Tektronix 7A18 Service manual

INSTRUCTION
7A18
IAi8/tA"%H
DUAL
TRACE
AMPLIFIER
7A18
TABLE
7A18
SECTION 1 SPECIFICATION Page OPTION INFORMATION
OF
CONTENTS
lntroduction Electrical Characteristics 7A18 And Mainframe Frequency
Response 1-3 Environmental Characteristics 1-3 Physical Characteristics 1-3
SECTION
SECTION
SECTION 4 MAINTENANCE Accessories
2
OPERATING INSTRUCTIONS
Installation
Front Panel Controls and Connectors 2-1 General Operating Information Basic Applications 2-4
3
CIRCUIT DESCRiPTlON
Introduction Block Diagram Description
Detailed Circuit Description 3-1 Index
Preventive Maintenance Troubleshooting Replacement Parts Component Replacement
I-
I
1-1 SECTION 6 ELECTRICAL PARTS LIST
Abbreviations and Symbols Parts Ordering lnformation
SECTION 7
2-
1
2-2
SEC'TION 8 MECHANICAL PAR1'S LIST
3-1 3-1 Mechanical Parts List
4-3
4-4
CHANGE INFORMATION
DIAGRAMS AND CIRCUIT BOARD
ILLUSTRATIONS
Symbols and Reference Designators Voltage and Waveform Conditions
of
Mechanical Parts Illustrations
Mechanical Parts List
Information
SECTION 5 CALIBRATION
Recalibration Interval 5-1 or taken directly from 'Test Equipment Required 5-1
I
-.-
Part Part
Performance Check 5-3 the electronics industry. Change information, if any,
I
I
-
Adjustment
5-7
REV.
Abbreviations and symbols used in this manual are based on
IEEE Standard
Symbols for Units", MIL-STD-12B and other standards of.
located at the rear of this manual.
JUNE
1974
260
"Standard
is
7A18
Change information, if any, affecting this section will be found at the rear of the manual.
7A18
t
ntroduction
The 7A18 and 7A18N Dual Trace Amplifier plug-in
units are designed for use with Oscilloscopes. The tical except that readout encoding capabilities and an "IDENTIFY" function are provided only in the references made to the unless othewise noted. The 7A18 is a dual-channel, medium-bandwidth amplifier. Internal gain and compensa­tion circuits are automatically switched to correspond to
7A18 and 7A18N are electrically iden-
7A18 apply equally to the 7A18N
l-ektronix 7000-Series
7A18. All
TABLE 1-1
ELECTRICAL
i
wrViDiv
to
5
VIDIV;
the setting of the inverted for differential measurements. The operated in any plug-in Oscilloscopes.
The following electrical characteristics are
stated environmental range for instruments calibrated at an
ambient temperature of minute
tcrt
warmup unless otherwise noted.
step5
111
6
VOL-I"S/DIV
i-20"~ to +30°c, and after a five
I,
switch. Channel 2 can be
7A18 can be
compartment of the 7000-Series
valid over the
Deflectran
Uncalibrated
(VARIABLE)
GAlN
Frequency
Syslorn
(8
Dependent
dtv
refc~ence
Upper
DC
{Direct)
ower
Factor
Respar7se
stgnal)
Bandwrritlr
CuuplecJ
ljandwiclth
Accuracy
Vrrhrri
0
mV/Div,
:ontinuously teps;
extends
East
12.5
-
296
ViDiv
with
GAlN
vat
iablc
deflection
adjusted
between
factor
at
cal~brart
tn
at
Pern-rits
for
cdlrbrdted
set
rcs
uscrltuzcopes,
adjustrrrcnt
or>erdtiot~
of
tic:flect~nn
wit17
all
factor
7000-
-
--
Specif ication-7A1817AI 8N
7A18
Characteristic
Maximum Input Voltage
DC Coupled
AC Coupled
Channel Isolation
Input R and
---
C
Resistance
Capacitance
RC
Product
Displayed
No~se
TABLE
""
Performance
1-1
(contj
Requiremen_t-.
-
---
--
-
-
--
Supplemental
Information
!50 volts, (DC + Peak AC); AC com­Ionent 500 volts peak-to-peak maxi­num. one kilohertz or less.
j00 volts, (DC + Peak AC); AC compo­lent 500 volts peak-to-peak maxi­num, one kilohertz or
less.
50: 1 display ratio up to 50
megahertz.
---
--
-----
Approximately 20.0 pf
---*-----~"."
Nithin
:k
.."."-"".----,-....-*-
1%
between
all
deflection
(Tangentially Measured)
".
"-
,"
......
Overdrive Recovery Time
Common Mode Rejection
Ratio
DC Drift
Dr~ft with Time (ambient temperature and line voltage
constant)
Dr~fr w~th Temperature
(line voltage constant)
Time Delay between Channels
---
----
-
--
-*
----
-
Display Modes
--
---
-----
.,"-'"--
At least 10:
1
up to 59 megahertz.
-
Zhannel 1 only. Dual-trace, alternate between channels.
Added
algebraically. Dual-trace chapped between channels. Shannel 2 only.
$00 microvolts or
less
at
5
mV/Div in
7000-Series Oscilloscope.
1.1 ms or
less
to recover to within one livision after the removal of an over­jrive signal of up to -1-75 divisions or
-75 divisions regardless of overdrive
;ignal duration.
"
.,..----..-..-..
1.02 divisron
fter one hour
-
--
- - --
do
more than 0.01 d~vis~on per degree C.
'00 picoseconds or
~
or
less
warmup.
--
..-"
in
any one minute,
--
"-
less.
---
TABLE
7A18
7A78
TABLE A
AND
With
60
1500
2.1
(cant)
MAINFRAME
Scr
res
MHz
Wutl7
7400
50
blE4r
Sas
its$
Refer
$0
st~c
Specifaca!~un
fw
tbw
il"iocidt~C1
OSCI~~OSCQ~~~
7
ABLE
$3
SECTlON
7A18
2
Tilt;
7A18
ss
caltkrratasrt
cdn
hi3
rn~tdiltxt
Srrrtes
ns~alio"~;~~~~~~~~
vrrttcal
aid
iiilly
rrwtllost
dt
aqair~si
~nscrihaj
dud
sIt~df+
Iowa*r
sr-as~pr-r-t
the
iraariicn
tkrtt
To
leanewe
thtc
111~
b~liisa
OJIC
vvsrh
lAIt3
7A
9i-1
vn
ccrlnjlartiuxnrarl":
ro~ll,
01:
SIW
11
The
~Uieli~
ille
left
trt?n$
~,sr.ar*i.
the
1117~
imri
WD~V
ux~lar~h
18
OL~T
arrd
81193
\;ilasirpartn$tilllI
IJ~IT
I**
7A
1;R
Irrrv~t
w~ll
7A
18
corrtcal
7AiH,
iitef~frll~~flo~t
the
~~w~ili~dscope
rr;arly
~~iiaenrbd
Tn
~rsq"iail,
livi~lli
tiia:
~(ii:~tin~
Isp
fllisi~
ns
liriiy
of
1111<7A18
i-xtitE
081
Ccinlifrup
TING
for
taw
as
rrcecirs:rd
uf
Tektrcrnix
P~ir
/J~H~CPIJSJ
~kiqtl
ojve
w11lb
fhc
rrrwrti~t~
wtril
1110
Lntah
'"A
18"
pullrriq
rrn
~fia"
liar
Iron1
artd
!JIP
E~~wl~ilacl.r
'rrr
"17A
rhe
\is#-
ki
tile
rn
9,rrch
INSTRUCTIONS
[jolt
wrll
ire
forrtrtl
as
II
7CIC)O-
ln
iilirjrdx"
nrad
rrl
IRc?
i&ch
ji1dc-e
v,
i8ftl'd)
YQ
t!ttv~^l%cli
E
01
~!TP
iilrnti?ilctl
I
FRONT
AND
PAMEL
CONTROLS
CONNECTORS
r"lIS17
ION
Con
trcS5
llrrnrnrj
Wlsirlay
POSSTXUN
CIS.
Mode
lposi
iihc
ronjroj:
tlojl
of
r91rb
trarlc-.
in
1s
i--,rr\?anlied
rsr.riy
tr
ilip
dr.i.
Pobi
"ADD""
by
CIH
I
Operating
7A18
Instructions--7A18/7A18N
VOLTSIDIV Selects calibrated deflection factors
from
5
mV/Div to 5 VIDiv; ten
steps in a
VARIABLE Provides continuously variable
(VOLTSiDI V) calibrated settings between cali-
brated steps. Extends the deflection factor range to 12.5 or more.
GAlN Adjustment When the VARIABLE control is
pushed in, screw-driver adjustment for cali­bration of deflection factor.
DISPLAY MODE Selects one of the following modes
of operation: CH
ALT--A dual-trace display of the
ADD-Algebraically adds the signals
CHOP---A dual-trace display of the
CH single-trace display of the
TRIGGER SOURCE
Selects source of the trigger signal. The trigger signals provide internal triggering for the oscilloscope
base units.
CH
MODE-.-,I nternal trigger signal auto-
1-2-5 sequence.
un-
voltsidivision
it
becomes a front-panel
I.--A single-trace display of the
signal applied to Channel
signal applied to both channels. 'The channels are alternately dis­played, and switching occurs at the end of each time-base sweep.
applied to the CH 1 and CH 2
input connectors, and the algebraic sum is displayed on the CRT. The CH 2 POLARITY switch allows the display to be
+
CH 1 Position of the trace in this dis-
play mode is controlled by CH 1
POSITION control only.
signals applied to both channels.
The two channels time-share the
sweep as determined by the in-
dicator oscilloscope.
signal applied to CH 2.
obtained from signal applied to
CH
matically follows DISPLAY
MODE selection. In ADD or
CH 2 or CH
1
-
i nternal triggering signal
1.
1
1.
-
CH 2.
time-
CHOP display modes, the trigger signal is the algebraic sum of CH
1
and CH 2 trigger.
CH 2---Internal trigger signal
obtained from signal applied to CH 2.
CH 2 POLARITY
Provides
-WP--.A positive-going signal at the
INVERT--.A positive-going signal at
means of inverting the CH
2 display.
CH
2
input connector deflects
the
CRI
display upward.
2
the CH flects the CRT display down­ward.
input connector de-
GENERAL OPERATING INFORMATION
Introduction
For single-trace operation, either of the two identical amplifier channels can be used independently by setting the DISPLAY MODE and TRIGGER SOURCE switches to CH
1 or CH 2 and connecting the signal to be observed to the
appropriate input. In the discussions to follow, single-trace
1
operations using CH
only apply equally to CH 2 only.
Signal Connections
In general, probes offer the most convenient means of connecting a signal to the input of the uator probe offers a high input impedance and allows the circuit under test to perform very close to normal operating conditions.
The Tektronix ring, was designed specifically for use with 'Tektronix 7A­series amplifier units equipped with readout. The readout
coding ring on the probe connects to fier unit which automatically corrects the readout displayed
on the CRT to the actual deflection factor at the tip of the probe being used. For probes to be used with amplifier units without readout, see the Tektronix, Inc. catalog.
P6053A probe, with its readout coding
7A18. A 10X atten-
a
circuit in the ampli-
Vertical Gain Check and Adjustment
TO check the gain of either channel, set the VOLTSlDfV switch to 10 mV and connect oscilloscope calibrator to the input connector of the chan­nel being checked. The vertical deflection should be exactly
four divisions. If not, adjust the front-panel GAIN for exactly four divisions of deflection. The
is
engaged by pressing in the GAIN control knob and turn-
ing the knob with a narrow-blade screwdriver (see Front
40
mV, 1 kHz signal from the
GAlN adjustment
Operating
7A18
Instructions--~7A'18/7A18N
Panel Controls and Connectors), Turn the knob clockwise, then counterclockwise, until the GAIN control is engaged. When the will change as the knob is turned.
knob with the screwdriver until the deflection is set to
exactly four divisions, then remove
GAIN control is engaged, the vertical deflection
Turn the GAIN control the screwdriver.
Input Coupling
The Channel 1 and Channel 2 coupling (AC-GND-DC) switches allow a choice of input coupling methods. The type of display desired and the applied signal will determine the coupling to use.
DC coupling position must be used to display the
The
It
DC component of the signal,
AC signals below about
probe) and square waves with low-frequency components as these signals are attenuated in the AC position.
In the AC coupling position, the DC component of the
is
signal coupling position provides the
DC component much larger than the AC components. The
precharge feature should be used with large DC inputs. use this feature, first set the coupling to GND. Connect the probe to the circuit and wait about two seconds for the coupling capacitor to charge. Then set the coupling to AC.
input of the amplifier input connectors. However, the signals connected to the inputs are not grounded, and the same DC load is presented to the
blocked by a capacitor in the input circuit. Ihe AC
The GND position provides a ground reference at the
signa! source.
must also be used to display
30
hertz (ten hertz with a
best display of signals with a
without externally grounding the
10X
70
VOLTSIDIV and VARIABLE Controls
The amount of vertical deflection produced by a signal is
determined by the signal amplitude, the attenuation factor
of the probe, the setting of the
the setting of the
tion factors indicated by the settings of the witch apply only when the VARIABLE control is in the
calibrated (CAL IN) position.
The VARIABLE control provides variable, uncalibrated settings between the calibrated steps of the switch. With the VARIABLE control fully counterclock-
wise and the
brated vertical deflection factor
volts/division. By applying a calibrated voltage source to the input connector, any specific deflection factor can be set within the range of the VARIABLE control,
VARlABLE control. Calibration deflec-
VOtTSlDlV set to 5 voltsldiv the uncali-
VOLTSlDlV switch, and
VOLTS/DIV
VOLTS/DIV
is
extended to at least 12.5
CH
2
POLARITY
2
The CH displayed waveform of the signal applied to the CH This
is
when differential measurements are to be made. The CH POLARITY switch has two positions, +UP and INVER1. In
+UP position, the displayed waveform will have the
the same polarity as the applied signal and a positive will move the CRT trace up. In the INVERT position, positive-going waveform at the CH 2 input will be displayed on the CRT in inverted form and a positive move the trace down.
POLARITY switch may be used to invert the
particularly useful in added operation of the 7A18
Switch
2
input.
2
DC voltage
a
DC voltage will
DISPLAY MODE Switch
For single-trace operation, apply tire signal either to the
1
inpttt or the CH 2 input and set the DISPLAY MODE
CH
f
or CH
switch to the corresponding position: CH
To display a signal in one channel independently when a signal is also applied to the other channel, simply select the desired channel by setting the DISPLAY MODE switch to
1
the appropriate CH
Alternate Mode. The
MODE switch produces a display which alternates between
channel 1 and channel
Although the ALT mode can be used at all sweep rates, the CHOP mode provides a more satisfactory display at sweep rates below about rates alternate mode switching becomes visually per-
ceptible,
or CH 2 position.
ALT
position of the DISPLAY
2
with each sweep on the CRT.
0.2
millisecond/division. At slow sweep
Add Mode. "The ADD position of the DISPLAY MODE
switch can be used to display the sum or difference of two signals, for common-mode rejection to remove an undesired
signal, or for DC offset (applying a DC voltage to one channel to offset the DC component of a signal on the other channel). The overall deflection factor in the ADD
mode with both
position is the deflection factor indicated by either VOLTS/DIV switch. However, if the CH 1 and CH
VOLTSlDIV switches are set to different deflection factors,
the resultant amplitude is difficult to determine from the CRT display. In this case, the voltage amplitude of the
resultant display can be determined accurately only if the amplitude of the signal applied to one channel is known.
the ADD mode, positioning of the trace is contro:led by the
1
channel
Chop Mode. MODE switch produces a display which switched between channels at approximately a hertr rate (controlled by mainframe). In general the CHOP mode provides the best display at sweep rates slower than
POSITION control only.
VOLTSlDIV switches set to the same
The CHOP position of the DISPLAY
2.
is
electronically
500
2
In
kilo-
REV.
APR
1974
2-3
Operating
7A18
about repetitive phenomena is to be displayed.
instructions~--7Al8/7A18N
0.2
miilisecondIdivision or whenever dual-trace, non-
TRIGGER SOURCE Switcl~
CH
1.
'The CH 1 position of the 1-RIGGER SOURCE
switch provides a trigger signal obtained from the signal
1
applied to the CH display of the signal applied to the CH
CH
2.
The CH 2 position of the TRIGGER SOURCE switch provides a trigger signal obtained from the signal applied to the CH display of the signal applied to the CH
MODE. In this position of the TRIGGER SOURCE switch, the trigger signal for on the setting of the DISPLAY MODE switch. The trigger source for each position of the DISPLAY MODE switch is as follows:
MODE TRIGGER
CH
1
CH
2
ADD Algebraic sum of channel 1 and channel CHOP Algebraic sum of channel 1 and channel A LT Alternates between channel 1 and channel
inp~~t connector. This provides a stable
1
input connector.
2
input connector. This provides a stable
2
input connector.
the time-base unit is dependent
Channel Channel
SIGNAL
1
2
SOURCE
2
2
2
2.
Set the DISPLAY MODE and TRIGGER SOURCE
switches to display the channel used.
3.
Set the co~lpling switch to AC.
For low-
the
4.
divisions of the waveform vertically.
5.
play. Set the time-base unit to a sweep rate which displays several cycles of the waveform.
6.
tion of the waveform coincides with one of the
lines below the center hori~ontal line, and the top of the waveform is within the viewing area. With the time-base Position control, move the display so one of the upper peaks lies near the center vertical line (see Fig.
7.
peak. Check that the VARIABLE
in the CAL IN position.
frequenc y signals helo w about
DC
position to prevent attenuation of the signal.
Set the VOL'T'SIDIV switch to display about five
Set the time-base Triggering controls for a stable dis-
Turn the 7Af8 POSITION control so the lower por-
Measure the divisions of vertical deflection peak-to-
(VOLTSIDIV) control is
30
hertz use
2-2).
graticuie
'Trace Identification (7A18 only)
When the IDENTIFY button is pressed, the trace
0.2
flected about feature is particularly useful when multiple traces are dis­played. tion factor readout with the word "IDFNTIFY".
In instruments with readout, also replaces deflec-
division to identify the 7A18 trace. This
is
de-
BASIC APPLICA'TIONS
General
The following information describes the procedures and techniques for making basic the associated Tektronix oscilloscope and time-base. applications are not described in detail since each applica­tion must be adapted to the requirements of the individual measurements. This instrument can applications not described in this manual. Contact your
'l-ektronix Field Office or representative for assistance
local in making specific measurements
meastlrerrlents with a 7A18 and
These
also be used for many
with this instrument.
Peak-to-Peak Voltage Measurements (AC)
To
make peak-to-peak voltage measurements,
following procedure:
I.
Apply the signal to either input connector
use the
NOTE
This techrtique can also be used to make measore-
ments between two points on the wavefom, rather
titan peak to peak.
Ftg.
2-2.
Measurtng the peak-to-peak voltage of a waveform.
8.
7A18
Multiply the deflection measured in step 7 by the VOLTSiDlV switch setting. Include the attenl~ation factor of the probe if used.
Operating
7.
Measure the distance in divisions between the
reference line
DC
level
measurement is between the reference line and point A.
anrf the point on the waveform at which the
is
to be measured. For example, in Fig.
Instrtictions--7Al8/7A18N
2-3
the
EXAMPLE: Assume that the
is
flection uator probe, and the VO LTSIDIV switch is set to 1 V.
Peak to Peak
Substituting the given values:
The peak-to-peak voltage is 45 volts.
Instantaneous Voltage Measurements (DC)
Po measure the
proceed as follows:
1.
2.
switches to display the channel used.
3.
divisions of
4.5 divisions (see Fig.
Volts
Connect the signal to either- input connector.
Set the DISPLAY MODE and TRIGGER SOURCE
Set the VOLTSIDIV switch to display about five
vertical probe
-
deflection X (divisions) factor
Volts Peak-to-Peak
DC
level at a given point on a waveform,
the waveform.
peak to peak vertical de-
2-2)
'OL
=
4.5 X
rS'DIV
setting
using a
1
X
x
10
10X
attenuation
atten-
8.
Establish the polarity of the waveform. With the POLARITY switch in the +UP position, any point above the reference line is ~ositive.
9.
Multiply the distance measured in step 7 by the
VOLTS/DIV setting. Include the attentlation factor of the
if
probe,
divisions (see Fig. reference line usinq a of 0.5 V.
Using the formula:
Instan- vertical VOLTS! probe taneous Voltage (divisions) setting factor
Substitllting
used.
EXAMPLE: Assume
2-3)
=
distance X polarity
given
Instantaneous
Voltage
the vertical distance measured is
and the waveform is above the
10X probe wit11 a VOLTSIDIV setting
X
DIV X attenuation
values:
3.6
X i 1
X
0.5
V
X
CH
3.6
10
2
4. Set the couplinq switch to GND and trace to the voltage trace to the top control after this reference line has been establ~shed.
line can be checked at any time by position.
display. Set the time-base sweep rate for an display of the waveform.
thc lnottom graticule line or other reference line. If
is
ncqative with respect to ground, position the
graticitle line. Do not move the POSl PION
NOPE
Po
n~easrire a voltage Icvel with respect to a voltage
orher ti~afi grout?d,
step
4.
Set tl?~ COLI/?IIII~ switch to DC ar~d
reference voltage to the inpiit connector. Then
position the trace to the refere17ce ltne.
5.
Set the coupling switch to DC. The ground reference
6.
Set the time-base
make
the followirig c/)at?ges
sw~tclr~nq to the GND
'I
r~ggering controls for a stable
pos~tion the
apply
optinium
to
tlte
The instantaneous voltage
Fig.
2-3.
ence.
Measuring instantaneous voltage
1s
18
volts.
with
respect
to
some
refer-
Operating Instructions-7Al8/7A18N
7A18
Comparison Measurements
In some applications it may be desirable to establish
arbitrary units of measurement other than those indicated
VOL-ISiDIV switch. This is particularly useful when
by the
a
comparing unknown signals to use for the comparison-measurement facilitate calibration of equipment where the desired arnplitude does not produce an exact number of of deflection. The adjustment will be easier and more accurate if arbitrary units of measurement are established so that the correct number of divisions of deflection. The following procedure
describes
conlparison measurcmc?nts.
To estatrtlish an arbitrary vertical deflection factor- based
upon a specific refert?ncc amplitude, proceed as follows:
1. Connect the reference signal to the input connector.
Set the rime-base
tile signal.
of
2.
control to produce a display vertical VARIABLE control
3. To
the amplitude of an unknown signal can he measured accurately at any setting of the ai-nl,litude of the t-eierence signal must be icnown. If it is not known, it VOLTS'DIV control is set in step
how to establish arbitrary units of measure for
Set tile VOLTS'DIV switch and the VARIABLE
div~sions in atnplitude. Do not cf.tange the
estat,lish an arbitrary vertical deflection factor so
adjiistment is indicated by an exact
unit sweep rate to display several cycles
after obta~nrng the desiretl cfeflect~on.
can
be
measured before the VARIABLE
reference arnpiitcide. One
technitlue is to
which
is
an exact number of
VOLT-S/DIV switch, the
2.
tlivisions
Signal VOLT'SIDIV vertical vertical
Amplitude
EXAMPLE: Assume
volts, a
control divisions. Substituting these values in conversion factor formula (step
]-hen with a VOLTS/DIV setting of 2 V, the peak to-peak amplitude of an unknown signal which produces a vertical deflection of five divisions can be determined by using
-
setting X conversion X deflection
factor (divisions)
a
reference signal atnplitude of 30
VOLTSIDIV setting of 5 V and the VARIABLE
adjusted to provide a vertical deflection of four
tlie vertical
4)
:
2
-
V X
30 V
4 X 5
1.5
V = 1.5
X
5
=
15
6):
volts
Vertical
Conversion
Factor
Signal
Amnlitude
tlie signal amplitut.le formilla (step
=
Dual-Trace Phase Difference Measurements
Phase conlparison between two signals of the same fre­quency can be made 7A18. 'Th~s method of phase dilference measurement can be used system. To make the comparison, use the following pro­cedure:
position, clepending on the type of
up to the frequency limit ot the oscilloscope
1.
Set the CH 1 and CH 2 coi~pling switches to the same
using the dual-trace feature of the
cot~plir~g desired.
4.
Divide the atnplitude of the reference signal (volts) by
the product of tho vertical deflection (divisions) established
2
in step the vertical conversion factor. TRIGGER SOURCE to CH 1.
disconnect the signal to the input connector. Set the to a setting that provides sufficient vertical cleflection to make an accurate measurement. Do not readjust the VARIABLE control.
6.
calculate the amplitude of the unknown signal using the controls of the two channels so the displays are equal and
following
and the settirig of the VOLTS/DIV switch. 1-his is ALT is more suitable for high frequencies. Set the
Vertical reference signal
Conversion amplitude (volts)
Factor vert~cal VOLTSIDIV
5.
To measure the amplitude of an unknown signal,
Measure the vertical deflection in divisions and 5. Set the VOLTS/DIV switches and the VARIABLE
formula. about five divisions in amplitude.
--
-
.-
--
--
-
deflection (divisions) setting
reference signal and connect the unknown
X
-
--
switch
VOLTSI'DIV switch
2.
Set the DISPLAY MODE to ALT or CHOP. In
general, CHOP is more suitable for low frequencies and
3.
Connect the reference signal to the CH 1 input and
2
the comparison signal to the CH or probes which have similar time delay characteristics to connect the signals to
4.
If the signals are of opposite polarity, set the CH
POLARITY switch to invert the channel 2 display.
(Signals may be of opposite polarity due to
difference; if so, take this into account in culation.)
the input connectors.
input. Use coaxiat cables
180' phase
the
final cal-
2
Operating
7A18
Instructions--7A18/7A18N
6. Set the time-base unit to a sweep rate which displays
about one cycle of the waveforms. Set the Triggering
controls for a stable display.
7. Center the waveforms on
POSFTION controls.
the graticule with the 7A18
Substitiitinq the given values:
Phase Difference
The phase difference is 13.5".
-
0.3 X 45"
High Resolution Phase Measure~nents
8. Adjust the time-base Variable TitneiDiv control until one cycle of the reference signal occupies exactly eight horizontal divisions between the second and tenth vertical lines of the graticule (see Fig. graticuie represents 45"idivision). The sweep rate can now be stated in terms of degrees as
sponding points on the waveform.
45Oidivision to obtain the exact amount of phase dif- zontal difference of 3 divisions, tile phase difference is. ference.
sion with a sweep rate
45'/division.
9.
Measure the horizontal difference behveen corre-
10. Multiply the measured distance (in divisions) by Fig.
EXAMPLE: Assume a horizontal difference of 0.3 divi- Phase Difference
45O
of
of
2-4).
Each division of the
the cycle (360°
45°/division as shown in
+
8
divisions
=
Fig.
2-4. (divisions) (cfegreesi'division)
More accurate dual-trace phase made by increasing the sweep rate Variable
crease the sweep rate is Set the Magnifier to
sweep rate by the amount of sweep magnification.
the Magnifier, the magnified sweep rate is
-
TirneIDiv control). One of the easiest ways to in-
with the time-base Magnifier switch.
XI0
and determine the magnified
t~y dividing the sweep rate obtained previously
EXAMPLE. If the sweep rate is increased 10 times by
4.5°/division. Fig. 2-5 shows the same signals as used in 2-4 but with the Magnifier set to X10. With a iiori-
horizontal magnified
-
difference
measuremenis can be
(witho~lt changing the
45Oidivision : 10
X
sweep rate
Using the formula:
Phase Difference
Substiti~tintj the given values:
hori7ontal sweep rate Phase Difference - 3 X 4.5"
-
difference X (degrees/division)
(divisions) The phase difference is 13.5".
Fig.
2-4.
Measuring phase difference between two signals.
Fig.
f
ier.
2-5.
High resolution phase measurement using time-base
magrii-
2-7
Operating
7A18
Instructions--7A18/7A18N
Common Mode Rejection
The ADD feature of the 7A18 can
signals
which contain undesirable components. These un­desirable components can be eliminated through common­mode rejection. The procedure is as follows:
1.
Set the DISPLAY MODE s~zlitcli to ALT or CHOP
and the TRIGGER SOURCE switch to MODE,
be
used to display
2. Connect the signal containing
undesired information to the CH
3.
Connect a signal similar to the unwanted portion of
both the desired and
1
input connector.
tlie CH 1 signal to the CH 2 input connector. For example,
2-6
a
in Fig.
line-freclcrency signal is connectetl to Channel
to cancel out the line-frequency component of the Channel
1
signal.
4.
Set botii coupling switches to the same setting, DC or
depericling on the applied signal.
AC,
5.
Set the VOLTS/DIV switches so the signals are about
equal in amplit~~de.
6.
Set tlie DISPLAY MODE switch to ADD. Set the CH
2
POLAR1 rY switch to INVERT so the common-mode
sign;ils are
7,
VAR
of
opposite polarity.
Adjust the Channel
l
ABLE
control for maxiinurn cancellation of the
2
VOLTSiDlV switch and
cor~7mon-i~lode signal. The signal which remains shocrld be
tlie
only
desired portion of the Channel 1 signal.
2
EXAMPLE: An example of this mode of operation is
shown in Fig.
2-6. The signal applied to Charinel 1 contains unwanted line frequency components (Fig. 2-6A). A corre­sponding line frequency signal is connected to Channel (Fig. 2-GB). Fig. 2-6C shows the desired portion of tlie signal as displayed when common-mode rejectioli
The
above procedure can also be used for examining a
siiperimposed on some DC level when DC coupling is
signal
is
used.
usecf. A DC voltage of the Proper polarity Channel 2 can be used to cancel out the DC portion of the signal applied to Channel
1.
2
Fig.
Channel frequency component.
(Cf
~estrltant
2-6.
Using the
3
CRT
ADD
mode for common-mode rejection.
contains desired information along with line-
(B)
Channel 2 contains line frequency only.
display using common-mode rejection.
(A)
Change information, if any, affecting this section will
7A18
be
found at the rear of this manual.
l
ntroduction
This section of the manual contains a description of the
circuitry used in the
tion begins with block diagram shown in the Diagrams section. 'Then, each
circuit is described in detail using block diagrams to show
the interconnections between stages in each major circuit
and the relationship of the front-panel controls to the indi­vidual stages.
Complete schematics of each circuit are given in the Diagrams section. Refer to these schematics throughout the following circuit description for electrical values and relationship.
7A18 dual-trace amplifier. The descrip-
a
discussion of the instrument using the
BLOCK DIAGRAM
The following discussion standing the overall concept of the 7A18 before the indivi­dual circuits are discussed in detail. Only connections between the individual blocks are shown on the block diagram (see Diagrams section). Each block represents a major circuit within the instrument. The number on each block refers to the schematic on
is
complete circuit
The signal to be displayed on the CRT is applied to the
input connector. The signal passes through the input coupling switch, where the appropriate coupling to the attenuators. The correct amount of attenuation and the signal the input amplifier.
The Channel
setting, variable gain control, and trace positioning. The
2
Channel in addition to gain setting, variable gain control, and trace positioning. The outputs of these circuits are applied push­pull to the Signal and Trigger Channel Switches.
The Channel Switches select the proper signal and trigger
as determined by the DISPLAY MODE and TRIGGER
Input Amplifier provides signal polarity inversion
found.
1
Input Amplifier circuit provides gain
is
provided to aid in nnder-
the basic inter-
which the
is
selected,
VOLTSIDIV switch selects the
is
passed to
SOURCE switches. The signal and trigger outputs are provided to the oscilloscope via the Interface Connector.
The Readout Encoding circuit readout logic for the oscilloscope readout system. Data supplied to the mainframe readout system identifying the polarity, deflection factor, the uncalibrated symbol (when the VARIABLE control is in the outward position), and the plug-in mode. When the trace is deflected about factor readout is replaced by the word
IDEN-SIFY button
0.3
(7A18 only) provides
IS
pressed, the
division and the deflection
"IDEN1-IFY".
DETAILED CIRCUIT' DESCRIP'TION
ATTENUATOR
General
The Attenuator circuit determines the input coupling
and the 7A18 deflection factor.
NOTE
The
Cff
I
and
Cff
2
Attentlator circuits are identicai.
To minimize duplication, only
detail
throi~ghout tt~is discussion.
CH
7
is
described in
AC-GND-DC Switch
Input signals connected to the input connector can be AC-coupled, DC-coupled, or internally disconnected. SlOOA is a cam-type switch; a contact-closure chart
is
showing the operation this chart indicate when the associated contacts are position switch directly to the input Attenuator stage. In the AC position, the input signal passes through capacitor acitor prevents the DC component of the signal from passing to the amplifier. The path and connects the input circuit of the amplifier to ground. This provides a ground reference without the need to disconnect the applied signal from the input connector. Resistor
allows trace remains on screen when switching to the AC position
if the applied signal has a high DC level.
shown (open or closed). When the AC-GND-DC
is
in the DC position, the input signal
R
102, connected across the AC-GND-DC switch,
C10 to be precharged in the GND position so the
given on Diagram
GND position opens the signal
1.
The dots on
III
the
is
cot~pled
C10. This cap-
IS
Circuit Description--7A18/7Al8N
7A18
lnput Attenuator
The ef.fective overall deflection factor of the 7A18 is determined by the setting of the SIOOB. The basic deflection factor sion of CRT deflection. To increase the basic deflection factor to the values indicated on the front panel, precision attenuators are switched into the circuit. These attenuators
are hybrid devices which contain the necessary resistances and capacitors. Each attenuator is
SIOOB is a cam-type switch and the dots on the closure chart (see Diagram 1) indicate when the associated contacts are in the position shown (open or closed). In the
5
mV/Div position, input attenuation
signal is connected directly to the input amplifier.
For switch positions above five millivolts, the uators are switched produce the deflection factor indicated on the front panel. These a dividers. For DC and low-frequency signals, the attenuators are primarily resistance dividers and the voltage attenuation is determined by the resistance ratio in reactance of the capacitors in the circuit frequencies that their effect higher frequencies, the reactance of the capacitors decreases
and the attenuator becomes primarily a capacitance divider.
frequencies within
input attenuators are designed to maintain the same input RC characteristics (one megohm
of the adjustable series capacitor to provide correct attenuation at high frequencies and an adjustable shunt capacitor to pro-
vide correct input capacitance.
ttenuators are frequency-compensated voltage
In addition to providing constant
VOLTS/DIV switch. Each attenuator contains an
into the circuit singly or in pairs to
the bandwidth of the instrument, the
CHANNEL
VOLTSIDIV switch,
is
five millivolts per divi-
replaceahie as a unit.
is
not used; the input
the circuit. The
is
is
negligible. However, at
atLenuation at all
X
20 pF) for each setting
contact-
so high at low
1
atten-
INPUT AMPLIFIER
General
The Channel 1 lnput Amplifier converts the single-ended signal applied to the Channel 1 input connector to a dif­ferential (push-pull) output. Fig. 3-1 shows a detailed block
diagram of the Channel 1 lnput Amplifier. A schematic of
this circuit is shown on Diagram 2 in the Diagrams section.
lnput Source Follower
The Input Source Follower Q210A provides a high input impedance with a low impedance drive for the following
stage.
R210 limits the current drive to the gate of Q210A.
Dual-diode CR210 provides circuit protection by limiting
the voltage swing at the gate of or negative) 15 volts.
Q210B provides a constant current
0210A to about 4 (positive
source for the same case so that circuit.
0210A. 0210A and Q210B are encapsulated in
0210B temperature-compensates the
Paraphase Cascode Amplifier
Paraphase amplifier 0220-0320, in conjunction with 0225-0325, forms a the single-ended input signal to a differential output signal. Diodes 0220 close to ground to limit the voltage swing to about
40.6
isolation between the paraphase amplifier and the
VARIABLE controls. 'Phe gain of the Channel 1 lnput Amplifier is set in this stage by front-panel
R237A with the CAL IN switch pressed in. When the CAL IN switch is in the outward (uncalibrated) position and
turned fully gain of the amplifier is reduced by a factor of at least 2.5. Adjustment 1 R321 varies the base level of Q320 to provide the same voltage levels at the collectors of This prevents a position when varying the
CR220-CR221 hold the voltage level at the base of
volt. Common-base connected 0225-Q325 provide
co~~nterclockwise to minimum resistance, the
cascode amplifier. Q220-Q320 convert
GAIN-
GAlN control
(2225 and 0325.
rerouolt reference trace from changing
GAlN or VARIABLE controls.
Second Cascode Amplifier
The Second Cascode Amptifier stage provides a signal gain of TION control and, in the
circuitry. The emitters of common-base connected (2250-0350 provide a low-impedance point for injection of
the POSITION control and
Position of the trace is determined by the setting of the POSITION control, R 11. This control changes the current drive to
impedance point in the circuit, there change at these points. However, the change in current from the voltage difference at the collectors to change the position of the trace. Trace identification is accomplished by
inserting resistor R357 from ground through junction of the emitter current of Q250 to cause the trace to move. This aids in identifying the channel 1 trace when multiple traces are displayed.
vides high frequency compensation. work provide high-frequency response adjustment for this stage.
approximateiy two. This stage includes the POSI-
7A18 only, trace IDENTIFY
IDENrlFY switch currents.
Q250-0350. Since the emitters are a very low-
is
negligible voltage
POSI1"ION control produces a resultant DC
CR357 to the
R11-R256. This results in a slight increase in
The network
C246-C345-C245-R246-R345-R245
R245-C245 in this net-
pro-
Emitter Follower
Emitter Follower stage 0260-(2360 provides a low out-
put impedance to drive the Signal and Trigger Channel Switches,
between the Second Cascode Amplifier and
U270-U470. This stage also provides isolation
U270-U470.
Circuit Description-7Al8/7A18N
7A18
Fig.
3-1.
Channel 1 lnput Amplifier detailed block diagram.
CHANNEL 2 INPOUAMPLIFIER
General
The Channel 2 lnput Amplifier circuit
1
same as the Channel
lnput Amplifier circuit. Only the differences between the two circuits are described here. Portions of this circuit not described in the following des­cription operate in the same manner as for the Channel
lnput Amplifier circuit (corresponding circuit numbers
assigned in the
400 599
block diagram of the Channel
schematic of this circuit
range). Fig.
2
lnput Amplifier circuit. A
is
shown on Diagram 3 in the
Diagrams section.
is
basically the
3-2
shows a detailed
Paraphase Cascode Amplifier
The Paraphase Cascode Amplifier for Channel 2 consists
of
Q420, Q520, (3425, Q525,0426, to the functions described under Channel fier, the Channel provides a means of inverting the displayed signal. With the
1
CH
2
POLARI1-Y switch set to
biased on and the signal
Amplifier stage as for the Channel
CH
2
the
Q525
POLARITY switch set to
are biased off and
provide signal inversion.
CHANNEL
-+
and
Q526.
2
Paraphase Cascode Amplifier stage
+UP,
0425
is
passed to the Second Cascode
1
lnput Amplifier. With
INVERT,
Q426-Q526
are turned on to
SWITCH
U470
In addition
1
lnput Ampli-
and
0525
Q425
are
and
FROM
INPUT
ATTEN
REV.
D, DEC.
1976
PARAPHASE
Fig.
FROM
S23
17A18
ontv)
3-2.
Channel 2 lnput Amplifier detailed block diagram.
TO
SIGNAL
GMANNEL
SWITCH
U270
CHANNEL
SWiTCW
U470
Circuit Description-7Al8/7A18N
7A18
Second Cascode Amplifier
Ttie Second Cascode Amplifier for Channel 2 consists of 0440, 0540, by the POSITION control, R21 or by network R455-H555 as determined by the DISPLAY MODE switch. In any DIS­PLAY MODE switch position other than ADD, applied to the center arm of the POSl I^ION control through R32. The POSITION control varies the current drive to the emitters of impedance point in the circuit, there is negligible voltage change at these points. However, the change in current from the POSITION control produces a resultant DC voltagt-: difference at the collectors to change the position of the trace. When the DISPLAY MODE switch is in the ADD position, +50 volts tors
R455-R555 through R32 to balance the current drive
to the emitters of
volts (approximately) difference between the collectors.
Since +50 volts is not applied to the the ADD position of the DISPLAY MODE switch, the
control setting has no
Q450, and Q550. Position of the trace
Q450-0550. Since the emitters are a very low-
is
applied to the junction of resis-
(2450-0550. This results in a fixed zero
POSII-ION control in
effect on the circuit operation.
i
50 volts
is
set
is
CHANNEL SWITCHES
General
Ttie Channel Switches circuit provides Signal and Trigger outputs to the oscilloscope via the determined by the DISPLAY MODE and SOURCE switches. A schematic of this circuit is given on Diagram 4 in the Diagrams section.
Interface Connector as
THIC<GER
Signal Channel Switch
The Signal Channel Switch stage consists of integrated
circuit
U270 and its external components. This staqe selects
one, or mixes two input analog signals in response to inputs
from tlie DISPLAY MODE switch. The Signal Cliannel
Switch stage determines which input (CH 1 or CH 2) pro­vides
the signal to the oscilloscope as controlled by the DISPLAY MODE switch setting. Resistors R376-R377 set the current gain for each channel. Networks
C274-R274-C215-R275 and C374-R374-C375-R375 pro- vide the DC level shifting necessary to drive the vide high-frequency compensation for each channel. C275 and
C315 in these networks are high-frequency compen.
sation adjustments.
R276-R277 and
DISPLAY MODE
SELECTED
*Level is switched between the HI-level and LO-level
0.5
mate **Level is switched between the Hl-level and LO-level at a rate
determined by the setting
megahertz rate.
of
the time-base unit sweep rate.
mm
I
Fig.
3-3.
U270
input combinaliotis for DISPLAY MODE selection
'Trigger Channel Switching
The Trigger Channel Switch U470 Signal Channel Switch. This stage determines which (CH 1 or CH 2) provides the trigger stgnal for internal triggering of the time-base signal
is
controlled by inputs from the TRIGGER SOURCE switch. Resistors rent gain for each channel. Networks C474-R474-C475-R475 and C574-R574-C575-R575 pro-
vide high-frequency compensation for each channel.
An
inputloutput table for this stage is shown in Fig. 3-4.
Wlien the level at pin 14 is LO, the output of U470 is
determined
the level at pin 4 LO, the channel 1 and channel 2
and triggers are added algebraically.
Signal
Output stage, of a pair of common-base connected transistors which pro-
and
'Phe Signal Output stage, Q280-Q380, and the Trigger
circuits.
R476-R477 and R576-R577 set the cur-
by
the level at pin 4. With the level at pin 14 HI
0480-0580, are similar. Each stage consists
unit. 7 he selection of the trigger
Output
is
identical to the
DISPLAY MODE AND
---
at
an approxi-
input
mainframe
I
Fig. 3-3 shows the U270 input combinations for each position of the DISPLAY MODE switch. When the pin 14 is LO the output of U270 is determined by at pin
4.
With the level at pin 14 HI and the level
LO, the signals from both channel 1 and channel 2 are
passed to the Signal Output stage. This condition occurs
only when the DISPLAY MODE switch is set to ADD. In
is
this operating mode the signal output
of channel 1 and channel
determines the mainframe deflection.
2
signals and the resultant signal
the algebraic sum
level at
the level
at
pin 4
TRIGGER SWI"T'CHING
General
The Display Mode and Trigger Switching circuit deter-
mines
which input signal (Channel 1 or Channel the Signal and Trigger outputs to the mainframe as selected by the DISPLAY MODE and TRIGGER SOURCE switches.
This circuit also provides plug-in mode information to the
mainframe chop blanking circuit, and readout control infor­mation for proper
CRT display.
2)
providcs
Circuit Description--7A18/7A18N
7A18
DISPLAY MODE
'The DISPLAY MODE switch provides logic level outputs to the Signal Channel Switch stage Switches diagram 4). A table of the outputs for each position of the DISPLAY MODE switch is shown in Fig. 3-3.
'TRIGGER SOURCE
The TRIGGER SOURCE switch provides logic level out-
puts to the Trigger Channel Switch
Switches diagram 4).
position is shown in Fig.
ALT
1
I
Fig.
'TRIGGER
-G6-D-E
3-4.
Input/Output combinations for
SOURCE
Switch
(U270, Channel
Switch
(U470, Channel
A
table of the outputs for each switch
3-4.
Trigger S~gnaI
--"---------
..f-"T--i5
switch selections.
*f-.
-cG"-t""Ei
;-me..----
ernates between
and
CH
2
+-*
-"------
CY"?
-...
cw_
GM
-"
?.
----
1
and
DISPLAY
Source
------
--*
-----"
CW
2
MODE
--
-- - -
GW
.--*-
----
-"
----
added
---
1
-
and
CONNECTORS AND READOUT
General
The Connectors and Readout circuit consists of the power supply and signal distribution from the lnterface Connector and the Readout Encoding circuit. A schematic
is
of this circuit section.
shown on Diagram 6 in the Diagrams
Connectors
All the connections made to the mainframe by the 7A18 are shown on the Connectors portion of Diagram 6. Also shown are the power
sopply decoupling components.
Readout Encoding (7A18 only)
'The Readout Encoding circuit consists of switching
resistors and probe sensing stage
2,
the Channel 1 and readout of deflection factor, uncalibrated deflection factor (VARIABLE) information, and signal inversion (channel 2 only). Data is encoded on these output lines by switching resistors between them and the time-slot input lines, or by adding current through 0620.
R647-CR647 are switched between time-slot three
(TS-3) and Column output line when the CAL
uncal position. This results in the symbol > (greater
in the than) being displayed preceding the deflection factor read­out. R648 (Channel the Column output line when the
is in the INVERT position.
(inverted) being displayed preceding the deflection factor
readout.
Switching resistors are used to indicate the setting of the
VOLTSiDIV switch to the mainframe readout system. The VOLTSIDIV switch is a cam-type switch. The dots on the contact-closure chart (see Diagram 6) indicate when the associated contacts are closed. select the number 1, combination that prefix and R639 selects the symbol V (volts) in the 5 mV through switch. R638 selects the symbol V in the positions. stage (0620) select the decimal point (number of zeroes) again depending on the resistor combination switched in by the
factor of the probe connected to the input connector by sensing the amount of current flowing from the current sink through the probe coding resistance. The output of this circuit corrects the mainframe readout system to in­clude the probe attenuation factor. input connector provides the input to the probe sensing stage from the probe coding resistance (coded probes only; see used for the voltage divider with R621 through supply. determines, along with emitter resistor current. When the -15 volt time-slot pulse is applied to
Interface Connector B33, 0620 is interrogated and collector current through Interface Connector A37.
.5
V (500 mV) positions of the VOLTSiDIV
R630, R631, and the output of the probe sensing
VOLTSiDIV switch.
Probe sensing stage Q620 identifies the attenuation
0pera;ing Instructions). The third contact is also
IDENTIF-Y input. The coding resistor forms a
The resultant voltage sets the bias on Q620 and
Row and Column output lines for
2
only) is switched between TS-2 and
2,
is
switched in. R637 selects the m (milli-)
is
added to the column current output
Q620. This circuit encodes
IN switch is
CH
2 POLARITY switch
This results in the symbol
R633, R634, and R635
or 5 depending on the resistor
1,
2, and 5 V
'The third contact of the
CR621 to the
R622,
--I5
the collector
its
1
V
REV.
D,
JAN.
1975
Circuit Description--7A1817A18N
7A18
With a connector, 0620 is turned off. The deflection factor read­out
10X probe connected, the bias on 0620 will allow 100
a microamperes of collector current to flow. the deflection factor readout by a factor of 10.
The IDENTIFY button (S13 or 523 on Diagram two things when pressed
I.
channel of the discussion on the Channel 1 or Channel 2
1X
probe (or no probe) connected to the input
is
determined by the VOLTS!DIV switch position. With
1-his increases
I)
:
It
causes the trace representing the appropriate
7A18 to move about 0.3 division (see the
Input Amplifier).
does
2.
Forward biases CR621 and Q620 to result in a sufficient amount of collector current which, when added to the column current output, replaces the deflection factor readout with the word "IDENTIFY".
These two actions aid in identifying the
when multiple traces are displayed. When the IDENTIFY
is
button position are restored.
For further information on the operation of
system, see the oscilloscope instruction manual.
released, the deflection factor readout and trace
7A18 trace
the readout
Change information, if any, affecting this section will be found at the rear of this manual.
7A18
l
ntroduction Lubrication
This section of the manual contains maintenance infor-
mation for use in preventive maintenance, corrective main- connecting plug contacts, and switch contacts. Lubricate
tenance, and troubleshooting of the
Further maintenance information relating to component
color codes and soldering techniques can be found in the
instruction manuals for the 7000-series oscilioscopes.
7A18. switch detents with a heavier grease. A lubrication kit
PREVEN'TIVE MAINTENANCE
Use a cleaning-type lubricant on shaft bushings, inter-
containing the necessary lubricating materials and instruc-
is
tions
Tektronix Part Number 003-0342-01.
available through any Tektronix Field Office. Order
Recalibration
To ensure accurate measurements, the 7A18 should be checked after each months if used infrequently. A complete performance check procedure
1000 hours of operation or every six
is
given in Part I for Section
5.
General
Preventive maintenance, consisting of cleaning, visual inspection, lubrication, etc., performed on a regular basis, will improve the reliability of this instrument. Periodic checks on the semiconductor devices used in the unit are not recommended as a preventive maintenance measure. See semiconductor-checking information given under Troubleshooting.
The performance check procedure can be helpful in
isolating
troubles not apparent during regular operation may be
revealed
major
and
corrected.
troubles
in
the
unit.
Moreover,
minor
TROUBLESHOOTING
Cleaning
Avoid the use of chemical cleaning agents which
might damage the plastics in this instrument. Avoid
chemicals containing benzene, toluene, xylene, ace-
tone, or similar solvents.
Front Panel. Loose dust may be removed with a soft
cloth or a dry brush. Water and mild detergent may be
used; however, abrasive cleaners should not be used.
Interior. Cleaning the interior of the unit should pre-
cede calibration, since the cleaning process could alter the
settings of the calibration adjustments. Use low-velocity compressed air to blow dirt can be removed with a soft, dry brush, cotton-tipped
swab, or cloth dampened with a mild detergent and water
solution.
off the accumulated dust. Hardened
General
The following contained in other sections of this manual when trouble­shooting the 7A 18. The schematic diagrams, Circuit Description, and Calibration sections should be used to full advantage. information on circuit behavior and output requirements.
The Circuit Description section gives detailed
is
provided to augment information
Troubleshooting Aids
Diagrams. Circuit diagrams are given on foldout pages in Section 7. component in this instrument are shown on the diagrams. important voltages are also shown.
Circuit Board. The circuit board used in the outlined on the schematic diagrams, and a photograph of the board board-mounted electrical component photograph by
The circuit number and electrical value of each
7A18
is
shown on the back of Diagram
is
identified on the
its
circuit number.
1.
Each
is
2.
7A18
Component and Wiring Color Code. Colored stripes or dots on resistors and capacitors signify electrical values, tolerances, etc., according to the EIA standard color code. Components not color coded usually have the value printed on the body.
The insulated wires used for interconnection in the
7A18 are color coded to facilitate tracing a wire from one
point to another in the unit.
4-1
Semiconductor Lead Configuration. Fig.
lead configuration of the semiconductor devices used in this instrument.
shows the
DC Voltmeter and Ohmmeter -A voltmeter for checking voltages within the circuit and an ohmmeter for checking resistors and diodes are required.
3.
Test Oscilloscope-.A test oscilloscope is required to
view waveforms at different points in the circuit.
A Tektronix 7000-series Oscilloscope equipped with a readout system, Time-Base unit, and a 7A-series Amplifier unit with a probe will meet the needs for items 2 and
7D13 Digital Multimeter unit, 76-series
10X
3.
Troubleshooting Procedure
This troubleshooting procedure is arranged in an order
Troubleshooting Equipment
The following equipment is useful for troubleshooting
the 7A 18.
1. Semiconductor Tester--Some means of testing the transistors, diodes, and helpful. A transistor-curve tracer such as the Tektronix tion of a control or front-panel connector, see the
Type 576 will give the most complete information.
FET's used in this instrument is
which checks the simple trouble possibilities before pro­ceeding with extensive troubleshooting.
1. Check Control Setting. An incorrect selting of the 7A18 controls can indicate a trouble that does not exist. If there is any question about the correct function or opera-
Operating Instructions section.
Fig.
4-1.
Electrode configuration for semiconductors used in this instrument.
2.
7A18
Check Associated Equipment. Before proceeding with troubleshooting of the ment used with this instrument is operating correctly. If possible, substitute an amplifier unit known to be operating
correctly into the indicator unit and
persists. Check that the inputs are properly connected and that the interconnecting cables are not defective.
3.
Visual Check. Visually check the portion of the instrument in which the trouble is suspected. Many troubles can be located by visual indications, such as unsoldered connections, broken wires, damaged circuit boards,
damaged components, etc.
4.
Check Instrument Performance. Check the cali­bration of the unit, or the affected circuit by performing
I
--
Part trouble may only be a result of misadjusrment and may be corrected by calibration. Complete calibration instructions are given in Part
5. component or stage can be located by checking for the correct voltage or waveform in the circuit. Typical voltages and waveforms are given on the diagrams; however, these are not absolute and may vary slightly between instru­ments. To obtain operating conditions similar to those used to take these readings, see the instructions in the Diagrams section.
6. Check methods are provided for checking the individual compo-
nents in the are best checked by disconnecting one end to isolate the measurement from the effects of surrounding circuitry.
A. 'TRANSISTORS AND INTEGRATED CIRCUITS. The best check of transistor and integrated circuit operation is actual performance under operating conditions.
If a transistor or integrated circuit is suspected of being
defective,
ponent known to be good; however, be sure that circuit conditions are not damaged. If substitute transistors are not available, use a dynamic tester (such as Tektronix Type 576). Static-type testers may be used, but since they do not check operation under simulated operating conditions some defects may go unnoticed. Fig. of semiconductor devices. Be sure the power is off before attempting to remove or replace any transistor or integrated circuit.
Performance Check of Section
II
of Section
Check Voltages and Waveforms. Often the defective
l
ndiv idual Components. The following
7A18. Components which are soldered in place
it
can best be checked by substituting a com-
sl.ich that a replacement might also be
4-1
shows base pin and socket arrangements
7A18, check that the equip-
see if the problem
5.
The apparent
5.
Integrated circuits can be checked with a voltmeter, test oscilloscope, or by direct substitution. A good under­standing of the circuit description is essential to trouble­shooting circuits using integrated circuits. Use care when checking voltages and waveforms around the integrated circuits so that adjacent leads are not shorted together. An integrated-circuit test clip provides a convenient
14-
clipping a test probe to the cuits. Phis device also doubles as an integrated-circuit extraction tool.
A
B. DIODES.
a
short circuit by measuring the resistance between termi­nals with an ohmmeter set to the R resistance should be very high in one direction and very low when the meter leads are reversed. Do not check tunnel diodes or back diodes with an ohmmeter.
Do not use an ohmmeter scale that has a high internal
current. High currents may damage the diodes.
C.
RESISTORS. Check resistors with an ohmmeter. Resistor tolerance is given in the Electrical Parts List. Resistors normally do not need to be replaced unless the measured value varies widely from the specified value.
D. CAPACITORS. A leaky or shorted capacitor can be detected by checking resistance with an ohmmeter on the highest scale. Use an ohmmeter which will not exceed the voltage rating of the capacitor. The resistance reading should be high after initial charge of the capacitor. An open capacitor can best be detected with a capacitance meter, or by checking whether the capacitor passes AC signals.
7.
Repair and Readjust the Circuit. Special techniques required to replace components in this unit are given under Component Replacement. Be sure to check the perfor­mance of any circuit that has been repaired or that has had any electrical components replaced. Recalibration of the
affected circuit may be necessary.
diode can be checked for an open or for
and 16-pin integrated cir-
X I k scale. The diode
rneans of
REPLACEMEN-T PARTS
Standard Parts
Ail electrical and mechanical part replacements for the
7A18
can be obtained through your local Tektronix Field
Office or representative. However, many of the standard electronic components can be obtained locally in less time than is required to order them from Tektronix, Inc. Before purchasing or ordering replacement parts, check the parts
lists for value, tolerance, rating, and description.
REV.
D,
MAY,
1975
COMPONENT REPLACEMENT
7A18
When selecting replacement parts, it is important to remember that the physical size and shape of the component may affect its performance in the instru­ment. All replacement parts should be direct replace­ments unless it is known that a different component
will not adversely affect the instrument performance.
Special Parts
Some parts are manufactured or selected by Tektronix
to satisfy particular requirements, or are manufactured for Tektronix to our specifications. These special parts are indicated in the parts list by an asterisk preceding the part number. Most of the mechanical parts used in this instru­ment have been manufactured by Tektronix. Order all special parts directly from your local Tektronix Field Office or representative.
Ordering Parts
When ordering replacement parts from Tektronix, Inc.,
refer to the Parts Ordering Information and Special Notes and Symbols on the page immediately preceding the Elec­trical Parts List section. Include the following information:
1. lnstrument type
2.
Instrument Serial Number
3.
A description of the part (if electrical, include the
circuit number)
4.
Tektronix Part Number
(7A18)
General
The exploded-view drawing associated with the Mechan­ical Parts List may be helpful when disassembling or re­assembling individual components or sub-assemblies.
Circuit Board Removal
In general, the circuit boards used in the 7A18 need
never be removed unless they must be replaced. Electrical
connections to the boards are made by soldered con-
it
is
nections. If
assembly, use the following procedures.
A.
READOUT CIRCUIT BOARD REMOVAL (7A18 only)
1. Disconnect the wires connected to the outside of the
board.
2.
Remove the seven screws holding the board to the
mounting surface.
3.
Disconnect the wires connected to the inside of the
board.
4.
Remove the board from the unit.
5.
To replace the board, reverse the order of removal.
B. ATTENUATOR CIRCUIT BOARD REMOVAL
1. Remove the readout board as outlined in the previous
procedure.
necessary to replace a circuit board
Soldering Techniques
Attenuator Circuit Boards. The Attenuator circuit boards are made from polyphenylene oxide because of excellent electrical characteristics. Use more than normal care when cleaning or soldering this material. The following rules should be observed when removing or replacing parts:
1.
Use a very small soldering iron (not over 15 watts)
2.
Do not apply more heat, or apply heat for a longer
time, than is absolutely necessary.
3.
Use a vacuum-type desoldering tool to remove the
excess solder from the circuit board.
4.
Do not apply any solvent containing ketones, esters,
or halogenated hydrocarbons.
5. To clean, use only water-soluble detergents, ethyl,
methyl, or isopropyl alcohol.
its
2.
Disconnect the resistor/capacitor connected to the
rear of the board.
3.
Loosen the front set screw on the VARIABLEIGAIN
control shaft coupling (use a 0.050-inch hex-key wrench).
4.
Retnove the red VARIABLE control knob and rod
from the control shaft.
5. Remove the remaining front-panel knobs using a
111 6-inch hex-key wrench.
6. Remove the front panel from the instrument.
7.
Remove the attenuator shields.
8. Disconnect the wires and resistor from the input BNC
connector.
9.
Remove the input BNC connector.
10. Remove the POSITION control using a 5116-inch
7A18
nut driver.
11. Remove the attenuator board with cam switch from
the instrument.
12. To replace the board, reverse the order of removal.
C. AMPLIFIER CIRCUIT BOARD REMOVAL
1. Remove the Readout circuit boards as given pre-
viously.
2.
Remove the plastic plug-in guide from the rear of the
instrument.
3.
Disconnect the wires connected to the board from
the front-panel controls.
B. ROTARY SWITCHES
Single wafers on the DISPLAY MODE and TRIGGER SOURCE switches are not normally replaced. If any part of these switches should be replaced. A new switch can be ordered through your Tektronix Field Office.
When disconnecting or connecting leads to a wafer-
type rotary switch, do not let solder flow around and
beyond the rivet on the switch terminal. Excessive solder can destroy the spring tension of the contact.
is
defective, the entire switch assembly
Transistor and Integrated Circuit Replacement
4.
Loosen the front hex-socket screw in the front
coupling of the VARIABLE control shaft using a
0.050-inch hex-key wrench. Pull the VARIABLE knob and shaft from the front of the instrument.
5. Loosen the front hex-socket screw in the coupling
between the DISPLAY MODE and 'TRIGGER SOUCE
switch sections. Pull the T'RIGGER SOURCE knob and
long shaft from the front of the instrument.
6. Loosen the front hex-socket screw in the coupling of
the DISPLAY MODE switch shaft using a 5116-inch
key wrench. Pull the DISPLAY MODE knob and long shaft
from the front of the instrument.
7.
D i scon nect the resistor-capacitor combinations
connected to the ceramic strips at the front of the board.
8.
Remove the screws and nuts securing the board to
the chassis or other mounting surface.
9.
Remove the board from the instrument.
10. To replace, reverse the order of removal.
hex-
Switch Replacement
Several types of switches are used in the 7A18. The following special maintenance information is provided for the cam-type switches and rotary switches.
A. CAM-TYPE SWITCHES
CAUTION
m
Repair of cam-type switches should be undertaken
only by experienced maintenance personnel. Switch alignment and spring tension of the contacts must be carefully maintained for proper operation of the switch. For assistance in maintenance of the cam-type switches, contact your local Tektronix Field Office or
resen ta tive.
rep
Transistors and IC's should not be replaced unless they are actually defective. If removed from their sockets during routine maintenance, return them to their original sockets. Special care must be given to integrated circuit leads, because they can easily be damaged in removal from sockets. Unnecessary replacement or switching of components may affect the calibration of the instrument. When a transistor part of the instrument that may be affected.
is
replaced, check the operation of that
Recalibration After Repair
After any electrical component has been replaced, the calibration of that particular circuit should be checked, as well as the calibration of other closely related circuits. Performance Check instructions given in Part I of Section provide a quick and convenient means of checking the instrument operation. The Calibration Procedure in Part of Section 5 can then be used to adjust the operation to meet the Performance Requirements listed in Section 1.
The
5
II
Repackaging for Shipment
If the 1-ektronix instrument is to be shipped to a 1-ektronix Service Center for service or repair, attach a tag showing: owner (with address) and the name of an individual firm that can be contacted, complete instrument serial number and a description of the service required.
Save and re-use the package in which your instrument was shipped. If the original packaging available, repackage the instrument as follows:
Surround the instrument with polyethylene sheeting to protect the finish of the instrument. Obtain a corrugated cardboard of the correct carton strength and having inside dimensions of no than the instrument dimensions. Cushion the instrument by tightly packing three inches of dunnage or urethane foam between carton and instrument, on all sides. Seal carton with shipping tape or industrial stapler
The carton
test
strength for your instrument is 200 pounds.
is
unfit for use or not
less
than six inches more
at
your
caiton of
REV.
E
DEC
1976
4-5
Change information, if any, affecting this sec
7A18
,tion will be found at the rear of the manual.
Recalibration l nterval
To assure instrument accuracy, check the calibration of
7A18 every 1000 hours of operation, or every six
the
months if used infrequently. Before complete calibration, thoroughly clean and inspect this instrument as outlined in the Maintenance section.
Tektronix Field Service
Tektronix, Inc. provides complete instrument repair and
recalibration at local Field Service Centers and the Factory Service Center. Contact your local Tektronix Field Office
or representative for further information.
Using 'This Procedure
General. This section provides several features to facili
tate checking or adjusting the 7A18. These are:
Index. To aid in locating a step in the Performance
Check or Adjustment procedure, an index is given pre-
----
ceding Part I
ment procedure.
Performance Check and Part II
--
Adjust-
instrument is both correctly adjusted and performing
within all Performance Requirements as given in Section
TEST'
EQUIPMEN'T REQUIRED
General
The following test equipment and accessories, or its equivalent, is required for complete calibration of the 7A18. Specifications given for the test equipment are the minimum necessary for accurate calibration. Therefore, some of the specifications listed here may be somewhat less precise than the actual performance capabilities of the test equipment. All test equipment is assumed to be correctly calibrated and operating within the listed specifications.
The Performance Check and Adjustment procedures are based on this recommended equipment. If other equipment is substituted, control settings or calibration need to be altered to meet ment used. Detailed operating instructions for the test equipment are not given in this procedure. Refer to the instruction manual for the test equipment if more infor­mation is needed.
tlie requirements of the equip-
setup may
1.
Performance Check. The performance of this instru­ment can be checked without removing the side shields or making internal adjustments by performing only Part I Performance Check. This procedure checks the instrument
against the tolerances listed in the Performance Requirement column of Section 1. In addition, a cross-reference is pro-
II
---
vided to the step in Part the instrument to correct calibration. In most cases, the adjustment step can be performed without changing control settings or equipment connections.
Adjustment Procedure. To return this instrument to
correct calibration with the minimum number of steps, perform only Part cedure gives the recommended calibration procedure for all circuits in this instrument.
Complete Performance
pletely check and adjust all parts of this instrument, per-
form both Parts I and I I. Start the complete procedure by performing the Adjustment procedure and follow this with the Performance Check. 'i'his method will assure that the
II
Adjustment which will return
--
Adjustment. The Adjustment pro-
Check/Adjustment. To com-
Calibration Equipment Alternatives
All of the test equipment is required to completely check and adjust this instrument. However, some of the items used only for the Performance Check can be deleted without compromising capabilities. For example, the low-frequency amplitude signal generator is used on1 y in the Performance Check and may be deleted if the user does not desire to
check the lower frequency response or operation. Equipment used only for the Performance Check procedure is indicated by note only for note
tlie Adjustment procedure are indicated by
2.
the instrument's measurement
constant-
triggel- source
1
;
items required
Test Equipment
1.
7000-series oscilloscope, referred to Oscilloscope in this procedure. Tektronix 7403 recom­mended.
2.
T'ime-Base plug-in unit, Tektronix 7B50.
as
the Indicator
Loading...
+ 60 hidden pages