4-6.
4-J. Grid Current Check. 33
4-a. DC Amplifier Adjustment 31
4-9.
4-10. Calibration of Ohms Ranges. 35
4-11. Meter Zero Calibration. . 3j
4-12. Accuracy Check. . . . . . 35
4-13. Drift Check . 36
5. ACCESSORIES . . . . . . 43
5-l. Model 6101A Shielded Test
5-2.
5-3.
5-4.. Model 6103A Voltage Divider
5-5. Model 6104 Test Shield. . . 44
5-6.
5-7.
5-8.
5-9.
5-10. Model 370 Recorder. . . . . . 48
Calibration Procedures. . . . 31
High-Megohm Resistor Veri-
fication. . . 35
Lead. . . . . . . . . 43
Model 61018 Gripping Probe. 43
Model 6102A Voltage Divider
Probe . . . . . . . . . 43
Probe . . . . . . . . . 43
Model 6105 Resistivity
Adapter . . . . . , . . . 44
Models 2501 and 2503 Static
Detector Heads. . . . . . . . 45
Model 6106 Electrometer
Connection Kit. . . . . . . . 46
Model 6107 pH Electrode
Adapter . . . . . . . . . . . 4i
? :I '2 2
3.
CIRCUIT DESCRIPTION
3-l. General ...........
3-2. Voltmeter Operation
3-3. Voltmeter circuit ......
3-4. Ammeter Operation ......
3-5.
3-6. Coulombmeter Operation.
3-7. Power Supply. ........
4.
UAINTENANCE
4-l. General
4-2. Parts Replacement ......
4-3. Troubleshooting .......
4-4. Procedures to Guide Trouble4-5.
O268R
Ohmmeter Operation.
............
...........
shooting. ..........
234-Volt Ooerarion.
........
21
21
..... 21
22
24
..... 24
... 25
26
29
29
29
29
30
..... 32
6. REPLACEABLE PARTS . . . . . . . 51
6-l.
6-2. How to Order Parts. . . . . . 51
* Change Notice Last Page
* Yellow Change Notice sheet is included
only for instrument modifications affect-
ing the Instruction Xanual.
Replaceable Parts List. . . . 51
Models 6108, 610BR Replaceable
Parts List. . . . . . . . 52
Model 6108 Schematic Diagram
177953. . . , . . . . . . . 59
Green Repair and
Calibration Forms . . 61
1
GENERAL DESCRIPTION
MODELS 6lOB, 610BR ELECTROMETERS
FIG
FIGURE 2.
610BR is the rack version of the Model 6108.
Keithley Instruments Model 610BR Electrometer. Model
ii
1064R
i
MODELS 6108, 610BR ELECTROMFPERS
GENERAL DESCRIPTION
SECTION 1.
l-l.
measures wide ranges of dc voltages, currente, resistances and charges. It
is a highly improved form of conventional dc vecuum tube voltmeters that
ueee an electrometer tube input to provide greeter than lo14 ohm Fnput resistance. T’he Model 6lOB has all the capabilities of conventional VTVMe,
but it ten also make many more measurements without loading circuits.
100 volts, 28 ~urm-ie: ranges from Lo-14
25 linear resistance ranges from 100 ohms full scale to 1014 ohms, and 15
couLomb ranges from LO-12 coulomb full scale to LO-5 coulomb.
The output stage uses a vacuum tube; othewiee.solid-state devices are
used for all amplifier stages and the power supply. This increases re-
liability end stability and aLlows
l-2.
GENERAL.
a. The Keithley Model 610B Electrometer is a versatile instrument which
b. The Electrometer has 11 voltage ranges from 0.001 volt full scale to
c. The Model 6LOB input stage uses a matched pair of electrometer tubes.
FEATURES.
a. Voltmeter accuracy is Cl% of full scale, exclusive of noise and drift.
GENERAL DESCRIPTION
ampere full scale to 0.3 ampere,
greater
regulation of the power supply.
Zero drift of the Model 6108 is 200 microvolts per hour maxirmun averaged
b.
over any 24-hour period after warm-up.
is no more then 2 millivolts after the first hour.
c. Two amplifier outputs are available.
either 23 volts or 21 milliampere for full-scale meter deflection. The current
output is variable 25% with 1.4-kilohm recorders.
output is equal to the input voltage within 50 ppm or 100 microvolts, exclusive
of zero drift.
Current measurements can be made by one of two methods, the normal method
d.
in which the current is determined by measuring the voltage drop across a resistor shunting the input, or the fast method in which negative feedback is
appliedthrough the shunt resistor.
end greatly increases the response speed on the low-current ranges.
1-3.
Model 610B. The circuits, specifications,
procedures for the two models are the same. Besides the outside dimensions,
the main difference between them is the Model 6LOBR’s COARSE ZERO Control
is on the front panel, not on the rear panel.
Model 610B is mentioned. Any differences are identified for the correct model.
DIFFERENCES BETWEEN THE MODELS 610B AND 610BR.
The Model 610BR
a.
b. The instructions in the Manual are for both models, although only the
is
the rack
version
During the 2-hour warm-up, zero drift
A switch on the rear panel allows
The unity-gain amplifier
The latter method reduces ehe input drop
of the cabinet configuration,
electrical parts end operating
t064R
1
GENERAL DESCRIPTION
1-4. SPECIFICATIONS.
AS A VOLTMETER:
?,ODELS 6108, 610BR ELECTROMTERS
RANGE:
ACCURACY:
,001 volt full scale to 100 volts in eteven Lx and 3x ranges.
?L% of full scale on all ranges exctusive of noise and drift.
ZERO DRIFT: After L-hour warm-up no more than 2 millivolts in the second hour, and in any
subsequent 24-hour period, the average drift wilt not exceed 200 microvolts per hour.
HETER NOISE:
INPUT IMPEDANCE:
+tO microvotes with input shorted.
Greater than Lot4
ohms shunted by 22 picofarads on the VOLTS position of
the Range Switch. Input resistance may be selected in decade steps from LO to LOLL ohms
with the Range Switch.
AS AN AMMETER:
RANGE:
AccuRAcY:
10-14
ampere full scale to 0.3 ampere in twenty-eight Lx and 3x ranges.
?2% of full scale on 0.3 to 10-11 ampere ranges using smattest available Ilulti-
ptiee Switch setting; 14% of full scale on 3 x tOeL2 to tOdL4 ampere ranges.
METER NOISE: Less than i3 x lO-L5 ampere.
GRID CURRENT: Less than 2 x lO-L4 ampere
AS AN OHMMETER:
RANGE:
‘KcIJRAcY: ?3% of full scale on 100 to log ohm ran es using highest available Multiplier
Switch setting;
100 ohms full scale co 1014 ohms on twenty-five Linear Lx and 3x ranges.
K4
?5% of full scale on 3 x LO9 to
to
ohm ranges.
AS A COULOMBMETER:
RANGE :
ACCURACY: ?5% of full scale on all ranges.
LO-L2 coulomb full scale to tom5 coulomb Ln fifteen lx and 3x ranges.
Drift due to grid current does not exceed
2 x 10eL4 coulomb per second.
AS AN AMPLIFIER:
INPUT IHPEDANCE: Greater than tot4 ohms shunted by 22 picofarads on the VOLTS position
of the Range Switch.
Input resistance may be selected in decade steps from LO to LOLL
ohms with the Range Switch,
OLTPUTS: Unity-gain output and either voltage or current recorder output.
Unity-Gain Output:
exclusive of zero drift,
At dc,
output is equal to input within 50 ppm or 100 microvolts.
for output currents of LOO microamperes or Less.
up to one
milliampere may be drawn for input vottages OE LO votes or Less.
.*
* .
t066R .
MODELS 6108, 610BR ELECTROMETERS
GEXERAL 3ESCRI?TIO”
Voltage Recorder Output:
+3 volts for full-scale
input.
Internal resistance is 3 kil-
ohms. Output polarity is opposite input polarity.
Gain: 0.03, 0.1, etc., to 3000
Frequency Response (Within 3 db):
dc to 300 cps at a gain of 3000, rising to 25 kc
at a gain of 30, decreasing to 2.5 kc at a gain of 0.03.
Noise: 3% rms of full scale at a gain of 3000, decreasing to 1% at gains below 100.
Current Recorder Output:
tl milliampere for full-scale input, variable T5% with LLiOO-
ohm recorders.
GENERAL :
POLARITY: METER Switch selects left zero (positive or negative) or center-zero scales
Output polarity is not reversed.
LINE STABILITY: A 10% change in line voltage will ca”se less than a LO-microvolt merer
defLection on aLL ranges.
CONNECTORS : Input:
output:
POWER:
105-125 or 210-250 volts (switch selected): 50 to 1000 cps; IO Watts
uhf type; ground binding post.
Voltage or current, AmphenoL 80-PCZF; Unity-gain, binding posts
DIMENSIONS, WEIGHT: Model 6108:
net weight, 12 pounds. Model 610BR:
10-l/2 inches high x 6-S/8 inches wide x 10 inches dee?;
5-l/4 inches high x 19 inches wide x 10 inches deep;
net weight, 12 pounds.
ACCESSORIES SUPPLIED: Mating input and output plugs; 3:2 power line adapter; binding plug.
l-5. APPLICATIONS.
a. Voltmeter applications include directly measuring potentials of vacuum tube plates
and grids,
pH electrodes, piezo-electric crystals,
capacitors and electrochemical cells,
and the gate potentials of field effect transistors. With the Model 2501 or 2503 Static
Detector Probe, the Model 610B can measure electrostatic voltages.
b. As a picoammeter,
the Electrometer can measure mass spectrograph currents.
It can
be used with photo multiplier tubes, flame and beta ray and lithium ion-drift detectors.
Other uses are in gas chromatograph work, nuclear studies, plasma physic studies and
vacuum studies.
Also, it can be used as a current null detector with an accurate current
reference source.
c. As an ohmeter, its uses include measuring diode characteristics, insulation resistance, and resistor voitage coefficients. With the Model 6105 Resistivity Adapter and a
power supply,
the Node1 610B can measure volume and surface resistivities.
d. In addition to measuring charge directly, other coulombmeter uses are measuring charge
current over a period and obtaining integral curves of time-varying currents. The Model
6108 can measure the total energy output from a pulsed laser on its coulomb ranges when
used with a calibrated photo tube or photo diode.
It also can be used as a charge ampli-
fier to measure pieza-electric crystal outputs.
1066R
3
GENERAL DESCRIPTION
MODELS 61OB, 610B~ ELECTROMETERS
1-6.
ACCESSORIES.
a. Three accessory probes, fully described in Section 5,
facilitate measurements end
extend the Electrometer’s voltage range to 30 kilovolts.
Model 6104 Test Shield is suitable for resistance measurements with either 2 or 3-
b.
terminal guarded connections, as well as voltage and current tests.
c. Model 6105 Resistivity Adapter ts a guarded test fixture for measuring volume and
surface resistivitirs of materials when used with the Model 6108 and the Keithley Node1
240A High Voltage Supply.
Models 2501 and 2503 Static Detector Probes are capacitive voltage dividers with a
d.
10,OOO:l ratio, when used with the probe 3/B inch from the charged surface.
e. Model 6106 Electrometer Connection Kit contains a group of the mwt useful leads
and adapters for electrometer measurements.
Model 6107 pH Electrode Adapter has a 2-Eoot cable and coaxial caxxector and accepts
f.
a Beckman and Coleman (B-C) or a Leeds and Northrup (L &N) connector.
The Adapter allows
accurate and convenient pH potential measurements with the Model 6lOB.
Model 370 Recorder is uniquely compatible with the Model 610B es well es other
g.
Keithley microvoltmeters, electrometers and picoarmneters.
The recorder is a high quality
economical instrument that maximizes the performance of the Model 6108 and many other
Keithley instruments, even in the most critical applications.
l-7.
EQUIPMENT SHIPPED.
ped with all components in place.
mating plugs for the input and output receptacles.
The Model 6108 Electrometer is factory-calibrated and is ship-
The shipping cartnn contains the Instruction Manual and
The Model 610BR is shipped with the
rgles and screws for rack mounting packed separately within the shipping carton.
FIGURE 3. Model 610ER Front Panel Controls and Terminals.
Circuit designations refer to
Replaceable Parts List and the schematic diagram.
I
0667R
MODELS 610B, 610BR ELECTROMETERS
OPERATION
SECTION 2.
2-1. FRONT PANEL CONTROLS AND TERMINALS (See Figures 3 and 4)
a.
Range Switch.
range.
ranges and four COULOMBS ranges.
the dial skirt.
b. Multiplier Switch.
sitivity of the dc amplifier and sets the full-scale voltage range when the
Range Switch is set to VOLTS. The Multiplier Switch may also be used to
multiply the AMPERES (3x maxinolm setting above LO-3), OHMS and COULOMBS
ranges on the Range Switch.
c.
the instrument. OFF disconnects only the meter during recorder operation.
The + and - positions determine the polarity of the meter.
sets the instrument for center zero operation (lower meter scale).
d. ZERO Controls.
Switch (outer knob) and a lo-turn FINE potentiometer (center knob).
allow precise meter zeroing.
It is divided fnto a VOLTS position, 11 AMPERES ranges, eight OHMS
METER Switch. The Switch has five positions:
The Range Switch selects the measuring mode and the
The range used is indicated by an X above
The Multiplier Switch determines the voltage sen-
Two ZERO Controls are on the front panel: a MEDIUM
OPERATION
POWER OFF shuts off
CENTER ZERO
These
0764R
Circuit designations refer to Replaceable Parts List and the
schematic diagram.
OPERATION
ZERO CHECK Button. Depressing the Button effectively removes all input
e.
MODELS 6108, 610BR ELECTROMETERS
signal from the instrument by shunting the input and amplifier through 10
megohms.
This allows meter zeroing on any range.
The Button is locked in
the zero check position when the line is horizontal.
f. FEEDBACK Switch.
the feedback connections within ehe instrument.
made only in NORMAL.
with lower input voltage drops and faster response speeds.
The FAST and NORMAL positions of the Switch determine
Voltage measurements are
With the Switch et FAST, current measurements are made
The FAST position
is also used to make guarded resistance measurements, for couLomb measurements,
and to increase response speed.
INPUT Receptacle.
!z.
connector.
A dust
The INPUT Receptacle fs a Teflon-insulated uhf-type
cap is provided.
The ground post is below the recep-
tacle.
2-2. RFAR PANEL CONTROLS AND TERMINALS (See Figure 5).
a. COARSE ZERO Switch (On front panel of Model 610BR). The COARSE ZERO
Stitch has 11 positions to extend the zeroing capability of the front panel
ZERC Controls.
Output Switch. The Switch fs a two-position slide switch for the output.
b.
In the 1 MA position, the instrument will drive l-milliampere recorders. In
the 3 V position, the output is 3 voles for full-scale meter deflection.
Source resistance is 3 kllohms.
c. 1 MA CAL Control. The Control varies the output from 0.95 to 1.05 milliampere for 14CO-ohm recorders , so the recorder scale will correspond with the
Electrometer panel meter.
OUTPUT Receptacle. A 2-terminal microphone-type receptacle provides
d.
3 volts or 1 milliampere for full-scale meter deflection. Pin no. 2 is at
case ground potential if the FEEDBACK Switch is at NORMAL.
e. Xl OUTPUT and GUARD Terminals.
The potential between the Xl OUTPUT
Terminal and the CUARD Terminal (case ground for the FEEDBACK Switch in NORMAL)
Fs eaual to the input voltage with 0.005% linearity or LOO microvolts. when
the FEEDBACK Switch is at FAST, the Xl OUTPUT Terminal is et case ground end
rhe,GUARD Terminal is floating.
f. FUSE.
fuse.
117-234 Switch.
g.
For 210-250 volt operation, use a L/8 ampere, 3 AC Slow Blow fuse.
For 105-125 volt operation, use a l/4 ampere, 3 AG Slow Blow
The screwdriver-operated,slide switch sets the Model
610B for 117 or 234-volt ac power lines.
h. Power Cord. The 3-wire power cord with the NEMA approved 3-prong plug
provides a ground connection for the cabinet. An adapter for operation from
2- tenninal outlets is provided.
2-3. INPUT CONNECTIONS.
a. The accessories described in Section 5 are designed to increase the
designations refer to Replaceable Parts List and the schematic
diagram.
accuracy and convenience of input connections.
Use them to gain the maximm
capability of the Model 6LOB.
NOTE
Using the accessories and coaxial cables is the best
way to make input connections.
b. Carefully shield the input connection and the source being measured,
since power Line Erequencies are well within the pass band of the Electrometer
Unless the shielding is thorough,
any alteration in the electrostatic field
near the Input circuitry will cause definite meter disturbances.
c. Use high resistance, Low-Loss materials -
polyethylene or polystryene - for insulation.
such as Teflon (reconrmended),
The insulation leakage resis-
tance of test fixtures and Leads should be several orders magnitude higher
than the internal resistance of the source. If it is not,
leakage Losses
will cause Lowered readings. Coaxial cables used should be a Low-noise type
which employ a graphite or other conductive coating between the dielectric
and the surrounding shield braid.
Amphenol-Borg Electronics Corporation,
Hicrodot , Inc. , and Simplex Wire and Cable Company make satisfactory types.
1064R
OPERATION MODELS 6108, 610BR ELECTROMETERS
Using the Model 6LOLA Probe is a simple way to insure.good input connections.
NOTE
Clean and dry connections and cables are very important
to maintain the value of all insulation materials.
Any change in the capacitance of the measuring circuit to ground will
d.
cause extraneous disturbances.
and tie down connecting cables to prevent their movement.
vibration is present, it may appear at the output as a sinusoidal signal and
other precautions may be necessary t" isolate the instrument and the connecting
cable from the vibration.
Make the measuring setup as rigid as possible,
If a continuous
e. For Low impedance measurements,
adapter may be used. However, keep the leads short.
f. When measuring currents LO-l4 ampere or less with the FEEDBACK switch
at FAST, some insulators - such as Teflon show up as erratic meter deflections. Insulation used in the Model 6108 is
carefully selected to minimize these signals.
Connect the M"deL 610B to the circuit only when a reading is being made.
.s.
I" some cases, the grid current can charge the external test circuitry when
it is connected to the Electrometer input. One example is measuring a
capacitor's Leakage resistance by observing the decay of its terminal voltage.
If
the
Leakage currant
the terminal voltage when the electrometer is Left connected across the
capacitor's terminals.
that the capacitor dielectric has a negative resistance.
Techniques and applications are thoroughly discussed
in thebrochure, Electrometer Measurements, by Joseph
F. Keithley. It is available by writing to Keithley
I"stnlae"ts, 1°C.
2-4. Preliminary Procedures.
is
less than the grid
Instead, there is a build-up which seems to indicate
unshielded Leads and a binding post
may produce random signals which
current, there is
NOTE
no
decay
of
a. Check the 117-234 Switch and the Fuse for the proper ac Line voltage.
NOTE
Make sure MRTRR Switch is sst to POWER OFF before connecting
or disconnecting the power cord.
Set the controls as follows:
b.
METER Switch
Range Switch
Multiplier Switch
FEEDBACK Switch
ZERO CHECK Button
a
POWER OFF
VOLTS
1
NORMAL
LOCK
0667R
MODELS 61.0~. 610B~ ELECTROMETERS
c.
Connect the power cord and turn the XF,TER Switch to CEpiTER ZERO. Within
ten seconds, the meter pointer should come to the center zero position. If
not,
adjust to meter zero with the MEDIUM and FINE ZERO Controls.
there is no need fo use the COARSE ZERO Switch.
d. After a few moments increase the voltage sensitivity by advancing the
Multiplier Switch to .3, .L, etc.
Continue zeroing with the FINE ZERO Control.
NOTE
OPERATION
Normally,
Always turn the Model 6lOB off using the METER Switch.
shut the instrument off by disconnecting the power cord or otherwise
externally shutting off the power supply.
externally, set the Multiplier Switch to 1 or lower.
cautions are not followed, the Electrometer may drift for several
hours when it is used again.
e. After Long periods of storage or after an overload, the Model 610B
may drift excessively.
severe jolt can also cause a zero offset.
trols. Drifting, however, can occur for a few hours.
Although the grid current of the Electrometer is much below that found
f.
in conventional voltmeters, it can be observed on the meter. A small voltage
results from the grid current charging the input capacity, and the Electro-
meter appears to drift when the input is open. Use the ZERO CHECK Button
to discharge the build-up.
Keep the protective cap on the INPDT Receptacle when the
Electrometer is not in a circuit.
Follow the particular procedures for measuring voltage, current, re-
g.
sistance and charge in the next four paragraphs.
Although the electrometer tubes
NOTE
If it w be turned off
This is corrected with the Zero Con-
Do not
If these
are
shock mounted, a
2-5. Voltugo Measurements.
a. The Model 610B’s high input impedance allows circuit measurements without causing circuit loading. For low resistance in-circuit tests, the input
resistance can be Lowered to avoid pick-up problems,
NOTE
Make all voltage measurements with the FEEDBACK Switch &
in the NORMAL position.
High Impedance Measurements (LO14 ohms, 22 picofarads). Follow the
b.
instructions of paragraph 2-4.
METER Switch
Range Switch
Multiplier Switch
FEEDBACK Switch
ZERO CHECK Button
t064R Y
Set the controls as follows:
CENTER ZERO
VOLTS
100
NORMAL
LOCK
I
OPERATION
8fom.s 610~) 610~~ ELECTROMETERS
Connect the unknown source to the INPUT Receptacle and unlock the ZERO CHECK
Button.
Set the METER Switch to + or -, as necessary.
with the Multiplier Switch,
Recheck the zero setting after increasing the sen-
Increase the sensitivity
sitivity.
c. Low Impedance Measurements.
set the Range Switch to one of the AMPERES ranges.
the reciprocal of the current range. For instance,
of 107 ohms
to the 10-7
Switch.
To measure sources more than 100 volts, use the Model 6102A 1O:l Divider
d.
- which is normal for conventional VTVMs - set the Range Switch
AMPERES range.
Set the full-scale voltage range with the Multiplier
Operating procedures are the same as subparagraph b above.
Probe or the 6103A 1OOO:l Divider Probe.
610B’s range to 1000 volts;
1010 ohms.
The Model 6103A extends the Model 610B’s range to 30 kilovolts;
overall accuracy is *3% and input resistance is
overall accuracy is C5% and inputresistanceis 1012 ohms.
operating procedures with the dividers as in subparagraph b.
To decrease the input resistance from 1014 ohms,
The input resistance is now
to obtain an input resistance
The Model 6102A extends ths Model
Follow the same
The full-scale
voltage range is the divider ratio times the Wltiplier Switch setting.
Accuracy decreases 0.5% when the center zero scales are used
because the scale span is shorter.
2-6.
Currant Mearuremantr.
The Model 610B can measure currents three ways.
a.
In the normal method - used on any range
1.
-. the current is determined
by measuring the voltage drop across a resistor shunting the amplifier input.
This method is useful when lower noise is more important than faster response
speeds or if some damping is needed.
In the fast method -
2.
for use only below the lO-5 ampere range the shunt resistor is between the amplifier output and input in the feedback
loop.
This circuit largely neutralizes
greatly increases the response speed.
the effect of input capacity and
Also, the input voltage drop is reduced
to a maximum of one millivolt on any range.
For galvanometric current measurements,
3.
the Model 610B acts as a null
indicator between a very accurate current source and the unknown current source.
Rise time varies primarily with the currant range, theinput capacity
b.
and the method used.
50 picofarads across the input.
On most ranges, the rise time is less than one second with
Even with much larger shunt capacities, the
negative feedback maintains a short rise time. Given a choice, it is better to
place the Electrometer near to the current source than to the data reading instrument.
Transmitting the fnput signal through long coaxial cables greatly de-
creases the response speed and increases noise due to the cable capacitance.
c. Normal Method (0.3 to LO-l4 ampere ranges).
1. Follow the instructions of paragraph 2-4. Set the controls as follows:
10
1064R
MODELS 610B, 6lOBR ELECTROMETERS
OPERATION
METER Switch CENTER ZERO
Range Switch
10-l AMPERES
Multiplier Switch 1
FEEDBACK Switch
NORMAL
ZERO CHECK Button LOCK
Connect the unknown source to the INPUT Receptacle and unlock the ZERO CHECK
Button. Set the METER Switch to + or -, es necessary.
ity with the Range Switch and the Multiplier Switch.
Switch higher then 3 for Range Switch settings lo-3 and above.
Increase the sensitiv-
Do not set the Multiplier
Check zero with
the ZERO CHECK Button.
2. The full-scale current range is the Range Switch setting times the tilti-
plier Switch setting.
Use the smallest Multiplier Switch setting possible to
obtain the best accuracy. The input resistor varies with the Range Switch setting, from 10 ohms et 10-l AMPERES to lo11 ohms for lo-l1 AMPERBS.
The input
voltage drop is the meter reading times the Multiplier Switch setting.
3. On the 10-S to lo-11 AMPERES settings of the Range Switch, the high-megohm
resistors used have a voltage coefficient of a nominal y.0277 per volt.
If the
input voltage drop reaches 100 volts, en error of 100 x .02% or 2% occurs to
lower the specified accuracy. Therefore,
use a low filtiplier Switch setting
- best is 0.1 to 1 - so neither voltage coefficient nor zero stability presents a problem.
On the low current ranges, balance out the grid current
with the Zero Controls or subtract the value from the
reading. To find the amount of grid current, cap the
INPUT Receptacle and read the meter.
Fast Method (ranges below 10e5 ampere).
d.
Follow the instructions of paragraph 2-4.
1.
Set the controls as follows:
METER Switch CENTER ZERO
Range Switch 10-6 AMPERE.5
Wltiplier Switch
1
FEEDBACK Switch FAST
ZERO CHECK Button
LOCK
Connect the unknown source to the INPUT Receptacle and unlock the ZERO CHECK
Button. Set the METER Switch to + or -, as necessary. Increase the sensitivity
with the Range Switch end the Multiplier Switch. Do not set the Range Switch
to 10-S AMPERES or higher. Check zero with the ZERO CHECK Button.
NOTE
Use only the ZERO CHECK Button to check zero for the
East method. Do not short the input, because this will
remove the feedback from the circuit.
1264R
OPERATION
The full-scale current range is the Range Switch setting times the Multi-
2.
plier Switch setting.
smell settings permit lower current source resistance, and larger settings im-
prove instrument zero stability.
With the fast method, the input drop is reduced and the response speed
3.
is increased at least 100 times,
a) The internal impedance of the unknown current source should not be less
than .l of the value of the feedback resistor being used.
full feedback voltage cannot be developed at the input, and zero instability
results.
of the Range Switch.
b) The low side of the OUTPUT Receptacle is no longer grounded. Therefore,
do not ground recorders or oscilloscopes to the Elecerometer’s case, such as
through the ground lead of the power cord.
unity-gain output (paragraph Z-10).
The feedback resistor value is the reciprocal of the AMPERES range
When selecting the Multiplier Switch setting, remember
Check the caution in subparagraph 3a below.
However,
HOODELS 610B, 610BR ELECTROMETERS
follow these cautions:
Otherwise, the
Another alternative is using the
Do not use the fast method to measure capacitor leakages.
C)
a capacitor to the input causes the circuit to be transformed into a differ-
entiator, resulting in extreme sensitivity to very smell voltage transients.
Galvanometric Method.
e.
Operate the Model 610B as a picoarmneter in the fast method of operation.
1.
Use en accurate reference current source to buck out the unknown current source.
Connect as shown in Figure 6.
Set the METER Switch to CENTER ZERO and use the higher current ranges.
2.
Adjust the buckout current to indicate null on the l&de1 610B. Increase the
Electrometer’s sensitivity as needed.
as possible, the known reference current source equals the unknown source f
?hn Mrvi!a1 6,“R c-l,Yv-cxnr r!s,<rl,no
>
When the Model 610B is as close to null
current
Source
Connecting
FIC;URE b.
reference current source to buckout the unknown current source, I,. The
Xodel 6LOB, on its current ranges, serves es a null detector.
fitting at the Xodel 6lOB input.
with coaxial cable.
graph Z-3).
12
Measuring Current by the Galvanometric Nethod. Use an *ccurs.te
Use a uhf-tee
Connect the Electrometer to the two sources
Select cable carefully for very low currents (see para-
1064R
MODELS bLOB, 61OBR ELECTROMETERS OPERATION
2-7. Resistance Measurements.
a. The Model 610B can measure resistances by three methods.
In the normal or two-terminal method (annneter-voltmeter) , the ELec-
1.
trometer measures the voltage drop across the unknown sample as a known,
constant current flows through it.
the resistance of the sample.
The voltage drop is proportional to
This method is the simplest for the 100
to LO11 ohm ranges.
Above LOLL ohms or to prevent leakage,
2.
the guarded method is better.
It results in faster response speeds and also nullifies Leakage errors
across the Electrometer input,
since the potential across the input ter-
minal is small.
3. In the preceeding methods,
arbitrarily set.
In some cases,
the voltage across the sample cannot be
as in measuring capacitor leakage, these
methods involve much more time than if a Larger voltage could be applied.
In the external voltage method the Model 6LOB Fs used as a fast picoammeter.
The unknown resistance sample is connected to an external known voltage
source and the current through the sample is measured,
or fast method may be used.
The resistance is calculated from the readings.
Either the normal
NOTE
Discharge any capacitor before removing it from the circuit.
Depressing the ZERO CHECK Button shorts the fnput through a
lo-megohm resistor, providing a discharge path.
b. Normal Method (100 to 10” ohm ranges).
1.
Follow the instructions of paragraph 2-4.
Set the controls as follows:
METER Switch
Range Switch To5 OHMS
Multiplier Swftch 1
FEEDBACK Switch NORMAL
ZERO CHECK Button LOCK
Connect the resistance sample to the INPUT Receptacle. Unlock the ZERO
CHECK Button. Check zero with only the ZERO CHECK Button,
Do not open-circuit the Electrometer on the OHMS ranges; the
input will develop a Large voltage due to its constant current
characteristic. Keep the input shorted or the ZERO CHECK
Button locked.
2. The full-scale ohms range is the Range
Switch setting times the
Multiplier Switch setting. Use the Largest Multiplier Switch setting
possible to obtain the best accuracy.
1064R
13
OPERATION
Before making a final reading, manipulate the Multiplier end Range
3.
MODELS 6LOB, 6LOBR ELECTROMETERS
Switches, so the sample is tested at a number of test potentials. The
applied test voltage is the meter reading times the Multiplier SwFtch setting.
NOTE
Shield the input if the resistance sample exceeds one megohm.
Guarded Method (to 10L4 ohm ranges).
C.
Follow the instructions of paragraph 2-4.
1.
Set the controls as follows:
METER Switch
Range Switch
Multiplier Switch
;011 OHMS
1
FEEDBACK Switch FAST
ZERO CHECK Button LOCK
Connect the low impedance side of the resistance sample to the Model 610B
GUARD Terminal, and the high impedance side to the INPUT Receptacle. Unlock
the ZERO CHECK Button.
2. Read the resistance as outlined for the normal method, subparagraph b.
External Voltage Method (To ~016 ohms).
d.
1. Turn the ZERO CHECK Switch to L,OCK.
Connect the sample between the
INPUT Receptacle and the power supply. (See Figure 7.) Put a switch in
the high voltage line to ground the low impedance end of the sample when
it is disconnected from the ootential.
from a known source, V, is applied to the unknown resistance sample, &. The
Model 610B measures the current through &,
lated.
Switch S grounds s when no potential Fs applied.
from which the resistance is calcu-
2. Set the FEEDBACK Switch to NORMAL. Usually this method is best, since
instabilities can arise if the resistance sample is Less than .L the value
of the feedback resistor.
3. Apply a potential to the sample before releasing the ZERO CHECK Button.
Set the Range Switch to .3 AMPERES and increase sensitivity until a reading
is obtained.
14
1064R
MODELS 6108, 610BR ELECTROMETERS OPERATION
4. If the potential applied is at least 100 times the full-scale input drop (Multiplier Switch setting), the resistance is equal to the applied potential divided by the
current reading.
The high voltage sensftivity of the Model 61OB, therefore, permits
external voltages of .l volt or more to be used.
5. If the potential applied is less than 100 times the input drop, the resistance is
e~qual to the difference between the applied potential and the input drop, all divided
by the current reading.
6. If the current is read by the fast method,
need not be included in the calculation.
If the capacity shunted across the sample is
the input drop is so slight that it
large, such as encountered in capacitor leakage measurements, the fat method increases
response speed and this connection is reconrmended.
2-&Charge Measurements.
a. Follow the instructions of paragraph 2-4.
Set the controls as follows:
METER Switch CENTER ZERO
Range Switch 10-7 COULOMBS
Multiplier Switch
.Ol
FEEDBACK Switch FAST
ZERO CHECK Button
LOCK
Unlock the ZERO CHECK Button and then connect the unknown source to the INPUT Receptacle.
If the Electrometer reads off scale, increase the Multiplier Switch setting.
If the sensi-
tivity is not enough, decrease the Multiplier Switch setting until the reading is on scale.
Changing the Multiplier Switch setting does not affect the transfer of charge from the un-
known source to the instrument. If increasing sensitivity with the Multiplier Switch does
not .bring the reading on scale,
increase sensitivity with the Range Switch and repeat the
above steps.
The full-scale charge range is the Range Switch setting times the Multiplier Switch
b.
setting. Grid current contributes 2 x LO-l4 coulomb per second maximum.
NOTE
Because of the instrument’s RC time constant,
internal capacitance on the 10.’
coulomb range before making another measurement.
wait 20 seconds after discharging
On the 10-B coulomb range, wait at least two seconds.
2-g. Recorder Outputs.
a.
For recording with the Model 610B, use the Keithley Model 370 Recorder for ease,
~COUXlly, versatility and performance. The Model 370 is a pen recorder with 10 chart speeds
and 1% Linearity. The Model 370’s input cable has a connector which matee directly with
the OUTPUT Connector on the Model 610B; this avoids interface problems often encountered
between a measuring instrument and a. recorder. No special wiring is needed. No recorder
preamplifier is required. (See paragraph 5-B Eor more on the Model 370.)
b. Other recorders, oscilloscopes and similar instruments can be used with the Model 6108.
The Model 610B has two variable gain outputs,
nals within 112% for recorders,
oscilloscopes and similar instruments.
?3 volts and *l milliampere, to amplify sig-
These can be used
on all ranges of the Model 610B.
0667R
I.5
OPERATION
MODELS 6lOB, 610BR ELECTROMETERS
c. 3-Volt output.
Receptacle. Pin no.
Switch to 3 V.
any range.
The Model 6lOB output is now +3 volts for full-scale meter deflection on
Internal resistance is 3 kilohms.
300 cps at a gain of 3000,
gain of 0.03.
Noise is 3% rms of full scale at a gain of 3000, decreasing to 1% at gains
Connect oscilloscopes and pen recorder amplifiers to the OUTPUT
1 is the negative terminal; pin no. 2 is grounded.
Set the Output
The frequency response ($3 db) is dc to
rising
to
25 kc at a gain of 30, and decreasing to 2.5 kc at a
below 100. The METER Switch does not reverse the output polarity.
NOTE
The Model 610B may be used with the FEEDBACK Switch fn FAST position with other
instruments.
meter us3
However,
end the other instrument.
d. l-Milliampere Output.
General Electric or Texas Instrument Rectiriter, to the OUTPUT Receptacle.
the negative terminal.
make sure there is no common ground between the Electro-
Connect l-milliampere instruments, such as Esterline Angus,
Pin no. 1 is
Set the Output Switch to 1 MA.
The output is approximately 1 mil-
liampere for full-scale meter deflection on any range. For exact output, adjust the meter
on the .003-volt range with the FINE ZERO Control for full-scale deflection. Then adjust
the 1 MA CAL Control until the recorder reeds full scale.
zero end repeat adjustment if necessary.
The METER Switch does not reverse the output
Check the recorder and meter
polarity.
For servo rebalance recorders,
e.
See Figure 8.
Set the Output Switch to 1 MA.
for full-scale recorder deflection.
use a divider
Operation is the same as for current outputs.
across
the Model 610B OUTPUT Receptacle.
Use the 1 MA CAL Control to trim the output
1-$
2)
Model
6108
Output
(J105)
100~millivolt output
1 kn
100 n
loo-mv
Recorder
:
FIGURE 8.
A
Model
610B
1 kn
output
*
(JlO5)
50 0
It 1
50-mv
Recorder
50-millivolt output
Divider Circuits Across Model 610~ Output. The dividers are for driving 50
and 100-millivolt recorders. Use 1% resistors in the dividers.
f. When the FEEDBACK Switch is in the NORMAL position,
terminal is grounded to the instrument case.
Therefore, no difficulty will be experienced
using oscilloscopes end recorders with the Model 61OB set for normal operation.
position, however, neither side is grounded.
If this is used, make sure there is no com-
the negative side of the output
In FAST
mon ground between the recorder or oscilloscope end the Model 6108 case, or use the unity-
gain output.
16
;:
10&R
MODELS 6108, 610BR ELECTROMETERS
OPERATION
2-10. Unity-Gain Output. The unity-gain amplifier can be used as an
impedance matching device to minimize circuit loading errors or for convenient
connections to a recorder when the FEEDBACK Switch is in FAST position.
The unity-gain output is equal to the input within 50 ppm or 100 microvolts
a.
when the load resistance is 100 kilohms or better below 10 volts or 1 megahm or
better above 10 volts. By placing the Model 610B between a 1010 ohm source, for
example, and a 0.01% voltmeter with a 1-megohm input resistance, overall accuracy
better than 0.025% can be achieved.
1. Connect the voltmeter to the Xl OUTPUT and GUARD Terminals as shown in
Figure 9. The GUARD Terminal is at ground with the FEEDBACK Switch in NORMAL.
Maximum output amplitude is 100 volts peak-to-peak.
2. Adjust the Model 610B Zero Controls to obtain a zero-voltage reading on
the instrument using the unity-gain output. Make sure the latter’s sensitivity
is high enough for a precise zero adjustment. This adjustment is necessary
because a slight zero shift may occur when the Model 6108 is changed from the
O.l-volt range or lower to a range above 0.1 volt. The shift, caused by a
gain-reducing network switched in by the amplifier on the 0.3-volt and higher
ranges, is too slight to be read on the meter, but it can cause an error in
accurate measurements using the unity-gain output.
It 000000
‘Vx
t
cl
6108 y1
+ -
0.01% VM
V
I
GUARD
$-
FIGURE 9. Measuring Potential of High Resistance Source with 0.025% Accuracy.
The Node1 6108 is used between a high-resistance source, V,, and a 0.01% voltmeter to obtain high accuracy without causing circuit loading.
The digital volt-
meter or, as above, the Keithley Model 662 Differential Voltmeter connects to
the Model 6108 unity-gain terminals.
b. When the FEEDBACK Switch is in FAST position, the unity-gain terminals per-
mit more convenient connections to recorders without special precautions. In
this mode, the Xl OUTPUT Terminal is grounded and the GUARD Terminal delivers a
full-scale output equal to the hltfplier Switch setting or the input.
2- 11, CUP.RENT SOURCE. The Model 6108 can be used as a -4% current source from
10-j to lo-11 ampere.
Set
a.
the FEEDBACK Switch to NORMAL, the Range Switch to OHMS and the DIETER
Follow these
procedures:
Switch to + or -
b. The current supplied at the INPUT Receptacle is the reciprocal of the OHWS
Setting on the Range Switch. (For example, log OHMS indicates LO-9 ampere current
at the INPUT Receptacle.)
c.
It
does
1264R
The ,%ltiplier Switch does not affect the current at the INPUT Receptacle
afEect the maximum input voltage drop,
which is equal to the ?lultiplier
;:
OPERATION
XODELS 6108,
6tOBR EL&CTRO>IETL:{S
Switch setting.
Z-12.
STATIC CHARGE MXASUREMENTS.
For accurate output current, check the meter zero on the .Ol-volt range
Electrometers are very sensitive to static charges
and can readily make qualitative or quantitative measurements.
Zero the Model 6108; set the FEEDBACK Switch to NORMAL and the Range Switch to VOLTS
(l:.or 30 volts full scale).
PDT Receptacle.
Depending upon the distance between the charge and the instrument, a voltage will be induced on the input terminals which can be read on the meter.
frequently,
since accumulation of charge due to the electrometer tube grid current wilt
Bring the charged object near the uncovered, unshielded I!J-
Check zero
cause a slow drift of the inwt voltage.
b. Connecting a capaci’tor acroee the input reduces the drift due to grid current and
also the sensitivity to charge. Therefore,
set the Range Switch to COULOMBS position.
capacitor value connected across the input in farads is equal to the COULOMBS Range.
2-13. CAPACITANCE MEASUREMENTS.
The Model 610B can measure capacitance from 500 pico-
farads to LOO microfarads. The Electrometer charges the capacitor to a known potential
and then measures the charge. The resulting capacitance is easily figured.
a.
Charge the capacitor as follows:
1. Set the Model 610B front panel controls to:
METER Switch
+
Range Switch VOLTS
Multiplier Switch
.OOl to 100
FEEDBACK Switch NORMAL
ZERO CHECK Button
LOCK
The
2. Connect the unknown capacitor to the INPUT Receptacle. Unlock the ZERO CHECK Button and charge the capacitor to a known voltage by setting the Range Switch to the O~DIS
ranges. As the capacitor charges, the meter will advance up scale to show the voltage
(meter reading times the Multiplier Switch setting) acroee the capacitor at any given time
3.
Charge the capacitor to a convenient voltage, such as 0.1, 1, 10, etc. Charge
Large capacitors to a Lower voltage than that used for low value capacitors. The OHMS
ranges control the charging rate: start with the Lower ranges and increase the setting
as the voltage acroes the capacitor reaches the desired value.
4. When the capacitor reaches the desired value,
VOLTS position, then disconnect the capacitor from the INPDT Receptacle.
quickly set the Range Switch to the
Or, if more
convenient, first disconnect the capacitor from the INPUT Receptacle, then set the Range
Switch to the VOLTS position.
NOTE
Be careful handling charged capacitors.
to 100 volts.
If the capacitance is above O.OL microfarad, a dangerous potential
The Model 6108 can charge a capacitor
could exist.
b. Measure the charge on the capacitor following the instructions in paragraph 2-B
18
1066R
MODELS 610B, 610BR ELECTROMETERS
OPERATION
c. The value of the unknown capacitance is the stored charge divided by the
initial voltage:
c (farads) = Q (=‘ulo~S~
v (volts)
Z-14. APPLICATION NOTES.
The versatility of the Model 610B is such that it
is almost a complete dc laboratory in itself.
follow;
the purpose of these is to suggest uses and techniques to increase the
usefulness of the Model 610B.
a. current 1nceerator.
currents, Figure 10 shows the circuit.
The Model 6lOB works as an integrator for time-varying
Set the Range Switch to COULOMBS.
the unity-gain output. Output voltage, V,ut, is
T
Vout
=+
/
0
i dt
where C is the coulomb range setting in farads
610B 610B
FIGURE 10. Current Integrator.
current integrator.
Model 610B input.
A square wave from a current source, I,, is applied to the
Using the COULOMBS ranges, the output through the
The diagram shows the Model 610B acting as a
unity-gain terminals is shown.
Some particular applications
Xl Xl
b b
GUARD GUARD
L L
Use
b. Potentiometric Voltage Measurement%. The high input impedance, l-millivolt
sensitiv~ity and low zero drift make the Model 6108 useful as a potentiometer.
The circuit shown in Figure 11 is useful when no loading of the source voltage,
on or off null, can be tolerated. Make the measurements with the center zero
scale, following the procedures in paragraph 2-5. When the Model 610B is at null
- no meter deflection - the voltage from the known source equals the unknown
voltane.
Voltage Supply Voltage Supply
+ - + -
,
FIGURE 11. Potentiomecric Voltage Measurements.
used to buck out the potential from the unknown source, V,.
An accurate voltage supply is
The Node1 610B on
the voltage ranges and center-zero scales acts as a null detector.
1264~
19
OPERATION
Measuring Diode Characteristics.
C.
The Model 610B can accurately measure
diode characteristics in one simple step
without using any other equipment. The
circuit is shown in Figure 12. The
measurements are made on the Model 610B
OHMS ranges
currents d&ns~~e~,“~~ z$y:n decade
steps and reading the voltage drop
across the diode. The characteristic
curve can then be plotted in volts
and amperes or volts and ohms.
Read the voltage from the meter.
d. peak-Readina Voltmeter. The Model
610B easily converts to a peak-reading
voltmeter. Set the Range Switch to
the COULOMBS ranges and apply the input voltage to the INPUT Receptacle
through an external diode. In the
circuit shown Fn Figure 13, if the input resistance (RF) and diode leakage
resistance (Rd) are high and if the time constant of the RC combination is
high compared to the period of the signal, the capacitor will charge to the
peak value of the applied ac voltage minus the small drop across the diode.
The Model 6LOB input resistance on the VOLTS or COULOMBS ranges is 1014 ohms
or, greater, and may therefore be neglected compared to the leakage resistance
of most diodes. The capacitor (C) Fs selected using the COULOMBS ranges with
the FEEDBACK Switch in the NORMAL position. C in farads is equal to the
COULOMBS range setting.
MODELS 6 LOB , 6 LOBR ELECTROMETERS
Itl
610B
10
I
FI :Gum 12. Measuring Diode Characte
istics with Model 610B.
ranges of the Electrometer.
Use the OHMS
0
0 hl
VI
Input
.I101 c
0 ST
I
FIGURE.. 13. Diagram of Model 610B as Peak-Reading Voltmeter.
is used on its COULOMBS ranges.
is the capacitor selected with the Range Switch; Ri is the Model 610B input
resistance,
of the diode used.
20
Accuracy of the measurements depends mainly upon the quality
1
0 L
1
The diode is external to the voltmeter, C
Amplifier
. II.
and
Meter
Ri
The Model 610B
1064R
MODELS 610B. 610BR ELECTROMEl’ERS
CIRCUIT DESCRIPTION
SECTION 3.
3-1.
meter with a full-scale sensitivity of 1 millivolt and an input impedance of
LOL4 ohms shunted by 22 picofarads.
resistors and capacitors are selected to make measurements over a total of
79 voltage, CUTrent, resistance and coulomb ranges. Current and resistance
are measured using precision resistance standards, from lo-ohm wirewound
resistors to 1OLL ohm glass-sealed, deposited carbon resistors.
are measured using close tolerance polystyrene film capacitor standards.
the necessary amplifier power.
3-2.
by three differential transistor stages and a vacuum tube output stage. A
Large amount of negative feedback ts used for stability and accuracy.
Figure 14 shows the block diagram for the voltmeter mode of operation.
GEINEML.
a. The Keithley Model 61OB is basically an extremely stable Linear de volt-
A super-regulated, low voltage transistorized power supply furnishes
b.
VOLTMETER OPERATION.
a. The voltmeter amplifier has matched electrometer input tubes followed
Input
CIRCUIT DESCRIPTION
By using the front panel controls, shunt
COUlOmbS
)I-
a01 t
%
i
I
Outnut Gnd.
F1GlJR.B 14.
refer to schematic diagram.
across the amplifier.
for a given Multiplier Switch setting. e, is the voltage drop across F,,,.
When thereis no input signal and the zero is set, the current (i,) in the
b.
upper Loop of Figure 14 equals that in the lower Loop (ib): La = ib. With an
input signal,
When a positive voltage, cei, is applied to the input, the amplifier
1.
the circuit reacts as follows:
drives the V103 grid less negative, and the plate current, i,, is increased.
Increasing ia
Rm. At equilibrium this drop equals ei,
drop, ea,
when a negative voltage, -ei, is applied to the input, the amplifier
2.
causes a voltage drop through the E6rLtipLier Switch resistor,
and therefore, it keeps the voltage
across the amplifier to a fraction of the input voltage.
drives the VL03 grid more negative with respect to the cathode, and ia is
decreased. The current in the Lower Loop, Fb, becomes greater than i,,
producing a VOLtage drop, eo, across R,,,. The drop is sufficient to null the
negative input voltage, ei, again keeping the voltage drop across the ampli-
fier to a fraction of the input voltage.
The voltage drop across the amplifier is
3.
ea =
=i
kc1
where k is the Loop gain, approximately LO5 on the 0.001 to 0.3-volt ranges.
c. The output stage, V103, drives the amplifier ground at the same poten-
tial as the input signal. This means the ground terminal of the amplifier is
not at chassis ground, but is connected directly to the cathode of V103. Ikillg
this circuit, the input can accept voltages up to rlO0 volts without input
dividers. This maintains high input impedance and eliminates instabilities
which can occur with high resistance dividers.
NOTE
On the schematic diagram, amplifier ground is called “floating ground”
and the output amplifier ground, “output ground.”
All power supplies are floating with respect to the input or chassis
d.
ground. The +250 and -265 volt supplies are provided for the output voltage
amplifier, V103. Separate regulated -10 and +21 volt supplies provide power
for the amplifier and are.referred to floating ground.
NOTE
Refer to Schematic Diagram 17795E for circuit designations.
3-3. VOLTMETER CIRCDIT.
a. Two balanced 5886 electrometer tubes are at the voltmeter amplifier input.
Their filaments are operated in parallel from the regulated ~21 volt supply
through dropping resistors, R119 and R122.
control grid of VlOl, the active electrometer tube,
due to overloads.
Capacitors Cl01 and Cl10 are high-frequency bypasses. The
Resistors RlOL and R113 protect the
from excessive grid currents
control grid of VL02 is returned to floating ground,
22
1064R
MODELS 610~, 6LOBR ELECTRO~BRS
CIRCDIT
DESCRIPTION
Depressing the ZERO CHECK Button,
b.
to output ground. This effectively removes all signal from the grid of VLOL,
and the input impedance is reduced to 10 megohms.
c. An emitter follower stage, transistors QLOL and Q102, matches the relatively high impedance output of the electrometer tube stage to the Low input
impedance of the differential amplifier stage formed by transistors 4103 and
9104.
sistors QLOS and QlO6. Resistors R152 and RL53 prevent Lock-up of the amplifier under overload conditions.
Resistor RL84 protects V103 from drawing excessive grid current; capacitor
Cl13 is a high-frequency bypass.
tube screen grids. The screen grids of VlOL and VLOZ are returned, in
effect, to the emitters of transistors Q103 and 4104 through the COARSE
and MEDIUM ZERO Switches, SL04 and SLO5. The emitter voltage of Q103 and
9104 can vary, resulting in a negative feedback loop for signals in phase at
the electrometer tubes through the 9103 and Q104 emitter circuit back to the
VLOL and VlOZ screen grids. This connection stabilizes the electrometer plate
potential and tube operating points. Also, for signals arriving at the VZOL
control grid, the first stage gain will be much greater than spurious signals.
RL66, determines the voltmeter sensitivity. A l.l-milliampere current
through the resistors produces a full-scale meter deflection. On the higher
voltage ranges, resistors R171 through Rl73 are across the collectors of Q103
and QL04 to reduce the amplifier gain.
when maxiawa gain is unnecessary.
This latter stage drives a second differential amplifier stage, tran-
d. Transistor Q106 drives the grid of the output voltage amplifier, VL03.
e. The zero balance controls adjust the dc voltages of the electrometer
f. The voltage drop across the l%.zLtiplier Switch resistors, Rl56 through
SlO3, connects resistors RLOL and Ill13
This assures the amplifier stability
The recorder output is derived from the current flow from VlO3 through
8.
the ~Ltiplier Switch resistor, With the Output Switch, S108, on 3V, ?3 volts
for full-scale deflection are obtained at the output connector, 3105, by r1.L
milliamperes flowing through resistor R180.
and R179 are connected across 5105, allowing 21 milliampere *5% to pass through
an external Load.
h. The second triode section of V103, connected as a diode, provides warmup compensation for the -265 volt supply.
conditions during warm-up and cool-down, which could cause excessive eLectro-
meter tube grid current to flow and result in poor drift characteristics.
The "normal" and "fast" referred to below are only the
positions of the FEEDBACK Switch. "Normal method" is
when the Switch is set to NORMAL; "fast method" is when
the Switch is set to FAST.
1064R
With S108 at 1 MA, resistors R174
This prevents severe out-of-balance
23
CIRCUIT DESCRIPTION
MODELS 610B, 6LOBR ELECTROMETERS
.I101
Amplifier
Rs (
I
FIGURE 15.
refer to schematic diagram.
for a given setting. Rs is the resistor selected by the Range Switch, SlOl.
5102 is the FEEDBACK Switch. Ground references correspond to schematic.
3-4. AMMEIER OPERATION.
a. Normal Methc.4.
Switch in NORMAL position), one of the Range Switch resistors, R102 through
Rl12, shunts the input. (See Figure 15.) The Modal 6lOB than measures the
voltage drop acros8 the resistor. The meter is calibrated to read the current
in amperes for the appropriate range.
Block Diagram of Model 610B as a Picoanuneter. Circuit designations
In the normal method of current measurements (FEEDBACK
.
FAST
SlO6 is the tiltiplier Switch; I&, is the resistor
?
Floating Gnd
+250 v z
L
I
-265 v C
b. East Method.
in FAST position), the Model 610B functions as an
The differential amplifier output is divided by the tiltiplier Switch resistors,
R156 to R166, and fed back to the amplifier input through a feedback resistor
selected'with the Range Switch.
to the low impedance side of the input, and the output ground is floating.
method increases the
it also reduces the input drop to less than 1 millivolt.
3-5. OHhMEX'ER OPERATION.
a. NoMl Method In the normal method of resistance measurements (FEEDBACK
Switch in NORMALposition), the Model 610B uses a constant-current, voltage-drop
circuit. Refer to Figure 16. & is the unknown resistor. A voltage source, E,
applies a potential acros8 &.
through the resistor divider network, Rll4 through Rll8.
10 volts, depending upon the OHMS range used.
between floating ground and the input grid of V103 through R,, the range re-
sistor. Rs
24
is one of the resistors, R102 through R112.
In the fast method 0f current measurements (FEEDBACK Stitch
response
anmeter
(See Figure 15.) Floating ground is connected
speed by minimizing the effects of input capacity;
floating ground keepa the grid et nearly the same potnetial as the cathode,
the current, I, through Q and Rs is constant.
of the value of G,, as long as the voltage drop across R/ does not exceed the
Multiplier Switch setting.
the input.
indicate the resistance value on its calibrated meter.
The Model 610B can then measure the voltage drop across Rx and
This circuit provides a true source regardless of
I is equal to E/R,, regardless
+25ovz
I
t
I
I
Amplifier
Output Gnd
I (
1
I
1
FIGURE 16. Block Dianram of Model 610B for Normal Method of Measurine Re-
sistance. Circuit designations refer to schematic diagram.
mltiplier Switch; R,,, is the resistor for a given setting.
known resistance being measured; E is the voltage source (see paragraph 3-5);
R, 1s the range resistor selected with the Range Switch.
correspond to the schematic.
b. Guarded Method In the guarded method of resistance measurements (FEED-
BACK Switch inF&ition and the sample resistance connected between the
input terminal, JlOl, and the guard terminal, J104), feedback is applied through
the sample. Refer to Figure 17. The circuit is similar to the normal method,
except for the feedback. This reduces the slowing effect of the instrument's
input capacity.
input terminal is small.
low impedance side of the input and the output ground is floating.
terminal is at output ground porenfial.
Leakage error is also reduced since the potential across the
In this mode, floating ground is connected to the
SlO6 is ;he
G is the un-
Ground references
The guard
3-6.
similar to that used for an annneter with the fast method. A negative feedback
is applied around a shunt capacitor, Cl03 to C106, selected with the Range
Switch.
is proportional to the voltage across the capacitor, which is measured by
the Model 610B voltmeter circuits.
1064R 25
COULOMBMETER OPERATION.
The shunt capacitor replaces R, in Figure 15. The stored charge
The Model 610B circuit for measuring charge is
CIRCUIT DESCRIPTION
MODELS 6108, 610BR ELECTROMETERS
FIGURE 17. Block Diagram of Model 610B for Guarded Method of Measuring Resis-
tame. Circuit designatfons refer to schematic diagram.
Switch; R, is the resistor for a given setting.
G is the unknown resistor being
5106 is the Multiplier
measured; E is the voltage source (see paragraph 3-5); R, is the range resistor
selected with the Range Switch. 5104 is the GUARD terminal. Ground references
correspond to the schematic.
3-7. POWER SUPPLY.
nishes +250,
a.
A 24-volt
-265, +21 and -10 volts to the differential amplifier.
rectified by diodes D203 end D204 and is filtered by capacitor C204.
The power supply operates from the line voltage and fur-
rms output
from the power transfomer, T201, is full-wave
The dc
voltage across C204 is approximately 30 volts dc.
b. To obtain a stable, accurate voltage, the output of the series trms-
istor, Q202, is regulated by comparing e sample voltage from dividers R212
and R214 to the zener diode reference, D210.
If e voltage difference exists,
ft is amplified by a differential amplifier consisting of transistors 4204
and 4205. The Si@‘ld is further amplified by transistor 4203. The output of
4203 is applied to Q202 to nullify input and load variations. Capacitor
C208 prevents high-frequency oscillations.
The +21 volts at the output of
the regulator powers the amplifier, the filament of the electrometers, and
is used as the reference supply for the OHMS ranges.
c. Transistor 4203 operates at a high gain by conneceing its collector load
to a negative regulated supply. The circuit permits linear operation of Q202
with widely varying input voltages. To supply Q203, a 50-volt output of trans-
former T201 is half-wave rectified by diode D201 and is filtered by capacitors
C203 and C205. Resistor R201 is a dropping resistor. The voltage across the
zener diode D202 is a stable -10 volts, which is supplied to resistor R209,
the collector load resistor of 9203, and the differential amplifier.
26 1060
MODELS 6LOB, 610BR ELECTROMETERS
Transistor 4201 is an emitter follower whose function is to increase the
d.
current gain of the series transistor,
4202, forming a Darlington pair.
CIRCUIT DESCRIPTION
Resistors R208 and R209 and diode D208 provide current overload protection.
Excessive current drawn from the power supply causes an increased voltage
drop across R208, which forward biases diode D208, thus preventing the
collector of Q203 from going more negative. Since the collector voltage
cannot rise, further amplification is prevented and, therefore, further
current increase is prevented.
e. The ~250 volts and -265 volts for the output stage are obtained from
a single winding on transformer T20L. 210 volts from the transformer is
half-wave rectified by diodes D205 and D206, and filtered by capacitors
C206 and C207, respectively.
Resistor R203 provides short-circuit protection;
resistor R204 discharges the capacitors when the instrument is turned off.
1064R
MODELS 6108, 6106~ ELECTROMETERS
SECTION 4.
4-1.
the Model 610~ Electrometer.
ly as possible to maintain the accuracy of the instrument.
high-quality electronic equipment. The value of the high-megohm resistors, R110, RI11,
R112, should be checked every two or three years for specified accuracy. AlSJ,
amplifier balance requires occasional adjustment; see paragraph 4-8.
4-2. PARTS REPLACEMENT.
Electrometer.
meet the specifications.
from Keithley Instruments, Inc. In normal use,
10,OpO hours of operation. They can be checked only by replacement. Standard 5886 tubes
could be used in an emergency, but the drift, noise and grid current specifications may
not be met. When replacing the electrometer tubes with a matched set from Keithley, fur-
ther ageing of the tubes within the Model 6106 is required; a minimum of 3 days is recommender
GENERAL.
a. Section 4 contains the maintenance,
It is recommended that these procedures be followed a5 cLos?-
b. The Model 6108 requires no periodic maintenance beyond the normal care required of
a. The Replaceable Parts List in Section 6 describes the electrical components of the
Replace components only as necessary.
b. The electrometer tubes, VlOL and VlO2, are specially matched and aged; order only
MAINTENANCE
troubleshooting and calibration prticedures for
the dc
Use only reliable replacements which
they should not need replacement before
NOTE
When replacing the electrometer tubes, do not touch the glass base where the
leads emerge. Increased leakage will result from any contamination.
c. Transistor pairs QlOl, Q102 end Q103, QlO4 are matched for dc current (hFE). Order
only Ecom Keichley Instruments, Inc.; replace only as a pair.
4-3. TROUBLESHOOTING.
a. The procedures which follow give instructions for repairing troubles which might
occur in the Model 6108.
parts. Table 1 lists equipment recommended for troubleshooting.
Instrument
dc voltmeter,
input resistance,
Erom one volt to 300 volts.
Tektronic Type 561A Oscilloscope Observe wave forms in
TABLE 1. Equipment Recommended for Node1 610B Troubleshooting.
or their equivalents.
with minimum 100-megohm
LO% accuracy, range
Use the procedures outlined and use only specified repl.acement
If the trouble cannot
Use
Circuit checking
power supply
Use these instruments
1066R
‘9
MAINTENANCE
MODELS 6108, 610BR ELECTROMETERS
c
,
Difficulty
Excessive zero drift.
Excessive grid current.
microphonics. Defective electrometer
Cannot meter zero on
any range.
Meter
Multiplier Switch
setting.
10-l’ to lo-l4 ampere
current ranges are out
of specifications.
TABLE 2.
will not zero on
i.fodel 610B Troubleshooting.
I
I
Probable Cause
Electrometer tubes may be
defective.
DC amplifier not balanced. Check per paragraph 4-8
Excessive humidity or
&bfEctive electrometer
See paragraph 4-4.
Faulty resistor for set-
ting of mltiplier Switch.
Defective high megohm
resistors.
See paragraph 4-2 for checking VlOl and V102.
I
Check VlOl and VlO2;
replace if faulty.
Check VlOl and V102;
replace if faulty.
I
Check VlOl and V102;
replace if faulty.
I
3ee paragraph 4-4.
Check resistors; replace ff faulty.
Check per paragraph 4-9.
Solution
-
‘.
d
be readily located or repaired, Keithley Instruments, Inc., can service
the fnstrument et its complete service Facilities. Contact your nearest
representative.
Table 2 contains the more conrmon troubles which might occur. If the re-
b.
pairs indicated in the table do not clear up the trouble, find the difficulty
through a circuit-by-circuit check,
the circuit description in Section 3 to find the more critical components and
to determine their function in the circuit.
177953, is found in Section 6.
4-4. PROCEDURES TO GUIDE TROUBLESHOOTING.
If the instrument will not operate, check the fuse, line cord and power
a.
source. If these are all found satisfactory, use the following procedures to
isolate the trouble.
The schematic diagram indicates all tube element voLtages.and transistor
b.
terminal voltages referenced to floating or output ground; check Notes on the
diagram.
Multiplier Switch at 1, and the meter zeroed. Measurements are with a LOO-megohm dc voltmeter.
c. At times,
Adjust the COARSE ZERO Switch, S104 (Figure S),
balance. If this does not work, continue to check the circuits.
The control settings Eor these values are the Range Switch at VOLTS,
the meter will not zero on any range with the input shorted.
such as given in paragraph 4-4. Refer to
The complete circuit schematic,
to bring the Model 610B into
30
--r
1064R
MODELS 6108, 610BR ELECTROMETERS MAINTENANCE
Check the filaments of the electrometer tubes, VLOL and V102, by measur-
1.
ing the voltage drop across the filament dropping resistor, Rll9 (Figure 22).
IF both filaments are operating properly,
the voltage will read 2.0 volts dc
?lO%.
Check the output tube, V103 (Figure 23 ), to see that both filaments
2.
are lit.
Inspect the leads from the shock-mounted input printed circuit board,
3.
PC-84, to the amplifier printed circuit board, PC-85, for possible breaks.
d.
Su~oly. Power
1. Set the FEEDBACK Switch to FAST to connect the instrument's chassis to
floating ground.
supply and the chassis.
Connect a 100~megobm dc voltmeter between the -10 volt
(-10 V is indicated on PC-86, Figure 18.) The
voltage should be -10 volts dc 220%.
2. Connect the voltmeter between the +21 volt supply and the chassis.
(~21 V is indicated on PC-86, Figure 18.) The +21 volt supply can,not
operate unless the -10 volt supply is operating.
~'21 volts dc 210%.
A faulty ~21 volt supply can indicate a faulty zener,
The voltage should be
D210 (Figure 23).
Check the +2SO end -265 volt supplies; these should be correct within
3.
10%.
For this measurement, use output ground as the co@anon connection for
both supplies.
Amplifier.
e.
To check the amplifier, disconnect the feedback 150~ by removing the
1.
output tube, V103. This allows each stage of the amplifier to be individual-
ly checked.
It also eliminates the possibility of applying excessive voltage
to the electrometer tube grids, causing serious damage.
NOTE
Use a LOO-megohm dc voltmeter as a null detector to check the amplifier
stages.
2.
Adjust the COARSE and MEDIUM ZERO Controls for null.
Do not ground its low side.
Connect the voltmeter between the plates of VlOl and VlO2 (Figure 22).
If a null cannot be
reached, check VlOl, V102, the COARSE end MEDIUM ZERO Control circuits
(resistors R128 to RlSl), and transistors QlOl and Q102. Check the tran-
siscors by removing them and adjusting for null again.
replace the transistor pair with a new pair.
ed,
Check the next stage by connecting the voltmeter across the emitters
3.
lf null is now reach-
of QlOl and 9102 (Figure 21) end adjusting the COARSE and MEDIUM ZERO
Controls for null. If a null is not reached, check this stage end the base
circuit of the next stage. Check the base circuit by removing transistors
9103 and 4104 end again adjust for null. If null is now reached, replace
the transistor pair with a new pafr.
1064~
31
MAINTENANCE MODELS 610B, 610BR ELECTROMETERS
Check the next stage by connecting the voltmeter across the collectors
4.
of 9103 end Q104 (Figure 21), end adjusting the COARSE and MEDIUM ZERO Controls
for null.
base circuit of Q105 and QlO6.
5.
the FINE ZERO Control for null.
operating correctly and the trouble is in the output stage or the feedback stage.
6.
citor, Cl13 (Figure 23), for a possible short.
7.
the recorder output resistors, R180 on 3V position
position,
ing of the meter. An open multiplier resistor, however, prevents zeroing
for only that particular multiplier setting.
If a null is
Connect the voltmeter across the collectors of Q105 and Ql06. Adjust
Check the output stage cathode resistor, Rl8S (Figure 24)) end capa-
The feedback Loop includes the multiplier resistors, RlS6 through Rl66,
and the meter.
not
reached, check this stage and for shorts in the
If null is reached,
An opening of any of these components prevents zero-
the dc amplifier is
or
R174 and R179 on 1 MA
4-s.
power source unless otherwise ordered.
use a screwdriver to change the slide switch on the beck panel to 234.
the fuse from l/4
To switch from 234 to 117-volt operation, reverse the procedures.
4-6.
the mdel 610B.
are not avaflable or if difficulty is encountered, contact Keithley Instruments,
Inc., or its representatives to arrange for factory calibration.
fier balance adjustment, high-megohm resistor verification, calibration of the
ohms ranges, meter zero calibration and
Keithley Instruments Model 241
Regulated High Voltage Supply
Keithley Instruments Model 261 Picoampgre spurce* fr#m LO-L4 to LOe4 ampere,
234-VbLT OPEBATION.
ampere
CALIBRATION PROCEDURES.
The following procedures are recoranended for calibrating and adjusting
'a.
Use the equipment reconrmended in Table 3. If proper facilities
b. Procedures
Check grid current (paragraph 4-7) et regular intervals to make sure
1.
the electrometer tubes are functioning
0 1.5/,.
are
covered for the following: grid current check, dc ampli-
Instrument
The Model 610B is shipped for use with a 117-volt
To convert
to l/8 ampere. No other adjustment is necessary.
accuracy
correctly.
Calibrate meter zero and verify
voltage range accuracy.
Calibrate current, resistance
and charge ranges.
it
for 234-volt sources,
check.
Use
Change
Keithley Instruments Model 370 Recorder
Keithley Instruments Model 515
Megohm Bridge
KeTCh1eyRrsrrnment.s Model 662
TABLE 3.
ments or their equivalents.
32
Equipment Recommended for Model 610B Calibration. Use these inetru-
Zero drift check.
Verify high-megohm resistors in
Range Switch.
Calibrate resisrmrre-ranges
and meter zero.
0667R
MODELS 610B, 610BR ELECTROMETERS
Control
MAINTENANCE
Circuit
Designation
Ohms Calibration
Meter Center Zero Calibration
Meter Calibration
DC Amplifier Balance
TABLE 4. Model 610B Internal Controls.
Rll5 18 4-10
R167
R175
RL77
18 4-11
18 4-11
19 4-8
The Table Lists all internal controls,
the figure picturing the location and the paragraph describing the adjustment.
2. The dc amplifier balance adjustment (paragraph 4-8) is necessary about
every six months or when amplifier components are replaced.
3. Verify the value of the high-megohm resistors (paragraph 4-9) about
every six months to check on drifting.
4. Calibrate the ohms ranges (paragraph 4-10) about once a year or when
range resistor or power supply components are replaced.
5. Calibrate the meter zero (paragraph 4-11) about once a year or when
components are replaced.
Check the Model 610B (paragraph 4-12) once a year, after the other
6.
adjustments, or if improper operation is suspected.
c. If the tide1 610B is not within specifications after the calibrations
and adjustments, follow the troubleshooting procedures or contact Keithley
Instruments, Inc.,
or its nearest representative.
4-7. GRID CURRENT
CHECK.
Check grid current whenever excessive noise or drift
is suspected. To read the grid current of the Model 610B, set the front panel
controls to:
METER Switch
+
tiltiplier Switch .Ol
Range Switch
ZERO CHECK Button
FEEDBACK Switch
Cap the INPUT Rece
less than 2 x 10-L
tacle.
e
ampere. If this is exceeded, check the +Zl volt supply
The grid current indicated on the meter should be
10-11 AMPERES
Unlocked
FAST
and the electrometer Tube, VlOl.
4-8.
DC AMPLIFIER BALANCE ADJUSTMENT.
an occasional adjustment of the amplifier balance.
components are replaced.
This adjustment sets the gain to allow the meter to
Gradual aging of components may require
Also adjust if amplifier
receive the correct signal on all ranges.
1064R
33
MAINTENANCE
MODELS 610B. 6lOBR ELECTROMETERS
FIGURE 18.
Model 610B Internal Controls.
+21 volt power supplies are also shown.
Test points for the -LO volt and
Refer to Figure 19 for the DC AMP
BAL Control.
a. Set the front panel controls to:
METW Switch
Multiplier Switch
Range Switch
ZERO CHECK Button
FEEDBACK Switch
CENTER ZERO
0.1
VOLTS
LQCK
NORMAL
Zero the meter on the 0.1~volt range.
b. Connect the floating Model662 across the collectors of transistors QlO5
and Q106 (this is across the ends of resistors Rl81 and R182, Figure 21). Make
the connections from the component side of the printed circuit board.
Float
the Model 662 ground terminal.
c. Adjust the DC AMP BAL potentiometer,
R177 (Figure 19)) until the meter
shows a null.
4-9. HIGH-MEGOHM RESISTOR VERIFICA!l!ION.
a. About every six months, it is necessary to check the value of the high-
megobm resistors, RllO to R112, on the Range Switch.
The instrument should be
within its rated accuracy for two or three years from the time It leaves the
factory. After this, smne of the resistors may drift cut of tolerance and should
be replaced.
Faulty high-megohm resistors will affect the accuracies of measure-
ments for the 10-9 to LO-LL AMPERES and the LOB to 10L2 OHMS settings of the
Range Switch.
MODELS 610B, 6lOBR ELECTROMETERS
MAINTENANCE
To check these
resistors
thban 1% accuracy up to 1011 oh&.
it is necessary to use a bridge capable of better
An accurate megobm bridge, such as the Keithley
Instruments Model 515 Megohm Bridge which is accurate to 0.25% for these
ranges, is therefore necessary.
cedures
are
recommended to check out the resistors:
Return the complete instrument to the factory for resistor calibration.
1.
If such equipment is not available, two pro-
2. Replace the high-megohm resistors periodically with a certified set
from Keithley Instruments to assure absolute calibration accuracy.
Connect the Model 662 to the junction of resistors RlL5 and R116
(Faigure 24) and the conrmon lead to the + terminal of capacitor C203
(Figure 23). Adjust the OHMS CAL potentiometer, RlL5 (Figure 18), for 10
volts dc 20.5%.
If the control does not have enough latitude, check the -21 volt
b.
supply (paragraph 4-4).
4-11. METER ZERO CALIBRATION.
Check meter zero whenever components are
replaced or other adjustments are made.
a. Turn the METER Switch to METER OFF.
Set the mechanical Zero Meter adjust-
ment for zero meter reading (top-scale zero).
b. Turn the Model 610B on.
Zero the meter on the .OOl-volt range. Then
switch to the LO-volt range. Set the Multiplier Switch to 10; apply 10
volts ?0.05% to the tide1 6108 INPUT Receptacle. Adjust the METER CAL
potentiometer, R175 (Figure 18), for full-scale meter reading.
Set the center zero by first zeroing the meter on the .OOl-volt range.
c.
Then switch to the l-volt range. Set the METER Switch to CENTER ZERO and
adjust the CENTER ZERO CAL potentiometer, R167 (Figure 181, for exact
center-scale meter
4-12.
operation;
ACCURACY CHECK.
a.
Checking the accuracy is the quickest way to spot improper Electrometer
Perform the check about once a year, if components are replaced,
or if other adjustments made.
zero.
If accuracy is verified over all ranges,
the Model 610B should be able to meet all specifications. If the accuracy
must be checked often, check for zero drift.
b.
Voltaee.
Connect the Model 610B to the Model 241 Power Supply. First,
set the Model 610B for the lo-volt range. Increase the input voltage in l-
volt steps.
full scale.
The Model 610B should indicate the input voltage to ?l% of
Check the other voltage ranges for accuracy at full scale.
Current.
c.
Connect the Model 610B to the Model 261 Picoampere source.
Check the full-scale accuracy of all current positions on the Range Switch,
For the
1 to lOA
ampere ranges,
set the FEEDBACK switch to XORMAL.
For
066iR
35
MAINTENANCE
MODELS 6108, 610BR ELECTROMETERS
the 1O-5 to lo-l1 ampere ranges,
set the FEEDBACK Switch to FAST.
The Model
6108 should indicate the input current to ?2% of full scale from the .L to 10-B
ampere ranges,
%% of full scale from the 10-g to IO-l1 ampere ranges.
d. u. All resistance ranges will be within specifications if the current
ranges are within specifications and if the ohms range calibration (paragraph
4-10) is correct.
coulombs. Set the FEEDBACK Switch to FAST and the tiltiplier Switch
e.
to 3.
Use the current source to provide an input current. With the Range
No other procedures are necessary.
Switch set to the COULOMBS ranges given in Table 5, apply the corresponding
input current. Use a stop watch to time the rise to full-scale deflection.
Minilmml
COULOMBS Range
(Range Switch)
Input Current Rise Time
To Model 610B (Zero to Full Scale)
10-7 10-B ampere 30 seconds
10-B
10-g
10-10
TABLE 5.
Coulomb Ranges Accuracy Check, The table gives the input current
needed to check the rise time for each coulomb range.
4
4-13. DRIFT CHECK.
Refer to the accuracy check before performing the drift
check.
Set the front panel controls to
a.
METER Switch +
tiltiplier Switch
.03
Range Switch VOLTS
ZERO CHECK Button LOCK
FEEDBACK Switch
NORMAL
Set the Output Switch on the back panel to 3 V.
Connect the Model 610B OUTPDT Receptacle to the Model 370 Recorder.
b.
recorder to 1 volt.
Make sure the Model 610B chassis cover is attached with
Set the
at least two screws.
Make the drift run for at least 24 hours.
c.
The zero drift specification is
200 microvolts per hour maximum averaged over any 24-hour period after warm-up;
during two-hour warm-up, no more than 2 millivolts after the first hour(Refer to
Chart 1).
Change electrometer tubes if the instrument does not meet the zero drift
d.
specification.
36
1066R
X~DELS 610B, 610BR ELECTROMETERS
NAINT'ZNILUCE
CHART 1. Typical drift run of Keithley Model 610B.
of the Electrometer.
second hour.
In any subsequent 24-hour period,
After l-hour warm-up the drift is no more than 2 millivolts in the
the average drift will not exceed 200
microvolts per hour.
The drift run shows the stability
0166R
FIGURE 19.
Side View of Model 610B Chassis.
Front panel faces right.
of components, printed circuits and switches is shown.
for circuit designations.
Figure 20 shows ehe opposite side.
Location
Refer to the Parts List
37
MAINTENANCE
MODELS 6108, 610~~ ELECTROMEXERS
~rinced tircuir 86
23, 24
FIGURE 20.
Side View of Model 610B Chassis. Front panel faces left. Locatfon
of components and printed circuits is shown.
circuit designations.
Figure 19 shows the opposite side.
38
Refer to the Parts List for
1064R
XODELS 610B, 610BR, ELECTROMETERS
HAINTEXANCC
si. ‘& ; ~
1.
.
FIGURE 21. Component Locations on Printed Circuit 85.
,’ ,, ,;.,o
i &:,~3
FIGURE 22 (left).
Printed Circuit 84.
Component Locations on
The red dots on VlOl
and V102 are toward each other.
MODELS 6108, 610BR ELECTROMETERS
RIO5
MAINTENANCE
FIGURE 25. Component Locations on Range Switch, SlOl.
high-megohm resistors.
Capacitors Cl03 and C104, not shown, are parallel to
capacitors Cl05 and C106.
RI12
Cl02
Figure 26 shows the
\
\
1064R
FIGURE 26.
Component Locations on Range Switch, 5101.
shows the other components.
Figure 25
MAINTENANCE
MODELS 610B, 610BR ELECTROMETERS
FIGURE 27. Component Locations
of MEDIUM and FINE ZERO Controls.
Also refer to Figure 28.
Rl’jl
FIGURE 28. Component Locations
of MEDIUM and FINE ZERO Controls.
Also refer to Figure 27.
R147
RI46
42
Rlj7
Rl58
Rl59
FIGURE 30.
of lbltiplier Switch, SlO6.
Also refer to Figure 29.
Component Locations
1064R
MODEL 6108, 610~R ELECTRO~TERS
I
SECTION 5. ACCESSORIES
hCCESSORIES
5-1.
more convenient.
with 30 inches of low-noise cable terminated by a uhf-type plug. The plug connects direce-
Ly into the Node1 610B INPUT Receptacle.
Its operation with the Model 610B is explained in paragraph 2-5.
5-2.
meaSulementS. The gripping feature is useful for attaching the probe at a single point :~r
repeated measurements.
plug.
5-3.
can be used with the Model 6108 up to 1000 volts with an accuracy of 3%.
nished with 30 inches of low-noise cable,
Model 6102A with the Model 6108 are given in paragraph 2-5.
MODEL 6101~ SHIELDED TEST LEAD.
a.
The Model 6101A is a prbbe and shielded Lead to make measurements with the Xodel 6i’?B
It does not alter any Node1 610B specification. The Lead is furnished
b. Using the Model 6101A Lead insures better input connections (see paragraph J-3)
MODEL 61018 GRIPPING PROBE.
It is furnished with a 30-inch cable that is terminated with a uiir
The Model 61016 connects directly into the Model 6108 INPUT Receptacle.
MODEL 6102~ VOLTAGE DIVIDER PROBE.
a.
The Model 6102A is a LO:1 voltage divider with an input resistance of 1010 ohms. it
b. The plug connects directly into the INPUT Receptacle.
The Model 6lOlB is a gripping probe to permit fast, easy
‘The Probe is fur-
terminated by a uhf-type plug.
Instructions for using the
FIGURE 3L.
5-4.
It can be used with the Node1 6108 up to 30,000 volts.
decreasing to :Lo% at 30 kilovolts,
cable,
using the Xodel 6103A with the 4lodel 6108 are given in paragraph 2-5.
HODEL 6103A VOLTAGE DIVIDER PROBE.
The ~lodel 6103A is a 1OOO:l voltage divider with an input resistance sf 10Lz ohms.
a.
The plug connects directly into the Node1 6108 INPUT Receptacle.
b.
Keithley Model 6101A and 6102~.
terminated by 3 uhf-type plug.
Both Xodels have identical external appearance
Accuracy is =5X to 15 kilovolt;.
The Probe is furnished with 30 inches of low-noise
Instructions for
ACCESSORIES
MODELS 610B, 610BR ELECTROMETERS
Model 6103A 1OOO:l Voltage Divider Probe.
5-5.
FIGLIRB 32.
MODEL 6104 TEST SHIELD.
a. The Model 6104 is useful for making guarded measurements or to avoid pick-up
problems.
or three-terminal connections.
For resistance measurements, it can be used with either two-teminal
The Model 6104 is also suitable for voltage and
current tests.
2-Terminal Operation.
b.
Connect the high-impedance side of the test sample to the Model 6104
1.
INPUT terminal. Connect the low-impedance side to either black jack marked
GROUND, whichever is more convenient.
Lock the enclosute and connect the mdel 6104 bnc-type receptacle to the
2.
Model 6LOB INPUT Receptacle.
Use a short coaxial cable lead.
Operate the
Electrometer as explained in Section 2.
c. 3-Terminal Operation.
For guarded measurements,
1.
make the connections as above with some
additional steps.
Connect the Ebdel 6104 black external plug to the Model 6lOB GUARD terminal.
2.
Connect either terminal marked EXTBBNAL SOURCE to the test sample at
3.
its high-impedance side. Lock the enclosure and connect the Model 6104 to
the Electrometer.
Connect the external power source to the l@del 6104 banana plugs, positive
4.
to red and negative to black.
5-6.
MODEL 6105 RESISTIVITY ADAPTER.
Operate the Electrometer as explained in Section 2.
a. The Model 6105 is a guarded test fixture for measuring resistivities of
materials in conjunction with the Node1 6LOB and the Model 240A Regulated High
;Jl:"~l;~gPlY.
The complete system directly measures volume resistivities up
ohm-cm and surface resfstivities up to 5 x 1018 ohms, in accordance
with the procedures of the American Society for Testing and Materials. The
shielded Model 6105 is a means for maintaining good sample contact with uniform
44
0268R ~,
MODELS 610B, 610BR ELECTROMETERS
ACCESSORIES
FIGURE 33. Model 6104 Test Shield.
FIGUU 34. Elodel 6105 Resistivity Adapter
pressure. The Adapter holds samples up to 3.5 inches in diameter and .25
inch thick. Maximum excitation voltage is 1000 volts.
The model 6105 Instruction Manual fully explains the operating proce-
b.
dures for the system, using the Models 24.0A and 6108.
c. The Model 610B alone can also be used with the Model 6105. Operate the
Model 6lOB as a constant current source (paragraph 2-11) with the FEEDBACK
Switch in FAST position. Connect the GUARD Terminal on the rear panel to
the center terminal of the Model 6105 Power Receptacle.
procedures are as for the 3-unit system.
The current from the Model 6lOB
Other operating
is the reciprocal of the OHMS setting of the Range Switch; the voltage is
the Multiplier Switch setting times the meter reading.
j-7.
NODELS 2501 AND 2503 STATIC DETECTOR HEADS.
a.
The Model 2501 is a lO,OOO:l capacitive voltage divider when used
with the head 3/8 inch from the charged surface. Measurements are accurate
within 10%. The Model 2501 consists of a 3-inch diameter head and 10 feet of
cable.
It connecta directly into the Model 610B.
The Model 2503 is a similar probe. The difference between them is the
b.
Hodel 2501 has a 3-inch diameter head and the Model 2503, l/2-inch diameter
head.
0268R
ACCESSORIES
MODELS 610B, 6LOBR ELECTROMETER
The Models 250/2501 Instruction Manual fully explains
c.
the system.
I
4
5-g. MODEL 6106 ELECTROMETER CONNECTION KIT.
a. The Model 6106 Kit contains a group of the most useful Leads and adapters.for use in
electrometer measurements with the Model 6LOB.
wide x 8 inches deep with polyethylene foam compartments.
pounds.
FIGURE 35.
Model 2501 Static Detector Head.
The kit case is 2 inches high x 12 inches
the operating procedures for
It weighs approximately three
The Kit contains two Leads made with 30-inch low-noise coaxial shielded cable (RG
b.
58A/W.
clips for connection to the circuit under test,
Table 2 for model numbers.)
c.
ed together, a uhf "tee" connector, and a bindlng post adapter.
Fig. 36
t
One lead has a uhf plug for connection to the electrometer, and two alligator
The second lead has two uhf plugs. (See
Adapters are also supplied for uhf-to-bnc conversion.
Other accessories include two uhf female couplers which permit cables to be connect-
Item
Description
1 Test Lead, two uhf plugs
2
3 Two Connectors, female uhf to female uhf
4
5
6
7
Binding Post Adaptor, uhf pLug HodeL 6106-2
Two Connector Adaptors, female uhf to male bnc
connector, male uhf to female bnc
Connector Adapter Tee,
Test Lead, uhf plugs and two alligator clips Model 6106-L
TABLE 6. Contents of Hodel 6106 Electrometer Conneccian Kit,
two male uhf and one female uhf
Keithley
Part No.
. 18265C
cs-5
CS-172
CS-115
cs-171
46
3268~
MODELS 6108, 610BR ELECTROMETERS
j-9.
MODEL 6107 pH ELECTRODE ADAPTER.
a. The Keithley Model 6107 pH Electrode
Adapter is designed to allow accurate and
convenient pH potential measurements with
the Keithley Model 6108 Electrometer and
other Keithley electrometers having a uhf
input receptacle. The Adapter accepts a
Beckman and Coleman (B-C) or a Leeds and
Northrup (L&N) connector.
The two terminals on the Model 6107 marked RRF accept a
banana jack or a phone tip jack.
b. The Adapter uses Teflon insulation
on all high impedance connections.
Guard-
ed measurements can be made with the i%deL
6107 by connecting the Xl OR GND Terminal
on the Model 6L07 to the X1 OUTPUT Terminal on the rear panel of the Model 6108.
The ZERO CHECK Button on the Adapter allows
checking zero on the Model 610B without
shorting the pH electrodes.
LGlJU 3b.
Connection Kit.
Keitniey ?lodeL bLUb Electrometer
Refer to Table 6 for com-
ponents.
c. To use the Model 6107 with the Model
6LOB, set both the ZERO CHECK Button on
the NodeL 6LOB and the ZERO CHECK Button on the Model 6107 to the LOCK Position.
FEEDBACK Switch on the Model 6108 can be in either the FAST or NORMAL Position.
The
d. Connect the Adapter cable directly into the Model 6lOB INPUT Receptacle.
Connect
the Xl OR GND Terminal on the Model 6107 to the Ground Post on the front panel of the
Model 6LOB if guarded measurements are not desired.
For guarded measurements connect the
XL OR GNU Terminal to the XL OUTPUT Terminal on the rear panel of the Model 6108.
a.
Insert the pH electrodes into the Model 6107 and open the ZERO CHECK Button on the
To check zero on the Model 610B, use only the ZERO CHECK Button on the Model
6107. Do not use the ZERO CHECK Button on the Model 610B.
The Button on the
Adapter isdesigned to disconnect the Hi electrode before shorting the Model
610B leads. If the Model 610B ZERO CHECK Button is used to check zero the
electrodes will be shorted and may be damaged.
f. To obtain a reading, set the Model 6108 METER Switch to CENTER ZERO.
Switch to VOLTS.
CHECK
Button on the Model 6107. Read the voltage off the lower meter scale. To convert
Use the appropriate range for the Multiplier Switch. Open the ZERO
Set the Range
the voltage to pH, consult the Handbook of Chemistry and Physics.
NOTE
If the Adapter becomes contaminated, clean it with distilled water followed by
a thorough swabbing with pure methyl alcohol, CH30H. This will suffice for most
contaminants.
5-10.
XODEL 370 RECORDER.
a. The Model 370 Recorder is uniquely compatible with the Model 6108 as well as other
Keithley electrometers,
microvoltmeters and picoammeters. The Recorder is a high quality
economical instrument that maximizes the performance of the Model 6108, and many other
Kefthley instruments,
even in the most critical applications. The Model 370 can be used
with the Model 6108 to record voltage, resistance, currents and charge over the Node1
6108’~ entire range.
b. The Model 6108 has the output necessary to drive the Recorder directly (1 volt, 1
milliampere), thus eliminating the need for a pre-amplifier.
volts off ground.
The Recorder is specially shielded to avoid pickup of extraneous sig-
The Model 370 floats 1500
nals. The response time of the Model 370 Recorder is 0.5 second; linearity is ‘1% of full
scale.
able with front panel controls. The 6-inch chart has a rectilinear presentation.
Ten chart
speeds -
from 314 inch per hour to 12 inches per minute - are select-
The
Node1 370 Recorder has a self-priming inking system. Chart paper and Fnk refills are easy
to install.
c. The Model 370 is very easy to use with the Model 6108. All that is necessary is connecting the two units, setting the Model 6108 Output Switch to 1MA and adjusting an easily
accessible control for full-scale recorder deflection. The furnished Model 3701 Input
Cable mates wtth the output connector on the Model 610B.
MODELS 6108, 610BR ELECTROMETER
ACCESSORILY
FIGURE 38.
Especially Designed Eor use with Keithley Electrometers and Other Instruments.
370 can be directly connected to the Model 610B output with the Recorder’s accessory cable
Response time and other specifications of the Model 370 Recorder are compatible with those
of the Xodel 6108 Electrometer.
Maximum Recording Convenience is Obtained Using the Keithley Xodel J/U,
The XodeL
MODELS 610B, 610BR ELECTROMETERS
REPLACEABLE PARTS
SECTION 6.
6-l.
ponents of the Models 610B
the part description, a suggested manufacturer, the manufacturer's part number
and the Keithley Part Number.
the part. The name and address of the manufacturers listed in the "Mfg. Code"
column are contained in Table 7.
6-2.
Keithley Part Number, the circuit designation and a description of the part.
All structural parts and those parts coded for Keithley manufacture (80164)
must be ordered from Keithley Instruments, Inc.
in the Replaceable Parts List, completely describe the part, its function
and its location.
Service Department, Keithley Instruments, Inc.
amp
CerD Ceramic, Disc
Comp
REPLACEABLE PARTS LIST.
HOW TO ORDER PARTS.
For parts orders,
a.
b.
Order
parts through your nearest Keithley distributor or the Sales