We warrant each of our products to be free from defects in material
and workmanship. Our obligation under this warranty is to repair
or replace any instrument or part thereof which, within a year after
shipment, proves defective upon examination. We will pay local
domestic surface freight costs.
To exercise this warranty, write or call your local Keithley repre-
sentative, or contact Keithley headquarters in Cleveland, Ohio.
You will be given prompt assistance and shipping instructions,
REPAIRS AND
CALIBRATION
Keithley Instruments maintains a complete repair and calibration
service as well as a standards laboratory in Cleveland, Ohio. A
service facility is also located in Los Angeles for our west coast
customers.
A Keithley service facility at our Munich, Germany office is
available for our customers throughout Europe. Service in the
United Kingdom’can be handled at our office in Reading. Addition-
ally, Keithley representatives in most countries maintain service
and calibration facilities.
To insure prompt repair or recalibration service, please contact
your local field representative or Keithley headquarters directly
before returning the instrument. Estimates for repairs, normal
recalibrations and calibrations traceable to the National Bureau of
Standards are available upon request.
France: 44 Rue Anatole France. F-91121 Palaiseau (01) 928-00-48
FOR YOUR SAFETY
Before undertaking any maintenance procedure, whether it
be a specific troubleshooting or maintenance procedure
described herein or an exploratory procedure aimed at
determining whether there has been a malfunction, read the
applicable section of this manual and note carefully the
WARNING and CADTlON notices contained therein.
The equipment described in this manual contains voltages
hazardous to human life and safety and which is capable of
inflicting personal injury. The cautionary and warning
notes arc included in this manual to alert operator and
maintenance personnel to the electrical hazards and thus
prevent personal injury and damage to equipment.
If this instrument is to be powered from the AC Mains
through an autotransformer (such as a Vartac or equivalent)
ensure that the instrument common connector is connected to the ground (earth) connection of the power
mains.
Before operating the unit ensure that the protective conductor (green wire) is connected to the ground (earth)
protective conductor of the power outlet. Do not defeat
the protective feature of the third protective conductor in
the power cord by using a two conductor extension cord or
a three-prong/two-prong adapter.
Maintenance and calibration procedures contained in this
manual sometimes call for operation of the unit with power
appliedandprotective covers removed. Read the procedures
carefully and heed Warnings to avoid “live” circuit points
to ensure your personal safety.
Before operating this instrument,
1. Ensure that the instrument is configured to
operate on the voltage available at the power
source See Installation section.
Ensure that the proper fuse is in place in the
2.
instrument for the power source on which the
instrument is to be operated.
3. Ensure that all other devices connected to or in
proximity to this instrument are properly grounded
or connected to the protective third-wire earth
ground.
COPYRIGHT
0
Copyright
States of America.
not be reproduced in any form without written permission of the
publishers.
1976 by Keithley Instruments, Inc. Printed in the United
All rights reserved. This book or parts thereof may
PROPRIETARY NOTICE
This document and the technical data herein disclosed, are proprietary
to Keithley Instruments, Inc., and shall not, without express written
permission of Keithley Instruments, Inc., be used, in whole or in part
to solicit quotations from a competitive source or used for manufacture
by anyone other than Keithley Instruments, Inc. The information
herein has been developed at private expense, and may only be used for
operation and maintenance reference purposes or for purposes of
engineering evaluation and incorporation into technical specifications
and other documents which specify procurement of products from
Keithley Instruments, Inc.
TABLE OF CONTENTS
Section
1
1.1
1.7
1.9
1.11
1.13
1.15
1.17
1.19
I .26
I .32
2
2.1
2.4
2.6
2.8
2.12
2.13
2.16
2.18
2.19
2.21
2.23
2.24
2.28
2.42
2.44
2.46
2.48
2.50
2.54
2.56
2.59
2.60
2.62
2.64
2.66
2.68
2.70
2.72
2.75
2.77
GENERALDESCRIPTION
Introduction
Options .’
Model 42 Remote Programming
Rear Input Options(-I, -1s)
Rack-Mounting Flanges (403402)
High-Voltage Probe
Current Shunt Set (651)
Measurement Capability
Specifications
Operating Controls
Maximum Input Voltage
Positive True Logic Relationships
Range Codes(Printer Output)
Function Programming
Range Codes(Programmet)
Timeouts
Maximum Input Voltage
Required Equipment
DC Range Check(Low Ranges)
DC Range Check (High Ranges)
3.Wire Ratio Check
4.Wire Ratio Check
DC Input Resistance.
Model 33 AC Converter Range Check
Model 32 AC Converter Range Check
Ohms-Megohms Range Check.
Common Mode Rejection (In DC Volts Function)
Normal Mode Noise Rejection (In DC Volts Function)
Common Mode Rejection (In AC Volts Function)
Switching Board Range Decode
Range Switch Code
Autorange Logic
Relay Logic Coding
Annunciator Logic
“NO” Annunciator Logic
Program Logic Conversion
Required Calibration Equipment
Fixed Voltage Dividers
DC Source Accuracies
AC Source Accuracies
Power Supply Check
OperationalCheck
Trollblesllooting Guide
Troubleshooting Chart-Digitizer and 1OV Reference Amplifier
Troubleshooting Chart. Isolator and Attenuator Boards
Troubleshooting Chart
Table7.1
List of Suppliers
................
...............
....................
...............................
.............................
..........................
.............. 1 1 : : 1 1 1 1 1 1 1 1 1 :
..........................
............... : : : : : : : : : : : ,: : : : :
..................
................
.......... : : : : : : : : : : : : : : : :
.....................
........................
...............
...............
...........................
......................
................
..................
....................
...................
............................
............................
..........................
..........................
.......................
...........................
...........................
...........................
.........................
...........................
Optional Accessories
...............
..............
...................
: : : : : : : : : : : : : : :
............. : : :
: : : : : : : : : : : : : : :
................
: : : : : : : : :
: : : : : : : : : : : : :
: : : : : :
: : : :
: : : : : : : : :
.............
Page
1-1
1-4
2.3
2.4
2-a
2.15
2.17
2.17
2-18
2.18
3-1
3.2
3-3
3-4
3-s
3.6
3-7
3-8
3-9
3-10
3-11
3.12
4-1
4.16
4-16
4.16
4.17
4.17
4-2 1
5-a
S-l
s-3
s-4
S-II
s-1 1
s-12
s-13
s-14
s-14
7-1
7-l
980453
ix
SECTION 1
GENERAL DESCRIPTION
1.1 INTRODUCTION.
The Model 5900 Digital Multimeter is a five-decade
1.2
instrument with a sixth digit providing 60% overrange. The
basic instrument is equipped for dc and dc/dc ratio measurements on five ranges. With the addition of the optional
AC Converter, a-c and ac/dc ratio measurements on four
ranges are available. The Ohms Converter, also optional,
adds ohms measurements on eight ranges. Complete measurement capability of a fully equipped instrument is tabulated in table 1 .l.
Table 1 .I - Measurement Capability
r
DC & Dc/Dc
RATIO
Range (Basic 5900)
.lV I x I
1vI x I
I
1ov X
FUNCTION
OhIllS
(Model 52
Ohms
Converter)
I
AC & AC/DC
RATIO
(Model 32
or 33
AC Converter)
1
IOOV I x I
ooov I x I
100.
I
.I KS2 X
I x
until a single reading is commanded by an external command. The new measurement is then held until the next
external command. In Periodic mbde (RATE control CW),
measurements are made automatic+ly at the rate of approximately four per second.
1.5 The basic Model 5900 includes an analog output
voltage that is proportional to the parameter being measured (except ratio). The voltage, at 20 volts maximum, is
available at a rear panel connector.
Also included as standard equipment is a solid-state
1.6
isolated BCD output. TTL-compatible output levels of the
reading, function, range, etc., plus a print command are
provided. An additional line enables a new reading to be
commanded externally. An optional isolated remote pro.
gramming unit (Model 42) allows all operating commands to
be made extcmally.
1.7 OPTIONS.
1.8 All optional accessories having model numbers are
plug-in circuit boards that may be added at any time. A
calibrated accessory board can be installed without affecting the d-c calibration of the basic instrument. An
instrument shipped without PCB accessories will not be
equipped with a Function Switching PCB assembly. This
board must be added when accessory boards are installed.
Analog accessories are identified in table 1 .l.
1Kn
10Kn 1
100 K$-l 1
000 KS2 1
lOMa/
lOOMa 1
1.3 Range can be selected manually or automatically
(autorange).
ular measurement is selected automatically (full scale is defined as “100000” on any range). The instrument “up
ranges” at 16U% of full scale and “downranges” at 15% of
full scale. Polarity selection is also automatic and is displayed on the readout.
1.4
(RATE control on EXT), a measurement is held (displayed)
In AUTO range, the proper range for a partic-
Two operating modes are provided. In Hold mode
X
I x
I x
I x
I x
I x
Model 42 Remote Programming.
1.9
The Model 42 Remote Programming accessory al-
1.10
lows the selection of function, range, filter, read command,
etc., to be made externally. Auto range selection is also
provided and appropriate timeouts are generated internally
when ranging takes place. Remote Programming “overrides”
all manual control settings to prevent erratic selections.
Complete isolation of the programming unit is achieved by
the use of photoauplers and pulse transformers.
1.11 Rear Input Options (-1, -lB, -1s. -1SB).
1.12
5900 DMM. The option designated -1 or -IB consists of
connector 5204 on the back panel with input lines +
INPUT, + CURRENT, and GUARD wired in parallel with
the front panel input terminals; the option designated -1s
or -1SB is the same as the -1 or -lB except that the front or
rear inputs are selectable by a switch on the front panel.
Two rear input options are available for the Model
b
I-I
980453
T T
1.13 Rack-Mounting Flanges (403402).
1.14 Rack.Mounting Flatages are used where 1111’ IIIS~~U-
ment is to be installed in a relay-rack or cabinet.
1.15 High-Voltage Probe (641).
1.16 The High.Voltage Probe extends the voltage range of
the instrument up to 10,000 volts (or 7SOOV rms). It is an
insulated prube containing a 1000: 1 voltage divider.
l-2
1.17 Current Shunt Set (651).
I .I8 The Current Shunt Set consists of six precision
sIllill assemblies witb values selected to produce a voltage
drop that, n~e:tsured in millivolts. Ins a numerical value
cqual to 111~ current Ilow in milliamps or micro;tmps.
1.19 ELECTRICAL DESCRIPTION.
1.70
1be Model 5900 DMM is a dual slope integration
instrument consisting of three main functional areas: signal
9804.53
conditioning, integrating, and control/display. A block diagram of the instrument is shown in figure I, 1.
1.21 The signal conditioning section includes the
Switching p-c board, AC Converter, Ohms Converter,
Attenuator, and Isolator. The function of these circuits is
to convert the incoming signal to 10 MC, full scale into the
integrator.
1.22 The Integrating section consists of the Integrator
amplifier, Null Detector and + reference voltages. The
function of these circuits is to convert the conditioned in.
put signal to an equivalent time period and to transmit this
iime period to the display portion of the DMM.
1.23 Dual slope integration operates as follows in a
sequence of program (PCM) states:
a. Signal Integration (KM-A). The integrator capaci-
tor charges to a voltage proportional to the input
voltage during a 20 msec sampling period.
b. Reference Integrate (KM-C). During this period,
the integrator capacitor discharges at a constant current. The time that the integrator requires to discharge (full discharge detected by the Null Detector)
is measured by the counter. The data in the counter
at the end of PCM.C is proportional to the input
voltage.
1.28 The Function Switching board is used only when
either or both of the options (AC and Ohms) are installed.
With no options installed, the Function Switching board is
replaced with the Switching Bypass board. The Switching
Bypass merely connects the + Input (from input connector)
directly to the Isolator input and the - Input to ground.
1.29 At the rear edge of the Logic and Interconnection
assembly is a PCB connector that extends to the rear panel
and serves as the BCD output connector 1201. If the
optional Remote Program board is installed, it is mounted
on stand-offs above the Logic and Interconnection board
with the PROGRAM INPUT connector (J202) available at
the rear panel above the BCD OUTPUT connector.
1.30 The POWER input connector 1203, the power transformer, and power transistors for the power supply are
mounted on the rear panel of the instrument. Other power
supply components are mounted on the Logic and Inter-
connection assembly. Also mounted on the rear panel, in
addition to .I201 and 1202, is the rear INPUT connector
J204, the ANALOG OUTPUT connector and common, the
EXTernal REFerence connector and common, and the line
fuse F201,
1.31 A dimensional outline of the Model 5900 is shown
in tigure 1.2.
c. Strobe @CM-D). At this time, data in the counter
is strobed into the storage latches and displayed -
the print pulse is inhibited, however if an uprange
or downrange command is generated by the Autorange logic. If a range change is required, the
counter is reset and the program returns to PGM-A.
d. Reset @‘CM-E). At PCM-E, all internal logic is reset
in preparation for the next reading.
1.24 An additional control state, PGM.B, occurs after
PGM-A and is a delay to allow for propagation time of the
counter.
1.25 The Control/Display section generates the control
signals necessary to operate the signal conditioning and
integrating circuits.
1.26 MECHANICAL DESCRIPTION.
1.21 The ohms measurement option consists of a single
printed-circuit board. The AC options both consist of two
boards. The accessory boards plus the Digitizer, Isolator,
and Function Switching board all plug into the Main Logic
board called the Logic and Interconnection assembly. This
board also carries much of the instrument logic.
Figure 1.2 Dimensions
1.32 SPECIFICATIONS.
1.33 Specifications are listed in table 1.2.
1-3
980453
Table 1.2 - Specifications
SPECIFICATIONS, MODEL SSOO
AS A DC VOLTMETER IBASIC INSTRUMENTl
RATIO tdcldc. mVldc. acldc)
SPECIFICATIONS CONTINUED NEXT PAOE
Table 1.2 -Specifications (continued)
980453
l-5
SECTION 2
2.1 UNPACKING AND INSPECTION.
INSTALLATION & OPERATION
2.2
plastic-foam form within a cardboard carton for shipment.
The plastic form holds the DMM securely in the carton and
absorbs any reasonable external shock normally encountered
in transit. Prior to unpacking, examine the exterior of the
shipping carton for any signs of damage. Carefully remove
the DMM from the carton and inspect the exterior of the
instrument for any signs of damage. If damage is found,
notify the carrier immediately.
2.3 Included with the instrument in the packing container are the instruction manual, power cord, and rear input and BCD output mating connectors. With instruments
equipped with remote programming, a mating connector
for that accessory is included.
2.4 BENCH OPERATION.
2.5 Each Model 5900 is equipped with a tilt bail or
“kickstand” to enable the front of the instrument to be
elevated for convenient bench use. The tilt bail is attached
to the two front supporting “feet” at the bottom of the
instrument. For use, the bail is pulled down to its supporting position.
The Model 5900 DMM is packed in a molded
2.6 RACK MOUNTING.
2.1 The instrument can be mounted in a standard 19.
inch rack with the optional rack-mounting flanges (403402,
includes attaching hardware). To install the flanges, proceed as follows:
a. With instrument on its side, remove four Phillips-
head screws holding bottom cover. Remove cover.
Remove screws holding feet (and bail) in place. Replace bottom cover.
b. Place one of the supplied screws through each of
the two holes in the mounting flange (figure 2.1).
Thread a securing nut onto each screw just enough
to attach it to the screw (approximately one turn).
c. Place the mounting flange onto the mounting slot
in the instrument side panel so that the securing
nuts fit entirely into the slot. Be sure the rackmount slots on the flange are toward the front of
the instrument.
d. Tighten screws. The securing nuts will rotate and
hold the flange securely in place.
2.6 POWER CONNECTIONS.
2.9 Standard units operate on either 1 IS volts or 230
volts, SO to 60 Hz (400 Hz available). Power consumption
is less than 40 watts. Operation on either of the two line
voltages is selectable by a slide switch on the rear panel.
Operation on lOO/ZOO volts or 1201240 volts is possible by
simple rewiring of the power transformer secondary wires:
WARNING
Disconnect the instrument from the AC Power source
before attempting to change power connections.
Potentially lethal voltages are exposed when covers
are removed.
a. For operation on 100/200 volts, cut the brown wire
1” from the transformer and splice it to the red
wire on the transformer; cut the blue wire 1” from
the transformer and splice it to the violet wire.
b. For operation on 120/240 volts, cut the brown
wire 1” from the transformer and splice it to the
black wire; cut the blue wire and splice it to the
yellow wire on the transformer.
2-I
980453
2.10 A standard power cable having a three-pin plug is
supplied with the instrument. It connects to POWER con.
nectar 5203. The ground pin (round) is attached to the
instrument case. It is important that this pin be connected
to a good quality earth ground.
2.1 I
Fuse receptacle F20l on the rear panel is equipped
with a .5 amp fuse in domestic units.
2.12 INPUT/OUTPUT CABLING.
2.13 Binding Posts.
2.14 Several connectors on the Model 5900 consist of a
pair of binding posts spaced so as to accept standard
“banana” plugs. The connectors are:
I
Front Panel
I
Rear Panel
+ INPUT f. ANALOG OUTPUT
+ OHMS CURRENT
?1 REFerence INput
2.15 Input cables to fit this type of connector can be
ordered from Keithley (P/N 5900-402190). Figure 2.2 is a
wiring diagram of this cable included for assistance to users
desiring to construct their own cables.
panel binding posts. The rear-panel input lines are wired in
parallel with the front-panel input lines. It is recommended
that the cable for the mating connector be constructed as
shown in figure 2.3 using two two-conductor shielded
cables. Other configurations may be desirable depending
on the ohms measuring method to be used (see paragraph 2.28).
I
2.18 MANUAL OPERATION.
2.19 Controls.
2.16 Rear Input Connector.
2.17 Instruments equipped with the -I or -IS rear input
option are supplied with 5204 7.pin input connector
(Keithley P/N 5900-600673) and a mating connector
(Keithley P/N 5900-600616). The instrument acceptsinputs
applied to this connector or inputs applied to the front-
2.20 All operating controls are located on the front panel
of the instrument. They are shown in figure 2.5 and their
operation described in table 2.1. Description of Systems
operation begins in paragraph 2.44.
2.21 DISPLAY.
2.22 The Pisplay consists of 6 LED decimal readout
devices with moving decimal point. The decimal point
moves in conjunction with the range switchor automatically
in auto range. Maximum usable readout with overrange is
159999. Overload is indicated by a NO and 160000 readout. A non-compatible range and function is indicated by
a NO. However, mechanical interlocks are provided to prevent illegal combinations from the front panel. Figure 2.4
illustrates the readout, NO indicator, polarity sign and the
DC MO
+
r /--- r-7 ,---I ,---I ,--1 AC
- I L-, L-4 L-4 L-,’ L-4 PGM
NO,’ --J. ,’
Fiwe 2.4 . Readout
: : :
KO
n
FIL R
2-2
Figure 2.5 -Front Panel
Table 2.1 . Operating Controls
Control
POW1
(rocker switch)
Function Select
(rotary switch)
Range Select
(rotary switch)
DATA OUTPUT
(pushbutton)
Position
ON (UP)
OFF (down)
ACT
DC
AUTO
Other Positions
Depressed
Function
Applies power to instrument
Removes power from instrument
Selects the measurement of AC voltages on the I, IO, 100, and
1000 volt ranges (max. input, 1OOOV RMS)
Selects the measurement of DC voltages on the .l, 1, IO, 100,
and 1000 volt ranges (max. input, IlOOV)
Selects the measurement of resistance on the 10 ohm range; on
the.1,1,10,100,andlOOOkilohmsranges;oronthe10or100
Megohm ranges
Selects Auto Range in which the optimum range is selected automatically by internal circuits. Uprange occurs at 160% of full
scale; downrange occurs at 15% of full scale
Enables manual selection of fiied ranges. Ranges permissible for
each function are inscribed on the parlel
,Enables the print pulse causing BCD data at P201 to be recorded
by printer, or other output device. (Output data is present at
P201 regardless of the position of this switch.)
PROGRAM CONTROL
(pushbutton)
Depressed
Enables the selection of range, function, and mode to be made
externally through the remote programming connector and dis.
ables all front panel controls (requires programming option)
RATIO
(pushbutton)
Depressed
Selects a ratio measurement in which the readout represents the
ratio of the input to an external d-c reference voltage (applied at
terminals on the rear panel) multiplied by 10: Ein/ERafx 10
FILTER
Depressed
Adds an active four.pole filter across the input circuit
(pushbutton)
RATE EXT (ccw)
(pot)
cw
FRONT/REAR FRONT
Selects the Hold mode. A new reading is initiated through the
remote program input
Increase periodic read rate to a maximum of four readings/second
Connects front panel data input terminals to instrument
(slide switch)
REAR
Connects rear panel data input terminals to instrument
*NOTE: Ohms input terminals are open in AC or DC function.
TNOTE: For inputs greater than lSOV, Filter should be “IN”.
980453
2.23 MEASUREMENT CONNECTIONS.
NOTE
Before taking any measurements, refer, to the list of
maximum input voltages, table 2.2.
Table 2.2 - hlaximum Input Voltage
CAUTION
Do not exceed the following maximum inputs:
DC
AC
1lOOVDCor 1OOOVRMSAC
1OOOV RMS decreasihg to 20V RMS at
1 MHz (see figure 2.7)
n
+SOOV peak between +I and -I(lOOOV
RMS if in DC or AC function.)
RATIO
Input: same as function selected
Reference: +lO.SV, -0.5V
achieved by ‘placing a shorting bar between - INPUT and
GUARD and shorting the single banana plug (shield) to the
-
INPUT side of the double banana plug at the input connector. This arrangement is adequate for measuring all but
low voltage (mV) levels and/or in high-noise environments.
2.27 When making “floating”voltage measurements(both
measurement points above ground potential), do not connect GUARD to measurement ground without making sure
that the voltage between GUARD and - INPUT does not
exceed 250 volts.
Y
‘RMS
GUARD
Voltage between GUARD and - INPUl
must not exceed 250 volts or damage
to the instrument may result
2.24 Basic Voltage Measurement.
2.25 An ac or dc voltage measurement connection recommended to minimize the effects of noise requires a twoconductor shielded cable connected as shown in figure 2.6.
t
Figure 2.6. Basic Voltage Measurement Connections
2.26 For all voltage measurements, the GUARD lead ar
Id
the - INPUT lead are connected to the measurement point
nearest ground potential.
Somewhat less shielding is
2.28 Ohms Measurement.
2.29
Ohms measurement in the Model 5900 consists of
the application of a known current through the unknown
resistance (Rx) and measuring the ratio of the voltage drop
across Rx to the drop across an internal “full-scale” resistor
(Eh/EFS). Current through Rx is applied through leads
from the f. OHMS CURRENT terminals. The voltage drop
is sensed by the + INPUT terminals.
2.30 TWO.WIRE MEASUREMENTS.
2.31
Connections for a simple two-wire shielded ohms
measurement are shown in figure 2.8a. It consists simply of
a single-conductor shielded cable with the conductor serving
as both the t CURRENT and + INPUT leads and the shield
2.4
(4
(b)
Y&v453
lNP”T CVRRENT
Figure 2.8 - Two Wire Ohms Measurements
RX
2-s
980453
2-6
Figure 2.9 - Four Wire Ohms Measurement
980453
carrying - CURRENT and - INPUT. While reasonably
accurate measurements can be made with this method,
shunt leakage problems result from the parallel combinations
of Rx and the cable impedance. This causes loss of accuracy,
especially at high resistance (100 Ma range). Also, lead
resistance becomes a factor in the 10 and 100 ohms ranges;
the four wire measurement system is recommended for
these ranges.
2.32 A more accurate two-wire measurement connection
is shown in figure 2.6b. The + INPUT and + CURRENT,
-
INPUT and - CURRENT terminals are again tied to-
gether. But now, the positive side is a single-conductor,
shielded cable with the shield tied to Ohms Guard. Ohms
Guard is the low ANALOG OUTPUT terminal on the rear
panel of the Model 5900 when ohms is selected. The negative
side is a single wire connected as shown. Guard current is
present in the low side, but the leakage problems of the
first configuration are eliminated.
2.33 In high noise-level environments, the configuration
shown in figure 2.8~ is recommended. This method also
eliminates error due to shunt leakage, but provides more
complete shielding. The positive terminals are tied together
and carried in a single-conductor, double-shielded cable with
the inner shield tied to OhmsGuard (-ANALOG OUTPUT).
The outer shield is tied to GUARD. The negative terminals
are tied together and carried in a single-conductor shielded
cable with the shield tied to GUARD. This configuration
eliminates guard current sensitivity, thereby increasing
guarding characteristics.
2.36 This configuration, although shielded, places the
shield capacitance and cable leakage in parallel with Rx.
This results in loss of accuracy and slow measurements. In
addition, it is very responsive to the triboelectric effect at
high resistance measurements.
2.39 Better guarding is achieved by the use of the configuration shown in figure 2.9b. Here again, RG196U teflon
dielectric cable (either single-conductor shielded or twoconductor shielded) is used on the positive terminals. The
shield(s) are connected to Ohms Guard (low ANALOG
OUTPUT terminal). The negative leads are single wires with
the - INPUT terminal tied to GUARD.
2.40 This eliminates much of the shunt leakage problem
of the previous configuration since guard current now flows
through the low side of the measurement circuit. Measurement is much faster since the shield capacity is driven by
the guard current.
2.41 A high-noise environment calls for the “super”
configuration shown in figure 2.9~. Here, a two-conductor,
double-shielded cable is used as the positive leads. The inner
shield is tied to Ohms Guard. A twoanductor shielded
cable is used as the negative leads. Its shield is tied to
GUARD and to the outer shield of the positive cable. The
shield is also tied to CURRENT at the measurement point.
This configuration maintains high guarding characteristics
while eliminating guard current sensitivity.
2.42 Ratio Measurements.
2.34 FOUR-WIRE MEASUREMENTS
2.35 In most system applications, the device to be measured is located at B remote location requiring interconnection
by cables of lengths from several to possibly hundreds of
feet. When measuring low resistance values over long cables,
most lead resistance problems can be solved by the use of a
four-wire measurement system.
2.36 For high resistance measurements over long cables,
other problems are encountered: noise pick-up, leakage
resistance, and capacitive loading of the system. These
problems can be minimized by proper shielding and the use
of ohms guard.
2.37 Figure 2.9a shows a basic shielded four-wire ohms
measurement configuration. This method uses two singleconductor shielded teflon cables. The conductors carry the
positive sides of the INPUT and CURRENT lines while each
shield carries the low side.
2.43 Ratio measurements are made by applying a positive
d-c voltage to the reference input terminals on the rear panel
and an input signal of any function at the front input
terminals. For DC/DC or AC/DC ratios, the reference
voltage must be within the range of +IV to +lO.SV. Input
signal limitations (numerator) are the same as tbosc given
for conventional measurement of the particular function
(table 2.2). The readout is the ratio multiplied by ten:
Einput/Ereference x 10. In the standard instrument the
INPUT terminal is internally connected to the REF input terminal; in instruments equipped with the option 62
4-wire ratio, both reference inputs arc floating (IO Ma
between - REF and -SIGNAL).
2.44 SYSTEM CAPABILITIES.
2.45 The 5900 has two system interface connectorsdesignated as 3201 (PRINTER OUTPUT) and 5202 (PROGRAM
INPUT) mounted on the rear panel of the instrument (fig
ure 2.10). The following is a brief description of the capabilities of each connector.
2-7
980453
Printer Output - 5201.
2.46
2.47 Through this connector the 5900 supplies BCD
representations of the decimal display; various flags or
indicators of the mode of operation, function and range;
and a print command. Provision has also been made for 60
Hz instruments to accept a fast (20 readings per second
maximum) or a superfast (I 01 readings per second minimum)
read command. In 50 Hz units, the fast command obtains
I7 readings per second, minimum, and the superfast com-
mand 93 readings per second.
2.48 Program Input.
2.49
Through this connector the 5900 receives externally
generated signals that select the function, range, mode of
operation, and initiate the read commands.
2.50 LOGIC LEVELS AND ELECTRONIC IN-
TERFACE.
TTL-compatible positive-true logic levels are used
2.5 1
in the 5900. In some instances, however, complementary
signals are used. These terms are more specifically defined
below:
other words, the complement of X is x. The truth table
shows that the two signals X and x, are by definition, in
opposite logic states (see table 2.3).
Table 2.3 - Positive True Logic Relationships
Voltage Level of
Voltage Level of
Signal Logic State Output Line “x” Output Line “p
“X” True or “1” 2.4 - 5.0
False or “0” 1 0.0-0.4
2.53
As seen above, if gate A has a true or “1” level on
Volts 0.0.0.4
Volts 1
2.4.5.0
output X, its voltage level is the most positive of the two
ranges present, and output x must be in a false or “0”
state with the lowest or most negative voltage range present.
The reverse would be true for a false or “0” level on output
X.
2.54 Driving the Inputs.
2.55 All inputs are TTL compatible and most are the
equivalent of one 7400 series TTL input with a pull-up
resistor for contact closure operation.
Signals and Their Complements -
2.52
If the non-inverting output of gate A is defined as
signal X, then it follows that the inverting output is ???; in
Figure 2. IO - Rear Panel
0
2.56 TTL Loading Conditions,
2.57 To input a “I” level the pull.up resistor will supply
the necessary source current (40 /.tA) to maintain the minimum 2.4 volts. In fact, the pull-up resistor will maintain a
2-8
980453
Figure 2.11 Measurement Sequence
2-9
980453
one level as long as the input source resistance (RI) to
ground is greater than ISK ohms.
2.58 To input a “0” level, at least 2.0 ma of current must
be sinked maintaining the input voltage below 0.4 volts.
This requires a resistance to ground of 200 ohms or less.
Exceptions to Input Loading Conditions.
2.59
a.
Program Storage Input (5202, pin B.15) is the
equivalent of 3 TTL inputs and requires a minimum
5.8 ma sinking current, or 68 ohms or less to common.
b.
Maximum input voltage level, referenced to common,
must not exceed 5.5 volts peak. Otherwise, gate
destruction will occur.
c.
The Direct and Time Out Commands are AC coupled
with pullap resistors to t5 volts. These inputs are
compatible with TTL outputs or contact closures
to ground. The AC coupling does require that rise
and fall times be less than 100 peconds. This input
circuit is illustrated below:
ts” +5V
P 9
2.62 Timing Sequence.
2.63 The standard remote mode of operation of the 5900
is to initiate a reading sequence with each Direct Command
received through the programmer, providing
time has been allowed between commands for the reading
to be completed. This reading sequence is illustrated in
figure 2. I I
TI .To
T2-TI
T3-T2
T4.T3
During this period the input signal must finish settling to within the desired accuracy.
Any control changes involving the 25 msec
relay settling time (a) can he completed;
other logic control inputs (b) can also be
changed.
The Direct Comma signal, which is AC
coupled, must meet the following conditions:
a. Rise and fall times less than 100 /.wc.
b.
Signal must stay in the logical “0”
state for at least 3.4 /&ec. If these
conditions are met, the internal read
command is sustained at T2 and the
signal integrate period is started.
The period of signal integration lasts for
16.2/3 mw (60 Hz line frequency; 20 mw
in 50 Hz units). During this time the integrator charges to a voltage proportional to
the input voltage. This is the input sampling
period.
During this period, the integrator is isolated
from the input signal, and is discharged at a.
precise current. The time the integrator requires to discharge to a level equal to its
voltage at T2 is proportional to the input
voltage. This time is measured by an internal
counter and stored.
that
sufficient
d. Digital output common can be floated as high as
200 VDC above power line ground.
2.60. TTL Output Capabilities.
2.61 The 5900 electrical outputs are specified to drive
two ‘ITL inputs such as described in the TTL loading section.
Summary:
False: 0 to tO.4V
True: t2.4 to +S.OV
Fanout: 2minimum ,
Maximum Capacitance Load: 500 pF
2-10
T5.T4
To-TS
TI -To
This I.7 /J.W period is required to strobe the
new reading from the internal counter into
the readout latches.
This 5 mseconds (&IO%) is required to reset
the internal logic for the next reading.
If the next read command is a Direct
Command, this period must be made long
enough to allow for the condition covered in
the first cycle; however, if the next command
is a Timeout Command, this period can
approach zero since the necessary timeout to
satisfy these conditions are automatically
programmed.
980453
Figure 2.12. Command Timing
2.64 Other Read Command Options.
2.65 In addition to the Direct Command, there are two
other programmable read commands, as illustrated in figure 2.12.
a.
Time Out Command: Again, this is an AC coupled
input which must have rise and fall times of less
than 100 weconds but must remain in a “0” state
for at least 0.1 @second. The timeouts given in
table 2.7 for various combinatioris of ranges and
functions ranging from 30 mseconds to 500 msec.
onds will be automatically inserted before the internal read command is generated. If this command
is wired to the m on J2Ol - pin
All, fully automatic reading with timeouts is
achieved.
The 5 msec internal delay is not adequate for settling
time on the 100 VDC, 1000 Ka 10 MS1, IO0 MS1
ranges, or any AC range. Therefore, the timeout command, providing timeout delays listed in table 2.7,
must be used to initiate accurate readings on these
ranges unless a fixed range and function have been
programmed and the input has been present longer
than the timeout period.
System Direct Command: While the other two read
b.
commands were AC coupled and programmed
through the Program Input Connector (J202), this
command is DC coupled and programmed through
2-11
980453
Figure 2.13. Minimum Read Rate vs. Input
the Printer Output Connector (5201). The $&%
Direct Command must remain in the “0” state for
at least 3.4 /.&seconds to generate the internal read
command. This delay, as in the Direct Command, is and internal reset remain fixed while the reference integrate
to prevent noise from triggering the readings. If this period can vary from 0 to 32 mseconds. Therefore the
command is tied to ground, the 5900 will recycle at maximum read rate could very from 17.4 to 45 reading per
its maximum reading rate with no timeouts. second.
2.66 Reading Rates.
2.67 In figure 2.13, integrator operation with three different input signal levels is illustrated: half scale, full scale,
2-12
and 160% of full scale (full scale is defined as 100000 on any
range). The figure shows that the maximum reading rate is
a function of the input signal. The signal integrate period
2.66 Superfast.
2.69 The Superfast reading mode (programmed through
either the PRINTER OUT or PROGRAM INPUT connector)
980453
Figure 2.14
- Superfast
increases the minimum reading rate from 20.5 to 102
readings per second (50 Hz: 17.4 to 93 r/s). This is done at
the expense of losing the least-significant digit which is reset to zero (blanked out on readout). The signal integration
period is reduced from 16.2/3 msec to I -2/3 msec (SO Hz:
20
msec to 2 msec). This and the resulting reference integrate period reduce the maximum recycle time from 48.8
msec to 9.8 msec (SO Hz: 57.5 msec to 10.7 msec), thereby
yielding the 107 reading per second figure (93 r/s with 50
Hz). Timing changes are shown in figure 2.14.
2.70 PRINTER OUTPUT.
2.71 The printer output connector is a double-edged
PCB connector (extension of Interconnection and Logic
board) with pins Al through A22 on the bottom edge and
pins B I through B22 on the top edge. Pin assignments are
shown in figure 2.15. All outputs are referenced to digital
ground pin B I.
Read Rate (Worst Case)
2.72 Numerical Data
2.73 Numerical data appears as positive true, four-line
BCD code, as shown in figure 2.15. The designator of each
line identifies the digit and weight. For example: Pins Al8,
19, 20, and 21 are designated 12. Ig, 14. and I, consecutively. The 1 indicates these lines correspond to the units
or least significant display; the 2, 8, 4, and I subscripts
indicate the hinary weight of each line.
CAUTION
True output lines arc not short-circuit proof. Accidental grounding may damage the output circuitry.
2.13
Loading...
+ 134 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.