All TEKTRONIX instruments are warranted against
defective materials and workmanship for one year.
Any questions with respect to the warranty should
be taken up with your TEKTRQNIX Field Engineer or
representative.
All requests for repairs and replacement parts
should be directed to the TEKTRONIX Field Office
or representative in your area. This will assure you
the fastest possible service. Please include the
instrument Type Number or Part Number and Serial
Number with all requests for parts or service.
Specifications and price change privileges reserved.
Copyright
@
1976
by Tektronix, Inc., Beaverton,
Oregon. Printed in the United States of America. All
rights reserved. Contents of this publication may not
be reproduced in any form without permission of
Tektronix, Inc.
-5403-D41 Readout
Table 3-7 -D41 Miscellaneous
Table 3-8 -5403-041 Environmental
Table 3-9 -5403-D41 Physical
Electrical Parts List
Controls and Connectors, and Symbols
Adjustment Operational Checks (Back Side of Foldout)
Block Diagram
Vert Parts Location Grid and Adjustments (Back Side of F
OF
old out)
CONTENTS
(cont)
Page
Vertical Amplifier
Horiz Parts Location Grid and Adjustments (Back Side of Foldout)
Horizontal Amplifiers
High-Voltage Parts Location Grid and Adjustments (Back Side of Foldout)
2-Axis Amplifier & CRT Circuit
Storage Parts Location Grid and Adjustments (Back Side of Foldout)
Storage Circuit
Interconnect and Voltage Distribution Diagram
Mechanical Parts List
Fig. 1 Exploded View
Accessories
0
REV.
B,
MAR.
1977
INSTALLATIO PROCEDURE
Before you start
1.
Check the power supplylamplifier module rear panel markings. If the factory settings are compatible with the available
line voltage and frequency, insert the desired plug-ins. Use the bail to raise the front of the instrument.
...g
o to Operating Instructions...
2.
If a change is needed, follow these steps:
Fig.
instrumen
aged
by
V
Line
a
220
Selr
V
nominal
line
with
voltage.
0-1.
Illustration showing a portion of power supply/amplifier module's
L.V.
power supply circuit board.
a. Line Selector Block(s)
Remove the Scope-Mobile retainer blocks and their screws, then remove the bottom dust cover from the power
supply/amplifier module by turning the four slotted fasteners a quarter turn counterclockwise. This gives easy access to
the Line Selector blocks located on the LV power supply circuit board.
Reaulating Ranges for Power Transformer
Line
Selector
Block
Position
L
Line Fuse
Data
-
b.
Line Range Taps
120 Volts Nominal
1.25 A slow-blow
Regulating Ranges
220 Volts Nominal
1
0.7
A
slow-blow
Fig.
0-2.
c.
Rear Panel
3.
Replace the bottom dust cover and the two Scope-Mobile retainer blocks on the power supply/amplifier module.
4.
If necessary, change the line cord power plug to match the power source receptacle or use an adapter.
L.V. power supply circuit board.
5.
6.
7.
Primary taps area of power supplylarnplifier module's
Plug the cord into the power source.
Insert the desired plug-ins.
Use the bail to raise the front of the instrument.
Fig.
0-3.
module.
View showing rear-panel of power supplylamplifier
Operating
Instructions..
.
0-2
Operating I nstructions-D41
Position controls so that the display
and starts at the left edge of the graticule.
7.
Adjust the FOCUS control for a sharp, well-defined
display over the entire trace length.
8.
Disconnect the input signal and position the trace
vertically so that
of the graticule.
9.
If the trace
ine, see Trace Alignment Adjustment in this section.
10. Rotate the GRAT
range and notice that the graticule lines are illuminated as
the control
lines are illuminated as desired.
it
coincides with the center horizontal line
is
not parallel with the center horizontal
l
LLUM control throughout its
is
turned clockwise. Set control so graticule
is
centered vertically
Calibration Check
11. Move the trace two divisions below graticule center
and reconnect the calibrator signal to the amplifier plug-in
input connector.
12. The display should be four divisions in amplitude
with six complete cycles (five complete cycles for 50-hertz
line frequency) shown horizontally. An incorrect display
indicates that the oscilloscope mainframe or plug-ins need
to be recalibrated.
Readout
13. Turn the READOUT INTENS control clockwise
until an alpha-numeric display is visible within the top or
bottom division of the CRT (reset the FOCUS adjustment
if necessary for best definition of the readout). Change the
VoltsIDiv switch of the amplifier plug-in that is selected for
display. Notice that the readout portion of the display
is
changes as the deflection factor
change the Sec/Div switch of the time-base unit that
selected for display. Notice that the readout display for the
time-base unit changes also as the sweep rate
14. Set the time-base unit for magnified operation.
Notice that the readout display changes to indicate the
correct magnified sweep rate. If a readout-coded 1 OX probe
is
available for use with the vertical unit, install it on the
input connector of the vertical plug-in. Notice that the
deflection factor indicated by the readout
10 times when the probe is added. Return the time-base
unit to normal sweep operation and disconnect the probe.
changed. Likewise,
is
changed.
is
increased by
15. Notice that the readout for a particular plug-in
occupies
the vertical plug-in units
readout for channel
the CRT below the readout for channel 1.
a
specific location on the display area. If either of
is
a dual-trace unit, notice that the
2
appears within the lower division of
Beam Finder
16. Move the display off-screen with the vertical position control.
17. Push the BEAM FINDER button and observe that
the
display
display to screen center and release the BEAM FINDER
button.
compresses
into
the
screen
area.
External Intensity Input
18.
Connect a 5 volt, 1 kHz sine-wave or square-wave
signal to the EXT INTENSITY INPUT connector on rear
panel of D41. Also, use the signal to externally trigger the
time-base plug-in.
19. Slowly rotate the INTENSITY control counterclockwise until the trace appears to be a series of dimmed
and brightened segments. The brightened segments correspond with the tops of the square waves.
Storage Operation (Variable Persistence)
20. Disconnect the 5 volt, 1 kHz sine-wave or square-
wave signal from the EXT INTENSITY INPUT connector.
Reset the time-base plug-in triggering controls for an
internally-triggered display from the vertical plug-in.
21. After making sure the calibrator signal
connected to the amplifier plug-in input connector, set the
controls on the time-base plug-in for single-sweep operation. Press the STORE button and note that the CRT will
momentarily be flooded positive (becoming quite bright).
is
22. Press the Reset button on the time-base plug-in. One
sweep will be generated, and a display should be visible, but
will fade out within a few seconds. lncrease the VARI-
ABLE PERSIST control setting if
retention
to increase the background level for viewing lightly written
information. If there is no display visible, increase the
INTENSITY control setting and press the Reset button.
positive and erase. Press the Reset button on the time-base
plug-in to obtain another stored display on the CRT. Press
is
required. lncrease the WRITING SPEED setting
23. Press the ERASE button; the CRT should turn
a
display of longer
Reposition
is
the
still
Operating I nstructions-D41
the SAVE button; the display is stored and no other sweep
or change in position will affect the display. The SAVE
button, when pushed, locks out the sweep erase circuit and
cuts off the current to the flood gun cathode as long as the
VARIABLE PERSIST control
(display stored at the maximum time). Turning the
PERSIST (Save Time) control counterclockwise will
decrease the time a stored display
When the display is not being used, turn the PERSIST (Save
Time) control fully clockwise; the display will not be visible
on the screen,
destroyed. Turn the PE RSlST (Save Time) control counterclockwise until
instrument
Time) control
turned off without losing any stored information. Turn the
instrument on again and turn the PERSIST (Save Time)
control counterclockwise until display appears.
The
04
ence control are combined into one control. If the
PERSIST (Save Time) control has been adjusted to
view the display in the 'Save' mode and the
then taken out of the 'Save' mode, the display will
fade rapidly, since the persistence is now set for a
shorter time.
but the information has not been lost or
a
display
is
in the "SAVE" mode and the PERSIST (Save
is
fully clockwise, the instrument can be
7
Save Time control and the Variable Persist-
is
turned fully clockwise
is
retained on the CRT.
is
visible when needed. When the
NOTE
047
is
Storage Operation (Without Variable Persistence)
24. Rotate the VARIABLE PERSIST control to
fully counterclockwise position and release the SAVE
its
button to
25. Press the Reset button on the time-base plug-in. One
sweep will be generated, and a display should be visible.
Increase the WRITING SPEED setting if a brighter display
is
required. If there
INTENSITY control setting and press the Reset button.
26. Press the ERASE button and note that the stored
waveform
pressing the Reset button on the time-base plug-in.
27.
to the point where the stored readout information doesn't
spread.
This completes the description of the basic operating
procedure for the D4l. Instrument operations not
explained here, or operations that need further explanation
are discussed under General Operating Information.
off (button out) position.
is
no display visible, increase the
is
erased. Another display can be stored by
Turn READOUT I NTENS control slowly clockwise
its
GENERAL OPERATING INFORMATION
Graticule
The graticule of the D41
faceplate of the CRT to provide accurate, parallax-free
measurements. The graticule
and ten horizontal divisions. Each division
0.9
cm. In addition, each major division
minor divisions. The vertical gain and horizontal timing are
calibrated to the graticule so that accurate measurements
can be made from the graticule. The illumination of the
graticule lines can be varied with the GRAT ILLUM
control.
Intensity Control
The intensity of the display on the CRT
the INTENSITY control. This control
display
probably require readjustment for different displays or
sweep rates. Particular care should be exercised when only
spot
phosphor and cause permanent damage to the CRT if
allowed to remain too long.
is
easily visible but not overly bright.
is
displayed. A high-intensity spot may burn the CRT
is
internally marked on the
is
marked with eight vertical
is
0.9
is
divided into five
is
controlled by
is
adjusted so the
cm by
It
will
Display Focus
If a well-defined display cannot be obtained with the
FOCUS control, even
of the internal astigmatism control may be required.
To check for proper setting of the Astig control, slowly
turn the FOCUS control through the optimum setting with
a signal displayed on the CRT screen. If the Astig control
correctly
trace will come into sharpest focus
the FOCUS control.
set,
the vertical and horizontal portions of the
at
low intensity settings, adjustment
at
the same position of
Trace Alignment Adjustment
If a free-running trace
graticule lines,
a
adjustment) as follows: Position the trace to the center
horizontal line and adjust the Trace Rotation adjustment so
that the trace
set
is
parallel with the horizontal graticule lines.
is
not parallel with the horizontal
the Trace Rotation adjustment (internal
is
Operating I nstructions-D41
Beam Finder
The BEAM FINDER switch provides a means of locating
a display that overscans the viewing area either vertically or
horizontally. When the BEAM FINDER switch
the display
display intensity
overscanned display, use the following procedure:
1.
increase the vertical and horizontal deflection factors until
the display
2. Adjust the vertical and horizontal position controls
to center the display about the vertical and horizontal
centerlines.
3. Release the BEAM FINDER switch; the display
should remain within the viewing area.
is
compressed within the graticule area and the
is
increased. To locate and reposition an
Press the BEAM FINDER switch, hold
is
within the graticule area.
is
pressed,
it
in, then
Readout (Works Only With 5400-Series Plug-In
Units)
The readout system of the power supply/amplifier and
display modules allows alpha-numeric display of information on the CRT, along with the analog waveform
displays. The information displayed by the readout system
is
obtained from the plug-in units that are installed in the
plug-in compartments. The characters of the readout
display are written by the CRT beam on a time-shared basis
with the signal waveforms.
I
The READOUT
sity of only the readout portion of the display independent
of the other traces. The readout system
fully counterclockwise OFF position. This may be desirable
when the top and bottom divisions of the graticule are to
be used for waveform display, or when the trace inter-
ruptions necessary to display characters do not allow a
satisfactory waveform display to be obtained.
Option
3,
NTENS control determines the inten-
is
inoperative in the
Externally Programmed Seventh and
Eighth Readout Words
This option adds a 25-pin connector to the rear panel of
the display module, through which two ten-character
readout words can be displayed on the CRT,
see
Fig.
1-1.
Care of Storage Screen
The following precautions will prolong the useful storage
life of the CRT screen used in this instrument.
1.
Use the minimum beam intensity required to produce
a clear, well-defined display. A too-high beam intensity may
permanently damage the CRT screen, particularly if a
is
bright spot
area.
allowed to remain stationary on the display
Left Right
Vertical
Channel 1 Channel
Vertical
Horizontal Horizontal
1
A
Or
Main B or Delayed
The Readout System operates in
interrupt the waveform display to present characters. The
is
waveform display
seconds for each character that
The readout information from each plug-in
word. Up to six (eight with option
information can be displayed on the display module (a
seventh and eighth word are available when option 3 is
installed). The location at which each readout word
presented
and channel from which
area of the graticule where the readout from each plug-in
unit channel
available only with option 3). Notice that the readout from
channel
division of the graticule and the readout from channel 2
displayed directly below within the bottom division of the
graticule. Only the readout from plug-in channels that are
selected by display switches, or by the mode switches of
dual-channel plug-ins, appears in the readout display.
is
fixed and
is
1
of each plug-in unit
interrupted for only about 20 micro-
is
directly related to the plug-in unit
it
originated. Fig.
displayed (external readout programming
a
free-running mode to
is
displayed.
is
called a
3)
words of readout
1-1
shows the
is
displayed within the top
is
is
is
Left Right External External
Vertical Vertical Word Word
Channel 2 Channel 2 (Channel) 1 (Channel)
2
External readout
information when
option is installed.
Fig.
1-1.
plug-in unit and channel (and external, if Option
Location of readout on the crt identifying the originating
f
3
is installed).
2
2.
Avoid repeated use of the same area of the screen. If
a
particular display
vertical position occasionally to use other portions of the
display area.
is
being stored repeatedly, change the
Operating
obtained by applying signals between these levels. Maximum safe input voltage
range of the Z-Axis circuit
is
+
or -50 volts. Usable frequency
is
DC to two megahertz.
Instructions-D41
3.
Do not leave a stored display on the screen when
no longer needed.
4. Operate the instrument in the non-store mode unless
is
storage
required.
it
Storage Operation
The storage feature greatly increases the versatility of
the 5403-D41 Oscilloscope. The storage cathode-ray tube
allows a display to be retained for a longer period of time.
When the STORE button
a conventional oscilloscope.
When the STORE button
operates in a storage mode. Two modes of storage are
available. They are Variable Persistence, where the persistence of the CRT
control; and a SAVE mode, that allows longer retention of
the displayed information. When the SAVE button
pressed in, a lockout function prevents accidental erasure of
the stored display.
A stored display
In the SAVE mode, the erase function is disabled.
is
is
out, the instrument operates
is
pressed in, the instrument
electrically controlled by the PERSIST
is
erased by pressing the ERASE switch.
as
is
is
Time markers applied to the EXT INTENSITY INPUT
provide a direct time reference on the display. With
uncalibrated horizontal sweep or X-Y operation, the time
markers provide a means of reading time directly from the
display. However, if the markers are not time-related to the
displayed waveform, a single-sweep display should be used
(for internal sweep only) to provide a stable display.
Cali brator
The internal calibrator of the D41 provides a convenient
signal source for checking basic vertical gain and sweep
is
timing. The calibrator signal
probe compensation, as described in the probe instruction
manual. The output square-wave voltage
within
within 1%. The frequency of the square-wave signal
the power-line frequency. The signal
the probe to the loop.
I%,
and the square-wave current
also very useful for adjusting
is
is
is
obtained by clipping
Display Photography
A permanent record of the CRT display can be obtained
with an oscilloscope camera system. The CRT bezel of the
D41 provides integral mounting for a Tektronix oscilloscope camera. The instruction manuals for the Tektronix
oscilloscope cameras include complete instructions for
obtaining waveform photographs.
400 millivolts,
4 milliamperes,
is
twice
lntensity Modulation
Intensity (Z-Axis) modulation can be used to relate a
third item of electrical phenomena to the vertical (Y-Axis)
and the horizontal (X-Axis) coordinates without affecting
the waveshape of the displayed signal. The Z-Axis modu-
lating signal, applied to the EXT INTENSITY INPUT,
changes the intensity of the displayed waveform to provide
this type of display. The voltage amplitude required for
visible trace modulation depends on the setting of the
INTENSITY control. About +5volts will turn on the
display to a normal brightness level from an off level, and
about -5 volts will turn the display off from a normal
brightness level. "Gray scale" intensity modulation can be
Oscilloscope Applications
The 5400-series oscilloscope, including
display module and plug-in units, provides
measurement system. Specific applications for the individual plug-ins are described in the manuals for those units.
Refer to the Operating Instructions section of the 5403
instruction manual for basic oscilloscope applications,
including peak-to-peak AC voltage measurements, instan-
taneous DC voltage measurements, comparison measure-
ments, time duration measurements, determining
frequency, risetime measurements, and phase-difference
measurements.
its
associated
a
very flexible
Theory of Operation-D41
in a DC level across R422 equal to the peak-to-peak R370. In order to obtain optimum spot size and shape,
excursion at the anode of CR430.
both the FOCUS and Astig controls are adjusted to provide
the proper electrostatic lens configuration in the CRT.
CRT Control Circuits
In addition to the INTENSITY control discussed The Geom adjustment R365 varies the positive level on
previously, front-panel FOCUS and internal astigmatism the horizontal deflection plate shields to control the overall
controls have been incorporated for arriving
CRT display. FOCUS control R440 provides the correct permits adjustment of the DC current through
voltage for the second anode in the CRT. Proper voltage for beam-rotation coil L375
is
the third anode
obtained by adjusting Astig control horizontal graticule lines.
at
an optimum geometry of the display. The trace rotation control, R375,
to align the display with the
HORIZONTAL AMPLIFIER
The horizontal amplifier amplifies the push-pull
horizontal deflection signal from the interface circuit board
and applies
CRT.
it
to the horizontal deflection plates of the
lnput Amplifier
The horizontal signal from the interface circuit board
connected to the bases of 0200 and 021 5. Under no-signal
conditions, the bases of 0200 and 021 5 are within 150 mV
of ground. Resistive network
between the emitters of 0200 and 0215, controls the
emitter degeneration of this stage. R212 provides
of adjusting the emitter degeneration of the input amplifier
and thereby controls the gain of the horizontal amplifier,
within +lo%.
To compress an off-screen display so that
viewed on the CRJ, the BEAM FINDER reduces the
dynamic range of the input amplifier. This is done by
disconnecting CR208 in the emitter circuitry of
0200-021 5, and supplying
current setting resistors R205, R208, and R213.
Resistors R202 and R217 provide thermal compensation
for the input amplifier, while R222 provides
correcting for differential unbalance in the amplifier or
CRT.
R205-R207-R210-R212-R213,
a
means
it
may be
a
reduced current through
a
means of
Output Amplifier
Transistors 0240-0244-0250 and 0270-0274-0280 are
as
connected
amplifiers. lnput transistor 0240 (in the left output
amplifier) is
positive-going
right output
is
negative-goi
Negative feedback
output transistors 0244-0250-0274-0280 to the base of
input transistors 0240 and 0270 through feedback
networks C242-R242 and C272-R272. Variable capacitors
C242 and C272 adjust the transient response of the
feedback networks to provide good linearity at fast sweep
rates. The Zener diode-fast switching series diode,
CR242-VR240 and CR272-VR270 (across each of the
feedback networks in the output amplifier), turn on when
the sweep passes the right edge of the CRT. This action
stops the collectors of the output transistors and shunts out
the feedback networks, thus current limiting the output
amplifier. Capacitors C240, C250, and C280 are speed-up
capacitors to improve the amplifier response to fast
changes. Diodes CR246 and CR274 prevent 0244 and
0274 from going into saturation.
two separate current-driven feedback
an NPN transistor for better response to
signals, while input transistor 0270 (in the
amplifier)
signal response.
ng
is
a PNP transistor for better
is
provided from the collectors of
VERTICAL AMPLIFIER
The vertical amplifier provides the final amplification for
the vertical signal before
deflection plates of the CRT. The vertical amplifier
circuitry includes the delay line and part of the beam finder
circuit, which reduces the final drive to compress an
over-scanned display to within the viewing area of the
it
is applied to the vertical
CRT.
Delay Line
Delay line DL100 provides approximately 140 ns of
delay for the vertical signal. This allows the time-base
circuits time to initiate a sweep before the vertical signal
reaches the CRT deflection plates. This delay of the vertical
signal allows the leading edge of the signal originating the
trigger pulse to be displayed when using internal triggering.
Theory
of
Operation-D41
The delay line has a characteristic input impedance of
about 75 ohms, or about 150 ohms from side-to-side.
Amplifier
The vertical amplifier consists of a high bandpass
three-stage paraphase amplifier having an input sensitivity
of approximately 35 mV/division and a voltage gain of
about 11 5. The amplifier
of 0100 and 0125 by the input signal from the delay line.
Rl00 and R 125 terminate the delay line.
The first amplifier stage consists of 01 00, Q106, Q125,
and 0130. The gain of this stage
of the feedback resistors R 104-R 103 or R 128-R 129 and
the emitter resistor R 11 1. The networks parallel to the
emitter
delay line. R135
compensates for resistive tolerance errors and CRT
electrical center error in the vertical amplifier, and allows
the mainframe input to be standardized.
resistor compensates for the signal losses in the
acts
is
differentially driven
is
determined by the ratio
as
a
DC
centering control, which
at
the bases
STORAGE CIRCUIT
The next stage of amplification consists of 0148, 0170,
Q165, and Q172. Thermistor RTI 57 resistor R 157 varicap
CR146 and capacitor C160 between the emitters of 0148
and 0165 comprise a thermal compensation network to
correct for frequency loss with temperature changes. The
two RC networks (R 151 -Cl56 and
the emitters of 0148 and 0165, and the RCL network in
the collectors of 0148 and Q165 provide high frequency
compensation.
The final amplifier stage consists of Q180, 0188, 0182,
and 0190. R 175 provides a means of adjusting the vertical
amplifier gain within a 220% range.
Pushing the BEAM Fl NDER compresses an off-screen
display to determine
turning off Q140, when the BEAM FINDER
which reduces the standing current in the second amplifier
stage. This lowers the voltage drop across
which lowers the standing current in the final amplifier
stage. The lower final amp1 ifier stage standing current
reduces the possible scan on the CRT.
its
location. This
R
l55-Cl53-Cl55) in
is
accomplished by
is
pushed,
R
172 and R 176,
The cathode-ray tube
tube. The collector mesh is a coarse mesh, which accelerates
electrons toward the target area. The target (storage mesh)
is
a fine mesh with a highly insulative dielectric layer
deposited on it.
occurs. The flood guns cover the entire storage target with
continuous stream of low velocity electrons; these electrons
are prevented from reaching the phosphor screen unless a
display has been written on the storage mesh.
The collimation bands (electrodes) are used
which uniformly distributes flood-gun electrons over the
storage target area. In the non-store mode, the collimation control fully clockwise), or if the PERSIST control
bands have no control over the CRT.
Switching from the non-store mode to the store mode
done by pushing in the STORE pushbutton. When changing
modes from non-storage to storage operation, an erase cycle
is
generated automatically.
It
Flood-Gun Filaments
The flood-gun filament supply for the CRT consists of
VR690-VR691 and emitter follower 0690. The power to
heat the filaments is derived from the
VR690-VR691 sets the base voltage of 0690 and provides
regulation.
is
a
transmission halftone storage
is
in the dielectric layer that storage
as
-20
V power supply.
a lens
Flood-Gun Cathode
is
The flood-gun cathode
switching transistors 0675, 0680 and 0685. Two signals
control whether the flood-gun cathode
are:
a
a. The STORE switch pushed in signal turns on the
flood-gun cathode.
b. The SAVE pushbutton, S675, when depressed either
the
flOOd-gun
turned away from its fully clockwise position, allows the
flood-gun cathode to be pulsed
is
When the D41 power
microsecond 0700 turns off. This eliminates the negative
voltage that would result from the slow decay of the power
supply filter capacitors, leaving only the positive voltage
present at the base of 0675. 0675
0680 and 0685 to be off. This prevents damage to any
stored information if the save mode is being utilized, since
there is not any flood-gun cathode current available.
switched either on or off by
is
turned on; they
at
a low duty factor.
is
turned off, in less than
is
turned on causing
Flood-Gun Anode
The flood-gun anode circuit consists of operational
amplifier 0670-0672. The operational amplifier is used
(PERSIST
as
is
a
a
Theory of Operation-D41
current-to-voltage conversion amplifier. A composite erase
pulse is supplied to the base of 0670 to hold the flood-gun
anode high during the erase cycle.
Collimation Electrodes (CE)
The CE 1 and 2, and the CE 3 circuits consist of an
emitter follower, 0655 and 0600, with base-emitter reverse
voltage protection furnished by CR655 or CR600. R655,
CE 1 and R600, CE 3 adjustments control the voltage level
of the CE 1 and 2, and
voltage varies between store and non-store modes of
operation. R655, CE 1 adjustment sets the CE 1 and 2 level
in the store mode, while R654 and
and CE 2 level in the non-store mode. R655-R657 divider
allowed to float in the non-store mode.
CE
3 respectively. CE 1 and 2
VR655 sets the CE 1
Collector Mesh
Emitter follower 0595, base-emitter reverse voltage
protection diode CR595 and Zener diode VR595 make up
the collector mesh circuitry. VR595
for 0595.
sets
the base voltage
Storage Mesh
0580, 0575 and Q585 form an operational amplifier
is
that
Internal adjustments, R572, Prep Level, and R590, Op
Level, set the quiescent operating levels for the CRT in the
storage mode. R585, WRITING SPEED control, allows a
small adjustment to be made in the operate, level, which
increases the CRT's ability to store a fast waveform. During
the erase cycle the storage mesh is made a high positive for
50 ms, returned to its operate level for 10 ms, then made
slightly positive level for 440 ms. The two positive-going
erase signals (see Fig. 2-1) are supplied to the storage mesh
by the collectors of 0515 (50 ms) and 0555 (440 ms).
used as a current-to-voltage conversion amplifier.
Erase Generator
Pushing the ERASE pushbutton creates a negative pulse
at the base of Q510. R665-R500-C500
remover network for the ERASE pushbutton with R500
used to dampen any ringing. 0510 and 0515 form
monostable multivibrator; C514 and R507 are the timing
components. Coupling capacitor C528 couples the 50 ms
multivibrator to the input of the I0 ms monostable
multivibrator consisting of 0530-0535. The timing
components for 0530-0535 are C534 and R530. The
output from the 10 ms multivibrator
input of the 440 ms monostable multivibrator by C538.
0550-0555, along with timing components C554 and R545
form the 440 ms monostable multivibrator. The
is
monostable multivibrators are cascaded to provide
sequential pulse train. Diodes CR560, CR562 and CR564
make up an OR gate to supply a composite erase signal to
the flood-gun anode circuit.
The signal available from the collector of 0550 is used
to retrigger the erase generator when the ERASE
is
pushbutton
held depressed.
is
a contact bounce
is
coupled to the
Variable Persistence and Save
0625 and 0640 form a monostable multivibrator that
triggered by the output of the 10 ms pulse generator
consisting of programmable unijunction transistor 0610.
The timing components for the monostable multivibrator
are C625, R622 and R620 PERSIST control. The 10 ms
pulse generator timing components are C610, R608 and
R609. 0635 acts as a switch for R620 PERSIST control.
a
When S610 STORE pushbutton is pushed in, the 10 ms
pulse generator feeds pulses to the storage mesh to increase
the background level slightly for better storage and to the
Q625-0640 monostable multivibrator for triggering.
a
50 ms
a
is
This level controlled by R590
/
Op
Level
This distance set by R572
Prep Level and is not affected by
R590 Op Level adjustments
R585A WRITING SPEED control changes
distance Dreo level is above 0
V
remains constant.
Fig. 2-1. Erase cycle waveform as seen at the storage mesh-connection to the crt.
Prep Level
Op
level, but
Q~~~~o~~e~~o~~]
Signal from
0535 collector
A
Signal from
0555 collector
440 ms
S675 SAVE
If
pushbutton
position), then the output of
CR640
to the storage mesh. a sweep.
Q640
When
output of
is
not pushed in (out cathode. Pushing
pins
(Fig. 2-2)
is
supplied via
16A
S675 SAVE
0640
(Fig. 2-2)
S675
and
17A
of the time-base plug-in high, preventing
Theory
of
Operation-D41
pushbutton is pushed in, then the
is
supplied to the flood-gun
in forward-biases
CR650
and holds
I
I
-
Fig.
Variable from
to
1
control
2-2.
Variable Persistence
2
ms by varying
10
10
ps
ms-
and
Save circuit output waveform.
r
I
I
Service Information-D41
TABLE 3-3
D41 2-Axis Amplifier
Characteristics
External lnput
lnput Voltage
Usable Frequency Range
lnput l mpedance
Maximum Safe lnput
Characteristics
Geometry
Orthogonality
Stored Writing Rate
Performance Requirements
+5 V turns CRT beam on from off condition
-5
V turns CRT beam off from on con-
dition
I
DCto 2 MHz
Resistance: 10
Capacitance: 40 pF
Bowing or tilt < 0.1 division
5 divlps
1 divlps at 60 second viewing time
ki-2
TABLE 3-4
D41 Cathode-Ray Tube
Performance Requirements
at
15 second viewing time
Supplemental Information
Supplemental Information
Phosphor
Deflection
Acceleration Potential
Characteristics
Power Line Input
Line Voltage (RMS)
Line Frequency
Input Power
Fuse Data
Cali brator
Voltage
Current
I
P31 standard
Electrostatic
I
8.5
kV
TABLE 3-5
5403-D41 Power Supply and Calibrator
I
Performance Requirements
I
/
Nominal IOOV, llOV, 120". 200V.
1
50to400Hz
100
W
maximum at 120 VAC, 60
A
slow blow (1 20 VAC)
1.25
0.7 A slow blow (240 VAC)
/
4mAf21%
Hz
I
I
I
Supplemental Information
I
I
I
I
Frequency
I
Twice the power line frequency
I
REV.
B,
MAR.
1977
Service Information-D41
Intensity Range
Characteristics
Graticule
Scale
-
Scale Color and Type
Normal
1
TABLE
5403-D41
Performance Requirements
TABLE
D41
Miscellaneous
Performance Requirements
8
x
10 divisions with 0.9 cmldiv
3-6
Readout
3-7
Supplemental Information
Off to full brightness. Readout inoperative when READOUT
fully counterclockwise in detent position.
Top words are displayed in top major
graticule division between left and
right extreme graticule lines. Bottom
words are displayed in bottom major
graticule division between left and
right: extreme graticule lines.
Supplemental Information
l NTENS
I
Beam Finder
Characteristics
Temperature
Operating
Storage
Altitude
Operating
Storage
Vibration
Operati
Operati
ng and Nonn g
Brings trace within viewing area and in-
tensifies trace
TABLE 3-8
5403-D4 1
Performance Requirements
To 15,000 feet
To
50,000 feet
With the instrument complete and operating,
vibration frequency swept from 10 to 50 to
10
Hz
at
1 minute per sweep. Vibrate 15
minutes in each of the three major axes
0.015" total displacement. Hold 3 minutes
at
any major resonance, or if none,
Total time. 54 minutes
Environmental
at
50 Hz.
Supplemental lnformation
at
REV.
JULY
1974
Loading...
+ 46 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.