The Model 4299-1 Single-Unit Rack-Mount Kit contains the hardware required to mount one half-rack
instrument in a standard 48.3 cm (19 in.) rack. Typical installation is for one Series 2600A or 2600B System
®
SourceMeter
Parts list
Refer to the table below for a list of parts included in the kit.
Quantity Description Keithley part number
instrument.
Tools required
Medium (#2) Phillips screwdriver
3/8-inch wrench
Flat-bladed screwdriver or cage-nut installation tool
Installation overview
Review the Installation precautions (on page 2).
Select a location in the rack:
a. Clear as much space as possible around the selected rack location.
b. Remove the cabinet side panels from the rack.
Prepare the instrument (on page 2
Install mounting hardware (on page 3).
Prepare the rack (on page 6).
Install the instrument (on page 7) in the rack.
PA-908 Rev. F April 2021
).
*PPA-908F*
1
Model 4299-1 Single-Unit Rack-Mount Kit Installation Instructions
Installation precautions
Observe the following safety precautions while installing instruments.
Turn off instrument power and unplug all cables before installing an instrument in a rack.
Failure to remove power before installation may cause personal injury or death from
electrical shock.
Heat sources should be mounted away from the instrument, cabling, and accessories, with
sufficient space provided between the instrument and heat source for airflow and cooling.
To prevent damaging heat build-up and ensure specified performance, make sure there is adequate
ventilation and airflow around the instrument to ensure proper cooling. Do not cover the ventilation
holes on the sides or back of the instrument. The rear vents and both side vents must be
unobstructed to properly dissipate heat.
Mount precision equipment as low as possible in the rack. Operating temperatures are cooler lower
in the rack.
Prepare the instrument
To remove the handle, mounting ears, and rear feet:
1. Gently pull the handle (1) away from the sides of the instrument case and swing it up or down until the
orientation arrows on the handles line up with the orientation arrows on the mounting ears, as shown in the
following figure.
Figure 1: Rack-mount kit handle removal
2 PA-908 Rev. F April 2021
Model 4299-1 Single-Unit Rack-Mount Kit Installation Instructions
2. After you align the arrows (2), pull the ends of the handle (1) away from the case.
3. Remove the screw that secures each mounting ear (3).
4. Pull down and out to remove each ear.
Figure 2: Mounting ear removal
5. From the bottom of the instrument, pull out the rubber feet and remove the screws (4), see the following
figure.
Figure 3: Instrument rubber feet removal
6. Store the handle, rubber feet, and hardware for future use.
Install mounting hardware
The instrument may be installed on the either side of the rack. Attach the half-rack front panel (2602-
363) to the side without the instrument.
PA-908 Rev. F April 2021 3
Model 4299-1 Single-Unit Rack-Mount Kit Installation Instructions
To install the mounting hardware on the instrument:
1. Select the side that will be closest to the rack rail (see note above).
2. Secure the rear rack-mount bracket (2602-334) to the side of the instrument using two Phillips pan-head
SEMS screws (#10-32 × 3/8 PPHSEM).
Figure 4: Rear rack-mount bracket
3. Secure the front rack-ear (2602-355) to the rear holes in the rack ear if you are installing one Series 2600A
®
or 2600B System SourceMeter
instrument (refer to the following figure) and to the same side of the
instrument as the rear rack-mount bracket using two Phillips pan-head screws (#8-32 × 3/8 PPH).
If you are not installing a Series 2600A or 2600B instrument, use the front holes to attach the rack
ear and half-rack front panel. See the note in the following figure.
4 PA-908 Rev. F April 2021
Model 4299-1 Single-Unit Rack-Mount Kit Installation Instructions
4. Secure the half-rack front panel (2602-363) to the other side of the instrument using two Phillips pan-head
screws (#8-32 × 3/8 PPH).
Figure 5: Attach the rack ear and half-rack front panel
PA-908 Rev. F April 2021 5
Model 4299-1 Single-Unit Rack-Mount Kit Installation Instructions
Prepare the rack
1. Install the cage nuts (FA-274) into the front rack holes for the front rack-ear and the half-rack front panel
(see the following figure).
2. Install the cage nuts (FA-274) into the rear rack holes for the rack-mount extension arm (2602-335) on the
same side as the front rack-ear (2602-355). Make sure the locations selected for the rear rack holes
correspond to the location selected for the front rack holes.
Cage nuts for the front rack-ear, half-rack front panel, and rack-mount extension arm should be
installed into the rails at the same hole spacing (vertical height). Cage nuts for the rack-mount
extension arm should be installed on the same side as the cage nuts for the front rack-ear (refer to
Install the instrument topic for more detail).
Figure 6: Rack-mount extension arm
3. Secure the rack-mount extension arm (2602-335) to the rear of the rack rail using two Phillips truss-head
screws (#10-32 × 5/8 PHTRSH).
6 PA-908 Rev. F April 2021
Model 4299-1 Single-Unit Rack-Mount Kit Installation Instructions
Install the instrument
When rack mounting restricts access to the main power cord, provide a separate main power
disconnect device, located close to the equipment and within easy reach of the operator.
Failure to disconnect all power when needed may expose personnel to hazardous voltages,
which, if contacted, could cause personal injury or death.
To complete this procedure, two people are recommended.
1. Slide the instrument into the rack by aligning the rear rack-mount bracket (2602-334) with the rack-mount
extension arm (2602-335)(see the following figure).
2. While supporting the instrument, secure it to the front of the rack with four Phillips truss-head screws (#1032 × 5/8 PHTRSH).
Figure 7: Install the instrument
PA-908 Rev. F April 2021 7
Safety precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by personnel who recognize shock hazards and are familiar with the safety precautions required
to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the
product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained
service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low
transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient
overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient
overvoltages often associated with local AC mains connections. Certain Keithley measuring instruments may be connected to
mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any
unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If
the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth)
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the
voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or
accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be
impaired.
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are defined in the
specifications and operating information and shown on the instrument panels, test fixture panels, and switching cards.
Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground)
connections.
The WARNING heading in the user documentation explains hazards that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure.
8 PA-908 Rev. F April 2021
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or minor
injury or damage the instrument. Always read the associated information very carefully before performing the indicated
procedure. Damage to the instrument may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley. Standard fuses with applicable national safety
approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may
only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other
suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through
Keithley to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement
component, call a Keithley office for information.
Unless otherwise noted in product-specific literature, Keithley instruments are designed to operate indoors only, in the following
environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2.
To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the
instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that
consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board
should be returned to the factory for proper cleaning/servicing.
Safety precaution revision as of June 2017.
PA-908 Rev. F April 2021 9