Tektronix 7B53N Instruction Manual

TABLE
OF
CONTENTS
Introduction Simplified
Block
Diagram Discussion Main
Sweep
Delayed
Sweep
Mixed
Sweep
3-1 3-1 3-1
3-3
7B53N
Abbreviations
and
symbols
usedinthis
manual
are
based
on or
taken
directly
from IEEE
Standard
260
"Standard
Symbols
for
Units",
MIL
STD-12B,
and
other
standardsofthe
electronics
industry
.
SECTION
1
SPECIFICATION
Page
SECTION 3
CIRCUIT
DESCRIPTION
(cont)
Page
Introduction
1-1
Circuit
Operation
3-3
Table
1-1
Electrical
1-1
General Main
Trigger
Preamp
3-3 3-4
Sweep
Triggering
1-1
Trigger
Generator
3-5
Delayed
Sweep
Triggering 1-3
Sweep
Generator
3-8
Sweep
Generator
1-4
Delayed
Trigger
Preamp
Delayed
Sweep
Generator
1-4
Delayed
Trigger
Generator
Mixed
Sweep,
Variable
Delay
1-5
Delayed
Sweep
Generator
Amplifier Output
Signals
1-5 1-6
Horizontal
Preamp
Table
1-2
Environmental
Characteristics
1-7
SECTION 4
MAINTENANCE
Table
1-3
Physical
1-7
Introduction
4-1
SECTION
2
OPERATING
INSTRUCTIONS
Preventive
Maintenance
4-1
Troubleshooting
4-2
General
2-1
Corrective
Maintenance
4-6
Installation
2-1
Controls
and
Connectors
2-2
SECTION
5
CALIBRATION
General
2-2
Triggering
Controls
2-2
Introduction
5-1
Sweep
Controls
2-4
Tektronix
Field
Service
5-1
Delayed
Trigger
Controls
2-4
Using
Procedure
5-1
Front-Panel
Connectors
2-5
Equipment
Required
5-1
Sweep
Calibration
2-6
Preliminary
Control
Settings
5-2
Simplified
Operating
Instructions
2-6
Part1Performance
Check
5-4
Sweep
2-6
Index
to Part
I-Performance
Check
5-4
Magnified
Display
2-6
Preliminary
Procedure
for
Delayed
Sweep
Display
2-6
Performance
Check
5-4
Mixed
Sweep
Display
2-7
Trigger
System
Check
5-5
General
Operating
Instructions
2-7
Horizontal
System
Check
Pushbutton
Switch
Logic
2-7
Output
Signals
Check
Triggered
Light
2-7
Part11Adjustment
Trigger
2-7
IndextoPart
11-Adjustment
Trigger
Coupling
2-8
Preliminary
Adjustment
Procedure
Trigger
Source
2-9
Trigger
System
Adjustment
Trigger
Slope
Horizontal
System
Adjustment
Trigger
Level
Output
Signals
Selecting
Sweep
Rates
Measurement
SECTION
6
ELECTRICAL
PARTS
LIST
Sweep
Magnifier
Abbreviation
and
Symbols
Sweep
Operation
Parts
Ordering Information
Delayed
Sweep
Operation
IndextoElectrical
Parts
Mixed
Sweep
Operation
Electrical
Parts
Delayed
Sweep
Triggering
Delayed
Output
SECTION
7
DIAGRAMS
AND
CIRCUIT
BOARD
Applications
ILLUSTRATIONS
General
Comparison
Measurement
Symbols
and
Reference
Designators
7-1
Techniques
Diagrams
Duration
Measurements
Circuit
Board
Illustrations
Determining
Frequency
Risetime
Measurements
SECTION 8
MECHANICAL
PARTS
LIST
Delayed
Sweep
Measurements
Mechanical
Parts
Information
Delayed
Sweep
Magnification
IndexofMechanical
Parts
Triggered
Delayed
Sweep
and
Illustrations
Magnification
Mechanical
Parts
Pulse
Jitter
Measurements
Accessories
SECTION 3
CIRCUIT
DESCRIPTION
Introduction
The
7B53N
isadual
time
base unit
designed
for use
in
Tektronix
7000-Series
Oscilloscopes
withoutareadout
system
;
however
it is
compatible
with
all
7000-Series
Oscilloscopes
.
The
7B53N
features
calibrated
sweeps
from
5
seconds/division
to
50
nanoseconds/division
(5
nanoseconds/division
with
X10
magnification).The
Main
or
Delayed
sweep
rates
may
be
varied
continuously
(uncali-
brated)
between
calibrated
steps.Sweep
triggeringispro-
videdtoat least
100
megahertz,
Separate
trigger
controls
are
provided
to
select
the
desired
triggering
for
the
main
and
delayed
sweeps
.
O
SECTION
l
SPECIFICATION
TABLE
1-1
ELECTRICAL
MAIN
SWEEP TRIGGERING
7B53N
Four
display
modes
are
provided.These
include
separate
displayofthe
main
sweep,
an
intensified
display,
delayed
sweep,
and
mixed
sweep
.
The
7B53N
can
alsobeusedasan
amplifier
for
X-Y
operation
.
This
instrument
will
meet
the
electrical
characteristics
listed
under
Performance
Requirement
in
Table
1-1, fol-
lowing
complete
calibration
.
The
following
electrical
char-
acteristics
apply
over an
ambient
temperature
rangeof0°C
to
+50°C,
exceptasotherwise
indicated.Warmup
time
for
given
accuracyis20
minutes
.
Characteristic
Performance
Requirement
Supplemental
Information
Source
Internal
from
associated
vertical
unit
.
SelectedbySOURCE
switch
.
Internal
from
AC
power
source
.
External
External
dividebyten
.
Coupling
AC
SelectedbyCOUPLING
switch
.
AC
low-frequency
reject
.
AC
high-frequency
reject
.
DC
Polarity
Sweep
can
be
triggered
from
positive-
Selected
by
LEVEL/SLOPE
control
.
going
or
negative-going
portionoftrigger
signal
.
Internal
Trigger
Sensitivity
AC
0.3divisionofdeflection,
minimum, 30
The
specified
upper-3dB
frequency
hertz
to 10
megahertz;increasingto1
.5
of
the
vertical
system
supersedes
the
divisions
at
100
megahertz
.
frequency
limits
giveninthe
Internal
Trigger
Sensitivity
table
when
the
numberinthe
tableisgreater
than
the
upper-3dB
frequencyofthe
vertical
unit
.
Specification-7B53N
Performance
Requirement
0
.3
divisionofdeflection,
minimum,
30
kilohertz
to10megahertz;increasing
to
1
.5
divisionsat100
megahertz
.
0.3division
of
deflection,
minimum,
30
hertzto50
kilohertz
.
0.3division
of
deflection,
minimum,
DC
to 10
megahertz;increasingto1
.5
divi-
sion
at
100
megahertz
.
100
millivolts,
minimum,
30
hertz
to
10
megahertz
;
increasingto500
millivolts
at
100
megahertz
.
100
millivolts,
minimum,
150
kilohertz
to
10 megahertz;increasingto500
milli-
voltsat100
megahertz
.
100
millivolts,
minimum,
30
hertzto50
kilohertz
.
100
millivolts,
minimum,
DCto10
mega-
hertz;increasingto500
millivolts
at
100
megahertz
.
Stable
display
presented
with
signal
amplitudes
given
under
Internal
and
Ex-
ternal
Trigger
Sensitivity
above30hertz
.
Presentsafree-running
sweep
for
lower
frequenciesorin
absenceofa
trigger
signal
.
Main
Sweep
Generator
produces
only
one
sweep
when
triggered.Further
sweeps
are
locked
out
until
RESET
button
is
pressed.Trigger
sensitivity
sameasgiven
for
internal
and
external
sensitivity
.
1
nanosecondorless
at
75 megahertz
.
AC
LF
REJ
AC
HF
REJ
DC
External
Trigger
Sensitivity
AC
AC
LF
REJ
AC
HF
REJ
DC
Auto
Triggering
Single
Sweep
Characteristic
Internal
Trigger
Jitter
External
Trigger
Input
InputRand
C
Maximum
Safe
Input
Voltage
TABLE
1-1
(cont)
Supplemental
Information
SOURCE
switch
settoEXT
.
Trig-
gering
signal
requirements
increased
10
times
for
EXT=10
position
.
Approximately1megohm
paralleled
by 20 pF
.
500
volts
(DC
+
Peak
AC).500
volts
peak-to-peakACat1kilohertzorless
.
Level
Range
EXT
TABLE
1-1
(cont)
Characteristic

I

Performance
Requirement
EXT-10
At
least
+
and
-1.5volts
.
At
least
+ and
-15
volts
.
DELAYED
SWEEP
TRIGGERING
Specification-71353N
Supplemental
Information
Source
Internal
from
associated
vertical
unit
Selected
by
Delayed
Triggering
SOURCE
switch
.
External
Coupling
AC
Selected
by
Delayed
Triggering
COUPLING
switch
.
DC
Polarity
Sweep
can be
triggered
from
positive-
Selected
by
Delayed
Triggering
SLOPE
goingornegative-going
portionoftrigger
switch
.
signal
.
Internal
Trigger
Sensitivity
AC
0.3divisionofdeflection,
minimum,
30
hertzto10
megahertz
;
increasingto1
.5
divisionsat100 megahertz
.
DC
0.3divisionofdeflection,
minimum,
DC
to
10
megahertz,
increasing
to1.5
divi-
sionsat100
megahertz
.
External
Trigger
Sensitivity
AC
100
millivolts,
minimum, 30
hertzto10 Delayed
Triggering
SOURCE
switch
megahertz;increasingto500
millivolts
at
settoEXT
.
100
megahertz
.
DC
100
millivolts,
minimum,
DC
to
10
mega-
hertz;increasing
to
500
millivoltsat100
megahertz
.
Internal
Trigger
Jitter
1
nanosecondorlessat75
megahertz
.
External
Trigger
Input
Maximum
Safe
Input
Voltage
500
volts
(DC+Peak
AC).500
volts
peak-to-peakACat1kilohertz
or
less
.
InputRand
C
1
megohm
paralleled
by 20
picofarads
.
Level
Range
At
least
+
and
-1.5volts
.
Specification-71353N
TABLE
1-1
(cont)
MAIN
SWEEP
GENERATOR
DELAYED
SWEEP
GENERATOR
Characteristic
Performance
Requirement
Supplemental
Information
Sweep
Accuracy
Measured
in
7400-
and
7500-Series
Oscilloscopes
Time
Interval
+15°C
to
+35
°
C
0°Cto
+50°C
Over
Center
8
Divisions
Unmagnified
Magnified
Unmagnified
Magnified
50
milliseconds/division
to
Within
2%
Within2.5%
Within
3%
Within
4%
0.5microsecond/division
5
seconds/division
to
0
.1
second/division
and0.2
Within
3%
Within
3
.5%
Within
4%
Within
5%
microsecond/division
to
0
.05
microsecond/division
Sweep
Linearity
Over
any2division
portion
with-
in
center
eight
divisions
(all
Within
5%
Within
7%
sweep
rates)
VARIABLE
Sweep
Rate
Range
Continuously
variable
between
calibrated
Extends
sweep
rate
to at least12.5
sweep
rates
.
seconds/division.VARIABLE
control
internally
switchable
between
Main
and
Delayed
Sweeps
.
Sweep
Hold-Off
Time
5
seconds/division
to
10 micro-
1.5times
the
TIM
E/DIVsettingorless
seconds/division
5
microseconds/division
to0.5
2.5microsecondsorless
microsecond/division
Normal-Mag
Registration
Within0.5
division
Sweep
Accuracy
Time
Interval
Measured
in
Oscilloscopes
.
+15°C
7400-Series
or
to
+35°C
7500-Series
00Cto50
°
C
Over
Center8Divisions
Unmagnified
Magnified
Unmagnified
Magnified
50
milliseconds/division
to
Within
3%
Within
3.5%
Within
4%
Within
5%
0.5microsecond/division
0.5second/divisionto0
.1
second/division
and0.2
Within
4%
Within4.5%
Within
5%
Within
6%
microsecond/division
to
0.5microsecond/division
Within2%± measured
Main
Sweep
error
Within
2%
Within
1%
Within0.2%offull
scale(1minor
divi-
sion)
Less
than1partin20,000
of the maxi-
mum
available
delay
time
(10 times
the
Main
Sweep
TIME/DIV
switch
setting)
.
Sweep
Accuracy
DELAY
TIME
MULT
Delay
Accuracy
over
center
eight
divisions
5
seconds/division to
1
second/division
0
.5
second/divisionto1
microsecond/division
Multiplier
Linearity
Delay
Time
Jitter
TABLE
1-1
(cont)
MIXED
SWEEP,
VARIABLE
TIME
DELAY
AMPLIFIER
Specification-71353N
Exclude
the following
portions
of
Mixed
Sweep
:
First
0
.5
division
after
startofMAIN
sweep
display
and0.2
divisionor1
microsecond
(whicheverisgreater)
after
transitionofMAIN
to
DLY'D
sweep
10
millivolts/division
within
10%
.
100
mill
ivolts/division
within
10%
.
1
volt/division
within
10%
.
Deflection
Factor
EXT,
MAG
X10
EXT,
MAG
off
EXT=
10,
MAG
off
Characteristic
Performance
Requirement
Supplemental
Information
Sweep
Linearity
Over
any2division
portion
with-
in
center
eight
divisions
(all
Within
6%
Within
8%
sweep
rates)
VARIABLE
Sweep
Rate
Range
Continuously
variable
between
calibrated
Extends
sweep
ratetoat
least
1
.25
sweep
rates
.
seconds/division.VARIABLE
control
internally
switchable
between
Main
and Delayed
Sweeps
.
Sweep
Hold-Off
Time
Determined
by
the
Main
Sweep
TIME/
DIV
setting
.
Specification-71353N
Performance
Requirement
Characteristic
TABLE
1-1
(cont)
.
OUTPUT
SIGNALS
Supplemental
Information
Delayed
Sweep
Gate
Availableatfront-panel
DLY'D
TRIG
IN
connector
when
operatingina
Waveshape
Rectangular
pulse
delayed
sweep
mode
with
the Delayed
Triggering
SOURCE
switch
settoINT
.
Amplitude
+3.5volts
within
40°%
with
baselineat0
to-1volt
.
Polarity
Positive-going
Duration
For
the
timeofthe
delayed
sweep
Output
Resistance
Approximately1kilohm
Loading
At
least10kilohms
shuntedby100
picofaradsorless
.
Composite
Sweep
Gate
Coupled
to
associated
indicator
oscilloscope
by
way
of
pinA1on
In-
Waveshape
Rectangular
pulse
.
terface
circuit
board
.
Polarity
Positive-going
Duration
DISPLAY
MODE
MAIN
SWP
Coincident
with
the
main
sweep
interval
.
INTEN
Coincident
with
the
main
sweep
interval
.
DLY'D
SWP
Coincident
with
the
delayed
sweep
in-
terval
.
MIXED
Coincident
with
the
main
sweep
interval
plus
the
delayed
sweep
interval
.
Nominal
Frequency
Response
System
-3 dB
pointsin7400-Series
or
7500-Series
Oscilloscope
.
AC
Lower-3dB
40
hertz
Upper-3dB
2 megahertz
AC
LF
REJ
16
kilohertz
2
megahertz
AC HF
REJ
40
hertz
100
kilohertz
DC
DC
2 megahertz
Performance
Requirement
Rectangular
pulse
Positive-going
Coincident
with
the
delaying
sweep
(all
DISPLAY MODES)
Sawtooth
signal
with
slope
determined
by
settingofTIME/DIV
OR
DL'Y
TIME
switch
.
Sawtooth
signal
with
slope
determined
by
setting
of
DELAYED
SWEEP
Time/
Division
switch
.
Composite
Sawtooth
signal
with
slope
determinedbythe
settingofTIME/DIV
OR
DL'Y
TIME
switch
during
the main
sweep
portionofdisplay,
and
by
the
setting
of
DELAYED
SWEEP
Time/
Division
switch during
delayed
sweep
por-
tion of
display
.
Negative-going
Coincident
with
the main
sweep
interval
Ramp
duration
coincident
with
the
de-
layed
sweep
interval;total
duration
coin-
cident
with
the
delaying
sweep
interval
.
Ramp
duration
coincident
with
the
main
sweep
interval
plus
the
delayed
sweep
in-
terval;total
duration
coincident
with
the
delaying
sweep
interval
.
Auxiliary
Gate
Waveshape
Polarity
Duration
Composite Sawtooth
Waveshape
DISPLAY
MODE
MAIN
SWP
O
Polarity
Characteristic
INTEN
DLY'D
SWP
MIXED
Duration
DISPLAY
MODE
MAIN
SWP
INTEN
DLY'D
SWP
MIXED
TABLE
1-1
(cont)
TABLE
1-2

TABLE
1-3
ENVIRONMENTAL
CHARACTERISTIC
Refer
to
the
Specification
for
the
associated
oscillo-

Size
scbpe
.

Weight
Specification-7B53N
Supplemental
Information
Coupled
to
associated
indicator
oscil-
loscope
by
way
of
B5
on
Interface
board
.
Coupled
to
associated
indicator
oscilloscope
by
way
of pinsA3and
133
on
Interface
board
PHYSICAL
Fits
all
7000-series
plug-in
compartments
.
3
.25
pounds.(1
.48
kilograms)
General
Installation
SECTION
z
OPERATING
INSTRUCTIONS
The
7B53N
Dual
Time
Base
plug-in
Unit
operates
with
a
Tektronix
7000-Series
indicator
oscilloscope
and
a
7A-Series
vertical
plug-in
unit
to
form
a
complete
oscilloscope
system.To
effectively
use the
71353N,
its
operation
and
capabilities
should
be
known.This
section
describes
the
operation
of the
front-panel
controls,
gives
simplified
operating
instructions
and
general
operating
information,
and
lists
some
basic
applications
for
this
instrument
.
The
7B53Nisdesignedtooperate
in
the horizontal
plug-
in
compartment
of the
indicator
oscilloscope.This
instru-
ment
can
also
be
installedinthe
vertical
plug-in
compart-
ment
to
provideasweep
that
runs
vertically
on
the
CRT
.
However,
when
usedinthis
manner,
there
arenoretrace
blanking
or
internal triggering
provisions,
and
the
unit
may
not
meet
the
specifications
given
in
Section1.
The
instruc-
tions
in this
manual
are
written
for
use
of the
7B53N
in
the
horizontal
plug-in
compartment
.
The
7B53N
canbeoperatedinan
indicator oscilloscope
with
four
plug-in
compartments
either
independently,
in
the Alternate or
Chopped
Horizontal
Modes,
or
as
a
delayed
sweep
unit.However,
when
the
7B53N
is
operated
asadelayed
sweep
unit,itmust
be
triggered for
a
CRT
display.It
cannot
delay
another
time
base
unit,
butitcan
delay
its
own
internal
delayed
sweep
.
Before
proceeding
with
installation,
it is
necessary
to
check
the
setting
of
the
internal
Variable
Selector
switch
on
the
right
side
of
the
instrument
(see
Fig.2-1
for
location).The
Variable Selector
switch
determines
whether
the
front-panel
VARIABLE
control
operates
in
conjunction
with
the
main
or
delayed
sweeps
(see
controls
and
connectors
discussion
in
this
section
for
VARIABLE
control
operation)
.
To
install
the
7B53N
in
a
plug-in
compartment,
push
it
in
untilitfits
firmly
into
the
compartment.The
front-panel
of
the
7B53N
shouldbeflush
with
the
front-panel
of the
indicator
oscilloscope.Even
though
the gain of the
indica-
tor
oscilloscopeisstandardized
to
minimize
adjustment
when
inserting
plug-in
units,
the
sweep
calibration
of the
7B53N
should
be
checked
when
it is
installed
.
The
pro-
cedure
for
checking
the
unitisgiven
under
Sweep
Calibra-
tion
Check
in
this
section
.
7B53N
Fig.2-1.Location
of Variable Selector switch
.
To
remove
the
7B53N,
pull
the
release
latch
(see
Fig
.
2-2)todisengage
the unit
from
the
indicator
oscilloscope
and
pullitoutofthe
plug-in
compartment
.
Fig.2-2.Locationofrelease
latch
.
Operating Instructions-71353N
General
To
make
full
use
of
the
capabilitiesofthis
instrument,
the
operator
should
be
familiar
with
the
function
and
use
of
each
of
the
controls.A
brief
descriptionofthe
controls
and
connectorsisgiven
here
.
More
detailed
information
is
given
under
General
Operating
Information.Fig.2-3
shows
the
front-panel
and
external controls
and
connectors
of
the
7B53N
.
Main
Triggering
Controls
LEVEL/SLOPE

Selects
the amplitude
point
and
r
signal
on
which
the
gered
.
When
the
indi
the
outer
ringisto
the
the
sweepistriggered
ive-going
slopeofthe
trigger
signal, as
shown
by
the
g
waveform.To
the
ter,
the
sweepistrig-
negative-going
slope
of
signal as
shown
by
the
negative-going
waveform
.
2-2
CONTROLS
AND
CONNECTORS
slopeoftrigg sweepistri cator
line
on
left
of
cente
on
the
posi
positive-goi rightofce
geredonth the
trigger
TRIG'D

Lamp
indicates
that
the
sweep
is
triggered
and
will
produceadisplay
with
correct
setting
of
the
POSITION
control
and
the
controls
on
the
associated
vertical
unit
and
indicator
oscilloscope
.
MODE

Four
pushbutton
switchestoselect
the
desired
trigger
mode
.
AUTO
:
Sweep
initiatedbythe
applied
trigger
signal at
point
selectedbythe
LEVEL/SLOPE
control
when
the
trigger
signal
repetition
rateisabove about
30
hertz
and
within the
frequency
range
selected
by
the
COU-
PLING
switch.Triggered
sweep
can be
obtained only over
the
amplitude
rangeofthe
applied
trigger
signal
.
When
the
LEVEL/
SLOPE
controlisoutside
the
amplitude
range,
the
trigger
repetition
rateisoutside
the
fre-
quency
range
selected
by the
COUPLING
switch,orthe
trig-
ger
signalisinadequate,
the
sweep
free-runsatthe
sweep
rate
selected
by
the
TIME/DIV
OR
DL'Y
TIME
switch
.
COUPLING
NORM
:
Sweep
initiatedbythe
applied
trigger
signalatpoint
se-
lected
by
the
LEVEL/SLOPE
control
over the
frequency
range
selected
by
the
COUPLING
switch
.
Triggered
sweep
can
be
obtained only over
the
ampli-
tude
range
of the
applied
trigger
signal.Thereisno
trace
when
the
LEVEL/SLOPE
control
is
outside
the
amplitude
range,
the
trigger
repetition
rateisoutside
the frequency
range
selected
by
the
COUPLING
switch,
or the
trigger
signalisinadequate
.
SINGLE
SWP:After
a
sweepisdis-
played, further
sweeps
cannot
be
presented
until
the
RESET
but-
ton
is
pressed.Displayistrig-
geredasfor
NORM
operation,
using
the
MAIN
TRIGGERING
controls
.
RESET-READY
:
Pushbutton
to
reset
the
Sweep
Generator
for
the
next
sweep
in
the
SINGLE
SWP
mode.The
RESET-READY
button
remains
illuminated
to
indicate
that
the
unitisready
to
be
triggered.After
the
unit
is
triggered
andasweepiscom-
pleted,
the
RESET-READY
light
goes
out
until
the
button
is
pressed again
.
Four pushbutton
switchestoselect
trigger
coupling
.
AC
:
RejectsDCand
attenuates
AC
signals
below about
30
hertz
.
Accepts
signals
between
30
hertz
and 100
megahertz
.
AC
LF
REJ:RejectsDCand
atten-
AC
HF
REJ:Accepts
signals
be-
tween about
30
hertz
and
50
kilohertz
.
RejectsDCand
atten-
uates
signals
from
50
kilohertz
to
100
megahertz
.
O
uates
signals
below
about
30
kilohertz.Accepts
signals
be-
tween
30
kilohertz
and 100
megahertz
.
SOURCE
O
INT:Trigger
signal
obtained
in-
ternally
from
vertical
unit
by
way
of
associated
indicator
oscilloscope
.
Fig.2-3.Front-panel
controls
and
connectors
.
DC:Accepts
all
trigger
signals
from
DC
to
100
megahertz
.
Four
pushbutton
switches
to
select
the
trigger
source
.
Operating
Instructions-71353N
LINE:Trigger
signal
obtained
in-
ternally
from
a
sample
of the
line
voltage appliedtoassociated
indicator
oscilloscope
.
EXT:Trigger
signal
obtained
from
an
external
signal
applied to the
MAIN
TRIG
IN
connector
.
EXT
=10:Trigger
signal
obtained
from
an external
signal
applied
2-
3
Operating
Instructions-71353N
Sweep
Controls
POSITION
X10
MAG
SWP
CAL
TIME/DIV
OR
DL'Y
TIME
VARIABLE
2-
4
to
the
MAIN
TRIG
IN
con-
nector.In
this
position
the
ex-
ternal signalisattenuated
10
times
beforeitis
appliedtothe
trigger
circuit
.
Controls
the
horizontal position
of
trace
.
Increases
sweep
rateofboth
the
main and
delayed
sweeps
ten
times
by
horizontally expanding
the
center
division
of
display
.
Light
indicates
when
magnifierison
.
Screwdriver
adjustmenttoset
hori-
zontal
gain
of unit.Usedtoset
the
basic
timingofthe
7B53N
and
to
compensate
for
differencesinCRT
deflection factor
when
changing
indicator oscilloscopes
.
Combination
switch
selects
the
sweep
rates
for
both
the
main
and
delayed
sweep
generators
.
The
clear
plastic
flange
indicates
the
sweep
rateofthe main
sweep
circuit
for
main
sweep
display
only
and
indi-
cates
the
basic
delay
time
.(to
be
multipliedbythe
DELAY
TIME
MULT
dial
setting)
for
delayed
or
mixed sweep
operation.The
DELAYED
SWEEP
knob,
when
pulled
out and
rotated
clockwise,
selects
the
sweep
rateofthe
de-
layed
sweep
circuit
for
mixed
or
delayed
sweep
operation.The
VARIABLE
control
mustbein
the
CAL
position
and the
X10
MAG
switch
must
be
off
for
indicated
sweep
rate
.
Two
position switch
actuated
by
the
VARIABLE
knobtoselect
cali-
brated
or uncalibrated
sweep
rates
.
In
the
CAL
position
(pushed
in)
the
VARIABLE
controlisinoperative
and
the
sweep
rateiscalibrated
.
When
pressed
and
released,
the
knob
moves
out
to
activate
the
VARIABLE
control
for
uncali-
brated
sweep
rates.The
sweep
rate
in
each
TIME/DIV
switch
position
DISPLAY
MODE
DELAY
TIME
MULT
Delayed
Trigger
Controls
LEVEL
can be reducedatleasttothe
sweep
rate
of
the
next
slower
position
.
The
VARIABLE
control
can
be
switchedtoeither
the main or
de-
layed
sweeps by
meansofthe
inter-
nal
Variable
Selector
switch
(see
In-
stallationinthis
section)
.
Four
pushbutton
switchestoselect
the
desired display
mode
.
MAIN
SWP:Sweep
rates
deter-
mined
by
the
TIME/DIV
OR
DL'Y
TIME
switch
.
INTEN:In
this
mode,aportion
of
the
main
sweep
is
intensified
during
the
time
that
the
delayed
sweepisin
operation,asdeter-
mined
by
the
DLY'D
SWEEP
Time/Division
switch
.
D
LY'D
SWP:The
DLY'D
SWP
mode
isafunction
of
the
delayed
sweep.In this
mode,
the
delayed
sweepisdisplayedata
rate
determined
by
the
DLY'D
SWP
Time/Division
switchatthe
end
of
each
delay
period,
as
determined
by
the
TIME/DIV
OR
DL'Y
TIME
switch
and
the
DELAY
TIME
MULT
dial
set-
tings
.
MIXED:The
MIXED
mode
is
a
combination
of
the
main
and
delayed
sweeps.The
main
sweep
is
displayedonthe
CRT
to
a
point
determined
by
the
DELAY
TIME
MULT
dial;the
remainderofthe
sweepisat
the
rate
determined
by the
delayed
sweep
.
Provides
variable
delayof0to10
times
the
basic
delay
time
selected
by
the
TIME/DIV
OR
DL'Y TIME
switch
.
Dual
function
controltodetermine
the
mode
and
level
for
delayed
sweep
display
.
O
SLOPE
COUPLING
SOURCE
IN-RUNS
AFTER
DLY
TIME
:
The
delayed
sweep
runs
immedi-
ately
following
the
delay
time
selectedbythe
TIME/DIV
OR
DL'Y
TIME
switch
and
the
DE-
LAY
TIME
MULT
dial.Delayed
SLOPE,
COUPLING,
SOURCE,
and
LEVEL
functions
inoperative
.
OUT-DLY'D
SWP
TRIGGER-
ABLE:When
the
LEVEL
con-
trolispressed
and
released
the
delayed
sweepistriggerable.The
LEVEL
control
can
now
be
ro-
tated
to
select
the
amplitude
point
on
the
trigger
signal at
which
the
delayed
sweepistrig-
gered.In
the
OUT-DLY'D
SWEEP
TRIGGERABLE
posi-
tion
the
delayed
SLOPE,
COUPLING,
and
SOURCE
func-
tions
are
activated
.
Two
position
switchtoselect
the
portion
of
trigger
signal
which
starts
the
delayed
sweep
.
+
:
The
delayed
sweep
canbetrig-
gered
from
positive-going
por-
tionoftrigger
signal
.
-:The
delayed
sweep
canbetrig-
gered
from
negative-going
por-
tion
of
trigger
signal
.
Two
position
switch
to
determine
the
method
of
coupling
delayed
trigger
signaltothe
delayed
trigger
circuit
.
AC:RejectsDCand
attenuates
AC
signals
below
about
30
hertz
.
DC:Accepts
trigger signals
from
DCto100
megahertz
.
Two
position
switch to
select
sourceofthe
delayed
trigger.Also
determines
the
function
of
the
DLY'D
TRIG
IN connector
.
INT
:
The
delayed
trigger
signal
is
obtained
from
the
vertical
ampli-
fier
unitbyway
of
the
associ-
ated
indicator
oscilloscope
.
Also
Front-Panel
Connectors
MAIN TRIG
IN
DLY'D
TRIG
IN
Operating
Instructions-71353N
connects
the
Delayed
Gate
Out
signaltothe
DLY'D
TRIG
IN
connector
for
external
use
.
EXT:The
delayed
trigger
signal
is
obtained
fromanexternal
signal
appliedtothe
DLY'D
TRIG
IN
connector
.
Front-panel
BNC
connector
serving
two
different
input
functions,
depending
upon
the
settingofthe
TIME/DIV
OR
DL'Y
TIME
switch
and
the
MAIN
TRIGGERING
SOURCE
switch
.
MAIN
TRIG
IN:External
trigger
input
for Jhe
Main
Triggering
cir-
cuit.The
SOURCE
switch
for
MAIN
TRIGGERING
must
be
set
to
EXT
or
EXT
=10
and
TIME/DIV
OR
DL'Y
TIME
switch
settoany
position
except
AMPL
.
AMPL:When
the
TIME/DIV
OR
DL'Y
TIME
switchisset
to
AMPL
and
the
MAIN
TRIG-
GERING
SOURCE
switchisset
to
the
EXT
or
EXT
=10
posi-
tion,
this
connector
serves
as an
External
Horizontal
Input
.
Front-panel
BNC
connector
serving
two
different
functions
depending
upon
the
settingofthe
Delayed
Triggering
SOURCE
switch
.
DLY'D
TRIG
IN:When
the
Delayed
Trigger
SOURCE
switch
is
settoEXT,
this
connector
servesasan
external
trigger
in-
put
for the
delayed
triggering
circuit
.
Delayed
Gate
Output:When
the
Delayed
Trigger
SOURCE
switch
is
settoINT
the
DLY'D TRIG
IN
connector
servesasa
Delayed
Gate
Output.The
Delayed
Gate
signal
is
a
rectangular
positive-
going
pulse
with
approximately
3.5volts
amplitude
and
pulse
width
coincident
with
the
delayed
sweep
.
2-5
Operating
Instructions-71353N
Sweep
Calibration
Check
Whenever
the
7B53Nisinserted intoaplug-in
compart-
ment
of an
indicator oscilloscope
other
than
the
one
in
which
it
was
originally
calibrated,
the
sweep
calibration
must
be
checked,
and
readjusted
if
necessary.Install
the
7B53N
into
the
plug-in
compartment
of
the
indicator
oscilloscope
and
allowatleast
20
minutes
warmup
before
proceeding
with
the following
:
For
accurate
sweep
timing,
applyasignal
of
known
fre-
quency
or time
period
(time-mark
signal,
calibrator
square-
vyave,
60-hertz
line,
etc.)to
the
associated
vertical
amplifier
unit
and
adjust
the
TIME/DIV
OR
DL'Y
TIME
switch
and
the
SWP CAL
controltocalibrate
that
signaltothe
oscillo-
scope
graticule.The
following
method,
using
a
Tektronix
2901
Time Mark
Generator,
is
recommended
.
1.Connecta1
millisecond
time-mark
signal
from
the
time-mark
generator
through a
50-ohm
BNC
coaxial
cable
to
the
Input
of
the
associated
vertical
unit
.
2.Set the
7B53N
TIME/DIV
OR
DL'Y
TIME
switch
to
1msand
press
the
VARIABLE
controltothe
CAL
posi-
tion.Press
the
MAIN
TRIGGERING AUTO,
AC,
and
INT
switches
.
3.Set
the
vertical
unit
foraCRT
display
amplitude
of
twotofour
divisions
usingDCcoupling
.
4.Rotate
the
LEVEL/SLOPE
control
foratriggered
display
.
5
.
Check
the
CRT
display
for
one
marker
each
major
graticule
division
over
center
eight
divisions
.
6.Adjust
SWP
CAL
(front-panel
screwdriver
adjust-
ment),
for
one
marker
each
major
division.Use
the
POSITION
controlasnecessarytoalign
the
display
with
the
vertical
graticule
lines
.
The
following
informationisprovidedtoaidinquickly
obtaining the
correct
settings
for the
7B53N
to
present
a
display.The
operator
shouldbefamilar
with the complete
function
and
operationofthe
unit
describedinthis
section
before
using
this
procedure
.
Main
Sweep
The
following
procedure
will
provideastable
display
for
most main
sweep
applications
:
2-6
SIMPLIFIED
OPERATING
INSTRUCTIONS
1.Select
the
MODE,
COUPLING,
and
SOURCE
push-
buttonofthe
MAIN
TRIGGERING
switches
which
fit
the
requirements
of
the
signal
to be
displayed
.
2.Press
the
MAIN
SWP
pushbuttonofthe
DISPLAY
MODE
switch
.
3.Turn
the
LEVEL/SLOPE
control
to
the
desired
slope
.
4.If
the
green
TRIG'D
lightisnot
on,
adjust
the
LEVEL/SLOPE
control
throughout
the
range
of
the
selected
SLOPE
until
the
TRIG'D
light
comes
on
.
If
the
light
does
not
come
onatany
setting
of
the
LEVEL/
SLOPE
control,
the
trigger
signalisnot
adequate
or
the
COUPLING
and
SOURCE
switches
are
set
incorrectly
.
5.Set
the
TIME/DIV
or
DL'Y
TIME
switch
and
POSITION
control
foradisplay
which
remains
on
the
dis-
play area horizontally
(VARIABLE
control
shouldbein
CAL
position
for
calibrated
sweep
rates)
.
6
.
If
the
display
does not
startatthe
correct
point
on
the
waveform,
readjust
the
LEVEL/SLOPE
control
for
the
desired
triggering
.
Magnified
Display
A
magnified-sweep
display
canbeobtainedasfollows
after
a
normal-sweep
displayisobtained
:
1.Adjust
the
POSITION
control
to
move
the
area to
be
magnified
within the
center
divisionofthe
CRT
.
2.If
necessary,
change
the
TIME/DIV
or
DELAY
TIME
switch
settingsothe
complete
area tobemagnified
is
within
the
center
graticule
division
.
3.Press
and
release
the
X10
MAG
switch.Light
indi-
cates
when
magnifierison
(the
indicator
oscilloscope
Inten-
sity
may
need to be
increasedtoview
magnified
display)
.
4.Use the
POSITION
control
to
establish
precise
positioningofthe
magnified
display
.
Delayed
Sweep
Display
The
following
procedureisrecommended
for use
of
INTEN
and
DLY'D
SWP
DISPLAY
MODES.Obtainadis-
play as previously
given
under
Main Sweep,
then
proceed
as
follows
:
O
1.Press
the
X10
MAG
switchin(off)
and
press
the
Delayed
Triggering
LEVEL
control
to
the
IN-RUNS
AFTER
DLY
TIME
position
.
2
.
Press
the
INTEN
pushbuttonofthe
DISPLAY
MODE
switch.Pull
out
the
DLY'D
SWEEP
Time/Division
switch
and
rotate
clockwisetoobtain the
amountofmagnification
desired.Note
that
a normal
sweep
(Main Sweep)
with
an
intensified
portion
(Delayed
Sweep)isdisplayedonthe
CRT.The
indicator
oscilloscope
Intensity
may
have
to be
decreased
to
view
the
intensified
display
.
3.Rotate
the
DELAY
TIME
MULT
dial
and
note
that
the
intensified
portionofthe
displayiscontrolledbythe
DELAY
TIME
MULT
.
4.Press
the
DLY'D
SWP
button
of
the
DISPLAY
MODE
switch
.
Note
the
magnified
display,asdetermined
by the
DLY'D
SWEEP
Time/Division
switch.The
Intensity
may
needtobe
increasedtosee the
delayed
sweep
display
.
5.Foradelayed
sweep
with
less
jitter,
pressinand
release
the
Delayed
Triggering
LEVEL
control.Delayed
triggering
SLOPE,
COUPLING,
and
SOURCE
switches
shouldbesetto+,
AC,
and
INT
respectively
.
Rotate the
Delayed
Triggering
LEVEL
control
forastable
delayed
sweep
display
.
Mixed
Sweep
Display
The
procedure
that
follows
can be
used
for
basic
mixed
sweep
operation.Obtain a
display
as
previously
given
under
Main
sweep.Then
proceedasfollows
:
1.Press
the
MIXED
pushbutton
of
the
DISPLAY
MODE
switch
and
press
and Delayed
Triggering
LEVEL
controltothe
IN-RUNS
AFTER
DLY
TIME
position
.
2.Pull
out
the
DLY'D
SWEEP
Time/Division
switch
and
turn
clockwisetoset
the
sweep
rate
for
the
delayed
portionofdisplay
.
3.Adjust
DELAY
TIME
MULT
dialtovary
the
point
at
which
the
display
switches
from
the
MaintoDelayed
sweep
rate
.
4.If
less
jitter
is
desired
for
the
delayed
sweep
portion
of
the
mixed
sweep
display,
pressinand
release
the De-
layed
Triggering
LEVEL
control.Delayed
Triggering
SLOPE,
COUPLING,
and
SOURCE
switches
shouldbeset
to
+,
AC,
and
INT
respectively.Rotate
the
Delayed
Triggering
LEVEL
controltotrigger
the
delayed
sweep
portionofdisplay
.
GENERAL
OPERATING
INSTRUCTIONS
Pushbutton
Switch
Logic
Operating
Instructions-71353N
The
MODE,
COUPLING,
and
SOURCE
pushbuttons
of
the
MAIN
TRIGGERING
switches
and
the
DISPLAY
MODE
pushbuttons
are
arrangedina
sequence
which
places
the
most-often
used
position at
the
topofeach
series
of
pushbuttons.With
this
arrangement,astable
display
can
usuallybeobtained
by
pressing
the
top pushbuttons
:
AUTO,
AC,
INT,
and
MAIN
SWP.When
an
adequate
trigger
signalisapplied,
the
unitistriggered
as
indicated
by the
illuminated
TRIG'D
light,
with
the
correct
settingofthe
LEVEL/SLOPE
control.If
the
TRIG'D
lightisnot
on,
the
LEVEL/SLOPE
controlisatasetting
outside
the
range
of
the
trigger
signal
appliedtothis
unit
from
the
vertical
unit
;
the
trigger
signalisinadequate,orits
frequency
is
below
the
lower
frequency
limit
of
theACCOUPLING
switch
posi-
tion.If
the
desired display
is
not
obtained with
these push-
buttons
pushed
in,
other
selections
mustbemade.Refer
to
the following
discussionsorthe
instruction
manuals
for the
associated
indicator oscilloscope
and
vertical
unit
for
more
information
.
Triggered
Light
The
TRIG'D
light
provides
a
convenient
indication
of
the
condition
of
the
triggering
circuits.If
the
MAIN
TRIGGERING
controls
are
correctly
set
andanadequate
trigger
signalisapplied,
the
TRIG'D
lightison.Under
certain
conditions,
the
TRIG'D
light
maybeoff,
indicating
that
the
sweepisnot
triggered.The
cause
might beamis-
adjusted
LEVEL/SLOPE
control,
incorrectly
set
COUPLING
or
SOURCE
switches,
low
trigger
signal
ampli-
tude,ora
trigger
signal
repetition
rate
outside
the
accept-
able
frequency
range.This
light
can be
usedasa
general
indicationofcorrect
triggering.It is
particularly
useful
when
settingupthe
trigger circuits
whenatrigger
signal
is
available
withoutadisplayonthe
CRT
.
Main
Trigger
Mode
The
pushbuttons
located
under
the
MODE
title
select
the
modeinwhich
the main
sweepistriggered
.
AUTO.When
the
AUTO
pushbuttonispressed,atrig-
gered
display
is
presented
with
the
correct
settingofthe
LEVEL/SLOPE
control
(see
Trigger
Level
discussion)
whenever
an adequate
trigger
signalisapplied.The
TRIG'D
light
indicates
when
the
displayistriggered
.
When
the
trigger
repetition
rateisbelow about30hertz
(or
outside
the
frequency
range
selectedbythe
COUPLING
switch)orwhen
the
trigger
signalisinadequate,
the
sweep
free
runsatthe
sweep
rate
indicated
by the
TIME/DIV
or
DELAY
TIME
switch
(TRIG'D
light
off)
.
When
an
ade-
quate
trigger
signal
is
again
applied,
the
free-running con-
2-
7
Operating Instructions-71353N
dition
ends
and
a
triggered
displayispresented.When
the
LEVEL/SLOPE
controlisatasetting
outside
the
amplitude
rangeofthe
trigger
signal,
the
sweep
also
free
runs
at
the
sweep
rate
indicatedbythe
TIME/DIV
or
DELAY
TIME
switch.This
type
of free-running
display
can be
useful
when
it is
desiredtomeasure
only
the
maximum
peak-to-
peak
amplitude
ofasignal
without
observing
the
waveshape
(suchasin
bandwidth
measurements)
.
When
the
display
is
ofamuch
greater
amplitude
than
canbedisplayedonthe
CRT,
the
sweep
willbetriggeredinall
positionsofthe
LEVEL/SLOPE
control
and
will
not
free-run
.
NORM.When
the
NORM
pushbuttonispressed,atrig-
gered display
is
presented
with
the
correct
settingofthe
LEVEL/SLOPE
control
wheneveranadequate
trigger
signal
is
applied
.
The
TRIG'D
light
indicates
when
the
display
is
triggered
.
The
NORM
trigger
mode
mustbeusedtoproduce
trig-
.
gered
displays
with
trigger
repetition
rates
below about
30
hertz.When
the
LEVEL/SLOPE
controlisatasetting
out-
side
the amplitude
rangeofthe
trigger
signal,
when
the
trigger
repeitition rateisoutside
the
frequency
range
selectedbythe
COUPLING
switch,
or
when
the
trigger
signalisinadequate,
thereisno
trace
(TRIG'D
lightisoff)
.
SINGLE
SWEEP.When
the
signal
to be
displayed
is
not
repetitiveorvariesinamplitude,
waveshape,
or
repetition
rate,aconventional
repetitive
type
display
may
produce
an
unstable
presentation.A
stable
display
can
often
be
obtained
under
these
circumstancesbyusing
the
single-
sweep
feature
of
this
unit.The
single-sweep
modeisalso
useful
to
photograph
non-repetitiveorunstable
displays
.
To
obtainasingle-sweep
displayofa
repetitive
signal,
first
obtain the
best
possible
displayinthe
NORM
MODE
.
Then
without
changing
the
other
MAIN
TRIGGERING
switches,
press
the
SINGLE
SWP
pushbutton.When
ready
to
view
the
single-sweep
display,
press
the
RESET-READY
pushbutton.A
single
traceispresented
each time
the
RESET-READY
pushbutton
is
pressed
(as
longasthe
repetitive
signal
remains
connected
to
the
system
and
MAIN
TRIGGERING
switches
are
correctly
set);further
sweeps
cannot be
presented
until
the
RESET-READY
push-
buttonispressed again.If
the
displayed
signalisa
complex
waveform composed
of pulses of
varying
amplitude,
successive
single-sweep
displays
may
not
start at
the
same
pointofthe
waveform.To
avoid
confusion
duetothe
CRT
persistence,
allow
the
displaytodisappear
before
pressing
the
RESET-READY
pushbutton
again.At
fast
sweep
rates,
it
may
be
difficulttoview
the
single-sweep
display.The
apparent
trace
intensity
canbeincreasedbyreducing
the
ambient
light
levelorusingaviewing
hoodasrecommended
in
the
indicator
oscilloscope
instruction
manual
.
Non-repetitive,
random
signals
can be
displayedinthe
single-sweep
mode
by
first
obtaining,the
best possible
dis-
2-
8
playinthe
NORM MODE
withasignal
whichisabout
the
same
amplitude
and
frequencyasthe
random
signal.Then
without
changing
the
other
MAIN
TRIGGERING
controls,
press
the
SINGLE-SWP
pushbutton.When
ready
for
the
random
signal,
press
the
RESET-READY
pushbutton.The
RESET-READY
pushbutton
remains
illuminated
to
indi-
cate that
the
unit
has
been
reset
andisreadytoproduce
a
sweep
.
The
light
goes
out
after
the
single
sweep
has
been
displayed.To
prepare
the
unit
for
another
single-sweep
dis-
play, press
the
RESET-READY
pushbutton
.
When
using
the
single-sweep
mode
to
photograph wave-
forms, the
graticule
mustbephotographed
separatelyinthe
normal
manner
to
prevent over-exposing
the
film.Be
sure
the
camera
systemiswell
protected
against
stray
light,
or
operate
the
system
inadarkened
room.For
repetitive
waveforms,
press
the
RESET-READY
pushbutton
only
once
for
each
waveform
unless
the
signal
is
completely
symmetrical.Otherwise,
multiple
waveforms
may
appear
on
the
film.For
random
signals,
the
lens
can be
left
open
until
the
random
signal triggers
the
unit
(RESET-READY
push-
button
illuminated).Further informationonphotographic
techniqueisgiveninthe
appropriate
camera
instruction
manual
.
RESET-READY
.
The
RESET-READY
pushbutton
re-
sets
the
main sweep
generator
for
the
next
sweep
when
operatinginthe
SINGLE
SWP
MODE.See
the
preceeding
Single
Sweep
discussion
for
more
information
.
Main
Trigger
Coupling
The
MAIN
TRIGGERING
pushbuttons
located
below
the
COUPLING
title
select
the
method
in
which
the
trigger
signalisconnectedtothe
trigger
circuits.Each
position
permits
selection
or
rejectionofthe
frequency
components
of
the
trigger
signal
which
trigger
the
sweep.Fig.2-4 graph-
ically
illustrates
the
bandoffrequencies
coveredbyeach
positionofthe
COUPLING
switch
.
AC.In
the
AC
position
of the
COUPLING
switch,
the
DC
component
of
the
trigger
signal
is
blocked.Signals
with
low-frequency
components
below about
30
hertz
are
attenuated.In
general,ACCOUPLING
can be
used
for
most
applications.However,ifthe
signal
contains
unwanted
frequency
componentsorif
the
sweepisto
be
triggeredata
low
repetition
rateorDC
level,
one
of
the
remaining
COUPLING
switch
positions
will
provideabetter
display
.
The
triggering
point
in
the
AC
position
of the
COU-
PLING
switch
depends
upon
the
average voltage
levelofthe
trigger
signal.If
the
trigger
signal
occurs
randomly,
the
average
voltage
level will
vary,
causing
the
triggering
point
to vary
also.This
shift
of
the
triggering
point
may
be
enough
so
it is
impossibletomaintainastable
display.In
such
cases,
useDCcoupling
.
O
AC
LF
REJ.In
the
AC
LF
REJ
position
of the
COU-
PLING
switch,
DC
is
rejected
and
low-frequency
trigger
signals
below about
30
kilohertz
are
attenuated.Therefore,
the
sweepistriggered
only
by
the
higher-frequency
com-
ponents
of the
trigger
signal.This
positionisparticularly
useful
for
providing
stable
triggeringifthe
trigger
signal
contains
line-frequency
components.Also, the
AC
LF REJ
position
provides
the
best
alternate-mode
vertical
displays
at fast
sweep
rates
when
comparing twoormore
unrelated
signals
.
AC
HF
REJ.The
AC HF
REJ
position
of the
COU-
PLING
switch
passes
all
low-frequency
signals
between
about
30
hertz
and
50
kilohertz
.
DCisrejected
and
signals
outside the
above
range
are
attenuated.When
triggering
from
complex
waveforms,
this
positionisusefultoprovide
a
stable
display
of the
low-frequency
components
.
DC.The
DC
positionofthe
COUPLING
switch
can
be
usedtoprovide
stable
triggering
with
low-frequency
signals
which
would
be
attenuatedinthe
other
modes,
or
with
low-repetition
rate
signals.It
can
alsobeusedtotrigger
the
sweep
when
the
trigger
signal
reaches aDClevel
selected
by
the
setting
of the
LEVEL/SLOPE
control.When
using
internal
triggering,
the
settingofthe
vertical
unit
position
controls
affects
theDCtriggering
point
.
Fig.2-4.Frequency
rangeofeach
COUPLING
switch position
.
Main
Trigger
Source
Operating
Instructions-71353N
The
MAIN
TRIGGERING
pushbuttons
located
below
the
SOURCE
title
select
the
source
of the
trigger
signal
whichisconnectedtothe
main
trigger
circuits
.
INT.In
the
INT
positionofthe
SOURCE
switch, the
trigger
signalisderived
from
the
associated
vertical
unit
.
Further
selection
of the
internal
trigger
signal
may
be pro-
vided
by
the
associated
vertical
unit
or
indicator
oscillo-
scope;see the
instruction
manuals
for these
instruments
for
information.For
most
applications,
theINT
position
of the
SOURCE
switch
canbeused.However,
some
applications
require
special
triggering
which
cannot
be
obtainedinthe
INT
position
of the
SOURCE
switch.In
such
cases
LINE
or
EXT
positions of the
SOURCE
switch
mustbeused
.
LINE.The
LINE
position
of the
SOURCE
switch con-
nectsasample
of the
power-line
voltage
from
the
indicator
oscilloscope
to the
trigger circuit.Line
triggeringisuseful
when
the
input
signalistime-related (multiple
or
sub-
multiple)
to
the
line
frequency.Itisalso
useful
for
pro-
viding a
stable
displayofa
line-frequency
component
in a
complex
waveform
.
EXT.An
external
signal
connected
to the
MAIN
TRIG
IN
connector
canbeused to
trigger
the
sweepinthe
EXT
2-
9
Operating
Instructions-71353N
position of
the
SOURCE
switch.The
external
signal
must
be
time-relatedtothe
displayed
waveform
forastable
dis-
play.An
external
trigger
signal
can be
usedtoprovide
a
triggered
display
when
the
internal
signalistoo
low
in
amplitude
for
correct
triggering,orcontains
signal
com-
ponentsonwhichitis
not
desiredtotrigger.Itisalso
useful
when
signal
tracinginamplifiers,
phase-shift
networks,
wave-shaping
circuits,
etc.The
signal
fromasingle
point
in
the
circuit
under
test
can be connectedtothe
EXT
TRIG
IN connector throughaprobeorcable.The
sweepisthen
triggered
by the
same
signalatall
times
and
allows
ampli-
tude,
time
relationship,orwaveshape
changesofsignals
at
various
points
in
the
circuittobe
examined
without
re-
setting
the
MAIN
TRIGGERING
controls
.
EXT
=10.Operationinthe
EXT
=10
position of
the
SOURCE
switchisthe
same
as
described for
EXT
except
that
the
external
signalisattenuated10times.Attenuation
of
high-amplitude
external
trigger signalsisdesirable
to
broaden
the
rangeofthe
LEVEL/SLOPE
control
.
Main
Trigger
Slope
The
LEVEL/SLOPE
control
determines
whether
the
trigger circuit
responds
on
the
positive-going
or
negative-going
portion
of
the
trigger
signal.When
the
indi-
cator
line
on the
outer
ring
of the
LEVEL/SLOPE
control
istothe
leftofcenter,
the
dispaly
startsonthe
positive-
going
portionofthe
waveform
(notice
positive-going
wave-
formtoleftofcontrol).To
the
right
of
center,
the
display
startsonthe
negative-going
portion
of
the
waveform
(notice negative-going
waveform).Fig.2-5
illustrates
the
operationofthe
LEVEL/SLOPE
controlatdifferent
levels
and
slopesofthe displayed
waveform.When
several
cycles
of
a
signal
appearinthe
display,
the
selectionof-the
trigger
slopeisoften
unimportant.However,ifonlyacertain
port-
ionofa
cycleistobedisplayed,
correct
settingofthe
LEVEL/SLOPE
controlisimportanttoprovideadisplay
which
startsonthe
desired
slopeofthe
input-signal
.
Main
Trigger
Level
In
addition to
selecting
the
trigger
slope,
the
LEVEL/
SLOPE
control
determines
the
voltage
levelonthe
trigger
signalatwhich
the
displayistriggered.The
horizontal
lines
marked
on
the
waveforms
to the
left
and
rightofthe
LEVEL/SLOPE
control represent
the
zero-volt
levelofthe
triggering
signal.When
the
LEVEL/SLOPE
controlisset
to
the
lineoneither
the
positive-goingornegative-going
wave-
form,
the
sweepistriggered
near
the
zero-volt
levelofthe
trigger
signal
.
As
the
LEVEL/SLOPE
controlisrotated
away
from
this
line,
the
displayed
waveform
startsata
point
correspondingtothe
position of
the
indicator
line
on
the
associated
slope
waveform.For
example,ifthe
LEVEL/
SLOPE
controlisturned
clockwise
from
the
line
on
the
positive-going
slope,
the displayed
waveform
startsata
more
positive
level
.
2-10
Before
setting
the
triggering
level,
the
desired
SLOPE,
MODE,
COUPLING,
and
SOURCE
should
be
selected
for
MAIN
TRIGGERING
.
Then
adjust
the
LEVEL/SLOPE
control
so
the displayed
waveform
starts
from
the
desired
point
.
The
triggering
slope
canbechangedatany
time
by
rotating
the
LEVEL/SLOPE
control
to the
corresponding
pointonthe
other
slope
waveform
.
Selecting
Sweep
Rates
The
TIME/DIV
OR
DL'Y
TIME
switch
selects
calibrated
sweep
rates
for
the
Main
Sweep
Generator,
and
the
DE-
LAYED
SWEEP
Time/Division
switch
selects
calibrated
sweep
rates
for
the
Delayed
Sweep
Generator
.
The
sweep
rateofthe Main
Sweep
Generatorisbracketedbythe
black
lines
on the
clear
plastic
flange of
the
TIME/DIV
OR
DL'Y
TIME
switch
(see
Fig
.
2-6).Sweep
rateofthe Delayed
Sweep
Generatorisindicatedbythe
white
line
on
the
DE-
LAYED
SWEEP
knob.When
the
white
line
on
the
outer
knobissettothe
same
positionasthe
linesonthe
inner
knob,
the
two
knobs
lock
together
and the
sweep
rate
of
both
generatorsischangedatthe
same
time.However,
when
the
DELAYED
SWEEP
Time/Division
knobispulled
outward,
the
clear
plastic
flangeisdisengaged
and
only
the
Delayed
Sweep
Generator
sweep
rate
is
changed.This
al-
lows
changing
the
delayed
sweep
rate
without
changing
the
delay
time
determinedbythe
Main
Sweep
Generator
.
A
VARIABLE
control
is
provided
concentric
with
the
TIME/DIV
OR
DELAY
TIME
and
DELAYED
SWEEP
Time/Division
switches.This
control
can be
used
either
with
the
main
or
delayed
sweep
generator.The
internal
Variable Selector switch
(see
Fig.2-1
for
location)
deter-
mines
which
sweep
generator the
VARIABLE
control oper-
ates
with.The
VARIABLE
control
also
incorporatesatwo
position
switch
to
determine
if
the
applicable
sweep
rate
is
calibrated
or uncalibrated.When
the
VARIABLE
knob
is
pressed
in,
it is
inoperative.However,
when
pressed
and
released,
the
VARIABLE
controlisactivated
for
uncalib-
rated
sweep
rates.The
sweep
rate
canbereturnedtothe
calibrated
position
by
pressing
the
VARIABLE
control
.
This
featureisuseful
whenaspecific
uncalibrated
sweep
rate
has
been
obtained
and
it is
desiredtoswitch
between
calibrated
and
uncalibrated
sweep
rates.Switching
from
un-
calibratedtocalibrated
and
vice-versa
does
not
affect
the
settingofthe
VARIABLE
control.The
VARIABLE
control
allows
the
sweep
rateineach
Time/Division
switch
position
to
be
reducedtoat least
the
next
adjacent
switch
position
.
Time
Measurement
When
making
time
measurements
from
the
graticule,
the
center
eight
graticule divisions
provide
the
most
linear
time
measurements
(see
Fig.2-7).Position
the
startofthe timing
area to
the
second
vertical
line
and
adjust
the
TIME/DIV
OR
DL'Y
TIME
switchsothe
end
of
the
timing
area
falls
between
the
second
and
tenth
vertical lines
.
CRT
displays
obtained
with
LEVEL/SLOPE
control
in
positive-slope
region
CRT
displays
obtained with
LEVEL/SLOPE
control
in
negative-slope
region
TRIG'D
EVELISLOPE
ma
mmmomm
umummmummm
mvxmuuxxmu mmummmmrrrmm
ammummummm mmmmmmmmmm
NIMENEEMENIN
mmmmm
youffffmuffm mummummumm offiammmmamm
"K1"
1/\VAR
"M
"
1
MrEWEENEEN
mommommomm
Negative
(-)slope
Fig.2-5.Effect of
LEVEL/SLOPE
controlonCRT
display
(AUTOorNORM
MODE)
.
Operating
Instructions-7B53N
2-
1 1
Operating
Instructions-71353N
Fig.2-6.Main
and
delayed
time/division
switch
.
Sweep
Magnifier
2-1
2
Brackets
enclose
Sweep
ndicates
delayed
rat'!
Mainordelayed
sweep
VARIABLE
Second-vertical
Tenth-vertical
line line
Time Measurement
.
Fig
.
2-7.Area
of
graticule
for
accurate
measurements
.
The
sweep
magnifier
canbeused to
expand
each
sweep
rate
ten times.The
center
division
of the
unmagnified
dis-
playisthe
portion
visibleonthe
CRT
in
the
magnified
form
(see
Fig.2-8).Equivalent
length
of the
magnified
sweepismore
than
100
divisions;any
10
division
portion
canbeviewed
by
adjusting
the
POSITION
controltobring
the
desired
portion
onto
the
viewing
area
.
To
use
sweep
magnification,
first
move
the
portion
of
the
display
whichisto be
expanded
to the center of
the
graticule.Then
press
and
release
the
X10
MAG
pushbutton
(concentric
with
POSITION
control);the
X10
MAG
lamp
indicates
that the
sweep
rateismagnified.Whenever
the
X10
MAG
indicatorisilluminated,
the
equivalent
magnified
sweep
rate
can
be
determined
by
dividing
the
TIME/DIV
Fig
.
2-8.OperationofSweep
Magnifier
.
OR
DL'Y
TIME
switch
settingby10.For
example,ifthe
TIME/DIV
OR
DL'Y
TIME
switchissetto.5
us,
the
equi-
valent
magnified
sweep
rateis5
nanoseconds/division.The
equivalent
magnified
sweep
rate
must
be
used
for
all
time
measurements
when
the
X10
MAG
indicatorisilluminated
.
The
equivalent
magnified
sweep
rateiscalibrated
when
the
VARIABLE
controlisin
the
calibrated
position
.
Main
Sweep
Operation
For
main
sweep
displays,
press
the
MAIN
SWP
push-
button
of
the
DISPLAY
MODE
switch.In this
mode
the
TIME/DIV
OR
DL'Y
TIME
switch
selects
the
sweep
rate
for
the
main
sweep
circuit.Calibrated
sweep
rates
from
5
seconds
to
50
nanoseconds/division
are
provided
by
the
TIME/DIV
OR
DL'Y
TIME
switch
(5
nanoseconds
with
X10
magnification).By
using the
VARIABLE
control
(internal
Variable Selector switch
must
beinthe
Main
Time/Div
Variable
position)
uncalibrated
sweep
rates
to
12.5seconds/division
are
available
.
Delayed
Sweep
Operation
The
delayed
sweepisoperableinthe
INTEN,
DLY'D
SWP,
and
MIXED
positions
of the
DISPLAY
MODE
switch
.
ONE
so
-
MAG
switch
off
to
X10
Magnified
waveform
Calibrated
delayed
sweep
rates,
as
determined
by
the
DLY'D
SWEEP
Time/Division
switch,
are
available
0.5seconds
to
50
nanoseconds/division
(5
nanoseconds/
division
X10
magnification).Uncalibrated
delayed
sweep
rates
to1.25
seconds/division
canbe
obtained
by
using
the
VARIABLE
control
(internal
Variable Selector
switch
beinthe
Delayed
Time/Div
Variable
position)
.
The
INTEN
position
of
the
DISPLAY
switch
provides
an
intensified
portion
on
the
main sweep
during
the
the
delayed
sweep
isinoperation.The
amount
of
delay
between
the
start
of
the
main sweep
and the
intensified
portionisdetermined
by
the
TIME/DIV
OR
switch
and
the
DELAY
.
the
DLY'D
SWP
pushbutton
is
pressed,
the
inten-
sified
portion,
as
viewed
the
DISPLAY
switch
isinthe
INTEN
position,isdisplayed
on
the
CRT
at
the
sweep
indicated
by
the
DLY'D
SWEEP
Time/Division
switch
Fig
.
.
ENNEENEEN
Fig.2-9.(A)
INTEN
display
(DL'Y TIME/DIV,
.5ms;
DLY'D
SWEEP
Time/Division,
50
)is),
(B)
Delayed
Sweep
display
.
Operating
Instructions-71353N
I

SweepI
FE---
(Delay)-1.1

Delayed
Sweep
U
Fig.2-10
.
A
typical
mixed
sweep
display
(TIME/DIV
OR
setto1
ms,
DELAYED
SWEEP
Time/Division
set
to
.1
ms,
and
DELAY
setto3
.
Mixed
Sweep
Operation
The
display
in
the
MIXED
position of
the
DISPLAY
switch
isafunction
of both
the
and
delayed
sweeps.In this
mode,
the
main sweep
is
displayed
at
the
determined
by
the
TIME/DIV
OR
switch
followed
by
the
delayed
sweep
atarate
determined by
the
DLY'D
SWEEP
Time/Division
switch.The
amount
of
allocatedtoeach
sweepisdetermined
by
the
setting
of
the
DELAY
dial.A
typical
mixed sweep
playisshown
in
Fig.2-10
.
Delayed
Sweep
Triggering
A
LEVEL
control
and
SLOPE,
COUPLING,
and
SOURCE
switches
are
provided
for
delayed
sweep
trig-
gering.When
the
LEVEL
controlispressed
to
the
IN
-
AFTER
DLY
position
the
delayed
sweep
starts
immediately
after
the
delay
time.The
delayed
sweep
LEVEL
control
and
the
SLOPE,
COUPLING,
and
SOURCE
switches
are
inoperative.This
permits
the
selection
of
continuously
variable
delay times
(by
varying the
DELAY
MULTdial)
.
the
Delayed
Triggering
LEVEL
controlispressed
in
and
releasedtothe
OUT
-
DLY'D
SWP
TRIGGERABLE
position,
the
delayed
sweep
not
start at
the
pletion
of
the
delay
time.Instead,itwaits
until
a
trigger
pulseisreceived
by
the
delayed
sweep
triggering
circuit
.
The
delay
in
is
dependent
not
on
the
settings
of
the
delay-time
controls,
but
on
the
delayed
sweep
triggering
controls
and
the
occurrence
of
the
delayed
sweep
triggering
signal
as well.The
primary purpose
of
istoeliminate
jitter
the
displayed
delayed-
sweep
waveform.Since
the
delayed
sweep
is
triggered
by
2-13
rotensinea
-
portion
MEN--
NEEM
ME
mmmm
no
oil
M
-
__
-
WON
EMENNEEN EMENEENN
Operating
Instructions-71353N
the
input
waveform,
jitteriseliminated
from
the
delayed
sweep
display
even
thoughitmaybe
inherent
in
the
input
waveform
.
In
the
DLY'D
SWP
TRIGGERABLE
mode
(LEVEL
control
in
OUT
position)
the
Delayed
Sweep
Triggering
LEVEL
control
and
SLOPE,
COUPLING,
and
SOURCE
switches
are
activated
.
The
Delayed
Sweep
Triggering
LEVEL
control
determines
the voltage
levelonthe
trigger
signal at
which
the
delayed
sweep
is
triggered.The
SLOPE
pushbutton
determines
whether
the
delayed
trigger
circuit
responds
on
the
positive-going
or
negative-going
portion
of
the
trigger
signal.The
selected
SLOPE
is
indicated
by
the
illuminated
portion
of the
pushbutton
.
COUPLING
and
SOURCE
lights
are
also
provided
for
delayed
triggering
.
The
positions,asmarked
on
the
pushbuttons, have
the
same
functions
for
delayed
triggering
as the
identically
marked
COUPLING
and
SOURCE
switches
for
MAIN
TRIGGERING
(see
Main
Triggering
Coupling
and
Source
discussions
given
in this
section).The
selected
Delayed
Triggering
COUPLING
and
SOURCE
positions
are
indi-
catedbyilluminated
portionsofthe
pushbuttons
.
Delayed
Gate
Output
When
the
Delayed
Triggering
SOURCE
switch
is
set
to
INT,
the
DLY'D
TRIG
IN
connector
serves
as a
Delayed
Gate
Output.The
Delayed
Gate
Outputisa
positive-going
rectangular
pulse
with
approximately
3.5volts
amplitude
.
The
pulse
width
is
coincident
with
the
time
that
the
delayed
sweep
runs
.
Therefore,
conditions
for
a
triggered
delayed
sweep,
as
described
under Delayed
Sweep
Trig-
gering
in this
section,
must
be
met
foraDelayed
Gate
Output
.
General
APPLICATIONS
The
following
information
describes
the
procedure
and
techniques
for
making
basic
measurements
witha7B53N
installedina
7000-Series
Oscilloscopq.These
applications
are
not
describedindetail,
since
each
application
must
be
adapted
to the
requirements
of the
individual
measurement
.
This
instrument
can
alsobeused
for
many
applications
not
described
in
this
manual.Contact
your
local
Tektronix
Field
Office or
representative
for
assistanceinmaking
spe-
cific
measurements
.
Also,
the
following
books
describe
oscilloscope
measurement
techniques
which
canbeadapted
for
use
with
this
instrument
:
Harley
Carter,
"An
Introduction
to the
Cathode
Ray
Oscilloscope",
Philips
Technical
Library,
Cleaver-Hume
Press
Ltd.,London,
1960
.
J.Czech,
"Oscilloscope
Measuring
Technique",
Philips
Technical
Library,
Springer-Verlag,
New
York,
1965
.
2-14
Robert
G.Middleton
andL.
Donald
Payne, "Using the
OscilloscopeinIndustrial
Electronics",
HowardW.
Sams
&
Co.Inc.,The
Bobbs-Merill
Company
Inc
.,
Indianapolis,
1961
.
John
F.Rider
and
Seymour
D.Uslan,
"Encylopedia
of
Cathode-Ray
Oscilloscopes
and
Their
Uses",
John
F.Rider
Inc.,New
York,
1959
.
JohnF.
Rider,
"Obtaining
and
Interpreting
Test
Scope
Traces",
John
F.Rider
Publisher
Inc
.,
New
York,
1959
.
Rufus
P.Turner,
"Practical
Oscilloscope
Handbook",
Volumes1and2,John
F.Rider
Publisher Inc.,New
York,
1964
.
Comparison
Measurement
Techniques
Sweep
RatesTo
establishanarbitrary
horizontal
sweep
rate
based
upon
a
specific
reference
frequency,
proceed
as
follows
:
1.Connect
the
reference
signal
to the
input of the
ver-
tical
unit.Set the
Volts/Division switch
of
the
vertical
unit
for
four
or
five
divisionsofvertical
deflection
.
2
.
Set the
TIME/DIV
OR
DL'Y
TIME
switch
and
the
VARIABLE
control so
one
cycleofthe
signal
covers an
exact
number
of
horizontal divisions.Do
not
change
the
VARIABLE
control
after
obtaining
the
desired
deflection
.
This
display
canbeusedasa
reference
for
frequency
com-
parison
measurements
.
3.To
establishanarbitrary
sweep
rate
so the
period
(time for
one
complete
cycle)ofan
unkown
signal
can
be
measured
accurately
at
any
setting
of
the
TIME/DIV
OR
DL'Y
TIME
switch,
the
period
of the reference
signal
must
be
known.If it is
not
known,
it
canbemeasured
before
the
VARIABLE
switch
is
setinstep 2
.
4.Divide
the
period
of
the reference
signal
(seconds)
by
the
productofthe
horizontal deflection
established
in
step
2
(divisions)
and
the
setting
of the
TIME/DIV
OR
DL'Y
TIME
switch.Thisisthe
horizontal
conversion
factor
:
Horizontal Conversion Factor
reference
signal
period
(seconds)
horizontal

TIME/DIV
OR
deflectionXDL'Y
TIME
(divisions)

switch
setting
5.To
measure
the period ofanunknown
signal,
discon-
nect
the
reference
signal
and
connect
the
unknown
signal
to
O
the
vertical
unit.Set
the
TIME/DIV
OR
DL'Y
TIME
switch
toasetting
that
provides
sufficient
horizontal
deflection
to
make
an
accurate
measurement.Do
not
readjust
the
VAR
(-
ABLE
control
.
6
.
Measure
the
horizontal
deflection
in
divisions
and
cal-
culate
the
period of
the
unknown
signal
using
the following
formula
:
TIME/DIV
OR

horizontal

horizontal
Period
-
DLY'
TIME
X
conversion
X
deflection
(Seconds)
switching
factor
(divisions)
NOTE
If
the
horizontal
magnifier
is
used,besure to
use
the
magnified
sweep
rate
in
placeofthe
TIME/DIV
DL'Y
TIME
switch
setting
.
Example
.
Assume
a
reference
signal
frequency
of
455
hertz
(period
2
.19
milliseconds),
a
TIME/DIV
OR
DL'Y
TIME
switch
setting
of.2ms,
and
the
VARIABLE
control
adjusted
to
provideahorizontal
deflection
of
eight
divi-
sions
.
Substituting
these
valuesinthe
horizontal
conversion
factor
formula
(step
4)
:
Horizontal
Conversion
=

=
1
.37
.2msX8
Factor
2
.19
milliseconds
Then,
with a
TIME/DIV
OR
DL'Y
TIME
switch
setting
of
50
ps,
the periodofan
unknown
signal
which
completes
one
cycleinseven
horizontal
divisions
can
be
determined
by
using
the
period
formula
(step6):
Period
-
(Seconds)

50psX
1
.37X7 =
480
ps
This
answer
can be
converted
to
frequencybytaking
the
reciprocal
of
the period
in
seconds
(see
application
on
Determining
Frequency
Measurements)
.
Time
Duration
Measurements
To
measure
time
between
two
pointsona
waveform,
use
the
following
procedure
:
1
.
Connect
the
signal
tobedisplayed
to
the
inputofthe
vertical
unit
.
2.Set the
Vertical
and
Horizontal
Mode
switches
on
the
indicator
oscilloscope
to
display
the
plug-in
units
used
.
Operating
Instructions-71353N
3.Set
the
Volts/Division
switch
of
the
vertical
unit
to
display
about
four
divisions
of
waveform
.
4.Set
the
MAIN
TRIGGERING
controls to
obtain
a
stable
display
.
5
.
Set the
TIME/DIV
OR
DL'Y
TIME
switch to
the
fastest
sweep
rate
that
displays
less
than
eight
divisions
be-
tween
the
time
measurement
points
(see
topic
entitled
Time
Measurements
and
Fig
.
2-7)
.
6.Adjust
the
vertical
unit
position
controltomove
the
points
between
which
the
time
measurement
is
made
to
the
center horizontal
line
.
7.Adjust
the
horizontal
POSITION
control
to position
the
time-measurement
points
within the
center
eight
divi-
sionsofthe
graticule
.
8.Measure
the
horizontal
distance
between
the
time
measurement
points.Be
sure
the
VARIABLE
controlisset
to
CAL
.
9
.
Multiply
the
distance
measured
in
step8by
the
set-
tingofthe
TIME/DIV
OR
DL'Y
TIME
switch
.
If
sweep
magnification
is
used,
divide
this
answer by10.
Example.Assume
that
the
distance
between
the
time
measurement
points
is
five
divisions
(see
Fig.2-11),
and
the
TIME/DIV
OR
DL'Y
TIME
switch
is
setto.1ms.
Fig.2-11
.
Measuring
the
time
duration
between
pointsona
wave-
form
.
2-15
Operating
Instructions-71353N
Using
the
formula
:
Time
Duration
Time
Duration
=
Substituting
the
given
values
:
The
time
durationis0
.5
millisecond
.
Determining
Frequency
Use
the
following
procedure
:
horizontal

TIME/DIV
OR
distanceX

DL'Y
TIME
(divisions)
setting
magnification
5X0.1ms
1
The
time
measurement
technique
can
also
be used to
determine
the
frequencyofa
signal.The
frequencyofa
periodically
recurrent
signalisthe
reciprocalofthe
time
duration
(period)ofone
complete
cycle
.
1.Measure
the
time duration
of
one
complete
cycle
of
the
waveformasdescribedinthe
previous
application
.
2.Take
the
reciprocalofthe
time
durationtodetermine
the
frequency
.
Example
.
The
frequencyofthe
signal
showninFig.2-11
which
hasatime
periodof0.5millisecondis:
1

1
Frequency
=
time
period0.5
millisecond=
2
kilohertz
Risetime
Measurements
Risetime
measurements
employ
basically
the
same
tech-
niquesastime-duration
measurements.The
main
difference
is
the
points
between
which
the
measurementismade
.
The
following
procedure
gives
the
basic
method
of
measuring
risetime
between
the
10%
and
90%
pointsofthe
waveform
.
Falltime
can
be measured
in
the
same
manner
on
the
trail-
ing
edgeofthe
waveform
.
1 .
Connect
the
signal
to be
displayedtothe
inputofthe
vertical
unit
.
2-16
2.Set
the
Vertical
and
Horizontal
Mode
switchesonthe
indicator oscilloscopetodisplay
the
plug-in'unit
used
.
3.Set
the
Volts/Division
switch
and
the
Variable
Volts/
Division
controlofthe
vertical
unittoproduceasignal
an
exact
numberofdivisionsinamplitude
.
4.Center
the
display
about
the center
horizontal
line
with
the
vertical
unit Position
control
.
5.Set
the
MAIN
TRIGGERING
controls to
obtain
a
stable
display
.
6.Set the
TIME/DIV
OR
DL'Y
TIME
switch
to
the
fastest
sweep
rate
that
displays
less
than
eight
divisions
be-
tween
the
10%
and
90%
pointsonthe
waveform
.
7
.
Determine
the
10%
and
90%
pointsonthe
rising
portionofthe
waveform
.
The
figures
given
in
Table
2-1 are
for
the
points
10%upfrom
the
startofthe
rising
portion
and
10%
down
from the
topofthe
rising
portion
(90%
point)
.
TABLE
2-1
Risetime
Measurements
8.Adjust
the
horizontal
POSITION
controltomove
the
10%
pointofthe
waveform
to
the
second
vertical
line
of
the
graticule
.
For example,
withafive-division
display
as
shown
in
Fig.2-12,
the
10%
pointis0.5divisionupfrom
the
startofthe
rising
portion
.
9.Measure
the
horizontal
distance
between
the
10%
and
90%
points.Be
sure
the
VARIABLE
control
is
settoCAL
.
10.Multiply
the
distance
measured
in
step9by
the
setting
of
the
TIME/DIV
OR
DL'Y
TIME
switch.If
sweep
magnificationisused,
divide
this
answer by 10
.
O
Divisions
vertically
Vertical
between
display
10%
and
90%
10%
and
90%
(divisions)
points
points
4
0
.4
and3.6
divisions
3
.2
5
0.5and4.5
divisions
4
.0
6
0.6and5.4
divisions
4
.8
7
0
.7
and5.4
divisions
5
.6
8
0.8and7.2
divisions
I
6.4
Fig.2-12.Measuring
risetime
.
Example
.
Assume
that
the horizontal
distance
between
the
10%
and
90%
pointsisfour
divisions
(see
Fig.2-12)
and
the
TIME/DIV
OR
DL'Y
TIME
switchissetto1 Its
with
the
MAG
switch
settoX10.Applying
the
time
duration
formulatorisetime
:
horizontal

TIME/DIV
OR
Time
Duration
=

distance

DL'Y
TIME
(Risetime)
(divisions)
setting
magnification
Substitute
the
given
values
:
Risetime
= 4
X1microsecond
The
risetimeis0.4microsecond
.
Delayed
Sweep
Measurements
The
delayed
sweep
mode
canbeusedtomake
accurate
time
measurements.The
following
measurement
determines
the
time
difference
between
two
pulses
displayed
on
the
same
trace.This
application
may
alsobeusedtomeasure
time
difference
from
two
different
sources
(dual-trace)
or
to
measure
time duration
of
a
single
pulse.See
Section
1
for
measurement
accuracy
.
1.Connect
the
signaltobe
displayed
to the
input
of the
vertical
unit
.
2.Set
the
vertical
and
horizontal
Mode
switchesonthe
indicator oscilloscope
to
display the
plug-in
units
used
.
3.Set the
Volts/Division
switch of the
vertical
unit to
produce
a display
about4divisionsinamplitude
.
4.Adjust
the
MAIN
TRIGGERING
controls
forastable
display
.
Operating
Instructions-7B53N
5.If
possible,
set
the
TIME/DIV
OR
DL'Y
TIME
switch
toasweep
rate
which
displays
about
eight
divisions
be-
tween
pulses
.
6.Press
the
INTEN
pushbutton
of the
DISPLAY
MODE
switch
and
press the
Delayed
Triggering
LEVEL
control to
theIN-
RUNS
AFTER
DL'Y
TIME
position
.
7.Set the
DLY'D
SWEEP
Time/Division
switch
to
a
setting
1/100ofthe
TIME/DIV
OR
DL'Y
TIME
sweep
rate
.
This
produces
an
intensified
portion
approximately0.1
di-
vision
in
length
.
NOTE
Measurement
accuracy
willbeaffectedifthe
LEVEL
control
setting
for
MAIN
TRIGGERING
or the
hori-
zontal
POSITION
control
settingischanged
.
8.Rotate the
DELAY
TIME
MULT
dial
to
move
the
intensified
portion of the
trace
to the
first
pulse
.
9.Press
the
DLY'D
SWP
pushbutton
of the
DISPLAY
MODE
switch
.
10.Adjust
the
DELAY
TIME
MULT
dial
to
move
the
pulse
(or
the
rising
portion)tothe center
vertical
graticule
line.Note
the
exact
setting
of the
dials
.
11.Turn
the
DELAY
TIME
MULT
dial
clockwise
until
the
second
pulseispositioned
to the
same
pointasthe
first
pulse.(If
several
pulses
are
displayed, returntothe
INTEN
positionofthe
DISPLAY
MODE
switchtolocate
the
cor-
rect
pulse).Again
note
the
exact
dial
setting
.
12.Subtract
the
first
dial
setting
from
the second
and
multiply
by
the
delay
time
shown
by
the
TIME/DIV
OR
DL'Y
TIME
switch.This
figureisthe
time
interval
between
pulses
.
Example.Assume
the
first
dial
setting
is 1
.31
and
the
second
dial
settingis8
.81
with
the
TIME/DIV
OR
DL'Y
TIME
switch
setto0.2microsecond
(see
Fig.2-13)
.
Time
Difference

__
(Delayed
Sweep)
(second
dial
setting-first
dial setting)
X
Substituting the
given
values
:
Time
Difference=(8
.81-1
.31)X0.2!ts
The
time
differenceis1.5ps
delay
time
(TIME/DIV
OR
DL'Y
TIME
switch
setting)
2-17
RPM
MEMONE
t I
jam-
Horizontal
distance
}i
11
MMMOMMMMMM
(A)
Intensified
sweep
display
.
(B)
Delayed
sweep
display
.
Fig.2-13
.
Measuring
difference using
delayed
sweep
.
Delayed
Sweep
Magnification
The
delayed
sweep
feature
of
the
7B53N
can be
used
to
provide
higher
apparent
magnification
than
is
provided
by
the
MAG
switch.The
sweep
rate
of
the
delayed
sweep
is
not
actually
increased
;
the
apparent
magnificationisthe
resultofdelaying
the
Delayed
Sweep
an
amountoftime
selected
by
the
TIME/DIV
OR
DL'Y
TIME
switch
and
the
DELAY
TIME
MULT
dial
before
the
display
is
presented
at
the
sweep
rate
selected
by
the
DLY'D
SWEEP
Time/
Division switch
.
The
following
method
uses
the
IN-RUNS
AFTER
DL'Y
TIME
Delayed
Trigger
Mode
to
allow
the
delayed
portion
of
the
display
to be
positioned
with
the
DELAY
TIME
MULT
dial
.Ifthereistoo
much
jitter
in
the
delayed
sweep
display,
use
the
Triggered
Delayed
Sweep
Magnification
procedure
which
follows
this
procedure
.
1
.
Connect
the
signal
tobedisplayed
to
the
input
con-
nector of
the
vertical
unit.Set
the
Vertical
and
Horizontal
Mode
switches
on
the
indicator
oscilloscope
to
display
the
plug-in
units
used
.
2-
1
8
4.Set
the
TIME/DIV
OR
DL'Y
TIME
switch to
a
sweep
rate
which
displays
the
complete
waveform
(see
Fig
.
2-14A)
.
5.Press
the
INTEN
pushbuttonofthe
DISPLAY
MODE
switch
and
press
the
Delayed
Trigger
LEVEL
control
to
IN-RUNS
AFTER
DL'Y
TIME
.
6.Position
the
startofthe
intensified
portion
with
the
DELAY
TIME
MULT
dialtothe
partofthe
displaytobe
magnified
.
71-
1
(A)
sweep
display
.
(B)
Delayed
sweep
display
.
Pulsetobe
magnified
MOMMMON
ME
amm
I
Fig.2-14
.
Using
delayed
sweep
for
magnification
.
Operating
Instructions-712153N
2.Set
the
Volts/Division
switch
of
the
vertical
unit
to
DELAY-TIME
DELAY-TIME
:
produce
a
display
about
4
divisionsinamplitude
.
:
1
.31
8
.81
1
MINE
WOMEN
HNIN
3.Adjust
the
MAIN
TRIGGERING
controls
forastable
display
.
EKE
7.Set the
DLY'D
SWEEP
Time/Division
switchtoa
setting
which
intensifies
the
full
portionofthe
display
to
be
magnified
.
The
startofthe
intensified
trace
will
remain
as
positionedinStep6.
8.Press
the
DISPLAY
MODE
switchtoDLY'D
sweep
.
9
.
Time
measurements
canbemade
from
the
display
in
the
conventional
manner
.
Sweep
rateisdeterminedbyset-
tingofthe
DLY'D
SWEEP
Time/Division
switch
.
10.The
apparent
sweep
magnification
can be
calculated
by
dividing
the
TIME/DIV
OR
DL'Y
TIME
switch
setting
by
the
DLY'D
SWEEP
Time/Division
switch
setting
.
Example.The
apparent
magnificationofthe
display
shown
in
Fig.2-14
with a
TIME/DIV
OR
DL'Y
TIME
settingof.1msandaDLY'D
SWEEP
Time/Division
switch
settingof1
microsecond
is
:
Apparent
Magnification
=
TIME/DIV
OR
DL'Y
TIME
setting
DLY'D
SWEEP
Time
Division
setting
Substituting
the
given
values
:
Apparent

=
1
X
10-4
Magnification

1
X~
10Z-
The
apparent
magnificationis100
times
.
Triggered
Delayed
Sweep
Magnification
The
delayed
sweep
magnification
method
just
described
may
produce too
much
jitterathigh
apparent
magnifi-
cation ranges
.
The
OUT-DLY'D
SWP
TRIGGERABLE
mode
providesamore
stable
display,
since
the
delayed
sweep
displayistriggeredatthe
same
point
each
time
.
1.Setupthe
displayasgiveninsteps1through 7inthe
Delayed
Sweep
Magnification
procedure
.
2.Pressinand
release
the Delayed
Triggering
LEVEL
controltothe
OUT-DLY'D
SWP
TRIGGERABLE
posi-
tion.Select
the
desired
Delayed
Triggering
SLOPE,
COUPL-
ING,
and
SOURCE
.
3.Adjust
the
Delayed
Triggering
LEVEL
controltopro-
duce an
intensified
portiononthe
display
.
4.Inabilitytoproduceanintensified
zoneonthe
dis-
play
indicates
that
the Delayed
Triggering
controls
are
Operating
Instructions-71353N
incorrectly
set,orthat
the
signal
does not
meet
triggering
requirements
.
If
the
condition
cannotberemedied
with
the
Delayed
Triggering controls or
by
increasing
the
display
amplitude
(lower Volts/Division
setting),
externally
trigger
the
delayed
sweep
.
5.When
the
correct
portionofthe
displayisintensified,
set
the
DISPLAY
MODE
switch to
DLY'D
SWP.Slight
re-
adjustmentofthe
Delayed
Triggering
LEVEL
control
may
be
necessary
to
produce
a
stable
display
.
6.Measurement
and
magnification
areasdescribed
aboveinDelayed
Sweep
Magnification
discussion
.
Displaying
Complex
Signals
Using
Delayed
Sweep
Complex
signals
often
consist
of
a
numberofindividual
events of
differing
amplitudes.Since
the
trigger
circuits
are
sensitivetochangesinsignal
amplitude,astable
display
can
normallybeobtained only
when
the
sweepistriggered
by
the
event(s)
having the
greatest
amplitude.However,
this
may
not
produce
the
desired displayofa
lower-amplitude
portion
which
follows
the
triggering
event.The
delayed
sweep
feature
provides
a
meansofdelaying
the
startofthe
delayed
sweep
byaselected
amount
following the
event
which
triggers
the
Main
Sweep
Generator.Then, the
part
of
the
waveform
which
contains
the
informationofinterest
can be
displayedatthe
delayed
sweep
rate
.
Use
the following
procedure
:
1 .
Set
up
the
displayasgiveninSteps1through8of
Delayed
Sweep
Magnification
.
2.Time
measurements
can be
made
from
the
display
in
the
conventional
manner.Sweep
rateisdetermined by
the
settingofthe
DLY'D
SWEEP
Time/Division
switch
.
Example
:Fig.2-15
showsacomplex
waveformasdis-
playedonthe
CRT.The
circled
portionofthe
waveform
cannotbeviewedinany
greater
detail
because
the
sweep
is
triggeredbythe
larger
amplitude
pulses at
the
startofthe
display
andafaster
sweep
rate
moves
this
area
of
the
wave-
form
off
the
viewing
area.The
second
waveform
shows
the
area
of
interest
magnified
10 times
using
Delayed
Sweep
.
The
DELAY
TIME
MULT
dial
hasbeen
adjusted so
the de-
layed
sweep
starts
just
before
the
area of
interst
.
Pulse
Jitter
Measurements
In
some
applications
it is
necessarytomeasure
the
amount
of
jitteronthe
leading
edgeofa
pulseorjitter
between
pulses
.
2-19
Operating
Instructions-71353N
(A)
This portionofdisplay
cannot be viewed
adequately
be-
cause
the
main
sweep
is
triggeredonlarger
amplitude
signals
at
the
startofdisplay
.
NEENEENEEN NEEMEMENNN NESPEENNEN
"
EMEMEMEN
"
MENNEENEEN
ENEENEEMEN
(B)
Areaofinterest
displayedbydelaying
the
main
sweep
(DLY'D
SWP
mode)
.
Fig.2-15.Displaying
a
complex
signal
using
delayed
sweep
.
1.Connect
the
signaltobe
displayedtothe
input
con-
nectorofthe
vertical
unit.Set
the
Vertical
and
Horizontal
Mode
switches
on
the
indicator oscilloscope
to
display
the
plug-in
units
used
.
2.Set the
Volts/Division
switchofthe
vertical
unit
to
produceadisplay
about
4
divisionsinamplitude
.
3.Adjust
the
MAIN
TRIGGERING
controls
forastable
display
.
4.Set the
TIME/DIV
OR
DL'Y
TIME
switchtoa
sweep
rate
which
displays
the complete
waveform
(see
Fig
.
2-14A)
.
5.Press
the
INTEN
pushbuttonofthe
DISPLAY
MODE
switch
.
Press
and
release
the Delayed
Triggering
LEVEL
controltoOUT-DLY'D
SWP
TRIGGERABLE,
and
adjust
the
LEVEL
control
forastable
intensified
display
.
2-
20
6
.
Position
the
start
of
the
intensified
portion
with
the
DELAY
TIME
MULT
dialtothe
part of
the
display
to be
magnified
.
7.Set the
DLY'D
SWEEP
Time/Division
switch
to
a
setting
which
intensifies
the
full
portionofthe
display
to
be
magnified.The
startofthe
intensified
trace
will
remain
as
positionedinStep 6
.
8
.
Press
the
DISPLAY
MODE
switch
to
DLY'D
SWP
.
9.Slight
readjustment
of
the
Delayed
Triggering
LEVEL
control
maybenecessarytoproduceasstable
dis-
playaspossible
.
10
.
Pulse
jitterisshownbyhorizontal
movement
on
the
pulse
(take
into
account
inherent
jitter
of
Delayed Sweep)
.
Measure
the
amount
of
horizontal
movement.Be
sure that
both
vertical
and
horizontal
VARIABLE
controls
are
set
to
CAL
.
11.Multiply
the
distance
measuredinStep 10
by
the
DLY'D
SWEEP
Time/Div
switch
setting
to
obtain
pulse
fit-
terintime
.
Example.Assume
that
the
horizontal
movementis0
.5
division
(see
Fig.2-16)
and
the
DLY'D
SWEEP
Time/
Division
switchis.5
microsecond
.
Using
the
formula
:
horizontal

DLY'D
SWEEP
Pulse
Jitter=jitter
X
Time/Division
(division)
setting
Substituting
the
given
values:Pulse
Jitter=0.5X0.5
microsecond
The
pulse
jitteris0
.25
microsecond
.
MMMMMMMMMM MMMMMr=M== MMMMMIJIMMMM MMMMMI
MMMM! MMMMMAMMMM MMMMEMMMM
MENEAMMEN
EMENEREENE
I
Jitter
.mod
1
.0-
Fig.2-16.Measuring
pulse
jitter
.
Introduction
SECTION
3
CIRCUIT
DESCRIPTION
This
section
of
the
manual
contains
a
descriptionofthe
circuitry
usedinthe
7B53N
Dual
Time
Base.The
descrip-
tion
begins
withadiscussion
of
the
major
circuit
functions
usingasimplified
block
diagram
.
SIMPLIFIED
BLOCK
DIAGRAM
The
Simplified
Block
Diagram,
Fig.3-1,
shows
intercon-
nectionofthe
basic
circuit
blocksinthe
7B53N
.
In
some
cases,
suchasthe
Main
Sweep
Trigger,
the
block
includes
a
number
of
separate
circuits.The
individual
circuits
are
dis-
cussed
in
detail
later
in this
section
.
Main
Sweep
Mode
When
the
DISPLAY
MODE
switchissettoselect
MAIN
SWP,
operationisas
follows
:
Main
Sweep
Trigger.This
block
includes
circuitry
for
selecting
the
trigger
source,
type
of
coupling,
triggering
mode,
and
point
on
the
trigger
signal
where
triggering
oc-
curs.Also,
regardlessofthe
trigger
signal
shape
or ampli-
tude
(within
specification),
this
circuitry
providesafast-
rise,
uniform-amplitude
pulsetothe
Main
Sweep
Start
Multi
.
Terminationofthe
pulse
(or
gate)
occursatthe
rise
of
Main
Sweep
Holdoff
.
Main
Sweep
Start
Comparator.This
circuitisactivated
by
the
positive
gate
from
the
Main
Sweep
Trigger.The
output
signal
coupledtothe
Main
Sawtooth
Generatorisa
positive gate
with
the
same
durationasthe
sweep
.
This gate
is
also
coupled
to
the
Sweep
Gate
Out
.
A
negative-going
gate
(coincident
with
the
positive
gate)
is
coupledtothe
Delayed
Sweep
Lockout
Multi
and
the
Delayed
Sweep
Start
Control
.
Main
Sawtooth
Generator
.
The
main
sweep
signalisde-
velopedbythe
Main
Sawtooth
Generator.Whenapositive
gate
from
the
Main
Sweep
Start
Multiisapplied,
a
sawtooth
waveform
is
generated
.
The
sawtooth
durationisdeter-
minedbythe
positive
gate
duration.Rateofchange
of
the
sawtooth
is
set
by Ct and
Rt,
selected
by
the
TIME/DIV
switch
.
Delayed
Sweep
Mode
7B53N
Sweep
Stop
Comparator
.
One
sideofthis
comparator
is
drivenbythe
main
sweep
sawtooth
signal,
and the
other
sideissetbythe
Main
Swp
Stop
adjustment
.
When
the
sawtooth
waveform
passes
through
the
settingofthe
Main
Swp
Stop
adjustment,
the
output
of
the
Sweep
Stop
Com-
parator
switchestoa
positive
level.This
positive step
is
applied
to
the
Main
Sweep
Holdoff and
by
wayofa
small
capacitancetothe
Main
Sweep
Start
Multi.This
resets
the
Main
Sweep
Start Multisothat
it is
readytoreceive
another
trigger
signal
.
Main
Sweep
Holdoff.This
circuit
developsagate
which
is
usedtoprevent
generation
ofatrigger
signal until
the
sweep
circuits
have
stabilized
afterasweep
.
The
positive
step
from
the
Sweep
Stop
Comparator
initiates
the
positive
holdoff
gate.The
durationofthe
holdoff
gate
is
variable,
depending
on
the
settingofthe
TIME/DIV
switch.Holdoff
timing
capacitors
are
separate
trom
sweep
timing
capacitors.Holdoffislonger
for
slower
sweep
rates
.
Output
from
the
Main
Sweep
Holdoff
is
coupledtothe
Main
Sweep
Trigger
and the Delayed
Sweep
Trigger.A
trig-
ger
signal
cannotbegenerated
during the
holdoff
interval
.
The
holdoff
servestoreset
the
trigger circuits
so that
they
are
readytoreceiveaninput
trigger
signal
after
holdoff
.
Horiz
Output.The
Horiz
Output
block
includes
the
Ext
Horiz
Amp,
Position
Amp,
Horiz
Display
Selector,
and
Horiz
Out
Amp
circuits
.
With the
DISPLAY
MODE
switch
settoMAIN
SWP,
this
circuit selects
the
signal
from
the
Main
Sawtooth
Gener-
ator,
amplifies
the
signal,
and
converts
the
single-ended
inputtoa
push-pull
output
signal.A
DC
positioning
level
is
also
applied
to
this
block
.
To
generate
the
delayed
sweep,
the
Main
Sawtooth
Gen-
erator
must
first
be
gatedon(see
Main
Sweep Mode)
.
Delay
Pickoff.This
circuit
suppliesapositive gate
which
starts
when
the
main
sawtooth
signal
passes
through
the
level
selected
by
the
DELAY
TIME
MULT
control.The
gate
ends
with
the
main
sawtooth
signal.The
output
signal
is
coupled
to
the Delayed
Sweep
Trigger
.
INT
Fig.3-
1 .
7B53N
Simplified
Block
Diagram
.
Delayed
Sweep
Trigger
.
When
the
Delayed
Trigger
LEVEL
is
pushed
in,
the
output
triggerisgeneratedassoon
as
the
Delay
Gate
signalisapplied.If
the
LEVEL
control
is
out,
the
output
trigger
is
initiatedbythe
next
input
trigger
after
the
Delay Gateisapplied
.
The
Delayed
Sweep
Trigger
outputisa
positive gate
whichisterminated
by
the
Holdoff
signal
and/or
the
pos-
itive
step
from
the
Delayed
Sweep
Stop
circuit.The
pos-
itive
output
gateiscoupled
to
the
Delayed
Sweep
Start
Multi
.
Delayed
Sweep
Start
Multi.The
signal
from
the
Delayed
Sweep
Trigger
causes the
Delayed
Sweep
Start
Multitoflip
so
that
a
positive
gateiscoupled
to
the
Delayed
Sweep
Start
Control,
andanegative
gateisappliedtothe
Mixed
Sweep
Comparator
.
The
output
gates
are
the
same
in
dura-
tionasthe
positive
gate
from
the Delayed
Sweep
Trigger
.
Delayed
Sweep
Start
Control
.
For
DLY'D
SWP
mode
of
operation
the
Delayed
Sweep
Start
Control
servestocouple
the
positive gate
from
the
Delayed
Sweep
Start
Multitothe
Delayed
Sawtooth
Generator and
the
Sweep
Gate
Out
.
Input
signals
from
the
Main
Sweep
Start
Multi
and
the
Delayed
Sweep
Lockout
Multi
are
not
effective
in this
mode
.
Delayed
Sawtooth
Generator
.
The
delayed
sweep
signal
is
developed
by the
Delayed
Sawtooth
Generator.The
saw-
toothisgenerated
during
the
time
thatapositive
gate
is
applied
from
the
Delayed
Sweep
Start
Control.Rate
of
changeofthe
sawtoothisset
by Ct
and
Rt,
selectedbythe
TIME/DIV
(Dly'd)
switch
.
The
sawtooth
output
signaliscoupled
to
the
Mixed
Sweep
Comparator
and
the
Horiz
Output
circuits
.
Delayed
Sweep
Stop
Circuit
.
A
positive step
occurs
at
the
outputofthe
Delayed
Sweep
Stop
circuit
when
the
delayed
sawtooth
passes
through
the
level
selectedbythe
Dly'd
Swp
Length
adjustment.This
stepiscoupledtothe
Delayed
Sweep
Trigger
and the Delayed
Sweep
Lockout
Multi
.
Mixed
Sweep
Mode
In this
modeofoperation,
the
sweepisfirst
running
at
the
MAIN
SWP
rate
and
then,
after
the
selected
delay
in-
terval,
runsatthe
DLY'D
SWP
rate.The
main
sweep
and
delayed
sweep
are
initiated
as previously
described
.
Oper-
ationofother
circuit
blocks
follows
.
External
Horiz
Input
General
CIRCUIT
OPERATION
Circuit
Description-71353N
Mixed
Sweep
Comparator.This
circuit
determines
which sweep
signaliscoupledtothe
Horiz
Output
stage
.
First,
the
main
sweep
sawtooth
is
coupled through
the
Mixed
Sweep
Comparator
and
the
Delayed
Sawtooth
Gen-
eratortothe
Horiz
Output
stage.These
stages
perform
as
an
operational
amplifier
during the
time
that
the
main
sweep
is
being
displayed
.
When
a
positive
gate
from
the
Delayed
Sweep
Trigger
is
appliedtothe Delayed
Sweep
Start
Multi,anegative
gate
is
generated
and coupled
to
the
Mixed
Sweep
Comparator
.
This
opens the
Mixed
Sweep
Comparator
circuit,
preventing
the
main
sweep
sawtooth
from
being
coupled
to
the
Horiz
Output
circuit
.
Simultaneously,
the
positive
gate
from
the
Delayed
Sweep
Start
Multi
is
coupled through
the
Delayed
Sweep
Start
Control
to
the
Delayed
Sawtooth
Generator.The
de-
layed
sweep
sawtoothisgenerated
and
coupled
to
the
Horiz
Output
stage
.
Delayed
Sweep
Lockout
Multi.The
positive
step
from
the
Delayed
Sweep
Stop
circuitisinverted
by
the
Delayed
Sweep
Lockout
Multi
and coupledtothe
Delayed
Sweep
Start
Control,
thus
turning
off
the
Delayed
Sawtooth
Gen-
erator
.
Sweep
Gate
Out
.
Depending
on the
selection
of
the
DISPLAY
MODE
switch,
this
stage
couples
the
positive
gate
from
either
the
Main
Sweep
Start
Multi
or
the
Delayed
Sweep
Start
ControltoconnectorAl.
The
Sweep
Gate
signal
servestounblank
the
CRT
in
the
Indicator
Oscillo-
scope
during the
sweep
.
When
the
TIME/DIV
switch
is
settoAMPL,
partofthe
Main
Sweep
Trigger
circuitry
becomes
the
Horiz Input
Amp.An
external
signal
connectedtothe
MAIN
TRIG
IN
or
AMPL
input
is
amplified
and
then
coupled
to
the
Horiz
Output
stage
.
The
main and
delayed
sawtooth
generators
are
disabled to
prevent
intensity
modulationofthe
CRT
tracebythe
unblanking
waveforms
.
This
section
provides
a
detailed
descriptionofthe
elec-
trical
operation
and
relationshipofthe
circuits
in
the
7B53N
.
The
theoryofoperation
for
circuits
unique
to this
instrumentisdescribed
in
detailinthis
discussion.Circuits
which
are
commonly
usedinthe
electronics
industry
are
not
describedindetail.If
more
information
is
desired
on
these
commonly
used
circuits,
refertothe
following
text-
books
.
3-3
Circuit
Description-71353N
Tektronix
Circuit
Concepts
Books
(order
from
your
local
Tektronix
Field
Office or
representative)
.
Horizontal
Amplifier
Circuits,
Tektronix
PartNo.
062-1144-00
.
Oscilloscope Trigger
Circuits,
Tektronix
Part
No
.
062-1056-00
.
Sweep
Generator
Circuits,
Tektronix
PartNo.
062-1098-01
.
Phillip
Cutler,
"Semiconductor
Circuit
Analysis,'
McGraw-
Hill,
New
York,
1964
.
Lloyd P
.
Hunter (Ed
.),
"Handbook
of
Semiconductor
Electronics,"
second
edition,
McGraw-Hill,
New
York,
1962
.
Jacob
Millman
and
Herbert
Taub,
"Pulse,
Digital,
and
Switching
Waveforms,"
McGraw-Hill,
New
York,
1965
.
The
main
headings
in this
circuit
analysis
refertosche-
maticsinthe
diagrams
section
with
the
same name.The
sub-headings
indicate
the
individual
circuit
being described
.
The
main
block
diagraminthe
last
sectionofthe
manual
shows
interconnection
between
circuits
.
MAIN TRIGGER
PREAMP
The
Main
Trigger
Preamp
selects trigger
source
and
coupling
for
the
Main
Trigger
Generator
and
provides
am-
plification
for
external horizontal
signals
when
the
TIME/
DIV
OR
DL'Y
TIME
switchissettoAMPL
.
This
circuit
consists
of four
stages:Trigger
Source
Switching, External
Trigger
Preamp
or External
Input
Am-
plifier,
Balanced-to-Single-Ended
Converter,
and
Trigger
Coupling.Fig.3-2
showsadetailed
block diagramofthe
Main
Trigger
Generator
circuit
and the
schematicofthis
circuitisshown
on diagram1at
the
rearofthis
manual
.
Trigger
Source
Switching
U330
receives
trigger
inputs at
pins2and15for
internal
triggering
andatpin7for
external
trigger signals.Pin4of
U330
determines
which
input
signalisselectedbymeans
of
J16
IN or
EXT-10
Trigger
Signal
-0
0
Frame
S7
IJV
RCE
SOU
f1W
T
Internal
.
_±__~
ger
Signal
Vertical
S6
COUPLING
Balanced-to- Single-Ended
Converter 0352,
12354,
To
Slope
Selector and
Level
\
Comparator
01
Trigger Coupling 0362,
12364
12366
3-4
3-2.Main
Trigger
Preamp
detailed
block
diagram
.
External
Trigger
Preamp
or
Horizontal
Input
Am-
plifier
SOURCE
switchS7at
the
input
selects internal,
ex-
ternal,orline
signals
for
triggering.The
external
trigger
(or
horizontal
input)
signal
may
be
attenuated
to
one-tenth
amplitudebyselecting
EXT-10
.
R13
and
R14
(paralleled
by
R302)
forma10:1attenuator
.
The
input
impedance
for
the
trigger
(or amplifier)
input
is1megohm,
consisting
primarilyofR5
and
R302.This
resistor
pair
also
causes
a2Xattenuationofthe
input
signal
as
seenatthe
gateofQ308
.
C301
compensates
the
input
stage
and
C10
compensates
the
10X
attenuator
.
CR303
and
CR305
protect
Q308
from
excessive
input
signalbyclamping
the
gateifthe
signal at
the
input
con-
nector
exceeds
approximately+or-2.5
volts.The
signal at
the sourceofQ308iscoupled through
emitter-followers
0312
and
Q316
to
pin7of
U330.The
signal at
pin
7
of
U330isterminatedinapproximately50ohms
by
R319
to
preserve
the
high-frequency
characteristics
.
a
digital signal
(voltage
level).A
Low
Levelatpin4acti-
vates
pins2and 15
for
internal
triggering,
whileahigh
level
on
pin
4
switches
U330
to
activate pins7and10for
ex-
ternal
triggering
.
To
further
examine
U330,
assume
that
pin4is
low,
activating
pins
2 and 15
(internal
triggering).This
input
is
a
relatively
high
impedance
differential
configuration.Pin
15
receives
the
positive-going
trigger
signal
and
pin2is
the
negative-going
input.The
inputs
are
biasedatthe
center
of
their
dynamic
range,
and
signal-limitinginthe
trigger
pick-
off
circuitry
(in
the
indicator
oscilloscope)
assures
that
the
inputs
will
not be
driven
into
either
cutofforsaturation
.
R336
and
R337
terminate
the
internal trigger
signal
from
the
indicator oscilloscope.The
current
source for
internal
triggeringisby
wayofPins1and16and
R343
.
The
output
current
appears
at pins12and13.
A
positive-going
signal at
pin15causes an
increase
in
current
into
pin13and
out
through
pin
16,
R341
and
R343
.
Si-
multaneously,
the
negative-going
signalatpin2causes
a
decreaseincurrent
into
pin12and
out through
pin
1,
R342,
and
R343
.
The
net
resultisthat
the
total
current
through
pins 12
and13and
R343
remains
constant
.
The
current
source
for
external
triggeringisby
way
of
pins8and9and
R346.Operationisthe
sameasinternal
triggering
.
Circuit
Description-71353N
R330
sets
theDClevelatpin
10ofU330,
whichisthe
negative
sideofthe
external
trigger
differential
input.This
matches
the
DC
balanceofthe
external
trigger
input
of
U330
to that
of the
internal
trigger
input
.
Balanced-to-Single-Ended
Converter
The
balancedtosingle-ended
converter
changes
the
out-
putofU330toa
single-ended
signal at
the
emitterofQ358
.
The
trigger signal
through
U330
causesadecreaseincurrent
into
pin
12
from
R350
and
R354
andanincreaseincurrent
into
pin13from
R351.This
would
normally
cause
the
voltageatpin
12toswing
in
a
positive
direction,
while
pin
13
goes
inanegative
direction.However,
the
current
through
R350
and
R354
actually
increases
duetothe
feed-
back
via
R355
and
Q354,
causing
the
voltageatpin12to
swing
negative
along
with
pin13.
Q354isconnectedasa
diode
andisenclosedinthe
same
heat-sink
with
Q352,
providing
good
DC
stability.The
DC
Balance
adjustment
R350
sets
the
quiescentDClevelofthe
BalancedtoSingle
Ended
Converter
to
zero
volts
allowing
theDClevelatthe
outputtocorrespondtotheDClevel
at
the
input
.
Trigger
Coupling
When
DC
coupling
is
selected
by the
front-panel
COUPLING
switch,
Q362
is
turned
on by
the
+15
volts
supply through
R18,
S6,
and
R361.The
triggering
signal
is
then
coupled through
R359
and
Q362
to
the
baseofQ402
(on Main
Trigger
Generator
diagram)
.
Q364isturnedonwhen
AC
couplingisselected
.
The
triggering
signal
then
passes
through
Q364
and
C364tothe
base of
Q402.For
AC
LF
REJ
coupling,
Q364isoff
and
the
triggering
signaliscoupled through
C362
and
C364,
attenuating
low-frequency
signals
.
For
AC
HF
REJ
coupling,
both
Q364
and
Q366
are
turnedon.
The
high-frequency
components
are
coupled
through
C367
and
Q366
to
ground,
while
the
desired
trig-
gering
componentiscoupled through
Q364
and
C364
(as
in
AC
coupling)
.
MAIN
TRIGGER
GENERATOR
The
Main
Trigger
Generator
provides
selection
of
the
level
and
slope
where
triggering
occurs
and
suppliesafast-
rise
uniform-amplitude
pulsetothe
Main
Sweep
Start
Multi
.
The
Main
Trigger
Generator
includes
the
Level
Com-
parator,
Slope
Selector,
Trigger
TD
and
Driver,
and
Main
Trigger
Generator
circuits.Fig.3-3
shows
a
detailed
block
diagramofthe
Main
Trigger
Generator
circuit
and
the
sche-
matic
is
shown
on
diagram 2
at
the
rearofthe
manual
.
3-
5
Circuit
Description-71353N
R2
LEVEL
(Main)
Main Trigger Signal
I I
1
S2
SLOPE
(Main)
Holdoff Pulse
Slope
Selector
and
Level
Comparator
Q402
and
0404
are
connectedasa
differential
com-
parator.The
reference
voltage
for
the
comparator
is
selected
by
the
settingofLEVEL
controlR2.
The
Main
Trig
Level
Center
adjustment,
R410,
sets
the
levelatthe
base of
0404sothe
sweepistriggeredatthe0volt
point
of
the incoming
trigger
signal
when
the
LEVEL
controlisset
to
the
centerofthe
positiveornegative
slope
region
.
The
LEVEL
control
varies
the
voltage
on
the
base of
Q404
to
select
the
point
on
the
trigger
signal at
which
triggering
occurs
.
0408
and
R408
establish
the
emitter current
for
Q402
and
Q404
.
Priortothe
arrivalofa
trigger
signal,
with
the
LEVEL
control
settothe
centerofthe
positive
or negative
slope,
Q402
and
Q404
are
passing
equal
currents
.
Assume
that
a
positive-going
signalisappliedtothe
MAIN
TRIG
IN
connector
and
that
the
LEVEL/SLOPE
controlisset
to centeronthe
positive
slope.The
signal at
the
MAIN
TRIG
IN connectorisinverted
by the Main
Trig-
ger
Preamp,
appearingatthe
baseofQ402asa
negative
going
signal.This causesadecreaseincurrent
through
Q402,
and
becauseofthe
common
emitter
source
(Q408
and R408),
the
current
through
Q404
increases.The
de-
creased
collector
currentofQ402
biases
Q418ina
reverse
direction,
while
Q416
becomes
more
forward
biased
due
to
the
increased
current
through
Q404
.
With
SLOPE
switchS2in
the+position,
the cathode
of
CR424isgrounded,
forward
biasing
CR424,
which
reverse
biases
CR423
.Atthe
same
time,
the
baseof0428isat
3-
6
Fig.3-3.Main
Trigger
Generator
detailed
block
diagram
.
ground and
Q428isoff.This
causes
CR421tobe
reverse
biased
and
CR422isforward
biased
through
Q416.An
in-
creased
currentisappliedtothe
TriggerTDand
Driver
circuit
through
Q416
and
0422
(see Fig.3-4)
.
When
the
SLOPE
switchissettothe-position,
Q428
and
CR421
are
forward
biased
and
CR422isreverse
biased
.
CR424isreverse
biased
and
CR423isforward
biased so
that
current
flows
through
Q418
and
CR423tothe
Trigger
TD
and
Driver
circuit
.
TriggerTDand
Driver
The
Trigger
TD
stage
shapes
the
outputofthe
com-
parator
to
provideatrigger
pulse
withafast
leading
edge
.
Tunnel diode
CR430isquiescently
biased
so that
it is
in
its
low-voltage
state.Increased
trigger
current
from
Q416
and
CR422or0418
and
CR423
through
R432,
L432
and
CR430
causes
CR430toswitchtothe
high-voltage
state
.
The
resulting
fast-rise
positive
stepiscoupled
through
emitter-follower
Q434toC451
and
C461inthe
Main
Trig-
ger
Generator
stage
.
Main
Trigger
Generator
The
Main
Trigger
Generator
includes
Q454,
Q466,
CR470
and
CR475.The
functionofthis
stageisto
supply
a
fast-rise
trigger
signal to
the Main
Sweep
Start Multi.For
normal
triggering,
this signalisdeveloped
after
receiptofa
fast-rise
transition
from
the
TriggerTDand
Driver
stage,
except
during
holdoff
.
O
Level
Slope
Trigger
Main
Compar-
Selector
To
and
Trigger
ator
Driver
Generato
Triggered
Q402,
Q416,
Q466,
Sweep
Enable
0404, Q418,
CR430,
CR470
Q408 0428
Q434
0454,
CR475
+5
V
+15
V
1
L__________
+5V
i
i
i
LEVEL/SLOPE
- -
+15
V
+5
V
n
n
v
I
VWCT1
W
Fig
.
3-4
.
Trigger current
for
positive
slope
triggering
.
Z
Circuit
Description-71353N
For
the
following
description
of operation,
assume
that
the
MODE
switchissettoNORM
and
that
a
trigger
signal
is
appliedtothe
MAIN
TRIG
IN
connector
.
CR470
and
CR475
are
bothintheir
high
state-,
until
the
holdoff
signal
switches
themtothe
low
state
.
The
holdoff
signal
from
pin17of
U720isa
positive
pulse
which
for-
ward
biases
both
0454
and
Q466
(see
the
Holdoff
Circuit
discussion
under
Main
Sweep
Generator
in this
section)
.
When
these
transistors
are
forward
biased,
they
divert
cur-
rent
from
CR470
and
CR475
which
causes the tunnel
diodes
to
switchtothe
low
state
.
The
next
trigger
after
holdoff appearsasa
positive
tran-
sition
at
C451
and
C461.The
positive
transition,
coupled
through
R461
and
R462,
causes
CR470
to
switchtoits
high
state.This higher
level,
through
R472,
brings
CR475
up
to near
its
switching
current.The
positive
transition
is
also
coupled through
C451
and
R451,
and
after3.5
nano-
seconds
of delay,
through
R474toCR475.The
short
delay
assures
that
CR470
has
had
timetoswitchtoits
high
state,
arming
CR475
before
arrivalofthe
switching
signal
at
CR475.This
prevents
extraneous
noise
from
prematurely
activating
CR475.CR475
then
switches to
its
high
state
.
The
fast-rise
positive
trigger
from
CR475iscoupledtothe
Main
Sweep
Start
Multi,
Q722/Q726
.
MAIN
SWEEP
GENERATOR
The
Main
Sweep
Generator
circuit
producesasawtooth
voltage
which
is
amplifiedbythe
Horizontal
Amplifier
circuittoprovide
horizontal
sweep
deflection
on
the
CRT
of
the
indicator
oscilloscope.This
output
signalisgenerated
on
command
(trigger
pulse)
from
the Main
Trigger
Gener-
ator
.
The
Main
Sweep
Generator
also
producesaMain
Sweep
Gate
pulse
coincident
with
the
time
that
the
Main
Sweep
runs.The
Main
Gate
pulseisprocessedbythe
Sweep
Gate
Out
circuit
and the
indicator oscilloscope
for
CRT
unblanking
and
Auxiliary
Gate
output.In
addition,
the
Main
Sweep
Generator
produces
several
control
signals
for
other
circuits
within
the
instrument
.
Fig.3-5
showsade-
tailed
block
diagram
of
the
Main
Sweep
Generator
and
the
schematic
is
shownondiagram 3atthe
rear
of
the
manual
.
The
MAIN
TRIGGERING
MODE
switch allows
three
modes
of
operation.When
the
NORM
button
is
pressed,
a
sweepisproduced
only
when
a
trigger
pulseisreceived
from
the
Main
Trigger
Generator
circuit.When
the
AUTO
buttonispressed,
a
sweepisproducedasin
NORM
except
thatafree-running
traceisdisplayed
whenatrigger
pulse
is
not
present
.
SINGLE
SWEEP
MODE
operationisalso
sim-
ilartoNORM
operation
except
that
the
sweepisnot
re-
current.The
RESET
button
must
be
pressedtoview
another
trace
.
The
following
circuit
descriptionisgiven
with
the
MAIN
TRIGGERING
MODE
switch
pressed
to
3-
8
NORM.Differenceinoperation
for
the
other
two
modes
is
discussed
later
.
Main
Sweep
Start
Comparator
Q722, Q726,
and
Q728
comprise
the
Main
Sweep
Start
comparator.In
the
absenceofa
trigger,
Q722
is
off
and
Q726isheld
on
by
the
high
level
from
pin
3ofU720.The
collector
of
Q726islow
and
this
lowiscoupled
through
emitter
follower
Q728
to
pin1of
U750,
thus
preventing
a
sweep.When
the
Main
Trigger
Generator
supplies
a
trigger,
the
positive
transitioniscoupled
to
the
base of
Q722.The
base
of
Q722
rises
above
the
level at
the
baseofQ726
and
the
current
through
common
emitter
resistor
R724
is
di-
verted
from
Q726
to
Q722.The
collectorofQ726
rises
and
the
positive
step
is
coupled
through
emitter-follower
Q728
.
The
positive
step
appears
across
divider
R731/R732
causing
pin1of
U750togo
positive,
thus
starting
the
sweep
.
Sawtooth
Generator
The
lower
half
of
the
U750
diagram
symbol
constitutes
a
Miller
Integrator.When
pin1is
positive,alinear
sawtooth
(positive-going)
is
generated
and
appears
at
pin8.
The
timing
components,
Rt and Ct
connected
to pins8and
9,
determine
the
rate
of
changeofthe
sawtooth
waveform
.
Q774
prevents
high-speed
error
currents
from
being
coupled
into
U750
by
wayofC754
and
pin9.
Sweep
Stop
Comparator
The
Sweep
Stop
Comparator
consists
of
Q794
and
Q796.In
the
absenceofa sawtooth
signalatpin8of
U750,
Q796
is
conducting
and
Q794
is
held
offbythe
positive
level
setatits
basebyR795,
the
Main
Sweep
Length
adjust-
ment.When
the
sawtooth
voltageatpin8of
U750
raises
the
base of
Q796
higher
than
the
base of
Q794,
Q796
turns
off
and
Q794
turnson.
The
collector
of
Q794
rises
and
the
positive stepiscoupled
through
emitter
follower
Q792
to
pin16of
U720
and
sweep
holdoff
begins
.
The
positive
step
at
pin16will
also
be seenatpin
17
of
U720,
and
conse-
quentlyatthe
bases
of
Q454
and
Q466
(Main
Sweep
Generator
Circuit).These
transistors
are
forward
biased,
which
diverts
the
current
from
CR470
and
CR475
and
causes
the tunnel
diodestoswitchtotheir
low
state.As
a
result,
Q722
turns
off
and
Q726
turnson.
The
collector
of
Q726
drops
and
the
negative
stepiscoupled
through
emitter
follower
Q728,
thus
ending
the
sweep
.
Holdoff
Circuit
The
Holdoff
Circuit
consists
of pins8,10,
16,
and
17 of
U720
plusRandCtime
constants
selectedbythe
TIME/
DIV
switch
.
The
holdoff
prevents
re-triggering
the
sweep
generator
until
after
the
sweep
timing
capacitor(s)
has
dis-
charged
and
sweep
circuits
are
again
ready
to generate
a
sweep
.
O
DISPLAY
MODEI
Fig
.
3-
5
.
Sweep
Generator
detailed
block
diagram
.
M
TIME/
DIV
To
Composite
Sweep
Out
and
Mixed
Sweep
Comparator
Holdoff
to
Dly'd
Sweep
Trigger
Generator
Holdoff
to
Trigger
Generator
Circuit
Description-71353N
At
the
endofthe
sawtooth
waveform,apositive step
is
coupledtopin
16 of
U720,
by
way
of
the
Sweep
Stop
Comparatoraspreviously
described.The
positive
pulse
seen
at
pin
16 of
U720iscoupled
internally
through
U720
to
pin
17
and
in
turntoQ454
and
Q466inthe
Main
Trigger
Generator.Q454
and
Q466
are
turned
on,
thus
robbing
the
current
from
CR470
and
CR475,
switching
themtothe
low
state.This
prevents
the Main
Sweep
Start Multi
from
gener-
atingasweep
.
Afteratime
determined
by the
timing
components
at
pin8,internal
circuitry
within
U720
switches
pin
17
to
its
low
state,
ending the
holdoff
gate.The
Main
Trigger
Gener-
atorisreleasedtogenerateatrigger
signal
.
A
negative
gate
coincident
with
the
positive
holdoff
gate,
appears
at
pin
10ofU720.This
negative
gateisin-
vertedbyQ748
and
coupledtothe
Delayed
Sweep
Gener-
ator
for
composite
holdoff
functions
.
Trig'd
Lamp
Driver
When
the
main
sweep
gateishigh
and
the
sweep
is
running,
the
TRIG'D
lampison.At
all
other times
the
lampisoff
.
Delay
Pickoff
The
upper
halfofthe
diagram
symbol
for
U750
includes
the
Delay
Pickoff
circuitry.Inside
U750,
the
main
sweep
sawtooth
signalisappliedtoone
sideofa comparator
circuit
.
Pin
6isconnectedtothe
other
sideofthe
com-
parator
.
The
setting
of
the
DELAY
TIME
MULTIPLIER
control,
R19,
determines
the
pointonthe
main
sweep
saw-
toothatwhich
the
comparator
switches
.
When
the
comparator
switches
(delay
pickoff
occurs),
a
positive
gate
appearsatpin4of
U750.This
gate
terminates
at
the endofthe
main
sweep
sawtooth
.
The
positive-going
gateatpin4of
U750
is
coupled
through
emitter-follower
Q762
to
the
Delayed
Trigger
Generator
.
Auto
Triggering
Mode
Operationofthe
Main
Sweep
Generator
circuitinthe
AUTO
positionofthe
MAIN
TRIGGERING
MODE
switch
is
the
sameasfor
NORM
position
just
described
when
a
trigger
pulseisapplied.However,
whenatrigger
pulse
is
not
present,afree-running
reference
traceisproducedinthe
AUTO
position.This
occursasfollows
:
The
Auto
Triggering
circuit
consistsofpinsI,3, 6,
and
19ofU720
.
When
the
AUTO
buttonofthe
MAIN
TRIG-
3-
1
0
GERING
MODE
switchispressed,alowatpin19of
U720
enables
the
Auto
Circuit.Whenarepetitive trigger
signal
above
30
Hz,
andofadequate
amplitudeisappliedtothe
Main
Sweep
Start
Comparator
and
pin1of
U720,
the
in-
ternal
Auto
Multiatpin6of
U720
charges
towards
five
volts
through
C786
and
R786,
butisdischargedbyeach
incoming
trigger
pulse
.
In
the
absenceofa
trigger
pulse,
C786
charges
towards
+5
V,
switching pin6to
its
high
state
and
pin3to
its
low
state.Q726
turns
off,
its
collector
rises
andahigh
is
coupled through
emitter
follower
Q728topin1of
U750,
causing
the
sweeptorun
.
Single
Sweep
Operation
Operationofthe Main
Sweep
Generatorinthe
SINGLE
SWEEP
position of
the
MAIN
TRIGGERING
MODE
switch
is
similartooperationinthe
NORM
positionaspreviously
described.However,
after
one
sweep
has
run,
all
other
sweeps
are
inhibited
until
the
RESET
buttonispressed.A
READY
lamp
is
providedtoindicate
when
the
sweep
is
readytoacceptatrigger
.
The
Single
Sweep
circuit
consists
of pins
11, 12, 14, 15,
and
17ofU720.For
SINGLE
SWEEP
operation,
the +5
volt
supplyisappliedtopin12of
U720.The
holdoff
pulse
at
pin17of
U720
goes
positive,
preventing
generationofa
sweep.When
the
RESET
buttonispressed,
pin15is
mo-
mentarily
held to
ground,
pin
17
goes
lowtoallow
the
Main
Trigger
Generatortoacceptatrigger.The
holdoff
line
(pin
17
of
U720)
stays
low
untilasweep
has been
com-
pleted.At
this
time,
the
holdoff
pulse
rises at
pin17and
staysinthe
holdoff
state
until
the
RESET
button
is
pressed
.
0744
actsasa
switch for the
READY
lamp.When
the
holdoff
gateatpin17is
high,
preventing
the
sweep
gener-
ator
from
acceptingatrigger,
pin
11ishigh
and
Q744
and
the
READY
lamp
are
off.When
the
RESET
button
is
pressed,
the
holdoff
gateatpin17goes
low
and
allows the
Main
Sweep
Generatortoacceptatrigger.Pin11rises
and
turnson0744, which
provides
the
current
to
turnonthe
READY
lamp
.
Sweep
Lockout
Q714,
Q716
and
pins3,16,
and18comprise
the
Sweep
Lockout
circuit.The
Sweep
Lockout
circuitisfunctional
when
the
7B53NisinstalledintheBHorizontal
compart-
ment
of an
indicator oscilloscope
which
accommodates
two
horizontal
plug-in
units,
anditis
desiredtooperateinthe
Alternate
Horizontal
Mode,orto
operate
the
71353N
as
a
delayed
sweep
unit.Lockoutisappliedtothe
7B53N
during the
time
that
the
sweep
from
the
associated
time
base
is
displayed
.
The
indicator
oscilloscope
controls
initiation
of
a
sweep
by
supplying
current
to
the
base of
Q714
when
lockout
is
required.This
current
causesapositive step
at
pin18of
U720.Pin3of
U720
steps
positive
and
Q726
turnson.
The
collector
of
Q726
falls
and
the
lowiscoupled
through
emitter
follower
Q728
to
pin
1ofU750,
thus
preventing
the
sweep.If
lockout
is
initiated
while
the
sweep
is
running,
the
leading
edge
of
the
lockout
pulseisdifferentiated
through
C719
and
D719,
and
appearsasa
high
at
pin16of
U720
.
This
starts
the
holdoff
cycle.(The
holdoff
cycleisas
described
previously)
.
Delayed
Mode
Control
When
the
7B53N
is
installedintheBHorizontal
com-
partment
of an
indicator
oscilloscope
with
two
horizontal
compartments,
the
Delayed
Mode
Control
determines
whether
the
7B53N
operatesasan
independent timebase
or
as
a
delayed
sweep
unitinthe
triggerable
after
delay
time
mode.When
approximately+3to4.5
voltsispresent
at
interface
connector B1
(and
therefore
pin13of
U720),
the
Auto
Circuit
(previously
described)isdisabled.A
sweep
can
be
enabled
only
by a
trigger
pulsetothe
Sweep
Start
Comparator.During
delay
time determined
by
the
settings
of
the
delaying
sweep
unit,
sweep
lockout
(previously
de-
scribed)
inhibits
the
sweep
.
After delay
time,
the
7B53N
J18
DLY'D
TRIG
IN
OR D
LY'D GATE OUT
+5
V
(Dly'd)
T
I l
Ext
Int
SOURCE
S18
i
10
Int
I I
I
I
l
Delayed
Gate
from Delayed
Sweep
Start
Control
Inter
Trigger
Signal
from Vertical Unit
Dly'd External
Trigger
Preamp 0508,
0512,
0516
S10B

(MAINSw1

S15
IME/DIV
I
DLY'D
TRIGGER
I
MODE
(AMPL)
I
RUNS
AFTER)
Fig.3.6.
Delayed
Trigger
Preamp
detailed
block
diagram
.
Circuit
Description-712153N
can
be
triggered.An
approximate
zero
volt
level at
pin
13
of
U720
enables
the
Auto
Circuit,
causing
the
7B53N
to
operateasan
independenttime
base
.
DELAYED
TRIGGER
PREAMP
The
delayed
Trigger
Preampisvery
similar
to
the
Main
Trigger
Preamp
previously
described
.
Therefore,
only
those
portions
that
are
different
will
be
describedindetail.This
circuitry
selects
trigger
source
and
coupling
for the
signal
driving
the
Delayed
Trigger
Generator.Fig.3-6
shows
a
detailed
block
diagram
of
the
Delayed
Trigger
Preamp
circuit
and
the
schematic
of
this
circuitisshown
on
diagram4at
the
rear of
the
manual
.
Dly'd
Trigger
Source
Switching
U530
initiates
the
Delayed
Trigger
Mode.When
the
DISPLAY
MODE
switchisset
to
INTEN,
DLY'D
SWP,
or
MIXED
position
and
the Delayed
Triggering
LEVEL
controlisin
the
OUT-DLY'D
SWP
TRIGGERABLE
position,
a low
level
is
coupled through
08
to
pin6of
U530,
which
allows
U530toacceptatrigger
.Ahighatpin
6
(DISPLAY
MODE
switch
settoMAIN
SWP,
TIME/DIV
S12
DISPLAY
MOD
Trigger
Inhibit
08
S17
COUPLING
(Dly'd)
Dly'd
Trigger
Coupling
0562
~
J
To
Delayed Trigger Slope
Selector and
Level
Comparator
F
3-11
Circuit
Description-71353N
OR
DL'Y
TIME
switch
set
to
AMPL,
or
the
Delayed
Triggering
LEVEL
controlatthe
IN-RUNS
AFTER
DL'Y
TIME
position)
causes
U530torefuse
a
trigger
.
U530
also
performs
the
functionofselecting
either
the
Dly'd
Internal
Trig
Amp
or the
Dly'd
External
Trig
Amp
as
the
sourceoftrigger
.
When
pin4of
U530ispositive,
pins
7
and10are
activated
andanexternal
trigger
must
be applied
to the
DLY'D
TRIG
IN
connector.When
pin4is
low
(near
ground),
pins2and
15
are
active
and
an
internal
trigger
sourceisselected
.
Dly'd
External
Trig
Amp
Q508,
Q512
and
Q516
comprise
the
Dly'd
External
Trig
Amp.This
circuitisidenticaltothe
Main
External
Trigger
Preamp.The
amplifier
providesacurrent
gain
and
is
terminated
by
R519
at
pin
10 of
U530
.
Balanced-to-Single-Ended
Converter
This
circuit
includes
Q552,
Q554
and
Q558.Except
for
minor
differences
in
component
values,
the
circuitry
is
identical
with
the
Balanced-to-Single-Ended
Converter
in
the
Main
Trigger
Preamp.The
output
signal
at the emitter
of
Q558
is
inverted
from
the
signal at
the
DLY'D
TRIG
IN
connector
.
R15
LEVEL
(Dly'd)
Trigger
Signal
Dly'd
Level
Comparator
Q602,0604
Dly'd
Slope
Selector
0616,
0618
0828
SLOPE
(Dly'd
)
Delay
Gate
from
Delay
Pickoff
Dly'd
Swp
Holdoff
Dly'd
Trigger
Generator
Q654, Q666,
Q862,
CR670
CR675
DELAYED
TRIGGER
GENERATOR
To
Dly'd
Sweep
-~
Holdoff
Dly'd
Sweep
Enable
3-12
Dly'd
Trigger
Coupling
Fig.3-7.Delayed
Trigger
Generator
detailed
block
diagram,
When
COUPLING
switch
S17
is
settoDC,
0562
is
forward
biased
by wayofR56
and
R561
.
The
trigger
signal
is
direct
coupled
between
the
emitter
of
Q558
and
the
base
of
Q602
(Slope
Selector
and
Level
Comparator)
.
If
the
COUPLING
switchissettoAC,
Q562isreverse
biased
and
the
trigger
signaliscoupled
through
C562
.
DELAYED
TRIGGER
GENERATOR
The
Delayed
Trigger
Generator
circuitry
is
essentially
the
same
as
the
Main
Trigger
Generator,
except
thereisno
provision
for
automatic
mode
of
triggering
.
Therefore,
only
the
circuits
that
are
different
willbedescribed
in
detail.For
a
detailed
description of
the
rest
of
the
circuitry,
refer
to
the
Main
Trigger
Generator
.
Fig.3-7
showsadetailed
block
diagram
of
the
Delayed
Trigger
Generator,
and
the
schematicisshown
on
diagram5at
the
rear
of the
manual
.
Slope
Selector
and
Level
Comparator
This
circuit
consistsofQ602, Q604,
Q616,
Q618
and
Q628
.
Operation
is
identical
to
the
Slope
Selector
and
Level
Comparator
on
the
Main
Trigger
Generator
diagram
.
O
When
the
input
signal
at
the
base of
Q602
passes
through
the
level
set
at
the
base of
Q604,
an
increaseincurrent
occursatthe
output
.
Trigger
TD
The
Trigger
TD
is
CR630.The
increased
current
caused
by
applying
a
trigger
signaltothe
Slope
Selector
and
Level
Comparator
circuitiscoupled
through
R632
and
CR630,
switching
CR630toits
high
state
.
Delayed
Trigger
Generator
The
Delayed
Trigger
Generator
circuit
includes
Q654,
Q666, Q862,
CR670
and
CR675.Operationofthe tunnel
diodes,
CR670
and
CR675,
is
identicaltooperationofthe
TD'sinthe
Main
Trigger
Generator
.
If
the
DLY'D
LEVEL
controlispushed
to
the
IN-RUNS
AFTER
DL'Y
TIME
position,
S15
isinthe
"open"
position
and
current
through
R871,
CR655,
and
R674
to
the
Sweep
StartTD(CR675)
biases
CR675
just
below
the
switching
level.When
the Delay Gate
is
generated
(at
the
Trigger
Pickoff),
the
positive stepatthe
junction
of
CR866
and
CR869
forward
biases
Q862.This
increases
current
through the
Sweep
Start
TD,
causingitto
switch
to
the
high
state.This
occurs
immediately
upon
arrival
of
the
Delay
Gate,
without
need
foradelayed
trigger
input
.
When
the
DLY'D
LEVEL
controlisin
the "out"
position
(DLY'D
SWP
TRIGGERABLE),
S15
is
closed,
forward
biasing
CR71.Q954
becomes
forward
biased
and
Q862isreverse
biased.The
static
current
through
CR675
is
at
a low
level.Q862
becomes
forward
biased
upon
arrival of
the
Delay
Gate
signal at
its
emitter
(via
CR866).The
resulting
current
biases
the
Sweep
StartTDjust
below
the
switching
level.A
trigger
signal
from
the
Trigger
TD
(CR630)
then
causes
the
Sweep
StartTDto
switch to
the
high
state
.
The
Dly'd
Swp
Holdoff coupledtothe
basesofQ654
and
Q666
prevents
the
Sweep
Start
TD
from
switching
until
after
the
main
sweep
has
occurred
.
DELAYED
SWEEP
GENERATOR
The
Delayed
Sweep
Generator
produces
a
sawtooth
voltage
which
is
amplified
by
the
Horizontal
Amplifier
circuitstoprovide
a
delayed
sweep
CRT
display.The
sawtooth
output
voltageisgeneratedoncommand
of
the
Delayed
Trigger
Generator.The
Delayed
Sweep
Generator
also
producesaDelayed
Sweep
Gate
pulse,
coincident
with
the
time
that
the
Delayed
Sweep
Generator
runs,
to
be
processed
by
the
Sweep
Gate
Out
circuit
and
the
indicator
oscilloscope
for
CRT
unblanking.Fig
.
3-8
showsadetailed
Dly'd
Swp
Start
Multi
Dly'd
Swp
Start
Control
Mixed
Swp
Comparator
Circuit
Description-71353N
block diagramofthe
Delayed
Sweep
Generator and
the
schematicisshown
on
diagram6at
the
rearofthe
manual
.
Q882
and
Q886
comprise
the
Dly'd
Swp
Start
Multi
.
This
circuitisconnectedasa
bistable
multivibrator,
with
Q886
normally conducting
and
Q882
off
.
When
the
Sweep
StartTDswitches
to
its
high
state,
the
positive
step
appearsatthe
base
of
Q882.This
causes the
multitoflip,soQ882ison and
Q886
off.The
collector
of
Q886
goes
positive.The
Sweep
Start
TDisheldinits
high
state
for
the
durationofthe
Delay
Gate
.Atthe
endofthe
Delay
Gate,
the
Dly'd
Sweep
Start
Multi
reverts
to
its
original
state
with
Q882
off
and
Q886on.
The
Dly'd
Swp
Start
Control
circuit
includes
Q902,
0904
and
Q906.This
circuit
couples
a
positive gatetopin
1ofU930
(Miller
Integrator)tocontrol
the
period
during
whichasawtooth
is
generated
.
In
all
Positions
of
the
DISPLAY
MODE
switch
except
MIXED,
0902
and
Q904
are
inactive
due
to
reverse bias
current
via
CR901,
S12
and
the+5volt
supply
.
When
the
collectorofQ886
(Dly'd
Swp
Start
Multi)
goes
positive,
Q906
couples
the
positive gate
to
pin1of
U930,
initiating
the
generationofa
delayed
sawtooth.At
the
endofthe
Delay
Gate,
the
collectorofQ886
drops.This ends
the
positive
gate to
pin1of
U930,
terminating
the
delayed
sawtooth
.
When
the
DISPLAY
MODE
switchissettoMIXED,
the
anode
circuitofCR901
is
open
.
The
gate
from
the
Main
Swp
Start
Multi
is
negative-goingatthe
baseofQ904.The
resulting
current
from
-Q904 forward
biases
Q906,
and
a
positive gateiscoupledtopin1of
U930
.
Q888, Q892,
Q896
and
Q898
comprise
the
Mixed
Swp
Comparator
circuit
.
This
circuit
determines
whether
U930
is
runningatthe
main
sweepordelayed
sweep
rate
.
With
the
DISPLAY
MODE
switch
settoMIXED,
Q892
is
forward
biased.The
main
sweep
sawtooth
at
the
emitter
(and
thus,
the
collector)ofQ892isa
positive-going
ramp
.
This
causesaramp
of
increasing
current
through
0896
.
During
the
time
thataDelay Gate
is
not
being
generated,
Q882
(Dly'd
Swp
Start
Multi)isbiased
off
and
Q888ison
.
In this
condition,
U930, Q888,
Q896
and
Q898
form
a
operational
amplifier.The
negative-going
ramp
at
the
collectorofQ896
becomesapositive-going
rampatpin
8
of
U930,
runningatthe
main
sweep
rate
.
3-13
i
Mainframe
Mode
-L
A76
Mainframe
B7
,
Information
Swp
Start
Comparator
Sweep
Sawtooth
S12
DISPLAY
MIXED
TIME/DIV
DLY'D
Aux
Z
Control
Q984,Q988
Dly'd
Sweep
I
Start
Control
+5
V
Q902,
10
I
Mixed
Sweep
Comparator
Q888,
Q896,
DISPLAY
Aux
Z
",
Out
Dly'd
Swp
P/O
Swp
Sweep
Holdoff
Dly'd
Sweep
Lockout
Multi
Q942,
Dly'd
Sweep
Holdoff
Holdoff
to
Delayed
--I
"
Trigger
Generator
S12
DISPLAY
A3
B3
Fig
.
3-
8
.
Delayed
Sweep
Generator
detailed
block
diagram
.
c
TIME/DIV
-1
-
DISPLAY
v
(DLY'D)
Composite
Sweep
~'
Composite
S
Out
Al
Out
Q922,
0924,
V
Aux
Z-Axis
Out
A17
W
Ci1
Z
LEVEL
S15
Dly'd
T
In
-15
V
-0
410
When
the
Delay
Gate
is
generated,
the
Delayed
Trigger
Generator
forward
biases
Q882.The
collector
current
through
R888
reverse
biases
Q888,
opening
the
operational
amplifier
loop.U930isreleased
to
runatthe
delayed
sweep
rate.Therefore,
the
sawtooth
at pin8of
U930
will
first
run
at
the
main
sweep
rate
and
then
change
to
the
delayed
sweep
rate
when
the
Delay
Gate
is
generated
.
Dly'd
Swp
Stop
Circuit
The
upper
half
of
the
diagram
symbol
for
U930
constitutes
the Dly'd
Swp
Stop
Circuit.The
setting
of
the
Dly'd
Sweep
Length
adjust
(R930)
determines
the
point
on
the
delayed
sawtooth
at
which
pin4of
U930
will
go
positive
.
Dly'd
Swp
Lockout
Multi
Q942
and
Q944
form
the
Dly'd
Swp
Lockout
Multi
.
This
circuit
serves
to
terminate
the
delayed
sweep
as
determined
by
the
setting
of the
Dly'd
Sweep
Length
adjust.When
pin
4
of
U930
goes
positive,
Q942
becomes
forward
biased
.
The
negative-going
stepatthe
collectorofQ942
forward
biases
Q902
(Dly'd
Swp
Start
Control
circuit).Q904
and
Q906
become
reverse
biased,
dropping
the
levelatpin1of
U930
and
terminating
the
sweep
.
Dly'd
Swp
Holdoff
The
Dly'd
Swp
Holdoff
circuit
includes
Q954
.
The
holdoff
gateatconnectorGis
a
composite
of the
positive
gate
from
the
Dly'd
Swp
Lockout
Multi,
the
Main
Swp
Holdoff
Gate
via
R952;and,
when
the
DLY'D
LEVEL
controlissettoDLY'D
SWP
TRIGGERABLE,
the
positive
level
setby0954
.
With
the
DLY'D
LEVEL
control
set
to
DLY'D
SWP
TRIGGERABLE,
Q954
is
forward
biased
until
the
Delay
Gateisgenerated.This
pullsupthe
holdoff
linetoprevent
the
Sweep
Start
TD
from
switching to
its
high
state
with
a
trigger
signal
until after
the
Delay
Gateisgenerated
.
Composite
Swp
Out
Q962,
Q966
and
Q968
form
the
Composite
Swp
Out
circuit
.
When
the
DISPLAY
MODE
switch
is
settoMAIN
SWP
or
INTEN,
Q966isforward
biased,
coupling
the
main
sweep
sawtooth
to the
base of
0968.Q968
is
an
emitter-follower
stage
which
couples
the
signaltooutput
terminalsA3andB3.
If
DLY'D
SWP
or
MIXED
is
selectedbythe
DISPLAY
MODE
switch,
Q962
if
forward
biased
and
couples
the
delayed
sweep
or
mixed
sweep
sawtooth
to
the
base of
Q968
.
Composite
Swp
Gate
Out
Aux
Z
Axis
Control
AuxZAxis
Out
Circuit
Description-71353N
0966
and
Q968
or
Q962
and
Q968
(depending
on
DISPLAY
MODE
setting)
are
connected
as
an operational
amplifier,
providing a high
degree
of
gain
stability
.
The
Composite
Sweep
Gate
Out
circuit
includes
Q922,
Q924
and
Q928.The
output
at
the
collector
of
Q928
connects
to
interface
connector
pin Al
for useinthe
indicator
oscilloscope.In
the
AMPL
position
of the
TIME/DIV
OR
DL'Y
TIME
switch,
connector
A1isset
to
approximately
+4.3volts
(via
CR100)
to
unblank
the
CRT
.
Q928
serves
as
the
output
stage
.
With
the
DISPLAY
MODE
switch
settoeither
MAIN
SWP
or
INTEN,
Q922
couples
the
main
sweep
gate to
the base of
Q928
.
When
either
DLY'D
SWP
or
MIXED
is
selected,
Q924ison
.
The
gate
signalatthe
emitter of
Q906
(Dly'd
Swp
Start
Control)iscoupled
to the
baseofQ928
.
The
AuxZAxis
Control
circuit
includes
Q984
and
0988
.
This
circuit
uses the
indicator
oscilloscope
mode
and
switching
levels
to
determine
when
the
sweep
signal
from
the
7B53N
is
being
displayedonthe
CRT
.
Information
of
this
type
is
normally
used
only
when
operating
the
7B53N
in
a
four
plug-in
indicator
oscilloscope
.
Typical
levelstocause
the
Aux
Z
Axis
Control
to
intensify
the
CRT
are+5voltsatterminal
A16
and-0.6
volt
at terminal
B7.This
forward
biases
Q988,
resultingina
positive
levelatits
emitter
.
When
the
7B53N
is
used
in
a
three
plug-in
indicator
oscilloscope
and the
DISPLAY
MODE
is
set
to
INTEN,
0984
is
off
and
0988
is
forward
biased
.
Q992
is
the
Aux
Z-Axis
Out
stage
.
The
output
at
connector
DZ
is
connected
to pin
A17
on
the
interface
connector
and
then
to
the
Z-axis
circuit
in
the
indicator
oscilloscope
.Areductionincurrent
through
Q992
causes
the
CRT
trace
to brighten
.
For
this
description,
assume
that
the
7B53N
is
used
in a
three
plug-in
indicator
oscilloscope
.
As
described
under
AuxZAxis
Control,
when
INTEN
is
selected
by
the
DISPLAY
MODE
switch,
Q988
is
turned
on
.
The
positive
levelatthe
emitter of
Q988
reverse
biases
CR991,
which
reduces
conduction
of
Q992.The
positive
gate
appearingatthe
emitterof0906
(Dly'd
Swp
Start
5
Circuit
Description-7B53N
Control)
during the
delayed
sweep
further
reduces current
through
Q992,
causing
the
CRT
tracetointensify
beyond
the
normal
levelofunblanking
.
In
all
other
selections
of
the
DISPLAY
MODE
switch,
Q984isforward
biased
through
CR66.This
turns
off
Q988,
which
diverts
current
through
CR991.Q992
isinsaturation
and the
CRT
trace
brightnessisnow
setbythe
unblanking
signal
(Sweep
Gate)
.
HORIZONTAL PREAMP
The
Horizontal
Preamp
selects
the
sourceofthe
output
signal
(main
or
delayed sweep)
and
supplies
an
amplified
sawtooth
signaltothe
horizontal
circuits in
the
indicator
oscilloscope.In
addition,
this
circuit
contains
the
horizontal
magnifier
circuit
and
the
horizontal
positioning
network
.
Fig.3-9
showsadetailed
block diagram
of
the
Horizontal
Preamp
and
the
schematic
is
shown
on diagram
8atthe
rear of
the
manual
.
Ext
Horiz
Amp
The
Ext
Horiz
Amp
consistsofQ1004
and
Q1006
connectedasan
operational
amplifier.When
the
TIME/DIV
(DL'Y)
switch
is
in
any
setting
except
AMPL,
the
+5
volts
coupled through
CR
113,CR1016,
andR1016tothe
base
of
Q1006
holds
this
transistorinsaturation.Therefore,
any
incoming
external
horizontal
signalisby-passed
to
ground
(by
way
of
-15
volt
supply)
through
Q1006
and
U1020D
.
When
the
TIME/D
IV
(DL'Y)
switch
is
settoAMPL,
the
+5
voltsisremoved
from
the
base of
Q1006
allowing
Q1004
and
Q1006
to
function
as
an
operational
amplifier.The
output
signal,
inverted
from
the
inputatconnectorA,is
coupled through
R1007
.
Horiz
Display
Selector
Q1024
and
U1020A,
B,
C,D,andEcomprise
the
Horiz
Display
Selector
circuitry.Depending
upon
the
setting
of
the
DISPLAY
MODE
switchorthe
TIME/DIV
(DL'Y)
switch,
this
circuit
determines
which
signaliscoupled
to
the
Horiz
Out
Amp
.
When
the
TIME/DIV
(DL'Y)
switch
is
settoAMP,
U1020C
is
forward
biased
and
couples
the
signal
from
the
Ext
Horiz
Amp
to
the
Horiz
Out
Amp
.
Simultaneously,
+5
volts
is
disconnected
from
the
DISPLAY
MODE
switch,
assuring
thatnointernally
generated
sweep
signaliscoupled
through
at
this
time
.
In
all
other
positionsofthe
External Horizontal from Main
Trigger
Coupling
R8
POSITION
-15
V
DISPLAY
MODE
DS8
X10
MAG
X10
MAG
3-16
Fig.3-9.Horizontal
Preamp
detailed
block
diagram
.
TIME/DIV
(DL'Y)
switch,+5volts
is
connected
to
the
DISPLAY
MODE
switch
.
When
MAIN
SWPorINTEN
is
selectedbythe
DISPLAY
MODE
switch,
+5
voltsisapplied
to
the
anodeofCR
111
or
CR110
.
This forward
biases
U1020A,
which
couples
the
main
sweep
sawtooth
to
the
Horiz
Out
Amp
.
Q1024isalso
forward
biased so that
any
signal
developed
by
the
Delayed
Sweep
Generator
is
by-passedtoground
(via
the
-15
volt
supply)
.
Any
output
from
the
Ext Horiz
Ampiscoupled
to
ground
through
U1020D
.
If
the
DISPLAY
MODE
switch
is
settoDLY'D
SWP
or
MIXED,
+5
voltsisappliedtothe
anodeofCR108
or
CR107
.
This
forward
biases
U1020B,
which
couples
the
delayed
sweeporthe
mixed sweep
signaltothe
Horiz
Out
Amp.U1020E
is
also
forward
biased,
coupling
the
main
sweep
signal
to
ground
.
Position
Amp
The
POSITION
controlR8sets
the
bias
on
Q1014,
thus
setting
the
DC
current
coupledtothe
Horiz
Out
Amp
.
Horiz
Out
Amp
Circuit
Description-7853N
The
Horiz
Out
Amp
includes
Q1038,
Q1046,
and
U
1034A,
B, C,
andD.U1034B
andU1034C
are
connected
asanoperational
amplifer,
with
Rfbeing
R1052
andRi
the
the
Swp
Cal
adjust,
R60
.
U1034C
and
U1034D
formaparaphase
amplifier.This
stage
converts
the
single-ended
input
signaltoa
push-pull
output
signal
whichisnecessarytodrive
the
horizontal
output
stage
in
the
indicator oscilloscope
.
This
stage
also
provides
the
X10
magnification
and
Mag
Gain
adjustment.When
the
X10
MAG
switchisactivated,
R1045
and
R1055
are
connectedinparallel
with
R1046
and
R1056,
decreasing
the
emitter
degeneration
of
the
stage
.
This
increases
gain of
the
stage
10 times
.
The Mag
Gain
adjustisset
to
provideacalibrated
gain
when
magnified.A
contactofK1055
compretes
the
circuit
for
the
X10
MAG
indicator
lamp
when
the
X10
MAG
switch
is
activated
.
Q1038
and
U1034A
set
the operating
bias
for
the
output
Stage
.
Q1046
servesasa
constant-current
source for
U
1034C
andU1034D
.
NOTES
Loading...