Tektronix 466, DM44, 464 Instruction Manual

466/464
OSCILLOSCOPES AND
DM44
DIGITAL MULTIMETER
'I
lektron~
COMMITTED
TO
EXCELLENCE
]
J
]
J
Tektronix, P.O. Beaverton, Oregon
070-2037-00 Product Group 40
Box
Inc.
500
97077
PLEASE CHECK FOR CHANGE INFORMATION OF THIS MANUAL.
INSTRUCTION
Serial Number
______
AT
THE REAR
MANUAL
First Printing JUL 1976 Revised FEB 1982
Copyright reserved. Contents of this publication may not be reproduced in any form permission of Tektronix, Inc.
Products of Tektronix, Inc. and its subsidiaries are covered by
pending patents.
© 1976 Tektronix, Inc. All rights
without
U.S.
and foreign patents
the
written
and/
or
E
TEKTRONIX, TEK, SCOPE-MOBILE, are registered trademarks of Tektronix, Inc.
PM
TELEOUI Tektronix
Printed in U.S.A. Specification and price change privileges are reserved.
ENT is a registered trademark of
U.K. Limited.
and~
J
J
l
l
LIST OF ILLUSTRATIONS LIST OF TABLES
BEFORE OPERATING
INTRODUCTION SAFETY INFORMATION OPERATING VOLTAGE
Line Voltage Selection Regulating Range Selection
OPTIONS
TABLE
OF
CONTENTS
Page
YEO
iv DELA v MIXED SWEEP DISPLAYS
X-Y SINGLE SWEEP DISPLAY
1 2 2 2 3
STORAGE DISPLA
STORAGE SAVE STORAGE SAVE RANDOM EVENTS REDUCED SCAN STORAGE (466 ONLY)
SWEEP DISPLA
DISPLAYS
VS
VS
Page 20 20 21 21
22
22 22 23 23
:J
]
J
l
J
___
._J
CONTROLS, CONNECTORS, AND
INDICATORS VERTICAL DISPLAY AND STORAGE DM44
TRIGGER
HORIZONTAL
REAR PANEL RIGHT SIDE PANEL
LEFT SIDE PANEL
BASIC
REV FEB 1982
_L"'
OSCILLOSCOPE DISPLAYS NORMAL MAGNIFIED SWEEP DISPLAYS
___
L-
AND POWER
SWEEP DISPLAY
----
s .. J
..
ltl2iL_J
UL.LL
DM44 DISPLAYS AND MEASUREMENTS 4 4 VOLTS 6 9 10 Low Temperature 13 High Temperature
15 TIME AND 1/TIME 16 16
18 18 19
466/464 DM44 Operators
..
----X_
RESISTANCE
TEMPERATURE Accuracy Check
APPLICATIONS AND MEASUREMENTS
PRELIMINARY
Signal Ground Input Coupling Capacitor Precharging
d--
LL.
ti-.-
24 24 26 28 28 28 31 31
32 32 32 32
L
.L
.L.
k.J,
··-
JLU
au
TABLE
APPLICATIONS AND MEASUREMENTS (cont)
OPERATOR'S ADJUSTMENTS AND CHECKS
Trace Rotation Adjustment Probe Compensation Vertical Gain Check Basic 466/464
Line Only) DM44 Timing Check (60 Hertz Line Only) External Horizontal Gain Check
PEAK-TO-PEAK VOLTAGE
MEASUREMENTS-AC
INSTANTANEOUS VOLTAGE
MEASUREMENTS-DC
ALGEBRAIC ADDITION COMMON-MODE AMPLITUDE COMPARISON
MEASUREMENTS
TIME DURATION MEASUREMENTS
FREQUENCY MEASUREMENTS RISETIME MEASUREMENTS
TIME-DIFFERENCE BETWEEN TWO
PULSES FROM DIFFERENT SOURCES
TIME COMPARISON MEASUREMENTS
PHASE DIFFERENCE MEASUREMENTS
Timing
Check (60 Hertz
REJECTION 39
OF
CONTENTS
Page 33
33 33 33
34 34 35
35
36 38
40 41 42 42
43 44 46
APPLICATIONS AND MEASUREMENTS
HIGH
RESOLUTION PHASE DIFFERENCE
MEASUREMENTS PULSE JITTER MEASUREMENTS 48 DELA
YEO
OR MIXED SWEEP MAGNIFICATION Magnified Sweep Starts Triggered Delay Sweep Magnification 50
BASIC
466/464 DELA YEO OR MIXED SWEEP TIME MEASUREMENTS TIME
DIFFERENCE BETWEEN REPETITIVE
PULSES (BASIC 466/464)
TIME
DURATION MEASUREMENTS
(BASIC 466/464)
FREQUENCY MEASUREMENTS
(BASIC 466/464)
TIME
DIFFERENCE BETWEEN TWO PULSES FROM DIFFERENT SOURCES (BASIC 466/464)
RISETIME (BASIC 466/464) MIX (BASIC 466/464)
(cont)
After
Delay
(cont)
Page
47
48 48
51
51
52
53
54 55 57
!
I
r
E
I I I
REV
ii
466/464 DM44 Operators
A AUG 1980
I
J
l
J
l
J
J J
TABLE OF CONTENTS (cont)
DM44 DELAYED OR MIXED SWEEP TIME
MEASUREMENTS
TIME
DIFFERENCE BETWEEN REPETITIVE
PULSES (DM44)
TIME
DURATION FREQUENCY MEASUREMENTS (DM44) TIME
DIFFERENCE BETWEEN TWO PULSES
FROM DIFFERENT SOURCES (DM44) RISETIME {DM44) MIX {DM44)
OPERATOR'S SPECIFICATIONS
466/464
OSCILLOSCOPE
STORAGE
Stored
Storage View VERTICAL TRIGGERING
Sensitivity
External
Level
Trigger DIFFERENTIAL
{BASIC DIFFERENTIAL
{DM44)
Writing
Control
View
466/464)
MEASUREMENTS (DM44)
Speed
Time
Trigger
Input
Range in Ext 67
TIME
MEASUREMENT
TIME
MEASUREMENT
Page 58
60 61 62
62 64 64
65 65 65 65 65 65
66 66 66
67
67
67
OPERATOR'S SPECIFICATIONS (cont)
HORIZONTAL
X-Y
CALIBRATOR Z AXIS INPUT OUTPUTS
Out
CH 1 A+
and
B+
Gates AC POWER SOURCE ENVIRONMENTAL
DM44
DIGITAL
RESISTANCE
TIME 1/TIME TEMPERATURE DC VOLTAGE
ACCESSORIES
STANDARD ACCESSORIES OPTIONAL ACCESSORIES
OPTIONS
OPTION 1 OPTION 4 OPTION 7
MULTIMETER
INCLUDED
Page 67
67 68 68 68 68 68 68 68
69 69 69 69 69 69
70 70 70
71 71 71 71
@
$_
466/464 DM44 Operators
.
d
..
£
,(
.z
iii
,£.
LIST OF
Fig. 1
Fig. 2 Vertical controls, connectors, and
Fig. 3 Display and storage controls, Fig.
Fig. 4
Fig. 5
Fig. 6 Trigger, horizontal, and
Fig. 7 Horizontal and power controls,
Fig. 8 Rear panel controls, connectors, and Fig.
Fig. SA Fig. 9
Fig. 10 Resistance. Fig. Fig. 12 Fig.
Fig. 14 Basic 466/464 timing check. Fig. 15 Fig. 16
Regulating range selector and line
fuse.
indicators. 4
connectors, and indicators.
OM
44 controls, connectors, and
indicators. 8
Trigger controls, connectors, and
indicators. 10
controls, connectors, and indicators.
connectors, and indicators.
indicators.
Left
side panel controls 17
Tilted operation.
11
Volts. Temperature Probe compensation.
13
DM44 timing Peak-to-peak voltage
check
of
power
a waveform.
ILLUSTRATIONS
Page
Fig. 17
3
6 Fig.
12
14
16
18 25 27 29
33
34 34
36
Fig. 18 Algebraic addition. Fig. 19 Fig.
Fig. Fig. 24 High-resolution phase difference.
Fig. Fig. Fig.
Fig.
Fig. 30 Risetime. Fig.
Fig. 32
Fig. 33
Fig. 34
Instantaneous voltage measurement.
Common-mode rejection. Time duration.
20 21
Risetime. Time difference between
22
from different sources.
Phase difference.
23
Pulse jitter.
25 26
Delayed sweep magnification.
27
Time difference between repetitive
pulses.
28
Time duration.
29
Time difference between
from different sources.
31
Time difference between repetitive
pulses.
Time duration and frequency
measurements.
Time difference between
from different sources.
Risetime.
two
two
two
pulses
pulses
pulses
Page 37 38
39 41 43
44 46 47 48 49
52
53
54 56
60
61
63
64
m
r
E
E
E
I
I
iv
466/464 DM44 Operators
REV FEB
1982
..
l
J
]
]
J
]
l
TABLE TABLE TABLE
1 Regulating Ranges. 2 Resistance Ranges. 3 Volts Ranges.
LIST OF TABLES
Page
3 TABLE 4 Centigrade to Fahrenheit Conversion 24
26
TABLE 5 DM44 Delayed Sweep Displays
Page
30
58
J
l
@
466/464 DM44 Operators
v
. JI.<•··
THE ACCESSORY POUCH MAIN
ATTACHED SIGNED AND TION
THE
TO
TO
AFFORD
IN
TRANSIT.
ADJUSTABLE TO OFFER TIONS, SUCH AS THOSE
TO THE OSCILLOSCOPE.
KEEP ACCESSORIES
THE OSCILLOSCOPE SOME PROTEC-
CARRYING
CONVENIENT
ILLUSTRATED.
ILLUSTRATED
READILY
HANDLE
STABLE,
OPERATING
SHOULD RE-
IT
IS
DE-
AVAILABLE
IS
DESIGNED
POSI
2037-16
STORAGE OR OPERATION OF THE OSCILLOSCOPE ON
NOT
-
,
ITS FOUR
STABLE
REAR FEET IS
OPERATING POSITIONS
DESIRABLE
ARE
. THE MOST
ILLUSTRATED.
2037-18
466/464 DM44
Operators
@
J
J
]
J
J
l
J
INTRODUCTION
The
Tektronix oscilloscopes. for
displays divisions/microsecond The
464
storage
mode
and
the
cond
in
the identical. full-stored minutes
provides volts/division. stable vertical system
0.05 features magnifier nanoseconds/division. wide power
Storage
using
The
dual-channel
calibrated
triggering
deflection
provides
microsecond/division
for
variation
consumption
466 and 464 are
The
466
with
system does
writing
Fast mode.
viewtime is
display
accurate
extends the
intensity-extending
reduced
deflection
The
sweep
over
system.
calibrated sweep rates
of
line voltages and frequencies.
BEFORE OPERATING
dual-channel
storage
a
(in the Fast-Reduced Scan mode).
speed is
intensity
dc-to-100
the
full
relative-time measurements. A X10
The
is
about
system
writing
not
up
to
The
remaining
greater
in the Save mode.
factors
trigger
bandwidth
The
along
calibrated
instrument
100 watts.
provides
speed
have the reduced scan
110
divisions/microse-
than
to
MHz
vertical system
from 5 millivolts
circuits
capabilities
horizontal
from
with
delayed sweep
sweep rate
operates over a
portable
storage
up
to
3000
features are
15 seconds at
more than 6
to
are capable
of
deflection
0.5
second
to
Maximum
of
the
to
The
Tektronix
ohm
to 20 megohms, 0 de (using the temperature probe) measurement is displayed the
oscilloscope operates
The digital provide a digital readout points ment 1/TIME frequency.
5
phase power source with conductors potential. Operation from current-carrying
5
ground is over-current
on the oscilloscope display.
points are displayed on the
The instrument is designed
(such as phase-to-phase on a
not
recommended, since
multimeter
function
SAFETY INFORMATION
(the Neutral
(fuse)
DM44
Digital
volt
on a 3-1/2 normally.
and of
can
provide
Conductor)
conductors
protection
Multimeter
to
1200
-55°C
oscilloscope
the
time
crt direct
to
operate
one
of
power
are live
only
the
within
measures 0
de
volts(+ to
digital
between any
Both
at
the
measurement
the
current-carrying
at
sources where both
with
three-wire
Line
the
or-),
+150°C.
display
time
same time.
from a single-
ground
Conductor
instrument.
while
combine
measure-
(earth)
respect
system)
or
The
to
two
The
of
to
has
,,
l
@
UL
..
_. l .......
JLLJLk
"~
.....
J.
466/464 DM44 Operators
li
.fa.
..
.
Li
. L
..
Sit.
1
The
instrument three-terminal polarized plug source and safety-earth. The the plug electric-shock protection, insert this plug outlet
is
directly
with a safety-earth
has a
three-wire
connected
contact.
power
for
connection
ground
to
(earth) terminal
the
instrument frame. For
cord with a
to the power
only
in a mating
panel)
to
indicate the 230 volt adapter may be required sure to use the
of
selected (see Table 1).
correct
correct
line fuse
nominal
for
the
for
voltage. A 115-to-
line-cord
the line voltage
plug. Be
.
OPERATING VOLTAGE
This
instrument
the line voltage
selector
range or
if
the
wrong
Line Voltage Selection
This
instrument operates 230-volt nominal line voltage source, 48 to 440 hertz. To convert move the Line Voltage Selector
2
from one nominal line voltage range
may
selector
set
for
line fuse
be
damaged
the
wrong
is
switch
used.
from
switch
if
operated with
or
the regulating
applied
either
line voltage
a 115-volt
(located on side
or
to
the other,
466/464 DM44 Operators
Regulating Range Selection
The
Regulating Range the rear panel) is set shown
in Table
protection.
To
change
1.
Disconnect
2.
Loosen
on
the
selector assembly; then
a
3.
Pull Select a range average line voltage. Slide the bar and plug it in. Push the
the
the
out
the range
1.
It also
regulating range:
the
two
from
Selector
for
one
contains
instrument
captive screws that
selector
Table 1
cover
assembly (located on
of
the line voltage ranges
the line fuse
from the
pull
to
switch
which
is centered
to
the desired
on and
tighten
for
overload
power
source.
hold
the cover
remove the cover.
bar (see Fig. 1
about
the
position
the screws.
@
I
).
TABLE 1
Regulating Ranges
Regulating Range
Range Selector 115-Volts 230-Volts
Switch Position
LO
(Switch bar in lower holes)
Nominal Nominal
99 to
121
volts
198 to 242 volts
@
Fig. 1. Regulating range selector & line fuse.
466/464
1737·01
M (Switch bar in
HI (Switch bar in
to meet particu lar slot indicates a
DM44
Operators
middle holes)
upper holes)
Fuse Size
Options
are available to
(see Item 60, C
modified
104 to 126
108 to 132
1.5
A
3AG
Fast-blow
OPTIONS
alter
applications
ontrol
s,
Connectors
oscilloscope
volts
volts
207 to 253 volts
216 to 264 volts
0.75 A 3AG Fast-blow
oscilloscope
. A
number
, and Indi
.
performance
in
either MOD
cator
s)
3
CONTROLS, CONNECTORS AND
1.
CH 1 and
deflection
the
calibrated detent
in
2.
VOL
for
each channel, located beneath
VOL
TS/DIV
indicate the scale-switching connector
3.
VAR-Provides deflection VOL
TS/DIV deflection position).
CH 2 VOLTS/DIV-Selects
factor
in a 1-2-5 sequence (VAR
TS/DIV
knob. One
correct
connector
lights
the
factors between the calibrated settings
switch, and extends the
to
at least 12.5 volts per
INDICATORS
VERTICAL
control
for
the indicated
Readout-Consists
or
the
other
deflection
X1
continuously
is
lamp.
factor
used. A
deflection
of
two
the
lamp
will
when a
probe
variable
maximum
division
the vertical
must factor).
small lamps
skirt
of
each
light
up
probe
with
without
uncalibrated
this
of
vertical
(5 volt
be
to
the
I
I
I
a
E
I
Fig.
2.
Vertical controls, connectors, and indicators.
4
4.
UNCAL
VOL
TS/DIV vertical
X-Y mode, the CH 2 POSITION axis (vertically) and the positions on
466/464 DM44 Operators
deflection
5.
POSITION-Positions
Lamp-Indicates
control
is
out
of
factor
is uncalibrated.
Horizontal
the
X-axis (horizontally).
when
the calibrated
the
display
control
positions
POSITION
the
VAR
detent
and the
vertically. In the
on the Y
control
@
I
I
I
I
J
J
l l
J
l
J
6.
CH
1 OR X application amplifier. connected tal
deflection
connector
7. signal position, amplifier.
the
In disconnected allow position, the
input
8.
tical
CH
of In to
provides vertical deflection.
AC-GND-DC-Selects
to
the signals are capacitively The
GND
position,
the
input
all
components
amplifier.
VERT
MODE-Selects
amplifier
1:
Channel 1 only
and
CH
2 OR
Y-lnput external signals to the inputs the
X-Y
mode
of
operation, the signal
the
CH 1 OR X
and the signal connected
input
of
de
component
the
from
the
coupling
system.
connector
the method used
the vertical amplifier. In the
of
the
input
of
input
connector
capacitor
of
the
input
mode
is displayed.
provides to
coupled
input
signal is blocked.
the vertical
and
to
precharge. In the DC
signal are passed to
of
operation
connectors
of
the vertical
the CH 2 OR Y
to
to the vertical
amplifier
grounded
for
horizon-
couple
AC
for
ver-
is
to
ALT: Provides dual-trace display
is
channels. Display of
each sweep. Useful at sweep rates faster than
microseconds/division.
ADD: Signals applied to the connectors are algebraically added, and the algebraic sum
is
displayed on the crt. 2 allows the display to be CH 1 plus CH 2 CH
2.
a
Useful undesired signal
CHOP: Provides dual-trace both channels. Display is switched between channels repetition rate rates slower than about 50 when a dual-trace, single-sweep
2:
Channel 2
CH
in X-Y operation.
9.
20
MHz BW/TRIG limits the bandwidth approximately causes the signal applied to A displayed on the crt.
10.
INVERT-Channel 2 display
INVERT (button in) position.
switched between channels at the end
The
for
common-mode
or
for
de offset.
of
approximately 250 kHz. Useful at sweep
only
is
displayed.
VIEW-Dual-purpose
of
the vertical
20
MHz when pulled,
of
the signals
CH
1 and CH 2
INVERT switch in Channel
or
rejection
display
microseconds/division,
display
Trigger
to
of
the
is required.
It
must be selected
amplifier
or
when pressed,
Generator
is
inverted in the
of
both
about
input
CH 1
minus
remove an
signals
switch
system
to
50
of
at
or
that
to
be
a
l
@
1 I
XL.L
..
ULUAJlJ!
LLlllqi
466/464 DM44 Operators
.J$£lUJJtJ
..
faL.
LJ
5
j
..
.iJ
£
....
CJ
..
L.JLJH
..
ktl.L
....
24.
A
L
DISPLAY AND STORAGE
I
I
I
m
I
I
Fig.
3.
Display and storage controls,
6
466/464 DM44 Operators
connectors,
and
indicators.
@
I
J
]
l
11a.
Internal amplitude left-hand
11b. Reduced Scan Graticule (466 operating set
by
the VOL referenced measurements should be made
operating
BEAM
12.
the
graticule
applied
signals and provides a visible viewing level.
Graticule-Eliminates
and measurement points are indicated at the
graticule
in the reduced scan
to
in the reduced scan mode.
edge.
mode
TS/DIV
the smaller divisions
FINDER-Compresses
area independently
and
TIME/DIV within
parallax. Risetime,
Only)-When
the deflection factors,
switches, are
of
this
graticule. All
this
graticule
the display
of
display position
to
within
when
19. REDUCED SCAN
reduced scan mode
20. STORAGE
crt
in the FAST and
21.
accidental erasure
22. SAVE
mode.
or
23.
as
a conventional oscilloscope in the NON STORE mode.
is
LEVEL-Varies
VAR
SAVE-Provides
of
INTEN-Varies
NON
STORE-Allows
LAMP-Indicates
in operation. 466 only.
the
writing
PERS storage modes.
longer viewing time. It prevents
the stored display.
the intensity
operation
of
that the
rate
of
the SAVE
the instrument
of
the
J
J
l
J
l
13.
INTEN-Controls FOCUS-Adjusts
14.
15.
SCALE
16.
ASTIG-Used control readjustment
the
Reduces display
an increase in
@
to
TRACE
17. horizontal
18. REDUCED SCAN
ILLUM-Controls
obtain a well-defined display. It does
in normal use.
ROTATION-Adjusts
graticule
to
writing
brightness
for
optimum
in
conjunction
lines.
(PULL
the small reduced scan graticule with
rate. 466
of
the
display definition.
graticule
with the FOCUS
trace
- INTEN
only.
crt
display.
illumination.
not
require
to
align with
knob)-
466/464 DM44 Operators 7
24. VAR
stored display.
25. disables the TRIG MODE switch. It automatically se­quences
in the full clockwise detent (manual). determined by the VIEW TIME sweep in a single-sweep mode.
26. VIEW sistence) varies the time between erase cycles in FAST mode.
27. SAVE
PERS-Permits
FAST-Used
an
erase cycle, unless the VIEW TIME
of
the stored display in the VAR PERS mode. It
ERASE-Erases
mode when ERASE
for fast-writing displays. FAST mode
TIME-Varies
variable retention of. the
It waits a time period
control;
the retention time (per-
the stored display, except in
is
disabled.
then resets the
control
is
Fig.
4.
DM44 controls, connectors, and indicators.
I I
I
I
I
I I
I
8
466/464 DM44 Operators
@
I
J
J
l
J J
DM44
28.
Input
Connectors­(red) and only.
29.
temperature
30. matic polarity display ed
by operation the
31. (1200 volts is the maximum safe mode);
32. 1/TIME,
COM
(black) inputs
Probe
Readout-Negative
for
indicates overrange. The decimal
the
TIME
RANGE-Selects
from
FUNCTION-Selects
Connector-Input
probe.
negative de voltage and temperature with
indication
FUNCTION
and
or
or
TIME
for
by
the oscilloscope A
1/TIME
200 ohms
functions
Two
for
polarity
positive measurement. A
and RANGE
modes.
from
.2 V to
to
20 megohms in 6 ranges.
VOL TS, OHMS, TEMP C),
of
the DM44.
banana jacks provide +
voltage and resistance
connector
indication
point
controls
TIME/DIV
1.2 kV de in 5 ranges
input
in the 1.2 kV de
for
is
blinking
is
position-
for
multimeter
switch in
the
auto-
no
In the 1 /TIME function the Readout and Scale Factor
of
Lamps indicate the number of
time. One lamp
per
ms
(1/ms lamp) lamp lit indicates intervals per second. If the duration one event Factor Lamps indicate frequency. The
dicates kHz, the 1/µs lamp indicates MHz, and
indicates Hz.
34. ll.
TIME-Used
TIME
POSITION control in the
tions. The ll. TIME
measurement point. The DELAY moves both the reference measurement point. With the time-measurement
the left
negative time difference.
of
or
the other will
or
is
being measured, the Readout and Scale
the reference
measured intervals per unit
light
to
indicate intervals
intervals
in
conjunction
control
point
per
µs
(1/µs lamp). No
1/ms
with the DELAY
TIME
and
moves
TIME
point
the Readout indicates a
only
POSITION
and the time-
lamp in-
no
1/TIME
the time-
point
lamp lit
func-
control
of
to
l
33. Scale Factor the
indicate
light
TIME
zones on the
to
In Lamps intensified will seconds.
@
Lamps-
function,
the time difference between the
indicate ms
Two
lamps.
the
Readout and Scale Factor
crt
display. One lamp or
µs.
No
lamp lit indicates
..
t
You
two
or
the
other
466/464 DM44 Operators
TIME POSITION
point.
located in the Maintenance section
Service manual.
NOTE
can
modify
The
your
instrument
control
procedure for making this
to make the DELAY
move
only
modification
the reference
is
of
the DM44
9
TRIGGER I
Fig.
5.
Trigger
controls,
connectors,
and indicators.
35. TRIG
operation
AUTO: Sweep
In
the absence trigger sweep free runs and provides a
NORM: Sweep
In the absence
trace. When the
NORM.
SINGL Sweep operates in the single sweep mode. sweep until
the SINGL SWP when the signal to amplitude, shape tional display. It can also nonrepetitive signal.
36. READY upon receipt single-sweep display.
MODE-Determines
for
A Sweep. I
is
initiated by the applied
of
an adequate
repetition rate
of
SWP: When this
is
displayed,
Lamp-Indicates
of
an adequate
is
less than
is
initiated by the applied
an
adequate
trigger
be
rate is
pushbutton
further
or
sweeps cannot be presented
button
displayed is
time causing an unstable conven-
the mode of
trigger
bright
trigger
too
is
again pushed. It
not
be
used
A Sweep is
trigger
signal,
about
reference trace.
signal, there is
low
is
repetitive
to
signal, will present a
trigger
trigger
signal.
or
if the
20 hertz, the
trigger
for
pushed, the A
photograph
signal.
AUTO use
After
a single
is
or
varies in
"armed"
useful
and,
I
no
I
a
I
I
10
466/464 DM44 Operators
@
I
J
J
]
l
J
J
l
37.
TRIG
Lamp-Indicates
and
will
produce
the
trigger without a display external triggers).
TRIG
38. A
of
time
between sweeps. signals (such as clockwise of
B sweep
rate
for
delayed-sweep presentations and rate signals. Use the A possible control.
signals
trigger 30 Hz are attenuated.
of about stable complex
display
39.
COUPLING-Determines
to
AC: Signals are capacitively
generator. De is rejected and signals
LF REJ: Signals are capacitively coupled
the
trigger
50 kHz are attenuated. It
display
waveform.
a stable display. It is useful
circuits
position
to
provide the fastest possible sweep repetition
the
when a
on
the
HOLDOFF-Provides
complex (BENDS
before using the A TRIG
trigger
circuit.
of
the
generator
De is rejected and signals
high-frequency
that
A Sweep is
trigger
crt
(for
example, when using
Allows
triggering
digital
A), A sweep is reset at the end
trigger
words). In the
controls
method used
circuit.
coupled
is
useful
triggered
for
setting up
signal is available
continuous
to
the
for
components
control
on
aperiodic
fully
low-repetition
for
the best
HOLDOFF
to
couple
input
of
the
below
about
to
the
input
below
providing
of
REJ:
HF
of
the trigger circuit. De is rejected and signals below about useful for providing a stable components of a complex waveform.
DC: All components the input stable display signals, except the mode
40.
starts the sweep.
+:
portion
-:
portion
Correct SLOPE setting display when
41.
signal at which the sweep
a
adjusted
a
COUPLING and SLOPE have been selected.
Signals are capacitively coupled
30
Hz and above
of
the trigger
of
with the trigger SOURCE
SLOPE-Selects
Sweep can be triggered
of
a trigger signal.
Sweep can
of
a trigger signal.
only a portion
LEVEL-Selects
for
the desired display after
about
50 kHz are attenuated. It
display
of a trigger
circuit.
low-frequency
combination
the slope
be
triggered
the
It is useful
is
important
of a cycle
amplitude
is
to
the
input
of
the
low-frequency
signal are coupled to
for
providing a
or
low-repetition rate
of
the
ALT
(dual trace)
switch
from
from
in NORM.
of
the
trigger
the positive-going
the negative-going
is being displayed.
point
triggered. It is usually
trigger
signal that
in obtaining a
on the
trigger
SOURCE,
is
l
@
466/464 DM44 Operators
11
42.
SOURCE-Determines
to
the
signal coupled
NORM:
not
indicate time relationship between CH 1 and CH 2 signals. However, stable signals usually can be obtained ALT, SOURCE Carefully
CH used as a time-related.
CH used as a time-related.
Trigger
adjust LEVEL
1: A sample
trigger
2:
A sample trigger
input
source is displayed signal(s). It does
to
NORM and
for
of
the signal available in Channel 1
signal. CH 2 signal
of
the signal available in Channel 2
signal. CH 1 signal
the source
of
the
trigger
triggering
a stable display.
of
by
setting VERT MODE to
COUPLING
is
unstable if it
is
unstable if it
of
the trigger
circuit.
non-time-related
to
LF REJ.
is
not
is
not
I
I
is
is
I
LINE (A
line
frequency
the
input the line frequency stable display waveform.
EXT: Signals connected connectors be time-related It is useful when the internal signal undesired signals that is useful when operating in
Trigger
signal is time-related
are used
circuit
only): A sample
is used
of a line-frequency
as a trigger
(multiple
or
when it
to
the External
for
triggering. External signals must
to
the displayed signal
could
cause unstable triggering. It
CHOP
is
12
of
the power-
signal. It is useful when
or
submultiple) to
desirable to provide a
component
is
too
mode.
in a complex
Trigger
for
a stable display.
small
or
contains
466/464 DM44 Operators
Input
Fig.
6.
Trigger,
indicators.
horizontal, and
power
controls,
connectors
and
@
I
I
I
J
l
J
J J
l
EXT
-;-
10 (A
Trigger
signal is attenuated
STARTS Sweep the
DELAY-TIME differential obtain with
a DM44
measurements
43. External Trigger
ternal
AFTER DELAY
runs
immediately
time
measurements
valid measurements. On
you
when using the
trigger
signals.
HORIZONTAL
44a. A AND B
TIME/DIV
rate
selects TIME TIME/DIV
sweep rate
operation
detent
switches
the
horizontal
switch
of
the
A Sweep
the
POSITION
switch
only. A VAR
for
calibrated A Sweep rates. When
are
.(clear plastic skirt) selects the sweep
basic delay
control} (pull
for
the
fully
(X-axis}
circuit
by a factor
after the delay
POSITION
must use this mode
Inputs-Input
only}: External
of
10.
(8
Trigger
control.
you
must use this
instruments
TIME
or
AND POWER
TIME/DIV
circuit
time
out
B Sweep
control
counterclockwise
display
AND DELAY
for
A sweep
(to be used
for
delayed sweep operation. B
and rotate
to
circuit
must be in
to
is the
CH 1 input.
trigger
circuit
time
1/TIME connectors
only
unlock}
for
the X-Y position,
only}: B
selected
When making
mode
equipped
to
obtain valid
functions.
for
TIME-A
operation and
by
the
DELAY-
selects the
delayed sweep
the
calibrated
both
TIME/DIV
by
to
ex-
44b. DM44 A AND B
The controls operate in the same manner The A TIME/DIV switch also
point
decimal or
1/TIME MODE.
45.
FINE/POSITION-Positions
ly
for
A sweep and B Sweep
mode.
46. X10
factor nanosecond/division. The magnified sweep
division
of
the center graticule line}.
47.
VAR-Provides between the calibrated settings It extends the slowest A Sweep rate seconds/division. The A Sweep rate is calibrated when the control be in the detent position measurements. On instruments equipped VAR control must accurate measurements in the tions.
48. UNCAL
uncalibrated (VAR
location when the DM44 is in the
MAG-Increases
of
10.
It extends fastest sweep rate
of
the unmagnified
is
set
fully
Lamp-Indicates
TIME/DIV
continuously
clockwise
to
make accurate differential
be
in the detent
control
AND DELAY
as
controls
or
displayed sweep rate
display
to
out
the
TIME
the display
on
the
X-axis in the X-Y
(0.5 division
variable sweep rates
of
the A TIME/DIV
to
at least 1.25
the calibrated detent.
with
position
TIME
and
1/TIME
the A Sweep rate
of
the calibrated detent}.
TIME-
37a above.
lamps and
TIME
horizontal-
by
to
is
the
center
either
side
switch.
Must
time
a DM44 the
to
make
func-
a 1
is
J
@
J . .
....
& .
!1
L .
466/464 DM44 Operators
..
!LA
l&JJL
Xi!S!.~M!Jl
......
L.J1!kJL.U3
J.L3i
LJ
,.<JJ
.a
SL
.JJL!
LLB
13
a·;
49. X10
is
on.
MAG
Lamp-Indicates
that the X10
magnifier
I
Fig.
7.
Horizontal and
indicators.
power
controls, connectors, and
50a. DELAY sweep delay by the A
same manner as 41a. Also, when the DM44 is in the or with the a TIME control. The DELAY TIME POSITION control measurement point. The a TIME time-measurement point. With the time-measurement point indicates a negative time difference.
TIME/DIV
SOb. DM44 DELAY TIME
1/TIME
moves
to the left
You can
TIME
POSITION
point. The
located in the DM44 Service manual.
TIME
to
more than 10 times the delay time indicated
function, this
both
modify
procedure
Operating
POSITION-Provides
switch.
POSITION-Operates
control
the reference
of
the re.terence
NOTE
your
instrument
control
move
for
making
Information
operates in
point
control
moves
point
to make the only
the reference
this
modification
section
variable
in the
TIME
conjunction
and the
the Readout
time-
only
DELAY
is
of
the
the
I I
I I I I
14
466/464 DM44 Operators
@
I
J
l
J
51.
POWER-Turns
52.
HORIZ
tion
for
A: Horizontal deflection provided
sweep rate determined
switch. B sweep is inoperative.
MIX: at a rate set the sweep at a rate set amounts
rates are determined POSITION
DISPLAY-Determines
the horizontal deflection system.
The
first part
by
the A
of
the display allocated
control.
instrument
by
the setting
of
the horizontal sweep is displayed
TIME/DIV
by
the B
by
the setting
power
the
by
of
switch
and the latter part
TIME/DIV
to
each
of
the DELAY-TIME
on
and off.
mode
of
opera-
A Sweep at a
the A
TIME/DIV
switch. Relative
of
the
two
sweep
REAR PANEL
53. A
+GATE-Provides pulse coincident with the A Sweep time, which can be used to trigger the signal source (with TRIG MODE switch set to AUTO).
54. B
+GATE-Provides pulse coincident with the B Sweep time, used to trigger the signal source after a selected delay
of
time, providing that A Sweep is triggered internally.
55.
CH
1 VERT SIGNAL
the signal applied to the CH 1
a positive-going rectangular
a positive-going rectangular
OUT-Provides
input
connector.
which
a sample
can
be
of
l
J
l
J
l
A INTEN: The sweep rate is determined by the A TIME/DIV display the B provides a check Sweep (delayed sweep) with respect
(delaying sweep).
B DL YD: The sweep rate is determined by the B TIME/DIV setting
DELAY-TIME
REV FEB 1982
switch. An intensified
during
TIME/DIV
of
the
B Sweep time,
switch setting.
of
the
switch
the DELAY
with the delay time determined
TIME
POSITION control.
duration
(A
TIME/DIV)
portion which
appears
is
about
This
switch
and position
to
the A Sweep
switch and the
10 times
position
of
466/464 DM44 Operators
on
the B
by
the
the
56. EXT
crt
display. Does not affect display waveshape. Signals with fast rise and fall provide the most change. Signals must be time-related stable display. Useful
calibrated modes
57.
square-wave voltage compensate voltage probes and check vertical gain, current probes and oscilloscope operation.
intended to verify time-base calibration.
Z-AXIS-Permits
of
operation.
CALIBRATOR-A
output
intensity modulation
abrupt
to
the
for
adding time markers in
combination
that permits the operator
current
intensity
display
It
of
the
for
un-
loop/
is
not
15
a
to
58. Line Fuse
Holder-Contains regulating range selector. See Table tion .
the line fuse and the
1
for
change
informa-
Fig.
8.
Rear panel controls,
connectors,
1737·08
and indicators.
59. Regulating Range
regulating range. See Table 1
60.
Line
Cord-Makes
Selector-Shown
for
change information .
the
connection
in Medium
between the oscilloscope and the power source. The cord may conveniently stored by
wrapping
it around the feet on the
rear panel.
61. MOD
the instrument contains an
SLOTS-A
number
option
in either slot indicates
or
other modifications.
RIGHT SIDE PANEL
62. LINE VOLTAGE SELECTOR
either 115 V
or
230 V nominal line voltage.
SWITCH-Selects
LEFT SIDE PANEL
63. Vertical Gain Controls
side
panel)-Screwdriver
each vertical channel.
64. Variable Balance Controls
left
side
panel)-Screwdriver
of
each vertical channel.
(accessible
through
adjustments to set the gain
(accessible
through
adjustments to set balance
be
left
of
16
466/464 DM44 Operators
REV FEB 1982
I
J J
l
J J
l l
Figure
BA.
Left
side panel controls.
(2906)2037-01
:I
REV FEB 1982
466/464 DM44 Operators
17
BASIC OSCILLOSCOPE
DISPLAYS
These
most
instructions
commonly
NORMAL SWEEP DISPLAY
1. Set the
VERT MODE VOLTS
VOL TS/DIV VAR AC-GND-DC Vertical POSITION
20
MHz
INVERT
TIME A
TIME HORIZ X10
POSITION
controls
/DIV
BW
/DIV Switches
/DIV VAR
DISPLAY A
MAG
permit
used basic
as
follows
Vertical
Horizontal
the
operator
displays
:
CH
1
Positi
amplitude
be applied Calibrated AC Midrange Not
not Button
Locked Calibrated
Off
(button
Midrange
to obtain the
.
on
determined by
detent
limited
(yellow band
visible)
out
together
detent
out)
of
signal to
at 1
ms
THE
ADJUSTABLE
TO
OFFER
SITIONS. THE
HANDLE
OPERATING
PROVIDES
CARRYING
CONVENIENT,
NEXT-TO-LAST
THE
MOST
POSITION.
Fig. 9. Tilted operation.
HANDLE
STABLE
LOCKING
STABLE, NEAR-UPRIGHT,
IS
DESIGNED
OPERATING
POSITION
OF
1737·15
PO-
THE
18
466/464 DM44 Operators
@
]
l
Storage
REDUCED SCAN (PULL) STORAGE SAVEINTEN SAVE NON VIEW
INTENSITY
LEVEL
STORE
TIME
Display
Trigger
(Both A and B if applicable)
Pushed in (466 only) NORM
ccw
Button Button NORM
Fully
out
in
counterclockwise
pushbutton and adjust the CH 1 VOL reduce the vertical display size. display with the vertical and trols; release the BEAM FINDER pushbutton.
4.
Set the CH 1 VOL POSITION area.
5.
display.
6. POSITION control to locate the area. Then adjust the FOCUS
control
Adjust the A
Set the A
to locate the
Trigger
TIME/DIV
horizontal
TS/DIV
display
LEVEL
switch and the horizontal
display
control
TS/DIV
Center
the compressed
POSITION
switch and vertical
within
control
within
as
needed.
switch
con-
the
display
for
a stable
the display
to
J J
J
J
SLOPE LEVEL SOURCE COUPLING TRIG
MODE
A
TRIG
2.
Pull the POWER
the CH 1
display INTENSITY
@
3.
Adjust
brightness. If
input
(A only)
HOLDOFF
connector.
the
INTENSITY
control
at midrange, press the BEAM FINDER
switch
the
display is
+
0
NORM AC AUTO NORM
(on).
control
Connect
for
not
visible with the
1.
2. area to
(0.5 division on each side
the signal
the desired
466/464 DM44 Operators
to
line). setting.
3. horizontal POSITION
the
magnified display. Divide the
to
determine the magnified sweep rate.
MAGNIFIED SWEEP DISPLAYS
Obtain a Normal Sweep Display.
Adjust the horizontal POSITION
be
magnified to
It may
be
Push the
necessary
X10
within
the center graticule division
of
the
to
change the
MAG
switch (on) and adjust the
control
for
control
center
vertical graticule
TIME/DIV
precise
TIME/DIV
to move the
positioning
setting
switch
of
by
10
19
DELAYED SWEEP DISPLAY
1.
Obtain a Normal Sweep Display.
2.
Set the
B
Trigger
3.
Pull
the intensified zone
INTENSITY
zone
distinguishable
instrument switch delayed sweep. Dual delayed
6.
step
HORIZ
SOURCE switch
out
on the B
control
is equipped
to a function
DISPLAY switch
to
TIME/DIV
is
the
as
needed
from the rest
with
a DM44, set the FUNCTION
other
than
to
A INTEN and the
STARTS AFTER DELAY.
knob
and turn cw until
desired length. Adjust the
to
make the intensified
of
TIME
or
displays
the display. If
1/TIME
are discussed in
for
your
a single
displays set
MODE control control
Trigger Trigger for
switch
positions
positions
7.
For
a delayed
SOURCE
SOURCE switch and
a stable display.
Differential using
the TIME
invalid
when the B
to
STARTS
the
FUNCTION
to ALT. The DELAY
both delayed displays. only
switch
time
measurements
or 1 /TIME
AFTER DELAY.
switch
the CH 2 delayed display.
display
to
NOTE
functions
Trigger
SOURCE
to
with
less jitter, set
the same
adjust
the B LEVEL
and
TIME
and the VERT
TIME
POSITION
The
ALT
position
measurements
of
the DM44 are
switch
is
as
not
DELAY
the
B
the A
control
set
I
I I
4.
Adjust
the DELAY the intensified zone to cover the be displayed in delayed form.
5.
Set the intensified zone noted in steps 3 and 4 is delayed form. The delayed sweep rate is indicated by the dot
on the B
6.
If displays The DM44 will indicate the time beginning
20
HORIZ
TIME/DIV
your
instrument is
of
two
signals can be obtained at the same time.
of
the delayed displays.
TIME
DISPLAY
knob.
equipped
POSITION
portion
switch
to
with
difference
To
obtain
control
of
now
a DM44 delayed
to
move
the display to
B DLY'D. The
displayed in
between the
two
delayed
466/464 DM44 Operators
MIXED SWEEP DISPLAYS
1.
Obtain
2.
Pullout desired sweep rate. If DM44 set the TIME displays are discussed in step
3.
Set the now
contains
display
a normal sweep display.
on the B
your
FUNCTION
or
1/TIME
is
at
for a single
HORIZ
two
sweep rates.
the
A sweep rate
DISPLAY
TIME/DIV
instrument
switch to a
while
knob
and
turn
cw
is
equipped
function
mixed display. Dual
4.
switch
to
The
first
the latter
other
MIX. The
portion
portion
to the
with
than
mixed
display
of
the
of
the
@
I
a
I
I
l
l
J
J
display
portion POSITION
sweep time. the To switch
DELAY
displays.
mixed
Normal on.
MODE CH the
visible. If
is at the B sweep rate. The start
can be changed
4.
If
your
displays
The
beginning
obtain
to
TIME
display.
1.
Preset the
Sweep Display, then turn the
Allow
2.
Set the
switch 2 OR Y CH
1 OR X
3.
Advance
by
control.
instrument
of
DM44
will
of
the
two
mixed sweep displays set
TIME
and the VERT
POSITION
The
fl.
TIME
is equipped with a DM44, mixed
two
signals can be obtained at the same
indicate the time difference between
B sweep
control
control
X-Y DISPLAY
instrument
several minutes
TIME/DIV
to
CH
input
connector
input
the
INTENSITY
the
display
controls
for
switches
2.
Apply
connector.
is
not visible
of
the B sweep rate
adjusting the DELAY
portions
MODE
instrument
and the horizontal signal to
of
the
two
the
switch
positions
positions
to
the vertical signal to the
control
only
as
given in step 1
instrument
warm-up.
X-Y and
until
the
with
the INTENSITY
TIME
displays.
FUNCTION
to
ALT.
The
both
mixed
the CH 2
of
power
the
VERT
display
is
control and adjust the CH 1 and CH 2 VOL the display tally. Center the compressed
controls (CH 2 POSITION vertically, Horizontal · POSITION horizontally); release the BEAM FINDER pushbutton. Adjust the FOCUS display.
at midrange, press the BEAM FINDER pushbutton
TS/DIV
is
reduced in size, both vertically and
display
control
switches until
with
the POSITION
for
a well-defined
horizon-
SINGLE SWEEP DISPLAY
1.
Obtain a Normal Sweep Display. For random signals, set the approximately the same random signal.
2.
Push the SINGL SWP switch. The next trigger pulse starts the sweep and displays a single trace. If READY lamp lights, indicating the A Sweep Generator circuit
is
3.
After the sweep out"
and the READY lamp is out. Press the button display.
to prepare the
trigger
set and waiting
circuit
is
to
trigger
amplitude
button
no
triggers are present, the
to
be triggered.
complete, the
circuit
for
on a signal that
and
frequency
on the A TRIG
circuit
is
SINGL
another
single-sweep
is
as
the
MODE
"locked
SWP
@
466/464 DM44 Operators
21
STORAGE DISPLAYS
STORAGE
1.
Obtain a normal display
(see Normal Sweep Display
of
Displays section
2.
Set the STORAGE LEVEL and VIEW TIME controls NORM. Set the SAVE switch button out. Select either PERS
or
FAST mode.
this manual).
of
the signal to
in
the Basic Oscilloscope
be
measured
to
VAR
4.
Set the INTENSITY control brightness. Set display (depends on the amplitude and speed of the signal to be viewed). Set the VIEW TIME display retention (persistence) in the VAR PERS mode for
the time between erase cycles in the FAST mode.
5.
Push the ERASE button. The display is erased, except in the SAVE mode, and the storage and circuits are reset.
the
STORAGE LEVEL
for
adequate
control
control
for
optimum
for
the desired
display
or
trigger
SAVE STORAGE
1.
Obtain a Storage Display using steps 1
through
4.
I
a.
In general,
VAR
PERS
100 µs/division. b.
When using the B DL Y'D control rate.
3.
Push the ERASE button. storage and display.
22
better
mode
horizontal
to
BENDS A for
trigger
NOTE
results are
for sweep speeds slower than
combination
display, set A TRIG HOLDOFF
maximum
circuits
obtained
of
FAST storage
display repetition
Allow
a few seconds for the
to
automatically provide a
using the
466/464 DM44 Operators
and
2.
Immediately after the event is ·stored, push in the
SAVE button, providing a longer viewing time. The SAVE
INTENS
the display brightness decreases the viewing time.
be TIME setting), this reduces maximum viewing time.
control control
control
Pushing the SAVE button, after the event is stored, can delayed
Maximum viewing time occurs when the SAVE INTENS
is
only
3.
Erase
adjusts the display brightness. Increasing
up
to
several seconds (depending on the VIEW
set
fully
counterclockwise, advancing the
long enough
is
disabled in the SAVE mode.
to
view the display.
@
I
]
]
J J
l l
SAVE RANDOM EVENTS
1.
Obtain a Storage Display using steps 1 through
with a test signal similar
not
mode,
2.
3.
4.
5.
random event occurs, push the SAVE button.
VAR PERS.
Set
the
SAVE INTEN control Replace the test signal with the random event. Push the ERASE button.
When the READY lamp comes on,
to
a random event. Use the FAST
fully
ccw.
but
4,
before the
6.
After the random event storage mode.
circuitry
will automatically
occurs
and is stored, the
switch
to
the SAVE
REDUCED SCAN STORAGE (466 ONLY)
To
obtain a greatly increased the INTEN control to operate the 466 in the reduced scan mode. In the reduced scan mode the instructions obtaining storage displays are unchanged. The difference in operation referenced to the smaller graticule in the center (see item section within this smaller graticule when operating in the reduc­ed scan mode.
11
b in the Controls, Connectors, and Indicators
of
this manual). All measurements should be made
is
writing
that the deflection factors are
rate, pull
of
out
the
on
for
only
crt
J
l
@
466/464 DM44 Operators
23
Except TIME under
for
and
1/TIME
Delayed
DM44 DISPLAYS AND MEASUREMENTS
the
TIME
and
1/TIME
functions
or
Mixed Sweep Measurements.
RESISTANCE
functions,
are discussed in the Oscilloscope
the DM44 is
independently
usable
Applications
anytime
and Measurements section
the
oscilloscope
TABLE
Resistance Ranges
is
turned
of
this
on.
The
manual
E
2
The
meter measure voltage mode
of
operation
ed
in)
and
rms.
1.
Push in the
20
MO RANGE
2.
Connect
resistance.
3.
Reduce
proper
readout is obtained.
24
may
if
the
applied
OHMS
button. the + and
the
range,
be
damaged
the
meter
(OHMS
voltage
FUNCTION
See Fig. 10.
COM
using
by
attempting
is in the resistance
FUNCTION
is
the
button
in excess
button and the
leads
to
following
to
push-
of
120 V
A
the
unknown
table, until a
466/464 DM44 Operators
any
tion
A
is normal.
RANGE
20
MO
2 MO
200 kO
20
kO
2 kO
200 0
blinking unknown and a need
blinking
.
READOUT
20.00-02.00
2.000-0.200
200.0-20.00
20.00-02.00
2.000-0.200
200.0-000.0
readout on
resistance, indicates an overrange
to
readout, when meter leads are
select a
any
higher
MEASUREMENT
20
M0-2
2
M0-200
200
k0-20
20
k0-2
2
k0-200
200
0-0
range, when connected
range.
disconnected
MO
kO kO
kO
0
0
condi-
to
@
I
I
:I
:J
l
1]
J
l l
J
@
RANGE
Fig. 10. Resistance.
466/464 DM44 Operators
SELECTION
(465/DM-0-2)2039-9
25
I
VOLTS
The
maximum
peak
ac)
+
input
the
maximum
The peak
ac) to chassis. meter
The measure voltage mode
of
excess
If
(indicating to
of
the reading exceeds 1200 V
prevent
safe
between
and
COM
may
operation
120 V rms.
overrange)
possible
input
the+
chassis.
floating
be
damaged
if
the
and
disconnect
meter
and
voltage
COM
voltage
meter
the
damage.
is
±1200 V (de+
inputs
or
between
is
±500 V (de+
by
attempting
is
in the resistance
applied
or
voltage
the readout blinks
the+
lead at once
3.
Reduce the range, using Table
readout is obtained.
NOTE
When
no
voltage ranges, the elements
meter
2 V ranges.
to
is
in
A overrange range.
may
leads
blinking
condition
readout
blink. Also,
may
increase the
readout
is
applied
is
0000
and
noise
readout
on
any
and a need
TABLE
Volts Ranges
3
3,
until a proper
in the 20 V to 1.2
individual
picked
range indicates an
to select a
readout
up
by
in the .2 V
higher
kV
the
and
1.
Push in the VOL TS
1.2 kV RANGE button. See Fig.
2.
Connect the {usually a unknown
ground
voltage
COMMON
or
to
be measured and observe the reading.
FUNCTION
test point) and the HIGH lead to the
26
11.
lead
button and the
to
the reference point
466/464 DM44 Operators
RANGE
1.2 kV 200 v
20 v
2V
200 mV
READOUT
1.200-0.200
200.0-020.0
20.000-02.00
2.000-0.200
0.200-0.000
MEASUREMENT
1.2
kV-200
200
20
2 .2
V-20 V-2
v-.2
V-0
V
V
V v V
@
I
J J
J
]
1]
]
l
J
J
@
466/464
Fig. 11. Volts.
DM44
Operators
(465/DM-0-3)2039-10
27
TEMPERATURE
the device removed
These items
probe
from
manual,
to
be measured, a
the
device
and
other
which
should
by
probe
thermal
the
information
be reviewed.
gradient,
probe,
etc.
heat
I
are in the
The
maximum
surface
ground. The
sensor subjected sensor
If the
the
probe
tip
reading
safe voltage
is
±400
V (de +
tip
is fragile
to excessive stress.
should
exceeds
at
once
not
to
prevent
and
exceed
-55°
on
peak
may
20
C
probe
the
ac)
above chassis
break
Force
exerted
pounds.
or
+150°
damage.
measurement
if
dropped
C,
or
on
the
remove
The DM44 is calibrated to its accurate readings The DM44 probe.
In
the
following
to
verify water temperature.
the
melting
by
changes in
Accuracy Check
within
2°C
for
should
temperature and the
be recalibrated
checks, use an accurate
Anything
altitude
and
barometric
original
examples that
boiling
probe,
to
any replacement
thermometer
in
solution
point
pressure.
giving
follow.
affects
is affected
Low Temperature
1.
Push in the TEMP
Fig. 12.
2.
Apply
the temperature
measured so the the device Table 4 to convert the reading fahrenheit.
Temperature
such
as the
flat
to
be measured and observe the reading. See
accuracy
angle
28
l __________________________________________________________________________
I
surface
NOTE
is
of
contact
C)
FUNCTION
probe
of
the
affected
between
to
the device to
probe
tip mates against
from
by
several factors
the
button. See
to
be
centigrate to
bottom
probe
and
466/464 DM44 Operators
indicating
Allow a container
melt
until
there are
To
prevent
sealed
Put the
possible
portion
of
of
probe
tip the container. Wait the probe has reached the water temperature.
(preferably insulated)
only
a few pieces
of
crushed ice
of
ice remaining.
~
probe
damage,
the
probe
is
into the water, avoiding the side
for
be
immersed
the readout
sure
only
(see Fig. 12).
to
stabilize,
the
or
@
I
I
__
J J J
l
J
l
]
J
APPLY
SILi
TO
THE
MEASURED
su~~~E
(W
GREASE
CE
TO
HEN
BE
POSSIBLE).
J
SLM
@
~
JXJ
F"
ig. 12. Tempe rature.
466/464 DM44
Operators
29
CENTIGRADE
FAHRENHEIT
-
50
I I
11111111111
l1111l'l'
-
50
CENTIGRADE
-
40
liiljilill"'ll'"
'I''''''''
-40
-30
TABLE
TO
FAHRENHEIT
-
30
'II'
-20
4
liilll'l'liljlii
'II''''''
-10
CONVERSION
-
20
111111111111 111111111 I
I'
0
-10
'I' '11j11111
10
20
11
0
"I'
11I
30
I I
CENTIGRADE
FAHRENHEIT
CENTIGRADE
FAHRENHEIT
CENTIGRADE
FAHRENHEIT
1738-35
30
0
1
,,,,
,
1
1
""
1
40
50
1·1111· '''I'''''''''
130
100
I··
1·11
11r111
'I
220
10
••
,,,,,
11111111111
50
,,,
1111
60
1'"'1" 11
60
l""I
140
150
110
•1·1··1•11•111111111· 1·I·1111111111111111111
11
11
t
11
n
1111
n I
Iii
230
240
466/464 DM44 Operators @
20
,,
,1,,
,
11111
1
70
70
,,,.
1111H111111111q 111111111111111111111r11111111
,,,,
1
160
120
11111111
1 1 T 1 1 n
250
30
1,
,·11111111111!•
11
1"'
80
I
"'1
90
1111
1111
100
80
.,,
1.,1''
170 180
190
130
11
11
11
1 1
260
270
280 290
40
111
1
1
11
90
''I''
200
140
111
1111 1111Ir11
1·11
"1"
110
''II''
1
11
"'I'
120
210
11
n I '
300
50
,
,1
100
'I 1111111111q1111111r111
150
I:
E
E
I:
I:
I
I
J
J
]
The
readout should be ice remaining after the test to verify that inserting the probe
did
not
raise the water temperature.
High Temperature
Bring
water
to
a slow boil (to prevent splattering). Put
the
probe
tip
into the water, avoiding the side
the
container. Wait
the
probe
has reached the water temperature.
for
-2°
C to 2° C. There should
the readout
or
bottom
to
stabilize, indicating
be
The readout should be between
clean water at sea level.
TIME AND
of
See
DM44
MEASUREMENTS in this manual.
DELAYED
98°C
1/TIME
OR
MIXED
and 102°C
SWEEP
for
J J
@
466/464 DM44 Operators
31
APPLICATIONS AND MEASUREMENTS
Refer
to
Basic Oscilloscope Displays
as
needed
to
obtain
correct
displays.
I
PRELIMINARY
Signal Ground
~
Avoid
cross-grounding
may occur.
The most reliable measurements are made when the oscilloscope under ground on the with
the
ground
test by the
lead can be connected to the
oscilloscope
unit
ground
under
Input Coupling Capacitor Precharging
In the
GND
position, ground precharging network. Coupling level voltage transients applied
through
capacitor
of
the signal applied
to
the
a one-megohm resistor
amplifier
because instrument damage
is
connected
strap on the signal probe. Also a
chassis
test.
the
This
to
charge to the average de voltage
accidentally
input. The precharge network also
to
the
ground
ground
to
establish a common
input
signal
network
to
the probe. Thus, any large
allows the Input
generated will not be
of
banana jack
is
connected to
to
the
ground
form a
unit
of
provides a measure by
reducing the
external
probe de level than that previously applied, especially level difference is more than 10 times the VOL setting.
set the
Wait several seconds discharge.
capacitor
remain on the screen so the ac
can be measured in the
circuitry
The
following
tip
is connected
1.
Before
connecting
Input
Coupling
2.
Touch
the probe
3.
Connect
4.
Wait several seconds
5.
Set the
the probe
to charge.
Input
protection
current
levels that can be
during
procedure
to
switch to GND.
tip
for
Coupling
normal
capacitor
a signal source having a
the probe
to
the
tip
to the external
drawn
charging.
should be used whenever
tip
to
a signal source,
oscilloscope
Input
to the signal source.
for
switch
component
manner.
chassis
Coupling
the
Input
to
AC.
The
capacitor
Coupling
display
of
the
circuitry from the
the
different
if
the de
TS/DIV
ground.
to
will
signal
I I
I
~
I I
32
466/464 DM44 Operators
@
I
J
J
J
J
J
OPERATOR'S
ADJUSTMENTS
AND CHECKS
To
verify
measurement accuracy, perform the
ing
checks ment. See the for
calibration
Trace Rotation Adjustment
Normally using switch vertical deflection. Adjust driver the trace
and adjustments before
Calibration information.
not required. Obtain a
only
steps 1
to
GND
adjustment
with
through
to display a
located below the
the center horizontal
section
3.
the
making
of
the Service manual
Normal
Set the CH 1
free-running TRACE
ROTATION
crt
graticule)
graticule
follow-
a measure-
Sweep Display
input
Coupling
trace with no
(screw­to
align
line.
--
I~-+--+--+-~--+--+--+---+--+----,
100... . ..
I
--
~~-+---+--~,___,.-+-~-+-..,--+-.--+--.--+--+-~
'--
I
10~...j_-+..-L....j.___J_-+---L--+--L-........,-+---'--+---+-~
r ·
r.~
IL__--+--+--+-----1--+---+--+---t--t-___J
· ·
'--
--
r.-
_.,_
,_
.C'~
'--
--
....
'
· ·
--
,
__
---
.c-.
· · · . C UNDER COM-
CORRECT
FLAT
OVER COM-
PENSATED
(OVERSHOOT)
PENSATED
(ROLLOFF)
1737·17
l
J
J J
L.
Probe Compensation
Miscompensation operator adjustments Probe
calibrator VOL to rolloff,
@
-
error. Most attenuator probes are equipped
to ensure
compensation
Obtain
DC.
a Normal Sweep Display presentation of the
square-wave voltage. Set the appropriate
TS/DIV
switch to
Check
the waveform presentation
and if necessary, readjust
is
one
of
the greatest sources of
optimum
is
accomplished
.1 V position
with
measurement accuracy.
as
follows:
and the
compensation
Input
coupling
for
overshoot
466/464 DM44 Operators
for
or
flat
Fig. 13. Probe
tops
on
the waveforms.
for
method
Vertical Gain Check
Obtain a Normal Sweep Display presentation calibrator square-wave voltage. Set· the appropriate VOL TS/DIV switch to the 50 mV Coupling control deflection
of
compensating the probe being used.
switch to DC. Make sure
is
in the calibrated detent.
is
5.8
to
6.2 divisions.
compensation.
See
Fig.
13.
Refer to probe manual
position
the
Check
and the
VAR VOL
that the vertical
of
the
Input
TS/DIV
33
Basic 466/464
Obtain a Normal 5 ms position. Set the A Push the TRIG VIEW sample control the position
14). Verify the
1
of
to
display
the left peak
~-------
._,.-+-~
O<?
...•
Timing
Display. Set
switch
the line voltage. Use
vertically
area. Use
position
horizontal
I I
6.53
TO
\
\
10------~--_,_
I
~..
. . . . . .
..........
..__..__'"""--l,__..
...
Fig. 14. Basic 466/464
Check
Trigger
and
the
the
horizontal
to
the
distance
9.8
TO 10.2
6.79
~--+-._..
..
...
,
...
..
...
...
..
. .......... .
...
..
(60 Hertz Line Only)
the A TIME/DIV SOURCE
hold
top
left
graticule
_____
timing
switch
it in. This displays a
the A Trigger
of
the
display to
position
edge (see Fig.
between the first and
_____
---+-+-'--+-~+--+-1
(
1738-14)2037-14
check.
switch
to
LINE.
LEVEL
within
control
...,.
to
to
the
fourth
peaks is 9.8
not
visible, verify the
is
first
and
the
third
peaks is 6.53
DM44
accuracy
Check
Timing
Perform the Basic 466/464
Leave
except
TIME/DIV
B
HORIZ
B SOURCE FUNCTION
6.
TIME
Push the TRIG VIEW DELAY point Point measurement on is 16.49
to A). Use
the
TIME
next
to
Check (60 Hertz Line
of
the
horizontal
controls
DISPLAY
the
16.84 and
set
as
follows:
POSITION
center
the
ALT
point
cycle
(see Fig. 15,
horizontal
to
the
to
10.2 divisions. If
horizontal
to
Timing
deflection
as
for
the
Basic 466/464
5µs
AINT STARTS TIME To
move measurement right point.
button
DELAY
the
ms lamp is I it.
control
graticule
control
center
Point
and
to
horizontal
the
fourth
distance
6.79 divisions.
between
Only)
Check system.
of
move the reference
B). Verify
to
verify
Timing
AFTER DELAY
the
time-
point
to
the
reference
hold
it in. Use
line (see Fig. 15,
to
move the
graticule
the
readout
peak
the
the
the
the
time-
line
I I
I
I
I
34
466/464 DM44
Operators
@
I
J
J
]
l l
J
J
r
100····
External Horizontal Gain Check
the Basic Oscilloscope Displays procedure Normal voltage waveform; then, set the With the
...
~-··
9a-\-+-
(If
X-Y
A__)
't..
operation
Sweep Display
calibrator
....
····~
\
\ :: I\:
....
Fig. 15. DM44
is
to
be used.) Use
of
signal connected to the CH 1
....
~
.._~,......_-+-~---11
1
..
.. ..
,_·
...
..
..
timing
check.
~teps
the
calibrator
TIME/DIV
(1738-15)
2037-15
1through3
for
obtaining a
square-wave
switch
to
OR
of
X-Y.
input
connector and the
50
mV,
the crt display
horizontally by 5.75 to 6.25 divisions.
CH
1 VOL
should
TS/DIV
be
two
dots separated
PEAK-TO-PEAK VOLTAGE
MEASUREMENTS-AC
Obtain a Normal Display. Make sure the VAR
VOL TS/DIV
position the display so the
horizontal graticule line (see Fig. 16, position the display so one with
the center vertical
Measure the vertical deflection from peak-to-peak (Point A
to
Point B).
If
trace signal.
tained
of
X
the measurement.
control
the amplitude measurement
is
thick
a more accurate
by
measuring from the
a valley. This
as
a result
will
is
in
the
graticule
NOTE
subtract
calibrated detent. Vertically
lower
portion
of
the
line (see Fig.
of
hum
measurement
top
the trace thickness from
coincides
Point
A).
upper
peaks coincides
is
critical
and/or
noise on the
can be
of
a peak to the top
switch set to
with
a
Horizontally
16,
Point B).
or
if
the
ob-
l
@
466/464 DM44 Operators
35
POSITION CENTERLINE
TO
Example: The peak-to-peak vertical
measured is 4.6 divisions (see Fig. 16)
switch
setting
Using
the
Volts
Peak-to-peak
of
.5.
formula:
vertical deflection (divisions)
X
deflection
with
a VOL
VOLTS/DIV
setting
TS/DIV
I I
10-
I
0
•...
Multiply
TS/DIV
VOL factor factor
36
........
~-.....--~------""-------~--
MEASURE
Fig. 16. Peak-to-peak voltage
the vertical
switch
of
the
switching
setting. Also
probe
connector.
FROM A
deflection
if the
AMPLITUDE
TO
probe
..........
-..+~-+
B (1738-16)2037-16
of
a waveform.
just
measured by the include does
the attenuation
not
have a scale-
466/464 DM44
Substituting
Volts Peak-to-Peak
the given values:
INSTANTANEOUS
MEASUREMENTS-DC
Obtain a Normal
TS/DIV
VOL
To
measured, set the
vertically
input
the above waveform moves is negative.
Operators
control
determine
position
coupling
the
center
= 4.6 X 0.5 V = 2.3 volts.
VOLTAGE
Display. Make sure the VAR
is in the calibrated detent.
the
polarity
input the baseline to switch
to
of
the crt, the
to
below
the
of
coupling
the
DC. If
voltage
center
the
switch
center
the
waveform moves
is positive. If
of
the
voltage
crt, the voltage
to
of
GND
the
to
be
and
crt. Set
the
to
@
I
I
J
l
J
]
l l
Set
the
input
coupling baseline voltage baseline
divisions and Multiply setting. probe
4.6 reference line, and the
to a convenient
to
be measured is positive, then
to
the
Switch
EXAMPLE:
Using
the
of
the
desired
the
Include
does
divisions
the
Instan-
taneous
Voltage
vertical
not
bottom
Input Coupling
deflection
point
veritcal
formula:
deflection
the
attenuation
have a
The
(see Fig. 17), the waveform is above the
vertical
distance
(divisions)
switch
to
GND
and
position
reference line. For
graticule
on the waveform (see Fig. 17).
scale-factor
vertical
VOL
line.
Switch
between the reference line
by
the VOL
factor
switching
distance
TS/DIV
X
polarity
examplE:l,
position
to
DC. Measure the
TS/DIV
of
the
probe connector.
measured is
switch
is set
VOLTS/
X DIV
switch
to
setting
the
if
the the
if the
2.
£._
:~GATIVE
REFERENCE
LINE
©--/
I
100
..
I
90-+----+-...---+®
--+--+--t--+-+-+--r+--f+11r-t--+-t--+--+-++-,-t--
T----
VERTICAL C
DE
FjLECTI
---
---
:r_+----+-.........,__._-+--"""---+~-+--'--+-~+----l
-~
=~"-~·
Fig. 17. Instantaneous voltage measurement.
B I MEASURE
,__
..,.\,-t--t-t-+-1-+--t--+-t
l
f-----t-
·-·~~~-~~-~~~
1737·05
AMPLITUDE
A TO B
A~~ti~~v~E
TO
B
J J
l
Substituting:=
If a negative voltage is
trace
to
the top
graticule
Fig. 17).
The
by
switching
@
ground
reference
to
the
4.6 X
(+1)
X 2 V = 9.2 volts.
to
be measured,
line and measure from C
line
can be checked at
GND
position.
position
466/464 DM44 Operators
to
any
the
B (see
time
This method can also be used
with
to
a convenient reference line. Measure the vertical deflection between the reference line and other given.
respect
to
another. Position one
voltage level.
Substitute
to
measure one voltage
of
the voltage levels
this value in the
divisions
formula
of
the
just
37
ALGEBRAIC
In the
ADD
position
waveform displayed
to
applied
CH 2 INVERT switch is pushed, the waveform displayed
the difference inputs (CH 1 - CH 2). ADD either switches are set to-the same position).
the CH 1 and CH 2 inputs (CH 1 +
mode
is equal
VOL
TS/QIV switch (when both VOL TS/DIV
is
of
the signals applied
to
ADDITION
of
the VERT MODE switch, the
the algebraic sum
The
total
deflection
the
deflection
of
the signals
C~
to
the CH 1 and CH 2
factor in the
factor
indicated
2). If
th.e
1s
by
The
following
when using the
1.
Do
oscilloscope.
2.
Do
not
about
eight
example, voltage applied four
with
volts. Large voltages may
general precautions
ADD
not
exceed the
apply
times the VOL
a VOL
to
that
mode.
signals
TS/DIV
channel
should
input
voltage rating
that
exceed an equivalent
TS/DIV
switch
switch setting
should
distort
not
exceed
the display.
be observed
of
the
of
settings. For
of
0.5, the
about
I
I
38
ov
100
90
1?
0
I
r
I
L
·~
t
I
I I I
POSITIVE
LEVEL
(A)
CHANNEL 1 SIGNAL
WITH 3 POSITIVE
v
DIVISIONS
DC
I
LEVEL.
OF
/'
i
I
100
I
90
OV
l
I
'
I
1?
0
I
Fig.
466/464 DM44
NEGATIVE
OFFSET -
I
I
(B)
CHANNEL 2 DISPLAY
WITH 3 NEGATIVE
18.
Algebraic addition.
'--1--->--
I
I
DIVISIONS
OFFSET.
Operators
OF
I
100
I
90
I
I
10
I
0
I
I
I
L_i
)
/'
/
I
(C)
)
.
j
RESULTANT
I/
.
I/
DISPLAY
I
j
I
I
I
---
I
@
I
J
J
J
]
J
l
l
l
l
common
A
for
offset
EXAMPLE: The Channel 1 signal is on a 3 division, positive Fig. 18A.
setting de level of The operating
to
the
value determined, to Channel 2 input. See Fig. 188.
ADD
COMMON-MODE
The
ADD contain components rejection. should
contains 19A). following
INVERT
be observed.
EXAMPLE: The signal applied
To
1.
Connect
2.
Set
@
use
of
the
ADD
mode is
a signal riding on a de level.
de
level (using the center line Multiply determine
(or
a positive level using the CH 2
mode
range
3 divisions by the VOL
the de-level value.
puts the resultant
of
the POSITION controls.
to
provide a de
as
zero volts). See
TS/DIV
Apply
a negative
INVERTswitch),
display
within
REJECTION
mode can be used to display signals that
undesirable components. These undesirable
can be eliminated
The
precautions given
unwanted
remove the undesired
procedure.
the
switch
line frequency
a line
frequency
VERT MODE switch
to on (button in).
through
under
algebraic
to
components
components
signal
to
to
ALT
Adjust
common
addition
the
CH 1 input
(see Fig.
use the
the
CH
2 input.
and the CH 2
the CH 2
466/464 DM44 Operators
switch
the
mode
10
CH 2
SIGNAL
FROM
LINE
FREQUENCY
SOURCE
(INVERTED)
SIGNAL
LI
WITH
~~~~6~~~~CY
CANCELED
OUT
//
0...
I
"
(A) CH 1
I
10<?····
~--+---+-+---+-+--+-+----+--~--1
----··-··--
l+H++++++l'++-++4+-~++++l+l+H+++~l++++++~"++-i
--
._
_____
10--+----+--+----+--+---+--+---+--+----I
I
~
...
(B)
RESULTANT
Fig. 19. Common-mode rejection.
'-;'-
. . . . . . . . . .
AND
CH
2 SIGNALS.
-
...
-------
-------- ------·
SIGNAL.
--
1738-19
39
meas.
ll.
,_ 2 ...
L
L.il. . L
.L
C.£$4
LJ
..
J.
.•
J
VOL
TS/DIV display portion
3.
readjust the CH cancellation
198).
and VAR VOL
is
about
of
the CH 1 display (see Fig. 19A).
Set the VERT
of
AMPLITUDE COMPARISON
TS/DIV
the
same
MODE
2 VAR VOL
the undesired signal
TS/DIV
controls
amplitude
switch
control
component
MEASUREMENTS
so the CH 2
as
the undesired
to
ADD. Slightly
for
maximum
(see Fig.
Determine the vertical conversion factor using this
formula:
reference signal amplitude (volt)
Vertical
Conversion
Factor
Determine the
formula:
vertical
deflection
(divisions)
arbitrary
VOLTS/DIV
X switch
setting
deflection factor using the
I
If comparisons signal are repetitious (e.g., on an assembly line test) it possible to obtain more accurate easily read measurements if
to
set the reference signal
unknown
The compared with divisions.
Other disturbing establishing a vertical conversion deflection factor. The must established.
40
be
known
unknown
the setting
of
an
unknown
the
VAR VOL
to
signal can then be
or
adjusted to an exact
signals may be measured
of
the VAR VOL
amplitude
before a vertical conversion factor can be
signal with a reference
TS/DIV
an exact
control is adjusted
number
quickly
TS/DIV
factor
and
of
the reference signal
of
divisions.
and easily
number
without
control
an
466/464 DM44 Operators
by
arbitrary
is
of
Arbitrary
Deflection
Factor
To
measure the
VOL
TS/DIV vertical not
readjust the VAR VOL vertical of
the
switch to a setting that provides
deflection
deflection
unknown
Signal
Amplitude
signal using the
Vertical
Conversion X
Factor
amplitude
to
make an accurate measurement.
in divisions and calculate the
of
an
TS/DIV
Arbitrary
deflection
factor
VOLTS/DIV
switch
setting
unknown
control. Measure
following
X deflection
signal, set
formula:
Vertical
(divisions)
the
sufficient
Do
the
amplitude
@
I
I
J J
Signal
Amplitude
EXAMPLE:
volts
30 VOL
TS/DIV
tion
of
4 divisions.
=
with
Vertical
Conversion
Factor
The
reference signal
a VOL
TS/DIV
control
adjusted
or
VOL
TS/DIV
X
switch
setting (divisions)
switch
to
provide
X
amplitude
setting
of
a vertical deflec-
deflection
5 and the VAR
Vertical
measured is
TIME/DIV display horizontal graticule line (see Fig. 20).
Measure the horizontal distance between the time measurement points. the
setting
is used, divide this answer
switch
to place the time measurement
of
the
TIME/DIV
for a single
Multiply
switch. If sweep by
event and
the
distance
10.
position
points
to
measured
magnification
the
the
center
by
J
l
l
l
J
..
Jl
X
Substituting
factor
formula:
Vertical
Conversion
Then
with
to-peak
can be
amplitude
determined
Signal
Amplitude:
TIME-DURATION
Obtain a Normal
TIME/DIV
@
kJ,,
...
control
,.JUJLL"'hbLJ
these values in the vertical conversion
Factor
a VOL
TS/DIV
of
an
by using the signal
30
4XS
switch
unknown
setting of
signal, 5
1.5
1,
divisions
amplitude
the peak-
= 1 V X 1.5 X 5 = 7.5 volts.
MEASUREMENTS
Sweep display. Be sure the VAR
is set to the
k.
JU
calibrated
L.." , ..
.EL.
detent. Set the
466/464
..
l
high
formula:
DM44
.&!Z
I
100
I
90
I
I
®""
/
,V
I
0
I
I
Operators
I
1/·
-
~.-
..
~
r\'
\
\
\
HORIZONTAL
DISTANCE
Fig.
20.
~
\
~
\
\
"-
-I
Time duration.
. '
®~
ti
V!
j I
I
v
I/
MEASURE
TIME FROM
ATO
1737-18
B
41
EXAMPLE: The distance between the time measure-
ment
points
is
8.3 divisions (see Fig. 20) and the TIME/DIV
switch is set to 2 ms
with
the
magnifier
off.
Frequency
= time
duration
16
.6 ms
= 60 hertz
I
Using the formula:
Time
Duration
Substitute the given values:
Time
Duration
horizontal
distance X
(divisions)
magnification
= 8.3 X 2 ms = 16.6 milliseconds
FREQUENCY MEASUREMENTS
Time
Duration measurements can be used to determine
the
frequency
1.
Measure one cycle
Time-Duration
2.
Take the reciprocal
mine the frequency.
EXAMPLE: The
which
has a time
of
a recurrent signal
of
the
Measurements.
of
frequency
duration
of
waveform
the
time
of
the signal shown in Fig. 20
16.6 milliseconds is:
Time/Div
setting
as
follows:
as
described in
duration to deter-
RISETIME MEASUREMENTS
Risetime measurements use the same methods duration made between the 10% and 90% Falltime is measured between the
a sweep speed setting events (if possible) and be sure the VAR is in the calibrated detent. Set the VOL VAR division display. Set vertical
bottom
touches
the risetime spread Horizontally
waveform intersects the second vertical graticule line (see
Fig.21).
measurements,
trailing
Obtain a Normal Sweep Display. Set A SLOPE
Set the
edge
control
(or signal
touches the
the 100%
TIME/DIV
position the
of
graticule
except
the
the waveform.
that
displays several cycles
amplitude)
positioning
0%
graticule
line (see Fig.
switch
for
a single-event display
horizontally
display
the measurements are
points
of
90% and 10% points on
TIME/DIV
TS/DIV
for
line and the
as
much
so the
10%
as
time-
the waveform.
to+.
Use
control
switch and
exactly
so the
21
display
).
as
possible.
point
a five-
display
top
with
on the
or
I I
I
42
466/464 DM44 Operators
@
Using the time
duration
formula
to
find
risetime:
l
J
l
l
l
J J
I
...
I
w
..Jo
<{
::::>
z!::::
(.!)...J
-
Q.
c:n::!:
<{
~@,
0
I
L
90% setting
and TIME/DIV
lJ
~-
Measure the
points
of
the
EXAMPLE: The
90%
points switch
..
)'
,,
/
horizontal
and
multiply
TIME/DIV
is 5
is set
I
--··
···®~
~
~,,,,,.
V:
~""
)""
..
HORIZONTAL
DISTANCE -
Fig. 21. Risetime.
distance
the
distance measured
switch.
horizontal
distance
divisions
to
1
µs.
/
....
MEASURE
TIME
....
1737•14
between
(see Fig. 21) and
the
between
10% and
the
ATO
by
FROM
B
the
10%
the
Time
Duration
(risetime)
Substituting the given values:
Risetime
= 5 X 1
horizontal
distance
(divisions)
µs
= 5
TIME/DIV
X
microseconds.
setting
TIME-DIFFERENCE BETWEEN
TWO PULSES FROM
DIFFERENT SOURCES
Obtain a Normal Sweep Display. Make sure the VAR TIME/DIV Trigger signal to CH 1 and the the with
control
SOURCE switch
signals
to
the
input
equal time delay.
is
in the
to
comparison
connectors
calibrated
CH
1.
Connect
signal
using
detent. Set the A
the reference
to
CH
2.
Connect
probes
or
cables
l
@
466/464 DM44 Operators
43
Set the VERT general, and the signals.
signals.
the
divide
CHOP
ALT
Center
Measure the
Multiply
TIME/DIV
this answer by 10.
CHANNEL
I
....
I
0
9
I
LEVEL
. . . .
50%
100
A MPLITUDE
\
0
I
0
....
I
Fig. 22. sources.
Time
MODE
is more
position
each
of
horizontal
the
measured
switch; if sweep
1 (REFERENCE)
\
. . . .
·\
\
~@
F:.
I I
. . .
. . . .
.
...
HORIZONTAL
~
-
difference
DIFFERENCE -
switch
suitable
is
more
the
displays
....
-
-
-
-
. . . .
,
between
to
either
for
low-frequency
suitable
vertically
difference
difference
magnification
. . . .
. .
,......-
)
®
.
...
. . . .
I
I
r-t
I
two
CHOP
or
ALT. In
signals
for
high-frequency
(see Fig. 22).
between the two
by the setting
CHANNEL2
/
of
is used,
I
L
. . . . .
...
MEASURE
TIME FROM
ATO
B
. . . .
1737
15
pulses from different
EXAMPLE:
MAG
switch
waveforms
Using
Difference
Substituting
Difference
TIME
If
comparisons signal are possible measurements if the VAR set the reference signal to an
unknown
The
compared divisi_ons.
The
TIME/DIV
to
is 4.5
the
Time
Time
X10 and
formula:
the
given values:
the
divisions
Time/Div
60 µs X 4.5
horizontal (see Fig. 22).
setting
10
COMPARISON MEASUREMENTS
of
an
unknown
repetitious
to
otbain
signal can then be
with,
(e.g., on assembly line test) it is
more
TIME/DIV
or
adjusted to, an
switch
magnification
exact
is set
to
50
µs,
difference
horizontal
x
difference (divisions)
22.5
signal
with
accurate, easily read
control
number
quickly
exact
between
microseconds
a reference
is adjusted
of
divisions.
and easily
number
the
I
I
'
to
of
44
466/464 DM44 Operators
@
J
J
Other disturbing establishing a horizontal bitrary reference signal conversion
unknown
the setting
deflection
factor
signals may be measured
of
factor. The time must can
ba
the VAR
conversion
be
known
established.
TIME/DIV
factor
duration
before a
without
control
and an ar-
horizontal
of
by
the
horizontal deflection in duration
using the formula:
Time
Duration
divisions
arbitrary
deflection
factor
and calculate the
horizontal
X
deflection (divisions)
time
J
J
J
l
J
J
l
Determine
formula:
Horizontal
Conversion
Factor
Determine
formula:
Arbitrary
Deflection
Factor
To
measure the
the
TIME/DIV horizontal Do
not
readjust the VAR
@
the
horizontal
the
arbitrary
time
switch
deflection
conversion
reference
(seconds)
Horizontal
deflection
(divisions)
deflection
horizontal
conversion
factor
duration
to
a setting that provides
to
make an accurate measurement.
of
TIME/DIV
X
X
an
unknown
control.
factor
using the
TIME/DIV
switch
setting
factor
using this
TIME/DIV
switch
setting
signal, set
sufficient
Measure the
466/464 DM44 Operators
or
Time
Duration
EXAMPLE: The reference signal
455 hertz (time duration: 2.19
TIME/DIV control divisions. Substituting these values in the conversion factor formula:
duration
adjusted
Horizontal
Conversion
Factor
Then
with a TIME/DIV
of
horizontal
= conversion X
factor
switch setting
to
provide a
8 X 0.2 ms
an
unknown
2.19 ms
signal
of
.2 ms, and
horizontal
switch
that
TIME/DIV
switch
setting
frequency
milliseconds)
the
deflection
setting
of
completes
X
VAR
1.37
50 µs,
horizontal deflection
(divisions)
measured is
with
TIME/DIV
of
eight
horizontal
the
time
one
cycle
in
45
a
seven
horizontal
time
duration
Time
Duration = 1.37 X 50 µs X 7 = 480 µs
This answer can be converted to the reciprocal Determining
divisions, can be determined
formula:
frequency
of
the time
Frequency). ·
duration
(see applications on
by
using the
by taking
PHASE DIFFERENCE
MEASUREMENTS
Use
either
the CHOP
SOURCE switch
precede the comparison signal in time. Use coaxial cables
or
probes
to
pushbutton
of take this into and CH 2 VOL
VAR
that
have equal
the
input
connectors.
If the signals are
to invert the Channel 2
opposite
controls
polarity due to 180° phase difference; if so,
account
TS/DIV
so the displays are equal in amplitude.
or
ALT
mode. Set the A TRIGGER
to
CH
1.
The reference signal should
time
delay
to
connect the signals
of
opposite
in the final calculation). Set the CH 1
switches and
polarity, set the INVERT
display
(signals may be
the
CH 1 and CH 2
CHANNELl
(REFERENCE)
r
100
..
I
7
....
··I\···
)/
V1
v
..
.......
V'"
,
/
v
-
--0-~
I
HORIZONTAL 1
DIFFERENCE-~
90
/'i
~
~
1.
Fig.
represents 45° 45° /division). The sweep rate can be stated in terms degrees
as
of
45° /division.
CHANNEL 2
(LAGGING)
.
. .
··V
~
K
"\
r\
8 DIVISIONS
(360)
23.
the cycle (360°
...
..
'I\.
,@
\
\.
N
'!\
~
iN
'\
I
i'\.
'\
I
0
Phase difference.
: "
I
·!--··
.....
"'-.......
.
. .
/
..
..
I
v
. . .
...;-
..
..
.
...
ii
J
MEASURE
v/
TIME FROM
v
8 divisions =
ATOB
. .
..
1737·20
I
I
I
of
Set the
waveform. Position the
TIME/DIV eight
46
TIME/DIV
control
divisions (see Fig. 23). Each
for
1 reference signal cycle in exactly
switch
display
for
about
and turn the A VAR
division
a one-cycle
of
the graticule
466/464 DM44 Operators
Measure the horizontal difference between
ding
points on the waveforms and measured (in divisions) by 45° obtain the
amount
of
phase difference.
multiply
/division
correspon-
the distance
(sweep rate) to
@
I
l
J
l
EXAMPLE:
a sweep rate
Using the formula:
Substituting
Phase Difference
of
Phase
Difference
The
horizontal
45°
/division
horizontal
difference
divisions
the given values:
= 0.6 X 45° = 27°.
difference
as
shown
X sweep
is 0.6
division
in Fig.
23.
(degrees/div)
with
I
CHANNELl
100
(REFERENCE)
I
90
,..._,_
-
'
----©)
0
I
....
0
..
I
I
J
I I
I
· · · ·
CHANNEL
2 · · · ·
I
I
I
®-
...
J
.
...
.
...
~
-
-
....
MEASURE
TIME
FROM
ATOB
l
l
l
J
l
HIGH
RESOLUTION PHASE
DIFFERENCE MEASUREMENTS
Make more accurate phase measurements
the
creasing TIME/DIV
sweep
EXAMPLE: If the sweep rate were increased
with
the magnifier, the magnified sweep rate
-;-
10 = 4.5°
used in Fig. 23
@
sweep rate
control)
magnification
/division.
but
by
using
may also
Fig.
with
(without
the
24
the X10
changing
X10
MAG
be
used (see Fig. 24).
shows the same signals
MAG
switch
by
in-
the A VAR
mode. Delayed
10
times
would
be 45°
as
set
to
X10.
466/464 DM44 Operators
----
Fig.
With a horizontal difference difference is:
Phase
Difference
Substituting the given values:
Phase Difference
HORIZONTAL
DIFFERENCE
24.
High-resolution phase difference.
horizontal difference (divisions)
= 6 X 4.5° = 27°.
----
of
6 divisions, the phase
magnified
X sweep rate
(degrees/div)
1737-21
47
MEASUREMENTS
Be sure the VAR detent. Set the B rising toB
portion
DLY'D.
100
of
the
i
I
0
I
I
0
I
0-
L
PULSE JITTER
TIME/DIV
TIME/DIV
pulse. Set
JITTER
l
i
@
switch
switch
the
MEASURE
TIME
TO B
A
·,
!'--
. I
--i
to
HOR IZ DISPLAY switch
FROM
,~
©
I
_,
I
I
I
I
I
I
I
I
I
I
,,
I
!---
is in the calibrated
intensify the full
r-
1737•23
Pulse
jitter
is
shown pulse and includes Sweep (see Fig. 25). TIME/DIV
switch
the
setting
by
inherent
Multiple
to
obtain
DELAYED OR MIXED
SWEEP
The
delayed sweep features
provide
the
(B sweep) is magnification
amount
DELAY-TIME presented at the sweep rate selected
switch. The
DELAY the TIME delayed display, use operation.
Magnified Sweep Starts After Delay
VERT
I
about
higher
X1 O MAG
of
time selected
position
delayed
POSITION
1.
Connect
MODE
2.
Set the VOL
four
divisions
switch.
POSITION
following
portion
switch
MAGNIFICATION
apparent
not
is the result
of
the signal
magnification
The
sweep
actually
by
the A TIME/DIV
control
method
the B Trigger
to
be
positioned
control.
the
to
to
display
TS/DIV
in
amplitude.
If
there
Triggered
either
switch
horizontal
jitter
the distance by
pulse
of
the
475 can be used
than is
rate
of
the
increased;
of
delaying
before
by
uses
the
SOURCE
is
too
B Sweep
input
the
channel
to
produce a display
movement
of
jitter
delayed sweep
the
the
switch
the
the B
STARTS AFTER
switch
with
much
connector.
of
the
Delayed
the
time.
provided
apparent
B sweep an
and display TIME/DIV
to
allow
the
DELAY-
jitter
in
mode
Set
used.
the
B
to
by
the
is
the
of
the
I I
E
48
Fig. 25. Pulse jitter.
466/464 DM44 Operators
@
I
J
J
J
l
3.
Set
the A TIME/DIV
displays
Trigger
Instrument
TION
DELAY-TIME to
intensifies intensified
magnify
26).
the
4.
Set
the
SOURCE switch
equipped
switch
5.
Position the start
be magnified.
6.
Set the B
the zone remains
7.
Set the
the
switch
complete waveform.
HORIZ
is
POSITION
portion
DISPLAY switch
with
not
set
TIME/DIV
full
portion
HORIZ
of
to DM44. Verify that the FUNC-
to
TIME
of
the intensified zone with the
control
switch to a setting which
to
be magnified. The start
as
positioned
DISPLAY switch to B DLY'D
A sweep
to
a sweep rate which
to A INT
STARTS AFTER DELAY.
or
1/TIME.
to the part
above (see Fig. 26).
that
is intensified (see Fig.
and the B
of
the display
of
the
to
INTENSIFIED
TO
101~-+1--+---+--+--+--+._+..-J
I
~
....
(Al A INTENSIFIED
1~
••.
90
BE
MA,9NIFIED
DISPLAY
ZONE
'
·\
'
.....
-+~
l
J
J
l
8.
Time
measurements can in the conventional manner. The sweep rate is determined by
the
setting
9.
The calculated the B TIME/DIV
@
of
the B
TIME/DIV
apparent sweep magnification can be
by
dividing
the A
switch setting.
be
made from the display
switch.
TIME/DIV
switch setting by
46~/464
DM44 Operators
1~
0
..
I
(Bl B
Fig.
7
,/
""'
DELAYED
DISPLAY
26.
Delayed sweep
v
1738-31
magnification.
49
EXAMPLE:
with
an A
TIME/DIV
Substituting
The
apparent
The
apparent
TIME/DIV
switch
setting
Apparent
Magnification
(Delayed - B
Sweep)
the
given values:
Apparent
Magnification
magnification
switch
of
_ A
magnification
setting
1
µs
is:
TIME/DIV TIME/DIV
1 x 10-
1 x 10-6
is 100 times.
of
.1
setting setting
4
Triggered Delayed Sweep Magnification
The
delayed sweep
ed
may
produce magnification triggered delayed
under
display
1.
Set
"Magnified
mode
up
the
magnification
too
much
ranges.
provides a is
Operating
triggered
display
Sweep Starts
method
jitter
at high apparent
the
B Sweep
more
stable
display
at
the
same
point
as given in steps 1
After
Delay."
of a display
ms and a B
just
describ-
in
since the
each time.
through
a
6
2.
Set
the B Trigger
position
on obtained,
that If
controls
VOL
HORIZ of display.
are previously
as
the A
Trigger
3.
Adjust
the
B LEVEL
the
trace is stable.
see step 4.)
4.
Inability
the
the
condition
TS/DIV
5.
When
DISPLAY
the
B LEVEL
6.
Measurements are made and
calculated
to
signal does
cannot
or
by
increasing
setting),
the
correct
switch
control
as
given.
intensify
not
in the
SOURCE
SOURCE
control
(If
an
meet
be remedied
trigger
portion
to B DLY'D.
may be necessary
STARTS
intensified
the
desired
the
triggering
the
display
B Sweep
so
is intensified, set
switch
switch.
the
with
Slight
magnification
AFTER DELAY
to
the
intensified
zone
cannot
portion
the B Triggering
amplitude
externally.
indicates
requirements.
readjustment
for
a stable
same
I
zone
be
I I
(lower
the
factors
mode
I
I
50
466/464 DM44 Operators
@
I
l]
J.
J
l
l l
J
BASIC 466/464 DELAYED OR MIXED SWEEP TIME MEASUREMENTS
The
delayed
TIME
DIFFERENCE BETWEEN
REPETITIVE PULSES (BASIC 466/464)
Obtain
a Delayed Sweep Display. For measurement, sweep
speed
With
the
DELAY
zone
to
the
Set
DELAY
portion)
Note
the second
pulses
correct tal
DELAY
formula:
TIME
the
DELAY
pulse
are
pulse).
POSITION
TIME
Determine
set
the B TIME/DIV
that
gives
the
HORIZ TIME
the
first
HORIZ
to
some vertical reference line (see Fig.
setting
TIME
displayed,
DISPLAY
POSITION
pulse
DISPLAY
POSITION
of
the
DELAY
POSITION
to
the same vertical reference line (if several
Do
not
change
and FINE
POSITION
the
time
sweep
switch
usable
(see Fig. 27 A).
return
(visable)
switch
dial
switch
dial
to
move
TIME
dial
to A INTEN
the
controls.
dial.
difference
to
to B DL
POSITION
clockwise
settings
Note
modes
the
intensified
set
to
move
Y'D.
the
pulse
of
the
using
can be used
most
accurate
to
the
fastest zones.
A INTEN, use
the
intensified
Adjust
the
(or
rising
278).
dial.
Turn
to
move
the
to locate
the
setting
the
the
horizon-
of
the
following
to
make more accurate time measurements.
Time Difference =
second
dial
[
setting
EXAMPLE: Assume the second dial setting is to
0.2
µs
(see Fig. 27).
Using the formula:
Time Difference
second
[
Substituting the given values:
Time Difference
The
time
dial_
setting
difference
first
l
dial
setting
8.81
with
X A
first
dial
the A TIME/DIV
=
first
dial
X (A
setting
= (8.81 - 1.31) X 0.2 µs.
is
]
1.5
microseconds.
delay
TIME/DIV
setting
setting
delay
TIME/DIV
setting)
time
is
1.31
time
and
switch
the
set
l
@
466/464
DM44
Operators
51
DELAY-TIME
POSITION
1:31
DIAL:
DELAY-TIME
POSITION
8:81
DIAL:
TIME
DURATION MEASUREMENTS
(BASIC 466/464)
Fig.
(A) A
100
DISPLAY
I
90
I
?
VERTICAL
(REFERENCE
I I
@oR®
'
LINE
I
l
I
I
I
I
MEASURE TIME FROM A TO B
(B)
B
DL
Y'D
DISPLAY
27.
Time difference between repetitive pulses.
Obtain a Delayed Sweep Display. Set the A switch
TIME/DIV
accurate measurement, set the B
to
display a single event. Be sure the VAR
control
is in the calibrated detent. For the most
TIME/DIV
TIME/DIV
switch
I
to the
fastest sweep speed that gives a usable (visible) inten-
sified zone. Vertically position the measurement points to the
center
display
to
horizontal
place the time
graticule
line
(see Fig. 28).
Use the DELAY (left-hand edge) intersection graticule
of
line (see Fig.
POSITION dial
of
the intensified zone to just
the signal and the center horizontal
28,
Point A).
Note
to
move the start
touch
the DELAY
the
TIME
TIME
POSITION dial setting.
Use the DELAY TIME POSITION dial of the intensified zone to the second point
I
j
··-
~
(see Fig. 28, Point B). Note the DELAY
POSITION dial setting.
Substitute into the time difference
the DELAY
TIME
POSITION dial settings
formula
to
move the start
time
measurement
TIME
to obtain the time
duration.
1737·10
EXAMPLE: The A the B
TIME/DIV
TIME/DIV
switch
is
set to
switch is set
.1
ms.
to
2 ms and
52
466/464 DM44 Operators
@
]
J J
]
l
l
l
l
I
. . . .
100.
I
90
I
I
@"
(Fig. 28) is 1.20. B (Fig. 28) is 9.53.
formula:
/,
'~
I
INTENSIFIED
lv
I
I
I
I .
The
DELAY
Time
Difference
[second
setting
ir·
dial
. .
I\.
"
I I
TIME The
. .,.,
---·
Fig.
. . . .
,
...
~
'
\
\
~'\
i.-
'
\
ZONE
HORIZONTAL
DISTANCE
28.
POSITION
DELAY
To
~
I I
Time duration.
TIME
find
Time
Duration
l
first
dial
setting
....
L,.....-
_L,.....-
+---
'
dial POSITION
the
time
x
. .
@~
~
I
V.
setting
duration
delay
(A
TIME/DIV
setting)
....
1
~
J :
I I
I
..
.1.
I
I
I
I I
dial
time
I
MEASURE
TIME FROM
ATO
1737·11
at
Point at
use
B
A
Point
the
Substituting the given values:
Time
Duration=
(9.53 - 1.20) X 2
FREQUENCY MEASUREMENTS
(BASIC 466/464)
The
frequency
by
taking the reciprocal
EXAMPLE: The
Point
B,
Fig. 28)
Using the formula:
Substituting
of a recurrent
time
is
16.66
Frequency
the given values:
Frequency
signal can be
of
the
time
duration
duration
milliseconds.
= . .
=
time
16.66
of
one
1
duration
ms
ms
of
event
60 hertz
=16.66
one
(Point A to
ms
calculated
event.
l
@
466/464 DM44 Operators
53
TIME DIFFERENCE BETWEEN TWO
PULSES FROM DIFFERENT SOURCES
(BASIC 466/464)
Make sure
calibrated detent. Set the A
1.
Connect parison signal connectors
CHANNEL 1 (REFERENCE)
I
....
100
I
90
I
50%
A MPLITUDE
LEVEL
0
I
0
....
I
(A)
the
VAR
the reference signal
to
CH
2.
Connect
using probes
or
INTENSIFIED
\
....
\
. . . .
....
~\
/ -
/
v
,
~
F:::::
-0
I
. . . . . . . .
. . . .
HORIZONTAL
~
- DIFFERENCE -
TIME/DIV
Trigger
cables
I I I
ZONE
control
SOURCE
to
CH
1 and the
the signals
with
equal time delay.
/~i...
:@
I
. . . .
'
--i
')
. .
-I
I
-
-
-
-
'
is in the
switch to CH
to
the
CHANNEL2
v
I
/_
.
...
MEASURE
TIME FROM
.
...
com­input
. . . .
ATO
Set the VERT general, and signals.
29A).
Trigger TIME/DIV switch
CHOP
the
ALT
Center
Set the
SOURCE
(when possible).
VERTICAL
r
.
...
100
9
~
MODE
is
more
position
HORIZ
switch
is
each
DISPLAY
to
STARTS AFTER DELAY. Set the B
20
times faster than the A
REFERENCE
-
"""
. . . . . . . . . .. .
switch
suitable
more
suitable
of
the displays
switch
LINE
~
-
.
...
. .. .
to
either
CHOP
for
low-frequency
for vertically
to A INT
CH
2
Sl~~AL
CH 1 SIGNAL
or
ALT. In
signals
high-frequency
(see Fig.
and the B
TIME/DIV
E
E
B
CH 1 SIGNAL
10
I
~
)
(B
.....
.....
-
-
--
-
-
~It.
....,i'....:A
FOR
B
FOR
. . . . . . . .
CH CH
.....
1
2
. . . .
1737·12
I
54
Fig.
29.
Time difference between two pulses
466/464 DM44
from
Operators
different
sources.
@
J
Use
the
DELAY intensified DISPLAY
POSITION some
DELAY
zone
switch
dial
vertical reference line (see Fig. 298).
TIME
TIME
POSITION dial
to
the CH 1 pulse. Set the
to B DL
to
move the CH 1 pulse
POSITION
Y'D. Readjust the DELAY
dial setting.
or
to
rising
move the
HORIZ
TIME
portion
Note
to
the
To
find the time difference use
T tsecond
ime = dial
difference tt·
Substituting the given values:
se
mg
the
first
dial
setting
formula:
J
X (A
delay time
TIME/DIV
setting)
J
]
l l
l l
l
Use
the
DELAY
pulse
or
rising
note
the
DELAY
Substitute
the
time
difference
EXAMPLE: The A
the B TIME/DIV
POSITION line.
Use pulse dial
setting
@
The
the
to
dial
DELAY
DELAY
the
reference line.
is 7.10.
TIME
POSITION dial
portion
the DELAY
switch to
to the same reference line. Again
TIME
POSITION dial setting.
TIME
formula
move the CH 1 pulse
TIME
TIME
to
TIME/DIV
is set
POSITION dial setting is 2.60.
POSITION dial
The
POSITION
find the
switch
to
2 µs. Use
DELAY
to
move
dial settings in
time
difference.
is set
to
50
the
DELAY
to
the
reference
to
move
TIME
POSITION
466/464 DM44 Operators
the
µsand
the
CH 2
TIME
CH 2
Time
Difference=
(7.10 - 2.60) X 50
RISETIME (BASIC 466/464)
R isetime measurements use the same d u ration measurements, made between the Falltime the
or control switch
is
measured between the 90% and 10%
trailing edge
Use a sweep speed setting
events (if possible) and be sure the VAR
is
and the VAR
of
in the calibrated detent. Set the VOL
except
10%
and 90%
the waveform.
control
the
points
that
displays several
(or
signal
µs
= 225
µs
methods
measurements are
of
as
the
waveform.
points
cycles
TIME/DIV
TS/DIV
amplitude)
time-
on
for
55
I
I
INTENSIFIED
..
.
I
r=
w
....10
<{
::>
z
t:
(!)...I
-
4.
I/):!:
<{
~@"
0
L
f-J
exactly a five-division display. Set vertical positioning so the display bottom touches the display
ZONE
~
top
touches the 100% graticule line (see Fig. 30).
_C
.........,;,.;.
~""
J,
~,
~'
,.
....
HORIZONTAL
DISTANCE
Fig. 30. Risetime.
I
~®~'\.
~-
~"
/
.............
i...,..o""
...
. '
---
0% graticule line and the
MEASURE
TIME FROM
ATO
B
1737·13
Use the DELAY
of
the intensified zone (left-hand edge) intersection Fig.
30,
Point
setting.
Use the DELAY
of
the intensified zone signal and the
Note the DELAY
Substitute the DELAY
the time difference
EXAMPLE: The A
DELAY
2.50 The DELAY
(Fig.
30)
TIME
is
7.50.
TIME
POSITION dial
of
the signal and the 10%
A).
Note
the DELAY
TIME
POSITION dial
to
just
touch
190%
graticule
TIME
formula
TIME/DIV
POSITION dial setting at
TIME
POSITION dial setting at
To
find the risetime use the formula:
line (see Fig. 30,
POSITION dial setting.
TIME
POSITION dial settings in
to
find the risetime.
switch is set
to
move
the
to
just
touch
graticule
TIME
to
the intersection
point
line (see
POSITION
move
the
Point
to
1 µs. The
A (Fig. 30) is
start
dial
start
of
point
the
the
B).
B
I I
E
E
Set the A
with
the risetime spread Horizontally waveform intersects the second vertical graticule line (see Fig.
30). Set the B
speed that provides a usable (visible) intensified zone.
56
TIME/DIV
position the
TIME/DIV
switch horizontally
for
display
switch
so the 10%
a single-event display
as
much
as
possible.
point
of
the
to
the fastest sweep
466/464 DM44 Operators
Time
.ff =
1 erence
D
second first ]
dial dial X
[
setting setting
. .
R1set1me
=
delay time
(A
TIME/DIV
setting)
@
I I I
]
Substituting
Risetime
= (7.50 - 2.50) X 1
the given values:
µs
= 5 ms.
switch, it is ensure location
not
necessary
of
the
correct
to
pulse.
switch
display
modes
to
l
]
]
l l
]
J
MIX {BASIC 466/464)
For
the
MIX procedures a sweep rate set part
of
can be used. With
the
display
mode
of
operation,
the
by
the A TIME/DIV
at a sweep rate set
first
switch
the
same general
part
of
the
and
by
the B Tl
display
the
second
ME/DIV
However, inaccuracies are measurement DL
at
Y'D recommended mode measurements.
mode
by
the transition from A
is the
most
of
making
introduced
to
accurate
into
B sweeps. The B
and therefore
differential
time
the
J
@
466/464 DM44 Operators
..
.. : a
LU,
57
.t
JU
DM44 DELAYED OR MIXED SWEEP TIME MEASUREMENTS
Most measurements
TIME
the operating modes and the
function
of
the DM44 and the delayed-sweep mode of the oscilloscope. Table 5 lists the DM44 and
of
time, time-duration, frequency, time difference and risetime are
crt
display
obtained in these modes.
TABLE 5
DM44 Delayed Sweep Displays
more
easily performed using
oscilloscope
I
I
DM44
FUNCTION
VOLTS OHMS,
OR
TEMP
1
1n
the B DLY'D mode, the
58
HORIZ
DISPLAY
1
A
INT
MIX
intensified
CH
OR
CH
OR
ALT
VERT
MODE
1,
CH
ADD
ALT
or
CHOP
1,
CH
ADD
OR
CHOP
zones
DISPLAY
2,
One intensified zone, DELAY
Two
intensified zones, one on each channel display. Intensified zones are coincident in time. DELAY zones.
2,
One mixed display. Position TIME POSITION.
Two mixed displays, one on each channel display. Transition points are coincident
TIME POSITION.
(that
are displayed in the A
466/464 DM44 Operators
in
time. Position
INTEN
of
mode)
OBTAINED
TIME
POSITION moves intensified zone.
TIME
POSITION moves both intensified
of
transition
transition points determined by DELAY
will
point
determined by DELAY
be displayed at the B sweep rate.
@
E
E
I
I I
I
]
]
DM44
FUNCTION
HORIZ
DISPLAY
TABLE 5
DM44 Delayed Sweep Displays
VERT
MODE DISPLAY OBTAINED
Two
CH
OR
1,
ADD
CH 2
intensified zones. DELAY TIME POSITION moves both intensified
zones. A TIME moves
(cont)
only
one intensified zone.
]
]
J J
l
J
J
2
INT
A
TIME
OR
1/TIME
MIX
2
Your
instrument
instructions
@ 466/464 DM44 Operators 59
.!
for
may
making
l .! LJU
be
CH
OR
modified
this
modification
@ts!,.,
ALT
CHOP
1,
CH
ADD
ALT
CHOP
to
j .. ,
..
Two
intensified zones. Reference Measurement moves both intensified zones. A TIME moves (on the CH 2 display).
Two
pairs
are
coincident
four
intensified zones.
Two
2,
make
,~
.... au
are
mixed displays on the same trace. displays overlap.
Two
mixed displays, one on each channel display. DELAY TIME POSITION moves transition CH 2 display only.
Four mixed displays, since mixed displays overlap.
the
DELAY
located
..
,
jJ)LL
point
appears on CH 2 display. DELAY
of
intensified zones, one pair on each channel display. The pairs
in time with each other. DELAY
!::i.
TIME moves two intensified zones.
point
of
both displays. A
two
on each channel display.
TIME
in
POSITION
the Maintenance section
and ll.
point
appears on CH 1 display.
only
the measurement
TIME
Not
generally used since mixed
TIME
moves transition
Not
TIME
controls
of
the DM44 Service manual.
operate
a
..
TIME
POSITION
point
POSITION moves all
point
of
generally used
independently. The
j
,.
it"
TIME
DIFFERENCE BETWEEN
REPETITIVE PULSES (DM44)
Set the
31A. Use the DELAY
adjusted. Use the
control
controls
FUNCTION HORIZ B SOURCE
TIME/DIV
A B
TIME/DIV
fl
TIME
as
follows:
DISPLAY
TIME AINTEN ST ARTS AFTER DELAY To
display
3
or
cw
To
move the time­measurement the
2 pulses
4 positions more
than A TIME/DIV
point
right
of
the ref-
to
erence point
VAR
TIME/DIV
Position the display
approximately
TIME
point
to
reference
the first pulse.
will move when the DELAY
fl
TIME
measurement
Set the readjust the waveforms (see Fig.
point
HORIZ
fl
to the second pulse.
DISPLAY switch to B DLY'D. Slightly
TIME
31
B). The DELAY TIME POSITION
may need to be
Detent position
POSITION
TIME control
control
slightly
readjusted to keep the
as
shown in Fig.
control
Both
POSITION
to move the
intensified zones
control
to move the time-
to
superimpose the
display on screen. The Readout and Scale Factor Lamps now
indicate
the time
difference
between the pulses.
I
100
-~----+-----+---+--+----+-~~
(A) A
DISPLAY
I i
100
90
is
(8) B
Fig.
31.
!
I
I
i
i
I
i I
I
!
I
I
I
I
0 0
:
i
I
I
I
DL
Y'D
DISPLAY
Time difference between repetitive pulses.
I I
: J:: I I J
'-0
AND©~
SUPERIMPOSED
i
I
I
I
:
--
1
I
i
I
I
i
I
+--
i I
(1737-10)2037-31
I
I
I
E
I
60 466/464 DM44 Operators
@
]
To
find
waveforms
1/TIME.
indicate
The Readout and Scale Factor Lamps
the
the
pulse repetition rate,
as
above and set the
repetition rate.
superimpose
FUNCTION
switch
the
to
now
J
J
]
l
J
l
l
TIME
Set
controls
FUNCTION HORIZ
SOURCE
B A
TIME/DIV
B
TIME/DIV
!),.
VAR
Use the DELAY reference Point measurement line
on
B}.
@
DURATION
DISPLAY
TIME
TIME/DIV
point
A}. Use
the
next
as
follows:
TIME to a horizontal the
!),.
point
to
cycle
of
MEASUREMENTS
(DM44)
TIME AINTEN STARTS AFTER DELAY To
display event. or 4 positions
3
cw than A
To
move the measurement the erence point.
Detent
POSITION
graticule
TIME
control
near
the
same
the waveform (see Fig. 32A,
a single
TIME/DIV
time-
right
of
the ref-
position
control
horizontal
to
line (see Fig. 32A,
to
move
the
466/464 DM44 Operators
more
point
to
move the
time-
graticule
Point
(A) A SWEEP
(B) B
Fig.
32. Time duration and frequency measurements.
DLY'D
DISPLAY
SUPERIMPOSE
PORTIONS OF
DISPLAY
L
WAVEFORM
2039-31
61
Set the readjust waveform POSITION keep the Lamps
the
display
indicate
HORIZ
fl.
TIME
portions
control
on screen.
the
DISPLAY
control
(see Fig.
may need
time
duration.
switch
to
The
to
8 DLY'D.
superimpose
328).
to
be
slightly
Readout and Scale Factor
the displayed
The
DELAY TIME
readjusted
Slightly
to
A Time/Div Set
Multiples of Zones
2
5
To
Decade Spacing Between Intensified
Less
Than
0.25 DIV
0.5 DIV
1.0 DIV
I
I
FREQUENCY MEASUREMENTS (DM44}
To
measure frequency, use the same procedure
Time
Duration switch Readout and Scale Factor Lamps
The Scale Factor Lamps follows:
This
62
to
1/ms Lamp
OFF
ON
OFF ON
A
blinking occurs
Measurements
1/TIME. With the
1/µs Lamp
display
on measurements as follows:
indicates an over-range condition.
except
display
indicate
indicate
OFF OFF
set the FUNCTION
superimposed the
the frequency.
the scale
Multiplier
as
factor
Hz
kHz
MHz
466/464 DM44 Operators
for
as
TIME DIFFERENCE BETWEEN
DIFFERENT SOURCES (DM44}
Set
controls
A SOURCE
MODE
VERT HORIZ
SOURCE
8 8
TIME/DIV
VAR
TIME/DIV
Connect comparison to
the
inputs
delays.
TWO PULSES FROM
as
follows:
CH 1 ALT
DISPLAY
the
reference signal
signal
to
with
the CH 2 input.
cables
AINT STARTS AFTER DELAY 3
or 4 positions
cw than A
Detent
to
or
probes having equal time
position
the CH 1
Connect
more
TIME/DIV
input
and the
the signals
@
I
I
I
I
]
] ]
Adjust
the
DELAY
point
the
to
point
/:::,,.
point
Point
reference display the time-measurement
measurement (see Fig. 33A,
(see Fig. 33A,
reference
Adjust
TIME
POSITION
the desired
Point appears point
appears on
TIME
control
to
the desired
B).
control
spot
on the reference (CH
A). In the
on
the CH 1
ALT
the
to
spot
on the CH 2
display
CH 2 display.
move
to
move the
1)
vertical mode
while the
the
time-
display
Set the readjust the DELAY to
superimpose the waveforms (see Fig. Readout and Scale difference.
HORIZ
DISPLAY
TIME
POSITION
Factor
switch
Lamps
to
B DLY'D.
and/:::,,.
indicate
TIME
338).
Slightly
controls
The
the
time
l
l l
l
l
l
MA.lU.
CHANNEL
I
100
I
90
I
MPLITUDE
A
LEVEL
. . . .
50%
1 (REFERENCE)
\
. . .
·\
-
/
I'
INTENSIFIED
ZONES
.
...
/..
/
/
~
\
~
;;:::::::
0
I
. .
0
..
I
(A)
@
j
•'a
...
J
..
. .
tl!l$£
. . . .
i----
....
£2#Cti2
t--
~@
- DIFFERENCE -
~
t--
~
I I
.
...
. . . .
HORIZONTAL
Fig.
...
J.MtJ
=@
. . . .
..
k.
...
33.
\
a.l,
CHANNEL2
v
I
/_
....
....
....
-
MEASURE
TIME
)
.
.........
.
...
I
I
--i
Time difference between
..
J,J,,Jl~
.....
FROM
ATOB
466/464 DM44 Operators
Jfa#~t.~,.,ua
....
two
JUJ@.lk
I
. .. .
.
100
9
~
CH 1 SIGNAL
0
I
0
I
... . ...
. .. . . . . .
8)
pulses from different sources.
a
..
21
2££L&JLZ~,JJ&,,~
..
L
..
2.l
....
-
"It.
...
...
. .
-
-
... · LL
CH
2
CH 1 SIGNAL
"i'.
SUPERIMPOSE
LEADING
.
...
..
Sl~~AL
I
EDGE -
. . . .
.. . . . . . .
(1737
12)2037-33
63
Set
controls
FUNCTION HORIZ B SOURCE
TIME/DIV
B
!:::..
TIME
RISETIME (DM44}
as
follows:
DISPLAY
TIME AINTEN
ST ARTS AFTER DELAY
or
4 positions more
3
cw
than A
To
move the time­measurement the
erence
right
TIME/DIV
of
the ref-
point
point
to
Set the A several events. Set the VOL so
the Vertically 0%
graticule line.
Adjust
reference
A).
Adjust
ment
point
The
Readout and Scale
risetime.
TIME/DIV
amplitude
position
line and the
the DELAY
point
to
the!:::..
to
the 90%
of
the
the
TIME
the 10%
TIME
switch
display
display
top
control
graticule
to
a setting that
TS/DIV
POSITION
graticule
and VAR
is
so the
touches
to
move the
line (see Fig. 34,
Factor
displays
VOL
TS/DIV
exactly
bottom
line (see Fig. 34
Lamps
the
100%
control
time
5 divisions.
touches
graticule
to
move the
measure-
Point
indicate
the
Point
B).
the
I
I
I
I
w
..Jo
<(
:J
z!::
(!)..J
- 0..
en::!:
<(
~@"
0
I
L
o-
I
64
I I I_
I
NT
ENS I Fl ZONES
,J"
..
"
LI
----
I
ED
~®·~
-...:..;.;,..
,J'7
......
.
~
v
~"'
,J,
_
H~~~~~~:L
Fig. 34. Risetime.
-
~
(1737-13)2037-34
For the MIX
MEASURE
TIME
FROM
ATOB
....
466/464 DM44 Operators
procedures a sweep rate set by part switch, it is ensure location
However, inaccuracies are measurement by DL Y'D recommended measurements.
of
mode
can be used. With
the
display
not
necessary
of
the
mode
is the
mode
MIX (DM44)
of
operation,
the
first part the A TIME/DIV at a sweep rate set
the
correct
transition
most
accurate and
of
switch, and
to
switch
pulse.
introduced
from A
making differential
the
same general
of
the
by
the B
display
to
B sweeps.
therefore
display
the
second
TIME/DIV
modes
into
the
The
the
time
at
to
B
@
I
I
I
I
]
]
+40°C
Refer
unless
to
the
otherwise
service
OPERATOR'S SPECIFICATIONS
manual
stated.
for
complete
specifications.
Specifications
given are
for
an
operating
range
of
0° C
to
] ]
J
l l
J
J
STORAGE
Stored Writing Speed
VAR PERS Full Scan Speed: 0.5
X 8
divisions. control time.
466 and 464.
FAST
divisions
FAST Full Scan Speed: 150
divisions
FAST Reduced Scan Speed: 3,000
center
VAR PERS Reduced Scan Speed: 3
center
@
It
is necessary
to
compromise
Full Scan Speed:
(0.9
cm/div).
(0.9
cm/div).
8 X 10
divisions
8 X 10
divisions.
to
use
between
11 O div/µs
464
only.
div/µs
466
only.
(0.45
cm/div).
It is necessary
writing
466/464 OSCILLOSCOPE
STORAGE speed and view time. 466
div/µs
over
the
the
STORAGE
speed and view
over
the
over
the
div/µs
466
only.
div/µs
center6
LEVEL
center
center
over
over the
to
use the
466/464
6 X 8
6 X 8
the
Storage View Time
FAST and VAR PERS View Times:
These times are at full stored extended at least the
SAVE Display mode.
Deflection Factor Accuracy:
position.
Frequency Response: De
and
CH
or
less. Use
1 Hz.
DM44
Operators
LEVEL
2.
Ac-coupled,
of
a 10X
control
25
to
times
compromise
only.
by
VERTICAL
to
low-frequency
probe
extends
between
At
least 15 seconds.
display
using
Within
at least 100 MHz
intensity.
reduced
3% in
the
response is 10 Hz
frequency
response
writing
They
intensity
calibrated
for
CH 1
are
in
to
65
..
J
Risetime: 3.5 nanoseconds
-:-
bandwidth in MHz).
0.35
Maximum Input Voltage:
or
500 V p-p ac at 1 kHz
ac) + peak ac)
or
500 V
p-p
or
less (calculated from
De-coupled
or
less; ac-coupled, 500 V (de
ac 1 kHz
or
250 V
less.
(de+
peak
TRIGGERING
Sensitivity
Ac Coupled Signal: 0.3 div internal
to
25
from 30 Hz
150 mV external at 100 MHz.
MHz; increasing
or
50 mV external,
to
1.5 div, internal
I
or
I
Positive-Going Step Aberrations: Less than +3%,
not
to
exceed
Common-Mode Rejection Ratio
Inverted): At least
signals CMRR at
Input Gate Current: 0.5
5 mV/DIV), from
Channel Isolation: At least 100:1 at
CHOP
Cascaded Operation
Approximately 50
MHz, coupled, using a 50 50
n at CH 2 Input.
66
3%
peak-to-peak,
10:1
of
6 divisions
50
kHz.
Mode Repetition Rate: Approximately 250 kHz.
with
CH 2
or
-15°C
1 mV/DIV.
OUT
0,
42-inch
excluding
at 20 MHz
less
with
nA
to
+30°C.
(CH 1 OUT
Bandwidth
connected
BNC
the ADD mode.
(ADD
Mode with CH 2
for
common mode
GAIN adjusted
or
less
25
MHz.
int.o
CH
is de to at least
to
CH 1 input. AC-
cable, terminated in
-3%,
for
best
(0.1
div at
2)
sensitivity:
466/464 DM44 Operators
LF
external, from
internal
signals below about 50 kHz.
HF
external, from 30 Hz to 50 kHz. Attenuates signals below about
DC Coupled Signal: 0.3 div internal from de to external at 100 MHz.
EXT
tiplied
External Trigger Input
Maximum Input Voltage: 250 V
p-p ac
REJ
Coupled Signal: 0.5 div internal
50
kHz to
or
300 mV external at 100 MHz. Attenuates
REJ
Coupled Signal: 0.5 div internal
50 Hz and above
25
MHz; increasing
-:-
10 Signal:
by
10.
(1
kHz
or
less).
25
MHz; increasing to 1.5 div
about
50
to
Amplitude
kHz.
or
50 mV external,
1.5 div internal
requirements are
(de+
peak ac)
or
or
or
or
100 mV
50 mV
150 mV
mul-
250 V
@
I
I
I I
I
I
I
J
]
]
J
l l
J
J
Level
Control
At
least+
-20
V,
40
Trigger
Deflection Factor:
500 mV/div in EXT-:-
Risetime: 5 ns Trigger
center.
Range in Ext
and
-2
V p-p.
View
or
Centering Point: Within 1.0 division of screen
DIFFERENTIAL
MEASUREMENT
Accuracy Dial Division: Within ±2.5% from
Accuracy
Division: Within ±0.01
+35°
C.
+55°C.
for
Measurements Greater than One
-15°C for
Measurements Less than One Major Dial
Within ±0.03
to
DIFFERENTIAL TIME
MEASUREMENT (DM44)
Refer to TIME
MULTIMETER.
AND
V,
4 V p-p; EXT-:- 10
About
50 mV/div in EXT and about
10.
less.
(BASIC
±1%
from
+15°C
+55°C.
major
dial divisions from
major
dial divisions from
1/TIME
is
at
TIME
466/464)
to
+35°C.
under DM44
least+
Major
Within
+15°
-15°
DIGITAL
and
C to C to
HORIZONTAL
Sweep Rate Accuracy:
magnified, from
Mixed Sweep Accuracy: With in Sweep inaccuracy, when viewing the A Sweep portion remains the same
Trigger time
by at least a factor
Delay Time Jitter: 1 part,
10 times the A TIME/DIV switch setting.
Calibrated Delay Continuous of
A (delaying) sweep.
X-Axis Sensitivity (X10
vertical system.
X-Axis Bandwidth:
10 division reference signal.
Phase Difference Between X-
Within 3° from de to 50 kHz.
Deflection Accuracy: Within
+20°
Holdoff
from
0.2
Within
2%,
unmagnified, and
C to
+30° C for
Variable: Increases A Sweep
of
10.
or
Time
µstoat
(VAR
least 5 seconds after the start
A and B Sweeps.
2%
plus the measured A
as
above.
less, in 50,000 (0.002%)
control
X-V
MAG
turned
De
to at least 4 MHz,
and
4%.
3%
portion
off):
Y-Axis Amplifiers:
only. B
to
Same
holdoff
of
CAL):
as
the
with
a
l
@
466/464 DM44 Operators
67
CALIBRATOR
Output Voltage: 0.3 V
+20° C to
Output Current: 30
+30°C.
Repetition Rate:
+30°
C.
mA
About
within
1 kHz.
Z AXIS
Sensitivity:
intensity signal decreases intensity.
Frequency Range (Usable): De
Noticeable
settings,
by
intensity
a 5 V
1%
and
within
0.3% from
within
2%
from
+20°C
INPUT
modulation,
p-p
signal. A positive-going
to
50 MHz.
to
at normal
AC POWER SOURCE
Regulating Ranges:
115
v 230 v
Low Medium High
Line Frequency: From 48 Hz
Maximum Power Consumption: 100 watts at 115
60 Hz, medium range.
99 V 104 V 108 V
to
121
V
to
126 V 207 V
to
132 V 216 V
to
440 Hz.
198 V
to to to
242 V 253 V 264 V
V,
I
I
I
E
Maximum Input Voltage: 100 V (de plus peak ac)
100 V p-p ac at 1 kHz
or
less.
OUTPUTS
CH
1 Out
Output Voltage:
25
mV/div
Bandwidth: De DC
A+
Output Voltage:
Output Resistance:
68
Level:
and
into
B+
50
About
Gates
At
least 50
ohms.
to
at least
0 volts.
About
5.5 V
About
mV/div
50
500 ohms.
into
1 MO;
MHz
into
50
of
positive-going pulse.
ohms.
466/464 DM44 Operators
to
at least
or
ENVIRONMENTAL
Operating Temperature:
Operating Altitude:
temperature decreases 1° C/1,000 feet, above 5,000 feet.
Humidity (Operating and Storage): 5 cycles (120 hours)
to
referenced
MIL-E-16400F.
-15°C
To
15,000 feet.
to
+55°C.
Maximum
operating
@
'
I
I
I
DM44
DIGITAL
MULTIMETER
J J
]
l l
RESISTANCE
Maximum Safe Input Voltage: 120 V rms
COM inputs.
Resistance Accuracy: 200 0
count, plus probe resistance; 2 kO, MO
ranges-within
0.3%
±1 count.
Temperature Dependence:
ranges-250 350
ppm/°C.
Resolution: Recycle Time: At least 3 measurements/second. Response Time: All ranges within 1 second except 2
MO and
20
MO
0.25%
ppm/°C;
0.1
200
0.
(within 5 seconds).
range-within
±1count;20
20
kO,
0,
2
kO
20
and
kO,
MO
200
between+
0.25% ±1
200
kO,
range-within
kO
and 2
20
MO
ranges-
TIME
Accuracy:
count.
-15°C
+15°C
to
+35°C;
to 55°C; within
within
2.5%
1%
of
reading ±1
of
reading ±1 count.
1/TIME
Accuracy:
count.-15°C
+15°C
to
+55°C;
to
+35°C;
within
within
3.5%
2%
of
reading ±1
of
reading ±1 count.
and
and 2
MO
TEMPERATURE
Range:
Accuracy: Original
+125°C.
Replacement Probe: Accuracy will equal original probe
accuracy after DM44 is compensated.
Maximum Safe Voltage on Measurement Surfaces:
±100
Temperature (Storage and Operating): Probe Body and
Cable,
+150°C.
-55°C
Within 3°C,
V (de + peak ac) above chassis ground.
-55°
to
+150°C
+125°C
C to +105°
in 1 range.
Probe-Within
to
+150°C.
C.
Probe Sensor Tip,
C,
-55°
-55°
DC VOLTAGE
Maximum Safe Input Voltage:
between + and chassis.
Common Floating Voltage:
chassis.
DC Voltage Accuracy: Within 0.1%
count.
±1200 V (de+
±500 V (de+
of
peak ac)
peak ac) to
reading,
C to
C to
±1.
J
..
Jdfah
REV.
A,
DEC. 1977
JJi!!iiJZJUtL0$2lk i . !AQMIJ#k.
ct
t.
bJ#IJl#ZHJJLL. J
466/464 DM44 Operators
..
.l!JL
$%,k,J
I Jut!
69
..
,J
L.
d.#L.444
4L
U&
t ..
!l#U!!
4Jt
Temperature Dependence:
Resolution: 100 µV. Recycle Time: At least 3 measurements/second. Response Time: Within 0.5 second. Normal/Common Mode Rejection Ratio:
Normal Common
50
Hz and 60 Hz.
Input
Mode-At
Mode-At
Impedance: 1 O MO.
44
least 60
least 100
ppm/°
dB
at
dB
C.
50
at de;
Hz and
60
80
ACCESSORIES
Hz.
dB at
Service Manual (466) 070-1753-01
1 Service Manual (464) 070-1653-01
2 Fuses, 1.5 A 3AG fast-blow
0.
75
2 Fuses,
Filter, Blue Plastic (installed) CRT Filter, Clear Plastic 337-1674-01 Adapter, Ground Wire 134-0016-01 Pair, Test Leads (DM) 003-0120-00 Service Manual (DM44) P6430 Temperature Probe(DM44)
A 3AG fast-blow 159-0042-00
159-0016-00
337-1674-00
070-2036-00 010-6430-00
OPTIONAL ACCESSORIES
I
I
I
STANDARD ACCESSORIES
INCLUDED
2 P6062B Probe Packages 010-6062-13
Accessory Pouch, Snap (w/o DM) 016-0535-02 Accessory Pouch (DM) 016-0594-00 Accessory Pouch, Operator's Manual 070-2037-00
70
Zipper
016-0537-00
466/464 DM44 Operators
C-5A Option 2 low cost fixed focus
C-5A Option
Protective
0365-00.
Folding Polarized Viewing
Folding Viewing Hood,
0592-00.
2. Cover-Waterproof,
Hood-Order
light-shielding-Order
Camera.-Order
blue
vinyl-Order
016-
I
016-0180-00.
016-
REV FEB 1982
tttmrre
T·trnrtrmrtttrtr
twY····r··
r r
·:··
[ [
[ [ [ [ [ [
Folding Viewing Hood,
0566-00.
Mesh
Filter-Improves
378-0726-01.
SCOPE-MOBILE
aisle space, has storage area in
Cart-Occupies
light-occluding-Order
contrast and
EMI
less than
base-Order
filter-Order
18
200.
016-
inches
OPTIONS
Your instrument may contain the following options:
OPTION 1
Deletes DM44 temperature probe from the DM44.
.OPTION 4
The instrument specifications on radiated interference requirements. There
is
no change
is
modified to meet certain
in
operating instructions.
Test Lead
grounding
Includes retractable hook tip and Cl tester probe cover May be used with other miniature probe
Order 012-0427-00.
Set-1
clip, 1 red lead with banana
black lead with banana plug and
plug
and probe.
tip
accessories.
OPTION 7
At
the time
do
not have the Option 7 available.
Option 7 permits operation on performance deterioration. against damage due to connection mode
of
with conventional de power (marine and aircraft).
of
this writing, instruments having the DM44
12
Circuitry
operation. The 24-voltexternal
of
is 24
provided
or
24 Vdc with no
to
protect
V when in the 12 V
input
permits use
..
@
466/464 DM44 Operators
71
Loading...