The following safety precautions should be observed before using this product and any associated instrumentation.
Although some instruments and accessories would normally be used with non-hazardous voltages, there are
situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety
precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance
information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring
that the equipment is operated within its specifications and operating limits, and for ensuring that operators are
adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper
use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example,
setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user
documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be
performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly
trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I
and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC
60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly
connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II
connections require protection for high transient over-voltages often associated with local AC mains connections.
Assume all measurement, control, and data I/O connections are for connection to Category I sources unless
otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks
or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage
levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous
voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that
operators are prevented access and/or insulated from every connection point. In some cases, connections must be
exposed to potential human contact. Product operators in these circumstances must be trained to protect
themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive
part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedancelimited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards,
install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power recep tacle.
Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate
main input power disconnect device must be provided in close proximity to the equipment and within easy reach of
the operator.
11/07
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the
!
circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before:
connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes,
such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line
(earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of
withstanding the voltage being measured.
The instrument and accessories must be used in accordance with specifications and operating instructions, or the
safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and
operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground
connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The symbol on an instrument indicates that the user shoul d refer to the operating instructions located in the
documentation.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined
effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these
voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp
must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death.
Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such
damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protecti on from electri c shock and fire, re placement components in mains circuits - including the power
transformer, test leads, and input jacks - must be purchased from Keithley Instruments. Standard fuses with
applicable national safety approvals may be used if the rating and type are the same. Other components that are
not safety-related may be purchased from other suppliers as long as they are equivalent to the original component
(note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and
functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley
Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water-based cleaner . Clean the exterior of the instrument only . Do
not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist
of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected,
the board should be returned to the factory for proper cleaning/servicing.
Table of Contents
1Getting Started
General information ................................................................... 1-2
Warranty information .......................................................... 1-2
Contact information ............................................................ 1-2
•General Information — Covers general information that includes warranty informa-
tion, contact information, safety symbols and terms, inspection, and available options
and accessories.
•Product Overview — Summarizes the features of the Model 6430 Sub-Femtoamp
Remote SourceMeter.
•Mainframe and Remote PreAmp Familiarization — Summarizes the controls and
connectors on the mainframe and Remote PreAmp.
•Power-up — Covers line power connection, line voltage settings, fuse replacement,
and the power-up sequence.
•Display — Provides information about the Model 6430 display.
•Default Settings — Covers factory default setups and saving and recalling user setups.
•Menus — Covers the main and configuration menus as well as rules to navigate menus.
1-2Getting Started
!
General information
Warranty information
Warranty information is located at the front of this manual. Should your Model 6430 require
warranty service, contact the Keithley representative or authorized repair facility in your area
for further information. When returning the instrument for repair, be sure to fill out and include
the service form at the back of this manual to provide the repair facility with the necessary
information.
Contact information
Worldwide phone numbers are listed at the front of this manual. If you have any questions,
please contact your local Keithley representative or call one of our Application Engineers at
1-800-348-3735 (U.S. and Canada only).
Manual addenda
Any improvements or changes concerning the instrument or manual will be explained in an
addendum included with the manual. Be sure to note these changes and incorporate them into
the manual.
Safety symbols and terms
The following symbols and terms may be found on the instrument or used in this manual.
The symbol on an instrument indicates that the user should refer to the operating
instructions located in the manual.
The symbol on the instrument shows that high voltage may be present on the terminal(s). Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading used in this manual explains dangers that might result in personal
injury or death. Always read the associated information very carefully before performing the
indicated procedure.
The CAUTION heading used in this manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Inspection
The SourceMeter was carefully inspected electrically and mechanically before shipment.
After unpacking all items from the shipping carton, check for any obvious signs of physical
damage that may have occurred during transit. (There may be a protective film over the display
lens, which can be removed.) Report any damage to the shipping agent immediately. Save the
original packing carton for possible future shipment. The following items are included with
every Model 6430 Sub-Femtoamp Remote SourceMeter order:
Getting Started1-3
•SourceMeter mainframe with line cord.
•SourceMeter Remote PreAmp with cable to mainframe.
•Safety high voltage dual test leads (Model 8607).
•Accessories as ordered.
•Certificate of calibration.
•Instruction Manual.
•Support Software Disk including TestPoint instrument library for GPIB and LabV
indows driver.
for W
•Manual addenda, containing any improvements or changes to the instrument or manual.
IEW
If an additional manual is required, order the appropriate manual package (for example,
6430-901-00). Th
e manual packages include a manual and any pertinent addenda.
Options and accessories
The following options and accessories are available from Keithley for use with the
Model 6430.
Triax cables and adapters (for Remote PreAmp)
Model 6430-322-1A — This low-noise 8-inch cable is terminated with a 3-slot male triax
connector on one end, and three booted alligator clips on the other end.
Model 7078-TRX-1
3-slot male triax connectors.
Model 237-TRX-BAR Barrel Adapter
nect two triax cables together. Both ends of the ad
connectors.
CS-1053 Barrel Adapter — This barrel adapter is
triax connectors.
Model 237-BNC-TRX Adapter — This is a male
disconnected). It is used to terminate a triax cable with a BNC plug.
— This low-noise 12-inch triax cable is terminated at both ends with
— This is a barrel adapter that allows you to con-
apter are terminated with 3-lug female triax
terminated at both ends with 3-slot male
BNC to 3-lug female triax adapter (guard
1-4Getting Started
Model 237-TRX-T Adapter — This is a 3-slot male to dual 3-lug female triax tee adapter
for use with triax cables.
Model 7078-TRX-BNC Adapter — This is a 3-slot male
adapter lets you connect a BNC cable to the triax input of the Model 6430.
Model 237-TRX-TBC Connector
cap for assembly of custom panels and interface connections.
— This is a 3-lug female triax bulkhead connector with
triax to female BNC adapter. This
General purpose probes
Model 8605 high performance modular test leads — Consists of two high voltage
(1600V) test probes and leads. The test leads are terminated with a banana plug that has a
retractable sheath on each end.
Model 8606 high performance probe tip kit —
clips, and two spring hook test probes. (The spade lugs and alligator clips are rated at 30V
RMS, 42.4V peak; the test probes are rated at 1000V.) These components are for use with high
performance test leads terminated with banana plugs, such as the Model 8605.
Model 8607 High Performance Banana Cables —
banana cables. The cables are terminated with banana plugs that have retractable sheaths.
The following test leads and probes are rated at 30V RMS, 42.4V peak:
Models 5805 and 5805-12 Kelvin probes — Consists
probes with banana plug termination. Designed for instruments that measure four-terminal
resistance. The Model 5805 is 0.9m long; the Model 5805-12 is 3.6m long.
Model 5806 Kelvin clip lead set
plug termination. Designed for instruments that measure four-terminal resistance. A set of
replacement rubber bands is available (Keithley P/N GA-22).
— Includes two Kelvin clip test leads (0.9m) with banana
Consists of two spade lugs, two alligator
Consists of two high voltage (1000V)
of two spring-loaded Kelvin test
Model 8604 SMD probe set — Consists
surface mount device “grabber clip” on one end and a banana plug with a retractable sheath on
the other end.
of two test leads (0.9m), each terminated with a
Low thermal probes
Model 8610 low thermal shorting plug — Consists of four banana plugs mounted to a
1-inch square circuit board, interconnected to provide a short circuit among all plugs.
Model 8611 low thermal patch leads
a plug that has a retractable sheath at each end. These leads minimize the thermally-
banan
induced offsets that can be created by test leads.
Model 8612 low thermal spade leads
with a spade lug on one end and a banana plug that has a retractable sheath on the other end.
These leads minimize the thermally-induced offsets that can be created by test leads.
— Consists of two test leads (0.9m), each with a
— Consists of two test leads (0.9m), each terminated
Getting Started1-5
Cables and adapters
CA-176-1D Preamp Cable — Connects the REMOTE Preamp to the Model 6430
mainframe.
Models 7007-1 and 7007-2 shielded GPIB cables —
bus using shielded cables and connectors to reduce Electromagnetic Interference (EMI). The
Model 7007-1 is 1m long; the Model 7007-2 is 2m long.
Models 8501-1 and 8501-2 trigger link cables —
ments with Trigger Link connectors (e.g., Model 7001 Switch System). The Model 8501-1 is
1m
long; the Model 8501-2 is 2m long.
Model 8502 trigger link adapter —
the SourceMeter to instruments that use the standard BNC trigger connectors.
Model 8503 DIN to BNC trigger cable — Le
meter Complete) and two (External Trigger) of the SourceMeter
trigger connectors. The Model 8503 is 1m long.
Lets you connect any of the six Trigger Link lines of
Connect the SourceMeter to the GPIB
Connect the SourceMeter to other instru-
ts you connect Trigger Link lines one (Volt-
to instruments that use BNC
Rack mount kits
Model 4288-1 single fixed rack mount kit — Mounts a single SourceMeter in a standard
instrument (Models 195A, 196, 220, 224, 230, 263, 595, 614, 617, 705, 740, 775, etc.) side-byside in a standard 19-inch rack.
nts two instruments (Models 182, 428,
nts a SourceMeter and a Model 199
nts a SourceMeter and a 5.25-inch
Model 4288-5 dual fixed rack mounting kit —
high instrument (Model 182, 428, 486, 487, 2000, 2010, 2400, 2410, 2420, 2430, or 7001)
side-by-side in a standard 19-inch rack.
Mounts a SourceMeter and another 3 -inch
Calibration standards
Model 5156 Electrometer Calibration Standard Set — This calibration fixture contains
standardized resistors and capacitors needed to calibrate the Model 6430.
Carrying case
Model 1050 padded carrying case — A carrying case for a SourceMeter. Includes handles
and a shoulder strap.
1-6Getting Started
Product overview
The SourceMeter combines a precise, low-noise, highly stable DC power supply with a lownoise, highly repeatable, high-impedance multimeter and a remote preamplifier for ultra low
current measurements. It has 0.012% basic accuracy with 5 -digit resolution. At 5 digits, the
SourceMeter delivers 520 readings/second over the IEEE-488 bus. At 4 digits, it can read up to
2000 readings/second into its internal buffer. The unit has broad source and measurement
ranges:
•Source voltage from 5µV to 210V; measure voltage from 1µV to 211V.
•Source current from 0.5fA to 105mA; measure current from 10aA to 105.5mA.
•Measure resistance from 100µΩ (<100µΩ in manual ohms) to 21.1TΩ.
•Maximum source power is 2.2W
klqbThe Model 6430 is Y2K compliant.
.
Some additional capabilities of th
•Perform measurements at the DUT using the small
•Concurrent measurements of all three functions over the remote interface.
•Source-measure sweep capabilities (linear and log
eep list of up to 1000 points, memory sweep of up to 100 instrument setups).
sw
•6-wire ohms measurement with programmable I-source or V-source with V or I clamp.
•2.2W, 4-quadrant source and sink operation.
•Up to 11 stages of limit testing with a b
•Digital I/O for stand-alone binning operations or interface to component handler.
•Programming language and remote interfaces — The SourceMeter uses the SCPI pro-
gramming language and two remote interfa
•Trigger-Link interface to Keithley Series 7000 switching hardware.
•Math expressions — 5 built-in, up to 5 user-defined (bus only).
•Reading and setup storage — Up to 2500 readings and seven setups (five user defaul
actory default, *RST default) can be stored and recalled.
f
•Closed-cover calibration — The instrument can be calibrated
or remote interface.
panel
e SourceMeter include:
Remote PreAmp.
arithmic staircase sweeps, source
uilt-in comparator for pass/fail testing.
ce ports (IEEE-488/GPIB and RS-232C).
either from the front
ts,
Getting Started1-7
6430 SUB-FEMTOAMP REMOTE SourceMeter
OUTPUT
EDIT
TOGGLE
POWER
RANGE
DISPLAY
ON/OFF
AUTO
RANGE
®
EXIT ENTER
CONFIG MENU
SWEEP
TRIG
REL
LOCAL
FILTER
LIMIT
DIGITS SPEED
V
Ω
MEAS
I
FCTN
V
I
SOURCE
230
1
67
89
4
+/-
5
STORE
RECALL
EDIT
cбЦмкЙ=NJN
Front panel
Mainframe and Remote PreAmp familiarization
The following information should be reviewed before operating the instrument and is orga-
nized as follows:
•Mainframe front panel summary — Provides an overview of front panel controls
the display
.
•Mainframe rear panel summary — Provides an overview of rear panel connectors.
•Remote PreAmp summary — Covers the Remote PreAmp connectors.
Mainframe front panel summary
The front panel of the Model 6430 is shown in Figure 1-1.
and
Measurement (MEAS) function keys:
VMeasure volts.
IMeasure amps.
Ω
Measure ohms.
FCTNPerform math functions.
SOURCE function keys:
VSource voltage (V-Source).
I Source current (I-Source).
▲Increase source or compliance value.
▼Decrease source or compliance value.
1-8Getting Started
▲
▲
Operation keys:
EDITSelect source or compliance reading for editing.
TOGGLEToggle display positions of source and measure readings, or display V and I
LOCALCancel remote operation.
REL Enable/disable relative reading on present function.
FILTERDisplay digital filter status for present function and toggle filter on/off.
LIMITPerform configured limit tests.
TRIGTrigger a measurement from the front panel.
SWEEPStart configured sweep.
DIGITSChange number of digits of display resolution.
SPEEDChange measurement speed by selecting accuracy or specifying NPLC.
STORESet buffer size and enable reading storage.
RECALLDisplay stored readings and timestamp.
CONFIGPress CONFIG and then appropriate key to configure function or operation.
MENUAccess and configure Main Menu selections. When entering numeric data, use to
EXITCancels selection. Use to back out of menu structures.
ENTERAccepts selection.
RANGE keys:
▲Moves to next higher range, increments digit, moves to next selection.
▼Moves to next lower range, decrements digit, moves to previous selection.
AUTOEnables or disables measurement auto range.
measurements.
and Move through parameter values or selections within functions and operations.
clear reading to minimum absolute value.
Annunciators:
EDITInstrument in edit mode.
ERRQuestionable reading, invalid cal step.
REMInstrument in GPIB remote mode.
TALKInstrument addressed to talk over GPIB.
LSTNInstrument addressed to listen over GPIB.
SRQService request over GPIB.
REAROn = Remote Preamp not connected. Off = Remote Preamp connected.
RELRelative measure reading displayed.
FILTDigital filter enabled.
MATHMath function enabled.
4WRemote sensing enabled.
AUTOAutoranging enabled.
ARMSource-measure operations being performed.
TRIGExternal trigger source selected.
* Reading being stored.
Source control:
ON/OFFTurns the source on or off.
Handle:
Pull out and rotate to desired position.
Mainframe rear panel summary
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V, Ω,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
INTERLOCK-
DIGITAL I/O
cбЦмкЙ=NJO
Model 6430
rear panel
The rear panel of the Model 6430 is shown in Figure 1-2.
Getting Started1-9
Source-measure connectors:
`^rqflkWhen using the
Remote PreAmp, DO NOT use the INPUT/OUTPUT HI
and 4-WIRE SENSE HI banana jacks. Using these source-measure
banana jacks will result in errors and/or noise, and possible damage to the
Remote PreAmp. Use the input/output HI and sense HI terminals on the
Remote PreAmp.
INPUT/OUTPUT HI and LOUse to source-measure volts, amps, and ohms.
4-WIRE SENSE HI and LOUse for 4-wire remote sensing.
V, Ω GUAR
GUARD SENSEUse to correct for IR drops in Guard Output lead.
t^okfkdINPUT/OUTPUT LO is not internally connected to the chassis and can
The maximum allowable voltage differentials between the various terminals are provided on
the
rear panel of the mainframe (see Figure 1-2):
Lo
ever, to achieve rated accuracy, do not exceed 1V peak on the 100µA through 100mA ranges.
F
or the lower current ranges (1pA through 10µA), do not exceed the maximum allowable resis-
tance. The maximum resistance for each of these current ranges are provided in Section 2 (see
Con
nections to DUT, 4-wire sensing specification clarifications).
DDriven guard for guarded measurements.
Earth (chassis) ground screw.
float up to 42V peak above chassis ground.
w voltage differentials — The maximum low voltage differentials are 5V peak. How-
1-10Getting Started
SENSE
KEITHLEY
250V
Peak
HI
HI
40V
Peak
40V
Peak
GUARD
250V
Peak
250V
Peak
IN/OUT
LO
42V
Peak
REMOTE Preamp
MAINFRAME
IN/OUT
HIGH
cбЦмкЙ=NJP
Remote preamp
High voltage differentials — Exceeding the high voltage differentials (42V peak and 250V
peak) could damage the instrument and create a shock hazard.
Remote PreAmp connector:
REMOTE PREAMPConnect the Remote PreAmp to the mainframe using the supplied
Interlock and digital input/output port:
INTERLOCK — DIGITAL I/OMale DB-9 connector for digital output lines, interlock, and
Power module:
Contains the AC line receptacle and the power line fuse.
Trigger link connector:
TRIGGER LINK8-pin micro-DIN connector for sending and receiving trigger pulses.
RS-232 connector:
RS-232Connector for RS-232 remote operation. Use a straight through (not
preamp cable (CA-176-1D).
component handler signals.
Use a trigger link cable or adapter, such as Models 8501-1, 8501-2,
8502, 8504.
null modem) DB-9 cable such as Keithley Model 7009-5.
GPIB connector:
IEEE-488 INTERFACEConnector for GPIB remote operation. Use a shielded cable
Remote PreAmp summary
The Remote PreAmp is shown in Figure 1-3 and terminal identification for the IN/OUT
HIGH and SENSE triax connectors is provided in Figure 1-4.
(Model 7007-1 or 7007-2).
Getting Started1-11
HI
(Input/Output High)
GUARD
(Cable Guard)
LO
(Input/Output Low)
HI
(4-Wire Sense High)
GUARD
(Cable Guard)
LO
(Input/Output Low)
SENSE Triax Connector
IN/OUT HIGH Triax Connector
cбЦмкЙ=NJQ
Tri ax
connectors
Preamp connector:
MAINFRAMEConnect the Remote PreAmp to the mainframe using the supplied preamp cable.
Triax connectors:
`^rqflkWhen using the Remote PreAmp, DO NOT use the INPUT/OUTPUT HI
and 4-WIRE SENSE HI banana jacks on the mainframe. Using these
source-measure banana jacks will result in errors and/or noise, and possi
ble damage to the Remote PreAmp.
-
IN/OUT HIGHHICenter conductor – Input/Output HI
Guard Inner shell – Driven cable guard for guarded measurements
LOOuter shell – Input/Output LO
SENSEHICenter conductor – 4-wire Sense HI
GuardInner shell – Driven cable guard for guarded measurements
LO Outer shell – Input/Output LO
klqbCable guard is always available on the Guard terminals of IN/OUT HIGH and
SENSE regardless of the guard mode setting. Ohms guard can only be accessed at
Ω
GUARD banana jack on the mainframe.
the V,
The diagram on the Remote PreAmp (Figure 1-3) shows the maximum allowable voltage
differentials between the various terminals. Exceeding the voltage differentials could damage
the SourceMeter and create a shock hazard.
`^rqflkInstrument damage caused by excessive voltage differentials is not covered
by the warranty.
1-12Getting Started
Power-up
t^okfkdTo prevent electric shock, power must be off when connecting the Remote
PreAmp to the mainframe. If you wish to connect the Remote PreAmp at
this time, proceed to Section 2, `зееЙЕнбеЦ=oЙгзнЙ=mкЙ^гй=нз=нЬЙ=ã~áåÑê~ãÉ. Important safety precautions are included with the connection
procedure.
t^okfkdDuring
SourceMeter. These voltage spikes could be at hazardous levels (>42.4V
peak) and could damage sensitive DUTs. Never touch external circuitry or
the test leads when powering up the SourceMeter. It is good practice to
always disconnect DUTs from the SourceMeter before powering up the
unit.
t^okfkdT
o prevent electric shock, test connections must be configured such that
the user cannot come in contact with conductors or any DUT that is in
contact with the conductors. Safe installation requires proper shields, barriers, and grounding to prevent contact with conductors.
tion and safety are the responsibility of th
Line power connection
The SourceMeter operates from a line voltage in the range of 85 to 250V at a frequency of
50 or 60Hz. The SourceMeter will also operate at 400Hz; accuracy specifications are not guaranteed however. Line voltage and line frequency are automatically sensed. Therefore, there are
no switch
`^rqflkOperating
es to set. Check to be sure the operating voltage in your area is compatible.
possibly voiding the warranty.
Perform the following steps to connect the SourceMeter to line power and turn it on:
1.Before plugging in the power cord, make sure the front panel power switch is in the of
(0) position.
2.Connect the female end of the supplied power cord to the AC receptacle on the rear
pane
l.
the power-up, voltage spikes may appear on the terminals of the
Operator protec-
e person installing the product.
the instrument on an incorrect line voltage may cause damage,
f
t^okfkdThe
3.Turn on the instrument by pressing the front panel power switch
power cord supplied with the SourceMeter contains a separate ground
for use with grounded outlets. When proper connections are made, instrument chassis is connected to power line ground through the ground wire in
th
e power cord. Failure to use a grounded outlet may result in personal
injury or death due to electric shock.
to the on (1) position.
Power-up sequence
On power-up, the SourceMeter performs self-tests on its EPROM and RAM and momentarily lights all segments and annunciators. If a failure is detected, the instrument momentarily
displays an error message, and the ERR annunciator turns on. Error messages are listed in
Appendix B.
klqbIf a problem develops while the instrument is under warranty, return it to Keithley
Instruments, Inc., for repair.
If the instrument passes the self-tests, the firmware revision levels are displayed. For
example:
REV A01 A02
where: A01 is the main board ROM revision.
A02 is the display board ROM revision.
Also displayed is the line frequency (50, 60, or 400Hz). (If the wrong frequency is displayed, it can be manually set as covered below). The communication interface status is briefly
displayed. If the IEEE-488 bus is the presently selected interface, the identification message
will include the primary address. For example, if the primary address is 24 (factory default), the
“IEEE Addr=24” message is displayed. If the RS-232 interface is selected, the “RS-232” mes
sage is displayed.
Getting Started1-13
-
After the power-up sequence, the instrument goes to its normal display state with the output
off (red OUTPUT indicator light off). With the output off, the “OFF” message is displayed, and
dashes replace the reading. The “OFF” message indicates that the NORMAL output-off state is
selected. See Section 12, Output configuration for details.
System identification
To obtain the serial number and revision information, use the MENU/GENERAL/SERIAL #
selection or the *IDN? query via remote.
Line frequency setting
At the factory, the SourceMeter is configured to sense the power line frequency and automatically select the frequency setting. If, however, the line power source is noisy, the SourceMeter may select the wrong setting on power-up. If this situation occurs, noisy measurement
readings will result, and accuracy may be affected. You can manually set the line frequency
from the front panel or via remote as covered below.
Front panel line frequency
Set the line frequency from the front panel as follows:
1.Press the MENU key to display MAIN MENU.
1-14Getting Started
2.Using the right arrow key, select AD-CTRL then press ENTER to display A/D
3.Select LINE-FREQ, then press ENTER to display LINE FREQUENCY.
4.Place the cursor on 50Hz, 60Hz, or AUTO, and press ENTER. (Use 50Hz for 400Hz
5.Press EXIT to return to normal display.
Remote command line frequency
Table 1-1 summarizes remote commands used to control line frequency. See Section 17 for
A rear panel fuse protects the power line input of the SourceMeter. If the line fuse needs to
be replaced, perform the following steps:
1.The fuse is located in a drawer above the A
of the fuse
the fuse dra
2.Slide the fuse drawer out to gain access to the fuse. Note that the fuse drawer doe
pull all
`^rqflkFor
3.Snap the fuse out of the drawer and replace it with the same type (250V, 2.5A
5 × 20mm). The Keithley part number is FU-72.
4.Push the fuse drawer back into the power module.
Select line frequency (freq = 50 or 60).
Enable/disable auto frequency (state = ON or OFF).
C receptacle. See Figure 1-2. At the bottom
drawer is a small tab. At this location, use a small bladed screwdriver to pr
wer open.
s not
the way out of the power module.
continued protection against fire or instrument damage, replace the
fuse only with the type and rating listed. If the instrument repeatedly
blows fuses, locate and correct the cause of the problem before replacing
the fuse.
,
y
Display
Display format
Getting Started1-15
The SourceMeter display is used primarily to program source and compliance values and
display measured readings. Annunciators, which are located along the top of the reading/
message display, indicate various states of operation, as covered previously in Front panel
summary.
On power-up, the top (primary) display is used for measurements when the output is on
(with the output off, “OFF” or “ZER” is displayed). The bottom-left display is used for the pro
grammed source value (Vsrc or Isrc), and the bottom-right display is used for the programmed
compliance (Cmpl) limit.
Reading information can be displayed using either engineering units or scientific notation in
either fixed- or floating-point format. Use the GENERAL/NUMBERS selection of the main
MENU to select the display format, as discussed under Menus later in this section.
Engineering units example: 1.23456µA
Scientific notation example: 1.23456e -6
-
EDIT key
The SourceMeter must be in the edit mode to set source and compliance values. The edit
mode is selected by pressing the EDIT key (EDIT annunciator on). The editing cursor (flashing
digit) appears for the source or compliance reading. If a value is not edited within six seconds,
the edit mode is cancelled. While in the edit mode, the EDIT key toggles between the source
value and compliance value. See Section 3 for details on setting source and compliance values.
TOGGLE key
The TOGGLE key manipulates readings on the top display and on the bottom-left display. It
has no effect on the compliance reading (Cmpl), which is located on the bottom right. Each
press of the TOGGLE key sequences through the display options.
With the voltage (V) or current (I) measurement function selected, the TOGGLE key lets
you display both the current and voltage measurements at the same time. It also allows you to
toggle display positions of the source and measure readings.
With the ohms (Ω) measurement function selected, the ohms measurement is always dis-
played on the top display. The TOGGLE key lets you display either the programmed source
value, the current measurement, or the voltage measurement on the bottom-left display.
The TOGGLE key is also used to display statistical data on readings stored in the data store.
This function is performed from the data store RECALL mode.
klqbIf FCTN, REL, or Limits is enabled, the TOGGLE key is disabled.
1-16Getting Started
Status and error messages
Status and error messages are displayed momentarily. During SourceMeter operation and
programming, you will encounter a number of front panel messages. Typical messages are
either status or error in nature and are listed in Appendix B.
Remote display programming
The display can also be controlled by various SCPI :DISPlay subsystem commands.
Table 1-2 summarizes basic commands. See :D
mation on using these commands and also Disabling
q~ÄäÉ=NJO
Basic display commands
CommandDescription
:DISPlay:ENABle <state>Enable/disable display (state = ON or OFF).
:DISPlay:CNDisplayReturn to source-measure display.
:DISPlay:DIGits <n>Set display resolution (n = 4 to 7).
ISPlay subsystem in Section 17 for more infor-
front panel display later in this section.
Front panel tests
Use the TEST/DISPLAY TESTS selection of the main MENU to test various aspects of the
front panel. Test selections include:
•KEYS — Front panel keys are tested. Pressing a key displays a message that identif
ey. Pressing EXIT twice cancels this test.
that k
•DISPLAY PATTERNS — Use this selection to turn on all display pixels and annuncia-
tors. Subsequent key presses cycle through tests that turn off annunciators an
els of each digit, turn on the rows of the top-left display digit, and turn on all annun-
pix
ciators and pixels of each digit in a sequential manner. Press EXIT to cancel this test.
•CHAR SET — This test displays special characters. Press
See Menus later
in this section for more menu information.
Disabling front panel display
Front panel display circuitry can be disabled to allow the instrument to operate at a higher
speed. While disabled, the display is frozen with the following message:
FRONT PANEL DISABLED
Press LOCAL to resume.
As reported by the message, all front panel controls (except LOCAL, TRIG, and OUTPUT
ON/OFF) are
disabled.
ies
d corner
EXIT to cancel the test.
Getting Started1-17
▲
▲
Front panel control
Front panel display circuitry is controlled from the DISABLE DISPLAY configuration
menu, which is accessed by pressing CONFIG and then EDIT (or TOGGLE). To select an
option (NOW, NEVER, SWEEP, or STORE), use the
desired option, then press ENTER.
The options for DISABLE DISPLAY are explained as follows:
NOW — Select this option to disable the display now.
NEVER — Select this option if you do not want the display to disable.
SWEEP — Select this option if you want the display to disable while performing a sweep.
The display will disable as soon as sweep is started. The display will automatically re-enable
after the sweep is completed.
STORE — Select this option if you want the display to disable when storing source-measure
readings in the buffer. The display will disable as soon as the buffer is enabled. The display
will automatically re-enable after the storage process is completed. Note that with this
option, the display will disable while performing a sweep. Sweep readings are automatically
stored in the buffer.
and keys to place the cursor on the
Remote command programming
Use the following SCPI commands to enable or disable the front panel display circuitry:
:DISPlay:ENABLe OFFDisable the display
:DISPlay:ENABLe ONEnable the display
Default settings
By using appropriate menu selections, you can save and recall various instrument setups,
define the power-on configuration, or restore factory defaults as outlined below.
Saving and restoring user setups
You can save and restore up to five of your own user setups using the following procedures.
Saving setups
1.Select the various instrument operating modes you wish to save.
2.Press the MENU key, select SAVESETUP, then press ENTER.
3.From the SAVESETUP menu, select GLOBAL, then press ENTER.
4.From the GLOBAL SETUP MENU, select SAVE, then press ENTER.
5.Select the setup position (0-4) to save, then press ENTER to complete the process.
1-18Getting Started
Restoring setups
1.Press the MENU key, select SAVESETUP, then press ENTER.
2.From the SAVESETUP menu, select GLOBAL, then press ENTER.
3.From the GLOBAL SETUP MENU, select
4.Select the setup position (0-4) to restore, then pr
Power-on configuration
You can also define which of the stored setups (factory default or user) the instrument
assumes as the power-on configuration as follows:
RESTORE, then press ENTER.
ess ENTER to complete the process.
1.Press the MENU key, select SA
2.From the SAVESETUP menu, select GLOBAL, then press ENTER.
3.From the GLOBAL SETUP MENU, select
4.From the SET POWER-ON DEFAULT menu, choose the power-on configura
BENCH
5.If you chose to use a user setup as the power-on configuration, select th
number
or GPIB (see below), or USER-SETUP-NUMBER.
, then press ENTER.
Factory default settings
As summarized in Table 1-3, there are two sets of factory defaults, BENCH (front panel)
and GPIB (remote). You can restore either of these default conditions as follows:
1.Press the MENU key, select SA
2.From the SAVESETUP menu, select GLOBAL, then press ENTER.
3.From the GLOBAL SETUP MENU, select
4.Select BENCH or GPIB defaults as desired, then press ENTER to
process.
VESETUP, then press ENTER.
POWERON, then press ENTER.
tion:
e user setup
VESETUP, then press ENTER.
RESET, then press ENTER.
complete the
q~ÄäÉ=NJP=
Factory default settings
SettingBENCH defaultGPIB default
A/D Controls:
Auto-zeroOnOn
Line frequencyNo effectNo effect
BeeperOnOn
Data StoreNo effectNo effect
Digital output*15 or 715 or 7
Digits5.55.5
FCTNPower (off)Power (off)
Filter:OnOn
Source rangingBest fixedBest fixed
Voltage protectionNONENONE
Triggered source:
ControlDisabledDisabled
Scale factor+1.0+1.0
Triggering:
Arm layer:
Trigger layer:
EventImmediateImmediate
Count11
Output triggerLine #2, OffLine #2, Off
EventImmediateImmediate
Count11
Output triggersLine #2, All offLine #2, All off
Delay0.0 sec0.0 sec
Remote setups
You can also save and recall setups via remote using the following SCPI commands:
•Save and recall user setups using *SAV and *RCL (Section 15).
•Restore GPIB defaults using *RST (Section 15).
•Restore bench defaults using :SYSTem:PRESet (Section 17).
•Save the power-on configuration using :SYSTem:POSetup (Section 17).
Menus
Main menu
q~ÄäÉ=NJQ==
Main menu
Getting Started1-21
The following paragraphs discuss the main menu, configuration menus, and rules to navi-
gate menus.
Use the MENU key to access the Main Menu to select, configure, and/or perform various
instrument operations. These include default setup conditions, communications (GPIB or
RS-232), calibration, front panel tests, digital output states, auto zero and NPLC caching,
timestamp, numeric display format, and the beeper.
The Main Menu structure is summarized in Table 1-4. Use the Rules to navigate menus to
check and/or change menu options. Figure 1-5 shows the main menu tree.
Menu item
SAVESETUP
GLOBAL
SOURCE MEMORY
COMMUNICATION
GPIB
RS-232
1
SAVE
RESTORE
POWERON
BENCH
GPIB
USER SETUP NUMBER
RESET
SAVE
RESTORE
BAUD
BITS
PA RI TY
TERMINATOR
FLOW CTRL
DescriptionParameters
Configure setup conditions.
Control instrument settings.
Save present SourceMeter setup to memory location.
Return the SourceMeter to setup saved in memory.
Select the power-on default setup.
Powers-on to BENCH defaults.
Powers-on to GPIB defaults.
Powers-on to user setup.
Returns unit to BENCH or GPIB defaults.
Control memory sweep source setup configurations.
Save present setup configuration to memory location.
2
Return to configuration saved in memory location.
Select and configure remote interface.
Select GPIB (IEEE-488 Bus), set primary address, GPIB
protocol (see Appendix G).
Select the RS-232 interface, set parameters.
Select baud rate.
Select number of data bits.
Select parity.
Select terminator.
Select flow control.
0 to 4
0 to 4
See Table 1-3.
See Table 1-3.
0 to 4
See Table 1-3.
1 to 100
1 to 100
0 to 30
(Default: 24)
57600, 38400,
19200, 9600,
4800, 2400,
1200, 600, 300
7 or 8
NONE, ODD,
EVEN
CR, CR+LF, LF,
or LF+CR
NONE or
XON/XOFF
1-22Getting Started
q~ÄäÉ=NJQ==EÅçåíKF
Main menu
Menu item
CAL
1
3
TEST
DISPLAY TESTS
KEYS
DISPLAY PATTERNS
CHAR SET
A/D CTRL
AUTO ZERO
5
DISABLE
ENABLE
ONCE
LINE FREQ
NPLC CACHE
DISABLE
ENABLE
REFRESH
RESET
GENERAL
DIGOUT
SERIAL#
TIMESTAMP
NUMBERS
BEEPER
DescriptionParameters
Calibrate SourceMeter. (See Section 19.)
4
Perform tests on SourceMeter.
Test front panel keys and display digits.
Test front panel keys.
Test display pixels and annunciators.
Test special display characters.
Control auto zero, line frequency, and NPLC caching.
Control auto zero.
Disable auto zero.
Enable auto zero.
Force auto zero immediate update.
Set the line frequency.
50 or 60Hz, or
AU
TO
Control NPLC caching.
Disable NPLC caching.
Enable NPLC caching.
Update all NPLC values in cache immediately.
Clear NPLC cache of all NPLC values.
Select general operations.
Set Digital I/O port bit pattern.
Display serial number, firmware revision, SCPI version.
0-15
YES or NO
Reset timestamp.
Select engineering units or scientific notation display
format.
ENGR,
SCIENTIFIC
Enable or disable beeper.
NOTES
1. Top level menu choices indicated in bold. Indent
2. When the remote operation interface selection (GPIB or RS-232) is
check or change options of the selected interface, you must re-enter the menu structure.
3. Password is required to unlock calibration. (See Section 19.)
4. Press EXIT key to cancel test.
5. Disabling auto zero will reduc
e measurement accuracy.
ation identifies each lower submenu level.
changed, the SourceMeter performs a power-on reset. To
Press MENU key (Use and to select item, then press ENTER)
▲
▲
SAVESETUP
GLOBAL
SAVE
RESTORE
POWERON
BENCH
GPIB
USER-SETUP-NUMBER
RESET
SOURCE MEMORY
SAVE
RESTORE
COMMUNICATION
GPIB
RS-232
BITS
PARITY
TERMINATOR
FLOW-CTRL
CAL*
UNLOCK
EXECUTE
VIEW-DATES
SAVE
LOCK
CHANGE-PASSWORD
TEST
DISPLAY-TESTS
BAUD
KEYS
DISPLAY-PATTERNS
CHAR-SET
A/D CTRL
AUTO-ZERO
LINE-FREQ
GENERAL
DIGOUT
SERIAL#
TIMESTAMP
NUMBERS
BEEPER
NPLC-CACHE
* Without PreAmp connected
cбЦмкЙ=NJR
Main menu tree
Getting Started1-23
1-24Getting Started
Rules to navigate menus
Many source-measure functions and operations are configured from the front panel menus.
Use the following rules to navigate through these configuration menus:
klqbRules to edit
measure procedure.”
•A menu item is selected by placing the cursor on it
tion is denoted by the blinking menu item or option. The left and right arrow keys control cursor position.
•A displayed arrow on the bottom line indicates
(or messag
•A source or parameter value range is changed by placing the cursor on the range designator (i.e., k, M, G, etc.)
that when the ne
ses by a decade.
decrea
•A parameter value is keyed in by placing the cursor on the digit to be cha
one of
the following methods:
klqbY
ou can clear a parameter value by pressing the MENU key.
-Use the SOURCE
digit.
-Use the number keys (0 through 9) to key in the value at the selected digit.
-Use the ± key to change source value polarity, regardless of cursor position.
•Boolean selections (such as ON/OFF and HIGH/LOW) are toggled by placing the cursor on the selection and pressing a SOURCE or RANGE up or down arrow key.
•A change is only executed when ENTER is pressed. Entering an invalid parameter generates an error, and the entry is ignored. However, entering an out-of-range value
sma
ll or too large) selects the lower or upper limit, respectively.
•The EXIT key is used to back out of the menu structure. Any change that is not entere
is cancelled when EXIT is pressed.
source and compliance values are found in Section 3, “Basic source-
and pressing ENTER. Cursor posi-
there are one or more additional items
es) to select from. Use the appropriate cursor key to display them.
and using the SOURCE ▲ or ▼ or RANGE ▲ or ▼ keys. Note
xt higher or lower source range is selected, the reading
or ▼ or RANGE ▲ or ▼ keys to increment or decrement the
▲
increases or
nged and using
(too
d
Editing source and compliance values
▲▲▲
▲
Use the following keys to edit source and compliance values:
Getting Started1-25
•EDIT: selects the source or compliance display
ar in the field to be edited. If no key is pressed within a few sec
appe
be cancelled automatically.
will
field for editing. A blinking cursor will
onds, the edit mode
•EDIT and : places the display cursor on the display digit to be changed.
•SOURCE
pressing either
•RANGE
or ▼: increments or decrements the source or compliance value. Note that
▲
of these keys will automatically enable the source edit mode.
or ▼: selects the source or compliance range.
▲
•Numeric keys (0-9): allow you to directly enter source or compliance values.
•EXIT: exits the edit mode without waiting for the time-out period.
The basic procedure for editing source and compliance values is outlined below. See Bas
source-measure procedure in Section 3 for more details.
1.Press the EDIT key until the blinking cursor is
in either the source or compliance dis-
play field to be edited.
2.If desired, use the RANGE
and ▼ keys to select the desired source or compliance
▲
range.
3.To simply increment or decrement the display
value, use the EDIT and keys to
place the blinking cursor on the digit to be changed, then increment or decrem
v
alue with the SOURCE
updated immediately; you need not press ENTER to complete the process.
be
and ▼ keys. Note that the source or compliance value will
▲
4.To enter the source or compliance value directly, simply key in the desired value w
the numeric k
eys while the cursor is blinking. Again, the source or compliance value
will be updated immediately.
ic
ent the
ith
Toggling the source and measure display fields
Normally the measured reading value will appear in the upper, main display line, while the
source and compliance values will appear in the left and right fields respectively of the lower
display line. You can toggle the source and measure display fields by pressing the TOGGLE
key to place the source and measure values in the desired positions.
1-26Getting Started
Configuration menus
There are a number of configuration menus that can be accessed by pressing the CONFIG
key followed by the appropriate function or mode key. For example, you can configure the volt
age source by pressing CONFIG then SOURCE V. Configuration menus, which are summarized in Table 1-5 through Table 1-10, are available for the following operating modes:
These various configuration menus are covered in detail in the pertinent sections of this
manual.
-
•Measure functions (Ω, FCTN): Table 1-5.
•SOURCE V, SOURCE I, and RANGE: Table 1-6.
•REL, FILTER, and LIMIT: Table 1-7.
•TRIG: Table 1-8.
•SWEEP, DIGITS, SPEED, and STORE: Table 1-9.
•ON/OFF OUTPUT and Display (EDIT or TOGGLE): Table 1-10.
Getting Started1-27
q~ÄäÉ=NJR
Measurement configuration menus
Configuration menu itemDescription
CONFIG MEAS ΩConfigure ohms measure.
CONFIG FCTNConfigure functions.
CONFIGURE FUNCTION
POWEREnable power function.
OFF COMP OHMSEnable offset-compensated ohms, program parameters.
VO LT-C OE FFEnable voltage coefficient, program parameters.
VA R AL P HAEnable varistor alpha, program parameters.
%DEVEnable percent deviation, program parameters.
1-28Getting Started
q~ÄäÉ=NJS
Source and range configuration menus
Configuration menu itemDescription
CONFIG SOURCE VConfigure V source.
CONFIGURE V SOURCE
PROTECTIONSelect voltage protection.
DELAYProgram delay between source and measure.
AUTO DELAYEnable/disable auto delay.
GUARDSelect ohms or cable guarding.
TRIGControl triggered source.
CONFIG SOURCE IConfigure I source
CONFIGURE I SOURCE
GUARDSelect ohms or cable guarding.
DELAYProgram delay between source and measure.
AUTO DELAYEnable/disable auto delay.
TRIGControl triggered source.
DISABLE
ENABLE
OHMS
CABLE
CONTROLEnable/disable triggered source.
DISABLE
ENABLE
SCALE FACTORProgram scale factor.
OHMS
CABLE
DISABLE
ENABLE
CONTROLEnable/disable triggered source.
DISABLE
ENABLE
SCALE FACTORProgram scale factor.
CONFIG ▲ RANGEProgram upper range limit.
CONFIG ▼ RANGEProgram lower range limit.
CONFIG AUTO RANGE
AUTO RANGE TYPESelect auto range type.
SINGLE SRC MTRSelect single SourceMeter operation.
MULTIPLESelect multiple SourceMeter, program soak time.
Getting Started1-29
q~ÄäÉ=NJT
Rel, filter, and limit configuration menus
Configuration menu itemDescription
CONFIG RELProgram REL value.
CONFIG FILTERConfigure filter.
AUTO FILTEREnable/disable auto filter.
DISABLE
ENABLE
CONFIG LIMITConfigure limit tests.
CONFIGURE LIMITS MENU
DIGOUTProgram Digital I/O bit patterns for pass/fail.
IN Fail when in compliance, program bit pattern.
OUTFail when out of compliance, program bit pattern.
S/W LIMITSControl LIM2, 3, 5-12 tests limits and bit patterns.
CONTROLEnable/disable limit tests.
DISABLEDisable test.
ENABLEEnable test.
LOLIMSet low limit.
HILIMSet high limit.
PA SSSet pass Digital I/O bit pattern.
PA SSSet limit test pass conditions.
DIGIO PATTERNSet pass conditions Digital I/O bit pattern.
SRC MEM LOCSet pass conditions next source memory location.
NEXTUse next location.
LOCATION#Specify location number.
EOT MODESet Digital I/O line 4 to act as EOT or BUSY signal.
BUSYSet line 4 HI while unit is busy (3-bit mode).
/BUSYSet line 4 LO while unit is busy (3-bit mode).
EOTOutput line 4 HI pulse at end of test (3-bit mode).
/EOTOutput line 4 LO pulse at end of test (3-bit mode).
1-30Getting Started
q~ÄäÉ=NJU
Trigger configuration menu
Configuration menu itemDescription
CONFIG TRIGConfigure triggering.
CONFIGURE TRIGGER
ARM LAYERConfigure trigger model arm layer.
TRIG LAYERConfigure trigger layer of trigger model.
HALTReturn unit to idle state.
ARM INSelect arm layer detection event.
IMMEDIATEImmediate event detection.
GPIBGPIB GET or *TRG.
TIMERAfter timer interval elapses, enter interval.
MANUALFront panel TRIG key.
TLINKEnter TLINK line and state.
BEST FIXEDBest fixed range based on maximum value.
AUTO RANGEAuto range during sweep.
FIXEDSet fixed source range.
CONFIG DIGITSSet display number of digits.
DISPLAY DIGITSSelect 3.5, 4.5, 5.5, or 6.5.
CONFIG SPEEDSet measurement speed.
SPEED ACCURACY MENU
FASTFast speed.
MEDMedium speed.
NORMALNormal.
HI ACCURACYMaximum accuracy.
OTHERProgram NPLCs (number power line cycles).
CONFIG STOREConfigure data store timestamp.
STORE TIMESTAMP
ABSOLUTEAbsolute timestamp.
D ELTADelta timestamp.
1-32Getting Started
q~ÄäÉ=NJNM
Output and display configuration menus
Configuration menu itemDescription
CONFIG ON/OFF OUTPUTConfigure output.
CONFIGURE OUTPUT
OFF STATESet up output off state.
AUTO OFFEnable disable auto off mode.
INTERLOCKEnable/disable interlock.
CONFIG EDIT or TOGGLEEnable/disable display.
DISABLE DISPLAY
NOWDisable display immediately.
NEVERNever disable display.
SWEEPTurn display off during sweep.
STORETurn display off during buffer store.
NORMALNormal off state.
ZEROZero off state.
GUARDGuard mode off state.
DISABLEKeep output on.
ENABLETurn output off after each measurement.
DISABLEDisable interlock.
ENABLEEnable interlock.
2
Connections
•Connection Overview — Explains how to connect the Remote PreAmp to the main-
frame, provides basic information on the input/output co
ixture interlock.
test f
nnectors, and discusses using a
•Connections to
including 4-wire remote sensing, 2-wire local sensing, cable and ohms guard, as we
guard sele
•Guard
ing Methods — Discusses different guarding methods including cable guard,
ohms gua
DUT — Covers various methods for making connections to the DUT,
ction.
rd, and guard selection.
ll as
2-2Connections
Connection overview
t^okfkdTo prevent electric shock, test connections must be configured such that
the user cannot come in contact with conductors or any DUT that is in
contact with the conductors. Safe installation requires proper shields, barriers, and grounding to prevent contact with conductors.
tion and safety are the responsibility of th
Operator protec-
e person installing the product.
t^okfkdDuring power
SourceMeter. These voltage spikes could be at hazardous levels (>42.4V
peak) and could damage sensitive DUTs. Never touch external circuitry or
the test leads when powering up the SourceMeter. It is good practice to
always disconnect DUTs from the SourceMeter before powering up the
SourceMeter.
t^okfkdUp to 21
electrical shock that could cause injury or death, NEVER make or break
connections to the SourceMeter while it is on or is connected to an external
source.
-up, voltage spikes may appear on the terminals of the
0V may be present on the output and guard terminals. To prevent
Connecting Remote PreAmp to the mainframe
t^okfkdPotentially hazardous source voltage is routed from the mainframe to the
Remote PreAmp via the preamp cable. Adhere to the following safety precautions to prevent electric shock:
SourceMeter must be turned off before connecting (or disconnect-
√ The
ing) the Remote PreAmp to the mainframe.
√ When
√ AL
not using the Remote PreAmp, disconnect the preamp cable at
rear panel of the mainframe. DO NOT leave the preamp cable con-
the
nected to the mainframe if the other end is not
eAmp.
Pr
WAYS re-install the plastic safety cover onto the mainframe preamp
connec
tor whenever the Remote PreAmp is not being used.
connected to the Remote
Use the supplied preamp cable to connect the Remote PreAmp to the mainframe as follows:
1.From the front panel of the SourceMeter, turn the POWER off.
2.Connect the preamp cable to the Remote PreAmp. The preamp conne
Remote P
reAmp is labeled “MAINFRAME.”
ctor on the
3.At the rear panel of the mainframe, remove the plastic safety cover from the preamp
ctor. This connector is labeled “REMOTE PreAmp.” The plastic cover is
conne
connector with two screws. Hold on to the plastic cover and the retaining scre
to the
ver the Remote PreAmp is not being used, the plastic safety cover must be re-
Whene
installed on the mainframe preamp connector.
4.Connect the other end of the preamp cable to the mainframe.
Source-measure terminals
The SourceMeter can be used with or without the Remote PreAmp. However, when not
using the Remote PreAmp, the lower current ranges and higher resistance ranges are not available. The small Remote PreAmp can be positioned ne
cables. Short triax cables help minimize cable capacitance which could adversely affect the
response time of low current measurements.
Connections2-3
secured
ws.
ar the DUT allowing the use of short triax
When not using the Remote PreAmp, all connections to the S
rear panel of the mainframe using cables terminated with banana plugs. These terminals are
summarized in Section 1 (Figure 1-2). When using the Remote PreAmp, not all source-measure terminals are available at the preamp and must be accessed at the mainframe. Table 2-1
lists each
q~ÄäÉ=OJN
Terminal equivalency between mainframe and Remote PreAmp
INPUT/OUTPUT HI
INPUT/OUTPUT LO
4-WIRE SENSE HI
4-WIRE SENSE LO V,
Ω GUARD GUARD
SENSE Chassis
ground (screw)
N/A = Not available at Remote PreAmp.
klqbWhene
mainframe terminal and the equivalent Remote PreAmp terminal.
IN/OUT HI
IN/OUT LO
SENSE HI
N/A
GUARD (cable guard only)
N/A
N/A
ver the Remote PreAmp is connected to the mainframe, adhere to the follow-
ing rules to achieve best performance:
√ Do not
√ Access input/output low at
√ Do not
use INPUT/OUTPUT HI and 4-WIRE SENSE HI on the mainframe.
Access these terminals at the Remote PreAmp.
the Remote PreAmp or at the mainframe, but not both.
est circuit common should be tied to one point to avoid ground loops whic
T
enerate error currents.
could g
use guard from the mainframe and the Remote PreAmp at the same time.
Use one or the other.
ourceMeter are made at the
h
2-4Connections
cбЦмкЙ=OJN
Basic input/output
configurations
Center Conductor
Inner Shield
Outer Shield
HI (In/Out or Sense)
GUARD (Cable)
LO (In/Out)
A) Remote PreAmp T
riax Connector Configuration
6430 Mainframe
INPUT/OUTPUT
HI
LO
Zener
Clamp
42V Peak DC Max
or
10.5mA Max
Chassis
Ground
C) Mainframe Input/Output Configuration
Chassis Ground Screw
(on Rear Panel)
Triax to Alligator Clip
HI (center conductor)
Red boot
Guard (inner shield)
LO (outer shell)
Black boot
Green boot
B) T
erminal identification for 6430-322-1A cable
Remote PreAmp triax connectors
conductor of the connector (and triax cable) is HI (input/output or sense), and the inner shield
is cable guard. The outer shield (shell) of each triax connector is input/output LO.
The electrical configuration of each triax connector is shown in Figure 2-1A. The center
klqbThe 6430-322-1A tri
triax connector on one end and booted alligator clips on the other end. (See Figure
2-1B.) When connected to the Remote PreAmp, the alligator clip with the red b
HI, the one with the blac
The outer shells of the triax connectors are connected to Input/Output LO. Therefore, for
oating source-measure operations, a voltage potential will be present on the shells of the triax
fl
connectors. Even with the output of the SourceMeter off, voltage could be applied from the
external test circuit. As a general rule, do not touch the triax cables while any power is present.
t^okfkdT
ax cable (which is a supplied accessory) is terminated with a
oot is
k boot is GUARD and the one with the green boot is LO.
o prevent injury from electric shock, DO NOT touch the triax cables of
the Remote PreAmp while the SourceMeter is turned on or any external
source is turned on.
Input/output LO and chassis ground
Input/Output LO is not directly connected to chassis ground. For test circuits that require
Input/Output LO connected to chassis ground, you can use the supplied chassis ground plug.
Connect the lug end of the cable to the chassis ground screw on the rear panel of the main
frame, and plug the other end into the INPUT/OUTPUT LO banana jack.
With no Input/Output LO-to-chassis ground connection, floating source-measure operations
(up to 42V peak) can be performed. Inside the mainframe, a zener clamp (Figure 2-1C) is used
to isolate Input/Output LO from the chassis.
t^okfkdExceeding 42V between Input/Output LO and chassis ground creates a
shock hazard and could cause damage to the SourceMeter that is not cov
ered by the warranty.
`^rqflkDo not connect any external sources between Input/Output LO and chassis
ground. Current exceeding 10.5mA will damage the zener clamp
(Figure
Test fixture interlock
A test fixture interlock switch can be used with the SourceMeter to help protect the DUT.
The SourceMeter output will turn off when the lid of the test fixture is opened. However, you
must ALWAYS assume that power is present until you verify that the SourceMeter output is
off.
Connections2-5
-
-
2-1C). Such damage is not covered by the warranty.
t^okfkdTo prevent electric shock, test connections must be configured such that
the user cannot come in contact with conductors or any DUT that is in
contact with the conductors. Safe installation requires proper shields, bar
riers, and grounding to prevent contact with conductors. Operator protection and safety are the responsibility of the person installing the product.
See Section 12 for complete details on using the interlock and output configuration
information.
-
2-6Connections
Noise Shield
Triax Cable
Test
Fixture
HI
LO
Connect to earth safety
ground using #18 AWG
wire or larger
DUT
WARNING
Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (≥42V peak) is
present on input/output high, it is also
present on the guard terminals of the
Remote PreAmp and mainframe.
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
Connect to REMOTE
PreAmp connector on rear
panel of mainframe
IN/OUT
HIGH
Preamp Cable*
*Keithley part number: CA-176-1D.
cбЦмкЙ=OJO
Two-wire
sense
connections
Connections to DUT
klqbConnection drawings in this manual are shown using the Remote PreAmp. If not
using the Remote PreAmp, make connections to the equivalent banana jack termi
nals on the rear panel of the mainframe.
t^okfkdTo prevent injury from electric shock, DO NOT touch the triax cables of
the Remote PreAmp while the SourceMeter is turned on or is connected to
an external source that is turned on.
`^rqflkWhen using the Remote PreAmp, DO NOT use the INPUT/OUTPUT HI
and 4-WIRE SENSE HI banana jacks on the mainframe. Using these ter
minals while the Remote PreAmp is plugged in will result in errors and/or
noise, and possible damage to the Remote Amplifier.
Sensing methods
Basic source-measure operations are performed using either 2-wire sense connections
(Figure 2-2) or 4-wire sense connections (Figure 2-3). See Section 4, Ohms sensing for
additional information.
-
-
Connections2-7
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V, Ω,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Noise Shield
Triax
Cable
Test
Fixture
HI
Connect to earth safety
ground using #18 AWG
wire or larger
DUT
WARNING
Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (≥42V peak) is
present on input/output high, it is also
present on the guard terminals of the
Remote PreAmp and mainframe.
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
Triax
Cable
LO
Sense
LO
Sense HI
Banana Plug Cable
6430 Mainframe
(rear panel connectors)
IN/OUT
HIGH
Preamp Cable*
*Keithley part number: CA-176-1D.
cбЦмкЙ=OJP
Four-wire sense
connections
t^okfkdThere is no direct internal connection between earth ground and the
INPUT/OUTPUT LO terminal of the SourceMeter. Therefore, hazardous
voltages (>30V rms) can appear on the LO terminal. Typically, this can
occur when the SourceMeter is operating in any mode where the output
changes rapidly, such as quick, pulsed waveforms that can be generated
using the ZERO, AUTO-OFF output state, or fast pulse sweep operations.
To eliminate the shock hazard, connect the INPUT/OUTPUT LO terminal
to earth ground. You can connect the LO terminal to the chassis ground
screw terminal on the rear panel using the supplied chassis ground plug, or
connect it to a known safety earth ground.
t^okfkdFor floating source-measure operations, voltage will be present on the
outer shell (IN/OUT LO) of the triax cable(s). If IN/OUT LO is connected
to a noise shield as shown in Figures 2-2 and 2-3, the floating voltage will
also be present on the shield. To prevent contact with the floating voltage,
enclose the test circuit in a metal test fixture that is connected to safety
earth ground (as shown in Figures 2-2 and 2-3), and do not touch the triax
cable(s).
2-8Connections
2-wire sensing
drops in the test leads could be significant. The ratio between the test lead resistance and DUT
resistance determines the error that is introduced. If the error introduced by the IR drop of the
test leads is not acceptable, use 4-wire sensing.
When using 2-wire sensing to source and/or measure voltage, the error associated with IR
For example, assume test lead resistance (RL) is 1Ω and the DUT resistance (R
10kΩ. The error is calculated as follows:
Error = RL / R
1Ω / 10kΩ
= 0.0001
= 0.01%
Notice that as the resistance of the DUT increases, the error decreases. For DUT above
1GΩ, guarding should also be used. See Cable guard.
Since current in a series circuit is the same at all points in the loop, remote sensing does not
improve I-Source or I-Measure accuracy. Thus, if sourcing current and measuring current, you
can use local sensing.
klqb For Measure Only (V or I) operation, 2-wire sensing must be used.
DUT
=
DUT
) is
4-wire sensing
Voltage source and measure accuracy are optimized by using 4-wire sense connections.
When sourcing voltage, 4-wire sensing ensures that the programmed voltage is delivered to the
DUT. When measuring voltage, only the voltage drop across the DUT is measured.
Use 4-wire sensing for the following source-measure conditions:
•The error contributed by test lead resistance for local sensing is n
•Optimum Ohms, V-Source, and/or V
-Measure accuracy are required.
ot acceptable.
4-wire sensing specification clarifications
•There is no hardware configurations needed to enable 4-wire sense. Simply hook up the
sense wires; otherwise, the Model 6430 will sense the voltage locally through resistors.
•Specified accuracies for both source and measure are only achieved using 4-wire
sensing.
•Sense wires must be no more than 10Ω per lead.
•The Model 6430 will perform to rated specification with up to 1V drop per source lead
Guard
Shield
Triax Cable
Test Fixture
HI
LO
Connect to earth
safety ground using
#18 AWG wire or
larger
DUT
WARNING
Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (≥42V peak) is
present on input/output high, it is also
present on the guard terminals of the
Remote PreAmp and mainframe.
6430
REMOTE
PreAmp
SENSE
MAINFRAME
Connect to
REMOTE PreAmp
connector on rear
panel of mainframe
Guard
IN/OUT
HIGH
Preamp Cable*
KEITHLEY
*Keithley part number: CA-176-1D.
cбЦмкЙ=OJQ
High-impedance
guarding
on the 100µA through 100mA ranges. On the 10µA range and below (when using the
Remote PreAmp), the allowable voltage drop in each source lead is limited as follows:
Current range
10µA
1µ
A
100nA
10nA
1nA
100pA
10pA
1pA
Guarding methods
Cable guard
Use the high-impedance (cable) guard connection scheme shown in Figure 2-4 for the fol-
lowing source-measure condition:
Connections2-9
Maximum allowable resistance
per source lead
10Ω
50Ω
500Ω
5kΩ
50kΩ
1MΩ
50MΩ
1GΩ
•Test circuit impedance is >1GΩ.
Cable guard is always available at the Remote PreAmp. If not using the Remote PreAmp,
use
the V, Ω GUARD banana jack on the mainframe with CABLE guard selected. See Guard
selection to select cable guard.
reduces noise and protects the user from a potentially hazardous voltage on the guard shield (or
A test fixture is typically used when testing hig
plate). See Section 5, Guard for details on using guard.
Note that the test fixtu
h-impedance devices. The test fixture
re chassis is connected to In/Out LO to reduce noise.
2-10Connections
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V, Ω,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCKDIGITAL I/O
REMOTE
PreAmp
!
Triax Cable
HI
DUT
WARNING
Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (≥42V peak) is
present on input/output high, it is also
present on the guard terminals of the
Remote PreAmp and mainframe.
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
LO
Banana Plug Cable
6430 Mainframe (rear panel connectors)
Resistor
Network
Ohms
Guard
I
G
IN/OUT
HIGH
Preamp
Cable*
≥1kΩ
R
G
*Keithley part number: CA-176-1D.
cбЦмкЙ=OJR
Guarded ohms
connections (basic)
Ohms guard
Use ohms guard for the following source-measure operation:
•In-circuit resistance measurements on the D
sent.
are pre
UT where other parasitic leakage devices
Note that ohms guard must be selected for this connection scheme. See Gu
select ohms guard.
klqbOhms guar
d cannot be accessed from the Remote PreAmp. It is only available at the
rear panel of the mainframe.
Figures 2-5, 2-6, and 2-7 show how to make connections to measure the resistance of a
sing
le resistor (DUT) in a resistor network. See Section 4, 6-wire ohms measurements and
Section 5, Guard for more information on guarded ohms measurements.
ard selection to
Connections2-11
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V, Ω,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Triax Cable
HI
DUT
WARNING
Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (≥42V peak) is
present on input/output high, it is
also present on the guard terminals
of the Remote PreAmp and mainframe.
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
LO
Banana Plug Cables
6430 Mainframe (rear panel connectors)
Resistor
Network
Ohms
Guard
I
G
<1kΩ
Guard
Sense
IN/OUT
HIGH
Preamp
Cable*
R
G
*Keithley part number: CA-176-1D.
cбЦмкЙ=OJS
Guarded ohms
connections
(guard sense)
The basic connection scheme for guarded ohms measurements is shown in Figure 2-5. If the
gu
ard resistance path (RG) is <1kΩ, IR drop in the GUARD test lead could be high enough
that the guard voltage at the resistor network is significantly less than the output voltage at the
DUT. This results in leakage current and will corrupt the measurement. To cancel the effect of
IR drop in the GUARD test lead, connect GUARD SENSE to the resistor network as shown
in Figure 2-6. Guard sense ensures that the guard voltage at the resistor network is the same as
the output voltage at the DUT.
2-12Connections
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V, Ω,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Triax
Cable
HI
DUT
WARNING
Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (≥42V peak) is
present on input/output high, it is also
present on the guard terminals of the
Remote PreAmp and mainframe.
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
LO
Banana Plug Cables
6430 Mainframe (rear panel connectors)
Resistor
Network
Ohms Guard
<1kΩ
Guard Sense
Sense HI
Sense LO
<1kΩ
Triax
Cable
IN/OUT
HIGH
Preamp
Cable*
R
G
*Keithley part number: CA-176-1D.
cбЦмкЙ=OJT
Guarded ohms
connections
(6-wire ohms)
sens-ing) by connecting SENSE HI and LO to the DUT (Figure 2-7).
Note that if the DUT is <1kΩ, you should use the 4-wire measurement method (remote
klqbGuard current (IG) must never exceed 50mA. If it does, guard voltage will become
less than the output voltage and corrupt the measurement.
Guard selection
▲
Cable guard is used for high-impedance guarding for cables (i.e., coax and triax) and test
fixtures. Ohms guard provides a high-current guard output, which allows in-circuit guarded
ohms measurements.
The guard setting (cable or ohms) only applies to mainframe guard (V,Ω GUARD banana
jack). Cable guard is always available at the Remote PreAmp, regardless of the guard setting.
Ohms guard is only available at the mainframe. On power-up, the mainframe defaults to cable
guard.
klqb When the guard selection is changed, the OUTPUT will turn OFF.
Front panel guard selection
Perform the following steps to check or change the guard selection:
1.Press CONFIG and then the SOURCE V, SOURCE I, or Ω. Changing guard in one
configuration menu changes it in all of the others.
2.Using the and ▲ keys, place the cursor (flashing menu item) on GUARD and press
ENTER.
klqbCursor position indicates the present guard selection (OHMS or CABLE). To retain
the present selection, use the EXIT key to back out of the menu structure and skip the
next two steps.
Connections2-13
3.To change the guard selection, place the cursor on the alternate se
ER.
ENT
4.Use the EXIT key to back out of the menu structure.
klqbDo
not connect ohms guard using coaxial cabling, or oscillations may occur.
lection and press
Remote command guard selection
Use the :SYSTem:GUARd command in Section 17 to choose between cable and ohms guard
mode via remote. For example, send the following command to enable ohms guard:
:SYST:GUAR OHMS
Conversely, send this command to enable cable guard:
:SYST:GUAR CABL
2-14Connections
3
Basic Source-Measure
Operation
•Operation Overview — Discusses source-measure capabilities, compliance limit, and
fundamenta
l source-measure configuration.
•Operation Considerations —
source delay.
•Basic Source-Measure
SourceMeter for source-mea
output v
turning the
•Me
•Sink Operation —
alues, and compliance limits; choosing measurement range and function; a
output on and off.
asure Only — Covers how to use the SourceMeter for measurements only.
Describes sink operation.
Covers warm-up, auto zero, V-source protection, and
Procedure — Describes the basic procedure for setting up the
sure operations, including selecting the source
function,
nd
3-2Basic Source-Measure Operation
CAUTION
Excessive heat could damage the SourceMeter and at the very least, degrade its perfor-
mance. The SourceMeter must be operated in an environment where the ambient temperature
t exceed 50°C.
does no
The SourceMeter uses a heat sink to dissipate heat.
expose the black, finned heat sink. To prevent damaging heat build-up and thus, ensure specified performance, adhere to the following precautions:
√ Keep the
heat could become impaired.
√ Keep the
feet
SourceMeter when it is sitting
could block the botto
√ Do not position any
unheated)
compro
√ When
e bottom and sides to ensure proper cooling. Adequate airflow enables air temper-
th
atures within approximately one inch of the SourceMeter surfaces
specified
√ Rack mounting high
could cause
maintained around the
√ A
only i
Precision equipment,
the
SourceMeter
heat sink free of dust, dirt, and contaminates, since its ability to dissipate
bottom cooling vents from becoming blocked. NEVER remove the plastic
and place the SourceMeter directly on a flat surface. NEVER operate
on a conformable surface (such as a ca
m cooling vents.
devices adjacent to the SourceMeter that force air (heated or
into or onto its surfaces or cooling vents. This
mise accuracy performance.
rack mounting the SourceMeter, make sure there is adequate airflow around
limits under all operating conditions.
power dissipation equipment adjacent to the SourceMeter
excessive heating to occur. The specified ambient temperatures
surfaces of the SourceMeter to specified accuracies.
good measure to ensure proper cooling in rack situations with convection cooling
s to place the hottest equipment (i.e., power supply) at the top of the
such as the SourceMeter, should be pla
rack where temperatures are the coolest. Adding spacer panels be
will help ensure adequate airflow.
The left side of the case is cut out to
the
rpet). This
additional airflow could
to remain within
must be
rack.
ced as low as possible in
low the
Operation overview
Source-measure capabilities
From the front panel, the SourceMeter can be configured to perform the following
operations:
Basic Source-Measure Operation3-3
•Source v
•Source current
•Me
•Me
Voltage —
1fA to 105mA. The V-Meter can measure voltage from ±1µV to ±211V. The V-Source and
V-Meter each have four ranges; 200mV, 2V, 20V, and 200V.
Current — With the Remote PreAmp, the I-Source can output current from ±0.5fA to
±105mA and limit voltage from 200µV to 210V. The I-Meter can measure current from ±10aA
to ±105.5mA. The I-Source and I-Meter each have 12 ranges; 1pA, 10pA, 100pA, 1nA, 10nA,
100nA, 1µA, 10µA, 100µA, 1mA, 10mA, and 100mA.
Without the Remote PreAmp, the 1pA, 10pA, 100pA, 1nA, 10nA, and 100nA ranges are not
available. Therefore, without the Remote PreAmp, the I-Source can output current from ±50pA
to ±105mA and measure current from ±10pA to ±105.5mA.
Resistance — With the Remote PreAmp, the SourceMeter can make resistance measurements from <100µΩ to >20TΩ. When using the auto ohms measurement method, the following
ohms ranges are available, 20Ω, 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 20MΩ, 200MΩ, 2GΩ, 20GΩ,
200GΩ, 2TΩ, and 20TΩ. When using the manual ohms measurement method, you can-not
select ohms ranges. The ohms reading is the mathematical result of V/I.
Without the Remote PreAmp, the 200MΩ, 2GΩ, 20GΩ, 200GΩ, 2TΩ, and 20TΩ auto ohms
ranges and, as previously pointed out, the six lowest current ranges are not available. Therefore, without the Remote PreAmp, the SourceMeter can measure resistance from <100µΩ to
>20MΩ.
oltage — Display current and/or voltage measurement
— Display voltage and/or current measurement
asure resistance — Display voltage or current component of measurement
asure only (V or I) — Display voltage or current measurement
The V-Source can output voltage from ±5µV to ±210V and limit current from
Measure only
(V-Meter) or an ammeter (I-Meter). When used as a V-Meter only, the I-Source is set to output
0A. When used as an I-Meter only, the V-Source is set to output 0V.
Range of operation — The maximum power output of the SourceMeter is 2.2W. At
maximum source levels, you can output 210V at 10.5mA or 21V at 105mA. The full range of
operation is explained in Section 5, Operating boundaries.
(V or I) — The SourceMeter can be used exclusively as a voltmeter
3-4Basic Source-Measure Operation
klqbLoad regulation – The voltage specification for V-source mode load changes is
0.01% +100µV. This means that on the 200mV range, the load current can be
changed from zero to full scale with less than 1.02mV of error. Calculation:
error = (0.01% × 0.2V) + 100µV
= 0.02mV + 100µV
= 0.12mV
Assuming a 0 to 100mA change in current, the output impedance equates to 1.2mΩ
(0.12mV/100mA = 1.2mW). This level can only be achieved using 4-wire sensing.
Compliance limit
When sourcing voltage, the SourceMeter can be set to limit current (from 1fA to 105mA).
Conversely, when sourcing current, the SourceMeter can be set to limit voltage (from 200µV to
210V). The SourceMeter output will not exceed the compliance limit.
Table 3-1 summarizes compliance limits according to range.
on compliance limits.
q~ÄäÉ=PJN
Compliance limits
Maximum compliance
Measurement range
200mV
002V
020V
200V
001pA*
010pA*
100pA*
001nA*
010nA*
100nA*
001µA
010µA
100µA
001mA
010mA
100mA
value
210mV
002.1V
021V
210V
001.05pA
010.5pA
105pA
001.05nA
010.5nA
105nA
001.05µA
010.5µA
105µA
001.05mA
010.5mA
105mA
See Section 5 for more details
* Only available when using the Remote PreAmp.
Setting the compliance limit
▲
▲
Front panel compliance limit
Set the compliance limit from the front panel as follows:
Basic Source-Measure Operation3-5
1.Select the desired source and measure function
2.Press the EDIT key until the cursor flashes in
3.Select the desired compliance range using the RANGE
4.To increment or decrement the compliance value, use the EDIT and ke
the cursor over the digit to be changed, then press the SOURCE ▲ or ▼ key to increment or decrement the compliance value.
5.To change the compliance value directly, simply enter the value using the numeric ke
while the cursor is flashing in
the compliance display field.
s using the MEAS and SOURCE keys.
the compliance (Compl:) display field.
▲ and ▼ keys.
ys to place
ys
Remote compliance limit
Table 3-2 summarizes basic commands to program the compliance limit. See Section 17, Set
compliance limit for more details on these commands. To program the compliance, simply send
the command using the desired parameter. For example, the following command sets the current compliance to 50mA:
:SENS:CURR:PROT 50E-3
Similarly, the following command sets the voltage compliance to 4V:
Set current compliance (n = compliance).
Set voltage compliance (n = compliance).
3-6Basic Source-Measure Operation
Mainframe
I-Meter
Source
V-Meter
REMOTE
PreAmp
HI
LO
A
IN/OUT
HIGH
cбЦмкЙ=PJN
Fundamental
source-measure
configuration
Basic circuit configuration
The fundamental source-measure configuration for the SourceMeter (with Remote PreAmp)
is shown in Figure 3-1, where the Source is either the V-Source or the I-Source. If not using the
Remote PreAmp, Input/Output HI and LO is accessed at the rear panel of the mainframe.
klqbWhen using the Remote PreAmp, nothing should be connected to INPUT/OUTPUT
HI banana jack on the mainframe. If using mainframe INPUT/OUTUT HI, the cable
to the Remote PreAmp should be disconnected.
Operation considerations
source delay.
Warm-up
rated accuracies. See Appendix A for specifications.
Auto zero
measurements. With auto zero enabled, all three of these measurements are performed for each
reading to achieve rated accuracy. With auto-zero disabled, zero and reference are not mea
sured. This increases measurement speed, but zero drift will eventually corrupt accuracy. With
auto zero disabled, periodically change measurement speed.
The following paragraphs discuss warm-up period, auto zero, V-source protection, and
The SourceMeter must be turned on and allowed to warm up for at least one hour to achieve
Every A/D conversion (reading) is calculated from a series of zero, reference, and signal
-
Basic Source-Measure Operation3-7
Temperature changes across components within the instrument can cause the reference and
zero values for the A/D converter to drift due to thermo-electric effects. Auto zero acts to
negate the effects of drift in order to maintain measurement accuracy over time. Without auto
zero enabled, measurements can drift and become erroneous.
Front panel auto zero
Set the auto zero from the front panel as follows:
1.Press the MENU key.
2.Select A/D CTRL from the main menu, then press ENTER.
3.Select AUTO ZERO, then press ENTER.
4.Select ENABLE or DISABLE as appropriate, then press ENTER.
5.Press EXIT as necessary to return to normal display.
Remote command auto zero
Use the :SYSTem:AZERo command to enable or disable auto zero via remote. For example,
send the following command to disable auto zero:
:SYST:AZER ON
Conversely, send this command to disable auto-zero:
:SYST:AZER OFF
NPLC caching
NPLC caching speeds up source memory sweeps by caching A/D reference and zero values.
When NPLC caching is enabled (using the NPLC-CACHE/ENABLE menu selection), the A/D
reference and zero values will be saved for up to the 10 most recent voltage, current, and resis
tance measurement function settings. Whenever the integration rate is changed via the SPEED
key, a recalled user setup (using the SAVESETUP/RESTORE menu selection), or during a
source memory recall (either with the SOURCE-MEMORY/RESTORE menu or during a
source memory sweep), NPLC caching will occur. If the integration rate is already stored in the
cache, the stored reference and zero values are recalled and used. Otherwise, a reference and
zero value are acquired and stored in the cache. If there are already 10 NPLC values stored, the
oldest one will be overwritten by the newest one.
klqbAuto zero should be disabled for maximum source memory sweep speed; otherwise
-
the cache is of little use. With auto zero enabled, new A/D reference and zero values
are taken for every reading and saved into the cache, slowing down sweep operation.
However, with auto zero disabled, measurements may drift and become erroneous.
To minimize drift when using NPLC caching with auto zero disabled, periodically
select AUTO-ZERO/ONCE in the A/D-CTRL menu to force an immediate auto zero
update.
3-8Basic Source-Measure Operation
NPLC cache setup
Follow the steps below to enable and use NPLC caching with a source memory sweep:
1.Press the MENU key, select
2.Select AUTO-ZERO, then press ENTER.
3.Choose DISABLE, then press ENTER to disable auto zero.
4.From the A/D CONTROLS menu, selec
5.Select ENABLE, then press ENTER to enable NPLC caching.
6.Use the EXIT key to back out of the main menu structure.
7.Set up the source memory parameters, and run the source memory sweep. (See Sour
ory sweep in Section 9.)
mem
A/D-CTRL, then press ENTER.
t NPLC-CACHE, then press ENTER.
ce
Typical NPLC cache test times
Typically, NPLC caching will decrease source memory sweep times by a factor of three.
Table 3-3 shows typical averaged times for a test consisting of 10 sweeps of four source memory locations with NPLC values for successive me
respectively.
q~ÄäÉ=PJP
Typical NPLC cache test times
NPLC cache conditionsTime
Auto Zero OFF, Caching OFF5.89s
Auto Zero ON, Caching OFF5.89s
Auto Zero OFF, Caching ON2.05s
Auto Zero ON, Caching ON5.89s
Auto Zero OFF, Caching ON, Cache empty2.5s
mory locations set to 10, 1, 0.1, and 0.01
V-source protection
Use V-source protection to select the maximum voltage level the SourceMeter can output.
Available limit values include 20V, 40V, 60V, 80V, 100V, 120V, 160V, and NONE (allows voltage >160V). These are absolute values with 5% toleranc
t^okfkdEve
NEVER touch the triax cable(s) or anything connected to the terminals of
the SourceMeter when it is on or connected to an external source. Always
assume that a hazardous voltage (>30V rms) is present when the power is
on.
To prevent damage to DUT (devices under test) or external circuitry, DO
NOT progra
limit.
e. The power-on default is NONE.
n with the voltage protection limit set to the lowest value (20V),
m the V-Source to levels that exceed the voltage protection
Front panel V-source protection
To program V-source protection from the front panel:
1.Press CONFIG then SOURCE V.
2.Select PROTECTION from the displayed choices, then press ENTER.
3.Select the desired protection value, then press ENTER.
4.Press EXIT to return to normal display.
Remote command V-source protection
Use the :SOURce:VOLTage:PROTection command to program the V-source protection
value via remote. See Section 17 for details. For example, the following command sets the pro
tection value to 20V:
:SOUR:VOLT:PROT 20
Source delay
The source delay options are used to set the settling time for the source. This source delay is
the delay phase of the Source-delay-measure cycle. See Section 5. The auto delay period is cur
rent range dependent (Table 3-4).
Basic Source-Measure Operation3-9
-
-
q~ÄäÉ=PJQ
Auto source delay
SourceMeter
I-RangeAuto Delay
1pA1sec
10pA350msec
100pA50msec
1nA35msec
10nA10msec
100nA6msec
1µA5msec
10µA5msec
100µA3msec
1mA3msec
10mA3msec
100mA3msec
Manual delay can be set from 0000.00000 to 9999.99800 seconds. Manually setting the
delay disables auto delay.
3-10Basic Source-Measure Operation
Front panel source delay
To set the manual source delay from the front panel:
1.Press CONFIG then SOURCE V (or SOURCE I).
2.Select DELAY from the displayed choices, then press ENTER.
3.Enter the desired DELAY value, then press ENTER.
4.Press EXIT to return to normal display.
To set the auto source delay state from the front panel:
1.Press CONFIG then SOURCE V (or SOURCE I).
2.Select AUTO DELAY from the displayed choices, then press ENTER.
3.Select ENABLE or DISABLE as desired, then press ENTER.
4.Press EXIT to return to normal display.
Remote command source delay
Use the :SOURce:DELay or :SOURce:DELay:AUTO commands to program the source
delay via remote. See Section 17 for details. For example, the following command sets the
source delay to 500ms:
:SOUR:DEL 0.5
Similarly, send the following command to enable auto delay:
:SOUR:DEL:AUTO ON
Basic source-measure procedure
Output control
Use the ON/OFF OUTPUT key to turn the SourceMeter output on or off for basic sourcemeasure situations. With the output on, the red ON/OFF OUTPUT indicator light will be on.
The indicator light turns off when the output is turned off.
You can also control the output off state (normal, zero, or guard) and program the instrument
for auto output off operation. See Section 12 for complete details on these aspects.
t^okfkdTo prevent electric shock, do not make or break connections to the
SourceMeter while it is on.
Current measurements and capacitive loads
When measuring current in a test circuit that has high capacitance, ringing may occur. Ringing is the fluctuation of current readings that is initiated b
eventually decays to a settled current reading. The higher the capacitance, the more ringing that
will occur.
Table 3-5 lists the maximum capacitive loads that the Model 6430 can accommodate effectively. For the higher current ranges (1nA range a
est capacitance in which ringing created by a voltage step will decay in less than
cycle. For the lower current ranges (100pA and lower), the listed values represent the largest
capacitance that causes ringing that will settle in a reasonable amount of time.
Current measurements can be performed for higher capacitive loads as long as the increased
ringing can be
the expense of longer settling times. See Source capacitance in Appendix F for more
information.
q~ÄäÉ=PJR
Maximum capacitive loads
Current range
100mA0.2µF
100µA0.04µF
100nA100pF
100pA100pF
tolerated. Ringing can be reduced by adding a resistor in a series with the load at
Maximum
Capacitive Load
10mA0.2µF
1mA0.2µF
10µA3300pF
1µA470pF
10nA100pF
1nA100pF
10pA100pF
1pA100pF
nd higher), the listed values represent the larg-
Basic Source-Measure Operation3-11
y a voltage step. This fluctuation
one power line
3-12Basic Source-Measure Operation
▲
▲
Front panel source-measure procedure
Refer to Section 4 to measure ohms.
klqbThe following pr
ocedure assumes that the SourceMeter is already connected to the
DUT as explained in Section 2.
Step 1: Select source.
Press SOURCE V to select the V-Source or press SOURCE I to select the I-Source. The
presently programmed source value (V
SRC
or I
) and compliance level (Cmpl) are
SRC
displayed.
Step 2: Set source level and compliance limit.
The source level is the voltage or current setting of the selected source (V-Source or
I-Source). A compliance limit is set to protect the DUT from damaging currents or vo
using the V-Source, a current compliance is set. When using the I-Source, a vo
When
compliance is set. Complian
ce defines the maximum absolute value the SourceMeter can out-
put.
Note that compliance can also be determined by the measurement range. Depending on
value is lower, compliance occurs at the programmed value (real compliance) or at the
which
maximum compliance value for the present fixed measurement range (range compliance). For
example, with compliance set to 2V and the 200mV measurement range selected, compliance
will occur at 210mV. On the 20V measurement range, compliance will occur at 2V. See
Section 5, Compliance limit for details
klqbThe S
ourceMeter must be in the edit mode (EDIT annunciator ON) to set source and
on real and range compliance.
compliance values. The edit mode is selected by pressing the EDIT key. The flashing
digit for the source or compliance reading indicates that the SourceMeter is in the
edit mode. If no editing operation is performed within six seconds, the edit mode
times out and is cancelled. To return to the edit mode, press EDIT again. While in the
edit mode, the EDIT key toggles between the source value and the compliance value.
The SOURCE
▲ and ▼
and and keys also enable the edit mode. They choose
the last selected field.
When editing the source value, the source is updated immediately, allowing you to
the source value while the output is on.
adjust
The source value cannot be changed while the SourceMeter is performing a sweep.
This oc
curs with Output ON and either the SWEEP key is pressed, Offset Compensa-
tion is enabled under Ohms, or OFF-COMP-OHMS,
VOLT-COEFF, or VAR-
ALPHA functions are enabled.
When editing the compliance value, compliance is not updated until ENTER is
pr
essed or the edit mode is allowed to time out.
EDIT always goes to the source field first, except while sweeping, in which case it
goes into
the compliance field.
ltages.
ltage
Basic Source-Measure Operation3-13
Source and compliance values cannot be edited in AUTO OHMS mode.
Perform the following steps to edit the source and compliance values:
1.Press EDIT to enter the edit mode. The flashing digit indicates which reading (source or
complia
EDIT ag
2.Use the RANGE
w
nce) is presently selected for editing. If you wish to edit the other field, pr
ain.
▲ and ▼ keys to select a range that will accommodate the value you
ant to set. For best accuracy, use the lowest po
ssible source range.
ess
3.Enter the desired source or compliance value. There are two methods to edit the value:
v
alue adjust and numeric entry.
klqbT
o clear the source value to 0V or 0A, press the MENU key while in the edit source
field.
alue adjust — To adjust the value, use the EDIT cursor keys to place the cursor
•V
the appropriate position, and use the SOURCE
at
decreme
•Nume
sign
nt the value.
ric entry — When the edit mode is entered, the cursor is located on the most
ificant digit of the value. From this position, you can key in the value using the
▲ and ▼ keys to increment or
number keys (0 through 9). After each number is keyed in, the cursor move
position
sor on a digit to be changed, and press the appropriate number key.
does no
MENU
to the right. If desired, you can use the EDIT cursor keys to place the cur-
The cursor
t have to be on the polarity sign of the value to change polarity
. If the
key is pressed, the Source Value will be clear to 0V or 0A.
4.To edit the other field, press EDIT to select it, and repeat steps 1 and 2.
5.When finished editing the source and compliance values, press ENTER or wait six seconds to exit from the edit mode.
klqbWhen a
compliance limit value is entered, the SourceMeter automatically goes to the
lowest (most sensitive) compliance range that will accommodate that value.
The lowest compliance levels that can be set are 0.00100pA (1fA) and 000.200mV
(200µV).
Step 3: Select measurement function and range.
Select the desired measurement function by pressing MEAS V (voltage) or MEAS I
(current).
s one
When measuring the source (i.e., Source V Measure V), you cannot se
lect the range using
the measurement RANGE keys. The selected source range determines the measurement range.
When not measuring the source (i.e., Source V Mea
sure I), measurement range selection can
be done manually or automatically. When using manual ranging, use the lowest possible range
for best accuracy. In autorange, the SourceMeter automatically goes to the most sensitive range
to make the measurement.
3-14Basic Source-Measure Operation
Measuring voltage — When sourcing current, you can use the RANGE ▲ and ▼ keys to
manually select the voltage measurement range. You can also press AUTO to select autorang
ing. When sourcing voltage, the RANGE keys are inoperative.
Measuring current — When sourcing voltage, you can use the RANGE ▲ and ▼ keys to
manually select the current measurement range. You can also press AUTO to select autorang
ing. When sourcing current, the RANGE keys are inoperative.
klqbWith the 200V V-Source range selected, the highest current measurement range is
10mA. With the 100mA I-Source range selected, the highest voltage measurement
range is 20V.
Measurement range is also limited by the compliance setting. For example, if voltage
compliance is 1V (2V compliance range), the highest voltage measurement range
that can be selected is 2V.
Step 4: Turn output on.
Turn the output on by pressing the ON/OFF OUTPUT key. The OUTPUT indicator will turn
on to indicate the output is on.
Step 5: Observe readings on the display.
-
-
The SourceMeter is in compliance if the “Cmpl” label or the units label (i.e. “mA”) for the
displayed compliance setting is flashing. If the “Cmpl” label is flashing, real compliance has
occurred. The output is clamped at the displayed compliance value. If the units label is flash
ing, range compliance has occurred. The output is clamped at the maximum compliance value
for the present fixed measurement range. For example, if presently on the 2V measurement
range, a flashing units label for the voltage compliance value indicates that the output is
clamped at 2.1V.
The SourceMeter can be taken out of compliance by going into the edit mode and decreasing
the source value or increasing the compliance value. Note that increasing the compliance limit
may compromise protection for the DUT. If in range compliance, selecting a higher measure
ment range may take the SourceMeter out of compliance.
klqbSee Section 5,“Compliance limit” for details on real and range compliance.
-
-
Step 6: Turn output off.
When finished, turn the output off by pressing the ON/OFF OUTPUT key. The OUTPUT
indicator light will turn off.
Basic Source-Measure Operation3-15
Remote command source-measure procedure
Basic source-measurement procedures can also be performed via remote by sending appropriate commands in the right sequence. The following paragraphs summarize the basic commands and give a simple programming example.
Basic source-measure commands
Table 3-6 summarizes basic source-measure commands. See Section 17 for more information on using these commands.
Select source function (function = VOLTage or CURRent).
Select fixed sourcing mode for I-source.
Select fixed sourcing mode for V-source.
Select I-source range (n = range).
Select V-source range (n = range).
Set I-source amplitude (n = amplitude in amps).
Set V-source amplitude (n = amplitude in volts).
Select measure function (function = “VOLTage” or
“CURRent”).
Set current compliance (n = compliance).
Set voltage compliance (n = compliance).
Set current measure range (n = range).
Set voltage measure range (n = range).
Select output state (state = ON or OFF)
Trigger and acquire reading.
3-16Basic Source-Measure Operation
Source-measure programming example
Table 3-7 summarizes the command sequence for a basic source-measure procedure. Note
that the steps correspond to those listed previously in Front panel source-measure procedure.
These commands set up the SourceMeter as follows:
•Source function: volts
•Source mode: fixed
•Source range: 20V
•Source output level: 10V
•Current compliance: 10mA
•Measure function: current
•Measure range: 10mA
q~ÄäÉ=PJT
Basic source-measure command sequence
Step1ActionCommands
2,3
Comments
*RST
1
Select source function,
mode.
2
Set source range,
level, compliance.
:SOUR:FUNC VOLT
:SOUR:VOLT:MODE FIXED
:SOUR:VOLT:RANG 20
:SOUR:VOLT:LEV 10
:SENS:CURR:PROT 10E-3
3
Set measure function,
range.
4
Turn on output.
5
Read data.
6
Turn off output.
1
Steps correspond to front panel steps listed previously in Front panel source-measure procedure.
2
Commands must be sent in order given.
3
Instrument must be addressed to talk after :READ? to acquire data.
:SENS:FUNC “CURR”
:SENS:CURR:RANG 10E-3
:OUTP ON
:READ?
:OUTP OFF
Restore GPIB defaults.
Select voltage source.
Fixed voltage source mode.
Select 20V source range.
Source output = 10V.
10mA compliance.
Current measure function.
10mA measure range.
Output on before measuring.
Trigger, acquire reading.
Measure only
Front panel measure only
In addition to being used for conventional source-measure operations, the SourceMeter can
also be used to measure only voltage or current. Perform the following steps to use the
SourceMeter to measure voltage or current:
1.Select source-measure functions.
2.Set source and compliance levels.
Basic Source-Measure Operation3-17
Measure voltage only (voltmeter) —
press MEAS V
Measure current only (ammeter) —
press MEAS I
Use the editing procedure provided in step 2 of Basic sour
the source and compliance levels as follows:
a.Select the lowest source range and set the source level to zero (0.00000pA
000.000mV).
b.Set compliance to a level that is higher than the expected measurement.
to select the voltage measurement function.
to select the current measurement function.
Press SOURCE I to select the I-Source, and
Press SOURCE V to select the V-Source, and
ce-measure procedure to edit
or
`^rqflkWhen using th
higher than the voltage that is being measured. Failure to do this could
result in instrument damage due to excessive current that will flow into the
SourceMeter.
3.Select range.
Use the RANGE ▲ and ▼ keys to select a fixed measurement range that will accommodate the expected reading. Use the lowest possible range for best accuracy.
When measuring current, AUTO range can be used instead. Th
matically go to the most sensitive range. W
range. See the following CAUTION.
`^rqflkWhen using th
range and NEVER select a measurement range that is below the applied
signal level. For these conditions, high current will be drawn from the
external source. This high current could damage the external source or test
circuit.
4.Connect voltage or current to be measured. Con
2-wire c
5.Turn output on. Press the ON/OFF key to turn the output on.
6.Take a reading from the display.
7.When finished, turn output off.
onnections. See Figure 2-2.
e SourceMeter as a voltmeter, V-Compliance must be set
e SourceMeter will auto-
hen measuring voltage, DO NOT use AUTO
e SourceMeter as a voltmeter only, DO NOT use AUTO
nect the DUT to the SourceMeter using
3-18Basic Source-Measure Operation
Remote command measure only
Table 3-8 summarizes the basic command sequence for measure only. The steps outlined
correspond to those in the Front panel measure only sequence above. These commands set up
the SourceMeter for measure only voltage measurements up to 20V as follows:
:SOUR:CURR:MODE FIXEDFixed current source mode.
:SENS:FUNC “VOLT”Volts measure function.
2Set source and compliance.:SOUR:CURR:RANG MINLowest source range.
:SOUR:CURR:LEV 00μA
:SENS:VOLT:PROT 2525V compliance.
3Select volts measure range.:SENS:VOLT:RANG 2020V range.
5Turn on output.:OUTP ONOutput on before measuring.
6Read data.:READ?Trigger, acquire reading.
7Turn off output.:OUTP OFFOutput off after measuring.
1
Steps correspond to front panel steps listed previously in Front panel measure only. DUT should be connected to SourceMeter before
running program.
2
Commands must be sent in order given.
3
Instrument must be addressed to talk after :READ? to acquire data.
2,3
Comments
source level.
Sink operation
Overview
When operating as a sink (V and I have opposite polarity), the SourceMeter is dissipating
power rather than sourcing it. An external source (i.e., battery) or an energy storage device (i.e.,
capacitor) can force operation into the sink region.
For example, if a 12V battery is connected to the V-Source (In/Out HI to battery high) that is
programmed for +10V, sink operation will occur in the second quadrant (Source +V and
measure -I).
`^rqflkWhen using the I-Source as a sink, ALWAYS set V-Compliance to a level
klqbThe sink operating limits are shown in Section 5, “Operating boundaries.”
Sink programming example
Basic Source-Measure Operation3-19
that is higher than the external voltage level. Failure to do so could dam
age the instrument due to excessive current that will flow into the
SourceMeter.
-
Table 3-9 lists a command sequence to program the SourceMeter for sink operation. These
commands set up the unit as follows:
•Source function: volts
•Measure function: current
•Source voltage: 0V
•Measure range: auto
•Compliance (discharge current): 100mA
q~ÄäÉ=PJV
Sink programming example
CommandDescription
*RST
:SOUR:FUNC VOLT
:SOUR:VOLT:MODE FIXED
:SENS:FUNC “CURR”
:SENS:CURR:RANG:AUTO ON
:SENS:CURR:PROT 100E-3
:OUTP ON
:READ?
Restore GPIB defaults.
V-source function.
Fixed source mode.
Current measure function.
Auto measure range.
100mA compliance (discharge current).
Turn on output.
Trigger and acquire reading.
3-20Basic Source-Measure Operation
4
Ohms Measurements
•Ohms Configuration Menu — Outlines the ohms configuration menu that allows you
to set up various ohms measurement aspects.
•Ohms
•Ohms Sensin
•Offset-compensated ohms — Describes of
•Ohms Sourc
•6-wire
•Remote Ohms
Measurement Methods — Discusses auto and manual ohms measurement
methods
to
SourceMeter for 6
netw
program the
ming examples.
and how to select them.
g — Covers 2-wire and 4-wire ohms sensing.
fset-compensated ohms, which can be used
overcome the effects of offsets when making low-resistance measurements.
e Readback — Covers enabling and disabling ohms source readback.
Ohms Measurements — Describes the basic procedure for setting up the
-wire ohms measurement, which can be used for measuring
orks and hybrid circuits.
Programming — Summarizes the basic remote commands required to
SourceMeter for ohms measurements and gives several typical program-
resistor
4-2Ohms Measurements
CONFIG
Ω
SOURCE
MANUALAUTO
GUARD
CABLEOHMS
OFFSET
COMPENSATION
DISABLEENABLE
SRC
RDBK
DISABLEENABLE
cбЦмкЙ=QJN
Ohms configuration
menu tree
Ohms configuration menu
Press CONFIG then Ω to access the ohms configuration menu. Use the Rules to navigate
menus in Section 1 to select the various items in the menu tree, which is shown in Figure 4-1.
Menu items include:
•SOURCE — Select AUTO or MANUAL source mode.
•GUARD — Choo
•SRC RDBK
•OFFSET COMPENSA
The following paragraphs discuss each of these aspects in detail.
se OHMS or CABLE guard.
— Enable or disable source readback mode.
TION — Enable or disable offset-compensated ohms.
Ohms measurement methods
There are two methods to measure ohms: auto ohms and manual ohms. When using auto
ohms, the SourceMeter operates as a conventional constant current source ohmmeter. To use
this method, simply select an ohms measurement range (or use autorange), and take the reading
from the display. When using auto ohms, the default test current varies with the ohms range, as
summarized in Table 4-1.
Ohms Measurements4-3
klqbY
With the manual ohms mode you can select either source V or source I to make ohms measurements, and the unit will automatically compute the resistance reading using the V/I measurement method. After configuring the desired source and selecting a voltage or current
measuring range, select the Ω measurement method to display the calculated V/I ohms
reading.
klqb To achieve optimum accuracy, the SourceMeter measures both V and I and uses
q~ÄäÉ=QJN
Auto ohms default test currents
Auto ohms range*Default test current
ou cannot change the test current in the auto ohms mode. If you attempt to change
the source current in auto ohms, the SourceMeter will display an error message.
these values in ohms calculations (with source readback enabled). The measured
source value is more accurate than the programmed source value. For remote opera
tion, the user specifies the functions to measure. See the resistance measurement
racy specifications in Appendix A for more information.
accu
020Ω
200Ω
002kΩ
020kΩ
200kΩ
002MΩ
020MΩ
200MΩ
002GΩ
020GΩ
200GΩ
002TΩ
020TΩ
100mA
010mA
001mA
100mA
010mA
001mA
001mA
100nA
010nA
001nA
100pA
010pA
001pA
-
*200MΩ maximum without PreAmp
4-4Ohms Measurements
Selecting ohms measurement method
On power-up, auto ohms is the default method for the ohms function. Perform the following
steps to check and/or change the ohms measurement method:
1.Press CONFIG and then Ω to display the ohms configuration menu.
2.Using left and right arrow EDIT keys, place the cursor (flashing menu item) on
SOURCE and press ENTER.
klqbCursor position indicates the presently selected ohms measurement method. To
retain this selection, use the EXIT key to back out of the menu structure and skip the
next two steps.
3.To change the measurement method, place the cursor on the alternate selection (AUT
MANUAL), and press ENTER.
or
4.Press EXIT to exit from the menu
Auto ohms measurements
Perform the following steps to perform auto ohms measurements.
klqbThe following pr
DUT as explained in Section 2.
t^okfkdT
1.Select ohms measurement function.
Press MEAS Ω to select the ohms measurement function.
2.Select auto ohms measurement method.
•Press CONFIG then Ω.
•Select SOURCE, then press ENTER.
•Select AUTO, then press ENTER.
•Press EXIT to return to normal display.
Note that the SourceMeter will be configured to Source I and Measure V. Also, the
I-Source level and V-Compliance limit are based on the measurement range and cannot
be edited.
o prevent electric shock, do not make or break connections to the
SourceMeter with the output on. If on, press the ON/OFF OUTPUT key to
turn the output off.
O
structure.
ocedure assumes that the SourceMeter is already connected to the
klqbUse the manual ohms mode and the V-source method when high-speed settling is
required.