Tektronix 617 Reference manual

MODEL 617
PROGRAMMABLE ELECTROMETER
QUICK REFERENCE GUIDE
INTRODUCTION
This reference and programming guide contains condensed specifica­tions, descriptions of the various features and information for using the Model 617.
Included with descriptions of each feature are brief operating instruc­tions and some examples of typical uses.
Included with the programming information are several example pro­grams utilizing some commonh, used controllers to get the Model 617 “up and running” on the IEEE-488 bus.
0 1984, Keithlay Instruments, Inc.
Cleveland, Ohio, U.S.A.
Document Number 617903-N Rev. A
1
CONTENTS
CONDENSED SPECIFICATIONS MODEL617FEATURES IEEE-488 PROGRAMMING PROGRAMS.
DATA STORE OPERATION
USING HP-85 IBM PC orXT Personal Computer. [Using Keiffiley Model 8573 Interface) APPLE II {Using Apple Interface).
HP-s5...............-.--------~-.-*..-~---..........-
HP-9825A _........~..........._.._.........__...~.__~ _ 47
HP.9816..........-.......-....-.~......~...--.....-.~. 48
DECLSIll ............................................
PETlCBM 2001
...........................................
.........................................
...........................
...................................
...............................
..............................
.........................
.......................... 45
3
7 35 41
42
43
.
46
49
51
2
CONDENSED SPECIFICATIONS
I
INPUT IMPEDANCE: >2OOTG in parallel with 2OpF (~2pF guarded).
Range Resolution
2PJ4 O.lfA
20 PA 1 fA
200 pA 10 fA
2nA 1oofA
20 nA 1 PA
2oanA 10 pA
2fi loopA
2:: 1 10 nA nA
2mA
20mA
INPUT VOLTAGE BURDEN: <lmV except 3mV on 2OmA mnge.
VOLTS
AMPS
1 year, 18-28X
f(%rdg + counts)
Accurecy
1.6 % + 66
1.6 % + 7
1.6 % + 1
0.25% + 5
0.25% + 1
0.25% + 1
0.15% + 4
0.15% 0.15% + + 1 1
0.15% + 4
0.15% t 1
I
3
OHMS
Accuracy
Range
aokn 10 n
200Mn 10 kn
2ooGn IOMQ
MAXIMUM OPEN CIRCUIT VOLTAGE: 300VDC.
Resolution
2kn 0.1 n
20 kn 1 n
2;:
2Gn loo kn
2cGn IMn
loo n 0.25% t 1
IkQ
1 year, 18-W??. Current
f (%rdg t counts)
0.2 % t 4 1fJW
0.15% + 1 l@W
0.25% + 1 IOd
0.25% + 1 IOOIIA
0.3 % t 1 IonA
1.5 % t 1 1flA
1.5 % + 1 IllA
1.5 % + 1 lti
Test
fl.576
Ifi
VOLTAGE SOURCE
OUTFWT: - 102V to t 102V in 50mV steps. ACCURACY: 1 year, 18+.WC; f(0.2% t 5&W)
MAXIMUM OUTPUTCURRENT: &2mA, active current limit at <4mA
with annunciation.
4
GENERAL
CONVERSION TIME: 33omsec. MAXIMUM INPUT: 25W peak, DC to 80Hz sinewave, 10s per
minute max on m4 range. ELECTROMETER MAX COMMON-MODE VOLTAGE: 5OOV peak. V-SOURCE MAX COMMON-MODE VOLTAGE: IOOV psak. ENVIRONMENT:
Operating: O°C to 50°C. 70% relative humidity up to 35OC.
storage: -2FPc to t65oc. WARMUP: 1 hour to rated accuracy.
ENGINEERING UNITS CONVERSION
PREFIX 1
femto-
pico-
nano-
micro-
milli-
kilo-
mega-
giga­tera­oeta-
5
EXPONENT
IO- ‘6 lo- ‘2 10-s 10-e 10-3 103 106 109 10’2 10’5
SAFETY PRECAUTIONS
1. Before operation, ground the instrument through a properly earth grounded power receptacle.
2. Before servicing, disconnect the instrument from the power line, all other equipment and consult the Model 617 instruction Manual.
3. Do not touch any terminals while the instrument is turned on or con­nected to any other test equipment.
DISPLAY ANNUNCIATORS
Readings can be displayed with either alpha exponents or numeric ax­ponents. A front panel program toggles the exponent mode (see PRO­GRAM FEATURES).
ALPHA EXPONENTS mV or V: millivolts or volts.
kfl, MQ, Gg, M or PG: kilohms, megohms, gigohms, teraohms or petaohms.
pA, nA, fl or mA: picoamps, nanoamps, microamps or milliamps.
nC or PC: nanocoulombs or picocoulombs.
NUMERIC EXPONENTS Positive Exponents: 0 through 18.
Negative Exponente: -01 through - 12. Example: A displayed exponent of - 12 = lo- 12.
Display Status-Three indicator lights define what is being read on the display.
METER Indicator Light-Indicates that electrometer readings are be­ing displayed.
SOURCE Indicator Light-Indicates that the voltage source W-SOURCE) value is being displayed.
DATA Indicator Light-Indicates that the data store is in the recall mode end a previously stored reading is being displayed. A flashing DATA light indicates that the data store is full.
NOTE: The DISPLAY button toggles the display between source and the cur­rent mode (METER or DATA).
IEEE Status Indicators-The TALK, LISTEN and REMOTE indicators
are used when programming the instrument over the IEEE-488 bus.
These status lights are not operational during front panel instrument
operation.
DISPLAY MESSAGES
The following massages will be displayed to flag errors or conditions as they occur:
MESSAGE
OL r.r.
n Err
b Err
t EN
. .
IEEE “address” Program message; in IEEE ad-
dlSP Program message; in display
CAL
31inking AMPS light 3linking ZERO CHECK light In V/I zero check.
DESCRIPTION Overrange input applied. Displayed briefly on power up;
RAM test passed. Displayed con­stantly; RAM test failed.
Number Error. Calibration or voltage source value exceeds limits.
Bus Error. Instrument pro­grammed while not in remote or
an illegal command (IDDC) sent or an illegal command option UDDCO) sent.
Trigger Overrun Error. Instrument triggered while processing reading from previous trigger.
flashing decimal points on expo­nent LEDs; not using stored cal constants. Constant decimal points; instrument in calibration male
dress program.
mode program. Program massage; in cal constant
program. In V/I current overload.
8
SHIK FEATURE
DESCRIPTION
Several of the front panel buttons have a secondary function. These functions are labeled on the front panel in the same color as the shit? button. A secondary function is selected by first pressing the SHIFT button and then pressing the appropriate button. More information on these secondary functions can be found in the following text.
ZERO CHECK
and ZERO CORRECT
ZERO CHECK
ZERO CORRECT
DESCRIPTION
The zero check feature provides a means for internal zero verification. The zero correct feature, when used with zero check, corrects internal offsets in the instrument 50 that the display raads zero. In the zero check mode, the input impedance of the Model 617 is changed as outlined in the following table.
INPUT IMPEDANCE WITH ZERO CHECK ENABLED
IFUNCTION I RANGE 1 IMPEDANCE IN ZERO CHECK
Volts Reistance
Coulombs Current
All All kg, 2MG
20MQ, 2OOMS, 9.09MG II 47pF All GG
All 1OMQ I 102OpF All pA 1OMg II 25pF
All nA 9.09Mn II 47pF AlI fi
All mA
IOMg II 2CpF SOSkG U 112OpF
99kn II 102OpF lOOn II 2OpF
OPERATION
Zero Verification and Correction
1. Allow the instrument to warm up for one hour after power up.
2. Turn zero check on by pressing the ZERO CHECK button. The ZERO CHECK indicator light will turn on.
3. Select the function and range that will be used. If unsure of range, select the lowest range.
4. If the display does not read zero, press the ZERO CORRECT button. The ZERO CORRECT indicator light will turn on and the display will read zero.
5. Take the instrument out of zero check by again pressing the ZERO CHECK button. The ZERO CHECK indicator light will turn off.
8. The instrument is now zeroed and ready to make measurements.
NOTES
1. The instrument can be rezeroad at any time by turning zero correct off (ZERO CORRECT indicator light off) and repeating steps 1 thmugh 5.
2. Correction is automatically scaled when range is changed. However, it is recommended that zero be checked after range or function is changed. If range or function is to be changed without rezeroing, in­itially zero correct on the lowest range of the function to be used.
3. Check zero every 24 hours, or whenever there is a > 1 OC temperature change.
4. On power up, zero check is on.
TYPICAL USE
Zero the 2pA range to ensure accurate measurements on that range.
11
MEASUREMENT MODES
VOLTS
I
OHMS
El
1 AMPS
El
DESCRIPTION
These buttons place the Model 617 in the desired electrometer measure­ment mode. The VOLTS and OHMS buttons have sacondan/functions: external feedback and VII modes respectively. In the external feedback mode, the internal feedback elements are electrically removed from the circuit so that an external element can be used instead. The VII mode utilizes the V-Source to extend the resistance measurement capabilities of the Model 617 to lOtaD. In this mode the Model 617 automatically calculates the resistance from the programmed voltage value and measured current, and then displays it.
OPERATION
Volts, Ohms and Amps:
1. Turn zero check on.
2 Determine the signal to be measured.
cl
V/I
clcl
q
12
3. Press the appropriate function button.
4. Select an appropriate range or autorange.
5. Zero correct the instrument if it is not reading zero.
6. Apply the signal to the triax input connector on the mar panel.
7. Turn zero check off.
Coulombs:
1. Turn zero check on.
2. Press the COUL button.
3. Select an appropriate range or autorange.
4. Zero correct the instrument if it is not reading zero.
5. Turn zero check off.
6. Turn suppress on to cancel the small amount of charge error in­troduced by turning zero check off.
7. Apply the signal to the triax input connector on the rear panel.
‘GRN
I
“SUPPLIED INPUT CABLE
Fig 1. Voltage Resistance, Current and Charge
Measurements
VII:
1. Turn zero check on.
MODEL 617
2. Select the current function and an appropriate range fautorenge suggested).
3. Zero correct the instrument lf it is not reading zero.
4. Set the voltage source (V-Source) to an appropriate level. (See page 20 for V-Source operation).
5. Connect the resistance to be measured to the Model 617 as shown.
6. Press the V-Source OPERATE button (OPERATE light on).
7. Turn zero check off.
13
8. Note the current reading on the display. If this current is to be sup pressed. press the SUPPRESS button (SUPPRESS light on).
9. Select the V/I function by pressing the SHIFT button and then the OHMS button. The V/I indicator light will turn on.
10. If the displayed resistance measurement is to be suppressed, press the SUPPRESS button. Otherwise, take the resistance measure­ments, in ohms, from the display.
*NOT USED
“SUPPLIED INPUT CABLE
SOURCE VOLTAGE
RX = MEASURED CURRENT
Pig 2. V/I Measurements
NOTES
1. To exit the VII mode, press any function button.
2. Suppress can be used with V/I ohms in two different ways. If sup-
press is enabled with the instrument in the V/i ohms mode, the displayed resistance value (not the current being measured) will be
suppressed. In this case, the instrument can be toggled between
amps and V/I ohms without cancelling the suppressed resistance value. However, suppress will be temporarily ignored by switching to amps, but it will be reinstated by returning to V/I ohms.
The current can be suppressed by placing the Model 617 in the amps
mode and then enabling suppress. Entering the V/I ohms mode after suppress is enabled will not cancel the suppressed currant, and the displayed resistance value will reflect the suppressed current (not a suppressed resistance). Again, the instrument can be toggled bet­ween VII ohms and amps without cancelling suppress, but the sup­pressed value will be a current for both amps and VII ohms under
these conditions.
14
RANGE
RANGE
I
A
AUTO
III
DESCRIPTION
The range buttons select the desired range of measurement for the Model 617. The range may be selected manually or automatically. With manual range selection always use the range that allows the greatest amount of resolution.
OPERATION
AUTO The AUTO button toggles autoranging and manual ranging. The instru­ment is in autoranging when the AUTO indicator light is on. If the light is off, press the AUTO button once to get into the autorange mode.
15
Loading...
+ 37 hidden pages