0.05
features
magnifier
nanoseconds/division.
wide
power
Storage
using
The
dual-channel
calibrated
triggering
deflection
provides
microsecond/division
for
variation
consumption
466 and 464 are
The
466
with
system does
writing
Fast mode.
viewtime is
display
accurate
extends the
intensity-extending
reduced
deflection
The
sweep
over
system.
calibrated sweep rates
of
line voltages and frequencies.
BEFORE OPERATING
dual-channel
storage
a
(in the Fast-Reduced Scan mode).
speed is
intensity
dc-to-100
the
full
relative-time measurements. A X10
The
is
about
system
writing
not
up
to
The
remaining
greater
in the Save mode.
factors
trigger
bandwidth
The
along
calibrated
instrument
100 watts.
provides
speed
have the reduced scan
110
divisions/microse-
than
to
MHz
vertical system
from 5 millivolts
circuits
capabilities
horizontal
from
with
delayed sweep
sweep rate
operates over a
portable
storage
up
to
3000
features are
15 seconds at
more than 6
to
are capable
of
deflection
0.5
second
to
Maximum
of
the
to
The
Tektronix
ohm
to 20 megohms, 0 de
(using the temperature probe)
measurement is displayed
the
oscilloscope operates
The digital
provide a digital readout
points
ment
1/TIME
frequency.
5
phase power source with
conductors
potential. Operation from
current-carrying
5
ground
is
over-current
on the oscilloscope display.
points are displayed on the
The instrument is designed
(such as phase-to-phase on a
not
recommended, since
multimeter
function
SAFETY INFORMATION
(the Neutral
(fuse)
DM44
Digital
volt
on a 3-1/2
normally.
and
of
can
provide
Conductor)
conductors
protection
Multimeter
to
1200
-55°C
oscilloscope
the
time
crt
direct
to
operate
one
of
power
are live
only
the
within
measures 0
de
volts(+
to
digital
between any
Both
at
the
measurement
the
current-carrying
at
sources where both
with
three-wire
Line
the
or-),
+150°C.
display
time
same time.
from a single-
ground
Conductor
instrument.
while
combine
measure-
(earth)
respect
system)
or
The
to
two
The
of
to
has
,,
l
@
UL
..
_. l .......
JLLJLk
"~
.....
J.
466/464 DM44 Operators
li
.fa.
..
.
Li
. L
..
Sit.
1
The
instrument
three-terminal polarized plug
source and safety-earth. The
the plug
electric-shock protection, insert this plug
outlet
is
directly
with a safety-earth
has a
three-wire
connected
contact.
power
for
connection
ground
to
(earth) terminal
the
instrument frame. For
cord with a
to the power
only
in a mating
panel)
to
indicate the
230 volt adapter may be required
sure to use the
of
selected (see Table 1).
correct
correct
line fuse
nominal
for
the
for
voltage. A 115-to-
line-cord
the line voltage
plug. Be
I
I
.
OPERATING VOLTAGE
This
instrument
the line voltage
selector
range
or
if
the
wrong
Line Voltage Selection
This
instrument operates
230-volt nominal line voltage source, 48 to 440 hertz. To
convert
move the Line Voltage Selector
2
from one nominal line voltage range
may
selector
set
for
line fuse
be
damaged
the
wrong
is
switch
used.
from
switch
if
operated with
or
the regulating
applied
either
line voltage
a 115-volt
(located on side
or
to
the other,
466/464 DM44 Operators
Regulating Range Selection
The
Regulating Range
the rear panel) is set
shown
in Table
protection.
To
change
1.
Disconnect
2.
Loosen
on
the
selector assembly; then
a
3.
Pull
Select a range
average line voltage. Slide the bar
and plug it in. Push the
the
the
out
the range
1.
It also
regulating range:
the
two
from
Selector
for
one
contains
instrument
captive screws that
selector
Table 1
cover
assembly (located on
of
the line voltage ranges
the line fuse
from the
pull
to
switch
which
is centered
to
the desired
on and
tighten
for
overload
power
source.
hold
the cover
remove the cover.
bar (see Fig. 1
about
the
position
the screws.
@
I
I
).
I
TABLE 1
Regulating Ranges
Regulating Range
Range Selector 115-Volts 230-Volts
Switch Position
LO
(Switch bar in
lower holes)
Nominal Nominal
99 to
121
volts
198 to 242 volts
@
Fig. 1. Regulating range selector & line fuse.
466/464
1737·01
M (Switch bar in
HI (Switch bar in
to meet particu lar
slot
indicates a
DM44
Operators
middle holes)
upper holes)
Fuse Size
Options
are available to
(see Item 60, C
modified
104 to 126
108 to 132
1.5
A
3AG
Fast-blow
OPTIONS
alter
applications
ontrol
s,
Connectors
oscilloscope
volts
volts
207 to 253 volts
216 to 264 volts
0.75 A 3AG
Fast-blow
oscilloscope
. A
number
, and Indi
.
performance
in
either MOD
cator
s)
3
CONTROLS, CONNECTORS AND
1.
CH 1 and
deflection
the
calibrated detent
in
2.
VOL
for
each channel, located beneath
VOL
TS/DIV
indicate the
scale-switching
connector
3.
VAR-Provides
deflection
VOL
TS/DIV
deflection
position).
CH 2 VOLTS/DIV-Selects
factor
in a 1-2-5 sequence (VAR
TS/DIV
knob. One
correct
connector
lights
the
factors between the calibrated settings
switch, and extends the
to
at least 12.5 volts per
INDICATORS
VERTICAL
control
for
the indicated
Readout-Consists
or
the
other
deflection
X1
continuously
is
lamp.
factor
used. A
deflection
of
two
the
lamp
will
when a
probe
variable
maximum
division
the vertical
must
factor).
small lamps
skirt
of
each
light
up
probe
with
without
uncalibrated
this
of
vertical
(5 volt
be
to
the
I
I
I
a
E
I
Fig.
2.
Vertical controls, connectors, and indicators.
4
4.
UNCAL
VOL
TS/DIV
vertical
X-Y mode, the CH 2 POSITION
axis (vertically) and the
positions on
466/464 DM44 Operators
deflection
5.
POSITION-Positions
Lamp-Indicates
control
is
out
of
factor
is uncalibrated.
Horizontal
the
X-axis (horizontally).
when
the calibrated
the
display
control
positions
POSITION
the
VAR
detent
and the
vertically. In the
on the Y
control
@
I
I
I
I
J
J
l
l
J
l
J
6.
CH
1 OR X
application
amplifier.
connected
tal
deflection
connector
7.
signal
position,
amplifier.
the
In
disconnected
allow
position,
the
input
8.
tical
CH
of
In
to
provides vertical deflection.
AC-GND-DC-Selects
to
the
signals are capacitively
The
GND
position,
the
input
all
components
amplifier.
VERT
MODE-Selects
amplifier
1:
Channel 1 only
and
CH
2 OR
Y-lnput
external signals to the inputs
the
X-Y
mode
of
operation, the signal
the
CH 1 OR X
and the signal connected
input
of
de
component
the
from
the
coupling
system.
connector
the method used
the vertical amplifier. In the
of
the
input
of
input
connector
capacitor
of
the
input
mode
is displayed.
provides
to
coupled
input
signal is blocked.
the vertical
and
to
precharge. In the DC
signal are passed to
of
operation
connectors
of
the vertical
the CH 2 OR Y
to
to the vertical
amplifier
grounded
for
horizon-
couple
AC
for
ver-
is
to
ALT: Provides dual-trace display
is
channels. Display
of
each sweep. Useful at sweep rates faster than
microseconds/division.
ADD: Signals applied to the
connectors are algebraically added, and the algebraic
sum
is
displayed on the crt.
2 allows the display to be CH 1 plus CH 2
CH
2.
a
Useful
undesired signal
CHOP: Provides dual-trace
both channels. Display is switched between channels
repetition rate
rates slower than about 50
when a dual-trace, single-sweep
2:
Channel 2
CH
in X-Y operation.
9.
20
MHz BW/TRIG
limits the bandwidth
approximately
causes the signal applied to A
displayed on the crt.
10.
INVERT-Channel 2 display
INVERT (button in) position.
switched between channels at the end
The
for
common-mode
or
for
de offset.
of
approximately 250 kHz. Useful at sweep
only
is
displayed.
VIEW-Dual-purpose
of
the vertical
20
MHz when pulled,
of
the signals
CH
1 and CH 2
INVERT switch in Channel
or
rejection
display
microseconds/division,
display
Trigger
to
of
the
is required.
It
must be selected
amplifier
or
when pressed,
Generator
is
inverted in the
of
both
about
input
CH 1
minus
remove an
signals
switch
system
to
50
of
at
or
that
to
be
a
l
@
1 I
XL.L
..
ULUAJlJ!
LLlllqi
466/464 DM44 Operators
.J$£lUJJtJ
..
faL.
LJ
5
j
..
.iJ
£
....
CJ
..
L.JLJH
..
ktl.L
....
.£
24.
A
L
DISPLAY AND STORAGE
I
I
I
I
m
I
I
I
Fig.
3.
Display and storage controls,
6
466/464 DM44 Operators
connectors,
and
indicators.
@
I
J
]
l
11a.
Internal
amplitude
left-hand
11b. Reduced Scan Graticule (466
operating
set
by
the VOL
referenced
measurements should be made
operating
BEAM
12.
the
graticule
applied
signals and provides a visible viewing level.
Graticule-Eliminates
and measurement points are indicated at the
graticule
in the reduced scan
to
in the reduced scan mode.
edge.
mode
TS/DIV
the smaller divisions
FINDER-Compresses
area independently
and
TIME/DIV
within
parallax. Risetime,
Only)-When
the deflection factors,
switches, are
of
this
graticule. All
this
graticule
the display
of
display position
to
within
when
19. REDUCED SCAN
reduced scan mode
20. STORAGE
crt
in the FAST and
21.
accidental erasure
22. SAVE
mode.
or
23.
as
a conventional oscilloscope in the NON STORE mode.
is
LEVEL-Varies
VAR
SAVE-Provides
of
INTEN-Varies
NON
STORE-Allows
LAMP-Indicates
in operation. 466 only.
the
writing
PERS storage modes.
longer viewing time. It prevents
the stored display.
the intensity
operation
of
that the
rate
of
the SAVE
the instrument
of
the
J
J
l
J
l
13.
INTEN-Controls
FOCUS-Adjusts
14.
15.
SCALE
16.
ASTIG-Used
control
readjustment
the
Reduces display
an increase in
@
to
TRACE
17.
horizontal
18. REDUCED SCAN
ILLUM-Controls
obtain a well-defined display. It does
in normal use.
ROTATION-Adjusts
graticule
to
writing
brightness
for
optimum
in
conjunction
lines.
(PULL
the small reduced scan graticule with
rate. 466
of
the
display definition.
graticule
with the FOCUS
trace
-INTEN
only.
crt
display.
illumination.
not
require
to
align with
knob)-
466/464 DM44 Operators 7
24. VAR
stored display.
25.
disables the TRIG MODE switch. It automatically sequences
in the full clockwise detent (manual).
determined by the VIEW TIME
sweep in a single-sweep mode.
26. VIEW
sistence)
varies the time between erase cycles in FAST mode.
27.
SAVE
PERS-Permits
FAST-Used
an
erase cycle, unless the VIEW TIME
of
the stored display in the VAR PERS mode. It
ERASE-Erases
mode when ERASE
for fast-writing displays. FAST mode
TIME-Varies
variable retention of. the
It waits a time period
control;
the retention time (per-
the stored display, except in
is
disabled.
then resets the
control
is
Fig.
4.
DM44 controls, connectors, and indicators.
I
I
I
I
I
I
I
I
8
466/464 DM44 Operators
@
I
J
J
l
J
J
DM44
28.
Input
Connectors(red) and
only.
29.
temperature
30.
matic
polarity
display
ed
by
operation
the
31.
(1200 volts is the maximum safe
mode);
32.
1/TIME,
COM
(black) inputs
Probe
Readout-Negative
for
indicates overrange. The decimal
the
TIME
RANGE-Selects
from
FUNCTION-Selects
Connector-Input
probe.
negative de voltage and temperature with
indication
FUNCTION
and
or
or
TIME
for
by
the oscilloscope A
1/TIME
200 ohms
functions
Two
for
polarity
positive measurement. A
and RANGE
modes.
from
.2 V to
to
20 megohms in 6 ranges.
VOL TS, OHMS, TEMP (° C),
of
the DM44.
banana jacks provide +
voltage and resistance
connector
indication
point
controls
TIME/DIV
1.2 kV de in 5 ranges
input
in the 1.2 kV de
for
is
blinking
is
position-
for
multimeter
switch in
the
auto-
no
In the 1 /TIME function the Readout and Scale Factor
of
Lamps indicate the number
of
time. One lamp
per
ms
(1/ms lamp)
lamp lit indicates intervals per second. If the duration
one event
Factor Lamps indicate frequency. The
dicates kHz, the 1/µs lamp indicates MHz, and
indicates Hz.
34. ll.
TIME-Used
TIME
POSITION control in the
tions. The ll. TIME
measurement point. The DELAY
moves both the reference
measurement point. With the time-measurement
the left
negative time difference.
of
or
the other will
or
is
being measured, the Readout and Scale
the reference
measured intervals per unit
light
to
indicate intervals
intervals
in
conjunction
control
point
per
µs
(1/µs lamp). No
1/ms
with the DELAY
TIME
and
moves
TIME
point
the Readout indicates a
only
POSITION
and the time-
lamp in-
no
1/TIME
the time-
point
lamp lit
func-
control
of
to
l
33. Scale Factor
the
indicate
light
TIME
zones on the
to
In
Lamps
intensified
will
seconds.
@
Lamps-
function,
the time difference between the
indicate ms
Two
lamps.
the
Readout and Scale Factor
crt
display. One lamp
or
µs.
No
lamp lit indicates
..
t
You
two
or
the
other
466/464 DM44 Operators
TIME POSITION
point.
located in the Maintenance section
Service manual.
NOTE
can
modify
The
your
instrument
control
procedure for making this
to make the DELAY
move
only
modification
the reference
is
of
the DM44
9
TRIGGER I
Fig.
5.
Trigger
controls,
connectors,
and indicators.
35. TRIG
operation
AUTO: Sweep
In
the absence
trigger
sweep free runs and provides a
NORM: Sweep
In the absence
trace. When the
NORM.
SINGL
Sweep operates in the single sweep mode.
sweep
until
the SINGL SWP
when the signal to
amplitude, shape
tional display. It can also
nonrepetitive signal.
36. READY
upon receipt
single-sweep display.
MODE-Determines
for
A Sweep. I
is
initiated by the applied
of
an adequate
repetition rate
of
SWP: When this
is
displayed,
Lamp-Indicates
of
an adequate
is
less than
is
initiated by the applied
an
adequate
trigger
be
rate is
pushbutton
further
or
sweeps cannot be presented
button
displayed is
time causing an unstable conven-
the mode of
trigger
bright
trigger
too
is
again pushed. It
not
be
used
A Sweep is
trigger
signal,
about
reference trace.
signal, there is
low
is
repetitive
to
signal, will present a
trigger
trigger
signal.
or
if the
20 hertz, the
trigger
for
pushed, the A
photograph
signal.
AUTO use
After
a single
is
or
varies in
"armed"
useful
and,
I
no
I
a
I
I
10
466/464 DM44 Operators
@
I
J
J
]
l
J
J
l
37.
TRIG
Lamp-Indicates
and
will
produce
the
trigger
without a display
external triggers).
TRIG
38. A
of
time
between sweeps.
signals (such as
clockwise
of
B sweep
rate
for
delayed-sweep presentations and
rate signals. Use the A
possible
control.
signals
trigger
30 Hz are attenuated.
of
about
stable
complex
display
39.
COUPLING-Determines
to
AC: Signals are capacitively
generator. De is rejected and signals
LF REJ: Signals are capacitively coupled
the
trigger
50 kHz are attenuated. It
display
waveform.
a stable display. It is useful
circuits
position
to
provide the fastest possible sweep repetition
the
when a
on
the
HOLDOFF-Provides
complex
(BENDS
before using the A TRIG
trigger
circuit.
of
the
generator
De is rejected and signals
high-frequency
that
A Sweep is
trigger
crt
(for
example, when using
Allows
triggering
digital
A), A sweep is reset at the end
trigger
words). In the
controls
method used
circuit.
coupled
is
useful
triggered
for
setting up
signal is available
continuous
to
the
for
components
control
on
aperiodic
fully
low-repetition
for
the best
HOLDOFF
to
couple
input
of
the
below
about
to
the
input
below
providing
of
REJ:
HF
of
the trigger circuit. De is rejected and signals below
about
useful for providing a stable
components of a complex waveform.
DC: All components
the input
stable display
signals, except the
mode
40.
starts the sweep.
+:
portion
-:
portion
Correct SLOPE setting
display when
41.
signal at which the sweep
a
adjusted
a
COUPLING and SLOPE have been selected.
Signals are capacitively coupled
30
Hz and above
of
the trigger
of
with the trigger SOURCE
SLOPE-Selects
Sweep can be triggered
of
a trigger signal.
Sweep can
of
a trigger signal.
only a portion
LEVEL-Selects
for
the desired display after
about
50 kHz are attenuated. It
display
of a trigger
circuit.
low-frequency
combination
the slope
be
triggered
the
It is useful
is
important
of a cycle
amplitude
is
to
the
input
of
the
low-frequency
signal are coupled to
for
providing a
or
low-repetition rate
of
the
ALT
(dual trace)
switch
from
from
in NORM.
of
the
trigger
the positive-going
the negative-going
is being displayed.
point
triggered. It is usually
trigger
signal that
in obtaining a
on the
trigger
SOURCE,
is
l
@
466/464 DM44 Operators
11
42.
SOURCE-Determines
to
the
signal coupled
NORM:
not
indicate time relationship between CH 1 and CH 2
signals. However, stable
signals usually can be obtained
ALT, SOURCE
Carefully
CH
used as a
time-related.
CH
used as a
time-related.
Trigger
adjust LEVEL
1: A sample
trigger
2:
A sample
trigger
input
source is displayed signal(s). It does
to
NORM and
for
of
the signal available in Channel 1
signal. CH 2 signal
of
the signal available in Channel 2
signal. CH 1 signal
the source
of
the
trigger
triggering
a stable display.
of
by
setting VERT MODE to
COUPLING
is
unstable if it
is
unstable if it
of
the trigger
circuit.
non-time-related
to
LF REJ.
is
not
is
not
I
I
is
is
I
LINE (A
line
frequency
the
input
the line frequency
stable display
waveform.
EXT: Signals connected
connectors
be time-related
It is useful when the internal signal
undesired signals that
is useful when operating in
Trigger
signal is time-related
are used
circuit
only): A sample
is used
of a line-frequency
as a trigger
(multiple
or
when it
to
the External
for
triggering. External signals must
to
the displayed signal
could
cause unstable triggering. It
CHOP
is
12
of
the power-
signal. It is useful when
or
submultiple) to
desirable to provide a
component
is
too
mode.
in a complex
Trigger
for
a stable display.
small
or
contains
466/464 DM44 Operators
Input
Fig.
6.
Trigger,
indicators.
horizontal, and
power
controls,
connectors
and
@
I
I
I
J
l
J
J
J
l
EXT
-;-
10 (A
Trigger
signal is attenuated
STARTS
Sweep
the
DELAY-TIME
differential
obtain
with
a DM44
measurements
43. External Trigger
ternal
AFTER DELAY
runs
immediately
time
measurements
valid measurements. On
you
when using the
trigger
signals.
HORIZONTAL
44a. A AND B
TIME/DIV
rate
selects
TIME
TIME/DIV
sweep rate
operation
detent
switches
the
horizontal
switch
of
the
A Sweep
the
POSITION
switch
only. A VAR
for
calibrated A Sweep rates. When
are
.(clear plastic skirt) selects the sweep
basic delay
control}
(pull
for
the
fully
(X-axis}
circuit
by a factor
after the delay
POSITION
must use this mode
Inputs-Input
only}: External
of
10.
(8
Trigger
control.
you
must use this
instruments
TIME
or
AND POWER
TIME/DIV
circuit
time
out
B Sweep
control
counterclockwise
display
AND DELAY
for
A sweep
(to be used
for
delayed sweep operation. B
and rotate
to
circuit
must be in
to
is the
CH 1 input.
trigger
circuit
time
1/TIME
connectors
only
unlock}
for
the X-Y position,
only}: B
selected
When making
mode
equipped
to
obtain valid
functions.
for
TIME-A
operation and
by
the
DELAY-
selects the
delayed sweep
the
calibrated
both
TIME/DIV
by
to
ex-
44b. DM44 A AND B
The controls operate in the same manner
The A TIME/DIV switch also
point
decimal
or
1/TIME MODE.
45.
FINE/POSITION-Positions
ly
for
A sweep and B Sweep
mode.
46. X10
factor
nanosecond/division. The magnified sweep
division
of
the center graticule line}.
47.
VAR-Provides
between the calibrated settings
It extends the slowest A Sweep rate
seconds/division. The A Sweep rate is calibrated when the
control
be in the detent position
measurements. On instruments equipped
VAR control must
accurate measurements in the
tions.
48. UNCAL
uncalibrated (VAR
location when the DM44 is in the
MAG-Increases
of
10.
It extends fastest sweep rate
of
the unmagnified
is
set
fully
Lamp-Indicates
TIME/DIV
continuously
clockwise
to
make accurate differential
be
in the detent
control
AND DELAY
as
controls
or
displayed sweep rate
display
to
out
the
TIME
the display
on
the
X-axis in the X-Y
(0.5 division
variable sweep rates
of
the A TIME/DIV
to
at least 1.25
the calibrated detent.
with
position
TIME
and
1/TIME
the A Sweep rate
of
the calibrated detent}.
TIME-
37a above.
lamps and
TIME
horizontal-
by
to
is
the
center
either
side
switch.
Must
time
a DM44 the
to
make
func-
a
1
is
J
@
J . .
....
& .
!1
L .
466/464 DM44 Operators
..
!LA
l&JJL
Xi!S!.~M!Jl
......
L.J1!kJL.U3
J.L3i
LJ
,.<JJ
.a
SL
.JJL!
LLB
13
a·;
49. X10
is
on.
MAG
Lamp-Indicates
that the X10
magnifier
I
Fig.
7.
Horizontal and
indicators.
power
controls, connectors, and
50a. DELAY
sweep delay
by the A
same manner as 41a. Also, when the DM44 is in the
or
with the a TIME control. The DELAY TIME POSITION
control
measurement point. The a TIME
time-measurement point. With the time-measurement
point
indicates a negative time difference.
TIME/DIV
SOb. DM44 DELAY TIME
1/TIME
moves
to the left
You can
TIME
POSITION
point. The
located in the
DM44 Service manual.
TIME
to
more than 10 times the delay time indicated
function, this
both
modify
procedure
Operating
POSITION-Provides
switch.
POSITION-Operates
control
the reference
of
the re.terence
NOTE
your
instrument
control
move
for
making
Information
operates in
point
control
moves
point
to make the
only
the reference
this
modification
section
variable
in the
TIME
conjunction
and the
the Readout
time-
only
DELAY
is
of
the
the
I
I
I
I
I
I
14
466/464 DM44 Operators
@
I
J
l
J
51.
POWER-Turns
52.
HORIZ
tion
for
A: Horizontal deflection provided
sweep rate determined
switch. B sweep is inoperative.
MIX:
at a rate set
the sweep at a rate set
amounts
rates are determined
POSITION
DISPLAY-Determines
the horizontal deflection system.
The
first part
by
the A
of
the display allocated
control.
instrument
by
the setting
of
the horizontal sweep is displayed
TIME/DIV
by
the B
by
the setting
power
the
by
of
switch
and the latter part
TIME/DIV
to
each
of
the DELAY-TIME
on
and off.
mode
of
opera-
A Sweep at a
the A
TIME/DIV
switch. Relative
of
the
two
sweep
REAR PANEL
53. A
+GATE-Provides
pulse coincident with the A Sweep time, which can be
used to trigger the signal source (with TRIG MODE switch
set to AUTO).
54. B
+GATE-Provides
pulse coincident with the B Sweep time,
used to trigger the signal source after a selected delay
of
time, providing that A Sweep is triggered internally.
55.
CH
1 VERT SIGNAL
the signal applied to the CH 1
a positive-going rectangular
a positive-going rectangular
OUT-Provides
input
connector.
which
a sample
can
be
of
l
J
l
J
l
A INTEN: The sweep rate is determined by the A
TIME/DIV
display
the B
provides a check
Sweep (delayed sweep) with respect
(delaying sweep).
B DL YD: The sweep rate is determined by the B
TIME/DIV
setting
DELAY-TIME
REV FEB 1982
switch. An intensified
during
TIME/DIV
of
the
B Sweep time,
switch setting.
of
the
switch
the DELAY
with the delay time determined
TIME
POSITION control.
duration
(A
TIME/DIV)
portion
which
appears
is
about
This
switch
and position
to
the A Sweep
switch and the
10 times
position
of
466/464 DM44 Operators
on
the B
by
the
the
56. EXT
crt
display. Does not affect display waveshape. Signals
with fast rise and fall provide the most
change. Signals must be time-related
stable display. Useful
calibrated modes
57.
square-wave voltage
compensate voltage probes and check vertical gain,
current probes and oscilloscope operation.
intended to verify time-base calibration.
Z-AXIS-Permits
of
operation.
CALIBRATOR-A
output
intensity modulation
abrupt
to
the
for
adding time markers in
combination
that permits the operator
current
intensity
display
It
of
the
for
un-
loop/
is
not
15
a
to
58. Line Fuse
Holder-Contains
regulating range selector. See Table
tion .
the line fuse and the
1
for
change
informa-
Fig.
8.
Rear panel controls,
connectors,
1737·08
and indicators.
59. Regulating Range
regulating range. See Table 1
60.
Line
Cord-Makes
Selector-Shown
for
change information .
the
connection
in Medium
between the
oscilloscope and the power source. The cord may
conveniently stored by
wrapping
it around the feet on the
rear panel.
61. MOD
the instrument contains an
SLOTS-A
number
option
in either slot indicates
or
other modifications.
RIGHT SIDE PANEL
62. LINE VOLTAGE SELECTOR
either 115 V
or
230 V nominal line voltage.
SWITCH-Selects
LEFT SIDE PANEL
63. Vertical Gain Controls
side
panel)-Screwdriver
each vertical channel.
64. Variable Balance Controls
left
side
panel)-Screwdriver
of
each vertical channel.
(accessible
through
adjustments to set the gain
(accessible
through
adjustments to set balance
be
left
of
16
466/464 DM44 Operators
REV FEB 1982
I
J
J
l
J
J
l
l
Figure
BA.
Left
side panel controls.
(2906)2037-01
:I
REV FEB 1982
466/464 DM44 Operators
17
BASIC OSCILLOSCOPE
DISPLAYS
These
most
instructions
commonly
NORMAL SWEEP DISPLAY
1. Set the
VERT MODE
VOLTS
VOL TS/DIV VAR
AC-GND-DC
Vertical POSITION
20
MHz
INVERT
TIME
A
TIME
HORIZ
X10
POSITION
controls
/DIV
BW
/DIV Switches
/DIV VAR
DISPLAY A
MAG
permit
used basic
as
follows
Vertical
Horizontal
the
operator
displays
:
CH
1
Positi
amplitude
be applied
Calibrated
AC
Midrange
Not
not
Button
Locked
Calibrated
Off
(button
Midrange
to obtain the
.
on
determined by
detent
limited
(yellow band
visible)
out
together
detent
out)
of
signal to
at 1
ms
THE
ADJUSTABLE
TO
OFFER
SITIONS. THE
HANDLE
OPERATING
PROVIDES
CARRYING
CONVENIENT,
NEXT-TO-LAST
THE
MOST
POSITION.
Fig. 9. Tilted operation.
HANDLE
STABLE
LOCKING
STABLE, NEAR-UPRIGHT,
IS
DESIGNED
OPERATING
POSITION
OF
1737·15
PO-
THE
18
466/464 DM44 Operators
@
]
l
Storage
REDUCED SCAN (PULL)
STORAGE
SAVEINTEN
SAVE
NON
VIEW
INTENSITY
LEVEL
STORE
TIME
Display
Trigger
(Both A and B if applicable)
Pushed in (466 only)
NORM
ccw
Button
Button
NORM
Fully
out
in
counterclockwise
pushbutton and adjust the CH 1 VOL
reduce the vertical display size.
display with the vertical and
trols; release the BEAM FINDER pushbutton.
4.
Set the CH 1 VOL
POSITION
area.
5.
display.
6.
POSITION control to locate the
area. Then adjust the FOCUS
control
Adjust the A
Set the A
to locate the
Trigger
TIME/DIV
horizontal
TS/DIV
display
LEVEL
switch and the horizontal
display
control
TS/DIV
Center
the compressed
POSITION
switch and vertical
within
control
within
as
needed.
switch
con-
the
display
for
a stable
the display
to
J
J
J
J
SLOPE
LEVEL
SOURCE
COUPLING
TRIG
MODE
A
TRIG
2.
Pull the POWER
the CH 1
display
INTENSITY
@
3.
Adjust
brightness. If
input
(A only)
HOLDOFF
connector.
the
INTENSITY
control
at midrange, press the BEAM FINDER
switch
the
display is
+
0
NORM
AC
AUTO
NORM
(on).
control
Connect
for
not
visible with the
1.
2.
area to
(0.5 division on each side
the signal
the desired
466/464 DM44 Operators
to
line).
setting.
3.
horizontal POSITION
the
magnified display. Divide the
to
determine the magnified sweep rate.
MAGNIFIED SWEEP DISPLAYS
Obtain a Normal Sweep Display.
Adjust the horizontal POSITION
be
magnified to
It may
be
Push the
necessary
X10
within
the center graticule division
of
the
to
change the
MAG
switch (on) and adjust the
control
for
control
center
vertical graticule
TIME/DIV
precise
TIME/DIV
to move the
positioning
setting
switch
of
by
10
19
DELAYED SWEEP DISPLAY
1.
Obtain a Normal Sweep Display.
2.
Set the
B
Trigger
3.
Pull
the intensified zone
INTENSITY
zone
distinguishable
instrument
switch
delayed sweep. Dual delayed
6.
step
HORIZ
SOURCE switch
out
on the B
control
is equipped
to a function
DISPLAY switch
to
TIME/DIV
is
the
as
needed
from the rest
with
a DM44, set the FUNCTION
other
than
to
A INTEN and the
STARTS AFTER DELAY.
knob
and turn cw until
desired length. Adjust the
to
make the intensified
of
TIME
or
displays
the display. If
1/TIME
are discussed in
for
your
a single
displays set
MODE
control
control
Trigger
Trigger
for
switch
positions
positions
7.
For
a delayed
SOURCE
SOURCE switch and
a stable display.
Differential
using
the TIME
invalid
when the B
to
STARTS
the
FUNCTION
to ALT. The DELAY
both delayed displays.
only
switch
time
measurements
or 1 /TIME
AFTER DELAY.
switch
the CH 2 delayed display.
display
to
NOTE
functions
Trigger
SOURCE
to
with
less jitter, set
the same
adjust
the B LEVEL
and
TIME
and the VERT
TIME
POSITION
The
ALT
position
measurements
of
the DM44 are
switch
is
as
not
DELAY
the
B
the A
control
set
I
I
I
4.
Adjust
the DELAY
the intensified zone to cover the
be displayed in delayed form.
5.
Set the
intensified zone noted in steps 3 and 4 is
delayed form. The delayed sweep rate is indicated by the
dot
on the B
6.
If
displays
The DM44 will indicate the time
beginning
20
HORIZ
TIME/DIV
your
instrument is
of
two
signals can be obtained at the same time.
of
the delayed displays.
TIME
DISPLAY
knob.
equipped
POSITION
portion
switch
to
with
difference
To
obtain
control
of
now
a DM44 delayed
to
move
the display to
B DLY'D. The
displayed in
between the
two
delayed
466/464 DM44 Operators
MIXED SWEEP DISPLAYS
1.
Obtain
2.
Pullout
desired sweep rate. If
DM44 set the
TIME
displays are discussed in step
3.
Set the
now
contains
display
a normal sweep display.
on the B
your
FUNCTION
or
1/TIME
is
at
for a single
HORIZ
two
sweep rates.
the
A sweep rate
DISPLAY
TIME/DIV
instrument
switch to a
while
knob
and
turn
cw
is
equipped
function
mixed display. Dual
4.
switch
to
The
first
the latter
other
MIX. The
portion
portion
to the
with
than
mixed
display
of
the
of
the
@
I
a
I
I
l
l
J
J
display
portion
POSITION
sweep
time.
the
To
switch
DELAY
displays.
mixed
Normal
on.
MODE
CH
the
visible. If
is at the B sweep rate. The start
can be changed
4.
If
your
displays
The
beginning
obtain
to
TIME
display.
1.
Preset the
Sweep Display, then turn the
Allow
2.
Set the
switch
2 OR Y
CH
1 OR X
3.
Advance
by
control.
instrument
of
DM44
will
of
the
two
mixed sweep displays set
TIME
and the VERT
POSITION
The
fl.
TIME
is equipped with a DM44, mixed
two
signals can be obtained at the same
indicate the time difference between
B sweep
control
control
X-Y DISPLAY
instrument
several minutes
TIME/DIV
to
CH
input
connector
input
the
INTENSITY
the
display
controls
for
switches
2.
Apply
connector.
is
not visible
of
the B sweep rate
adjusting the DELAY
portions
MODE
instrument
and the horizontal signal to
of
the
two
the
switch
positions
positions
to
the vertical signal to the
control
only
as
given in step 1
instrument
warm-up.
X-Y and
until
the
with
the INTENSITY
TIME
displays.
FUNCTION
to
ALT.
The
both
mixed
the CH 2
of
power
the
VERT
display
is
control
and adjust the CH 1 and CH 2 VOL
the display
tally. Center the compressed
controls (CH 2 POSITION vertically, Horizontal ·
POSITION horizontally); release the BEAM FINDER
pushbutton. Adjust the FOCUS
display.
at midrange, press the BEAM FINDER pushbutton
TS/DIV
is
reduced in size, both vertically and
display
control
switches until
with
the POSITION
for
a well-defined
horizon-
SINGLE SWEEP DISPLAY
1.
Obtain a Normal Sweep Display. For random
signals, set the
approximately the same
random signal.
2.
Push the SINGL SWP
switch. The next trigger pulse starts the sweep and
displays a single trace. If
READY lamp lights, indicating the A Sweep Generator
circuit
is
3.
After the sweep
out"
and the READY lamp is out. Press the
button
display.
to prepare the
trigger
set and waiting
circuit
is
to
trigger
amplitude
button
no
triggers are present, the
to
be triggered.
complete, the
circuit
for
on a signal that
and
frequency
on the A TRIG
circuit
is
SINGL
another
single-sweep
is
as
the
MODE
"locked
SWP
@
466/464 DM44 Operators
21
STORAGE DISPLAYS
STORAGE
1.
Obtain a normal display
(see Normal Sweep Display
of
Displays section
2.
Set the STORAGE LEVEL and VIEW TIME controls
NORM. Set the SAVE switch button out. Select either
PERS
or
FAST mode.
this manual).
of
the signal to
in
the Basic Oscilloscope
be
measured
to
VAR
4.
Set the INTENSITY control
brightness. Set
display (depends on the amplitude and speed of the signal
to be viewed). Set the VIEW TIME
display retention (persistence) in the VAR PERS mode
for
the time between erase cycles in the FAST mode.
5.
Push the ERASE button. The display is erased,
except in the SAVE mode, and the storage and
circuits are reset.
the
STORAGE LEVEL
for
adequate
control
control
for
optimum
for
the desired
display
or
trigger
SAVE STORAGE
1.
Obtain a Storage Display using steps 1
through
4.
I
I
a.
In general,
VAR
PERS
100 µs/division.
b.
When using the
B DL Y'D
control
rate.
3.
Push the ERASE button.
storage and
display.
22
better
mode
horizontal
to
BENDS A for
trigger
NOTE
results are
for sweep speeds slower than
combination
display, set A TRIG HOLDOFF
maximum
circuits
obtained
of
FAST storage
display repetition
Allow
a few seconds for the
to
automatically provide a
using the
466/464 DM44 Operators
and
2.
Immediately after the event is ·stored, push in the
SAVE button, providing a longer viewing time. The SAVE
INTENS
the display brightness decreases the viewing time.
be
TIME setting), this reduces maximum viewing time.
control
control
control
Pushing the SAVE button, after the event is stored, can
delayed
Maximum viewing time occurs when the SAVE INTENS
is
only
3.
Erase
adjusts the display brightness. Increasing
up
to
several seconds (depending on the VIEW
set
fully
counterclockwise, advancing the
long enough
is
disabled in the SAVE mode.
to
view the display.
@
I
]
]
J
J
l
l
SAVE RANDOM EVENTS
1.
Obtain a Storage Display using steps 1 through
with a test signal similar
not
mode,
2.
3.
4.
5.
random event occurs, push the SAVE button.
VAR PERS.
Set
the
SAVE INTEN control
Replace the test signal with the random event.
Push the ERASE button.
When the READY lamp comes on,
to
a random event. Use the FAST
fully
ccw.
but
4,
before the
6.
After the random event
storage
mode.
circuitry
will automatically
occurs
and is stored, the
switch
to
the SAVE
REDUCED SCAN STORAGE (466 ONLY)
To
obtain a greatly increased
the INTEN control to operate the 466 in the reduced scan
mode. In the reduced scan mode the instructions
obtaining storage displays are unchanged. The
difference in operation
referenced to the smaller graticule in the center
(see item
section
within this smaller graticule when operating in the reduced scan mode.
11
b in the Controls, Connectors, and Indicators
of
this manual). All measurements should be made
is
writing
that the deflection factors are
rate, pull
of
out
the
on
for
only
crt
J
l
@
466/464 DM44 Operators
23
Except
TIME
under
for
and
1/TIME
Delayed
DM44 DISPLAYS AND MEASUREMENTS
the
TIME
and
1/TIME
functions
or
Mixed Sweep Measurements.
RESISTANCE
functions,
are discussed in the Oscilloscope
the DM44 is
independently
usable
Applications
anytime
and Measurements section
the
oscilloscope
TABLE
Resistance Ranges
is
turned
of
this
on.
The
manual
E
2
The
meter
measure voltage
mode
of
operation
ed
in)
and
rms.
1.
Push in the
20
MO RANGE
2.
Connect
resistance.
3.
Reduce
proper
readout is obtained.
24
may
if
the
applied
OHMS
button.
the + and
the
range,
be
damaged
the
meter
(OHMS
voltage
FUNCTION
See Fig. 10.
COM
using
by
attempting
is in the resistance
FUNCTION
is
the
button
in excess
button and the
leads
to
following
to
push-
of
120 V
A
the
unknown
table, until a
466/464 DM44 Operators
any
tion
A
is normal.
RANGE
20
MO
2 MO
200 kO
20
kO
2 kO
200 0
blinking
unknown
and a need
blinking
.
READOUT
20.00-02.00
2.000-0.200
200.0-20.00
20.00-02.00
2.000-0.200
200.0-000.0
readout on
resistance, indicates an overrange
to
readout, when meter leads are
select a
any
higher
MEASUREMENT
20
M0-2
2
M0-200
200
k0-20
20
k0-2
2
k0-200
200
0-0
range, when connected
range.
disconnected
MO
kO
kO
kO
0
0
condi-
to
@
I
I
:I
:J
l
1]
J
l
l
J
@
RANGE
Fig. 10. Resistance.
466/464 DM44 Operators
SELECTION
(465/DM-0-2)2039-9
25
I
VOLTS
The
maximum
peak
ac)
+
input
the
maximum
The
peak
ac) to chassis.
meter
The
measure voltage
mode
of
excess
If
(indicating
to
of
the reading exceeds 1200 V
prevent
safe
between
and
COM
may
operation
120 V rms.
overrange)
possible
input
the+
chassis.
floating
be
damaged
if
the
and
disconnect
meter
and
voltage
COM
voltage
meter
the
damage.
is
±1200 V (de+
inputs
or
between
is
±500 V (de+
by
attempting
is
in the resistance
applied
or
voltage
the readout blinks
the+
lead at once
3.
Reduce the range, using Table
readout is obtained.
NOTE
When
no
voltage
ranges, the
elements
meter
2 V ranges.
to
is
in
A
overrange
range.
may
leads
blinking
condition
readout
blink. Also,
may
increase the
readout
is
applied
is
0000
and
noise
readout
on
any
and a need
TABLE
Volts Ranges
3
3,
until a proper
in the 20 V to 1.2
individual
picked
range indicates an
to select a
readout
up
by
in the .2 V
higher
kV
the
and
1.
Push in the VOL TS
1.2 kV RANGE button. See Fig.
2.
Connect the
{usually a
unknown
ground
voltage
COMMON
or
to
be measured and observe the reading.
FUNCTION
test point) and the HIGH lead to the
26
11.
lead
button and the
to
the reference point
466/464 DM44 Operators
RANGE
1.2 kV
200 v
20 v
2V
200 mV
READOUT
1.200-0.200
200.0-020.0
20.000-02.00
2.000-0.200
0.200-0.000
MEASUREMENT
1.2
kV-200
200
20
2
.2
V-20
V-2
v-.2
V-0
V
V
V
v
V
@
I
J
J
J
]
1]
]
l
J
J
@
466/464
Fig. 11. Volts.
DM44
Operators
(465/DM-0-3)2039-10
27
TEMPERATURE
the device
removed
These items
probe
from
manual,
to
be measured, a
the
device
and
other
which
should
by
probe
thermal
the
information
be reviewed.
gradient,
probe,
etc.
heat
I
are in the
The
maximum
surface
ground.
The
sensor
subjected
sensor
If
the
the
probe
tip
reading
safe voltage
is
±400
V (de +
tip
is fragile
to excessive stress.
should
exceeds
at
once
not
to
prevent
and
exceed
-55°
on
peak
may
20
C
probe
the
ac)
above chassis
break
Force
exerted
pounds.
or
+150°
damage.
measurement
if
dropped
C,
or
on
the
remove
The DM44 is calibrated to its
accurate readings
The DM44
probe.
In
the
following
to
verify water temperature.
the
melting
by
changes in
Accuracy Check
within
2°C
for
should
temperature and the
be recalibrated
checks, use an accurate
Anything
altitude
and
barometric
original
examples that
boiling
probe,
to
any replacement
thermometer
in
solution
point
pressure.
giving
follow.
affects
is affected
Low Temperature
1.
Push in the TEMP
Fig. 12.
2.
Apply
the temperature
measured so the
the device
Table 4 to convert the reading
fahrenheit.
Temperature
such
as the
flat
to
be measured and observe the reading. See
accuracy
angle
28
l __________________________________________________________________________
I
surface
NOTE
is
of
contact
(°
C)
FUNCTION
probe
of
the
affected
between
to
the device to
probe
tip mates against
from
by
several factors
the
button. See
to
be
centigrate to
bottom
probe
and
466/464 DM44 Operators
indicating
Allow a container
melt
until
there are
To
prevent
sealed
Put the
possible
portion
of
of
probe
tip
the container. Wait
the probe has reached the water temperature.
(preferably insulated)
only
a few pieces
of
crushed ice
of
ice remaining.
~
probe
damage,
the
probe
is
into the water, avoiding the side
for
be
immersed
the readout
sure
only
(see Fig. 12).
to
stabilize,
the
or
@
I
I
__
J
J
J
l
J
l
]
J
APPLY
SILi
TO
THE
MEASURED
su~~~E
(W
GREASE
CE
TO
HEN
BE
POSSIBLE).
J
SLM
@
~
JXJ
F"
ig. 12. Tempe rature.
466/464 DM44
Operators
29
CENTIGRADE
FAHRENHEIT
-
50
I I
11111111111
l1111l'l'
-
50
CENTIGRADE
-
40
liiljilill"'ll'"
'I''''''''
-40
-30
TABLE
TO
FAHRENHEIT
-
30
'II'
-20
4
liilll'l'liljlii
'II''''''
-10
CONVERSION
-
20
111111111111 111111111 I
I'
0
-10
'I' '11j11111
10
20
11
0
"I'
11I
30
I
I
CENTIGRADE
FAHRENHEIT
CENTIGRADE
FAHRENHEIT
CENTIGRADE
FAHRENHEIT
1738-35
30
0
1
,,,,
,
1
1
""
1
40
50
1·1111· '''I'''''''''
130
100
I··
1·11
11r111
'I
220
10
••
,,,,,
11111111111
50
•
,,,
1111
60
1'"'1" 11
60
l""I
140
150
110
•1·1··1•11•111111111· 1·I·1111111111111111111
11
11
t
11
n
1111
n I
Iii
230
240
466/464 DM44 Operators @
20
,,
,1,,
,
11111
1
70
70
,,,.
1111H111111111q 111111111111111111111r11111111
,,,,
1
160
120
11111111
1 1 T 1 1 n
250
30
1,
,·11111111111!•
11
1"'
80
I
"'1
90
1111
1111
100
80
.,,
1.,1''
170 180
190
130
11
11
11
1 1
260
270
280 290
40
111
1
1
11
90
''I''
200
140
111
1111 1111Ir11
1·11
"1"
110
''II''
1
11
"'I'
120
210
11
n I '
300
50
,
,1
100
'I 1111111111q1111111r111
150
I:
E
E
I:
I:
I
I
J
J
]
The
readout should be
ice remaining after the test to verify that inserting the
probe
did
not
raise the water temperature.
High Temperature
Bring
water
to
a slow boil (to prevent splattering). Put
the
probe
tip
into the water, avoiding the side
the
container. Wait
the
probe
has reached the water temperature.
for
-2°
C to 2° C. There should
the readout
or
bottom
to
stabilize, indicating
be
The readout should be between
clean water at sea level.
TIME AND
of
See
DM44
MEASUREMENTS in this manual.
DELAYED
98°C
1/TIME
OR
MIXED
and 102°C
SWEEP
for
J
J
@
466/464 DM44 Operators
31
APPLICATIONS AND MEASUREMENTS
Refer
to
Basic Oscilloscope Displays
as
needed
to
obtain
correct
displays.
I
PRELIMINARY
Signal Ground
~
Avoid
cross-grounding
may occur.
The most reliable measurements are made when the
oscilloscope
under
ground
on the
with
the
ground
test by the
lead can be connected to the
oscilloscope
unit
ground
under
Input Coupling Capacitor Precharging
In the
GND
position,
ground
precharging network.
Coupling
level
voltage transients
applied
through
capacitor
of
the signal applied
to
the
a one-megohm resistor
amplifier
because instrument damage
is
connected
strap on the signal probe. Also a
chassis
test.
the
This
to
charge to the average de voltage
accidentally
input. The precharge network also
to
the
ground
ground
to
establish a common
input
signal
network
to
the probe. Thus, any large
allows the Input
generated will not be
of
banana jack
is
connected to
to
the
ground
form a
unit
of
provides a measure
by
reducing the
external
probe
de level than that previously applied, especially
level difference is more than 10 times the VOL
setting.
set the
Wait several seconds
discharge.
capacitor
remain on the screen so the ac
can be measured in the
circuitry
The
following
tip
is connected
1.
Before
connecting
Input
Coupling
2.
Touch
the probe
3.
Connect
4.
Wait several seconds
5.
Set the
the probe
to charge.
Input
protection
current
levels that can be
during
procedure
to
switch to GND.
tip
for
Coupling
normal
capacitor
a signal source having a
the probe
to
the
tip
to the external
drawn
charging.
should be used whenever
tip
to
a signal source,
oscilloscope
Input
to the signal source.
for
switch
component
manner.
chassis
Coupling
the
Input
to
AC.
The
capacitor
Coupling
display
of
the
circuitry
from the
the
different
if
the de
TS/DIV
ground.
to
will
signal
I
I
I
~
I
I
32
466/464 DM44 Operators
@
I
J
J
J
J
J
OPERATOR'S
ADJUSTMENTS
AND CHECKS
To
verify
measurement accuracy, perform the
ing
checks
ment. See the
for
calibration
Trace Rotation Adjustment
Normally
using
switch
vertical deflection. Adjust
driver
the trace
Obtain a Normal Sweep Display presentation
calibrator square-wave voltage. Set· the appropriate
VOL TS/DIV switch to the 50 mV
Coupling
control
deflection
of
compensating the probe being used.
switch to DC. Make sure
is
in the calibrated detent.
is
5.8
to
6.2 divisions.
compensation.
See
Fig.
13.
Refer to probe manual
position
the
Check
and the
VAR VOL
that the vertical
of
the
Input
TS/DIV
33
Basic 466/464
Obtain a Normal
5 ms position. Set the A
Push the TRIG VIEW
sample
control
the
position
14). Verify the
1
of
to
display
the left peak
~-------
._,.-+-~
O<?
...•
Timing
Display. Set
switch
the line voltage. Use
vertically
area. Use
position
horizontal
I I
6.53
TO
\
\
10------~--_,_
I
~..
. . . . . .
..........
..__..__'"""--l,__..
...
Fig. 14. Basic 466/464
Check
Trigger
and
the
the
horizontal
to
the
distance
9.8
TO 10.2
6.79
~--+-._..
..
...
,
...
..
...
...
..
. .......... .
...
..
(60 Hertz Line Only)
the A TIME/DIV
SOURCE
hold
top
left
graticule
_____
timing
switch
it in. This displays a
the A Trigger
of
the
display to
position
edge (see Fig.
between the first and
_____
---+-+-'--+-~+--+-1
(
1738-14)2037-14
check.
switch
to
LINE.
LEVEL
within
control
...,.
to
to
the
fourth
peaks is 9.8
not
visible, verify the
is
first
and
the
third
peaks is 6.53
DM44
accuracy
Check
Timing
Perform the Basic 466/464
Leave
except
TIME/DIV
B
HORIZ
B SOURCE
FUNCTION
6.
TIME
Push the TRIG VIEW
DELAY
point
Point
measurement
on
is 16.49
to
A). Use
the
TIME
next
to
Check (60 Hertz Line
of
the
horizontal
controls
DISPLAY
the
16.84 and
set
as
follows:
POSITION
center
the
ALT
point
cycle
(see Fig. 15,
horizontal
to
the
to
10.2 divisions. If
horizontal
to
Timing
deflection
as
for
the
Basic 466/464
5µs
AINT
STARTS
TIME
To
move
measurement
right
point.
button
DELAY
the
ms lamp is I it.
control
graticule
control
center
Point
and
to
horizontal
the
fourth
distance
6.79 divisions.
between
Only)
Check
system.
of
move the reference
B). Verify
to
verify
Timing
AFTER DELAY
the
time-
point
to
the
reference
hold
it in. Use
line (see Fig. 15,
to
move the
graticule
the
readout
peak
the
the
the
the
time-
line
I
I
I
I
I
I
I
34
466/464 DM44
Operators
@
I
J
J
]
l
l
J
J
r
100····
External Horizontal Gain Check
the Basic Oscilloscope Displays procedure
Normal
voltage waveform; then, set the
With the
...
~-··
9a-\-+-
(If
X-Y
A__)
't..
operation
Sweep Display
calibrator
....
····~
\
\ :: I\:
....
Fig. 15. DM44
is
to
be used.) Use
of
signal connected to the CH 1
....
~
.._~,......_-+-~---11
•
1
..
..
..
,_·
...
..
..
timing
check.
~teps
the
calibrator
TIME/DIV
(1738-15)
2037-15
1through3
for
obtaining a
square-wave
switch
to
OR
of
X-Y.
input
connector and the
50
mV,
the crt display
horizontally by 5.75 to 6.25 divisions.
CH
1 VOL
should
TS/DIV
be
two
dots separated
PEAK-TO-PEAK VOLTAGE
MEASUREMENTS-AC
Obtain a Normal Display. Make sure the VAR
VOL TS/DIV
position the display so the
horizontal graticule line (see Fig. 16,
position the display so one
with
the center vertical
Measure the vertical deflection from peak-to-peak (Point
A