The servicing instructions are for use by qualified personnel only. To avoid
personal injury, do not perform any servicing unless you are qualified to do
so. Refer to all safety summaries prior to performing service.
Product Firmware V1.0 and above. Revision B; released September 13,
Check AFG Sine and Ramp Frequency......................................................................................100
Check AFG Square and Pulse Frequency Accuracy..................................................................101
Check AFG Signal Amplitude Accuracy...................................................................................102
Check AFG DC Offset Accuracy...............................................................................................103
Check DVM Voltage Accuracy (DC)........................................................................................104
Check DVM Voltage Accuracy (AC)........................................................................................105
Check DVM Frequency Accuracy and Maximum Input Frequency..........................................107
This completes the Performance Verification procedures..........................................................108
ii3 Series MDO Specifications and Performance Verification
Important safety information
This manual contains information and warnings that must be followed by the user
for safe operation and to keep the product in a safe condition.
To safely perform service on this product, see the Service safety summary that
follows the General safety summary.
General safety summary
Use the product only as specified. Review the following safety precautions to
avoid injury and prevent damage to this product or any products connected to it.
Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you follow
generally accepted safety procedures in addition to the safety precautions
specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove
the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is
operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.
3 Series MDO Specifications and Performance Verification iii
Important safety information
To avoid fire or personal
injury
Use proper power cord. Use only the power cord specified for this product and
certified for the country of use. Do not use the provided power cord for other
products.
Ground the product. This product is grounded through the grounding conductor of
the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded. Do not
disable the power cord grounding connection.
Power disconnect. The power cord disconnects the product from the power
source. See instructions for the location. Do not position the equipment so that it
is difficult to operate the power cord; it must remain accessible to the user at all
times to allow for quick disconnection if needed.
Connect and disconnect properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source. Use only insulated voltage
probes, test leads, and adapters supplied with the product, or indicated by
Tektronix to be suitable for the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and
markings on the product. Consult the product manual for further ratings
information before making connections to the product. Do not exceed the
Measurement Category (CAT) rating and voltage or current rating of the lowest
rated individual component of a product, probe, or accessory. Use caution when
using 1:1 test leads because the probe tip voltage is directly transmitted to the
product.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do not operate without covers. Do not operate this product with covers or panels
removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or
operates incorrectly. If in doubt about safety of the product, turn it off and
disconnect the power cord. Clearly mark the product to prevent its further
operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical
damage and replace when damaged. Do not use probes or test leads if they are
damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing
pieces.
Use only specified replacement parts.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a
unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
iv3 Series MDO Specifications and Performance Verification
Important safety information
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
Provide proper ventilation. Refer to the installation instructions in the manual for
details on installing the product so it has proper ventilation.
Slots and openings are provided for ventilation and should never be covered or
otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location
convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads.
Improper or prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an
ergonomics professional to avoid stress injuries.
Use care when lifting and carrying the product. This product is provided with a
handle or handles for lifting and carrying.
WARNING. The product is heavy. To reduce the risk of personal injury or damage
to the device get help when lifting or carrying the product.
Probes and test leads
Use only the Tektronix rackmount hardware specified for this product.
Before connecting probes or test leads, connect the power cord from the power
connector to a properly grounded power outlet.
Keep fingers behind the protective barrier, protective finger guard, or tactile
indicator on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude,
and amperage rated probes, test leads, and adapters for any measurement.
3 Series MDO Specifications and Performance Verification v
Important safety information
Terms in the manual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on the product
These terms may appear on the product:
■
DANGER indicates an injury hazard immediately accessible as you read the
marking.
Symbols on the product
■
WARNING indicates an injury hazard not immediately accessible as you
read the marking.
■
CAUTION indicates a hazard to property including the product.
When this symbol is marked on the product, be sure to consult the
manual to find out the nature of the potential hazards and any actions
which have to be taken to avoid them. (This symbol may also be used
to refer the user to ratings in the manual.)
The following symbols may appear on the product:
vi3 Series MDO Specifications and Performance Verification
Specifications
This chapter contains specifications for the 3 Series MDO oscilloscopes. All
specifications are guaranteed unless noted as "typical." Typical specifications are
provided for your convenience but are not guaranteed. Specifications that are
marked with the symbol have associated procedures listed in the PerformanceVerification section.
All specifications apply to all 3 Series MDO models unless noted otherwise. To
meet specifications, the following conditions must first be met:
■
This instrument must have been calibrated/adjusted at an ambient
temperature between +18 °C and +28 °C.
■
The instrument must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in this section.
■
The instrument must be powered from a source maintaining voltage and
frequency within the limits described in this section.
■
The instrument must have had its signal-path-compensation routine last
executed after at least a 20-minute warm-up period at an ambient temperature
within ±5 °C of the current ambient temperature.
■
The instrument must have had a warm up period of at least 10 minutes.
3 Series MDO Specifications and Performance Verification 1
Specifications
MDO32 and MDO34
Standard spectrum
analyzer frequency
range
Optional spectrum
analyzer frequency
range with 3-SA3
option
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
Analog channel input and vertical specifications
Number of input channels
MDO344 analog, BNC, digitized simultaneously
MDO322 analog, BNC, digitized simultaneously
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
9 kHz 1 GHz
9 kHz 3 GHz
Input CouplingDC, AC
Input termination selection1 M
Input termination 1 MΩ DC-
coupled
Input termination, 50 Ω, DC-
coupled
Input capacitance 1 MΩ, typical13 pF ± 2 pF
Input VSWR, 50 Ω, DC-coupled,
typical
Maximum input voltage (50 Ω)5 V
Ω or 50 Ω
Ω, ±1%
1 M
50 Ω ± 1%
BandwidthVSWR
For instruments with 1 GHz bandwidth≤ 1.5:1 from DC to 1 GHz, typical
For instruments with 500 MHz bandwidth≤ 1.5:1 from DC to 500 MHz, typical
For instruments with 350 MHz bandwidth≤ 1.5:1 from DC to 350 MHz, typical
For instruments with 200 MHz bandwidth≤ 1.5:1 from DC to 200 MHz, typical
For instruments with 100 MHz bandwidth≤ 1.5:1 from DC to 100 MHz, typical
with peaks ≤ ±20 V, (DF ≤ 6.25%)
RMS
There is an over-voltage trip circuit, intended to protect against overloads that might damage
termination resistors. A sufficiently large impulse can cause damage regardless of the over-voltage
protection circuitry, due to the finite time required to detect the over-voltage condition and respond
to it.
Maximum input voltage (1 MΩ, DC
coupled)
The maximum input voltage at the BNC, 300 V
Installation Category II.
De-rate at 20 dB/decade between 4.5 MHz and 45 MHz, De-rate 14 db between 45 MHz and
450 MHz. Above 450 MHz, 5 V
RMS
RMS
.
23 Series MDO Specifications and Performance Verification
Analog channel input and vertical specifications (cont.) Specifications
Maximum peak input voltage at the BNC, ±424 V
DC balance
0.2 div with the input DC-50Ω coupled and 50 Ω terminated
0.25 div at 2 mV/div with the input DC-50 Ω coupled and 50 Ω terminated
0.5 div at 1 mV/div with the input DC-50 Ω coupled and 50 Ω terminated
0.2 div with the input DC-1 MΩ coupled and 50 Ω terminated
0.3 div at 1 mV/div with the input DC-1 MΩ coupled and 50 Ω terminated
All the above specifications are increased by 0.01 divisions per °C above 40 °C.
Number of digitized bits8 bits
Displayed vertically with 25 digitization levels (DL) per division, 10.24 divisions dynamic range
"DL" is the abbreviation for "digitization level." A DL is the smallest voltage level change that can be
resolved by an 8-bit A-D Converter. This value is also known as the LSB (least significant bit).
Sensitivity range (coarse)
1 M Ω:1 mV/div to 10 V/div in a 1-2-5 sequence
50 Ω:1 mV/div to 1 V/div in a 1-2-5 sequence
Sensitivity range (fine)Allows continuous adjustment from 1 mV/div to 10 V/div, 1 MΩ
Allows continuous adjustment from 1 mV/div to 1 V/div, 50 Ω
Sensitivity resolution (fine), typical ≤ 1% of current setting
DC gain accuracy±2.5% for 1 mV/Div, derated at 0.100%/°C above 30 °C
±2.0% for 2 mV/Div, derated at 0.100%/°C above 30 °C
±1.5% for 5 mV/Div and above, derated at 0.100%/°C above 30 °C
±3.0% Variable Gain, derated at 0.100%/°C above 30 °C
Offset rangesInput Signal cannot exceed Max Input Voltage for the 50 Ω input path.
Volts/div settingOffset range
1 MΩ input50 Ω input
1 mV/div - 50 mV/div±1 V±1 V
50.5 mV/div - 99.5 mV/div±0.5 V±0.5 V
100 mV/div - 500 mV/div±10 V±5 V
505 mV/div - 995 mV/div±5 V±5 V
1 V/div - 10 V/div
1
±100 V±5 V
Position range±5 divisions
1
For 50Ω path, 1V/div is the maximum vertical setting.
3 Series MDO Specifications and Performance Verification 3
Specifications Analog channel input and vertical specifications (cont.)
Offset accuracy±[0.005 X | offset - position | + DC Balance]
NOTE. Both the position and constant offset term must be converted to volts by multiplying by the
appropriate volts/div term.
Number of waveforms for average
acquisition mode
DC voltage measurement accuracy
Average acquisition mode
Sample acquisition mode,
typicalNOTE. Offset, position and the constant offset term must be converted to volts by multiplying by the
2 to 512 waveforms, Default of 16 waveforms
NOTE. Offset, position and the constant offset term must be converted to volts by multiplying by the
appropriate volts/div term.
Measurement TypeDC Accuracy (In Volts)
Average of > 16 waveforms±((DC Gain Accuracy) X |reading - (offset -
position)| + Offset Accuracy + 0.1 div)
Delta Volts between any two averages of
16 waveforms acquired with the same setup and
ambient conditions
The basic accuracy specification applies directly to any sample and to the following measurements:
High, Low, Max, Min, Mean, Cycle Mean, RMS, and Cycle RMS. The delta volt accuracy
specification applies to subtractive calculations involving two of these measurements.
The delta volts (difference voltage) accuracy specification applies directly to the following
measurements; Positive Overshoot, Negative Overshoot, Pk-Pk, and Amplitude.
appropriate volts/div term.
±(DC Gain Accuracy X |reading| + 0.05 div)
Analog bandwidth limit filter
selections
Measurement TypeDC Accuracy (In Volts)
Any Sample±(DC Gain Accuracy X |reading - (offset -
position)| + Offset Accuracy + 0.15 div + 0.6 mV)
Delta Volts between any two samples acquired
with the same setup and ambient conditions
For instruments with 1 GHz, 500 MHz or 350 MHz analog bandwidth: 20 MHz, 250 MHz, and Full
For instruments with 200 MHz and 100 MHz analog bandwidth: 20 MHz and Full
±(DC Gain Accuracy X |reading| + 0.15 div +
1.2 mV)
43 Series MDO Specifications and Performance Verification
Analog channel input and vertical specifications (cont.) Specifications
Analog bandwidth, 50 Ω, DC
coupled
1 GHz instruments:
Volts/Div settingBandwidth
10 mV/div - 1 V/divDC - 1.00 GHz
5 mV/div - 9.98 mV/divDC - 500 MHz
2 mV/div - 4.98 mV/divDC - 350 MHz
1 mV/div - 1.99 mV/divDC - 150 MHz
500 MHz instruments:
350 MHz instruments:
200 MHz instruments:
100 MHz instruments:
Analog bandwidth, 1 MΩ input
termination, typical
1 GHz, 500 MHz, and 350 MHz
instruments
Volts/Div settingBandwidth
5 mV/div - 1 V/divDC - 500 MHz
2 mV/div - 4.98 mV/divDC - 350 MHz
1 mV/div - 1.99 mV/divDC - 150 MHz
Volts/Div settingBandwidth
5 mV/div - 1 V/divDC - 350 MHz
2 mV/div - 4.98 mV/divDC - 350 MHz
1 mV/div - 1.99 mV/divDC - 150 MHz
Volts/Div settingBandwidth
2 mV/div - 1 V/divDC - 200 MHz
1 mV/div - 1.99 mV/divDC - 150 MHz
Volts/Div settingBandwidth
1 mV/div - 1 V/divDC - 100 MHz
The limits are for ambient temperature of ≤ 30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Volts/DivBandwidth
2 mV/div - 10 V/divDC - 350 MHz
1 mV/div - 1.99 V/divDC - 150 MHz
200 MHz instruments
100 MHz instruments
Volts/DivBandwidth
2 mV/div - 10 V/divDC - 200 MHz
1 mV/div - 1.99 V/divDC - 150 MHz
Volts/DivBandwidth
1 mV/div - 10 V/divDC - 100 MHz
3 Series MDO Specifications and Performance Verification 5
Specifications Analog channel input and vertical specifications (cont.)
Analog Bandwidth, 1 MΩ with
standard probe, typical
1 GHz instruments:The limits are for ambient temperature of ≤ 30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Volts/Div settingBandwidth
100 mV/div - 100 V/divDC - 1.00 GHz
50 mV/div - 99.8mV/divDC - 400 MHz
20 mV/div - 49.8 mV/divDC - 250 MHz
10 mV/div - 19.9 mV/divDC - 150 MHz
500 MHz instruments:
350 MHz instruments:
200 MHz instruments:
100 MHz instruments:
Volts/Div settingBandwidth
100 mV/div - 100 V/divDC - 500 MHz
50 mV/div - 99.8mV/divDC - 400 MHz
20 mV/div - 49.8 mV/divDC - 250 MHz
10 mV/div - 19.9 mV/divDC - 150 MHz
Volts/Div settingBandwidth
50 mV/div - 100 V/divDC - 350 MHz
20 mV/div - 49.8 mV/divDC - 250 MHz
10 mV/div - 19.9 mV/divDC - 150 MHz
Volts/Div settingBandwidth
20 mV/div - 100 V/divDC - 200 MHz
10 mV/div - 19.9 mV/divDC - 150 MHz
Volts/Div settingBandwidth
10 mV/div - 100 V/divDC - 100 MHz
63 Series MDO Specifications and Performance Verification
Analog channel input and vertical specifications (cont.) Specifications
Calculated rise time, typical
50 ΩCalculated Rise Time (10% to 90%) equals 0.4/BW. The formula accounts for the rise time
contribution of the oscilloscope independent of the rise time of the signal source.
All values in the table are in ps.
Instrument
bandwidth
1 GHz2666 1333 800 400
500 MHz2666 1333 800 800
350 MHz2666 1333 1143 1143
200 MHz2666 2000 2000 2000
100 MHz3500350035003500
TPPxxx0 ProbeAll values in the table are in ps. 1 GHz BW models assume the TPP1000 probe. 500 MHz and
350 MHz models assume the TPP0500B probe. 200 MHz and 100 MHz models assume the
TPP0250 probe.
Instrument
bandwidth
1 GHz2666 1600 1000 400
500 MHz2666 1600 1000 800
350 MHz2666 1600 1143 1143
200 MHz2666 2000 2000 2000
100 MHz3500350035003500
Volts per division
1 mV-1.99 mV2 mV-4.99 mV5 mV-9.98 mV10 mV-1 V
Volts per division
1 mV-1.99 mV2 mV-4.99 mV5 mV-9.98 mV10 mV-1 V
Lower frequency limit, AC coupled,
typical
Upper frequency limit, 250 MHz
bandwidth limit filter, typical
Upper frequency limit, 20 MHz
bandwidth limit filter, typical
Measurements made using the scopes automated measurement feature may read slower rise time
values than those determined by the above equation. This is because the automated
measurements do not take interpolation into account. Measuring using cursors on the interpolated
waveform gives a more accurate result.
< 10 Hz when AC to 1 MΩ coupled
The AC coupled lower frequency limits are reduced by a factor of 10 when 10X passive probes are
used.
250 MHz, +25%, and –25% (all models, except 100 MHz and 200 MHz)
50 Ω and 1 M
Ω, DC coupled: 20 MHz, ±25% (all models)
3 Series MDO Specifications and Performance Verification 7
Specifications Analog channel input and vertical specifications (cont.)
Pulse response, peak detect, or
envelope mode, typical
Random noise, sample
acquisition mode, 50 Ω termination
setting, full bandwidth, typical
Random noise, sample
acquisition mode, 50 Ω termination
setting, full bandwidth, guaranteed
Instrument bandwidthMinimum Pulse Width
1 GHz> 1.5 ns
500 MHz> 2.0 ns
350 MHz> 3.0 ns
200 MHz> 5.0 ns
100 MHz> 7.0 ns
1 mV/div100 mV/div1 V/div
1 GHz-1.98 mV17.07 mV
500 MHz-1.54 mV13.47 mV
350 MHz-1.7 mV12.7 mV
200 MHz111 µV1.6 mV15.19 mV
100 MHz98 µV1.38 mV15.87 mV
1 GHz500 MHz350 MHz200 MHz100 MHz
1 mV, Full BW 0.13 0.13 0.157 0.162 0.125
2 mV, Full BW 0.24 0.15 0.14 0.143 0.11
5 mV, Full BW 0.36 0.2 0.18 0.16 0.15
10 mV, FullBW0.39 0.29 0.3 0.3 0.3
20 mV, FullBW0.58 0.53 0.7 0.57 0.55
50 mV, FullBW1.5 1.4 1.6 1.5 1.4
100 mV, FullBW3.1 3.1 3.3 3.25 2.85
200 mV, FullBW6.2 5.5 6.7 6.75 5.5
500 mV, FullBW15.5 14.5 15.4 16.4 17
1 V, Full BW31 25.8 25 30.5 35
1 mV,
250 MHz BW
2 mV,
250 MHz BW
5 mV,
250 MHz BW
10 mV,
250 MHz BW
20 mV,
250 MHz BW
50 mV,
250 MHz BW
0.13 0.162 0.162 --
0.126 0.12 0.12 --
0.165 0.155 0.155 --
0.3 0.3 0.3 --
0.63 0.7 0.7 --
1.6 1.58 1.58 --
83 Series MDO Specifications and Performance Verification
Analog channel input and vertical specifications (cont.) Specifications
1 GHz500 MHz350 MHz200 MHz100 MHz
100 mV,
250 MHz BW
200 mV,
250 MHz BW
500 mV,
250 MHz BW
1 V, 250 MHzBW30 30 30 --
1 mV, 20 MHzBW0.078 0.078 0.078 0.078 0.078
2 mV, 20 MHzBW0.084 0.086 0.086 0.086 0.086
5 mV, 20 MHzBW0.16 0.17 0.17 0.17 0.17
3.4 3.3 3.3 --
6.5 6.5 6.5 --
16 16 16 --
10 mV,
20 MHz BW
20 mV,
20 MHz BW
50 mV,
20 MHz BW
100 mV,
20 MHz BW
200 mV,
20 MHz BW
500 mV,
20 MHz BW
1 V, 20 MHzBW30 28 28 28 28
Delay between channels, full
bandwidth, typical
Deskew range–125 ns to +125 ns
≤ 100 ps between any two channels with input termination set to 50 Ω, DC coupling
NOTE. All settings in the instrument can be manually time aligned using the Probe Deskew function
0.32 0.3 0.3 0.3 0.3
0.63 0.55 0.55 0.55 0.55
1.6 1.5 1.5 1.5 1.5
3.4 3.25 3.25 3.25 3.25
6.4 6 6 6 6
17 15 15 15 15
Digital-to-Analog skew1 ns
Crosstalk (channel isolation),
typical
TekVPI InterfaceThe probe interface allows installing, powering, compensating, and controlling a wide range of
1 MΩ100:1 30:1
50 Ω100:1 30:1
probes offering a variety of features.
The interface is available on CH1-CH4 front panel inputs. Aux In is available on the front of twochannel instrument only and is fully VPI compliant. Four-channel instruments have no Aux In input.
≤100 MHz>100 MHz
3 Series MDO Specifications and Performance Verification 9
Specifications
Digital channel acquisition specifications
Number of input channels16 Digital Inputs
Input resistance, typical101 KΩ to ground
Input capacitance, typical8 pF
Specified at the input to the P6316 probe with all 8 ground inputs connected to the user's ground.
Use of leadsets, grabber clips, ground extenders, or other connection accessories may
compromise this specification.
Minimum input signal swing,
typical
Maximum input signal swing,
typical
DC input voltage range+30 V, -20 V
Maximum input dynamic range50 Vpp (threshold setting dependent)
Channel to channel skew (typical)500 ps
Threshold voltage range–15 V to +25 V
Digital channel timing resolutionMinimum: 2 ns
Threshold accuracy± [130 mV + 3% of threshold setting after calibration]. Requires valid SPC.
500mV peak-to-peak
Specified at the input to the P6316 probe with all 8 ground inputs connected to the user's ground.
Use of leadsets, grabber clips, ground extenders, or other connection accessories may
compromise this specification.
+30 V, -20 V
Digital Channel to Digital Channel only
This is the propagation path skew, and ignores skew contributions due to bandpass distortion,
threshold inaccuracies (see Threshold Accuracy), and sample binning (see Digital Channel Timing
Resolution).
Minimum detectable pulse2.0 ns
Specified at the input to the P6316 probe with all eight ground inputs connected to the user's
ground. Use of lead sets, grabber clips, ground extenders, or other connection accessories may
compromise this specification.
103 Series MDO Specifications and Performance Verification
Specifications
Horizontal specifications
Sample Rate Range
Sample rate range (with 3 or 4 channels enabled)
CharacteristicDescription
Sample rate
range (Analog
Channels)
Time/Div10 M record5 M record1 M record100 K record10 K record1 K record
Seconds/division range<1 GHz instruments models: 1 ns/div to 1000 sec/div
1 GHz instruments models: 400 ps/div to 1000 sec/div
Maximum triggered acquisition
rate
Aperture uncertainty, typical (also
called "sample rate jitter")
Long-term sample rate and delay
time accuracy
Timebase delay time range-10 divisions to 5000 s
Delta time measurement
accuracy
Bandwidth1 and 2 channels3 and 4 channels
FastAcqDPOFastAcqDPO
1 GHz> 280,000
wfm/sec
< 1 GHz> 230,000
wfm/sec
≤ (5 ps + 1 × 10-6 x record duration)RMS, for records having duration ≤ 1 minute
Record duration = (Record Length) / (Sample Rate)
±10 ppm over any ≥ 1 ms time interval
The formula to calculate delta-time measurement accuracy (DTA) for a given instrument setting
and input signal is given below (assumes insignificant signal content above Nyquist).
> 60,000 wfm/sec > 230,000 wfm/
> 50,000 wfm/sec > 230,000 wfm/
> 50,000 wfm/sec
sec
> 50,000 wfm/sec
sec
143 Series MDO Specifications and Performance Verification
Horizontal specifications (cont.) Specifications
SR1 = Slew Rate (1st Edge) around the 1st point in the measurement
SR2 = Slew Rate (2nd Edge) around the 2nd point in the measurement
N = input-referred noise (voltsrms, refer to the Random Noise, Sample acquisition mode
specification)
tsr = 1 / (Sample Rate)
TBA = timebase accuracy (refer to the Long-term sample rate and delay time accuracy
specification)
tp = delta-time measurement duration
RD = (Record Length) / (Sample Rate)
N
2
N
DTApp= ±5 2
DTA
rms
= 2
+ 2
(
)
SR
1
N
2
+ 2
(
)
SR
1
2
+(5ps + 1x10-6xRD
(
)
SR
2
N
2
+(5ps + 1x10-6xRD
(
)
SR
2
2
)
+ 2tsr+ TBA xt
2t
sr
2
)
(
+
12
2
)
+ TBA xt
p
p
Frequency response tolerance,
typical
Assumes that error due to aliasing is insignificant.
The term under the square-root sign is the stability, and is related to the TIE (Time Interval Error).
The errors from this term occur throughout a single-shot measurement. The second term is a result
of both the absolute center-frequency accuracy and the center-frequency stability of the timebase,
and varies between multiple single-shot measurements over the observation interval (the amount of
time from the first single-shot measurement to the final single-shot measurement).
±0.5 dB from DC to 80% of nominal bandwidth
3 Series MDO Specifications and Performance Verification 15
Specifications
Trigger specifications
Aux In
Number of channelsMDO32 - 2 channel instruments: One (1) channel
MDO34 - 4 channel instruments: Zero (0) channels
Input impedance, typical1 MΩ ±1% in parallel with 13 pF ± 2 pF.
Maximum input voltage300 V RMS, Installation Category II; derate at 20 dB/decade above 3 MHz to 30 V RMS at 30 MHz;
10 dB/decade above 30 MHz.
Based upon sinusoidal or DC input signal. Excursion above 300 V should be less than 100 ms
duration and the duty factor is limited to < 44%. RMS signal level must be limited to 300 V. If these
values are exceeded, damage to the instrument may result.
Bandwidth, typical> 250 MHz
Trigger bandwidth, edge, pulse,
and logic, typical
Edge trigger sensitivity, typical
Edge trigger, DC coupled
Edge trigger, not DC coupled
Instrument bandwidthTrigger bandwidth
1 GHz≥1 GHz
500 MHz≥500 MHz
350 MHz≥500 MHz
200 MHz≥200 MHz
100 MHz≥200 MHz
Trigger sourceSensitivity
Any Analog Channel1 mV/div to 4.98 mV/div: 0.75 div from DC to
50 MHz, increasing to 1.3 div at instrument
bandwidth.
≥ 5 mV/div: 0.40 divisions from DC to 50 MHz,
increasing to 1 div at instrument bandwidth.
Aux In (External)200 mV from DC to 50 MHz, increasing to
500 mV at 200 MHz
LineThe line trigger level is fixed at about 50% of the
line voltage.
Trigger couplingSensitivity
AC1.5 times the DC Coupled limits for frequencies
above 10 Hz. Attenuates signals below 10 Hz.
Noise Rej2.5 times the DC Coupled limits
HF Reject1.5 times the DC Coupled limits from DC to
50 kHz. Attenuates signals above 50 kHz.
LF Reject1.5 times the DC Coupled limits for frequencies
above 50 kHz. Attenuates signals below 50 kHz
Trigger modesAuto, Normal, and Single
Trigger typesEdge, sequence (B trigger), pulse width, timeout, runt, logic, setup & hold, rise/fall time, video, and
bus (serial or parallel).
163 Series MDO Specifications and Performance Verification
Trigger specifications (cont.) Specifications
Video trigger
Formats and field ratesTriggers from negative sync composite video, field 1 or field 2 for interlaced systems, any field,
specific line, or any line for interlaced or non-interlaced systems. Supported systems include NTSC,
PAL, SECAM.
Standard Video formats are: Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60,
1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/30, 1080p/50, 1080p/60, and custom
bi-level and tri-level sync video standards.
Sensitivity, typical
SourceSensitivity
Any Analog Input Channel0.6 to 2.5 divisions of video sync tip
Aux In (External)Video not supported through Aux In (External)
input.
Lowest frequency for successful
45 Hz
set level to 50%, typical
Logic, logic-qualified, and Delay-
≥1.0 division, from DC to maximum bandwidth.
by-events sensitivities, DC
coupled, typical
Pulse width trigger sensitivity,
≥1.0 division, from DC to maximum bandwidth.
typical
Runt trigger sensitivity, typical≥1.0 division, from DC to maximum bandwidth.
Logic trigger minimum logic or
Triggering typePulse widthRearm timeTime between
rearm time, typical
LogicNot applicable2 ns2 ns
Time qualified logic4 ns2 ns2 ns
Setup/Hold violation trigger,
typical
Minimum clock pulse width,
typical
Time ranges
Minimum pulse width, clock active
User's hold time +2.5 ns
1
FeatureMinimumMaximum
2
Setup time-0.5 ns1.024 ms
Hold time1 ns1.024 ms
Setup + hold time0.5 ns2.048 ms
channels
1
Minimum pulse width, clock inactive
2 ns
2
1
For Logic, time between channels refers to the length of time a logic state derived from more than one channel must exist to be recognized. For Events,
the time is the minimum time between a main and delayed event that will be recognized if more than one channel is used.
3 Series MDO Specifications and Performance Verification 17
Specifications Trigger specifications (cont.)
Minimum pulse width, rearm time,
and transition time
Rise/Fall time, delta time range4 ns to 8 seconds
Pulse width or time-qualified runt
trigger time range
Identifier Trigger:11 bits of user-specified data, equal to (=), not equal to <>, less than (<), greater than (>), less than
or equal to (<=), greater than or equal to (>=), Inside Range, or Outside Range
Cycle Count Trigger:6 bits of user-specified data, equal to (≤), greater than or equal to (≥), Inside Range, Outside
Range
Header Fields Trigger:40 bits of user-specified data comprising Indicator Bits, Identifier, Payload Length, Header CRC,
and Cycle Count, equal to (=)
Data Trigger:1 - 16 bytes of user-specified data, with 0 to 253, or don't care bytes of data offset, including
qualifiers of equal to (=), not equal to <>, less than (<), greater than (>), less than or equal to (≤),
greater than or equal to (≥), Inside Range, and Outside Range.
End Of Frame:User-chosen types Static, Dynamic (DTS), and All
3 Series MDO Specifications and Performance Verification 19
Trigger on:Start of Frame, Indicator Bits, Identifier, Cycle Count, Header Fields, Data, Identifier & Data, End of
Frame, or Error
I2S triggering, optional
Data Trigger:32 bits of user-specified data in a left word, right word, or either, including qualifiers of equal to (=),
not equal to <>, less than (<), greater than (>), less than or equal to (≤), greater than or equal to
(≥), inside range, outside range
Trigger on:SS Word Select or Data
Maximum Data Rate:12.5 Mb/s
Left Justified triggering, optional
Data Trigger:32 bits of user-specified data in a left word, right word, or either, including qualifiers of equal to (=),
not equal to <>, less than (<), greater than (>), less than or equal to (≤), greater than or equal to
(≥), inside range, and outside range
Trigger on:Word Select or Data
Maximum Data Rate:12.5 Mb/s
Right Justified triggering, optional
Data Trigger:32 bits of user-specified data in a left word, right word, or either, including qualifiers of equal to (=),
not equal to <>, less than (<), greater than (>), less than or equal to (≤), greater than or equal to
(≥), inside range, outside range
Trigger on:Word Select and Data
Maximum Data Rate:12.5 Mb/s
MIL-STD-1553 triggering, optionalFor MIL-STD-1553, trigger selection of Command Word will trigger on Command and ambiguous
Command/Status words. Trigger selection of Status Word will trigger on Status and ambiguous
Command/Status words.
Bit Rate:1 Mb/s
Trigger on:Sync
Word Type (Command, Status, and Data)
Command Word (set RT Address (=, ≠, <, >, ≤, ≥, inside range, outside range), T/R, Sub-address/
Mode, Data Word Count/Mode Code, and Parity individually)
Status Word (set RT Address ( =, ≠, <, >, ≤, ≥, inside range, outside range), Message Error,
Instrumentation, Service Request Bit, Broadcast Command Received, Busy, Subsystem Flag,
Dynamic Bus Control Acceptance (DBCA), Terminal Flag, and Parity individually)
Data Word (user-specified 16-bit data value),
Error (Sync, Parity, Manchester, Non-contiguous data), Idle Time (minimum time selectable from
2 μs to 100 μs; maximum time selectable from 2 μs to 100 μs; trigger on < minimum, > maximum,
inside range, and outside range)
203 Series MDO Specifications and Performance Verification
Trigger specifications (cont.) Specifications
TDM triggering, optional
Data Trigger:32 bits of user-specified data in a channel 0-7, including qualifiers of equal to (=), not equal to <>,
less than (<), greater than (>), less than or equal to (≤), greater than or equal to (≥), inside range,
outside range.
Trigger On:Frame Sync or Data
Maximum Data Rate:25 Mb/s
USB triggering, optional
Data Rates Supported:Full: 12 Mbs, Low: 1.5 Mbs
Trigger On:Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet, Handshake
Packet, Special Packet, or Error
Display specifications
Display
TypeDisplay Area - 256.32 mm (H) x 144.18 mm (V), 29 cm (11.6 inch) diagonal TFT active matrix,
Display luminance is specified for a new display set at full brightness
Color Support16,777,216 (8-bit RGB) colors
3 Series MDO Specifications and Performance Verification 21
Specifications
Input/Output port specifications
Ethernet interfaceAn 8-pin RJ-45 connector that supports 10/100 Mb/s
GPIB interfaceAvailable as an optional accessory that connects to USB Device and USB Host port, with the TEK-
USB-488 GPIB to USB Adapter
Control interface is incorporated in the instrument user interface
HDMI connectorAn 19-pin, HDMI type connector
USB interfaceTwo USB host ports on the front of the instrument: two USB 2.0 High Speed ports.
One USB host port on the rear of the instrument: USB 2.0 High Speed port.
One USB 2.0 High Speed device port on the rear of the instrument providing USBTMC support.
Also Supports Full Speed and Slow Speed modes
Probe compensator output voltage
and frequency, typical
Output voltage:0 to 2.5 V amplitude
Source Impedance:1 KΩ
Frequency1 kHz
Auxiliary output (AUX OUT)
Selectable Output:Main Trigger, Event, or AFG
Main Trigger:HIGH to LOW transition indicates the trigger occurred
Event Out:The instrument will output a negative edge during a specified trigger event in a test application.
A falling edge occurs when there is a specified event in a test application (i.e. the waveform
crosses the violation threshold in the limit / mask test application).
A rising edge occurs when the trigger system begins waiting for the next test application event.
AFG:The trigger output signal from the AFG.
223 Series MDO Specifications and Performance Verification
Specifications
Data storage specifications
Nonvolatile memory retention time,
typical
Real-time clockA programmable clock providing time in years, months, days, hours, minutes, and seconds
Memory capacity
Front panelA 64 Kbit EEPROM on the LED board that stores the USB vendor ID and device ID for the internal
Analog boardThe PMU includes 64 KB of nonvolatile memory for storage of its own binary executable
Probe interfaceA microcontroller is used to manage probe communication as well as power state for the instrument
Main acquisitionTwo eMMC 4 GB ISSI devices contain the U-Boot, kernel, CAL constants, scope application, and
Mass storage deviceLinux: ≥4 GB. Form factor is an embedded eMMC BGA. Provides storage for saved customer data,
Host processor system4 Gb of DDR3-1600 DRAM. The host processor utilizes two matched DDR3 non-ECC embedded
No time limit for front-panel settings, saved waveforms, setups, and calibration constants
front panel controller
user data storage
all calibration constants and the Linux operating system. Not customer serviceable. Partition on the
device, with a nominal capacity of 4 GB, is available for storage of saved customer data.
modules
Power source specifications
Power consumption130 W maximum
Source voltage100 V to 240 V ±10%
Source frequency100 V to 240 V: 50/60 Hz
115 V: 400 Hz ±10%
Fuse ratingT3.15 A, 250 V
The fuse is not customer replaceable.
3 Series MDO Specifications and Performance Verification 23
Specifications
Mechanical specifications
Weight
InstrumentMDO34 1GHz: 11.7 lbs (5.31 kg)
MDO32 1GHz: 11.6 lbs (5.26 kg)
With accessoriesProtective front cover: + 1.0 lbs (0.45 kg)
Pouch: + 0.2 lbs (0.09 kg)
Soft case (SC3): + 4.0 lbs (1.81 kg)
Instrument when packaged for shipping: 17.4 lbs (7.89 kg)
Dimensions
Height252 mm (9.93 in.)
Width370 mm (14.57 in.)
Depth148.6 mm (5.85 in.)
Clearance requirementsThe clearance requirement for adequate cooling is 2.0 in (50.8 mm) on the right side (when looking
at the front of the instrument) and on the rear of the instrument
Acoustic noise emission
Sound power level38 dBA - 40 dBA typical in accordance with ISO 9296
243 Series MDO Specifications and Performance Verification
Specifications
Environmental specifications
Temperature
Operating0 °C to +55 °C (+32 °F to +131 °F)
Non-operating–40 °C to +71 °C (–40 °F to +160 °F)
Humidity
Operating5% to 90% relative humidity (% RH) at up to +40 °C
5% to 60% RH above +40 °C up to +55 °C, non-condensing, and as limited by a maximum wetbulb temperature of +39 °C
Non-operating5% to 90% relative humidity up to +40 °C,
5% to 60% relative humidity above +40 °C up to +55 °C
5% to 40% relative humidity above +55 °C up to +71 °C, non-condensing, and as limited by a
maximum wet-bulb temperature of +39 °C
Altitude
Operating3,000 m (9,843 feet)
Non-operating12,000 m (39,370 feet)
Random vibration
Non-operating:2.46 G
Operating:0.31 G
Meets IEC60068 2-64 and MIL-PRF-28800 Class 3
Shock
Operating:50 G, 1/2 sine, 11 ms duration, 3 drops in each direction of each axis, total of 18 shocks
Meets IEC 60068 2-27 and MIL-PRF-28800 Class 3
Non-operating50 G, 1/2 sine, 11 ms duration, 3 drops in each direction of each axis, total of 18 shocks
Exceeds MIL-PRF-28800F
, 5-500 Hz, 10 minutes per axis, 3 axes, 30 minutes total
RMS
, 5-500 Hz, 10 minutes per axis, 3 axes, 30 minutes total
RMS
3 Series MDO Specifications and Performance Verification 25
Specifications
RF input specifications
Center frequency range9 kHz to 3.0 GHz (with 3-SA3 installed)
9 kHz to 1.0 GHz (Any model at 1 GHz BW without 3-SA3 installed)
9 kHz to 500 MHz (Any model at 500 MHz BW without 3-SA3 installed)
9 kHz to 350 MHz (Any model at 350 MHz BW without 3-SA3 installed)
9 kHz to 200 MHz (Any model at 200 MHz BW without 3-SA3 installed)
9 kHz to 100 MHz (Any model at 100 MHz without 3-SA3 installed)
Resolution bandwidth range for
Windowing functions
Kaiser RBW Shape Factor60 db/3 db Shape factor ≤ 4:1
Reference frequency error,
cumulative
Marker frequency measurement
accuracy
Kaiser (default): 30 Hz – 150 MHz
Rectangular: 20 Hz – 150 MHz
Hamming: 20 Hz – 150 MHz
Hanning: 20 Hz – 150 MHz
Blackman-Harris: 30 Hz – 150 MHz
Flat-Top: 50 Hz – 150 MHz
Adjusted in 1-2-3-5 sequence
Cumulative Error: ±10 x 10
Includes allowances for aging per year, reference frequency calibration accuracy, and temperature
stability.
Valid over the recommended 1 year calibration interval, from –10 °C to +55 °C .
NOTE. The RF and analog channels share the same reference frequency. Reference frequency
accuracy is tested by the Long-term Sample Rate and Delay Time Accuracy checks.
±(([Reference Frequency Error] x [Marker Frequency]) + (span / 750 + 2)) Hz
Reference Frequency Error = 10 ppm (10 Hz/MHz)
-6
Example, assuming the span is set to 10 kHz and the marker is at 1,500 MHz, this would result in a
Frequency Measurement Accuracy of ±((10 Hz/1 MHz x 1,500 MHz) + (10 kHz / 750 + 2)) =
±15.015 kHz.
Marker Frequency with Span/RBW ≤ 1000:1
Reference Frequency Error with Marker level to displayed noise level > 30 dB
Phase noise from 1 GHz CW
10 kHz< -81 dBc/Hz, < -85 dBc/Hz (typical)
100 kHz< -97 dBc/Hz, < -101 dBc/Hz (typical)
1 MHz< -118 dBc/Hz, < -122 dBc/Hz (typical)
263 Series MDO Specifications and Performance Verification
Example: an input channel set up with +2 V offset and 1 V/div measuring a –5 V signal would have
±( 2 mV + [((( 4 * 1 ) / 5 ) + 1 )% of 5 V] + [0.5% of 2 V] ) = ±( 2 mV + [1.8% of 5 V] + [0.5% of 2 V] )
= ±( 2 mV + 90 mV + 10 mV ) = ±102 mV. This is roughly ±2% of the input voltage.
±2% (20 Hz to 10 kHz) typical
For AC measurements, the input channel vertical settings must allow the Vpp input signal to cover
between 4 and 8 divisions.
Frequency: 5 digits
100 MHz for 100 MHz models
150 MHz for all other models
Trigger Sensitivity limits must be observed for reliable frequency measurements.
, AC+DC
rms
(reads out in volts or amps); frequency count
rms
323 Series MDO Specifications and Performance Verification
Performance verification
This chapter contains performance verification procedures for the specifications
marked with the symbol. The following equipment, or a suitable equivalent, is
required to complete these procedures.
Table 3: Required equipment
DescriptionMinimum requirementsExamples
DC voltage source3 mV to 100 V, ±0.1% accuracyFluke 9500B Oscilloscope Calibrator with a
Leveled sine wave generator9 kHz to 3,000 MHz, ±4% amplitude
accuracy
Time mark generator80 ms period, ±1 ppm accuracy, rise time <
50 ns
50 Ω BNC cableMale-to-male connectorsTektronix part number 012-0057-01 (43 inch)
BNC feed-through termination50 ΩTektronix part number 011-0049-02
RF signal generator9 kHz to 3 GHz, –20 dBm to +10 dBmAnritsu MG3690C series with options 2, 3, 4,
Power meterUse with Power sensorRhode & Schwarz NRX
Power sensor–30 dBm to +10 dBmRhode & Schwarz NRP-Z98
Frequency counter0.1 Hz to 50 MHz, 5 ppm accuracyTektronix FCA3000
DMMDC Voltage: 0.1% accuracy
AC RMS Voltage: 0.2% accuracy
9530 Output Module
An appropriate BNC-to-0.1 inch pin adapter
between the Fluke 9530 and P6316 probe
15, 22
Tektronix DMM4040
You may need additional cables and adapters, depending on the actual test
equipment you use.
These procedures cover all 3 Series MDO models. Please disregard any checks
that do not apply to the specific model you are testing.
Print the test record on the following pages and use it to record the performance
test results for your oscilloscope.
NOTE. Completion of the performance verification procedure does not update the
stored time and date of the latest successful adjustment. The date and time are
updated only when the adjustment procedures in the service manual are
successfully completed.
The performance verification procedures verify the performance of your
instrument. They do not adjust your instrument. If your instrument fails any of
the performance verification tests, you should consult the factory adjustment
procedures described in the 3 Series MDO Service Manual.
3 Series MDO Specifications and Performance Verification 33
Performance verification
Upgrade the Firmware
For the best functionality, you can upgrade the oscilloscope firmware.
To upgrade the firmware of the oscilloscope:
1. Open up a Web browser and go to www.tektronix.com/software/downloads.
Proceed to the software finder. Download the latest firmware for your
oscilloscope on your PC.
2. Unzip the files and copy the firmware.img file into the root folder of a USB
flash drive or USB hard drive.
3. Power off your oscilloscope.
4. Insert the USB flash or hard drive into the USB port on the front panel of
your oscilloscope.
5. Power on the oscilloscope. The instrument automatically recognizes the
replacement firmware and installs it.
NOTE. Do not power off the oscilloscope or remove the USB drive until the
oscilloscope finishes installing the firmware.
If the instrument does not install the firmware, rerun the procedure. If the
problem continues, try a different model of USB flash or hard drive. Finally,
if needed, contact qualified service personnel.
6. When the upgrade is complete, power off the oscilloscope and remove the
USB flash or hard drive.
7. Power on the oscilloscope.
8. Tap Help and select About. The oscilloscope displays the firmware version
number.
9. Confirm that the version number matches that of the new firmware.
343 Series MDO Specifications and Performance Verification
Performance verification
Test Record
Print this section for use during the Performance Verification.
Model numberSerial numberProcedure performed byDate
DVM Frequency Accuracy Tests and Maximum Input Frequency
DVM Frequency Accuracy
Tests and Maximum Input
Channel 1
Channel 2
Channel 3
27
Frequency
NominalLow LimitTest ResultHigh Limit
9.0000 Hz8.9998 Hz9.0002 Hz
99.000 Hz98.998 Hz99.002 Hz
999.00 Hz998.98 Hz999.02 Hz
99.000 kHz98.998 kHz99.002 kHz
999.00 kHz998.98 kHz999.02 kHz
150 MHz
1
149.99 MHz150.01 MHz
9.0000 Hz8.9998 Hz9.0002 Hz
99.000 Hz98.998 Hz99.002 Hz
999.00 Hz998.98 Hz999.02 Hz
99.000 kHz98.998 kHz99.002 kHz
999.00 kHz998.98 kHz999.02 kHz
150 MHz
3
2
149.99 MHz150.01 MHz
9.0000 Hz8.9998 Hz9.0002 Hz
99.000 Hz98.998 Hz99.002 Hz
999.00 Hz998.98 Hz999.02 Hz
99.000 kHz98.998 kHz99.002 kHz
999.00 kHz998.98 kHz999.02 kHz
150 MHz
4
149.99 MHz150.01 MHz
27
Channels 3 and 4 are only on four-channel oscilloscopes.
1
Verifies the maximum frequency.
2
Verifies the maximum frequency.
3
Channels 3 and 4 are only on four-channel oscilloscopes.
4
Verifies the maximum frequency.
603 Series MDO Specifications and Performance Verification
Performance verification
DVM Frequency Accuracy Tests and Maximum Input Frequency
Channel 4
5
9.0000 Hz8.9998 Hz9.0002 Hz
99.000 Hz98.998 Hz99.002 Hz
999.00 Hz998.98 Hz999.02 Hz
99.000 kHz98.998 kHz99.002 kHz
999.00 kHz998.98 kHz999.02 kHz
150 MHz
6
149.99 MHz150.01 MHz
Performance Verification Procedures
The following three conditions must be met prior to performing these procedures:
1. The oscilloscope must have been operating continuously for ten (10) minutes
in an environment that meets the operating range specifications for
temperature and humidity.
2. You must perform a signal path compensation (SPC). (See Self Tests —
System Diagnostics and Signal Path Compensation section below.) If the
operating temperature changes by more than 5 °C (41 °F), you must perform
the signal path compensation again.
3. You must connect the oscilloscope and the test equipment to the same AC
power circuit. Connect the oscilloscope and test instruments into a common
power strip if you are unsure of the AC power circuit distribution.
Connecting the oscilloscope and test instruments into separate AC power
circuits can result in offset voltages between the equipment, which can
invalidate the performance verification procedure.
The time required to complete all the procedures is approximately one hour.
WARNING.
Some procedures use hazardous voltages. To prevent electrical shock,
always set voltage source outputs to 0 V before making or changing any
interconnections.
5
Channels 3 and 4 are only on four-channel oscilloscopes.
6
Verifies the maximum frequency.
3 Series MDO Specifications and Performance Verification 61
Performance verification
Self Tests, System Diagnostics, and Signal Path Compensation
These procedures use internal routines to verify that the oscilloscope functions
and passes its internal self tests. No test equipment or hookups are required. Start
the self test with these steps:
Run the System Diagnostics (may take several minutes):
1. Disconnect all probes and cables from the oscilloscope inputs.
2. Push Default Setup on the front-panel to set the instrument to the factory
default settings.
3. Tap Utility > Self Test. This displays the Self Test configuration menu.
4. Tap the Run Self Test button.
5. Wait while the self test runs. When the self test completes, a dialog box
displays the results of the self test.
6. Verify that the status of all tests is Passed.
7. Cycle the oscilloscope power off and back on before proceeding.
NOTE. Remember to cycle the oscilloscope power off and back on before
proceeding.
Run the signal-path compensation routine (may take 5 to 15 minutes per
channel):
1. Push Default Setup on the front panel.
2. Tap Utility > Calibration. This displays the Calibration configuration
menu.
3. Tap the Run SPC button to start the routine.
4. Signal-path compensation may take 5 to 15 minutes to run per channel.
5. Verify that the SPC Status is Passed.
6. Return to regular service: Tap anywhere outside the menu to exit the
Calibration menu.
The self test procedures are completed. If any of the above tests failed, run
the tests again. If there are still failures, contact Tektronix Customer Support.
NOTE. You cannot run the remaining performance tests until the self tests
pass and the SPC has successfully run.
623 Series MDO Specifications and Performance Verification
Check Input Termination DC Coupled (Resistance)
This test checks the Input Termination for 1 MΩ or 50 Ω settings.
1. Connect the output of the oscilloscope calibrator (for example, Fluke 9500)
to the oscilloscope channel 1 input, as shown below.
WARNING. Be sure to set the generator to Off or 0 volts before connecting,
disconnecting, and/or moving the test hookup during the performance of this
procedure. The generator is capable of providing dangerous voltages.
NOTE. Impedance measuring equipment that produces a voltage across the
channel that exceeds the measurement range of the instrument may report
erroneous impedance results. A measurement voltage exceeds the
measurement range of the instrument when the resulting trace is not visible
on the graticule.
Performance verification
2. Push Default Setup on the front panel to set the instrument to the factory
default settings.
3. Push the channel button on the front panel for the oscilloscope channel that
you are testing, as shown in the test record (for example, 1, 2, 3, or 4).
4. Confirm that the oscilloscope termination and calibrator impedance are both
set to 1 MΩ.
5. Turn the Vertical Scale knob to set the vertical scale, as shown in the test
record (for example, 10 mV/div, 100 mV/div, 1 V/div). See Input
Termination Tests on page 35.
6. Measure the input resistance of the oscilloscope with the calibrator. Record
this value in the test record.
7. Repeat steps 5 and 6 for each volt/division setting in the test record.
3 Series MDO Specifications and Performance Verification 63
Performance verification
Check DC Balance
8. Change the oscilloscope termination to 50 Ω and repeat steps 5 through 7.
9. Repeat steps 4 through 9 for each channel listed in the test record and
relevant to the model of oscilloscope that you are testing, as shown in the test
record (for example, 2, 3, or 4).
This completes the procedure.
This test checks the DC balance.
You do not need to connect the oscilloscope to any equipment to run this test.
The only piece of equipment needed is a BNC feed-through 50 Ω terminator.
1. For 50 Ω coupling, attach a 50 Ω terminator to the channel input of the
oscilloscope being tested.
2. Push Default Setup on the front panel to set the instrument to the factory
default settings.
3. Double-tap the Horizontal badge on the Settings bar and set the Horizontal
Scale to 1 ms/div.
4. Tap the channel 1 button on the oscilloscope Settings bar to display a channel
badge.
5. Double tap the Ch 1 badge to open its menu.
6. Set the Vertical Scale to 1 mV/div.
7. Set the channel Termination to 50 Ω.
8. Tap Bandwidth Limit and select 20 MHz, 150 MHz, or Full, as given in the
test record.
9. Tap outside the menu to close it.
643 Series MDO Specifications and Performance Verification
Performance verification
10. Double-tap the Acquisition badge and set the Acquisition Mode to
Average.
11. Set the Number of Waveforms to 16.
12. Tap outside the menu to close it.
13. Double-tap the Trigger badge and set the Source to AC line. You do not
need to connect an external signal to the oscilloscope for this DC Balance
test.
14. Tap outside the menu to close it.
15. Add a Mean amplitude measurement for channel 1 to the Results bar:
a. Tap the Add New... Measure button to open the Add Measurements
menu.
b. Set the Source to Ch 1.
c.In the Amplitude Measurements panel, double-tap the Mean button to
add the Mean measurement badge to the Results bar.
16. View the mean measurement value in the display and enter that mean value
as the test result in the test record. See DC Balance Tests on page 36.
NOTE. Translate the mean value into divisions for use in the test record. To
do this, divide the voltage value by the vertical scale value. (e.g. 0.2 V / (1 V /
division) = 0.2 divisions)
17. Repeat step 6 and step 16 for each volts/division value listed in the results
table.
18. Repeat step 6 and step 17 for each bandwidth setting in the test record table.
19. Repeat the channel tests at 1 MΩ impedance as follows:
a. Double-tap the channel 1 badge.
b. Set the Termination to 1 MΩ.
c.Repeat steps 7 through 18.
20. Repeat the procedure for all remaining channels as follows:
a. Move the 50 Ω terminator to the next channel input to be tested.
b. Double-tap the channel badge of the channel that you have finished
testing and set Display to Off.
3 Series MDO Specifications and Performance Verification 65
Performance verification
21. Tap outside the menu to close it.
Check Analog Bandwidth
This test checks the bandwidth for each channel.
1. Connect the output of the leveled sine wave generator (for example, Fluke
c.Tap the channel button on the Settings bar of the next channel to test.
d. Starting from step 6, repeat the procedures until all channels have been
tested. To change the source for the Mean measurement for each channel
test:
a. Double-tap the Mean measurement badge.
b. Tap the Configure panel.
c.Tap the Source 1 field and select the next channel to test.
9500) to the oscilloscope channel 1 input as shown below.
2. Push Default Setup on the front panel to set the instrument to the factory
default settings.
3. Double-tap the Acquisition badge and set the Acquisition mode to Sample.
4. Tap outside the menu to close it.
5. Add the peak-to-peak measurement as follows:
a. Tap the Measure button.
b. Set the Source to the channel under test.
c.In the Amplitude Measurements panel, tap the Peak-to-Peak
measurement button and then tap the Add button to add the measurement
badge to the Results bar.
d. Tap outside the menu to close it.
663 Series MDO Specifications and Performance Verification
Performance verification
6. Set the channel under test settings:
a. Double-tap the badge of the channel under test to open its configuration
menu.
b. Set Vertical Scale to 1 mV/div.
c.Set Termination to 50 Ω.
d. Tap outside the menu to close it.
7. Adjust the leveled sine wave signal source to display a waveform of
8 vertical divisions at the selected vertical scale with a set frequency of
10 MHz. For example, at 5 mV/div, use a ≥40 mV
a ≥16 mV
signal.
p-p
signal; at 2 mV/div, use
p-p
NOTE. At some V/div settings, the generator may not provide 8 vertical
divisions of signal. Set the generator output to obtain as many vertical
divisions of signal as possible.
8. Double-tap the Horizontal badge in the Settings bar.
9. Set the Horizontal Scale to 1 ms/division.
10. Tap outside the menu to close it.
11. Record the Peak-to-Peak measurement in the V
entry of the test record.
in-pp
12. Double-tap the Horizontal badge in the Settings bar.
13. Set the Horizontal Scale to 4 ns/division.
14. Adjust the signal source to the maximum bandwidth frequency for the
bandwidth and model being tested.
15. Record the peak-to-peak measurement as follows:
a. Record the Peak-to-Peak measurement at the new frequency in the
V
entry of the test record.
bw-pp
Table 4: Maximum Bandwidth Frequency worksheet
TerminationVertical ScaleMaximum Bandwidth Frequency
For instruments with 1 GHz bandwidth
50 Ω10 mV/div1 GHz
50 Ω5 mV/div500 MHz
50 Ω2 mV/div350 MHz
50 Ω1 mV/div150 MHz
For instruments with 500 MHz bandwidth
50 Ω5 mV/div500 MHz
3 Series MDO Specifications and Performance Verification 67
Performance verification
TerminationVertical ScaleMaximum Bandwidth Frequency
50 Ω2 mV/div350 MHz
50 Ω1 mV/div150 MHz
For instruments with 350 MHz bandwidth
50Ω5 mV/div350 MHz
50Ω2 mV/div350 MHz
50Ω1 mV/div150 MHz
For instruments with 200 MHz bandwidth
50 Ω2 mV/div200 MHz
50 Ω1 mV/div150 MHz
50 Ω1 mV/div100 MHz
For instruments with 100 MHz bandwidth
50 Ω1 mV/div100 MHz
16. Use the values of Vbw-pp and Vin-pp recorded in the test record, and the
following equation, to calculate the Gain at bandwidth:
Gain = Vbw-pp/Vin-pp.
17. To pass the performance measurement test, Gain should be ≥ 0.707. Enter
Gain in the test record.
18. Repeat steps 6 through 16 for all combinations of Vertical Scale and
Horizontal Scale settings listed in the test record.
19. Repeat the tests at 1 MΩ impedance as follows:
a. Set the calibrator output to Off or 0 volts.
b. Change the calibrator impedance to 1 MΩ.
c.Double-tap the badge of the channel under test to open its menu.
d. Set the Termination to 1 MΩ.
e.Repeat steps 6 through 18.
20. Repeat the test for all remaining channels as follows:
a. Set the calibrator to 0 volts and 50 Ω output impedance.
b. Move the calibrator output to the next channel input to be tested.
c.Press the channel button of the channel that you have finished testing to
turn the channel off.
d. Tap the channel button on the oscilloscope Settings bar of the next
channel to test.
e.Double-tap the Peak-to-Peak measurement badge.
f.Tap the Configure panel.
g. Tap the Source 1 field and select the next channel to test.
h. Starting from step 6, repeat the procedure until all channels have been
tested.
683 Series MDO Specifications and Performance Verification
This completes the procedure.
Check DC Gain Accuracy
This test checks the DC gain accuracy.
1. Connect the oscilloscope to a DC voltage source. If using the Fluke
Performance verification
9500 calibrator, connect the calibrator head to the oscilloscope channel to
test.
2. Push Default Setup on the front panel to set the instrument to the factory
default settings.
3. Double-tap the Acquisition badge and set Acquisition Mode to Average.
4. Set the Number of Waveforms to 16.
5. Tap outside the menu to close the menu.
6. Double-tap the Trigger badge and set the trigger Source to AC line.
7. Tap outside the menu to close the menu.
8. Add the Mean measurement to the Results bar:
a. Tap the Measure button to open the Add Measurements menu.
b. Set the Source to Ch 1.
c.In the Amplitude Measurements panel, tap the Mean button and then
tap the Add button to add the Mean measurement badge to the Results
bar.
9. Tap the channel button of the channel to test, to add the channel badge to the
Settings bar.
3 Series MDO Specifications and Performance Verification 69
Performance verification
10. Double tap the channel to test badge to open its menu and set the channel
settings:
a. Set Vertical Scale to 1 mV/div.
b. Set Termination to 50 Ω.
c.Tap Bandwidth Limit and set to 20 MHz.
d. Tap outside the menu to close it.
11. Record the negative-measured and positive-measured mean readings in the
Gain expected worksheet as follows:
a. On the calibrator, set the DC Voltage Source to the V
listed in the 1 mV row of the worksheet.
b. Double-tap the Acquisition badge and tap Clear to reset the
measurement statistics.
c.Enter the Mean reading in the worksheet as V
d. On the calibrator, set the DC Voltage Source to V
the 1 mV row of the worksheet.
e.Double-tap the Acquisition badge (if not open) and tap Clear.
f.Enter the
Mean
Table 5: Gain Expected worksheet - channel 1
Oscilloscope
Vertical Scale
Setting
1 mV/div7 mV–3.5 mV+3.5 mV
2 mV/div14 mV–7 mV+7 mV
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10 mV70 mV–35 mV+35 mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
50 mV350 mV–175 mV+175 mV
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
V
diffExpected
V
negative
V
positive
reading in the worksheet as V
V
negative-measuredVpositive-measuredVdiff
negative
negative-measured
value as listed in
positive
positive-measured
value as
.
.
Test
Result(Gain
Accuracy)
703 Series MDO Specifications and Performance Verification
Performance verification
Table 6: Gain Expected worksheet - channel 2
Oscilloscope
V
diffExpected
V
negative
V
positive
Vertical Scale
Setting
1 mV/div7 mV–3.5 mV+3.5 mV
2 mV/div14 mV–7 mV+7 mV
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10 mV70 mV–35 mV+35 mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
50 mV350 mV–175 mV+175 mV
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
Table 7: Gain Expected worksheet - channel 3
Oscilloscope
Vertical Scale
Setting
1 mV/div7 mV–3.5 mV+3.5 mV
2 mV/div14 mV–7 mV+7 mV
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10 mV70 mV–35 mV+35 mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
50 mV350 mV–175 mV+175 mV
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
V
diffExpected
V
negative
V
positive
V
negative-measuredVpositive-measuredVdiff
V
negative-measuredVpositive-measuredVdiff
Test
Result(Gain
Accuracy)
Test
Result(Gain
Accuracy)
Table 8: Gain Expected worksheet - channel 4
Oscilloscope
Vertical Scale
Setting
1 mV/div7 mV–3.5 mV+3.5 mV
2 mV/div14 mV–7 mV+7 mV
V
diffExpected
V
negative
V
positive
V
negative-measuredVpositive-measuredVdiff
Test
Result(Gain
Accuracy)
3 Series MDO Specifications and Performance Verification 71
Performance verification
Oscilloscope
Vertical Scale
Setting
4.98 mV34.86 mV–17.43 mV+17.43 mV
5 mV35 mV–17.5 mV+17.5 mV
10 mV70 mV–35 mV+35 mV
20 mV140 mV–70 mV+70 mV
49.8 mV348.6 mV–174.3 mV+174.3 mV
50 mV350 mV–175 mV+175 mV
100 mV700 mV–350 mV+350 mV
200 mV1400 mV–700 mV+700 mV
500 mV3500 mV–1750 mV+1750 mV
1.0 V7000 mV–3500 mV+3500 mV
V
diffExpected
V
negative
V
positive
12. Calculate Gain Accuracy as follows:
a. Calculate V
b. Enter V
diff
in the worksheet.
diff
c.Calculate Gain Accuracy as follows: Gain Accuracy = ((V
V
diffExpected
)/V
d. Enter the Gain Accuracy value in the worksheet and in the test record.
V
negative-measuredVpositive-measuredVdiff
as follows: V
diffExpected
= | V
diff
) × 100%
negative-measured
- V
positive-measured
diff
Test
Result(Gain
Accuracy)
|
-
13. Repeat steps 10 through 12 for all vertical scale settings in the work sheet
and the test record.
14. Repeat tests at 1 MΩ impedance as follows:
a. Set the calibrator to 0 volts and 1 MΩ output impedance.
b. Double-tap the badge of the channel being tested.
c.Set the Termination to 1 MΩ
d. Repeat steps 10 through 13 for all vertical scale settings in the test
record.
15. Repeat the procedure for all remaining channels:
a. Set the calibrator to 0 volts and 50 Ω output impedance.
b. Move the calibrator output to the next channel input to be tested.
c.Press the channel button of the channel that you have finished testing to
turn off the channel.
d. Double-tap the Mean measurement badge.
e.Tap the Configure panel.
f.Tap the Source 1 field and select the next channel to test.
g. Starting from step 10, set the values from the test record for the channel
under test, and repeat the above steps until all channels have been tested.
16. Touch outside a menu to close the menu.
723 Series MDO Specifications and Performance Verification
Check Offset Accuracy
Performance verification
This completes the procedure.
This test checks the offset accuracy.
1. Connect the oscilloscope to a DC voltage source to run this test. If using the
Fluke 9500 calibrator as the DC voltage source, connect the calibrator head
to the oscilloscope channel to test.
WARNING.
disconnecting, or moving the test hookup during the performance of this
procedure. The generator is capable of providing dangerous voltages.
Push Default Setup on the front panel to set the instrument to the factory
2.
default settings.
3. Double-tap the Acquisition badge and set Acquisition Mode to Average.
4. Set the Number of Waveforms to 16.
5. Tap outside the menu to close the menu.
6. Double-tap the Trigger badge and set the trigger Source to AC line.
7. Double-tap the Horizontal badge and set Horizontal Scale to 20 ms/div.
8. Add the Mean measurement to the Results bar:
a. Tap the Measure button to open the Add Measurements menu.
b. Set the Source to Ch 1.
c.In the Amplitude Measurements panel, tap the Mean button and then
Set the generator output to Off or 0 volts before connecting,
tap the Add button to add the Mean measurement badge to the Results
bar.
3 Series MDO Specifications and Performance Verification 73
Performance verification
9. Tap the channel button (starting with channel 1) on the Settings bar to add
the channel under test to the Settings bar.
10. Double-tap the channel under test badge to open its configuration menu and
change the vertical settings:
a. Set Vertical Scale to 1 mV/div.
b. Set Offset to 900 mV.
c.Set Position to 0 by tapping Set to 0.
d. Set Termination to 50 Ω.
e.Tap Bandwidth Limit and set to 20 MHz.
f.Tap outside the menu to close it.
11. Set the calibrator output to +900 mV, as shown in the test record, and turn
the calibrator output On.
12. Enter the Mean measurement value in the test record.
13. Double-tap the channel under test badge to open its configuration menu andchange the Offset to -900 mV.
14. Set the calibrator output to -900 mV, as shown in the test record.
15. Enter the Mean measurement value in the test record.
16. Repeat step 10 through 15, changing the channel vertical settings and the
calibrator output as listed in the test record for the channel under test.
17. Repeat the channel tests at 1 MΩ impedance as follows:
a. Set the calibrator output to Off or 0 volts.
b. Change the calibrator impedance to 1 MΩ and voltage to +900 mV.
c.Turn the calibrator output On.
d. Repeat steps 10 through 15, changing the channel Termination to
1 MΩ and the vertical Offset value and the calibrator output as listed in
the 1 MΩ test record for the channel under test.
18. Repeat the procedure for all remaining channels as follows:
a. Double-tap the Mean measurement badge.
b. Tap the Configure panel.
c.Tap the Source 1 field and select the next channel to test.
d. Set the calibrator to 0 volts and 50 Ω output impedance.
e.Move the calibrator output to the next channel input to test.
f.Press the channel button of the channel that you have finished testing to
turn the channel off.
743 Series MDO Specifications and Performance Verification
g. Tap the channel button on the oscilloscope Settings bar of the next
channel to test.
h. Starting from step 2, repeat the procedure until all channels have been
tested.
19. This completes the procedure.
Check Long-term Sample Rate and Delay Time Accuracy
This test checks the sample rate and delay time accuracy (time base).
1. Push Default Setup on the oscilloscope front panel to set the instrument to
the factory default settings.
2. Connect the output of the time mark generator to the oscilloscope channel
1 input using a 50 Ω cable. Use the time mark generator with a 50 Ω source
with the oscilloscope set for internal 50 Ω termination.
Performance verification
3. Set the time mark generator to 80 ms. Use a time mark waveform with a fast
rising edge.
4. Set the mark amplitude to 1 Vpp.
5. Set the channel under test settings:
a. Double-tap the Channel 1 badge to open its configuration menu.
b. Set Vertical Scale to 500 mV/div.
c.Set Termination to 50 Ω.
d. Tap outside the menu to close it.
6. Double-tap the Horizontal badge in the Settings bar.
7. Set the Horizontal Scale to 20 ms/div.
8. Double-tap the Trigger badge in the Settings bar.
9. Adjust the Trigger Level for a triggered display.
3 Series MDO Specifications and Performance Verification 75
Performance verification
10. Adjust the vertical Position knob to center the time mark on center screen.
11. Adjust the Horizontal Position knob counterclockwise to set the delay to
exactly 80 ms.
12. Set the Horizontal Scale to 400 ns/div.
13. Compare the rising edge of the marker to the center horizontal graticule. The
rising edge should be within ±2 divisions of the center graticule. Enter the
deviation in the test record. See Sample Rate and Delay Time Accuracy on
page 44.
NOTE. One division of displacement from graticule center corresponds to a
5 ppm time base error.
This completes the procedure.
Check Random Noise Sample Acquisition Mode
This test checks random noise. You do not need to connect any test equipment to
the oscilloscope for this test.
1. Disconnect everything connected to the oscilloscope inputs.
2. Push Default Setup on the front panel to set the instrument to the factory
default settings. This sets the oscilloscope to Channel 1, Full Bandwidth,
1 MΩ input termination, 100 mV/div, and 4.00 μs/div.
3. Double-tap the Horizontal settings badge.
4. Set Horizontal Scale to 10 ms/div.
5. Double-tap the Channel badge of the channel being tested.
6. Set Termination to 50 Ω.
7. Set the Bandwidth Limit to the desired bandwidth.
8. Add the AC RMS measurement:
a. Tap the Measure button.
b. Set the Source to the channel being tested.
c.In the Amplitude Measurements panel, tap the AC RMS measurement
button and then tap the Add button to add the measurement badge to the
Results bar.
d. Double-tap the AC RMS measurement badge and tap Show Statistics in
Badge to display statistics in the measurement badge.
e.Tap outside the menu to close it.
9. Add the Mean measurement:
a. Tap the Measure button.
b. Set the Source to the channel being tested.
763 Series MDO Specifications and Performance Verification
Performance verification
c.In the Amplitude Measurements panel, tap the Mean measurement
button and then tap the Add button to add the measurement badge to the
Results bar.
d. Double-tap the Mean measurement badge and tap Show Statistics in
Badge to display statistics in the measurement badge.
e.Tap outside the menu to close it.
10. Record the measurements.
11. Calculate RMS noise voltage = Square root of (RMS2 – Mean2), and record
the result.
12. The calculated RMS noise voltage from step 11. should be less than the high
limit in the test record (the calculated maximum RMS noise).
13. Repeat the above test for the other bandwidths listed in the test record.
14. Repeat the above test for all other input channels. Channels 3 and 4 are only
available on four channel oscilloscopes.
This completes the procedure.
3 Series MDO Specifications and Performance Verification 77
Performance verification
Check Delta Time Measurement Accuracy
This test checks the Delta time measurement accuracy (DTA) for a given
instrument setting and input signal.
1. Set the sine wave generator output impedance to 50 Ω.
2. Connect a 50 Ω coaxial cable from the signal source to the oscilloscope
channel being tested.
WARNING.
disconnecting, or moving the test hookup during the performance of this
procedure. The generator is capable of providing dangerous voltages.
Push the oscilloscope front-panel Default Setup button.
3.
4. Double-tap the badge of the channel under test to open its configuration
menu.
5. Set Termination to 50 Ω.
6. Set the Vertical Scale to a value in the test record being tested.
7. Tap outside the menu to close it.
8. Double-tap the Trigger badge, and then, if necessary, set the Trigger Source
to the channel being tested:
9. Tap outside the menu to close it.
10. Double-tap the Horizontal badge.
11. Set the Horizontal Scale to a value in the test record being tested.
12. Tap outside the menu to close it.
Set the generator output to Off or 0 volts before connecting,
13. Add a Burst Width measurement for the channel under test:
a. Tap the Measure button.
b. Tap the Time Measurements panel.
783 Series MDO Specifications and Performance Verification
Performance verification
c.Tap the Burst Width measurement and then tap the Add button to add
the measurement badge to the Results bar.
d. Tap outside the menu to close it.
14. Double-tap the Burst Width results badge to open the measurement
configuration menu.
15. Tap Show Statistics in Badge to display the measurement statistics in the
results badge.
16. Tap outside the menu to close it.
17. Refer to the Test Record Delta Time Measurement Accuracy table. See Delta
Time Measurement Accuracy Tests on page 50. Set the oscilloscope and the
signal source as directed there.
18. Wait five or 10 seconds for the oscilloscope to acquire all the samples before
taking the reading.
19. Verify that the Std Dev is less than the upper limit shown for each setting,
and note the reading in the Test Record.
20. Repeat steps 4 through 19 for each setting combination shown in the Test
Record for the channel being tested.
21. Push the channel button on the front panel for the current channel to shut off
the channel. Push the channel button for the next channel to be tested, and
move the coaxial cable to the appropriate input on the oscilloscope. Only the
channel being tested should be enabled
22. Repeat steps 4 through 21 until all channels have been tested.
NOTE. For this test, enable only one channel at a time. If additional channels
are enabled at the same time, the maximum sample rate is reduced and the
limits in the Test Record are no longer valid.
This completes the procedure.
3 Series MDO Specifications and Performance Verification 79
Performance verification
Check Digital Threshold Accuracy (with 3-MSO option)
For models with the 3-MSO option only, this test checks the threshold accuracy
of the digital channels. This procedure applies to digital channels D0 through
D15, and to channel threshold values of 0 V and +4 V.
1. Connect the P6316 digital probe to the instrument.
2. Connect the P6316 Group 1 pod to the DC voltage source to run this test.
You will need a BNC-to-0.1 inch pin adapter to complete the connection.
NOTE. If using the Fluke 9500 calibrator as the DC voltage source, connect
the calibrator head to the P6316 Group 1 pod. You will need a BNCto-0.1 inch pin adapter to complete the connection.
3. Push Default Setup on the front panel to set the instrument to the factory
default settings.
4. Display the digital channels and set the thresholds as follows:
a. Tap the D15-D0 button on the Settings bar.
b. Double-tap the D15-D0 badge on the Settings bar.
c.Tap the D15-D8 Turn All On button to turn all bits on.
d. Tap the D7-D0 Turn All On button to turn all bits on.
e.Tap the D15-D8 Thresholds field at the bottom of the menu and set the
value to 0 V.
f.Tap the D7-D0 Thresholds field at the bottom of the menu and set the
value to 0 V. The thresholds are set for the 0 V threshold check.
g. Tap outside the menu to close it.
803 Series MDO Specifications and Performance Verification
Performance verification
5. You need to record the test values in the test record row for 0 V for each
digital channel. See Digital Threshold Accuracy Tests (with 3-MSO option)
on page 54.
6. Double-tap the Trigger badge.
7. Tap Slope and change the slope to rising edge.
8. Set the Source to the appropriate channel, such as D0.
By default, the Type is set to Edge, Coupling is set to DC, Slope is set to
Rising, Mode is set to Auto, and Level is set to match the threshold of the
channel being tested.
9. Tap outside the menu to close it.
10. Set the DC voltage source (Vs) to -400 mV. Wait 3 seconds. Check the logic
level of the corresponding digital channel in the display.
If the channel is a static logic level high (green), change the DC voltage
source Vs to -500 mV.
11. Increment Vs by +20 mV. Wait 3 seconds and check the logic level of the
corresponding digital channel in the display. If the channel is at a static logic
level high (green), record the Vs value as in the 0 V row of the test record.
If the channel is a logic level low (blue) or is alternating between high and
low, repeat this step (increment Vs by 20 mV, wait 3 seconds, and check for
a static logic high). Continue until a value for Vs- is found.
NOTE. In this procedure, the channel might not change state until after you
pass the set threshold level.
12. Double-tap the Trigger badge.
13. Tap Slope and change the slope to falling edge.
14. Tap outside the menu to close it.
15. Set the DC voltage source (Vs) to +400 mV. Wait 3 seconds. Check the logic
level of the corresponding digital channel in the display.
If the channel is a static logic level low (blue), change the DC voltage source
Vs to +500 mV.
16. Decrement Vs by -20 mV. Wait 3 seconds and check the logic level of the
corresponding digital channel in the display. If the channel is at a static logic
level low, record the Vs value as Vs+ in the 0 V row of the test record.
If the channel is a logic level high (green) or is alternating between high and
low, repeat this step (decrement Vs by 20 mV, wait 3 seconds, and check for
a static logic low). Continue until a value for Vs+ is found.
17. Find the average, V
= (Vs- + Vs+)/2. Record the average as the test result
sAvg
in the test record.
Compare the test result to the limits. If the result is between the limits,
continue with the procedure to test the channel at the +4 V threshold value.
3 Series MDO Specifications and Performance Verification 81
Performance verification
18. Repeat the procedure starting with step 6 for each remaining digital channel.
19. Double-tap the Trigger badge.
20. Set the Source to the appropriate channel, such as D0.
21. Tap Slope and change the slope to falling edge.
22. The remaining part of this procedure is for the +4 V threshold test.
a. Double-tap the D15-D0 badge on the Settings bar.
b. Tap the D15-D8 Turn All On button to turn all bits on.
c.Tap the D7-D0 Turn All On button to turn all bits on.
d. Tap the D15-D8 Thresholds field at the bottom of the menu and set the
value to 4.00 V.
e.Tap the D7-D0 Thresholds field at the bottom of the menu and set the
value to 4.00 V.
f.Tap outside the menu to close it.
23. Set the DC voltage source (Vs) to +4.4 V. Wait 3 seconds. Check the logic
level of the corresponding digital channel in the display.
If the channel is a static logic level low (blue), change the DC voltage source
Vs to +4.5 V.
24. Decrement Vs by -20 mV. Wait 3 seconds and check the logic level of the
corresponding digital channel in the display. If the channel is at a static logic
level low, record the Vs value as Vs+ in the 4 V row of the test record.
If the channel is a logic level high (green) or is alternating between high and
low, repeat this step (decrement Vs by 20 mV, wait 3 seconds, and check for
a static logic low). Continue until a value for Vs+ is found.
25. Double-tap the Trigger badge.
26. Tap Slope and change the slope to rising edge.
27. Tap outside the menu to close it.
28. Set the DC voltage source (Vs) to +3.6 V. Wait 3 seconds. Check the logic
level of the corresponding digital channel in the display.
If the channel is a static logic level high (green), change the DC voltage
source Vs to +3.5 V.
29. Increment Vs by +20 mV. Wait 3 seconds and check the logic level of the
corresponding digital channel in the display. If the channel is at a static logic
level high, record the Vs value as in the 4 V row of the test record.
If the channel is a logic level low (blue) or is alternating between high and
low, repeat this step (increment Vs by 20 mV, wait 3 seconds, and check for
a static logic high). Continue until a value for Vs- is found.
30. Find the average, V
= (Vs- + Vs+)/2. Record the average as the test result
sAvg
in the test record.
823 Series MDO Specifications and Performance Verification
Compare the test result to the limits. If the result is between the limits, the
channel passes the test.
31. Repeat the procedure starting with step 19 for each digital channel.
This completes the procedure.
Check Displayed Average Noise Level (DANL)
This test does not require an input signal.
The test measures the average internal noise level of the instrument, ignoring
residual spurs.
It checks these ranges:
■
9 kHz to 50 kHz
■
50 kHz to 5 MHz
■
5 MHz to BW (3-SA1 not installed)
Performance verification
■
5 MHz to 2 GHz (3-SA1 installed)
■
2 GHz to 3 GHz (3-SA3 installed)
NOTE. If the specific measurement frequency results in measuring a residual spur
that is visible above the noise level, the DANL specification applies not to the
spur but to the noise level on either side of the spur. Please refer to the Spurious
Response specifications.
1. Initial oscilloscope setup:
a. Terminate the RF input in 50 Ω with no input signal applied.
b. Push the Default Setup button on the front panel.
c.Tap the RF button to turn on the RF channel.
d. Turn on the average trace as follows:
a. Double-tap the RF badge to open the RF VERTICAL SETTINGS
configuration menu.
b. Tap TRACES to open the TRACES panel.
c.Tap Spectrum Traces Normal to turn off Normal.
d. Tap Spectrum Traces Average to turn on Average.
e.Turn on the average detection as follows:
a. Tap the Detection Method Manual button.
b. For the Average Spectrum Trace touch Detection Type and select
Average from the drop-down list.
3 Series MDO Specifications and Performance Verification 83
Performance verification
f.Set the reference level to –15 dBm as follows:
a. Tap Vertical Settings to open the Vertical Settings panel.
b. Tap Reference Level and set the Ref Level to –15.0 dBm.
g. Set the start and stop frequency as follows:
a. Double-tap the Horizontal badge.
b. Tap Start Frequency and set the start frequency to 9 kHz.
c.Tap Stop Frequency and set the stop frequency to 50 kHz.
2. Check from 9 kHz to 50 kHz:
a. Set Manual Marker (a) at the frequency with the highest noise level as
follows: Tap the Cursors button. Turn Multipurpose knob a to move the
marker to the frequency at the noise threshold (highest point of noise),
ignoring any spurs. For this span, it should be near 9 kHz on the far left
of the screen. See the following figure.
b. Record the noise threshold value (in dBm/Hz) in the test record and
compare it to the instrument specification.
3. In the test record, enter the result at this frequency (9 kHz).
843 Series MDO Specifications and Performance Verification
4. Check from 50 kHz to 5 MHz:
a. Double-tap the Horizontal badge.
b. Tap Stop Frequency and set the stop frequency to 5 MHz.
c.Tap Start Frequency and set the start frequency to 50 kHz.
d. Tap Span and set the Span to 10 MHz.
e.Set marker (a) at the frequency of the highest noise, ignoring any spurs.
f.Tap Center Frequency and set the frequency to 2.525 MHz.
a. Record the highest noise value (in dBm/Hz) in the test record and
compare it to the instrument specification.
Performance verification
5. In the test record, enter the result at this frequency (50 kHz).
6. Check from 5 MHz to BW (3-SA3 not installed):
a. Set the Stop Frequency to the maximum bandwidth.
b. Set the Start Frequency to 5 MHz.
c.Set marker (a) at the frequency of the highest noise, ignoring any spurs.
d. Tap Center Frequency and set the frequency to half the maximum
bandwidth.
e.Set the span to 10 MHz as follows: Tap Span and set the Span to
10 MHz.
f.Record the highest noise value (in dBm/Hz) in the test record and
compare it to the instrument specification.
7. Check from 5 MHz to 2 GHz (3-SA3 installed).
a. Set the Stop Frequency to 2 GHz.
b. Set the Start Frequency to 5 MHz.
c.Set marker (a) at the frequency of the highest noise, ignoring any spurs.
d. Tap Center Frequency and set the frequency to 1 GHz.
3 Series MDO Specifications and Performance Verification 85
Performance verification
e.Set the span to 10 MHz as follows: Tap Span and set the Span to
10 MHz.
f.Record the highest noise value (in dBm/Hz) in the test record and
compare it to the instrument specification.
8. Check from 2 GHz to 3 GHz (3-SA3 installed).
a. Set the Stop Frequency to 3 GHz.
b. Set the Start Frequency to 2 GHz.
c.Set marker (a) at the frequency of the highest noise, ignoring any spurs.
d. Tap the Center Frequency and set the frequency to1.5 GHz.
e.Set the span to 10 MHz as follows: Tap Span and set the Span to
10 MHz.
f.Record the highest noise value (in dBm/Hz) in the test record and
compare it to the instrument specification.
This completes the procedure.
Check Residual Spurious Response
This check verifies that the oscilloscope meets the specification for residual
spurious response. This check does not require an input signal.
1. Initial Setup:
a. Terminate the oscilloscope RF input in 50 Ω with no input signal applied.
b. Push Default Setup.
c.Tap RF. Double-tap the RF badge.
d. Tap TRACES to open the Traces panel
e.Tap Spectrum Traces Average to select Average. Tap Spectrum Traces
Normal to turn off Normal.
f.Tap VERTICAL SETTINGS to open the panel.
g. Tap Reference Level and set Ref Level to –15 dBm.
2. Check in the range of 9 kHz to 50 kHz (all models).
a. Double-tap the Horizontal badge.
b. Tap Start Frequency and set the start frequency to 9 kHz.
c.Tap Stop Frequency and set the stop frequency to 50 kHz.
d. Observe any spurs above –78 dBm and note them in the test record.
863 Series MDO Specifications and Performance Verification
Performance verification
3. Check in the range of 50 kHz to 5 MHz .
a. Set Stop Frequency to 5 MHz.
b. Set Start Frequency to 50 kHz.
c.Observe any spurs above –78 dBm and note them in the test record.
4. Check in the range of 5 MHz to Maximum Bandwidth (3-SA3 not installed):
a. Set Stop Frequency to the maximum bandwidth.
b. Set Start Frequency to 5 MHz.
c.Set RBW to 100 kHz.
d. Observe any spurs above –78 dBm and note them in the test record.
5. Check in the range of 5 MHz to 2 GHz (3-SA3 installed):
a. Set Stop Frequency to 2 GHz.
b. Set Start Frequency to 5 MHz.
c.Set RBW to 100 kHz.
d. Check the spur level at 1.25 GHz, if present. Turn the Multipurpose a
knob to line up the marker on the 1.25 GHz spur, if it is present. Adjust
the marker until the horizontal dash on the marker sits on top of the spur.
Note the spur level in the test record.
e.Observe any spurs above –78 dBm in the rest of the span, and note them
in the test record.
6. Check in the range of 2 GHz to 3 GHz (3-SA3 installed):
a. Set Stop Frequency to the 3 GHz.
b. Set Start Frequency to 2 GHz.
c.Set RBW to 100 kHz.
d. Check the spur level at 2.5 GHz, if present. Turn the Multipurpose a
knob to line up the marker on the 2.5 GHz spur, if it is present. Adjust
the marker until the horizontal dash on the marker sits on top of the spur.
Note the spur level in the test record.
e.Observe any spurs above –78 dBm in the rest of the span, and note them
in the test record.
This completes the procedure.
3 Series MDO Specifications and Performance Verification 87
Performance verification
Check Level Measurement Uncertainty
This test checks the level measurement uncertainty at three reference levels:
+10 dBm, 0 dBm, and –15 dBm. This check uses the generator to step
frequencies across four spans to verify that the instrument meets the
specification.
For this check, you will need the following equipment, which is described in the
Required Equipment table. See Table 3: Required equipment on page 33.
■
RF signal generator
■
Power meter
■
Power sensor
■
Power splitter
■
Adapters and cables as shown in the following figure.
WARNING. The generator is capable of providing dangerous voltages. Be sure to
set the generator to off or 0 volts before connecting, disconnecting, and/or
moving the test hookup during the performance of this procedure.
NOTE. Use an SMA connector with the RF signal generator. Equipment damage
will result if an N connector is used.
1. Connect the equipment as shown in the following figure.
2. Initial oscilloscope setup:
a. Push the Default Setup button on the front panel.
b. Tap RF to turn on the RF channel.
883 Series MDO Specifications and Performance Verification
Performance verification
3. Check at +10 dBm:
a. Double-tap the RF badge.
b. Set the reference level to +10 dBm as follows: Tap Reference Level and
set the Reference Level to +10 dBm.
c.Set the frequency range as follows:
■
Double-tap the Horizontal badge.
■
Tap Start Frequency and set the Start Frequency to 0 Hz.
■
Tap Stop Frequency ans set the stop frequency to 1 MHz.
d. Set the generator to provide a 9 kHz, +10 dBm signal.
e.At 9 kHz, determine the test result as follows:
■
Note the reading on the power meter and the readout for the
Reference marker on the oscilloscope. See the following figure.
■
Calculate the difference between the two readings. This is the test
result.
f.In the test record, enter the result at this frequency (9 kHz).
g. Set the generator to provide a 50 kHz, +10 dBm signal.
h. At 50 kHz, determine the test result as follows:
3 Series MDO Specifications and Performance Verification 89
Performance verification
■
Note the reading on the power meter and the readout for the
Reference marker on the oscilloscope. See the following figure.
■
Calculate the difference between the two readings. This is the test
result.
i.In the test record, enter the result at this frequency (50 kHz).
j.Step the generator, in 100 kHz intervals, through frequencies from
100 kHz to 900 kHz. At each interval, determine the test result as
follows:
■
Note the reading on the power meter and the readout for the
Reference marker on the oscilloscope.
■
Calculate the difference between the two readings. This is the test
result.
k. In the test record, enter the greatest result determined within this
frequency range (100 kHz – 900 kHz).
l.Change the frequency range as follows:
■
Change the stop frequency to 9.2 MHz.
■
Change the start frequency to 980 kHz.
m. Set the generator to provide a 1 MHz, +10 dBm signal.
n. Step the generator, in 1 MHz intervals, through frequencies from 1 MHz
to 9 MHz. At each interval, determine the test result as follows:
■
Note the reading on the power meter and the readout for the
Reference marker on the oscilloscope.
■
Calculate the difference between the two readings. This is the test
result.
o. In the test record, enter the greatest result determined within this
frequency range (1 MHz to 9 MHz).
903 Series MDO Specifications and Performance Verification
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.