This reference and programming guide contains condensed specifications, descriptions of the various features and information for using the
Model 617.
Included with descriptions of each feature are brief operating instructions and some examples of typical uses.
Included with the programming information are several example programs utilizing some commonh, used controllers to get the Model 617
“up and running” on the IEEE-488 bus.
1. Before operation, ground the instrument through a properly earth
grounded power receptacle.
2. Before servicing, disconnect the instrument from the power line, all
other equipment and consult the Model 617 instruction Manual.
3. Do not touch any terminals while the instrument is turned on or connected to any other test equipment.
DISPLAY ANNUNCIATORS
Readings can be displayed with either alpha exponents or numeric axponents. A front panel program toggles the exponent mode (see PROGRAM FEATURES).
ALPHA EXPONENTS
mV or V: millivolts or volts.
kfl, MQ, Gg, M or PG: kilohms, megohms, gigohms, teraohms or
petaohms.
pA, nA, fl or mA: picoamps, nanoamps, microamps or milliamps.
nC or PC: nanocoulombs or picocoulombs.
NUMERIC EXPONENTS
Positive Exponents: 0 through 18.
Negative Exponente: -01 through - 12.
Example: A displayed exponent of - 12 = lo- 12.
Display Status-Three indicator lights define what is being read on the
display.
METER Indicator Light-Indicates that electrometer readings are being displayed.
SOURCE Indicator Light-Indicates that the voltage source
W-SOURCE) value is being displayed.
DATA Indicator Light-Indicates that the data store is in the recall
mode end a previously stored reading is being displayed. A flashing
DATA light indicates that the data store is full.
NOTE:
The DISPLAY button toggles the display between source and the current mode (METER or DATA).
IEEE Status Indicators-The TALK, LISTEN and REMOTE indicators
are used when programming the instrument over the IEEE-488 bus.
These status lights are not operational during front panel instrument
operation.
DISPLAY MESSAGES
The following massages will be displayed to flag errors or conditions as
they occur:
MESSAGE
OL
r.r.
n Err
b Err
t EN
. .
IEEE “address” Program message; in IEEE ad-
dlSP Program message; in display
CAL
31inking AMPS light
3linking ZERO CHECK light In V/I zero check.
DESCRIPTION
Overrange input applied.
Displayed briefly on power up;
RAM test passed. Displayed constantly; RAM test failed.
Number Error. Calibration or
voltage source value exceeds
limits.
Bus Error. Instrument programmed while not in remote or
an illegal command (IDDC) sent or
an illegal command option
UDDCO) sent.
Trigger Overrun Error. Instrument
triggered while processing reading
from previous trigger.
flashing decimal points on exponent LEDs; not using stored cal
constants. Constant decimal
points; instrument in calibration
male
dress program.
mode program.
Program massage; in cal constant
program.
In V/I current overload.
8
SHIK FEATURE
DESCRIPTION
Several of the front panel buttons have a secondary function. These
functions are labeled on the front panel in the same color as the shit?
button. A secondary function is selected by first pressing the SHIFT
button and then pressing the appropriate button. More information on
these secondary functions can be found in the following text.
ZERO CHECK
and ZERO CORRECT
ZERO CHECK
ZERO CORRECT
DESCRIPTION
The zero check feature provides a means for internal zero verification.
The zero correct feature, when used with zero check, corrects internal
offsets in the instrument 50 that the display raads zero. In the zero check
mode, the input impedance of the Model 617 is changed as outlined in
the following table.
INPUT IMPEDANCE WITH ZERO CHECK ENABLED
IFUNCTION I RANGE 1 IMPEDANCE IN ZERO CHECK
Volts
Reistance
Coulombs
Current
All
All kg, 2MG
20MQ, 2OOMS, 9.09MG II 47pF
All GG
All 1OMQ I 102OpF
All pA 1OMg II 25pF
All nA 9.09Mn II 47pF
AlI fi
All mA
IOMg II 2CpF
SOSkG U 112OpF
99kn II 102OpF
lOOn II 2OpF
OPERATION
Zero Verification and Correction
1. Allow the instrument to warm up for one hour after power up.
2. Turn zero check on by pressing the ZERO CHECK button. The ZERO
CHECK indicator light will turn on.
3. Select the function and range that will be used. If unsure of range,
select the lowest range.
4. If the display does not read zero, press the ZERO CORRECT button.
The ZERO CORRECT indicator light will turn on and the display will
read zero.
5. Take the instrument out of zero check by again pressing the ZERO
CHECK button. The ZERO CHECK indicator light will turn off.
8. The instrument is now zeroed and ready to make measurements.
NOTES
1. The instrument can be rezeroad at any time by turning zero correct
off (ZERO CORRECT indicator light off) and repeating steps 1
thmugh 5.
2. Correction is automatically scaled when range is changed. However,
it is recommended that zero be checked after range or function is
changed. If range or function is to be changed without rezeroing, initially zero correct on the lowest range of the function to be used.
3. Check zero every 24 hours, or whenever there is a > 1 OC temperature
change.
4. On power up, zero check is on.
TYPICAL USE
Zero the 2pA range to ensure accurate measurements on that range.
11
MEASUREMENT MODES
VOLTS
I
OHMS
El
1 AMPS
El
DESCRIPTION
These buttons place the Model 617 in the desired electrometer measurement mode. The VOLTS and OHMS buttons have sacondan/functions:
external feedback and VII modes respectively. In the external feedback
mode, the internal feedback elements are electrically removed from the
circuit so that an external element can be used instead. The VII mode
utilizes the V-Source to extend the resistance measurement capabilities
of the Model 617 to lOtaD. In this mode the Model 617 automatically
calculates the resistance from the programmed voltage value and
measured current, and then displays it.
OPERATION
Volts, Ohms and Amps:
1. Turn zero check on.
2 Determine the signal to be measured.
cl
V/I
clcl
q
12
3. Press the appropriate function button.
4. Select an appropriate range or autorange.
5. Zero correct the instrument if it is not reading zero.
6. Apply the signal to the triax input connector on the mar panel.
7. Turn zero check off.
Coulombs:
1. Turn zero check on.
2. Press the COUL button.
3. Select an appropriate range or autorange.
4. Zero correct the instrument if it is not reading zero.
5. Turn zero check off.
6. Turn suppress on to cancel the small amount of charge error introduced by turning zero check off.
7. Apply the signal to the triax input connector on the rear panel.
‘GRN
I
“SUPPLIED INPUT CABLE
Fig 1. Voltage Resistance, Current and Charge
Measurements
VII:
1. Turn zero check on.
MODEL 617
2. Select the current function and an appropriate range fautorenge
suggested).
3. Zero correct the instrument lf it is not reading zero.
4. Set the voltage source (V-Source) to an appropriate level. (See page
20 for V-Source operation).
5. Connect the resistance to be measured to the Model 617 as shown.
6. Press the V-Source OPERATE button (OPERATE light on).
7. Turn zero check off.
13
8. Note the current reading on the display. If this current is to be sup
pressed. press the SUPPRESS button (SUPPRESS light on).
9. Select the V/I function by pressing the SHIFT button and then the
OHMS button. The V/I indicator light will turn on.
10. If the displayed resistance measurement is to be suppressed, press
the SUPPRESS button. Otherwise, take the resistance measurements, in ohms, from the display.
*NOT USED
“SUPPLIED INPUT CABLE
SOURCE VOLTAGE
RX = MEASURED CURRENT
Pig 2. V/I Measurements
NOTES
1. To exit the VII mode, press any function button.
2. Suppress can be used with V/I ohms in two different ways. If sup-
press is enabled with the instrument in the V/i ohms mode, the
displayed resistance value (not the current being measured) will be
suppressed. In this case, the instrument can be toggled between
amps and V/I ohms without cancelling the suppressed resistance
value. However, suppress will be temporarily ignored by switching to
amps, but it will be reinstated by returning to V/I ohms.
The current can be suppressed by placing the Model 617 in the amps
mode and then enabling suppress. Entering the V/I ohms mode after
suppress is enabled will not cancel the suppressed currant, and the
displayed resistance value will reflect the suppressed current (not a
suppressed resistance). Again, the instrument can be toggled between VII ohms and amps without cancelling suppress, but the suppressed value will be a current for both amps and VII ohms under
these conditions.
14
RANGE
RANGE
I
A
AUTO
III
DESCRIPTION
The range buttons select the desired range of measurement for the
Model 617. The range may be selected manually or automatically. With
manual range selection always use the range that allows the greatest
amount of resolution.
OPERATION
AUTO
The AUTO button toggles autoranging and manual ranging. The instrument is in autoranging when the AUTO indicator light is on. If the light is
off, press the AUTO button once to get into the autorange mode.
15
MANUAL
The instrument is in manual range when the AUTO indicator light is off.
If the light is on, press the AUTO button once to get into the manual
ranging mode. To uprange, press the button showing the arrowhead
pointing up. To downrange press the button showing the arrowhead
pointing down.
16
SUPPRESS
DESCRIPTION
The suppress feature serves as a means of baseline suppression allowing
a stored offset value to be subtracted from subsequent readings. When
the SUPPRESS button is pressed, the instrument triggers a conversion
and stores the reading as a baseline value. All subsequent readings
represent the differences between the applied signal level and the stored
baseline. The stored baseline can be as small as the resolution of the in-
strument will allow or as large as full range.
A suppress level can be established for any measurement function and is
effective only on that function. Changing functions, except between
amps and VII (see V/I operation), will cancel the suppress level already
established. Once a suppress level is established for a measurement
function, that stored level will be scaled according to what range the
Model 617 is on.
It is important to note that the suppress mode reduces the dynamic
range of measurement by that stored level. For example, assume that
the suppress level is + lfi end the w range is selected. The maximum
positive displayed reading, before overranging, would be +0.9999fi.
This is because the A/D converter would be seeing 1.9999fi (maxi-
mum) from the input. The dynamic range of measurement is
-1.9999fl to +0.9999@ (2.9998fi) as compared to the normal
-1.9999fi to + 1.9999fi (3.9998rN span. The dynamic range of
measurement has been reduced by lfi (10,060 counts). The effects of
dynamic range can reduced by selecting a higher range or using
autorange.
17
OPERATION
1. Select function and range.
2. Input the offset or baseline level to the Model 617.
3. Press the SUPPRESS button. The SUPPRESS indicator light will
turn on and the display will zero.
4. The stored level will be algebraically subtracted from all subsequent
measurements.
NOTE: Suppress can be turned off by again pressing the SUPPRESS
3. Null out circuit outputs to display only drift or changes.
18
TRIG (TRIGGER)
DESCRIPTION
As a bench instrument, the Model 617 has two trigger modes; con-
tinuous and single. In the continuous mode, the instrument constantly
triggers three readings per second. In the single mode, one reading is
triggered every time the TRIG button is pressed or an external trigger is
received. When the SGL light is off, the Model 617 is in the continuous
trigger mode. When the light is on, the instrument is in the single trigger
mode.
OPERATION
Changing From Continuous To Single Trigger:
1. Press the SHIFT button and then the TRIG button. The SGL indicator
light will turn on.
2. To trigger a reading, press the TRIG button once. The display will
blank momentarily to indicate that one reading was triggered.
Changing Fmm Single To Continuous Trigger:
1. Press the SHIFT button and then the TRIG button. The SGL indicator
light will turn off.
2. Readings are now triggering continuously.
TYPICAL USE
The front panel TRIG button can be used to prompt another instrument
to perform e specific function, such as outputting a current or voltage
level, or changing channels on a scanner. To use in this configuration,
the METER COMPLETE OUTPUT of the Model 617 must be connected
to the external trigger input of the other instrument.
V-SOURCE
V-SOURCE
DISPLAY
I
ADJUST
El
A
OPERATE
rl
7mA MAX
cl
DESCRIPTION
This feature delivers - 102.35V to + 102.4OV at a maximum of f2mA
to the V-SOURCE OUTPUT binding posts on the mar panel of the
Model 617.
OPERATION
Hazardous voltages may be present at the output.
Make sure that the OPERATE indicator light is off
before connecting the V-SOURCE OUTPUT to a
load.
WARNING
20
V-SOURCE OUTPUT
I
1 PEAK
1. Connect a suitable load to the V-SOURCE OUTPUT connectors.
2. Press the DISPLAY button and verify that the SOURCE indicator
light is on. This button toggles the meter and source display modes.
If the DATA indicator light is on (not blinking), the
DISPLAY button will toggle the display between source
and data display modes.
3. Using the two ADJUST buttons, set the displayed value to read the
desired voltage level. The reading adjust rate will increase if an ADJUST button is held in. For an even faster adjust rate, press the
SHIFT button and then hold in an ADJUST button.
4. Press the OPERATE button to output the displayed voltage to the
load. The OPERATE indicator light will turn on.
5. To turn off the V-Source function, press the OPERATE button again.
The OPERATE light will turn off and the output will be held at OV
25OmV.
NOTES
1. Current limit will occur at lass than 4mA and will cause the OPERATE
indicator light to flash.
2. On power up, V-Source will be set to OO.OW.
103V MAX
NOTE
21
TYPICAL USE
1. The V-Source is used with the VII function to extend the resistance
measurement capabilities of the Model 617 to lOl%.
2. The V-Source can also be used to deliver test voltages to external circuits.
22
DATA STORE with MIWMAX
DATA STORE
RECALL EXIT
DESCRIPTION
The data store feature saves up to 100 points of data and holds the
lowest (Anin) and highest (Max) readings while data store is on. The 100
points of data are stored at the rate of one of six selectable rates from
every reading (3 per second) to one reading per hour. Also, manual data
entry is available (RBI.. In the continuous trigger mode (SGL light off),
the minimumlmaximum readings are continuously updated at the rate
of three readings per second (regardless of the selected rate) while data
store is on.
100 POINT DATA STORAGE
1. Press and hold in the Data Store ON/OFF button. The following
storage rates will scroll on the display:
r=O (every reading)
r= 1 (1 reading per second)
r=2 (1 reading per 10 seconds)
r=3 (1 reading per minute)
r=4 (1 reading per 10 minutes)
r=5 (1 reading per hour)
r =6 (every time TRIG button is pressed)
2. When the desired storage rate is displayed, release the Data Store
ON/OFF button.
23
NOTE: If you are not ready to store data, turn the data store off.
When ready, press and release the Data Store ON/OFF button. The
preselected storage rate will be displayed briefly, the indicator light
will turn on and the store cycle will start.
3. Recalling Data-Data can be recalled at any time, but the flashing
DATA indicator light signals that the buffer is full (100 points stored).
A. Press and hold in the RECALL button. The DATA indicator light
will turn on and the display will scroll through the data and
minimum/maximum points. The first data point displayed is the
last data point that was stored. The next two points are the maximum (HI) and minimum (LO) readings made during that store cycle.
B. Release the RECALL button at the desired data point and note the
reading on the display. The data pointer can be incremented by
steps of one by momentarily holding in the RECALL button.
C. Exit from the recall mode by pressing the SHIFT button and then
the RECALL MIT button. The DATA indicator light will turn off.
4. Turn off the data store cycle by pressing the ON/OFF button. The indicator light will turn off. All stored data is retained until a new store
cycle is started, power to the instrument is turned off, or a DCL or
SDC is received over the IEEE bus.
TYPICAL USE
1. Saving data points of current drift of a low leakage device such as a
FET.
2. Monitoring and saving levels that fluctuate over a period of time.
MIN/MAX
1. Press and release the Data Store ON/OFF button. The indicator light
will turn on. In the continuous trigger mode, minimum and maximum
readings will continue to update at the rate of three readings per se
cond for as long as the data store is on and storing readings.
2. To recall the minimum/maximum readings:
A. Press and hold in the RECALL button. The DATA indicator light
will turn on.
B. Release the RECALL button when HI is displayed and note the
maximum reading.
24
C. Again press the RECALL button and release it when LO is
displayed. Note the minimum reading.
D. Exit from the recall mode by pressing the SHIFT button and then
the RECALL EXIT button. The DATA indicator light will turn off.
3. Turn off the data store cycle by pressing the ON/OFF button. The indicator light will turn off. The minimum/maximum readings, along
with the other data, will be retained until a new store cycle is started.
There must be at least one reading in data store in order
to recall the minimum/maximum values.
NOTE
TYPICAL USE
Detect surges, dips and peak drifts over a period of time.
25
PROGRAM FEATURES
The Model 617 has front panel programs to:
1. Display and change the IEEE-488 address of the instrument or select
the talk only mode.
2. Toggle the display for either alpha exponents or numeric exponents.
The program mode is selected by pressing the SELECT button. Program
parameters are changed with the use of the V-Source ADJUST buttons
(V-Source is disabled during program select). The adjust buttons increment and decrement the displayed value.
Program Storage and Exit
Exit the program mode and store the parameters by pressing the SHIFT
and SELECT EXIT buttons. The instrument will return to the previous
mode.
IEEE ADDRESS AND TALK ONLY
To display and change the IEEE address:
1. Press and release the SELECT button until IEEE and the present ad-
dress number are displayed.
2. To change the address, adjust the display to read the desired primary
address (O-30) using the ADJUST buttons.
3. Press the SELECT button to store the new address and exit the program mode.
To select the talk only mode:
1. Press and release the SELECT button until IEEE is displayed.
2. Using the adjust buttons, adjust the display to read 40 for talk only
mode with prefuc or to 41 for talk only mode without prefuc.
3. Press the SELECT button to store the talk only mode and exit the
program mode.
For more information on talk only operation, see page
40 of this booklet.
NOTE
26
DISPLAY EXPONENTS
Do the following to toggle the display for alpha exponents (i.e. fi, kn)
or numeric exponents (i.e. -06, 03).
1. Press and release the SELECT button until dlSP is displayed.
2. Press and release either of the ADJUST buttons. The exponent
mode will toggle.
3. Press the SELECT button to store the displayed exponent mode and
axit the program mode.
27
2V ANALOG OUTPUT
DESCRIPTION
A voltage level that is proportional to the incoming input level can be
monitored using the 2V ANALOG OUTPUT and COM binding posts on
the rear panel. All ranges wil[ output 2v at full range. The output is in-
verted when in volts and ohms.
TYPICAL USE
Connect a chart recorder to the analog output to monitor drii of the input signal czar an extended period of time.
28
PREAMP OUT
DESCRIPTION
The PREAMP OUT connector can be used to either configure the Model
617 for external feedback or to monitor the output voltage of the electrometer preamp up to k2OOV (proportional to the input level).
The table below lists the full range output voltages available at the
PREAMP OUT and COM terminals. The output is inverted when in
amps or coulombs.
Hazardous voltage levels may be present on the
PREAMP OUT terminal.
FULL RANGE PREAMP OUT LEVELS
FUNCTION
Amps
Ohms
Coulombs
RANGE
2JOmV
2v
20V
2oov
*A. m. &A r&-d+
ZOpA, 2OnA, ZO/rA, 20mA
2OOpA, 200nA, 2OOfi
2kfl
2OkQ - 2Gg
20Gg
2OilGll
2oopc
2nC
2OnC
WARNING
PREAMP OUT
(Absolute Value1
2oOmv
2v
20V
200 v
2oOmv
2v
20V
200mV
2v
20V
200 v
200mV
2v
20 v
29
OPERATION
External Feedback
1. Turn zero check on.
2. Press the SHIFT button and than the VOLTS button. The VOLTS
and AMPS indicator lights will turn on.
3. Connect the external feedback element from the PREAMP OUT binding post to input HI as shown.
4. Apply the signal to the triax input connector.
5. Turn zero check off.
“SUPPLIED INPUT CABLE
MODEL 617
Fig 3. External Feedback
NOTES
1. The negative of the voltage drop across the feedback element will be
displayed.
2. To exit the external feedback mode, turn zero check on, remove the
feedback element and prass any FUNCTION button.
TYPICAL USE
External feedback can be used if non-decade or logarithmic current gain
is required, or added coulombs capability is necessary.
30
V, hl GUARD
V, R GUARD
DESCRIPTION
Using the Model 617 in the guarded configuration IV, ‘2 GUARD switch
on) reduces response time and the effects of cable leakage when making volts and ohms measurements. The guard should be used when
making resistance measurements above lOOMl2 or voltage measurements that have high source resistances. See Fig 4 for guarded
measurement corrections.
00
w uo
*IF COM TO CHASSIS LINK IS IN, THE GRN LEAD CAN BE USED FOR
LOW. OTHERWISE. THE GRN LEAD IS UNUSED.
*‘SUPPLIED INPUT CABLE
Fig 4. Guard Measurements
MODEL 617
cl
31
INPUT
Hazardous voltage may be present on the inner
shield of triax input connector (black test lead)
when V. fl GUARD is on.
WARNING
1. Turn zero check on.
2 Place the V, Q GUARD switch (on rear panel) to the on position.
3. Select the volts or ohms function.
4. Select appropriate range or autorange.
5. Connect input as shown in the illustration.
6. Release zero check and take the reading.
TYPICAL USE
Measure Insulation Resistance-The guard configuration is recom-
mended to allaw quick and accurate measurements regardless of cable
capacitance or cable leakage.
32
EXTERNAL TRIGGER INPUT
DESCRIPTION
When the Model 617 is in the single trigger mode, external trigger pulses
can be used to initiate measurement conversions. Each pulse will trigger
one reading. The external trigger requires a falling edge, debounced
pulse at TTL logic level (see Figure 5). A debounced switch can be connected to the EXTERNAL TRIGGER input and used to trigger readings.
Every time the switch is closed, one reading will be triggered.
OPERATION
1. Set the Model 617 to the single trigger mode (sea TRIG).
2. Input an appropriate debounced trigger pulse source to the MTERNAL TRIGGER INPUT.
NOTE:
Trigger pulses that are not debounced will cause trigger errors ft Err).
TYPICAL USE
Have a programmable source trigger the Model 617 to take a reading
when the source outputs a new level.
TRIGGER ON
NEGATIVE
GOING
lTL HIGH
+2v to c5v
PULSE
TTL LOW
-z-u
ov to +imv
Fig 5. Trigger Pulse Specifications
33
The METER COMPLETE OUTPUT is a 4O@seo wide positive going
pulse that signifies completion of the measurement cycle. The mater
complete line can be used to trigger another instrument or to inform an
instrument that the measurement has bean completed.
OPERATION
Using an appropriate cable, connect the METER COMPLETE OUTPUT
of the Model 617 (locatacl on rear panel) to the atierna! trigger input of
another instrument.
TYPICAL USES
1. Trigger a .scanner to switch to the next available channel when the
Model 617 measurement cycle has been completed.
2. Tn’gger a programmable source to output the nex3 programmed level
when the Model 617 measurement cycle has been completed.
34
IEEE-488 PROGRAMMING
Function:
RO = Auto on
Rl = 200mV
R3 = 20 V
R4 = 200 V
R5 = 206 V
R6 = 200 v
R7 = 200 V
BO = Electrometer
El = Data Store*
82 = Highest (Max) Buffered Reading
83 = Lowest (Min) Buffered Reading
84 = Voltage Source
*See page 42 for program example.
X Fdbk
Auto on
200mV
2v
20 v
20 v
20 v
20 v
20 v
20 v
20V
20 v
20 v
V/I
Auto or
2oom
2OTB
2m
2Gn
200Mn
20Mn
2Mn
Auto of
35
Data Store
‘-source Value: Vfn.nnnnnE?cnn
‘-Source Operats: CXJ = Dff
lalrbration VaIue: A~n.nnnonErtnn
‘tore:
Rigger:
tatus:
‘ata Format: GO = Send prefix with data.
R(1:
Control:
=
Reading
Every
Ql = 7 reading per second.
M = 1 reading per 10 seconds.
Q3 = 1 reading per minute.
Q4 = 1 reading per 10 minutes.
Cl5 = 1 reading per hour.
06 = Every time Tl?lG button is pressed.
Q7=Off
n = Voltage Source Value
01 = On
A = Calibration Value
Ll = Store Calibration Constants
TO = Continuous, Talk
Tl = Single, Talk
T2 = Continuous, GET
T3 = Single, GET
T4 = Continuous, “X”
T!! = Single, “x”
T6 = Continuous, External
l7 = Single, External
“Trigger always starts a new reading,
regardless of mode (singlelcontinuousl.
UO = Send machine status word.
Ul = Send error status word.
U2 = Send data status word.
Gl = Do not send prefix with date.
G2 = Send prefix and buffer address with
data (if in 81).
MO = Clear SRQ Mask
Ml = Reading Overflow
M2 = Buffer Full
M8 = Reed’& Done
Ml6 = Ready
M32 = Error
Example: Ml0 programs the 617 to SRQ
when a reading is done or the buffer is full.
36
EOI: KO = Send EOI and hold off on “X”.
\
TRIGGER TO FIRST BYTE OUT: 350ms.
Kl = Do not send EOI and hold off on
“X
K2 = Send EOI and no hold off on 9”.
K3 = Do not send EOI and no hold off on
‘X”.
Ym = One terminator character.
Ymn = Two terminator characters
X = Execute Device-Dependent Commands
“Use “X” to force a sequence. Example: To
POWER UP DEFAULT CONDITIONS
MODE
Function
Range
Zero Check
zero cormct
supprass
Trigger
V-Source Operate
Read Mode
Data Format
Display
Data Store Control
SRQ
EOI
Teninator
V-Source Value
St&US
Cal Value
STATUS (see machine status word)
Fo = Volts
RO = Autorange
Cl = On
M=Off
NO = Off
T6 = Continuous, External
OO=Off
BO = Electrometer
GO = Send prefix with data
W = Electrometer
Q7 = off
MO = Clear SRQ mask
KO = Send EOI and hold off on “X”
YO = CR LF
vo = 0 Volts
U not active
A not active
A front panel program is used to display end change the IEEE-488
primary address of the Model 617. See page 26 for instructions.
39
TALK ONLY OPERATION
DESCRIPTION
Talk only may be used to send data to a listen only device such as a
printer. Data may be sent with the prefix (i.e. NDCA) or without the
prefix. When the Model 617 is in the talk only mode, it ignores com-
mands g’nren over the bus. The talk only mode is enabled through a front
panel program. The default talk rate is three readings per second (every
reading). However, a different talk rats can be selected.
OPERATION
1. Select Program. Press and release the SELECT button until IEEE is
displayed.
2. Select Talk Only Mode. Adjust the display, using the ADJUST buttons to mad 40 for talk only mode with prafix or to 41 for talk only
mode without prefix.
3. Store Program. Press the SHIFT button end then the SELECT EXIT
button.
4. To select a different talk rate:
A. Press and hold in the Data Store ON/OFF button. The following
talk ratas scroll on the display:
r = 0 (every reading)
r = 1 (1 reading per second)
r = 2 (1 reading par 10 seconds)
r = 3 I1 reading per minute1
r = 4 (1 reading per 10 minutes)
r = 5 (1 reading per hour)
r = 6 (1 reading every time TRIG button is pressed)
B. Release the ON/OFF button when the desired talk rate is
displayed.
NOTE: The instrument remains in the selected talk rate until a new rate
is salacted, the data store cycle is turned nff (DATA STORE in-
dicator light off) or power is cycled.
40
PROGRAMS
The following programs are designed to be a simple aid to the user.
They are not intended to suit specific needs. Detailed programming information can be found in the manual.
The first program demonstrates how to read the contents of the buffer
(data store) using the HP-85 computer. All 100 readings stored in the
buffer will be displayed.
The rest of the programs display one reading at the output of the controller. Each program provides an ASCII string variable output of the
form:
The note at the end of each program indicates modiications to provide
a numeric variable (A) in exponential form:
NDCA + 0.00000 + 0 CR LF
+ 0.00000
41
DATA STORE OPERATION
USWIG HP-85
The following program will enable data store (buffer) operation and obtain and display 100 readings on the computer CRT.
DIRECTIONS
1. Using the front panel program feeture, set the primary address on the
Model 617 to 27.
2. Connect the Model 617 to the HP82937A IEEE interface.
3. Enter the Following program using the END LINE ksyaffar each line is
typed.
4. Press the HP-95 RUN key. The program wiU enable the buffer, wait
for the buffer to fill, turn on the buffer output, and then request and
display all 100 readings.
PROGRAM
COMMENTS
10 DIM AsI
20 REMOTE 727
30 OUTPUT 727”QOX”
40 S=SPOLL rin,
60 IF NOT BlT(S,l) THEN 40
60 OUTPUT 727”BltXX”
70 FOR I =l TO’100
BO ENTER 727;M
90 DISP AS
100 NEXT I
..- -..-
110 END
Send remote enable.
Enable buffer at conversion rate.
Serial Poll the 617.
If buffer not full, wait.
Set read mode to buffer.
Loop 100 times.
Get a buffer reading.
Display it.
Loop back and get next reading
42
IBM XT or PC
(Keithley 8573 GPIB Interface)
The following program sends a command string to the Modal 617 from
an IBM PC or XT computer and displays the instrument reading on the
CRT. The computer must be equipped with a Keithley Instruments
Modal 8573 GPIB IEEE interface and the DOS 2.0 operating system. The
GPIB software and hardware must be configured per the Keithley Instruments Modal 8573 Instruction Manual.
DIRECTIONS
1. Using the front panel program feature, set the primary address of the
Model 617 to 27.
2 With the power off, connect the Model 617 to the IEEE-488 interface
installed in the IBM computer.
3. Type in the command BASICA on the IBM keyboard to get into the
IBM interpretive BASICA language.
4. Place the interface software disc in the default drive, type LOAD
“DECL”, and press the return key.
5. Enter the following program into the computer, pressing the return
key after each line is type. Lines l-6 are part of the DECL program
previously loaded and naed not be typed in. However, lines 1 and 2
will have to be altered. See the Model 8573 Instruction Manual.
6. Run the program and type in the desired command string when
pmmpted. For example: to place the Model 617 into the one shot on
talk trigger mode and in the 2@ range, type in FlR7TlX and press
the return key.
7. The display will show the Model 617 reading string on the CRT. For
example, when the Model 617 is in zero check, the display will read
NDCA + O.OOOOOE-06.
8. To exit the program type EXIT and press return.
NOTE: Lines l-6 of this program are not shown and need not be typed
in. They are contained in the floppy disc. When the command
LOAD”DECL” is entered these lines are already there.
NOTE: If conversion to numeric variable isdesired, change lines 120 and
130 as follows:
120 RD=VAL(MIDS(RDS,5.16))
130 PRINT RD
COMMENTS
Find the board number.
Find the 617 number.
Set timeout delays.
Change to primary address 27.
Set REN true.
Prompt for ccmmand string.
Sea if pmgram is to be halted.
If null command string, go back
and get another.
Address 617 to listen and send
command string.
Assign reading input buffer.
Address 617 to Talk and input
reading string.
Trim string to proper size.
Display the reading on the CRT.
Repeat.
Close the board file.
Close the instrument file.
44
APPLE II (APPLE Interface)
The following program obtains one reading from the Model 617 Elactrometar and displays the reading on the APPLE II screan, using an
APPLE IEEE-488 interface.
DIRECTIONS
1. Using the front panel program feature, set the primary address of the
Model 617 to 27.
2. Connect the Model 617 to the APPLE II and APPLE IEEE-488 interface.
3. Enter the following program using the RETURN key after each line.
4. Type in RUN and depress the RETURN key.
5. The display will read “TEST SETUP”.
6. To program the Model 617 to the &A range and take a reading, type
in FlR7TlX and depress the RETURN key.
7. The display will read NDCA + O.WOOOE-06 when the Model 617 is
in zero check.
NOTE: If conversion to numeric variable is desired, change lines 40 and
50 as follows:
40 ENTER 727;B
50 PRINT B
COMMENTS
Set to remote.
Prompt for test setup.
Send command string to 617.
Get data string from 617.
Display data string.
48
The following program obtains one reading from the Model 617 Etectrometer and displays the reading on the DEC LSI 11 microcomputer
CRT terminal. The LSI 11 must be configured with lf% words of RAM
and an IBV 11 IEEE interface. The software must be configured with IB
software as well as the FORTRAN and the RT 11 operating system.
DIRECTIONS
1. Using the front panel program features, set the primary address of
the Model 617 to 27.
2. Connect the Model 617 tothe IBV 11 IEEE cable.
3. Enter the following program, using the editor under RT 11 end the
name IPHILD.
4. Compile using the fortran compiler as follows: FORTRAN IPHILD.
5. Link with the system and IB libraries as follows: LINK IPHILD,IBLIB.
6. Type RUN IPHIL and depress the RETURN key.
7. The display will read “ENTER ADDRESS”.
8. Type in 27 and depress the RETURN key.
9. The display will read “TEST SETUP”.
10. To program the Model 617 to the &A range and take a reading, type
in FIRTTIX and depress the RETURN key.
11. The display will read NDCA + 0.000OOE -06 when the Mode! 617 is
in zero check.