THIS MANUAL IS AN AUTHENTICATION OF THE MANUFACTURER’S COMMERCIAL LITERATURE WHICH,
THROUGH USAGE, HAS BEEN FOUND TO COVER THE
DATA REQUIRED TO OPERATE AND MAINTAIN THIS
EQUIPMENT. SINCE THE MANIJAL WAS NOT PREPARED
IN ACCORDANCE WITH MILITARY SPECIFICATION, THE
FORMAT HAS NOT BEEN STRUCTURED TO CONSIDER
LEVEL OF MAINTENANCE NOR TO INCLUDE A FORMAL SECTION ON DEPOT MAINTENANCE STANDARDS.
WARNING
DANGEROUS VOLTAGES
EXIST IN THIS EQUIPMENT
DON’T TAKE CHANCES!
CAUTION
Special 3% silver solder is required on the ceramic terminal
strips in this equipment. A 40- to 75-watt soldering iron
should be used and it should be tinned with the same special
solder. Additional quantities of the solder may be procured
under FSN 3439-912-8698. Ordinary solder may be used
only in dire emergency.
This Manual Contains Copyrighted Material Reproduced
Tektronix, Inc. All Rights Reserved
Permission Of
By
TM 11-6625-1722-15
T
ECHNICAL MANUAL
HEADQUARTERS
DEPARTMENT OF THE ARMY
NO. 11–6625–1722–15
ASHINGTON
, D.C., 10 January 1972
W
Operator’s Organizational, Direct Support, General Support, and Depot
Maintenance Manual Including Repair Parts and Special Tools Lists
If your Oscilloscope AN/USM-273 needs improvement, let us know.
You,
the user,
are the only one who can tell us what you
Send
Let us know why you don’t like the
Tell us why a procedure is hard to perform.
Put it on an SF 368
Mail it to Commander, US Army Communications
DRSEL-ME-MQ, Fort
We’ll send you a reply.
Administrative Storage
Administrative storage of equipment issued to and used by Army activi-
ties shall be in accordance with TM 740-90–1.
0-6
Destruction of Army Electronics Materiel
Destruction of Army electronics materiel to prevent enemy use shall be
in accordance with TM 750-244-2.
0-7.
Reporting Errors and Recommending Improvements
You can help improve this manual.If you find any mistakes or if you
know of a way to improve the procedures, please let us know.Mail your
letter or DA Form 2028 (Recommended Changes to Publications and Blank
Forms) to
Readiness Command, ATTN:
either case,
Commander, US Army Communications and Electronics Materiel
DRSEL-ME-MQ, Fort Monmouth, NJ 07703. In
a reply will be furnished direct to you.
0-2
Change 1
SECTION 1
CHARACTERISTICS
TM 11-6625-1722-15
Introduction
The Tektronix Type 453 Oscilloscope is a transistorized
portable oscilloscope designed to operate in a wide range
of environmental conditions. The light weight of the Type
453 allows it to be easily transported, while providing the
performance necessary for accurate high-frequency measurements.
provides calibrated deflection factors from 5 millivolts to
10 volts/division. Channels 1 and 2 can be cascaded using
an external cable to provide a one millivolt minimum defection factor (both VOLTS/DIV switches set to 5 mV).
The trigger circuits provide stable triggering over the full
range of vertical frequency response. Separate trigger controls are provided to select the desired triggering for the A
and B sweeps. One of three sweep modes can be selected
for the A sweep; automatic, normal or single sweep. The
horizontal sweep provides a maximum sweep rate of 0.1
microsecond/division (10 nanosecond/division using 10X
magnifier) along with a delayed sweep feature for accurate
relative-time measurements.
can be made with Channel 2 providing the vertical deflection,
and Channel 1 providing the horizontal [deflection. (TRIGGER
switch set to CH 1 ONLY, HORIZ DISPLAY switch set to EXT
HORIZ). The regulated DC power supplies maintain con-
The dual-channel DC-to-50 MHz vertical system
Accurate X-Y measurements
ELECTRICAL CHARACTERISTICS
VERTICAL DEFLECTlON SYSTEM
stant output over a wide variation of line voltages and frequencies.
approximately 90 watts.
Information given in this instruction monual applies to the
Type R453 also unless otherwise noted. The Type R453 is
electrically identical to the Type 453 but is mechanically
adapted for mounting in a standard 19-inch rack. Rackmounting instructions, a mechanical parts list and a dimensional drawing for the Type R453 are provided in Section
10 of this manual.
The electrical characteristics which follow are divided
into two categories. Characteristics listed in the Performance
Requirement column are checked in the Performance Check
and Calibration sections of this manual. Items listed in the
Operational Information column are provided for reference
use and do not directly reflect the measurement capabilities of this instrument. The Performance Check procedure
given in Section 5 of this manual provides a convenient
method of checking the items listed in the Performance
Requirement column. The following electrical characteristics
apply over a calibration interval of 1000 hours at an ambient temperature range of -15°C to +55°C, except as
otherwise indicated. Warm-up time for given accuracy is
20 minutes.
Total power consumption of the instrument is
. .
Characteristic
Deflection Factor
Deflection Accuracy
Variable Deflection Factor
Bandwidth at Upper -3 dB
point (with or without P6010
Probe)
20 mV to 10 VOLTS/DIV
10 mV/DIV
5 mV/DIV
Channels land 2 cascaded
Risetime (calculated). With or
without P6010 Probe.
20 mV to 10 VOLTS/DIV
10 mV/DIV
5 mV/DIV
Channels 1 and 2 cascaded
Performance Requirement
5 millivolts/division to 10 volts/division in 11
calibrated steps for each channel. One millivolt/
division when Channel 1 and 2 are cascaded.
Within ±3Y% of indicated deflection with VARlABLE control set to CAL. Cascaded deflection
factor uncalibrated.
Uncalibrated deflection factor at least 2.5 times
the VOLTS/DIV switch indication. This provides
a maximum uncalibrated deflection factor of 25
volts/division in the 10 volts position.
DC to 50 MHz or greater
DC to 45 MHz or greater
DC to 40 MHz or greater
DC to 25 MHz or greater
Less than 7 nanoseconds
Less than 7.8 nanoseconds
Less than 8.75 nanoseconds
Less than 14 nanoseconds
Operational Information
Steps in 1-2-5 sequence
With gain correct at 20 mV
Driven from 25-ohm source
Measured at one millivolt/division
Risetime calculated from bandwidth
measurement using the formula:
Where:
= Risetime in nanoseconds.
t
r
BW = Bandwidth in megahertz.
1-1
TM 11-6625-1722-15
Characteristic
Input RC Characteristics
Maximum lnput Voltage
lnput Coupling Modes
AC Low-Frequency Response
(lower -3 dB point)
Without probe
With P6010 Probe
Trace Shift Due to Input Gate
Current (at 25°C)
Vertical Display Modes
Chopped Repetition Rate
Attenuator Isolation
Common Mode Rejection Ratio
Linear Dynamic Range Useful
for Common-Mode Relection
in ADD Mode
Polarity Inversion
Signal Delay Line
Low-Frequency
Vertical Linearity
Trace Drift (after 20 minute
warm up)
20 mV to 10 VOLTS/DIV
10 mV/DIV
5 mV/DIV
VERTICAL (cont)
Performance Requirement
AC or DC, selected by front-panel switch
Negligible
Channel 1 only
Channel 2 only
Dual-troce, alternate between channels
Dual-trace, chopped between channels
Added algebraically
Approximately one-microsecond segments from
each channel dispiayed at repetition rate of 500
kHz, ±20%.
Greater than 10,000:1, DC to 20 MHz
Greater than 20:1 at 20 MHz for common-mode
signals less than eight times VOLTS/DIV switch
setting.
Signal on Channel 2 can be inverted
Less than 0.15 division compression or expansion
of two division signal when positioned to vertical
extremes of display area
Operational Information
Typically 1 megohm (±2%), paralleled
by 20 pF (±3%)
600 volts DC + peak AC (one kilohertz
or less). Peak-to-peak AC not to exceed
600 volts.
Typicaily 1.6 Hz, Input Coupling switch
set to AC
Typically 0.16 Hz
With optimum GAIN
frequency
Less than 10% incremental signal dis-
tortion for instantaneous input voltage
-10 or +10 times VOLTS/DIV
of
switch setting
Approximately 140 nanoseconds
Includes CRT linearity. Measured with
one-kilohertz square wave.
Time
Typically less than Typically less than
0.03 division/hour
Typically less than
0.05 division/hour
Typically less than
0.08 division/hour
adjustment at low
Temperature
0.0075 division/degree C
Typically less than
0.0125 division/degree C
Typically less than
0.02 division/de-
gree C
Source
Coupling
Polarity
1-2
TRIGGERING (A AND B SWEEP)
Internal from displayed channel or from Channel
1 only
Internal from AC power source
External
External divide by 10
AC
AC low-frequency reject
AC high-frequency reject
DC
Sweep can be triggered from positive-going or
negative-going portion of trigger signal
TRIGGERING (cont)
TM 11-6625-1722-15
Characteristic
Internal Trigger Sensitivity
AC
LF REJ
HF REJ
DC
External Trigger Sensitivity
AC
LF REJ
HF REJ
DC
Auto Triggering (A sweep only)
Single Sweep (A sweep only)
Display Jitter
Maximum Input Voltage
External Trigger Input RC
Characteristics (approximate)
LEVEL Control Range
Performance Requirement
0.2 division of deflection, minimum, 30 Hz to 10
MHz; increasing to 1 division at 50 MHz
0.2 division of deflection, minimum, 30 Hz to 10
0.2 division of deflection, minimum, 30 kHz to 10
0.2 division of deflection, minimum, 30 Hz to 50
kHz
0.2 division of deflection, minimum, DC to 10
MHz; increasing to 1 division at 50 MHz
50 millivolts, minimum, 30 Hz to 10 MHz; increasing to 200 millivolts at 50 MHz
50 millivolts, minimum, 30 kHz to 10 MHz; increasing to 200 millivolts at 50 MHz
50 millivolts, minimum, 30 Hz to 50 kHz
50 millivolts, minimum, DC to 10 MHz; increas-
ing to 200 millivolts at 50 MHz
Stable display presented with signal amplitudes
given under Internal and External Trigger Sensitivity above 20 Hz. Presents a free-running sweep
for lower frequencies or in absence of trigger
signal.
A Sweep Generator produces only one sweep
when triggered. Further sweeps are locked out
until RESET button is pressed. Trigger sensitivity
same as given above.
Less than 1 nanosecond at 10 nanoseconds/division sweep rate (MAG switch set to X10)
At least ±2 volts, SOURCE switch in EXT position. At least ±20 volts, SOURCE switch in EXT
÷10 position
Operational Information
Typical -3 dB point, 16 Hz
Typical -3 dB point, 16 kHz
Typical -3 dB points, 16 Hz and 100
kHz
Typical -3 dB point, 16 Hz
Typical -3 dB point, 16 kHz
Typical -3 dB points, 16 Hz and 100
kHz
600 volts DC + peak AC (one kilohertz
or less). Peak-to-peak AC not to exceed
600 volts.
1 Megohm paralleled by 20 pF, except
in LF REJ
Sweep Rates
A Sweep
B sweep
Sweep Accuracy-A and B
Sweep
5 s to 0.1 s/DIV
50 ms to 0.1 µs/DIV
Variable Sweep Rate
HORIZONTAL DEFLECTION SYSTEM
A and B Sweep Generator
0.1 microsecond/division to 5 seconds/division in
24 calibrated stem
0.1 microsecond/division to 0.5 second/division
in 21 calibrated steps
0°C to +40°C
Within ±3% of indi- Within ±5% of indi-
cated sweep rate
Within ±3% of indi- Within ±4% of indi-
cated sweep rate
Uncalibrated sweep rate to at least 2.5 times the
TIME/DIV indication, or a maximum of at least
12.5 seconds/division in the 5 s position (B sweep,
maximum of 1.25 seconds/division in the .5 s
position.
-15°C to +55°C
cated sweep rate
cated sweep rate
A sweep is main and delaying sweep
B sweep is delayed sweep
A VARIABLE and B TIME/DiV VARlABLE controls set to CAL
modulation
Usable Frequency Range DC to greater than 50 MHz
Input Resistance at DC
Input CouplingDC coupled
Polarity of Operation
1 megohm, paralleled by 20 pF
Less than 3°
Selected by CALIBRATOR switch on side
panel
±1.5%
Approximately 200 ohms in 1 V position.
Approximately 20 ohms in .1 V position.
Approximately 47 kilohms
Positive-going input signal decreases
trace intensity
Negative-going signal increases trace
Maximum Input Voltage
A and B Gate
Waveshape
Amplitude
Polarity
Duration
Output resistance
Vertical Signal Out (CH 1 only)
Output voltage
Bandwidth
Output coupling
Output resistance
OUTPUT SIGNALS
—
Rectangular pulse
Posltlve-going with baseline at about -0.7 volts.
I Same duration as the respective sweep
I
25 millivolts, or greater/division of CRT display
into 1 megohm load.
DC to 25 MHz or greater when cascaded with
Channel 2 or into 50-ohm load.
DC coupled
200 volts combined DC and peak AC
A GATE duration variable between
about 4 and 11 times the A TIME/DIV
switch setting
with the A SWEEP
LENGTH control.
Approximately 1.5 kilohms
Approximately 50 ohms
1-5
TM 11-6625-1722-15
Characteristic
Line Voltage
Voltage Ranges (AC, RMS)
115-volts nominal
230-volts nominal
Line Frequency
Maximum Power Consumption
at 115 Volts, 60 Hz
Tube Type
Phosphor
Accelerating Potential
Graticule
Type
Area
Illumination
Unblinking
Raster Distortion
Trace Finder
POWER SUPPLY
Performance Requirement
115 volts nominal or 230 volts nominal
90 to 110 volts
104 to 126 volts
112 to 136 volts
180 to 220 volts
208 to 252 volts
224 to 272 volts
———
48 to 440 Hz
CATHODE-RAY TUBE (CRT)
I
Internal
Six divisions vertical by 10 divisions horizontal.
Each division equals 0.8 centimeter.
0.1 division or less total
Limits display within graticule area when pressed.
Operational Information
.—
Line voltage and range selected by Line
Voltage Selector assembly on rear
panel. Voltage ranges apply for waveform distortion which does not reduce
the peak line voltage more than 5%
below the true sine-wave peak value.
92 watts (105 volt-amperes)
Tektronix T4530-31-1 rectangular
P31 standard. Others available on
special order.
Approximately 10 kV total (cathode
potential -1.95 kV).
Variable edge lighting
Bias-type, DC coupled to CRT grid.
Adjustable with Geometry and Y Axis
Align adjustments.
Characteristic
Temperature
Operating
Non-operating
Altitude
Operating
Non-operating
Humidity
Non-operating
Vibration
Operating and
non-operating
Shock
Operating and
non-operating
ENVIRONMENTAL CHARACTERISTICS
The following environmental test limits apply when tested in accordance with the recommended test procedure. This instrument will meet the electrical characteristics given in this
section following environmental test.
including failure criteria, etc., may be obtained from Tektronix, Inc. Contact your local
Tektronix Field Office or representative.
Performance Requirement
-15°C to +55°c
-55° to +75°C
15,000 feet maximum
50,000 feet maximum
Five cycles (120 hours) of Mil-Std-202C, Method
106B
15 minutes along each of the three major axes
at a total displacement of 0.025-inch peak to peak
(4 g at 55 c/s) with frequency varied from 10-5510 c/s in one-minute cycles. Hold at 55 c/s for
three minutes on each axis.
Two shocks of 30 g, one-half sine, 11 millisecond
duration each direction along each major axis.
Complete details on environmental test procedures,
Supplemental Information
Fan at rear circulates air throughout instrument.
cutout protects instrument from overheating.
Derate maximum operating tempera-
ture by 1°C/1000 feet change in altitude
above 5000 feet.
Exclude freezing and vibration
Instrument secured to vibration platform
during test. Total vibration time, about
55 minutes.
Guillotine-type shocks.
Total of 12 shocks
Automatic resetting thermal
1-6
ENVIRONMENTAL CHARACTERISTICS (cont)
TM 11-6625-1722-15
Characteristic
Transportation
Package vibration
Package drop
Type 453
Type R453
MECHANICAL CHARACTERISTICS
Characteristic
Construction
Chassis
Panel
Cabinet
Circuit boards
Overall Dimensions, Type
453 (measured ot maximum points)
Height
Width
Length
Overall Dimensions, Type
R453 (measured at maximum points)
Height
Performance Requirement
Meets National Safe Transit type of test when
packaged as shipped from Tektronix, Inc.
One hour vibration slightly in excess of 1 g.
30-inch drop on any corner, edge or flat surface.
18-inch drop on any corner, edge or flat surface.
Width
Information
Length
Aluminum alloy
Aluminum alloy with ano-
dized finish
Blue vinyl-coated aluminum
Glass-epoxy laminate
Connectors
Z AXIS INPUT
All other connectors
Net Weight
Type 453 (includes front
cover without accessor-
Type R453 (without ac-
handle positioned for carrying.
Standard accessories supplied with the Type 453 and R453
are listed on the last pullout page of the Mechanical Parts
List illustrations.
7 inches
Operational Information
Package should just leave vibration surface
19 inches
panel;
Binding post
BNC
Approximately 29 pounds.
ies)
Approximately 32 pounds.
cessories)
STANDARD ACCESSORIES
1-7
SECTION 2
OPERATING INSTRUCTIONS
General
O effectively use the Type 453, the operation and capa-
T
bilities of the instrument must be known. This section describes the operation of the front-, side- and rear-panel
controls and connectors, gives first time and general operating information and lists some basic applications for this
instrument.
Front Cover and Handle
The front cover furnished with the Type 453 provides a
dust-tight seal around the front panel. Use the cover to
protect the front panel when storing or transporting the
instrument. The cover also provides storage space for probes
and other accessories (see Fig. 2-1).
TM 11-6625-1722-15
Fig. 2-1. Accessory storage provided in front cover.
The handle af the Type 453 can be positioned for carrying
or as a tilt-stand for the instrument. To position the handle,
press in at both pivot points (see Fig. 2-2) and turn the
handle to the desired position. Several positions are provided for convenient carrying or viewing. The instrument
may also be set an the rear-panel feet for operation or
storage.
Operating Voltage
The Type 453 can be operated from either a 115-volt or
a 230-volt nominal line-voltage source. The Line Voltage
Fig. 2-2. Handle positioned to provide a stand for the instrument
Selector assembly on the rear panel converts the instrument
from one operating range to the other. In addition, this
assembly changes the primary connections of the power
transformer to allow selection of one of three regulating
ranges. The assembly also includes the two line fuses. When
the instrument is converted from 115-volt to 230-volt nominal
operation, or vice versa, the assembly connects or disconnects one of the fuses to provide the correct protection for
the instrument. Use the following procedure to convert this
instrument between nominal line voltages or regulating
ranges.
1. Disconnect the instrument from the power source.
2. Loosen the two captive screws which hold the cover
onto the voltage selector assembly; then pull to remove the
cover.
3. To convert from 115-volts nominal to 230-volts nominal line voltage, pull out the Voltage Selector switch bar
(see Fig. 2-3]; turn it around 180° and plug it back into the
remaining holes. Change the line-cord power plug to match
the power-source receptacle or use a 115- to 230-volt
adapter.
4. To change regulating ranges, pull out the Range
Selector switch bar (see Fig. 2-3); slide it to the desired
position and plug it back in. Select a range which is centered about the average line voltage to which the instrument is to be connected (see Table 2-1).
5. Re-install the cover and tighten the two captive screws.
6. Before applying power to the instrument, check that
the indicating tabs on the switch bars are protruding through
the correct holes for the desired nominal line voltage and
regulating range.
2-1
TM 11-6625-1722-15
Fig. 2-3. Line Voltage Selector assembly on the rear panel (shown
with cover removed).
CAUTION
The Type 453 should not be operated with the
Voltage Selector or Range Selector switches in the
wrong positions for the line voltage applied.
Operation of the instrument with the switches in
the wrong positions may either provide incorrect
aperotion or damage the instrument.
TABLE 2-1
more frequently. The air filter should be cleaned occasion-
ally to aII
OW the maximum amount of cooling air to enter
the instrument. Cleaning instructions are given in Section 4.
The Type 453 can be operated where the ambient air
temperature is between
-15°C and +55°C. Derate the
maximum operating temperature 1°C for each additional
1000 feet of altitude above 5000 feet. This instrument can
be stored in ambient temperatures between –55°C and
+75°C. After storage at temperatures beyond the operating
limits, allow the chassis temperature to come within the
operating limits before power is applied.
Rackmounting
Complete information for mounting the Type R453 in a
cabinet rack is given in Section 10 of this manual.
CONTROLS AND CONNECTORS
A brief description of the function or operation
front-, side- and rear-panel controls and connectors
(see Fig. 2-4). More detailed information is given
of the
follows
in this
section under General Operating Information.
Cathode-Ray Tube
INTENSITY
FOCUS
Controls brightness of display.
Provides adjustment for a well-defined dis-
play.
SCALE ILLUM
TRACE FINDER
Controls graticule illumination.
Compresses display within graticule area
independent of display position or appli-
ed signals.
Operating Temperature
The Type 453 is cooled by air drawn in at the rear and
blown out through holes in the top and bottom covers. Adequate clearance on the top, bottom and rear must be provided to allow heat to be dissipated away from the instru-
ment. The clearance provided by the feet at the bottom and
rear should be maintained. If possible, allow about one
inch of clearance on the top. Do not block or restrict the air
flow from the air-escape holes in the cabinet.
A thermal cutout in this instrument provides thermal protection and disconnects the power to the instrument if the
internal temperature exceeds a safe operating level. Operation of the instrument for extended periods without the covers
may cause it to overheat and the thermal cutout to open
Vertical (both
VOLTS/DIV
VARIABLE
UNCAL
POSITION
GAIN
Input Coupling
(AC GND DC)
channels except as noted)
Selects vertical deflection factor (VARlABLE control must be in CAL position for
indicated deflection factor).
Provides continuously variable deflection
factor between the calibrated settings of
the VOLTS/DIV switch.
Light indicates that the VARIABLE control
is not in the CAL position.
Controls vertical position of trace.
Screwdriver adjustment to set gain of the
Vertical Preamp. Line between adjustment and 20 mV VOLTS/DIV position indicates that gain should be set with
VOLTS/DIV switch in this position.
Selects method of coupling input signal
to Vertical Deflection System.
AC: DC component of input signal is
blocked. Low frequency limit -3 dB
point) is about 1.6 hertz.
GND: Input circuit is grounded (does not
ground applied signal).
2-2
TM 11-6625-1722-15
Fig. 2-4. Front-, side- and rear-panel controls and connectors.
2-3
TM 11-6625-1722-15
DC: All components of the input signal are
passed to the Vertical Deflection System.
STEP ATTEN
BAL
INPUT
MODE
TRIGGER
INVERT (CH 2
only)
A and B Triggering (both where applicable)
EXT TRIG
INPUT
SOURCE
Screwdriver adjustment to balance Vertical Deflection System in the 5, 10 and 20
mV positions of the VOLTS/DIV switch.
Vertical input connector for signal.
Selects vertical mode of operation.
CH 1: The Channel 1 signal is displayed.
CH 2: The Channel 2 signal is displayed.
ALT: Dual trace display of signal on both
channels. Display switched at end of
each sweep.
CHOP: Dual trace display of signal on
both channels. Approximately one-microsecond segmerlts from each channel
displayed at a repetition rate of about
500 kilohertz.
ADD: Channel 1 and 2 signals are alge-
braically added and the algebraic sum
is displayed on the CRT.
Selects saurce of internal trigger signal
from vertical system.
NORM: Sweep circuits triggered from dis-
played channel(s). Channel 1 signal
available at CH 1 OUT connector.
CH 1 ONLY: Sweep circuits triggered only
from signal applied to the Channel 1
INPUT connector. No signal available
at CH 1 OUT connector. CH 1 lights,
located beside A and B SOURCE switches indicate when the TRIGGER switch is
in the CH 1 ONLY position.
Inverts the Channel 2 signal when pulled
out.
Input connector for external trigger signal.
Connector in B Triggering section of front
panel also serves as external horizontal
input when HORIZ DISPLAY switch is in
EXT HORIZ position and B SOURCE switch
is in EXT position.
Selects source of trigger signal.
INT: Internal trigger signal obtained from
Vertical Deflection System. When CH
1 light is on, trigger signal is obtained
only from the Channel 1 input signal;
when the light is off, the trigger signal
is abtained from the displayed chan-
nel(s). Source of internal trigger signal
is selected by the TRIGGER switch.
LINE: Trigger signal obtained from a sam-
ple of the line voltage applied to this
instrument.
EXT: Trigger signal obtained from an ex-
ternal signal applied to the EXT TRIG
INPUT connector.
nal approximately 10 times.
CH 1
COUPLING
SLOPE
LEVEL
HF STABDecreases display jitter for high-frequency
(A Triggering only)sweep rates.
A and B Sweep
DELAY-TIME
MULTIPLIER
A SWEEP
TRIG’D
UNCAL A
OR B
A AND B
TIME/DIV
AND DELAY
TIME
Light indicates that the internal trigger signal is abtained only from the signal connected to the Channel 1 INPUT connector
(see TRIGGER switch).
Determines method of coupling trigger
signal to trigger circuit.
AC: Rejects DC and attenuates signals be-
low about 30 hertz. Accepts signals
between abaut 30 hertz and 50 megahertz.
LF REJ: Rejects DC and attenuates signals
below about 30 kilohertz. Accepts signals between about 30 kilohertz and 50
megahertz.
HF REJ: Accepts signals between about 30
hertz and 50 kilohertz; rejects DC and
attenuates signals outside the above
range.
DC: Accepts all trigger signals from DC to
50 megahertz or greater.
Selects portion of trigger signal which
starts the sweep.
+: Sweep can be triggered from positive-
going portion of trigger signal.
-: Sweep can be triggered from negativegoing portion of trigger signal.
Selects amplitude point on trigger signal
at which sweep is triggered.
signals. Has negligible effect at lower
Provides variable sweep delay between
0.20 and 10.20 times the delay time indi-
cated by the A TIME/DIV switch.
Light indicates that A sweep is triggered
and will produce a stable display with
correct INTENSITY and POSITION control
settings.
Light indicates that either the A or B
VARIABLE control is not in the CAL posi-
tion.
A TIME/DIV switch (clear plastic flange)
selects the sweep rate of the A sweep
circuit for A sweep only operatian and
selects the basic delay time (to be multi-
plied by DELAY-TIME MULTIPLIER dial
setting) for delayed sweep operation.
B TIME/DIV (DELAYED SWEEP) switch
selects sweep rate of the B sweep circuit
Attenuates external trigger sig-
2-4
TM 11-6625-1722-15
A VARIABLE
B SWEEP
MODE
HORIZ
DISPLAY
MAG
A SWEEP
MODE
for delayed sweep operation only. VARlABLE controls must be in CAL positions for
calibrated sweep rates.
Provides continuously variable A sweep
rate to at least 2.5 times setting of the
A TIME/DIV switch. A sweep rate is calibrated when control is set fully clockwise
to CAL.
Selects B sweep operation mode.
TRIGGERABLE AFTER DELAY TIME: B
sweep circuit will not produce a sweep
until a trigger pulse is received following the delay time selected by the
DELAY TIME (A TIME/DIV) switch and
the DELAY-TIME MULTIPLIER dial.
B STARTS AFTER DELAY TIME: B sweep
circuit runs immediately following delay
time selected by the DELAY TIME switch
and DELAY-TIME MULTIPLIER dial.
Selects horizontal mode of operation.
A: Horizontal deflection provided by A
sweep.
B sweep inoperative.
A INTEN DURING B: Sweep rate deter-
mined by A TIME/DIV switch. An inten-
sified portion appears on the sweep
during the B sweep time. This position
provides a check of the duration and
position of the delayed sweep (B) with
respect to the delaying sweep (A).
DELAYED SWEEP (B): Sweep rate deter-
mined by B TIME/DIV switch with the
delay time determined by the setting of
the DELAY TIME (A TIME/DIV) switch
and the DELAY-TIME MULTIPLIER dial.
Sweep mode determined by B SWEEP
MODE switch.
EXT HORIZ: Horizontal deflection pro-
vided by an external signal.
Increases sweep rate to ten times setting
of A or B TIME/DIV switch by horizontally
expanding the center division of the display. Light indicates when magnifier is
on.
Determines the operating mode for A
sweep.
AUTO TRIG: Sweep initiated by the ap-
plied trigger signal using the A Trig-
gering controls when the trigger signal
repetition rate is above about 20 hertz.
For lower repetition rates or when there
is no trigger signal, the sweep free runs
at the sweep rate selected by the A
TIME/DIV switch to produce a bright
reference trace.
NORM TRIG: Sweep initiated by the ap-
plied trigger signal using the A Trig-
gering controls. No trace is displayed
when there is no trigger signal.
RESET
A SWEEP
LENGTH
POSITION
FINE
1 kHz CAL
POWER ON
Side Panel
ASTIG
B TiME/DIVVARIABLE
PROBE LOOP
A GATE
B GATE
CH 1 OUT
SINGLE SWEEP: After a sweep is display-
ed, further sweeps cannot be presented
until the RESET button is pressed. Display is triggered as for NORM operation using the A Triggering controls.
When the RESET button is pressed (SINGLE SWEEP mode), a single display will
be presented (with correct triggering) when
the next trigger pulse is received. RESET
light (inside RESET button) remains on
until a trigger is received and the sweep is
completed. RESET button must be pressed
before another sweep can be presented.
Adiusts length of A sweep. In the FULL
position (clockwise detent), the sweep is
about 11 divisions long. As the control is
rotated counterclockwise, the length of A
sweep is reduced until it is less than four
divisions long iust before the detent in
the fully-counterclockwise position is reached. In the B ENDS A position (counterclockwise detent), the A sweep is reset
at the end of the B sweep to provide the
fastest possible sweep repetition rate for
delayed sweep displays.
Controls horizontal position of trace.
Provides more precise horizontal position
adjustment.
Calibrator output connector.
Light: Indicates that POWER switch is
on and the instrument is connected to
a line voltage source.
Switch: Controls power to the instrument.
Screwdriver adjustment used in conjunction with the FOCUS control to obtain a
well-defined display. Does not require
readjustment in normal use.
Provides continuously variable sweep rate
to at least 2.5 times setting of B TIME/DIV
switch. B sweep rate is calibrated when
control is set fully clockwise to CAL.
Current loop providing five-milliampere
square-wave current from calibrator circuit.
Output connector providing a rectangular
pulse coincident with A sweep.
Output connector providing a rectangular
pulse coincident with B sweep.
Output connector providing a sample of
the signal applied to the Channel 1 lN-
PUT connector when the TRIGGER switch
is in the NORM position.
2-5
TM 11-6625-1722-15
CALIBRATOR
TRACE
ROTATION
Rear Panel
Z AXIS INPUT
Line Voltage
Selector
The following steps will demonstrate the use of the controls and connectors of the Type 453. It is recommended
that this procedure be followed completely for familiarization with this instrument.
Setup Information
1. Set the front-panel controls as follows:
CRT Controls
INTENSITY
FOCUS
SCALE ILLUMCounterclockwise
Vertical Controls (both channels if applicable)
VOLTS/DIV20 mV
VARIABLE
POSITIONMidrange
INPUT COUPLING
MODE
TRIGGER
INVERT
Triggering Controls (both A and B if applicable)
LEVEL
SLOPE
COUPLING
SOURCE
Sweep Controls
DELAY-TIME
MULTIPLIER
A and B TIME/DIV
A VARIABLE
B SWEEP MODE
Switch selects output voltoge of Calibrator.
1-volt or 0.1-volt square wave available.
Screwdriver adjustment to align trace with
horizontal graticule lines.
Input connector for intensity modulation
of the CRT display.
Switching assembly to select the nominal
operating voltage and the line voltage
range.
line fuses.
Voltage Selector: Selects nominal operat-
Range Selector: Selects line voltage range
FIRST-TIME OPERATION
The assembly also includes the
ing voltage range (115V or 230V).
(low, medium, high).
Counterclockwise
Midrange
CAL
DC
CH 1
NORM
Pushed in
Clockwise (+)
+
AC
INT
0.20
.5 ms
CAL
B STARTS AFTER
DELAY TIME
HORIZ DISPLAY
MAG
POSITION
A SWEEP LENGTH
A SWEEP MODE
POWER
Side-Panel Controls
B TIME/DIV VARIABLE
CALIBRATOR
2. Connect the Type 453 to a power source that meets
the voltage and frequency requirements of the instrument.
If the available line voltage is outside the limits of the Line
Voltage Selector assembly position (on rear panel), see
Operating Voltage in this section.
3. Set the POWER switch to ON. Allow about five minutes
warmup so the instrument reaches a normal operating temperature before proceeding.
CRT Controls
4. Advance the INTENSITY control until the trace is at
the desired viewing level (near midrange).
5. Connect the 1 kHz CAL connector to the Channel 1
INPUT connector with a BNC cable.
6. Turn the A LEVEL control toward 0 until the display
becomes stable. Note that the A SWEEP TRIG’D light is on
when the display is stable.
7. Adiust the FOCUS control for a sharp, well-defined
display over the entire trace length. (If focused display can-
not be obtained, see Astigmatism Adjustment in this section.)
8. Disconnect the input signal and move the trace with
the Channel 1 POSITION control so it coincides with one of
the horizontal graticule lines.
the graticule line, see Trace Alignment Adjustment in this
section.
9. Rotate the SCALE ILLUM control throughout its range
and notice that the graticule lines are illuminated as the
control is turned clockwise (most obvious with mesh or smokegray filter installed). Set control so graticule lines are
illuminated as desired.
Vertical Controls
10. Change the CH 1 VOLTS/DIV switch from 20 mV to 5
mV. If the vertical position of the trace shifts, see Step
Attenuator Balance in this section.
11. Set the CH 1 VOLTS/DIV switch to 20 mV and set the
Channel 1 Input Coupling switch to AC. Connect the 1 kHz
CAL connector to both the Channel 1 and 2 INPUT connectors with two BNC cables and a BNC T connector.
If the BNC cables and BNC T connector are not
available, make the following changes in the procedure. Place the BNC jack post (supplied accessory) on the 1 kHz CAL connector and connect
A
OFF
Midrange
FULL
AUTO TRIG
OFF
CAL
.1 V
If the trace is not parallel with
NOTE
2-6
TM 11-6625-1722-15
the Channel 1 and 2 INPUT connectors. Connect
the probe tips to the BNC jack post. Set the CALIBRATOR switch (on side-panel) to 1 V.
12. Turn the Channel 1 POSITION control to center the
display. The display is a square wave, five divisions in
amplitude with about five cycles displayed on the screen. If
the display is not five divisions in amplitude, see Vertical
Gain Adjustment in this section.
13. Set the Channel 1 Input Coupling switch to GND ond
position the trace to the center horizontal line.
14. Set the Channel 1 Input Coupling switch to DC. Note
that the baseline of the waveform remains at the center
horizontal line (ground reference).
15. Set the Channel 1 Input Coupling switch to AC. Note
that the waveform is centered about the center horizontal
line.
16. Turn the Channel 1 VARIABLE control throughout its
range. Note that the UNCAL light comes on when the VARl-
ABLE control is moved from the CAL position (fully clockwise). The deflection should be reduced to about two divisions. Return the VARIABLE control to CAL.
17. Set the MODE switch to CH 2.
18. Turn the Channel 2 POSITION control to center the
display. The display will be similar to the previous display
for Channel 1. Check Channel 2 step attenuator balance and
gain as described in steps 10 through 12. The Channel 2
input Coupling switch and VARIABLE control operate as
described in steps 13 through 16.
19. Set both VOLTS/DIV switches to 50 mV.
20. Set the MODE switch to ALT and position the Channel
1 waveform to the top of the graticule area and the Chan-
nel 2 waveform to the bottom of the graticule area. Turn
the A TIME/DIV switch throughout its range. Note that the
display alternates between channels at all sweep rates.
Triggering
25. Set the CALIBRATOR switch to 1 V. Rotate the A LEVEL
control throughout its range. The display free runs at the
extremes of rotation. Note that the A SWEEP TRIG'D light
is on only when the display is triggered.
26. Set the A SWEEP MODE switch to NORM TRIG. Again
rotate the A LEVEL control throughout its range. A display
is presented only when correctly triggered. The A SWEEP
TRIG'D light operates as in AUTO TRIG. Return the A SWEEP
MODE switch to AUTO TRIG.
27. Set the A SLOPE switch to -. The trace starts on the
negative part of the square wave. Return the switch to +;
the trace starts with the positive part of the square wave.
28. Set the A COUPLING switch to DC. Turn the Channel 1 POSITION control until the display becomes unstable
(only part of square wave visible). Return the A COUPLING
switch to AC; the display is again stable. Since changing
trace position changes DC level, this shows how DC level
changes affect DC trigger coupling. Return the display to
the center of the screen.
29. Set the MODE switch to CH 2; the display should be
stable. Remove the signal connected to Channel 1; the dis-
play free runs.
Set the TRIGGER switch to NORM; the dis-
play is again stable. Note that the CH 1 lights in A and B
Triggering go out when the TRIGGER switch is changed to
NORM.
30. Connect the Calibrator signal to both the Channel 2
INPUT and A EXT TRIG INPUT connectors. Set the A
SOURCE switch to EXT. Operation of the LEVEL, SLOPE
and COUPLING controls for external triggering are the same
as described in steps 25 through 28.
31. Set the A SOURCE switch to EXT
is the same as for EXT. Note that the A LEVEL control has
less range in this position, indicating trigger signal attenuation. Return the A SOURCE switch to INT.
32. Operation of the B Triggering controls is similar to
A Triggering.
21. Set the MODE switch to CHOP and the A TIME/DIV
Note the switching between channels as
shown by the segmented trace. Set the TRIGGER switch to
CH 1 ONLY; the trace should appear more solid, since it is
no longer triggered on the between-channel switching trans-
ients. Turn the A TIME/DIV switch throughout its range. A
dual-trace display is presented at all sweep rates, but unlike
ALT, both channels are displayed on each trace on a timesharing basis. Return the A TIME/DIV switch to .5 ms.
22. Set the MODE switch to ADD. The display should be
four divisions in amplitude. Note that either POSITION control moves the display.
23. Pull the INVERT switch. The display is a straight line
indicating that the algebraic sum of the two signals is zero
(if the Channel 1 and 2 gain is correct).
24. Set either VOLTS/DIV switch to 20 mV. The square-
wave display indicates that the algebraic sum of the two
signals is no longer zero.
Return the MODE switch to CH 1
and both VOLTS/DIV switches to .2 (if using 10X probes,
set both VOLTS/DIV switches to 20 mV). Push in the INVERT
switch.
Normal and Magnified Sweep
33. Set the A TIME/DIV switch to 5 ms and the MAG
switch to X10. The display should be similar to that obtained with the A TIME/DIV switch set to .5 ms and the MAG
switch to OFF.
34. Turn the horizontal POSITION control throughout its
range; it should be possible to position the display across
the complete graticule area. Now turn the FINE control. The
display moves a smaller amount and allows more precise
positioning. Return the A TIME/DIV switch to .5 ms, the
MAG switch to OFF and return the start of the trace to the
left graticule line.
Delayed Sweep
36. Pull the DELAYED SWEEP knob out and turn it to 50
(DELAY TIME remains at .5 ms). Set the HORIZ DISPLAY
switch to A INTEN DURING B. An intensified portion, about
one division in length, should be shown at the start of the
trace. Rotate the DELAY-TIME MULTIPLIER dial throughout
its range; the intensified portion should move along the display.
2-7
TM 11-6625-1722-15
37. Set the B SWEEP MODE switch to TRIGGERABLE
AFTER DELAY TIME. Again rolate the DELAY-TIME MULTI-
PLIER dial throughout its range and note that the intensified
portion appears to jump between posltive slopes of the display. Set the B SLOPE switch to -;
begins on the negative slope.
the intensified portion
Rotate the B LEVEL control;
the intensified portion of the display disappears when the B
LEVEL control is out of the trlggerable range. Return the
B LEVEL control to 0.
38. Set the HORIZ DISPLAY switch to DELAYED SWEEP
(B). Rotate the DELAY-TIME MULTIPLIER dial througout its
range; about one-half cycle of the waveform should be
displayed on the screen (Ieading edge visible only at high
INTENSITY control setting). The display remains stable on
the screen, indicating that the B sweep is triggered.
39. Set the B SWEEP MODE switch to B STARTS AFTER
DELAY TIME. Rotate the DELAY-TIME MULTIPLIER dial
throughout its range; the display moves continously across
the screen as the control is rotated.
40. Rotate the DELAYTIME MIULTIPLIER dial fully counter-
clockwise and set the HORIZ DISPLAY switch to A INTEN
DURING B. Rotate the A SWEEP LENGTH control counter-
clockwise; the Iength of the display decreases. Set the con-
trol to the B ENDS A position; now the display ends after
the intensified portion. Rotate the DELAY-TIME MULTIPLIER
dial ond note that the sweep length increases as the display
moves across the screen. Return the A SWEEP LENGTH con-
trol to FULL and the HORIZ DISPLAY switch to A.
Single Sweep
41. Set the A SWEEP MODE switch to SINGLE SWEEP.
Remove the Calibrator signal from the Channel 2 INPUT
connector. Press the RESET button; the RESET light should
come on and remain on. Again apply the signal to the Channel 2 INPUT connector; a single trace should be presented
and the RESET light should go out. Return the A SWEEP
MODE switch to AUTO TRIG.
External Horizontal
42. Connect the Calibrator signal to both the Channel 2
INPUT and EXT HORIZ (B EXT TRIG lNPUT] connectors. Set
the B SOURCE switch to EXT, B COUPLING switch to DC
and the HORIZ DISPLAY switch to EXT HORIZ. lncrease the
INTENSITY control setting until the display is visible (two dots
displayed diagonally). The display should be five divisions
vertically and about 3.7 divisions horizontally. Set the B
10. The display should be reduced
ten times horizontally. The display can be positioned horizontally with the horizontal POSITION or FINE control and
vertically with the Channel 2 POSITION control.
43. Connect the Calibrator signal to both the Channel 1
and 2 INPUT connectors. Set the TRIGGER switch to CH 1
ONLY and the B SOURCE switch to INT.
44. The display should be five divisions vertically and
horizontally.
The display can be postioned horizontally
with the Channel 1 POSITION control and vertically with the
Channel 2 POSITION control.
45. Change the CH1 VOLTS/DIV switch to 5. The display
play is reduced to two divisions horizontally. Now set the
CH 2 VOLTS/DIV switch to 5. The display is reduced to two
divisions vertically.
Trace Finder
46. Set the CH 1 and CH 2 VOLTS/DlV switches to 10
mv. The display i
S not visible since it exceeds the scan area
of the CRT.
47. Press the TRACE FINDER button. Note that the disis returned to the display area. While holding the
play
TRACE FINDER button depressed, increase the vertical and
horizontal deflection factors until the display is reduced to
about two divisions vertically ond horizontally. Adjust the
Channel 1 and 2 POSITION controls to center the display
about the center lines of the graticule. Release the TRACE
FINDER and note that the display remains within the viewing
area. Disconnect the applied signal.
48. Reduce the INTENSITY control setting to normal, B
SOURCE switch to INT and set the HORIZ DISPLAY switch
to A.
Z-Axis Input
49. If an, External signal is available (five volts peak to
peak minimum] the function of tlhe Z AXIS INPUT circuit can
be demonstrated. Connect the external signal to both the
Channel 2 INPUT connector and the Z AXIS INPUT binding
posts. Set the A TIME/DIV switch to display about five
cycles of the waveform. The positive peaks of the waveform
should be blanked and the negative peaks intensified, indi-
cating intensify modulation.
50. This completes the basic operating procedure for the
Type 453. lnstrument operation not explained here, or operations which need further explanation are discussed under
General Operating Information.
CONTROL SETUP CHART
Fig. 2-5 shows the front, side and rear panels of the Type
453. This chart can, be reproduced and used as a test-setup
record for special measurements, applications or procedures,
or it may be used as a training aid for familiarization with
this instrument.
GENERAL OPERATING INFORMATION
Intensify Control
The setting of the INTENSITY control may affect the correct focus of the display. Slight readjustment of the FOCUS
control may be necessary when the intensity level is changed.
To protect the CRT phosphor, do not turn the INTENSITY
control higher than necessary to provide a satisfactory dis-
play. The light filters reduce the observed light output from
the CRT. When using these filters, avoid odvancing the
INTENSITY control to a setting that may bum the phosphor.
When Ihe highest intensity display is desired, remove the
filters and use the clear faceplate protector. Also, be careful that the INTENSITY control is not set too high when
changing the TIME/DlV switch front a fast to a slow sweep
rate, or when changing the HORIZ DISPLAY switch from EXT
HORIZ operation to the norrmal sweep mode.
Astigmatism Adjustment
If a well-defined trace cannot be obtained with the FOCUS
control, adjust the ASTIG adjustment (side panel) as fol-
lows.
2-8
TM 11-6625-1722-15
Fig. 2-5. Control setup chart for the Type 453.
2-9
TM 11-6625-1722-15
NOTE
To check for proper setting of the ASTIG adjustment, slowly turn the FOCUS control through the
optimum setting. If the ASTIG adjustment is cor-
rectly set, the vertical and horizontal portions
of the trace will come into sharpest focus at the
same position of the FOCUS control. This setting
of the ASTIG adjustment should be correct for any
display. However, it may be necessary to reset
the FOCUS control slightly when the INTENSITY
control is changed.
1. Connect a 1 V Calibrator signal to either channel and
set the VOLTS/DIV switch of that channel to present a twodivision display. Set the MODE switch to display the chan-
nel selected.
2. Set the TIME/DIV switch to .2 ms
3. With the FOCUS control and ASTIG adjustment set to
midrange, adjust the INTENSITY control so the rising portion
of the display can be seen.
4. Set the ASTIG adjustment so the horizontal and vertical portions of the display are equally focused, but not
necessarily well focused.
5. Set the FOCUS control so the vertical portion of the
trace is as thin as possible.
6. Repeat steps 4 and 5 for best overall focus. Make final
check at normal intensity.
Graticule
The graticule of the Type 453 is internally marked on the
faceplate of the CRT to provide accurate, no-parallax measurements. The graticule is marked with six vertical and 10
horizontal divisions. Each division is 0.8 centimeter square.
In addition, each major division is divided inta five minor
divisions at the center vertical and horizontal lines. The
vertical gain and horizontal timing are calibrated to the
graticule so accurate measurements can be made from the
CRT. The illumination of the graticule lines can be varied
with the SCALE ILLUM control.
Fig. 2-6 shows the graticule of the Type 453 and defines
the various measurement lines. The terminology defined here
will be used in all discussions involving graticule measurements.
Fig. 2-6. Definition of measurement lines on Type 453 graticule.
remove the filter, press
down at the bottom of the frame
and pull the top of the filter away from the CRT faceplate
(see Fig. 2-7).
The tinted light filter minimizes light reflections from the
face of the CRT to improve contrast when viewing the display under high ambient light conditions. A clear plastic
faceplate protector is also provided with this instrument for
use when neither the mesh nor the tinted filter is used. The
clear faceplate protector provides the best display for waveform photographs. It is also preferable for viewing high
writing rate displays.
A filter or the faceplate protector should be used at all
times ta protect the CRT faceplate from scratches. The faceplate protector and the tinted filter mount in the same holder.
Trace Alignment Adjustment
If a free-running trace is not parallel to the horizontal
graticule lines, set the TRACE ROTATION adjustment as fol-
lows. Position the troce to the center horizontal line. Adjust
the TRACE ROTATION adjustment (side panel) so the trace
is parallel with the horizontal graticule lines.
Light Filter
The mesh filter provided with the Type 453 provides shielding against radiated EMI (electro-magnetic interference) from
the face of the CRT. It also serves as a light filter to make
the trace more visible under ambient light conditions. To
2-10
Fig. 2-7. Removing the filter or faceplate protector.
TM 11-6625-1722-15
To remove the light filter or faceplate protector from the
holder, press it out to the rear. They can be replaced by
snapping them back into the holder.
Trace Finder
The TRACE FINDER provides a means af locating a display
which overscans the viewing area either vertically or horizontally. When the TRACE FINDER button is pressed, the
display is compressed within the graticule area. T
O locate
and reposition an overscanned display, use the following
procedure.
1. Press the TRACE FINDER button.
2. While the TRACE FINDER button is held depressed,
increase the vertical and horizontal deflection factors until
the vertical deflection is reduced to about two divisions and
the horizontal deflection is reduced to about four divisions
(the horizontal deflection needs to be reduced only when in
the external horizontal mode of operation).
3. Adjust the vertical and horizontal POSITION controls
to center the display about the vertical and horizontal cen-
ter lines.
4. Release the TRACE FINDER button; the display should
remain within the viewing area.
Vertical Channel Selection
Either of the input channels can be used for single-trace
displays. Apply the signal to the desired INPUT connector
and set the MODE switch to display the channel used. However, since CH 1 ONLY triggering is provided only in Chan-
nel 1 and the invert feature only in Channel 2, the correct
channel must be selected to take advantage of these features. For dual-trace displays, connect the signals to both
INPUT connectors and set the MODE switch to one of the
dual-trace positions.
Vertical Gain Adjustment
To check the gain of either channel, set the VOLTS/DIV
switch to 20 mV. Set the CALIBRATOR switch to .1 V and
connect the 1 kHz CAL connector to the INPUT of the chan-
nel used. The vertical deflection should be exactly five
divisions.
If not, adjust the front-panel GAIN adjustment
for exactly five divisions of deflection.
NOTE
If the gain of the two channels must be closely
matched (such as for ADD mode operation), the
ADJUSTMENT procedure given in the Calibration
section should be used.
The best measurement accuracy when using probes is provided if the GAIN adjustment is made with the probes installed (set the CALIBRATOR switch to 1 V). Also, to provide
the most accurate measurements, calibrate the vertical gain
of the Type 453 at the temperature at which the measurement
is to be made.
Step Attenuator Balance
To check the step attenuator balance of either channel,
set the Input Coupling switch to GND and set the A SWEEP
MODE swich to AUTO TRIG to provide a free-running trace.
Change the VOLTS/DIV switch from 20 mV to 5 mV. If the
trace moves vertically, adjust the front-panel STEP ATTEN
BAL adjustment as follows (allow at least 10 minutes warm
up before performing this adjustment].
1. With the Input Coupling switch set to GND and the
VOLTS/DIV switch set to 20 mV, move the trace to the center
horizontal line of the graticule with the vertical POSITION
control.
2. Set the VOLTS/DIV switch to 5 mV and adjust the
STEP ATTEN BAL adjustment to return the trace to the center
horizonal line.
3. Recheck step attenuator balance and repeat adjustment
until no trace shift occurs as the VOLTS/DIV switch is changed from 20 mV to 5 mV.
Signal Connections
In general, probes offer the most convenient means of connecting a signal to the input of the Type 453. The Tektronix
probes are shielded to prevent pickup of electrostatic interference. A 10X attenuator probe offers a high input impedance and allows the circuit under test to perform very close
to normal operating conditions. However, a 10X probe
also attenuates the input signal 10 times.
The Tektronix
P6045 Field Effect Transistor probe and accessory power
supply offer the same high-input impedance as the 10X
probes. However, it is particularly useful since it provides
wide-band operatian while presenting no attenuation (1X
gain) and a low input capacitance. To obtain maximum
bandwidth when using the probes, observe the grounding
considerations given in the probe manual. The probe-toconnector adapters and the bayonet-ground tip provide the
best frequency response. Remember that a ground strap
only a few inches in length can produce several percent of
ringing when operating at the higher frequency limit of this
system. See your Tektronix, Inc. catalog for characteristics
and compatibility of probes for use with this system.
In high-frequency applications requiring maximum overall
bandwidth, use coaxial cables terminated at both ends in
their characteristic impedance. See the discussion on coaxial cables in this section for more information.
High-level, low-frequency signals can be connected directly
to the Type 453 INPUT connectors with short unshielded
leads. This coupling method works best for signals below
about one kilohertz and deflection foctors above one volt/
division. When this method is used, establish a common
ground between the Type 453 and the equipment under test.
Attempt to position the leads away from any source of interference to avoid errors in the display. If interference is
excessive with unshielded leads, use a coaxial cable or a
probe.
Loading Effect of the Type 453
As nearly as possible, simulate actual operating conditions in the equipment under test. Otherwise, the equipment
under test may not produce a normal signal. The 10X
attenuator probe and field effect transistor probe mentioned
previously offer the least circuit loading. See the probe
instruction manual for loading characteristics of the individual probes.
2-11
TM 11-6625-1722-15
When the signal is coupled directly to the input of the
Type 453, the input impedance is about one megohm
paralleled by about 20 pF. When the signal is coupled to
the input through a coaxial cable, the effeclive input capacitance depends upon the type and Iength of cable used.
See the following discussion for
inforrnation on obtaining
maximum frequency responspe with coaxial cables.
The signal cables used to connect the signal 10 the type
453 INPUT connectors have a Iarge effect on the accuracy
of the displayed high-frequency waveform.
To maitain the
high-frequency characteristics of the applied signal, highquality low-loss coaxial cable should be used. The cable
should be terminated at the Type 453 INPUT connector in its
characteristic impedonce.
with differing characteristic impedances,
If it is necessary to use cables
use suitable imped-
ance-matching devices to provide the correct transition, with
minimum loss, from one impedance to the other.
The characteristic impedance, velocity of propagation and
nature of signal lOSSeS in a coaxial cable are determined
by the physical and electrical characteristics of the cable.
Losses caused by energy dissipation in the dielectric are
proportional to the signal frequency. Therefore, much of
the high-frequency information in a fast-rise pulse can be
lost in only a few feet of interconnecting cable if it is not
the correct type. To be sure of the high-frequency response
of the system when using cables longer than about five feet,
observe the transient response of the Type 453 and the
interconnecting cable with a fast-rise
pulse generator (gen-
erator risetime Iess than 0.5 nanoseconds).
DC components. The pre-charging network incorporated
in this unit aII
OWS the input-coupling capacitor to charge to
the DC source voltage level when the Input Coupling switch
is set to GND. The procedure for using this feature is as
follows:
1. Before connecting the signal containing a DC component to the Type 453 INPUT connector, set the Input Coupling
switch to GND. Then connect the signal to the INPUT
connector.
2. Wait about one second for the coupling capacitor to
charge.
3. Set the Input Coupling switch to AC. The trace (display) will remain on the screen and the AC component of
the signal can be measured in the normal manner.
Deflection Factor
The amount of vertical deflection produced by a signal
is determined by the signal amplitude, the attenuation factor
of the probe (if used), the setting of the VOLTS/DIV switch
and the setting of the VARIABLE VOLTS/DIV control. The
calibrated deflection factors indicated by the VOLTS/DIV
switches apply only when the VARIABLE control is set to
the CAL position.
The VARIABLE VOLTS/DIV control provides variable
(uncalibrated) vertical deflection between the calibrated
settings of the VOLTS/DIV switch. The VARIABLE control
extends the maximum vertical deflection factor of the Type
453 to at least 25 volts/division (10 volts position).
Input Coupling
The Channel 1 and 2 lnput Coupling switches allow a
choice of input caupling. The type of display desired will
determine the coupling used.
The DC position can be used for most applications. However, if the DC component of the signal is much larger than
the AC component, the AC position vvill probably provide a
better display. DC coupling should be used to display AC
signals below about 16 heltz as they will be attenuated in
the AC position.
In the AC position, the DC component of the signal is
blocked by a capacitor in the input circuit. The low-frequency response in the AC position is about 1.6 hertz (–3 dB
point). Therefore,
some low-frequency distortion can be
expected near this frequency limit. Distortion will also
appear
in square waves which have low-frequency com-
ponents.
The GND position provides a ground reference at the
input of the Type 453. The signal applied to the input con-
nector is internally disconnected but not grounded. The
input circuit is held at ground potential, eliminating the need
to externally ground the input to establish a DC ground
reference.
The GND position can also be used to pre-charge the
coupling capacitor to the average voltage level of the signal
applied to the INPUT connector. This allows measurement
of only the AC component of signals having both AC and
Dual-Trace Operation
Alternate Mode. The ALT position of the MODE switch
produces a disploy which alternates between Channel 1 and
2 with each sweep of the CRT. Although the ALT mode
can be used at all sweep rates, the CHOP mode provides
a more satisfactory display at sweep rates below about 50
microseconds/division. At these slower sweep rates, alternate
mode switching becomes visually perceptible.
Proper internal triggering in the ALT mode can be ob-
tained in either the NORM or CH 1 ONLY positions of the
TRIGGER switch. When in the NORM position, the sweep is
triggered from the signal on each channel. This provides a
stable display of two unrelated signals, but does not indicate
the time relationship between the signals. In the CH 1 ONLY
position, the two signals are displayed showing true time
relationship. If the signals are not time related, the Channel
2 waveform will be unstable in the CH 1 ONLY position.
Chopped Mode. The CHOP position of the MODE switch
produces a display which is electronically switched between
channels. In general, the CHOP mode provides the best
display at sweep rates slower than about 50 microseconds/
division, or whenever dual-trace, single-shot phenomena are
to be displayed. At faster sweep rates the chopped switching becomes apparent and may interfere with the display.
Proper internal triggering for the CHOP mode is provided
with the TRIGGER switch set to CH 1 ONLY. If the NORM
position is used, the sweep circuits are triggered from the
between-channel switching signal and both waveforms will
2-12
TM 11-6625-1722-15
be unstable. External triggering provides the same result as
CH 1 ONLY triggering.
Two signals which are time-related can be displayed in
the chopped mode showing true time relationship. If the
signals are not time-related, the Channel 2 display will
appear unstable. Two single-shot, transient, or random
signals which occur within the time interval determined by
the TIME/DIV switch (10 times sweep rate) can be compared
using the CHOP mode. To correctly trigger the sweep for
maximum resolution, the Channel 1 signal must precede the
Channel 2 signal. Since the signals show true time relation-
ship, time-difference measurements can be made.
Channel 1 Output and Cascaded Operation
If a lower deflection factor than provided by the VOLTS/
DIV switches is desired, Channel 1 can be used as a wideband preamplifier for Channel 2. Apply the input signal to
the Channel 1 INPUT connector. Connect a 50-ohm BNC
cable (18-inch or shorter cable for maximum cascaded fre-
quency response) between the CH 1 OUT (side panel) and
the Channel 2 INPUT connectors. Set the MODE switch to
CH 2 and the TRIGGER switch to NORM. With both VOLTS/
DIV switches set to 5 mV, the deflection factor will be less
than one millivolt/division.
To provide calibrated one millivolt/division deflection
factor, connect the .1 volt Calibrator signal to the Channel 1
INPUT connector. Set the CH 1 VOLTS/DIV switch to .1
and the CH 2 VOLTS/DIV switch to 5 mV. Adjust the Channel 2 VARIABLE VOLTS/DIV control to produce a display
exactly five divisions in amplitude. The cascaded deflection
factor is determined by dividing the CH 1 VOLTS/DIV switch
setting by 5 (CH 2 VOLTS/DIV switch and VARIABLE control
remain as set above). For example, with the CH 1 VOLTS/
DIV switch set to 5 mV the calibrated deflection factor will
be 1 millivolt/division; CH 1 VOLTS/DIV switch set to 10 mV,
2 millivolts/division, etc.
The following operating considerations and basic appli-
cations may suggest other uses for this feature.
1. If AC coupling is desired, set the Channel 1 Input
Coupling switch to AC and leave the Channel 2 Input
Coupling switch set to DC. When both Input Coupling
switches are set to DC, DC signal coupling is provided.
2. Keep both vertical POSITION controls set near mid-
range.
one of the POSITION controls being turned away from
midrange, correct operation can be obtained by keeping
the Channel 2 POSITION control near midrange and using
the Channel 1 POSITION control to position the trace near
the desired locatian. Then, use the Channel 2 POSITION
control far exact positioning. This method will keep both
Input Preamps operating in their linear range.
least 25 millivolts/division of CRT display in all CH 1 VOLTS/
DIV switch positions.
DIV control have no effect on the signal available at the
CH 1 OUT connector.
ance matching stage with or without voltage gain. The
If the input signal has a DC level which necessitates
3. The output voltage at the CH 1 OUT connectar is at
4. The MODE switch and Channel 1 VARIABLE VOLTS/
5. The Channel 1 Input Preamp can be used as an imped-
input resistance is one megohm and the output resistance is
about 50 ohms.
6. The dynamic range of the Channel 1 Input Preamp is
equal to about 20 times the CH 1 VOLTS/DIV setting. The
CH 1 OUT signal is nominally at 0 volt DC for a 0 volt DC
input level [Channel 1 POSITION control centered). The
Chanel 1 POSITION control can be used to center the output signal within the dynamic range of the amplifier.
7. If dual-trace operation is used, the signal applied to
the Channel 1 INPUT connector is displayed when Channel
1 is turned on. When Channel 2 is turned on, the amplified
signal is displayed. Thus, Channel 1 trace can be used to
monitor the input signal while the amplified signal is dis-
played by Channel 2.
8. In special applications where the flat frequency re-
sponse of the Type 453 is not desired, a filter inserted
between the CH 1 OUT and Channel 2 INPUT connector
allows the oscilloscope to essentially take on the frequency
response of the filter. Combined with method 7, the input can
be monitored by Channel 1 and the filtered signal displayed
by Channel 2.
9. By using Channel 1 as a 5X low-level voltage preamplifier (5 mV position), the Channel 1 signal available
at the CH 1 OUT connector can be used for any application
where a low-impedance preamplifier signal is needed.
Remember that if a 50-ohm load impedance is used, the
signal amplitude will be about one-half.
Algebraic Addition
General. The ADD position of the MODE switch can be
used to display the sum or difference of two signals, for
common-mode rejection to remove an undesired signal or for
DC offset (applying a DC voltage to one channel to offset
the DC component of a signal on the other channel).
The common-mode rejection ratio of the Type 453 is
greater than 20:1 at 20 megahertz for signal amplitudes up
to eight times the VOLTS/DIV switch setting. Rejection ratios
of 100:1 can typically be achieved between DC and 5 megahertz by careful adjustment of the gain of either channel
while observing the displayed common-mode signal.
Deflection Factor. The overall deflection in the ADD
position of the MODE switch when both VOLTS/DIV switches
are set to the same position is the same as the deflection
factor indicated by either VOLTS/DIV switch. The amplitude of an added mode display can be determined directly
from the resultant CRT deflection multiplied by the deflection factor indicated by either VOLTS/DIV switch. However, if the CH 1 and CH 2 VOLTS/DIV switches are set to
different deflection factors, resultant voltage is difficult to
determine from the CRT display. In this case, the voltage
amplitude of the resultant display can be determined accurately only if the amplitude of the signal applied to either
channel is known.
Precautions.
be observed when using the ADD mode.
1. Do not exceed the input voltage
453.
The following general
precautions should
rating of the Type
2-13
TM 11-6625-1722-15
2. Do not apply signals that exceed on equivalent of
about 20 times the VOLTS/DIV switch setting. For example,
with a VOLTS/DIV switch setting of .5, the voltage applied
to that channel should not exceed about 10 volts. Larger
voltages may distort the display.
3. Use vertical POSITION control settings which most
nearly position the signal of each channel to mid-screen
when viewed in either the CH 1 or CH 2 positions of the
MODE switch. This insures the greatest dynamic range for
ADD mode operation.
4. For similiar response from each channel, set both Input
Coupling switches to the same position.
Trigger Source
INT. For most applications, the sweep can be triggered
internally. In the INT position of the Triggering SOURCE
switch, the trigger signal is obtained from the Vertical
Deflection System. The TRIGGER switch provides further
selection of the internal trigger signal; obtained from the
Channel 1 signal in the CH 1 ONLY position, or from the
displayed signal when in the NORM position. For singletrace displays of either channel, the NORM position provides the most convenient operation. However, for dualtrace displays special considerations must be made to provide
the correct display. Set Dual-Trace Operation in this section
for dual-trace triggering information.
LINE. The LINE position of the SOURCE switch connects
a sample of the power-line frequency to the Trigger Gen-
erator circuit. Line triggering is useful when the input signal
is time-related to the line frequency. It is also useful for
providing a stable display of a line-frequency component
in a complex waveform.
EXT. An external signal conected to the EXT TRIG INPUT
connector can be used to trigger the sweep in the EXT
position of the Triggering SOURCE switch. The external signal must be time-related to the displayed signal for a stable
display. An external trigger signal can be used to provide
a triggered display when the internal signal is too low in
amplitude for correct triggering, or contains signal components on which it is not desired to trigger. It is also useful
when signal tracing in amplifiers, phase-shift networks, waveshaping circuits, etc. The signal from a single point in the
circuit under test can be connected to the EXT TRIG INPUT
connector through a signal probe or cable. The sweep is
then triggered by the same signal at all times and allows
amplitude, time relationship or waveshape changes of
signals at various points in the circuit to be examined without resetting the trigger controls.
frequency components of the trigger signal which can trigger
the sweep.
AC. The AC position blocks the DC component of the
trigger signal. Signals with low-frequency components below
about 30 hertz are attenuated. In general, AC coupling can
be used for most applications. However, if the trigger signal
contains unwanted components or if the sweep is to be
triggered at a low repetition rate or a DC level, one of the
remaining COUPLING switch positions will provide a better
display.
The triggering point in the AC position depends on the
average voltage level of the trigger signal. If the trigger
signals occur in a random fashion, the average voltage level
will vary, causing the triggering point to vary also. This shift
of the triggering point may be enough so it is impossible to
maintain a stable display. In such cases, use DC coupling.
LF REJ. In the LF REJ position, DC is rejected and signals
below about 30 kilohertz are attenuated. Therefore, the
sweep will be triggered only by the higher-frequency com-
ponents of the signal. This position is particularly useful
for providing stable triggering if the trigger signal contains
line-frequency components. Also, in the ALT position of the
MODE switch, the LF REJ position provides the best display
at high sweep rates when comparing two unrelated signals
(TRIGGER switch set to NORM).
HF REJ. The HF REJ position passes all low-frequency
signals between about 30 hertz and 50 kilohertz. DC is
rejected and signals outside the given range are attenuated.
When triggering from complex waveforms, this position is
useful for providing stable display of low-frequency components.
DC. DC coupling can be used to provide stable triggering
with low-frequency signals which would be attenuated in
the AC position, or with low-repetition rate signals. The
LEVEL control can be adjusted to provide triggering at the
desired DC level on the waveform. When using internal
triggering, the setting of the Channel 1 and 2 POSITION
controls affects the DC trigger level.
DC trigger coupling should not be used in the ALT dualtrace mode if the TRIGGER switch is set to NORM. If used,
the sweep will trigger on the DC level of one trace and then
either lock out completely or free run on the other trace.
Correct DC triggering for this mode can be obtained with
the TRIGGER switch set to CH 1 ONLY.
Trigger Slope
same as described for EXT except that the external triggering
signal is attenuated 10 times. Attenuation of high-amplitude
external triggering signals is desirable to broaden the range
of the Triggering LEVEL control.
switch is set to LF REJ, attenuation is about 20:1.
Trigger Coupling
Four methods of coupling the
circuits can be selected with
switches. Each position permits
2-14
When the COUPLING
trigger signal to the trigger
the Triggering COUPLING
selection or rejection of the
The triggering SLOPE switch determines whether the trigger
circuit responds on the positive-going or negative-going
portion of the trigger signal. When the SLOPE switch is in
the + (positive-going) position, the display starts with the
positive-going portion of the waveform; in the - (negativegoing) position, the display starts with the negative-going
portion of the waveform (see Fig. 2-8]. When several cycles
of a signal appear in the display, the setting of the SLOPE
switch is often unimportant. However, if only a certain
portion of a cycle is to be displayed, correct setting of the
SLOPE switch is important to provide a display which starts
on the desired slope of the input signal.
Loading...
+ 229 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.