Tektronix 7078-TRX-BNC User manual

Keithley Instruments
Model 7078-TRX-BNC (CS-633)
500 V
center conductor to inner shield
30 V
inner shield to outer shell
1 A
0 °C to 50 °C, up to 35 °C at 70% relative humidity
<0.5 Ω
1015 Ω, center conductor to inner shield (500 V test voltage,23 °C at <4% relative humidity)
Brass with nontarnish finish and Teflon® insulation
28775 Aurora Road Cleveland, Ohio 44139 1-800-833-9200
tek.com/keithley
3-Slot Triaxial to BNC Adapter Information

Description

The Model 7078-TRX-BNC (CS-633) is a three-slot triaxial to socket BNC adapter. It can be used to connect BNC cables to equipment with three-lug triaxial socket connectors.
Figure 1: 7078-TRX-BNC (CS-633) 3-Slot Triaxial to BNC adapter
To maintain maximum insulation resistance, keep the connectors free of contaminants (such as dirt or oil). If the connections become contaminated, clean them thoroughly with methanol. Allow them to dry completely before use.

Characteristics

Working voltage
Maximum working current Operating environment Storage environment Contact resistance Insulation resistance
Material
PEAK
PEAK
PEAK
–25 °C to 65 °C
PA-153 Rev. E September 2022
*PPA-153E*
1
Model 7078-TRX-BNC (CS-633) 3-Slot Triaxial to BNC Adapter Information
igure 2: 7078-TRX-BNC (CS-633) adapter dimensions
F
To prevent electric shock, do not float inner shield more than 30 V above earth ground. Failure to do so may result in equipment damage, personal injury, or even death due to electric shock.
The case of the adapter is connected to the inner shiel d o f the triaxial connector. When used w
ith an electrometer or other instrument, input LO or guard voltage will be present on the
case of the adapter and any attached BNC connectors.
Figure 3: Model 7078-TRX-BNC (CS-633) electrical schematic
2 PA-153 Rev. E September 2022