Tektronix 561b schematic

TECHNICAL MANUAL
OPERATOR’S, ORGANIZATIONAL, DIRECT SUPPORT, AND
GENERAL SUPPORT MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST)
TM 9-6625-963-14-1
GRAPHICAL DISPLAY SYSTEM,
TEKTRONIX TYPE 561 SERIES
This copy is a reprint which includes current pages from Change 1.
HEADQUARTERS, DEPARTMENT OF THE ARMY
OCTOBER 1972
NOTE
This manual is an authentication of the manufacture’s commercial literature which, through usage, has been found to cover the data required to operate and maintain this equipment. Since the manual was not prepared in accordance in with military specifications, the format has not been structured to considered level of maintenance nor to include a formal section on depot overhaul standards.
This manual contains copyrighted information. Reproduced by permission of Tektronix, Incorporated. All rights reserved.
CHANGE TM 9-6625-963-14-1
Change 2
No. 2 HEADQUARTERS
DEPARTMENT OF THE ARMY
WASHINGTON, DC, 12 August 1986
OPERATOR’S ORGANIZATIONAL, DIRECT SUPPORT AND
GENERAL SUPPORT MAINTENANCE MANUAL,
INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST:
GRAPHICAL DISPLAY SYSTEM
TEKTRONIX TYPE 561 SERIES (NSN 4931-0(1-910-8164)
Current as of 16 April 1986
TM 9-6625-963-14-1, 11 October 1972, is changed as follows:
1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page.
Remove pages Insert pages
0 -1 and 0 -2 0 -1 and 0 -2
12 - 1 through 12 - 11 12 - 1 through 12 - 16
2. File this change sheet in front of the publication for reference purposes.
By Order of the Secretary of the Army:
JOHN A. WICKHAM, JR.
General, United States Army
Chief of Staff
Official:
R. L. DILWORTH
Brigadier General, United States Army
The Adjutant General
Distribution:
To be distributed in accordance with DA Form 12-34C, Block No. 319, requirements for calibration procedures publications.
TM 9-6625-963-14-1
CHANGE 1
CHANGE HEADQUARTERS
DEPARTMENT OF THE ARMY
NO. 1 WASHINGTON, D.C. 20 MARCH 1975
TM 9-6625-963-14-1, 11 October 1972, is changed as follows:
1. The title is changed to read as shown above.
2. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page.
}
OPERATOR’S, ORGANIZATIONAL, DIRECT SUPPORT,
AND SUPPORT MAINTENANCE MANUAL,
INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST:
GRAPHICAL DISPLAY SYSTEM, TEKTRONIX
TYPE 561 SERIES (NSN 4931-00-910-8164)
Current as of 11 September 1974
Remove pages Insert pages
i and ii i and ii
12-1 through 12-11
3. File this change sheet in front of the publication for reference purposes.
TM 9-6625-963-14-1
TECHNICAL MANUAL HEADQUARTERS
DEPARTMENT OF THE ARMY
NO. 9-6625-963-14-1 Washington, D.C.
SECTION 0 INTRODUCTION................................................................................................................................................0-1
1A TYPE R561B SPECIFICATION......................................................................................................................1-1A
2A OPERATING INSTRUCTIONS R561B...........................................................................................................2-1A
3A CIRCUIT DESCRIPTION R561B....................................................................................................................3-1A
4A MAINTENANCE R561B..................................................................................................................................4-1A
5A PERFORMANCE CHECK VIRGULE MAINTENANCE CALIBRATION R561B...............................................5-1A
10 PREVENTIVE MAINTENANCE INSTRUCTIONS............................................................................................10-1
11 RACKMOUNTING............................................................................................................................................11-1
12 REPAIR PARTS LIST......................................................................................................................................12-1
Appendix A REFERENCES..................................................................................................................................................A-1
B BASIC ISSUE ITEMS LIST AND ITEMS TROOP INSTALLED OR AUTHORIZED LIST..................................B-1
C MAINTENANCE ALLOCATION CHART...........................................................................................................C-1
}
OPERATOR’S ORGANIZATIONAL, DIRECT SUPPORT, AND
GENERAL SUPPORT MAINTENANCE MANUAL
INCLUDING REPAIR PARTS SPECIAL TOOL LIST
GRAPHICAL DISPLAY SYSTEM,
TEKTRONIX TYPE 561 SERIES
(NSN 4931-00-910-8164)
1 CHARACTERISTICS RM561A...........................................................................................................................1-1
2 OPERATING INSTRUCTIONS RM561A............................................................................................................2-1
3 CIRCUIT DESCRIPTION RM561A....................................................................................................................3-1
4 MAINTENANCE RM561A..................................................................................................................................4-1
5 PERFORMANCE CHECK RM561A...................................................................................................................5-1
6 MAINTENANCE CALIBRATION RM561A..........................................................................................................6-1
PARTS LIST ABBREVIATIONS......................................................................................................................7-0.1
7 ELECTRICAL, PARTS LIST...............................................................................................................................7-1
8 MECHANICAL, PARTS LIST.............................................................................................................................8-1
9 DIAGRAMS........................................................................................................................................................9-1
11 October 1972
Page
Change 1 i
TM 9-6625-963-14-1
Figure 1-1A. Type RM561A Oscilloscope
ii

SECTION 0

INTRODUCTION
TM 9-6625-963-14-1
Scope
The manual includes installation and operation instructions and covers organizational, direct support (DS), and general support (GS) maintenance. It describes Graphical Display System, Tektronix Type 561 Series. The basic issue items list appears in appendix B. Appendix ’B is current as of 1 September 1972.
Indexes of Publications
DA Pam 310-4.
to determine if there are any new editions, changes, or additional publications pertaining to the equipment.
DA Pam 310-7.
whether there are Modification Work Orders (MWO) pertaining to the equipment.
Forms and Records
Refer to the latest issue of DA Pam 310-4
Refer to DA Pam 310-7 to determine
Reports of Maintenance and Unsatisfactory Equipment.
Use equipment forms and records in accordance with instructions given in TM 38750.
Change 2 0-1
Report of Packaging and Handling Deficiencies
6). Fill out and forward DD Form 6 as prescribed in AR 700-8 (Army), NAVSUP Pub 378 (Navy), AFR 71-41 (Air Force), and MCO P4030.29 (Marine Corps).
Discrepancy in Shipment Report.
Discrepancy in Shipment Report (SF 361) as prescribed in AR 55-38 (Army), NAV SUPINST 4610.33 (Navy), AFM 7-18 (Air Force), and MCO P4610.19A (Marine Corps).
Fill out and forward
Reporting of Equipment Publication Improve
reporting of errors, omissions, and recommendations for improving this publication by the individual user is encouraged. Reports should be submitted on DA Form 2028 (Recommended Changes to Publications) and forwarded direct to: Commander, U. S. Army TMDE Support Group, ATTN: AMXTM-LML, Redstone Arsenal, AL 35898-5400.
(DD Form
ments. The
Fig. 1-1. Type R561B Oscilloscope.
0-2

SECTION 1

CHARACTERISTICS
General Information
The Tektronix Type RM561A Oscilloscope is essentially an indicator unit with provision for two plug-in units. The plug­in unit in the right-hand opening controls the horizontal (X-axis) deflection, and the plug-in unit in the left-hand opening controls the vertical (Y-axis) deflection. The plug-in units con be selected from any of the Tektronix ’2’ Series or ’3’ Series groups to provide the desired oscilloscope performance.
Cathode-Ray Tube
Type - T5032-31-1 (rectangular ceramic-envelope).
Phosphor-P31.
Unblanking - Deflection type, dc-coupled, with grid intensification.
Graticule - Variable edge lighted "no parallax" internal graticule. Marked in 8 vertical and 10 horizontal divisions with each major division divided into 5 minor divisions on centerlines.
Accelerating voltage - 3.5 kv.
Useable viewing area - 8 divisions vertical by 10 divisions horizontal.
Deflection meters - 18.5 to 20.5 volts per centimeter vertical, and 17.5 to 19.3 volts per centimeter horizontal.
Calibrator
Waveform - Square waves at line frequency.
RM561A
Output voltage - 1 millivolt to 100 volts, peak-to-peak, in 6 steps. The 1 v position provides an output of 0.1 volt into 50 ohms.
Accuracy - Peak-to-peak amplitude within 3% of indicated voltage.
Rise-time - Typically 5 microseconds.
Power Supplies
Electronically regulated for stable operation with normal line voltage variations and widely varying loads. The low voltage supplies hold regulation within ± 1% of value.
Line voltage requirements - 105 to 125 volts, or 210 to 250 volts, rms, 50 to 60 cps, single-phase ac.
Fuse requirements - 3-amp slow-blowing type for 117 volts,
1.25-amp slow-blowing type for 234 volts.
Ventilation
Forced air cooled. Automatic-resetting thermal cutout interrupts instrument power if internal temperature exceeds about 123° F.
Construction
Aluminum-alloy chassis.
Photo-etched anodized front panel.
Weight - 30-1/2 pounds, indicator unit only.
Dimensions - see Dimension Drawing at rear of manual.
1-1

SECTION 1A

TYPE R561B SPECIFICATION
Introduction
The Type R561B Oscilloscope is compatible with Tektronix 2-Series and 3-Series plug-in units (see following table and Section 2 for exceptions); thus, it con be used in a variety of applications including differential, multi-trace, wide­band, delayed sweep, sampling and spectrum analysis.
This instrument will perform to the specifications listed in this section in a laboratory environment with ambient
temperature range between 0°C and +50°C, except as indicated. Warm-up time for rated accuracy is 5 minutes (certain plug-ins may require additional warmup time). The Performance Check instructions outlined in Section 5 provide a convenient method of checking the performance of this instrument.
ELECTRICAL CHARACTERISTICS
DISPLAY
Characteristics Performance
Plug-in Compatibility 2- and 3-Series plug-ins, except the
3S6, 3T6, 385 (horizontal only) CRT Type Electrostatic deflection Graticule Area 8 x 10 cm Phosphor P31 Typical CRT Accelerating Voltage
EXTERNAL INPUTS AND OUTPUTS
CRT Cathode Input (AC Coupled)
Low - 3 dB Frequency 1.8 kHz or less Z Axis Modulation 10 V or less (P-P) gives useful
Maximum Input Voltage
Calibrator
Output Voltages Into High Impedance (1 MΩ or greater)
Into 50.0
Current Loop 10mA DC or 10mA (P-P) square
3.5 kV
intensity variation
150 V
+4 mV, +40 mV, +0.4 V, +40 V
(ground to peak) square wave and
40 VDC (within 1-1/2%, +20°C to
+30° ;2 %, 0°C to +50°C
+2mV, +20mV, +0.2V (ground to
panel) square wave (within 1-1/2%,
+20°C to +30° C; 2%, 0° C to
+50°C
wave (within 1-1/2%, +20°C to
+30°C; 2%, 0°C to +50°C
Type R561B
Frequency 1 kHz within 1% Duty Factor 48% to 52% Rise-time and Fall-time
Line Voltage 115 VAC
Low 90 V to 110 V 180 V to 220 V Medium 104 V to 126 V 208 V to 252 V
High 112 V to 136 V 224 V to 272 V Line Frequency Range 48 Hz to 66 Hz Maximum Power Consumption at 115 VAC, 60 Hz
ENVIRONMENTAL CHARACTERISTICS
Characteristics Performance
Temperature
Non-operating
Operating Altitude
Non-operating To 50,000 feet
Operating To 15,000 feet
Finish Lacquered aluminum panels. Dimensions
Height
Width
Length
Accessories
Standard accessories supplied with the Type R561B are listed on the last pullout page of the Mechanical Parts List illustrations. For optional accessories available for use with this instrument, see the current Tektronix, Inc. catalog.
1 µs or less at all voltages with load capacitance of 100pF or less, except 40 V, 2.5µs or less at 40 V with load capacitance at 100 pF or less
POWER SUPPLY
230 VAC
186 W, 2.02 A
-40°C to +65°C 0°C to +50°C
PHYSICAL
Anodized aluminum front panel
7 inches 19 inches 20-1/2 inches
(A)
1-1A

SECTION 2

OPERATING INSTRUCTIONS
RM561A
Introduction
Before operating the Type RM561A Oscilloscope, be sure that the instrument will cool properly, the proper line voltage is used and the crt deflection potentials are correct. Function of front-panel controls and operating considerations are given in this section.
Cooling
A fan at the rear of the Type RM561A provides cooling. The entire fan assembly is in a snap-in mounting that can be mounted to exhaust air at the rear of the instrument or draw air in at the rear and blow it throughout the instrument. Direction of air flow con be changed to meet the operating conditions.
The Type RM561A can operate in ambient temperatures up to 50°C. If the instrument overheats, the thermal cutout turns off the power. When the internal temperature drops to a safe operating level, power is automatically restored. If wired for 117-volt operation, the fan will continue to operate when the thermal cutout opens. On 234-volt operation, power for the fan is turned off when the thermal cutout opens.
Line Voltage
The Type RM561A can be wired for either 117-volt or 234-volt operation. It will operate properly between 105 and 125 volts when wired for 117-volt operation and between 210 and 250 volts when wired for 234-volt operation. Converting from one operating voltage to the other requires a change in the power transformer primary connections, fan connections, fuse and line cord plug. Figs. 2-1 and 2-2 indicate the transformer primary and fan connections for
117-volt and 234-volt operation. A 3-amp slow-blow type fuse is required for 117-volt operation and a 1.25-amp slow-blow type fuse is required for 234-volt operation.
Front Panel Controls
The POWER ON switch and SCALE ILLUM. control permit turning the instrument power on or off and provide adjustment of the brightness of the graticule markings.
The FOCUS control adjusts the trace or spot focus.
The ASTIG. control is used in conjunction with the FOCUS control, to assure proper focus over the entire crt display.
The INTENSITY control adjusts the crt display brightness.
The ALIGNMENT control permits electronic alignment of the crt trace to match the horizontal graticule markings.
The 7-position CALIBRATOR switch determines the peak-to­peak amplitude of the signal at the CAL. OUT connector. It also turns the Calibrator on or off.
Rear Panel Controls
The CRT CATHODE SELECTOR switch permits the operator to select either (1) internal DUAL-TRACE CHOPPED BLANKING to remove dual-trace chopped mode switching transients from the crt display, or (2) Z-axis intensity modulation by external signals.
Operation
Operation of the Type RM561A Oscilloscope with two plug-in units in place is much the same as that of a Tektronix oscilloscope with corresponding vertical and horizontal deflection systems built into the main frame. The plug-in
Fig. 2-1. Power transformer primary connections for operation at (a) 117 volts or (b) 234 volts.
(A)(A)
Fig. 2-2. Fan terminal strip located near thermal cutout. (a) 117 volts, (b) 234 volts
.
2-1
Operating Instructions--Type RM561A
units are the vertical and horizontal deflection systems; their outputs connect directly to the deflection plates of the cathode­ray tube. The plug-in units can be selected to give the Type RM561A Oscilloscope the degree and type of performance required for a particular application.
The controls and switches on the Type RM561A affect the display; however, the plug-in units determine the major characteristics of the system.
Calibrator Output
The Calibrator output signal of the Type RM561A provides a convenient signal source for setting the gain of an amplifier plug-in unit or the basic timing of a time-base plug-in unit. The 1 V position also provides a 0.1 volt amplitude signal when connected to a 50-ohm system. This is very useful for calibrating sampling plug-in units. See the plug-in unit Instruction Manual for complete adjustment procedure.
In addition to these applications, the Calibrator output signal can be used as a convenient square-wave signal source for other applications.
NOTE When using the Calibrator output signal as a signal source for a system sharing a common ground with the Type RM561A (including plug-ins used in the instrument) be sure that the outer conductor of the CAL. OUT connector is connected to the chassis (or ground). Otherwise, the current through the Calibrator ground resistor, R899, may cause an incorrect Calibrator output signal.
CRT CATHODE SELECTOR Switch
The CRT CATHODE SELECTOR switch provides two modes of operation; CHOPPED BLANKING for dual-trace amplifier plug-ins and EXT CRT CATHODE for intensity modulation of the display.
The CRT CATHODE SELECTOR switch should be left in the CHOPPED BLANKING position except when external intensity modulation is used. For normal operation the shorting plug should connect the EXT CRT CATHODE jack to ground. Remove the plug for Z-axis modulation:
Display Combinations
The plug-in unit in the right opening of the Type RM561A controls the horizontal (X-axis) deflection of the crt beam. The unit in the left opening controls the vertical (Y-axis) deflection. Conventional Display. To obtain the conventional display of a horizontal sweep with vertical deflection by the input signal, insert a time-base plug-in unit in the right-hand opening and
an amplifier unit in the left. If a vertical sweep is desired with the input signal displayed horizontally, reverse the plug-ins. However, in this manner, the sweep retrace will not be blanked. Retrace blanking is provided only when a time-base unit is inserted in the right side of the Type RM561A.
X-Y Display. To obtain an X-Y display, insert amplifier plug-in units in both plug-in openings. Plug-in units with equal characteristics should be used.
Intensity Modulated Display. The crt beam can be intensity modulated, (Z-axis modulated) by applying a signal to the EXT CRT CATHODE jack on the rear panel. Depending on the setting of the INTENSITY control, the crt beam can be turned on with a negative pulse, or off with a positive pulse.
To intensity modulate the display remove the shorting strap from between the EXT CRT CATHODE and GND jacks. Set the CRT CATHODE SELECTOR switch to the EXT CRT CATHODE position and apply the modulating signal to the EXT CRT CATHODE jack.
Changing Plug-in Units
Before inserting a plug-in unit into the Type RM561A Oscilloscope, make sure the latching bar (bottom front) is not pointing down., Then, push the plug-in unit all the way into the opening. Turn the aluminum knob at the bottom center of the unit clockwise until it is tight. To remove the unit, turn the knob counterclockwise several turns and pull the unit out of the plug-in compartment.
CAUTION Although most plug-in units can be inserted or removed without damage when the power is on, best protection is provided for all units by turning the power off before changing units.
Different plug-in units apply slightly different dc voltages to the crt deflection plates. The Type RM561A crt has higher horizontal deflection sensitivity than vertical deflection sensitivity. Also, changing one plug-in unit con affect the accuracy of the other unit installed. Therefore, if one or both of the units are replaced, check the gain of both units. If a plug-in needs adjustment, follow the directions in the applicable Instruction Manual to adjust gain or sweep timing.
When the plug-in units are changed, the FOCUS and ASTIG, controls on the Type RM561A may need readjustment.
The Type RM561A can be operated with only one plug-in unit if desired. For example, moving film recording may be used in place of a sweep plug-in unit. To operate with only one plug-in unit, it will be necessary to elevate the unconnected crt deflection plates to about +170 to +210 volts dc vertical or +150 to +180 volts dc horizontal. This will provide proper action of the FOCUS and ASTIG. controls.
(A)(A)
2-2

SECTION 2A

OPERATING INSTRUCTIONS R561B
Type R561B
Introduction
To effectively use the Type R561B, the operation and capabilities of the instrument must be understood. This section of the manual describes the operation of the front­and rear-panel controls and connectors, and gives first time and general operating information.
Rack-mounting
Complete information for rack-mounting installation of the Type R561B is given on the Rack-mounting fold-out pages at the back of this manual.
Line Voltage
The Type R561B can be operated from either a 1 5-volt or a 230-volt nominal line-voltage source. The Line Voltage Selector assembly on the rear panel converts the instrument from one operating range to the other. In addition, this assembly changes the primary connections of the power transformer to allow selection of one of three regulating ranges. The assembly also includes two fuses to provide the correct protection for the instrument; both fuses are connected for 230-volt nominal operation, and only one fuse is connected for 115-volt nominal operation. Use the following procedure to obtain the proper line voltage and regulating range settings of the Line Voltage Selector.
1. Disconnect the instrument from the power source.
2. Loosen the two captive screws which hold the cover onto the voltage selector assembly; then pull to remove the cover.
3. To convert from 115 volts nominal to 230 volts nominal line voltage, pull out the Voltage Selector switch bar (see Fig. 2-1); turn it around 180- and plug it back into the remaining holes. Change the line-cord power plug to match the power-source receptacle or use a 115- to 230-volt adapter.
4. To change regulating ranges, pull out the Range Selector switch bar (see Fig. 2-1); slide it to the desired position and plug it back in. Select a range which is centered about the average line voltage to which the instrument is to be connected (see Table 2-1).
5. Re-install the cover and tighten the two captive screws.
6. Before applying power to the instrument, check that the indicating tabs on the switch bars are protruding through the correct holes for the desired nominal line voltage and regulating range.

TABLE 2-1

Regulating Ranges
Regulating Range, Range Selector 115-Volts 230-Volt Switch Position Nominal Nominal
LO (switch bar in 90 to 110 volts 180 to 220 volts
left holes)
M (switch bar in 104 to 126 volts 208 to 252 volts
middle holes)
HI (switch bar in 12 to 13 volts 224 to 272 volts
right holes)
Operating Temperature
The Type R561B can be operated where there ambient air
temperature is between 0°C and +50°C. The instrument can be
stored in ambient temperatures between -40°C and +65°C. After storage at a temperature beyond the operating limits, allow the chassis temperature to come within the operating limits before power is applied.
A fan at the rear of the Type R561B provides forced air cooling of the instrument. For proper circulation of air the instrument should normally be operated with the top and bottom covers in place. Do not block or restrict the air flow through the instrument.
A thermal cutout provides thermal protection and disconnects the power to the instrument if the internal temperature exceeds a safe operating level. This device will automatically reapply power when the temperature returns to a safe level.
Fig. 2-1. Line Voltage Selector assembly on the rear panel
(shown with cover removed).
CAUTION Damage to the instrument may result from incorrect Line Voltage Selector settings.
(A)
2-1A
Operating Instructions--Type R561B
SELECTION OF PLUG-IN UNITS
General Information
The Type R561B is designed to use Tektronix 2-series
and 3 series amplifier and time-base plug-in ’units for amplifying the vertical input signal and producing the time­base sweep. The use of plug-in permits the selection of display modes, bandwidth, sensitivity and number of inputs so that the oscilloscope performance can be changed to meet changing measurement needs.
NOTE Programmable plug-in Types 3S6 and 3T6 are not compatible with the Type R561B. However, the capability of these plug-ins is available (without programmability) with Types 3S5 and 3T5. Use the Type 3B5 in the right-hand (horizontal) compartment only. Refer to the instruction manual of the plug-in unit for specific information.
To install a plug-in unit in the Type R561B, push it all the way in to the plug-in compartment, then turn the locking knob ,at the bottom of the front panel) clockwise until it is tight. To remove the unit, turn the knob counterclockwise until the latch releases, then pull the unit out.
The accuracy of measurements made with the Type R561B depends on the calibration of the plug-in units used. Since the plug-in units drive the deflection plates directly, each unit must be adjusted to match the deflection sensitivity of the particular CRT that it drives. Therefore, the gain or sweep timing adjustment must be checked each time a plug-in unit, is changed. On most units, gain or timing calibration is made with a front-panel screwdriver adjustment. (Sampling units require special adjustment procedures.) Refer to the plug-in unit instruction manual for the required procedure. Since the various plug-in units present different output voltage levels to the deflection plates, the FOCUS and ASTIGMATISM controls will also require re-adjustment.
Plug-in units can be changed without turning off the instrument power, but it is recommended that the power be turned off while the change is made.
Either or both of the plug-in units can be operated on special extension cables for troubleshooting. For normal operation, power, the units must be installed in the oscilloscope. If the instrument is operated with the plug-in units on extension cables, the high-frequency response and fast sweep rates will be affected.
Display Combinations
The plug-in unit in the left plug-in compartment controls the vertical (Y-axis) deflection of the CRT beam, and the unit in the right plug-in compartment controls the horizontal (X-axis) deflection. The following paragraphs discuss some of the display combinations that con be obtained.
Time-Base Displays. To produce a conventional time­base, or Y-T display, an amplifier plug-in unit is used in the vertical (left) plug-in compartment and a time-base plug-in unit is used in the horizontal (right) plug-in compartment.
If a vertical sweep is desired with the input signal displayed horizontally, the time-base unit is inserted in the left compartment
and the amplifier unit in the right compartment. However, if a vertical sweep is used, there is no retrace blanking, no chopped blanking multi-trace displays and the delaying-sweep intensification does not operate because these circuits are associated with the horizontal (right) plug-in compartment.
X-Y Displays. Two amplifier units may be used to produce either a single or a multiple X-Y display (for example, for phase comparison measurement). Plug-in units with equal phase shift will produce an accurate X-Y display; however, for high-frequency X-Y operation, use of two units of the some type is recommended. Careful factory adjustment of deflection-circuit capacitance to a standard value in the Type R561B minimizes the high-frequency phase-shift between two plug-in units of the same type. For multiple X-Y displays, both synchronization and automatic pairings are provided for some amplifier units. Refer to the instruction manual of the plug-in unit to be used.
Multi-Trace Displays. The use of a dual-trace or multi­trace amplifier unit permits almost simultaneous viewing of two or more signals. For a multi-trace amplifier unit with single channel trigger capability ("Channel 1 Trigger"), the trigger signal is applied through the Type R531B to the trigger circuit in the time-base unit. This permits triggering from a single input signal and the CRT display shows the time relationship between the various signals. When using a plug-in without single-channel triggering, an external trigger is recommended to establish time relationship in multi-trace displays. In chopped mode, the multi-trace blanking pulses from the amplifier unit are applied internally through the Type R561 B to the CRT cathode to blank the CRT beam while it is switched from one channel to another (with the amplifier unit in the left compartment and the time-base unit in the right compartment, and with the rear panel CRT CATHODE SELECTOR switch in the CHOPPED BLANKING position).
Delayed Sweep. A delayed-sweep time-base unit is convenient for detailed viewing of pulse-train segments that occur a relatively long time after the maximum-amplitude (triggering) portion of the signal. The portion of the pulse train to be displayed by the delayed sweep may be intensified on the delaying-sweep display. A delayed-sweep unit that has a calibrated time delay can also be used for making very accurate (within 1%) time measurements
Raster Generation. A roster display can be presented by using two time-base units, one in each compartment. Intensity modulation can be achieved through the Z-axis of the CRT by applying the signal to the EXT INPUT connector on the rear panel and setting the CRT CATHODE SELECTOR to the EXT INPUT position.
Sampling. The apparent bandwidth of the oscilloscope can be increased to as much as one Gigahertz through the use of sampling plug-in units. A sampling time-base unit must be used with a sampling amplifier unit in the Type R561 B in order to produce the sampling display (even for X-Y operation). Generally, sampling and conventional plug-in units cannot be used together in the oscilloscope. However, Type 3S1 and 3S2 sampling amplifiers do have limited compatibility with conventional time-base units and the Type 3T5 sampling time-base unit con generate a real-time staircase sweep usable with conventional amplifiers.
Spectrum Analysis. Spectrum analyzer plug-in units can be used in conjunction with conventional time-base units
2-2A (A)
Operating Instruction--Type R561 B
to produce a spectral display ( a graph of the relative amplitude distribution as a function of frequency).
FUNCTIONS OF CONTROLS AND
CONNECTORS
A brief description of the function or operation of the front­and rear-panel controls and connectors follows. See Fig. 2-2 for locations.
Front Panel
ASTIGMATISM Used in conjunction with FOCUS
(Screwdriver control to obtain a well-defined display.
adjustment) FOCUS Control Used to optimize focus. INTENSITY Control Controls display brightness. TRACE ALIGNMENT Permits alignment of the trace with
(Screwdriver respect to the horizontal graticule lines.
adjustment) SCALE ILLUM Varies illumination of the graticule grid
Control lines. POWER Switch Used to apply or remove instrument
input power.
POWER Indicator Lamp bulb which indicates that AC
power is applied to the instrument.
CALIBRATOR Switch Provides selection of one of several
values of square wave voltage or a calibrated DC voltage. A calibrated DC or square wave current can also be selected.
CAL OUT Connector BNC connector at which calibrator
output voltage is available.
10 mA Current Loop Convenient means for calibrating
current probes.
Rear Panel
CRT CATHODE Permits .election of normal CRT SELECTOR operation, chopped blanking (blanking
Switch of the between-channel switching
transients when using multi-channel plug-in units in the chopped mode) and external CRT cathode input (permitting intensity modulation of the CRT by an external signal).
EXT INPUT BNC connector by which an external signal
Connector can be applied to the CRT cathode.
Line Voltage Provides quick method of changing
Selector transformer taps to allow instrument to
operate over a wide range of line voltages.
Horizontal and Permit installation of auxiliary inputs
Vertical and outputs through rear panel. Connector Holes
FIRST TIME OPERATION
The following procedure, using normal single-channel time­base mode, will demonstrate the basic operation of this instrument and its plug-in units.
1. Install a 2-Series or 3-Series amplifier plug-in unit in the vertical (left) plug-in compartment and a 2-Series or 3-Series time­base plug-in unit in the horizontal (right) plug-in compartment. Lock the plug-in units in place with their locking screws.
2. Set the POWER switch to the off position (pushed in).
3. Connect the power cord from the Type R561B to the proper line voltage.
NOTE The LINE VOLTAGE SELECTOR assembly on the rear panel should be checked to be sure the Voltage Selector and Range Selector switch bars are in the proper positions for the line voltage applied.
4. Set the instrument controls as follows:
Type R561B
INTENSITY Counterclockwise FOCUS Centered SCALE ILLUM Counter-clockwise ASTIGMATISM Centered (Screwdriver adjustment) CALIBRATOR 4 V CRT CATHODE SELECTOR NORM
(rear panel)
Amplifier Unit
(For example: Type 3A6)
Position Centered Mode Normal (Channel 1) Volts/Div 2 Variable (Volts/Div) Calibrated Input Coupling DC
Time-Base Unit
(For example: Type 3B3)
Position Centered Time/Div .5 ms Variable (Time/Div) Calibrated Magnifier Off Sweep Mode Normal
(A)
2-3A
Operating Instructions--Type R561 B
Fig. 2-2. Front-and rear-panel controls and connectors.
2-4A
(A)
Normal-Single Sweep Normal Level Free run (clockwise) Triggering Source Internal Slope + Coupling Auto
5. Connect a patch cord from the CAL OUT connector to
the Channel 1 input connector of the amplifier plug-in unit.
6. Set the POWER switch to ON (pulled out) and allow a
few minutes for warm up.
7. Adjust the INTENSITY control to obtain a display of moderate brightness. The amplifier unit Position control may have to be adjusted to position the free running trace on the CRT screen. Do not turn the intensity higher than is necessary for adequate observation of the display.
8. Trigger the display by adjusting the time-base Level control.
9. Set the SCALE ILLUM control so the graticule illumination is approximately equal to the intensity of the display.
10. Adjust the time-base Position control to position the start of the trace at the left edge of the graticule.
11. Adjust the FOCUS and ASTIGMATISM controls for a sharp well-defined display over the entire trace length. (If a focused display cannot be obtained, see Astigmatism Adjustment later in this section under General Operating Information.)
12. Adjust the TRACE ALIGNMENT screwdriver adjustment to align the display with the graticule lines.
13. Check the gain and DC balance of the amplifier unit and the timing adjustment (Sweep Cal) of the time-base unit as given in the instruction manuals for those units before making any voltage or time measurements. (In this demonstration, the calibrator waveform should be displayed as two divisions per cycle and two divisions in amplitude.)
GENERAL OPERATING INFORMATION
General
Refer to the amplifier unit manual for the following: Compensation of probes, selection and use of input cables, coupling and attenuation, and measurement of input signal voltage and phase (X-Y). Refer to the time-are unit manual for selection of triggering sources and coupling, and for measurements of time intervals, frequency and phase (linear measurement).
Scale Illumination
The CRT graticule is edge-lighted by three small lamps at the bottom. The lighting can be adjusted to suit the ambient light conditions by means of the SCALE ILLUM control. Rotating the control clockwise increases the brightness of the graticule scale markings.
Intensity Control
The setting of the INTENSITY control may affect the correct focus of the display. Slight adjustment of the FOCUS control may be necessary when the intensity level is changed.
(A)
Operating Instruction--Type R561B
Do not leave a bright, sharply focused spot on the CRT screen for a prolonged period. An excessively bright stationary spot may damage the CRT phosphor.
Astigmatism Adjustments
For most displays, the trace can be adequately focused using only the front-panel FOCUS control. However, whenever a large change is made in the beam intensity (to offset large changes in sweep rates or triggering repetition rates), or when plug-in units are changed, adjustment of the ASTIGMATISM control may also be required for a sharp display.
To check for proper setting of the ASTIGMATISM adjustment, slowly turn the FOCUS control through the optimum setting with a signal displayed on the CRT screen. If the ASTIGMATISM adjustment is correctly set, the vertical and horizontal portions of the trace will come into sharpest focus at the same position of the FOCUS control.
To set the ASTIGMATISM adjustment, use the following procedure:
1. Connect a 4 V Calibrator signal to the vertical input and set the corresponding Volts/Div switch to present 2.5 divisions of vertical deflection.
2. Set the Time/Div switch .2 ms.
3. Adjust the INTENSITY control so that the rising portion of the display can be seen.
4. Alternately adjust the FOCUS and ASTIGMATISM controls so that the horizontal and vertical portions of the display are equally focused.
Graticule
The graticule of the Type R561B is marked with eight vertical and 10 horizontal divisions. Each division is one centimeter square. In addition, each major division is divided into five minor divisions on the center vertical and horizontal lines. With the vertical gain and horizontal timing calibrated to the graticule, accurate measurements can be made from the CRT. The illumination of the graticule lines can be varied with the SCALE ILLUM control.
Trace Alignment Adjustment
If a free-running trace is not parallel to the horizontal graticule lines, set the TRACE ALIGNMENT adjustment as follows: position the trace to the center horizontal line, and adjust the TRACE ALIGNMENT adjustment so the trace is parallel with the horizontal graticule lines.
1 kHz Calibrator
The 1 kHz Calibrator provides a convenient source of square waves of known amplitude at an accurate frequency of one kilohertz. The output square-wave voltages available at the CAL OUT connector are 4 mV, 40 mV 0.4 V, 4 V and 40V. The loading of a terminated 50 system at the CAL OUT connector will provide output square-wave voltages of 2 mV, 20 mV and 0.2 V. A constant 40-volt DC level is also provided.
2-5A
Operating Instructions--Type R561 B
The current link provides 10 milliamperes, available as either DC or a square-wave current signal, which can be used to check and calibrate current probe systems. This current signal is obtained by clipping the probe around current loop. The arrow indicates conventional current (i.e., positive to negative).
Intensity (Z-Axis) Modulation
Intensity modulation can be used to relate other voltage information to the display signal without changing the shape of the waveform. The modulating signal is AC-coupled to the CRT cathode through the rear-panel EXT INPUT connector and the CRT CATHODE SELECTOR switch. With the INTENSITY
control set correctly, a positive excursion will dim or blank the CRT beam, and a negative excursion will brighten the beam (see Section 1 for amplitudes).
Time markers may be applied for direct time reference of the display or for establishing the sweep rate when uncalibrated deflection is used. Fast-rise pulses of short duration provide best resolution with respect to time. If the markers are not time-related to the displayed wave-form, a single sweep display is required. If sine waves are used for Z-axis modulation, the minimum usable frequency is about 250 hertz, due to AC coupling at the input. Be sure the CRT CATHODE SELECTOR is in the NORMAL position when the EXT INPUT connector is not in use, to avoid random intensity modulation from stray signals.
(A)
2-6A

SECTION 3

CIRCUIT DESCRIPTION
Introduction
The Tektronix Type RM561A Oscilloscope has a low­voltage power supply circuit, a cathode-ray tube circuit, and a calibrator.
The low-voltage power supply circuit provides the regulated and unregulated power used by the instrument and the plug-in units.
The crt circuit has the necessary controls and input facilities needed to give a sharp trace of useable intensity.
Two negative high-voltage power supply outputs are used for the cathode, focus element, and control grid of the crt.
The calibrator produces amplitude-calibrated square waves.
LOW-VOLTAGE POWER SUPPLY
The low-voltage power supply circuits have regulated outputs of -100, -12.2, +125, and -300 volts, and unregulated output of +420 volts. These circuits use silicon diode rectifiers and series-regulators. Each regulator circuit has a vacuum tube (or transistor, in the case of the -12.2-volt supply) in series with the load. This tube’s series plate resistance (and current) is controlled to maintain a constant voltage drop across the load. For example, if the load increases the series tube plate resistance decreases to pass more current to the load. If the load decreases it passes less current. In both cases, the voltage across the load remains the same.
The -100-, +125-, and +300-volt regulated supplies require shunt resistors when supplying more current than can be handled by the series tube. A shunt allows some of the load current to bypass the supply-series-regulator tube. The size of the shunt is very important. If the shunt resistance is too high, the series tube can overheat; if the resistance is too low, the supply can fail to regulate. If a plug-in draws an amount of current that calls for a supply shunt, the correct shunt resistance is located in that plug-in unit. When the unit is plugged into the Type RM561A, the shunt is connected around the series regulator tube. The plug-in portion of the shunt is always in series with a resistor located in each power supply circuit.
Power for the Type RM561A Oscilloscope and its plug-in units is supplied through the power transformer T601. The two primary windings of T601 are connected in parallel for 117-volt operation, or in series for 234-volt operation, as shown on the schematic diagram.
RM561A
The series tube plate resistance changes to hold the load voltage constant. R616, -100 VOLTS adjustment, determines the percentage of the total divider voltage applied to the grid of V634A and thus controls the output voltage. When this control is properly adjusted, the output is exactly -100 volts.
Should the output voltage tend to change because of a change in input voltage or a change in load current, the potential at the grid of V634A will change a proportional amount. Any change at the grid of V634A is amplified by V634A and Q624 and applied to the grid of V627. The resulting grid change at V627 will cause its plate resistance to change in the direction needed to bring the output back to --100 volts. C616 improves the response of the regulator to sudden changes in output voltage R628 is part of the series tube shunt. connected by some plug-in units.
+ 125-Volt Supply
The -100-volt supply serves as a reference for the +125 volt supply. With the R651 end of the divider R650-R651 fixed at -100 volts, any change in the +125-volt output produces a proportional change at the grid of V654. This change is amplified and supplied to the grid of the series regulator tube, V667A. The change at the grid of V667A changes its plate resistance to bring the output voltage back to 4-125 volts. R656, the +125 VOLTS control, determines the percentage of total divider voltage applied to the grid of V654 and permits adjustment of the output voltage. W hen this control is properly adjusted, the output is exactly +125 volts. C650 improves the response of the regulator to sudden changes in output voltage. .R666 is port of the series tube shunt.
+ 300-Volt Supply
The + 300-volt supply works the some as the + 125-volt supply. To supply the voltage for the + 300-volt regulator, rectified voltage from the transformer pins 21 and 22 is added to the voltage supplying the + 125-volt regulator. R676, the -+ 300 VOLTS control, adjusts the output voltage. The + 300-volt supply has an unregulated output of + 420-volts for the crt circuit.
- 12.2-Volt Supply
Operation of the --12.2-volt regulating circuit is essentially the same as that of the other regulating circuits, except that transistors are used instead of vacuum tubes. The base of Q734 is fixed near ---12 volts by the voltage divider R731-R732 between ---100 volts and ground. Any variation of the --12 2-volt output at the emitter of Q734 is amplified by 0734 and 0744 to change the collector resistance of 0757 which is in series with the lead R730, the -12.2 VOLTS control, allows adjustment of the voltage applied to the base, of Q734 and thus the output voltage.
- 100 Volt Supply
Reference voltage for the -100-volt supply is established by the gas diode, V609. The constant voltage drop across V609 establishes a fixed potential of about --85 volts at the grid of V634B. Voltage at the grid of V634A is established by the divider R616, R617, and R618. The difference in voltage between the two grids of V634 determines the plate current cf V634A. Plate current of V634A determines the base voltage of transistor Q624 which in turn determines the grid voltage of the series tube, V627.
(A)(A)
3-1
Circuit Description--Type RM561A
When this control is properly adjusted, the output is exactly -
12.2 volts. F720 protects Q757 in case of an overload on the -
-12.2-volt supply.
CRT CIRCUIT
The crt circuit contains the cathode-ray tube and two high­voltage supplies (one for the crt and focus element, the other for the control grid). The circuit also contains the necessary controls and signal input facilities.
Cathode-Ray Tube
A Tektronix T5032-31-1 ceramic-envelope cathode-ray tube is used in the Type RM561A. The accelerating voltage is approximately 3500 volts, developed by about -3300 volts at the cathode and an average deflection plate voltage of about +200 volts. With this accelerating voltage, the nominal vertical and horizontal deflection factors are 19.5 and 18.4 volts per centimeter respectively.
Deflection blanking of the crt beam is used in the Type RM561A. The crt contains a special set of deflection plates, pins 6 and 7, for this purpose. Both plates are connected to +125 volts; however pin 6 is also driven by the horizontal plug­in unit.
During sweep time, or if no plug-in unit is installed, both plates rest at +125 volts and permit the electron beam to pass on to the crt phosphor. During sweep retrace, however, pin 6 is driven considerably away from +125 volts by the right-hand plug-in unit. This scatters the beam and prevents it from being displayed.
High-Voltage Supplies
Energy for both high-voltage supplies is furnished by T801. V800, the primary of T801, and the stray circuit capacitance form a Hartley oscillator which operates at about 45 kc.
The output of one secondary winding of T801, rectified by V822, provides voltage for the crt cathode and focus element. This voltage is about -3300 volts at the crt cathode, and between about -2200 and -3000 volts at the focusing element, depending on the setting of the FOCUS control. The 6.3-volt crt heater is elevated to the cathode potential by R851.
The output of the other secondary winding of T801 is rectified by V832 for the control grid. The grid voltage ranges from -3200 to -3450 volts, depending on the setting of the INTENSITY control. The reference to ground for this supply is determined by the voltage at the junction of diodes D838 and D839. The voltage at this junction, plus the setting of the INTENSITY control, determines the crt bias and therefore the intensity of the display.
Two neon bulbs, B856 and B857, keep the voltage between the grid and cathode of the crt within safe limits. If the voltage exceeds about 140 volts, the neons fire and the voltage reduces to about 120 volts.
Regulation of the -3300-volt supply is accomplished through feedback from the arm of R841. The -3450-volt supply is regulated indirectly by mutual coupling in T801. If, because of loading or a change in input voltage, the output of the -3300-volt supply changes, a proportionate change occurs at the arm of R841. This change is amplified by V814 and is coupled to the screen of V800. The resulting change in screen voltage of V800 will increase or decrease the amplitude of oscillations in V800, changing the output voltage of T801 in the direction needed to return the high voltage to the correct level. The HIGH VOLTAGE control, R841, permits adjustment of the output voltage by setting the bias on V814B.
Deflection Signals
Push-pull signals for the deflection plates appear at pins 17 and 21 of the plug-in connector. The effective deflection circuit capacitance these signals see at the connector affects the band­pass and phase shift of each plug-in unit. C760 and C761 (Interconnecting Socket diagram) are set at the factory.
Intensifying Signals
Two signals may be used to modulate the intensity of the crt display. First, intensifying signals from a two-sweep (delaying­sweep) time-base plug-in unit are applied to the grid supply through pin 14 of the right-hand Interconnecting Socket. W hen the overall display intensity is reduced with the INTENSITY control, positive intensifying pulses from a two-sweep time-base plug-in unit will brighten any desired portion of the display.
The ground return for the crt grid supply can be either through D838 in the case where the plug-in unit does not supply intensifying signals or through D839 to a negative voltage in the intensifying circuit of the plug-in unit. In the second case D838 is back-biased and the junction of D838 and D839 is at a low negative voltage. With the same setting of the INTENSITY control, the first case will provide a brighter display. The second case provides a slightly dimmer display due to the plug-in unit negative voltage. Intensification results when the plug-in unit positive pulse turns D839 off and the crt grid supply return again becomes D838. R837 and C837 then couple the leading edge of the intensifying pulse directly to the crt grid.
Other external intensifying signals can be fed to the crt cathode through the EXT CRT CATHODE jack. Depending on the setting of the INTENSITY control, a negative pulse of 5 volts or more will turn the crt beam on.
Crt Controls
The INTENSITY control, R833, has a range of about 250 volts to control the crt bias and permit changing the intensity of the display.
The FOCUS control, R844, adjusts the focus of the crt by varying the voltage at the focusing anode through a range from about -2200 to -3000 volts.
The ASTIG. control, R864, has a 300-volt adjustment range.
3-2 (A)(A)
Circuit Description--Type RM561A
The GEOMETRY control, R865, adjusts the geometry by varying the voltage of the crt isolation shield through a range from +180 to +246 volts.
The TRACE ALIGNMENT control, R860, rotates the display so it can be aligned with the graticule.
CALIBRATOR
The calibrator for the Type RM561A Oscilloscope produces line-frequency amplitude-calibrated square waves. The 6.3-volt (approximately 18 volts peak-to-peak) ac heater voltage for V884 is supplied through C876 to the cathode of V884A. The signal at the plate of V884A is coupled to the grid of V884B to turn that tube on and off.
(A)(A)
Regenerative feedback from the plate of V884B to the grid of V884A speeds up the switching action, and drives V884A into and out of cutoff.
The voltage present at the cathode of V884B during the time that V884B is conducting is adjusted to exactly 100 volts with the CAL. AMPL. adjustment, R871.
The voltage divider in the cathode circuit of V884B contains precision resistors to provide an output accuracy of 3% or better at the various settings of the CALIBRATOR control.
When the CALIBRATOR control is set to the IV position, there will be a 0.1-volt output when the CAL. OUT connector is terminated in 50 ohms.
3-3

SECTION 3A

CIRCUIT DESCRIPTION
Type R561B
R561B
Introduction
This section of the manual contains a description of the circuitry used in the Type R561B Oscilloscope. Each circuit is described in detail using a detailed block diagram to show the interconnections between the stages in each major circuit and the relationship of the front-panel controls to the individual stages. Complete schematic diagrams are located at the rear of this manual.
LOW-VOLTAGE POWER SUPPLY
General
The Low-Voltage Power Supply circuit provides the operating power for this instrument from four regulated supplies. Electronic regulation is used -to provide stable, low­ripple output voltages. Each regulated supply contains a short protection circuit to prevent instrument damage if a supply is inadvertently shorted to ground or to another supply. The voltage input stage includes the Voltage Selector Assembly which allows selection of the nominal operating voltage and regulating range for the instrument. Fig. 3-1 shows a detailed block diagram of the Low-Voltage Power Supply.
Power Input
Power is applied to the primary of transformer T1 through fuse F1, POWER switch SW1, thermal cutout TK1, Voltage Selector switch SW2 and Range Selector switch SW3. SW2 connects the split primaries of T1 in parallel for 115-volt nominal operation, or in series for 230-volt nominal operation. SW3 allows three ranges of regulation by changing the taps on the primary windings to fit different line requirements. A second fuse, F2, is connected into the circuit when SW2 is set to the 230V position to provide the correct protection for 230­volt operation.
Thermal cutout TK1 provides thermal protection by interrupting the power if the instrument overheats. When the temperature returns to a safe level, TK1 automatically closes to re-apply the power.
- 100 Volt Supply
The -100-Volt Supply provides the reference voltage for the remaining supplies. The output from the secondary of T1 is rectified by bridge rectifier D8A-D. This voltage is filtered by C9, then applied to the -100-Volt Series Regulator stage to provide a stable output voltage. The Series Regulator can be compared to a variable resistance which is changed to stabilize the output voltage. The conductance
of the Series Regulator stage is controlled by the Error Amplifier to provide the correct regulated output voltage.
The Error Amplifier consists of Q12 and Q14, which are connected as a comparator. The output at the collector of Q14 indicates any voltage variations which occur at the base of Q14 relative to the fixed voltage at the base of Q12. Zener diode D10 maintains a fixed 9-volt drop, setting the base of Q12 at about -9 volts. The base level of Q14 is determined by the voltage divider network R18-R19-R20-R21-R23. R23, the -100 Volts adjustment, allows the operating point of the Error Amplifier to be adjusted to set the output voltage of the supply at -100 volts. R13 is the emitter resistor for both comparator transistors and the current through it divides between Q12 and Q14. The output current of the Error Amplifier stage controls the conduction of the Series Regulator stage. This is accomplished as follows: Assume that the output voltage increases (becomes more negative) because of a change in load or an increase in line voltage. This negative­going voltage change at the output is applied to the base of Q14, reducing the conduction of Q14. As current through Q14 is reduced, Q24 base current increases. This results in increased Q24 collector current, increasing the voltage drop across R25 and R26 and pulling the base of Q28 negative. The emitter of Q28 follows the base; hence, the base of Q32 is also pulled negative. Reduced current through Series Regulator Q32 decreases current through the load, causing the output voltage to decrease (become less negative) to its correct level. These changes occur rapidly, and the effect is to maintain unchanged output voltage. In a similar manner, the Series Regulator and Error Amplifier stages compensate for output changes due to ripple. As will be seen in subsequent paragraphs, R33 determines the limit current for the Series Regulator stage, and thus for the load. Transients beyond the frequency range of the regulator are filtered by C31 to prevent their appearance on the output voltage.
When the power switch is activated, diode D25 provides a base current path for Q28, allowing the -100-Volt Supply to turn on first, since all the other supplies are dependent upon its output. As the -100-Volt Supply output builds up to its correct level, D25 is reverse biased and remains off during normal operation of the instrument.
The Short-Protection Amplifier stage, Q30, protects the -100­Volt Supply if the output is shorted, and also serves to limit the current demanded from the Series Regulator under excessive load. During normal operation, divider R30-R31 sets the base of Q30 to a point below the turn-on level of the transistor. When excess current is demanded from Series Regulator Q32 due to an overload or short circuit, the additional current through R33 raises the emitter of Q32 more positive. This produces a corresponding change at the base of Q32, which is connected through R30
(A)
3-1A
Circuit Description--Type R561B
Fig. 3-1. Power Supply detailed block diagram.
3-2A
(A)
Circuit Description--Type R561B
to the base of Q30. This positive-going change biases Q30 into conduction. As a result, less current is available to Q28, to Q32 and to the load, thus causing the supply to lose regulation. R31 senses the decrease in load voltage and adds to increasing base current of Q30. As the collector of Q30 goes negative, conduction of Q28 and Q32 is further decreased. Thus the output current is decreased and remains low until the excessive load is removed. D19, together with divider R18-R19-R20, provides protection to the --12.2-Volt supply in the event the -100-Volt Supply is shorted to the -
12.2-Volt Supply by causing the -100-Volt Supply to lose regulation, and therefore lose reference voltage for the -12.2­Volt Supply. D31 protects the -100-Volt Supply from damaging polarity reversal if it is shorted to either the +125­Volt Supply or the +300-Volt Supply.
- 12.2-Volt Supply
Rectified voltage for operation of the -12.2-Volt Supply is provided by D35A-B, filtered by C36 and applied to the -12.2­Volt Supply Series Regulator stage. Reference voltage for this supply is provided by voltage divider R42-R43 between the regulated -100-Volt Supply and ground. If the -12.2-volt output changes, a sample of the change appears at the base of Q46 as an error signal. Regulation of the output voltage is controlled by Error Amplifier Q44-Q46-Q49 and Series Regulator Q51 in a manner similar to that described for the ­100-Volt Supply. Transients beyond the frequency range of the regulator are filtered by C47.
Short protection is provided by Q38 and R38. For normal operation, the emitter-base voltage of Q38 is not enough to bias it into conduction. However, when the output is shorted, the high current demanded from the -12.2-Volt Supply is drown through R38, producing a voltage drop sufficient to forward bias Q38. Q38 collector current then produces an increased voltage drop across R40, reducing the conduction of both Q49 and Q51 to limit the output current. R39 protects Q38 from sudden current surges by limiting the base current. D47 protects the -12.2-Volt Supply from damage if it is shorted to either the +;125-Volt Supply or the +300-Volt Supply.
+ 125-Volt Supply
Rectified voltage for operation of the +125-Volt Supply is provided by D53A-D, filtered by C54 and applied to the +125­Volt Supply Series Regulator stage. The +125-volt output is summed with the -100-volt reference through divider R62-R63, and the summation is applied through R61 to the base of Q60 and compared to the grounded base of Q58. If the +125-volt output changes, a sample of the change appears at the base of Q60 as an error signal. Regulation of the output voltage is controlled by Error Amplifier Q58-Q60-Q66-Q68 and Series Regulator Q74 in a manner similar to that described for the ­100-Volt Supply. Transients beyond the frequency range of the regulator are filtered by C97B-C to prevent their appearance on the output voltage.
Short protection for this supply is provided by the Short­Protection amplifier stage, Q70, which functions in a manner similar to that described for Q30 in the --10Volt Supply. D62 protects the Error Amplifier from damage if the output of the +125-Volt Supply collapses or goes negative, causing C62 to
rapidly discharge and reverse bias 060. Diode D75 protects electrolytic capacitor C97B-C and the transistors in the circuit from damaging polarity reversals in the event the +300-volt output is shored to ground or to one of the negative supplies. D76 causes the +125-Volt Supply to go into current limiting when the +300-Volt Supply is shorted to ground or to one of the negative supplies.
+ 300-Volt Supply
Rectified voltage for operation of the +300-Volt Supply is provided by D77A-D, filtered by C78 and applied from the negative side of the rectifier to the +300-Volt Supply Series Regulator stage. The +300-volt output is summed with the -100­volt reference through divider R80-R81, and the summation is applied through R83 to base of Q84 and compared to the grounded base of Q86. If the +300-volt output changes, a sample of the change appears at the base of Q84 as an error signal. Regulation of the output voltage is controlled by Error Amplifier Q84-Q86-Q88-Q90 and Series Regulator Q96 in a manner similar to that described for the -100-Volt Supply. Transients beyond the frequency range of the regulator are filtered by C97A. The load current through Series Regulator Q96 also posses through the +125-Volt Supply Series Regulator, Q74. However, this does not affect the limit current of the +125-Volt Supply.
Shorting protection for this supply is provided by the Short­Protection Amplifier stage, Q91, which functions in a manner similar to that described for Q30 in the -100-Volt Supply. D80 protects the Error Amplifier from damage if the output of the +300­Volt Supply collapses or goes negative, and D95 protects the transistors in the circuit from damaging polarity reversals in the event the +125-Volt Supply is shorted to ground or to one of the negative supplies.
6.5-Volt RMS AC Source
The four 6.5-volt RMS secondary windings of T1 provide power for the CRT heater, the plug-in heaters via J11 and J21, the pilot light, B7, and the scale illumination lights, B4, B5, and B6. Current through the scale illumination lights is controlled by the SCALE ILLUM control, R4, to change the brightness of the graticule lines.
DEFLECTION CIRCUITS
horizontal and vertical signals for deflecting the CRT beam are received through pins 17 and 21 of each plug-in unit and applied to the respective deflection plates of the CRT. The effective deflection circuit capacitance encountered by each of these signals at the plug-in connector affects the bandwidth and phase shift of the plug-in unit. Compensating capacitors C102 and C109 (shown on the Plug-In Connectors diagram) are factory-adjusted to set the effective capacitance at a standard value of 14.3 picofarads to ensure plug-in compatibility.
1 kHz CALIBRATOR
General
The 1 kHz Calibrator circuit produces a square-wave output with accurate amplitude and frequency. This output is available as a square-wave voltage at the CAL OUT connector or as a square-wave current through the 10 mA probe current
(A)
3-3A
Circuit Description--Type R561B
Fig. 3-2. 1 kHz Calibrator detailed block diagram.
loop. An accurate +40 volts DC level is also available. The CALIBRATOR switch selects the attenuation of the output signal to provide square-wave voltage outputs between 40 volts and 4 millivolts (between 0.2 volts and 2 millivolts into 50 ohms) peak to peak. Fig. 3-2 shows a detailed block diagram of the 1 kHz Calibrator circuit.
Calibrator Multivibrator
The Calibrator Multivibrator is comprised of Q151 and Q159, and is a free-running emitter-coupled multivibrator. The circuit operates in a symmetrical manner and the output is an accurate one-kilohertz square wave. Only an approximate 9­volt change is exhibited at the emitters of Q151 and Q159, so that an essentially constant current of about 0.8 mA is maintained through resistors R150 and R1 58.
Refer to the wave shapes shown in Fig. 3-3 for this discussion. With the CALIBRATOR switch, SW150, in all positions except 1 mA DC and OFF, the emitters of Q151 and Q159 are returned to the +125-volt supply through D151-R150 and D159-R158. Assume that the multivibrator has just switched states; Q151 is off and Q159 is on. This is To in Fig. 3-3. The base potential of Q159 is set to about -11.0 volts by voltage divider R153-R154-R156 to ensure that Q159 will not saturate. The voltage at the anode of D159 is about -9.8 volts because of the voltage drop across two forward biased junctions. Capacitor C157 had about a 2-volt charge as switching occurred; thus, the voltage at the anode of D151 is about -7.8 volts, cutting off Q151. C157 begins to charge toward the +125-volt supply via R150. Total current through Q159 is about 1.6 mA; 0.8 mA through R158 and 0.8 mA through C157 and R150.
After about 0.5 milliseconds (corresponding with T, in Fig. 3-3), C157 has charged to the turn-on level of Q151 and D151. At this point, the capacitor has a charge of about 11 volts and the potential at the anode of D151 is about +1.2 volts. The capacitor charging current through Q159 ceases as Q151 and D151 begin
to conduct. As the collector of Q151 (hence the base of Q159) rises, Q159 and D159 are switched off and C157 begins to discharge through R158. The C157-R158 current sums with R150 current through Q151, producing an approximate 9-volt positive­going step at the base of Q159.
C157 continues to discharge, and after 0.5 milliseconds (T, in Fig. 3-3), the voltage at the anode of D159 has risen to forward-bias Q159 and D159. As Q159 begins to conduct, the anode of D159 is clamped at about 0.8 volts and discharge action of C157 is halted. The current through Q151 decreases, causing its collector to introduce a negative-going step, which is connected through the Q159 base-emitter junction and D159 to C157. Because C157 cannot obtain an instantaneous charge, the anode of D151 is pulled negative to reverse bias D151 and Q151. Q151 turns off, and its collector falls rapidly to about -11.0 volts, resulting in an approximate 9-volt negative-going step applied through Q159 and D159 to C157. The anode of D151 is pulled down to about -7.8 volts, completing the cycle.
The Calibrator Multivibrator circuit has been designed to repeat the preceding sequence at an accurate one-kilohertz frequency. However, since a tolerance range of the passive components does exist, the frequency can be adjusted by varying slightly the amplitude across C157 during the charge-discharge cycle. This is accomplished by adjustment of R154, Frequency, which determines the potential on D159 anode at the instant the diode turns on. For example, with greater amplitude, longer charge and discharge times are required, thus lowering the frequency.
Output Stage
The output stage consists of the Current Switch, Q162, and the Divider Network. During the half cycle that Q159 is conducting, current is injected into the base of Q162. Q162 saturates and its collector drops to about -12 volts, reverse biasing D168. With D18 off, there is no current through R170 and R171, and the output level at the cathode of D168 drops to zero volts.
(A)
3-4A
When Q159 turns off, Q162 turns off and D161 turns on to protect the Q162 base-emitter junction from reverse-bias breakdown. D164 and D168 turn on, and the output of the circuit (at D168 cathode) is dependent upon voltage divider R166-R167-R170-R171 between +125 volts and ground. This output level is set to exactly +40 volts by adjustment of R166, Amplitude. When this adjustment is made, the current through the divider is an accurate 10 mA, which is available at the current probe loop in the 10 mA positions of the CALIBRATOR switch.
The signal voltage available at the CAL OUT connector is determined by the divider network (made up of precision resistors) and the setting of the CALIBRATOR switch. In the 10 mA DC (40 VDC) position, the Calibrator Multivibrator is inoperative so that a +40-volt DC output level is produced. R173 is placed in series with the R166-R167-R170-R171 resistance to obtain an effective resistance of 450 ohms with 4 volts applied, as seen by the CAL OUT connector in the 4 V position of the switch. This effective resistance becomes port of the output voltage divider in the positions of 0.4 V and below (these positions have an accurate 50- ohm output resistance, which when terminated by 50 ohms can further divide the outputs by two, providing outputs of 0.2V, 20 mV and 2 mV). In the 10 mA position, the CAL OUT connector is grounded.
R183, which is about ten times the resistance of the braid of a 42-inch coaxial cable, cancels any ground loop current that may exist between the CAL OUT connector and some other instrument chassis.
CRT CIRCUIT General
The CRT Circuit provides the high voltage and control circuits necessary for operation of the cathode-ray tube (CRT). Fig. 3-4 shows a detailed block diagram of the CRT Circuit.
High Voltage Oscillator
Q219 and its associated circuitry comprise a class C oscillator to provide the drive for the high-voltage transformer, T220. When the instrument is turned on, conduction of Q214 provides a base current path for Q219. The collector current of Q219 increases and a voltage is developed across the collector winding of T220. This produces a corresponding voltage increase in the feedback winding of T220 which is connected to the base of Q219, causing it to conduct harder. While Q219 is conducting, C217 charges negatively to the peak to peak voltage of the feed-back winding. Eventually the rate of collector current increase in Q219 becomes less than that required to maintain the voltage across the collector winding and the voltage drops.This turns off Q219 by way of feedback voltage to the base. During the interval that Q219 is not conducting, the negative charge on C217 is partially removed through Q214. Q219 remains off until the feedback voltage on the base is near the peak positive value again. The cycle repeats at a frequency of 40 to 50 kilohertz. The amplitude of sustained oscillation depends upon the average current delivered to the base of Q219, and finally, the average Q219 collector current.
Circuit Description--Typo R561 B
Fig. 3-3. Calibrator Multivibrator waveforms.
High Voltage Regulator
Feedback from the secondary of T220 and +125 volts is summed through the voltage divider network consisting of R200, R201, R206, R208 and R233 through R238, and the difference is applied to the gate of Field-Effect Transistor Q211.. This sample of the output voltage is compared to the regulated --12.2-volt level at the source of Q211. It is then inverted and amplified by Q211 and applied to the base of Q214. Amplitude of the oscillations at the collector of Q219 is determined by the average collector current of Q214.
Regulation. is accomplished as follows: If the output voltage at the -330Q V test point starts to go positive (becomes less negative), a sample of this positive-going voltage is applied to the gate of Q211. Conduction of Q211 is increased, and as its drain goes negative because of the voltage dropped across R211, the base current of Q214 is increased. An increase in conduction of Q214 increases the average collector current, which is applied through the feedback winding of T220 to the base of Q219. Q219 conducts harder, increasing the collector current to produce a larger induced voltage in the secondary of T220. This increased voltage appears as more negative voltage at the -3300 V test point to correct the original positive-going change. By sampling the output from the cathode supply in this manner, the total output of the high-voltage supply is held constant.
(A)
3-5A
Circuit Description--Type R561B
Fig. 3-4. CRT Circuit detailed block diagram.
Output voltage level of the high-voltage supply is controlled by the High Voltage adjustment, R206, in the gate circuit of Q211. This adjustment sets the effective divider ratio, which in turn determines the voltage necessary to satisfy the quiescent condition of Q214 and Q219 in the manner described for a change in output voltage. Neon bulb B209 and diode D212 protect the FET, Q211, from damage due to excessive voltage.
High Voltage Rectifiers and Output
The high-voltage transformer, T220, has two output windings. These windings provide the negative CRT cathode potential and the CRT control grid bias.
The accelerating potential for the CRT cathode is supplied by the half-wave rectifier D221 and held constant by the High­Voltage Regulator stage in the primary of T220. The output level is adjustable to about -3300 volts on the cathode by the High Voltage adjustment mentioned previously. (The 6.5-volt CRT cathode heater is also elevated to the cathode potential through R246.)
Half-wave rectifier D260 provides a negative voltage for the control grid. The voltage applied to-the-control-grid is determined by the setting of the INTENSITY control (to be
(A)
3-6A
Circuit Description--Type R561B
discussed in the next paragraph), the CRT Grid Bias control (R269) and any intensification signals received from the time­base plug-in unit (delayed sweep and sampling units only). Reference to ground for this supply is set by the conduction of D272.
Beam current is controlled by R225, INTENSITY. As the control is rotated clockwise, the wiper arm moves toward -100 volts. This more negative DC reference voltage is applied to the secondary winding controlling the CRT cathode, reducing the voltage demanded of the winding to maintain -3300 volts at the -3300 V test point. This is accomplished by the regulator circuit. The voltage across the grid winding is also reduced, which results in a more positive voltage applied to the CRT control grid, thus increasing beam current. Beam current is reduced in a like manner by rotating R225 counterclockwise.
Neon bulbs B277, B278 and B279 provide protection to the CRT if the voltage difference between the control grid and the cathode exceeds about 135 volts.
CRT Control Circuits
In addition to the INTENSITY control discussed previously, the FOCUS and ASTIGMATISM controls have been incorporated for arriving at the optimum CRT display. FOCUS control R237 provides the correct voltage for the second anode in the CRT. Proper voltage for the third anode is obtained by adjusting ASTIGMATISM control R257. In order to obtain optimum spot size and shape, both the FOCUS and ASTIGMATISM controls are adjusted to provide the proper electrostatic lens configuration in the CRT. The TRACE ALIGNMENT control, R259, permits adjustment of the DC current through beam-rotation coil L259 to align the display with the horizontal graticule lines. The Geometry adjustment, R256, controls the overall geometry of the display.
Blanking
The CRT beam is blanked by a special set of deflection plates in the CRT. One of the plates (pin 7) is connected directly to the +-125-volt supply. The second plate (pin 6) is connected through plug-in connector J21 to the horizontal plug-in unit. When there is no sweep, a quiescent voltage is
applied from the horizontal unit to create a difference of potential between the two plates. This voltage can either be positive or negative with respect to the +125 volts on the other plate. The potential difference created is sufficient to deflect the CRT beam so that it is absorbed in the deflection structure and does not reach the screen.
The CRT beam is unblanked whenever the two deflection plate voltages become equal. For example, if a sweep occurs or if the horizontal plug-in unit is removed, the voltages are made equal and the beam is allowed to pass through to the CRT screen. Sweep unblanking is produced by either a positive or negative gate pulse (depending on the quiescent level) applied to pin 6, equaling the +125 volts normally present at pin 7. In a like manner, when the horizontal plug-in unit is removed, the two deflection plates are equalized through R244 at +125 volts.
Intensity Modulation
The intensity of the CRT display may be modulated by applying signals to either the grid or the cathode of the CRT.
Intensifying signals from a delayed sweep time-base plug-in unit are applied to the grid supply via pin 14 of the horizontal plug­in interconnecting socket, J21. These signals brighten the delayed-sweep portion of the delaying-sweep display. When the time-base unit is set to Intensified, the control grid supply is referred to a negative voltage in the intensifying circuit through D275, reducing the overall display intensity. At this time, D272 is reverse biased by the negative voltage at the juncture of the two diodes. Intensification results when the positive-going pulse from the time-base unit reverse biases D275 and the grid supply is referred to ground through D272 (as for normal operation). The positive-going pulse is then coupled through R275 and C275 to the CRT control grid. Thus the brightened portion of the display is the same intensity as a normal display and the background trace is dimmed.
External modulating signals may also be applied to the CRT by way of the cathode, through the rear-panel EXT INPUT connector, J255, and the CRT CATHODE SELECTOR, SW255. With the INTENSITY control adjusted properly, a positive or negative pulse between 3 and 50 volts in amplitude will produce dimming or intensification of the CRT beam.
When a multi-channel vertical plug-in amplifier that provides dual-trace chopped blanking pulses is used, the blanking pulses are applied via the interconnecting socket J11 and the CRT CATHODE SELECTOR to the CRT cathode circuit. These pulses are about 5 volts in amplitude, and at normal intensity levels are sufficient to cut off the CRT beam during the time the amplifier channels in the vertical plug-in unit are being switched.
(A)
3-7A

SECTION 4

MAINTENANCE
R561A
PREVENTIVE MAINTENANCE
Cleaning
Occasionally blow the dust out of the instrument with a low-velocity dry air stream. Remove persistent dirt with a small paint brush or damp cloth. The screen on the fan con be snapped out and should be cleaned as needed. Check it frequently.
Fan Oiling
The fan should be oiled with a few drops of oil about every six months. An industrial hypodermic syringe and needle is used to insert oil through a protective rubber cap located under the fan label. The oil recommended is Anderol L826 from the Lehigh Company or Rotron distributors, but if not available, a good light machine oil may be used. If a syringe and needle cannot be obtained locally, you can order them through the NICP by specifying Hypodermic Syringe, Tektronix Part No. 003-282 and Hypodermic Needle, Tektronix Part No. 003-285.
Fig. 4-1 shows how to oil the fan. Place the needle at the point on the label shown in Fig. 4-1. With the needle at about 45’, pierce the label and rubber cap (located under the label); insert the needle about 1/4" and depress the syringe-plunger to inject a few drops of oil.
CORRECTIVE MAINTENANCE
Replacing Parts
Most parts in the Type RM561A Oscilloscope can be replaced without detailed instructions. Some parts, however, should be replaced by using definite procedures. These procedures are described in the following paragraphs.
A replaced part may affect instrument calibration. Check and adjust where needed.
Soldering Precautions and Procedure
In Tektronix instruments, parts are connected to ceramic terminal strip notches with solder containing about 3% silver. The bond between the notch and ceramic strip may be broken by repeated use of ordinary 60 40 tin-lead solder or by excessive heat. Therefore, when resoldering parts to a ceramic strip use solder containing 3% silver and do not overheat the work. Occasional use of ordinary solder, however, will not break the bond. Usually 3% silver solder is available locally; or one-pound rolls may be ordered from the NICP.
The following soldering procedure may be used to remove and replace parts on a ceramic terminal strip.
1. Use 50-to-70-watt soldering iron with a wedge shaped tip. (With this type tip you can heat the solder slot without overheating the strip.)
2. Tin the soldering iron tip with silver-bearing solder.
3. Heat the parts soldered to the ceramic strip only enough to make the solder flow freely. Do this by touching one corner of the soldering iron tip to the notch. (Be careful: excessive pressure will break the ceramic strip.)
4. When you remove a part, pull its lead out of the notch while the solder is hot.
5. When you replace a part use only enough solder to cover the wires and form a small fillet in the notch.
6. Clip excess leads of parts replaced and be sure to remove all clippings from the instrument.
Fig. 4-1. Fan motor oiling with hypodermic
(A)(A)
Replacing Ceramic Terminal Strips
Damaged ceramic terminal strips can be replaced by the following procedure. Fig. 4-2 shows how ceramic strip parts are assembled.
1. Unsolder all connections to damaged ceramic strip.
2. Cut off one side of each plastic yoke holding the old
ceramic strip.
3. Remove old ceramic strip.
4. Remove remainder of old yokes from spacers.
4-1
Maintenance--Type RM561A
Fig. 4-2. Ceramic terminal strip assembly.
5. Replace old spacers with new ones. (If not damaged,
spacers can be reused.)
6. Set new ceramic strip yoke pins into spacers.
7. Drive new yoke pins completely into spacers by pressing or lightly tapping the ceramic strip directly above the yokes. Be careful, don’t break strips.
8. Cut off portion of new yoke pins protruding through spacers on side of chassis opposite the ceramic strip.
9. Resolder connections to new ceramic strip using the information headed Soldering Precautions and Procedure.
Replacing Calibrator Switch
The entire switch should be replaced and can be ordered
either wired or unwired.
falls directly under the grounding strip. Tighten the base clamp screw.
After the crt is replaced, it may be necessary to adjust the TRACE ALIGNMENT and HIGH VOLTAGE controls and the deflection-circuit capacitance (C760 and C761) according to the calibration procedure. Also check the calibration of time-base and amplifier plug-in units.
Cathode-Ray Tube
WARNING When replacing crt, wear a plastic face mask and protective gloves for protection in case tube implodes.
To remove the cathode-ray tube, disconnect the four leads connected to the neck of the tube, the tube socket, and loosen the tube clump on tube base. Remove the crt bezel, light reflector and light shield. Pull the crt straight out through the front panel, being careful not to bend or break the crt neck pins. The rubber gasket and implosion shield can be removed from the crt face after it is out of the instrument.
Install the new crt by the reverse of the preceding procedure. When replacing the implosion shield be sure that the notched side is down. Follow the color-code information on the tube shield when the crt neck pin leads are replaced. When the crt is properly installed, the back of its faceplate is flush with the instrument front panel.
Correct position of the shield extension, grounding strip and Mylar sleeve is important when re-installing the crt. Replace these components on the positions shown in Fig. 4-
3. Note that the gap between the ends of the shield extension
Fig. 4-3. Correct position of shield extension and Mylar sleeve.
TROUBLESHOOTING
If trouble occurs in the Type RM561A Oscilloscope, a five­step procedure may be used for repair. First, confirm the trouble. Next, isolate trouble to a plug-in unit or to the Type RM561A. Localize a trouble in the Type RM561A to the Power Supply, Crt Circuit, or Calibrator. Troubleshoot the correct circuit to find the defective parts. Replace defective parts.
Confirmation of Trouble
Improper control settings may at times give indications of trouble. Therefore, you should be sure that the apparent trouble is not caused by improper front panel control settings. For example, an improper setting of the SOURCE or COUPLING switch on a time-base unit can cause apparent trigging troubles: an improper setting of the VARIABLE
(A)(A)
4-2
MaintenanceType RM561A
control on an amplifier unit can cause an apparent decrease in sensitivity.
When a trouble exists, it must be isolated to one of he
plug-in units or to the Type RM561A indicator.
Isolating Trouble
Isolating trouble to a plug-in unit or to the indicator can be done in one of two ways, depending on whether or not there are spare plug-in units available.
If trouble is found to be in the plug-in unit, refer to the instruction manual for that unit. A plug-in extension (Part Number 013-034, available through the NICP) allows the units to be operated while extended out through the front of the plug-in opening.
Spare Plug-In Units Available
If more than two plug-in units are available, the easiest way to isolate trouble is to replace the plug-in units one at a time. The faulty unit can then be isolated by noting when proper operation is restored. If plug-in replacement
does not correct the trouble, the Type RM561A is probably faulty.
NOTE A non-sampling plug-in unit cannot be operated with a sampling plug-in unit.
Spare Plug-In Units Not Available
When spare plug-in units are not available, the crt display will be of great help. Trouble will normally show up as an erroneous display, or as no display at all. Since the crt display will help isolate the trouble, the following procedure is divided according to the type of display.
No Trace Or Spot. If no trace can be obtained on the crt, remove both plug-in units and vary the INTENSITY control. A spot should appear. If no spot appears, the trouble is in the Type RM561A. If a spot does appear when both plug-in units are removed reinsert each unit separately. After warm up, vary its POSITION control. If the spot or trace cannot be returned to the approximate center of the crt when a single plug-in unit is in the indicator, the trouble
Fig. 4-4. Power-Supply test points (1) through (17). CRT Circuit test points, (20) through (26). Calibrator test points (4)
through (45).
4-3
MaintenanceType RM561A
is probably In that unit. (if a lime-base unit is installed, be sure it is set for free-running operation to unblank the crt.)
Insufficient Vertical Deflection or Improper Sweep
Timing. If the plug in unit front-panel gain controls do not
permit adequate vertical deflection or proper sweep timing, check the low-voltage and high-voltage power supply ,voltages (see schematics).
If the power-supply voltages are not correct, remove both plug-in units and check the voltages again. If they still are not correct, the trouble is in the Type RM561A indicator If the power-supply voltages are correct with both plug-in until removed, but incorrect with only one of the two inserted, the trouble is in that unit. If the power-supply voltages are correct with both plug-in units removed, but incorrect when either is inserted, the trouble is probably the power supply of the Type RM561A.
Improper Triggering. If external triggering and line triggering are satisfactory, but internal triggering is not, the trouble is probably in the Trigger Pick-off Circuit of the amplifier plug-in units. If satisfactory triggering cannot be obtained from any of the three sources external, internal or line) the trouble is probably in the time-base plug-in unit.
Waveform Distortion. If there is any waveform distortion, but no other indications of malfunction (such, as improper sweep timing) the trouble is probably in the vertical plug-in unit
Localizing Trouble In Type RM561A Indicator
If it has been determined that trouble exists in the Type RM561A, the location of the trouble can be further localized to the Power Supply, the Crt Circuit, or the Calibrator.
To localize the trouble to either the Power Supply or the Crt Circuit, measure the output of each of the regulated power supplies. If incorrect supplies are measured, the trouble is in the Power Supply or the power source. If all of the measured voltages are proper. the trouble is in the Crt Circuit.
Trouble in the Calibrator is indicated by an erroneous output at the CAL. OUT connector.
When it is known which circuit contains the trouble, that circuit can be checked to isolate the defective component
Circuit Troubleshooting
In troubleshooting a circuit, broken wires, worn switches, scorched parts, and improperly seated tubes can be found by visual inspection. (Before a scorched part is replaced the cause of the scorching should be eliminated.)
Tube failure is the most prevalent cause of trouble in Tektronix instruments. Therefore, ii a visual check falls to reveal the cause of trouble, oil tubes should be checked as the next step. Tubes should be checked by substitution, one at a time. (Tube­testers cannot adequately check all parameters.) Be sure to return all good tubes to their original sockets; otherwise the instrument may have to be recalibrated unnecessarily because rf different characteristics of the same tube type. If tube substitution does not correct the trouble, then check the rest of the circuit by voltage and resistance measurements. Voltages at various test points throughout the instrument are indicated on the schematics. The test points can easily be located using Fig. 4-4. The voltages at the test points are typical and may vary considerably from instrument to instrument. Resistance measurements in a circuit will usually be point-to-point checks for which the proper values can be approximated from the schematic diagrams. Table 4-1 may be used as an aid in relating trouble symptoms to the probable cause and the procedure to use to locate the defective component. Voltages listed in the table were obtained with the Type RM561A operated without plug-in units.
All signal wiring in the Type RM561A is colored to make Circuit tracing easier. In addition, all regulated power supply leads are color coded to follow the standard EIA code. The first color (widest stripe) indicates the first number of the voltage. The second color indicates the second and the third color (smallest stripe) indicates the multiplier. The voltage is positive if the main color of the wire is white and negative if the main color of the wire is tan. Four colors would be required to indicate all the digits plus a multiplier for the 4-125-volt leads. To avoid this, the +125-volt leads are coded as +120; that is, brown, red, brown on a white base. A -100-volt lead is coded brown, black, brown on a ton base while +300-volt leads are coded orange, black, brown on white
The switch wafer shown on the Calibrator diagram is coded. The letters F and R indicate whether the front or rear of the wafer is used to perform the particular switching function.
Table 4-1.
RM561A Troubleshooting Table
Symptoms Checks
- 100 Volt Supply
(Any change in this supply may change instrument calibration).
1. Incorrect output level, ripple, or regulation.
2. Output voltage high (too negative).
a. Check line voltage. b. Check Setting of - 100 volts adjustment (see Calibration Procedure). c. Check V609, V634, V627 and Q624 by substitution. d. If output still incorrect go to whichever symptoms apply of 2 through 7. b. Measure V634 bias, between pin 7 of V609 and pin 8 of V634; about 2 v. Use meter that can be elevated. (If bias is excessive check R618 and R616). c. Check R618, R616, R634, R633, and R624.
4-4
(A)(A) 1
Symptoms Checks
3. Output voltage low (too positive).
4. Poor regulation at either high, low or normal line voltage. (Supply should hold regulation within ± 1% throughout normal line voltage range).
5. Poor regulation at low line voltage.
6. Poor regulation at high line voltage.
7. Excessive ripple.
(Any change in this supply may change instrument
8. Incorrect output level, ripple, or regulation.
9. Output voltage high.
10. Output voltage low.
a. Check voltage between pins 17 and 18 of T601. 137 vac ± 10% check T601 primary circuit). b. Check voltage between C640A terminals: about 175 vdc ± 10%. (Incorrect: check C640A, C640B, R640, and D640). c. Check test point (14): about -80 vdc. (Incorrect: check V609). d. Check R617, R616, C616, C640B. a. Check for correct output voltage before checking regulation - also, check output voltage after regulation trouble is fixed. b. Check voltage at test point (17): about +85 vdc. (Voltage low: check D660 and do checks 2A, 2B, and 2C). c. Check for excessive loading by either plug-in unit or RM561A circuits.
Check R635, R628 and shunt resistors in plug-in unit.
Substitute another plug-in unit. (Symptom occurs if load is reduced while shunt resistor remains the same in plug­in unit). a. Check for correct output voltage and regulation before checking ripple - also check output voltage and regulation after ripple trouble is fixed. b. Check voltage at test point (17): about +85 vdc. c. Check ripple at test point (17): should be less than 25 v peak-to-peak with load. (Excessive: check C640A and D640). d. Check tubes. e. Check C616, C611, C640A and C640B.
+125-Volt Supply
calibration). a. Check line voltage. b. Check -100 volt supply output test point (7). c. Check setting of -12.2 VOLTS adjustment (see Calibration Procedure). d. Check V654, V667. e. If output still incorrect go to whatever symptoms apply of steps 9 through 14. a. Check voltage at test point (1): about +68 vdc. (Incorrect: check R657, R658, and V654). b. Check R650. a. Check voltage between pins 19 and 20 of T601: 160 vac +10%. (Incorrect: check primary circuit).
MaintenanceType RM561A
Symptoms Checks
b. Check voltage at test point (2): +215 vdc. (Incorrect: check C642A, R642, D642). c. Check test point (1): about +68 vdc. (Incorrect: check R659, V654). d. Check C650, R651.
11. Poor regulation at either high, low, or normal line. (Supply should hold regulation within 1% throughout normal line voltage range).
12. Poor regulation at low line voltage.
13. Poor regulation at high line voltage.
14. Excessive ripple.
Symptom and Checks similar to those described for the +125
(Any change in this supply may change instrument calibration).
15. Incorrect output level, ripple or regulation
16. Output high (too negative).
17. Output low or zero, (too positive).
a. Check for correct output voltage before checking regulation - also check output voltage after regulation trouble is fixed. b. Check tubes. c. Check voltage at test point (2): about +215 vdc. (Incorrect: do checks 10A and 10B).
Check R653, C650, R666, and shunt resistors on plug-in unit.
a. Check C650. b. Substitute another plug-in unit. (Symptom occurs if load is reduced while shunt resistor remains the same in plug-in unit). a. Check for correct output voltage and regulation before checking ripple-also, check output voltage and regulation after ripple trouble is fixed. b. Check tubes. c. Check voltage at test point (2): about +215 vdc. (Incorrect: do checks 10A and 10B). d. Check voltage at test point (1): about +68 vdc. (Incorrect: check R657, R659, R658). e. Check C642A, and B, C650, D642.
+300 Volt Supply
volt supply.
-12.2 Volt Supply
a. Check line voltage. b. Check -100 volt supply output test point (7). c. Check setting of -12.2 VOLTS adjustment (see Calibration Procedure). d. Check Q734, Q744, Q757, and F720. e. If output still incorrect go to whichever symptoms apply to steps 16 through 20. a. Check test point (9): about -12 vdc. (Incorrect: check C732, R732). b. Check test point (10): about +7.9 vdc. (Incorrect: check R734, C737). c. Check C737, R734, R744. a. Check voltage between pins 14 and 15 then 125 and 16 of T60: 15 vac ±10% with 117 (234) vac line. (Incorrect: check transformer primary circuit).
(A)(A) 1
4-5
MaintenanceType RM561A
Symptoms Checks
b. Check voltage between C720A and B terminals, test points (8) and (12): about 18 vdc. (Incorrect: check C720A, C720B, D720, D721). c. Check voltage at fuse F720: about +8.2 vdc. (Incorrect: check C757 and F720). d. Check test point (9): about -12 vdc. (Incorrect: check R731). e. Check test point (10): about +7.9 vdc. (Incorrect: check R735). f. Check C757.
18. Poor regulation at either high, low, or normal line voltage. (supply should hold regulation within ± 1% throughout normal line voltage range.
19. Poor regulation at low or high line voltage.
20. Excessive Ripple
(Any change in this circuit may change display vertical and
21. Any incorrect operation of CRT circuit.
a. Check for correct output voltage before checking regulation-also, check output voltage after regulation trouble is fixed. b. Check for excessive loading by either plug-in units or RM561A circuits.
Check R734, R735, R744, and C737.
a. Check for current output voltage and regulation before checking ripple-also, check output voltage and regulation after ripple trouble is fixed. b. Check ripple at test point (12): should be less than 4v peak-to-peak with load. (Excessive: do checks 17A and B). c. Check D720, D721, C720A and B, C732, C757.
CRT Circuit
horizontal calibration).
a. Check line voltage. b. Check HIGH VOLTAGE adjustment, (­3300 volts at test point (21), see Calibration Procedure). If minor adjustment restores operation circuit may be normal. c. Check V822 and V832 heaters color. Dull orange - normal bight - heavy load on either tube dim - poor tube (replace) both dark - oscillator inoperative.
Symptoms Checks
d. If operation still incorrect, go to whichever symptoms apply of 22 through 29
22. Low Intensity
23. High Intensity.
25. Poor astigmatism and/or focus
26. Blooming (size of display increases with intensity increase).
27. No intensifying pulse.
28. No Z-axis modulation or no chopped blanking
29. Intensity varies (unwanted Z­axis modulation).
30. Incorrect outputs.
a. Check intensity control setting. b. Check V822 and R835. c. Unblanking pulse doesn’t return pin 6 of crt fully to +125 v. d. Check crt. a. Check intensity control setting. b. Make sure spot or trace is not deflected off screen. (Spot should appear near center with both plug-in units removed). c. V822 heater open, (B856 and B857 glowing brightly show V822 open). If V822 is open check V814A. d. Check oscillator tubes and circuit including T801). e. Check C822, C842, and crt. a. Check FOCUS and ASTIG. Control settings. b. Check HIGH VOLTAGE at pin 2 of crt, test point (20): -3300 vdc. (Incorrect: check R852). c. Check voltage at pin 9 of crt: 0 to +300 vdc when ASTIG. Control is rotated. (Incorrect: check R864 and +300 vdc to R864). Check V800 and V822 (either may be weak).
a. D838 shorted, D839 open. b. C837 or R837 open (leading edge of long intensifying pulse missing). C853 or R853 open. If R853 open check for shorted C853.
D838 and/or D839 open.
CALIBRATOR
a. Check CAL. AMPL. Adjustment (see Calibration Procedure). b. Check V884. c. Check voltages at test points (4) through 45). d. Use Calibration Procedure, check divider in V884B cathode.
4-6
(A)(A) 1

SECTION 4A

Type R561B
MAINTENANCE
Introduction
This section of the manual contains information for use in preventive maintenance, corrective maintenance and troubleshooting of the Type R561B.
Access to the Interior
The top and bottom dust covers of the Type R561B can be easily removed for access to the internal circuitry. The covers should be re-installed on the instrument for normal operation to keep dust out and provide proper distribution of the air flow.
PREVENTIVE MAINTENANCE
General
Preventive maintenance consists of periodic inspection and cleaning at regular intervals. The Type R561B should be checked approximately every 500 hours of operation, or every six months, whichever occurs first. If the instrument is subjected to adverse environmental conditions, such as excessive dust, high temperatures or high humidity, the frequency of the checks should be increased. A convenient time to perform preventive maintenance is preceding re­calibration of the instrument.
Cleaning
The Type R561B should be cleaned as often as operating conditions require. Accumulation of dirt in the instrument can cause overheating and component breakdown. Dirt on components acts as an insulating blanket and prevents efficient heat dissipation. It also provides an electrical conduction path.
The top and bottom covers provide protection against dust in the interior of the instrument. Operation without covers in place necessitates more frequent cleaning.
CAUTION Avoid the use of chemical cleaning agents which might damage the plastics used in this instrument. Avoid chemicals which contain benzene, toluene, xylene, acetone or similar solvents.
Exterior. Loose dust accumulated on the outside of the
Type R561B can be removed with a soft cloth or small point brush. The point brush is particularly useful for dislodging dirt on and around the front-panel controls. Dirt which remains can be removed with a soft cloth dampened in a
R561B
mild detergent and water solution. Abrasive cleaners should not be used.
Clean the graticule and CRT face with a soft, lint-free cloth
dampened with a mild detergent and water solution.
Interior. Dust in the interior of the instrument should be removed occasionally due to this electrical conductivity under high-humidity conditions. The best way !o clean the interior is to blow off the accumulated dust with dry, low-pressure air. Remove any dirt which remains with a soft paint brush or a cloth dampened with a mild detergent and water solution. A cotton-tipped applicator is useful for cleaning in narrow spaces and/or circuit boards.
The high voltage circuits, particularly parts located in the high­voltage comportment and the area surrounding the CRT socket receive special attention. Excessive dirt in these areas may cause high-voltage arcing and result in improper instrument operation.
Lubrication
The reliability of potentiometers, rotary switches and other moving parts can be maintained if they are kept properly lubricated. Use a cleaning-type lubricant (e.g., Tektronix Part No. 006-0220-001. The pot lubricant can also be used on shaft bushings. Do not over lubricate.
Fan Oiling. During periodic servicing, the fan motor should be lubricated with a few drops of light machine oil (Anderol L826 available from Lehigh Company or Rotan Distributors is recommended). An industrial hypodermic needle and syringe is used to insert the oil through the rubber seal, as shown in Fig. 4-
1. Hold the syringe at a 45° angle, pierce the rubber seal, then insert the needle about 1/4 inch and depress the plunger for enough to inject 3 or 4 drops of oil into the hearing. If a syringe and needle cannot be obtained locally, they may be ordered from NICP (Tektronix Part No. 003-0282-00 for the syringe; 003-0285­00 for the needle).
Visual Inspection
The Type R561B should be inspected occasionally for such defects as broken connections, improperly seated transistors, damaged circuit boards and heat-damaged ports.
The corrective procedure for most visible defects is obvious; however, particular care must be taken if heat-damaged components are found. Overheating usually indicates other
(A)
4-1A
MaintenanceType R561B
Fig. 4-1. Oiling the fan with a hypodermic.
trouble in the instrument; therefore, it is important that the cause of overheating be corrected to prevent a recurrence of the damage.
Transistor Checks
Periodic checks of the transistors in the Type R561B are not recommended. The best check of transistor performance is its actual operation in the instrument. More details on checking transistor operation are given under Troubleshooting.
Recalibration
To assure accurate measurements, check the calibration of this instrument after each 500 hours of operation or every six months if used infrequently. In addition, replacement of components may necessitate recalibration of the affected circuits. Complete calibration instructions are given in the Calibration section.
The calibration procedure can also be helpful in localizing certain troubles in the instrument. In some cases, minor troubles may be revealed and/or corrected by recalibration.
TROUBLESHOOTING
Introduction
The following information is provided to facilitate troubleshooting in the Type R561B. Information contained in other sections of this manual should be used along with the following information to aid in locating the defective components. When replacing a defective semiconductor, be certain that all associated components are in good condition
before application of power, a precaution which may prevent further damage. An understanding of the circuit operation is very helpful in locating troubles. See the Circuit Description section for complete information.
Troubleshooting Aids
Diagrams. Circuit diagrams are given on fold-out pages in
Section 8. The component number and electrical value of each component in this instrument are shown on the diagrams. Each main circuit is assigned a series of component numbers. Table 4­1 lists the main circuits in the Type R561B and the series of component numbers assigned to each. Important voltages and waveforms are also shown on the diagrams at the rear of this manual. The portion of the circuit mounted on the circuit board is enclosed with a blue line.
Table 4-1
Component Numbers
Component
Numbers on
Diagrams
1.99 1 Power Supply 100-110 2 Plug-in
150-199 3 1 kHz Calibrator 200-299 4 CRT circuit
Switch Wafer Identification. Switch wafers shown on the diagrams are coded to indicate the position of the wafer in the complete switch assembly. The numbered portion of the code refers to the wafer number counting from the front, or mounting end of the switch, toward the rear. The letters F and R indicate whether the front or rear of the wafer performs the particular switching function. For example, a wafer designated 2R indicates that the rear of the second wafer J (from the front) is used for this particular switching function.
Circuit Boards. Fig. 4-5 through 4-7 show the circuit boards used in the Type R561B. Each electrical component on each board is identified by its circuit number. The circuit board is also outlined on its schematic diagram with a blue line. These pictures used with the diagrams will aid in locating the components mounted on the circuit boards.
Wiring Color Code. All insulated wire and cable used in the Type RSS61B is color-coded to facilitate circuit tracing. Signal carrying leads are identified with three stripes to indicate the approximate voltage, using the EIA resistor color code. A white background color indicates a positive voltage and a tan background indicates a negative voltage. The widest color stripe identifies the first color of the code. Table 4-2 gives the wiring color code for the power supply voltages used in the Type R561B.
Diagram
Number Circuit
Connectors

TABLE 4-2

Power Supply Wiring Color Code
Back-
ground
Supply +300 V White Orange Black Brown +125 V White Brown Red Brown
-12.2 V Tan Brown Red Black
-100 V Tan Brown Black Brown
Color
First
Stripe
Second
Stripe
Third
Stripe
4-2A
(A)
Fig. 4-2. Standard EIA color code
MaintenanceType R561B
Resistor Color Code. In addition to the brown composition
resistors, some metal-film resistors and some wire-wound resistors are used in the Type R561B. The resistance values of wire-wound resistors are printed on the body of the component. The resistance values of composition resistors and metal-film resistors are color-coded on the components with EIA color code (some metal-film resistors may have the value printed on the body). The color code is read starting with the stripe nearest the end of the resistor. Composition resistors have four stripes which consist of two significant figures, a multiplier and a tolerance value (see Fig. 4-2). Metal-film resistors hove five stripes consisting of three significant figures, a multiplier and a tolerance value.
Capacitor Marking. The capacitance values of common disc capacitors and small electrolytic are marked in micro­farads on the side of the component body. The white ceramic capacitors used in the Type R561B are color coded in picofarads using a modified EIA color code (see Fig. 4-2).
(A)
Diode Color Code. The cathode end of a glass-enclosed diode is
indicated by a stripe, a series of stripes or a dot. For most silicon or germanium diodes with a series of stripes, the color code identifies the three significant digits of the Tektronix Part Number using the resistor color-code system (e.g., a diode color-coded blue-brown-gray-green (6, 1, 8, 5) indicates Tektronix Part Number 152-0185-00). The cathode and anode ends of metal diodes can be identified by the diode symbol marked on the body.
Troubleshooting Equipment
The following equipment is useful for troubleshooting the
Type R561B:
1. Transistor Tester Description: Tektronix Type 575 Transistor Curve Tracer
or equivalent.
Purpose: To test the semiconductors used in this
instrument.
2. Multimeter
Description: Electronic Voltmeter, 10 megohms, or greater, input resistance; 0 to 500 volts; 0 to 50 megohms. Accuracy, within 3% (1% is necessary to check power supply voltages). Test prods must be insulated to prevent accidental shorting.
Purpose: To check operating voltages and for general
troubleshooting in this instrument.
NOTE A 20,000 ohms/volt VOM can be used to check the voltages in this instrument if allowances are made for the circuit loading of the VOM at high impedance points.
4-3A
MaintenanceType R561B
3. Test Oscilloscope (with 1X and 10X probes). Description: DC to 1 MHz frequency response, 1 milli-
volt to 10 volts/division deflection factor.
Purpose: To check waveforms in the instrument.
Troubleshooting Techniques
This troubleshooting procedure is arranged in an order which checks the simple trouble possibilities before proceeding with the extensive troubleshooting. The first few checks assure proper connection, operation and calibration. If the trouble is located by these checks, the remaining steps aid in locating the defective component. When the defective component is located, it should be replaced following the replacement procedures given under Corrective Maintenance.
1. Check Control Settings. Incorrect control settings con indicate a trouble that does not exist. If there is any question about the correct function or operation of any control, see the Operating Instructions section of the manual.
2. Check Associated Equipment. Before proceeding with troubleshooting of the Type R561B, check that the equipment used with this instrument is operating correctly. Substitute another vertical or time-base plug-in which is known to be operating properly. Check that the signal is properly connected and that interconnecting cables are not defective. Also, check the power source.
3. Visual Check. Visually check the portion of the instrument in which the trouble is located or suspected. Many troubles can be located by visual indications such as unsoldered connections, broken wire, damaged circuit boards, damaged components, etc.
4. Check Instrument Calibration. Check the calibration of this instrument, or the affected circuit if the trouble exists in one circuit. The apparent trouble may only be a result of misadjustments or may be corrected by calibration. Complete calibration instructions are given in the Calibration section.
5. Isolate Trouble to a Circuit. To isolate trouble to a circuit, note the trouble symptom. The symptom often identifies the circuit in which the trouble is located. For example, poor focus indicates that the CRT circuit (includes high voltages) is probably at fault. When trouble symptoms appear in more than one circuit, check affected circuits by taking voltage and waveform readings.
Incorrect operation of all circuits often indicates trouble in the power supplies. However, a defective component elsewhere in the instrument con appear as a power supply trouble and may also affect the operation of other circuits.
NOTE Turn the instrument off before attempting to remove or replace connections to any circuit board.
Table 4-3 lists the tolerances of the power supplies in his
instrument. If a power supply voltage is within the listed tolerance, the supply can be assumed to be working correctly. If outside the tolerance, the supply may be misadjusted or operating incorrectly. Use the procedure given in the Calibration section to adjust the power supplies.

TABLE 4-3

Power Supply Tolerances
Power Supply Tolerance
-3300 V Within 3%
-100V Within 0.5%
-12.2V Within 1.2% +125V Within 1.5% +300V Within 1.5%
6. Check Circuit Board Interconnections. After the trouble has been isolated to a particular circuit, check the pin connectors on the circuit board for correct connection. Figs. 4-5 through 4-7 show the correct connections for each board.
The pin connectors used in this instrument also provide a convenient means of circuit isolation. For example, a short circuit in a power supply can be isolated to the power supply itself by disconnecting the pin connectors for that voltage at the remaining boards.
7. Check Voltages and Waveforms. Often the defective component can be located by checking for correct voltage or waveforms in the circuit. Ideal voltages are given on the diagrams.
NOTE Voltages given on the diagrams are calculated with the assumption that conditions are ideal (variable components at design center, etc.), and may vary slightly from actual measured voltages.
8. Check Individual Components. The following
procedures describe methods of checking the individual components in the Type R561B. Components which are soldered in place are best checked by disconnecting one end. This isolates the measurement from the effects of surrounding circuitry.
A. TRANSISTORS. The best check of transistor operation is actual performance under operating conditions. If a transistor is suspected of being defective, it can best be checked by substituting a new component or one which has been checked previously. However, be sure that circuit conditions are not such that a replacement transistor might also be damaged. If substitute transistors are not available, use a dynamic tester (such as Tektronix Type 575). Static type testers may be used, but since they do not check operation under simulated operating conditions, some defects may go unnoticed. Fig. 4-3 shows transistor base pin and socket arrangements. Be sure power is off before attempting to remove or replace any transistor.
B. DIODES. A diode can be checked for an open or shorted condition by measuring the resistance between terminals. With an ohmmeter scale having an internal source of between 800 millivolts and 3 volts, the resistance should be high in one direction and low when the leads are reversed.
C. RESISTORS. Check the resistors with an ohmmeter. See the Electrical Parts List for the tolerance of the resistors used in this instrument. Resistors normally need not be
(A) 1
4-4A
MaintenanceType R561B
Fig. 4-3. Transistor electrode configuration data.
replaced unless the measured value varies widely from the specified value.
D. CAPACITORS. A leaky or shorted capacitor can be determined by checking resistance with an ohmmeter on the highest scale. Use an ohmmeter which will not exceed the voltage rating of the capacitor. The resistance reading should be high after initial charge of the capacitor. An open capacitor can best be detected with a capacitance meter or by checking whether the capacitor passes AC signals.
9. Troubleshooting in the Low Voltage Power
Supplies. The low voltage supplies incorporate special
circuitry to (A)
prevent damage due to short circuits on the voltage output lines. When this circuitry is operating properly, the low voltage supplies are extremely reliable. If any trouble occurs in the low-voltage power supply (Diagram 1), be certain that all defective components are replaced before reapplication of power. Otherwise uncorrected problems could cause further damage, including damage to the new component. Although not every situation can be predicted, such occurrences will be unlikely if the following precautions are taken:
A. Turn the power off.
4-5A
MaintenanceType R561B
B. If the problem is possibly in the -100-, or +125-, or +300-volt supplies, visually check current sensing resistors R33, R75 and R94 for charring or cracking. For location of components, refer to Fig. 4-5.
C. Check current-sensing transistors Q30, Q38, Q70 and Q91 on a transistor checker.
D. If a current-sensing resistor or transistor has failed, check all remaining diodes and transistors in the corresponding section of circuitry (see Table 4-4). Most of the diodes mentioned can be checked in the circuit using an ohmmeter.

TABLE 4-4

Component
Failure Check
Q38 D35A, D35B, D47, Q44, Q46, Q49, Q51 Q30 or R33 D8A, D8B, D8C, D8D, D19, D25, D31,
Q12, Q14, Q24, Q28, Q30, Q32 Q70, Q91, R75 or R94
E. Make a careful visual check of R8, R52, R53, R61,
R69, R77 and R82 for charring or cracks.
F. Re-apply the power. If a line-voltage auto-transformer is available, gradually increase the line voltage from zero volts to 115 volts (or to the center of the regulating range to which the Voltage Selector Assembly is set) while monitoring the output of the low-voltage supply.
10. Repair and Readjust the Circuit. If any defective parts are located, follow the corrective maintenance procedures in this section. Be sure to check the performance of any circuit that has been repaired or that has had any electrical components replaced.
General
Corrective maintenance consists of component replacement and instrument repair. Special techniques required to replace components in this instrument are given here.
Obtaining Replacement Parts
D53A, D53B, D53C, D53D, D62, D65, D75 D76, D77A, D77B, D77C, D77D, D80, D95, Q58, Q60, Q66, Q68, Q74, Q84, Q86, Q88, Q90, Q96
CORRECTIVE MAINTENANCE
Fig. 4-4. Removing or replacing component on circuit
board.
component will not adversely affect instrument performance.
Soldering Techniques
WARNING Disconnect the instrument from the power source before soldering.
Circuit Boards. Use ordinary 60/40 solder and a 35- to 40-
watt pencil-type soldering iron on the circuit boards. The tip of the iron should be clean and properly tinned for best heat transfer to the solder joint. A higher wattage iron may separate the wiring from the base material. The following technique should be used to replace a component without removing the boards from the instrument:
1. Grip the component lead with long-nose pliers. Touch the soldering iron tip to the lead at the solder connection. Do not lay the iron directly on the board, as it may damage the board. See Fig. 4-4.
When the solder begins to melt, pull the lead out gently. This should leave a clean hole in the board. If not, the hole can be cleaned by reheating the solder and placing a sharp object such as a toothpick into the hole to clean it out.
See page 7-0.
NOTE When selecting replacement parts, it is important to remember that the physical size and shape of the component may affect its performance in the instrument. All replacement parts should be direct replacements unless it is known that a different
(A)
4-6A
MaintenanceType R561B
(A) 1
Fig. 4-5 Low Voltage Power Supply and 1 kHz Calibrator circuit board.
4-7A
MaintenanceType R561B
3. Bend the leads of the new component to fit the holes in the board. Insert the leads into the holes in the board so the component is firmly seated against the board (or as positioned originally). If it does not seat properly, heat the solder and gently press the component into place.
4. Touch the iron to the connection and apply a small amount of solder to make a firm solder joint. To protect heat­sensitive components, hold the lead between the component body and solder joint with a pair of long-nose pliers or other heat sink.
5. Clip the excess lead that protrudes through the board (if not clipped in step 3).
6. Clean the area around the solder connection with a flux-remover solvent. Be careful not to remove information printed on the board. Metal Terminals. When soldering metal terminals (e.g., switch terminals, potentiometers, etc.) ordinary 60/40 solder can be used. Use a soldering iron with a 40- to 75-watt rating and a 1/,-inch wide wedge-shaped tip. Observe the following precautions when soldering metal terminals:
1. Apply heat only long enough to make the solder flow freely.
2. Apply only enough solder to form a solid connection. Excess solder may impair the function of the part.
3. If a wire extends beyond the solder joint, clip off the excess.
4. Clean the flux from the solder joint with a flux-remover solvent.
Component Replacement
WARNING Disconnect the instrument from the power source before replacing components.
Circuit Board Replacement. If a circuit board is
damaged beyond repair, either the entire assembly including all soldered-on components, or the board only, can be replaced. Part numbers are given in the Mechanical Parts List for either the completely wired or the unwired board. Many of the components mounted on the circuit boards can be replaced without removing the boards from the instrument. Observe the soldering precautions given under Soldering Techniques in this section. However, if the bottom side of the board must be reached or if the board must be moved to gain access to other areas of the instrument, the mounting screws need to be removed and it may be necessary to disconnect some of the interconnecting wires from the pin connectors. Refer to Figs. 4-5 through 4-7 for component locations and interconnecting wire identifying colors.
GENERAL:
Most of the connections to the circuit boards are made with pin connectors; however, the connections to the High Voltage circuit boards are soldered. See the special instructions to remove these boards as units.
Use the following procedure to remove a circuit board:
1. Disconnect all the pin connectors.
2. Remove all screws that hold the board to the chassis.
3. Lift the board out of the instrument. Do not force or bend
the board.
4. To replace the board, reverse the order of removal. Replace the pin connectors carefully so they mate correctly with the pins. If forced into place incorrectly positioned, the pin connectors may be damaged.
HIGH VOLTAGE BOARD REMOVAL:
1. Remove the metal high voltage shield (secured to the chassis by two nuts on the opposite side of the chassis).
2. Remove the plastic cover on the high voltage compartment (held in place with three screws).
3. Unsolder the two diodes connected between the boards and the high-voltage transformer. Extra care should be used to avoid damaging the plastic compartment with the soldering iron.
4. Ease the board assembly out of the plastic comportment, while unsoldering the wires on the side nearest the transformer. Unsolder other wires as necessary until the entire assembly can be removed as a unit.
5. To replace the boards, reverse the order of removal.
Fan Assembly Replacement. The fan assembly may be
removed from its snap-in mounting using the following procedure:
1. Pry off the snap-in protective grille.
2. From the interior of the instrument, depress the holding clip on one side by inserting a small screwdriver or other tool between the fan assembly and the clip. While the clip is depressed, pull the fan toward the rear with the other hand until it is clear of the holding fingers (about 1/2 inch). At this stage, the fan assembly should be tilted from the centerline, since the fingers on the opposite clip will still be engaged.
3. Depress the opposite holding clip while pulling the fan assembly to the rear. When the fan assembly is free of the holding fingers of both clips, it may be pulled free of the snap-in mounting to the length of the power leads.
4. To completely disconnect the fan assembly from the instrument, unsolder the two power leads.
5. To replace the fan, reverse the order of removal. Very little pressure is required to properly seat the fan assembly in the holding clips, provided the alignment is correct. To assure proper alignment, observe that the upper and lower ends of the holding clips are curved inward to form flanges and that the sides of the fan assembly are grooved to accept the flanges when alignment is correct.
Cathode-Ray Tube Replacement. The following procedure
outlines the removal and replacement of the cathode-ray tube:
A. REMOVAL:
1. Remove the bezel (held in place with four knurled nuts) and the plastic graticule light conductor. Also remove any filters that may be in front of the graticule.
4-8A
(A)
MaintenanceType R561B
(A)1
Fig. 4-6. Upper High Voltage circuit board
4-9A
MaintenanceType R561B
2. Disconnect the deflection-plate leads. Be careful not
to bend the deflection-plate pins.
3. Remove the fan assembly as outlined in the preceding
paragraphs.
4. Remove the CRT base socket.
5. Loosen the three screws on the CRT clamp inside the CRT shield (base end). Do not remove the screws. (One of these screws is for the clamp; the other two permit positioning of the clamp.)
6. Pushing on the CRT base, slide the CRT forward. Pull the CRT cut of the instrument from the front. Be sure that the CRT neck-pins clear the shield edge as the CRT is pushed out.
WARNING High vacuum cathode-ray tubes are dangerous to handle. To prevent personal injury from flying glass in case of tube breakage, wear face mask or safety goggles, and gloves.
Handle the CRT with extreme care. Do not strike or scratch it. Never subject it to more than moderate force or pressure when removing or installing.
Always store spare CRT’s in original protective cartons. Save cartons to dispose of used CRT’s.
7. Remove the boot (shock mounting gasket) and
graticule from the CRT faceplate.
B. REPLACEMENT:
1. Make sure the faceplate and graticule are clean, then place the graticule on the faceplate (with the etched graticule lines against the faceplate). Place the boot around the CRT faceplate and graticule such that the two tabs on the graticule extend through the two slots in the boot, and the graticule is held firmly against the faceplate.
2. Check that the CRT base pins are straight (make a test installation of the CRT base socket onto the base pins outside the instrument), then insert the CRT into the shield. Guide the CRT base into the clamp and slide the CRT toward the rear of the instrument.
3. Tighten the clamp screw inside the CRT shield, leaving the two positioning screws loose. Recommended tightening torque: 4 to 7 inch-pounds.
4. Align the CRT faceplate square with the front of the instrument by positioning the CRT base. Tighten the two positioning screws.
5. Replace the light conductor, filter (if used), bezel and securing nuts.
6. Place the CRT base socket onto the CRT base pins.
7. Replace the deflection-plate pin connectors. Correct location is indicated on the CRT shield.
8. Replace the fan assembly.
9. Replacing the CRT will require instrument recalibration. Refer to Calibration, Section 5.
Scale Illumination Lamp Replacement. The procedure is
as follows:
1. Remove the CRT bezel (held in place with four knurled nuts) and the plastic graticule light conductor. Also remove any filters that may be in front of the graticule.
2. Remove the Storage board to gain access to the lamp sockets.
3. Loosen the nut holding the adjustable lamp-socket bracket and slide the bracket forward through the front panel as far at possible.
4. Remove the defective lamp by pulling straight out with the fingers.
5. Insert the new lamp. Be sure it is pushed all the way into the socket.
6. Adjust the lamp-socket bracket for proper protrusion through the front panel. It must fit into the cavity in the plastic light conductor.
7. Replace the Storage board, light conductor, filters (if any) and bezel.
Transistor Replacement. Transistors should not be replaced unless they are actually defective If removed from their sockets during routine maintenance, return them to their original sockets. Unnecessary replacement or switching of components may affect the calibration of the instrument When a transistor is replaced, check the operation of that part of the instrument which may be affected.
Any replacement component should be of the original type or a direct replacement. Remount the components in the same manner as the original. Fig. 4-3 shows the lead configurations of the transistors used in this instrument.
Transistors which are mounted on the heat sin., on the rear panel use a special thermal-joint compound to increase heat transfer. Replace the compound when replacing these transistors.
WARNING If silicone grease is used as a thermal-joint compound, handle the silicone grease with care. Avoid getting silicone grease in the eyes. Wash hands thoroughly after use.
Fuse Replacement. Table 4-5 gives the rating, location and
function of the fuses used in this instrument.

TABLE 4-5

Circuit
Number Rating Location Function
F1 3.2A Line Voltage Selector 115-volt line
Slo-blow assembly F2 2 A Line Voltage Selector 230-volt line F9 0.15 A Low Voltage Power High Voltage
Fast-blow Supply circuit board
(A)
4-10A
MaintenanceType R561B
Fig. 4-7. Lower High Voltage circuit board.
(A) 4-11A
MaintenanceType R561B
Rotary Switches. Individual wafers or mechanical parts
of rotary switches are normally not replaceable. If a switch is defective, replace the entire assembly. Replacement switches can be ordered either wired or unwired; refer to the Parts List for the applicable part numbers.
When replacing a switch, tag the switch terminals and leads with corresponding identification tags as a guide for installing the new switch. An alternate method is to draw a sketch of the switch layout and record the wire color at each terminal. When soldering to the new switch, be careful that the solder does not flow beyond the rivets on the switch terminal. Spring tension of the switch contact can be destroyed by excessive solder.
Power Transformer Replacement. The power transformer in this instrument is warranted for the life of the instrument. If the power transformer becomes defective, contact your local Tektronix Field Office or representative for a warranty replacement (see the Warranty note in the front of this manual). Be sure to replace only with a direct replacement Tektronix transformer.
When removing the transformer, tag the leads with the corresponding terminal numbers to aid in connecting the new transformer. After the transformer is replaced, check the performance of the complete instrument using the Performance Check instructions in Section 5, Calibration.
High-Voltage Compartment. The components located in the high-voltage compartment can be reached for maintenance or replacement by using the following procedure:
1. Remove the metal high-voltage shield by removing the
two hexagonal nuts on the opposite side. of the chassis.
2. Remove the plastic cover (held in place with three
screws).
3. To remove the complete wiring assembly from the high­voltage compartment, first unsolder the two diodes connected between the board assembly and the high-voltage transformer and then lift the board assembly out far enough to unsolder the leads connecting to the side of the assembly closest to the transformer. Unsolder other leads as necessary to allow the board assembly to be lifted free of the instrument.
4. To remove the high-voltage transformer, unsolder the leads connecting to the pins on the top of the transformer.
5. To replace the high-voltage compartment, reverse the order of removal.
NOTE All solder joints in the high-voltage compartment should have smooth surfaces. Any protrusions may cause high-voltage arcing at high altitudes.
Recalibration After Repair
After any electrical component has been replaced, the calibration of that particular circuit should be checked, as well as the calibration of other closely related circuits. Since the low­voltage supply affects all circuits, calibration of the entire instrument should be checked if work has been done in the low­voltage power supply or if the power transformer has been replaced. The Performance Check instructions outlined in Section 5 provide a quick and convenient means of checking the instrument operation.
4-12A
(A)

SECTION 5

PERFORMANCE CHECK
RM561A
Introduction
This section of the manual provides a means of rapidly checking the performance of the-Type RM561A. It is intended to check the calibration of the instrument without the need for performing the complete Calibration Procedure. The Performance Check does not provide for the adjustment of any internal controls. Failure to meet the requirements given in this procedure indicates the need for internal checks or adjustments, and the user should refer to the Calibration Procedure in this manual.
Recommended Equipment
The following equipment is recommended for a complete performance check. Specifications given are the minimum necessary to perform this procedure. All equipment is assumed to be calibrated and operating within the original specifications. If equipment is substituted, it must meet or exceed the specifications of the recommended equipment.
For the most accurate and convenient performance check, special calibration fixtures are used in this procedure. These calibration fixtures are available the NICP.
1. Tektronix 2- or 3-series amplifier plug-in unit.
2. Tektronix 2- or 3-series time-base plug-in unit.
3. Variable auto-transformer. Must be capable of supplying at least 350 volt-amperes over a voltage range of 105 to 125 volts (210 to 250 volts for 234-volt nominal line). For example, General Radio W10MT3 Metered Variac Auto­transformer.
4. Test oscilloscope. Frequency response, 10MHz; minimum deflection factor, 0.005 volts/division. Tektronix Type 561A/RM with Type 3A1 and 2A63 Amplifier Plug-ins and Type 3B4 Time-Base Plug-In recommended.
5. CRT deflection capacitance normalizer. Tektronix Calibration Fixture, 067-0500-00, recommended.
6. Standard amplitude calibrator. Amplitude accuracy,
0.25%; signal amplitude, 50 millivolts to 100 volts; output signals, 1-kHz square-wave. Tektronix Calibration Fixture, 067-0502-00, recommended.
7. Time-mark generator. Marker outputs, 1 µs and 100 µs; accuracy 0.001%. Tektronix Type 184 Time-Mark Generator recommended.
8. Termination. Impedance, 50-ohm; accuracy, -±3%; connectors, BNC. Tektronix Part No. 011-0049-00.
9. Cable (two). Impedance, 50-ohm; connectors, BNC Tektronix Part No. 012-0057-00.
10. Patch cord. Length, 18-inch; connectors, BNC-to­banana plug. Tektronix Part No. 012-0091-00.
1
Use maximum performance plug-in units. .For example, use a high-frequency unit In performance to low-frequency, dual­trace rather than single-trace, etc.
11. Patch cord. Length, 18-inch; connectors, BNC-to-
BNC Tektronix Part No. 012-0087-00.
PERFORMANCE CHECK PROCEDURE
General
In the following procedure, test equipment connection or control settings should not be changed except as noted. If only a partial check is desired, refer to the preceding step(s) for setup information.
The following procedure uses the equipment listed under Recommended Equipment. If substitute equipment is used, control settings or setup must be altered to meet the requirements of the equipment used.
Preliminary Procedure
1. Insert the amplifier plug-in in the left plug-in
compartment.
2. Insert the time-base plug-in unit in the right plug-in
compartment.
3. Connect the auto-transformer to a suitable power
source.
4. Connect the Type RM561A to the auto-transformer
output.
5. Set the auto-transformer to 115 (or 230) volts.
6. Set the Type RM561A POWER switch to ON. Allow at least 20 minutes warm up for checking the instrument to the given accuracy. Set the following controls:
Type RM561A
FOCUS Any position INTENSITY Midrange CALIBRATOR OFF SCALE ILLUM Clockwise CRT CATHODE Normal SELECTOR
Amplifier Unit
Position Midrange Ac Dc Gnd Ac Volts/Division 1 Variable Calibrated
Time-Base Unit
Position Midrange Magnifier Off Time/Division 1 millisecond Variable Calibrated Triggering controls Adjusted for free running trace
(A)(B)
5-1
Performance CheckType RM561A
1. Check CRT for Double Peaking
a. Requirement—There should be a constant increase in trace intensity as each INTENSITY control is rotated clockwise. All signs of double peaking must be gone within 3 minutes.
b. Turn the INTENSITY control counterclockwise and then slowly clockwise.
c. Check—For a constant increasing brightness of the trace. If the trace brightens, then decreases intensity and then constantly brightens again it is double peaking.
d. Turn the INTENSITY control counterclockwise from its fully clockwise position.
e. Check—For a constantly decreasing brightness. If the trace brightness decreases, then increases and then decreases once more, this again is double peaking.
f. Readjust the INTENSITY control for a usable trace brightness.
2. Checking Calibrator
a. Requirement—The signal amplitude accuracy must be within 3% of !he indicated peak-to-peak amplitude except at the .1V INTO 50 where it is ±3.5% when terminated with 50
ohms. The symmetry must be within ±20% of being a 50% duty cycle and the rise-time must not exceed 6 µs.
b. Set the test oscilloscope Volts/Div switch to 10.
c. Connect a 50-ohm coaxial cable from the test oscilloscope vertical input connector to the Output of the standard amplitude calibrator.
d. Connect a second 50-ohm coaxial cable from the Unknown Input of the standard amplitude calibrator to the CAL OUT connector of the Type RMS61A.
e. Set the Mode switches of the standard amplitude calibrator to Square-Wave and Mixed.
f. Set the CALIBRATOR to 100 VOLTS.
Fig. 5-1. Measuring rise time.
g. Using Table 5-1, switch the Amplitude control of the standard amplitude calibrator and the CALIBRATOR control of the Type RM561A simultaneously to each setting listed in the table.
h. Check—Top of the mixed square-wave presentation noting the amount of trace separation. The trace separation noted must not exceed the amount listed in Table 5-1.

TABLE 5-1

Calibrator
Switch
Settings
100 Volts 10 3 mm
50 Volts 5 3 mm 20 Volts 2 3 mm 10 Volts 1 3 mm
5 Volts .5 3 mm 2 Volts .2 3 mm
1 Volt .1 3 mm .5 Volt .05 3 mm .2 Volt .02 3 mm .1 Volt .01 3 mm
50 mVolts .005 3 mm
i. Set the standard amplitude calibrator Amplitude control to
.1 Volt and the Type RM561A CALIBRATOR control to .1 V INTO 50 .
j. Disconnect the 50-ohm coaxial cable from the Type
RM561A CAL OUT connector.
k. Connect a 50-ohm termination to the CAL OUT connector,
then connect the 50-ohm coaxial cable that was removed in part j to the 50-ohm termination.
I. Set the test oscilloscope Volts/Div switch to .01. m. Check—For not more than 3.5 mm of trace separation. n. Remove the coaxial cables and the 50-ohm termination
connections.
o. Connect a 50-ohm coaxial cable from the Type
RM561A CALIBRATOR to the test oscilloscope vertical input connector.
p. Set the CALIBRATOR for a two-volt output. q. Adjust the test oscilloscope Volts/Div and Position controls
to obtain a 4-cm high display centered about the center horizontal graticule line.
r. Set the Time/Div switch to 1 ms and the Triggering Level
control for a stable display.
s. Adjust the test oscilloscope Horizontal Position control and
Time/Div Variable control until one complete cycle of the square­wave presentation occupies the center 8 cm of the graticule.
t. Check—The width, in centimeters, of the positive-going
portion of the square-wave presentation. The width should not be less than 3.2 cm nor more than 4.8 cm for a square-wave whose duty cycle is 50% ±20%.
Test Oscilloscope
Volts/Div switch
Settings
Maximum
Trace Separation
5-2
(A)(B)
Performance Check Type RM561A
u. Return the Time/Div Variable control to its Calibrated
position and set the Time/Div switch to 1 µs.
v. Position the rising portion of the display to the center
vertical graticule line with the Horizontal Position control.
w. Check—The rise-time (see Fig. 5-1). The rise-time
should not exceed 6 µs.
x. Remove the coaxial cable and set the Type RM561A
CALIBRATOR to OFF.
3. Graticule Scale Illumination
a. Requirement—There should not be any flickering of the graticule scale illumination as the SCALE ILLUM control is rotated throughout its range.. When the SCALE ILLUM control is fully counterclockwise there should be no illumination, and with the control fully clockwise there should be maximum illumination.
b. Set the SCALE ILLUM control fully counterclockwise.
c. Check—For no graticule illumination.
d. Check—That as the control is slowly rotated clock-wise the graticule illumination does not flicker.
e. Check—For maximum graticule illumination when the control has been rotated to its fully clockwise position.
f. Set the SCALE ILLUM control for a usable graticule illumination level.
4. Trace Alignment
The trace should be parallel to the horizontal graticule lines, but due to the different effects that the earth's magnetic field will have at various locations, it is impossible to state that the trace should be aligned within plus or minus a specific tolerance.
If the trace is not parallel to the horizontal graticule lines it will be necessary to adjust the ALIGNMENT control until the trace is parallel to the horizontal graticule lines.
6. High-Voltage Regulation
a. Requirement-High-voltage supply should regulate (no spot blooming) between the high and low line voltages that the Type RM561A has been designed to operate from.
b. Set the auto-transformer for the low-line voltage 105 volts for 115 volts line voltage or 210 volts for 230 volts line voltage.
c. Rotate the FOCUS control fully clockwise.
d. Position the spot onto the graticule area with the positioning controls.
e. Rotate the INTENSITY control from its normal position to its fully clockwise position.
f. Check-For blooming (see Fig. 5-1).
g. Return the INTENSITY control to normal position.
h. Set the auto-transformer for the high-line voltage 125 volts for 115 volts line voltage or 250 volts for 230 volts line voltage.
i. Rotate the INTENSITY control fully clockwise from its normal position.
j. Check—For blooming (see Fig. 5-2).
5. Horizontal Position Control
a. Requirement--Clockwise rotation of the control should move the trace smoothly to the right, while counter-clockwise rotation of the control should move the trace smoothly to the left.
b. Rotate the horizontal position control clockwise.
c. Check—That the trace moved smoothly to the right.
d. Rotate the horizontal position control counter­clockwise.
e. Check—That the trace moved smoothly to the left. f. Return the horizontal position control to its midrange position.
(A)(B)
Fig. 5-2. Checking CRT display for spot blooming (high-voltage
not in regulation).
k. Return the INTENSITY control to its normal position and the auto-transformer to the design center line voltage, 115 volts or 230 volts.
I. Readjust the FOCUS control for the best focus.
7. Vertical Geometry
a. Requirement-The vertical display lines must not deviate more than 2 mm from being parallel to the vertical graticule lines.
b. Set the Time/Div switch of the time base plug-in to 1 ms.
5-3
Performance CheckType RM561A
c. Connect a 50-ohm coaxial cable from the Type 184
Marker Out connector to the amplifier plug-in input connector.
d. Apply 1 µs time markers from the Type 184 and adjust the display amplitude so the time markers extend beyond the graticule limits.
e. Adjust the vertical positioning control to move the bottom of the time marker display out of the display area.
f. Adjust the FOCUS control for the best focus.
g. Check—That the vertical time marker lines are within 2 mm of being parallel to the vertical graticule line. The time markers are 1 mm apart.
h. Remove the coaxial cable.
8. Horizontal Geometry
a. Requirement-The horizontal display lines must not deviate more than 2 mm from being parallel to the horizontal graticule lines.
b. Set the Time/Div switch of the time base plug-in to 0.1 ms and cause the trace to free run.
c. Set the amplifier plug-in Volts/Div switch to .5.
d. Connect 100 mVOLTS of CALIBRATOR signal through 50-ohm coaxial cable to the input connector of the amplifier plug-in.
e. With the vertical position control, place the display of two parallel lines to the horizontal center graticule line.
f. With the vertical position control place the bottom line of the two-line display upon the top graticule line.
g. Check—For not more than 2 mm of trace deviation in a horizontal distance of 10 cm. The lines of the two-line display are 2 mm apart.
h. With the position control place the top line of the two­line display upon the bottom graticule line.
i. Check—For not more than 2 mm of trace deviation in a horizontal distance of 10 cm. The lines of the two-line display are 2 mm apart.
i. Disconnect the coaxial cable.
9. Focus and Astigmatism
a. Requirement—That individual vertical time markers 1 mm apart can be distinguished within the graticule area when the FOCUS control has been properly adjusted. The FOCUS control when properly adjusted must not be at either end of adjustment range.
b. Set the Time/Div switch time base of the plug-in to 1 ms.
c. Apply 100-µs time markers from the Type 184 through a 50-ohm coaxial cable to the input connector of plug-in amplifier.
d. Adjust the display amplitude so the time markers go
beyond the graticule area.
e. Adjust the vertical position control so the bottom of the
time markers is positioned out of the viewing area.
f. Adjust the triggering controls of the time base plug-in for a
stable display.
g. Adjust the FOCUS control for the best focus throughout
the graticule area.
h. Check—That the individual vertical time markers 1 mm
apart can be distinguished anywhere within the graticule area.
i. Check—That the FOCUS control when adjusted is not at
either end of its range.
j. Remove the coaxial cable.
10.Vertical and Horizontal CRT Deflection-Plate Compensation
a. Requirement-Optimum square corner must be obtained
with the CRT Deflection Capacitance Normalizer for both the vertical and the horizontal plug-in compartments.
b. Remove the amplifier plug-in from the left plug-in
compartment.
c. Insert the CRT Deflection Capacitance Normalizer into the
left plug-in compartment.
d. Connect 100 VOLTS of CALIBRATOR signal through a
50-ohm coaxial cable to the Capacitance Normalizer input connector.
e. Adjust the time-base plug-in triggering controls so the
leading edge of the waveform is displayed.
NOTE The INTENSITY control may need to be advanced slightly to view the leading edge of the waveform.
f. Check-For optimum square corner (see Fig. 5-3A).
g. Interchange the Capacitance Normalizer and the time­base plug-in. Readjust the FOCUS control and the time-base plug-in triggering controls for a stable well focused display.
h. Check—For optimum square corner (see Fig. 5-4).
i. Disconnect the coaxial cable.
j. Remove the Capacitance Normalizer. Replace the amplifier plug-in in the left compartment and the time-base plug-in in the right compartment.
(A)(B)
5-4
Performance CheckType RM561A
Fig. 5-4. Typical CRT display showing correct adjustment of
horizontal compensation.
11. Alternate Trace and Chopped Blanking
a. Requirement—In the alternate mode two traces must show for the vertical plug-in at all sweep rates (TIME/CM switch settings).
In chopped mode the chopping transients (rising and falling portions of the waveform) must be blanked out.
NOTE This check can only be made with an amplifier plug-in which has both alternate trace and chopped blanking provisions.
Fig. 4. (A) Typical CRT display showing correct vertical
compensation adjustment, (B1 and (C) Incorrect adjustment.
b. Set the amplifier plug-in Mode switch to Alter. c. Set the amplifier plug-in Ac-Dc Gnd switches to Gnd. d. Position the traces about three divisions apart. e. Rotate the Time/Div switch of the time base plug-in
throughout its range.
f. Check—For an alternating two-trace display at all sweep
rates.
g. Set the amplifier plug-in Mode switch to Chop. h. Position the two traces about three divisions apart.
i. Set the time-base plug-in Time/Div switch to display several cycles of the waveform and adjust the triggering controls for a stable display.
j. With the CRT CATHODE SELECTOR switch in the No rmal position, notice the overshoot on the display (see Fig. 5-5A).
k. Set the CRT CATHODE SELECTOR switch to CHOPPED BLANKING.
I. Check—That the between-channel switching transients are removed from the display (see Fig. 5-5B).
m. Set the amplifier plug-in Mode switch to display only one channel.
n. Set the CRT CATHODE SELECTOR switch in the Normal position.
(A)(B)
5-5
Performance CheckType RM561A
Fig. 5-5. Alternate waveform showing IA) unblanked chopping transients and 1i5 blanked transients.
12. CRT Intensifier Circuit
a. Requirement—A portion of the trace must appear brighter (intensified) than the rest of the trace. The intensified portion of the trace should move along the trace as the delay time control is rotated.
NOTE This check can only be made with a time­base plug-in which has delayed sweep provisions.
b. Set the time-base plug-in Time/Div switch to 1 ms and the time-base plug-in delayed sweep Time/Div switch to 0.1 ms.
c. Adjust -the time-base plug-in triggering controls to produce a trace.
d. Adjust the FOCUS control and the tine-base plug-in triggering controls for a well-focused tree-running (or automatic triggered) trace.
e. Turn the time-base plug-in Mode switch to Intensified (not triggered).
f. Check—That an intensified portion appears on the trace (see Fig. 5-6). If the display appears only as a shortened trace, increase the INTENSITY control slightly until the complete trace, showing an intensified portion, appears.
g. Check—That as the delay time control is rotated throughout its range the intensified portion moves along the trace.
h. Set the time-base plug-in Mode switch for a normal sweep and the triggering controls for a free running trace.
13. Intensity (Z-Axis) Modulation
a. Requirement—The trace can be modulated when 10 volts or more of signal is applied to the EXT CRT CATHODE connector.
b. Remove the ground strap between the rear-panel banana jacks.
c. Set the CALIBRATOR for a 10-volt output signal.
Fig. 5-6. Typical CRT display showing correct Intensifier
circuit operation.
Fig. 5-7. Typical CRT display showing correct intensity
modulation (time/division, 10 milliseconds).
(A)(B)
5-6
Performance CheckType RM561A
d. Connect the CALIBRATOR signal to the EXT CRT CATHODE banana jack with the BNC-to-BNC and BNC-to­banana jack jumper leads.
e. Check—The display for intensified dots (see Fig. 5-7). It may be necessary to reduce the INTENSITY setting slightly to see the dots.
(A)(B)
f. Remove the jumper leads and replace the ground strap.
This completes the Performance Check of the Type RM561A. Disconnect all test equipment. The plug-in amplifier and time­base plug-in are checked separately.
5-7

SECTION 5A

PERFORMANCE CHECK/MAINTENANCE CALIBRATION
R561B
Type R561B
1
Introduction
To assure instrument accuracy, check the calibration of the Type R561B every 500 hours of operation, or every six months if used infrequently. Before complete calibration, thoroughly clean and inspect this instrument as outlined in the Maintenance section.
This section provides several features to aid in checking or calibrating this instrument. For example:
Index. The Short-Form Procedure lists the step numbers and titles of the Complete Calibration Procedure and gives the page on which each step begins. Therefore, the Short-Form Procedure can be used to locate a step in the complete procedure.
Calibration Record. The Short-Form Procedure can be reproduced and used as a permanent record of instrument calibration. Spaces are provided to check off each step as it is completed and to record performance data.
Abridged Calibration Procedure. The Short-Form Procedure lists the adjustments necessary for each step and the applicable tolerance for correct calibration. The experienced technician who is familiar with the calibration of this instrument can use this procedure to facilitate its checking or calibration.
Performance Check. The Complete Calibration Procedure con be used as a front-panel check of the instrument’s performance by doing all portions except the ADJUST part of a step. When used as a performance check procedure, the instrument is checked to the original performance standards without removing the covers or making internal adjustments. Screwdriver adjustments which are accessible without removing the covers can be adjusted. Some steps are not applicable to a performance checkout procedure. These steps have a note which gives the next applicable step.
Complete Calibration. Completion of each step in the Complete Calibration Procedure checks this instrument to the original performance standards and gives the procedure to return each adjustment to its optimum setting. Limits, tolerances and waveforms in this procedure are given as calibration guides and are not instrument specifications. Where possible, instrument performance is checked before an adjustment is made. For best overall instrument performance make each adjustment to the exact setting even if the CHECK­is within the allowable tolerance.
Partial Calibration. To check or adjust only part of this instrument, start with the nearest "equipment required" picture preceding the desired portion. To prevent recalibration of other parts of the instrument
1
For calibration procedure, see TB 750-236
when performing a partial calibration, re-adjust only if the tolerance given in the CHECK- part of the step is not met. If an adjustment is made, any steps listed in the INTERACTION- part of the step should also be checked for correct tolerance.
TEST EQUIPMENT REQUIRED
General
The following test equipment and accessories, or its equivalent, is required for complete calibration of the Type R561B. Specifications given are the minimum necessary for accurate calibration. Therefore, some of the recommended equipment may have specifications which exceed those given. All test equipment is assumed to be correctly calibrated and operating within the given specifications. If equipment is substituted, it must meet or exceed the specifications of the recommended equipment.
For the quickest and most accurate calibration, special Tektronix calibration fixtures are used where necessary. These special calibration fixtures are available from the NICP.
1. Vertical amplifier. Dual trace; bandwidth, DC to at least 10 MHz; deflection factor, 0.01 volts/division to 5 volts/division; chopped and alternate modes. Tektronix Type 3A6 Dual-Trace Amplifier recommended.
2. Time-base unit. Normal and delayed sweeps; sweep rates, one millisecond/division to one microsecond/division; 5X magnifier; single-sweep operation. Tektronix Type 383 Time-Base Unit recommended.
3. CRT deflection capacitance normalizer. (Normally required only if new cathode-ray tube has been installed or if deflection-plate compensation has been inadvertently maladjusted. Alternate method of adjustment is also given). Tektronix Calibration Fixture, 067-0500-00, recommended. (For alternate method, Tektronix Type 130 L-C Meter recommended.)
4. Variable auto-transformer. Must be capable of supplying at least 186 watts over a voltage range of 90 to 136 volts (180 to 272 volts for 230-volt nominal line). If auto-transformer does not have an AC (RMS) voltmeter to indicate output voltage, monitor output with an AC (RMS) voltmeter. For example, General Radio W10MT3W Metered Variac Auto-transformer.
5. DC volt-ohmmeter. Minimum sensitivity, 20,000 ohms/volt. For example, Triplett 630.
6. Precision DC voltmeter. Accuracy, within ±0.05%; meter resolution, 50 microvolts; range, zero to 3.5 kilovolts. For example, Fluke Model 825A Differential DC Voltmeter (with Fluke Model 80E-5 Voltage Divider to measure the high-voltage supply).
(A)
5-1A
Performance Check/CalibrationType R561B
7. Test oscilloscope, with 1 X probe. (Optional, for checking power supply ripple.) Bandwidth, DC to 300 kilohertz; minimum deflection factor, five millivolts/division; accuracy, within 3%. Tektronix Type 561B with 2A63 and 2B67 plug-in units, and P6028 Probe recommended.
8. Square-wave generator. Frequency, 100 kilohertz; output amplitude, three volts to ten volts. Tektronix Type 106 Square-Wave Generator recommended.
9. Time-mark generator. Marker output, one millisecond; marker accuracy, within 0.1 %. Tektronix Type 184 Time-Mark Generator recommended.
10. Cable. Impedance, 50 ohms, electrical length, five nanoseconds; connectors, GR874. Tektronix Part No. 017­0502-00.
11. Cable, coaxial. Impedance, 50 ohms; length, 42 inches; connectors, BNC. Tektronix Part No. 012-0057-00.
12. Adapter, GR to BNC male. Tektronix Part No. 017­0064-00.
13. Adapter, BNC to banana terminal. For example, Pomona # 1269 (Tektronix Part No. 103-0090-00).
14. T connector, BNC. Tektronix Part No. 103-0030-00.
15. Termination. Impedance, 50 ohms; accuracy, ±3%; connectors, BNC. Tektronix Part No. 011-0049-00.
16. Current-measuring probe with passive termination. Sensitivity, two milliamperes/millivolt; accuracy, within ±3%. Tektronix P6019 Current Probe with 011-0078-00 passive termination recommended.
SHORT-FORM PROCEDURE
Type R561B, Serial No. Calibration Date Calibrated By
1. Adjust -100-Volt Power Supply Page 5-5
-100 volts, ±0.5 volts
2. Check Low-Voltage Power Supplies Page 5-5 3. Check Low-Voltage Power Supply Page 5-5
Regulation and Ripple (Optional Check)
4. Adjust High Voltage Page 5-6
-3300 volts, ±99 volts
5. Check High Voltage Regulation Page 5-6 6. Adjust CRT Grid Bias Page 5-6
Coarse intensity adjustment.
7. Check Alternate Trace Page 5-7
Trace alternates at all sweep rates.
8. Check Dual-Trace Blanking Page 5-8
Switching transients (vertical lines) blanked between chopped segments.
9. Check External CRT Cathode Page 5-9
Intensity (Z-Axis) modulation with 3 volts input.
10. Adjust Trace Alignment Page 5-9
Trace parallel to horizontal graticule lines.
11. Adjust CRT Geometry Page 5-10
Best overall geometry
12. Check CRT Vertical Deflection Factor Page 5-10
18.5 to 20.5V/cm
13. Check CRT Vertical Electrical Center Page 5-10
Trace within 0.5 major division of graticule horizontal centerline.
14. Check CRT Horizontal Deflection Page 5-10
Factor 17.5 to 19.25V/cm
15. Check CRT Horizontal Electrical Page 5-11
Center Trace within 0.8 major division of graticule vertical centerline.
16. Check Delaying Sweep Intensification Page 5-11
Intensified portion on trace (with delayed sweep time­base unit).
17A. Adjust Vertical and Horizontal Page 5-12
Deflection-Plate Compensation Optimum square corner (or 14.3 picofarads effective capacitance).
17B. Alternate Method of Adjusting Page 5-14
Deflection-Plate Compensation Optimum square corner (or 14.3 picofarads effective capacitance).
18. Check and Adjust Calibrator Amplitude Page 5-15
+40 volts, ±0.6 volt.
19. Check and Adjust Calibrator Repetition Page 5-16
Rate One kilohertz, 10 hertz.
20. Check Calibrator Duty Factor Page 5-16
48% to 52%
21. Check Calibrator Rise-time Page 5-17
<2.5 microseconds at 40 volts; < other voltages
22. Check Current Through Probe Loop Page 5-18
Ten milliamperes
1 microsecond at all
5-2A
(A)
Performance Check/CalibrationType R561B
COMPLETE CALIBRATION PROCEDURE
General
The following procedure allows the Type R561B to be calibrated with the least interaction of adjustments and reconnection of equipment. An equipment required picture is shown for each group of checks and adjustments to identify the test equipment used. Following this picture is a complete list of front-panel control settings for the Type R561B. Controls which have been changed for the new group of checks and adjustments are printed in bold type. Each step following the test equipment picture continues from the equipment setup and control settings used in the preceding step(s) unless noted otherwise. External controls or adjustments of the Type R561B referred to in this procedure are capitalized (e.g., INTENSITY). Internal adjustment names are initial capitalized only (e.g., High Voltage).
All waveforms shown in this procedure were token with a Tektronix Oscilloscope Camera System and Projected Graticule. The following procedure uses the equipment listed under Test Equipment Required. If equipment is substituted, control settings or equipment setup may need to be altered to meet the requirements of the equipment used. Detailed operating instructions for the test equipment is not given in this procedure. If in doubt as to the correct operation of any of the test equipment, refer to the instruction manual for that unit.
NOTE
This instrument should be calibrated at an ambient temperature of +25C, ±5C. Its performance
can be checked at any temperature within
the 0°C to 50°C range. If the ambient temperature is outside the given range, see Section 1 for the applicable tolerances.
Preliminary Procedure for Performance Check Only
1. Connect the Type R561B to a power source which meets
the voltage and frequency requirements of this instrument.
2. Set the POWER switch to ON. Allow at least 5 minutes
warm-up before proceeding.
3. Begin the Performance Check with step 7.
Preliminary Procedure for Complete Calibration
1. Remove the top and bottom covers from the Type R561B.
2. Set the Line Selector to 115 V and the Range Selector to
Medium.
3. Connect the auto-transformer to a suitable power source.
4. Connect the Type R561B to the auto-transformer output.
5. Set the auto-transformer output voltage to 115 volts.
6. Set the POWER switch to ON. Allow at least 5 minutes
warm-up before proceeding.
(A)
5-3A
Performance Check/CalibrationType R561B
Fig. 5-1. Test equipment required for steps 1 through 6.
POWER SUPPLIES
Control Settings
Type R561B
INTENSITY Counter-Clockwise FOCUS Midrange ASTIGMATISM Midrange SCALE ILLUM As desired CALIBRATOR OFF CRT CATHODE SELECTOR NORM
(rear panel)
Vertical Amplifier Unit
Ch 1 Input Coupling AC Ch 1 Volts/Div 5 Mode Ch 1 Position Centered
Time-Base Unit
Mode Norm Time/Div 1 ms
Normal-Single Sweep Single Sweep Slope + Coupling Auto Source Int Level Clockwise
Test Oscilloscope (Optional — To check ripple)
Intensity Nominal brightness Focus and Astigmatism Well defined trace
Time Base
Mode Normal Time/Div 1 ms Normal-Single Sweep Norm Level 0 Slope + Coupling Auto Source Int
Vertical Amplifier
Mode Ch 1 Volts/Div 0.01 Input Coupling AC
(A) 1
5-4A
1. Adjust100-Volt Power Supply
For performance check only, proceed to step 7.
a. Test equipment required for steps 1 through 6 is shown in Fig. 5-1. The illustrated equipment is keyed to that listed under Test Equipment Required.
b. Connect the precision DC voltmeter between the ­100-volt test point and ground (see Fig. 5-2).
c. ADJUST—R23, -100 Volts (Fig. 5-2) for exactly – 100 volts.
d. INTERACTION—Operation of all circuits within the Type R561B is affected by the 100-volt supply.
Performance Check/CalibrationType R561B
Fig. 5-2. Locations of power supply test points and R23, -100
Volts adjustment.
2. Check Low-Voltage Power Supplies
a. Connect the precision DC voltmeter between each low-voltage test point and chassis ground. See Fig. 5-2 for test point locations.
b. CHECK—Each supply is within the tolerance listed in Table 5-1.
TABLE 5-1
Maximum Line
Supply Tolerance +300 V +295.5 to +304.5 V 5 mV +125 V +123.1 to 126.9 V 3 mV
-12.2 V -12.05 V to -12.35 V 2 mV
-100 V. -99.5 V to -100.5 V 2 mV
(A) 1
Frequency Ripple
Fig. 5-3. Location of (A) -3300-volt test point, and (B) R206,
High Voltage adjustment
3. Check Low-Voltage Power Supply Regulation and Ripple (Optional Check)
a. To check regulation, connect the precision DC volt-
meter between each low voltage supply test point and chassis ground. To check ripple, connect the 1X probe from the test oscilloscope Ch 1 input connector to each test point.
b. Set the auto-transformer output to 104 VAC c. CHECK—Each supply output and ripple amplitude must
be within the tolerance listed in Table 5-1.
NOTE Power supply voltage and ripple tolerances in this step are guides to correct instrument operation; not instrument performance requirements. Actual values may exceed listed tolerances with no loss in measurement accuracy if the instrument meets the performance requirements in Section 1 as tested in this procedure.
d. Set the auto-transformer output to 126 VAC.
.
5-5A
Performance Check/CalibrationType R561B
Fig. 5-4. Location of R269, CRT Grid Bias.
e. CHECK—Each supply output and ripple amplitude
must be within the tolerance listed in Table 5-1.
f. Return the auto-transformer output to 115 VAC and
disconnect the precision DC voltmeter and test oscilloscope.
4. Adjust High Voltage
a. Connect the DC voltmeter between ground and the -
3300-volt test point (Fig. 5-3A).
b. CHECK—Meter reading must be -3300 volts, ±99 V. c. ADJUST—R206, High Voltage (Fig 5-3B), for a
meter reading of exactly -3300 volt,.
5. Check High Voltage Regulation (Optional Check)
a. With the DC voltmeter connected between ground
and
the -3300-volt test point and the High Voltage within the limits stated in step 4b, adjust the auto-transformer for an output of 104 VAC and then 126 VAC to check the regulation of the high-voltage supply.
b. CHECK—Meter reading should not vary more than :-t 15
volts when checking regulation.
c. Remove the DC voltmeter and return the auto-
transformer output to 115 VAC.
6. Adjust CRT Grid Bias
a. Connect the precision DC voltmeter between pin AL of the power supply board (Fig. 5-2; this is connected to center terminal of the INTENSITY potentiometer) and ground.
b. Rotate the INTENSITY control clockwise until the meter reading is -40 volts. The knob index should be aligned with the control mid-range panel mark.
c. ADJUST—R269, CRT Grid Bias (Fig. 5-4) for a trace of medium brightness.
d. Rotate the INTENSITY control counterclockwise and observe that the trace disappears completely.
e. Disconnect the electronic voltmeter.
NOTE The Type R561B may now be connected directly to the power source for the remainder of the procedure, provided the Line Selector and Range Selector switches are set to the proper positions for the source line voltage.
(A) 1
5-6A
Performance Check/Calibration-Type R561B
Fig. 5-5. Test equipment required for steps 2 through 16.
CRT AND DEFLECTION CIRCUITS
Control Settings
Type R561B
INTENSITY Counterclockwise FOCUS. Midrange ASTIGMATISM Midrange SCALE ILLUM As desired CALIBRATOR OFF CRT CATHODE SELECTOR NORM
(rear panel)
Vertical Amplifier Unit
Input Coupling (Ch 1 & 2) AC Volts/Div (Ch 1 & 2) 5 Mode Alter Position (Ch 1 & 2) Centered Trigger Composite (pushed in)
Time-Base Unit
Mode Norm Time/Div 1 ms
(A)
CANNOT READ TEXT
5-7A
Performance Check/Calibration-Type R561B
Fig. 5-6. Typical CRT displays showing correct
blanking. (A) Chopped transients visible with CRT
CATHODE SELECTOR switch set to NORM Time/Div
µ
); B transients blanked with CRT CATHODE
2
SELECTOR switch set to CHOPPED BLANKING
compartment and pin 3 of the interconnecting socket in the horizontal plug-in compartment. Also check for infinite resistance between pins 3 and 4 of the vertical plug-in unit connector. Since the Type R561B interconnects the two plug­in units in the Alternate mode, the continuity check also checks the alternate-trace function.
c. Perform the substitute procedure for single-trace units given in step 8 before re-inserting the plug-in units and applying power to the instrument.
8. Check Dual-Trace Blanking
a. Reset the following controls: Mode (amplifier) Chop Time/Div (time-base) 2µs
b. Position the two traces about 2 major divisions apart
on the CRT screen and trigger the chopped waveform (see Fig. 5-6A) using the Triggering Level control.
.
Fig. 5-7. Typical CRT displays when checking intensity
modulation; (A) 10-volt signal applied to both vertical
amplifier and CRT cathode. (B) 3-volt signal applied only
to CRT cathode
c. Adjust the INTENSITY control so the vertical segments of the chopped waveform are barely visible. Adjust FOCUS and ASTIGMATISM as needed.
d. Move the CRT CATHODE SELECTOR switch to
CHOPPED BLANKING.
e. CHECK-The vertical segments should now be blanked and the horizontal segments should be slightly intensified. (See Fig. 5-6B.)
f. Return the CRT CATHODE SELECTOR switch to
NORM.
If a single-trace amplifier plug-in unit is used:
a. (With power off and amplifier plug-in unit removed), connect the ohmmeter between pin 24 of the interconnecting socket in the vertical plug-in compartment and the wire strap that diagonally connects the inboard terminals of switch SW255 (CRT CATHODE SELECTOR)
b. CHECK-Meter reading is infinity (open circuit) with the CRT CATHODE SELECTOR switch in the NORM and EXT INPUT positions.
.
5-8A
(A)
Performance Check/Calibration-Type R561B
Fig. 5-8. (A) Idealized waveforms showing good geometry with examples of poor geometry; (8) location of R256,
Geometry.
c. Set the CRT CATHODE SELECTOR to CHOPPED
BLANKING.
d. CHECK-Meter reading is zero ohms (closed circuit). Together with the following external-cathode intensity modulation check, this continuity check tests the chopped blanking function of the Type R561B.
e. Disconnect the ohmmeter and set the CRT CATHODE SELECTOR to NORM.
f. Re-insert the plug-in units into the Type R561B.
g. Set the POWER switch to ON and allow five minutes for the instrument to warm up.
9. Check External CRT Cathode
a. Change the following control settings:
Mode (Vertical Amplifier) Ch 1 Time/Div (Time-Base) 5 µs
b. Connect the square-wave generator high amplitude output to the amplifier unit Ch 1 input connector through a 5­nanosecond GR cable and BNC T connector (use a GR to BNC adapter to connect the GR cable to the T connector). Connect a coaxial cable from the T connector to the EXT INPUT connector at the rear of the Type R561B.
c. Set the square-wave generator for a two-division CRT display (10 volts peak to peak) at 100 kilohertz. Adjust the Triggering Level control for a stable display.
d. Decrease the intensity until the waveform is barely
visible.
e. Set the CRT CATHODE SELECTOR to EXT INPUT.
f. CHECK--The top portions of the waveform should be blanked completely and the bottom portions should be intensified (see Fig. 5-7A).
g. Set the CRT CATHODE SELECTOR to NORM and
the amplifier Volts/Div switch to 1.
h. Adjust the square-wove generator amplitude to produce a three-division CRT display (3 volts peak to peak). Use an attenuator if necessary.
i. Move the signal lead from the amplifier Ch 1 input to the time-base Ext Trig input (use a BNC to banana terminal adapter). Do not remove the signal from the EXT INPUT connector.
j. Set the time-base Source switch to Ext and the CRT CATHODE SELECTOR to EXT INPUT. Trigger the display.
k. CHECK-Intensity modulation should be visible with
the 3-volt signal applied (see Fig. 5-7B).
I. Remove the square-wave generator signal.
10. Adjust Trace Alignment
a. Change the following control settings: INTENSITY Nor m a l d i s p l a y b r i g h t n e s s CR T CA T H O D E S E L E C TO R NORM
(A)
5-9A
Fig. 5-9. Location of vertical and horizontal deflection-
plate neck pins
Time/Div (time-base) 1 ms Triggering Level (time base) Clockwise Triggering Source (time-base) Int
b.Position the trace to the horizontal centerline
c.CHECK--Trace should be parallel to the horizontal graticule lines. If necessary, adjust the TRACE ALIGNMENT adjustment front panel to align the trace to the horizontal graticule line.
.
Performance Check/Calibration-Type R561 B
i.Position the trace to the top graticule line. j.CHECK-Deviation from straight line should not
exceed 0.1 division.
For Performance Check only, proceed to step 16.
12. Check CRT Vertical Deflection Factor
This step is not applicable to Performance Check.
a.Connect the DC voltmeter (set to 300 volt scale between the two vertical deflection plate neck pins (BLUE and BROWN leads, see Fig 5-9.) Do not short to the CRT shield.
b.Position the trace to the top graticule line.
c.Note and record the meter reading.
d.Remove the meter leads and position the trace to the bottom graticule line.
e.Reconnect the DC voltmeter between the two vertical deflection plate neck pins, opposite in polarity to the connection in step a.
f.Again note and record the meter reading.
g Determine the voltage swing over the eight major divisions by adding the meter reading noted in step c to the meter reading noted in step f.
h.CHECK--Voltage swing over eight major divisions is between 148 and 164 volts. This indicates a vertical deflection factor of 18 5 to 20 5 volts per division.
13. Check CRT Vertical Electrical Center
This step is not applicable to Performance Check.
11. Adjust CRT Geometry
a.Connect the time-mark generator marker output to
the amplifier Ch 1 input connector with a coaxial cable.
b.Set the time-mark generator for 1 millisecond
markers.
c.Trigger the display, with the Triggering Level control and position the display baseline to a point below the bottom edge of the graticule.
d.Set the Volts/Div switch so that the time markers over-scan the graticule area.
e.ADJUST R256. Geometry (Fig 58B), for minimum bowing of markers at the left and right edges of the graticule.
f.CHECK- Deviation from straight line should not
exceed 0.1 division (see Fig. 5 8A).
g.Remove the time mark signal and position the trace to the bottom graticule line.
h.CHECK-Deviation from straight line should not exceed 0.1 division.
a.With the DC voltmeter connected between the vertical deflection-plate neck pins, position the trace towards graticule center while observing the meter. Continue to position the trace until the meter reading is zero volts. This is the CRT vertical electrical center:
b.CHECK- Trace must be within 0.5 major division of the graticule centerline.
c.Disconnect the DC voltmeter.
14. Check CRT Horizontal Deflection Factor
This step is not applicable to Performance Check.
a.Rotate the INTENSITY control fully counterclockwise.
b.Remove the two plug in units, then insert the time­base plug-in unit into the vertical (left) compartment and the amplifier unit into the horizontal (right) compartment.
c.Allow about 1/2 minute warmup, then increase the intensity to normal brightness and note the trace is now vertical.
d.Connect the DC voltimeter (set to 300-volt scale) between the two horizontal deflection plate neck pins (GREEN anti RED leads, see Fig. 5-9).
(A)
5-10A
e. Position the trace to the jeft edge of the graticule.
f. Note and record the meter reading.
g. Remove the meter leads and position the trace to the
right edge of the graticule.
h. Reconnect the DC voltmeter between the horizontal deflection-plate neck pins, opposite in polarity to the connection in step d.
i. Again note and record the meter reading.
j. Determine the voltage swing over the ten major divisions by adding the meter reading noted in step f to the meter reading noted in step i.
k. CHECK-Voltage swing over ten. major divisions is between 175 and 192.5 volts. This indicates a horizontal deflection factor of 17.5 to 19.25 volts per division.
15. Check CRT Horizontal Electrical Center
This step is not applicable to Performance Check.
a. With the DC voltmeter connected between the horizontal deflection-plate neck pins, position the trace towards graticule center while observing the meter. Continue to position the trace until the meter reading is zero volts. This is the CRT horizontal electrical center.
b. CHECK-Trace must be within 0.8 major division of the graticule vertical centerline.
c. Disconnect the DC voltmeter.
d. Rotate the INTENSITY control fully counterclockwise.
e. Remove the plug-in units, then re-insert the amplifier unit into the vertical (left) compartment and the time-base unit into the horizontal (right) compartment. Allow about 1/2 minute warmup.
Performance Check/Calibration-Type R561B
Fig. 5-10. Typical CRT display showing correct
intensifier circuit operation
16. Check Delaying Sweep Intensification
a. Set the time-base Delayed Sweep Time/Div switch to
0.1 ms and the Delay Time dial to 1.00. Position the display as needed.
b. Switch the time-base Mode switch to Intensified (not
Trig Intensified).
c. CHECK-It should be possible to adjust the INTENSITY control so that only the intensified portion of the delaying (normal) sweep is visible on the CRT screen. See Fig. 5-10.
.
5-11A
For Performance Check only, proceed to step 18.
(A)
Performance Check/Calibration-Type R561B
Fig. 5-11. Test equipment required for stop 17.
DEFLECTION-PLATE COMPENSATION
Control Settings
Type R561B
INTENSITY Counterclockwise FOCUS Well-defined trace ASTIGMATISM Well-defined trace SCALE ILLUM As desired CALIBRATOR OFF CRT CATHODE SELECTOR NORM
(rear panel)
Vertical Amplifier Unit
Input Coupling (Ch 1 & 2) AC
Volts/Div (Ch 1 & 2) 5
Mode Ch 1 Position Centered
Time-Base Unit
Mode Norm Time/Div 0.1 ms
Normal-Single Sweep Norm
Slope ---
Coupling Auto Source Int
Level Clockwise
17A. Adjust Vertical and Horizontal
Deflection-Plate Compensation
Omit this step unless the CRT has been replaced.
a.Test equipment required for step 17 is shown in Fig. 5-11. The illustrated equipment is keyed to that listed under Test equipment Required.
b.Remove the vertical amplifier plug-in unit and insert the CRT Deflection Capacitance Normalizer into the left plug­in compartment.
c.Connect a coaxial cable from the CAL OUT connector
to the Capacitance Normalizer input connector.
d.Set the CALIBRATOR switch to 40 V.
e.Increase the intensity to normal display brightness and adjust the Triggering Level control for a stable display.
(A)
5-12A
Performance Check/Calibration-Type R561 B
Fig. 5-12 (A) Typical CRT display showing correct vertical compensation adjustment; (B) and (C) incorrect adjustment, (D)
location of C109 (left side).
f.Turn the 51/2 Magnifier on and position the leading
edge of the square wave as shown in Fig. 5.12A.
g.ADJUST - C109 (Fig. 5-12) for optimum square
corner.
h.Rotate the INTENSITY control counterclockwise and interchange the Capacitance Normalizer and time-base unit. Readjust the INTENSITY, FOCUS and time-base triggering controls for a stable well-focused display.
i.ADJUST--C102 (Fig. 5-13B) for optimum square
corner. See Fig. 5-13A.
j.Turn the 5X Magnifier off and rotate the INTENSITY
control counter clockwise.
k.Remove the Capacitance Normalizer. Replace the amplifier unit in the left compartment and the time-base unit in the right compartment.
(A)
5-13A
Performance Check/Calibration-Type R561B
Fig. 5-13. (A) Typical CRT display showing correct adjustment of horizontal compensation; (B) location of C102 (right side).
17B. Alternate Method of Adjusting
Deflection-Plate Compensation
NOTE The following method can be used to adjust the deflection-plate compensation if a Capacitance Normalizer is not available.
The effective deflection-plate capacitance of the CRT is the capacitance seen by the plug-in unit when the deflection plates ale driven push-pull. Therefore, it cannot be measured directly with a capacitance meter. However, the individual circuit capacitances which make up the effective deflection plate capacitance can be measured. These individual circuit capacitances are shown schematically in Fig. 5-14. C1 and C2 represent the capacitance from each deflection plate to ground. C3 represents the variable capacitance
Fig. 5-14. Schematic representation of the effective
CRT deflection-plate capacitance
.
between the deflection plates. The variable capacitor is made up of the circuit capacitance, plus the capacitor added for adjustment.
The effective deflection-plate capacitance, C
can be
eff
expressed in terms of C1, C2 and C3 as follows:
CC12
eff
+
=
2
+2(C3)
Setting C
C
equal to 14.3 picofarads (value set at factory),
eff
the value of the variable capacitor can be calculated.
CC12
C3 = 7.15 pF -
+
4
a. Disconnect the power cord and isolate the Type
R551B from ground.
b. Remove both plug-in units.
c. Connect the capacitance meter guard voltage to pin 21 of the vertical (left) plug-in compartment and measure the capacitance between pin 17 and the oscilloscope chassis. This is C1.
d. Connect the capacitance meter guard voltage to pin 17 and measure the capacitance between pin 21 and the oscilloscope chassis. This is C2.
e. Substitute the measured capacitance values into the equation and solve for C3.
f. Connect the guard voltage to the oscilloscope
chassis and measure the capacitance between pins 17 and
21.
g. Adjust C109 until the measured capacitance equals the value calculated for C3 in step e.
h. Repeat steps c through g for the horizontal (right) plug-in compartment. Adjust C102 for the calculated value of C3.
i. Remove the capacitance meter.
j. Replace the plug-in units removed in step b and reapply power to the instrument. Allow about five minutes warmup before continuing.
5-14A
(A)
Performance Check/Calibration-Type R561 B
Fig 5-15. Test equipment required for steps 18 through 22.
1 kHz CALIBRATOR
Control Settings
Type R561B
INTENS!TY Normal brightness FOCUS Well defined trace ASTIGMATISM Well defined trace SCALE ILLUM As desired CALIBRATOR 10 mA DC (40 V DC) CRT CATHODE SELECTOR NORMA
(rear panel)
Vertical Amplifier
Input Coupling (Ch 1 & 2) DC Volts/Div (Ch 1) 2 Volts/Div (Ch 2) .5 Mode Alter Position (both) Centered
Time-Base
Mode Norm Time/Div 1 ms Normal-Single Sweep Norm
Slope +
Coupling Auto Source Int Level Clockwise
18. Check and Adjust Calibrator Amplitude
a. Test equipment required for steps 18 through 22 is shown in Fig. 5-15. The illustrated equipment is keyed to that listed under Test Equipment Required.
b. Connect the differential voltmeter between ground and the CAL OUT connector.
For Performance Check only:
CHECK--- Meter reading is -+40 volts, ±0.6V. Proceed to
step 19.
c. ADJUST--R166, Amplitude (Fig. 5-16) for meter
reading of exactly +40 volts.
d. Remove Q159 (Fig. 5-16) and check the remaining calibrator voltages as listed in Table 5-2. Connect a 50 ohm termination to the CAL OUT connector when checking the 0.2 V, 20 mV and 2 mV positions of the CALIBRATOR switch.
e. Set the CALIBRATOR witch to OFF.
f. Replace Q159 and remove the Differential Voltmeter
5-15A
(A)
Performance Check/Calibration-Type R561B
Fig. 5-16. Location of Calibrator controls
.

TABLE 5-2

Calibrator Meter Reading
40 V 4V
0.4 v 40 mV 40 mV
Into 50 (tolerance within 3% )
0.2 V 20 mV 2 mV
19. Check and Adjust Calibrator Repetition Rate
a.Connect the CAL OUT connector to the Ch 1 input
connector with a coaxial cable.
b.Set the CALIBRATOR to 4 V. c.Connect the time-mark generator marker output
connector to the Ch 2 input connector with o coaxial cable.
d.Set the time-mark generator for one-millisecond
markers.
e.Position the display so that the square-wave calibrator signal is superimposed on the time-mark signal. Adjust the Triggering Level control for a stable display.
f.Adjust the time-base Sweep Cal (front panel) for one time marker each major division, and adjust the amplifier Calib control (front panel) for exactly two major divisions of calibrator signal amplitude.
For Performance Check only:
CHECK--One cycle of calibrator waveform for each marker (see Fig 5-17). The error in 10 major divisions must be 0.5 minor divisions (the positive transition of the square wave at the right hand edge of the graticule must be no more than 0.5 minor divisions (1 mm) from the positive transitions of the corresponding time marker. Disconnect the time -mark
generator and proceed to step 20.
+40V, .±0.6V +4V, ±:0.06 V +0.4 V, ± 6 mV +40 mV, ±O.6 mV
-+4 mV, ±0.06 mV
+0.2 mV, -±4.5% +2 mV, ±4.5% +2 mV, ±4.5%
Fig. 5-17. Typical CRT display showing correct
calibrator repetition rate.
g.ADJUST--R154, Frequency (Fig. 5-16), for one cycle of calibrator waveform for each marker. See Fig. 5-17 (Since the sweep was calibrated in step 4, the display should also be one cycle of calibrator waveform for each major division.)
h.Position the leading edge of the tenth cycle of calibrator waveform to the center of the graticule and turn the 5X Magnifier on.
i.Make final adjustment of the Frequency control by aligning the positive-going transition of the square wave with the positive-going transition of the tine marker.
j.Turn the 5>X Magnifier off and position the sweep start to the left edge of the graticule.
k.Disconnect the time-mark generator.
20. Check Calibrator Duty Factor
a.Change the following control settings:
Mode (amplifier unit) Ch 1 Volts/Div (amplifier unit) 1 Time/Div (time-base unit) 0.1 ms
b.Center the display vertically with the Ch 1 Position
control.
c.Set the Triggering Level control so the display starts
on the rising portion of the waveform.
d.Turn the 5X Magnifier on. e.Position the 50% point on the falling edge of the
calibrator waveform to the center vertical line.
f.Set the Triggering Slope to
(A)
5-16A
Fig. 5-18. Typical CRT display when checking calibrator
duty cycle.
Performance Check/Calibration-Type R561B
g.CHECK-50% point on the rising edge is now displayed not more than two divisions from the center vertical line (indicates a duty factor of 48% to 52%; see Fig. 5-18).
21. Check Calibrator Risetime
a.Change the following control settings:
Volts/Div 0.5 5 X Mag Off Time/Div (Normal Sweep) 0.5 ms Time/Div (Delayed Sweep) 1 µs Triggering Slope +
b.Adjust the amplifier unit Variable Volts/Div control (concentric with Volts/Div switch) for exactly 5 divisions of vertical display.
c.Set the time-base Mode switch to Intensified and adjust the Delay Time dial to brighten the leading edge of the second cycle of display (approximately 2.00).
d.Set the time-base Mode switch to Delayed Sweep.
5-17A
(A)
Performance Check/Calibration-Type R561B
Fig. 5-19. Idealized waveform showing maximum
allowable calibrator risetime at sweep rate of 1
µ
s/division.
e. Position the 10% point on the leading edge to a vertical graticule line. It may be necessary to increase the intensity slightly to see the leading edge.
f. CHECK-CRT display for one division or less between the 10% and 90% points on the leading edge of the calibrator waveform (one microsecond, or less, risetime); see Fig. 5-19
g. Set the time-base Mode switch to Norm, amplifier
Volts/Div switch to 5 and the CALIBRATOR switch to 40V.
h. Repeat steps b through e to obtain the leading-edge
display of the 40-volt calibrator waveform.
i. CHECK-CRT display for 2.5 divisions or less between the 10% and 90% points on the leading edge of the calibrator waveform (2.5 microseconds, or less, risetime).
j. Disconnect the coaxial cable.
22. Check Current through Probe Loop
a. Change the following control settings:
CALIBRATOR 10 mA Mode (time-base) Norm Volts/Div 0.01 (Calibrated)
Fig. 5-20. Typical CRT display when checking calibrator
current
b. Connect the current-measuring probe and passive termination to the Ch 1 input connector. Set the passive termination for a sensitivity of 2 mA/mV.
c. Clip the current probe around the probe loop on the
front panel.
d. Position the display vertically so the amplitude of the square-wave current can be measured.
e. CHECK-CRT display is 0.5 division in amplitude (ten milliamperes; see Fig. 5-20).
This step checks for the presence of current in the probe loop. This current will remain within the stated 1% accuracy due to the tolerance of the divider resistors and tolerance of the calibrator output voltage. If it is necessary to verify the accuracy of the calibrator current, use a current measuring meter with an accuracy of at least 0.25%.
f. Disconnect all test equipment. This completes the
calibration procedure for the Type R561B Oscilloscope.
.
NOTE
5-18A
(A)

SECTION 6

MAINTENANCE CALIBRATION RIV1561A
5
Introduction
Complete calibration information for the Type RM561A is given in this section. This procedure checks the instrument against the performance requirements listed in the Characteristics section. The Type RM561A can be returned to original performance standards by completion of each step in this procedure. If it is desired to merely touch up the calibration, perform only those steps entitled "Adjust ...". A short-form calibration procedure is also provided in this section for the convenience of the experienced calibrator. It may also be used as a calibration record or as an index to the steps in the complete Calibration Procedure. The Type RM561A should be checked, and recalibrated if necessary, after 500 hours of operation, or every six months if used infrequently, to assure correct operation and accuracy.
EQUIPMENT REQUIRED
The following test equipment (shown in Figs. 6-1 and 6-2), or its equivalent, is required for complete calibration of the Type RM561A. Specifications given are the minimum necessary for accurate calibration of this instrument. All test equipment is assumed to be correctly calibrated and operating within the given specifications. If equipment is substituted, it must meet or exceed the specifications of the recommended equipment.
For the quickest and most accurate calibration, special calibration fixtures are used where necessary. All calibration fixtures listed here can be obtained from Tektronix, Inc. Order by part number through your local Tektronix Field Office or representative.
5. Test oscilloscope. Frequency response, 10OMHz; minimum deflection factor, 0.005 volts/division. Tektronix 530-, 540-series Oscilloscope with Type B Plug-In Unit or Tektronix Type 561A/RM with Type 3A1 and 2A63 Amplifier Plug-Ins and Type 3B4 Time-Base Plug-In recommended.
6. CRT deflection capacitance normalizer. Tektronix Calibration Fixture 067-0500-00 recommended.4
7. Standard amplitude calibrator. Amplitude accuracy, 0.25%; signal amplitude, 0.2 millivolts to 100 volts; output signals, 1-kHz square-wave and +DC. Tektronix Calibration Fixture 067-0502-00 recommended.
8. Termination. Impedance, 50-ohm; connectors, GR; accuracy, ±0.1%. Tektronix Calibration Fixture 067­0515-00.
9. 1X probe. BNC Connector. Tektronix P6028 Probe. 10. Cable (three). Impedance, 50-ohm; connectors, BNC. Tektronix Part No. 012-0057-00.
11. Patch cord. Length, 18 inches; connectors,
BNC-to-banana plug. Tektronix Part No. 012-0091-00.
12. Patch cord. Length, 18 inches; connectors,
BNC-to- BNC. Tektronix Part No. 012-0087-00.
13. Adjustment tool. Insulated screwdriver, non-
metallic. Tektronix Part No. 003-0000-00.
14. Adapter. GR to BNC female. Tektronix Part
No. 017-0063-00.
15. Adapter. GR to BNC male. Tektronix Part No.
017- 0064-00.
16. Jumper lead (not shown). Length, 6 inches; connectors, insulated alligator clips. Not available from Tektronix.
CALIBRATION RECORD AND INDEX
1. Tektronix 2- or 3-series amplifier plug-in unit.’
2. Tektronix 2- or 3-scries time base plug-in unit.’
3. DC voltmeter. Minimum sensitivity, 20,000 ohms/volt; accuracy, checked to 1% at -100 volts,-12.2 volts, +125 volts and +300 volts and ±3% at --3300 volts. For example, Simpson Model 262.
2
4. Variable autotransformer. Must be capable of supplying at least 350 volt-amperes over a voltage range of 105 to 125 volts (210 to 250 volts for 230-volt nominal line). For example, General Radio W10MT3W Metered Variac Auto­transformer.
1
Use maximum performance plug-in units. For example, use a
3
high-frequency unit in preference to low-frequency, dual-tracs rather than single-trace, etc.
2
When used with sampling plug-in units. The Type RM561A low-voltage power supplies must be adjusted to 0.5%. or better for correct plug-in operation. Use a DC voltmeter with accuracy better than 0.05 %.
3
Used only to check power-supply regulation and ripple in steps 6 and 7. May be deleted from list if these checks are not performed.
5
For calibration procedure, see TB 750-236
This short-form calibration procedure is provided to aid in checking the operation of the Type RM561A. It may be used as a calibration guide by the experienced calibrator, or it may be used as a record of calibration. Since the step numbers and titles used here correspond to those used in the complete Calibration Procedure, this procedure also serves as an index to locate a step in the complete Calibration Procedure. Performance requirements correspond to those given in the Characteristics section.
Type RM561A, Serial No. ___________________________ Calibration Date___________________________________ Calibration Technician______________________________
4
Normally required only If new cathode-ray tube has been installed or if deflection-plate compensation has been inadvertently misadjusted. An alternative method of adjustment is also given in step 16b.
6-1
Calibration-Type RM561A
Fig. 6-1. Recommended calibration equipment, items 1 through 7.
6-2
(A)(B)
Calibration-Type RM561 A
Fig. 6-2. Recommended calibration equipment, items 8 through 15.
1. Adjust -100-Volt Power Supply (Page 6-5) --100 volts,
±3 volts.
2. Adjust +125-Volt Power Supply (Page 6-5) 4-125
volts, ±3.8 volts.
3. Adjust -.-300-Voit Power Supply (Page 6-6) +300
volts, ±9 volts.
4. Adjust -.1 2.2-Voi Power Supply (Page 6-6) .--12.2
volts, ±0.37 volt.
5. Adjust High-Voltage Power Supply (Page 6-6) -3300
volts, ±100 volts.
6. Check Low-Voltage Power Supply Regulation (Page 6-
6)
--100 volt +1 volt +125 volt ±1.25 volts +300 volt ± 3 volts
--12.2 volt ±0.12 volt Check regulation at low and high line.
7. Check High-Voltage Power-Supply Regulation (Page
6-7) Within +100 volts at low and high line.
8. Adjust Calibrator Amplitude (Page 6-8)
+100 volts, ±3 volts.
9. Check D838-D839 Junction Voltage (Page 6-8)
Less than +0.6 volt DC.
10. Check Power-Supply Ripple (Page 6-9)
--100 volt 5 millivolts +125 volt 10 millivolts +300 volt 80 millivolts
--12.2 volt 3 millivolts
-3300 volt DO NOT MEASURE
11. Adjust Astigmatism (Page 6-10)
Sharp, well-defined display.
12. Adjust Trace .Alignment (Page 6-10)
Trace parallel to horizontal graticule lines
13. Adjust Cathode-Ray Tube Geometry (Page 6-11)
Best overall geometry.
6-3
(A)(B)
14. Check Calibrator Accuracy (Page 6-12)
Within ±3% of indicated voltage.
Calibration-Type RM561A
during calibration should be corrected using the techniques given in the Maintenance section.
15. Check Calibrator Risetime and Symmetry (Page 6-14)
Risetime, less than 6 microseconds; symmetry, 40% to 60%.
16. Adjust Vertical and Horizontal Deflection-Plate
Compensation (Page 6-15) Optimum square corner (or 14.3 picofarads effective
capacitance).
17. Check Dual-Trace Chopped Blanking (Page 6-17)
Blanking of between channel switching transients (with dual-trace amplifier unit).
18. Check CRT Intensifier Circuit (Page 6-18)
Intensified portion on trace (with delayed sweep time­base unit).
19. Check CRT Vertical Sensitivity (Page 6-19) 19.5
volts, ±1 volt.
20. Check CRT Horizontal Sensitivity (Page 6-19) 18.4
volts, ±0.9 volt.
21. Check Intensity (Z-Axis) Modulation (Page 6-19)
Intensity modulation with 10 volts input.
CALIBRATION PROCEDURE
General
The following procedure is arranged in a sequence which allows the Type RM561A to be calibrated with the least interaction of adjustments and reconnection of equipment. If desired, the steps may be performed individually or out of sequence. However, some adjustments affect the calibration of other circuits within the instrument. In this case, it will be necessary to check the operation of other parts of the instrument. When a step interacts with others, the steps which need to ’e checked will be noted.
Any needed maintenance should be performed before proceeding with calibration. Troubles which become apparent.
The Adjust . . . steps in the following procedure provide a check of instrument performance, whenever possible, before the adjustment is made. To prevent recalibration of other circuits when performing a partial calibration, readjust only if the listed tolerance is not not. However, when performing a complete recalibration, best overall performance will be provided if each adjustment is made to the exact setting, even if the CHECK . . . is within the allowable tolerance. The
symbol is made.
In the following calibration procedure, a test equipment setup is shown for each major group of adjustments and checks. Beneath each setup picture is a complete list of front-panel control settings for the Type RM561A. To aid in locating individual controls which have been changed during complete calibration, these control names are printed in bold type. If only a partial calibration is performed, start with the nearest setup preceding the desired portion.
The following procedure uses the equipment listed under Equipment Required. If substitute equipment is used, control settings or setup must be altered to meet the requirements of the equipment used.
Preliminary Procedure
rack, pull it out to the fully extended position. Type RM561 A. in compartment. plug-in compartment. power source. autotransformer output.
Allow at least 20 minutes worm it) for chcckin1ig the instrument the given accuracy.
is used to identify the steps in which an adjustment
1. If the Type RM561A is mounted in a cabinet
2. Remove the top and bottom covers from the
3. Insert the amplifier plug-in unit in the left plug
4. Insert the time-base plug-in unit in the right
5. Connect the autotransformer to a suitable
6. Connect the Type RMA561 A to the 7 Set the autotransformer to 15 (or 230) volts.
8 Set the Type RM561A POWER switch to ON.
6-4
(A)(B)
Calibration-Type RM561 A
Fig. 6-3. Test equipment setup for power supply adjustments
Control Settings
Type RM561A
FOCUS Any position INTENSITY Midrange CALIBRATOR OFF SCALE ILLUJM Clockwise CRT CATHODE SEI ECTOR Normal
Amplifier Unit
Position Midrange AC DC Gnd AC Volts/Division 1 Variable Calibrated
Time-Base Unit
Position Midrange Magnifier Off Time/Division 1 millisecond
.
Variable Calibrated Triggering controls Any Position
1. Adjust --100-Volt Power Supply It
a. Test equipment setup is shown in Fig 6-3. b. Connect the DC voltmeter from the - 100 volt test
point (Fig. 6-4A) to chassis ground.
c. CHECK –Meter reading; -- 100 volts, ±3 volts.
d. ADJUST - - 100V adjustment, R616 (Fig. 6-4B) for -
100 Volts.
e. INTERACTION - May affect operation of all circuits
within the Type RM561A.
2. Adjust + 125-Volt Power Supply
a. Test equipment setup is shown in Fig. 6-3. b. Connect the DC voltmeter from the +125,-volt test
point (Fig. 6-4A) to chassis, ground.
c. CHECK-Meter reading; +125 volts, ±3.8 volts.
(A)(B)
6-5
Calibration-Type RM561 A
4. Adjust -1 2.2-Volt Power Supply
a. Test equipment setup is shown in Fig. 6-3. b. Connect the DC voltmeter from the --12.2-volt test
point (Fig. 6-6) to chassis ground.
c. CHECK-Meter reading; -12.2 volts, ±0.37 volt.
d. ADJUST-12.2 Volts adjustment, R730 (Fig. 6-5), for -
12.2 volts. e. INTERACTION-May affect operation of all circuits
within the Type RM561A.
5. Adjust High-Voltage Power Supply
a. Test equipment setup is shown in Fig. 63. b. Connect the DC voltmeter from the high-voltage test
point (Fig. 6-7A) to chassis ground.
c. CHECK-Meter reading; -3300 volts, ±100 volts.
d. ADJUST-High Voltage adjustment, R841 (Fig. 6-7B),
for -3300 volts.
e. INTERACTION-Check steps 7, 11-13 and 16.
Fig. 6-4. (A) Location of -100-, +125- and +300-volt test
points (bottom view), (B) location of -100-volt adjustment
(top view).
d. ADJUST-+125 Volts adjustment, R656 (Fig. 6-5), for
+125 volts.
e. INTERACTION-May affect operation of all circuits
within the Type RM561A.
3. Adjust +300-Volt Power Supply
a. Test equipment setup is shown in Fig. 6-3. b. Connect the DC voltmeter from the +300-volt test
point (Fig. 6-4A) to chassis ground.
c. CHECK-Meter reading; +300 volts, ±9 volts. d. ADJUST-+300 Volts adjustment, R676 (Fig. 6-5), for
+300 volts.
e. INTERACTION-May affect operation of all circuits
within the Type RM561A.
Fig. 6-5. Location of + 125-, +300- and --12.2-volt
adjustment (bottom view).
6. Check Low-Voltage Power-Supply Regulation
a. Test equipment setup is shown in Fig. 6-3. b. Connect the DC voltmeter from the --100-volt test
point (Fig. 6-4A) to chassis ground.
c. With the autotransformer set to 117 volts (or 234
volts) note the test point voltage
(A)(B)
6-6
d.Set the autotransformer to 105 (210) volts and again check the voltage. Voltage should be within 1 volt of reading in step c.
e.Set the autotransformer to 125 (250) volts and check the voltage. Voltage should be within 1 volt of reading in step c.
f.Remove both plug-ins from the instrument (minimum
load).
g.With the autotransformer set to 125 (250) volts, the voltage should be within 1 volt of reading in step c.
h.Return the autotransformer to 105 (210) volts and check the voltage. Voltage should be within 1 volt of reading in step c.
Calibration-Type RM561A
Fig. 6-6. Location of -1 2.2-volt test point (bottom view).
i.Repeat this procedure for each low-voltage supply. Table 6-1 shows maximum voltage deviation allowable for each supply. Test point locations are shown in Figs. 6-4A and 6-6.
NOTE A 150-milliamp minimum load current is necessary for proper regulation of the --
12.2-volt supply. Minimum-load regulation of this supply can be checked by placing an 82-ohm, 5-watt resistor between pins 5 and 16 of either interconnecting plug.
TABLE 6-1.
Power Maximum Voltage
Supply Deviation
-100 volt 1 volt +125 volt 1.25 volts +300 volt 3 volts
-12.2 volt 0.12 volt
Fig. 6-7. (A) Location of high-voltage test point (bottom
view), location of high-voltage adjustment (top view).
7. Check High-Voltage Power-Supply Regulation
a.Test equipment setup is shown in Fig. 6-3. b.Connect the DC voltmeter from the high-voltage test
point (Fig. 6-7A) to chassis ground.
c.Set the autotransformer to 117 (234) volts and note
the voltage at the test point.
d.Set the autotransformer to 105 (210) volts and check
the voltage. It should be within 100 volts of reading in step c.
(A)(B)
6-7
Calibration-Type RM561 A
Fig 6-9. Location of D838-D839 test point (bottom
view).
f Return the autotransformer to 117 (234) volts and
disconnect the voltmeter.
8. Adjust Calibrator Amplitude
Fig. 6-8. (A) Location of calibrator test point bottom view), (B) location of calibrator adjustment (top view)
e.Set the autotransformer to 125 (250) volts and check the voltage. Voltage should be within 100 volts of reading in step c.
a.Test equipment setup is shown in Fig. 6-3. b.Connect the DC voltmeter from the calibrator test
point (Fig. 6-8A) to chassis ground
c.CHECK---Meter reading; 4-100 volts, ±3 volts Be
sure the CALIBRATOR switch is set to OFF.
d.ADJUST--Cal Ampl adjustment, R871 (Fig 6-8B),
for + 100 volts.
9. Check D838-D839 Junction Voltage
a.Connect the voltmeter from the D838-D839 test
point (Fig 6-9) to chassis ground.
b.CHECK--Voltage should be less than +0.6 volt DC. Higher voltages could damage a time-base unit with an intensifying circuit.
(A)(B)
6-8
Calibration Type RMS61 A
Fig. 6-10. Test equipment setup for power-supply ripple measurements.
Control Settings
Type RM561A
FOCUS Any position INTENSITY Midrange CALIBRATOR OFF SCALE ILLUM Clockwise CRT CATHODE SELECTOR Normal
Amplifier Unit
Position Midrange AC DC Gnd AC Volts/Division 1 Variable Calibrated
Time-Base Unit
Position Midrange Magnifier Off Time/Division 1 millisecond Variable Calibrated
Triggering controls Adjust so sweep is not
triggered
10. Check Power-Supply Ripple
b. Connect the 1X probe to the test oscilloscope input.
c. Set the test oscilloscope volts/division switch to
0.005. Set the input coupling switch to AC.
d. Measure the line-frequency ripple of each regulated power supply at the power-supply test points. Test points are shown in Figs. 6-4A and 6-6. Table 6-2 gives maximum ripple for each supply. Check ripple with autotransformer set at 105 and 125 (210 and 250 if wired for 234-volts nominal). Fig. 6­11 shows typical test oscilloscope display.
e. Disconnect the test oscilloscope and return the auto­transformer to line voltage (instrument may be connected directly to line).

TABLE 6-2

Maximum
Power Ripple
Supply (millivolts)
-100 volt I 5 +125 volt 10
+300 volt - 80
--12.2 volt 3
-3300 volt DO NOT MEASURE
a. Test equipment setup is shown in Fig. 6-10.
(A)(B)
6-9
Fig. 6-1 1. Typical test oscilloscope display showing
power-supply ripple (-100-volt supply). Volts/division,
0.005 volts, AC coupled; time/division, 5 milliseconds
11. Adjust Astigmatism
Calibration--Type RM561A
.
Fig. 6-12. Location of ASTIGMATISM and TRACE
ALIGNMENT controls(front panel).
a. Set the tine-base unit triggering controls to produce a
trace on the CRT.
b. ADJUST--FOCUS control and ASTIGMATISM adjustment, R864 (Fig. 6-12) for c sharp, well-defined display over the entire race length.
c. Slight readjustment of these controls may be necessary during normal operation.
12. Adjust Trace Alignment
a. Adjust the amplifier unit Position control to move the
trace to the horizontal graticule centerline.
b. CHECK-The trace should be parallel with the center-
line.
c. ADJUST-TRACE ALIGNMENT adjustment, R860 (Fig. 6-12), so the trace is parallel with the horizontal graticule lines.
(A)(B)
6-10
Calibration--Type RM561A
Fig. 6-13. Test equipment setup for geometry adjustment and CALIBRATOR check.
Control Settings
Type RM561A
FOCUS Adjust for focused
display
INTENSITY Midrange
CALIBRATOR 100 V
SCALE ILLUM Clockwise CRT CATHODE SELECTOR Normal
Amplifier Unit
Position Midrange AC DC Gnd AC Volts/Division 1 Variable Calibrated
Time-Base Unit
Position Midrange Magnifier Off Time/Division 1 millisecond Variable Calibrated
Triggering Controls Automatic
13. Adjust Cathode-Ray Tube Geometry
a. Test equipment setup is shown in Fig. 6-13.
b. Set the Standard Amplitude Calibrator controls as follows (if the test oscilloscope has a 1-kHz Calibrator, it may be used for this step):
Amplitude 10 Volts Mode Square wave Function Up Power On
c. Connect the Standard Amplitude Calibrator Output
connector to the amplifier unit input connector.
d. Adjust the time-base unit Position and Variable Time/ Division controls so a rising portion of the waveform coincides with the left and right graticule lines. The INTENSITY control may need to be advanced slightly to view the rising portions of the waveform (top and bottom of square wave should be positioned off of the viewing area).
e. CHECK-Good linearity at the left and right edges of the graticule (Fig. 6-14A shows typical CRT display as well as examples of poor geometry).
6-11
(A)(B)
Calibration-Type RM561 A
Fig. 6-14. (A) Idealized waveforms showing good geometry with examples of poor geometry, (B) location of Geometry
adjustment (top view).
f. ADJUST-Geometry adjustment, R865 (Fig. 6-14B), for best linearity at left and right edges of the graticule. If a setting cannot be found that provides best linearity on both sides of the graticule, a compromise setting will be necessary to provide best overall linearity.
g. Disconnect the signal from the amplifier unit input and
return the INTENSITY control to normal setting.
14. Check Calibrator Accuracy
a. Test equipment setup is shown in Fig. 6-13. b. Change the following Standard Amplitude Calibrator
controls:
Amplitude 100 Volts Mode +DC Function Mixed AC DC Gnd DC
c. Connect the CAL OUT connector to the Standard
Amplitude Calibrator Unknown Input with a 50-ohm cable.
d. Connect the standard Amplitude Calibrator Output connector to the test oscilloscope input (minimum deflection factor, 0.005 volts/division) with a 50-ohm cable.
e. Connect the six-inch jumper lead from pin 8, V884 (Fig. 6-15A), to chassis ground. This provides a DC calibrator output voltage.
f. The difference between the Standard Amplitude Calibrator +DC voltage and the Type RM561A CALIBRATOR output voltage should be less than 3%. Table 6-3 gives maximum allowable difference for 3% CALIBRATOR accuracy at each position of the CALIBRATOR switch. (The Standard Amplitude Calibrator output can be identified by setting the Type RM561A CALIBRATOR switch to a higher or lower position; the signal which remains is the Standard Amplitude Calibrator output.) Fig. 6-15B shows a typical test oscilloscope display. Check accuracy of all positions between 100V and 100 mV.
g. Set the CALIBRATOR switch to the 1 V INTO 50
position (1 V position).
h. Connect the 50-ohm termination (067-0515-00) between the CAL OUT connector and the 50 ohm table using the GR to BNC adapters.
i. Leave the Standard Amplitude Calibrator Amplitude switch set to .1 Volts and set the test oscilloscope volts/division switch to 0.005.
j. The test oscilloscope display should be 0.6 division or less in amplitude for 3% accuracy.
k. Remove the 50-ohm termination and -connect the
50-ohm cable to the CAL OUT connector.
I. Connect the Standard Amplitude Calibrator Output connector to the X 100 Amplifier Input connector with a 50­ohm cable.
(A)(B)
6-12
Calibration-Type RM561A
Fig. 6-15 (A) Location of Pin 8, V884 (bottom view), (B) typical test-oscilloscope display showing about +1.5% error in the
10V position (volts/division, 0.1 volts; time/division, 5 milliseconds), (C) typical test-oscilloscope display showing about
+1.2% error in the 10mV position (volts/division, 20 millivolts; time/division, 0.2 milliseconds; using X100 Amplifier section).
m. Connect the Output ±50 V Max connector to the input
of the test oscilloscope.
n. Set the test oscilloscope time/division switch to 0.2 milliseconds. Adjust the Standard Amplitude Calibrator Output DC. Level control to center the trace on the screen.
o. Check the accuracy of the 10mV and 1 mV CALIBRATOR switch positions. Table 6-3 gives maximum allowable difference and control settings. Fig. 6-15C shows a typical test oscilloscope display.
p. Disconnect all test equipment and remove the jumper lead from pin 8 of V884.
Fig 6-16. (A) Typical test oscilloscope display for checking Calibrator risetime (time/division, 1 microsceond), (B)
typical test oscillioscope display for checking Calibrator symmetry.
6-13
Calibration-Type RM561A

TABLE 6-3

Type RM561 A
CALIBRATOR
setting and
Standard
Amplitude
Calibrator setting
Test
Oscilloscope
Volts/Division
Switch Setting
100 V 1 Volt 3
Maximum
Display
Amplitude for
3% Accuracy
(divisions)
5
10 V .1 Volt 3
1 V 10 Millivolts 3
100 mV 5 Millivolts 0.6
10 mV 10 Millivolts 3
5
For maximum Calibrator accuracy, reset the Cal Ampl
1 mV 5 Millivolts 0.6
adjustment for minimum difference
15. Check Calibrator Risetime and Symmetry
a. Test equipment setup is shown in Fig. 6-13.
b. Connect the CAL OUT connector to the vertical input of the test oscilloscope with a 50-ohm cable.
c. Set the CALIBRATOR switch to 1 V. d Set the test oscilloscope time/division switch to 1 micro­second and the volts/division and variable control for exactly 5 divisions of display (four divisions if using 540A-series test oscilloscope).
e. Check for risetime of less than 6 microseconds as shown by less than 5 divisions between the 10% and 90% points on the test oscilloscope display (see Fig. 6-16A).
f. Set the test oscilloscope Time/Division switch and Variable control to display one complete cycle of the calibrator signal in 10 divisions. Adjust the triggering controls, If necessary, so the display starts on the leading edge of the waveform.
g. Check for Calibrator symmetry of ±20%. This is shown by each segment of the square wave having a length of 4 to 6 divisions (see Fig. 6-16B).
h. Disconnect the test oscilloscope.
Fig. 6-17. Test equipment setup for vertical and horizontal compensation.
6-14
(A)(B)
Calibration-...-Type RM561A
Fig. 6-18. (A) Typical CRT display showing correct vertical compensation adjustment, (1) and (C) incorrect adjustment,
(D) location of C760 (Ieft side).
Control Settings
Type RM561A
FOCUS Adjust for focused
display INTENSITY Midrange CALIBRATOR 100 V SCALE ILLUM Clockwise CRT CATHODE Normal SELECTOR
Amplifier unit
Position Midrange AC DC Gnd AC Volts/Division 1 Variable Calibrated
Time-Base Unit
Position Midrange Magnifier Off
Time/Division .2 millisecond
Variable Calibrated
Triggering Controls Adjust for triggered
sweep
16A. Adjust Vertical and Horizontal Deflection-
Plate Compensation
a. Test equipment is shown in Fig. 6-17. b. Insert the CRT Deflection Capacitance Normalizer
into the left plug-in compartment
c. Connect the CAL OUT connector to the Capacitance
Normalizer input connector.
d. Adjust the time-base triggering controls so the
leading edge of the waveform is displayed.
NOTE The INTENSITY control may need to be advanced slightly to view the leading edge of the waveform
e. CHECK-For optimum square corner (see: Fig. 6-
18A).
f. ADJUST--C760 (Fig. 6-18D) for optimum square
corner.
g. Interchange the Capacitance Normalizer and the time-base unit. Readjust the FOCUS control and the time-base triggering controls for a stable well-focused display.
h. CHECK--For optimum square corner (see Fig. 6­19A).
6-15
(A)(B)
Calibration-Type RM561 A
Fig. 6-19. (A) Typical CRT display showing correct adjustment of horizontal compensation, (B) location of C761 (right
side)
i. ADJUST-C761 (Fig. 6-19B) for optimum square
corner.
j. Remove the Capacitance Normalizer. Replace the amplifier unit in the left compartment and the time-base unit in the right compartment.
16B. Alternative Method of Adjusting Deflection Plate Compensation
NOTE The following method can be used to adjust the deflection-plate compensation if a Capacitance Normalizer is not available.
The effective deflection-plate capacitance of the CRT is the capacitance seen by the plug-in unit when the deflection plates are driven push-pull. Therefore, it cannot be measured directly with a capacitance meter However, the Individual circuit capacitances which make up the effective deflection­plate capacitance can be measured. These individual circuit capacitances are shown schematically in Fig. 6-20. C1 and C.2 represent the capacitance from each deflection plate to ground C3 represents the capacitance between the deflection plates. The variable capacitor is made up of the circuit capacitance plus the capacitor added for adjustment.
The variable capacitor for the vertical (left) plug-in compartment may be C3 (Fig. 6-20A) or C2 (Fig. 6-20B) depending on other circuit capacitance values.
Fig. 6-20. Schematic representation of the effective CRT deflection-plate capacitance; (A) configuration 1, (B)
configuration 2.
(A)(B)
6-16
Calibration-Type RM561A
Fig. 6-21. Typical CRT display showing correct chopped blanking. (A) Chopped transients visible with CRT CATHODE
SELECTOR switch up (time/division, 2 microseconds), (8) transients blanked out with CRT CATHODE SELECTOR switch
in the CHOPPED BLANKING position.
Check the instrument to determine the configuration and
formula to use for correct deflection-plate capacitance.
The effective deflection-plate capacitance, C
expressed in terms of C1, C2, and C3 as follows:
C
eff=
21 CC +
2 (C3)
eff
can be
2
Setting C
the value of the variable capacitor con be calculated. Configuration 1, Fig. 6-20A. C3 = 7.15 pF ­Configuration 2, Fig. 6-20B. C2 = 28.6 pF - 4 (C3) - C1
a. Disconnect the power cord and isolate the Type
RM561A from ground.
b. Remove both plug-in units.
c. Connect the capacitance meter guard voltage to pin 21 of the vertical plug-in compartment and measure the capacitance between pin 17 and the oscilloscope chassis this is C1.
d. Connect the capacitance meter guard voltage to pin 17 and measure the capacitance between pin 21 and the oscilloscope chassis-- this is C2. (For configuration 2, connect the guard voltage to the oscilloscope chassis and measure the capacitance between pin 17 and 21--this is C3.)
e. Substitute the measured capacitance values into the equation and solve for C3 configuration 2, solve for C2).
f. Connect the guard voltage to the oscilloscope
chassis and measure the capacitance between pins 17 and
21. (For configuration 2, connect the guard voltage to pin 17 and measure the capacitance between pin 21 and the oscilloscope chassis.)
g. Adjust C760 until the measured capacitance equals the value calculated for C3 in step e (configuration 2, adjust C760 for calculated value of C2).
h. Repeat steps c through g for the horizontal (right) plug-in compartment. Fig. 6-20A shows the individual capacitances for the horizontal plug-in comportment.
, equal to 14.3 picofarads (value set at factory),
eff
CC12
+
4
17. Check Dual-Trace Chopped Blanking
a. Set the amplifier unit Mode switch to Chopped (this
check can be made only with a dual-trace plug-in unit).
b. Set the amplifier unit AC DC Gnd switches to Gnd.
c. Position the traces about three divisions apart.
d. Set the time-base Time/Division switch to display several cycles of the waveform and adjust the triggering controls for a stable display.
e. With the CRT CATHODE SELECTOR switch in the normal position, notice the overshoot on the display (see Fig 6-21A).
f. Set the CRT CATHODE SELECTOR switch to CHOPPED BLANKING. The between-channel switching transients should be removed from the display (see Fig. 6­21B).
Fig. 6-22. Typical CRT display showing correct
intensifier circuit operation.
6-17
Calibration-Type RM561A
18. Check CRT Intensifier Circuit
a. Set the time-base Time/Division switch to 1 millisecond and the delayed sweep Time/Division switch to 0.1 millisecond (this check can be made only with a delayed sweep plug-in unit).
b. Set the amplifier unlit mode switch to display only one channel.
c. Adjust the time-base triggering controls to produce a trace.
d. Adjust the FOCUS control and the time-base triggering controls for a well-focused free-running (or automatic triggered) trace.
e. Turn the time-base Mode switch to Intensified (not triggered).
f. An intensified portion should appear on the trace (see Fig. 6-22). If the display appears only as a shortened trace, increase the INTENSITY setting slightly until the complete trace, showing on intensified portion, appears.
g. Turn the delay time control throughout its range and
check that the intensified portion moves along the trace.
Fig. 6-23. Test equipment setup for checking CRT vertical and horizontal sensitivity.
Control Settings AC DC Gnd AC
Type RM561 A Volts/Division 1
FOCUS Adjust for focused Variable Calibrated
display
INTENSITY Midrange Time-Base Unit CALIBRATOR 10 V
Position Midrange
SCALE ILLUM Clockwise
CRT CATHODE Normal Magnifier Off
SELECTOR Time/Division 10 milliseconds
Amplifier Unit Variable Calibrated
Position Midranqe Triggering controls Automatic
6-18
(A)(B)
CAUTION
In steps 19 and 20 be careful not to short the CRT deflection pins to chassis ground.
19. Check CRT Vertical Sensitivity
a. Test equipment setup is shown in Fig. 6-23. b. Connect the DC voltmeter between the vertical
deflection-plate pins. Be careful not to bend the pins.
c. T urn the am plifier unit Position control to m ove
the trace to the top graticule line.
d. Note the meter reading. e. Turn the amplif ier unit Position control to m ove
the trace to the bottom graticule line.
f. Again note the meter reading.
g. Calculate the difference in voltage between step d and f. Divide by 8 to determine the deflection voltage/ division (sensitivity).
h. Sensitivity should be 19.5 volts/division, +1 volt.
20. Check CRT Horizontal Sensitivity
a. Test equipment setup is shown in Fig. 6-23.
b. Interchange the plug-ins so the tim e-base unit is in the left compor tment and the amplifier unit is in the right compartment.
c. Connec t the DC voltmeter between the vertical deflection-plate pins. Be careful not to bend the pins.
d. Adjust the FOCUS control and the time-base triggering controls for a well-focused free-running (or automatic triggered trace).
e. Turn the amplif ier unit Position control to m ove the trace to the left graticule line.
f. Note the meter reading.
g. Turn the amplif ier unit Position control to m ove the trace to the right graticule line.
h. Again note the meter reading.
i. Calculate the difference in voltage between step f and h. Divide by 10 to determine the deflection voltage/ division (sensitivity).
Calibration—Type RM561
Fig. 6-24. Typical CRT display showing correct Intensity modulation (time/division. 10 milliseconds).
j. Sensitivity should be 18.4 volts/division, ±0.9
volt.
k. Disconnect the voltmeter and return the amplifier unit to the left com partment and the tim e-base unit to the right compartment.
21. Check Intensity (Z-Axis) Modulation
a. Remove the ground strap between the rear­panel binding posts.
b. Connect the CALIBRATOR signal to the EXT CRT CATHODE binding post with the BNC-to-BNC and BNC-to-banana jack jumper leads.
c. Chec k the display for intensified dots (s ee Fig. 6-24). It may be necessary to reduce the INTENSITY setting slightly to see the dots.
d. Remove the jumper leads and replace the ground strap.
This completes the c alibration of the T ype RM561A. Disconnect all test equipment. Replace the top and bottom panels. The gain of the amplifier unit and tim ing of the time-base unit should be c hecked before using the system.
(A)(B)
6-19
PARTS LIST ABBREVIATIONS
BHB binding head brass int internal BHS binding head steel Ig length or long cap. capacitor met. metal cer ceramic mtg hdw mounting hardware comp composition OD outside diameter conn connector OHB oval head brass CRT cathode-ray tube OHS oval head steel csk countersunk PHB pan head brass DE double end PHS pan head steel dia diameter plstc plastic div division PMC paper, metal cased elect. electrolytic poly polystyrene EMC electrolytic, metal cased prec precision EMT electrolytic, metal tubular PT paper, tubular ext external PTM paper or plastic, tubular, molded F & I focus and intensity RHB round head brass FHB flat head brass RHS round head steel FHS flat head steel SE single end Fil HB fillister head brass SN or S/N serial number Fil HS fillister head steel SW switch h height or high TC temperature compensated hex. hexagonal THB truss head brass HHB hex head brass thk thick HHS hex head steel THS truss head steel HSB hex socket brass tub. tubular HSS hex socket steel var variable ID inside diameter w wide or width incd incandescent WW wire-wound
7-0.1
PARTS PROVISIONING INFORMATION
REPLACEMENT PARTS
To obtain replacement parts, find the manufacturer’s part number and description in this manual and then refer to the appropriate Repair Parts and Spec ial Tools List (RPST L) TH. In the RPSTL, find the assembly or subassembly first and then the desc ription which corresponds with
that in this manual. Under the description in the RPSTL find the manuf actur er’s part num ber , and then order the part by the listed Federal stock number . If the part is not listed in the RPSTL, it should be requisitioned from the NICP in accordance with AR 725-50.
SPECIAL NOTES AND SYMBOLS
X000 Part first added at this serial number
00 X Part removed after this serial number
*000-0000-00 Asterisk preceding Tektronix Part Number indicates manufactured by
or for Tektronix, Inc., or reworked or checked components.
Use 000-0000-00 Part number indicated is direct replacement.
Screwdriver adjustment. Control, adjustment or connector.
7-0
7-0
Type RM561A

SECTION 7

ELECTRICAL PARTS LIST
TYPE RM561A
Values are fixed unless marked Variable.
Tektronix S/N Range
Ckt. No. Part No. Description
Bulbs
B601 150-001 Incandescent, G.E. #47, Graticule Light B602 150-001 Incandescent, G.E. #47, Graticule Light B603 150-004 Incandescent, G.E. #328, Pilot Light B856 150-025 Neon, NE-2E 5001-12899 B856 150-0030-00 Neon, NE-2V 12900-up B857 150-025 Neon, NE-2E 5001-12899 B857 150-0030-00 Neon, NE-2V 12900-up
Capacitors
Tolerance ± 20% unless otherwise indicated. Tolerance of all electrolytic capacitors are as follows (with exceptions):
3V - 50V = ––10%, +250%
51 V - 350V = ––10%, +100%
351 V - 450V = ––10%, +50%
C611 285-510 .01 µfMT400 v C616 285-510 .01 µfMT400 v C640A,B Use 290-224 340 µf x 10 µf EMC 250 v C644 Use 290-224 340 µf x 10 µf EMC 250 v C642A,B Use 290-224 340 µf x 10 µf EMC 250 v
C650 285-510 .01µfMT400 v C667 290-002 8 µf EMT 450 v C670 285-510 .01 µfMT400 v C720A Use 290-086 2000 µf EMC 30 v C720B Use 290-086 2000 µf EMC 30 v
C732 Use 290-201 100 µf EMT 15v C737 283-026 .2 µf Disc Type 25 v C757 Use 290-231 100 µf EMT 25 v C760 281-027 .7-3 µf Tub Var. C761 281-027 .7-3 µf Tub Var.
C801 283-006 .02 µf Disc Type 600 v C803 283-000 .001 µf Disc Type 500 v C807 285-502 .001 µfMT1000 v C822 283-071 .0068 µf Disc Type 5000 v C830 Use 283-0071-00 .0068 µf Disc Type 5000 v
C832 Use 283-0071-00 .0068 µf Disc Type 5000 v C837 Use 283-0034-00 .005 µf Disc Type 4000 v C841 285-519 .047 µfMT 400 v C842 283-071 .0068 µf Disc Type 5000 v C853 Use 283-0034-00 .005 µf Disc Type 4000 v C876 290-025 6.25 µf EMT 300 v C878 281-523 100 µf Cer 350 v C884 281-524 150 µf Cer 500 v (F)
7-1
Electrical Parts List-Type RM561A
Diodes
Tektronix S/N Range
Ckt. No. Part No. Description
D640A,B,C,D *152-047 Silicon Replaceable by 1N2862 5001-10309 D640A,B,C,D 152-0066-00 Silicon 1N3194 10310-up D642A,B,C,D *152-047 Silicon Replaceable by 1N2862 5001-10309 D642A,B,C,D 152-0066-00 Silicon 1N3194 10310-up D644A,B,C,D *152-047 Silicon Replaceable by 1N2862 5001-10309 D644A,B,C,D 152-0066-00 Silicon 1N3194 10310-up D663 152-0066-00 Silicon 1N3194 X12500-up D664 152-0066-00 Silicon 1N3194 X12500-up D720 152-035 Silicon 1N1563A D721 152-035 Silicon 1N1563A D745 *152-0107-00 Silicon Replaceable by 1N647 X8270-up D837 152-0066-00 Silicon 1N3194 X11270-up D838 *152-047 Silicon Replaceable by 1N2862 5001-10309 D838 152-0066-00 Silicon 1N3194 10310-up D839 *152-047 Silicon Replaceable by 1N2862 5001-10309 D839 152-0066-00 Silicon 1N3194 10310-up
Fuses
F601 159-005 3 Amp 3AG Slo-Blo 117v oper 50 & 60 cycle F601 159-041 1.25 Amp 3AG Slo-Blo 234v oper50 & 60 cycle F720 159-023 2 Amp 3AG Slo-Blo 5001-11449 F720 159-0021-00 2 Amp 3AG Fast-Blo 11450-11597 F720 159-0023-00 2 Amp 3AG SIo-Blo 11598-up
Inductors
L860 Use *108-285 Beam Rotator
Resistors
Resistors are fixed, composition, ± 10% unless otherwise indicated. R601 311-340 50 R601 311-0672-00 50 R602 308-142 30 R603 304-330 33 R609 302-106 10 meg 1/2 w
R610 302-104 100 k 1/2w R611 302-102 1 k 1/2 w R612 302-272 2.7 k 1/2 w R616 311-015 10k Var WW -100 VOLTS R617 308-186 80 k 1/2 w WW 1% 5001-10059 R617 308-0186-0180 k 1 w WW 1% 10060-up R618 308-226 10 k 1/2 W WW 1%
R619 302-224 220 k 1/2 w R624 302-473 47 k 1/2 w R625 302-222 2.2 k 1/2 w R626 302-184 180 k 1/2 w R627 302-102 1 k 1/2 w
Concentric with SW601. Furnished as a unit.
Ω Ω Ω Ω
3 w WW 5% 1 w
Var SCALE ILLUM 5001-12499 Var SCALE ILLUM 12500-up
7-2
(K)
Electrical Parts List—Type RM561A
Resistors (Cont'd)
Tektronix S/N Range
Ckt. No. Part No. Description
R628 308-176 4k 20 w WW 5% R632 302-102 1 k 1/2 w R633 302-473 47 k 1/2 w R635 301-302 3 k 1/2 w 5% R640 304-100 10
1 w
R642 304-100 10 R644 304-100 10 R650 309-101 330 k 1/2 w Prec 1% R651 309-162 250 k 1/2 w Prec 1% R652 302-102 1 k 1/2 w
R653 302-225 2.2 meg 1/2 w R654 302-474 470 k 1/2 w R655 302-685 6.8 meg 1/2 w R656 311-068 500 k 2 w Var. +125 VOLTS R657 302-684 680k 1/2 w R658 302-273 27 k 1/2 w R659 302-333 33k 1/2 w R663 302-102 1 k 1/2 w R664 302-102 1 k 1/2 w R666 308-176 4 k 20 w WW 5% R667 308-176 4k 20 w WW 5% R670 309-156 1.024 meg 1/2w Prec 1% R671 309-053 333 k 1/2 w Prec 1% R672 302-102 1 k 1/2 w R673 302-105 1 meg 1/2w R675 302-825 8.2 meg 1/2 w R676 311-068 500 k 2 w Var. +300 VOLTS R677 304-224 220 k 1 w R678 302-394 390k 1/2 w R679 302-333 33k 1/2 w R729 302-823 82 k 1/2 w X5160-up R730 311-068 500 k 2 w Var. -12.2 VOLTS R731 309-104 2.05 k 1/2 w Prec 1% R732 310-115 15k 1 w Prec. 1% R733 302-564 560 k 1/2 w 5001-5159 R733 301-394 390 k 1/2 w 5% 5160-up R734 302-334 330k 1/2 w R735 302-272 2.7k 1/2 W R737 302-151 1500 1/2w R744 308-231 220 0 3 w WW 5% R749 3030563-00 -56 k 1w 5% X8270-up R754 302-471 470 1/2 w R759 302-104 100 k 1/2 w R770 302-564 560k 1/2 w R781 316-470 47 0 1/4 w 5001-12499X R782 316470 470 1/4 w 5001-12499X R783 316470 47 1/4 w 5001-12499X R784 31670 470 1/4 w 5001-12499X R801 306-681 680 n 2 w R802 302-562 5.6 k 1/2 w R803 306-273 27 k 2 w
Ω Ω
1 w 1 w
(D)
7-3
Electrical Parts List-Type RMS61A
Resistors (Cont’d)
Tektronix S/N Range
Ckt. No. Part No. Description
R806 302-104 100 k 1/2 w R807 302-472 4.7 k 1/2 w R813 302-101 100 R815 302-474 470 k 1/2 w R816 302-102 1 k 1/2 w R831 302-104 100 k 1/2 w R832 302-106 10 meg 1/2 w R833 311-349 2 meg Var. INTENSITY R834 302-105 1 meg 1/2 w R835 Use *050-115 Replacement Kit 5001-5609 R835 (3) 306.565 5.6 meg 2w 5610-up
(2) 306485 6.8 meg .2w R836 302-223 22 k 1/2 w R837 302-471 470 a 1/2 w R838 301-242 2.4 k 1/2 w 5% R839 302-104 100k 1/2 w R840 301-125 1.2 meg 1/2 w 5% R841 311-042 2 meg Var. HIGH VOLTAGE R842 Use *050-147 Replacement Kit 5001-5609 R842†† (3) 306-395 3.9 mer 2w 5610-up R844 311-254 5 meg .2 w Var. FOCUS R846 304-225 2.2 meg 1 w R849 302-223 22k 1/ w R851 302-104 100 k 1/ w R852 302-273 27 k 1/ w R853 302-471 470 R854 302-105 1 meg 1/2 w R860 311-007 2 x 1 k Var. ALIGNMENT R861 302-680 68 R862 302-823 82 k 1/2 w R863 302-823 82k 1/ w R864 311-206 250 k Var. ASTIGMATISM R865 311-026 100 k Var. GEOMETRY R870 301-393 39 k 1/2 w 5% R871 311-315 20 k Var. CAL AMPL R872 301-154 150 k 1/2 w 5% R873 302-103 10 k 1/2 w R876 301-153 15 k 1/2 w 5% 5001-13129 R876 323-0306-00 15k 1/2 w Prec. 1% 13130-up R877 301-183 18k 1/2 w 5% 5001-13129 R877 323-0314-00 18.2 k 1/2 w Prec. 1% 13130-up R878 301-564 560 k 1/2 w 5% 5001-13129 R878 323-0455-00 536 k 1/2 w Prec. 1% 13130-up R879 Use 301-114 110k 1/2 w 5% 5001-13129 R879 3230389-00 110 k 1/2 w Prec. 1% 13130-up R883 305-223 22 k 2 w 5% R885 310-066 18k 1 w Prec. 1% R886 309-30 1.8 k 1/2 w Prec. 1% R887 309-072 180 1/2 w Prec 1% R88 8 3094064 20 1/2 w Prec 1%
1/2 w
1/ w
1/2 w
S/N 5610-up *050-115 may be used ††S/N 5610-up *050-147 may be used
(D)
7-4
Electrical Parts List-Type RM561A
Resistors (Cont)
R890 309-030 1.8 k 1/2 W Prec. 1% R891 309-072 180 R892 309-064 20 R898 309-178 250 R899 *308-090 1/4
1/2 w Prec. 1% 1/2 w Prec. 1% 1/2 w Prec. 1%
1 w WW 1%
Switches
Tektronix S/N Range
Ckt. No. Part No. Description
Unwired Wired SW601 311-340 POWER ON-OFF 5001-12499 SW601 311-0672-00 POWER ON-OFF 12500-up SW854 260449 Slide CRT CATHODE SELECTOR SW870 260-394 *262-515 Rotary CALIBRATOR
Thermal Cutout
TK601 260-246 Thermal Cutout 123
O
Transformers
T601 *120-280 L.V. Power T801 *120-297 H.V. Power
Transistors
Q624 151-087 Selected from 2N1131 Q734 151-040 2N1302 Q744 151-042 2N1378 Q757 Use 151-0165-00 2N3614
Electron Tubes
V609 154-291 0G3 V627 154-307 7233 V634 154-187 6DJ8 V654 154-022 6AU6 V667 154-056 6080
V674 154-022 6AU6 V800 154467 6CZ5 V814 154-046 12BH7 V822 154-051 5642 V832 154.051 5642
V859†† Use *154-449 T5032-31-1 CRT Standard Phosphor 5001-6885 V859 *154-449 T5032-31-1 CRT Standard Phosphor 6886-up V884 154-278 6BL8
Concentric with R601. Furnished as a unit. ††S/N 5001-6885 add ’050-214 kit.
(D)
7-5
Type RS61B
TYPE R561B
Values are fixed unless marked Variable.
Tektronics Serial/Model No.
Ckt. No. Part No. Eff Disc Description
Bulbs
B4 150-0047-00 Incandescent #CN8-398 B5 150-0047-00 Incandescent #CN8-398 86 150-0047-00 Incandescent #CN8-398 B7 150-0045-00 Incandescent #685 B209 150-0030-00 Neon, NE 2V
8277 150-0030-00 Neon, NE 2V B278 150-0030-00 Neon, NE 2V B279 150-0030-00 Neon, NE 2V
Capacitors
Tolerance ± 20% unless otherwise indicated. C9 290-0319-00 1300 µF Elect 150V +75%-10%
C11 293-0245-00 1.5µF Elect 10V 10% C16 281-0523-00 100 µF Cer 350V C24 281-0546-00 330 µF Cer 500V 10% C31 290-0271-00 B010100 B139999 9µF Elect 125V +20%-15% C31 290-0486-00 B140000 6.8 µF Elect 100V 10% C36 290-0398-00 10,000µF Elect 25 V +100%-10%
C42 290-0267-00 1 µF Elect 35 V C47 290-0201-00 100 µF Elect 15 V C50 281-0525-00 470 µF Cer 500 V C54 290-0318-00 650 µF Elect 250 V +75%-10% C62 283-0079-00 B010100 B129999 0.01 µF Cer 250 V C62 283-0267-00 B130000 0.01 µF Cer 500 V
C65 281-0638-00 B010100 B139999 240 pF Cer 500 V 5% C65 283-0104-00 B140000 2000 pF Cer 500 V 5% C75 290-0305-00 XB100000 B139999X 3 µF Elect 150 V
C78 290-0181-00 290 µF Elect 350 V C82 283-0079-00 B010100 B129999 0.01 µF Cer 250 V C82 283-0267-00 B130000 0.01 µF Cer 500 V C88 281-0536-00 1000 pF Cer 500 V 10%
C97 A, B 290-0089-00 3 X 20 µF Elect 350 V C102 281-0027-00 0.7-3 pF, Var Tub. C109 281-0027-00 0.7-3 pF, Var Tub. C157 *285-0758-00 0.05 µF MT 400 V 2% C203 283-0000-00 0.001 µF Cer 500 V C209 283-0057-00 0.1 µF Cer 200 V +80%-20%
C215 283-0092-00 B010100 B099999 0.03 µF Cer 200 V +80%-20% C215 285-0628-00 B100000 0.033 µF PTM 300 V C217 285-0572-00 0.1 µF PTM 200 V C221 283-0071-00 0.0068 µF Cer 5000 V C223 283-0008-00 0.1 µF Cer 500 V
7-6
Electrical Parts List-Type R561 B
Capacitors (cont)
Tektronics Serial/Model No.
Ckt. No. Part No. Eff Disc Description
C229 283-0071-00 0.0068 µF Cer 5000 V C232 283-0071-00 0.0068 µF Cer 5000 V C250 283-0071-00 0.0068 µF Cer 5000 V C260 283-0071-00 0.0068 µF Cer 5000 V C262 283-0071-00 0.0068 µF Cer 5000 V C275 283-0071-00 0.0068 µF Cer 5000 V
Semi-conductor Device, Diodes
D8A,B,C,D (4) 152-0066-00 Silicon 1N3194 D10 152-0212-00 Zener 1N936 9V, 5%, TC D19 *152-0185-00 Silicon Replaceable by 1N4152 D25 152-0333-00 Silicon High Speed and
Conductance D31 152-0066-00 Silicon 1N3194 D35A *152-0198-00 Silicon Replaceable by MR-1032A D35B *152-0198-00 Silicon Replaceable by MR-1032A D47 152-0066-00 Silicon 1N3194 D53A,B,C,D (4) 152-0066-00 Silicon 1N3194 D62 *152-0185-00 Silicon Replaceable by 1N4152 D65 *152-0107-00 Silicon Replaceable by 1N647 D68 *152-0107-00 XB040000 Silicon Replaceable by 1N647 D70 *152-0107-00 XB040000 Silicon Replaceable by 1N647 D75 152-0066-00 Silicon 1N3194 D76 152-0066-00 Silicon 1 N3194 D77A,D,C,D (4)152-0066-00 Silicon 1N3194 D80 *152-0185-00 Silicon Replaceable by 1N4152 D88 *152-0107-00 Silicon Replaceable by 1N647 D90 *152-0107-00 XBC40000 Silicon Replaceable by 1N647 D91 *152-0107-00 XB040000 Silicon Replaceable by 1N647 D95 152-0066-00 Silicon 1N3194 D151 *152-0185-00 Silicon Replaceable by 1N4152 D159 *152-0185-00 Silicon Replaceable by 1N4152 D161 *152-0185-00 Silicon Replaceable by 1N4152 D164 *152-0185-00 B010100 B019999 Silicon Replaceable by 1N4152 D164 *152-0107-00 B020000 Silicon Replaceable by 1N647 D168 152-0333-00 Silicon High Speed and
Conductance D204 *152-0107-00 Silicon Replaceable by 1N647 D212 *152-0185-00 Silicon Replaceable by 1 N4152 D214 *152-0107-00 XB100000 Silicon Replaceable by 1N647 D217 152-0333-00 Silicon High Speed and Conductance D221 152-0218-00 B010100 B029999 Silicon 10,000 V 20 mA D221 152-0408-00 B030000 Silicon 10,000 V 5 mA D260 152-0218-00 B010100 B029999 Silicon 10,000 V 20 mA D260 152-0408-00 B030000 Silicon 10,000 V 5 mA D271 152-0066-00 Silicon 1N3194 D272 152-0066-00 Silicon 1N3194 D275 152-0066-00 Silicon 1N3194
Fuses
F1 159-0026-00 3.2 A 3AG Slo-Blo F2 159-0023-00 2A 3AG Slo-Blo F9 159-0083-00 150 mA 3AG Fast-Blo
7-7
(E)
Electrical Parts List––Type R561
Connectors
Tektronics Serial/Model No.
Ckt. No. Part No. Eff Disc Description
J11 131-0148-00 24 Contact, Female J21 131-0148-00 24 Contact, Female J183 131-0274-00 BNC J255 131-0126-00 BNC
Inductor
L259 *108-0495-00 CRT Trace Rotator
Transistors
Q12 151-0190-00 Silicon 2N3904 Q14 151-0190-00 Silicon 2N3904 Q24 151-0190-00 Silicon 2N3904 Q28 *151-0136-00 Silicon Replaceable by 2N3053 Q30 151-0190-00 Silicon 2N3904 Q32 151-0149-00 Silicon 2N3441 Q38 151-0190-00 Silicon 2N3904 Q44 *151-0192-00 Silicon Replaceable by MPS-6521 Q46 *151-0192-00 Silicon Replaceable by MPS-6521 Q49 *151-0136-00 Silicon Replaceable by 2N3053
Q51 151-0165-00 Germanium 2N3614 Q58 151-0190-00 Silicon 2N3904 Q60 151-0190-00 Silicon 2N3904 Q66 *151-0253-00 B010100 B139999 Silicon Replaceable by 2N3439 066 *151-0150-00 B140000 Silicon Selected from 2N3440 Q68 *15140253-00 B010100 B139999 Silicon Replaceable by 2N3439 Q68 *151-0150-00 B140000 Silicon Selected from 2N3440
Q70 151-0190-00 Silicon 2N3904 Q74 *151-0256-00 Silicon Tek Spec Q84 151-0190-00 Silicon 2N3904 Q86 151-0190-00 Silicon 2N3904 Q88 *151-0253-00 B010100 B139999 Silicon Replaceable by 2N3439 Q88 *151-0150-00 B140000 Silicon Selected from 2N3440
Q90 *151-0253-00 B010100 B139999 Silicon Replaceable by 2N3439 Q90 ’151-0150-00 B140000 Silicon Selected from 2N3440 Q91 151-0190-00 Silicon 2N3904 Q96 *151-0256-00 Silicon Tek Spec Q151 *151-0216-00 Silicon Replaceable by MPS-6523 Q159 *151-0216-00 Silicon Replaceable by MPS-6523
Q162 151-0250-00 Silicon 2N5184 Q211 151-1005-00 Silicon FET Q214 *151-0228-00 Silicon Tek Spec Q219 151-0201-00 Silicon 2N3739
Resistors
Resistors are fixed, composition, ± 10% unless otherwise indicated. R4 311-0377-00 25, Var R7 302-0330-00 33 1/2 W R8 307-0093-00 1.2 1/2 W 5% R9 304-0333-00 33 1 W
7-8
Loading...