LIST OF ILLUSTRATIONS
LIST OF SABLES
OPERATORS SAFETY SUMMARY
SERVICE SAFETY SUMMARY
Section
Section
7Be remaining portion of this Table of Contents lists the
servicing instructions
use by qualified personnel only
do not perform any
operating instructions unless qualified to do so
Section
1
2
3
...........................
SPECIFICATION
OPERATING INSTRUCTIONS
Controls and Connectors
Operating Checkout
General Operating Instructions
Applications 2-1 3
Test Equipment Required
PART 1 PERFORMANCE CHECK
Trigger System Check
Horizontal System Check
PART 2 ADJUSTMENT
Trigger System Adjustment
Horizontal System Adjustment
Partial A6--Sweep circuit board
component locator (SN 8090000 and up).
Partial
component locator (SN 3089999 and
below).
on1 y) component locator.
A8---Readout circuit board
component locator.
A6-Sweep circuit board
Partial
Al--Interface circuit board (7653A
A1 --Interface circuit board
Partial
(7653AN only) component locator.
(7853A only)
Partial
component locator (SN 8090000 and up).
A6--Sweep circuit board
LIST
Table
No.
OF
8-21 Partial A6---Sweep circuit board
component locator (SN 8089999 and
below).
8-22
8-23
8-24
8-25
Partial AG.--Sweep circuit board
component locator
Location of adjustments in the
7653A/7853AN (SN 6209999 and below).
Location of adjustments in the
(SN 8210000 and up).
Location of adjustments in the
7853A/7 B53AN.
(SN
8090000 and up).
TABLES
Page
7853A
Electrical
Environmental
Physical
Deflection Factor of the X Signal
Risetime Measurements
Test Equipment 4-3
Main Sweep Timing
Delayed Sweep Timing
Main Sweep Magnifier Accuracy
Delayed Sweep Magnifier Accuracy
Relative Susceptibility to Static Discharge
Damage
.......................
...................
.......................
......
............
..................
...............
.............
......
....
........................
1-1
1-6
1-6
2-13
2-16
4-13
4-14
4-15
4-15
5-2
@?
JUN
1980
OPERATORS
7B53A
The general safety information in this part of the summary is for both operating and servicing personnel. Specific
warnlngs and cautions will be found throughout the manual where they apply, but may not appear in this summary.
IN THlS MANUAL
CAUTION statements identify conditions or practices that could result in damage to the equipment or other property.
WARNING statements identify conditions or practices that could result in personal injury or loss of life
AS MARKED ON EQUIPMENT
CAUTION indicates a personal injury hazard not immediately accessible as one reads the marking, or a hazard to
property including the equipment itself.
DANGER indicates a personal injury hazard immediately accessible as one reads the marking.
SAFETY
SUMMARY
SYMBOLS
IN THlS MANUAL
@
Static-Sensitive Devices.
This symbol indicates where applicable cautionary or other information is to be found
AS MARKED ON EQUIPMENT
DANGER-H~gh voltage.
@
Protective ground (earth) terminal.
A
ATTENTION----refer to manual.
WARNINGS
POWER SOURCE
This product is intended to operate in a mainframe connected to a power source that will not apply more than
rms between the supply conductors or between either supply conductor and ground.
way of the
GROUNDING THE PRODUCT
'This product is grounded throirgh the grounding conductor of the mainframe power cord. To avoid electrical shock, plug
the
mainframe power cord into a properly w~red receptacle before connecting to the product input or output term~nals. A
protective ground connection by way of the grounding conductor in the mainframe power cord is essential for safe
operation.
ground~ng conductor in the mainframe power cord is essential for safe operation.
A
protective ground connection by
250
volts
REV
JAN
1985
DANGER ARISING FROM
7B53A
Upon loss of the protectwe-ground connection, all access~bie condijctrve parts (~ncludrng knobs and controls that may
appear to be
rnsulat~ng), can render an electr~c shock
LOSS
OF
GROUND
DO NOT OPERATE IN EXPLOSIVE ATMOSPHERES
To avoid explosion, do not operate this product in an atmosphere of explosive gases unless it has been specifically
certified for such operation.
DO NOT OPERATE PLUG-IN UNIT
To avoid personal injury, do not operate this product without covers or panels installed.
unlt via a plug-in extender.
WITHOUT
COVERS
Do
not apply power to the plug-~n
@
JUN
1980
SERVICING
7B53A
SAFETY
SUMMARY
FOR
QUALIFIED
SERVICE
PERSONNEL
ONLY
Refer also to the preceding Operators Safety Summary
DO NOT SERVICE ALONE
Do not perform internal service or adjustment of this product unless another person capable of rendering first aid and
resuscitation is present.
USE CARE WHEN SERVICING WITH POWER ON
Dangerous voltages exist at several points in this product. To avoid personal injury, do not touch exposed connections
and components while power is on.
Disconnect power before
POWER SOURCE
This product is intended to operate in a mainframe connected to a power source that will not apply more than
rms between the supply conductors or between either supply conductor and ground.
way of the grounding conductor in the mainframe power cord is essential for safe operation.
removlng protective panels, soldering, or replacing components.
A
protective ground
250
connection
volts
by
@
JUN
1980
7B53A
The
7B53A
7B53A
alphanumeric readout system provided for the 7000-series oscilloscopes.
and
7B53AN
are electrically identical except that only the
785314
Section 1-.--7B53A/7B53AN
is compatible with the
This instrument will meet the electrical characteristics listed under Performance Requirement in Table
The
complete calibration.
except as otherwise indicated. Warmup time for given accuracy is 20 minutes.
-
-.--.
MAIN SWEEP
Sweep Rates 0.05 ys/div to 5 s/div in 25 calibrated
Sweep Accuracy
-
."
Over Center Eight Divisions
-
- - -
Variable Sweep Rates
Characteristics Performance Requirement Supplemental information
-
.
50 ms/div to 0.5 ysldiv
.-~
5 s/div to 0.1 s/div and
0.2
ys/div to 0.05 psldiv
Over any 2 div portion within
center eight div (all sweep
rates)
-
- - -
-
-
--
following electrical characteristics apply over an ambient temperature range of 0°C to +-50°C,
Table 1-1
to
-C35OC
[
:in
". .
1
3.5%
--
5%
---
- - -
ELECTRICAL
".
.
.
--
--
.
O°C to +50°C
------"
1
w;;n
Within
-
4%
T-".-'-."
-
..
.,
..
1..
I
1
w!;i
.
~
I
Within
5%
Within
7%
)variable to at least 2.5:l. VARIABLE
".'."
-
delayed sweep rates, and variable
main sweep
holdoff.
~
-,--
r.-,.-.---.--.-.w-
"
~
-
--
-
-
--
-
J
..4.--"
-A=-"
steps.
Measured in 7000-Series Oscilloscopes.
-115OC
I
Unmag Unmag Mag
--.m----....-..-."
1
w;itn
..
.".
~
Within W~th~n
3%
---
-
With~n
-
.
- - -
-
/
Continuously variable between cali-
/
brated sweep rates. Extends sweep rate /control internally switchable between
I
to at least 12.5 s/div. variable main sweep rates, variable
--
1-1,
^-"
.
following
Sweep Hold-Off Hold-off time may be varied at least
2:l
by front-panel VARIABLE control
when Variable Selector connector
(P140) is connected for variable
holdoff.
REV
D
JUN
1980
Table
7B53A
1-1
(cont)
..
DELAYED
Sweep Rates
Over Center
Variable Sweep Rate
Characteristics Performance Requirement Supplemental l nformation
.
SWEEP
50
ms/div to 0.5 psldiv
0.5 s/div to 0.1 sldiv and
Over any 2 div portion within
center eight div (all sweep
rates)
"
8
Divisions
__
-l.__."
_
-.,---
"^
-l_-....l...lI
.l-ll___.
.05
ps/div to .5 s/div in 22 calibrated
easured in 7000-Series Oscilloscopes.
Continuously variable between
calibrated sweep rates.
_l__-..--------"
"."
---.----,-.
1
-l_l._l_.l
division of the unmagnified sweep when
5%
--
-.
-~
....
"
measuring
Extends sweep rate to at least
1.25
2.51. VARIABLE control switchable
between variable main sweep rates,
variable delayed sweep rates, and
variable main sweep
-.--_
_-_
magnified
sldiv. Variable to at least
or
unmagnified
holdoff.
"--
MIXED
PIME
Mixed Sweep Accuracy
Variable Time Delay
SWEEP, VARIABLE
DELAY
. .- .
- - - - -
Main Sweep
Delayed Sweep
Delay Time Range
Differential Delay Time
Measurement Accuracy
(-+-15OC
Delay Time Jitter at
1
to +35'C)
5 s/div to 1 s/div
msfdiv
- -
-
-
I
Within
0 to 10 times DLY TIME/DIV
I
Within
I
1
the TlMVDlV setting.
2%
plus main sweep error.
settings from 5 s/div to 1 ps/div.
Within 1.4% of measurement plus
0.3% of full scale.
0.7%
0.3% of full scale.
Less than 1 part in 20,000 of 1 OX
...
"
of measurement plus
.....-..l._-.
""
_..__..._-,._l."__
--
I
Exclude the following portions of
the mixed sweep:
First 0.5 div after start of main
sweep display and 0.2 div or 0.1
(whichever is greater) after transition
of main to delayed sweep.
Full scale is 10 times the TlMWDlV
OR DLY TlMEsetting. Accuracy applies
over the center eight major DELAY TIME
MULT dial divisions.
""
I
I
I
.
. . .
.- . -.
-
ps
--
----
-
-
"
-
1-2
REV
A
JUN
1980
-"
7B53A
~
,
AMPLIFIER
Deflection Factor
EXT, MAG
EXT, MAG Xl
.
.
EXT
Nominal Frequency Response
Characteristics Performance Requirement Supplemental Information
.. - .
.-
.-i
10, MAG XI
AC
AC
LF REJ
Xi0
"..
......-----A-
""
"
"".
""
"*.."
i
1
10 mV/div within 10%.
100
"~.
mVfdiv within 10%.
1
v/div within 10%.
"
System
Oscilloscope.
.--3
Table
1-1
(cont)
~."
dB
points in 7000-Series
".
.
.
..
...
".
-.
"-
--
"
~
.-..
".
.-
."
-.
.-
.
--
-
.....
"
,.-"
"""
"."
.
-----
""
-
..
--
..
AC HF REJ 100 kHz
DC
MAIN TRIGGERING
Trigger Sensitivity
COUPLING
AC LF REJ 30 kHz to
"
~re~uentq Range
"
Hz
""
DC
Triggering
30
Hz
to
10 MHz
10 MHz
150 kHz
10 MHz
Minimum Trigger
I
.-
"-
/
I
10 MHz
to 100 MHz
AC HF REJ 30 Hz
1
50 kHz
to
1
1
~
""
2
MHz
Signal Required
--
---
0.3
1.5
0.3 100
EXT
1
100
/
500
---
I
1
Triggering signal
~ncreased 10 times for
ments
EXT
+
10 ooeration.
1
1
amplitude
requtre-
DC
-
-"
REV
JAN
---
1385
dc to
10 MHz
-
-
--
-
-
-
--
--
---
10 MHz to
100 MHz
--
0 3
-"-
-
--
1.5
d
-
--
100
,-
--
--
500
--
-
Table
7B53A
1-1
(cont)
---------".-
External Trigger Input
lnput
Maximum Safe Input Voltage
Level Range
Internal Trigger Jitter
DELAYED TRIGGERING
Trigger Sensitivity
Coupling
Characteristics Performance Requirement Supplemental Information
-
R and C
EXT
EXT
-,"---.----
-:-
10 At least + and 15 V.
--,.-,"
I
I
least + and - 1.5 V.
1 ns or less at 75 MHz.
Triggering
Frequency Range
--
30 Hz to
-
10 MHz
.
.-
10 MHz to
100 MHz
---
--.
Minimum Trigger
Signal Required
.-
- - -
-
Approximately 1 Mf2 paralleled by
20
pF.
500 V (dc
to-peak ac at 1 kHz or less.
-1
Peak ac). 500 V peak-
."..
-
-.
.
I
---.
-.
.-
----
. .
-
External 'Trigger lnput
Maximum Safe lnput
Voltage
SOURCE set to EXT)
--
Input R and C
Level Ranae
Internal Trigger Jitter
(DLY'D
TRIG
- - - - --
dc to
10 MHz
I
At least + and - 1.5 V.
-
1 ns or less at
1.
--_---p-....."
75
MHz.
500 V (dc
to-peak ac at 1 kHz or less.
11 Mf2 paralleled by 20 pF.
-1-
peak ac). 500 V peak-
I
1
l..lll...lll..-..-.----ll
--
---
..
-
REV A JUN
1980
OUTPUT
7B53A
Table
1-1
(cont)
Characteristics Performance Requirement Supplemental Information
SIGNALS
Delayed Sweep Gate
Maximum Safe Input
Voltage
--.".-..-
Amplitude
Output Resistance
Loading
Polarity
Duration
DISPLAY MODE
INTEN, DLY'D SWP
----.-----.---.--
MIXED
..".-,"---".~-
I
--
-....""~~.v",~-~~~-
Rectangular pulse.
23.0
V with baseline at
to
--I
V when loaded by at least
20.475
--0.2
Positive-goi ng.
ll--l~_l_____----.---~-~-.~-"--~--
For the time that the delayed sweep runs.
-----.--.-
--
Composite gate signal with timing de-
termined by the setting of the
DIV OR DLY TIME switch during the
main sweep portion of the display,
and by the setting of the DLY'D
Division switch during the delayed
sweep portion of the display.
with baseline at 0 to
V
when loaded by
--,~-...-----.---"
.
---0.2
50
0.
10
kfl.
----.-----
TIMV
lime/
rtlO
V
(dc i Peak ac).
to-peak ac at
Available at front-panel DLY'D
TRIG IN connector when
operating in the
SWP, or MIXED Display Modes.
The DLY'D
must be set to
be connected for Delayed Sweep
Gate Out.
Approximately
At least
or less.
-,--,."-----.....-....----
1
kHz
or less.
INTEN, DLY'D
TRIG SOURCE switch
IN-i and P613 must
350
---
10
kfl shunted
n.
-
t
I
20
V peak-
by
----
100
--
pF
Sweep Gate
Duration
DISPLAY MODE
MAIN SWEEP, INTEN
DLY'D SWP
MIXED
l_l-~l-..l"
Main Sweep Gate
Duration
@
JUN
1980
----.-.-._lll.-..---
Refer to associated oscilloscope manual.
I
1
Coincident with the main sweep interval.
""-
Coincident with the main sweep interval.
"_
-l.ll_,ll)....
Coincident with the main sweep interval
plus the delayed sweep interval.
.--_____-lll.-.,l_-l
Refer to associated oscilloscope manual.
i
Coincident with the main sweep (all
Display Modes).
""
-
__III^.._,.l
_
I
1
^..
"..
".-
"--.
"
-.-----.l"....X-..-.
X.
-
---
-----
--
.-
---
--
I
Table
7B53A
1-1
(cont)
"
Sawtooth
Waveshape
-
.-
Duration
a
For Internal Triggering only, the specified
when the number in the table is greater than the
Characteristics Performance Requirement Supplemental Information
-----
.-.....-~wp-.--"
"--
Refer to associated oscilloscope manual.
"
1
I
DISPLAY MODE
MAIN SWP, INTEN
DLY'D SWP
MIXED
I I
Sawtooth signal with slope determined
by setting of the
TIME switch.
Sawtooth signal with slope determined
by the setting of the DLY'D
Division switch.
Composite sawtooth signal with slope
determined by the setting of the
TlMWDlV OR DLY TIME switch during
the main sweep portion of display, and
by
Division switch during the delayed
sweep portion of display.
Coincident with the time that each
sweep is displayed.
-
I
TlMWDlV
the setting of the DLY'D Time/
--3
dB frequency of the vertical system replaces any frequencies in the above table
--3
dB frequency of the vertical system.
OR
---.-
Time/
DLY
--
--
__.."l-___-_.l_l_._
Characteristics
Size
Weight
Table
1-2
ENVIRONMENTAL
Refer to the Specification for the associated oscilloscope.
Table
1-3
Ibs
(1.5
PHYSICAL
kg).
-.."
Fits all 7000-series plug-in compartments.
3.3
@
JUN
1980
Section
7B53A
2-7B53N7B53AN
OPERAT
7B53N7B53AN
The 7B53A and 7B53AN Dual Time Base units provide
Main, Intensified, Delayed, and Mixed sweep operation for
'EKTRRONIX 7000-Series Oscilloscopes. Calibrated
sweep rates from 5
magnification) and triggering to 100 MHz are provided.
7B53A and 7B53AN are electrically identical except
The
that only the
readout system provided for 7000-Series Oscilloscopes.
Other features include
delay,
variable main and delayed sweeprates, and variable
main sweep holdoff. Separate triggering controls are
provided for main and delayed sweep triggering, and
when operating in the AUTO MAIN TRIGGERING MODE,
a bright base line is displayed in the absence of a trigger
signal. The
amplifier for
Features
s/div to 50 ns/div
7B53A is compatible with the alphanumeric
0 to 10 times continuous sweep
7B53N7B53AN can also be used as an
X-Y
operation.
NG
(5
ns with XI0
INSTRUCTIONS
To install the
ment, push it in until it fits firmly into the
The front panel of the 7653N7653AN should beftush with
the front panel of the indicator oscilloscope. Even though
the gain of the indicator oscilloscope is standardized to
minimize adjustment when inserting plug-in units, the
sweep calibration of the
ed when installed. The procedure for checking the unit is
given under Sweep Calibration in
procedure in this section.
To remove the
(see ~i~,
oscilloscope and pull it out of the plug-in compartment.
2-11
7B53N7B53AN in a plug-in compart-
compaitment.
7B53A17B53AN should becheck-
theoperating Checkout
7853N7B53AN, pull the release latch
to disengage the unit
from
the indicator
General
The 7B53A17B53AN operates with a TEKTHONIX 7000Series Oscilloscope and a 7A-Series Amplifier unit to form
a complete oscilloscope system. To
7B53A/7B53AN, its operation and capabilities should be
known. 'This section explains the operation of the
panel controls and connectors, provides an Operating
Checkout procedure, gives general operating information, and describes basic applications for this
effective1 y use the
front-
instrurnent.
Installation
The 7853A17853AN is designed to operate in the
horizontal compartment of the indicator oscilloscope.
Thisinstrument can also beinstalled
compartment to provideasweepthat runsvertically on the
crt. However, when used in this manner, there are no
retrace blanking or internal triggering provisions, and the
unit may not meet the specifications given in Section
The instructions in this manual are written for use of the
7B53N7B53AN in the horizontal plug-in compartment,
Before proceeding with installation it is necessary to
check the internal connections of the Variable Selector
connector and the Delayed Sweep Gate Out connector.
These connections should be checked against those
4-1
shown in Fig.
by qualified service personnel.
inthevertical plug-in
1.
Fig.
2-1.
Location
of
release latch.
CONTROLS AND CONNECTORS
General
All controls required for the operation of the
7B53N?B53AN, except the Variable Selector and the
Dly'd Sweep Gate Out connector are located on the front
panel of the instrument.
capabilities of this instrument, the operator should be
familiar with the function and use of each control. A brief
description of the front-panel controls and connectors is
given here. More detailed information is given under
'To make full use of the
REV 6 JUN
1980
2-
1
Operating
7B53A
Instructions.---9853A17B53AN
General Operating Information. Fig.
2-2
shows the front
panel and external controls and connectors of the
7
B53A/7 B53AN.
1.
Main Triggering Controls
LEVEL. Selects the amplitude point on the trigger
signal where sweep triggering occurs.
SLOPE. Two-position switch permits triggering on the
positive-going or negative-going portion of the main
triggering signal.
TRIG'D. Light indicates that the main sweep is
triggered and will produce a display.
MODE. Pushbutton switches select the operating mode
for the main triggering circuits.
COUPLING. Pushbutton switches select the method of
coupling the triggering signal to the main triggering
circuits.
SOURCE. Pushbutton switches select thesource of the
main triggering signal.
2. Sweep
Controls
TIMEIDIV OR DLY TIME. Selects the sweep rate of the
main sweep generator (see Fig.
2-5).
DLY'D Time/Division. Selects the sweep rate of the
MAIN
delayed sweep generator and selects the
INT'EN, and
or variable main sweep holdoff; depending on the setting
of the Variable Selector multi-pin connector
VARIABLE control also selects the MIXED Display Mode
(see Fig. 2-5).
SWP CAL. Screwdriver adjustment to match the gain of
7653A/7B53AN to the indicator oscilloscope for
the
calibrated sweep rates.
POSITION. Controls horizontal position of display.
FINE. Provides precise control of horizontal position
adjustment,
MAG. Pushbutton switch selects
magnification.
3.
Delay Time Control
DELAY l IME MULT. Provides variable sweep delay
between 0.00 and 10.0 times the delay time indicated by
TIMUDIV OR DLY rlME switch.
the
4.
Delayed Triggering Controls
XI or XI0 horizontal
(P140). The
OPERATING
CHECKOUT
Introduction
The following procedure checks the basic operation of
the
7B53A/7653AN. It may also be usedfor familiarization
with this instrument or as an incoming inspection. This
procedure is divided into two parts, Sweep Control
Functions, and Main and Delayed Triggering Functions. A
complete operating check of the
functions can be made by performing both parts, or each
part may be performed separately.
NOTE
For optimum performance, the
should
similar frequency and sweep rate capabilities.
be
instalied
in
an oscilfoscope system
7853N7853AN control
7853Ai7B53AN
with
Setup Procedure
1.
Install the 7B53A/7B53AN in the right horizontal
compartment of the indicator oscilloscope.
2.
Install a Vertical Amplifier unit in the left vertical
compartment.
LEVEL. Selects the RUNS AFTER DLY TIME or
Triggerable After Dly Time Modes, and the amplitude
point at which the delayed sweep is triggered.
SLOPE. Two-position switch permits triggering on the
positivegoing or negative-going portion of the delayed
triggering signal.
COUPLING. Two-position switch selects the method of
coupling the triggering signal to the delayed triggering
circuits.
SOURCE. Two-position switch to select the source of
the delayed triggering signal.
5.
InputlOutput Connectors
MAlN TRIG IN OR AMP IN. Front-panel bnc connector
serving as an external trigger input for the maintriggering
circuit or as an external horizontal input, depending upon
the setting of the
MAlN TRIGGERING SOURCE switch.
TRlG IN. Front-panel bnc connector serving as
DLY'D
an external trigger input forthedelayed triggering circuits
or as a delayed sweep gate output; depending upon the
DLY'D TRlG SOURCE switch, the position of
the DISPLAY MODE.
TIMVDIV OR DLY TlMEswitch and the
P613, and
3.
Turn on the indicator oscilioscopeand allow at least
20 minutes
4.
MAlN TRIGGERING
DLY'D TRlG
Sweep Controls
5. Set the indicator oscilloscope to display the plug-in
units and adjust for
warmup.
Set the 7B53N7653AN controls as follows:
SLOPE
MODE AUTO
COUPLING AC
SOURCE INT
I.
EVEL RUNS AFTER DLY
SLOPE
COUPLING AC
SOURCE
POSITION Midrange
MAG
TlMElDlV OR DLY FlME 1 ms
DLY'D TrmefDivision
VARIABLE CAL (fully clockwise)
Variable Selector Main Variable
DELAY[-IMEMULT 1.00
a
well defined display. See indicator
(-f
)
TIME (fully clockwise)
(f)
INT
X1-IN
1
ms
REV
C
JUN
1980
2-3
Operating
7B53A
oscilloscope and vertical unit instruction manuals for
detailed operating instructions.
Sweep
'The following procedure checks the operation of the
sweep controls and checks the Display Modes.
Normal Sweep
1.
Instructions~--7B53A/7B53AN
Control
Perform steps 1 through 5 of the Setup Procedure.
Functions
6.
Connect a 1 ms time-mark signal from the TimeMark Generator with a
connectors and 50
associated vertical unit. Adjust the vertical
switch for about four divisions of display and rotate the
MAIN TRIGGERING
Check the crt display for one complete time-mark per
division (position as necessary). If necessary, adjust the
SWP CAL screwdriver adjustment for one complete
mark per division over the center eight divisions of display.
Disconnect the "Time-Mark Generator.
50
Z;1
bnc coaxial cable with bnc
62
bnc termination to the Input of the
VoltdDiv
1-EVEL controt for a stable display.
time
NOTE
The TimeIDivision switch selects main sweep rates,
delayed sweep rates, and Display Modes (MAIN
SWP, INTEN. DL Y'D
Seiectrng Sweep Rates and Display Modes dis-
in
cussions
further information.
2.
To select the MAlN SWP Display Mode, press in the
Time/Division knob, and set the TIMVDIV OR DLY
DLY'D
TIME switch and the DLY'D TimelDivision switch to the
same sweep rate
3.
Connect the 1 kHz calibrator signal from the in-
dicator oscilloscope to the vertical amplifier unit Input.
Adjust the Calibrator and thevertical
for four divisions of display.
4.
Rotatethe MAlN TRIGGERING LEVEL control for a
stable MAIN
DELAY TIME MULT dial and note that it has no effect on
the display.
the General Operating
(1
SWP
SWP,
and MIXEL?). Refer lo
instructions
ms).
Volts/Division switch
display (non-delayed). Rotate the
for
Intensified, Delayed, and Mixed Sweep
7.
Reconnect the osciIloscope Calibrator signal to the
vertical unit and adjust for about four divisions of vertical
display. Pull out the DLY'D
clockwise to
(TIMVDIV OR DLY TlME switch remains at 1 msldiv).
Note that a delaying sweep with an intensified portion
(delayed sweep) is displayed on the crt (the oscilloscope
Intensity may need to be varied to view the intensified
display).
8.
Rotatethe DELAY'T'IMEMULTdial andnotethatthe
amount of delay
display is controlled by the DELAY TlME MULT dial.
9.
Rotate the VARIABLE control counterclockwise out
of switch detent and note that the sweep rate indicated by
TIME/DIV OR DLY TIME switch can be varied to at
the
least the sweep rate
(2
msl'divj. 'The internal Variable Selector connector must
be
set for variable main sweep rates. Return the
VARIABLE control to the GAL position.
.1 msi'div for the INTEN Display Mode
time before the intensified portion of
TimelDivision knob and rotate
of the next adjacent position
5.
Rotate the POSITlON control and note that
horizontally positions the trace, Rotate the FINE control
and note that it provides precise horizontal positioning. position
Disconnect the calibrator signal.
Sweep Calibration
NOTE
For accurate sweep timing, apply a signal of known
frequency or time period (time-mark signal,
60
Hz
calibrator square wave,
associated vertical amplifier unit and adjust the
TIMEIDIV OR DLY nME switch and the
control to calibrate the signal
graticule. Be sure that the sweep timing signa! is
0.5%.
accurate within
tion procedure uses
Generator.
The following sweep caiibra-
a
TEKTRONIX
line etc.) to the
SWP
to
the oscilloscope
Time-Mark
CAF
it
10. Press the DLY'D
magnified display with sweep rate determined by the
DLY'D
may need to be increased to view the delayed sweep
display.
11.
MIXED Display Mode (MIXED Display Mode can be
selected when operating in the DLY'D SWP Display Mode
only). Note that the main sweep is displayed at a rate
determined by the
followed by a delayed sweep display at the rate determined by the DLY'D
DELAY TlME MULT dial
display allocated to each sweep is determined by the
DELAY TlME
2-4
Time/Division switch to the inner
for DLY'D SWP
Time/Division switch. The oscilloscope Intensity
Pull out the VARIABLE "l"ime/Division knob for the
'TIMEIDIV OR DLY TlME switch
Time/Division switch. Rotate the
and note that the amount of
MULT
dial.
REV
C
JUN
the
1980
Main and Delayed Sweep Triggering Functions
7B53A
The following procedure checks the operation of the
main and delayed triggering controls:
Partial Procedure. To begin the Operating Checkout
1
with triggering, perform steps
Procedure provided at the beginning of the Operating
Checkout. Connect the
indicator
osciIioscope to the vertical unit Input and adjust
1
for about four divisions of vertical display.
12.
Set the TIME/DIV OR DLY TlME switch and the
DLY'D
DLY'D
Time/Division switch to 1 ms and press in the
TimeiDivision switch and the Variable control.
CHECK---that a stable display can be obtained with the
COUPLING switch for MAlN
HF REJ, and DC for both the positive and negative
positions of the SLOPE switch (MAIN TRIGGERING
LEVEL control may be adjusted as necessary to obtain a
stable main sweep display). Remove all connections from
the oscilloscope system.
13.
Connect a 60
n
cable or voltage probe to the vertical unit Input and
50
adjust the
VoltsIDiv switch for about four divisions of
Hz
signal (iine voltage, etc.) with a
display. Set the MAlN TRIGGERING COUPLING switch
to AC LF REJ and set the
10
ms. Rotate the MAIN TRIGGERING LEVEL control
to
"r^lMVDlV OR DLY TIME switch
throughout its range and check that a stable display
cannot be obtained (TRIG'D light off).
14.
Change the MAlN 'TRIGGERING SOURCEswitch
to LINE and the COUPLING switch to AC. Check that a
stable main sweep display can be obtained with the MAIN
TRIGGERING COUPLING switch set to AC, AC HF REJ,
and DC (MAlN TRIGGERING LEVEL control may be
adjusted as necessary). Remove all connections from the
oscilloscope system.
15.
Connect the indicator oscilloscope Calibrator
signal to the verticai unit Input connector and
TRIG IN connector. Set the Calibrator for a 0.4 V,
signal and adjust the vertical unit
about four divisions of display. Set
TIME switch and the DLY'D
Set the MAlN TRIGGERING SOURCE switch to EXT and
check that a stable display can be obtained with the
COUPLING switch for MAIN TRIGGERING set to AC, AC
HF REJ, and DC, for both the positive and negative
positions of the SLOPE switch (MAIN TRIGGERING
LEVEL control may be adjusted as necessary for a stable
display).
16.
Change the MAlN TRIGGERING SOURCE switch
-:--
to EXT
1
kHz and adjust the vertical unit VoltsiDiv switch for
10. Set the oscilloscope Calibrator for 4 V at
through 5 of the Setup
kHz calibrator signal from the
'T'RIGGERING set to AC, AC
tothe MAIN
1
kHz
Volts/Div switch for
the'TIMWDIV OR DLY
TimelDivision switch to 1 ms.
about
Operating
4
divisions of display. Check that a stable display
Instructions--~7B53A/7B53AN
can be obtained with the COUPLING switch for MAIN
TRIGGERING set to AC, AC HF REJ, and DC, for
boththe
positive and negative positions of the SLOPE switch
(MAIN TRIGGERING LEVEL control may be adjusted as
necessary to obtain a stable delaying sweep display).
Remove the Calibrator signal from the MAlN TRIG IN
connector.
17.
Set the MAlN "TRIGGERING COUPLING switch to
AC and the SOURCE switch to
INT. Adjust the MAlN
TRIGGERING LEVEL control for a stable display. Set the
MODE switch to NORM and check for a stable display.
Change the
MODEswitch to AUTO and adjust the LEVEL
control for a free-running display. Change the MODE
switch to NORM and check for no display.
18.
Adjust the MAlN TRIGGERING LEVEL control for
a stable display. Change the MAlN TRIGGERING MODE
switch to SINGLE SWP. Press the
check for one sweep. Remove the
RESET button and
1
kHz signal from the
vertical unit and press the RESET button. CHECK for no
1
display and that the READY light is on. Connect the
kHz
signal to the vertical unit and check for one sweep as the
signal is applied.
19.
Set the MAIN TRIGGERING MODE switch to
AUTO. With the
pull out the DLY'D
.2
ms (Intensified Display Mode). Rotate the MAlN
TIMWDIV OR DLY TIME switch
Time/Division switch and rotate to
at
1
ms,
TRIGGERING LEVEL control for a stable display
(oscilloscope Intensity may need to be varied to view the
intensified display). Rotate the
Df LAY TlME MULT dial
and note that the intensified portion of display is con-
tinuously variable.
20.
Rotate the DLY'D TRlG LEVEL control counterclockwise out of switch detent for a triggered delayed
sweep. With the DLY'D
SOURCE switches set to
TRIG SLOPE, COUPLING, and
(-f
),
AC, and INT; rotate the
DLY'D TRIG LEVEL control for astable intensified portion
of display. Rotate the DELAY
TIME MULT dial and note
that the intensified portion of display does not start at the
completion
of the
of the delay time (determined by the settings
TIMEIDIV OR DLY TIME switch and the DELAY
TIME MULT dial), but waits for the next trigger pulse.
21.
Press the DLY'D Time/Division switch in for the
DLY'D SWP Display Mode. Rotate the DLY'DTWIG LEVEL
control for a stable delayed sweep display (oscilfoscope
Intensity may need to be varied). Check that a stable
delayed sweep display can be obtained with the DLY'D
TRIG COUPLING switch set to AC and
-
positions of the SLOPE switch (DLY'D TRIG LEVEL
DCfor both
.I-
and
control may be adjusted as necessary for a stable delayed
sweep display).
REV
A
JUN
1980
2-5
Operating
7B53A
Instructions-.-7B53Ai1B53AN
22.
Change the DLY'D TRIG SOURCE switch to EXT.
Connect the indicator oscilloscope Calibrator signal to
the DLY'D
Calibrator for
Volts/Division switch for about 4 divisions of display.
Check that a stable delayed sweep display can be
obtained with the DLY'D TRlG COUPLING switch set to
AC and DC for both the
switch (DLY'D TRlG LEVEL control may be adjusted as
necessary for a stable delayed sweep display).
TRlG IN connector. Set the oscilloscope
0.4
V at 1 kHz
-t-
and adjust the vertical
and positions of the SLOPE
GENERAL OPERATING INSTRUCTIONS
Pushbutton Switch Logic
The MODE, COUPLING, and SOURCE pushbuttons of
the MAlN TRIGGERING switches are arranged in
sequence that places the most-often used position at the
top of each series of pushbuttons. With this arrangement,
a stable display can usually be obtained by pressing the
top pushbuttons: AUTO, AC, and INT. When an adequate
trigger signal is applied, the unit is triggered as indicated
by the illuminated
of the LEVEL control and SLOPE switch. If the
light is not on, the LEVEL control isat asetting outsidethe
range of the trigger signal applied to this unit from the
vertical unit, the trigger signal is inadequate, or its
frequency is below the lower frequency limit of the AC
COUPLING switch position. If the desired display is not
obtained with these buttons pushed in, other selections
must be made. Refer to the following discussions or the
intruction manuals for the associated oscilloscope and
vertical unit for more information.
Triggered Light
'IRIG'D light, with the correct settings
TRIG'D
switch are correctly set. The TRIG'D light indicates when
the display is triggered.
When the trigger repetition rate is below about
outside the frequency range selected by the COUPLING
switch, or when thetrigger signal is inadequate,
free-runs at the sweep rate indicated by the -TIMVDlV OR
DLY TIME switch (TRIG'D light off). When an adequate
trigger signal is again applied,
ends and a triggered display is presented. The sweep also
free-runs when the LEVEL control is at a setting outside
the amplitude range of the trigger signal.
running display can be useful when it is desired to
measure only the maximum peak-to-peak amplitude of a
signal without observing the waveshape (such as in
bandwidth measurements). When the display is of amuch
a
greater amplitude than can be displayed on the crt, the
sweep will be triggered in all positions of the LEVEL
control and will not free-run.
NORM.
triggered display is presented with the correct setting of
the LEVEL control and SLOPE switch whenever an
adequate trigger signal is applied.
indicates when the display is triggered.
The NORM trigger mode must be used to produce
triggered displays with trigger repetition rates below
about
outside the amplitude range of thetrigger signal,
trigger repetition rate is outside the frequency range
selected by the COUPLING switch, or when the trigger
signal is inadequate,
When the NORM pushbutton is pressed, a
30
Hz.
When the LEVEL control is at a setting
thereis no trace (TRIG'D light isoff).
the free-running
Thistype of free-
"The TRIG'D light
30
Hz,
thesweep
condition
whenthe
The TRIG'D light conveniently indicates the condition
of the triggering circuits. If the MAlN TRIGGERING
controls are correctly set and an adequate trigger signal is
applied, the TRIG'D light is on. Under certain conditions,
the TRIG'D light may be off, indicating that the sweep is
not triggered. The cause could be a misadjusted LEVEL
control, incorrectly set COUPLING or SOURCE switches,
low trigger signal amplitude, or a triggering signal repetition rate outside the acceptable frequency range. This
light can be used as a general indication of correct
triggering. It is particularly useful when setting up the
trigger circuits when a triggering signal is available
without a display on the crt.
Main Trigger Mode
The pushbuttons located under the MODE title select
the mode in which the main sweep is triggered.
AUTO.
triggered display is presented whenever an adequate
trigger signal is applied and the
When the AU'TO pushbutton is pressed, a
LEVELcontrol and SLOPE
2-6
SINGLE
be displayed is not repetitve or varies in amplitude,
waveshape, or repetition rate, a conventional repetitive
type display may produce an unstable presentation. A
stable display can often be obtained under these circumstances by using the SINGLE
The SINGLE SWP MODE is also useful
non-repetitive or unstable displays.
To obtain a single-sweep display of a repetitive signal,
first obtain the best possible display in the NORM MODE.
Then without changing the other MAlN TRIGGERING
switches, press the SINGLE SWEEP pushbutton. When
ready to view thesingle-sweep display, press the
READY pushbutton. A single trace is presented each time
the RESET-READY pushbutton is pressed
repetitive signal remains connected to the system and
MAlN TRIGGERING switches are correctly set); further
sweeps cannot be presented until the RESET-READY
pushbutton is pressed again. If the display signal is a
complex waveform composed of varying amplitude,
successive single-sweep displays may not start at the
SWEEP,
RESET-READY.
SWPfeature of this unit.
When the signal to
to
photograph
RESET-
(as long as the
REV
B
JUN
1980
7B53A
Operating
7B53A
Instructions---7B53A/7B53AN
AC
LF
REJ.
In the AC LF REJ position of the
COUPLING switch, dc is rejected and low-frequency
kHz
trigger signals below about 30
Therefore, the sweep is triggered only by the
are attenuated.
higher-
frequency components of the trigger signal. This position
is particularly useful for providing stable triggering if the
trigger signal contains line-frequency components. Also,
the AC LF REJ position provides the best alternate vertical
displays at fast sweep rates when comparing two or more
unrelated signals.
AC
HF
REJ.
The AC
HF
REJ position of theCOUPLlNG
switch passes all low-frequency signals between about
Hz
and 50 kHz. Dc is rejected and signals outside the
30
above rage are attenuated. When triggeringfrom complex
waveforms, this position is useful to provide a stable
display of the low-frequency components.
DC.
The DC position of the COUPLING switch can be
used to provide stable triggering with low-frequency
signals which would be attenuated in the other modes, or
with low-repetition rate signals. It can also be used to
trigger the sweep when the trigger signal reaches a dc
level selected by the setting of the LEVEL control. When
using internal triggering, the setting of the vertical unit
position control affects the dc triggering point.
Main Triggering Source
'TheMAIN TRIGGERING pushbutton located belowthe
SOURCEtitle select the source of the trigger signal which
is connected to the main triggering circuits.
INT.
In the IN% position of the SOURCE switch, the
trigger signal is derived from the associated vertical unit.
Further selection of the internal trigger signal may be
provided by the associated vertical unit or indicator
oscilloscope; see the instruction manuals for these instruments for information. For most applications, the
position of the SOURCE switch can be used. However,
some applications requirespecial triggering which cannot
be obtained in the
such cases the
INT position of the SOURCEswitch. In
LlNE or EXT positions of the SOURCE
switch must be used.
INT
be time-related to the displayed waveform for a stable
display. An external trigger signal can provide a triggered
display when the internal signal is too
low in amplitude for
correct triggering, or contains signal components on
which it is not desired to trigger. It is also useful when
signal tracing in amplifiers, phase shift networks,
waveshaping circuits, etc.
the circuit undertest can beconnected
Thesignal from a single point in
tothe EXTqI"'RIG IN
connector with a probe or cable. The sweep is then
triggered by the same signal at all times and allows
amplitude, time relationship, or waveshape changes of
signals at various points in the circuit to be examined
without resetting the MAIN TRIGGERING controls.
-i-
10.
EXT
Operation in the EXT t 10 position of the
SOURCE switch is the same as described for EXT except
that the external signal is attenuated
lotimes. Attenuation
of high-amplitude external trigger signals is desirable to
broaden the range of the LEVEL control.
Trigger Slope
The MAIN TRIGGERING SLOPE switch (concentric
with the MAIN TRIGGERING LEVEL control) determines
whether the trigger circuit responds on the positive-going
or negative-going portion of the trigger signal. When the
SLOPE switch is in the
-I-
position, the display startsonthe
positive-going portion of the waveform; in the position,
the display starts with the negative-going portion of the
2-4).
waveform (see Fig.
When several cycles of a signal
appear in a display, the setting of the SLOPE switch is
often unimportant. However, if only a certain portion of a
cycle is to be displayed, correct setting of the SLOPE
switch is important to provideadisplay which starts on the
desired slope of the input signal.
Trigger Level
"The MAlN TRIGGER LEVEL control determines the
voltage level on the trigger signal at which the sweep is
4-
triggered. When the LEVEL control is set in the
the trigger circuit responds at a more positive point on the
trigger signal. When the LEVEL. control is set in the
region, the trigger circuit responds at a more negative
2-4
point on the trigger signal. Fig.
illustrates this effect
with different settings of the SLOPE switch.
region,
-
LINE.
The LlNE ~osition of the SOURCE switch
connects a sample of the power-line voltage from the
indicator oscilloscope to the trigger circuit. Line
trigger-
ing is useful whentheinput signal istime-related (multiple
or submultiple) to the line frequency. It is also useful for
providing a stable display of a line-frequency component
in a complex waveform.
EX$.
An external signal connected to the MAIN TRIG The TIME/DIV OR DLY TIME switch selects calibrated
IN connector can be used to trigger the sweep in the EXT
position of the SOURCE switch.
The external signal must
2-8
Before setting the MAIN TRIGGERING LEVEL, set the
desired
the
clockwise
SLOPE,
LEVEL
MODE, COUPLING,
control
until
the
fully
display
and
counterclockwise
starts
at
the
desired
SOURCE,
and then
point,
Set
rotate
Selecting Sweep Rates
sweep rates for the main sweep generator and the DLY'D
I'imeiDivision switch selects calibrated sweep rates for
REV A JUN
1980
(LEVEL)
7B53A
Operating
Instructions--7B53A/7B53AN
CRT DISPLAY
OBTAINED
SLOPE SWITCH
SET
WITH
TO
POSITIVE(+)
REV
A
car
DISPLAY
OBTAINED
SLOPE SWITCH
SET
JUN
W1TH
TO NEGATIVE
1980
(-)
Fig.
2-4.
Effect of LEVEL control and SLOPE
switch
on CRT display.
Operating
7B53A
Instructions--7BS3A/7B53AN
the delayed sweep generator. The sweep rate of the main
sweep generator is bracketed by the black lines on the
clear plastic flange of the TIMVDIV OR DLY TIME switch
(see Fig. 2-5). The sweep rate of the delayed sweep
generator is indicated by the white line on the DLY'D
Time/Division knob. When the T'IMVDIV OR DLY TlME
switch and the DLY'D Time/Division switch are set to the
same sweep rate, the switches lock together and the
sweep rate of both generators are changed at the same
time. However, when the DLY'D
pulled outward, the clear plastic flange is disengaged and
the
all0ws changing
changing
generator. The DLY'D Time/Division switch also seiects
section for further information.
A VARIABLE control is provided concentric with the
T'IMVDIV OR DLY
switches (see Fig. 2-5). When the VARIABLE control is
rotated clockwise to the CAL position (into switch detent)
the variable function is inoperative and the VARIABLE
knob can be used only to select the MIXED Display Mode
(see Display Mode discussion in this section for more
information). However, when rotated counterclockwise
(out of switch detent), the VARIABLE control is activated
the
Modes.
sweep generator sweep
the
time
determined
See
TlME and the DLY'D Time/Division
Time/Division knob is
rate
is
changed'
sweep
Mode
rate
the
main
discussion
without
sweep
in
this
for variable (uncalibrated)
sweep rates, or main sweep
setting of the Variable Selector multi-pin connector. The
VARIABLE control allows the sweep rate in each
Time/Division switch position (main or delayed) to be
reduced to at least the next adjacent switch position and
holdoff time to be increased
the
main sweep rates, delayed
holdoff; depending upon the
2:l.
"Time Measurement
When making time measurements from the graticule,
the center eight divisions provide the most linear time
measurements (see Fig, 2-6).
timing
TIMUD,,,
area falls between the second and tenth vertical lines,
area
to the second vertical line and adjust the
OR
DLY
Position the start
switch
so
the
end
of
the
of
the
timing
Sweep Magnifier
The sweep magnifier can be used to expand the display
ten times. The center division of the unmagnified display is
the portionvisibleon thecrt in themagnifiedform (see Fig.
2-7).
Equivalent length of the magnified sweep is more
100
than
by adjusting the POSITION control to bring the desired
portion onto the viewing area.
divisions; any
10
division portion can be viewed
Fig.
2-5.
Composite Time/Division switch.
REV 6 JUN
1980
Operating
7B53A
Instructions---7B53A/7B53AN
Fig.
2-6.
Area of graticule used for accurate time measurements.
OUT
position, the equivalent magnified sweep rate can be
determined by dividing the
Forexample,
1
in the
delayed sweep rate as well as the main sweep rate is
increased 10 times.
Display
settings of the
DisIJlav Mode (non-delayed sweep) set the TIMVDIV OR
DLY
sweep rate and press in the DLY'D
Calibrated sweep rates in the MAIN SWP Display Mode are
available from 5
magnification). The VARIABLE control (Variable Selector
connector set to Main Variable) can be used to obtain
uncalibrated sweep rates to 12.5
MAlN SWP Display Mode is controlled by the MAlN
TRIGGER controls.
XI0 position. When the MAG switch is in the XI0
Time/Division setting by 10.
iftheTlMVDlVOR DLY TlMEswitch issetto
ms the equivalent sweep rate is 100 ps. When operating
INTEN, DLY'D SWP, or MIXED Display Modes the
Modes
Four Display Modes can be selected by appropriate
Time/Division switches (see Fig. 2-5).
Main Sweep Operation.
TIME
~~~'DLYD
~ime/~ivisibk switches to the same
s/div to 0.05 ps/div (5 ns/div with X10
To select the MAlN SWP
Time/Division switch.
sldiv. Triggering in the
Fig.
2-7.
Operation of Sweep Magnifier.
TO use the MAG switch, first move the portion of the
display which is to be expanded to the center of the
graticule. Then press and release the MAG switch to the
Intensified SweepDisplay.
Mode, pull out the DLY'D
to a desired sweep rate faster than the TIMVDIV OR DLY
TIME switch setting. The INTEN Display Mode provides
an intensified portion of the main sweep during the time
the delayed sweep is in operation (see Fig. 2-8A). The time
that the delayed sweep runs is determined by the DLY'D
Time/Division switch. The amount of delay time between
the start of the delaying sweep and the intensified portion
is determined by the
the DELAY
sweep portion of display is controlled by the MAlN
"TRIGGERING controls; triggering for the intensified
portion of display is controlled by the DLY'D TRIG
controls.
Delayed Sweep Display.
is selected when the DLY'D
out, rotated in the
and then pushed in, In this mode, the intensified portion of
display, as viewed in the
played on the crt at the sweep rate indicated by the DLY'D
Time/Division switch (see Fig. 2-88}. Calibrated sweep
rates in the DLY'D SWP Display Mode are available from
.5 s/div to 0.05 psldiv. Uncalibrated delayed sweep rates
to 1.25
control (Variable Selector connector set to Dly'd Variable
position). Triggering for the DLY'D SWP display is controlled by
TIME MULT dial, Triggering for the delaying
s/div can be obtained by using the VARIABLE
the DLY'D TRIG controls.
TiMEIDIV OR DLY 'TIMEswitch and
INTEN Modeforthedesiredsweeprate,
To select the INTEN Display
Time/Division knob and rotateit
The DLY'D SWP Display Mode
Time/Division switch is pulled
INTEN Display Mode, is dis-
REV
C
JUN
1980
2-1
1
Operating
7B53A
Instructions--78fi3Al7B53AN
Fig.
set to
2-9.
Typical Mixed Sweep display (DELAY TIME MULT dial
3.55)
Rg.
2-8.
display.
(A) Intensified Sweep display;
Mixed Sweep Operation.
The MIXED Display Mode is
(B)
Delayed Sweep
selected when the DLY'D SWP Display Mode is selected
and the VARIABLE knob is pulled out. In this mode, the
main sweep is displayed on the crt to a point determined
by the DELAY 'TIME MULT dial; the remainder of the
sweep is at a rate determined by the delayed sweep.
Triggering for the main sweep portion of display is
controlled by the MAIN TRIGGERING controls; and
triggering for the delayed sweep portion of display is
TRlG
controlled by the DLY'D
controls. A typical mixed
sweep display is shown in Fig. 2-9.
Delay Time Multiplier
'The DELAY TIME MULT dial (functional in the INTEN,
DLY'D SWP, and MIXED Display Modes) provides Oto 10
times continuous sweep delay. The time that the main
sweep runs before the start of the delayed sweep is
determined by the settings of the
switch and the DELAY TlME MULT dial.
For example, the delay time indicated by the DELAY
"TIME MULT dial in Fig.
3.55
crt divisions of main sweep. Thus 3.55 multiplied by
2-10
TlMVDlVOR DLY TIME
is 3.55; this corresponds to
Fig.
2-10.
DELAY TlME MULT dial. Reading shown: 3.55.
the delaying sweep rate, indicated by the 'TIMVDIV OR
DLY TlME switch, gives the calibrated delay time before
the start of the delayed sweep.
Delayed Sweep Triggering
A LEVEL control and SLOPE, COUPLING, and
SOURCE switches are provided for delay sweep triggering. When the LEVEL control is rotated to the RUNS
AFTER DLY TIME position the delayed sweep starts
immediately after the delay time. The DLY'D
control and the SLOPE, COUPLING, and SOURCE
switches are inoperative. This mode permits the selection
of continuously variable delay times (by varying the
DELAY TlME MULT dial).
TRlG LEVEL
2-12
REV C JUN
1980
Operating
7B53A
Instructions--7B53A/7B53AN
When the DLY'D TRlG LEVEL control is rotated
counterclockwise (out of switch detent), the delayed
sweep is triggerable. The delayed sweep does not start at
completion of the delay time but, waits until a trigger
the
pulse is received by the delayed sweep triggering circuit.
I'he delay time in this mode is dependent not only on the
settings of the delay-time controls, but on the delayed
sweep triggering controls and the occurrence of the
delayed sweep triggering signal. 'The primary purpose of
this mode is to eliminate jitter from the display delayed
sweep waveform. Since the delayed sweep is triggered by
the input waveform, jitter is eliminated from the delayed
sweep display even though it may be inherent in the input
waveform.
When the delayed sweep is triggerable, the LEVEL
control can be rotated to select the amplitude point on the
trigger signal at which the delayed sweep is triggered. The
DLY'D PRIG SLOPE, COUPLING, and SOURCE switches
are activated and their functions are the same for delayed
triggering as functions with the same title are for MAIN
'TRIGGERING (see Main Triggering Level, Slope, Coupling, and Source discussions in this section).
InputIOutput Connectors
Two dual-function bnc connectors are provided on the
instrument front-panel.
MAlN TRIG IN OR AMP IN.
external trigger input for the main triggering circuit when
the MAIN TRIGGERING SOURCE switch is set to EXT or
+
10 and the TIMVDIV OR DLY IIMEswitch is set to
EXT
any position except AMPL. When the
TIME switch is set to AMPL and the MAlN TRIGGERING
SOURCE switch is set to EXTor EXT
serves as an external horizontal input (see X-Y Operation).
'This connector is an
TlMWDlV OR DLY
1.-
10,
this connector
To apply an external signal (X-signal) to the horizontal
system, set the
7B53A/7B53AN TIM WDIV OR DLY TIME
switch to AMPL and the MAlN TRIGGERING SOURCE
EXT.
switch to
The MAlN TRIGGERING COUPLING
selected affects the frequency response of the X-signal.
Apply the external signal to the MAIN TRIG IN OR AMP IN
connector. Deflection factor of the X-signal is provided in
2-1.
multiples of 10 as shown in Table
The Y-signal can be
applied to the plug-in unit installed in the vertical compartment.
Table
2-7
DEFLECTION FACTOR OF
THE
X
SIGNAL
Main Deflection
Triggering
-,.-"--"
.------- "-
EXT 10 mV/div
EX7 700 mV/div
". .-
EXT + 10
.--.-.-.-
,"
----."
.-------.-.
""
--,-
"
1
---.
Vldiv
""
-.----.--
"-----
'To apply the X-signal internally via the triggering
system set the
S'IME/DIV OR DLY TlME switch to AMPL
and the MAlN TRIGGERING SOURCE switch to INT.
Apply the external signal to the Amplifier unit installed in
the vertical compartment. "The attenuator switch of the
Amplifier unit determines the horizontal deflection factor.
For information on X-Y applications see the oscilloscope manual. Also, the reference books listed under
Applications provide information on X-Y measurements
and interpreting the resultant
lissajous displays.
Mainframe Operating Modes
-
DLY'D TRIG IN.
This connector is an external input for
the delayed triggering circuit when the DLY'D TRlG
EXT.
SOURCE switch is set to
When the DLY'D T'RIG
SOUHCE switch is set to INT and P613 is properly
connected,
the DLY'D TRlG IN connector serves as a
Delayed Sweep Gate Out connector (not labeled). The
Delayed Sweep Gatesignal is arectangular positive-going
pulse with approximately 3.0 V amplitude and pulse width
coincident with the delayed sweep.
X-Y
Operation
Some applications required one signal displayed versus another rather than against time (internal sweep). The
amplifier function of the
7B53A/7B53AN allows an external signal to be appliedtothe horizontal deflection system
either externally via the MAIN TRlG IN OR AMP IN
connector or internally by way of the triggering system.
REV
A
JUN
1980
The 7B53A17B53AN can be operated in a 7000-Series
OsciIIsocope which has four plug-in compartments, either
independently, in the Alternate or Chopped Horizontal
Modes, or as a delayed sweep unit. However, when the
7B53N7B53AN is operated as a delayed sweep unit, it
must be triggered for a crt display.
It cannot delay another
time base unit but it can delay its own internal delayed
sweep. Refer to the appropriate oscilloscope manual for
additional mainframe horizontal operating information.
General
The following information describes the procedureand
techniques for making basic measurements with a
7B53N7B53AN installed in a 7000-Series Oscilloscope.
These applications are not described in detail, since each
application must be adapted to the requirements of the
2-13
Operating
7B53A
Instructions--7B53A/7B53AN
individual measurement. This instrument can also be used
for many applications not described in this manual.
Contact your local Tektronix
Fieldofficeor representative
for assistance in making specific measurements. The
following books describe oscilloscope measurement
techniques which can be adapted for use with this
instrument. horizontal
4. Divide the period of the reference signal (seconds)
by the product of the horizontal deflection established in
2
(division) and the setting of the TIMWDIV OR DLY
step
TIME switch. This is the horizontal conversion factor:
Horizontal
Conversion
Factor
reference signal period (seconds)
-
deflection
TIM WDlV OR
X
DLY TlME
(divisions) switch setting
J. Czech, "Oscilloscope Measuring Techniques",
Phillips Technical Library, Springer-Verlag, New York,
1965. 5. To measure the period of an unknown signal
disconnect the reference signal and connect the unknown
signal tothevertical unit. Set the TIMWDIVOR
DLY'TIME
switch to a setting that provides sufficient horizontal
John
D.
Lenkl
andbo book
and Applications", Prentice-Hall, Inc. Englewood
of Oscillosco~e
deflection to make an accurate measurement. Do not
the
VARIABLE
control,
Cliffs, N.J., 1968.
6. Measure the horizontal deflection in divisions and
Charles
H.
Roth, Jr., "Use of the Oscilloscope",
Programmed Text, Prentice-Hall, Inc., Englewood
cliffs, N.J., 1970.
J.H. Golding, "Measuring Oscilloscope", Tran-
satlantic, 1971.
calculate the period of the unknown signal using the
following formual:
Period
(Seconds)
-
--
7-IMVDIV OR horizontal horizontal
X
DLY TIME
setting
conversion
factor (divisions)
X
deflection
Comparison Measurement Techniques
Sweep Rates. To establish an arbitrary horizontal
Example. Assume a reference signal frequency of
455 Hz(period2.19
setting of
.2
ms),aTIMVDlVORDLYTIMEswitch
ms, and the VARIABLE control adjusted to
sweep rate based upon a specific reference frequency provide a horizontal deflection of eight divisions.
proceed as follows:
1. Connect the reference signal to the input of the
vertical unit. Set the
Volts/Division switch of the vertical
unit for four or fivedivisions of vertical deflection. Obtain a
Substituting these values in the horizontal conversion
factor formula (step 4):
Horizontal
Conversion
-
Factor
triggered display.
2.
Set the
TIME/DIV OR DL'Y TlME switch and the
VARIABLE control (Variable Selector connector set to
Main Variable) so one cycle of the signal covers an exact
number of horizontal divisions. Do not change the
VARIABLE control after obtaining the desired deflection.
Then, with a
ps, the period of an unknown signal which completes
50
one cycle in seven horizontal divisions can be determined
by using the period formual (step
Period
(Seconds)
TIME/DIV OR DLY TIMEswitch setting of
6):
=5Ops X 1.37 X 7= 48Ops
This display can be used as a reference for frequency
comparison measurements.
Phis answer can be converted to frequency by taking
the reciprocal of the period in seconds (see application on
3. To establish an arbitrary sweep rate so the period
Determining Frequency Measurements).
(time for one complete cycle) of an unknown signal can be
measured accurately at any setting of the
DLY TlME switch, the period of the reference signal must
be known. If it is not known, it can be measured beforethe
VARIABLE switch is set in step 2.
TIM WDlV OR
Time
Duration Measurements
Po measure time between two points on a waveform,
use the following procedure:
2-14
REV
A
JUN
1980
Operating
7B53A
Instructions--7B53A17B53AN
1. Connect thesignal to be displayedtotheinput of the
vertical unit.
Determining Frequency
The time measurement technique can also be used to
determine the frequency of a signal. The frequency of a
2,
set
the
Vertical
and
~~~~~~~~~l ~~d~
on
the indicator oscilloscope to display the plug-in units
periodically recurrent signal is the reciprocal of the time
duration
(period)
Of One
used.
Use the following procedure:
3.
Set the Volts/Division switch of the vertical unit to
display about four divisions of waveform.
4.
Set the MAIN TRIGGERING controls to obtain a
1. Measure the time duration of one complete cycle of
the waveform as described in the previous application.
2.
'Take the reciprocal of the time duration to deter-
stable display. mine the frequency.
Setthe TIMVDIVOR DLY TIMEswitchtothefastest
5.
sweep rate that displays less than eight divisions between
Example. The frequency of the signal shown in Fig. 2-
11 which has a time period of 0.5 ms is:
the time measurement points (see topic entitled "Time
Measurements" and Fig. 2-6).
6.
Adjust the vertical unit position control to move the
points between which the time measurement is made to
Frequency
-.-----"*--
time period 0.5 ms
--=
1
the center horizontal line.
X
cycle-
1
--*-.-
--
2
kHz
7.
Adjust the horizontal POSITION control to position
the time-measurement points within the center eight
divisions of the graticule.
8.
Measure the horizontal distance between the time
measurement points. Be sure the VARIABLE control is set
to CAL.
8
9. Multiply the distance measured in step
setting of the
TIME/DIV OR DLY TlME switch.
by the
Example. Assume that the distance between the time
measurement points is five divisions (see Fig. 2-11), and
"TIMVDIV OR DLY TIME switch is set to
the
.1
ms.
Using the formula:
horizontal TlME/DIV OR
Time Duration
--
distance
X
DLY TIME
(divisions) setting
Substituting the given values:
'Time Duration
The time duration
-
5
is
0.5 ms.
X
0.1 ms
i
Fig.
2-11.
waveform.
[s+~Horizontal
I
I
Measuring the
distance
time
duration between points
I
I
I
on
Risetime Measurements
Fiisetime measurements empfoy basically
techniques as time-duration measurements. The main
difference
made.
measuring
waveform.
isthe points between which the measurement is
The following procedure gives the basic method of
risetime between the 10% and 90% pointsof the
Falltime can be measured in the same manner
on the trailing edge of the waveform.
the
same
a
REV
B
JUN
1980
Loading...
+ 135 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.