator turns
Q468 on, ending the Auxiliary Gate. The positive pulse
from the Sweep Stop Comparator is also coupled
through CR811 to the base of 0812 and results in a
positive Hold-off Start pulse at the collector of Q816.
The Hold-off Start pulse is connected to the Horizontal
Logic which starts the Trigger Disable pulse.
Baseline Stabilizer. W hen the Trigger Disable
pulse resets the output of the Main Trigger Generator to
a low level, Q402 turns on and Q404 turns off. T he low
level on the base of 0410 turns on Q412, which
discharges the timing capacitors. With the timing
capacitors discharged, the Bas eline Stabilizer maintains
a constant level from which the ramp begins. The output
of Q450 is compared with the reference on the base of
Q420A. If the output is less than the reference, Q430
charges the timing capacitors through CR434 until the
output and reference voltages are equal. If the output is
greater than the reference, Q430 conduc ts less and the
timing capacitors discharge through Q412. When the
two voltages are equal, the currents through Q430 and
Q436 equal the current through Q412, holding the
voltage on the timing capacitors constant.
Delay Pick off. The Delay Pick off allows a
continuously variable delay of 0 to 9.8 times the
TIME/DIV OR DLY TIME control setting between the
start of the Delaying Sweep Generator and the start of
the Delayed Sweep Generator. The Delay Pick off uses
the Delaying Sweep ramp as a time reference by
comparing the ramp voltage to a voltage set by the
DELAY TIME MULT. When the ram p voltage exceeds
the DELAY TIME MULT voltage, a pulse is coupled to
the Delayed Trigger Generator.
When a trigger pulse is received from the Main
Trigger Generator, Q493 cuts off and Q492 conducts.
As the Delaying Sweep ramp exceeds the
voltage on the base of Q482B, Q482A cuts off and
Q482B conducts. The differential signal produced when
both Q492 and Q482B conduct is coupled to the Delayed
Trigger Generator.
Transistor Q494 is a constant current source
maintaining a constant voltage across DELAY TIME
MULT, R490. When the time bas e is set for a Normal
sweep mode, S490 grounds R499 and cuts off Q494.
With Q494 cut off, the voltage at the input (pin 3) of
voltage follower U492 is zero. Therefore, in the Normal
sweep mode, the Delay Pick off outputs a pulse as s oon
as the Main Trigger pulse cuts off Q493
.
Delayed Trigger Generator
When the Delayed Triggering LEVEL control is
in the RUNS AFTER DELAY TIME position, or the time
base is in the Normal Sweep mode, the Delayed Trigger
Generator outputs a trigger pulse to the Delayed Sweep
Generator as soon as the Delay Pick off pulse is
received. If the Delayed Triggering LEVEL contro l is out
of the RUNS AFTER DELAY TIME detent, and the time
base is in the Delayed or Alternate sweep modes, the
Delayed Trigger Generator is enabled by the Delay Pick
off pulse. The Delayed Trigger Gener ator then operates
much the same as the Main Trigger Generator.
External Source (SN B070000-above). The
external trigger signal is connected to the Delayed
Trigger Generator through the DLY’D TRIG IN
connector, J200. The input impedance at J200 can be
set to either 1 megohm or 50 ohms by TERM switch
S205.
Input signals below 100 megahertz are coupled
through Q222A and Q224 to pin 3 of U274. Input signals
above 100 megahertz are coupled through C220 to pin 4
of U274. (Pins 3 and 4 of U274 are internally
connected.)
External Source (SN B069999-below). The
external trigger signal is connected to the Delayed
Trigger Generator through the DLY’D TRIG IN
connector, J200. The input impedance at J200 can be
set to either 1 megohm or 50 ohms by TERM switch
S205.
Input signals below 100 megahertz are coupled
through Q222A and Q224 to pin 3 of U264. Input signals
above 100 megahertz are coupled through C226 to pin 4
of U264. (Pins 3 and 4 of U264 are internally
connected.)
Internal Source (SN B070000-above). When
the Delayed Trigger Generator is using an internal trigger
signal, the internal trigger signal from the vertical channel
of the oscilloscope is connected to the Delayed Trigger
Generator through J250. Signals above 30 k ilohertz are
coupled through C246 to pin 14of U274. (Pins 13 and
14of U274 are internally connected.) Signals below 30
kilohertz are connected to amplif ier U244 through R257.
The output of amplif ier U244 is coupled, along with the
offset from the LEVEL control, to pin 1 of U274.
Internal Source (SN B069999-below). When
the Delayed Trigger Generator is using an internal trigger
signal, the internal trigger signal from the vertical channel
of the oscilloscope is connected to the Delayed Trigger
Generator through J250. Signals above 30 k ilohertz are
coupled through C246 to pin 14 of U264. (Pins 13 and
14 of U264 are internally connected.) Signals below 30
kilohertz are connected to amplif ier U240through R253.
The output of amplif ier U240 is coupled, along with the
offset from the LEVEL control, to pin 1 of U264.
3-4
REV B, JUN 1978
Theory of Operation-7B92A
Slope Selector Amplifier (SN B070000-
above). Amplifier U322 selects the slope of the input
waveform on which triggering occ ur s . If pin 3 is high, the
time base will trigger on the negative slope. If pin 3 is
low, the time base will trigger on the positive slope. (The
internal signal is inverted by U274 before being applied
to U322.)
Slope Selector Amplifier (SN B069999-below).
Amplifier U284 selects the slope of the input waveform
on which triggering occurs. If pin 6 is high, the time base
will trigger on the negative slope. If pin 6 is low, the tim e
base will trigger on the positive slope. (The internal
signal is inverted by U264 before being applied to U284.)
The push-pull outputs (pins 7, 8 and 9, 10) of
U284 are connected to the inputs (pins 3 and 13) of
U304.
Output Amplifier (SN B069999-below).
Amplifier U304 provides a final gain stage befor e driving
the trigger output tunnel diodes. The push-pull outputs
of U304 (pins 8 and 9) drive the emitter and base of
Q312. Transistor Q312 c onverts the push-pull output of
U304 to a single ended output to drive tunnel diode
CR314. As the emitter base voltage of Q312 increases,
the current through CR314 increases. W hen the current
through CR314 reaches 10 milliamperes, it switches to
its high level. The fast rise of CR314 is coupled through
C341 to the Arming Tunnel Diode, CR372, which also
switches high. The high level at the anode of CR372
increases the current through the Gate Tunnel Diode,
CR376. The rise of CR314 is also coupled through C324
(and a 1 nanosecond delay line) to CR376. The
combination of the added current from CR372 and the
pulse from CR314 (1 nanosec ond later) s witches CR376
high. The fast-rise pulse from CR376 is connected to
the Delayed Sweep Generator.
Trigger Disable (SN B070000-above). At the
end of each Delaying Sweep, the Horizontal Logic
produces a Trigger Disable pulse to allow enough time
for the sweep generator to reset and stabilize before
another trigger pulse starts the next sweep. The Trigger
Disable pulse resets the main trigger output to a low
level. At the same time U820 pulls pin 3 AUTO high.
With pin CF low, and pin AH high (see Delaying Sweep
<> ),comparatorsQ402Q404, Q492-Q493 and Q482AQ482B reset to Q404, Q0492, and Q482B off and Q402,
Q493, and Q482A conducting. This resets the
differential comparators Q362, Q366 and Q342, Q346 to
Q362, Q342 off and Q366, Q346 on. W ith Q342 off , pin
2 of U322 is high resetting the Delayed Trigger.
Trigger Disable (SN B069999-below). At the
end of each Delaying Sweep, the Horizontal Logic
produces a Trigger Disable pulse to reset the Arming
and Gate Tunnel Diodes to a low level and allow enough
time for the sweep generator to reset and stabilize before
another trigger Theory of Operation-7B92A pulse starts
the next sweep. The Trigger Disable signal is connec ted
to the base of Q332. When the T rigger Disable pulse is
high, 0348 and 0358 turn on, removing the bias cur rent
from CR372 and CR376, respectively. With no bias
current, CR372 and CR376 both reset to a low level.
Delayed Triggering Modes (SN B070000-
above). The Delayed Trigger Generator operates in one
of three modes; Normal, Delayed (Runs After Delay
Time), and Delayed (Triggerable After Delay Time).
In the Normal mode (both Time/Division knobs
locked together), Q328 is cut off and U322 inputs are
disabled. When the Delay Pick off pulse goes high,
U322 gate goes high.
In the Delayed (Runs After Delay Time) mode,
the Delayed Trigger Generator operates the sam e as in
the Normal mode. However, Q328 is cut off by the
RUNS AFTER DELAY TIME switch, S250; but, there is a
delay between the time the Delaying Sweep Generator
starts and the Delay Pick off is generated. The Delay
Pick off pulse c aus es the gate of U322 to go high. In the
Delayed (Triggerable After Delay Time) mode, U322 is in
the Triggerable mode. Now the Delay Pick off pulse
enables U322 to trigger when a trigger input signal is
present.
Delayed Triggering Modes (SN B069999below). The Delayed Trigger Generator operates in one
of three modes; Normal, Delayed (Runs After Delay
Time), and Delayed (Triggerable After Delay Time).
In the Normal mode (both Time/Division knobs
locked together), Q334 is cut of f and U284 and U304 are
both disabled. Additional current for CR372 and CR376
is supplied by R335 and R337. When the Delay Pick off
pulse cuts off CR381, CR372 switches high. The fast
rise of CR382 is coupled through R373 and C373 to
CR376, which also switches high.
In the Delayed (Runs After Delay Time) mode,
the Delayed Trigger Generator operates the sam e as in
the Normal mode. However, Q334 is cut off by the
RUNS AFTER DELAY TIME switch, S250, and there is a
delay between the time the Delaying Sweep Generator
starts and the Delay Pick off pulse cuts off CR381.
In the Delayed (Triggerable After Delay Time)
mode, the entire Delayed Trigger Generator is operating,
but the Arming Tunnel Diode, CR372, is held low by the
Delay Pick off signal until after the delay time.
When the Delay Pick off pulse cuts off CR381,
the next positive pulse from CR314 sets CR372 high and
1 nanosecond later sets CR376 high.
3-5
Delayed Sweep Generator
The Delayed Sweep Generator produces a linear
ramp waveform when gated by the Delayed Trigger
Generator. The Delayed Sweep ramp is displayed as
the Normal or Delayed Sweep trace.
The linear ramp waveform is produced by
charging a capacitor from a constant current source.
The slope of the ramp determ ines the time/division of the
displayed trace.
Ramp Generator. When a trigger pulse is
received from the Delayed Trigger Generator, Q502 cuts
off and Q504 conducts, driving the base of Q510 high.
When Q510 turns on, Q512 turns off and the timing
current from Q 536 starts to charge the tim ing capacitors
in a positive ramp. The tim ing current is determined by
the timing resistors in the em itter circuit of Q536 and the
reference voltage at the base of 0536. Transistors
Q546A and B and Q550 form a unity-gain, voltage
follower for the delaying sweep signal. The output of
Q550 is attenuated by divider R557 and R558 before it is
connected to the Output Amplifier. At 0.5
nanosecond/division, the Delayed Sweep ramp is not
attenuated. The output of Q550 also drives the Sweep
Stop Comparator and the Baseline Stabilizer.
Sweep Stop Comparator. Transistors Q562
and Q564 control the sweep length. W hen the Delayed
Sweep ramp exceeds the voltage on the base of Q564,
Q562 turns off and Q564 provides a positive pulse to end
the Main Gate.
Main Gate Generator. When the trigger pulse
cuts off 0502, the negative pulse at the base of Q572
causes the Main Gate at the emitter of 0584 to go high.
The Main Gate signal rem ains high until a positive pulse
from the Sweep Stop Comparator turns Q568 on, ending
the Main Gate.
Baseline Stabilizer. W hen the Trigger Disable
pulse resets the output of the Delayed Trigger Generator
to a low level, 0502 turns on and Q504 turns off. The low
level on the base of Q510 turns on Q512, which
discharges the timing capacitors. With the timing
capacitors discharged, the Bas eline Stabilizer maintains
a constant output level.
The output of 0550 is compared with the
reference on the base of Q522. If the output is less than
the reference, 0530 charges the timing capacitors
through CR534 until the output and reference voltages
are equal. If the output is greater than the reference,
Q530 conducts less and the timing capac itors discharge
through 0512. When the two voltages are equal, the
currents through Q530 and Q536 equal the current
through Q512, which holds the voltage on the timing
capacitors constant.
Auxiliary Y and Z Axis. The Aux. Y and Aux. Z
outputs allow the 7B92A to control the trace separ ation,
intensity, and contrast of the Delaying Sweep trace when
the time base is operating in the Alternate mode.
The Aux. Y, Z Inhibit signal disables both outputs
except when the time base is in the Intensified or
Alternate modes, and the delaying sweep is being
displayed. The CONTRAST control is active when
operating in the ALT mode, and the Delayed Sweep
Generator is running. The CONTRAST control varies
the brightness of the intensified zone.
Horizontal Logic
The Horizontal Logic controls the different sweep
modes and functions of the time base (e.g., sweep
display, hold-off, auto-trigger, single-sweep, etc.). The
Horizontal Logic also generates control signals for the
oscilloscope mainframe.
Sweep Control IC. The Sweep Control IC,
U820, generates most of the control signals us ed in the
7B92A.
When the MAIN T RIGGERING MODE is set to
AUTO, the Sweep Control IC supplies a triggering gate
to the Delaying Sweep Generator when the Main Trigger
Generator is not triggered. The auto triggering circuit
starts to operate if pin 19 of U820 is held low by S100
and an Auto Disable Pulse has not been received for
about 40 milliseconds.
When the MAIN T RIGGERING MODE is set to
SINGLE SWEEP, the Sweep Control IC allows one r amp
to be displayed. The Trigger Disable signal then
prevents another ramp from running until U820 is
manually reset by pressing the SINGLE SWEEP-RESET
button.
Lockout. When the 7B92A is used in a fourchannel oscilloscope mainframe in an Alternate mode
with another time base, a Lockout signal prevents the
7B92A from running while the other time base is being
displayed. The Lockout signal is coupled through the
Lockout Amplifier, 0802, 0804, and 0806 to pin 18 of
U820. The Lockout signal drives the Trigger Disable
output (pin 17) high to reset and hold the trigger
generator outputs low.
The 7B92A also outputs a Hold off pulse (pin B4)
to the oscilloscope mainframe which controls the
Lockout pulse to the other time base. The Hold off pulse
occurs at the end of each Delaying Sweep ramp when
the time base is in the Normal, Intensified or Delayed
Sweep mode. When the time base is in the Alternate
mode, the Hold off pulse occurs after the Delayed Sweep
ramp is displayed.
REV A, JUN 1978
3-6
Theory of Operation-7B92A
NOTE
When operating in the Intens ified mode,
the intensified zone is controlled by the
intensity of the test oscilloscope instead
of the 7B92A CONTRAST and IN
TENSITY controls.
Sweep Display. The Sweep Display flip-flop,
U856A, determines which sweep ramp is displayed.
When the tim e base is in the Normal or Delayed Sweep
mode, S800 connects R861 and R862 to ground. W ith
the Set input of U856A low, pin 5 is held high and pin 6 is
held low. The high level on pin 5 of U856A inhibits the
Delaying Sweep signal at the Output Amplifier. T he low
on pin 6 allows the Delayed Sweep signal to be coupled
through the Output Amplifier to the oscilloscope
mainframe.
If the time base is in the Alternate mode, the Set,
J, and K inputs of U856A are all high. The end of each
Aux. Gate pulse toggles U856A, which allows alternate
display of the Delaying and Delayed Sweep ramps.
When the tim e base is in the Intensified Sweep
mode, pin 15 of U856A is held low by pin 4of U635,
clearing flip flop U856. With pin 5 of U856A low, the
Delaying Sweep signal is coupled to the output amplifier.
Output Amplifier
The Output Amplifier connects the sweep signal
to the oscilloscope mainframe and provides an offset
voltage to position the trace on the graticule.
The Delaying Sweep and Delayed Sweep ramps
are connected to Q900 and Q910, respectively. The
Delaying Sweep Inhibit and Delayed Sweep Inhibit
signals saturate either Q902 or Q912 to prevent that
sweep ramp from being amplified and coupled to the
oscilloscope mainframe.
The POSITION control, R930, offsets the ramp
waveform to horizontally position the displayed trace.
Transistors Q926 and 0932 tur n on when contact 52 is
closed and the Delaying Sweep Inhibit signal is high.
These transistors pr ovide additional off set at fast
sweep rates.
Theory of Operation-7B92A The positioning
offset voltage and the selected sweep ramp are
connected to 0942 and Q952. The push-pull output is
connected to the oscilloscope mainframe through pins All
and Bl1.
The Auxiliary Sweep Amplifier, Q456 and Q458,
is a unity-gain amplifier that couples the Delaying Sweep
ramp to the mainframe. This signal may be connected to
the + Saw-tooth output of the oscilloscope mainframe.
Readout
The oscilloscope readout system in 7000-Ser ies
Oscilloscopes provides alphanumeric display of
information encoded by the plug-in units. This dis play is
presented on the crt on a time-shared basis with the
waveform display.
The oscilloscope readout system produces a
pulse train consisting of 10 negative-going pulses called
times lots. Each pulse represents a possible character in
a readout word, and is assigned a time-slot number
corresponding to its position in the word. Each tim e-slot
pulse is directed to 1 of 10 lines, labeled TS-1
throughTS10 (time slots 1 through 10), which are
connected to the vertical and horizontal plug-in
compartments. Two output lines, row and column, are
connected from each channel (two channels per plug-in
compartment) back to the oscilloscope readout system.
Data is encoded on the output lines either by
connecting resistors between the output lines and the
time-slot input lines, or by generating equivalent currents.
The resultant output is a sequence of analog current
levels on the row and column output lines. The row and
column current levels address a c haracter matrix during
each time-slot; thus, selecting a character to be
displayed or a special instruction to be followed.
The encoding resistors are selected by the
TIME/DIV OR DLY TIME and DLY’D Time/Division
switches. Table 3-1 lists the resistors that control the
readout characters and functions.
3-7
Theory of Operation-7B92A
Table 3-1.
7B92A READOUT CHARACTER SELECTION
.
Characters Time-Slot Description Encoded
Channel
(Delaying Sweep)
Decimal TS-1 Determines decimal magnitude (number
of zeros displayed or prefix change
information).
Uncalibrated (>) TS-3 Indicates calibrated or uncalibrated sweep
rates
1, 2, 5 TS-4 Scaling R763, R764
m, /i, n, p TS-8 Defines the prefix which modifies the units
of measurement
s (seconds) TS-9 Defines the unit of measurement R784, R787 R789, R796
3-8
R761, R762
R781
R782 R792
R785
R765, R766
R783, R786
Channel
(Delayed Sweep)
R771, R772
R791
R773, R774
R793
R775, R776
R788, R795
Section 4-7B92A
MAINTENANCE
This section of the manual contains inform ation for performing preventive maintenanc e, troubleshooting, and corrective
maintenance for this instrument.
PREVENTIVE MAINTENANCE
Preventive maintenance consis ts of cleaning, visual inspection, lubric ation, etc. Preventive maintenance perf ormed on a
regular basis may prevent instrum ent breakdown and will improve the reliability of the instrument. The severity of the
environment to which this instrument is subjected determines the frequency of maintenance. A convenient time to perform
preventive maintenance is preceding adjustment of the instrument.
CLEANING
This instrument should be cleaned as often as operating
conditions require. Accumulation of dirt on components
acts as an insulating blanket and prevents eff icient heat
dissipation that can cause overheating and component
breakdown.
CAUTION
Avoid the use of chemical cleaning
agents which might damage the plastics
used in this instrument. In particular,
avoid chemicals that contain benzene,
toluene, xylem, acetone, or similar
solvents.
Exterior
Loose dust accumulated on the front panel can be
removed with a soft cloth or small brush. Dirt that
remains can be removed with a soft cloth dam pened with
a mild detergent and water solution. Abras ive cleaners
should not be used.
Interior
Dust in the interior of the instrum ent should be removed
occasionally due to its electrical conductivity under highhumidity conditions. The best way to clean the interior is
to blow off the accumulated dust with dry, low pressure
air. Remove any dirt that remains with a soft brush or a
cloth dampened with a mild detergent and water solution.
A cotton-tipped applicator is useful f or c leaning in narrow
spaces.
Switch Contacts
Switch contacts and pads are designed to operate dry for
the life of the switch. However, as the switches are not
sealed, dust attracted to the contact area may cause
switch contacts to become electrically noisy. Cleaning
may be accomplished by flushing the contact area with
isopropyl alcohol or kelite (1 part kelite to 20 parts water).
Do not use chemical cleaning agents that leave a film or
that might damage plastic parts. Do not use cotton
swabs or similar applicators to apply cleaning agents, as
they tend to snag and leave strands of cotton on switch
contacts. Should it become necessary to remove a
switch for replacement or cleaning, refer to Component
Removal and Replacement in this section.
VISUAL INSPECTION
This instrument should be inspected occasionally for
such defects as brok en connections, improperly seated
semiconductors, damaged circuit boards, and heatdamaged parts.
The corrective procedure for most visible defects if
obvious; however, particular care must be taken if heatdamaged components are found. Overheating usually
indicates other trouble in the instrument; therefore, it is
important that the cause of overheating be corrected to
prevent recurrence of the damage.
LUBRICATION
Generally, there are no components in this instrument
that requires a regular lubrication progr am during the life
of the instrument.
4-1
Cam Switch Lubrication
In most cases, factory lubrication should be adequate for
the life of the instrument. However, if the switch has
been disassembled for replacement of switch s ub-parts,
a lubrication kit containing the necessary lubricating
materials and instructions is available through any
Tektronix Field Office. Order Tektronix Part 003-0342-
01. General Electric Versilubee silicone grease should
be applied sparingly so that the lubricant does not get on
the contacts. Refer to Fig. 4-1 for lubrication
instructions.
SEMICONDUCTOR CHECKS
Periodic checks of the semiconduc tors in this instrument
are not recommended. The best check of
semiconductor performance is actual operation in the
instrument.
More details on checking semiconductor operation are
given under Troubleshooting.
ADJUSTMENT AFTER REPAIR
After any electrical component has been replaced, the
adjustment of that particular circuit should be checked,
as well as the adjustment of other closely-related circuits.
The Performance Check and Adjustment procedure in
this manual provides a quick and convenient means of
checking instrument operation. In some cases, minor
troubles may be revealed or corrected by adjustment.
Maintenance-7B92A
Fig. 4-1. Lubrication procedure for typical cam switch.
4-2
Maintenance-7B92A
TROUBLESHOOTING
The following inform ation is provided to help troubleshoot this instrument. Information contained in
other sections of this m anual should be used along with the following information to aid in locating the
defective component. An understanding of the circuit operation is very helpful in locating troubles,
particularly where integrated circuits are used.
TROUBLESHOOTING AIDS
Diagrams
Circuit diagrams are given on f old out pages in
Section 8. The component number and elec trical value
of each component in this instrument is shown on the
diagrams.
Components that are mounted on circuit boards
are outlined on the diagrams with a heavy black line.
Voltages and Waveforms
Typical operating voltages are shown on the
diagrams. Voltage Conditions given on the diagram
page indicate the test equipment used and the frontpanel control status necessary to obtain the given
voltages.
Typical operating waveforms are shown nex t to
the diagram where they were measured. Each
waveform is numbered to locate the point on the diagram
where the waveform was taken. Waveform Conditions
given on the diagram page list the test equipment used
and the front panel control status necessary to obtain the
given waveform.
Circuit Board Illustrations
Circuit board illustrations are shown on the f old
out page preceding the associated diagram. Eac h board
mounted electrical component is identified by its circuit
number, as are interconnecting wires and connectors.
Figure 8-1, in the front of the diagrams sec tion,
shows the location and assembly number of each circ uit
board in this instrument.
Diode Color Code
The cathode end of each glass- encased diode is
indicated by a stripe, a series of stripes, or a dot. The
cathode and anode ends of metal-encased diodes are
identified by the diode symbol marked on the c ase. For
most silicon or germanium diodes with a series of
stripes, the color code identifies the f our signific ant digits
of the JEDEC or vendor number using the resistor colorcode system.
Wiring Color Code
Insulated wire and cable used in this instrument
is color-coded to facilitate circuit tracing.
Semiconductor Basing
Figure 4-2 illustrates the basing configurations
for all semiconductors used in this instrument. Some
plastic case transistors have lead conf igurations that do
not agree with those shown here. If a replacement
transistor is made by a different manufacturer than the
original, check the manufacturer’s basing diagram. All
transistor sockets in this instrument are wired for the
standard basing used for metal-case transistors.
Inter-Board Pin Connector Identification
The inter-board pin connector sockets are
installed on circuit boards in groups of 5 sockets ( as in
Fig. 4-3). Socket number 1 is indexed on the circuit
board with either a triangular mark or the number 1.
Each group of sock ets is identif ied by its J (jac k ) num ber
etched on the circuit board. The J num bers correlate to
the J (jack) and P (plug) circuit numbers on the
schematic diagrams.
Switch Cam Identification
Switch cam numbers shown on diagrams
indicate the position of each cam in the com plete switch
assembly. The switch cam s are numbered f rom front to
rear.
Multi-pin connectors mate with groups of pins
soldered to circuit boards. Pin num ber 1 is indexed with
atriangular mark on the circuit board and m olded on the
holder of the multi-pin connector , as shown in Fig. 4-4.
Each group of pins is identified by its corresponding J
number etched on the circuit board. T he J num bers, on
the circuit boards, correlate to the J and P component
numbers on the schematic diagrams.
Interface Connector Pin Locations
The Interface circuit boar d couples the plug-in unit
to the associated mainfram e (oscilloscope). Figure 4-5
identifies the pins on the interface connector as shown
on Interface Connectors and Power Supply diagram 8 in
the Diagrams section.
Fig. 4-4. End-lead multi-pin connector assembly
REV B, JUN 1978
Fig. 4-5. Location of pin numbers of Interface
connector.
4-5
Maintenance-7B92A
Performance Check and Adjustment
The Performance Check and Adjustment
procedure, given in Section 5 of this manual, provides a
quick and convenient means of checking instrument
operation. In some cases, minor troubles may be
revealed or corrected by adjustment.
TROUBLESHOOTING EQUIPMENT
The following equipment, in addition to that listed
in the Performance Check and Adjustment section, is
useful for troubleshooting.
Transistor Tester.
Description: Dynamic-type tester.
Purpose: Test semiconductors.
Recommended T ektronix types: 576 Curve Tracer,
577/177 Curve Tracer system, 7CT 1N Curve T racer unit
and a 7000-Series Oscilloscope system, or a 5CT1N
Curve Tracer unit and a 5000-Series Oscilloscope.
Multimeter.
Description: Voltmeter, 10 megohm input impedance and
a range from 0 to at least 50 volts dc; accuracy, within
0.1%. Ohmmeter, 0 to 20 megohms. Test probes
should be insulated to prevent accidental shorting.
Purpose: Check voltage and resistance.
Test Oscilloscope.
Description’ Frequency response, dc to 100
megahertz minimum; deflection factor, 5 millivolts to 5
volts/division. A 10OX, 10 megohm voltage probe
should be used to reduce circuit loading.
Purpose: Check operating. waveforms.
TROUBLESHOOTING TECHNIQUES
The following troubleshooting procedure is
arranged to check the simple trouble possibilities
before proceeding with extensive troubleshooting. The
first few checks ensure proper connection, operation,
and adjustment. If the trouble is not located by these
checks, the rem aining steps aid in locating the defective
component. When the defective component is located, it
should be replaced using the replacement procedure
given under Corrective Maintenance
Troubleshooting Procedure
1.Check Control Settings. Incorrect control
settings can indicate a trouble that does not exist. If
there is any question about the correct function or
operation of any control, see the Operating Instructions,
Section 2.
2.Check Associated. Equipment. Before
troubleshooting, check that the equipm ent used with this
instrument is operating cor rectly. Check that the signal
is properly connected and that the interconnecting cables
are not defective. Also, check the power source. If the
trouble persists, the time-base unit is probably at fault.
3.Visual Check. Visually check the portion
of the instrument in which the trouble is located. Many
troubles can be located by visible indications such as
unsoldered connections, broken wires, damaged circuit
boards, damaged components, etc.
4.Check Instrument Adjustment. Check
the adjustment of this ins trument, or the affected cir cuit if
the trouble appears in one circuit. The apparent trouble
may be the result of misadjustment. Complete
adjustment instructions are given in the Performance
Check and Adjustment section.
5.Isolate Trouble to a Circuit. To isolate
trouble to a circuit, note the trouble symptom. The
symptom often identifies the c irc uit in which the tr ouble is
located. When trouble symptoms appear in more than
one circuit, check the af fected circuits by taking voltage
and waveform readings. Incorrect operation of all
circuits often indicates trouble in the power supply.
Check first f or corr ect voltages of the individual supplies.
However, a defective component elsewhere in the
instrument can appear as a power-supply trouble and
may also affect the operation of other circuits.
After the defective circuit has been located, proceed with
steps 6 and 7 to locate the defective component(s).
6.Check Voltages and Waveforms. Often
the defective component can be located by checking for
the correct voltages and waveforms in the circuit. Ref er
to the diagrams section at the rear of the manual for
typical voltages and waveforms.
NOTE
Voltages and waveforms given on the
diagrams are not absolute and may v ary
slightly between instruments. To obtain
operating conditions similar to those
used to take these readings, see the
voltage
4-6
REV A, JUN 1978
Maintenance-7B92A
and waveforms page adjacent to each
schematic diagram. Note the
recommended test equipment, frontpanel control settings, voltage and
waveform conditions, and test
equipment cable connection instructions.
7.Check Individual Components. The
following procedures describe methods for checking
individual components. Two-lead components that are
soldered in place are best checked by first disconnecting
one end.
This isolates the meas urement from the effects
of surrounding circuitry.
CAUTION
To avoid damage, disconnect the power
source before removing or replacing
semiconductors.
TRANSISTORS. The best check of transistor
operation is actual performance under operating
conditions. A transistor can be most ef fectively checked
by substituting a new component or one that has been
checked previously. However, be sure that circuit
conditions are not such that a replacement transistor
might also be damaged. If s ubstitute transistors are not
available, use a dynamic tester. Static-type testers ar e
not recommended, since they do not check operation
under simulated operating conditions.
INTEGRATED CIRCUITS. Check with a
voltmeter, test oscillosc ope, or by direct substitution. A
good understanding of circuit operation is desirable when
troubleshooting circuits using IC. Use care when
checking voltages and waveform s around the IC so that
adjacent leads are not shorted together. A convenient
means of clipping a test pr obe to the 14and 16-pin IC is
with an IC test clip. This device also serves as an
extraction tool. The lead configuration for the
semiconductors used in this instrument is shown on a
pullout page in the front of the diagrams section.
DIODES. A diode can be checked f or an open
or shorted condition by measuring the resistance
between terminals with an ohmmeter scale having a low
internal source current, such as the R X 1K scale. The
resistance should be very high in one direction and very
low when the meter leads are reversed.
The cathode end of each glass- encased diode is
indicated by a stripe, a series of stripes, or a dot. The
cathode and anode ends of metal-encased diodes are
identified by the diode symbol marked on the c ase. For
most silicon or germanium diodes with a series of
stripes, the color code identifies the f our signific ant digits
of the JEDEC or vendor number using the resistor colorcode system.
RESISTORS. Check resistors with an
ohmmeter. See the Replac eable Electrical Parts list for
the tolerance of the resistors used in this instrument.
Resistors normally do not need to be replaced unless the
measured value varies widely from that specified.
INDUCTORS. Check for open inductors by
checking continuity with an ohmmeter. Shorted or
partially shorted inductors can usually be found by
checking the waveform response when high-frequency
signals are passed through the circuit. Partial shorting
often reduces high frequency response.
CAPACITORS. A leaky or shorted capacitor c an
usually be detected by checking resistance with an
ohmmeter on the highest scale. Do not exceed the
voltage rating of the capacitor. The resistance reading
should be high after initial charge of the capacitor. An
open capacitor can best be detected with a capacitance
meter or by checking that the capacitor passes ac
signals.
8.Repair and Adjustment.
parts are located, follow the replacement procedures
given in Corrective Maintenance. Be sure to check the
performance of any circuit that has been repaired or had
any electrical components replaced.
If any defective
Do not use an ohmmeter s cale that has
a high internal current. High currents
may damage the diode.
REV A, JUN 1978
CAUTION
4-7
CORRECTIVE MAINTENANCE
Corrective maintenance consists of component replacement and instrument repair. Special
techniques required to replace components in this instrument are given here.
Maintenance-7B92A
OBTAINING REPLACEMENT PARTS
All electrical and mechanical part replacements
can be obtained through your Tektronix field Office or
representative. However, many of the standard
electronic components can be obtained locally in less
time than is required to order them from Tektronix, Inc.
Before purchasing or ordering replacement parts , check
the parts list for value, tolerance, rating, and description.
NOTE
When selecting replacement parts,
remember that the physical size and
shape of a component may affect the
performance of the instrument,
particularly at high frequencies. All parts
should be direct replacements unless a
different component will not adversely
affect instrument performance.
Some parts are manufactured or selected by
Tektronix, Inc. to satisfy particular r equirements, or are
manufactured to specifications for Tektronix, Inc. Most
of the mechanical parts used in this instrument have
been manufactured by Tektronix Inc. To determine the
manufacturer of parts, refer to parts lis t, Cros s Index Mfr .
Code Number to Manufacturer.
When ordering replacement parts from Tektronix,
Inc., include the following information: 1. Instrument
type.
2. Instrument serial number.
3. A description of the part (if electrical, include
circuit number).
4. Tektronix part number.
SOLDERING TECHNIQUES
WARNING
To avoid electrical shock, disc onnect the
instrument from the power source befor e
soldering.
The reliability and accuracy of this ins trument can
be maintained only if proper soldering techniques are
used when repairing or replacing parts. General
soldering techniques, which apply to maintenance of any
precision electronic equipment, should be used when
working on this instrument. Use only 60/40 rosin-core,
electronic grade solder. The choice of soldering iron is
determined by the repair to be made. When soldering on
circuit boards, use a 15to 40-watt pencil-type soldering
iron with a 1/8-inch wide wedge-shaped tip. Keep the tip
properly tinned for best heat transfer to the solder joint.
A higher wattage soldering iron may separate the wiring
from the base material. Avoid excess ive heat; apply only
enough heat to remove the component or to make a
good solder joint. Also, apply only enough solder to
make a firm solder joint; do not apply too much solder.,
CAUTION
All circuit boards, except the Readout
circuit board, in this instrument ar e multilayer type boards with a conductive
path(s) laminated between the top and
bottom board layers. All soldering on
these boards should be done with
extreme care to prevent breaking the
connections to the center conductor(s);
only experienced maintenance
personnel should attempt repair of thes e
boards.
For metal terminals (e.g., switch terminals,
potentiometers, etc.), a higher wattage-rating soldering
iron may be required. Match the soldering iron to the
work being done. For example, if the component is
connected to the chassis or other large heat-radiating
surface, it will required a 75-watt or larger soldering iron.
The following techniques should be used to
replace a component on a circuit board: 1. Grip the
component lead with long-nose pliers.
1.Touch the soldering iron to the lead at the
solder connection. Do not lay the iron directly on the
board, as it may damage the board.
REV A, JUN 1978
2.When the solder begins to melt, gently pull the
lead out. If unable to pull out the lead without using
force, try removing the other end of the component as it
may be more easily removed.
4-8
Maintenance-7B92A
NOTE
The reason some component leads ar e
troublesome to remove is due to a bend
placed on each lead during the
manufacturing process. The bent leads
hold components in place during a
process that solders many components
at one time.
If a component lead is extremely difficult to
remove, it may be helpful to straighten the leads on the
back side of the board with a sm all screwdriver or plier s
while heating the soldered connection.
Use only enough heat to remove the component
lead without removing the solder from the board. If it is
desired to remove solder from a circuit-board hole for
easier installation of a new component, a solderremoving wick should be used.
3.Bend the leads of the new component to fit the
holes in the board. If the component is replaced while
the board is mounted in the instrument, cut the leads so
they will just protrude through the board. Insert the leads
into the holes so the component is firmly seated against
the board (or as positioned originally). If it does not seat
properly, heat the solder and gently press the component
into place.
4.Touch the iron to the connection and apply a
small amount of solder to mak e a firm solder joint. To
protect heat-sensitive components, hold the lead
between the component body and the solder joint with a
pair of long nose pliers or other heat sink.
5.Clip any excess lead protruding through the
board (if not clipped in step 3).
6.Clean the area around the solder connection
with a flux-removing solvent. Be careful not to remove
information printed on the board.
Component Removal and Replacement
WARNING
Disconnect the instrument from the
power source before replacing
components.
Semiconductors. Semiconductor devices used in this
instrument should not be replaced unless actually
defective. If removed from their sockets during routine
maintenance, return them to their original sockets.
Ferrite beads must be reinstalled on the proper leads.
REV A, JUN 1978
Unnecessary replacement may affect the calibration of
this instrument. When a semiconductor is replaced,
check the operation of the part of the instrument that
may be affected.
Replacement devices should be of the original type
or a direct replacement. Ins tall in the same manner as
the original. Figure 4-2 shows the lead configurations of
the semiconductor devices used in this instrument.
When replacing, check the manufacturer’s basing
diagram for correct basing.
Interconnecting Pins. Two methods of
interconnection are used in this instrument to connect
the circuit boards with other boards and components.
When the interc onnec tion is made with a coaxial cable, a
special end lead connector plugs into a socket on the
board. Other interconnections are made with a pin
soldered onto the board. Two types of mating
connectors are used for these interconnecting pins. If
the mating connector is mounted on a plug-on circuit
board, a special socket is solder ed into the board. If the
mating connector is on the end of a lead, an end-lead pin
connector that mates with the interconnecting pin is
used. The following information provides the
replacement procedure for the various interconnecting
methods.
COAXIAL-TYPE END-LEAD CONNECTORS.
Replacement of the coaxial-type end-lead connectors
requires special tools and techniques . Only exper ienced
maintenance personnel should attempt replacement of
these connectors. It is recommended that the c able or
wiring harness be replaced as a unit. For cable or wiring
harness part numbers, see the Replac eable Mechanical
Parts list. An alternate solution is to refer the
replacement of the defective c onnector to your Tektronix
Field Office or representative.
CIRCUIT BOARD PINS AND PIN SOCKETS. A
circuit board pin replacement kit, including necessary
tools, instructions, and replacement pins, is available
from Tek tronix, Inc. (Tektronix Part Num ber 040-0542-
00.)
CAUTION
The following procedures are
recommended for single-layer circuit
boards only. Pin and socket
replacement on multi-layer circuit boards
should be performed only by qualified
service personnel. Refer to your local
Tektronix Field Office or Service Center.
The pin sockets on the circuit boards are s oldered
to the board. To replace one of these sockets, first
unsolder the pin socket (use vacuum-type desoldering
tool to
4-9
Maintenance-7B92A
remove the excess s older). Then straighten the tabs on
the socket and remove it from the hole in the board.
Place the new socket in the circuit board hole and press
the tabs down against the board. Solder the tabs of the
socket to the circuit board; be careful not to get solder
into the socket.
END-LEAD PIN CONNECTORS. The pin
connectors used to connect the wires to the
interconnecting pins are clamped to the ends of the
associated leads. To replace damaged end-lead pin
connectors, remove the old connector from the end of
the lead and clamp the replacement connector to the
lead.
Some of the pin connectors are grouped together
and mounted in a plastic holder; the overall res ult is that
these connectors are rem oved and installed as a multipin connector. To provide correct orientation of this
multi-pi n connector when it is replaced, an arrow (or dot)
is stamped on the circuit board and a matching arrow is
molded into the plastic housing of the multi-pin
connector. Be sure that these arrows are aligned when
the multi-pin connector is replaced. If the individual endlead pin connectors are removed f rom the plas tic holder,
note the color of the individual wires for replacement.
Switches. Pushbutton and cam-type switches are
used in the 7B92A. Contact alignment and spacing is
critical to the operation of these switches. Therefore,
defective switches should be replaced as a unit or
repaired only by personnel experienced with switches of
this type. Your local Tektronix, Inc. Field Office can
provide additional repair information and instructions.
CAM SWITCH. The cam switch (TIME/DIV OR
DLY TIME and DLY’d Time/Division) consists of two
rotating cams and the associated contacts mounted on
the Interface and Readout boards. The cam switch can
be disassembled for cleaning, repair, or replacement;
however, it is recomm ended that the cam assembly be
removed from the instrument as a unit.
Remove the cam switch as follows: 1. Set the
TIME/DIV OR DLY TIME and DLY’D Time/Division
switches to 0.2 s and press the knob in.
2.Press and release the VARIABLE knob.
3.Loosen the set screws and remove the
VARIABLE and DLY’D Time/Division knobs.
4.Loosen the set screw in the clear plastic flange
(behind sub-panel) and remove the TIME/DIV OR DLY
TIME flange.
5.Remove the pin connector from the microswitch (in front of the cam ) to the front of the Interfac e
board.
6.Remove the Sweep Logic board.
7.Remove the 11 mounting screws holding the
cam switch assembly to the Interface board. Do not
remove the mounting screws from the Readout board.
8.Carefully lift the cam-switch assembly and
Readout board from the interconnecting pins at the rear
of the Readout board.
9.To replace the cam-switch assembly, reverse
the above procedure. Be sure to replace the TIME/DIV
OR DLY TIME flange and the DLY’D Time/Division knob
in the same position from which they were removed.
TRIGGERING SWITCHES. Remove the Main and
Delayed Triggering switches as follows:
1.Perform steps 1 through 4 of the cam-switch
replacement procedure.
2.Loosen the set screws and remove all front
panel knobs except the DELAY TIME MULT dial.
3. Unsnap the front panel from the top and bottom
of the sub-panel.
4.Remove the spring from the 7B92A release
latch.
5.Remove the 4 screws holding the sub-panel to
the chassis and pull the sub-panel forward.
6.Remove all necessary multi-pin and coaxial
connectors.
7.Remove the mounting screws from the desired
switch(es).
8.The MAIN TRIGGERING switches must be
removed as an assembly (all three switches) and then
disassembled further.
REV A, JUN 1978
4-10
Maintenance-7B92A
9. To replace the triggering switches , reverse the
above procedure.
ADJUSTMENT AFTER REPAIR
After any electrical component has been replaced,
the adjustment of that particular circuit should be
checked, as well as other closely related circuits. See
Adjustment section for a complete adjustment
procedure.
REPACKAGING FOR SHIPMENT
If the Tektronix instrument is to be shipped to a
Tektronix Service Center for service or repair, attach a
tag ;showing: owner (with address) and the name of an
individual at your firm that can be contacted, complete
instrument serial number, and a description of the
service required.
Save and re-use the package in which your
instrument was shipped. If the original pack aging is unfit
for use or not available, repackage the instrument as
follows:
1.Obtain a carton of corrugated cardboard having
inside dimensions of no less than six inches m ore
than the instrument dim ensions; this will allow for
cushioning. Refer to the following table for carton
test strength requirements.
2.Surround the instrument with polyethylene
sheeting to protect the finish of the instrument.
3.Cushion the instrument on all sides by tightly
packing dunnage or urethane foam between
carton and instrument, allowing three inches on all
sides.
4.Seal carton with shipping tape or industrial stapler.
SHIPPING CARTON TEST STRENGTH
Gross Weight (lb)Carton Test Strength (lb)
0-10200
10-30275
30-120375
120-140500
140-160600
REV C, JUN 1978
4-11
PERFORMANCE CHECK/CALIBRATION
Section 5-7B92A
PRELIMINARY INFORMATION
Calibration Interval
To ensure instrument accuracy, check the
calibration of the 7B92A every 1000 hours of operation,
or every six months if used infrequently. Before
complete calibration, thoroughly clean and inspect this
instrument as outlined in the Maintenance section.
Tektronix Field Service
Tektronix, Inc. provides complete instrument
repair and recalibration at local Field Service Centers
and the Factory Service Center. Contact your local
Tektronix Field Office or representative for further
information.
Using This Procedure
Outline. To aid in locating a step in the
Performance Check or Calibration procedure, outlines
are given preceding Part I-Perform ance Check and Part
II Calibration procedure.
Performance Check. The performance of this
instrument can be check ed without removing the covers
or making internal adjustments by performing Part
Performance Check. This procedure does not check
every facet of the instrument’s calibration; but is
concerned primarily with those portions of the instr ument
essential to measurement accuracy and correct
operation.
Calibration Procedure. Completion of Part II
Calibration procedure ensures that this instr um ent m eets
the electrical specifications given in the Operators
manual. Where possible, instrument performance is
checked before an adjus tment is made. For best over all
instrument performance when performing a complete
calibration procedure, make each adjustment to the
exact setting, even if the Check is within the allowable
tolerance.
Partial Calibration. A partial calibration is often
desirable after replacing com ponents, or to touch up the
adjustment of a portion of the instrument between major
recalibrations.
The Calibration procedure is divided into
Triggering Calibration and Sweep Calibration. To
perform a partial calibration, star t at the beginning of the
desired section. To prevent unneces sary rec alibration of
other parts of the instrument, readjust only if the
tolerance given in the Check part of the step is not met.
TEST EQUIPMENT REQUIRED
The following test equipment and accessories, or
their equivalents, are required for c omplete c alibration of
the 7B92A. Specifications given for the test equipm ent
are the minimum necessary for accurate calibration.
Therefore, some of the specifications listed here may
differ from the actual perf ormanc e capabilities of the test
equipment. All test equipment is assumed to be
correctly calibrated and operating within the listed
specifications. Detailed operating instructions for the test
equipment are not given in this procedure. Ref er to the
test equipment instruction manual if more information is
needed.
If only a Performance Check procedure is
performed, not all of the listed test equipment will be
required. Items used only for the Calibration pr ocedure
are indicated by footnote 1. The remaining pieces of
equipment are items common to both the Performance
Check and the Calibration procedure.
Special Calibration Fixtures
Special Tektronix calibration fix tures are used only
where they facilitate instrument calibration. These
special calibration fixtures are available from Tektronix,
Inc. Order by part number through your local Tekt ronix
Field Office or representative.
Calibration Equipment Alternatives
The Calibration procedure is based on the first
item of equipment given as an example of applicable
equipment. When other equipment is substituted,
control settings or calibration setup may need to be
altered slightly to meet the requirements of the subs titute
equipment. If the exact item of test equipment is not
available, first check the Minimum Specifications column
in Table 5-1 carefully to see if any other equipment is
available that might suffice.
REV A, JUN 1978
5-1
Table 5-1.
TEST EQUIPMENT
Description Minimum Specifications Purpose
Performance Check/Calibration-7B92A
Examples of Applicable
Test Equipment
1. Oscilloscope Bandwidth 500 MHz Used throughout procedure
to provide a display.
2. Wide-Band Amplifier
Plug-In Unit
Bandwidth 500 MHz; Deflection
factor 50 mV to 5 V
Used throughout procedure
to provide vertical input to
oscilloscope system.
3. Fast-Rise Pulse
Generator
1
4. Time-Mark Generator Marker outputs, 2 ns to 0.5 s
200 mV positive-going output
pulse; rise time 1 ns or less
within 0.1%
Time position check and
adjustment
Sweep timing checks and
adjustments. Sweep delay
checks and adjustments.
5. Low-Frequency Sine-
wave Generator
Frequency, 30 Hz to 50 kHz;
output amplitude, variable from
Low-frequency triggering
checks and adjustments
200 mV to 8 V.
6. Medium-Frequency
Signal Generator
Frequency, 20 MHz to 100 MHz;
output amplitude, variable from
20 MHz triggering checks a. TEKTRONIX SG 503
100 mV to 500 mV.
7. High-Frequency
Signal Generator
Frequency, 245 MHz to 1000
MHz; output amplitude variable
from 0.5 V to 4 V
8. Digital Voltmeter Range 0 to 50 V; accuracy within
0.1% checks.
9. Termination Impedance 50 Ò accuracy, within
connectors, bnc.2%;
10. Plug-In Extender’ Provides access to 7B92A
adjustments
Used throughout procedure
for voltage
Output termination for fastrise generator
Used throughout procedure
adjustments and test points
to provide access to
internal
TEKTRONIX
79040scilloscope
TEKTRONIX 7A19
Amplifier plug-in unit.
TEKTRONIX PG 506
Calibration Generator.
2
a. TEKTRONIX TG 501
Time- Mark Generator.2 b.
TEKTRONIX 2901 TimeMark Generator
a. TEKTRONIX FG 503
Function Generator.
2
b. General Radio 1310-B
Oscillator.
Sine-Wave Generator.
2
b. TEKTRONIX 191 SineWave Generator.
a. TEKTRONIX SG 504
Leveled Sine-Wave
Generator.
2
b. Wavetek 1002
Sweep/Signal Generator.
c. General Radio 1362
UHF Oscillator with 1263-C
Amplitude-Regulating
Power Supply.
TEKTRONIX DM 501
Digital Multimeter.
Tektronix Part No. 0110049-01.
Tektronix Part No. 0670589-00 Calibration
Fixture.
Tektronix Part No. 1030030-00.
I
5-2
@
Performance Check/Calibration-7B92A
Table 5-1. (cont)
Examples of Applicable
Description Minimum Specifications Purpose Test Equipment
12. Cable Impedance, 50 0; type RG-58/U;
length, 18 inches; connectors,
Used throughout procedure
for signal interconnection.
Tektronix Part No. 0120076-00.
bnc.
13. Cable Impedance, 50 0; type RG-58/U,
length 42 inches; connectors, bnc.
14. Screwdriver Three-inch shaft, 3/32 inch bit. Used to adjust variable
Used throughout procedure
for signal interconnection
Tektronix Part No. 0120057-01.
Xcelite R-3323.
resistors.
15. Low Capacitance’
Screwdriver
1
Used for calibration only; NOT used for performance check.
2
Requires a TM 500-Series Power Module.
1-1/2 inch shaft. Used to adjust variable
capacitors
Tektronix Part No. 0030000-00.
@
5-3
Performance Check/Calibration-7B92A
PART I-PERFORMANCE CHECK
The following procedure is intended to be used f or inspection and periodic calibration checks to confirm that the
7B92A is operating within acceptable limits. T his proc edure is conc erned with those por tions of the instr ument calibration
that are essential to measurement accuracy and correct operation. Removal of the side covers is not necessary to
perform this procedure in that all checks are made from the front panel.
OUTLINE FOR PART I-
PERFORMANCE CHECK
A.Trigger SensitivityPage
1.Check Main Triggering Level Range5-5
2.Check Delayed Triggering Level Range5-6
3.Check Low-Frequency Triggering5-6
4.Check AC LF Reject Operation5-7
5.Check 20 MHz Triggering5-7
6.Check 500 MHz Triggering5-8
7.Check HF Sync Triggering5-9
8.Check Line Triggering5-9
9.Check Single Sweep Operation5-9
B.Horizontal System
1.Set Basic Sweep Calibration5-10
2.Check Trace Separation Range
(SN B059999-below)5-10
3.Check Trace Separation Range and
ALT OFF Function (SN B060000-above)5-10
4.Check Sweep Timing5-10
5.Check Sweep Linearity5-12
6.Check VARIABLE Time/Division Range5-12
7.Check Differential Delay Time Accuracy5-13
8.Check Fast Delay Time Accuracy5-14
9.Check Delay Jitter5-14
PRELIMINARY PROCEDURE
1. Install the amplifier plug-in unit in a vertical
compartment of the oscilloscope.
2. Install the 7B92A into the horizontal
compartment of the oscilloscope.
3. Set the oscilloscope vertical mode switch to
display the vertical unit and the horizontal mode switch to
display the horizontal unit.
4. Set the oscilloscope intensity controls fully
counterclockwise and set the trigger s ource switches to
vertical mode.
Page
5. Turn on the oscilloscope and allow at least 20
minutes warm up before beginning the procedure.
POSITIONMidrange
TIME/DIV OR DLY TIME20 /s
DLY’D Time/Division10,us/PULL FOR
ALT
VARIABLE (CAL IN)In
DELAY TIME MULT1.0
TRACE SEPClockwise
INTENSITYAs desired
CONTRASTAs desired
LEVEL RUNS AFTER
DELAY TIME
SLOPE+
COUPLINGAC
SOURCEINT
MAIN TRIGGERING
Sweep Controls
Delayed Triggering
Inputs
A1. Check Main Triggering Level Range
a. Set the vertical amplifier plug-in unit controls as
follows:
Positionmidrange
Polarity+ Up
Input Couplingdc
Volts/Division1 V
b. Connect the bnc tee connector to the input of
the vertical amplifier plug-in unit.
c. Connect a 50-n cable from the low-frequency
sine-wave generator to the tee connector.
d. Connect a 50-n cable from the tee connec tor to
the MAIN TRIG IN connector.
e. Set the oscilloscope intensity and focus controls
for the desired display.
f. Set the low-frequency sine-wave generator for
an eight-division display at 50 kHz. Center the display
vertically on the graticule.
@
TERM1 M Ò (in)
g. Check-that all levels of the positive slope may
be selected for the sweep starting point as the MAIN
TRIGGERING LEVEL control is rotated throughout its
range, and that triggering occurs at least 3.5 divisions
above and below the center horizontal graticule line.
See’ Fig. 5-1 for reference.
h. Check-set the MAIN TRIGGERING SLOPE to
and repeat part g for the negative slope of the waveform.
5-5
Performance Check/Calibration-7B92A
throughout its range, and that triggering occurs at least
3.5 divisions above and below the center horizontal
graticule line. See Fig. 5-2 for reference.
Fig. 5-1. Main Triggering level range.
i. Check-set the MAIN TRIGGERING SOURCE to
EXT and repeat part g of this step.
j. Check-set the MAIN TRIGGERING SLOPE to +
and repeat part g for the positive slope of the waveform.
A2. Check Delayed Triggering Level Range
a. Disconnect the 50-n cable from the MAIN T RIG
IN connector and reconnect it to the DLY’D TRIG IN
connector.
b. Set the Delayed Triggering SOURCE and MAIN
TRIGGERING SOURCE to INT.
c. Check-that all levels of the positive slope of the
intensified zone may be selected for the sweep starting
point as the Delayed Triggering LEVEL control is rotated
Fig. 5-2. Delayed Triggering level range.
d. Check-set the Delayed Triggering SLOPE to
and repeat part c for the negative slope of the waveform.
e. Check-set the Delayed Triggering SOURCE to
EXT and repeat part c of this step.
f. Check-set the Delayed Triggering SLOPE to +
and repeat part c for the positive slope of the waveform.
A3. Check Low-Frequency Triggering
a. Disconnect the 50-n cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector
@
5-6
Performance Check/Calibration-7B92A
b. Set the 7B92A controls as follows:
MAIN TRIGGERINGAC
COUPLING
Delayed Triggering+
SLOPE
Delayed TriggeringAC
COUPLING
Delayed TriggeringINT
SOURCE
Delayed TriggeringRUNS AFTER
LEVELDELAY TIME
TIME/DIV OR DLY TIME10 ms
DLY’D Time/Division2 ms/PULL FOR ALT
DELAY TIME MULT dial1.0
c. Set the vertical amplifier plug-in unit
volts/division switch to 0.2 V.
d. Set the low-frequency sine-wave generator for
a0.5 division (100 mV) display at 30 Hz.
e. Check-for stable display (TRIG’D light on) at all
settings of the LEVEL control with MAIN TRIGGERING
COUPLING set to:
(1.) AC
(2.) AC HF REJ
(3.) DC
(Set the MAIN TRIGGERING LEVEL control as
necessary.)
f. Check-set the MAIN TRIGGERING SOURCE to
EXT and repeat part e of this step.
g. Disconnect the 50-0 cable from the MAIN T RIG
IN connector and reconnect it to the DLY’D TRIG IN
connector.
h. Set the control settings as follows:
MAIN TRIGGERING INT
SOURCE
Delayed Triggering near "O"
LEVEL
Delayed Triggering EXT
SOURCE
i. Check-for stable display (TRIG’D light on) at all
settings of LEVEL controls with Delayed Triggering
COUPLING set to:
(1.) AC
(2.) DC
(Set the MAIN and Delayed Triggering LEVEL controls
as necessary.)
j. Check-set the Delayed Triggering SOURCE to
INT and repeat part i of this step.
A4. Check AC LF Reject Operation
a. Disconnect the 50-n cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b. Set the vertical amplifier plug-in unit
volts/division switch to 0.5 V.
c. Set the low-frequency sine-wave generator for
an eight-division display at 60 Hz. Center the display
vertically on the graticule.
d. Check-set the MAIN TRIGGERING LEVEL
control for a stable display (TRIG’D light on).
e. Set the MAIN TRIGGERING COUPLING to AC
LF REJ; set the MODE to NORM.
f. Check-for no display (TRIG’D light off ) when the
MAIN TRIGGERING LEVEL control is r otated throughout
its range.
g. Set the MAIN TRIGGERING COUPLING to AC;
set the SOURCE to EXT.
h. Reduce the output of the low-frequency sinewave generator to provide a six-division display (3.0 V) at
60 Hz.
(Adjust the MAIN TRIGGERING LEVEL control to
maintain a stable display.)
i. Set the MAIN TRIGGERING COUPLING to AC
LF REJ and repeat part f of this step.
A5. Check 20 MHz Triggering
a. Set the vertical amplifier plug-in unit
volts/division switch to 0.2 V.
b. Set the 7B92A controls as follows:
MAIN TRIGGERING AC
COUPLING
MAIN TRIGGERING AUTO
MODE
MAIN TRIGGERING INT
SOURCE
@
5-7
Performance Check/Calibration-7B92A
Delayed Triggering All pushbuttons in
COUPLING, SOURCE
and SLOPE
Delayed TriggeringRUNS AFTER
LEVELDELAY TIME
TIME/DIV OR DLY TIME50 ns
DLY’D Time/Division5 ns/PULL FOR ALT
c. Disconnect the 50-n cable from the lowfrequency sine-wave generator and connect the mediumfrequency signal generator to the vertical amplifier plugin unit input using the 50-n cable.
d. Set the medium-frequenc y signal generator for
a 0.5 division display (100 mV) at 20 MHz.
e. Check-for a stable display (TRIG’D light on)
with the MAIN TRIGGERING COUPLING switch set to:
(1.) AC
(2.) AC LF REJ
(3.) DC
(Set the MAIN TRIGGERING LEVEL control as
necessary.)
f. Check-change the MAIN TRIGGERING
SOURCE to EXT and repeat part e of this step.
g. Disconnect the 50-n cable from the MAIN T RIG
IN connector and reconnect it to the DLY’D TRIG IN
connector.
h. Set the Delayed Triggering LEVEL control near
"O".
i. Set the MAIN TRIGGERING SOURCE to INT
and MAIN TRIGGERING COUPLING to AC.
j. Check-for a s table display (TRIG’D light on) with
the Delayed Triggering COUPLING set to:
(1.) AC
(2.) DC
k. Set the Delayed Triggering SOURCE to EXT
and repeat part j of this step.
A6. Check 180 MHz Triggering
a. Connect the high-frequency leveled sine-wave
generator to the bnc tee connected to the vertical
amplifier plug-in unit input.
b. Disconnect the 50-n cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
c. Set the 7B92A controls as follows:
MAIN TRIGGERINGDC
COUPLING
MAIN TRIGGERINGEXT
SOURCE
TIME/DIV OR DLY TIME2 ns
DLY’D Time/Division2 ns (knob in)
d. Adjust the oscilloscope intensity control for
normal viewing.
e. Set the vertical amplifier plug-in unit
volts/division switch to 0.5 V.
f. Set the high-frequency leveled sine-wave
generator for a one-division display at 180 MHz. Center
the display vertically on the graticule.
g. Set the MAIN TRIGGERING LEVEL control for
a stable display (TRIG’D light on).
h. Check-for a stable display with no more than
0.05 division of jitter.
i. Check-change the MAIN TRIGGERING
SOURCE to INT and repeat parts g and h of this step.
j. Disconnect the 50-0 cable fr om the MAIN TRIG
IN connector and reconnect it to the DLY’D TRIG IN
connector.
k. Set the DLY’D Time/Division to PULL FOR ALT;
set the Delayed Triggering SOURCE to EXT.
I. Set the Delayed Triggering LEVEL control for a
stable display (near "O").
I. Disconnect the 50-n cable from the mediumfrequency signal generator; disconnect the 50-n cable
from the bnc tee at the vertical amplifier plug-in input
connector.
@
5-8
Performance Check/Calibration-7B92A
m. Check-f or a stable display with no more than
0.05 division of jitter.
n. Check-change Delayed Triggering SOURCE to
INT and repeat m of this step.
A7. Check HF Sync Triggering
a. Disconnect the 50-n cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b. Set the Delayed Triggering LEVEL control to
RUNS AFTER DELAY TIME.
c. Set the DLY’D Time/Division switch to Normal
mode (knob in).
d. Set the MAIN TRIGGERING MODE to HF
SYNC.
e. Set the vertical amplifier plug-in unit
volts/division switch to 1 V.
f. Set the MAIN TRIGGERING LEVEL control for
a stable display (TRIG’D light on).
g. Check-for a stable display with no more than
0.05 division of jitter.
h. Set the high-frequency leveled sine-wave
generator for a one-division display.
i. Check-change the MAIN TRIGGERING
SOURCE to EXT 10 and repeat part g of this step.
c. Check-set the MAIN TRIGGERING LEVEL
control near "0" and check that the TRIG’D light is on.
d. Check-that the display is not triggered (TRIG’D
light off) at either end of the MAIN T RIGGERING LEVEL
control rotation.
A9. Check Single Sweep Operation
a. Connect the time-mar k gener ator to the input of
the vertical amplifier plug- in unit, using a 50-n cable; set
the generator for 1 ms markers.
b. Set the vertical amplifier plug-in unit deflec tion
factor for approximately two divisions of display; and
adjust MAIN TRIGGERING LEVEL control for a stable
display.
c. Set the MAIN TRIGGERING SOURCE to INT,
MODE to SINGLE SWEEP, and SOURCE to EXT.
d. Press the SINGLE SWEEP RESET pushbutton;
it should be lit.
e. Check-change the MAIN TRIGGERING
SOURCE to INT and observe that only one sweep is
displayed and SINGLE SWEEP RESET pushbutton goes
out.
f. Set the TIME/DIV OR DLY TIME switch to
Alternate mode (PULL FOR ALT).
g. Set the MAIN TRIGGERING SOURCE to EXT;
press the SINGLE SWEEP RESET pushbutton. The
pushbutton should be lit.
j. Disconnect the 50-Ò cable from the high-
frequency leveled sine-wave generator.
A8. Check Line Triggering
a. Disconnect the 50-Ò cable from the MAIN T RIG
IN connector.
b. Set the MAIN TRIGGERING MODE to AUTO,
SOURCE to LINE, and COUPLING to AC; the TIME/DIV
OR DLY TIME switch to 1 ms; and DLY’D Time/Divis ion
to 1 ms (knob in).
@
h. Check-change the MAIN TRIGGERING
SOURCE to INT and observe that only one sweep is
displayed and SINGLE SWEEP RESET pushbutton goes
out.
i. Press SINGLE SWEEP RESET pushbutton
several times and observe that display alternates
between Delaying and Delayed sweeps.
POSITION Midrange
TIME/DIV OR DLY TIME 1 ms
DLY’D Time/Division 1 ms (knob in)
Delayed Triggering
LEVELRUNSAFTER
DELAY TIME
b. Connect the time-mark generator to the vertical
amplifier plug-in unit input with a50-Q cable. Set the time
mark generator for 1 ms markers.
c. Set the oscilloscope intensity and focus c ontrols for a
desired display.
d. Set the vertical amplifier plug-in unit for approxim ately
two divisions of display. Position the display in the center
graticule area. Set the MAIN TRIGGERING LEVEL
control for a stable display (TRIG’D light on).
e. Set the front panel SWP CAL adjustment for one
marker/division over the center eight divisions (position
as necessary).
B2Check Alt Off Function
a. Check-for Delaying sweep (with intensified
zone) positioned at least 3.5 divisions above Delayed
sweep.
b. Check-for an intensified por tion on the Delaying
sweep, and that the gate (checked at oscilloscope Gate
Out connector) is not divided by two.
B3 Check Sweep Timing
NOTE
The tolerances given in Table 5-2 are for
ambient temperature range of +150°C to
+35°C. If outside this range, see the
Specification section for applicable
tolerances.
a. Set the TIME/DIV OR DLY TIME, DLY’D
Time/Division in Alternate mode (PULL FOR ALT), and
time-mark generator as indicated in Table 5-2; set
DELAY TIME MULT dial to 0.0.
5-10
@
Performance Check/Calibration-7B92A
Table 5-2.
SWEEP TIMING
Tolerance
TIME/DIV Time-Mark (+15°C to +35°C)
OR DLY'D Generator
DLY TIME Time/Division Setting Delaying Sweep Delayed Sweep
0.2 s 0.2 s 0.2 s Within 0.16 div Within 0.16 div
0.1 s 0.1 s 0.1 s Within 0.16 div Within 0.16 div
50 ms 50 ms 50 ms Within 0.16 div Within 0.16 div
20 ms 20 ms 20 ms Within 0.16 div Within 0.16 div
10 ms 10 ms 10 ms Within 0.16 div Within 0.16 div
5 ms 5 ms 5 ms Within 0.16 div Within 0.16 div
2 ms 2 ms 2 ms Within 0.16 div Within 0.16 div
1 ms 1 ms 1 ms Within 0.16 div Within 0.16 div
0.5 ms 0.5 ms 0.5 ms Within 0.16 div Within 0.16 div
0.2 ms 0.2 ms 0.2 ms Within 0.16 div Within 0.16 div
0.1 ms 0.1 ms 0.1 ms Within 0.16 div Within 0.16 div
50 µs 50 µs 50 µs Within 0.16 div Within 0.16 div
20 µs 20 µs 20 µs Within 0.16 div Within 0.16 div
10 µs 10 µs 10 µs Within 0.16 div Within 0.16 div
5 µs5 µs5 µs Within 0.16 div Within 0.16 div
2 µs2 µs2 µs Within 0.16 div Within 0.16 div
1 µs1 µs1 µs Within 0.16 div Within 0.16 div
0.5 µs0.5 µs0.5 µs Within 0.16 div Within 0.16 div
0.2 µs0.2 µs0.2 µs Within 0.16 div Within 0.16 div
0.1 µs0.1 µs0.1 µs Within 0.16 div Within 0.16 div
50 ns 50 ns 50 ns Within 0.16 div Within 0.16 div
20 ns 20 ns 20 ns Within 0.16 div Within 0.16 div
10 ns 10 ns 10 ns Within 0.24 div Within 0.24 div
b. Check-using the settings given in Table 5-2,
check sweep accuracy for onetime mark/division over
the center eight divisions within the tolerance given in
Table 5-2 Set the POSITION control and MAIN
TRIGGERING LEVEL control as necessar y for a stable
display that is aligned with the vertical graticule lines.
NOTE
If the time-mark generator used does not have 1-2-5
sequence markers, apply 1 unit markers in place of 2
unit markers and check for 2 markers/division over the
center eight divisions of display, to the
tolerances given in Table 5-2.
c. Push in the DLY’D Time/Division knob and
position the display to the center of the graticule.
d. Check-using the settings given in Table 5-3,
check sweep accuracy over the center eight divisions
within the tolerance given in Table 5-3.
NOTE
The HF SYNC MODE switch may provide a better
display when checking the fastest sweep rates.
5-11
Performance Check/Calibration-7B92A
Table 5-3.
FAST TIMING
TIME/DIV Time-Mark
OR Generator Display Tolerance
DLY TIME Setting Markers/Division (+15°C to +35°C)
5 ns 5 ns 1 Within 0.24 div
2 ns 2 ns 1 Within 0.24 div
B4Check Sweep Linearity
a. Set the TIME/DIV OR DLY TIME and
DLY’D Time/Division switch to 1 ms and pull DLY’D
Time/Division knob out for Alternate mode.
b.Set the time-mark generator for 1 ms markers.
c. Position the Delaying (top) and Delayed-sweep
traces horizontally to start on the first graticule line.
d. Check-that the fourth marker of the Delaying
sweep coincides within 0.1 division with the fourth
vertical graticule line.
e. Check-continue linearity check, as in part d of
this step, for each successive two divisions of the
Delaying sweep.
f. Check-repeat parts d and e of this step to
check linearity of the Delayed (bottom) sweep.
g. Set the TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 10 ns and pull DLY’D Time/Division knob out for Alternate mode.
k. Check-continue linearity check for Delayed
sweep rates as shown in Table 5-4.
Table 5-4.
FAST SWEEP LINEARITY
Time-mark
Time/ Generator Markers/ Tolerance/
Division Setting 2 Divisions2 Divisions
5 ns 5 ns 2 0.2 div
2 ns 2 ns 2 0.2 div
B5Check Variable Time/Division Range
a. Remove the 7B92A from the oscillosc ope plug-
in compartment.
b. Place the Time/Division Variable Selector
multi-pin connector on P761 (Delaying Sweep). See Fig.
1-1 for location.
h.Set the time-mark generator for 10 ns markers.
i. Check-perform linearity chec k as in parts d, e,
and f of this step. Tolerance is 0.1 division for each two
divisions over center eight graticule divisions.
j. Set the DLY’D Time/Division to Normal mode
(knob in).
c. Insert the 7B92A into the oscilloscope plug-in
compartment; turn on the power to the oscilloscope.
d. Set the time-mark generator for 100 us
markers.
e. Set the TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 20 ,us and pull DLY’D
Time/Division knob out for Alternate mode. Check that
the DELAY TIME MULT dial is 0.0.
5-12
f. Press and release VARIABLE control and turn
it fully counterclockwise.
g. Check-that Delaying sweep markers (top) and
Delayed sweep markers are displayed as shown in Fig.
5-3; at least 2-1/2 Delaying sweep markers to one
Delayed sweep marker are visible.
Performance Check/Calibration-7B92A
c. Set the vertical amplifier plug-in unit deflection
factor for approximately two divisions of display.
d. Set the DELAY TIME MULT dial to 0.90.
e. Align the time-mark to the center vertical
graticule line, using the POSITION control.
f. Turn the DELAY TIME MULT dial to
approximately 8.9 to align the time-mark to the center
vertical graticule line.
g. Check-the DELAY TIME MULT dial to read
within three minor divisions of 8.90 (8.87 to 8.93).
h. Repeat parts d through g of this step for eac h
sweep rate given in Table 5-5. Maintain approximately
two divisions of display.
Table 5-5.
DIFFERENTIAL DELAY TIME ACCURACY
Fig. 5-3. Variable Time/Division range.
h. Remove the 7B92A from the oscillosc ope plugin compartment; reposition the Time/Division Variable
Selector multi-pin connector to P760 and re-install the
7B92A into the plug-in compartment. Turn on power to
the oscilloscope.
i. Check-with VARIABLE control released and
turned fully counterclockwise, at least two Delayed
sweep markers (bottom) to one Delaying sweep marker
is visible.
B6Check Differential Delay Time Accuracy
a. Set the TIME/DIV OR DLY TIME switch to 200
ms, DLY’D Time/Division switch (pull knob out) to 5 ms
(then push knob in), push in VARIABLE (CAL IN) control,
and adjust MAIN TRIGGERING LEVEL control for a
stable display (TRIG’D light on).
b. Set the time-mark generator for 100 ms
markers.
Time-mark
TIME/DIV OR DLY’D Generator
DLY TIME Time/Division Setting
.2 s 5 ms 0.1 s
.1 s 2 ms 50 ms
50 ms I ms 10 ms
20 ms 0.5 ms 10 ms
10 ms 0.2 ms 5 ms
5 ms 0.1 ms 1 ms
2 ms 50 µs1 ms
1 ms 20 µs 0.5 ms
.5 ms 10 µs 0.1 ms
.2 ms 5 µs 0.1 ms
.1 ms 2 µs50µs
50 µs1 µs 10 µs
20 µs.5 µs 10 µs
10 µs.2 µs5 µs
a. Set the TIME/DIV OR DLY TIME switch to 50
ns, DLY’D Time/Division switch (pull knob out) to 10 ns
(then push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
b. Set the time-mark generator for 50 ns markers.
c. Set the vertical amplifier plug-in unit deflection
factor for approximately two divisions of display.
d. Set the DELAY TIME MULT dial to 9.00.
e. Align the leading edge of theti me-mark to the
center vertical graticule line, using the POSITION
control. See Fig 5-4 for reference.
f. Check-set the DELAY TIME MULT dial to the
settings listed in Table 5-6, and check for maximum error
in dial reading.
Table 5-6.
FAST DIFFERENTIAL DELAY TIME ACCURACY
DELAY TIME MULT Maximum Error
(Minor Dial Divisions)
9.0 0 (step e)
8.0 Within 3.0
7.0 Within 3.5
6.0 Within 4.0
5.0 Within 4.5
4.0 Within 5.0
3.0 Within 3.0
2.0 Within 6.0
1.0 Within 6.5
g. Set the TIME/DIV OR DLY TIME switch to 20
ns, DLY’D Time/Division switch (pull knob out) to 2 ns
(then push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
h. Set the time-mark generator for 20 ns markers.
i. Repeat parts d, e, and f of this step.
j. Set the TIME/DIV OR DLY TIME switch to 10
ns, DLY’D Time/Division switch (pull knob out) to 2 ns
(then push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
Fig. 5-4. Fast delay time accuracy.
k. Set the time-mark generator for 10 ns markers.
I. Repeat parts d, e, and f of this step. Delete
check at DELAY TIME MULT dial setting of 1.0.
B8Check Delay Jitter
a. Set the TIME/DIV OR DLY TIME switch to 1
ms, DLY’D Time/Division switch (pull knob out) to 5 ns
(then push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
5-14
Performance Check/Calibration-7B92A
b.Set the time-mark generator for 1 ms markers.
c.Set the DELAY TIME MULT dial to
approximately 1.0 (position the time-mark to the center
of the graticule).
d.Set the DLY’D Time/Division switch to 0.2 us.
e.Turn the DELAY TIME MULT dial slowly to
position the time-mark near the graticule center (turn un
INTENSITY control as necessary).
f.Check-jitter must not exceed one division
(disregard slow drift).
g.Set the TIME/DIV OR DLY TIME switch to 20 ps,
DLY’D Time/Division switch (pull knob out) to 50 ns ( then
push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
h.Set the time-mark generator for 50 ns markers.
i.Set the DLY’D Time/Division switch to 5 ns.
j.Turn the DELAY TIME MULT dial slowly to
position the time-mark near the graticule center (turn up
INTENSITY control as necessary).
k.Check-jitter must not exceed 0.9 division.
I.Turn the DELAY TIME MULT dial to
approximately 9.0 and position the time-mark to the
center of the graticule.
m.Check-jitter must not exceed 0.9 division.
This completes the Performance Check procedure
of the 7B92A.
5-15
PART II-CALIBRATION
The following procedure returns the 7B92A to corr ect c alibration. All lim its and tolerances given in
this procedure are calibration guides and should not be inter preted as instrum ent spec ifications ex cept
as specified in the Specification section of this m anual. Where possible, instrum ent performance is
checked before an adjus tment is made. When perf orming a complete Calibration proc edure, make
each adjustment to the given setting, even if the Check- is within the allowable tolerance.
OUTLINE FOR
PART II--CALIBRATION
A.Trigger Sensitivity (SN B070000-above) Page
1.Preliminary Main Triggering Adjust-5-18
ment
2.Adjust Main Triggering Internal DC5-18
Balance
Performance Check/Calibration-7B92A
Page
2.Adjust Main Triggering Arming Tunnel
Diode Bias5-26
3.Adjust Main Triggering Gate Tunnel
Diode Bias5-26
4.Adjust Main Triggering Internal DC
Balance5-26
5.Adjust Main Triggering Sensitivity5-26
3.Adjust Main Triggering Sensitivity5-19
4.Adjust External DC Balance5-19
5.Check Main Triggering Level Range5-19
6.Preliminary Delayed Triggering Adjustment5-20
7.Adjust Delayed Triggering Sensitivity5-20
8.Adjust Delayed Triggering External DC
Balance5-20
POSITIONMidrange
TIME/DIV OR DLY TIME20 Ps
DLY’D Time/Division10 ps/PULL FOR ALT
VARIABLE (CAL IN)In
DELAY TIME MULT1.0
INTENSITYAs desired
CONTRASTAs desired
Delayed Triggering
LEVELRUNS AFTER
DELAY TIME
SLOPE+
COUPLINGAC
SOURCEINT
PositionMidrange
Polarity+ up
Input Couplingdc
Volts/Division50 mV
A1. Preliminary Main Triggering Adjustment
a.Connect the bnc tee connector to the input of
the vertical amplifier plug-in unit.
b.Connect a 50-Q cable from the low-frequency
sine-wave generator to the tee connector.
c.Connect a 50- Q cable from the tee connector
to the MAIN TRIG IN connector.
d.Set the oscilloscope intensity and focus
controls for the desired display.
e.Set the low-frequency sine-wave generator for
a four-division display at 50 kHz. Center the display
vertically on the graticule.
f.Adj ust-Int DC Bal, R75, Ext DC BAL, R99, and
Trig Sens, R175, to midrange.
Inputs
TERM1 MΩ (in)
A2. Adjust Main Triggering Internal DC Balance
a.Set the MAIN TRIGGERING LEVEL control to
"O".
5-18
b.Adjust-Int DC Bal, R75, for a sweep trigger
point at graticule centerline (TRIG’D light on). See Fig.
5-5 for reference.
A3. Adjust Main Triggering Sensitivity
a.Adjust-Trig Sens, R1 75, for a sweep trigger
point at approximately 0.3 division from bottom of
waveform as MAIN TRIGGERING LEVEL control is
rotated through negative (-) portion of its range. See Fig.
5-6for reference.
NOTE
Setting a trigger sensitivity so display is trigger ed at
less than 0. 1 division should be avoided to prevent
erratic triggering.
b.Adjust-Ext DC Bal, R99, for a sweep trigger
point at graticule centerline (TRIG’D light on). See Fig.
5-5 for reference.
c.Set the MAIN TRIGGERING SOURCE to INT.
A5. Check Main Triggering Level Range
a.Set the vertical amplifier plug-in unit
volts/division switch to 1 V.
Fig. 5-7. Main Triggering level range (Circle denotes
triggering limits).
Fig. 5-5. DC balance (Circle denotes correct trigger point)
5-19
Performance Check/Calibration-7B92A
b.Set the low-frequency sine-wave generator for
an eight-division display at 50 kHz. Center the display
vertically on the graticule.
c.Chec k-that all levels of the positive slope may
be selected for the sweep starting point as the MAIN
TRIGGERING LEVEL control is rotated throughout its
range, and that triggering occurs at least 3.5 divisions
above and below the center horizontal graticule line. See
Fig. 5-7 for reference.
d.Check-set the MAIN TRIGGERING SLOPE to
and repeat part c for the negative slope of the waveform.
e.Check-set the MAIN TRIGGERING SOURCE
to EXT and repeat part c of this step.
f.Check -set the MAIN TRIGGERING SLOPE to
+ and repeat part c for the positive slope of the
waveform.
A6. Preliminary Delayed Triggering Adjustment
a.Disconnect the 50-n cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
b.Set the 7B92A controls as follows:
MAIN TRIGGERINGINT
SOURCE
MAIN TRIGGERINGNear "0"
LEVEL
Delayed Triggering+
SLOPE
Delayed TriggeringAC
COUPLING
Delayed TriggeringINT
SOURCE
Delayed Triggering"0"
LEVEL
DELAY TIME MULT1.0
TIME/DIV OR DELAY20 µs
TIME
DLY’D Time/Division10 ps/PULL FOR
ALT
c.Adjust Ext DC Bal, R230, Int DC Bal, R255,
and Trig Sens, R320, to midrange.
d.Set the vertical amplifier plug-in unit
volts/division switch to 50 mV.
e.Set the low-frequency sine-wave generator for
a four-division display at 50 kHz. Center the display
vertically on the graticule.
f.Adjust-Int DC Bal, R255, for a sweep trigger
point at graticule centerline (TRIG’D light on). See Fig.
5-5 for reference.
A7. Adjust Delayed Triggering Sensitivity
a.Adjust-Trig Sens, R320, for a sweeptrigger
point at approximately 0.3 division from bottom of
waveform as Delayed Triggering LEVEL control is
rotated through negative (-) portion of its range. See Fig.
5-6 for reference.
NOTE
Setting trigger sensitivity so display is
triggered at less than 0. 1 division s hould be
avoided to prevent erratic triggering.
b.Reset the Delayed Triggering LEVEL control to
"0"
A8. Adjust Delayed Triggering External DC Balance
a.Set Delayed Triggering SOURCE to EXT.
b.Adjust-Ext DC Bal, R230, for a sweep trigger
point at graticule centerline (TRIG’D light on). See Fig.
5-5 for reference.
c.Set the Delayed Triggering SOURCE to INT.
A9. Check Delayed Triggering Level Range
a.Set the vertical amplifier plug-in unit
volts/divisior switch to 1 V.
b.Set the low-frequency sine-wave generator for
ar eight-division display at 50 kHz. Center the display
vertically on the graticule. (Lower the Delaying sweep
intensity with the oscilloscope intensity control.)
c.Chec k-that all levels of the positive slope of the
intensified zone may be selected for the sweep starting
point as the Delayed Triggering LEVEL control is rotated
throughout its range, and that triggering occurs at least
3.5 divisions above and below the center horizontal
graticule line. See Fig. 5-8 for reference.
@
5-20
Performance Check/Calibration-7B92A
Delayed Triggering+
SLOPE
Delayed TriggeringAC
COUPLING
Delayed TriggeringINT
SOURCE
Delayed TriggeringRUNS AFTER
LEVELDELAY TIME
TIME/DIV OR DLY TIME10 ms
DLY’D Time/Division2 ms/PULL FOR ALT
DELAY TIME MULT1.0
c.Set the vertical amplifier plug-in unit
volts/division switch to 0.2 V.
d.Set the low-frequency sine-wave generator for
a 0.5 division (100 millivolts) display at 30 Hz.
e.Check-for a stable dis play (TRIG’D light on) at
all settings of the MAIN TRIGGERING LEVEL control
with MAIN TRIGGERING COUPLING set to:
and repeat part c for the negative slope of the waveform.
e.Check-set the Delayed Triggering SOURCE to
EXT and repeat part c of this step.
f.Check -set the Delayed Triggering SLOPE to +
and repeat part c for the positive slope of the waveform.
A10.Check Low-Frequency Triggering
a.Disconnect the 50-) cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b.Set the 7B92A controls as follows:
MAIN TRIGGERINGAC
COUPLING
(Set the MAIN TRIGGERING LEVEL control as
necessary.)
f.Check-set the MAIN TRIGGERING SOURCE
to EXT and repeat part e of this step.
g.Disconnect the 50-4 cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
h.Set the control settings as follows:
MAIN TRIGGERINGINT
SOURCE
Delayed Triggeringnear "0"
LEVEL
Delayed TriggeringEXT
SOURCE
i.Check-for a s table display (TRIG’D light on) at
all settings of LEVEL controls with Delayed Triggering
COUPLING set to:
(1.) AC
(2) DC
(Set the MAIN and Delayed Triggering LEVEL controls
as necessary.)
@
5-21
Performance Check/Calibration-7B92A
j.Check-set the Delayed Triggering SOURCE to
INT and repeat part i of this step.
A11.Check AC LF Reject Operation
a.Disconnect the 50-C cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b.Set the vertical amplifier plug-in unit
volts/division switch to 0.5 V.
c.Set the low-f requency sine-wave generator for
an eight-division display at 60 Hz. Center the display
vertically on the graticule.
d.Check-set the MAIN TRIGGERING LEVEL
control for a stable display (TRIG’D light on).
e.Set the MAIN TRIGGERING COUPLING to AC
LF REJ; set the MODE to NORM.
f.Check-for no display (TRIG’D light off) when
the MAIN TRIGGERING LEVEL control is rotated
throughout its range.
g.Set the MAIN TRIGGERING COUPLING to
AC; set the SOURCE to EXT.
h.Reduce the output of the low-frequency sinewave generator to provide a six-division display (3.0 V) at
60 Hz. (Adjust the MAIN TRIGGERING LEVEL control
to maintain a stable display.)
i.Set the MAIN TRIGGERING COUPLING to AC
LF REJ and repeat part f of this step.
A12. Check 20 MHz Triggering
a.Set the vertical amplifier plug-in unit
volts/division switch to 0.2 V.
SOURCE
Delayed TriggeringAll pushbuttons in
COUPLING, SOURCE
and SLOPE
Delayed TriggeringRUNS AFTER
LEVELDELAY TIME
TIME/DIV OR DLY TIME50 ns
DLY’D Time/Division5 ns/PULL FOR ALT
c.Disconnect the 50-Q cable from the low-
frequency sine-wave generator and connect the mediumfrequency signal generator to the vertical amplifier plugin unit input using the 50-n cable.
d.Set the medium-frequency signal generator for
a 0.5 division display (100 mV) at 20 MHz.
e.Check-for a stable display (TRIG’D light on)
with the MAIN TRIGGERING COUPLING switch set to:
(1) AC
(2) AC LF REJ
(3) DC
(Set the MAIN TRIGGERING LEVEL control as
necessary.)
f.Check-change the MAIN TRIGGERING
SOURCE to EXT and repeat part e of this step.
g.Disconnect the 50-C cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
h.Set the Delayed Triggering LEVEL control near
"0".
i.Set the MAIN TRIGGERING SOURCE to INT
and MAIN TRIGGERING COUPLING to AC.
j.Check-for a stable display (TRIG’D light on)
with the Delayed Triggering COUPLING switch set to:
(1) AC
(2) DC
b.Set the 7B92A controls as follows:
MAIN TRIGGERINGAC
COUPLING
MAIN TRIGGERINGAUTO
MODE
MAIN TRIGGERINGINT
k.Set the Delayed Triggering SOURCE to EXT
and repeat part j of this step.
I.Disconnect the 50-C cable from the medium-
frequency signal generator; disconnect the 50-C cable
from the bnc tee at the vertical amplifier plug-in input
connector.
5-22
Performance Check/Calibration-7B92A
A-13. Check 180 MHz Triggering
a. Connect the high-frequency leveled sine-wave
generator to the bnctee connected to the vertical
amplifier plug-in unit input.
b. Disconnect the 50-n cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
c. Set the 7B92A controls as follows:
MAIN TRIGGERING DC
COUPLING
MAIN TRIGGERING EXT
SOURCE
TIME/DIV OR DLY TIME 1 ns
DLY’D Time/Division 1 ns (knob in)
d. Adjust the oscilloscope intensity control for
normal viewing.
e. Set the vertical amplifier plug-in unit
volts/division switch to 0.5 V.
f. Set the high-frequency leveled sinewave
generator for a one-division display at 180 MHz. Center
the display vertically on the graticule.
g. Set the MAIN TRIGGERING LEVEL control for
a stable display (TRIG’D light on).
h. Check-for a stable display with no more than
0.05 division of jitter.
i. Check-change the MAIN TRIGGERING
SOURCE to INT and repeat parts g and h of this step.
j. Disconnect the 50-Ω cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
k. Set the DLY’D Time/Division to PULL FOR
ALT; set the Delayed Triggering SOURCE to EXT.
I. Set the Delayed Triggering LEVEL control for a
stable display (near "0").
m. Check-for a stable display with no more than
0.05 division of jitter.
n. Check-change Delayed Triggering SOURCE to
INT and repeat m of this step.
A14. Check HF Sync Triggering
a. Disconnect the 50-n cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b. Set the Delayed Triggering LEVEL control to
RUNS AFTER DELAY TIME.
c. Set the DLY’D Time/Division switch to Normal
mode (knob in).
d. Set the MAIN TRIGGERING MODE to HF
SYNC.
e. Set the vertical amplifier plug-in unit
volts/division switch to 1 V.
f. Set the MAIN TRIGGERING LEVEL control for
a stable display (TRIG’D light on).
g. Check-for a stable display with no more than
0.05 division of jitter.
h. Set the high-frequency leveled sine-wave
generator for a one-division display.
I. Check-change the MAIN TRIGGERING
SOURCE to EXT + 10 and repeat part g of this step.
j. Disconnect the 50-n cable from the high-
frequency leveled sine-wave generator.
A15. Check Line Triggering
a. Disconnect the 50-Ω cable from the MAIN
TRIG IN connector.
b. Set the MAIN TRIGGERING MODE to AUTO,
SOURCE to LINE, and COUPLING to AC; the TIME/DIV
OR DLY TIME switch to 1 ms; and DLY’D Ti me/Division
to 1 ms (knob in).
5-23
c.Check-set the MAIN TRIGGERING LEVEL
control near "0" and check that the TRIG’D light is on.
d.Check-that the display is not triggered (TRIG’D
light off) at either end of the MAIN T RIGGERING LEVEL
control rotation.
A16Check Single Sweep Operation
Performance Check/Calibration-7B92A
e.Check-change the MAIN TRIGGERING
SOURCE to INT and observe that only one sweep is
displayed and SINGLE SWEEP RESET pushbutton goes
out.
f.Set the TIME/DIV OR DLY TIME switch to
Alternate mode (PULL FOR ALT).
a.Connect the time-mark generator to the input
of the vertical amplifier plug-in unit, using a 50-Ω cable;
set the generator for 1 ms markers.
b.Set the vertical amplifier plug- in unit deflection
factor for approximately two divisions of display; and
adjust MAIN TRIGGERING LEVEL control for a stable
display.
c.Set the MAIN TRIGGERING SOURCE to INT,
MODE to SINGLE SWEEP, and SOURCE to EXT.
d. Press the SINGLE SWEEP RESET
pushbutton; it should be lit.
@
g.Set the MAIN TRIGGERING SOURCEto EXT;
press the SINGLE SWEEP RESET pushbutton. The
pushbutton should be lit.
h.Check-change the MAIN TRIGGERING
SOURCE to INT and observe that only one sweep is
displayed and SINGLE SWEEP RESET pushbutton goes
out.
i.Press SINGLE SWEEP RESET pushbutton
several times and observe that display alternates
between Delaying and Delayed sweeps.
POSITIONMidrange
TIME/DIV OR DLY TIME20µs
DLY’D Time/Division10 µs/PULL FOR
ALT
VARIABLE (CAL IN)In
DELAY TIME MULT1.0
INTENSITYAs desired
CONTRASTAs desired
Delayed Triggering
LEVELRUNS AFTER
DELAY TIME
SLOPE+
COUPLINGAC
SOURCEINT
Inputs
TERM1 MΩ (in)
Vertical Amplifier Plug-In Unit
PositionMidrange
Polarity+ up
Input Couplingdc
Volts/Division50 mV
B1. Preliminary Main Triggering Adjustment
a.Connect the bnc tee connector to the input of
the vertical amplifier plug-in unit.
b.Connect a 50-Ω cable from the low-f requency
sine-wave generator to the tee connector.
c.Connect a 50-Ω cable from the tee c onnector
to the 7B92A MAIN TRIG IN connector.
d.Set the oscilloscope intensity and focus
controls for the desired display.
e.Set the low-frequency sine-wave generator for
a four-division display at 50 kHz (200 mV). Center the
display vertically on the graticule.
f.Adj ust-Int DC Bal, R25, Ext DC Bal, R59, and
Trig Sens, R125, to midrange.
g.Adjust-Arming TD Bias, R145, and Gate TD
Bias, R155, fully clockwise. The T RIG’D light should be
on and the displayed waveform should be unstable.
5-25
h.Adjust-Gate TD Bias, R155, counterclockwise
until a stable display occurs, then clock wise until display
is unstable.
i.Connect the digital voltmeter between TP145
and chassis ground.
B2. Adjust Main Triggering Arming Tunnel Diode
Bias
a.Adjust-Arming TD Bias, R145,
counterclockwise until a stable display just occurs.
Record the voltage at TP145; continue counterclock wise
adjustment of R145 until TRIG’D light just goes out.
Record the voltage at TP1 45.
b.Adjust-Arming T D Bias, R145, for the average
of the two recorded voltage readings; disconnect the
digital voltmeter.
Performance Check/Calibration-7B92A
Fig. 5-9. DC balance. (Circle denotes correct trigger
point.)
B5. Adjust Main Triggering Sensitivity
B3. Adjust Main Triggering Gate Tunnel Diode Bias
a.Disconnect the delay line cable at connector
J170.
b.Connect the digital voltmeter between TP155
and chassis ground.
c.If TRIG’D light is on and waveform display is
stable, adjust Gate TD Bias, R155, counterclockwise
until TRIG’D light just goes out.
d.Record the voltage at TP155.
e.Adjust-Gate TD Bias, R155, until TRIG’D light
just goes out; then clockwise until waveform display is
stable. Record the voltage at TP155.
f.Adj ust-Gate TD Bias, R155, for the average of
the two recorded voltage readings obtained in parts d
and e of this step; disconnect the digital voltmeter.
B4. Adjust Main Triggering Internal DC Balance
a.Set the MAIN TRIGGERING COUPLING to
AC; set the LEVEL control to "0".
a.Adjust-Trig Sens, R125, for a sweep trigger
point at approximately 0.5 division from bottom of
waveform as MAIN TRIGGERING LEVEL control is
rotated through negative (-) portion of its range. See Fig.
5-10 for reference.
B6. Adjust Main Triggering External DC Balance
a.Set the MAIN TRIGGERING SOURCE to EXT.
b.Adjust-Ext DC Bal, R59, for a sweep trigger
point at graticule centerline. See Fig. 5-9 for reference.
B7. Check Main Triggering Level Range
a.Set MAIN TRIGGERING SOURCE to EXT.
b.Adjust-Int DC Bal, R25, for a stable crt display
at graticule centerline (TRIG’D light on). See Fig. 5-9 for
reference.
b.Set the vertical amplifier plug-in unit
volts/division switch to 1 V.
c.Set the low-f requency sine-wave generator for
an eight-division display at 50 kHz. Center the display
vertically on the graticule.
d.Check-that all levels of the pos itive slope may
be selected for the sweep starting point as the MAIN
TRIGGERING LEVEL control is rotated throughout its
range, and that triggering occurs at least 3.5 divisions
above and below the center horizontal graticule line. See
Fig. 5-11 for reference.
e.Check-set the MAIN TRIGGERING SLOPE to
and repeat part d for the negative slope of the waveform.
Performance Check/Calibration-7B92A
f.Check-set the MAIN TRIGGERING SOURCE
to EXT and repeat part d of this step.
g.Check-set the MAIN TRIGGERING SLOPE to
+ and repeat part d for the positive slope of the
waveform.
B8. Preliminary Delayed Triggered Adjustment
a.Disconnect the 50-Ω cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
b.Set the 7B92A controls as follows:
MAIN TRIGGERINGINT
SOURCE
MAIN TRIGGERINGNear "O"
LEVEL
Delayed Triggering +
SLOPE
Delayed Triggering AC
COUPLING
Delayed TriggeringINT
SOURCE
Delayed Triggering"O"
LEVEL
DELAY TIME MULT1.0
TIME/DIV OR DLY TIME20 µs
DLY’D Time/Division10, µs PULL FOR
ALT
Fig. 5-11. Main Triggering level range. (Circle denotes
triggering limits.)
c.Adjust-Ext DC Bal, R225, Int DC Bal, R245,
and Trig Sens, R325, to midrange.
d.Adjust-Arming TD Bias, R345, and Gate TD
Bias, R355, fully clockwise.
e.Set the vertical amplifier plug-in unit
volts/division switch to 50 mV.
f.Set the low-f requency sine-wave generator for
a four-division display at 50 kHz. Center the display
vertically on the graticule.
NOTE
Intensified zone should start at the beginning of Delaying
(top) sweep.
g.Adjust-Gate TD Bias, R355, counterclockwise
until intensified zone moves to right and disappears.
a.Adjust-Arming TD Bias, R345,
counterclockwise until Delayed sweep starts again.
Record the voltage at TP345. Continue
counterclockwise rotation of R345 until Delayed sweep
again disappears. Record the voltage at TP345.
b.Adjust-Arming T D Bias, R345, for the average
of the two voltage readings recorded in part a of this
step. Delayed sweep should be visible.
a.Disconnect the digital voltmeter from TP345
and connect it to TP355; record the voltage.
b.Adjust-Gate Bias, R355, counterclock wise until
Delayed sweep disappears. Record the voltage at
TP355.
c.Adjus t-Gate Bias, R355, for the average of the
two voltage readings recorded in parts a and b of this
step.
B11.Adjust Delayed Triggering Internal DC
Balance
a.Set Delayed Triggering LEVEL control to "0".
b.Adjust-Int DC Bal, R245, for a Delayed sweep
trigger point at graticule centerline. See Fig. 5-9 for
reference.
B13.Adjust Delayed Triggering External DC
Balance
a.Set Delayed Triggering SOURCE to EXT.
b.Adjust-Ext DC Bal, R225, for a Delayed sweep
trigger point at graticule centerline. See Fig. 5-9 for
reference.
B14.Check Delayed Triggering Level Range
a.Set the Delayed Triggering SOURCE to INT.
b.Set the vertical amplifier plug-in unit
volts/division switch to 1 V.
c.Set the low-f requency sine-wave generator for
an eight-division display at 50 kHz. Center the display
vertically on the graticule. (Lower the Delaying sweep
intensity control.)
d.Check-that all levels of the pos itive s lope of the
intensified zone may be selected for the sweep starting
point as the Delayed Triggering LEVEL control is rotated
throughout its range, and that triggering occurs at least
3.5 divisions above and below the center horizontal
graticule line. See Fig. 5-12 for reference.
e.Check-set the Delayed Triggering SLOPEtoand repeat part d for the negative slope of the waveform.
f.Check-set the Delayed Triggering SOURCEto
EXT and repeat part d of this step.
g.Check-set the Delayed Triggering SLOPE to +
and repeat part d for the positive slope of the waveform.
B15.Check Low-Frequency Triggering
B12.Adjust Delayed Triggering Sensitivity
a.Adjust-Trig Sens, R325, for a Delayed sweep
trigger point at approximately 0.5 division from bottom of
waveform as Delayed Triggering LEVEL control is
rotated through negative (-) portion of its range. See Fig.
5-10 for reference.
NOTE
Setting trigger sensitivity so display is triggered at less
than 0. 1 division should be avoided to prevent erratic
triggering.
@
a.Disconnect the 50-Ω cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b.Set the 7B92A controls as follows:
MAIN TRIGGERINGAC
COUPLING
Delayed Triggering+
SLOPE
Delayed TriggeringAC
COUPLING
5-28
Performance Check/Calibration-7B92A
(2)AC HF REJ
(3)DC
(Set the MAIN TRIGGERING LEVEL control as
necessary.)
f. Check-set the MAIN TRIGGERING SOURCE
to EXT and repeat part e of this step.
g. Disconnect the 50-Ω cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
h. Set the control settings as follows:
MAIN TRIGGERING INT
SOURCE
Delayed Triggering near "0"
LEVEL
Delayed Triggering EXT
SOURCE
i. Check-for a stable display (TRIG’D light on) at
all settings of LEVEL controls with Delayed Triggering
COUPLING set to:
(1)AC
(2)DC
(Set the MAIN and Delayed Triggering LEVEL controls
as necessary.)
LEVEL DELAY TIME
TIME/DIV OR DLY TIME 10 ms
DLY’D Time/Division 2 ms/PULL FOR ALT
DELAY TIME MULT 1.0
c. Set the vertical amplifier plug-in unit
volts/division switch to 0.2 V.
d. Set the low-frequency sine-wave generator for
a 0.5 division (100 millivolts) display at 30 Hz.
e. Check-for a stable display (TRIG’D light on) at
all settings of the MAIN TRIGGERING LEVEL control
with MAIN TRIGGERING COUPLING set to:
(1)AC
j. Check-set the Delayed Triggering SOURCE to
INT and repeat part i of this step.
B16. Check AC LF Reject Operation
a. Disconnect the 50-Ω cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b. Set the vertical amplifier plug-in unit
volts/division switch to 0.5 V.
c. Set the low-frequency sine-wave generator for
an eight-division display at 60 Hz. Center the display
vertically on the graticule.
d. Check-set the MAIN TRIGGERING LEVEL
control for a stable display (TRIG’D light on).
5-29
Performance Check/Calibration-7B92A
e. Set the MAIN TRIGGERING COUPLING to AC
LF REJ; set the MODE to NORM.
f. Check-for no display (TRIG’D light off) when
the MAIN TRIGGERING LEVEL control is rotated
throughout its range.
g. Set the MAIN TRIGGERING COUPLING to
AC; set the SOURCE to EXT.
h. Reduce the output of the low-frequency sinewave generator to provide a six-division display (3.0 V) at
60 Hz. (Adjust the MAIN TRIGGERING LEVEL control
to maintain a stable display.)
i. Set the MAIN TRIGGERING COUPLING to AC
LF REJ and repeat part f of this step.
B17. Check 20 MHz Triggering
a. Set the vertical amplifier plug-in unit
volts/division switch to 0.2 V.
b. Set the 7B92A controls as follows:
MAIN TRIGGERING AC
COUPLING
MAIN TRIGGERING AUTO
MODE
MAIN TRIGGERING INT
SOURCE
Delayed Triggering All pushbuttons in
COUPLING, SOURCE
and SLOPE
Delayed Triggering RUNS AFTER
LEVEL DELAY TIME
TIME/DIV OR DLY TIME 50 ns
DLY’D Time/Division 5 ns/PULL FOR ALT
c. Disconnect the 50-Ω cable from the low-
frequency sine-wave generator and connect the mediumfrequency signal generator to the vertical amplifier plugin unit input using the 50-Ω cable.
e. Check-for a stable display (TRIG’D light on)
with the MAIN TRIGGERING COUPLING switch set to:
(1)AC
(2)AC LF REJ
(3)DC
(Set the MAIN TRIGGERING LEVEL control as
necessary.)
f. Check-change the MAIN TRIGGERING
SOURCE to EXT and repeat part e of this step.
g. Disconnect the 50-Ω cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
h. Set the Delayed Triggering LEVEL control near
"0".
i. Set the MAIN TRIGGERING SOURCE to INT
and MAIN TRIGGERING COUPLING to AC.
j. Check-for a stable display (TRIG’D light on)
with the Delayed Triggering COUPLING switch set to:
(1)AC
(2)DC
k. Set the Delayed Triggering SOURCE to EXT
and repeat part j of this step.
I. Disconnect the 50-Ω cable from the mediumfrequency signal generator; disconnect the 50-n cable
from the bnc tee at the vertical amplifier plug-in input
connector.
B18. Check 180 MHZ Triggering
a. Connect the high-frequency leveled sine-wave
generator to the bnc tee connected tothe vertical
amplifier plug-in unit input.
b. Disconnect the 50-Ω cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
d. Set the medi um-frequency signal generator for
a 0.5 division display (100 mV) at 20 MHz.
@
c. Set the 7B92A controls as follows:
MAIN TRIGGERING DC
COUPLING
MAIN TRIGGERING EXT
SOURCE
TIME/DIV OR DLY TIME 2 ns
DLY’D Time/Division 2 ns (knob in)
5-30
Performance Check/Calibration-7B92A
d. Adjust the oscilloscope intensity control for
normal viewing.
e. Set the vertical amplifier plug-in unit
volts/division switch to 0.5 V.
f. Set the high frequency leveled sine-wave
generator for a one-division display at 180 MHz. Center
the display vertically on the graticule.
g. Set the MAIN TRIGGERING LEVEL control for
a stable display (TRIG’D light on).
h. Check-for a stable display with no more than
0.05 division of jitter.
i. Check-change the MAIN TRIGGERING
SOURCE to INT and repeat parts g and h of this step.
j. Disconnect the 50-Ω cable from the MAIN
TRIG IN connector and reconnect it to the DLY’D TRIG
IN connector.
k. Set the DLY’D Time/Division to PULL FOR
ALT; set the Delayed Triggering SOURCE to EXT.
I. Set the Delayed Triggering LEVEL control for a
stable display (near "0").
m. Check-for a stable display with no more than
0.05 division of jitter.
n. Check-change Delayed Triggering SOURCE to
INT and repeat m of this step.
e. Set the vertical amplifier plug-in unit
volts/division switch to 1 V.
f. Set the MAIN TRIGGERING LEVEL control for
a stable display (TRIG’D light on).
g. Check-for a stable display with no more than
0.05 division of jitter.
h. Set the high-frequency leveled sine-wave
generator for a one-division display.
i. Check-change the MAIN TRIGGERING
SOURCE to EXT . 10 and repeat part g of this step.
j. Disconnect the 50-Ω cable from the high-
frequency leveled sine-wave generator.
B20. Check Line Triggering
a. Disconnect the 50-Ω cable from the MAIN
TRIG IN connector.
b. Set the MAIN TRIGGERING MODE to AUTO,
SOURCE to LINE, and COUPLING to AC; the TIME/DIV
OR DLY TIME switch to l ms; and DLY’D Time/Division
to 1 ms (knob in).
c. Check-set the MAIN TRIGGERING LEVEL
control near "0" and check that the TRIG’D light is on.
d. Check-that the display is not triggered (TRIG’D
light off) at either end of the MAIN TRIGGERING LEVEL
control rotation.
B19. Check HF Sync Triggering
a. Disconnect the 50-Ω cable from the DLY’D
TRIG IN connector and reconnect it to the MAIN TRIG IN
connector.
b. Set the Delayed Triggering LEVEL control to
RUNS AFTER DELAY TIME.
c. Set the DLY’D Time/Division switch to Normal
mode (knob in).
d. Set the MAIN TRIGGERING MODE to HF
SYNC.
B21. Check Single Sweep Operation
a. Connect the time-mark generator to the input
of the vertical amplifier plug-in unit, using a 50-Ω cable;
set the generator for 1 ms markers.
b. Set the vertical amplifier plug-in unit deflection
factor for approximately two divisions of display; and
adjust MAIN TRIGGERING LEVEL control for a stable
display.
5-31
Performance Check/Calibration-7B92A
c.Set the MAIN TRIGGERING SOURCE to INT,
MODE to SINGLE SWEEP, and SOURCE to EXT.
d. Press the SINGLE SWEEP RESET
pushbutton; it should be lit.
e.Check-change the MAIN TRIGGERING
SOURCE to INT and observe that only one sweep is
displayed and SINGLE SWEEP RESET pushbutton goes
out.
f.Set the TIME/DIV OR DLY TIME switch to
Alternate mode (PULL FOR ALT).
g.Set the MAIN TRIGGERING SOURCEto EXT;
press the SINGLE SWEEP RESET pushbutton. The
pushbutton should be lit.
h.Check-change the MAIN TRIGGERING
SOURCE to INT and observe that only one sweep is
displayed and SINGLE SWEEP RESET pushbutton goes
out.
i.Press SINGLE SWEEP RESET pushbutton
several times and observe that display alternates
between Delaying and Delayed sweeps.
POSITIONMidrange
TIME/DIV OR DLY TIME1 ms
DLY’D Time/Division1 ms (knob in)
Delayed Triggering
LEVELRUNS AFTER
DELAY TIME
SLOPE+
COUPLINGAC
SOURCEINT
Inputs
TERM1 MΩ (pushbutton in)
b.Connect the time-mark generator to the vertical
amplifier plug-in unit input with a 50-Ω cable. Set the
time-mark generator for 1 ms markers.
d.Set the vertical amplifier plug-in unit for
approximately two divisions of display. Position the
display in the center graticule area. Set the MAIN
TRIGGERING LEVEL control for a stable display
(TRIG’D light on).
e.Set the front-panel SWP CAL adjustment for
one marker/division over the center eight divisions
(position as necessary).
C2. Check Alt Off Function
a.Check-for Delaying sweep (with intensified
zone) positioned at least 3.5 divisions above Delayed
sweep.
c.Check-that Delaying sweep (with intensified
zone) is not alternating with Delayed sweep, and no
readout exists.
C3. Adjust DTM Scale and Delay Start
c.Set the oscilloscope intensity and focus
controls for a desired display.
a.Set the DLY’D Time/Division switch to 10 /s;
set the TRACE SEP control fully clockwise.
5-33
b.Set the DELAY TIME MULT dial to 9.00.
c.Position the first time-mark to the second
vertical graticule line, using the POSITION control. See
Fig. 5-13 for reference.
d.Adjust-DTM Scale, R495, to start ris ing portion
of tenth time-mark so Delayed (bottom) sweep display
coincides with second vertical gratic ule line. See Fig. 513 for reference.
e.Set the DELAY TIME MULT dial to 1.00.
f.Adj ust-Delay Start, R425, to start rising por tion
of second time-mark so Delayed (bottom) sweep display
coincides with second vertical gratic ule line. See Fig. 513 for reference.
g.Interaction-repeat the adjustm ent of R495 and
R425 as necessary.
Performance Check/Calibration-7B92A
b.Adjust-Sweep Registration, R525, to align first
time-mark of Delayed sweep with first time-mark of
Delaying sweep.
c.Interac tion-repeat the adjustment of R458 and
R525 as necessary.
C6. Adjust Position Centering
a.Set the time-mark generator for 2 ms markers.
b.Set the POSITION control fully
counterclockwise.
c.Adjust-Position Centering, R935, to align the
second time-mark with the left (first) vertical graticule
line.
d.Set the POSITION control fully clockwise.
e.Check-that the sweeps start to the right of
graticule center.
C7. Check Sweep Length
Fig. 5-13. Correct DTM Scale and Delay Start
adjustment (circle).
C-4.Adjust Delaying Sweep Calibration
a.Set TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 1 ms, and pull DLY’D
Time/Division knob out for alternate mode.
b.Adjust-Delaying Sweep Cal, R458, for one
time-mark per division over center eight graticule
divisions.
C-5.Adjust Sweep Registration
a.Set DELAY TIME MULT dial to 0.0.
a.Set the time-mark generator for 1 ms markers.
b.Check-both Delaying and Delayed sweeps
must be 10.5 divisions, within 0.3 division.
C8. Adjust 2 ns Timing
a.Set the TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 2 ns (knob pushed in).
b.Set the time-mark generator for 2 ns markers.
c.Set the vert ical amplifier plug-in unit def lection
factor for a two-division display.
d.Adjust the MAIN TRIGGERING LEVEL control
for a stable display (TRIG’D light on).
e.Align the second time-mark to the second
vertical graticule line, using the POSITION control.
f.Check-that the tenth time-mark is within 0.32
division of tenth vertical graticule line.
5-34
Performance Check/Calibration-7B92A
g.Adjust-2 ns Timing, C540, for one time-mark
per division over center eight graticule divisions.
C9.Adjust Delayed Sweep 20 ns Timing
a.Set TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 20 ns (knob pushed in).
b.Set the time-mark generator for 20 ns markers.
c.Set the vert ical amplifier plug-in unit def lection
factor for a two-division display.
d.Adjust the MAIN TRIGGERING LEVEL control
for a stable display (TRIG’D light on).
e.Align the second time-mark to the second
vertical graticule line, using the POSITION control.
f.Check-that the tenth time-mark is within 0.16
division of tenth vertical graticule line.
g.Adjust-20 ns Timing, C721, for one tim e-mark
per division over center eight graticule divisions.
C10.Adjust Delaying Sweep 20 ns Timing
a.Set the TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 2 ns, and pull DLY’D
Time/Division knob out.
h.Adjust-20 ns Timing, C440, to position the
leading edge of the time-mark at the intersection of
horizontal and vertical graticule centerlines.
i.Set the DELAY TIME MULT dial to 3.00.
j.Repeat part f of this step.
k.Interac tion-repeat parts f through i of this step
as necessary.
I.Disconnect the 50-n cable from the time-mark
generator.
C11.Adjust Time Position
a.Connect a 50-Ω termination to the fast-rise
output of the calibration generator; connect a 50-Ω cable
to the termination.
b.Set the vertical amplifier plug- in unit deflection
factor for a three-division display.
c.Remove the 7B92A from the extender and
remove the extender from the oscilloscope plug-in
compartment; install the 7B92A into the plug-in
compartment.
d.Set the TIME/DIV OR DLY TIME switch to 2 ns
(knob pushed in).
b.Set the DELAY TIME MULT dial to 3.00.
c.Align the second time-mark to the second
vertical graticule line, using the POSITION control.
d.Check-that the tenth time-mark is within 0.16
division of tenth vertical graticule line.
e.Adjust-20 ns Timing, C440, for one tim e-mark
per division over center eight graticule divisions.
f.Align the display to place leading edge of
Delayed sweep time-mark at intersection of horizontal
and vertical graticule centerlines, using the POSITION
control and the vertical amplifier position control.
g.Set the DELAY TIME MULT dial to 9.00.
e.Set the MAIN TRIGGERING LEVEL control f or
a stable display (TRIG’D light on), and set the
oscilloscope intensity control for desired display.
f.Set the POSITION control fully clockwise.
g.Check-the leading edge of the pulse for a
setting to the right of graticule center.
h.Remove the 7B92A from the osc illos c ope plugin compartment and re-install the 7B92A with extender
into the compartment.
i.Adjust-Time Positioning, R925, to position
leading edge of pulse 3.5 divisions from left side of
graticule.
j.Repeat parts c and g of this step.
5-35
Performance Check/Calibration-7B92A
k. Disconnect the 50-Ω cable from the calibration
generator.
C12. Check Sweep Timing
NOTE
The tolerances given in Table 5-7 are for ambient
temperature range of +150C to +350C. If outside
this range, see the Specification section for
applicable tolerances.
Table 5-7.
SWEEP TIMING
TIME/DIV Time-Mark (+15°C to +35°C)
OR DLY'D Generator
DLY TIME Time/Division Setting Delaying Sweep Delayed Sweep
0.2 s 0.2 s 0.2 s Within 0.16 div Within 0.16 div
0.1 s 0.1 s 0.1 s Within 0.16 div Within 0.16 div
50 ms 50 ms 50 ms Within 0.16 div Within 0.16 div
20 ms 20 ms 20 ms Within 0.16 div Within 0.16 div
10 ms 10 ms 10 ms Within 0.16 div Within 0.16 div
5 ms 5 ms 5 ms Within 0.16 div Within 0.16 div
2 ms 2 ms 2 ms Within 0.16 div Within 0.16 div
1 ms 1 ms 1 ms Within 0.16 div Within 0.16 div
0.5 ms 0.5 ms 0.5 ms Within 0.16 div Within 0.16 div
0.2 ms 0.2 ms 0.2 ms Within 0.16 div Within 0.16 div
0.1 ms 0.1 ms 0.1 ms Within 0.16 div Within 0.16 div
50 µs 50 µs 50 µs Within 0.16 div Within 0.16 div
20 µs 20 µs 20 µs Within 0.16 div Within 0.16 div
10 µs 10 µs 10 µs Within 0.16 div Within 0.16 div
5 µs5 µs5 µs Within 0.16 div Within 0.16 div
2 µs2 µs2 µs Within 0.16 div Within 0.16 div
1 µs1 µs1 µs Within 0.16 div Within 0.16 div
0.5 µs0.5 µs0.5 µs Within 0.16 div Within 0.16 div
0.2 µs0.2 µs0.2 µs Within 0.16 div Within 0.16 div
0.1 µs0.1 µs0.1 µs Within 0.16 div Within 0.16 div
50 ns 50 ns 50 ns Within 0.16 div Within 0.16 div
20 ns 20 ns 20 ns Within 0.16 div Within 0.16 div
10 ns 10 ns 10 ns Within 0.24 div Within 0.24 div
a. Set the TIME/DIV OR DLY TIME, DLY’D
Time/Division in Alternate mode (PULL FOR ALT), and
time-mark generator as indicated in Table 5-7; set
DELAY TIME MULT dial to 0.0.
b. Check-using the settings given in Table
5-7, check sweep accuracy for onetime mark/division
over the center eight divisions within the tolerance given
in Table 5-7. Set the POSITION control and MAIN
TRIGGERING LEVEL control as necessar y for a stable
display that is aligned with the vertical graticule lines.
Tolerance
@
5-36
Performance Check/Calibration-7B92A
NOTE
If the time-mark generator used does not have 1-2-5
sequence markers, apply 1 unit markers in place of 2
unit markers and check for 2 markers/division over the
center eight divisions of display, to the toleranc es given
in Table 5-7.
c. Push in the DLY’D Time/Division knob and
position the display to the center of the graticule.
FAST TIMING
TIME/DIV Time-Mark
OR Generator Display Tolerance
DLY TIME Setting Markers/Division (+15°C to +35°C)
5 ns 5 ns 1 Within 0.24 div
2 ns 2 ns 1 Within 0.24 div
C13. Check Sweep Linearity
a. Set the TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 1 ms and pull DLY’D
Time/Division knob out for Alternate mode.
b. Set the time-mark generator for 1 ms markers.
c. Position the Delaying (top) and Delayed-sweep
traces horizontally to start on the first graticule line.
d. Check-that the fourth marker of the Delaying
sweep coincides within 0.1 division with the fourth
vertical graticule line.
d. Check-using the settings given in Table 5-8, check
sweep accuracy over the center eight divisions within the
tolerance given in Table 5-8.
NOTE
The HF SYNC MODE switch may provide a better
display when checking the fastest sweep rates.
Table 5-8.
g. Set the TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 10 ns and pull DLY’D
Time/Division knob out for Alternate mode.
h. Set the time-mark generator for 10 ns markers.
i. Check-perform linearity chec k as in parts d, e,
and f of this step. Tolerance is 0.1 division for each two
divisions over center eight graticule divisions.
j. Set the DLY’D Time/Division to Normal mode
(knob in).
k. Check-continue linearity check for Delayed
sweep rates as shown in Table 5-9.
e. Check-continue linearity check, as in part d of
this step, for each successive two divisions of the
Delaying sweep.
f. Check-repeat parts d and e of this step to
check linearity of the Delayed (bottom) sweep.
Table 5-9.
FAST SWEEP LINEARITY
Time-mark
Time/ Generator Markers/ Tolerance/
Division Setting 2 Divisions2 Divisions
5 ns 5 ns 2 0.2 div
2 ns 2 ns 2 0.2 div
5-37
C14.Check Variable Time/Division Range
a.Remove the 7B92A from the osc illos c ope plugin compartment.
b.Place the Time/Division Variable Selector
multi-pin connector on P761 (Delaying Sweep). See Fig.
1-1 for location.
c.Insert the 7B92A into the oscilloscope plug-in
compartment; turn on the power to the oscilloscope.
Performance Check/Calibration-7B92A
d.Set the time-mark generator for 100 p
markers.
e.Set the TIME/DIV OR DLY TIME and DLY’D
Time/Division switch to 20 µs and pull DLY’D
Time/Division knob out for Alternate m ode. Check that
the DELAY TIME MULT dial is 0.0.
f.Pr ess and release VARIABLE control and turn
it fully counterclockwise.
g.Check-that Delaying sweep markers ( top) and
Delayed sweep markers are displayed as shown in Fig.
5-8; at least 2-1/2 Delaying sweep markers to one
Delayed sweep marker are visible.
h.Remove the 7B92A from the osc illos c ope plugin compartment; reposition the Time/Division Variable
Selector multi-pin connector to P760 and re-install the
7B92A into the plug-in compartment. T urn on power to
the oscilloscope.
i.Check-with VARIABLE control released and
turned fully counterclockwise, at least two Delayed
sweep markers ( bottom) to one Delaying sweep marker
is visible. Refer to Fig. 5-14.
C15.Check Differential Delay Time Accuracy
µ
Fig. 5-14. Variable Time/Division Range
b.Set the time-mark generator for 100 ms
markers.
c.Set the vert ical amplifier plug-in unit def lection
factor for approximately two divisions of display.
d.Set the DELAY TIME MULT dial to 0.90.
e.Align the time-mark to the center vertical
graticule line, using the POSITION control.
f.Turn the DELAY TIME MULT dial to
approximately 8.9 to align the time-mark to the center
vertical graticule line.
a.Set the TIME/DIV OR DLY TIME switch to 200
ms, DLY’D Time/Division switch (pull knob out) to 5 ms
(then push knob in), push in VARIABLE (CAL IN) control,
and adjust MAIN TRIGGERING LEVEL control for a
stable display (TRIG’D light on).
g.Check-the DELAY TIME MULT dial to read
within three minor divisions of 8.90 (8.87 to 8.93).
h.Repeat parts d through g of this step f or each
sweep rate given in Table 5-10. Maintain approxim ately
two divisions of display.
5-38
Performance Check/Calibration-7B92A
Table 5-10.
DIFFERENTIAL DELAY TIME ACCURACY
Time-mark
TIME/DIV OR DLY’D Generator
DLY TIME Time/Division Setting
.2 s 5 ms 0.1 s
.1 s 2 ms 50 ms
50 ms 1 ms 10 ms
20 ms 0.5 ms 10 ms
10 ms 0.2 ms 5 ms
5 ms 0.1 ms 1 ms
2 ms 50 µs1 ms
1 ms 20 µs 0.5 ms
.5 ms 10 µs 0.1 ms
.2 ms 5 µs 0.1 ms
.1 ms 2 µs 50 µs
50 µs1 µs 10 µs
20 µs.5 µs 10 µs
10 µs.2 µs5 µs
e. Align the leading edge of the time-mark to the
center vertical graticule line, using the POSITION
control. See Fig. 5-15 for reference.
Fig. 5-15. Fast delay time accuracy.
f. Check-set the DELAY TIME MULT dial to the
settings listed in Table 5-11,’and check vor maximum
error in dial reading.
Table 5-11.
FAST DIFFERENTIAL DELAY TIME ACCURACY
DELAY TIME MULT Maximum Error
(Minor Dial Divisions)
C16. Check Fast Delay Time Accuracy
a. Set the TIME/DIV OR DLY TIME switch to 50 ns,
DLY’D Time/Division switch (pull knob out) to 10 ns(then
push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
b. Set the time-mark generator for 50 ns markers.
c. Set the vertical amplifier plug-in unit deflection
factor for approximately two divisions of display.
d. Set the DELAY TIME MULT dial to 9.00.
9.0 0 (step e)
8.0 Within 3.0
7.0 Within 3.5
6.0 Within 4.0
5.0 Within 4.5
4.0 Within 5.0
3.0 Within 3.0
2.0 Within 6.0
1.0 Within 6.5
g. Set the TIME/DIV OR DLY TIME switch to 20 ns,
DLY’D Time/Division switch (pull knob out) to 2 ns (then
push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
h. Set the time-mark generator for 20 ns markers.
5-39
Performance Check/Calibration-7B92A
i.Repeat parts d, e, and f of this step.
j.Set the TIME/DIV OR DLY TIME switch to 10
ns, DLY’D Time/Division switch (pull knob out) to 1 ns
(then push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
k.Set the time-mark generator for 10 ns markers.
I.Repeat parts d, e, and f of this step. Delete
check at DELAY TIME MULT dial setting of 1.0.
C 17.Check Delay Jitter
a.Set the TIME/DIV OR DLY TIME switch to 1
ms, DLY’D Time/Division switch (pull knob out) to 5 ns
(then push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
b.Set the time-mark generator for 1 ms markers.
c.Set the DELAY TIME MULT dial to
approximately 1.0 (position the time-mark to the center
of the graticule).
d.Set the DLY’D Time/Division switch to 0.2 us.
f.Check-jitter must not exceed one division
(disregard slow drift).
g.Set the TIME/DIV OR DLY TIME switch to 20
us, DLY’D Time/Division switch (pull knob out) to 50 ns
(then push knob in), and adjust the MAIN TRIGGERING
LEVEL control for a stable display (TRIG’D light on).
h.Set the time-mark generator for 50 ns markers.
i.Set the DLY’D Time/Division switch to 5 ns.
j.Turn the DELAY TIME MULT dial slowly to
position the time-mark near the graticule center (turn up
INTENSITY control as necessary).
k.Check-jitter must not exceed 0.9 division.
I.Turn the DELAY TIME MULT dial to
approximately 9.0 and position the time-mark to the
center of the graticule.
m.Check-jitter must not exceed 0.9 division.
This completes the Calibration procedure of the 7B92A.
e.Turn the DELAY TIME MULT dial slowly to
position the time-mark near the graticule center (turn up
INTENSITY control as necessary).
5-40
Section 6-7B92A
OPTIONS
No Options available at the time of this printing. Ref er to the CHANGE INFORMATION in the back of this manual for
Options available after this printing.
REV A, JUN 1978
6-1
Section 7-7B92A
REPLACEABLE
ELECTRICAL PARTS
PARTS ORDERING INFORMATION
Replacement parts are available from or through your local Tektronix, Inc. Field Office or
representative.
Changes to Tektronix instruments are sometimes made to accommodate improved
components as they become available, and to give you the benefit of the latest circuit impr ovements
developed in our engineering department It is theref ore important, when ordering parts, to include
the following information in your order: Part number , ins trument type or number, serial number, and
modification number if applicable.
If a part you have ordered has been replaced with a new or improved part, your local Tek tronix,
Inc Field Office or representative will contact you concerning any change in part number
Change information, if any, is located at the rear of this manual.
SPECIAL NOTES AND SYMBOLS
X000Part first added at this serial number
00XPart removed after this serial number
ITEM NAME
In the Parts List, an Item Nam e is separated from the description by a colon (:). Because of
space limitations, an Item Name may sometimes appear as incomplete For further Item Name
identification, the U.S Federal Cataloging Handbook H6-1 can be utilized where possible.
00853SANGAMO ELECTRIC CO., S. CAROLINA DIV.P 0 BOX 128PICKENS, SC 29671
01002GENERAL ELECTRIC COMPANY, INDUSTRIAL
01121ALLEN-BRADLEY COMPANY1201 2ND STREET SOUTHMILWAUKEE, WI 53204
01295TEXAS INSTRUMENTS, INC., SEMICONDUCTOR
01963CHERRY ELECTRICAL PRODUCTS CORPORATION3600 SUNSET AVENUEWAUKEGAN, IL 60085
02111SPECTROL ELECTRONICS CORPORATION17070 EAST GALE AVENUECITY OF INDUSTRY, CA
91745
03508GENERAL ELECTRIC COMPANY, SEMI-CONDUCTOR
04222AVX CERAMICS, DIVISION OF AVX CORP.P O BOX 867, 19TH AVE. SOUTHMURTLE BEACH, SC 29577
04713MOTOROLA, INC., SEMICONDUCTOR PROD. DIV.5005 E MCDOWELL RD, PO BOX 20923 PHOENIX, AZ 85036
07910TELEDYNE SEMICONDUCTOR12515 CHADRON AVE.HAWTHORNE, CA 90250
08806GENERAL ELECTRIC CO., MINIATURE
10389CHICAGO SWITCH, INC.2035 WABANSIA AVE.CHICAGO, IL 60647
11237CTS KEENE, INC.3230 RIVERSIDE AVE.PASO ROBLES, CA 93446
12617HAMLIN, INC.GROVE & LAKE STS.LAKE MILLS, WI 53551
12697CLAROSTAT MFG. CO., INC.LOWER WASHINGTON STREETDOVER, NH 03820
14193CAL-R, INC.1601 OLYMPIC BLVD.SANTA MONICA, CA 90404
15818TELEDYNE SEMICONDUCTOR1300 TERRA BELLA AVE.MOUNTAIN VIEW, CA 94043
28480HEWLETT-PACKARD CO., CORPORATE HQ.1501 PAGE MILL RD.PALO ALTO, CA 94304
32293INTERSIL, INC.10900 N. TANTAU AVE.CUPERTINO, CA 95014
32997BOURNS, INC., TRIMPOT PRODUCTS DIV.1200 COLUMBIA AVE.RIVERSIDE, CA 92507
50347OPCOA, DIVISION OF IDS330 TALMADGE ROADEDISON, NJ 08817
56289SPRAGUE ELECTRIC CO.NORTH ADAMS, MA 01247
71590CENTRALAB ELECTRONICS, DIV. OF
72982ERIE TECHNOLOGICAL PRODUCTS, INC.644 W. 12TH ST.ERIE, PA 16512
73138BECKMAN INSTRUMENTS, INC., HELIPOT DIV.2500 HARBOR BLVD.FULLERTON, CA 92634
74970JOHNSON, E. F., CO.299 10TH AVE. S. W.WASECA, MN 56093
78488STACKPOLE CARBON CO.ST. MARYS, PA 15857
80009TEKTRONIX, INC.P 0 BOX 500BEAVERTON, OR 97077
81483INTERNATIONAL RECTIFIER CORP.9220 SUNSET BLVD.LOS ANGELES, CA 90069
90201MALLORY CAPACITOR CO., DIV. OF
91637DALE ELECTRONICS, INC.P. O. BOX 609COLUMBUS, NE 68601
AND POWER CAPACITOR PRODUCTS DEPARTMENTJOHN STREETHUDSON FALLS, NY 12839
GROUPP O BOX 5012, 13500 N CENTRAL
EXPRESSWAYDALLAS, TX 75222
PRODUCTS DEPARTMENTELECTRONICS PARKSYRACUSE, NY 13201